Abstract
Radiocapitellar arthroplasty has been shown to improve pain and function in patients with a degenerative joint. Due to problems with the loosening of the radial head component, one of the few available systems was removed from the global market. This offered specific challenges in terms of treatment strategies when one or both components of a system that is no longer available fail. Due to the very different geometry of the capitellar implant, revision of the capitellar component would require a complex procedure, likely requiring bone graft and a high chance of early failure, leaving resection or interposition arthroplasty as the only available option. Although implant mismatch is common practice in hip and knee arthroplasty with satisfactory results, it should remain a salvage option as the off-label use of components gives rise to several medicolegal implications. We report two cases of radiocapitellar arthroplasty partial revision, by replacing only the radial head component with an implant from another system while keeping the well-fixed original capitellar component in place. At a minimum of three-year follow-up, both cases improved from poor to good and excellent Mayo elbow performance scores. There were no signs of implant failure on standard radiographs.
Keywords: Radial head arthroplasty, radial head replacement, radial head prosthesis, revision radial head arthroplasty, elbow, osteoarthritis
Introduction
Isolated symptomatic radiocapitellar degenerative changes are relatively rare1,2 and are mostly idiopathic or post-traumatic in origin. 3 When conservative treatment fails, available surgical options are limited. 4
The first radio-capitellar arthroplasty (RCA), the Lateral Resurfacing Elbow (LRE) (LRE system, Oxford, UK), was introduced in 2005, followed soon after by the UniElbow System (UES) (Stryker, Morrisville, USA). The third RCA that was developed was a custom (Stryker, Morrisville, USA). In all implant designs, the RCA consists of a radial head (RH) component with a polyethylene (PE) surface that articulates with a metal capitellar component. The main difference between these systems is that the LRE can be considered as a resurfacing arthroplasty as it requires only reaming of degenerative joint surfaces, while the UES and custom system are replacement arthroplasties that entail both RH and capitellar resection (Figure 1).
Figure 1.
Radial head component with a polyethylene articulating surface from the CRF-II® prosthesis (Tornier, Montbonnot, France) (courtesy of More Foundation).
The advent of RCA was intended to fill the gap in the surgical management of young patients with degenerative or non-reconstructible joints.5–7 The literature on the outcome after RCA is scarce and mostly based on small case series.8–12 If the available studies show significant improvements in pain, range of motion and stability, the complication rate is also significant. Notably, the increased risk of early loosening of the UES RH component led to its removal from the global market in 2017. Since then, there have been no other options from the same system to revise an RCA failure which offers specific challenges in terms of treatment strategies.
We report two cases of loosening of the UES RH component treated by revision with a bipolar cemented long-stem RH implant with a PE articulating surface, from another system while retaining the well-fixed capitellar component. Patients gave their written informed consent to this procedure.
Case reports
Case 1: A 50-year-old male patient with ankylosing spondylitis was initially treated at another institution by arthroscopic debridement, synovectomy, and loose body removal for a painful loss of function of his right dominant elbow. Primary isolated end-stage RC osteoarthritis was confirmed arthroscopically. As his symptoms did not improve, an RCA with the UES was performed five months later. Postoperatively, the elbow remained painful. Successive standard radiographs showed an overlengthening and progressive loosening of the RH component (Figure 2). Fourteen months after RCA, it was decided to remove the loose RH component and, as a revision with the same system was no longer possible, the surgeon decided to perform an anconeus interposition arthroplasty. This did not resolve the patient's symptoms, and he was referred to the senior author (RVR) one year after the implant had been removed. At this point, the patient reported a visual analogue scale (VAS) for pain was 9, with a 20° extension deficit and flexion up to 120°. Pronation and supination were limited to 10° in both directions, with clear proximal radioulnar impingement and valgus instability (Figure 3). The Mayo elbow performance score (MEPS) was poor with a score of 25, We considered this to be a failure of the anconeus interposition arthroplasty and after discussion of available options with the patient, it was decided, to re-implant an RH component from another system, using a bipolar cemented long-stem implant with a polyethylene articulating surface (CRF-II® prosthesis (Tornier, Montbonnot, France), whilst retaining the original capitellar component. The patient was positioned supine with the arm on a side table and a sterile tourniquet was applied. The previous lateral incision was resumed and extended as needed. Access to the joint was obtained via an extensor split approach. The medullary canal was prepared according to the manufacturer's recommendations. The height of the implant was determined according to the small sigmoid notch. 13 The definitive implant was placed with the prosthesis neck oriented towards Lister's tubercle to ensure correct rotational alignment.
Figure 2.
Anteroposterior (A) and lateral (B) standard radiographs of the right elbow of case 1, 12 months after radiocapitellar arthroplasty with UniElbow system (Stryker, Morrisville, USA). While the capitellar component appears well fixed in the correct position, the radial head component shows signs of overstuffing and progressive loosening (courtesy of More Foundation).
