Skip to main content
ACS Omega logoLink to ACS Omega
. 2024 Nov 11;9(47):47350–47360. doi: 10.1021/acsomega.4c09084

Colorimetric Determination of Dissolved Oxygen: Assessment of Methodological Influences on Iodine Spectra, Isosbestic Point, Precision and Accuracy

Su-Cheng Pai 1, Cheng-Ho Li 1, Brandon M Stephens 1,*
PMCID: PMC11603272  PMID: 39619564

Abstract

graphic file with name ao4c09084_0006.jpg

The colorimetric method for determining dissolved oxygen concentrations in freshwater and seawater samples essentially relies on measuring the absorbance of released I2 and I3 in a mixed form. While this approach is relatively quick and convenient, it is susceptible to significant temperature effects during analysis, irrespective of field temperatures. Additionally, the influences of spectrophotometric absorbance wavelengths and iodine concentrations on oxygen concentration determinations are ambiguous. We found that, while iodine concentration and absorbance wavelength impart minor influences, temperature changes during sample analysis alter the fractionation of two major iodine species (I2 and I3) and affect their spectra, with the latter effect imparting the greatest influence on oxygen concentrations. An empirical molar extinction coefficient was defined for the mixture (consisting of 96% I3 and 4% I2), yielding values of 2282, 1115, and 792 M–1 cm–1 at 25 °C at the most commonly used wavelengths of 430, 456, and 466 nm, respectively. Altering the temperature led to variances in absorbance by 0.33%, 0.42%, and 0.51% °C–1 at 430, 456, and 466 nm, respectively. Therefore, for absorbances measured at 456 nm, a room temperature correction can be applied to the empirical molar coefficient by 1115 M–1 cm–1 × [1 + (room temperature – 25 °C) × 0.0042]. Reducing room temperature differences between samples and calibration solutions became the most important factor in maintaining the precision and accuracy of the measurement. With these precautions, a precision of ∼0.2% (coefficient of variation) over the normal surface water oxygen concentration of around 250 μM can be readily achieved at all tested wavelengths. This method offers a linear oxygen concentration range of 0–650 μM.

1. Introduction

The precise quantification of dissolved oxygen in freshwater and seawater is crucial to identifying subtle environmental differences in both temporal and spatial trends. Several methods are available to measure dissolved oxygen, including the Winkler titration,1,2 the colorimetric detection of iodine by spectrophotometer,35 and various oxygen sensors.6,7 Although oxygen sensors are becoming increasingly common in laboratory and field studies,6 data quality by this method still heavily relies on precise, accurate and reliable wet-chemical calibration. The classic Winkler titration is the first choice for this purpose. This method employs three analytical steps: oxygen fixation with a manganese hydroxide oxide precipitate, liberation of iodine by adding acid in the presence of excess iodide, and titration with thiosulfate. Since the titration process requires several hands-on steps, which can be time-consuming, several colorimetric methods4,5,8 have been proposed as alternatives to the titration procedure, especially for clean waters like open oceans and nonturbid freshwaters. The advantages of using colorimetric methods are not only due to their simplicity and high precision but also to the effectiveness of eliminating iodine vaporization loss during the measurement.

Although the colorimetric approach has been used as a routine tool for seagoing expeditions over the last three decades, several analytical concerns still need clarification.

  • (a)

    Regarding reagent strengths. Since the colorimetric method has been established for measuring a mixture, it is essentially empirical. It is assumed that the three species (i.e., I2, I3, and I) have reached equilibrium in the final solution. However, in reality, the oxygen within the sample and the added KIO3 both consume I, thus the ratio of iodine species can be altered from one sample to another. The fractionation of iodine species must be calculated to identify these influences.

  • (b)

    Regarding the wavelength chosen for detection. A wavelength of 456 nm was first proposed by Pai et al.4 and then 430 nm by Roland et al.5 However, Labasque et al.8 suggested that the measurement should be carried out at an “isosbestic” wavelength of 466 nm where both I2 and I3 species share the same molar extinction coefficient. According to this concept, the determination will not be affected if the iodine fraction is altered. As shown in Labasque et al.8 and as will be shown below, there was a significant temperature effect on the isosbestic point (the absorbance may change approximately 0.5% per 1 °C at 466 nm). The reason for this effect remains to be identified.

  • (c)

    Regarding how to minimize the temperature effect. Ocean water samples (i.e., as a depth profile) can be collected from in situ temperatures ranging between 2 and 35 °C. Additionally, adding sulfuric acid during the last step will suddenly release heat before detection. Analytical criteria need to be established to ensure all measurements (as well as the standardization) are made under the same conditions.

These concerns will be carefully examined in this study.

1.1. Equations Involved in the Colorimetric Winkler Reactions

While basic equations for the Winkler method are available in numerous literature sources,4,5,8 we present more detailed equations related to colorimetry, specifically, to focus on how the fractionation of iodine alters absorbance with increases in oxygen or iodate levels.

