Abstract
Purpose
This work proposes a “hybrid” RF pulse design method for parallel transmit (pTx) systems to simultaneously control flip angle and root‐mean‐squared (). These pulses are generally only designed for flip angle, however, this can lead to uncontrolled , which then leads to variable magnetization transfer (MT) effects. We demonstrate the hybrid design approach for quantitative imaging where both flip angle and are important.
Theory and Methods
A dual cost function optimization is performed containing the normalized mean squared errors of the flip angle and distributions weighted by a parameter . Simulations were conducted to study the behavior of both properties when simultaneously optimizing them. In vivo experiments on a 7T MRI system with an 8‐channel pTx head coil were carried out to study the effect of the hybrid design approach on variable flip angle (= 1/T1) mapping.
Results
Simulations showed that both flip angle and can be homogenized simultaneously without detriment to either when compared to an individual optimization. By homogenizing flip angle and , maps were more uniform (coefficient of variation 6.6% vs. 13.0%) compared to those acquired with pulses that only homogenized flip angle.
Conclusion
The proposed hybrid design homogenizes on‐resonance MT effects while homogenizing the flip angle distribution, with only a small detriment in the latter compared to a pulse that just homogenizes flip angle. This improved mapping by controlling incidental MT effects, yielding more uniform maps.
Keywords: inhomogeneity, magnetization transfer, parallel transmit, RF pulse design, ultra high field
1. INTRODUCTION
Ultra‐high field (UHF) imaging (≥7T) offers increased SNR, which can be translated to higher resolutions. 1 , 2 However, the higher associated Larmor frequency leads to spatially inhomogeneous RF magnetic () fields, 3 , 4 which can severely impact image quality. Several solutions have been proposed to address this 5 , 6 , 7 , 8 including parallel transmission 7 , 8 , 9 , 10 (pTx) whereby the RF system is split into multiple independently controlled channels. Optimizing the relative amplitude/phase of every channel to homogenize the created field, known as RF shimming, 11 , 12 , 13 can require large number of channels for full 3D coverage, 11 therefore, advanced RF pulse designs are usually combined with pTx. Conventional pulse designs 14 , 15 , 16 typically control the nutation of the magnetization for mobile protons (free‐water and fat) that exhibit prolonged transverse magnetization decay times (), that is, the flip angle produced by an RF pulse, which is based on dynamics described by the Bloch equation. 17
Biological tissues also contain a pool of “semisolid” magnetization associated with protons in macromolecules 18 and bound water. These protons typically have very short T2 (˜10 μs) making them unobservable to most MRI; their dynamics are also not well‐described by the Bloch equation because they exhibit non‐exponential decays, as captured by the recently proposed generalized Bloch model. 19 For RF pulses whose durations are significantly longer than the semisolid T2 (which is the case for most pulses on human imaging systems), a binary spin‐bath (BSB) formulation 20 , 21 provides a good model. According to the BSB model the semisolid magnetization has no transverse magnetization and its longitudinal magnetization is saturated according to the root‐mean‐squared () created by the RF pulses. Although in practice the semisolid magnetization is not directly observed, it exchanges magnetization with the free‐water pool, an effect known as magnetization transfer 22 , 23 (MT). Hence, RF pulses that produce spatially non‐uniform can produce spatially non‐uniform contrast, mediated by MT. The recently proposed parallel transmit pulse design for saturation homogeneity (“PUSH”) method 24 addressed this effect by controlling the distribution produced by the RF pulse. PUSH was applied to design saturation pulses for generation of MT contrast; in this scenario, the direct effect 23 on the free‐water magnetization, that is, the flip angle, was ignored because the pulses were applied off‐resonance.
In general, most MRI experiments apply pulses on‐resonance, whether for excitation, refocusing, or inversion. Although in most cases one just considers the flip angle of such pulses, they also saturate the semisolid magnetization according to . This effect is typically smaller than that of the flip angle, nevertheless studies have shown that such “incidental” on‐resonance MT effects are a confound for different sequences including quantitative MT, 25 , 26 , 27 , 28 single pool relaxometry () mapping using variable flip angle 28 , 29 , 30 , 31 (VFA) and inversion recovery 32 (IR), or even multiple spin echo 33 (MSE) sequences. In the case of the VFA method, it has been shown 30 that the apparent relaxation rate () varies with the applied :
(1) |
where and are the longitudinal exchange rates of the free‐water () and semisolid () pools, (and ) is the exchange rate from pool to (and from pool to ), is the semisolid absorption line‐shape at RF frequency offset and is the gyromagnetic ratio. The dependence of on leads to spatially varying if is itself spatially inhomogeneous.
In this work, we propose a generalized “hybrid” RF pulse design method that optimizes both flip angle and distributions simultaneously and demonstrate its utility for designing pulses for VFA mapping at 7T.
2. THEORY
2.1. Hybrid design
Conventionally, pTx pulse designs target a desired rotation, or flip angle , of the free‐water magnetization:
(2) |
where and are vectors representing the RF and gradient waveforms respectively, is a vector with the desired flip angle, and is a diagonal matrix with spatial error weightings. We calculate using the small tip angle (STA) approximation similarly to the spatial domain method from Grissom et al. 34 Discretizing space in samples and time in samples, the transverse magnetization vector is given by:
(3) |
where is a vector with the transmit sensitivities of the channel (transformed into a diagonal matrix), is a vector with the time samples of the RF waveform in the channel, and is the system matrix whose elements are given by:
(4) |
where is the imaginary unit, is the equilibrium magnetization, is the time discretization step, is the off‐resonance at coordinates , is the pulse duration and is a matrix with the excitation k‐space trajectory. 35 Note that under the STA, therefore, Eq. (2) targets the magnitude of the magnetization and ignores the excitation phase.
