Non-invasive Deep-Brain Imaging with 3D Integrated Photoacoustic Tomography and Ultrasound Localization Microscopy (3D-PAULM)

Yuqi Tang 1, Nanchao Wang 2, Zhijie Dong 3, Matthew Lowerison 4, Angela del Aguila 5, Natalie Johnston 6, Tri Vu 7, Chenshuo Ma 8, Yirui Xu 9, Wei Yang 10, Pengfei Song 11, Junjie Yao 12
PMCID: PMC11892115  NIHMSID: NIHMS2054422  PMID: 39383084

Abstract

Photoacoustic computed tomography (PACT) is a proven technology for imaging hemodynamics in deep brain of small animal models. PACT is inherently compatible with ultrasound (US) imaging, providing complementary contrast mechanisms. While PACT can quantify the brain’s oxygen saturation of hemoglobin (sO2), US imaging can probe the blood flow based on the Doppler effect. Further, by tracking gas-filled microbubbles, ultrasound localization microscopy (ULM) can map the blood flow velocity with sub-diffraction spatial resolution. In this work, we present a 3D deep-brain imaging system that seamlessly integrates PACT and ULM into a single device, 3D-PAULM. Using a low ultrasound frequency of 4 MHz, 3D-PAULM is capable of imaging the brain hemodynamic functions with intact scalp and skull in a totally non-invasive manner. Using 3D-PAULM, we studied the mouse brain functions with ischemic stroke. Multi-spectral PACT, US B-mode imaging, microbubble-enhanced power Doppler (PD), and ULM were performed on the same mouse brain with intrinsic image co-registration. From the multi-modality measurements, we further quantified blood perfusion, sO2, vessel density, and flow velocity of the mouse brain, showing stroke-induced ischemia, hypoxia, and reduced blood flow. We expect that 3D-PAULM can find broad applications in studying deep brain functions on small animal models.

Keywords: Photoacoustic imaging, stroke, ultrasound localization microscopy, functional imaging

I. INTRODUCTION

Photoacoustic computed tomography (PACT) is a hybrid imaging modality that combines the rich contrast of optical absorption and deep penetration of ultrasound detection [1], [2], [3]. Hemoglobin, as the most commonly used endogenous absorber, offers excellent PA contrast for visualizing blood vessels and their functions in the brain (Fig. 1a). A key application of PACT is mapping oxygen saturation of hemoglobin (sO2), by differentiating the oxygenated and deoxygenated hemoglobin using multi-spectral analysis of PA signals. However, without using exogenous contrast, PACT has difficulty in detecting blood flow, due to the lack of speckle contrast. Moreover, PACT with a partial detection aperture often suffers from the so-called ‘limited view problem’, in which coherent PA signal generation results in missing structures [4], [5]. These shortcomings of PACT can be mitigated by incorporating ultrasound (US) imaging (Fig. 1a). US imaging provides information about tissue anatomy and is sensitive to the blood flow based on either the Doppler effect or the speckle fluctuations induced by the flowing red blood cells. Detecting acoustic signals at similar frequencies, PACT and US imaging are highly compatible, and can be integrated into a single system to offer more comprehensive hemodynamic information than a single modality alone [6], [7], [8], [9].

Fig. 1. 3D integrated photoacoustic tomography and ultrasound localization microscopy.

Fig. 1.

(a) Radar chart illustrating the imaging capabilities of photoacoustic (PA) and ultrasound (US) imaging. (b) Schematic of the imaging system setup. A 2D hemispherical transducer array is immersed in water, with a central aperture for optical fiber insertion. The mouse is placed in a supine position, and the region of interest is positioned around the focal zone of the transducer array (approximately 40 mm from the transducer surface). The system is capable of multi-spectral PA imaging, US B-mode imaging, US power Doppler imaging (PD), and ultrasound localization microscopy (ULM).

Nevertheless, traditional PA and US imaging often use relatively low ultrasound frequency (<10 MHz) to accommodate the large penetration depth (> 1 cm), which results in limited spatial resolution for visualizing deep-tissue hemodynamics. Such limitation is particularly true for non-invasive deep-brain imaging on small animal models, in which the intact skull has strong acoustic attenuation at high frequencies and induces severe acoustic aberration. Recently, super-resolution ultrasound localization microscopy (ULM) has overcome the resolution limit [10], [11], [12]. Using gas-filled microbubbles as strong echoic tracers, ULM can track the positions of the microbubbles over time, and the super-resolved microvessel images can be reconstructed in the deep brain [13], [14]. 3D ULM can provide detailed microvascular structure and hemodynamic information (e.g., blood flow) in deep tissues[15], [16]. Thus, further incorporating ULM into the PA and US imaging system can significantly enhance microvasculature imaging capabilities. Robin et al. combined PA and US imaging with a spherical transducer array, and demonstrated contrast-enhanced power Doppler (PD) imaging at a volumetric rate of 100 Hz, which, however, is too slow for ULM [17]. Zhao et al. implemented 3D ULM with a multimodal PA and US imaging system based on raster scanning of a 1D linear transducer array (15 MHz central frequency), which would have difficulty imaging through the mouse brain with the skull intact [18]. To date, there has been no demonstration of integrated PACT and ULM with a high volume rate, high spatial resolution, and large imaging depth.

In this study, we have developed an integrated imaging system that seamlessly combines 3D PACT and ULM (3D-PAULM). 3D-PAULM works at a relatively low frequency of 4 MHz, which is beneficial for deep brain imaging with both the scalp and skull intact. Particularly, we employ a low-frequency spherical ultrasound transducer array, in which the normal of each individual element is pointing toward the spherical center of the array to efficiently collect signals from a wide compounding angle range. Such a focusing geometry not only improves the spatial resolution for PA and US imaging, but also reduces the limited-view issue in the PA imaging [5]. The relatively large element size of the spherical array leads to a low electrical impedance, further improving the signal-to-noise ratio (SNR) for transcranial deep-brain imaging. Besides multi-spectral PACT and ULM, 3D-PAULM can also perform US B-mode imaging and PD imaging, with all the modalities automatically co-registered. We characterized the key performance of 3D-PAULM by numerical simulations and phantom experiments. Then, using an ischemic stroke mouse model, we demonstrated 3D-PAULM for comprehensive functional imaging of the brain, including blood perfusion, microvasculature density, sO2, and blood flow velocity. Our results have highlighted the potential of 3D-PAULM as a non-invasive imaging tool for studying deep brain functions.