  • (i)
    Oxygen is fixed by adding Mn2+ and NaOH in a flared-mouth bottle, usually referred to as a biological oxygen demand (BOD) bottle, to form precipitates
    graphic file with name ao4c09084_m001.jpg 1
  • (ii)
    Iodine is released by dissolving the precipitates with acid in the presence of iodide
    graphic file with name ao4c09084_m002.jpg 2
  • (iii)
    Iodine further reacts with excess iodide to form triiodide
    graphic file with name ao4c09084_m003.jpg 3
  • (iv)
    The final fractions of each iodine species are dependent upon the concentration of iodide present and the stability constant K for I3
    graphic file with name ao4c09084_m004.jpg 4
    According to Burger and Liebhafsky9 the K value at various temperatures T (°C) can be estimated by an empirical equation
    graphic file with name ao4c09084_m005.jpg 5
  • (v)
    The molar extinction coefficient ε2 for I2 can be obtained by preparing a known concentration of pure molecular iodine solution and measuring its absorbance
    graphic file with name ao4c09084_m006.jpg 6
    where b is the cuvette light path (usually 1 cm).
  • (vi)
    The molar extinction coefficient ε3 for I3 can be calculated by measuring the absorbance of a mixed solution of I2 and NaI and solving for the triiodide fraction. Assuming the final concentration of triiodide is x, A and B are the initial concentrations of I2 and I, the equilibrium can be expressed as
    graphic file with name ao4c09084_m007.jpg 7
    where B > 2A. The final concentration of triiodide can be solved as
    graphic file with name ao4c09084_m008.jpg 8
    And the final concentration of I2 is
    graphic file with name ao4c09084_m009.jpg 9
    The absorbance measured refers to a combination of two major iodine species
    graphic file with name ao4c09084_m010.jpg 10
    Thus, the value of ε3 can be estimated following the equation below when ε2 is obtained previously
    graphic file with name ao4c09084_m011.jpg 11
  • (vii)
    In a mixed I2 + I3 solution, a molar ratio r can be defined as
    graphic file with name ao4c09084_m012.jpg 12
    The sum of I2 and I3 gives total iodine ∑I2
    graphic file with name ao4c09084_m013.jpg 13
    If r, b, ε2, and ε3 are all fixed values, then eq 10 can be simplified as
    graphic file with name ao4c09084_m014.jpg 14
    where εT can be regarded as the “empirical molar extinction coefficient” for total iodine (i.e., the sum of I2 and I3) at a given I concentration and a given temperature.
  • (viii)
    The calibration is usually made by adding a known amount of iodate to an iodide solution, which generates the brownish iodine color
    graphic file with name ao4c09084_m015.jpg 15
    In this reaction, each mole of iodate added will produce 3 mol of molecular iodine while consuming 5 mol of iodide, followed by a fraction of molecular iodine consuming another mole of iodide to form triiodide. Assuming the final concentration of triiodide is x, C and B are the initial concentrations of iodine and initial iodide (where B > 5C), the equilibrium can be expressed as
    graphic file with name ao4c09084_m016.jpg 16
    Accordingly, the final concentration of triiodide will be
    graphic file with name ao4c09084_m017.jpg 17
    The final concentration of molecular iodine is
    graphic file with name ao4c09084_m018.jpg 18
    Measuring the absorbance of this solution and calculating by eq 14 an εT value can be obtained in the spiking experiment
    graphic file with name ao4c09084_m019.jpg 19
  • (ix)

    There are two types of reagent blanks involved in the colorimetric method, namely RB1 and RB2, representing the iodine impurities in the R2 reagent (NaI + NaOH) and oxygen contents in both R1 (Mn2+) and R2 reagents. The contributions of RB1 to the absorbance can be obtained experimentally by carrying out the same procedure for oxygen but adding reagents in reverse order (i.e., add R3 and R2 first, then R1). In our experience, the absorbance of RB1 is very small, mostly 0.000 and rarely exceeds 0.001.

    RB2 refers to the oxygen content within the dense R1 (3 M MnCl2 with a density of 1.30 g mL–1) and R2 (4 M NaI and 8 M NaOH with a density of 1.68 g mL–1) reagents. It is difficult to measure directly but can be evaluated according to Murray et al.10 who measured the oxygen concentrations in these two dense reagents. By calculating the amount of reagent used in a sample
    graphic file with name ao4c09084_m020.jpg 20
    where V1, V2, and V3 are added volumes for R1, R2, and R3. The excess [O2] for this reason is approximately 0.5 μM, which can be subtracted directly from each measured result.
  • (x)
    Converting measured absorbance to oxygen concentration: During the oxygen fixation step, for a bottle with an inner volume of Vb, the actual volume of sample is VbV1V2 (e.g., 60 – 0.5 – 0.5 = 59 mL), and the final volume after adding R3 is Vb + V3 (e.g., 60 + 0.5 = 60.5 mL). Therefore, the resultant total iodine will be
    graphic file with name ao4c09084_m021.jpg 21
    By combining this equation with eq 13 and considering both reagent blanks, [O2]sample = (Abs – RB1)/2 × (Vb + V3)/(VbV1V2) – [O2]RB2 or
    graphic file with name ao4c09084_m022.jpg 22
    The dilution term of (Vb + V3)/(VbV1V2) is nearly constant, even though the bottle volume Vb may vary significantly. For example, in the Shibala procedure,4 the value is 60.5/59 = 1.0254; in the Labasque8 procedure, it is 146.9/144.2 = 1.019. A bottle volume error of ±5% will only give a final variation of ±0.1%. Thus, for example, if εT is measured to be 1115 M–1 cm–1 at 25 °C using the Shibala procedure,4eq 22 can be simplified as
    graphic file with name ao4c09084_m023.jpg 23

2. Experimental Section

2.1. Thermostat Spectrophotometer

Since the present study focuses on the temperature effect on the iodine color, most experiments were carried out under very precise temperature control. To do this, a double-beam spectrophotometer (Shimadzu 1800A, Japan) was equipped with a water-circulation type temperature-controlled flow system, including a sample-loading jacket and a cuvette holder (Figure 1a). The temperature of the water tank was controlled by a thermostat (Firstek, Taiwan) with a 1 KW heater and a 700 W cooler. A Hellma 1 cm dome-type flow cuvette (Hellma GmbH & Co, Germany) was used, which was surrounded by a thermo holder. The sample, contained within a 60 mL BOD bottle (Wheaton) was first placed in the thermo jacket and held for several minutes before being siphoned up to fill the cuvette volume. Then the sample was trapped in the cuvette to wait for another several minutes to reach a steady state before proceeding to either spectral scanning or to measure at a fixed wavelength. The ramping rate of the thermostat system was checked by an iodine color solution and found to be about 0.5 °C min–1 for both the heating and cooling processes (Figure 1b).

Figure 1.