The recently proposed PUSH design instead targets a desired saturation of semisolid magnetization by controlling (referred to as in the optimization):
(5) |
where is the desired , and is given by:
(6) |
where is the repetition time. This work considers simple repeating gradient‐recalled sequences where averaging over TR gives the sequence , although this could be extended to different sequences and pulse types as necessary.
Here, we explore a “hybrid pulse design” method that simultaneously controls and using a dual cost function optimization that also includes operational constraints on hardware and specific absorption rate (SAR) for a model subject:
(7) |
where is the global SAR limit, is the 10 g local SAR limit, is the maximum average power, is the maximum RF voltage per channel, is the maximum gradient amplitude and is the maximum slew rate. Local SAR is calculated using a virtual observation point (VOP) compressed model 36 with total observation points. The proposed design consists of one term that is the normalized mean squared error of the flip angle weighted by and another that is the normalized mean squared error of the weighted by .
Compared to the more conventional pulse design (Eq. [2]) that usually just targets a desired flip angle , the hybrid (HY) design (Eq. [7]) introduces two new parameters: and . The balancing parameter allows adjustment to a flip angle only (FA) optimization () or to a only (i.e., PUSH) optimization (). We explore HY optimizations where . In these cases, trades‐off between the flip angle and errors in what is called the Pareto front, and its choice depends on the acceptable error for each one.
2.2. Lower bound
The properties and from an RF pulse are related, so when choosing targets and in Eq. (7) for it is important to ensure that both targets are consistent. Consider an RF pulse applied on‐resonance without any gradients or off‐resonance and with duration and waveform ; its flip angle is:
(8) |
where is the peak and is the RF waveform normalized to unit duration and amplitude. At the same time the produced by this pulse is:
(9) |
Combining Eqs. (8) and (9) establishes a relationship between the two:
(10) |
Equivalence is only met if is real (i.e., pulse is on‐resonance), and there are no off‐resonance and/or applied gradient fields. In the more general case, magnetization can nutate back and forth between the longitudinal axis and transverse plane because of the aforementioned effects such that the final achieved flip angle is less than that given by Eq. (10), resulting in the following inequality:
(11) |
Therefore, a minimum is needed to achieve any target :
(12) |
For any design problem it will be necessary to choose to ensure that a feasible solution exists.
3. METHODS
3.1. Simulations
To explore the proposed design, excitation pulses were designed offline using the kT‐points 14 method (5 RF rectangular sub‐pulses of 200 μs duration and 100 μs gradient blips) for a variety of scenarios: all combinations from = 15°, in steps of ( = 0.415 μT), and in 51 logarithmically spaced steps (denser in the bounds and sparser in the middle) were optimized. These optimizations used 3D brain transmit maps from an 8‐channel pTx system (details below) and were solved using a multi‐start strategy with random seeds on a desktop computer (Intel i9‐10 900X @ 3.70 GHz, 64 GB of RAM, 16 cores) with MATLAB R2020b (The MathWorks). The solutions were analyzed in terms of their normalized root mean square error (NRMSE) for and .
For comparison, the circular polarized (CP) mode and RF shimming solutions for = 15° were also calculated and analyzed.
3.2. Experiments
All experiments were performed using a 7T scanner (MAGNETOM Terra, Siemens Healthcare) in prototype research configuration, with an 8Tx/32Rx head coil (Nova Medical).
3.2.1. Dual Flip Angle mapping
The proposed design was used to optimize excitation pulses for mapping in a dual flip angle (DFA) experiment. 37 This application was considered because of its sensitivity to incidental MT effects (Eq. [1]), such that the distribution is expected to affect the spatial distribution of .
The DFA method is a special case of VFA where only two spoiled gradient echo images (SPGR) are acquired, here, with flip angles = {3°, 15°} (optimal for 1300 ms and TR = 8 ms 38 ). When holding constant, as done here, the Ernst equation can be linearized and fitting can be performed via a voxelwise linear regression:
(13a) |
(13b) |
where is the slope of the linear regression, are the measured magnitude SPGR signals and are the flip angles. Voxelwise fitting was performed using calculated from a Bloch simulation of the RF pulses and applying the incomplete spoiling correction proposed by Baudrexel et al. 39
Six healthy volunteers (25–36 years old, 2 males) were scanned in accordance with local ethical approval. The SPGR volumes (sagittal orientation, = 240 × 240 × 192 mm3, resolution , TR = 8 ms, TE = 3 ms, bandwidth ), GRAPPA 40 acceleration factor of with (PE × SL directions, respectively) reference lines were acquired with 20 s of dummy pulses to reach steady state and stabilize the RF power amplifier output, 24 , 41 resulting in = 1 min 59 s per volume. All images were registered using FSL FLIRT 42 before fitting and an additional MP2RAGE 43 acquired at the same resolution was used for segmentation with SPM12. 44 mapping was repeated using different pulse types as described below, consisting of acquiring the two flip angle SPGRs for each pulse type (Figure 1).
