Abstract
Objectives
To implement and evaluate a super-fast and high-quality biparametric MRI (bpMRI) protocol for prostate imaging acquired at a new ultra-high gradient 3.0-T MRI system.
Methods
Participants with clinically suspected prostate cancer prospectively underwent a multiparametric MRI (mpMRI) on a new 3.0-T MRI scanner (maximum gradient strength: 200 mT/m, maximum slew rate: 200 T/m/s). The bpMRI protocol was extracted from the full mpMRI protocol, including axial T2-weighted and diffusion-weighted (DWI) sequences (b0/800, b1500). Overall image quality was rated by two readers on a five-point Likert scale from (1) non-diagnostic to (5) excellent. PI-RADS 2.1 scores were assessed by three readers separately for the bpMRI and mpMRI protocols. Cohen’s and Fleiss’ κ were calculated for PI-RADS agreement between protocols and interrater reliability between readers, respectively.
Results
Seventy-seven male participants (mean age, 66 ± 8 years) were included. Acquisition time of the bpMRI protocol was reduced by 62% (bpMRI: 5 min, 33 ± 21 s; mpMRI: 14 min, 50 ± 42 s). The bpMRI protocol showed excellent overall image quality for both the T2-weighted (median score both readers: 5 [IQR: 4–5]) and DWI (b1500) sequence (median score reader 1: 4 [IQR: 4–5]; reader 2: 4 [IQR: 4–4]). PI-RADS score agreement between protocols was excellent (Cohen’s κ range: 0.91–0.95 [95% CI: 0.89, 0.99]) with an overall good interrater reliability (Fleiss’ κ, 0.86 [95% CI: 0.80, 0.92]).
Conclusion
Ultra-high gradient MRI allows the establishment of a high-quality and rapidly acquired bpMRI with high PI-RADS agreement to a full mpMRI protocol.
Trials registration
Clinicaltrials.gov, NCT06244680, Registered 06 February 2024, retrospectively registered, https://classic.clinicaltrials.gov/ct2/show/NCT06244680.
Critical relevance statement
A novel 3.0-Tesla MRI system with an ultra-high gradient performance enabled high-quality biparametric prostate MRI in 5.5 min while achieving excellent PI-RADS agreement with a standard multiparametric protocol.
Key Points
Multi- and biparametric prostate MRIs were prospectively acquired utilizing a maximum gradient of 200 mT/m.
Super-fast biparametric MRIs showed excellent image quality and had high PI-RADS agreement with multiparametric MRIs.
Implementation of high gradient MRI in clinical routine allows accelerated and high-quality biparametric prostate examinations.
Graphical Abstract
Keywords: Multiparametric MRI, Biparametric MRI, Prostate cancer, PI-RADS, Ultra-high gradient strength
Introduction
Multiparametric MRI (mpMRI) of the prostate has become an important non-invasive diagnostic tool for the assessment of suspected prostate cancer and is considered the baseline for MRI-targeted biopsy [1, 2]. As the incidence of prostate cancer is high, with an estimated 290,000 new cases in 2023 in the United States alone [3], the need for widespread and high-quality provision of prostate MRI will continue to rise. However, current clinical MRI protocols are long, with acquisition times of > 30 min, potentially limiting the absolute number of MRI scans. According to the guidelines of the Prostate Imaging Reporting and Data System (PI-RADS), a sufficient mpMRI protocol must include diffusion-weighted imaging (DWI), T2-weighted (T2w) imaging, dynamic contrast-enhanced imaging, and T1-weighted pre- and post-contrast images [4, 5]. Different approaches have been developed to shorten the protocol itself or to accelerate acquisition times. For instance, a significant reduction of T2w-sequence acquisition times was achieved by employing deep learning reconstruction methods [6, 7] or the utilization of advanced compressed sensing [8]. Different studies showed equal performance of biparametric MRI (bpMRI) protocols compared to the standard multiparametric protocols, effectively reducing acquisition times [9–11]. This was done by focusing only on DWI and T2w imaging while omitting the dynamic contrast-enhanced sequence and T1-weighted sequences, as the additional diagnostic value of these is controversial [12].
Recent developments in MRI scanner hardware have allowed for ultra-high diffusion gradient strengths (up to 500 mT/m), thus reducing echo times and acquisition times by establishing faster diffusion gradients [13, 14]. Furthermore, these gradients can image at smaller scales with a higher signal-to-noise ratio, consecutively enhancing sensitivity for the detection of tissue microstructures by a higher spatial resolution [15, 16]. Due to the experimental nature of these gradients, this technique has only been investigated in research settings in healthy volunteers [17]. However, recently introduced high gradient 3.0-T MRI technology with clinical FDA clearance gives the opportunity to evaluate high gradient imaging in whole-body MRI, including prostate imaging.
Therefore, our study aimed to implement a super-fast and high-quality abbreviated bpMRI protocol for patients with suspicion of prostate cancer using ultra-high gradient DWI. Aside from assessing acquisition times and image quality, the main objective of this study was to compare PI-RADS scores in the bpMRI protocol with the full mpMRI protocol.