Figure 1

Layout of the water-circulation type thermostat spectrophotometer for the colorimetric Winkler method. (a) A thermo jacket outside the spectrophotometer holds a 60 mL BOD bottle, while a thermo cell holder inside the spectrophotometer contains a Hellma 1 cm dome-type flow cuvette. (b) The ramping rates for heating and cooling are roughly equal at 0.5 °C min–1. The efficiency of the water flow-circulating thermostat was demonstrated by filling the cuvette with an iodine solution and observing the variation of absorbance due to a change in the temperature setting. The absorbance indicates that when it becomes flat, the temperature has reached a steady state. The ramping rates for heating and cooling were similar at 0.5 °C min–1. If the temperature of the sample is different from the desired temperature, an appropriate waiting time is required before the scanning or detection operation.

2.2. Procedure for Routine Measurement Using an Ordinary Spectrophotometer

The routine procedure for measuring oxygen can be described graphically, as shown in Figure 2a. It is based on the protocol suggested by Pai et al.4 (also known as the Shibala procedure). For this, the water sample is filled into a BOD bottle (volume Vb = 60 mL), to which are added 0.5 mL of reagent R1 (containing 3 M MnCl2), 0.5 mL of reagent R2 (containing 8 M NaOH and 4 M NaI), the bottle is then stoppered, well mixed and allowed to stand for few hours until all precipitates are settled down to the bottom. Upon measurement, the bottle is opened and placed on a stir plate, and an aliquot of 0.5 mL of R3 (10 N sulfuric acid) is added while stirring. The brownish-colored solution is promptly sipped and transferred to fill a flow cuvette installed in a spectrophotometer, and the absorbance is recorded. The brownish color results from a mixture of molecular iodine (I2) and triiodide (I3). Since their molar ratio is fixed, the absorbance is directly proportional to the oxygen concentration of the sample. DDW refers to double distilled water.

Figure 2.

Figure 2

Illustration of the colorimetric Winkler method for oxygen measurement and calibration (following procedure outlined by Pai et al.4). Reagents used are R1 (containing 3 M Mn2+), R2 (containing 4 M NaI and 8 M NaOH) and R3 (10 N sulfuric acid). Potassium iodate (KIO3) is used as a standard. (a) Oxygen in a sample is fixed by adding R1 and R2 to form a precipitate. Upon measurement, the precipitate is dissolved by adding R3, releasing iodine and giving a brown color. The absorbance of the mixture is proportional to the [O2] in the original sample. (b) Distilled water (DW) is added with the same amounts of reagents but in reverse order. No precipitation is formed in the bottle, to which a known amount of KIO3 is added, which generates an iodine color. An empirical molar extinction coefficient for this mixture can be obtained for calibration purposes.

The procedure for the standardization is illustrated in Figure 2b. A BOD bottle filled with distilled water is treated with the same amounts of reagents but in reverse order. As the acidic R3 is added first, no precipitation will form; thus, the absorbance of this solution refers to the reagent blank, which has the same matrix as the oxygen measurement. The addition of known amounts of KIO3 to this solution produces an iodine color, and the absorbance obtained can be used to construct a calibration curve for the quantification of oxygen.

2.3. Definition of “Temperature”

In this study, two types of spectrophotometers were used. A thermostat spectrophotometer was employed to scan the spectra of pure I2 and I3 solutions to identify the isosbestic point, while an ordinary spectrophotometer was used for routine measurements. To avoid any potential misunderstandings or confusion for readers, we here define the term “temperature” more clearly and precisely as follows:

Tb: Temperature of the circulating water in the thermostat, which might slightly differ from the actual temperature of the flow cuvette if it is significantly higher or lower than room temperature.

Tr: Room temperature in the laboratory where routine measurements are conducted.

Tw: Temperature of the sample liquid after it has been left in the laboratory for at least 4 h to equilibrate with the surrounding environment. It is obtained by reading a thermometer placed in a bottle filled with water next to the spectrophotometer.

Tc: Temperature of the sample immediately after acidification with R3 sulfuric acid, which releases heat and temporarily raises the temperature by roughly 0.8 °C.

Ta: Actual temperature of the sample liquid trapped in the cuvette upon detection, which cannot be measured directly.

Since only Tw can be reliably recorded, in subsequent routine practice with an ordinary spectrophotometer, the term “temperature” should refer to Tw.

3. Results and Discussion

3.1. Comparison of Existing Protocols

The two most adopted colorimetric protocols4,8 and two titration methods11,12 were compared based on bottle volumes, reagent strengths, and possible conditions that may cause variations in the final iodine fractionation. For simplicity, these four methods are referred to as Shibala, Labasque, APHA and Grasshoff in text and tables.

According to eq 4, the iodine fraction is heavily related to two factors, i.e., the I concentration and the formation constant K; the latter is a function of temperature. Theoretical calculations on alterations to the fractionation of iodine due to temperature are made and shown in Table 1. The original data provided by Burger and Liebhafsky9 showed a clear linear relationship between ln K and the reciprocal of temperature, leading to an empirical equation shown in eq 5. In this study, the K values were calculated to be 806, 725, 655, and 593 at temperatures of 20, 25, 30, and 35 °C, respectively. Four protocols were compared using those K values, assuming the oxygen concentration was 250 μM. For example, for the Shibala protocol an aliquot of 0.5 mL of R2 (containing 4 M I) was added to a 60 mL bottle, resulting in an initial I concentration of 0.033 M in the final 60.5 mL solution. At 25 °C with a K value of 725, the final I3 fraction is 95.82%, which will be reduced to 94.94% if the temperature is raised to 35 °C, as calculated according to eqs 8 and 9. The reducing trend is estimated to be −0.09% per 1 °C. The Labasque protocol uses the same R2 but adds 0.9 mL to a 146 mL bottle to give an initial I concentration of 0.0245 M in the final solution. At 25 °C the final I3 fraction is 94.36% and will reduce to 93.49% at 35 °C. The decreasing trend is ca. −0.12% per 1 °C.