II. METHODS

A. The Overall 3D-PAULM System

As shown in Fig. 1b, the 3D-PAULM system comprises a programable ultrasound scanner (Vantage 256, Verasonics, Kirkland, WA), multiple pulsed lasers for PA excitation, and a customized 2D spherical ultrasound transducer array (Imasonics, France) for both PA and US imaging. The spherical transducer array has 256 elements, with a central opening for inserting an optical fiber bundle. The transducer array has a center frequency of 4 MHz, and -6 dB bandwidth of 45% in the transmit/receive mode, although it was fine tuned to maximize the receiving bandwidth. The transducer array has a curvature radius of 40 mm and total aperture size of 57 mm. The well-sampled field of view (FOV) of the transducer (or the focal zone) is 8×8×8 mm3. During the imaging, the transducer array was facing up inside a temperature-regulated water bath for acoustic coupling. The imaged animal was mounted on an optically and acoustically transparent sample holder at the transducer array’s focal zone. The sample holder can be raster scanned by a 2D motorized translation stage for an expanded FOV. The 3D-PAULM system can perform multi-spectral PACT, US B-mode imaging, PD imaging and ULM. The system setup, signal acquisition, and image reconstruction for each imaging modality are detailed below.

B. 3D US Pressure Field Simulation

Due to the large pitch size of the transducer array (3.57 mm on average), synthetic-aperture US imaging was employed to minimize grating lobes [19], instead of the traditional plane-wave compounding imaging often used by the linear array or matrix array with dense elements. By using the synthetic-aperture method, we encountered a trade-off between the number of individual transmitting elements (Tx elements) and the maximum volumetric frame rate. To investigate the transmitted beam pattern, we maximized the distribution of the Tx elements to cover the entire array aperture while limiting the maximum adjacent distance to suppress the grating-lobe artifacts. Moreover, we conducted a 3D k-Wave simulation to compare the transmission patterns with different maximum adjacent distances [20]. Based on our simulations, we did not observe significant improvements with more than 31 Tx elements. For the case of 31-Tx-element, the maximum distance between neighboring Tx elements was 7.5 mm, as shown in Fig. 2a. The pixel size in the simulation was 150 μm in x-, y-, and z-directions, with a simulation volume of 62×62×45 mm3. The element geometry in the simulation was consistent with that of the experimental transducer array, i.e., trapezoidal shape with a 40-mm curvature. Forward US transmission simulations were performed, and the maximum pressure field and root-mean-square (RMS) pressure were recorded for each transmission pattern. We recorded the transmitted pressure from a single element (Fig. 2b) and compared the accumulated pressure field from the 31-element and 256-element synthetic aperture transmission (Fig. 2c).

Fig. 2. System characterization and imaging sequence.

Fig. 2.

(a) Locations of the transducer elements projected in the lateral plane. Hollow orange elements represent the receiving aperture (256 elements), while solid green elements represent the 31 transmitting elements. The inset shows the dimension of a single element. (b) Simulated pressure field transmitted from a single element at the transducer focus. The pressure profiles along different directions are extracted. (c) Simulated pressure field with two different transmission patterns: a 256-element synthetic transmission aperture (top row) and a 31-element synthetic transmission aperture (bottom row). Pressure is displayed in dB. (d-e) Data acquisition sequence of PA and US imaging, with a respective volumetric frame rate of 10 Hz and 215 Hz.

C. System Characterization

The receiving electrical impulse response of the system was first measured by imaging a thin layer (~80 μm thick) of gold nanoparticles at 1064 nm. The gold nanoparticles had a diameter of ~80 nm and were highly concentrated, generating a Dirac PA signal. The signal detected by a single element at the center of the transducer array was sampled at 62.5 MHz and averaged over 100 times for noise reduction. Additionally, to characterize the system resolution, a crossed hair target was imaged using both PACT and US B-mode imaging. The hair phantom was placed at the transducer’s focal zone, approximately 40 mm from the transducer’s surface.

D. Stroke Animal Preparation

A male C57B1/6 mouse was used to demonstrate the 3D-PALUM’s in vivo imaging performance. The animal was 4 months old and had undergone permanent middle cerebral artery occlusion (pMCAO) surgery 72 hours prior to imaging, following the same surgical protocol as described in our previous work [21]. The head was shaved, and the body temperature was maintained at 37°C throughout the experiment. A femoral vein catheter was inserted right before the imaging. The imaging session lasted approximately 10 minutes, including both PA and US data acquisition. Microbubbles were injected via the catheter only for US imaging. Microbubbles were continuously infused using a syringe pump with a concentration of 7×108 microbubbles/mL and the injection rate was 10 μL/min. The total volume of microbubbles injected was less than 100 μL. All the animal experimental procedures were approved by Duke University Institutional Animal Care & Use Committee.

E. Data Acquisition

The PA and US data were acquired sequentially for the mouse brain. Due to the relatively small focal zone of the 2D transducer array, raster scanning of the mouse by the 2D motorized translation stage was performed to cover the whole mouse brain. A total of 6 scanning positions were used (step size: 3 mm), which results in an extended FOV of 14×11×8 mm3. In PA imaging, three excitation wavelengths at 700nm, 750 nm and 1064 nm were synchronized, and the laser pulse emissions were separated by 200 μs to form a single multi-wavelength PA frame with a volumetric frame rate of 10 Hz (Fig. 2d). A total of 10 PA frames were acquired at each scanning location, with a total acquisition time of 1 second per location.

The US data was acquired after the PA imaging was completed. The synthetic aperture imaging with 31-Tx-element transmission was applied, and the back-scattered echo signals from each transmission event were received with the full aperture, i.e., 256 elements (Fig. 2e). A total of 4800 3D US frames were acquired at each scanning position, with a total acquisition time of <3 minutes. To minimize reverberation artifacts, each US transmission event was separated by 150 μs. The received RF data was batch-transferred to ensure a US volumetric frame rate of 215 Hz.

F. 3D Multi-spectral PACT

PACT often uses near-infrared light for deep penetration. We first used 1064 nm light (DRL 100, Qutantel Laser) due to its ability to penetrate the entire brain. Two more wavelengths of 700 nm and 750 nm (Surelite OPO Plus, Amplitude Laser) were selected for PA imaging of sO2. The OPO laser performed pulse-by-pulse wavelength switching between 700 nm and 750 nm. Both lasers have an 8 ns pulse-width and a pulse repetition rate of 10 Hz. A small fraction of the pulse energy was sampled by a beam splitter and a power meter to record the pulse-by-pulse energy variation and calibrate the optical fluence for sO2 calculation. The OPO laser and the 1064-nm laser were synchronized with a 1-ms time interval. The light was delivered using the same optical fiber bundle (Dolan Jenner, 50% coupling efficiency) inserted into the central opening of the 2D spherical array transducer, illuminating the focal zone of the transducer array. The optical fluence on the sample surface at 700 nm, 750 nm, and 1064 nm was 13 mJ/cm2, 15 mJ/cm2, and 75 mJ/cm2, respectively, all of which were within the ANSI limit [22].

