. 2024 Dec 5;102(12):1009–1012. [Article in French] doi: 10.62438/tunismed.v102i12.5143

Cytologie urinaire: L'automate Iris DxU-850 versus examen microscopique

Urinary Cytology: The DxU-850 Iris automaton versus Microscopic Examination

Hasnae Lekfif 1,1,2, Abderrazak Saddari 1,1,2,3, Said Ezrari 1,1,3, Amjad Idrissi 1,1,2, Elmostafa Benaissa 4,4,5, Yassine Ben Lahlou 4,4,5, Mostafa Elouennass 4,4,5, Adil Maleb 1,1,2,3
PMCID: PMC11770784  PMID: 39748685

ABSTRACT

Introduction: The use of urine cytobacteriological examination is a common and essential practice in medicine which helps guide therapeutic management in case of urinary tract infection. The cytological examination of urine samples can be done using the manual (microscopic) or automated technique. The automated approach, which involves the use of artificial intelligence, is faster, more reliable, and more efficient for laboratories. Aim: This work aims to evaluate the performance of the DxU-850 Iris automate by comparing it to the microscopic method. Methods: Using a four-month prospective study from May to August 2023, we analyzed urine samples received during this period by the Microbiology laboratory of Mohammed VI University Hospital in Oujda. Strict quality standards were respected when obtaining the results. Several elements were analyzed, and various parameters calculated to evaluate the performance of this automaton. Results: The present study investigated 1000 CBEU samples, mainly from outpatient clinics (46.34%) and emergency departments (21.72%). Most patients were males (50.58%) and asymptomatic (72.9%). The samples’ culture showed varied results: concordance between the results of automaton and microscopic count was satisfactory except for crystals and yeasts in turbid samples. Conclusion: Automation has advantages like speed, traceability, error reduction, and time optimization. The results reveal a satisfactory concordance between the two methods despite discordances occurring mainly with crystals and yeasts, especially in turbid urine samples. The DxU-850 Iris automaton is an effective way to screen for urinary tract infections. We recommend that automated and manual techniques be considered complementary.

Introduction

Urinary tract infection (UTI) is one of the most common infections in humans 1.

It is characterized by the presence of significant bacteriuria and leukocyturia 1.

The annual incidence of UTIs is estimated at more than 150 million cases worldwide, with high morbidity and significant medical costs 2.

Diagnosis is based on urine cytobacteriological examination, which, due to the high frequency of this infection, is the most common test in a medical microbiology laboratory 37 .

The cytological component can be evaluated either manually using a microscope or automatically.

Despite the affordability of the microscopic approach, it has the disadvantage of being time-consuming and requires an experienced operator 6 , 811.

In this context, the use of artificial intelligence is a breakthrough for laboratories, offering automated solutions that optimize workflow by minimizing the need for manual sample processing.

This approach thus contributes significantly to reducing sample processing times and improving the overall efficiency of laboratory operations.

Besides, the automated method helps overcome these problems by providing faster results, offering enhanced traceability, reducing the risk of errors, and optimizing the technician’s time 5, 6, 8, 10, 12.

The objective of this study was to evaluate the performance of the DxU-850 Iris automate (Beckman Coulter, Danaher, USA) by comparing it to the microscopic method (reference method), in accordance with current guidelines and real-world conditions of daily practice at the Microbiology Laboratory of Mohammed VI University Hospital, Oujda.

Methods

This is a four-month prospective study from May to August 2023, focusing on urine samples received in the same period by the Microbiology laboratory of the Mohamed VI University Hospital, Oujda.

Samples came from different patients, including pregnant women, children, immunocompromised individuals, patients following antibiotic treatment, patients with urine collection devices, people admitted to the intensive care unit and/ or patients who have undergone transplantation.

All analyzes were performed under conditions of ambient temperature and relative humidity in accordance with the manufacturer's recommendations.

The study included all urine samples sent to our laboratory for a Urinary Cytobacteriological Examination.

All samples were sent to the laboratory within 30 minutes.

The standard sample processing protocol in our laboratory is to collect 10 µL of each urine sample, which is then seeded on the Brilliance UTI Agar chromogenic medium (Oxoid Brilliance™, UK) and then incubated from 24 to 48 hours.