Table 1. Theoretical Calcultion of Iodine Fractionation Affected by Increasing Temperature Using Various Protocolsa.

            fractionation
 
protocol vol. of bottles (mL) vol of R2 (mL) [I] in R2 (M) final volume (mL) temp (°C) (I3)% (I2)% change of (I3)% per °C
Shibala 60 0.5 4 60.5 20 96.22% 3.78%  
          25 95.82% 4.18% –0.09%
          30 95.39% 4.61%  
          35 94.94% 5.06%  
Labasque 146 0.9 4 146.9 20 94.89% 5.11%  
          25 94.36% 5.64% –0.12%
          30 93.79% 6.21%  
          35 93.19% 6.81%  
APHA 300 1 0.87 301 20 56.60% 43.40%  
          25 54.17% 45.83% –0.87%
          30 51.82% 48.18%  
          35 49.50% 50.50%  
Grasshoff 50 0.5 3.61 50.5 20 96.51% 3.49%  
          25 96.13% 3.87% –0.08%
          30 95.74% 4.26%  
          35 95.32% 4.68%  
a

Calculations were based on an assumed oxygen concentration of 250 μM.

For the APHA method, in which an aliquot of 1 mL of 0.87 M I is added to a 300 mL bottle, the starting I concentration is 0.0029 M. It does not provide enough [I] to give a high I3 fraction value. At 25 °C the I3 fraction is 54.17% and drops to 49.5% at 35 °C with a decreasing trend of −0.87% per 1 °C. The increase of I2 fraction may induce vaporization loss,13,14 therefore it is not recommended for the colorimetric approach. In contrast, the Grasshoff protocol provides a more reasonable high I concentration (0.0357 M) and results in high I3 fraction of 96.13% at 25 °C, similar to that of the two colorimetric protocols.

In general, raising the temperature will lead to a slight decrease in I3 and an increase in I2. However, by using an I concentration of >0.03 M, these variations can be effectively limited to a negligible level within the normal oxygen concentration range.

3.2. Change of Iodine Fractionation Due to Oxygen Consumption

When oxygen is fixed and converted to I2, it consumes I as indicated by eqs 2 and 7, thus causing a slight decrease in I concentration and changing the final I2 fractionation. A theoretical calculation is also made for the four protocols. The results are summarized in Table 2.

Table 2. Theoretical Calculation of Iodine Fractionation Affected by Increasing Oxygen Concentration Using Various Protocolsa.

            fractionation
 
protocol vol. of bottles (mL) vol. of R2 (mL) [I] in R2 (M) final volume (mL) assumed [O2] (μM) (I3)% (I2)% change of (I3)% per 100 μM O2
Shibala 60 0.5 4 60.5 100 95.93% 4.07%  
          200 95.86% 4.14% –0.07%
          300 95.78% 4.22%  
Labasque 146 0.9 4 146.9 100 94.55% 5.45%  
          200 94.42% 5.58% –0.14%
          300 94.29% 5.71%  
APHA 300 1 0.87 301 100 63.20% 36.80%  
          200 57.53% 42.47% –11.14%
          300 50.38% 49.62%  
Grasshoff 50 0.5 3.61 50.5 100 96.23% 3.77%  
          200 96.17% 3.83% –0.06%
          300 96.10% 3.90%  
a

Calculations were made at a temperature of 25 °C.

For the Shibala procedure, the starting I concentration is 0.033 M within the bottle. Assuming the oxygen concentration of a sample is 100 μM, the final concentrations for I, I2 and I3 will be 32855, 7.95, and 187.1 μM, respectively, with a [I3]/[I2] ratio of 23.55. This means that 95.93% of total iodine will be I3 and 4.07% will be I2 in the final solution. Increasing the oxygen concentration from 100 to 300 μM, I3 only drops slightly to 95.78%. The decreasing trend is ca. −0.07% per 100 μM O2.

For the Labasque procedure, at a dissolved oxygen concentration of 100 μM, the final concentrations of I, I2 and I3 will be 24299, 10.7, and 185.6 μM, with a molar ratio of 17.35 and a final I3 of 94.55%. Increasing the oxygen concentration from 100 to 300 μM decreases the fraction slightly to 94.29%, and the decreasing trend is also very small, at −0.14% per 100 μM O2.

Calculations of the two titration protocols are also given in Table 1. It is clear that the APHA procedure uses a comparatively larger bottle but weaker reagent strength for I, giving a very low [I3]/[I2] ratio of 1.72 at an [O2] of 100 μM. The ratio drops further as [O2] increases, which indicates that the I2 fraction turns gradually to nearly 50% at an [O2] of 300 μM. The Grasshoff procedure uses a smaller bottle and a stronger I strength, giving a much higher [I3]/[I2] ratio of 25.5 at an [O2] of 100 μM, this is comparable to the high ratios provided by the Shibala and Labasque procedures.

3.3. The Iodine Spectra

Although the absorbance measured by the colorimetric method refers to a mixture of I2 and I3, it is possible to obtain their contribution individually by observing the spectra of a pure I2 standard solution and a nearly pure I3 solution (following eqs 619). Two solutions were prepared for this purpose and they were scanned between 400 and 700 nm at 20, 25, 30, and 35 °C. The first solution contained 0.0563 g I2 flakes (purity 99.8%) in 1 L of water, giving an [I2] of 224 μM. The second solution contained 0.0527 g of I2 flakes and 35.54 g of NaI in 1 L giving a final [I] of 0.237 M, a total [I2] of 207 μM, and a I3 fraction of >99%.

The absorption spectra of the two solutions are shown in Figure 3a. There are two major spectra for the pure I2 and the I2+NaI mixture, each consisting of 4 lines revealing the spectra at 4 different temperatures. The two bunches of spectra have two crosses, matching the results reported by Morel15 and Labasque et al.,8 who described that the two species have identical absorptivities at 466 and 580 nm, respectively.

