Abstract
Purpose
Pulmonary MRI faces challenges due to low proton density, rapid transverse magnetization decay, and cardiac and respiratory motion. The fermat‐looped orthogonally encoded trajectories (FLORET) sequence addresses these issues with high sampling efficiency, strong signal, and motion robustness, but has not yet been applied to phase‐resolved functional lung (PREFUL) MRI—a contrast‐free method for assessing pulmonary ventilation during free breathing. This study aims to develop a reconstruction pipeline for FLORET UTE, enhancing spatial resolution for three‐dimensional (3D) PREFUL ventilation analysis.
Methods
The FLORET sequence was used to continuously acquire data over 7 ± 2 min in 36 participants, including healthy subjects (N = 7) and patients with various pulmonary conditions (N = 29). Data were reconstructed into respiratory images using motion‐compensated low‐rank reconstruction, and a 3D PREFUL algorithm was adapted to quantify static and dynamic ventilation surrogates. Image sharpness and signal‐to‐noise ratio were evaluated across different motion states. PREFUL ventilation metrics were compared with static 129Xe ventilation MRI.
Results
Optimal image sharpness and accurate ventilation dynamics were achieved using 24 respiratory bins, leading to their use in the study. A strong correlation was found between 3D PREFUL FLORET UTE ventilation defect percentages (VDPs) and 129Xe VDPs (r ≥ 0.61, p < 0.0001), although PREFUL FLORET static VDPs were significantly higher (mean bias = −10.1%, p < 0.0001). In diseased patients, dynamic ventilation parameters showed greater heterogeneity and better alignment with 129Xe VDPs.
Conclusion
The proposed reconstruction pipeline for FLORET UTE MRI offers improved spatial resolution and strong correlation with 129Xe MRI, enabling dynamic ventilation quantification that may reveal airflow abnormalities in lung disease.
Keywords: FLORET, hyperpolarized 129Xe MRI, lung MRI, motion compensation, PREFUL, UTE, ventilation imaging
1. INTRODUCTION
MRI of the lung is constrained by several challenges, including low proton density, which limits parenchymal signal; rapid transverse magnetization decay, which reduces high‐frequency signal information and increases blurring; and motion artifacts caused by cardiac and respiratory motion. 1 These challenges can be mitigated by using a sequence that (i) samples from the center of k‐space outward, to oversample low‐frequency signal regions and increase robustness to motion; (ii) has a short echo time, to minimize low‐frequency signal loss; (iii) has short readouts, to minimize T2* signal loss in high‐frequency regions of k‐space; and (iv) samples k‐space in a temporally incoherent manner, such that subsequent aliasing caused by retrospective subsampling of data propagates as incoherent noise (e.g., when dividing data into respiratory‐resolved bins).
One sequence that meets all these requirements is three‐dimensional (3D) fermat‐looped orthogonally encoded trajectories (FLORET), which samples 3D volumes using efficient Fermat spirals projected outward from the center of k‐space. Willmering et al. demonstrated that FLORET sampling of the lungs supports high parenchymal signal and increased sharpness compared with 3D radial kooshball, with increased sampling efficiency and thus reduced scan time. 2 , 3 Additionally, FLORET can be tuned to sample k‐space in a highly incoherent manner, through the use of multiple hubs and interleave ordering schemes to distribute the spiral arms. 4 Consequently, we reason that FLORET could be highly suitable for dynamic lung MRI, where continuously acquired data are retrospectively gated according to respiratory position and reconstructed into a series of images throughout the breathing cycle. However, such dynamic lung MRI reconstruction and analysis using FLORET sampling has yet to be explored in the literature.
Recently, a promising reconstruction technique for handling respiratory motion was proposed called motion‐compensated low rank (MoCoLoR) reconstruction. 5 This approach relies on the assumption that the matrix rank between reconstructed images at different respiratory states, when registered to each other, should be inherently low. Subsequently, the MoCoLoR algorithm converges to a solution that has regularized nuclear norm sparsity along the registered respiratory phase dimension. This approach demonstrated high‐resolution reconstructions of data acquired using 3D radial kooshball sampling, binned into up to 50 respiratory frames. However, this has not yet been applied to FLORET sampling, which may offer improved signal‐to‐noise ratio (SNR), image sharpness, and sampling efficiency.
Following reconstruction, the respiratory‐resolved image reconstructions can be compared against each other to assess pulmonary ventilation. 6 Numerous publications have focused on this topic, including Fourier decomposition, 7 SENCEFUL (self‐gated non‐contrast‐enhanced functional lung), 8 phase‐resolved‐functional lung (PREFUL),9 and matrix pencil MRI. 10 Except for PREFUL or SENCEFUL, the techniques mentioned here have been limited to two‐dimensional (2D) imaging. Three‐dimensional PREFUL MRI has been proposed previously using spoiled gradient‐echo sequence with stack‐of stars acquisition and golden‐angle increment, but has yet to be applied to 3D FLORET sampling where image quality and sampling efficiency are improved. 11
The objectives of this study were to optimize a binning algorithm and implement MoCoLoR image reconstruction for 3D FLORET ultrashort echo time (UTE) acquisitions. The resultant respiratory‐resolved images undergo a 3D PREFUL algorithm to extract static and dynamic ventilation parameters. The feasibility of the proposed methods was tested, and the static and dynamic ventilation information of 3D PREFUL FLORET UTE were compared with the direct static measurement of pulmonary ventilation derived by 129Xe MRI in volunteers with distinct pulmonary disease and healthy controls.