At each scanning position, RF data were acquired for 10 laser pulses at each wavelength with a sampling frequency of 20.83 MHz. The RF data from the 10 laser pulses were averaged to improve signal-to-noise ratio (SNR). A 3D delay-and-sum (DAS) method was used for PA image reconstruction, with a voxel size of 60 μm in all dimensions. The reconstructed images for all scanning positions were stitched to form the final 3D brain image.

PA images at 700 nm and 750 nm were used to estimate sO2. To account for frequency-dependent acoustic attenuation in deeper regions and improve the quantification accuracy, the RF data were deconvolved with the electrical impulse response of the transducer array and low-pass filtered at a cutoff frequency of 1 MHz. The relative concentrations of oxygenated and deoxygenated hemoglobin were calculated using the linear unmixing method [23], [24]. We did not perform wavelength-dependent optical fluence compensation for the sO2 calculation. Instead, we chose two relatively close optical wavelengths at 700 nm and 750 nm, at which the OPO laser has similar and stable output power. Also, the optical properties of the brain tissue at 700 nm and 750 nm are similar, allowing for a comparable fluence distribution inside the tissue and thus more accurate calculation of sO2 without fluence compensation.

G. 3D PD Imaging and ULM

The 3D US B-mode imaging was performed by using the synthetic aperture method, as shown in Fig. 2a. The transmitted waveform from each element had a central frequency of 4 MHz, 1 complete burst cycle, and a 10 V peak-to-peak driving voltage. In-phase and quadratic-phase (IQ) volumes were reconstructed using 3D DAS with the same voxel size as PACT (i.e., 60 μm in all dimensions). To mitigate the motion artifacts due to animal breathing, 2D cross-correlation was computed on the lateral maximum intensity projection (MIP) images, using the averaged MIP frame as the reference frame. A correlation threshold of 99% was applied to remove the moving frames, which corresponded to ~10% of total frames. From the US B-mode images, the PD and ULM processing workflow consists of four major steps to reconstruct the brain vasculature images, as shown in Fig. 3 [25].

Fig. 3. ULM processing workflows.

Fig. 3.

(a) US B-mode images are reconstructed by 3D DAS. (b) Spatial-temporal singular value decomposition (SVD) filtering is applied to remove the clutter signals. (c) Power Doppler image is formed by accumulating the filtered data. (d) Radial symmetry-based localization is performed to extract the positions of microbubble signals (green dots). (e) Trajectories retrieved using the 3D tracking algorithm are accumulated to form the ULM density map and flow speed map. (f) Illustration of MB localization and tracking.

Step 1. Spatial-temporal singular value decomposition (SVD) was applied within each block of 600 B-mode volumes to remove clutter signals and extract microbubble signals. The first 150 largest singular values were removed for each block, as demonstrated in Figs. 3ab [26].

Step 2. Power Doppler (PD) volumetric images were formed by integrating the SVD-filtered IQ data (Fig. 3c). To facilitate microbubble separation, directional filtering and bandpass filtering (20–107 Hz) were used [27].

Step 3. Following the filtering, a radial symmetry-based microbubble localization algorithm was applied to extract the sub-diffraction locations of the microbubbles (Fig. 3d) [15].

Step 4. To track the motion of microbubbles across consecutive volumes, a 3D particle tracking algorithm was applied [28]. We assigned a maximum paring/linking search radius of 10 voxels in our tracking algorithm, which corresponds to a maximum bubble velocity of ~130 mm/s. The maximum gap closing time window was set to 1 which disabled merging the discontinuous track in essence. For frame-to-frame linking we used the linear motion Kalman filter model with default initial state and applied a maximum linking angle of 45 degrees. The microbubble trajectories were then accumulated to generate the ULM density map, as shown in Fig. 3e. Additionally, a microbubble flow velocity map was calculated based on the trajectories, providing information on blood flows.

An example of localizing and tracking microbubbles in a water tank is shown in Fig. 3f. Similar to PACT, the reconstructed PD and ULM images at all scanning positions were stitched to the final 3D images.

H. Real-time Image Reconstruction and Display

We utilized a real-time image reconstruction method for multi-spectral PACT that allowed for the immediate display of PA images at all three wavelengths. In DAS, the delay calculation was performed between each element of the array and each pixel in the region of interest (ROI). These delays were then discretized to match the RF signal length, and the corresponding pixel values were added across the transducer array to reconstruct an average initial pressure distribution. To further speed up the DAS process, we employed a GPU-enhanced sparse-matrix-multiplication based method [29].

I. Vessel Segmentation

Unlike previous ULM techniques that often utilize tens of thousands of frames for microbubble localization, our ULM approach uses only 4800 frames at one scanning position, which may not provide continuous vessels in the final trajectory images [10]. Since vessel continuity is crucial for accurate quantitative analysis of the hemodynamics, we chose to use the 3D PD images for vessel segmentation, which have lower resolution but higher vessel continuity.

The vessel segmentation was performed using the TubeTK library in Python. The 3D PD image of the mouse brain vasculature was first divided into the left and right hemispheres to compare the stroke and healthy regions. To further improve the vessel continuity and connect potentially disconnected vessels, the PD image was repeatedly blurred by 3D Gaussian filters and sharpened by erosion. Then an empirical thresholding was applied to reduce background noise. The preprocessed PD images were subjected to a difference of gaussian blurring to highlight the vessel ridges. A Danielsson distance filter was applied to create a mask for the radii of the vessels. The filtered images were fed to the TubeTK ridge extraction algorithm. About 500 seeds were placed to extract initial vessel-like structures for subsequent discriminant analysis, which defined the image space according to vessel-like properties of curvature, roundness, and ridgeness [30]. The vessel ridges were identified if they have high values of curvature, roundness, and ridgeness. The extracted ridges were saved as a directed graph, which contained the coordinates and radii of the segmented vessels. The vessel length was calculated by integrating the Euclidean distance between all points of each vessel, and the vessel volume was calculated using the length and the radius of the vessels. The final extracted vessel parameters included vessel radius, vessel length, vessel volume, and vessel spatial coordinates. Additional parameters included total vessel length and total vessel number. The segmented individual vessels were also used as spatial masks to extract the functional information from the PA and ULM images.