Materials and methods
Study population
This prospective study was approved by the institutional review committee of the University Hospital Bonn (Clinicaltrials.gov: NCT06244680) and followed the principles of the Declaration of Helsinki from 2013. After obtaining written informed consent, male participants with suspicion of prostate cancer, defined by elevated levels of prostate-specific antigen (> 4 ng/mL), suspicious digital rectal and/or transrectal ultrasound exams, were continuously and consecutively enrolled in this study between September and November 2023. Participants with general contraindications for MRI at 3.0 T (e.g., total hip replacements of both sides) or contraindication for administration of gadolinium-based contrast media, as well as severe claustrophobia, were excluded from the study.
Image acquisition
All examinations were performed on a 3.0-T MRI scanner (Siemens MAGNETOM Cima.X, Siemens Healthineers) using an additional anterior phased array coil with 18 channels for signal reception. The whole-body MRI system employs a maximum gradient strength of 200 mT/m and a maximum slew rate of 200 T/m/s, whereby the maximum gradient amplitude might be limited due to adverse effects on the heart, e.g., cardiac stimulation. Participants with no contraindications received 1 mL hyoscine butylbromide (Butylscopolamin 20 mg/mL, Panpharma) once prior to MRI to reduce bowel peristalsis. The contrast agent gadoteric acid (Clariscan, GE Healthcare) was used for post-contrast T1-weighted sequences. A full mpMRI protocol was acquired as the reference standard, consisting of an axial, sagittal and coronal T2w turbo spin-echo sequence, an axial DWI sequence with b-values of 0 and 800 s/mm2 (the apparent diffusion coefficient map was calculated from these b-values), an axial DWI sequence with a b-value of 1500 s/mm2, a T1-weighted turbo spin-echo Dixon sequence pre and post-contrast media administration, and a dynamic contrast enhanced sequence (Time-resolved angiography With Stochastic Trajectories, TWIST). DWI sequences were acquired using echo-planar imaging (EPI) with zonally-magnified oblique multislice acquisition (ZOOM) and employed a restricted maximal gradient strength of 111 mT/m. Vendor-specific, commercially available deep learning techniques for denoising were utilized for reconstructions of T2w and DWI sequences (Deep Resolve Boost, Siemens Healthineers). Additionally, resolution upscaling was used for T2w sequences (Deep Resolve Sharp, Siemens Healthineers). Only the axial T2w sequence and DWI sequences, which were acquired during the mpMRI, were used for the interpretation of the biparametric protocol. Acquisition times per sequence were added to calculate the overall acquisition times of the bpMRI and mpMRI protocols. Detailed sequence parameters are specified in Table 1.
Table 1.
T2w TSE axial* | DWI axial (b0,b800)* | DWI axial (b1500)* | T2w TSE sagittal | T2w TSE coronal | T1w TSE axial | 3D T1w DCE | |
---|---|---|---|---|---|---|---|
Image matrix | 384 × 384 | 46 × 90 | 46 × 90 | 384 × 384 | 384 × 384 | 280 × 352 | 160 × 160 |
Field of view (mm2) | 200 × 200 | 102 × 200 | 102 × 200 | 200 × 200 | 200 × 200 | 255 × 320 | 200 × 200 |
Spatial resolution (mm3) | 0.5 × 0.5 × 3 | 2.2 × 2.2 × 3 | 2.2 × 2.2 × 3 | 0.5 × 0.5 × 3 | 0.5 × 0.5 × 3 | 0.9 × 0.9 × 2.5 | 1.3 × 1.3 × 3 |
Slice thickness (mm) | 3 | 3 | 3 | 3 | 3 | 2.5 | 3 |
Standard slice number (n) | 26 | 26 | 26 | 24 | 25 | 88 | 24 |
Echo time (ms) | 96 | 54 | 57 | 107 | 107 | 2.46 | 1.5 |
Repetition time (ms) | 8560 | 3200 | 3200 | 6990 | 7130 | 5.51 | 4.41 |
Flip angle (degree) | 90 | 90 | 90 | 90 | 90 | 10 | 15 |
Averages | 1 | 2 (b0), 8 (b800) | 7 | 1 | 1 | 1 | 1 |
Diffusion gradients | - | 4 directions (4-Scan Trace) | 4 directions (4-Scan Trace) | - | - | - | - |
Bandwidth (Hz/pixel) | 200 | 1684 | 1684 | 200 | 200 | 592 | 679 |
Acquisition time (s) | 105 | 117 | 104 | 99 | 101 | 70 | 203 |
Temporal resolution (s) | - | - | - | - | - | - | 4.22 |
Fat suppression | - | Fat-Sat | Fat-Sat | - | - | Dixon | - |
DL denoising | + | + | + | + | + | - | - |
DL resolution upscaling | + | - | - | + | + | - | - |
DCE dynamic contrast enhanced, DL deep learning, DWI diffusion-weighted imaging, SPIR spectral presaturation with inversion recovery, TSE turbo spin-echo, T1w T1-weighted, T2w T2-weighted
* Included in the biparametric protocol
Qualitative image analysis
Qualitative image rating of the bpMRI protocol for the T2w sequence and both DWI sequences was performed by two radiologists with 3 (L.M.B.) and 12 (J.A.L.) years of experience in prostate MRI on a five-point Likert scale for six different qualitative categories (artifacts, image sharpness, lesion conspicuity, capsule delineation, overall image sharpness, and diagnostic confidence). Grading was defined as follows: 1, non-diagnostic due to extensive artifacts, strongly impaired conspicuity of anatomical structures, and no diagnostic confidence; 2, several artifacts, difficult conspicuity of anatomical structures, and low diagnostic confidence; 3, moderate artifacts, fair conspicuity of anatomical structures, and moderate diagnostic confidence; 4, little artifacts, good conspicuity of anatomical structures, and good diagnostic confidence; 5, no artifacts, excellent conspicuity of anatomical structures, and high diagnostic confidence. The artifacts category included movement artifacts, metal artifacts, adjacent air artifacts of the rectum, and Gibbs Ringing artifacts. The results of both raters were averaged.