Figure 3.

Figure 3

(a) Direct wavelength scan (400–700 nm) of two iodine solutions at various temperatures (Tb = 20–35 °C). The pure I2 solution containing 0.0563 g I2 flakes in 1 L gives an [I2] of 221 μM. The NaI+I2 solution is a mixture of 0.0527 g I2 flakes and 34.54 g NaI in 1 L. (b) Molar extinction coefficients (400–480 nm) for I2 and I3 at different temperatures: (blue) 20 °C (green) 25 °C (yellow) 30 °C (red) 35 °C. (c) Molar extinction coefficients for I2 (no iodide) and I3 (with different iodide concentrations). (d) The molar extinction coefficients for I3 at 430, 456, and 466 nm show a linear correlation with the concentration of iodide. The intercept of the regression indicates a hypothetical value.

The influence of temperature on the absorption spectra was statistically significant, but the trends were opposite to that of the fractionation of iodine species. In Figure 3b, the absorbance scale was converted to molar extinction coefficient scale. For the pure I2 solution, the highest absorption occurred at 460 nm. The molar extinction coefficient (ε2) measured at 20 °C was higher than that at 35 °C, with values of 729 and 708 M–1cm–1, respectively (Table 3). The decreasing trend was −0.15% °C–1. For the predominantly I3 solution, the spectrum appeared to shift up or to the right as temperature increased. At 456 nm, the molar extinction coefficient (ε3) was 1145 M–1 cm–1 at 20 °C and 1222 M–1 cm–1 at 35 °C (Table 3), with an increasing trend of 0.44% °C–1.

Table 3. Effect of Temperature on the Molar Extinction Coefficients.

  temperature
 
wavelength (nm) 20 °C 25 °C 30 °C 35 °C effect of temp. (1/°C)
Molar Extinction Coeffient for I22, M–1cm–1)a
430 543 540 536 531 –0.15%
456 724 717 711 703 –0.19%
460 729 722 715 708 –0.19%
466 723 717 711 704 –0.18%
469 715 709 704 697 –0.17%
470 712 706 700 694 –0.17%
476 685 680 676 670 –0.15%
Molar Extinction Coefficient for I33, M–1cm–1)b
430 2444 2484 2525 2569 0.34%
456 1145 1171 1195 1222 0.44%
460 997 1022 1045 1070 0.48%
466 797 819 840 863 0.53%
469 709 730 749 771 0.57%
470 682 703 722 742 0.58%
476 538 556 573 592 0.64%
a

A solution containing 0.0563 g iodine flake in 1 L.

b

A solution containing 0.527g I2 flake and 35.542 g NaI in 1 L.

3.4. The Effect of Iodide Concentration

In this experiment, six solutions were prepared by transferring different amounts of NaI (i.e., 0.571, 1.103, 2.027, 3.025, 6.116, and 31.711 g) (mol. wt. = 149.89) with distilled water into a 100 mL flask. To each flask, 1 mL of 4972 μM KIO3 solution and 0.1 mL of 10 N sulfuric acid were added, and the final volume was adjusted to the mark. These solutions contained starting concentrations of 0.038, 0.074, 0.135, 0.202, 0.408, and 2.116 M of I. These solutions were loaded into the flow cuvette at a fixed temperature of 25 °C for scanning between 430 and 480 nm. The results are shown in Figure 3c and Table 4.

Table 4. Effect of I Concentration to the Molar Extiction Coefficient of I3–,a,b.

Experimental Condition
[I](M) 0.038 0.074 0.135 0.202 0.408 2.116
[I3](μM) 143.8 146.4 147.6 148.1 148.7 149.1
[I2](μM) 5.34 2.78 1.52 1.02 0.50 0.10
[I3]/[I2] 27 53 97 146 295 1533
(I3)% 96.42% 98.14% 98.98% 99.32% 99.66% 99.93%
Apparent Extinction Coefficient for I3 (M–1 cm–1)
ε3 (430 nm) 2376 2396 2427 2467 2599 3672
ε3 (456 nm) 1127 1136 1147 1163 1208 1629
ε3 (466 nm) 791 797 805 815 842 1131
Apparent Extinction Coefficient for ∑I2 (M–1 cm–1)
εT (430 nm) 2310 2362 2407 2454 2592 3670
εT (456 nm) 1112 1128 1143 1160 1207 1628
εT (466 nm) 788 796 804 814 842 1130
a

Each NaI solution was added with 1 mL of 0.4972 mM KIO3.

b

Experiments were carried out at 25 °C.

Under extremely high concentrations of I (e.g., >0.2 M), the I3 fraction exceeds 99% (Table 4), but the absorbance still increases with I concentration. This phenomenon could be attributed a further complexation of I3 to (I5).16 However, in practical applications, I concentrations typically do not exceed 0.04 M, rendering the formation of I5 negligible.

3.5. The Isosbestic Point

Labasque et al.8 suggested that the colorimetric measurements for [O2] should be made at an isosbestic point of 466 nm, at which the molar extinction coefficients for both I2 and I3 species are identical. However, the isosbestic point was found to be around 470 nm in the current study. Experimental results have demonstrated that the isosbestic point shifts toward higher wavelengths and lower intensities by increasing temperature or [I].

For instance, at constant [I], the isosbestic point (λiso) was found to be 468.7 nm at 20 °C, but drifted to 472.1 nm as the temperature increased to 35 °C (Figure 3b and Table 5). The corresponding molar extinction coefficient (εiso) dropped from 716.5 M–1 cm–1 at 20 °C and to 687.6 M–1 cm–1 at 35 °C.

Table 5. Isosbestic Point Affected by Temperature and I Concentration.