2. METHODS
2.1. Human participants
Thirty‐six participants were imaged after obtaining informed consent (and age‐appropriate assent for participants aged 11 years and older) following a protocol approved by the Institutional Review Board at Cincinnati Children's Hospital Medical Center (with Food and Drug Administration approval for 129Xe MRI IND‐123577). The studied population included healthy participants and participants with a range of lung conditions, including cystic fibrosis (CF), lymphangioleiomyomatosis (LAM), post–hematopoietic stem‐cell transplant recipients (post‐HSCT), bronchiolitis obliterans syndrome (BOS), and neuroendocrine cell hyperplasia of infancy. Table 1 describes the study participant demographics.
TABLE 1.
Study participants characteristics and quantitative ventilation measurements derived using the 129Xe and 3D PREFUL FLORET MRI.
All study participants (N = 36) | Healthy participants (N = 7) | CF (N = 14) | LAM (N = 7) | Post‐HSCT (N = 4) | BOS (N = 3) | NEHI (N = 1) | r Value a | |
---|---|---|---|---|---|---|---|---|
Age range | 7–55 | 8–27 | 7–22 | 39–55 | 11–21 | 14–21 | 13 | ‐ |
Sex | 21 M/15 F | 5 M/2 F | 8 M/6 F | 7 F | 4 M | 3 M | 1 M | – |
VDPXe [%] | 7.1 (4.6–14.9) | 4.2 (3.1–4.5) | 14.9 (4.4–21.3) | 8.1 (6.6–10.9) | 5.9 (5.4–7.1) | 11.8 (8.3–16.9) | 14.1 | – |
VDPRVent [%] | 18.6 (12.7–25.5) | 10.8 (6.7–11.5) | 18.9 (14.9–31.4) | 21.4 (19.1–29.8) | 16.0 (14.3–18.0) | 17.7 (16.1–29.9) | 25.1 | 0.76* |
VDPFVL‐CM [%] | 10.4 (4.1–15.5) | 1.9 (1.4–4.6) | 8.9 (7.2–16.1) | 13.6 (11.7–14.9) | 10.0 (6.1–12.4) | 12.6 (7.4–21.7) | 20.5 | 0.61* |
Mean VTTP [%] | 50.7 (48.1–53.4) | 49.1 (47.5–51.8) | 50.3 (49.4–51.9) | 51.7 (49.3–53.7) | 45.8 (43.7–49.3) | 54.0 (52.4–54.7) | 56.0 | 0.36* |
Note: Values of ventilation parameters are median with interquartile range in brackets.
Abbreviations: 3D, three‐dimensional; BOS, bronchiolitis obliterans syndrome; CF, cystic fibrosis; F, female; HSCT, hematopoietic stem‐cell transplant; LAM, lymphangioleiomyomatosis; M, male; NEHI, neuroendocrine cell hyperplasia of infancy; Post‐HSCT, post–hematopoietic stem cell transplantation; PREFUL, phase‐resolved functional lung; VDPFVL‐CM, ventilation defect percentage based on flow‐volume‐loop correlation metric; VDPRVent, VDP based on regional ventilation; VDPXe, VDP based on the Xenon; VTTP, ventilation time‐to‐peak.
Pearson correlation analysis (r) assessed the relationship between ventilation parameters derived using the 129Xe (VDPXe) and 3D PREFUL FLORET UTE MRI for all study participants.
Significant correlations: p < 0.05.
2.2. Image acquisition– 129Xe
2.2.1. Acquisition and reconstruction
Xenon gas (enriched to > 85% 129Xe) was hyperpolarized to 20%–40% using a continuous‐flow, Polarean 9820A 129Xe hyperpolarizer (Polarean Imaging PLC, Durham, NC, USA). 12 Polarized gas was administered to patients from volume‐matched Tedlar bags (Jensen Inert, Coral Springs, FL, USA). 129Xe gas was dosed at one‐sixth of the subject sex‐predicted and height‐ predicted total lung capacity (up to 1 L) for CF patients, 13 and one‐sixth of the predicted forced vital capacity (up to 1 L) for the remaining study participants. Doses were administered from expiration at functional residual capacity.
Breath‐hold (< 15 s) 129Xe ventilation images were acquired on a 3T Philips Ingenia‐Elition MRI scanner (Philips Healthcare, Best, Netherlands). 129Xe acquisitions were performed with a quadrature vest coil (Clinical MR Solutions, Brookfield, WI, USA). Images were acquired with a multislice 2D spoiled gradient‐echo sequence, field of view (FOVxy) = (300–400)mm2, FOVz = 150–200 mm, resolution = (3 × 3 × 15)mm3, slice gap = 0 mm, echo time (TE)/repetition time (TR) = < 5 ms/< 10 ms, and total scan duration < 15 s. 129Xe calibration scans were performed to calculate the optimal flip angle (8°–20°), depending on the number of excitations and receiver frequency for each subject. All scans were completed within < 60‐min total time, with 129Xe scans performed first, and FLORET UTE scans performed last (starting at about 45 min).