III. RESULTS

A. System Characterization

From the numerical simulations, we found that the spatial pressure fields in the US B-mode imaging were almost identical between the two transmission patterns (31 Tx elements vs. 256 Tx elements), so we used the 31-Tx-element transmission pattern for future experiments to achieve a higher volumetric rate for ULM. The receiving electrical impulse response of the transducer was shown in Fig. 4a. The transducer has a central frequency of 4 MHz and −6 dB receiving-only bandwidth of 75%, with slightly better sensitivity towards lower frequencies. It was assumed that all 256 elements had identical impulse responses, and the measured impulse response was used for RF data deconvolution prior to PA image reconstruction. Using the extracted 3D intensity profiles for both PA and US images of the crossed hair phantom, the measured lateral and axial resolution were 260 μm and 390 μm for PA, and 270 μm and 440 μm for US, respectively (Fig. 4b). US imaging has slightly worse resolutions than PA imaging. One possible reason is that the transducer was optimized to maximize the receiving sensitivity and bandwidth, which benefits PA imaging more. Another possible reason is the element position information limited by the manufacturing precision, which impacts the US imaging more. With relatively homogeneous light illumination, the PA imaging shares the same FOV as the US imaging. Consistent with previous literatures using similar transducer arrays [17], [31], [32], the effective field of view of the system was approximately 8×8×8 mm3 for both PA and US imaging, mainly determined by the focusing geometry of the spherical transducer array.(Fig. 3c).

Fig. 4. System characterization.

Fig. 4.

(a) The system’s receiving electrical impulse response and bandwidth, extracted from PA signals of a nanoparticle sheet. (b) PA and US imaging resolutions measured with a crossed hair phantom placed at the transducer focus. (c) The in-vivo single-MB images as the reference point source function (PSF) at the center and edge of the imaging field-of-view (FOV). (d) The estimated resolution based on the intersection of Fourier shell curve (FSC, solid blue line) and its corresponding 1/2 threshold (solid green line).

Furthermore, the in-vivo transcranial images of individual MBs were employed to evaluate local point spread function (PSF) within the FOV. The spatial variance of MB PSFs in different regions of the FOV could impact MB localization (Fig. 4c). This is a common issue for ULM imaging and is an active area of research. Raster scanning with overlapping to reconstruct the mouse brain vasculature can partially mitigate this issue. Future work may incorporate data driven strategies to localize MBs more efficiently throughout the imaging FOV, such as multi-template matching or deep-learning-based techniques for MB localization [33]. We also estimated 3D ULM resolution via Fourier shell correlation (FSC) which utilizes the independent reconstructions based on splitting microbubble tracking data to generate FSC curve [34]. These independent reconstructions were then used to estimate the FSC curve by comparing the spectrum of one reconstruction relative to another along iso-spatial frequency shell. In Fig. 4d, the resolution was estimated by finding the intersection of the FSC curve (solid blue line) with a 1/2 threshold (solid green line), yielding a resolution estimation of 48 μm.

B. 3D PA Imaging of Stroke Mouse

After the imaging session of the stroke mouse, we performed histology validation using 2,3,5-triphenyltetrazolium chloride (TTC) staining, which confirmed that the pMCAO procedure primarily affected the cortex region of the right hemisphere in the mouse (Fig. 5a). The impact on the hippocampus region was not as evident as in the cortex region.

Fig. 5. Multi-spectral PACT of mouse brain with ischemic stroke.

Fig. 5.

(a) TTC stained 1-mm-thick coronal slices of mouse brain (white area indicates stroke-induced infarct). (b) Maximum amplitude projection (MAP) images from the transverse, coronal, and sagittal planes of the 3D PA images at 700 nm, 750 nm, 1064 nm. (c) sO2 images of the coronal brain slices with a thickness of 1.5 mm. The white arrows mark the region surrounding the stroke area in the right hemisphere with relatively high sO2. (d) Representative frequency spectra of original RF signal, deconvolved signal, and low-pass filtered signal from PA measurement at 750 nm.

The PA images of the mouse brain at 1064 nm, 750 nm, and 700 nm are shown in Fig. 5b. No clear signal reduction in the stroke region is visible in the PA image. This may be explained by a compensation of the decrease in blood perfusion, and some remaining signals coming from the static blood clots, which may not be detected by the PD imaging or ULM that relies on the flowing blood (shown below).

To demonstrate the functional imaging capability of the imaging system, sO2 maps were extracted from coronal slices with a thickness of 1.5 mm at various positions (Fig. 5c). We would like to point out that high-frequency PA signals were more attenuated by tissues and more sensitive to the tissue’s optical and acoustic heterogeneity, and thus in this work low-frequency signals (<1 MHz) were used in the sO2 quantification. We have previously demonstrated that, in PACT, sO2 quantification can be improved by using low-frequency signals, which are less sensitivity to deteriorating factors such as speed of sound heterogeneity, motion artifacts, and under-sampling artifacts [35]. However, employing low-frequency signals results in degraded spatial resolution. Therefore, we utilized the PD image as the spatial mask for better sO2 visualization, which, however, does not improve the spatial resolution of the sO2 quantification. Moreover, we have observed that the region surrounding the stroke area in the right hemisphere had relatively high sO2, which was not observed in the left hemisphere, as indicated by the white arrows in Fig. 5c. It is possibly due to the overcompensation by the collateral blood flow from the left hemisphere [36].Figure 5d also shows the representative frequency spectra of original RF signal, deconvolved signal, and low-pass filtered signal from PA measurements at 750 nm.

The sO2 maps were consistent with the stroke-induced tissue infarct in the TTC staining results, as reflected by the lower sO2 levels in the cortex and hippocampus regions of the right hemisphere. The pMCAO surgery significantly reduced blood flow in the cortex regions and some of the deeper regions. In addition, the cortical region of the left hemisphere around Bregma −1.5 mm was also slightly affected, which was not clear on the TTC results but clearer on the PD and ULM results shown later. We also observed that the sO2 in the cortex region at Bregma −1.5 mm was higher than those at Bregma +1.5 mm. This suggests that the initially impacted region by the pMCAO surgery may be closer to the Bregma +1.5 mm.

