Abstract
Background
Basivertebral nerve radiofrequency ablations (BVNRFA) is a relatively new procedure that has demonstrated positive effects to treat chronic low back pain. Fluoroscopy guidance is utilized to access the vertebral body via the pedicle and confirm the correct location of the probe for ablation. Radiation exposure during this procedure has not been previously reported.
Objective
The purpose of this study was to evaluate the average fluoroscopic time and radiation exposure during BVNRFA.
Methods
Patients treated with BVNRFA that had failed conservative treatment, with primarily midline back pain, and corroborating Modic type I or Modic type II changes on MRI at a tertiary academic spine center were retrospectively analyzed. Chart review was conducted to obtain patient demographics, fluoroscopic time and radiation exposure, involvement of trainees, and vertebral levels treated. Average fluoroscopic and radiation exposure was calculated.
Results
A total of 55 patients were included in this study. The average fluoroscopic time was 152.5 s (±84.3 s). The average cumulative dose was 70.3 mGy (±53.0 mGy) and the average dose area product was 7.9 mGy·cm2 (±5.2 mGy·cm2).
Conclusions
Our study demonstrated that the average fluoroscopic time during BVNRFA to be about 2 and a half minutes.
Keywords: Basivertebral nerve ablation, Radiation, Fluoroscopy, Exposure, Radiofrequency
1. Introduction
Basivertebral nerve radiofrequency ablation (BVNRFA) is a relatively new procedure utilized to treat vertebrogenic back pain. BVNRFA has been demonstrated to be an effective procedure in providing long term pain relief [1]. In Conger et al.’s meta-analysis of 12 studies, they found that BVNRFA led to significantly improved pain, disability and function up to 2 years [2].
BVNRFA requires fluoroscopic guidance to access the vertebral body via the pedicle and decrease the risk of spinal nerve injuries. Fluoroscopy guidance also ensures that appropriate anatomical placement of the probe is achieved to provide ablation of the basivertebral nerve (BVN).
Cumulative radiation exposure and dose to patients, physicians, and staff have been associated with negative long-term health outcomes [[3], [4], [5], [6], [7], [8]] specifically cataracts [9] and cancer [[10], [11], [12]]. Protective equipment is utilized during the procedure to decrease and minimize these risks [[13], [14], [15], [16]]. Decreasing fluoroscopic time during procedure use also decreases the overall radiation risk [17]. Patient factors such as BMI [18] and procedural factors such as presence of a trainee [19] have also been shown to impact radiation exposure during spine procedures.
To our knowledge, fluoroscopy time and radiation dose have not been reported for patients undergoing BVNRFA. The purpose of this study was to report average fluoroscopy time and radiation dose for patients undergoing BVNRFA and assess the impact of patient and procedural factors on radiation exposure.
2. Methods
This retrospective study was approved by the Institutional Review Board at the University of Rochester. Consecutive adult patients undergoing BVNRFA between May 01, 2023 and September 30, 2024 were included in this study. All procedures were performed by the senior author (AA) who completed an accredited pain fellowship as well as additional training from the manufacturer of BVNRFA equipment. BVNRFA was indicated for patients if they had primarily midline low back pain without radicular symptoms and corroborating Modic type I or type II changes noted on MRI [20]. BVNRFA was performed as described in the literature [21]. In our study, all patients were treated with 7-min ablations. The months of the provider's practice that each BVN RFA was performed was broken up into 5 3-month time period stages. Of note, there were two months within the 17-month time frame where no BVNRFA was performed.
2.1. Statistical analysis
Mean and standard deviation were calculated for continuous variables. Number and percent were calculated for categorial variables. BMI was classified as normal (BMI <25.0 kg/m2), overweight (BMI = 25.0–29.9 kg/m2), and obese (BMI ≥30.0 kg/m2). A statistical software package (SPSS) was used to analyze the data. Univariate analysis was conducted to assess the impact of BMI classifications, levels treated, S1 involvement, stages in practice, and the presence of a trainee. Finally, controlling for all other variables, a multivariate linear regression analysis was conducted to assess the impact of age, sex, BMI (as a continuous value), involvement of a trainee, stages in practice, levels performed, and involvement of S1 for both fluoroscopy time and cumulative dose. The level of significance was set at 0.05.