2.2.2. VDP analysis
129Xe ventilation defect percentage (VDPXe) was calculated from the percentage of the lung voxels with signal intensity below 60% of the mean whole‐lung signal intensity. 14 , 15 Image segmentations were performed using an in‐house pretrained deep learning model that excludes the airways. 16 Ventilation images were corrected for B1 inhomogeneity using N4‐ITK bias correction before VDP calculation. 17
2.3. Image acquisition–proton
2.3.1. Acquisition
Proton lung images were acquired during free breathing using a 3D FLORET UTE sequence, with the following acquisition parameters: FOV = (320 40 mm)3, acquisition resolution = (1.6 mm)3, TE/TR = 0.1/3.76 ms, readout length = 0.95 ms, flip angle = 4.0°, total excitations = 140 000 20 000, hubs = 3, total scan duration = 7 2 min. The variable number of excitations were a result of controlling the Nyquist sampling percentage for each exam's FOV. Images were acquired with a 16‐channel anterior coil and a 12‐channel posterior coil with a respiratory bellows belt to acquire respiratory position with each excitation. Images were acquired within 30 min of the 129Xe ventilation acquisition.
2.3.2. Binning algorithm
The binning algorithm described by Capaldi et al. 18 was used and optimized in this study, as illustrated in Figure 1. First, the respiratory bellows signal was smoothed and filtered using a moving average function. Second, high‐amplitude and low‐amplitude oscillations were excluded (±2 standard deviations of the mean bellows reading). Third, the bellows reading indices (corresponding to each RF excitation) were sorted into [8, 12, 16, 20, 24, 32, 40] respiratory‐motion bins according to instantaneous phase calculated by Hilbert transformation and applied to raw k‐space data and trajectories. Equal numbers of excitations were sorted into each respiratory motion bin ([30 000, 20 000, 15 000, 12 000, 10 000, 7500, 6000] excitations per bin), such that the total amount of excitations equaled 240 000. Dynamic view sharing was permitted such that each respiratory bin shared up to 50% excitations.
FIGURE 1.
Respiratory binning workflow. (A) Waveform collected from respiratory bellows measurement (only first 20 000 excitations; ˜1 min shown). (B) Hilbert‐transform of respiratory bellows waveform to separate each respiratory cycle into phase position − to . Transformed data were binned into 24 windows containing 10 000 excitations each. Different colors refer to different respiratory bins. (C) The degree of overlap within each bin. Small white lines in (C) correspond to excluded excitations that fell beyond ±2 standard deviations of the mean respiratory bellows measurement.
2.3.3. Reconstruction
Proton data were reconstructed with multiple respiratory phases using a motion‐compensated low rank (MoCoLoR) reconstruction algorithm. 5 The following reconstruction problem was solved:
(1) |
where is the underlying image with respiratory states and image size; are the k‐space preconditioning weights for excitations (readouts) per respiratory phase and samples per readout; is the nonuniform fast Fourier transform; are the coil sensitivity maps for coils; are the binned k‐space measurements; λL is the low‐rank regularization parameter; is the nuclear norm; and are the motion field operators. were calculated using the single‐channel k‐space preconditioning approach by Ong et al., as it offered the best balance between computation time and image quality. 19 Motion field operators were estimated using registration software within the ANTs framework. 20 Sensitivity maps were estimated using the JSENSE/NLINV algorithm described by Uecker et al. 21 and Ying et al. 22 The full reconstruction workflow is visualized in Figure 2. Reconstructions were performed in Python 3.12, using linear algebra functionality from SigPy, and an NVIDIA‐A5000 (24GB) graphical processing unit on a Linux distribution. Reconstruction settings included resolution = (3 × 3 × 3)mm3, λL = 0.01, inner iterations = 5, outer iterations = 3, superior iterations = 3, and step size = 1.
FIGURE 2.
Three‐dimensional FLORET (fermat‐looped orthogonally encoded trajectories) reconstruction pipeline for this study. In chronological order: (1) Excitation indices are binned according to position in respiratory cycle and applied to raw k‐space data and trajectories; (2) sensitivity maps and k‐space preconditioners are calculated; (3) conjugate‐gradient sensitivity encoding (CG‐SENSE) is explored to initialize images corresponding to each respiratory phase; (4) motion field from initial images is estimated, iterated through the motion‐compensated low‐rank reconstruction (MoCoLoR) algorithm using the initial motion fields as inputs; and (5) reconstruction is complete.
2.3.4. PREFUL analysis
2.3.4.1. Image registration and postprocessing
Following image reconstruction, the respiratory‐resolved data were resorted such that the ventilation cycle started at end‐expiration, running to end‐inspiration, and ending back at end‐expiration. Subsequently, the reordered respiratory‐resolved images underwent bias‐field correction using the “N4BiasFieldCorrection” algorithm, 17 integrated within ANTs framework to alleviate intensity inhomogeneity.