The germs were identified using the BD Phoenix 100 system (Becton Dickinson, USA).

The results are interpreted based on the REMIC criteria (version 7; 2022) 13, helping in classifying the samples into one of the following three categories: Urinary tract infection (UTI), urinary colonization, and contamination.

Immediately after urine seeding, they were examined both by the DxU-850 Iris automaton and under the microscope using the Fast-Read 102® counting cell (Dominique Dutscher SAS, Fr) with a magnification of (40X).

The elements studied after analysis with the DxU850 Iris automaton were the number of white blood cells, the number of red blood cells, the number of epithelial cells, and the presence of yeast, crystals, and casts.

In parallel, the microscopic examination helped in counting of the same elements.

To compare the results of the two techniques, we determined positivity thresholds compatible with the clinical-microbiological requirements of the recommendations we follow 13 (Table1).

Table 1. Significant positivity thresholds in our study to assess the degree of concordance between the results of the DxU-850 Iris automaton and manual microscopy.


Table 1. Significant positivity thresholds in our study to assess the
degree of concordance between the results of the DxU-850 Iris
automaton and manual microscopy.

We calculated various parameters, including the number of true positives, true negatives, false positives, and false negatives.

Such calculation allowed us to determine the sensitivity, specificity, positive predictive value, negative predictive value, Youden's index, Yule's Q coefficient, and the Chi2 (χ2) index.

Results

Over the study period, a total of 1,000 urine cultures were analyzed.

They came mainly from outpatient clinics (46.34%; n=463), as well as from the emergency department (21.72%; n=217), cardiology (5.6%; n=56), intensive care (4.7%; n=47) and nephrology (4.5% n=45).

Of the patients whose urine cultures were analyzed, 509 (50.85%) were male.

The average age was 48.72 ±23.17 years.

The distribution of patients according to age group is presented in Table 2.

Table 2. Distribution of patients with urine cultures according to age group (n=1000).


Table 2. Distribution of patients with urine cultures according to age
group (n=1000).

In our series, 72.9% (n=729) were asymptomatic, while 27.1% (n=271) had symptoms of UTI.

There were 53 (5.3%) patients with indwelling urinary devices.

Requests for urine cultures in our series were motivated by an etiological diagnosis of UTI in 63.2% (n=632) of cases, by a search for urinary colonization in 36.3% (n=363), or by therapeutic monitoring in 0.5% (n=5).

The culture in our series was sterile in 615 (61.5%) cases; it showed contamination in 343 (34.3%) cases, colonization in 21 (2.1%) cases and UTI in 21 (2.1%) cases.

We isolated 45 germs in all UTI cases and colonization.

Among these germs, Escherichia coli was by far the most isolated (n=28; 62.2%), followed by Candida spp. (N=8; 17.8%).

As for urinary cytology, the results of the correlation between the automated and manual methods are shown in Table 3.

Table 3. Comparison of results of the automated DxU-850 Iris and microscopy (n=1000).


Table 3. Comparison of results of the automated DxU-850 Iris and
microscopy (n=1000).

Discussion

A Urinary Cytobacteriological Examination must necessarily examine urinary cytology and count urinary germs 6, 9.

Urinary cytology can be evaluated manually or automatically.

The manual method uses microscopic counting in a counting chamber

It is time-consuming, but its affordability makes it the most utilized in laboratories 811.

Automation allows for rapid results (higher throughput), better traceability (connection to the laboratory's computer system), reduced risk of error (manual transcription errors) and optimized technician time (reduction of repetitive tasks) 5, 6, 8, 10, 12.

The DxU-850 Iris automated urinalysis system is a fully automated urinalysis solution based on the innovative Digital Flow Morphology technology with the automated particle recognition software (APR) 14.

This automatic classification is first checked, then validated and saved by the operator.

If any abnormal particles are detected, the user can modify this classification by conducting a more detailed analysis of the images before validating them.

This ensures that the automaton (machine learning) accurately memorizes the updated information.

This technology allows laboratories to provide standardized results using Artificial Intelligence.

It allows for the isolation, identification, and characterization of digital images of particles on the screen, thus considerably reducing the need for manual microscopy, all with an impressive throughput of 101 samples per hour.