C. 3D PD imaging and ULM of Stroke Mouse

While the US B-mode images mainly show the mouse skull due to its strong acoustic reflection (Fig. 6a), the PD and ULM images clearly show a reduction in vessel density at the right hemisphere from both the transverse and coronal projections (Figs. 6bc). Both the PD and ULM images show a significant decrease in blood perfusion in the stroke region compared to the healthy region, from both the vessel density and velocity maps. It is worth noting that the PD and ULM images of blood vessels are based on the flow contrast, which requires the actual motion of the blood and the microbubbles. This contrast mechanism is different from the PA imaging of blood vessels that does not require the flow. Thus, the PD and ULM images may not detect the static blood clots formed in the stroke region. On the other hand, PD and ULM images clearly show the diving vessels in the mouse cortex, including the penetrating pia arteries and veins. These diving vessels are not visible in the PA images, mainly due to the aforementioned limited-view problem. Moreover, the effective penetration depth in PA imaging mainly depends on the optical attenuation by the tissue, which includes optical absorption and optical scattering. The effective optical attenuation coefficient of soft tissue at 1064 nm (~2.5 cm−1) is approximately twice as large as that at 700/750 nm (~1.2 cm−1), mostly due to the elevated optical absorption of water at the longer wavelengths. Since we can provide stronger optical energy at 1064 nm than that at 700/750 nm, the eventual imaging depths at all three wavelengths were comparable. Our results are consistent with other transcranial imaging using similar spherical arrays [17], [37]. Moreover, by tracking the relative motion of microbubbles, ULM can also report the blood flow velocities in absolute units (Fig. 6d), which is not available in either the PA or PD imaging.

Fig. 6. 3D US imaging of the mouse brain.

Fig. 6.

Coronal, transverse, and sagittal view of the US B-mode image (a), US PD image (b), ULM image (c), and the velocity map (d). (e) and (g) show close-up images of two distinct regions of the PD and ULM density map, which share the same color bars as (b) and (c). (f) and (h) depict representative vessel profiles by ULM and PD, respectively, illustrating the different spatial resolution.

Compared to the PD images, the 3D ULM images reveal more detailed vessel structures with an approximately 4±6-fold higher spatial resolution. For example, the spatial resolutions measured on two representative diving vessels were 269.1 μm and 444.9 μm in the PD image, and 64.3 μm and 68.1 μm in the ULM image, respectively (Figs. 6eg). Nevertheless, the PD images have better vessel continuity than the ULM images, mainly because the ULM images need to accumulate a large number of microbubble trajectories.

D. Muti-modal Imaging Fusion and Quantitative Analysis

Taking advantage of the complementary contrast and inherent co-registration of the PA, US B-mode, PD, and ULM images, we have explored the multi-modal imaging fusion, which can provide more comprehensive understanding of the brain functions than each single imaging alone.

Acute stroke results in irreversible damage to the ischemic core, while the surrounding tissue known as the penumbra can potentially be salvaged. Without reperfusion, the tissue in the penumbra region undergoes necrosis, leading to an expansion of the infarct [38]. In TTC results, the white areas correspond to the ischemic core, whereas the pink rim surrounding it represents the penumbra region. Existing literature demonstrates that the ischemic region affected by pMCAO undergoes constant changes following the surgery. Over a period of 24 to 72 hours, the ischemic volume significantly increases by approximately 40%, indicating the progression of the stroke [39]. Therefore, regions displaying a lack of blood perfusion on PD and ULM images, while still exhibiting relatively high sO2, are possibly penumbra regions. In these tissues, tissue viability is sustained, and there hasnt’ been sufficient time for the clots to become fully deoxygenated.

Moreover, using the vessel segmentation result from the PD image as the spatial mask, we can quantify various morphological and functional parameters on the same vessels from the ULM images (Table I). For example, the total number of vessels in the stroke (right) and healthy (left) hemispheres are 162 and 231 (Table I), respectively, showing a 30% stroke-induced reduction in the vessel density. We also analyzed vessel radius, vessel length, sO2, and flow speed of the two hemispheres, as shown in Fig. 7. We fitted the histograms with a kernel density estimation to estimate the probability density function of vessel distribution in the healthy and stroke hemisphere. Notably the relative sO2 statistical analysis was generated from the unmasked PACT results and reflected the regional impact of stroke. As expected, all of the quantified parameters showed significant differences in the stroke region. We performed two-sample Z-test between the two hemispheres and we observed that 1) the stroke hemisphere had a significant increase in mean vessel length and a reduction in vessel radius compared to the healthy hemisphere; 2) The vessels that were still perfused in the stroke hemisphere exhibited an ~8.1% decrease in sO2 compared to those in the healthy hemisphere; and 3) The vessels in the stroke hemisphere had a median flow speed less than 4.6 mm/s, while the median speed for the healthy hemisphere was 5.2 mm/s.

Table I.

Statistical analysis of the blood vessels extracted from stroke and healthy hemispheres.

Healthy
Parameter Meanhealthy stdhealthy Meadian
Radius (mm) 0.099 0.029 0.092
Length (mm) 3.533 2.753 2.720
Avg. speed 5.198 1.266 5.149
Stroke
Parameter Meanstroke stdstroke Meadian
Radius (mm) 0.090 0.026 0.081
Length (mm) 4.867 4.044 3.660
Avg. speed 4.577 1.296 4.571
Two-sample Z-test
Parameter Meanhealthy - Meanstroke Z-test statistic p-value
Radius (mm) 0.09 3.12 0.00181
Length (mm) −1.335 −3.69 0.000228
Avg. speed 0.621 4.43 9.41E-06

Fig. 7. Quantitative analysis of the brain hemodynamics.

Fig. 7.

Histogram and fitting of vessel radius, vessel length, sO2, and averaged flow speed of segmented vessels in the left (healthy) and right (stroke) hemispheres. Note that the sO2 analysis was based on the low-pass filtered PA images that reflect the regional volume-averaged sO2 values.

IV. Discussion

We have developed an integrated multi-modality imaging system, 3D-PAULM, that is inherently capable of both multi-spectral PA imaging and high-speed US imaging with complementary contrasts. The system utilizes a 2D spherical transducer array at a relatively low frequency of 4 MHz and can provide brain imaging with intact skin and skull. The system can achieve synthetic-aperture-based US imaging with a volumetric frame rate of 215 Hz and is capable of 3D PD imaging and ULM. To overcome the small focal zone of the transducer array, we stitched multiple locations for both PA and US imaging, and the entire imaging process with the extended field of view took ~10 minutes. We used the segmented vessels from the PD image as a mask to extract individual vessel information, including vasculature structure, sO2, and blood flow speed.