Figure 3.
Computed tomography scan illustrating the first revision of the Radiocapitellar arthroplasty of case 1 by radial head component removal and anconeus interposition arthroplasty. Clear radioulnar impingement is visible (courtesy of More Foundation).
At the final follow-up, 36 months postoperatively, the patient had a VAS for the pain of 2, MEPS of 85 (good), flexion was 120° and extension improved to 10°. Pronation and supination returned to normal with a range of 85° and 90°, respectively. The elbow was stable, the radioulnar impingement had resolved completely and there was no radiographic evidence of implant loosening (Figure 4).
Figure 4.
Anteroposterior (A) and lateral (B) standard radiographs of the right elbow of case 1 27 months after revision of the radial head component of the UniElbow system (Stryker, Morrisville, USA) with a bipolar cemented long-stem implant with a polyethylene articulating surface) CRF-II® prosthesis (Tornier, Montbonnot, France) whilst retaining the capitellar component. At follow-up, both components are in a good position without any signs of loosening (courtesy of More Foundation).
Case 2: A 58-year-old male patient presented with pain and stiffness of his left dominant elbow, six months after a minimally displaced radial head fracture. He underwent an arthroscopic debridement showing a diffuse full-thickness loss of RC cartilage. Postoperatively, ROM returned to normal, but the pain persisted. Intra-articular infiltration of hyaluronic acid resulted in a temporary decrease in pain. Radiographically there was a progression of degenerative changes and pain increased. Two years after the arthroscopy, a RCA was performed. After an initial positive result, a painful progressive loosening of the RH component occurred and led to a failure of the implant at 37 months after the index procedure (Figure 5). It was decided to revise the radial head component as described in the previous case. Before revision surgery, the patient had a VAS for pain of 7 and a MEPS of 55 (poor). The elbow was stable with a fixed flexion deformity of 5° and flexion to 130°. Pronation and supination were 80° in both directions. At the last follow-up at 37 months postoperatively, he had a VAS for pain of 0 to 1 and a MEPS of 100. Elbow ROM and stability were unchanged from the pre-revision status (Figure 6).
Figure 5.
Anteroposterior (A) and lateral (B) standard radiographs of the left elbow of case 2 36 months after radiocapitellar arthroplasty with UniElbow system (Stryker, Morrisville, USA). While the capitellar component appears well fixed in the correct position, the radial head component shows signs of progressive loosening (courtesy of More Foundation).
Figure 6.
Anteroposterior (A) and lateral (B) standard radiographs of the right elbow of case 2 36 months after revision of the radial head component of the UniElbow system (Stryker, Morrisville, USA) with a bipolar cemented long-stem implant (CRF-II® prosthesis (Tornier, Montbonnot, France) with a polyethylene articulating surface whilst retaining the capitellar component. At follow-up, both components remained in a good position without any signs of loosening (courtesy of More Foundation).
Discussion
These two cases illustrated significant improvement in pain and overall function more than three years after RCA partial revision by mismatched implants for painful loosening of the RH component.
RCA revision remains a challenge, as little is known about replacing one or both components. To add complexity, one of the previously available systems, the UES, has since been discontinued as the rHead prosthesis (Stryker, Morrisville, USA), corresponding to the UES RH component, was at increased risk of early loosening. 14 Unsurprisingly, painful loosening of the RH component was also the main complication of the UES, with rates of up to 29%.8–12 The rHead (Stryker, Morrisville, USA) and CRF-II (Tornier, Montbonnot, France) prostheses are both modular, with a bipolar CoCr head and an internal polyethylene coating. While the former has a short press-fit stem, the latter has a long-cemented stem.15,16 This difference in the design and fixation technique of the RH component could be at the root of the presented results.
Implant mismatch is a common practice worldwide in hip and knee arthroplasty. It is generally done in the interest of the patient, aiming to optimize treatment outcomes. However, off-label use of components gives rise to several medicolegal implications. 17 Furthermore, the EFORT implant and Patient Safety Initiative acknowledged that, under certain circumstances, patients may benefit from selective mismatch applications. 18 It is crucial to balance the risks and benefits to the patient, obtain informed consent, and document the decision decision-making process appropriately. Moreover, surgeons must be fully aware of the features of the components that they use during mismatch applications and only combine implants that are compatible. In the presented cases, our decision to perform a partial revision of the UES with an RH implant from another system was driven by patients’ age and functional demand, the absence of capitellar component involvement and UH OA as well as the failure of various other treatment modalities. The decision was taken jointly with the patient, who was fully informed that this was a salvage option.