3.2.2. Excitation pulse and data analysis
Before the pulse design, and mapping was performed using a combination of a quantitative actual flip angle imaging 45 , 46 (AFI) map ( = 2 min18 s; sagittal orientation, , resolution , = 25/125 ms, α = 60°, TE = {0.75, 1.75, 10}ms, bandwidth , GRAPPA 40 acceleration factor of with (phase encoding direction) reference lines, slice partial Fourier 6/8, elliptical scanning shutter, 10 s of dummy pulses) and relative per channel estimates via low flip angle SPGR images 47 ( = 45 s each; sagittal orientation, , resolution , TR = 10 ms, TE = 1 ms, bandwidth , 10 s of dummy pulses), as per Figure 1. These maps were then used in the pulse design as described in Section 3.3, before acquiring data for the DFA experiment with the appropriate pulses; pulse calculation was fully scanner‐integrated within a MATLAB framework (R2012b) from the scanner vendor (release Syngo.MR VE12U‐SP01). The optimization was solved using a multi‐start strategy with random starts on the system's console (Intel Xeon E5‐1620 v3 @ 3.5G Hz, 32 GB of RAM, 4 cores), taking s.
The DFA mapping experiment was conducted using three types of excitation pulses: (1) CP pulse; (2) flip angle optimized pulse with , referred to as FA pulse; (3) HY optimized pulse with and , referred to as HY pulse.
Their implementation is described in Section 3.3. For all six subjects an map was acquired using the above three types of excitation pulses. Additionally, for four of these subjects a test–retest of mapping was performed with the FA and HY pulses. For the retest, the pulses were redesigned by performing a new optimization from the same and maps, but using a different set of random starting points.
The maps and their distributions in white matter (WM), gray matter (GM) and CSF were compared across all subjects and pulse types. For a voxel‐based analysis, all maps were warped into a common Montreal Neurological Institute (MNI) space using DARTEL 48 as implemented in the hMRI toolbox, 49 and the coefficient of variation (CoV) across subjects was calculated for each pulse type. Furthermore, the test–retest map acquisitions were compared to assess reproducibility.
3.3. Pulse design implementation
The design in Eq. (7) was performed for the excitation pulse types described in Subsection 3.2.2. The CP pulse consisted of a single rectangular RF pulse 200 μs long, whose amplitude was optimized to minimize the flip angle mean squared error with respect to the target. The FA and HY pulses consisted of 5 kT‐points 14 as in the simulations (200 μs RF rectangular sub‐pulses and 100 μs gradient blips), resulting in = {0.08 μT, 0.42 μT} for the HY pulses with = {3°, 15°}, respectively. The optimization was solved using a multi‐start strategy in MATLAB with the interior‐point algorithm from the fmincon routine, providing first and second order analytical derivatives, and the following constraints: (using vendor provided SAR model with ) in first level SAR mode 50 ; , , . To speed‐up the optimization, the number of voxels was compressed using k‐means clustering similarly to Tomi‐Tricot et al. 51 Voxels inside a brain mask (obtained using FSL BET 52 ) were classified according to their transmit sensitivities, off‐resonance and position, in a total of 500 clusters. The centroid of each cluster was then used as representative of the cluster properties, and the cluster size was used in the weighting matrix (Eq. (7)).
4. RESULTS
4.1. Simulations
Figure 2 shows the flip angle and maps achieved with the different excitation pulses. As expected, the CP pulse has a central brightening inhomogeneity in both properties, whereas FA pulse has a very uniform flip angle map, but inhomogeneous distribution. When using HY pulse both flip angle and become uniform. Figure S1 shows that the maps obtained with a magnitude least squares RF shim are very similar to those of the CP pulse.
Figure 3A shows the Pareto fronts representing the trade‐off between the two terms of the proposed cost function (Eq. [7]). For the errors in and increase substantially when (see Figure 3B), but for the Pareto front resembles the L‐curve from regularization problems with most solutions located in the corner. This clustering of solutions occurs as the total NRMSE becomes practically constant for as can be seen in Figure 3B. Moreover, as increases up to , the total NRMSE decreases driven by lower NRMSE (α) and then plateaus until approaches 1 μT where it reaches SAR limits and increases the NRMSE of , as can be seen in Figure 3C. Figure S2 expands the visualization of the cost function by plotting the NRMSE of and for all and
Figure 4 shows the distributions of and for the excitation pulses studied. The distribution for CP pulse lies on the line defined by Eq. (10) showing large dispersion in both and , which are perfectly correlated because one is proportional to the other in this case. The FA pulse () produces a narrow distribution of , but wide distribution of because the latter is not controlled by the calculation. Moreover, the optimized pulse () has a narrow distribution of , but wide distribution of . The HY pulse () achieves both narrow and distributions.
4.2. Experiments
Figure 5 shows the maps for one subject estimated using different pulses; the corresponding and maps are given in Figure S3 and the maps for all subjects are in Figure S4. With CP mode the values are fairly uniform, but noisy in the center (green arrow; see zoom‐ins), corresponding to the area of higher and intensity. The noisy estimation is no longer present with the FA pulses, but severe shading is observed (blue arrows). Finally, using HY pulses estimation stays precise and no shading is observed. The histograms of the distribution in Figure 6 show tall and narrower distributions in WM and GM with CP and HY pulses, as well as similar mean in GM, but HY pulses having a higher mean in WM. Moreover, FA pulses cause a broader distribution in WM and GM, with mean also different from the other pulses. In CSF, all pulse types show similar mean and dispersion of .
