Abstract
Aims
18F-sodium fluoride (18F-NaF) positron emission tomography (PET) detects active microcalcification and predicts adverse outcomes including bioprosthetic valve deterioration. However, measuring small areas of 18F-NaF uptake within moving structures remains challenging, requiring further optimization. We developed a representative cardiac phantom to optimize 18F-NaF imaging of bioprosthetic valves.
Methods and results
We placed a bioprosthetic valve with two pockets sutured to the leaflets mimicking valvular lesions and a subvalvular ring mimicking the valve remnant into the phantom and injected each with 18F-radionuclide (1 μCi pockets, 4 μCi ring). We injected the cardiac chambers with iohexol and 18F-radionuclide (0.176 mCi) for background activity. PET and computed tomography (CT) images were acquired using a Siemens Biograph Vision high-resolution digital PET/CT scanner. We analysed target-to-background ratio (TBR) and signal-to-noise ratio (SNR) and subjective measures of image quality. We compared results with a human case of transcatheter aortic valve replacement. Initially the SNR and TBR in the phantom greatly exceeded those from human imaging. We reduced the scan duration used for reconstruction to 30 and 15 s, achieving comparable results (30 s vs. 15 s vs. patient: SNR 45.6 vs. 13.9 vs. 44.3, TBRmax 6.5 vs. 5.4 vs. 4.1, noise 10.2% vs. 8.8% vs. 12.0%). With motion correction, SNR and image quality improved in the phantom (30 s 135.8 vs. 45.6, 15 s 32.9 vs. 13.9) but remained similar in the human case (47.3 vs. 44.3).
Conclusion
A cardiac phantom can mimic clinical 18F-NaF valve bioprosthesis imaging, providing an opportunity to explore acquisition, reconstruction, and post-processing of 18F-NaF PET/CT for small mobile cardiac structures.
Keywords: positron emission tomography, cardiac phantom, cardiac computed tomography, multi-modality imaging
Graphical Abstract
Graphical Abstract.
Introduction
Positron emission tomography (PET) is an imaging modality that allows direct visualization of disease activity, subject to the availability of a radiolabelled tracer that targets the disease process of interest. Fluorine-18 labelled sodium fluoride (18F-NaF) is a radiotracer targeting active microcalcification.1 Its presence provides important prognostic information in many cardiovascular diseases in which calcification is a key driver, including coronary artery disease, aortic stenosis, and, most recently, degeneration and failure of aortic valve bioprostheses.2–7
Prosthetic valves are used to replace dysfunctional or damaged cardiac valves. This can be due to stenosis, where progressive calcification and narrowing of the valves impede the ejection of blood from or within the heart8,9; regurgitation, where blood leaks back due to failure of adequate valvular coaptation; or infection of the valvular apparatus due to infective endocarditis. A significant proportion of prosthetic valve replacements are implanted for stenosis of the aortic valve, a condition whose prevalence and severity increases with age.10
However, although it addresses the underlying pathology, valve replacements are not without their own complications. Bioprosthetic valve degeneration leading to failure is increasingly well-recognized.3,4 Factors predicting valvular failure are not completely understood, but calcification seems to be a key driver in their degeneration.3,4,11,12 In particular, as the population ages, there will be an increase in age-related conditions such as aortic stenosis where valve replacement remains the only treatment option.10 An increased prevalence of those with bioprosthetic valve replacements and therefore a larger population at risk of bioprosthetic degeneration and failure is anticipated. The ability to predict degeneration and future calcification with cardiovascular 18F-NaF PET imaging is therefore a highly relevant and important discovery in clinical cardiology.3,4,11,12
However despite these research advances, small, dynamic, and highly mobile structures, such as the heart valves, continue to present major technical challenges for PET imaging acquisition and analysis in clinical practice making accurate quantification of disease activity difficult.13 In addition, myocardial, respiratory, and patient motion can also degrade the PET signal quality.13
Phantoms are often used in optimization of imaging acquisition processes.14–16 The benefit of phantoms is the ability to repeat image acquisitions using different parameters on multiple occasions as unlike in human imaging, radiation burden and patient comfort and inconvenience are eliminated as limiting factors. Cardiovascular phantom models have been used effectively to investigate and to optimize PET image acquisition and reconstruction in nuclear cardiology, with some initial data concerning coronary imaging,14–16 but these studies did not include motion or valve imaging models.