It automatically identifies and processes samples in 10-position holders, presenting a sample sandwiched between enveloping layers of a membrane to a microscope coupled to a charged coupling device (CCD) video camera.

Individual particle images are isolated in each image.

The automated particle recognition software (APR), a highly trained neural network, uses size, shape, contrast, and texture features to classify each image into one of the following 12 categories: red blood cells, white blood cells, white blood cell clusters, hyaline casts, unclassified casts, squamous epithelial cells, non-squamous epithelial cells, bacteria, yeasts, crystals, mucus, and spermatozoa.

The particle concentration is calculated using the number of particle images and the analyzed volume.

User-defined release criteria are checked, and the results are sent to an operator review screen 14.

In our study, the concordance rates of results, whether true positives or true negatives, between DxU-850 Iris and the microscope, are very satisfactory for red blood cells, white blood cells, and casts.

The discordances in our series are slightly more frequent for false negatives for red blood cells and largely dominated by false negatives for yeasts and crystals.

We observed the high false negative rate for yeasts and crystals mainly in turbid urine samples.

Sensitivity and specificity remain satisfactory.

However, it is important to note that sensitivity is lower for yeasts and crystals compared to other parameters.

Several studies, in the literature, have reported the results of comparing the manual method with automated methods, including those using the flow cytometry technique (such as the Sysmex UF-1000i automated system).

They noted discordances slightly dominated by false negatives for white blood cells and red blood cells, and largely dominated by false positives for yeasts and crystals 15.

Sensitivity ranged between 70.9% and 98.8%, and specificity between 73.3% and 90.85% for white blood cells 5, 7, 16, 17.

As for other automated systems using the same technology as DxU-850 Iris (such as the iQ1200 ELITE automated system), one study showed a sensitivity ranging between 91.57% and 100%, and a high specificity ranging from 87.33% to 93.33% for white blood cells and casts, respectively. However, specificity for red blood cells was lower 18.

As for epithelial cells, their interpretation remains a crucial element in the bacteriological examination of urine samples.

The presence of Group 4 microorganisms, defined by their role in the etiology of urinary tract infections and considered contaminants from the urethral or genital flora, combined with the presence of urinary epithelial cells in the microscopic examination of urine, almost certainly indicates contamination during the collection process 13, 15.

In our study, concordance between the results of DxU-850 Iris automated system and manual microscopy for epithelial cells was perfect.

Youden's indices, Yule's Q coefficient and Chi2 (χ2) test for white blood cells, red blood cells and casts indicate that DxU-850 Iris is effective, that its results are strongly correlated with those of the microscope, and that this concordance is statistically significant.

Conclusion

Our study has found that DxU-850 Iris automated urinalysis system provides results in satisfactory concordance with those obtained by manual microscopy.

Having shown the predominance of discordance for crystals and yeasts, mainly in turbid urine samples, we recommend that the automated and manual techniques be considered as complementary.