3. Results
A total of 55 patients underwent BVNRFA during the inclusion time period. Demographic data is presented in Table 1. The average procedure length was 67.6 min (±25.8 min) and the fluoroscopic time of all the procedures was 152.5 s (±84.3 s). The average fluoroscopic exposure time per level performed was 50.7 s (±19.1 s). The average cumulative dose was 70.3 mGy (±53.0 mGy) and the average dose area product was 7.9 mGy·cm2 (±5.2 mGy·cm2). Univariate analyses evaluating the impact of the number of levels treated, presence of a trainee, patient BMI, stages in practice, and the inclusion of S1 on the average fluoroscopic time (seconds) and average cumulative radiation dose (mGy) are reported in Table 2, Table 3, respectively.
Table 1.
Variable | Mean/N | Standard Deviation/% |
---|---|---|
Age | 65.2 | 13.9 |
Sex | ||
Female | 28 | 51 % |
Male | 27 | 49 % |
BMI | 30.34 | 6.0 |
Normal | 10 | 18.2 % |
Overweight | 23 | 41.8 % |
Obese | 22 | 40.0 % |
Race | ||
White | 52 | 95 % |
Black | 2 | 4 % |
Not Disclosed | 1 | 2 % |
Area Deprivation Index | 58 | 18.0 |
Number of Levels | 3 | 1 |
Involvement of Trainee | 32 | 58 % |
S1 Involvement | ||
Yes | 43 | 78 % |
No | 12 | 22 % |
Procedure Time (minutes) | 67.44 | 25.6 |
Fluoroscopic Time (seconds) | 150.2 | 82.8 |
Dose Area Product (Gy·cm2) | 7.8 | 5.1 |
Cumulative Dose (mGy) | 68.6 | 51.5 |
Table 2.
Variable | Mean Fluoroscopy Time (Standard Deviation) | P value |
---|---|---|
Trainee Involvement | ||
Yes | 151.1 (85.6) | 0.924 |
No | 148.8 (78.4) | |
BMI Classifications | ||
Normal | 137.9 (53.8) | 0.379 |
Overweight | 168.7 (94.8) | |
Obese | 135.8 (74.9) | |
Number of Levels | ||
1 | 52 (5) | <0.001 |
2 | 97.4 (37.4) | |
3 | 125.3 (9.5) | |
4 | 207.2 (80.3) | |
Involvement of S1 | ||
Yes | 164.9 (86.0) | 0.009 |
No | 92.5 (23.8) | |
Stages of Practice | ||
1st | 264.7 (103.2) | <0.001 |
2nd | 153.9 (56.6) | |
3rd | 117.2 (55.9) | |
4th | 147.2 (76.7) | |
5th | 123.1 (54.0) |
Table 3.
Variable | Mean Cumulative Radiation Dose mGy (Standard Deviation) | P value |
---|---|---|
Trainee Involvement | ||
Yes | 61.7 (42.6) | 0.24 |
No | 78.7 (60.9) | |
BMI Classifications | ||
Normal | 47.3 (23.8) | <0.001 |
Overweight | 62.1 (34.2) | |
Obese | 84.7 (67.8) | |
Number of Levels | ||
1 | 19.1 (4.7) | 0.55 |
2 | 62.4 (60.8) | |
3 | 47.3 (5.7) | |
4 | 80.4 (42.8) | |
Involvement of S1 | ||
Yes | 76.0 (54.9) | 0.04 |
No | 39.7 (14.1) | |
Stages of Practice | ||
1st | 88.0 (40.0) | 0.19 |
2nd | 99.8 (77.2) | |
3rd | 69.1 (55.9) | |
4th | 55.8 (23.1) | |
5th | 57.3 (47.9) |
Regression analysis of factors correlating with fluoroscopy time had a R2 of 0.66. Compared to performing 4 levels, performing 1 level (β = −149.4, p = 0.002), 2 levels (β = −106.4, p < 0.001), and 3 levels BVN RFA (β = −82.1, p = 0.04) were significantly correlated with less fluoroscopy time. Compared to the fluoroscopy time in the first stage of practice, there was a significant difference in fluoroscopy time at the 2nd (β = −78, p = 0.02), 3rd (β = −104.8, p = 0.001), and 5th stage of practice (β = −109.9, p < 0.001). Additional multivariate data is presented in Table 4.