In this work, we aimed to mimic in vivo human cardiovascular imaging conditions of degenerative bioprosthetic aortic valve disease using a bespoke representative cardiac phantom. Our goals were to optimize imaging acquisition, reconstruction, and post-processing protocols to image simulated degenerating valve bioprostheses using 18F-radionuclide PET and computed tomography (CT) imaging of our cardiac phantom.
Materials and methods
Cardiac phantom and accessories
We developed our cardiac phantom in collaboration with Archetype BioMedical Inc. (London, Canada) who produced the final product according to our specifications and requirements. With two different models comprised of silicon or polyvinyl alcohol cryogel, respectively, the phantom consists of four chambers mimicking the left atrium, two ventricles, and proximal aorta and two valve modules in the mitral and aortic position (Figure 1). The ventricles contract in response to the device control panel which can be adjusted to reflect a range of physiological human heart rates. A customized pulsatile flow pump stationed adjacent to the phantom in a watertight container simulates blood flow through the chambers and thus passive motion of the valvular structures as occurs in vivo.
Figure 1.
Our realistic cardiac phantom is displayed with imaging accessories to mimic bioprosthetic valve degeneration. (A and B) Photographs of our bespoke cardiac phantom displaying the different chambers and components, as well as the transparent polycarbonate box that contains it and the electronic module that allows variation of the contraction rate and force. (C) A picture of bioprosthetic aortic valve, hollow silicon ring, and valve assembly. (D) A 3D-printed plaque with pockets onto a bioprosthetic valve leaflet to mimic degeneration and allow radiotracer infiltration for PET imaging sessions. (E) A technical drawing of the phantom.
To allow us to simulate regions of radiotracer uptake, 3D-printed silicon modules were also developed by Archetype BioMedical in order to incorporate one small (0.04 mL) and one large (0.12 mL) pocket on the valve bioprosthetic leaflets, and a hollow ring (0.5 mL) that was placed in the subvalvular position to mimic disease activity from the remnant of the native aortic valve as seen in patients who undergo valve replacement (Figure 1).
We utilized two Medtronic Mosaic tissue valves (Medtronic, Minneapolis, MN, USA). The two pocket modules were sutured with care to valve leaflets on one of the Medtronic Mosaic bioprostheses to mimic tracer-avid lesions often observed on degenerating aortic bioprostheses (Figure 1).
Image acquisition
All imaging was performed at Cedars-Sinai Medical Center, Los Angeles, on a Siemens Biograph Vision high-resolution digital PET/CT scanner (Siemens Healthcare, Germany) (Figure 2). In total, we performed five phantom imaging acquisitions using either contrast-enhanced cardiac CT (CCT, n = 3) or PET/CT (n = 2). Our objectives were as follows: (i) to establish appropriate iohexol volume to maintain a 350–500 Hounsfield units (HU) attenuation within the heart and aorta for optimal valve bioprosthesis visualization on the CCT acquisition, (ii) to isolate the atrial chamber from the transparent polycarbonate box containing the phantom, and (iii) to optimize physiological parameters and gating for CCT and attenuation correction sequence. The final two sessions also included a PET acquisition. Here, we aimed to determine tracer dosage to mimic (i) regions of interest (ROIs) of a degenerating aortic valve bioprosthesis and (ii) background activity.
Figure 2.
Preparation of the cardiac phantom for hybrid PET/CT imaging (A). Following the activity injection into the dedicated pockets, the chamber is filled with 700 mL of 0.9% sterile sodium chloride and 50 mL iodinated contrast (iohexol 350 mg iodine/mL, Omnipaque 350™) (B). Subsequently, the phantom is ready for scanning and emission scanning commences.