The WM distribution of for all subjects is in Figure 7. Both CP and HY pulses show consistent distributions across all subjects, with mean being smaller for CP pulses, whereas FA pulses exhibit a big variability in the location and dispersion of the distributions. The CoV of across subjects is shown in Figure 8 after diffeomorphic registration of all maps to the MNI space. The CP and HY pulses exhibit a smaller CoV in WM, whereas FA pulses have the largest CoV that changes over the WM (blue arrow) because of different shadings across subjects.
The maps from subject C for the scan‐rescan test are in Figure 9. With FA pulses the maps obtained in the two scans show considerable differences as indicated by the green and blue arrows, resulting in a different mean in WM (0.713s−1 vs. 0.834s−1). Moreover, HY pulses yielded similar maps in the two scans, with both looking equally uniform and practically the same mean in WM (0.783s−1 vs. 0.769s−1). Figure 10 shows the distributions in WM for all subjects across the two scans. Overall, FA designs yield a different distribution when scanning the same subject with new pulses and that is non‐unimodal in some cases. When using different HY pulses the distributions are unimodal and very close to each other. In three of four cases, we observed a shift toward larger values observed in the second HY pulse scan, which was performed later during each scanning session.
5. DISCUSSION
This work presents a novel pTx pulse design that simultaneously controls the spatial distributions of flip angle and produced by an RF pulse. These dictate the rotation of the free‐water magnetization and the saturation of the semisolid magnetization, respectively, being relevant to applications affected directly or indirectly by MT effects. Here, we showed how the proposed design improves consistency of mapping at 7T by controlling incidental MT effects.
Although RF shimming is a straightforward solution to homogenize both flip angle () and () distributions by targeting (or ), 12 , 13 performance for full 3D coverage using our 8‐channel Tx head coil is often similar to that of CP mode 11 (e.g., Figure S1). Current pTx pulse designs overcome this limitation by instead targeting the flip angle or distribution produced by an RF pulse, but not both—this results in a wide distribution of the disregarded property as shown in Figure 4. Although for off‐resonance saturation pulses it is sufficient to consider the distribution of , but not flip angle variation (as in the PUSH method 24 ), for on‐resonance pulses both properties are important in determining the magnetization dynamics of systems with MT. The proposed HY design is a generalization of conventional pulse designs that only consider and the PUSH design 24 that only considers by performing a dual cost function optimization.
5.1. HY optimization performance
The HY optimization introduces two new parameters compared to a FA optimization: (1) , the target ; and (2) , the balancing parameter. Theory shows that a lower bound on can be derived based on the desired on‐resonance flip angle . Simulation results in Figure 3 show that for HY optimizations () the NRMSE of and reduce considerably when increasing until , after which it plateaus until reaches the SAR limits and then the NRMSE of increases sharply. In this case, the NRMSE of does not increase because its target is achievable within , that is, . The threshold is likely to depend on several factors (e.g., coil properties) the volume being imaged and specific properties of the pulse design.
Simulations showed that the pulse design appears to be insensitive to balancing parameter , with the NRMSE of and only changing considerably when and , respectively. This implies that the optimization finds solutions with simultaneously uniform and without having to perform substantial trade‐offs; therefore, an even balancing was used for the experiments. The insensitivity to meant that most of the points in the Pareto front (Figure 3A) were concentrated in a corner, resembling the L‐curve behavior seen in other regularized optimizations. 34 The local SAR was also found be largely insensitive to when this is , depending mostly on (results not shown).
The performance of the proposed design in 3D is mostly limited by the optimization. This was observed in the PUSH study 24 that also reported worse homogeneity in 3D imaging because of the limited variability in the head‐foot direction provided by the coil used. Although for flip angle optimization this can be mitigated by using gradient encoding, only depends on the transmit sensitivities and RF waveforms. Nevertheless, this key difference provides greater control in achieving both and targets as gradients are a degree of freedom that differentiate the two.
5.2. DFA mapping
The HY pulse design was used for mapping with the DFA method that is known to be sensitive to incidental MT effects. 28 , 29 , 30 , 31 Although flip angle inhomogeneities can be corrected in the fit (Eq. [13]) and cause “noisy”/imprecise estimation because of ill‐conditioned fitting, incidental MT effects created by cannot be corrected in single pool VFA experiments and cause bias. 28 , 29 , 30 , 31 The impact of flip angle inhomogeneity was observed in Figure 5 where was noisy in the central regions with CP pulses, which disappeared when using FA and HY pulses. However, the inhomogeneous produced by the FA pulses introduced severe spatial inhomogeneity of . By homogenizing with the HY pulses the shading was suppressed while preserving a precise estimation because of a homogeneous . These findings were verified in the CoV of the normalized maps across all subjects (Figure 8), with HY pulses having a small SD similar to that of CP pulses, and last, FA pulses exhibiting the largest variability because of different shadings among subjects. The WM distribution of in Figure 7 also showed that FA pulses yielded the least consistent distributions across all subjects, whereas CP and HY pulses gave very similar distributions. This can be explained by the fact that both CP and HY pulses deliver a similar for each subject, whereas with FA pulses this can change as is not controlled by the optimization.