Table 4.
Variable | Coefficient (β) | P value |
---|---|---|
Age | −0.04 | 0.583 |
BMI Classifications | 1.97 | 0.17 |
Sex | ||
Female | Reference | |
Male | −8.4 | 0.6 |
Trainee Involvement | ||
No | Reference | |
Yes | 3.2 | 0.87 |
Number of Levels | ||
1 | −149.4 | 0.002 |
2 | −106.4 | <0.001 |
3 | −82.1 | 0.004 |
4 | Reference | |
Involvement of S1 | ||
No | Reference | |
Yes | 4.3 | 0.86 |
Stages of Practice | ||
1st | ||
2nd | −78 | 0.02 |
3rd | −104.8 | 0.001 |
4th | −60.7 | 0.052 |
5th | −109.9 | <0.001 |
Multivariate regression analysis of factors correlating with cumulative radiation dose exposure had a R2 of 0.58. In a multivariate regression analysis controlling all other variables (age, sex, BMI, levels performed, S1 involvement, stage in career, or trainee involvement), increased BMI (β = 5.15, p < 0.001) was significantly correlated with increased cumulative radiation dose exposure. Involvement of S1 was correlated with greater radiation exposure compared to no involvement of S1 (β = 38.5, p = 0.03). Additional multivariate data is presented in Table 5.
Table 5.
Variable | Coefficient (β) | P value |
---|---|---|
Age | −0.25 | 0.58 |
BMI Classifications | 5.15 | <0.001 |
Sex | ||
Female | Reference | |
Male | −20.7 | 0.07 |
Trainee Involvement | ||
No | Reference | |
Yes | −0.2 | 0.99 |
Number of Levels | ||
1 | −40.2 | 0.19 |
2 | −20.2 | 0.17 |
3 | −42.6 | 0.1 |
4 | Reference | |
Involvement of S1 | ||
No | Reference | |
Yes | 38.5 | 0.03 |
Stages of Practice | ||
1st | Reference | |
2nd | 12.8 | 0.57 |
3rd | −14.3 | 0.49 |
4th | −18.1 | 0.39 |
5th | −25.8 | 0.16 |
4. Discussion
This study reported the average fluoroscopic time for BVNRFA to be about two and half minutes. The average effective dose was 71.9 mGy.
The radiation exposure during fluoroscopic interventional spine procedures are risks for future complications for patients and the medical team [[3], [4], [5], [6],8]. Radiation exposure varies based on the spinal procedures. For example, Cohen et al. found that the average fluoroscopy time for lumbar facet injections was shown to be 16.5 s per site and 24.3 s per lumbar transforaminal epidural steroid injection site [22].
According to the US Nuclear regulatory commission the average American is exposed to 6.20 mGy of radiation annually [23]. Taken together, the median BVNRFA is equivalent to 8.5 years of average radiation exposure for Americans and is equivalent to roughly 7.6 chest CTs [23]. Crawley and Rogers originally described radiation exposure for common fluoroscopic orthopaedic surgeries to establish reference doses for radiological procedures to optimize patient care. Their study reported the median DAP for hip and spine procedures to be approximately 10 Gy · cm2 [24]. Cohen et al. proposed reference values for radiation dose for fluoroscopically guided lumbar procedures as 13 mGy for transforaminal/intralaminar injections, 7 mGy for RFA, and 4 mGy for medial branch blocks [22]. The DAP reported by Crawley and Rogers's study is similar the BVNRFA radiation exposure reported by this study. These findings suggest BVNRFA radiation exposure is on the order of orthopaedic surgeries and less similar to commonly performed interventional pain procedures for the lumbar spine.