To enable voxel‐wise analysis of the respiratory‐resolved data, it is necessary to implement motion‐compensation techniques to account for the movement of the heart and lungs. Consequently, following a method described by Klimeš et al., 11 all respiratory phases were subsequently stepwise‐registered to the phase at maximum inspiration using the ANTs toolkit. During registration, the initial affine transform was followed by the nonrigid flexible registration. For the nonrigid registration, a B‐spline symmetric normalization diffeomorphic registration algorithm was executed with the following parameters: gradient step = 1, updated field mesh size at base level = 40, total field mesh size at base level = 0, spline order = 3, metric = cross‐correlation, radius = 4, metric weight = 1, convergence parameter = 150 × 150 × 100 × 80/1 × 10−6/10, shrink parameters = 6 × 4 × 2 × 1, and smoothing parameters = 3 × 2 × 1 × 0. For both rigid and nonrigid transformations, a rectangular mask covering the entire 3D lung volume was used to reduce convergence time.
Finally, the respiratory images were interpolated into 16 equidistant phases using a Gaussian kernel with a sigma value of 0.3, inspired by the approach outlined by Voskrebenzev et al. 9 To mitigate signal variations unrelated to respiratory motion, a low‐pass filter with a cutoff frequency of 0.7 Hz was used on the interpolated phases. Additionally, a 3D edge‐preserving filter was applied to the images to maintain the structural integrity while removing noise. 23
2.3.4.2. Lung parenchyma and vessel segmentation
The lung cavity within the end‐inspiration image was used as a reference image for image registration. The cavity was segmented using an in‐house pretrained deep learning model 16 that excludes the airways and subsequent manual refinement by a postdoctoral researcher with 8 years of experience in pulmonary MRI. The lung‐cavity segmentation was followed by a lung‐vessel recognition. 24 The final lung parenchyma mask was generated by subtracting the lung‐vessel mask from the lung‐cavity mask.
2.3.4.3. Ventilation parameters
The ventilation parameters used in this study were based on those outlined in the original 3D PREFUL publication by Klimeš et al. 11 Regional ventilation (RVent) was quantified for each respiratory phase of the entire ventilation cycle. 25 RVent, a static ventilation parameter, considers two respiratory phases in its calculation and has been demonstrated to be influenced by both respiratory frequency and depth. 26 , 27 The RVent map of the end‐inspiration phase was then further considered for the image and statistical analysis.
The RVent cycle within each voxel was evaluated using flow‐volume loops (FVLs), where flow (the first derivative of RVent) was plotted on the y‐axis and volume (RVent) on the x‐axis. To establish a baseline, a high‐ventilation reference FVL was generated from a region where RVent values fall between the 75%–95% quantiles. The similarity between the FVL of each voxel and the high‐ventilation reference was then evaluated using the cross‐correlation metric (CM).
For both the RVent at end‐inspiration and the FVL‐CM parameters, maps depicting ventilation defects and VDP values were calculated using established thresholds. 9 , 28 Although a fixed threshold of 0.9 was used for FVL‐CM, resulting in VDPFVL‐CM values, RVent used a variable threshold. Specifically, the threshold for RVent was determined as 40% of the 90th percentile value, resulting in VDPRVent values. Both VDP parameters, derived using those published thresholds, have demonstrated a strong correlation with spirometry measurements, 11 129Xe ventilation maps, 29 and more recently, with VDP values obtained through direct visualization of ventilation using 19F MRI. 30
Subsequently, ventilation time‐to‐peak (VTTP) maps as a percentage of the ventilation cycle were generated. VTTP denotes the time at which ventilation attains its maximal magnitude within the whole ventilation cycle. In regions exhibiting normal physiology, ventilation ideally achieves its peak at approximately 50% of the cycle duration. Prolonged VTTP values may signify diminished or delayed ventilatory dynamics.
2.4. Statistical analysis
To establish the optimal regularization settings, a range of λL parameters (0.005–0.025) were used for MoCoLoR reconstruction. The quality of the ventilation images obtained from 3D PREFUL FLORET UTE was assessed both visually and quantitatively with Sobel image sharpness (∇), using data from a single healthy volunteer.
Postprocessing optimization was conducted in a subcohort consisting of 10 study participants (5 healthy participants, 2 patients with CF, 2 patients with BOS, and 1 patient with LAM). The 3D PREFUL analysis was assessed across different respiratory binning settings, including [8, 12, 16, 20, 24, 32, 40] respiratory bins. Significance testing using Kruskal‐Wallis analysis was performed to evaluate differences in ventilation parameters (mean RVent, standard deviation RVent, mean FVL‐CM, standard deviation FVL‐CM, mean VTTP, VDPRVent, and VDPFVL‐CM), as well as in signal‐to‐noise ratio (SNR) and mean sharpness of the RVent images across all settings. SNR was calculated in the central coronal slice at the level of the trachea bifurcation of a morphological image used as a fixed image for registration. At this location, regions of interest were selected inside the lung‐parenchyma mask and in the background. The same regions of interest were used for all tested reconstruction settings. The image sharpness was evaluated using the Sobel gradient operator in all three spatial directions. The final sharpness value was defined as an average gradient magnitude inside the lung‐parenchyma mask.
Subsequently Pearson correlation analysis was used to assess relationships between static VDPXe with static VDPRVent and dynamic VDPFVL‐CM in the whole study cohort, and the mean differences of the VDP values were quantified by Bland–Altman analysis.