CT acquisition and reconstruction
Initially, CT attenuation correction of the phantom was performed with a craniocaudal acquisition of 1-mm slice thickness over a large (780-mm) field of view (FoV). Following this, a prospectively gated axial CCT was acquired from the cranial to the caudal position with thin (1-mm) slice thickness over a 260-mm FoV. All acquisitions used a tube voltage of 120 kVp. CCT was acquired using prospective gating over 10 gates of the simulated cardiac cycle. Optimal physiological conditions in human imaging were mimicked by a simulated ventricular contraction rate of 60 bpm which provided excellent image quality with precise characterization of the aortic bioprosthesis at all stages of the cardiac cycle.
An iterative reconstruction algorithm (sinogram affirmed iterative reconstruction; SAFIRE)17 was then applied to the 10-gate CCT sequence obtained with soft tissue (Br38) kernels. Reconstructions of PET data were all performed using dedicated reconstruction from Siemens. The PET matrix size was 440 × 440 × 524. Gaussian post-filtering (2 mm SD) was applied using FusionQuant software (FusionQuant, Cedars-Sinai Medical Center, Los Angeles)18 with pixel size 0.83 × 0.83 × 0.5 mm.
CT contrast density determination
We filled the phantom with 700 mL of 0.9% sterile sodium chloride and introduced iodinated contrast agent (iohexol 350 mg iodine/mL, Omnipaque 350™) in 25 mL increments. Administration of 50 mL iohexol 350 mg iodine/mL achieved a cardiac chamber attenuation of 350–500 HU providing excellent visualization of the aortic valve bioprosthesis and the ventricular walls (Figure 3). These concentrations were maintained for the duration of image acquisition, unlike in human imaging when contrast quickly dissipates throughout the circulating volume.
Figure 3.
Hybrid PET/CCT acquisition of the entire silicon cardiac phantom model in horizontal and vertical long axes (top row), with short- and long-axis views focusing on the degenerative aortic valve module (bottom row). Tracer activity representative of remnant native valve disease (green arrows) and leaflet degeneration (blue arrows) is displayed.
PET radiotracer
Because our ex vivo phantom model incorporating aortic valve bioprostheses does not feature structural abnormalities, which are present in the aortic valve disease process of interest, we substitute these with hollow modules into which a radiotracer can be injected. Therefore, any tracer can be used with mutually applicable results provided it is radiolabelled identically to the tracer of interest. For this reason, we used excess clinical doses of 18F-fluorodeoxyglucose (18F-FDG) in our phantom experiments in place of 18F-NaF.
PET acquisition and reconstruction
In acquisitions where PET was included, PET images were acquired following acquisition of CT attenuation correction and CCT as described above. The injected dose to the phantom was corrected for the decay. A 1 μCi 18F-FDG was injected into each of the larger (0.12 mL) and smaller (0.04 mL) pockets on the bioprosthesis leaflets. The ring on the periphery of the valve was filled with 4 μCi (0.5 mL) of 18F-FDG. We injected 0.176 mCi of 18F-FDG into the 750 mL heart chambers to mimic background activity seen in in vivo imaging. This also allowed us to assess noise and the signal-to-noise ratio (SNR). A 10-min list-mode PET acquisition was performed 10 min following injection of the radiotracer to whichever module that was injected first (Figure 2). This allowed for several independent 15-s reconstructions. The data set was reconstructed using an ultra-high definition algorithm with time of flight on a 440 × 440 matrix (slice thickness 0.8 mm, pixel spacing 0.8 mm) using four iterations and five subsets. Rather than embedding Gaussian filtering within the reconstruction, we performed it post-reconstruction in a dedicated software package.