The mapping reproducibility improved substantially when using HY pulses (Figure 9). Both sets of HY pulses yielded similar maps, whereas with FA pulses test–retest maps showed large differences in the estimated values. For all subjects the WM distribution of changed considerably between the two scans with FA pulses, but retained a consistent unimodal distribution with HY pulses. However, with the latter there was generally a shift in toward longer values in the second scan. A separate phantom experiment revealed some small deviations in the amplitude of the RF power amplifiers (RFPA) (Figure S5) over time that could explain this shift because (by definition) the second scan was always performed later in the scanning session. We and others have reported some instability relating to RF power drifting specifically on this RFPA model and generation 24 , 41 — 20 seconds of dummycycles were used to minimize this effect as far as possible, but it seems this was not a complete solution.
Overall, the results indicated that homogenizing with traditional FA pulse design led to worse maps compared to CP pulses by introducing a strong shading across the maps. The HY design solves this problem by homogenizing , which homogenizes the MT bias and, therefore, results in uniform maps. Interestingly, despite produced by CP mode being inhomogeneous, maps using this mode showed less bias than FA design. One possible reason was that the CP pulses applied here were shorter and did not involve gradients blips compared to the FA and HY kT‐points pulses that could make them less sensitive to errors in the and maps 53 as well as eddy currents. 54 VFA experiments are sensitive to other sources of bias, such as incomplete spoiling. 55 We sought to minimize this by using the correction factor for proposed by Baudrexel et al., 39 and reduced scanning time to mitigate motion. Nevertheless, any intra‐scan and inter‐scan motion can affect mapping by changing the receive sensitivity, 56 but also the and fields, 57 which in turn changes the and maps produced by the RF pulses, and these effects would be less predictable for the pTx methods used than CP mode.
This said, we also believe that the relatively good performance of CP mode compared with the pTx methods in our results is likely down to the choice of flip angles and how the signal biases appearing in each of the two separate component images interacted to cause a bias in the measured . To test this we conducted an experiment using other nominal flip angles, which yielded less self‐consistent maps when using CP mode (Figure S6). Ultimately there will be circumstances in which CP mode can give reasonable results, and pTx is not necessary; this is a matter that must be addressed on a study specific basis. However, our results show that if a pTx design is used, then the presented HY method provides an effective means for mitigating incidental MT effects, and these effects can result in quite severe artefacts if using more standard flip angle based design methods.
5.3. Assumptions, extensions, and other applications
The impact of homogenizing is smaller than homogenizing because of (1) the semisolid magnetization being a fraction of the free‐water magnetization; and (2) the semisolid magnetization having virtually no transverse magnetization to create signal, so it is only via exchange with the free‐water magnetization that it produces a visible effect. Nevertheless, this can be important for quantitative methods like MPM, 58 quantitative MT, 25 and IR based mapping methods 32 , 59 where shading in mapping have also been observed with MP2RAGE. 60
Here, we explored DFA mapping with each pulse designed to use the smallest reasonable amount of (), but the design could also be used to create the same for all pulses as in the controlled saturation MT approach. 30 Moreover, the design approach can be used generally, both for alternative k‐space trajectories to kT‐points (e.g., spiral nonselective, 15 spokes, 61 , 62 etc.) or high tip angle designs such as for inversion or refocusing pulses, as well as for universal pulses. 63 For the case of high tip angle pulses, longer pulse durations may be necessary such that relaxation and exchange can no longer be neglected. In that scenario, the magnetization response of both pools becomes intertwined, violating the assumption in the PUSH design that the pools are independent during RF exposure. This scenario would require more complex designs that are beyond the scope of this study.
6. CONCLUSION
This work proposed a novel HY optimization of RF pulses to homogenize both their flip angle and distributions simultaneously. Simulations showed that it is possible to obtain pulses with both uniform flip angle and without big detriment of each property compared to an individual optimization. As an exemplar, we applied the approach to single pool mapping and showed that the HY optimized pulses produce clearly more uniform estimates than flip angle optimized pulses, because of this effect. In this application, the HY pulse design provides a way to spatially control MT effects, therefore, its bias, to achieve uniform estimation, as well as to control the flip angles that impact estimation precision. The HY design method is general and could be used in other scenarios such as other quantitative imaging methods, or for non‐quantitative sequences where incidental MT effects have a significant effect on contrast.
FUNDING INFORMATION
National Institute for Health Research (NIHR) Biomedical Research Centre based at Guy's and St Thomas' NHS Foundation Trust and King's College London and/or the NIHR Clinical Research Facility, funded by the King's College London & Imperial College London EPSRC Centre for Doctoral Training in Medical Imaging [EP/L015226/1], by core funding from the Wellcome/EPSRC Centre for Medical Engineering [WT203148/Z/16/Z] and by a Wellcome Trust Collaboration in Science Award [WT201526/Z/16/Z].
CONFLICT OF INTEREST STATEMENT
R.T.T., P.L., R.G., and D.R. work for Siemens Healthcare.
Supporting information
ACKNOWLEDGMENTS
This research was supported by the National Institute for Health Research (NIHR) Biomedical Research Centre based at Guy's and St Thomas' NHS Foundation Trust and King's College London and/or the NIHR Clinical Research Facility, funded by the King's College London & Imperial College London EPSRC Centre for Doctoral Training in Medical Imaging [EP/L015226/1], by core funding from the Wellcome/EPSRC Centre for Medical Engineering [WT203148/Z/16/Z] and by a Wellcome Trust Collaboration in science award [WT201526/Z/16/Z]. For the purpose of Open Access, the Author has applied a CC BY public copyright license to any Author Accepted Manuscript version arising from this submission. The views expressed are those of the author(s) and not necessarily those of the NHS, the NIHR or the Department of Health and Social Care.