One notable feature of our system is that it has achieved 3D ULM on the mouse brain with the skin and skull intact, which has always been a technical challenge for previous ULM methods. The key difference is that our imaging system uses a low-frequency transducer as opposed to the high-frequency ones typically employed. High-frequency ultrasound waves provide detailed images but are quickly attenuated by skull and tissue, thus limiting their depth of penetration. To compensate for potential resolution loss associated with lower frequencies, we utilized bandpass filtering and directional filtering for better microbubble localization [27]. This makes our system suitable for potential clinical applications where the imaging depth is critical. It is worth noting that merging imaging modalities with different spatial resolutions is a common practice in biomedical imaging, such as in PET/CT [40]. PA images cannot provide detailed vascular morphology (Fig. 5b), mostly due to the low ultrasound frequency and the limited-view issue. The quantitative analysis in Fig. 7 does not improve the spatial resolution of the sO2 measurement. To further improve the accuracy of ULM, the volumetric frame rate of the US imaging should be improved, for example, by optimizing the transmission pattern with fewer transmitting events. Nevertheless, further studies are warranted to assess the impact of these changes on the ULM image quality. Moreover, the 3D PA imaging still has the limited view problem (e.g., the missing diving vessels). We can potentially address this issue by rotating the transducer array or by using microbubbles as virtual point sources [41]. In this work, the PA imaging of the deep-brain region still suffers from the optical attenuation and limited-view problem that further impact the accuracy of sO2 measurements, which is a long-standing issue for quantitative PA imaging that will not be fully addressed in this work [23]. Nevertheless, we chose optical wavelengths that were relatively close to reduce this impact. Moreover, adopting the deconvolution method that highlights the low-frequency PA signals would help improve the accuracy of deep sO2 measurement, with the cost of spatial resolution. Similarly, localization-based techniques can be adapted by PA imaging to achieve super-resolution imaging [33], [42], [43], enhance imaging depth, and improve the quantitative accuracy, which can be integrated with ULM in future work. These methods track the central positions of flowing optical absorbers in the blood vessels, providing fine vessel structures and functional information such as flow speed. However, the localization-based PA imaging is much slower than ULM, mainly limited by the laser’s pulse repetition rate.

The current image reconstruction method is 3D DAS with a homogenous speed of sound (SOS). The image quality and subsequent processing accuracy can be further improved by adopting multi-SOS reconstruction [44]. This can be achieved by mapping the skull location and thickness using the US B-mode image, as well as estimating the speed of sound heterogeneity. The use of more accurate reconstruction methods is expected to enhance PD imaging, and more significantly, the ULM imaging, as microbubbles can be more accurately reconstructed and better matched with the system’s point spread function (PSF). Further improvement in localization accuracy can be achieved by incorporating the spatial variability of the system’s PSF, for example, through deep-learning-based methods. In our study, vessel segmentation was only performed on the PD images, as vessel continuity is important for accurate vessel extraction. However, more volumetric frames can be acquired for complete microbubble localization, and vessel segmentation may be performed on the ULM images.

Since our system is totally non-invasive and capable of real-time imaging, it could be applied for treatment guidance, such as monitoring thermal therapy of brain cancers [45], [46]. In addition, we can incorporate PA molecular imaging into our system, such as using photoswitchable phytochromes [47], [48]. With the combination of these inherently compatible image modalities, we believe that our technology can be a powerful tool for a broad range of preclinical and clinical studies.

Acknowledgments

This work was sponsored by National Institutes of Health (R21EB027981, R21 EB027304, RF1 NS115581, R01 NS111039, R01R01DK139109, R01 EB028143, R21 EB030072), National Science Foundation CAREER award 2144788,Duke University Pratt Beyond the Horizon Grant, Eli Lilly Research Award Program; and Chan Zuckerberg Initiative Grant (2020-226178).

Contributor Information

Yuqi Tang, Department of Biomedical Engineering, Duke University, Durham, NC 27703 USA.

Nanchao Wang, Department of Biomedical Engineering, Duke University, Durham, NC 27703 USA.

Zhijie Dong, Beckman Institute for Advanced Science and Technology and the Department of Electrical and Computer Engineering, University of Illinois Urbana-Champaign, Urbana, IL 61801 USA.

Matthew Lowerison, Beckman Institute for Advanced Science and Technology and the Department of Electrical and Computer Engineering, University of Illinois Urbana-Champaign, Urbana, IL 61801 USA.

Angela del Aguila, multidisciplinary Brain Protection Program, Department of Anesthesiology, School of Medicine, Duke University, Durham 27710, NC, USA.

Natalie Johnston, Department of Biomedical Engineering, Duke University, Durham, NC 27703 USA.

Tri Vu, Department of Biomedical Engineering, Duke University, Durham, NC 27703 USA.

Chenshuo Ma, Department of Biomedical Engineering, Duke University, Durham, NC 27703 USA.

Yirui Xu, Department of Biomedical Engineering, Duke University, Durham, NC 27703 USA.

Wei Yang, multidisciplinary Brain Protection Program, Department of Anesthesiology, School of Medicine, Duke University, Durham 27710, NC, USA.

Pengfei Song, Beckman Institute for Advanced Science and Technology and the Department of Electrical and Computer Engineering, University of Illinois Urbana-Champaign, Urbana, IL 61801 USA.

Junjie Yao, Department of Biomedical Engineering, Duke University, Durham, NC 27703 USA.