Other surgical treatment modalities could include implant removal 4 with or without a concomitant anconeus or Achilles interposition arthroplasty19,20 as well as revision to a total elbow arthroplasty (TEA). While the latter is not indicated in an active population without ulnohumeral (UH) OA, implant removal alone, similar to an RH resection, and interposition arthroplasty have proven effective in treating failed RH arthroplasties.21,22 In a retrospective multicentre study of 28 failed RH arthroplasties, Barret et al. 23 observed that implant removal alone and RH arthroplasty revision improved pain and function with similar results at a mean follow-up of more than 5 years. Similarly, Baghdadi et al. 24 found an improvement in pain and function in their retrospective analyses of 29 anconeus interposition arthroplasties at a mean follow-up of 10 years. In addition, the authors included two cases of RCA for which they reported mild and moderate pain and a MEPS of 70 and 85 points, respectively. Unfortunately, this procedure did not lead to an acceptable outcome in our first case, who developed persistent pain, valgus instability and a clear proximal radioulnar impingement, illustrating its potentially less predictable clinical outcomes.25,26
Compared to the latter, RCA has clear benefits like restoring elbow kinematics when the ulnar collateral ligament remains intact 27 and theoretically postpone the need for a TEA. However, the literature lacks comparative data and long-term outcomes between these different techniques. Pending evidence of the superiority of one strategy over another, the treatment plan should be tailored to the cause of RCA failure, as well as to surgeon and patient preferences. As the low volume of RCA makes it commercially less interesting to develop, partial revision by mismatched implants may become a definite solution in the absence of further options.
Conclusion
Failure of a RCA remains a challenge. Little is known about replacing one or both components and overall treatment options are limited. Although RCA partial revision by mismatched implants is a salvage option, it appears to be a viable alternative for retaining an RCA. To our knowledge, these are the first cases of RCA partial revision by implant mismatch. At a medium-term follow-up, both cases had mild pain, a good to excellent MEPS, a functional arc and a stable elbow. In addition, there were no signs of loosening on standard radiographs.
Acknowledgements
None.
Footnotes
Contributorship: TM: conceptualization, investigation, methodology and writing–original draft. PC: conceptualization, investigation, methodology and writing–review and editing. LV: conceptualization, investigation, methodology and writing–review and editing. RvR: conceptualization, investigation, methodology, supervision and writing–review and editing.
The author(s) declared the following potential conflicts of interest with respect to the research, authorship, and/or publication of this article: A declaration form must be submitted for all types of manuscripts. Shoulder & Elbow operates a double-blind peer review process, so please do not include names of any authors or institutions in this form as it will be seen by peer reviewers. The Shoulder & Elbow declaration policy document provides options for many of the declarations below. Please find this in the journal submission guidelines on the journal website.
Ethical approval: AZ MONICA does not require ethical approval for reporting individual cases.
Funding: The author(s) received no financial support for the research, authorship, and/or publication of this article.
Guarantor: RvR
Informed consent: Written informed consent was obtained from all patients for their anonymized information to be published in this article.
ORCID iDs: Tiago Martinho https://orcid.org/0000-0002-8138-0898
Pieter Caekebeke https://orcid.org/0000-0001-8627-9206
Trial registration: Not applicable.
References
- 1.Lim YW, van Riet RP, Mittal R, et al. Pattern of osteophyte distribution in primary osteoarthritis of the elbow. J Shoulder Elbow Surg 2008; 17: 963–966. [DOI] [PubMed] [Google Scholar]
- 2.Ahrens PM, Redfern DR, Forester AJ. Patterns of articular wear in the cadaveric elbow joint. J Shoulder Elbow Surg 2001; 10: 52–56. [DOI] [PubMed] [Google Scholar]
- 3.Ravalli S, Pulici C, Binetti S, et al. An overview of the pathogenesis and treatment of elbow osteoarthritis. J Funct Morphol Kinesiol 2019; 4: 30. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Kachooei AR, Baradaran A, Ebrahimzadeh MH, et al. The rate of radial head prosthesis removal or revision: a systematic review and meta-analysis. J Hand Surg Am 2018; 43: 39–53.e31. [DOI] [PubMed] [Google Scholar]