Statistical analyses were performed using JMP Pro 17 (SAS Institute, Cary, NC, USA), and the level of statistical significance was set at p < 0.05.
3. RESULTS
3.1. Image‐reconstruction optimization
Morphological end‐inspiration and static 3D PREFUL FLORET UTE ventilation images from the MoCoLoR reconstruction optimizations are shown in Figure 3. The λL parameters ranging from 0.005 to 0.025 were used for the dynamic MoCoLoR reconstruction. Using Figure 3 as a reference for the rest of the study, image reconstructions with λL = 0.01 were chosen, as they demonstrated the best balance between image sharpness of ventilation maps (∇ = 2.07 in the Figure 3) and mitigation of aliasing artifacts/noise. Visually examining the morphological end‐inspiration images, the λL = 0.01 and λL = 0.005 settings also showed pronounced sharpness in the diaphragm region as well as in the pulmonary vasculature, compared with larger λL values.
FIGURE 3.
Comparison of different regularization parameters and their influence on image quality of ventilation images derived from three‐dimensional PREFUL (phase‐resolved functional lung) FLORET (fermat‐looped orthogonally encoded trajectories) ultrashort echo time. Overview of morphological end‐expiration (left) and regional ventilation (RVent; right) maps derived using different reconstruction settings (regularization parameter, λL) for a healthy participant (male, 11 years old). Note the superior Sobel image sharpness (∇) of ventilation images and improved qualitative sharpness around diaphragm as well as well‐defined vessels of morphological images for image reconstructions with λL = 0.005 and 0.01. VDP, ventilation defect percentage.
3.2. Postprocessing optimization
Regardless of the number of respiratory bins, there were no significant differences in ventilation and image sharpness and SNR parameters across all evaluated reconstruction settings (Table 2; all p > 0.87 for all parameters). Despite being statistically insignificant, we observed a small decreasing trend in mean RVent values with increasing number of respiratory bins, ranging from 0.09 mL/mL for 8 bins to 0.07 mL/mL for 40 bins. However, standard deviations of RVent, mean FVL‐CM, standard deviations of FVL‐CM, VDPRVent, and VDPFVL‐CM values remained stable without any discernible trend. Additionally, we observed a small but statistically insignificant decrease in mean VTTP values with the increasing number of reconstructed bins, from 52.7% for 8 bins to 49.6% for 40 bins. We observed a statistically insignificant increase in SNR values with the number of reconstructed bins, from 14.0 for 8 bins to 14.9 for 40 bins, while image sharpness decreased slightly from 1.3 to 1.0 between the 8‐bin and 40‐bin settings.
TABLE 2.
Assessment of ventilation parameters depending on the applied setting for image reconstruction in the study subcohort consisting of 10 participants.
Parameter/setting | 8bin | 12bin | 16bin | 20bin | 24bin | 32bin | 40bin | p‐Value/chi square a |
---|---|---|---|---|---|---|---|---|
mean RVent [mL/mL] | 0.09 (0.06–0.13) | 0.09 (0.07–0.14) | 0.09 (0.07–0.14) | 0.08 (0.06–0.14) | 0.08 (0.07–0.14) | 0.08 (0.06–0.13) | 0.07 (0.06–0.13) | 0.96/1.55 |
std RVent [mL/mL] | 0.04 (0.03–0.06) | 0.05 (0.03–0.06) | 0.05 (0.02–0.06) | 0.05 (0.02–0.06) | 0.05 (0.03–0.06) | 0.05 (0.02–0.06) | 0.05 (0.02–0.06) | 0.99/1.01 |
mean FVL‐CM [−] | 0.98 (0.96–0.99) | 0.98 (0.96–0.99) | 0.98 (0.96–0.99) | 0.98 (0.96–0.99) | 0.98 (0.97–0.99) | 0.98 (0.96–0.99) | 0.98 (0.97–0.99) | 1.00/0.29 |
std FVL‐CM [−] | 0.10 (0.07–0.20) | 0.08 (0.07–0.19) | 0.09 (0.06–0.20) | 0.09 (0.07–0.18) | 0.09 (0.06–0.18) | 0.11 (0.06–0.20) | 0.10 (0.06–0.18) | 1.00/0.38 |
mean VTTP [%] | 52.68 (48.18–53.32) | 52.05 (48.31–53.17) | 51.30 (49.46–52.38) | 51.36 (48.08–53.16) | 50.87 (49.42–53.61) | 49.18 (48.59–51.97) | 49.60 (47.64–52.66) | 0.98/1.09 |
VDPRVent [%] | 14.02 (10.24–19.29) | 13.70 (9.47–16.37) | 14.39 (10.27–14.94) | 13.52 (9.48–15.11) | 13.89 (9.06–14.11) | 13.74 (9.59–15.22) | 12.61 (9.29–15.46) | 1.00/0.36 |
VDPFVL‐CM[%] | 2.50 (1.40–4.29) | 2.22 (1.48–4.29) | 2.50 (1.38–4.67) | 2.39 (1.86–4.69) | 2.45 (1.80–5.23) | 2.98 (1.65–5.14) | 2.74 (1.23–4.38) | 1.00/0.40 |
SNR [−] | 14.04 (9.66–17.68) | 14.85 (9.85–17.31) | 13.54 (8.50–19.26) | 14.35 (9.12–17.62) | 14.52 (7.99–18.15) | 15.00 (10.00–20.14) | 14.90 (10.70–22.22) | 0.91/2.14 |
Sharpness [−] | 1.26 (0.90–1.74) | 1.22 (1.01–1.81) | 1.17 (1.02–1.79) | 1.14 (0.98–1.79) | 1.13 (1.00–1.74) | 1.05 (0.94–1.64) | 0.99 (0.88–1.61) | 0.87/2.51 |
Note: Values are median with interquartile range in brackets.