Correction added after online publication 25th October 2024. The author has updated the Acknowledgments to include a funding source.
Leitão D, Tomi‐Tricot R, Bridgen P, et al. Parallel transmit hybrid pulse design for controlled on‐resonance magnetization transfer in R 1 mapping at 7T. Magn Reson Med. 2025;93(3):1090‐1103. doi: 10.1002/mrm.30333
DATA AVAILABILITY STATEMENT
According to United Kingdom research councils' Common Principles on Data Policy and Wellcome Trust's Policy on data, software and materials management and sharing, all simulated data supporting this study is openly available at https://github.com/mriphysics/HybridPTx (hash 98fc91d at time of submission). This will exclude proprietary code from SIEMENS, but that can be shared on request by agreement including the vendor, and in vivo MRI data because of the terms of the ethical approval under which they were acquired.
REFERENCES
- 1. Lüsebrink F, Mattern H, Yakupov R, et al. Comprehensive ultrahigh resolution whole brain in vivo MRI dataset as a human phantom. Sci Data. 2021;8:138. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Wang F, Dong Z, Tian Q, et al. In vivo human whole‐brain Connectom diffusion MRI dataset at 760 μm isotropic resolution. Sci Data. 2021;8:122. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Van De Moortele PF, Akgun C, Adriany G, et al. B1 destructive interferences and spatial phase patterns at 7 T with a head transceiver array coil. Magn Reson Med. 2005;54:1503‐1518. [DOI] [PubMed] [Google Scholar]
- 4. Collins CM, Liu W, Schreiber W, Yang QX, Smith MB. Central brightening due to constructive interference with, without, and despite dielectric resonance. J Magn Reson Imaging. 2005;21:192‐196. [DOI] [PubMed] [Google Scholar]
- 5. Tannús A, Garwood M. Adiabatic pulses. NMR Biomed. 1997;10:423‐434. [DOI] [PubMed] [Google Scholar]
- 6. Yang QX, Mao W, Wang J, et al. Manipulation of image intensity distribution at 7.0T: Passive RF shimming and focusing with dielectric materials. J Magn Reson Imaging. 2006;24:197‐202. [DOI] [PubMed] [Google Scholar]
- 7. Katscher U, Börnert P, Leussler C, Van den Brink JS. Transmit sense. Magn Reson Med. 2003;49:144‐150. [DOI] [PubMed] [Google Scholar]
- 8. Zhu Y. Parallel excitation with an array of transmit coils. Magn Reson Med. 2004;51:775‐784. [DOI] [PubMed] [Google Scholar]
- 9. Hoult DI, Phil D. Sensitivity and power deposition in a high‐field imaging experiment. J Magn Reson Imaging. 2000;12:46‐67. [DOI] [PubMed] [Google Scholar]
- 10. Padormo F, Beqiri A, Hajnal JV, Malik SJ. Parallel transmission for ultrahigh‐field imaging. NMR Biomed. 2016;29:1145‐1161. doi: 10.1002/nbm.3313 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Mao W, Smith MB, Collins CM. Exploring the limits of RF shimming for high‐field MRI of the human head. Magn Reson Med. 2006;56:918‐922. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Setsompop K, Wald LL, Alagappan V, Gagoski BA, Adalsteinsson E. Magnitude least squares optimization for parallel radio frequency excitation design demonstrated at 7 tesla with eight channels. Magn Reson Med. 2008;59:908‐915. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Katscher U, Vernickel P, Graesslin I, Börnert P. RF shimming using a multi‐element transmit system in phantom and in vivo studies. Proc Intl Soc Mag Reson Med. 2007;15:1693. [Google Scholar]
- 14. Cloos MA, Boulant N, Luong M, et al. kT‐points: short three‐dimensional tailored RF pulses for flip‐angle homogenization over an extended volume. Magn Reson Med. 2012;67:72‐80. [DOI] [PubMed] [Google Scholar]
- 15. Malik SJ, Keihaninejad S, Hammers A, Hajnal JV. Tailored excitation in 3D with spiral nonselective (SPINS) RF pulses. Magn Reson Med. 2012;67:1303‐1315. [DOI] [PubMed] [Google Scholar]
- 16. Cao Z, Donahue MJ, Ma J, Grissom WA. Joint design of large‐tip‐angle parallel RF pulses and blipped gradient trajectories. Magn Reson Med. 2016;75:1198‐1208. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Bloch F. Nuclear induction. Phys Rev. 1946;70:460‐474. [Google Scholar]
- 18. Orzylowska A, Slowik T, Chudzik A, Pankowska A, Lam W, Stanisz G. The CLARITY procedure of lipid removal from brain tissue sample reveals the lipid‐origin of MT contrast in CEST imaging experiment. Proceedings of the 28th Annual Meeting of ISMRM. ISMRM; 2020:497. [Google Scholar]
- 19. Assländer J, Gultekin C, Flassbeck S, Glaser SJ, Sodickson DK. Generalized Bloch model: a theory for pulsed magnetization transfer. Magn Reson Med. 2022;87:2003‐2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Henkelman RM, Huang X, Xiang QS, Stanisz GJ, Swanson SD, Bronskill MJ. Quantitative interpretation of magnetization transfer. Magn Reson Med. 1993;29:759‐766. [DOI] [PubMed] [Google Scholar]