Quantitative image analysis
An equal-sized region of interest (30 mm2) was drawn in the healthy peripheral zone of the prostate and in the internal obturator muscle on both the axial T2w sequence and on the DWI (b1500) sequence. The apparent signal-to-noise ratio (signal intensity in the peripheral zone divided by standard deviation of the signal intensity in the muscle) and the apparent contrast-to-noise ratio ((signal intensity of the peripheral zone minus signal intensity of muscle) divided by standard deviation of the signal intensity in the muscle) were calculated as previously described [6].
PI-RADS assessment and agreement
Three radiologists with 3 (L.M.B.), 12 (J.A.L.), and 8 (C.E.) years of experience in prostate MRI separately read the bpMRI protocol and graded the lesions according to the PI-RADS classification. Only the highest-graded lesion and its respective prostate zone were noted. If there were two distinct lesions with the same high PI-RADS score in both the peripheral and transition zone, both were noted. Participants were presented in a random order. All readers were blinded to any personal and clinical parameters (e.g., name, age, patient history, value of the prostate-specific antigen, clinical examination, and transrectal ultrasound). After a washout period of one month, all readers repeated the same assessment for the mpMRI protocol.
Statistical analysis
Statistical analysis was conducted by L.M.B. using SPSS (Version 27, IBM Corp.). The sample size of this study was chosen equivalent to previous biparametric prostate MRI studies (18–20). Continuous variables are given as mean ± standard deviation, whereas discrete variables are given as median and interquartile range (IQR) and binary variables as absolute percentages. Both, the agreement of bpMRI and mpMRI PI-RADS scores for the whole prostate, and for the specific zonal distribution (peripheral and transition zone) were assessed by calculation of Cohen’s κ, interpreted as follows: < 0.5 = poor; 0.5–0.75 = moderate; 0.75–0.9 = good; > 0.9 = excellent. Interrater reliability between readers for PI-RADS scores was assessed equivalently with Fleiss’ κ, while interrater reliability between raters for qualitative analysis was assessed with Cohen’s κ with equal interpretation as above. A p-value of < 0.05 was considered statistically significant in all cases.
Results
Clinical characteristics of participants
A total of 77 male participants with a mean age of 66 ± 7 years (range: 53 to 84 years) were included in the study after exclusion of four participants due to general MRI contraindications, fifteen participants due to refusal of study participation and six participants due to incomplete protocol acquisition (Fig. 1). 90% (69/77) of participants had an elevated prostate-specific antigen of > 4 ng/mL, whereas 16% (12/77) had a suspicious digital rectal exam, and 10% (8/77) had a suspicious transrectal ultrasound. 53% (41/77) of participants had an initial PI-RADS score of ≥ 3 and 38% (29/77) underwent subsequent biopsy with confirmed malignancy in 22% (17/77). Median International Society of Uropathology grade was 2 (IQR, 1–2). Detailed clinical characteristics of enrolled participants are given in Table 2. For detailed biopsy results, see Table S1.
Table 2.