REFERENCES

  • [1].Wu D, Huang L, Jiang MS, and Jiang H, “Contrast Agents for Photoacoustic and Thermoacoustic Imaging: A Review,” International Journal of Molecular Sciences, vol. 15, no. 12, Art. no. 12, Dec. 2014, doi: 10.3390/ijms151223616. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [2].Xia J, Yao J, and Wang LV, “Photoacoustic tomography: principles and advances,” Electromagn Waves (Camb), vol. 147, pp. 1–22, 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [3].Ntziachristos V and Razansky D, “Optical and Opto-Acoustic Imaging,” in Molecular Imaging in Oncology, Springer, Berlin, Heidelberg, 2013, pp. 133–150. doi: 10.1007/978-3-642-10853-2_4. [DOI] [PubMed] [Google Scholar]
  • [4].Beard P, “Biomedical photoacoustic imaging,” Interface Focus, Aug. 2011, Accessed: Sep. 20, 2019. [Online]. Available: https://royalsocietypublishing.org/doi/abs/10.1098/rsfs.2011.0028 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [5].Paltauf G, Nuster R, Haltmeier M, and Burgholzer P, “Experimental evaluation of reconstruction algorithms for limited view photoacoustic tomography with line detectors,” Inverse Problems, vol. 23, no. 6, pp. S81–S94, Nov. 2007, doi: 10.1088/0266-5611/23/6/S07. [DOI] [Google Scholar]
  • [6].Hester SC, Kuriakose M, Nguyen CD, and Mallidi S, “Role of Ultrasound and Photoacoustic Imaging in Photodynamic Therapy for Cancer,” Photochemistry and Photobiology, vol. 96, no. 2, pp. 260–279, Mar. 2020, doi: 10.1111/php.13217. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [7].Karmacharya MB, Sultan LR, and Sehgal CM, “Photoacoustic monitoring of oxygenation changes induced by therapeutic ultrasound in murine hepatocellular carcinoma,” Sci Rep, vol. 11, no. 1, Art. no. 1, Feb. 2021, doi: 10.1038/s41598-021-83439-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [8].Sangha GS and Goergen CJ, “Label-free photoacoustic and ultrasound imaging for murine atherosclerosis characterization,” APL Bioeng, vol. 4, no. 2, Jun. 2020, doi: 10.1063/1.5142728. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [9].Zhang Y and Wang L, “Video-Rate Ring-Array Ultrasound and Photoacoustic Tomography,” IEEE Transactions on Medical Imaging, vol. 39, no. 12, pp. 4369–4375, Dec. 2020, doi: 10.1109/TMI.2020.3017815. [DOI] [PubMed] [Google Scholar]
  • [10].Errico C et al. , “Ultrafast ultrasound localization microscopy for deep super-resolution vascular imaging,” Nature, vol. 527, no. 7579, pp. 499–502, Nov. 2015, doi: 10.1038/nature16066. [DOI] [PubMed] [Google Scholar]
  • [11].Couture O, Hingot V, Heiles B, Muleki-Seya P, and Tanter M, “Ultrasound Localization Microscopy and Super-Resolution: A State of the Art,” IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 65, no. 8, pp. 1304–1320, Aug. 2018, doi: 10.1109/TUFFC.2018.2850811. [DOI] [PubMed] [Google Scholar]
  • [12].Renaudin N, Demené C, Dizeux A, Ialy-Radio N, Pezet S, and Tanter M, “Functional ultrasound localization microscopy reveals brain-wide neurovascular activity on a microscopic scale,” Nat Methods, vol. 19, no. 8, Art. no. 8, Aug. 2022, doi: 10.1038/s41592-022-01549-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [13].Stride E and Saffari N, “Microbubble ultrasound contrast agents: A review,” Proc Inst Mech Eng H, vol. 217, no. 6, pp. 429–447, Jun. 2003, doi: 10.1243/09544110360729072. [DOI] [PubMed] [Google Scholar]
  • [14].Quaia E, “Microbubble ultrasound contrast agents: an update,” Eur Radiol, vol. 17, no. 8, pp. 1995–2008, Aug. 2007, doi: 10.1007/s00330-007-0623-0. [DOI] [PubMed] [Google Scholar]
  • [15].Heiles B et al. , “Volumetric Ultrasound Localization Microscopy of the Whole Rat Brain Microvasculature,” IEEE Open Journal of Ultrasonics, Ferroelectrics, and Frequency Control, vol. 2, pp. 261–282, 2022, doi: 10.1109/OJUFFC.2022.3214185. [DOI] [PubMed] [Google Scholar]
  • [16].Demeulenaere O et al. , “In vivo whole brain microvascular imaging in mice using transcranial 3D Ultrasound Localization Microscopy,” eBioMedicine, vol. 79, p. 103995, 2022, doi: 10.1016/j.ebiom.2022.103995. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [17].Robin J, Özbek A, Reiss M, Dean-Ben XL, and Razansky D, “Dual-Mode Volumetric Optoacoustic and Contrast Enhanced Ultrasound Imaging With Spherical Matrix Arrays,” IEEE Transactions on Medical Imaging, vol. 41, no. 4, pp. 846–856, Apr. 2022, doi: 10.1109/TMI.2021.3125398. [DOI] [PubMed] [Google Scholar]
  • [18].Zhao S, Hartanto J, Joseph R, Wu C-H, Zhao Y, and Chen Y-S, “Hybrid photoacoustic and fast super-resolution ultrasound imaging,” Nat Commun, vol. 14, no. 1, Art. no. 1, Apr. 2023, doi: 10.1038/s41467-023-37680-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [19].Jensen JA, Nikolov SI, Gammelmark KL, and Pedersen MH, “Synthetic aperture ultrasound imaging,” Ultrasonics, vol. 44, pp. e5–e15, Dec. 2006, doi: 10.1016/j.ultras.2006.07.017. [DOI] [PubMed] [Google Scholar]
  • [20].Treeby BE and Cox BT, “k-Wave: MATLAB toolbox for the simulation and reconstruction of photoacoustic wave fields,” JBO, vol. 15, no. 2, p. 021314, Mar. 2010, doi: 10.1117/1.3360308. [DOI] [PubMed] [Google Scholar]
  • [21].Menozzi L, del Águila Á, Vu T, Ma C, Yang W, and Yao J, “Three-dimensional non-invasive brain imaging of ischemic stroke by integrated photoacoustic, ultrasound and angiographic tomography (PAUSAT),” Photoacoustics, vol. 29, p. 100444, Feb. 2023, doi: 10.1016/j.pacs.2022.100444. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [22].Kelechava B, “ANSI Z136.1–2022: Safe Use of Lasers - ANSI Blog,” The ANSI Blog. Accessed: Jul. 14, 2023. [Online]. Available: https://blog.ansi.org/ansi-z136-1-2022-safe-use-of-lasers/ [Google Scholar]
  • [23].Li M, Tang Y, and Yao J, “Photoacoustic tomography of blood oxygenation: A mini review,” Photoacoustics, vol. 10, pp. 65–73, Jun. 2018, doi: 10.1016/j.pacs.2018.05.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [24].Cox B, Laufer JG, Arridge SR, and Beard PC, “Quantitative spectroscopic photoacoustic imaging: a review,” J Biomed Opt, vol. 17, no. 6, p. 061202, Jun. 2012, doi: 10.1117/1.JBO.17.6.061202. [DOI] [PubMed] [Google Scholar]