- 21. Graham SJ, Henkelman RM. Understanding pulsed magnetization transfer. J Magn Reson Imaging. 1997;7:903‐912. [DOI] [PubMed] [Google Scholar]
- 22. Balaban RS, Ceckler TL. Magnetization transfer contrast in magnetic resonance imaging. Magn Reson Q. 1992;8:116‐137. [PubMed] [Google Scholar]
- 23. Henkelman RM, Stanisz GJ, Graham SJ. Magnetization transfer in MRI: a review. NMR Biomed. 2001;14:57‐64. [DOI] [PubMed] [Google Scholar]
- 24. Leitão D, Tomi‐Tricot R, Bridgen P, et al. Parallel transmit pulse design for saturation homogeneity (PUSH) for magnetization transfer imaging at 7T. Magn Reson Med. 2022;88:180‐194. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Dortch RD, Moore J, Li K, et al. Quantitative magnetization transfer imaging of human brain at 7T. Neuroimage. 2013;64:640‐649. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Soustelle L, Troalen T, Hertanu A, et al. Quantitative magnetization transfer MRI unbiased by on‐resonance saturation and dipolar order contributions. Magn Reson Med. 2023;90:875‐893. doi: 10.1002/mrm.29678 [DOI] [PubMed] [Google Scholar]
- 27. Bagnato F, Hametner S, Franco G, et al. Selective inversion recovery quantitative magnetization transfer brain MRI at 7T: clinical and postmortem validation in multiple sclerosis. J Neuroimaging. 2018;28:380‐388. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Olsson H, Andersen M, Lätt J, Wirestam R, Helms G. Reducing bias in dual flip angle T1‐mapping in human brain at 7T. Magn Reson Med. 2020;84:1347‐1358. [DOI] [PubMed] [Google Scholar]
- 29. Teixeira RP, Neji R, Wood TC, Baburamani AA, Malik SJ, Hajnal JV. Controlled saturation magnetization transfer for reproducible multivendor variable flip angle T1 and T2 mapping. Magn Reson Med. 2020;84:221‐236. doi: 10.1002/mrm.28109 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Teixeira RPAG, Malik SJ, Hajnal JV. Fast quantitative MRI using controlled saturation magnetization transfer. Magn Reson Med. 2019;81:907‐920. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Ou X, Gochberg DF. MT effects and T1 quantification in single‐slice spoiled gradient echo imaging. Magn Reson Med. 2008;59:835‐845. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Rioux JA, Levesque IR, Rutt BK. Biexponential longitudinal relaxation in white matter: characterization and impact on T1 mapping with IR‐FSE and MP2RAGE. Magn Reson Med. 2016;75:2265‐2277. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Weigel M, Helms G, Hennig J. Investigation and modeling of magnetization transfer effects in two‐dimensional multislice turbo spin echo sequences with low constant or variable flip angles at 3 T. Magn Reson Med. 2010;63:230‐234. [DOI] [PubMed] [Google Scholar]
- 34. Grissom W, Yip CY, Zhang Z, Stenger VA, Fessler JA, Noll DC. Spatial domain method for the design of RF pulses in multicoil parallel excitation. Magn Reson Med. 2006;56:620‐629. [DOI] [PubMed] [Google Scholar]
- 35. Pauly J, Nishimura D, Macovski A. A k‐space analysis of small‐tip‐angle excitation. J Magn Reson (1969). 1989;81:43‐56. doi: 10.1016/0022-2364(89)90265-5 [DOI] [PubMed] [Google Scholar]
- 36. Eichfelder G, Gebhardt M. Local specific absorption rate control for parallel transmission by virtual observation points. Magn Reson Med. 2011;66:1468‐1476. [DOI] [PubMed] [Google Scholar]
- 37. Fram EK, Herfkens RJ, Johnson GA, et al. Rapid calculation of T1 using variable flip angle gradient refocused imaging. Magn Reson Imaging. 1987;5:201‐208. [DOI] [PubMed] [Google Scholar]
- 38. Wood TC. Improved formulas for the two optimum VFA flip‐angles. Magn Reson Med. 2015;74:1‐3. [DOI] [PubMed] [Google Scholar]
- 39. Baudrexel S, Nöth U, Schüre JR, Deichmann R. T1 mapping with the variable flip angle technique: a simple correction for insufficient spoiling of transverse magnetization. Magn Reson Med. 2018;79:3082‐3092. [DOI] [PubMed] [Google Scholar]
- 40. Griswold MA, Jakob PM, Heidemann RM, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002;47:1202‐1210. [DOI] [PubMed] [Google Scholar]
- 41. Aghaeifar A, Bosch D, Heule R, et al. Intra‐scan RF power amplifier drift correction. Magn Reson Med. 2024;92:645‐659. doi: 10.1002/mrm.30078 [DOI] [PubMed] [Google Scholar]
- 42. Jenkinson M, Smith S. A global optimisation method for robust affine registration of brain images. Med Image Anal. 2001;5:143‐156. [DOI] [PubMed] [Google Scholar]
- 43. Marques JP, Kober T, Krueger G, van der Zwaag W, van de Moortele PF, Gruetter R. MP2RAGE, a self bias‐field corrected sequence for improved segmentation and T1‐mapping at high field. Neuroimage. 2010;49:1271‐1281. doi: 10.1016/j.neuroimage.2009.10.002 [DOI] [PubMed] [Google Scholar]
- 44. Ashburner J, Friston KJ. Unified segmentation. Neuroimage. 2005;26:839‐851. [DOI] [PubMed] [Google Scholar]
- 45. Yarnykh VL. Actual flip‐angle imaging in the pulsed steady state: a method for rapid three‐dimensional mapping of the transmitted radiofrequency field. Magn Reson Med. 2007;57:192‐200. [DOI] [PubMed] [Google Scholar]
- 46. Nehrke K. On the steady‐state properties of actual flip angle imaging (AFI). Magn Reson Med. 2009;61:84‐92. [DOI] [PubMed] [Google Scholar]
- 47. Van De Moortele P‐F, Snyder C, DelaBarre L, Adriany G, Vaughan JT, Ugurbil K. Calibration tools for RF shim at very high field with multiple element RF coils: from ultra fast local relative phase to absolute magnitude B1+ mapping. Proceedings of the International Society for Magnetic Resonance in Medicine . Vol 15. ISMRM; 2007:1676. [Google Scholar]
- 48. Ashburner J. A fast diffeomorphic image registration algorithm. Neuroimage. 2007;38:95‐113. [DOI] [PubMed] [Google Scholar]
- 49. Tabelow K, Balteau E, Ashburner J, et al. hMRI – a toolbox for quantitative MRI in neuroscience and clinical research. Neuroimage. 2019;194:191‐210. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50. IEC‐60601‐2‐33 . Medical Electrical Equipment‐Part 2‐33: Particular Requirements for the Basic Safety and Essential Performance of Magnetic Resonance Equipment for Medical Diagnosis, IEC 60601‐2‐33: 2010/AMD2: 2015. 2010.
- 51. Tomi‐Tricot R, Sedlacik J, Endres J, et al. Fully integrated scanner implementation of direct signal control for 2D T2‐weighted TSE at ultra‐high field. Proceedings of the International Society for Magnetic Resonance in Medicine . Vol 29. ISMRM; 2021:621. [Google Scholar]
- 52. Smith SM. Fast robust automated brain extraction. Hum Brain Mapp. 2002;17:143‐155. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53. Samsonov AA, Yarnykh VL. Accurate actual flip angle imaging (AFI) in the presence of fat. Magn Reson Med. 2024;91:2345‐2357. doi: 10.1002/mrm.30000 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54. Boulant N, le Ster C, Amadon A, et al. The possible influence of third‐order shim coils on gradient–magnet interactions: an inter‐field and inter‐site study. MAGMA. 2024;37:169‐183. doi: 10.1007/s10334-023-01138-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55. Corbin N, Callaghan MF. Imperfect spoiling in variable flip angle T1 mapping at 7T: quantifying and minimizing impact. Magn Reson Med. 2021;86:693‐708. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56. Balbastre Y, Aghaeifar A, Corbin N, Brudfors M, Ashburner J, Callaghan MF. Correcting inter‐scan motion artifacts in quantitative R1 mapping at 7T. Magn Reson Med. 2022;88:280‐291. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57. Plumley A, Watkins L, Treder M, Liebig P, Murphy K, Kopanoglu E. Rigid motion‐resolved B1+ prediction using deep learning for real‐time parallel‐transmission pulse design. Magn Reson Med. 2022;87:2254‐2270. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58. Weiskopf N, Suckling J, Williams G, et al. Quantitative multi‐parameter mapping of R1, PD*, MT, and R2* at 3T: a multi‐center validation. Front Neurosci. 2013;7:95. doi: 10.3389/fnins.2013.00095 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59. Reynolds LA, Morris SR, Vavasour IM, et al. Nonaqueous magnetization following adiabatic and selective pulses in brain: T1 and cross‐relaxation dynamics. NMR Biomed. 2023;36:e4936. doi: 10.1002/nbm.4936 [DOI] [PubMed] [Google Scholar]
- 60. Bapst B, Massire A, Mauconduit F, et al. Pushing MP2RAGE boundaries: ultimate time‐efficient parameterization combined with exhaustive T1 synthetic contrasts. Magn Reson Med. 2023;91:1608‐1624. doi: 10.1002/mrm.29948 [DOI] [PubMed] [Google Scholar]
- 61. Saekho S, Yip CY, Noll DC, Boada FE, Stenger VA. Fast‐kz three‐dimensional tailored radiofrequency pulse for reduced B1 inhomogeneity. Magn Reson Med. 2006;55:719‐724. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62. Zelinski AC, Wald LL, Setsompop K, et al. Fast slice‐selective radio‐frequency excitation pulses for mitigating B1+ inhomogeneity in the human brain at 7 Tesla. Magn Reson Med. 2008;59:1355‐1364. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63. Gras V, Vignaud A, Amadon A, Le Bihan D, Boulant N. Universal pulses: a new concept for calibration‐free parallel transmission. Magn Reson Med. 2017;77:635‐643. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
According to United Kingdom research councils' Common Principles on Data Policy and Wellcome Trust's Policy on data, software and materials management and sharing, all simulated data supporting this study is openly available at https://github.com/mriphysics/HybridPTx (hash 98fc91d at time of submission). This will exclude proprietary code from SIEMENS, but that can be shared on request by agreement including the vendor, and in vivo MRI data because of the terms of the ethical approval under which they were acquired.