Abbreviations: FVL‐CM, flow volume loop correlation metric; NEHI, neuroendocrine cell hyperplasia of infancy; RVent, regional ventilation; SNR, signal‐to‐noise ratio; std, standard deviation; VDP, ventilation defect percentage; VTTP, ventilation‐time‐to‐peak.
Kruskal‐Wallis test was used to assess differences for all parameters across different settings. p‐Value < 0.05 was deemed significant.
3.3. Comparison of ventilation parameters derived using the 3D PREFUL FLORET UTE and 129Xe MRI
Example results demonstrating the correspondence of 3D PREFUL FLORET UTE‐derived static RVent ventilation maps (second row) with 129Xe MRI (third row) are shown in Figure 4. Note the homogeneous distribution of ventilation maps for the healthy‐volunteer participant in the first column. In contrast, ventilation maps for patients with disease (second‐fourth columns) depict ventilation heterogeneities for both techniques with good visual overlap.
FIGURE 4.
Comparison of static three‐dimensional PREFUL (phase‐resolved functional lung) FLORET (fermat‐looped orthogonally encoded trajectories) ultrashort echo time MRI (first–second row) and breath‐hold 129Xe ventilation MRI (third row). Four study participants are depicted: healthy volunteer (first column, female, 13 years old), subject with bronchiolitis obliterans syndrome (BOS; second column, male, 24 years old), subject with cystic fibrosis (CF; third column, female, 25 years old), and subject with lymphangioleiomyomatosis disease (LAM; fourth column, female, 39 years old). For both measurements, the whole‐lung ventilation defect percentage (VDP) values are listed below the respective ventilation map.
Dynamic ventilation cycles of 3D PREFUL FLORET UTE are depicted in Figure 5. A homogeneous ventilation cycle can be observed in the healthy volunteer (first row), whereas abnormal ventilation dynamics were evident in subjects with CF (second row), with BOS (third row) and with LAM (fourth row). In these diseases, airway narrowing may cause abnormal airflow, which is depicted as irregular filling throughout the ventilation cycle.
FIGURE 5.
Comparison of three‐dimensional PREFUL (phase‐resolved functional lung) FLORET (fermat‐looped orthogonally encoded trajectories) ultrashort echo time MRI–derived ventilation cycle. Healthy volunteer (first row, male, 13 years old), cystic fibrosis (CF; second row, female, 25 years old) patient, bronchiolitis obliterans syndrome (BOS; third row, male, 24 years old) patient, and lymphangioleiomyomatosis (LAM; fourth row, female, 39 years old) patient are presented. Note homogenous ventilation cycle for the healthy volunteer and no ventilation filling for consolidation in the right lung of the CF patient (cyan arrows). For both BOS and LAM subjects, the ventilation dynamics are much more heterogeneous. Dynamic filling asynchronization is indicated with yellow arrows for BOS and green arrows for LAM subjects, suggesting abnormal airflow related to airway narrowing.
Figure 6 presents a comparison of static and dynamic ventilation parameters of 3D PREFUL FLORET UTE and 129Xe ventilation images. Healthy subject (first column) showed homogeneous ventilation maps for both techniques. Significant findings were observed in BOS (second column) and CF (third column) patients, wherein dynamic ventilation parameters of 3D PREFUL FLORET UTE (FVL‐CM (third row) and VTTP (fourth row)) exhibit greater heterogeneity compared with the static RVent parameter (second row), aligning better with 129Xe ventilation MRI (fifth row) results. For the LAM patient (fourth column), the ventilation defects visible on the 129Xe ventilation maps were well‐matched using both the static and dynamic ventilation maps derived from 3D PREFUL FLORET UTE.
FIGURE 6.
Comparison of static and dynamic three‐dimensional (3D) PREFUL (phase‐resolved functional lung) FLORET (fermat‐looped orthogonally encoded trajectories) ultrashort echo time (UTE) MRI (first–fourth row) and breath‐hold 129Xe ventilation MRI (fifth row). Healthy volunteer (first column; male, 25 years old) subject with bronchiolitis obliterans syndrome (BOS; second column, male, 24 years old), subject with cystic fibrosis (CF; third column, male, 25 years old), and subject with lymphangioleiomyomatosis disease (LAM; fourth column, female, 39 years old) are depicted. The overlapping hypoventilation areas depicted in static regional ventilation of 3D PREFUL FLORET UTE (second row) and 129Xe ventilation imaging (fifth row) are marked with cyan arrows. In contrast, overlapping ventilation abnormalities derived with dynamic 3D PREFUL FLORET UTE parameters (third–fourth row) and 129Xe breath‐hold imaging are marked with red arrows. CM, cross‐correlation metric; FVL, flow‐volume loop; RVent, regional ventilation; VDP, ventilation defect percentage; VTTP, ventilation time‐to‐peak.