T (°C) [I] (M) λ(iso) (nm) ε(iso) (M–1 cm–1)
Change Temperature
20 0.237 468.7 716.5
25 0.237 469.9 705.9
30 0.237 470.9 698.1
35 0.237 472.1 687.6
Change [I]
25 0.038 468.6 710.2
25 0.074 468.9 709.4
25 0.135 469.2 708.0
25 0.202 469.6 706.9
25 0.408 470.5 704.2
25 2.116 479.8 657.4

Iodide concentrations can also influence the isosbestic point. Under a fixed temperature of 25 °C, the isosbestic point drifted in a similar way as described above. At an [I] of 0.038 M, the isosbestic wavelength was 468.6 nm and drifted to a higher wavelength as concentrations increased (Figure 3c and Table 5). At an extremely high [I] of 2.116 M, the isosbestic wavelength (λiso) was found to be 479.8 nm. The isosbestic molar extinction coefficient (εiso) dropped from 710.2 to 657.4 M–1 cm–1.

3.6. Effect of Ionic Strength

For samples with high ionic strength, such as seawater, there is a concern that the activity coefficients of different species might change, potentially altering the formation constant K. However, previous studies have shown that the iodine spectra are not affected by salinity between wavelengths of 430 and 470 nm.4 This may be explained according to eq 4, given that both I and I3 are monovalent ions and are present in the numerator and denominator, their activity coefficients tend to cancel each other out. This leaves only a small variation due to the activity of the nonvalent I2 species. In an I3-dominating mixed solution, such variation is very small and can be reasonably ignored.

3.7. Oxygen Measurement and Calibration of εT by Spiking with KIO3

The calibration of εT was made by spiking various amounts of KIO3 standard solution to a distilled water sample and adding all the reagents in reverse order. The bottle was placed on a stir plate, and known amounts of KIO3 were added while stirring (e.g., 1.00–6.50 mL of 5000 μM KIO3). The absorbances recorded (at 456 nm in this case) were used to calculate the empirical molar extinction coefficient εT following eqs 19 and 20. Results are shown in Table 6. Basically, the empirical molar extinction coefficients were very consistent in the lower concentration range (0–1000 μM ∑I2), to be 1115 ± 3 M–1 cm–1 at Tw = 25 °C. The upper linear threshold of the calibration curve was defined as the concentration at which the slope value is dropped by 1% comparing to that at low concentration range. In the present case, the linear range was 0–1300 μM ∑I2, equivalent to an oxygen concentration range of 0–650 μM.

Table 6. Standardization of the Colorimetric Protocol by Spiking Known Amounts of Potassium Iodate at Tw = 25 °Ca.

Vb (mL) spiked V (mL) final V (mL) [∑I2] final (μM) abs (456 nm) εT (M–1 cm–1) (I3) % (I2) %
60.02 0.00 60.52 0.0 0.001      
60.47 0.00 60.97 0.0 0.001      
60.71 0.00 61.21 0.0 0.001      
60.31 1.00 61.81 242.0 0.271 1116 95.7% 4.3%
59.68 1.00 61.18 244.4 0.274 1117 95.8% 4.2%
61.07 1.00 62.57 239.0 0.268 1117 95.7% 4.3%
60.75 1.50 62.75 357.5 0.400 1116 95.6% 4.4%
60.12 1.50 62.12 361.1 0.405 1119 95.6% 4.4%
60.32 1.50 62.32 360.0 0.403 1117 95.6% 4.4%
60.34 2.50 63.34 590.3 0.660 1116 95.3% 4.7%
59.77 2.50 62.77 595.6 0.665 1115 95.4% 4.6%
62.02 2.50 65.02 575.0 0.643 1116 95.2% 4.8%
61.09 3.00 64.59 694.6 0.775 1114 95.1% 4.9%
60.00 3.00 63.50 706.5 0.790 1117 95.2% 4.8%
59.96 3.00 63.46 707.0 0.791 1117 95.2% 4.8%
59.48 3.50 63.48 824.6 0.918 1112 95.1% 4.9%
59.93 3.50 63.93 818.7 0.914 1115 95.0% 5.0%
59.90 3.50 63.90 819.1 0.912 1112 95.0% 5.0%
60.98 4.00 65.48 913.6 1.018 1113 94.8% 5.2%
59.37 4.00 63.87 936.6 1.043 1113 94.9% 5.1%
61.76 4.00 66.26 902.8 1.005 1112 94.7% 5.3%
59.46 4.50 64.46 1044.0 1.160 1110 94.7% 5.3%
60.64 4.50 65.64 1025.3 1.142 1113 94.6% 5.4%
61.44 4.50 66.44 1012.9 1.126 1111 94.6% 5.4%
59.93 5.00 65.43 1142.8 1.265 1106 94.5% 5.5%
59.81 5.00 65.31 1144.9 1.267 1106 94.5% 5.5%
61.22 5.00 66.72 1120.7 1.240 1106 94.4% 5.6%
59.36 5.50 65.36 1258.5 1.390 1104 94.4% 5.6%
60.67 5.50 66.67 1233.7 1.367 1107 94.3% 5.7%
60.25 5.50 66.25 1241.5 1.375 1107 94.3% 5.7%
61.39 6.00 67.89 1321.7 1.460 1104 94.0% 6.0%
59.61 6.00 66.11 1357.3 1.500 1104 94.1% 5.9%
59.42 6.00 65.92 1361.2 1.508 1107 94.2% 5.8%
59.30 6.50 66.30 1466.2 1.616 1102 94.0% 6.0%
59.89 6.50 66.89 1453.2 1.604 1103 93.9% 6.1%
59.97 6.50 66.97 1451.5 1.602 1103 93.9% 6.1%
a

A stock solution of 5.00 mM KIO3 was used for spiking.