The International Atomic Energy Agency (IAEA) has developed a “dose conversion coefficient” which can be used in combination with quotient of the effective dose and dose-area product to estimate risk of cancer. The IAEA reports the dose conversion coefficient for fluoroscopically-guided diagnostic lumbar spine procedures is 0.21. According to the data published in this study and the IAEA, an average BVNRFA could theoretically increase risk of cancer by 1.9 % [25]. It is important to consider the age of the patient as an important factor in understanding this radiation risk as the risk of developing cancer following radiation exposure decreases as age, at time of exposure, increases.
Our study did not find a significant impact of BMI on fluoroscopic time. Although there are no BVNRFA studies to directly compare to, this finding is similar to findings for patients undergoing sacroiliac joint injections [26], lumbar medial branch radiofrequency ablations [19] and cervical interlaminar injections [27]. This differs from published work regarding lumbar transforaminal epidural steroid injection [18]. Our study found that increased radiation dose was associated with increased BMI. Our study also found no significant difference between cases performed with a trainee for radiation exposure time or cumulative dose. This is consistent with McCormick et al.’s study in the cervical spine [27] but differs from Wagner et al.’s study of lumbar radiofrequency ablations [19].
Unsurprisingly, patients who underwent BVNRFA of more levels had statistically significant greater fluoroscopic exposure times. S1 involvement was also associated with significantly greater radiation exposure. Although a similar transpedicular access technique is used access the S1 pedicle, patient anatomy dictates how the procedure is performed. There is variability in iliac crest height in respect to S1 pedicle. The initial obliqued trajectory can be limited in “deeper” sacrums where an obliqued view can be obscured due to high iliac crests. For this reason, additional radiation is used for initial image optimization and pedicle docking. This is not an issue in the lumbar spine where the osseous structures are not in the way with an oblique transpedicular view. Unsurprisingly, there was significantly longer fluoroscopy times during the first few months of performing BVNRFA compared to later in the year. Though not controlled for BMI or levels performed, the average fluoroscopy times of the later months were less than half the average times of the earlier month a BVNRFA was performed. This finding should be taken into consideration when interpreting the relatively large radiation exposure we reported. It is likely that over a course of a career, the average radiation exposure from BVNRFA will decrease for individual physicians. Though attendings who consistently teach new fellows each year and allow them to independently perform BVNRFA should particularly closely monitor their radiation exposure.
To the knowledge of the authors, this is the first study to publish reference ranges for radiation exposure for BVNRFA. The results of this study indicate that radiation exposure during BVNRFA is relatively high when compared to diagnostic imaging such as CTs but is comparable to orthopaedic surgeries. Given the relative novelty of BVNRFA, it would be beneficial for other institutions to report their average fluoroscopy exposure towards establishing reference doses for practitioner and trainees to strive towards and to aid the development of new techniques to optimize exposure.
5. Limitations
Our study has several limitations. The purpose of this study was to report the average fluoroscopic time and radiation dose. As such, this study did not report patient reported outcomes. Given that the purpose of our study was to report the average fluoroscopy time and radiation exposure during BVNRFA, we did not evaluate for any interventions used to decrease the exposure. Although this was a retrospective study, a prospective study would have been unlikely to have altered our results since we were not investigating specific variables. This study reported the fluoroscopic times of a single provider conducted within a single medical center. Our results may not be generalizable to other institutions that may use different procedure equipment and setting, physicians with variable years of experience, and patients with differing demographic makeup.