Variable | Value |
---|---|
No. of participants | 77 |
Age (years) | 66 ± 8 |
Prostate-specific antigen (ng/mL) | 8.7 ± 7.2 |
Suspicious digital rectal exam | 12 (16) |
Suspicious transrectal ultrasound | 8 (10) |
Prior biopsy | 17 (22) |
PI-RADS score | |
1 | 0 (0) |
2 | 36 (47) |
3 | 10 (13) |
4 | 19 (25) |
5 | 12 (16) |
Continuous data is reported as mean ± standard deviation and dichotomous data as number of participants with percentages in parentheses
PI-RADS Prostate Imaging Reporting and Data System
Evaluation of image quality
No sequence of the bpMRI protocol was rated as non-diagnostic in any category by any rater. All sequences of the bpMRI protocol had a good or excellent overall image quality (averaged median score T2w: 5 [4, 5]; averaged median score b0/800: 4 [4, 5]; averaged median score b1500: 4 [4, 5]), while diagnostic confidence was equally good or excellent (averaged median score T2w: 5 [4, 5]; averaged median score b0/800: 4 [4]; averaged median score b1500: 5 [4, 5]) (Fig. 2a). Separate median scores per rater are shown in Table 3. Interrater agreement between readers was good, with a Cohen’s κ of 0.78 [95% CI: 0.76, 0.81].
Table 3.
Rater | Category | T2w TSE | DWI b0/800 | b1500 |
---|---|---|---|---|
1 | Artifacts | 4 [IQR: 4–4] | 4 [IQR: 4–4] | 4 [IQR: 4–4] |
Image sharpness | 5 [IQR: 4–5] | 4 [IQR: 4–5] | 3 [IQR: 3–4] | |
Lesion conspicuity | 5 [IQR: 4–5] | 4 [IQR: 3–4] | 5 [IQR: 4–5] | |
Capsule delineation | 4 [IQR: 4–5] | 5 [IQR: 4–5] | 4 [IQR: 3–4] | |
Overall image quality | 5 [IQR: 4–5] | 4 [IQR: 4–5] | 4 [IQR: 4–5] | |
Diagnostic confidence | 5 [IQR: 4–5] | 4 [IQR: 4–4] | 5 [IQR: 4–5] | |
2 | Artifacts | 4 [IQR: 4–4] | 4 [IQR: 4–4] | 4 [IQR: 4–4] |
Image sharpness | 5 [IQR: 4–5] | 4 [IQR: 4–5] | 3 [IQR: 3–4] | |
Lesion conspicuity | 4 [IQR: 4–5] | 4 [IQR: 3–4] | 5 [IQR: 4–5] | |
Capsule delineation | 4 [IQR: 4–5] | 5 [IQR: 4–5] | 4 [IQR: 3–4] | |
Overall image quality | 5 [IQR: 4–5] | 4 [IQR: 4–4] | 4 [IQR: 4–4] | |
Diagnostic confidence | 4 [IQR: 4–5] | 4 [IQR: 4–4] | 5 [IQR: 4–5] |
Variables are shown as median and interquartile range (IQR)
DWI diffusion-weighted imaging, T2w TSE T2-weighted turbo spin echo
The mean total acquisition time of the bpMRI protocol was reduced to 5 min, 33 ± 21 s compared to the mpMRI protocol with 14 min, 50 ± 42 s. The mean apparent signal-to-noise ratio of the axial T2w sequence was 45.0 ± 7.4, while the mean contrast-to-noise ratio was 40.1 ± 7.1. The mean signal-to-noise and contrast-to-noise ratio of the DWI (b1500) sequence were 9.8 ± 3.0 and 5.3 ± 3.0, respectively (Fig. 2b). Representative images of participants are shown in Figs. 3 and 4.
Comparison of PI-RADS scores
Agreement of PI-RADS scores for the bpMRI and mpMRI protocols with respect to the whole prostate was excellent for all raters (Cohen’s κ reader 1: 0.95 [95% CI: 0.91, 0.99]; reader 2: 0.95 [95% CI: 0.91, 0.99]; reader 3: 0.94 [95% CI: 0.89, 0.99]). PI-RADS scores for bpMRI and mpMRI protocols were similar and their agreement for all lesions in the peripheral zone and the transition zone was equivalently excellent (Table 4). Overall interrater agreement was good (Fleiss’ κ, 0.86 [95% CI: 0.80, 0.92]). Contingency tables for the correlation of the bpMRI and mpMRI for all three readers are shown in Table 5.
Table 4.
Parameter | Reader 1 | Reader 2 | Reader 3 |
---|---|---|---|
Whole prostate | |||
PI-RADS score bpMRI | 2 [IQR: 2–4] | 2 [IQR: 2–4] | 2 [IQR: 2–4] |
PI-RADS score mpMRI | 2 [IQR: 2–4] | 2 [IQR: 2–4] | 2 [IQR: 2–4] |
Agreement | 0.95 (95% CI: 0.91, 0.99) | 0.95 (95% CI: 0.91, 0.99) | 0.94 (95% CI: 0.89, 0.99) |
Peripheral zone | |||
PI-RADS score bpMRI | 2 [IQR: 2–2] | 2 [IQR: 2–2] | 2 [IQR: 2–2] |
PI-RADS score mpMRI | 2 [IQR: 2–2] | 2 [IQR: 2–2] | 2 [IQR: 2–2] |
Agreement | 0.96 (95% CI: 0.92, 1.0) | 0.96 (95% CI: 0.91, 1.0) | 0.94 (95% CI: 0.88, 0.99) |
Transition zone | |||
PI-RADS score bpMRI | 3 [IQR: 2–4] | 2 [IQR: 2–4] | 3 [IQR: 2–4] |
PI-RADS score mpMRI | 3 [IQR: 2–4] | 2 [IQR: 2–4] | 3 [IQR: 2–4] |
Agreement | 0.96 (95% CI: 0.88, 1.0) | 0.95 (95% CI: 0.87, 1.0) | 1.0 (95% CI: 1.0, 1.0) |
PI-RADS scores are reported as median and interquartile range (IQR). Agreement is reported as Cohen’s κ with their respective 95% confidence intervals (CI)
bpMRI biparametric MRI, mpMRI multiparametric MRI
Table 5.