  • [25].Song P et al. , “Improved Super-Resolution Ultrasound Microvessel Imaging With Spatiotemporal Nonlocal Means Filtering and Bipartite Graph-Based Microbubble Tracking,” IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 65, no. 2, pp. 149–167, Feb. 2018, doi: 10.1109/TUFFC.2017.2778941. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [26].Demené C et al. , “Spatiotemporal Clutter Filtering of Ultrafast Ultrasound Data Highly Increases Doppler and fUltrasound Sensitivity,” IEEE Transactions on Medical Imaging, vol. 34, no. 11, pp. 2271–2285, Nov. 2015, doi: 10.1109/TMI.2015.2428634. [DOI] [PubMed] [Google Scholar]
  • [27].Huang C et al. , “Short Acquisition Time Super-Resolution Ultrasound Microvessel Imaging via Microbubble Separation,” Sci Rep, vol. 10, no. 1, Art. no. 1, Apr. 2020, doi: 10.1038/s41598-020-62898-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [28].Jaqaman K et al. , “Robust single-particle tracking in live-cell time-lapse sequences,” Nat Methods, vol. 5, no. 8, Art. no. 8, Aug. 2008, doi: 10.1038/nmeth.1237. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [29].Klippel P, “Enabling fast image reconstruction for photoacoustic tomography using GPUs,” 2022.
  • [30].Aylward SR and Bullitt E, “Initialization, noise, singularities, and scale in height ridge traversal for tubular object centerline extraction,” IEEE Transactions on Medical Imaging, vol. 21, no. 2, pp. 61–75, Feb. 2002, doi: 10.1109/42.993126. [DOI] [PubMed] [Google Scholar]
  • [31].Deán-Ben XL and Razansky D, “Portable spherical array probe for volumetric real-time optoacoustic imaging at centimeter-scale depths,” Opt. Express, vol. 21, no. 23, pp. 28062–28071, Nov. 2013, doi: 10.1364/OE.21.028062. [DOI] [PubMed] [Google Scholar]
  • [32].Luís Dean-Ben X and Razansky D, “Localization optoacoustic tomography,” Light: Science & Applications, vol. 7, no. 4, pp. 18004–18004, Apr. 2018, doi: 10.1038/lsa.2018.4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [33].Shin Y et al. , “Context-Aware Deep Learning Enables High-Efficacy Localization of High Concentration Microbubbles for Super-Resolution Ultrasound Localization Microscopy,” bioRxiv, 2023, doi: 10.1101/2023.04.21.536599. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [34].Hingot V, Chavignon A, Heiles B, and Couture O, “Measuring Image Resolution in Ultrasound Localization Microscopy,” IEEE Transactions on Medical Imaging, vol. 40, no. 12, pp. 3812–3819, 2021, doi: 10.1109/TMI.2021.3097150. [DOI] [PubMed] [Google Scholar]
  • [35].Vu T et al. , “On the Importance of Low-Frequency Signals in Functional and Molecular Photoacoustic Computed Tomography,” IEEE Transactions on Medical Imaging, vol. 43, no. 2, pp. 771–783, 2024, doi: 10.1109/TMI.2023.3320668. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [36].Regenhardt RW et al. , “Blood Pressure and Penumbral Sustenance in Stroke from Large Vessel Occlusion,” Frontiers in Neurology, vol. 8, 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [37].Estrada H, Özbek A, Robin J, Shoham S, and Razansky D, “Spherical Array System for High-Precision Transcranial Ultrasound Stimulation and Optoacoustic Imaging in Rodents,” IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 68, no. 1, pp. 107–115, Jan. 2021, doi: 10.1109/TUFFC.2020.2994877. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [38].Liu S, Levine SR, and Winn HR, “Targeting ischemic penumbra: part I - from pathophysiology to therapeutic strategy,” J Exp Stroke Transl Med, vol. 3, no. 1, pp. 47–55, Mar. 2010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [39].Shanbhag NC, Henning RH, and Schilling L, “Long-term survival in permanent middle cerebral artery occlusion: a model of malignant stroke in rats,” Sci Rep, vol. 6, no. 1, Art. no. 1, Jun. 2016, doi: 10.1038/srep28401. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [40].Townsend David W., Carney Jonathan P.J., Yap Jeffrey T., and Hall Nathan C., “PET/CT Today and Tomorrow,” J Nucl Med, vol. 45, no. 1 suppl, p. 4S, Jan. 2004. [PubMed] [Google Scholar]
  • [41].Tang Y et al. , “High-fidelity deep functional photoacoustic tomography enhanced by virtual point sources,” Photoacoustics, vol. 29, p. 100450, Feb. 2023, doi: 10.1016/j.pacs.2023.100450. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [42].Deán-Ben XL et al. , “Deep optoacoustic localization microangiography of ischemic stroke in mice,” Nature Communications, vol. 14, no. 1, p. 3584, Jun. 2023, doi: 10.1038/s41467-023-39069-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [43].Zhang P, Li L, Lin L, Shi J, and Wang LV, “In vivo superresolution photoacoustic computed tomography by localization of single dyed droplets,” Light: Science & Applications, vol. 8, no. 1, p. 36, Apr. 2019, doi: 10.1038/s41377-019-0147-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [44].Cui M, Zuo H, Wang X, Deng K, Luo J, and Ma C, “Adaptive photoacoustic computed tomography,” Photoacoustics, vol. 21, p. 100223, Mar. 2021, doi: 10.1016/j.pacs.2020.100223. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [45].Gonzalez EA, Jain A, and Bell MAL, “Combined Ultrasound and Photoacoustic Image Guidance of Spinal Pedicle Cannulation Demonstrated With Intact ex vivo Specimens,” IEEE Transactions on Biomedical Engineering, vol. 68, no. 8, pp. 2479–2489, Aug. 2021, doi: 10.1109/TBME.2020.3046370. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [46].Song H et al. , “Real-Time Intraoperative Surgical Guidance System in the da Vinci Surgical Robot Based on Transrectal Ultrasound/Photoacoustic Imaging With Photoacoustic Markers: An Ex Vivo Demonstration,” IEEE Robotics and Automation Letters, vol. 8, no. 3, pp. 1287–1294, Mar. 2023, doi: 10.1109/LRA.2022.3191788. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [47].Yao J et al. , “Multiscale photoacoustic tomography using reversibly switchable bacterial phytochrome as a near-infrared photochromic probe,” Nat Methods, vol. 13, no. 1, Art. no. 1, Jan. 2016, doi: 10.1038/nmeth.3656. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [48].Li L et al. , “Small near-infrared photochromic protein for photoacoustic multi-contrast imaging and detection of protein interactions in vivo,” Nat Commun, vol. 9, no. 1, Art. no. 1, Jul. 2018, doi: 10.1038/s41467-018-05231-3. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES