Skip to main content
IJID Regions logoLink to IJID Regions
. 2024 Dec 16;14:100504. doi: 10.1016/j.ijregi.2024.100504

Comparison of the Abbott Alinity m and Qiagen Artus assays for the quantification of BK virus in clinical samples

Honorine Fenaux 1, Eric Marchadier 1, Alexandra Champagne 1, Corinne Prégermain 1, Lina Mouna 1,
PMCID: PMC11745809  PMID: 39835184

Highlights

  • Monitoring of BK virus (BKV) viral load helps prevent BKV-associated nephropathy

  • The Abbott Alinity m BKV AMPL kit is suitable for use in a diagnostic context

  • The Alinity m platform allows testing in a “random-access” manner

Keywords: BK virus, Kidney transplant recipients, Viral load, Alinity m, Interstitial nephropathy

Abstract

Objectives

BK virus (BKV) is highly seroprevalent in humans. After primary infection, it remains latent in the urinary tract and can reactivate in immunocompromised individuals, leading to interstitial nephropathy or hemorrhagic cystitis. The BKV viral load (VL) in plasma correlates with the occurrence of nephropathy and can be monitored in kidney graft recipients; the early detection of BKV viremia can enable an early reduction of immunosuppressant drug doses and the prevention of BKV-associated nephropathy. BKV VL is measured using real-time reverse transcription-polymerase chain reaction. Our aim was to evaluate the Abbott Alinity m BKV AMPL kit versus the Artus BKV quantification kit (Qiagen, Hilden, Germany).

Methods

We compared the methods on 69 plasma samples, 54 urine samples, and 12 external quality control samples. To assess analytical sensitivity, serial dilutions of internal control were also tested.

Results

Specificity reached 100%. Fifty-seven of the 59 plasma samples were found to be positive (sensitivity: 96.6%), as were 46 of the 48 urine samples (sensitivity: 95.8%). In quantifiable samples, the VL was correlated with the findings on plasma (r = 0.965) and urine samples (r = 0.971). The mean differences were 0.78 log10 and 0.28 log10 in plasma and urine samples, respectively. A comparison of the BKV VL in external quality control samples produced a correlation coefficient of 0.930 and a mean difference of 0.07 log10 IU/ml. The calculated analytical sensitivity was 15 IU/ml with 95% probability.

Conclusions

The Alinity m BKV AMPL kit is compatible with diagnostic use and can be used in a “random-access” manner, thus reducing the time between sampling and the results being available to clinicians.

Background

BK virus (BKV) belongs to the Polyomaviridae family and has a high rate of worldwide adult seroprevalence at approximately 80%. After primary infection, which happens most frequently during childhood, the virus remains latent in the urinary tract [1]. In immunocompromised individuals, it can reactivate and lead to interstitial nephropathy (kidney graft recipients) [2] or hemorrhagic cystitis (hematopoietic stem cell recipients) [3]. BKV-associated nephropathy can ultimately lead to graft loss.

The BKV plasmatic viral load (VL) correlates with the occurrence of nephropathy, especially with values >10,000 copies/ml for 28 days [2]. BKV can be excreted in the urine, even in asymptomatic immunocompetent individuals. BKV viruria is not correlated with nephropathy [4] or hemorrhagic cystitis, but the presence of BKV in the urine of an immunocompromised patient, particularly with a high VL, can suggest that the virus might be reactivated and that monitoring of the blood is warranted. There is no specific antiviral treatment for BKV reactivation, but it is possible to lower the doses of immunosuppressant drugs, which may help to control BKV reactivation and prevent BKV-associated nephropathy [4]. The sooner the immunosuppressant drug doses are lowered, the better the control of BKV reactivation: the monitoring of BKV VL is important to detect any reactivation before the occurrence of nephropathy and graft malfunction.

BKV VL is measured using real-time reverse transcription-polymerase chain reaction (RT-PCR).