Reader 1 | bpMRI | ||||||
---|---|---|---|---|---|---|---|
n = 0 | n = 38 | n = 8 | n = 18 | n = 13 | |||
PI-RADS scores (n = 77) | 1 | 2 | 3 | 4 | 5 | ||
n = 0 | 1 | 0 | 0 | 0 | 0 | 0 | |
n = 38 | 2 | 0 | 38 | 0 | 0 | 0 | |
mpMRI | n = 4 | 3 | 0 | 0 | 4 | 0 | 0 |
n = 22 | 4 | 0 | 0 | 4 | 18 | 0 | |
n = 13 | 5 | 0 | 0 | 0 | 1 | 12 |
Reader 2 | bpMRI | ||||||
---|---|---|---|---|---|---|---|
n = 0 | n = 41 | n = 8 | n = 15 | n = 13 | |||
PI-RADS scores (n = 77) | 1 | 2 | 3 | 4 | 5 | ||
n = 0 | 1 | 0 | 0 | 0 | 0 | 0 | |
n = 41 | 2 | 0 | 41 | 0 | 0 | 0 | |
mpMRI | n = 4 | 3 | 0 | 0 | 4 | 0 | 0 |
n = 19 | 4 | 0 | 0 | 4 | 15 | 0 | |
n = 13 | 5 | 0 | 0 | 0 | 1 | 12 |
Reader 3 | bpMRI | ||||||
---|---|---|---|---|---|---|---|
n = 0 | n = 38 | n = 10 | n = 17 | n = 12 | |||
PI-RADS scores (n = 77) | 1 | 2 | 3 | 4 | 5 | ||
n = 0 | 1 | 0 | 0 | 0 | 0 | 0 | |
n = 38 | 2 | 0 | 38 | 0 | 0 | 0 | |
mpMRI | n = 4 | 3 | 0 | 0 | 4 | 0 | 0 |
n = 23 | 4 | 0 | 0 | 6 | 17 | 0 | |
n = 12 | 5 | 0 | 0 | 0 | 0 | 12 |
If the same PI-RADS score was given for the bpMRI and mpMRI protocol, the corresponding value (number of patients with this PI-RADS score) was written bold.
Data are absolute numbers
bpMRI biparametric MRI, mpMRI multiparametric MRI, PI-RADS Prostate Imaging Reporting and Data System
Discussion
New technological advancements in ultra-high gradient performance 3.0-T MRI can be used to shorten echo times and therefore accelerate sequence acquisition times (i.e., for diffusion imaging) [16]. Therefore, this study explored the use of a new clinical MRI scanner with a maximum gradient strength of 200 mT/m in combination with commercially available deep learning (DL) reconstruction to investigate the feasibility of a super-fast and high-quality biparametric MRI (bpMRI) of the prostate. This study aimed to compare the assigned Prostate Imaging Reporting and Data System (PI-RADS) scores of this bpMRI protocol with the scores of a standard multiparametric MRI (mpMRI) protocol. A second objective was to assess the image quality of the bpMRI protocol sequences. We found the abbreviated bpMRI protocol to have an excellent PI-RADS agreement with the standard mpMRI protocol, while both T2-weighted (T2w) and diffusion-weighted imaging (DWI) sequences had an excellent overall image quality with a low mean total acquisition time of 5 min and 33 ± 21 s.
In contrast to previous clinical scanners, this MRI system reaches ultra-high gradients of up to 200 mT/m. It should however be noted that these maximum values are not yet fully applicable for full body imaging in clinical routine due to cardiac stimulation limitations, thus the maximum gradient strength in our study was limited to 111 mT/m by the software. However, this gradient strength directly translated to a reduction in echo and repetition times. This becomes particularly clear when comparing the scan parameter adjustments of our sequence made by another 3-Tesla system of the same manufacturer (Siemens MAGNETOM Vida, Siemens Healthineers) after copying the sequence with otherwise identical parameter specifications. For instance, the echo time of the DWI (b1500) sequence could be reduced by 10% to 57 ms and the repetition time by 35% to 3200 ms. Furthermore, in two recent studies [6, 8], in which all qualitative and quantitative parameters were evaluated methodically equivalent by the same raters, the image quality in standard sequences was significantly worse compared to the current study. For instance, in the most recent study [6], the overall image quality of standard non-cartesian T2w sequences was rated as good with a median score of 4 [IQR: 3–4], while the T2w sequence in this study was rated as excellent with a median score of 5 [4, 5]. Equivalently, the aSNR was significantly lower in the standard non-cartesian T2w sequence (25.8 ± 4.6) versus the T2w sequence in the current study (45.0 ± 7.4).