PET cardiac motion correction
Motion correction was performed to align PET images from all 10 cardiac gates preserving all counts. This was done using a mass-preserving, diffeomorphic, anatomically guided Demons method to optimize global energy between PET frames. In the initial step, we created a sphere encompassing the ROI labelled as a valvular ROI. We utilized a non-linear registration algorithm with radial constraint to align PET data to the diastolic gate. PET signal [maximum target-to-background ratio (TBRmax), SNR] was quantified before and after the application of motion correction.19–21
Image analysis
Analysis of PET imaging was performed using FusionQuant. A ROI was drawn over each of the three bioprosthesis ROIs into which the 18F-labelled tracer had been administered, namely, the two silicon pockets sutured to the bioprosthesis leaflets and the subvalvular ring. A 2-mm 3D Gaussian filtering was performed within FusionQuant. We then measured the maximum standardized uptake value (SUVmax) for each ROI. These values were corrected for mean blood pool activity obtained from the atrial chamber to give a maximum TBR (TBRmax). Image noise was calculated by dividing bloodpool SUVmax by the standard deviation of the bloodpool SUVmean. The SNR was then calculated by dividing the maximum uptake in each ROI by the image noise. The two valvular lesions were measured using a 4 (small) and 5 (large) mm sphere, and the native valve uptake was assessed using a 4-mm centreline around the circumference of the valve.
Human comparison of bioprosthetic dysfunction disease
To provide a similar clinical comparator for our phantom imaging acquisitions, we performed a repeat analysis of a clinical case of transcatheter aortic valve replacement (TAVR) dysfunction from the RESOLVE study.22 The patient was imaged for 30 min commencing 1 h after administration of 3.4 mCi of 18F-NaF. This case was obtained as part of a separate research study, and the participants provided full informed consent in line with the study’s ethical approval.22
Results
PET image reconstruction and quantification
Reconstruction
Despite using low tracer activities, initial reconstructions using 5 min of list-mode data gave excellent PET signal with TBR values measured on the diastolic gates (gate 7 of 10) markedly higher than those of clinical bioprostheses 18F-NaF imaging (TBRmax 11.5 vs. 6.7). The duration of PET reconstruction was therefore reduced to 30 and then 15 s to match SNR seen in human imaging, and the same Gaussian post-filtering was applied. For the human TAVR bioprosthetic case, Gaussian filtering was already applied in the reconstruction and therefore was not repeated.
Quantification of PET signal
With a shortened PET reconstruction time of 30 s, we achieved quantifiable valvular PET signal similar to human TAVR imaging (Table 1 and Figure 3). This included SNR (45.6 vs. 44.3) and TBRmax (6.5 vs. 4.1) for the small valvular ROIs compared with the human case. Background noise within the cardiac chambers (10.2% vs. 12.0%) was also comparable. Other areas of interest within the phantom (large pocket and ring lesion) also showed high PET signal (Table 1 and Figure 3), markedly exceeding the comparable human case even with the very short (30 s) reconstruction time (Figure 4). We then reduced the duration of reconstructed PET data to 15 s. This yielded signal for both small (TBRmax 5.4) and large (TBRmax 6.0) valvular ROIs that was more comparable with human imaging (TBRmax 4.1), though SNR for both ROIs (small pocket 13.9, large pocket 15.6) was lower than in the clinical case of TAVR dysfunction (44.3).
Table 1.
Analysis of 30 s PET/CT acquisition of our cardiac phantom alongside values from a comparable human case of TAVR dysfunction also imaged using 18F-labelled radiotracer
| Cardiac chambers | Native valve remnant (30-s recon) | Native valve remnant (15-s recon) | Valvular lesions (30-s recon) | Valvular lesions (15-s recon) | Human TAVR lesion | |
|---|---|---|---|---|---|---|
| Volume | 750 mL | 0.5 mL | 0.5 mL | Large: 0.12 mL Small: 0.04 mL |
Large: 0.12 mL Small: 0.04 mL |
. |
| Iohexol 350 mg i/mL | 50 mL | . | . | . | . | . |
| 18F-tracer activity | 0.176 mCi | 4 µCi | 4 µCi | Large: 1 μCi Small: 1 μCi |
Large: 1 μCi Small: 1 μCi |
3.4 mCi |
| TBRmax | . | 21.9 | 19.9 | Large: 10.2 Small: 6.5 |
Large: 6.0 Small: 5.4 |
4.1 |
| SNR | . | 153 | 51.3 | Large: 71.1 Small: 45.6 |
Large: 15.6 Small: 13.9 |
44.3 |
| SNR (motion corrected) | . | . | 154.3 | Small: 135.8 | Large: 35.8 Small: 32.9 |
47.3 |
| Noise (background) | . | 10.2% | 8.8% | 10.2% | 8.8% | 12.0% |
Using these reconstructive parameters, comparable values of tracer expression were achieved, with good agreement in background activity and image noise. HUs within the phantom reached levels desirable for human imaging (300–350) using a 1:14 concentration of iohexol contrast agent relative to sterile 0.9% sodium chloride solution.