References

  1. García-Coca M, Gadea I, Esteban J. Relationship between conventional culture and flow cytometry for the diagnosis of urinary tract infection. J Microbiol Methods. 2017;137:14–18. doi: 10.1016/j.mimet.2017.03.010. [DOI] [PubMed] [Google Scholar]
  2. Mancuso Giuseppe, et al. Infections des voies urinaires : le scénario actuel et les perspectives d'avenir. Pathogens. 2023;12(4):623. [Google Scholar]
  3. Caron F, et al. Practice guidelines for the management of adult community-acquired urinary tract infections. Med Mal Infect. 2018;48(5):327–358. doi: 10.1016/j.medmal.2018.03.005. [DOI] [PubMed] [Google Scholar]
  4. Gessoni G, Saccani G, Valverde S, Manoni F, Caputo M. Does flow cytometry have a role in preliminary differentiation between urinary tract infections sustained by gram positive and gram negative bacteria? An Italian polycentric study. Clin Chim Acta. 2015;440:152–156. doi: 10.1016/j.cca.2014.11.022. [DOI] [PubMed] [Google Scholar]
  5. Íñigo M, Coello A, Fernández-Rivas G, Carrasco M, Marcó C, Fernández A, et al. Evaluation of the SediMax automated microscopy sediment analyzer and the Sysmex UF-1000i flow cytometer as screening tools to rule out negative urinary tract infections. Clin Chim Acta. 2016;456:31–35. doi: 10.1016/j.cca.2016.02.016. [DOI] [PubMed] [Google Scholar]
  6. Twizeyimana E. Automates et uroculture: la cytologie urinaire. Rev Francoph Des Lab. 2016;2016:25–33. [Google Scholar]
  7. March-Rosselló GA, Gutiérrez-Rodríguez MP, Simarro-Grande M, Orduña-Domingo A, Bratos-Pérez MÁ. Evaluación del analizador de orinas Sysmex UF-1000i como método de cribado en el diagnóstico de la infección del tracto urinario. Rev Del Lab Clínico. 2016;9:3–8. [Google Scholar]
  8. Sánchez-Mora C, Acevedo D, Porres MA, Chaqués AM, Zapardiel J, Gallego-Cabrera A, et al. Comparison of automated devices UX-2000 and SediMAX/AutionMax for urine samples screening: a multicenter Spanish study. Clin Biochem. 2017;50:714–718. doi: 10.1016/j.clinbiochem.2017.02.005. [DOI] [PubMed] [Google Scholar]
  9. Fabbro C, Darolles J, Rault J-P. Évaluation des performances de l’automate d’analyse urinaire UF-1000i®. Ann. Biol. Clin. (Paris) 2011;69:431–439. doi: 10.1684/abc.2011.0604. [DOI] [PubMed] [Google Scholar]
  10. Shang Y-J, Wang Q-Q, Zhang J-R, Xu Y-L, Zhang W-W, Chen Y. Systematic review and meta-analysis of flow cytometry in urinary tract infection screening. Clin Chim Acta. 2013;424:90–95. doi: 10.1016/j.cca.2013.05.014. [DOI] [PubMed] [Google Scholar]
  11. Manoni F, Tinello A, Fornasiero L, Hoffer P, Temporin V, Valverde S. Urine particle evaluation: a comparison between the UF-1000i and quantitative microscopy. Clin Chem Lab Med. 2010;48:1107–1111. doi: 10.1515/CCLM.2010.233. [DOI] [PubMed] [Google Scholar]
  12. de Frutos-Serna M, Asensio-Calle ML, Haro-Pérez AM, Blázquez-de Castro AM, Gutiérrez-Zufiaurre MN, Iglesias-García J. Evaluación del citómetro UF-1000i como método de cribado en el diagnóstico de infecciones del tracto urinario. Enferm Infecc Microbiol Clin. 2014;32:147–151. doi: 10.1016/j.eimc.2013.02.015. [DOI] [PubMed] [Google Scholar]
  13. SFM FS de M. Rémic: Réferenciel en microbiologie médicale. 2022 [Google Scholar]
  14. Coulter B. Manuel automate DxU-850 Iris. n.d. [Google Scholar]
  15. Maleb A, Sebbar E-H, Lahlou Y Ben, Frikh M, Mikou KA , Lemnouer A. Cytologie urinaire: UF-1000i versus examen microscopique, dans des conditions réelles d’exercice d’un laboratoire de microbiologie. IRBM News. 2017;38:150–154. [Google Scholar]
  16. Boonen KJM, Koldewijn EL, Arents NLA, Raaymakers PAM, Scharnhorst V. Urine flow cytometry as a primary screening method to exclude urinary tract infections. World J Urol. 2013;31:547–551. doi: 10.1007/s00345-012-0883-4. [DOI] [PubMed] [Google Scholar]
  17. De Rosa R, Grosso S, Bruschetta G, Avolio M, Stano P, Modolo ML. Evaluation of the Sysmex UF1000i flow cytometer for ruling out bacterial urinary tract infection. Clin Chim Acta. 2010;411:1137–1142. doi: 10.1016/j.cca.2010.03.027. [DOI] [PubMed] [Google Scholar]
  18. Dewulf G, Harrois D, Mazars E, Cattoen C, Canis F. Évaluation des performances de l’automate de cytologie urinaire Iris iQ® 200 ELITE et comparaison avec la méthode manuelle microscopique. Pathol Biol. 2011;59:264–268. doi: 10.1016/j.patbio.2009.10.006. [DOI] [PubMed] [Google Scholar]

Articles from La Tunisie Médicale are provided here courtesy of Tunisian Society of Medical Sciences