According to the PI-RADS classification [4], a multiparametric MRI of the prostate is mandatory for a thorough evaluation of prostate cancer and consists of T1-weighted, T2-weighted (T2w), DWI and dynamic contrast-enhanced imaging. However, a clear recommendation for the use of primarily biparametric protocols, consisting only of T2w imaging and DWI, is not given by the PI-RADS classification. Several studies already have shown similar accuracy and sensitivity of bpMRI compared to mpMRI [18–21], but the bpMRI protocols investigated in these studies still had long acquisition times of > 10 min due to conventional reconstruction techniques like wavelet transformation [22] or acquisition of several planes for T2w imaging. Furthermore, these studies did not investigate separate b-values in a biparametric approach or achieved both, high image quality and low acquisition times. In contrast to this, we utilized high gradients and slew rates to acquire a fast bpMRI with an axial T2w sequence, a low b-value DWI (b0,800) sequence for computing the apparent diffusion coefficient and a high b-vaIue DWI (b1500) sequence. This resulted in a low total mean acquisition time of 5 min and 33 ± 21 s for the bpMRI versus 14 min, 50 ± 42 s for the mpMRI.
Another advancement and reason for the reduction in acquisition time is the employment of DL denoising and resolution upscaling. While these techniques already have been investigated in T2w imaging of the prostate [6, 23], it has also been shown to be promising for the reconstruction of DWI sequences [24, 25]. Similar to these studies, we found our whole bpMRI protocol to have a mostly good or excellent image quality for both T2w and DWI sequences when reconstructed with this technique, e.g., in image sharpness of the axial T2w sequence or lesion conspicuity of the DWI (b1500) sequence. Quantitative analysis mainly confirmed these findings when compared to literature and showed similar high values for the apparent signal-to-noise ratio and apparent contrast-to-noise ratio in T2w sequences [6], but had slightly lower values for the apparent signal-to-noise ratio and apparent contrast-to-noise ratio in DWI sequences [24], which could be due to small differences in the anatomic region (pathologic versus physiologic) that is ultimately measured.
Finally, we found a high PI-RADS agreement among readers between the bpMRI and mpMRI protocols with changed PI-RADS scores in only 4–6 participants (5–8%). After stratifying assigned PI-RADS scores for the peripheral and transition zones, we found no difference between protocols for the transition zone, while the agreement was slightly lower for the peripheral zone. This is mainly explained by the role of the dynamic contrast-enhanced sequence, as it is only relevant for further evaluation of PI-RADS 3 lesions in the peripheral zone. Additionally, the incidence of PI-RADS 3 lesions in our study was low (bpMRI readings: 8–10 participants [10–13%]), whereas prior bpMRI studies reported higher incidences of > 30% [26, 27]. It remains unclear, if this is coincidently or due to the use of a high gradient MRI scanner with DL reconstruction, as we did not investigate intraindividual comparisons between different scanners.
Our study has some limitations. First, not all participants underwent subsequent biopsy due to refusal of the procedure or low initial PI-RADS scores. Thus, correlation of histopathologic findings with PI-RADS scores and calculation of accuracy of the bpMRI versus the mpMRI was not possible, as it would be strongly biased towards higher PI-RADS scores. Additional studies with biopsies of all participants are needed. Second, the most significant lesions (PI-RADS 3–5) were found in the peripheral zone and were underrepresented in the transition zone. This potentially led to a bias towards the peripheral zone in the evaluation of the bpMRI. However, as DCE is mainly necessary for further assessment of PI-RADS 3 lesions in the peripheral zone, we still had adequate sample sizes that reflect the real world. Third, the study was carried out on a new 3.0-T system with high gradient strength. While this scanner is affordable for large hospitals, smaller hospitals or outpatient centers may have difficulties to buy and establish this scanner, thus it may take a long time for widespread establishment of this technique. Fourth, although MRI acquisition time can be significantly reduced, sufficient time must be allotted for MRI patient preparation. However, productivity could be even more increased by improvements in workflow and modern scheduling software.
In conclusion, we established a high-quality, 5-min biparametric MRI protocol for imaging suspected prostate cancer patients utilizing ultra-high gradients and deep learning reconstructions on a novel MRI scanner. Furthermore, the PI-RADS agreement between the biparametric and the full multiparametric protocol was excellent, thus we firmly believe this technique and this focused examination approach make the detection of prostate cancer more efficient. However, full correlation with histopathological analysis in future studies must be further addressed for valid calculation of diagnostic performance and accuracy.