Figure 4.
Hybrid 18F-NaF PET/CT acquisition in a patient with known TAVR degeneration. Intense tracer signal is seen localized to the TAVR, in particular the position of the prior non-coronary cusp of the aortic valve. In this case, TBRmax was calculated at 4.1 (SUVmax value 5.7, mean left atrial blood pool 1.4±0.1) and SNR was 47.3 after application of motion correction.
Application of motion correction
Following motion correction of the PET data, we observed further improvements in SNR in the valvular small pocket relative to the non-motion corrected phantom acquisition (30 s reconstruction: 135.8 vs. 45.6, 15 s recon 32.9 vs. 13.9) and subjective image quality (Table 1 and Figure 5). In contrast for the human TAVR case, SNR was essentially unchanged following application of motion correction (47.3 vs. 44.3).
Figure 5.
Image quality before (left) and after application of valvular motion correction is displayed. From top to bottom, a short-axis view through the valves and vertical and horizontal axis views through the atrial and ventricular chambers are shown. Alongside subjective improvement in image quality, superior SNRs were achieved in the native valve remnant (green, larger arrows) and valvular lesions (blue, smaller arrows) with motion correction applied.
Discussion
To the best of our knowledge, this study is the first to test the utility of a realistic cardiac phantom for PET/CT imaging of bioprosthetic valve degeneration. Advanced cardiac phantoms provide an excellent opportunity to optimize the acquisition, reconstruction, and post-processing of 18F-NaF digital PET/CT assessments of small moving structures, as multiple repeated acquisitions with different tracer activity ratios can be performed.13,14 However, previous cardiovascular phantoms did not allow imaging of small moving structures, such as cardiac valves, with previous work focusing instead on imaging of the myocardium rather than the valves.15 This work proves the concept of a realistic cardiac phantom in optimizing PET image acquisition, reconstruction, and analysis in small, thin, highly mobile structures within the cardiovascular system. By employing a realistic phantom, we have established effective parameters for imaging valvular prostheses with simulated degeneration using PET/CT.
18F-NaF PET is an established technique that provides a non-invasive assessment of calcification activity within the bones and cardiovascular system.6,23,24 A major recent development is the demonstration of its ability to detect degeneration of aortic valve prostheses.3,4,11 Our findings are likely to be of significant utility given the rising number of bioprosthetic aortic valve replacements being performed worldwide, leading to a growing population in whom bioprosthetic valve failure is a possibility. Identifying those at increased risk for additional surveillance emerges as a critical clinical need, and 18F-NaF PET has the potential to address this need and become a vital tool in optimizing the management of this complex and growing patient group. Many other 18F-labelled PET tracers provide pathological insights in cardiovascular disease, such as thrombus formation in both the coronaries and cardiac valves (18F-GP1),25,26 and myocardial inflammation and viability assessments (18F-fluorodeoxyglucose).27,28 Our findings are, therefore, additionally likely to be useful in optimizing PET assessments in these settings.
Our use of very small valvular lesion sizes with minimum volumes of only 0.04 mL mimics the very small areas of valvular degeneration as can be seen in human disease in earlier stages, as well as larger lesions (0.12 mL) and the native valve remnant, making ours a highly novel and accurate model for aortic valve prosthetic dysfunction. Our parameters are likely to inform optimal techniques for clinical imaging of patients with degenerative aortic valve prostheses. In particular, we found that our phantom required very low levels of tracer activity with very short reconstruction lengths to achieve values comparable with patient imaging of the same condition. A natural next step would be imaging a large clinical cohort with degenerative aortic valvular prostheses using PET/CT with image acquisition and reconstruction parameters informed by our phantom experiments.