3.8. Temperature Variation in Field Application

During open ocean field work, samples are often from different depths, which may cover a wide temperature range. A test was made to quantify the equilibration time required to minimize the temperature effect in a batch of environmental samples with different temperatures. To do this, aliquots of distilled water were adjusted to 32, 25, 10, and 5 °C, filled into the bottles, and then reagents were added. The temperatures inside those bottles were monitored for 4 h. After 1 h, the maximum temperature difference among the four bottles was reduced to 4.1 °C (Figure 4). The gap was further narrowed to 1.2 °C after 2 h, then 0.5 °C after 3 h, and 0.2 °C after 4 h. Upon adding the sulfuric acid, the temperature of each bottle was suddenly raised by 0.8 °C, and then gradually cooled down to the room temperature of nearly 25 °C (Figure 4). If the final temperature difference can be limited to less than 0.2 °C, the temperature influence on sensitivity (0.42% °C–1) can be reasonably controlled to less than 0.1%.

Figure 4.

Figure 4

Changes of temperature in sampling operation. Four water samples with initial temperatures of 5, 10, 25, and 32 °C were filled into BOD bottles following standard procedures,4 and the temperatures were monitored for 240 min. The maximum difference was 4.1 °C at 60 min, but narrowed down to 1.2, 0.5, and 0.2 °C after 2, 3, and 4 h. After the addition of 0.5 mL of 10 N sulfuric acid at 4 h, the temperature in each bottle was raised by ca. 0.8 °C.

Although a thermostat spectrophotometer is always preferable, most laboratories are not equipped with such thermostat devices. It should also be noted that the temperature in the cuvette chamber is typically warmer than outside the instrument. If the laboratory environment is 25 °C, the actual detection temperature may be higher by 1–2 °C. Consequently, the empirical molar extinction coefficient obtained may be raised by ca. 1%. Analysts may not need to know the exact temperature during detection but can ensure all samples (including the calibration samples) undergo the same process. In this way, a highly precise measurement can still be expected without using a thermostat.

3.9. Precision and Accuracy for Routine Analysis

A replicate test was carried out on an air-saturated distilled water sample pickled at 22.0 °C and measured at Tw = 24.8 °C, the resultant absorbances (n = 6) were 1.097 ± 0.002, 0.536 ± 0.001 and 0.381 ± 0.001 at wavelengths of 430, 456, and 466 nm, respectively, with a coefficient of variation (c.v.) of ∼0.2% at normal oxygen saturation level. It should be noted again that the variation in bottle volume does not contribute more than 0.1% to the variation in oxygen concentration, as described in eq 22.

The accuracy of this colorimetric method depends solely on the spiking with known amounts of KIO3, deriving an empirical molar extinction coefficient ϵT. Before analyzing each batch of samples, calibration must be performed carefully under a constant temperature Tw. In our laboratory, after a decade of experience, we have consistently obtained an ϵT value of 1115 ± 5 M–1 cm–1 at 456 nm when Tw = 25 ± 0.5 °C. If the laboratory temperature deviates from 25 °C, a modified empirical equation can be used

3.9. 24

This equation is valid under the conditions where 20 °C < Tw < 30 °C and 0.000 < Abs < 1.400.

4. Conclusions

The colorimetric Winkler method indeed offers advantages in the amount of time required and simplicity over traditional titration methods for determining dissolved oxygen concentration in water samples. Three factors have been identified to have influences on the precision and accuracy. These include:

  • (i)

    Alteration of the fractionation due to change of temperature on the formation constant K. In the Shibala procedure,4 the magnitudes are −0.09% °C–1 for I3 and +2.05% °C–1 for I2. In the Labasque8 procedure, the magnitudes are −0.12% °C–1 for I3 and +2.02% °C–1 for I2.

  • (ii)

    Consumption of Iby oxygen leads to a decrease in the I3but an increase in the I2. The change of fractions in I3 and I2 for the Shibala procedure are −0.07% and +1.73% per 100 μM O2 change, respectively. Meanwhile, for the Labasque procedure, the changes are −0.14% and +2.34% per 100 μM O2 change, respectively.

  • (iii)

    Changes in the absorptivities of both iodine species due to temperature effects. At 456 nm, the molar extinction coefficient increases by approximately +0.44% °C–1 for I3, decreasing by −0.19% °C–1 for I2. At 466 nm, the trends are comparable, with an increase of +0.53% °C–1 for I3 and a decrease of −0.18% °C–1 for I2.

The combined result, as seen in absorbance, indicates an apparent temperature effect of +0.33%, +0.42%, and +0.51% °C–1 at wavelengths of 430, 456, and 466 nm, respectively. Figure 5 illustrates this effect by comparing absorbances measured at different room temperatures from those measured at 25 °C.

Figure 5.

Figure 5

Overall temperature effect on the sensitivity at various wavelengths (430, 456, 466, and 476 nm) demonstrated as relative absorbance (w.r.t. Tw = 25 °C). Slopes are 0.33, 0.42, 0.51, and 0.62% per 1 °C at the corresponding wavelengths, respectively.

Based on the above findings, temperature control emerges as the most critical factor influencing the sensitivity, precision, and accuracy of the colorimetric method. While the temperature effect may be unavoidable, it can be minimized by ensuring a sufficiently high [I] (>0.03 M) and allowing all samples, as well as standard solutions, to equilibrate with the laboratory environment for at least 4 h. By implementing these conditions, the impact of temperature fluctuations on the colorimetric method can be effectively mitigated, thus enhancing the reliability of the [O2] measurements.

Given that variations caused by iodine fractionation are no longer significant, it is reasonable to conclude that detection does not necessarily need to be at the isosbestic wavelength; a wide range from 430 to 476 nm can be selected for detection, providing flexibility in choosing the most suitable wavelength for different purposes.

Considering the naturally occurring [O2] range of 100–800 μM in most environmental applications, the sensitivity of detecting at 430 nm would be too high (A430/A456 = 2.05), as it could potentially exceed the linearity range in absorbance. Conversely, detecting at 466 nm might lose significant sensitivity (A466/A456 = 0.71) without necessarily improving precision. Therefore, detection made at 456 nm, as suggested by the Shibala procedure,4 offers a balanced compromise between sensitivity and linearity, as well as accounting for the influence of the temperature effects.