Supplementary information
Abbreviations
- bpMRI
Biparametric magnetic resonance imaging
- DWI
Diffusion-weighted imaging
- IQR
Interquartile range
- mpMRI
Multiparametric magnetic resonance imaging
- PI-RADS
Prostate Imaging Reporting and Data System
- T2w
T2-weighted
Author contributions
P.K., J.E. and N.K. were responsible for initial screening of prostate cancer screening in the Clinic of Urology. L.B., A.S., P.M. and J.L. were responsible for the hardware configuration and the MRI protocol implementation. L.B., C.E. and J.L. analyzed, evaluated and interpreted the patient data. All authors contributed to the writing of the manuscript. All authors read and approved the final manuscript.
Funding
Open Access funding enabled and organized by Projekt DEAL.
Data availability
Data reported in this study is available by the corresponding author upon reasonable request.
Declarations
Ethics approval and consent to participate
The institutional review committee (Ethikkommission Bonn) approved the prospective study under reference number 198/23-EP. Oral and written informed consent was acquired from any individual in this study before participation.
Consent for publication
Consent for publication was acquired from any individual in this study before participation.
Competing interests
A.I.: Grants from the Medical Faculty Bonn and the German Research Foundation. J.E.: Editorial Board member of Cancers. U.A.: Speakers bureau payments from Siemens; advisory board member for Bayer. Editorial Board member of Investigative Radiology and Die Radiologie. C.P.: Payment for lectures from Guerbet and Julius Zorn; support for attending meetings or travel from Guerbet and Julius Zorn; advisory board member for Guerbet. Editorial Board member of Investigative Radiology. J.L.: Consulting fees from Bayer; payment for presentation from Philips. Editorial Board member of Investigative Radiology, Radiology: Cardiothoracic Imaging, Die Radiologie and Frontiers in Cardiovascular Medicine. The other authors declare that they have no competing interests.
Footnotes
Publisher’s Note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
The online version contains supplementary material available at 10.1186/s13244-024-01862-x.
References
- 1.Hugosson J, Månsson M, Wallström J et al (2022) Prostate cancer screening with PSA and MRI followed by targeted biopsy only. N Engl J Med 387:2126–2137. 10.1056/NEJMoa2209454 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Eklund M, Jäderling F, Discacciati A et al (2021) MRI-targeted or standard biopsy in prostate cancer screening. N Engl J Med 385:908–920. 10.1056/NEJMoa2100852 [DOI] [PubMed] [Google Scholar]
- 3.Siegel RL, Miller KD, Wagle NS, Jemal A (2023) Cancer statistics, 2023. CA Cancer J Clin 73:17–48. 10.3322/caac.21763 [DOI] [PubMed] [Google Scholar]
- 4.Turkbey B, Rosenkrantz AB, Haider MA et al (2019) Prostate Imaging Reporting and Data System Version 2.1: 2019 Update of Prostate Imaging Reporting and Data System Version 2. Eur Urol 76:340–351. 10.1016/j.eururo.2019.02.033 [DOI] [PubMed]
- 5.Hegde JV, Mulkern RV, Panych LP et al (2013) Multiparametric MRI of prostate cancer: an update on state-of-the-art techniques and their performance in detecting and localizing prostate cancer. J Magn Reson Imaging 37:1035–1054. 10.1002/jmri.23860 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Bischoff LM, Peeters JM, Weinhold L et al (2023) Deep learning super-resolution reconstruction for fast and motion-robust T2-weighted prostate MRI. Radiology 308:e230427. 10.1148/radiol.230427 [DOI] [PubMed] [Google Scholar]
- 7.Gassenmaier S, Warm V, Nickel D et al (2023) Thin-slice prostate MRI enabled by deep learning image reconstruction. Cancers (Basel) 15:578. 10.3390/cancers15030578 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Bischoff LM, Katemann C, Isaak A et al (2023) T2 turbo spin echo with compressed sensing and propeller acquisition (sampling k-space by utilizing rotating blades) for fast and motion robust prostate MRI: comparison with conventional acquisition. Invest Radiol 58:209–215. 10.1097/RLI.0000000000000923 [DOI] [PubMed] [Google Scholar]
- 9.Weiss J, Martirosian P, Notohamiprodjo M et al (2018) Implementation of a 5-minute magnetic resonance imaging screening protocol for prostate cancer in men with elevated prostate-specific antigen before biopsy. Invest Radiol 53:186–190. 10.1097/RLI.0000000000000427 [DOI] [PubMed] [Google Scholar]
- 10.Scialpi M, Prosperi E, D’Andrea A et al (2017) Biparametric versus multiparametric MRI with non-endorectal coil at 3T in the detection and localization of prostate cancer. Anticancer Res 37:1263–1271. 10.21873/anticanres.11443 [DOI] [PubMed] [Google Scholar]