Limitations
While our phantom provides excellent simulation of cardiac motion caused by myocardial contraction and intracardiac blood flow, it does not simulate respiratory patient motion, another important factor that can impede accurate quantification of PET activity.13 We performed imaging on a single PET/CT scanner which may limit the generalizability of our findings to other makes and models. Future work should investigate different scanners for confirmation. Finally, although extreme care was taken to mimic as closely as possible human degenerative bioprosthetic valve disease and actual valve implants were used, there will always be differences between even the most sophisticated of phantoms and the unparalleled complexity of human pathology.
Conclusions
We utilized a realistic heart phantom to explore and successfully mimic clinical imaging conditions of aortic valve bioprosthetic dysfunction using hybrid 18F-NaF PET/CT.
Consent
The patient case in this study was obtained as part of a separate research study (RESOLVE study), and the participants provided full informed consent in line with the study’s ethical approval.
Contributor Information
Anna K Barton, Departments of Medicine (Division of Artificial Intelligence in Medicine) and Biomedical Sciences, Cedars-Sinai Medical Center, 6500 Wilshire Blvd, Los Angeles, CA 90048, USA; Centre for Cardiovascular Science, The University of Edinburgh, Edinburgh, UK.
Jacek Kwiecinski, Departments of Medicine (Division of Artificial Intelligence in Medicine) and Biomedical Sciences, Cedars-Sinai Medical Center, 6500 Wilshire Blvd, Los Angeles, CA 90048, USA; Division of Cardiology and Structural Heart Diseases, Medical University of Silesia, Katowice, Poland.
Hidenobu Hashimoto, Departments of Medicine (Division of Artificial Intelligence in Medicine) and Biomedical Sciences, Cedars-Sinai Medical Center, 6500 Wilshire Blvd, Los Angeles, CA 90048, USA.
Mark Hyun, Departments of Medicine (Division of Artificial Intelligence in Medicine) and Biomedical Sciences, Cedars-Sinai Medical Center, 6500 Wilshire Blvd, Los Angeles, CA 90048, USA.
Keiichiro Kuronuma, Departments of Medicine (Division of Artificial Intelligence in Medicine) and Biomedical Sciences, Cedars-Sinai Medical Center, 6500 Wilshire Blvd, Los Angeles, CA 90048, USA; Department of Cardiology, Nihon University, Tokyo, Japan.
Aditya Killekar, Departments of Medicine (Division of Artificial Intelligence in Medicine) and Biomedical Sciences, Cedars-Sinai Medical Center, 6500 Wilshire Blvd, Los Angeles, CA 90048, USA.
Aakriti Gupta, Departments of Medicine (Division of Artificial Intelligence in Medicine) and Biomedical Sciences, Cedars-Sinai Medical Center, 6500 Wilshire Blvd, Los Angeles, CA 90048, USA.
Nipun Manral, Departments of Medicine (Division of Artificial Intelligence in Medicine) and Biomedical Sciences, Cedars-Sinai Medical Center, 6500 Wilshire Blvd, Los Angeles, CA 90048, USA.
John Moore, Research and Development Division, Archetype BioMedical Inc., London, ON, Canada.
Marc R Dweck, Centre for Cardiovascular Science, The University of Edinburgh, Edinburgh, UK.
David E Newby, Centre for Cardiovascular Science, The University of Edinburgh, Edinburgh, UK.
Daniel S Berman, Departments of Medicine (Division of Artificial Intelligence in Medicine) and Biomedical Sciences, Cedars-Sinai Medical Center, 6500 Wilshire Blvd, Los Angeles, CA 90048, USA.
Damini Dey, Departments of Medicine (Division of Artificial Intelligence in Medicine) and Biomedical Sciences, Cedars-Sinai Medical Center, 6500 Wilshire Blvd, Los Angeles, CA 90048, USA.
Piotr Slomka, Departments of Medicine (Division of Artificial Intelligence in Medicine) and Biomedical Sciences, Cedars-Sinai Medical Center, 6500 Wilshire Blvd, Los Angeles, CA 90048, USA.