This work provides valuable analytical guidance for achieving very precise measurements of oxygen concentration using colorimetry. However, it is worth noting that the current application is still limited to oligotrophic waters, such as nonturbid freshwaters and open oceans. Corrections can be applied to turbid waters by adopting the procedures suggested by Roland et al.5 Further research will be continued to evaluate the potential of expanding the Shibala4 method to correct for turbidity and a suite of other interferences including nitrite, hydrogen peroxide, and iodate.

Acknowledgments

We would like to thank many past students who have contributed to this work. In particular, we would like to thank Kuan-Lun Huang, Gallon Cheng and Yu-Chen Yen. We would also like to thank Professors G.T.F. Wong and A.C.T. Chen for their invaluable suggestions during the early stages of this study. Part of the data presented in this manuscript were taken from the Master’s Thesis of C.H. Li at National Taiwan University.

This work was supported by Taiwan’s National Science and Technology Council (NSTC), awarded to S.-C.P. under project NSC 101-261-M-002-013 and to B.M.S. under project NSTC 113-2611-M-002-009.

The authors declare no competing financial interest.

References

  1. Winkler L. W. Die Bestimmung Des Im Wasser Gelösten Sauerstoffes. Ber. Dtsch. Chem. Ges. 1888, 21 (2), 2843–2854. 10.1002/cber.188802102122. [DOI] [Google Scholar]
  2. Carpenter J. H. The Accuracy of the Winkler Method for Dissolved Oxygen Analysis. Limnol. Oceanogr. 1965, 10 (1), 135–140. 10.4319/lo.1965.10.1.0135. [DOI] [Google Scholar]
  3. Broenkow W. W.; Cline J. D. Colorimetric Determination of Dissolved Oxygen at Low Concentrations. Limnol. Oceanogr. 1969, 14 (3), 450–454. 10.4319/lo.1969.14.3.0450. [DOI] [Google Scholar]
  4. Pai S. C.; Gong G. C.; Liu K. K. Determination of Dissolved Oxygen in Seawater by Direct Spectrophotometry of Total Iodine. Mar. Chem. 1993, 41 (4), 343–351. 10.1016/0304-4203(93)90266-Q. [DOI] [Google Scholar]
  5. Roland F.; Caraco N. F.; Cole J. J.; Giorgio P. d. Rapid and Precise Determination of Dissolved Oxygen by Spectrophotometry: Evaluation of Interference from Color and Turbidity. Limnol. Oceanogr. 1999, 44 (4), 1148–1154. 10.4319/LO.1999.44.4.1148. [DOI] [Google Scholar]
  6. Wei Y.; Jiao Y.; An D.; Li D.; Li W.; Wei Q. Review of Dissolved Oxygen Detection Technology: From Laboratory Analysis to Online Intelligent Detection. Sensors 2019, 19 (18), 3995. 10.3390/s19183995. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Wang X. D.; Chen H. X.; Zhao Y.; Chen X.; Wang X. R.; Chen X. Optical Oxygen Sensors Move towards Colorimetric Determination. TrAC, Trends Anal. Chem. 2010, 29 (4), 319–338. 10.1016/j.trac.2010.01.004. [DOI] [Google Scholar]
  8. Labasque T.; Chaumery C.; Aminot A.; Kergoat G. Spectrophotometric Winkler Determination of Dissolved Oxygen: Re-Examination of Critical Factors and Reliability. Mar Chem. 2004, 88 (1–2), 53–60. 10.1016/j.marchem.2004.03.004. [DOI] [Google Scholar]
  9. Burger J. D.; Liebhafsky H. A. Thermodynamic Data for Aqueous Iodine Solutions at Various Temperatures. Exercise in Analytical Chemistry. Anal. Chem. 1973, 45 (3), 600–602. 10.1021/ac60325a044. [DOI] [Google Scholar]
  10. Murray C. N.; Riley J. P.; Wilson T. R. S. The Solubility of Oxygen in Winkler Reagents Used for the Determination of Dissolved Oxygen. Deep Sea Research and Oceanographic Abstracts 1968, 15 (2), 237–238. 10.1016/0011-7471(68)90046-6. [DOI] [Google Scholar]
  11. Grasshoff K.; Kremling K.; Ehrhardt M.. Methods of Seawater Analysis; John Wiley & Sons, 2009. [Google Scholar]
  12. American Public Health Association; others . APHA Standard Methods for the Examination of Water and Wastewater. In Standard Methods for the Examination of Water & Wastewater; American Public Health Association: Washington, DC, 2005. [Google Scholar]
  13. Knapp G. P.; Stalcup M. C.; Stanley R. J. Iodine Losses during Winkler Titrations. Deep Sea Res. Part A. Oceanographic Res. Papers 1991, 38 (1), 121–128. 10.1016/0198-0149(91)90057-M. [DOI] [Google Scholar]
  14. Green E. J.; Carritt D. E. An Improved Iodine Determination Flask for Whole-Bottle Titrations. Analyst 1966, 91 (1080), 207–208. 10.1039/an9669100207. [DOI] [Google Scholar]
  15. Morel A. Mise Au Point d’une Méthode Spectrophotométrique Pour Le Dosage de l’oxygène Dissous Dans Les Eaux de Mer. Bull. Inst. Océanogr. Monaco 1965, 64 (1332), 1–31. [Google Scholar]
  16. Hayakawa K.; Nakamura S. The Determination of the Formation Constants of the Triiodide Ion in Water-Alcohol Mixed Solvents at Various Temperatures. Bull. Chem. Soc. Jpn. 1977, 50 (3), 566–569. 10.1246/bcsj.50.566. [DOI] [Google Scholar]

Articles from ACS Omega are provided here courtesy of American Chemical Society

RESOURCES