A BKV quantification kit has recently been developed on the Alinity m platform (Abbott Molecular, Des Plaines, IL, USA); this would enable “random-access” processing and the timely delivery of biologic results compared with the many platforms that operate on a “batch” basis.

Objectives

Our aim was to compare the Abbott Alinity m BKV quantification kit versus the Artus BKV quantification kit (Qiagen, Hilden, Germany).

Study design

Reference method

BKV quantitative RT-PCR assays were performed on plasma and urine samples for diagnostic purposes using the QIAsymphony/RotorGeneQ (Qiagen); this was considered as the reference method. Nucleic acid was extracted using the QIAsymphony DSP Virus/Pathogen Midi Kit and BKV quantitative RT-PCR with the Artus® BKV-QS-RGQ kit (Qiagen). The results were expressed in copies/ml (and log10 copies/ml) according to the manufacturer's instructions. The limits of detection and quantification were 200 copies/ml – 2.30 log10 copies/ml. The samples were then stored at –80°C.

Comparison of methods

We tested 69 plasma samples and 54 urine samples on the Alinity m platform using the Alinity m BKV AMPL kit. Urine samples were diluted with the Alinity m Urine Transport kit according to the manufacturer's instructions. The results were expressed in IU/ml (and log10 IU/ml). The limit of detection was 50 IU/ml – 1.7 log10 IU/ml, and the limits of quantification were 50 IU/ml – 1.7 log10 IU/ml and 1,000,000,000 IU/ml – 9 log10 IU/ml, according to the manufacturer.

In the event of any discrepancies between the two methods, the procedure was repeated using both of them.

We tested 15 external quality control (EQC) samples (Quality Controls for Molecular Diagnostics, QCMD, Glasgow, United Kingdom), for which “reference” values expressed in IU/ml were available, which thus enabled us to compare them using the same units as Alinity m.

Sensitivity

To assess analytical sensitivity, an Alinity m BKV CTRL Kit (09N85-080) was used. Serial dilutions of low positive control (2.32-3.82 log10 IU/ml) and high positive control (5.40-6.90 log10 IU/ml) were tested. According to Abbott kit instructions, concentrations were standardized against the 1st World Health Organization International Standard for BK Virus DNA (NIBSC code: 14/212). The low positive internal control was diluted in a negative urine sample (in triplicate 1:10, 1:100, 1:1000, and 1:10,000). The high positive internal control was also diluted in a negative urine sample (in duplicate 1:10, 1:100, 1:1000, 1:10,000; in quadruplicate 1:30,000; in triplicate 1:50,000).

Repeatability and reproducibility

Repeatability and reproducibility were assessed as a first step using positive plasma and urine samples tested three times on the same day and twice more on two different days. As a second step, a diluted plasma sample and a diluted urine sample (dilution in negative plasma and urine samples, respectively) were tested 10 times on the same day (repeatability) and five times on day 1, five times on the morning of day 2, five times in the afternoon of day 2, and five times on day 3 (reproducibility).

Statistics

The VLs obtained with the two kits were compared using Analyse-it v5.4 software.

Results

Comparison of methods

The results obtained for all samples can be found in Supplementary data (Tables S1 and S2).

All 10 plasma samples that were negative with the reference method also tested negative with the Alinity m kit. Of the 59 positive plasma samples, 57 were found to be positive (sensitivity: 96.6%), whereas the other two samples were low positive (<3 log10 copies/ml) (Table 1).

Table 1.

Results of plasma samples when comparing the methods.

Alinity m (log10 IU/ml)
Not detected Not detected Low positive Positive Total

Artus (log10 copies/ml)
10 0 0 10
Low positive 2a 15 0 17
Positive 0 14 28 42
Total 12 29 28 69
a

Low positive (<3 log10 copies/ml).

The six negative urine samples also tested negative with the Alinity m kit. Of the 48 positive urine samples, 46 were positive (sensitivity: 95.8%), whereas the other two samples were low positive (<3 log10 copies/ml) (Table 2).

Table 2.

Results of urine samples when comparing the methods.

Alinity m (log10 IU/ml)
Not detected Not detected Low positive Positive Total

Artus (log10 copies/ml)
6 0 0 6
Low positive 2a 12 0 14
Positive 0 2 32 35
Total 8 14 32 54
a

Low positive (<3 log10 copies/ml).

In quantifiable samples, VLs were correlated in both plasma (r = 0.965) and urine samples (r = 0.971) (Figure 1, Figure 2). The mean differences were 0.78 log10 and 0.28 log10 in plasma and urine samples, respectively.

Figure 1.

Figure 1

Comparison of viral loads in plasma samples between the reference method (Qia) and the tested method (Alinity). (a) Bland-Altman test. (b) Passing-Bablok test.

Figure 2.

Figure 2

Comparison of viral loads in urine samples between the reference method (Qia) and the tested method (Alinity). (a) Bland-Altman test. (b) Passing-Bablok test.

The comparison of BKV VL in EQC samples is shown in Table 3 and Figure 3; r = 0.930, the mean difference being 0.07 log10 IU/ml.

Table 3.

Results of external quality control samples.

Alinity m (log10 IU/ml)
Not detected Low positive Positive Total
Quality Controls for
Molecular
Diagnostics,
QCMD, Glasgow,
United Kingdom
(log10 IU/ml)
Not detected 3 0 0 3
Low positive 0 3 1 4
Positive 0 0 8 8
Total 3 3 9 15

Figure 3.

Figure 3

Comparison of viral loads in EQC samples between their consensus values (supplied by the QCMD organization) and those obtained with the Alinity test. (a) Bland-Altman test. (b) Passing-Bablok test.

EQC, external quality control.

Sensitivity

The probit analysis showed that a VL of 15 IU/ml could be detected with a 95% probability (probit analysis, P < 0.0001) (Supplementary data S3).

Repeatability and reproducibility

Inter-assay and intra-assay coefficients of variation were assessed in two steps, as explained in the study design.

Inter-assay coefficients of variation:

  • In plasma, the coefficient of variation varies between 0.40% (highest tittered sample) and 1.70% (lowest tittered samples).

  • In urine samples, the coefficient of variation varies between 0.38% (highest tittered sample) and 3.94%.

Intra-assay coefficients of variation:

  • In plasma, the coefficient of variation varies between 0.80% and 1.19% (lowest tittered samples).

  • In urine samples, the coefficient of variation varies between 0.43% (highest tittered sample) and 2.20% (Table 4).

Table 4.

Results of repeatability and reproducibility assays.

Image, table 4

VL = viral load.

Discussion

We tested a method for the quantification of BKV VL using an Abbott Alinity m platform. The results showed a high correlation, especially with those obtained on EQC, for which a “reference” value in IU/ml was available; a variation of less than 0.5 log10 was not considered significant.

The analytical sensitivity declared by the manufacturer is 50 IU/ml, which is higher than what we found (15 IU/ml). However, these very low VL values are not correlated with BKV-associated nephropathy; thus this finding would make no difference in terms of follow-up.

Repeatability and reproducibility vary among samples, particularly urine samples, but are still within the permitted and recognized limits. Of note, urine samples need to be diluted to obtain the required VL, which might have had consequences in terms of repeatability and reproducibility and could partly explain the results. In the context of molecular biology methods, a difference of less than 0.5 log10 is not considered to be significant; the probability of such an event occurring is 5% when the coefficient of variation is 40% [5].