Funding
This research was supported in part by grants R35-HL161195 and R01-HL135557 from the National Heart, Lung, and Blood Institute/National Institutes of Health (NHLBI/NIH) (PI: P.S.). The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health.
Data availability
To the extent allowed by data sharing agreements and institutional review board protocols, the data from this manuscript will be shared upon written request.
Lead author biography
Dr Anna Barton is a specialty trainee in cardiology in NHS Lothian, Scotland. She received her primary medical degree and Master of Science from the University of Edinburgh, where she is currently a PhD student. Dr Barton’s clinical interests include heart failure, multi-modality cardiac imaging, and preventative cardiology. Her research work focuses on the role of novel PET tracers in cardiovascular disease.
References
- 1. Czernin J, Satyamurthy N, Schiepers C. Molecular mechanisms of bone 18F-NaF deposition. J Nucl Med 2010;51:1826–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Kwiecinski J, Tzolos E, Adamson PD, Cadet S, Moss AJ, Joshi N et al. Coronary (18)F-sodium fluoride uptake predicts outcomes in patients with coronary artery disease. J Am Coll Cardiol 2020;75:3061–74. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Cartlidge TRG, Doris MK, Sellers SL, Pawade TA, White AC, Pessotto R et al. Detection and prediction of bioprosthetic aortic valve degeneration. J Am Coll Cardiol 2019;73:1107–19. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Kwiecinski J, Tzolos E, Cartlidge TRG, Fletcher A, Doris MK, Bing R et al. Native aortic valve disease progression and bioprosthetic valve degeneration in patients with transcatheter aortic valve implantation. Circulation 2021;144:1396–408. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Joshi NV, Vesey AT, Williams MC, Shah ASV, Calvert PA, Craighead FHM et al. 18F-fluoride positron emission tomography for identification of ruptured and high-risk coronary atherosclerotic plaques: a prospective clinical trial. Lancet 2014;383:705–13. [DOI] [PubMed] [Google Scholar]
- 6. Dweck MR, Jones C, Joshi NV, Fletcher AM, Richardson H, White A et al. Assessment of valvular calcification and inflammation by positron emission tomography in patients with aortic stenosis. Circulation 2012;125:76–86. [DOI] [PubMed] [Google Scholar]
- 7. Tzolos E, Kwiecinski J, Berman D, Slomka P, Newby DE, Dweck MR. Latest advances in multimodality imaging of aortic stenosis. J Nucl Med 2022;63:353–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Dweck MR, Boon NA, Newby DE. Calcific aortic stenosis: a disease of the valve and the myocardium. J Am Coll Cardiol 2012;60:1854–63. [DOI] [PubMed] [Google Scholar]
- 9. Everett RJ, Clavel MA, Pibarot P, Dweck MR. Timing of intervention in aortic stenosis: a review of current and future strategies. Heart 2018;104:2067–76. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Yadgir S, Johnson CO, Aboyans V, Adebayo OM, Adedoyin RA, Afarideh M et al. Global, regional, and national burden of calcific aortic valve and degenerative mitral valve diseases, 1990–2017. Circulation 2020;141:1670–80. [DOI] [PubMed] [Google Scholar]
- 11. Sawaya F, Jorgensen TH, Sondergaard L, De Backer O. Transcatheter bioprosthetic aortic valve dysfunction: what we know so far. Front Cardiovasc Med 2019;6:145. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Pibarot P, Ternacle J, Jaber WA, Salaun E, Dahou A, Asch FM et al. Structural deterioration of transcatheter versus surgical aortic valve bioprostheses in the PARTNER-2 trial. J Am Coll Cardiol 2020;76:1830–43. [DOI] [PubMed] [Google Scholar]
- 13. Slomka PJ, Pan T, Germano G. Imaging moving heart structures with PET. J Nucl Cardiol 2016;23:486–90. [DOI] [PubMed] [Google Scholar]
- 14. Alessio A, Adams J, Lassen ML, Slomka P. Characterization of partial volume errors in coronary plaque PET imaging. J Nucl Med 2020;61:450. [Google Scholar]
- 15. Kamphuis ME, Kuipers H, Liefers HR, van Es J, Simonis FFJ, Greuter MJW et al. A multimodality myocardial perfusion phantom: initial quantitative imaging results. Bioengineering (Basel) 2022;9:436. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. O'Doherty J, Sammut E, Schleyer P, Stirling J, Nazir MS, Marsden PK et al. Feasibility of simultaneous PET-MR perfusion using a novel cardiac perfusion phantom. Eur J Hybrid Imaging 2017;1:4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Healthineers S. SAFIRE Erlangen, Germany 2024. Available from: https://www.siemens-healthineers.com/en-uk/molecular-imaging/options-and-upgrades/software-applications/safire (1 October 2024, date last accessed).