6. Conclusion
Our study reports the average fluoroscopic time for basivertebral nerve radio frequency ablations to be about two and a half minutes.
Declaration of competing interest
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
References
- 1.De Vivo A.E., D'Agostino G., D'Anna G., et al. Intra-osseous basivertebral nerve radiofrequency ablation (BVA) for the treatment of vertebrogenic chronic low back pain. Neuroradiology. May 2021;63(5):809–815. doi: 10.1007/s00234-020-02577-8. [DOI] [PubMed] [Google Scholar]
- 2.Conger A., Burnham T.R., Clark T., Teramoto M., McCormick Z.L. The effectiveness of intraosseous basivertebral nerve radiofrequency ablation for the treatment of vertebrogenic low back pain: an updated systematic review with single-arm meta-analysis. Pain Med. Jul 20 2022;23(Suppl 2):S50–S62. doi: 10.1093/pm/pnac070. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Tapio S., Little M.P., Kaiser J.C., et al. Ionizing radiation-induced circulatory and metabolic diseases. Environ Int. Jan 2021;146 doi: 10.1016/j.envint.2020.106235. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Paulo G., Bartal G., Vano E. Radiation dose of patients in fluoroscopically guided interventions: an update. Cardiovasc Intervent Radiol. Jun 2021;44(6):842–848. doi: 10.1007/s00270-020-02667-3. [DOI] [PubMed] [Google Scholar]
- 5.Vera G.V., Aleksandra F., Dragan K., Andrija H. Assessment of genome damage in occupational exposure to ionising radiation and ultrasound. Mutat Res. Dec 12 1997;395(2–3):101–105. doi: 10.1016/s1383-5718(97)00149-6. [DOI] [PubMed] [Google Scholar]
- 6.Maffei F., Angelini S., Forti G.C., et al. Spectrum of chromosomal aberrations in peripheral lymphocytes of hospital workers occupationally exposed to low doses of ionizing radiation. Mutat Res. Mar 22 2004;547(1–2):91–99. doi: 10.1016/j.mrfmmm.2003.12.003. [DOI] [PubMed] [Google Scholar]
- 7.Chang Y.J., Kim A.N., Oh I.S., Woo N.S., Kim H.K., Kim J.H. The radiation exposure of radiographer related to the location in C-arm fluoroscopy-guided pain interventions. Korean J Pain. Apr 2014;27(2):162–167. doi: 10.3344/kjp.2014.27.2.162. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Manchikanti L., Cash K.A., Moss T.L., Pampati V. Radiation exposure to the physician in interventional pain management. Pain Physician. Oct 2002;5(4):385–393. [PubMed] [Google Scholar]
- 9.Milacic S. Risk of occupational radiation-induced cataract in medical workers. Med Lav. May-Jun 2009;100(3):178–186. [PubMed] [Google Scholar]
- 10.Mastrangelo G., Fedeli U., Fadda E., Giovanazzi A., Scoizzato L., Saia B. Increased cancer risk among surgeons in an orthopaedic hospital. Occup Med. Sep 2005;55(6):498–500. doi: 10.1093/occmed/kqi048. [DOI] [PubMed] [Google Scholar]
- 11.Roguin A., Goldstein J., Bar O., Goldstein J.A. Brain and neck tumors among physicians performing interventional procedures. Am J Cardiol. May 1 2013;111(9):1368–1372. doi: 10.1016/j.amjcard.2012.12.060. [DOI] [PubMed] [Google Scholar]
- 12.Chartier H., Fassier P., Leuraud K., et al. Occupational low-dose irradiation and cancer risk among medical radiation workers. Occup Med. Oct 27 2020;70(7):476–484. doi: 10.1093/occmed/kqaa130. [DOI] [PubMed] [Google Scholar]
- 13.Schueler B.A. Operator shielding: how and why. Tech Vasc Intervent Radiol. Sep 2010;13(3):167–171. doi: 10.1053/j.tvir.2010.03.005. [DOI] [PubMed] [Google Scholar]