A summary of quantitative ventilation measurements for all study participants and participants grouped by disease is shown in Table 1.
All 3D PREFUL FLORET UTE MRI‐derived VDP values showed significant strong correlations with 129Xe MRI VDP values (all r ≥ 0.61, all p < 0.0001; Table 1). The static VDPRVent exhibited a stronger correlation with VDPXe than the dynamic VDPFVL‐CM. For VDPRVent, the Bland–Altman analysis showed significant differences from VDPXe (mean bias of −10.1%, p < 0.0001; Figure 7A). The correlation plot showing the relationship between static VDPRVent and VDPXe is displayed in Figure 7B (r = 0.76, p < 0.0001). No significant bias was observed between VDPFVL‐CM and VDPXe (mean bias of −1.4%, p = 0.22; Figure 7C). Figure 7D displays the correlation plot showing the relationship between dynamic VDPFVL‐CM and VDPXe (r = 0.61, p < 0.0001). Moderate correlation was found between VTTP and VDPXe (r = 0.36, p = 0.03; Table 1).
FIGURE 7.
Correspondence of ventilation defect percentage (VDP) values derived by three‐dimensional (3D) PREFUL (phase‐resolved functional lung) FLORET (fermat‐looped orthogonally encoded trajectories) ultrashort echo time MRI and 129Xe imaging. Bland–Altman (A,C) and correlation analysis (B,D) of global VDP values derived using 3D PREFUL MRI postprocessing and 129Xe MRI. Study groups are visually marked with distinct colors (green, healthy volunteers; pink, patients with cystic fibrosis [CF]; blue, patients with lymphangioleiomyomatosis [LAM]; orange, patients with bronchiolitis obliterans syndrome [BOS]; purple, patients post–hematopoietic stem cell transplantation [Post‐HSCT]; cyan, patient with neuroendocrine cell hyperplasia of infancy [NEHI]). cross‐correlation metric; FVL, flow‐volume loop.
4. DISCUSSION
This study introduced an optimized image reconstruction pipeline specifically developed for 1H FLORET UTE acquisition, providing a series of temporally resolved lung images suitable for pulmonary ventilation assessment with 3D PREFUL. The proposed method offered ventilation images with improved spatial resolution compared with previously published 3D PREFUL results (proposed 3mm3 vs. previous 3.9mm3) 11 and demonstrated a strong global correlation with static 129Xe MRI. Furthermore, the pipeline enabled calculation of dynamic ventilation parameters, which could be useful for understanding dynamic lung abnormalities caused by respiratory disease.
The first objective of the study was to optimize the respiratory binning (i.e., the number of respiratory phases to include) and MoCoLoR reconstruction settings for subsequent 3D PREFUL analysis. First, we noticed that the PREFUL‐derived VDP did not vary significantly with the number of bins used in the reconstruction. However, despite being statistically insignificant, we noticed that the image sharpness had a small decreasing trend with the increasing number of bins. Second, the increasing number of bins increased the load on the GPU and reconstruction time. Therefore, we opted to use 24 respiratory bins with 10 000 readouts per bin, as a qualitatively optimal trade‐off between image sharpness and temporal resolution of the ventilation cycle. Additionally, a MoCoLoR reconstruction regularization parameter (λL) setting of 0.01 offered a good balance between image sharpness and dynamic signal resolution between respiratory phases for 24 respiratory phases—of which are both highly important when studying heterogeneous abnormalities in the lung. However, it is noted that this binning algorithm and reconstruction approach relies on the quality of the respiratory bellows signal. Subsequently, detection of bulk motion using this method may not be feasible. Alternatively, sorting based on the zeroth k‐space sampling point (DC‐gating) 31 or use of a convolutional neuronal network for motion compensation 32 could enhance the stability of this binning algorithm. Further sorting according to cardiac phase may improve the accuracy of ventilation mapping in regions strongly affected by cardiac motion during free‐breathing acquisition.
By using the proposed MoCoLoR reconstruction scheme and applying the 3D PREFUL postprocessing algorithm, ventilation analysis was feasible for all study participants. In contrast to current 3D PREFUL literature, 11 this study achieved improved spatial resolution of 3 mm (isotropic) compared with the original publication's 3.9 mm (isotropic), potentially enhancing sensitivity to lung‐ventilation abnormalities. Notably, it is possible to obtain higher‐resolution images, as data were acquired at 1.8 mm; however, the reconstruction and registration times during image processing would have taken considerably longer to handle the larger matrix size. Consequently, 3 mm was chosen as the reconstruction resolution to match the highest in‐plane resolution of the 129Xe images in this study. Additionally, the FLORET acquisition method enabled more efficient k‐space sampling than conventional stack‐of‐stars acquisition, as expected according to Willmering et al., 2 which reduces scan time and offers more respiratory bins to be resolved. Although current 3D PREFUL studies have shown feasibility in single‐center, single‐vendor settings, our findings demonstrate the translatability of 3D PREFUL to multiple vendors, thereby making the technique accessible to more centers.