- 11.Rais-Bahrami S, Siddiqui MM, Vourganti S et al (2015) Diagnostic value of biparametric magnetic resonance imaging (MRI) as an adjunct to prostate-specific antigen (PSA)-based detection of prostate cancer in men without prior biopsies. BJU Int 115:381–388. 10.1111/bju.12639 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.De Visschere P, Lumen N, Ost P, Decaestecker K, Pattyn E, Villeirs G (2017) Dynamic contrast-enhanced imaging has limited added value over T2-weighted imaging and diffusion-weighted imaging when using PI-RADSv2 for diagnosis of clinically significant prostate cancer in patients with elevated PSA. Clin Radiol 72:23–32. 10.1016/j.crad.2016.09.011 [DOI] [PubMed] [Google Scholar]
- 13.Huang SY, Witzel T, Keil B et al (2021) Connectome 2.0: developing the next-generation ultra-high gradient strength human MRI scanner for bridging studies of the micro-, meso- and macro-connectome. Neuroimage 243:118530. 10.1016/j.neuroimage.2021.118530 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Fan Q, Eichner C, Afzali M et al (2022) Mapping the human connectome using diffusion MRI at 300 mT/m gradient strength: methodological advances and scientific impact. Neuroimage 254:118958. 10.1016/j.neuroimage.2022.118958 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Setsompop K, Kimmlingen R, Eberlein E et al (2013) Pushing the limits of in vivo diffusion MRI for the Human Connectome Project. Neuroimage 80:220–233. 10.1016/j.neuroimage.2013.05.078 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.McNab JA, Edlow BL, Witzel T et al (2013) The Human Connectome Project and beyond: initial applications of 300 mT/m gradients. Neuroimage 80:234–245. 10.1016/j.neuroimage.2013.05.074 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Huang SY, Tian Q, Fan Q et al (2020) High-gradient diffusion MRI reveals distinct estimates of axon diameter index within different white matter tracts in the in vivo human brain. Brain Struct Funct 225:1277–1291. 10.1007/s00429-019-01961-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Stanzione A, Imbriaco M, Cocozza S et al (2016) Biparametric 3T magnetic resonance imaging for prostatic cancer detection in a biopsy-naïve patient population: a further improvement of PI-RADS v2? Eur J Radiol 85:2269–2274. 10.1016/j.ejrad.2016.10.009 [DOI] [PubMed] [Google Scholar]
- 19.Thestrup KC, Logager V, Baslev I, Møller JM, Hansen RH, Thomsen HS (2016) Biparametric versus multiparametric MRI in the diagnosis of prostate cancer. Acta Radiol Open 5:2058460116663046. 10.1177/2058460116663046 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Fascelli M, Rais-Bahrami S, Sankineni S et al (2016) Combined biparametric prostate magnetic resonance imaging and prostate-specific antigen in the detection of prostate cancer: a validation study in a biopsy-naive patient population. Urology 88:125–134. 10.1016/j.urology.2015.09.035 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Radtke JP, Boxler S, Kuru TH et al (2015) Improved detection of anterior fibromuscular stroma and transition zone prostate cancer using biparametric and multiparametric MRI with MRI-targeted biopsy and MRI-US fusion guidance. Prostate Cancer Prostatic Dis 18:288–296. 10.1038/pcan.2015.29 [DOI] [PubMed] [Google Scholar]
- 22.Agarwal S, Singh OP, Nagaria D (2017) Analysis and comparison of wavelet transforms for denoising MRI image. Biomed Pharmacol J. 10.13005/bpj/1174
- 23.Harder FN, Weiss K, Amiel T et al (2022) Prospectively accelerated T2-weighted imaging of the prostate by combining compressed SENSE and deep learning in patients with histologically proven prostate cancer. Cancers (Basel) 14:5741. 10.3390/cancers14235741 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Ueda T, Ohno Y, Yamamoto K et al (2022) Deep learning reconstruction of diffusion-weighted MRI improves image quality for prostatic imaging. Radiology 303:373–381. 10.1148/radiol.204097 [DOI] [PubMed] [Google Scholar]
- 25.Ursprung S, Herrmann J, Joos N et al (2023) Accelerated diffusion-weighted imaging of the prostate using deep learning image reconstruction: a retrospective comparison with standard diffusion-weighted imaging. Eur J Radiol 165:110953. 10.1016/j.ejrad.2023.110953 [DOI] [PubMed] [Google Scholar]
- 26.Cai GH, Yang QH, Chen WB, Liu QY, Zeng YR, Zeng YJ (2021) Diagnostic performance of PI-RADS v2, proposed adjusted PI-RADS v2 and biparametric magnetic resonance imaging for prostate cancer detection: a preliminary study. Curr Oncol 28:1823–1834. 10.3390/curroncol28030169 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Tamada T, Kido A, Yamamoto A et al (2021) Comparison of biparametric and multiparametric MRI for clinically significant prostate cancer detection with PI-RADS version 2.1. J Magn Reson Imaging 53:283–291. 10.1002/jmri.27283 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
Data reported in this study is available by the corresponding author upon reasonable request.