- 14.Dagal A. Radiation safety for anesthesiologists. Curr Opin Anaesthesiol. Aug 2011;24(4):445–450. doi: 10.1097/ACO.0b013e328347f984. [DOI] [PubMed] [Google Scholar]
- 15.Cheon B.K., Kim C.L., Kim K.R., et al. Radiation safety: a focus on lead aprons and thyroid shields in interventional pain management. Korean J Pain. Oct 2018;31(4):244–252. doi: 10.3344/kjp.2018.31.4.244. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Valentin J. Avoidance of radiation injuries from medical interventional procedures. Ann ICRP. 2000;30(2):7–67. doi: 10.1016/S0146-6453(01)00004-5. [DOI] [PubMed] [Google Scholar]
- 17.Kim J.H. Three principles for radiation safety: time, distance, and shielding. Korean J Pain. Jul 2018;31(3):145–146. doi: 10.3344/kjp.2018.31.3.145. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.McCormick Z.L., Choxi S.C., Lee D.T., et al. The impact of body mass index on fluoroscopy time during lumbar epidural steroid injection; A multicenter cohort study. Pain Med. Jan 1 2017;18(1):25–35. doi: 10.1093/pm/pnw050. [DOI] [PubMed] [Google Scholar]
- 19.Graham E., Wagner D.M.C., Teramoto Masaru, PhD Aaron Conger, Taylor Burnham, McCormick Zachary L. Correlation between body mass index and fluoroscopy time in lumbar medial branch nerve radiofrequency ablation. J Quant Res Rehabil Med. 2018;1(2):37–41. [Google Scholar]
- 20.McCormick Z.L., Sperry B.P., Boody B.S., et al. Pain location and exacerbating activities associated with treatment success following basivertebral nerve ablation: an aggregated cohort study of multicenter prospective clinical trial data. Pain Med. Jul 20 2022;23(Suppl 2):S14–S33. doi: 10.1093/pm/pnac069. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Becker S., Hadjipavlou A., Heggeness M.H. Ablation of the basivertebral nerve for treatment of back pain: a clinical study. Spine J. Feb 2017;17(2):218–223. doi: 10.1016/j.spinee.2016.08.032. [DOI] [PubMed] [Google Scholar]
- 22.Cohen S.L., Schneider R., Carrino J.A., Zeldin R., Pavlov H. Radiation dose practice audit of 6,234 fluoroscopically-guided spinal injections. Pain Physician. Mar 2019;22(2):E119–E125. [PubMed] [Google Scholar]
- 23.Commision USNR Doses in our daily lives. 2024. https://www.nrc.gov/about-nrc/radiation/around-us/doses-daily-lives.html Updated April 26, 2022.
- 24.Crawley M.T., Rogers A.T. Dose-area product measurements in a range of common orthopaedic procedures and their possible use in establishing local diagnostic reference levels. Br J Radiol. Jul 2000;73(871):740–744. doi: 10.1259/bjr.73.871.11089466. [DOI] [PubMed] [Google Scholar]
- 25.Agency I.A.E. Radiation doses in interventional procedures. https://www.iaea.org/resources/rpop/health-professionals/interventional-procedures/radiation-doses-in-interventional-fluoroscopy
- 26.McCormick Z.L., Cushman D., Lee D.T., et al. Is there a relationship between body mass index and fluoroscopy time during sacroiliac joint injection? A multicenter cohort study. Pain Med. Jul 2016;17(7):1241–1248. doi: 10.1093/pm/pnv051. [DOI] [PubMed] [Google Scholar]
- 27.McCormick Z.L., Mattie R., Ebrahimi A., et al. Is there a relationship between body mass index and fluoroscopy time during cervical interlaminar epidural steroid injections? Pain Med. Jul 1 2017;18(7):1326–1333. doi: 10.1093/pm/pnw264. [DOI] [PubMed] [Google Scholar]