- 18. Massera D, Doris MK, Cadet S, Kwiecinski J, Pawade TA, Peeters FECM et al. Analytical quantification of aortic valve 18F-sodium fluoride PET uptake. J Nucl Cardiol 2020;27:962–72. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Rubeaux M, Joshi NV, Dweck MR, Fletcher A, Motwani M, Thomson LE et al. Motion correction of 18F-NaF PET for imaging coronary atherosclerotic plaques. J Nucl Med 2016;57:54–9. [DOI] [PubMed] [Google Scholar]
- 20. Lassen ML, Tzolos E, Massera D, Cadet S, Bing R, Kwiecinski J et al. Aortic valve imaging using (18)F-sodium fluoride: impact of triple motion correction. EJNMMI Phys 2022;9:4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Doris MK, Rubeaux M, Pawade T, Otaki Y, Xie Y, Li D et al. Motion-Corrected imaging of the aortic valve with (18)F-NaF PET/CT and PET/MRI: a feasibility study. J Nucl Med 2017;58:1811–4. [DOI] [PubMed] [Google Scholar]
- 22. Chakravarty T, Søndergaard L, Friedman J, Backer D, Berman O, Kofoed D et al. Subclinical leaflet thrombosis in surgical and transcatheter bioprosthetic aortic valves: an observational study. Lancet 2017;389:2383–92. [DOI] [PubMed] [Google Scholar]
- 23. Dweck MR, Chow MW, Joshi NV, Williams MC, Jones C, Fletcher AM et al. Coronary arterial 18F-sodium fluoride uptake: a novel marker of plaque biology. J Am Coll Cardiol 2012;59:1539–48. [DOI] [PubMed] [Google Scholar]
- 24. Even-Sapir E, Metser U, Mishani E, Lievshitz G, Lerman H, Leibovitch I. The detection of bone metastases in patients with high-risk prostate cancer: 99mTc-MDP planar bone scintigraphy, single- and multi-field-of-view SPECT, 18F-fluoride PET, and 18F-fluoride PET/CT. J Nucl Med 2006;47:287–97. [PubMed] [Google Scholar]
- 25. Bing R, Deutsch MA, Sellers SL, Corral CA, Andrews JPM, van Beek EJR et al. 18F-GP1 positron emission tomography and bioprosthetic aortic valve thrombus. JACC Cardiovasc Imaging 2022;15:1107–20. [DOI] [PubMed] [Google Scholar]
- 26. Tzolos E, Bing R, Andrews J, MacAskill MG, Tavares AAS, Macnaught G et al. Noninvasive in vivo coronary artery thrombus imaging. JACC Cardiovasc Imaging 2023;16:820–32. [DOI] [PubMed] [Google Scholar]
- 27. Miller RJH, Cadet S, Pournazari P, Pope A, Kransdorf E, Hamilton MA et al. Quantitative assessment of cardiac hypermetabolism and perfusion for diagnosis of cardiac sarcoidosis. J Nucl Cardiol 2022;29:86–96. [DOI] [PubMed] [Google Scholar]
- 28. Khalaf S, Al-Mallah MH. Fluorodeoxyglucose applications in cardiac PET; viability, inflammation, infection, and beyond. Methodist DeBakey Cardiovasular Journal 2020;16:122–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
To the extent allowed by data sharing agreements and institutional review board protocols, the data from this manuscript will be shared upon written request.