The present study had several limitations, such as a modest number of samples. Further studies are needed to complete our results. In addition, developing novel biomarkers for the diagnosis of BKV-associated nephropathy, such as the measurement of virus-specific T cell level and the development of cellular therapy, is ongoing and will be complementary to BKV VL tests. The Alinity m platform is practical, and assays are already available on that platform for viruses other than BKV: hepatitis B and C viruses, human immunodeficiency virus, and human papillomaviruses. It can, therefore, be applied to test a range of viruses, which could serve as a criterion for choosing between different methods. The Alinity m platform enables “random-access,” unlike our reference method: Qiagen platform operates on a “batch” mode with little flexibility, which may cause delayed availability of results while pending a sufficient number of samples to complete a batch to avoid increased costs. The Alinity m platform is a fully automated “random-access” system with a time-to-first result of less than 115 minutes and a throughput of up to 300 samples in 8 hours. That allows for rapid time-to-first results for the benefit of patients. The potential of this platform is to significantly improve workflow efficiency in diagnostic laboratories.

To conclude, the Alinity m BKV AMPL kit is compatible with diagnostic use and can be used in a “random-access” manner, thus reducing the time between sampling and results being available to clinicians.

Declarations of competing interest

The authors have no competing interest to declare.

Acknowledgments

Funding

This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors. Abbott provided reagents used in this study.

Ethics statement

This study was conducted in the Virology Laboratory at Paul Brousse Hospital (Paris, France) in accordance with the requirements of the Declaration of Helsinki as a retrospective non-interventional study with no addition to standard-of-care procedures. The reclassification of biologic remnants into research material after completion of the virologic tests that had been ordered was registered under number DC 2009-965 and received ethical approval from CPP Ile de France 7 (N°CO-15-000) in accordance with French law. No need for individual consent.

Acknowledgments

We are indebted to Prof Anne-Marie Roque-Afonso, who assisted us with the statistics on the Analyse-it v5.4 software.

Author contributions

H.F. and L.M.: conceived and planned the experiments. designed the model and the computational framework and analysed the data. performed the calculations. wrote the manuscript. E.M., A.C. and C.P.: performed the measurements/ analysis.

Data availability statement

We provide all data that support the findings of this study.

Footnotes

Supplementary material associated with this article can be found, in the online version, at doi:10.1016/j.ijregi.2024.100504.

Appendix. Supplementary materials

mmc1.xlsx (33.9KB, xlsx)
mmc2.docx (16.8KB, docx)

References

  • 1.Verghese PS, Schmeling DO, Filtz EA, Matas AJ, Balfour HH. The impact of recipient BKV shedding before transplant on BKV viruria, DNAemia, and nephropathy post-transplant: a prospective study. Pediatr Transplant. 2017;21:e12942. doi: 10.1111/petr.12942. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Borriello M, Ingrosso D, Perna AF, Lombardi A, Maggi P, Altucci L, et al. BK virus infection and BK-virus-associated nephropathy in renal transplant recipients. Genes (Basel) 2022;13:1290. doi: 10.3390/genes13071290. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Zhou X, Zhu C, Li H. BK polyomavirus: latency, reactivation, diseases and tumorigenesis. Front Cell Infect Microbiol. 2023;13 doi: 10.3389/fcimb.2023.1263983. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Querido S, Weigert A, Pinto I, Papoila AL, Pessanha MA, Gomes P, et al. Detection of JCV or BKV viruria and viremia after kidney transplantation is not associated with unfavorable outcomes. J Med Virol. 2023;95:e28800. doi: 10.1002/jmv.28800. [DOI] [PubMed] [Google Scholar]
  • 5.Reed GF, Lynn F, Meade BD. Use of coefficient of variation in assessing variability of quantitative assays. Clin Diagn Lab Immunol. 2002;9:1235–1239. doi: 10.1128/cdli.9.6.1235-1239.2002. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

mmc1.xlsx (33.9KB, xlsx)
mmc2.docx (16.8KB, docx)

Data Availability Statement

We provide all data that support the findings of this study.


Articles from IJID Regions are provided here courtesy of Elsevier

RESOURCES