- 5.Momtaz D, Ahmad F, Cushing T, et al. Radiocapitellar arthroplasty: systematic review. J Shoulder Elb Arthroplast 2023; 7: 24715492231152735. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Giannicola G, Sacchetti FM, Antonietti G, et al. Radial head, radiocapitellar and total elbow arthroplasties: a review of recent literature. Injury 2014; 45: 428–436. [DOI] [PubMed] [Google Scholar]
- 7.Sandefur EP, Hansen BK, Hackley DT, et al. The clinical outcomes of radiocapitellar reconstruction with radiocapitellar arthroplasty: a systematic review and meta-analysis. J Shoulder Elb Arthroplast 2023; 7: 24715492231199339. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Spross C, Jak W, van Riet RP. Radiocapitellar arthroplasty: a consecutive case series with 2 to 6 years’ follow-up. J Shoulder Elbow Surg 2019; 28: 131–136. [DOI] [PubMed] [Google Scholar]
- 9.Kachooei AR, Heesakkers NAM, Heijink A, et al. Radiocapitellar prosthetic arthroplasty: short-term to midterm results of 19 elbows. J Shoulder Elbow Surg 2018; 27: 726–732. [DOI] [PubMed] [Google Scholar]
- 10.Heijink A, Morrey BF, Eygendaal D. Radiocapitellar prosthetic arthroplasty: a report of 6 cases and review of the literature. J Shoulder Elbow Surg 2014; 23: 843–849. [DOI] [PubMed] [Google Scholar]
- 11.Giannicola G, Angeloni R, Mantovani A, et al. Open debridement and radiocapitellar replacement in primary and post-traumatic arthritis of the elbow: a multicenter study. J Shoulder Elbow Surg 2012; 21: 456–463. [DOI] [PubMed] [Google Scholar]
- 12.Bigazzi P, Biondi M, Ceruso M. Radiocapitellar prosthetic arthroplasty in traumatic and post-traumatic complex lesions of the elbow. Eur J Orthop Surg Traumatol 2016; 26: 851–858. [DOI] [PubMed] [Google Scholar]
- 13.van Riet RP, van Glabbeek F, de Weerdt W, et al. Validation of the lesser sigmoid notch of the ulna as a reference point for accurate placement of a prosthesis for the head of the radius: a cadaver study. J Bone Joint Surg Br 2007; 89: 413–416. [DOI] [PubMed] [Google Scholar]
- 14.Laumonerie P, Reina N, Gutierrez C, et al. Tight-fitting radial head prosthesis: does stem size help prevent painful loosening? Int Orthop 2018; 42: 161–167. [DOI] [PubMed] [Google Scholar]
- 15.Viveen J, Kodde IF, Heijink A, et al. Why does radial head arthroplasty fail today? A systematic review of recent literature. EFORT Open Rev 2019; 4: 659–667. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Laumonerie P, Tibbo ME, Reina N, et al. Radial head arthroplasty: a historical perspective. Int Orthop 2019; 43: 1643–1651. [DOI] [PubMed] [Google Scholar]
- 17.Peters RM, Hiemstra JT, Zijlstra WP, et al. To mix or not to mix? Medicolegal implications of mixed components in total hip arthroplasty. Acta Orthop 2020; 91: 624–626. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Tucker K, Gunther KP, Kjaersgaard-Andersen P, et al. EFORT recommendations for off-label use, mix & match and mismatch in hip and knee arthroplasty. EFORT Open Rev 2021; 6: 982–1005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Morrey BF, Schneeberger AG. Anconeus arthroplasty: a new technique for reconstruction of the radiocapitellar and/or proximal radioulnar joint. J Bone Joint Surg Am 2002; 84: 1960–1969. [PubMed] [Google Scholar]
- 20.McLaughlin RE, 2nd, Savoie FH, 3rd, Field LD, et al. Arthroscopic treatment of the arthritic elbow due to primary radiocapitellar arthritis. Arthroscopy 2006; 22: 63–69. [DOI] [PubMed] [Google Scholar]
- 21.Kodde IF, Viveen J, The B, et al. Management of the failed radial head arthroplasty. EFORT Open Rev 2020; 5: 398–407. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Antuña SA, Sánchez-Márquez JM, Barco R. Long-term results of radial head resection following isolated radial head fractures in patients younger than forty years old. J Bone Joint Surg Am 2010; 92: 558–566. [DOI] [PubMed] [Google Scholar]
- 23.Barret H, Mansat P, Langlais T, et al. After failed radial head arthroplasty, what are the options? Risk factors and results of revisions in a multicenter study. J Clin Orthop Trauma 2023; 38: 102128. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Baghdadi YM, Morrey BF, Sanchez-Sotelo J. Anconeus interposition arthroplasty: mid- to long-term results. Clin Orthop Relat Res 2014; 472: 2151–2161. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.van Riet RP, Morrey BF. Delayed valgus instability and proximal migration of the radius after radial head prosthesis failure. J Shoulder Elbow Surg 2010; 19: e7–10. [DOI] [PubMed] [Google Scholar]
- 26.Schiffern A, Bettwieser SP, Porucznik CA, et al. Proximal radial drift following radial head resection. J Shoulder Elbow Surg 2011; 20: 426–433. [DOI] [PubMed] [Google Scholar]
- 27.Sabo MT, Shannon H, De Luce S, et al. Elbow kinematics after radiocapitellar arthroplasty. J Hand Surg Am 2012; 37: 1024–1032. [DOI] [PubMed] [Google Scholar]