Across all study participants, a strong correlation was observed between VDP values derived from the proposed 3D PREFUL FLORET UTE algorithm and 129Xe MRI. The strongest correlation was found for the VDP of the static RVent parameter (VDPRVent) and VDP of 129Xe MRI (VDPXe). Recent publications have observed similar correlations for the 3D PREFUL technique with both stack‐of‐stars acquisition 33 and 3D radial UTE acquisition. 34
Notably, a significant bias (of 10%) toward higher VDPRVent values was observed in comparison to VDPXe values. Furthermore, the mean difference of VDP measures between 3D PREFUL FLORET UTE and 129Xe MRI increased with ventilation defect severity. Several explanations exist for the different VDP values between VDPRVent from 3D PREFUL FLORET UTE and the VDPXe from 129Xe MRI. First, the 3D PREFUL proton ventilation surrogates were derived indirectly from parenchymal‐density changes, whereas 129Xe MRI provides direct ventilation measurements derived from the density of inhaled magnetized xenon gas. Second, there were differences in spatial resolution (3 3 3 mm3 for 3D PREFUL FLORET versus anisotropic 3 3 15 mm3 for 129Xe MRI). McAllister et al. demonstrated that low‐resolution 129Xe ventilation images (voxel size of 2.2 × 6.4 × 21.0 mm3) tend to underestimate VDP due to their susceptibility to partial volume artifacts. 35 In accordance, VDP values of static RVent derived by 3D PREFUL FLORET UTE consistently exceeded VDP values of 129Xe. Validity could be improved by comparing 3D PREFUL FLORET data against 129Xe ventilation data acquired at the same resolution (e.g., as demonstrated in high‐resolution breath‐hold studies 36 , 37 ) or against computed tomography or V/Q scans. Third, each technique used different thresholds for VDP calculation based on previous studies with different study populations, which could potentially influence our results. While conducting this study, we aimed to use already established thresholds for 129Xe as well as 3D PREFUL MRI. Nevertheless, better agreement might be achieved by harmonizing the thresholding techniques. A fourth potential source of difference between 3D PREFUL FLORET UTE and 129Xe could be the variation in lung volumes used for image acquisition. The proton images are reconstructed at multiple respiratory phases during tidal breathing, typically starting at functional residual capacity (FRC) plus a mean tidal volume of 0.5 L, a value reported in the literature for men and women. 38 In contrast, 129Xe images may represent a larger lung volume, such as FRC plus 1 L. Hughes et al. reported a trend toward increased ventilation inhomogeneities at higher lung volumes, suggesting that this difference in lung volumes may contribute to variations observed between the two methods. 39 Fifth, differences in VDP values may arise from segmentation errors of the lung parenchyma and inaccuracies in image registration during the PREFUL ventilation analysis. A final source of the VDPXe and VDPRVent discrepancy could be related to lung atelectasis, as all 129Xe images were acquired first, about 30–45 min before FLORET UTE images. It is possible that the prolonged time in the supine position could have impeded the regional ventilation function preferentially in the FLORET UTE acquisition.
4.1. Limitations
The proposed algorithm has several limitations. First, the reconstruction time for temporally resolved images is 1–2 h, which limits the feasibility of real‐time reconstruction on the scanner. However, this is largely restricted by the ANTs software package, which currently does not support GPU processing. This also lengthens the duration for the PREFUL postprocessing, which can take up to 4 h. Future registration software improvements will speed up both reconstruction and postprocessing.
Second, although 129Xe imaging is a static breath‐hold measure, 3D PREFUL FLORET UTE provides dynamic measurement during free tidal breathing. Therefore, in 129Xe imaging, areas where gas flow is obstructed are depicted as defects, whereas in 3D PREFUL FLORET UTE, defect areas are correlated to areas of decreased tissue compliance or abnormal ventilation filling throughout the breathing cycle. Consequently, future work will explore advantages and disadvantages of each approach at detecting static and dynamic changes in lung pathophysiology.
Third, compared with proton‐based 2D measurements, the main disadvantage of the FLORET UTE acquisition is the lack of perfusion‐weighted information. This issue arises because the inflow effect of fresh blood cannot be measured due to the absence of slice selection. Finally, the clinical importance and repeatability of the dynamic ventilation parameters should be studied in greater detail. It is not yet known how dynamic lung abnormalities vary between diseases under the scope of this 3D PREFUL FLORET UTE technique. Additionally, the correlations presented here could be validated further by comparing against pulmonary function testing data such as forced expiratory volume in 1 s, which has shown strong correlations in 2D 40 , 41 , 42 and 3D 11 , 30 but was not included in our study. However, this proof‐of‐concept work highlights that PREFUL‐derived ventilation maps at multiple time points throughout the breathing cycle can be a useful addition in understanding dynamic lung abnormalities and improving clinical outcomes.
5. CONCLUSION
Using the optimized binning‐reconstruction and image‐reconstruction algorithm, 3D PREFUL FLORET UTE MRI offers improved isotropic spatial resolution of 3mm3 as a result of a more efficient acquisition scheme compared with previous methods. Subsequently, the proposed algorithm demonstrated promising visual correspondence of ventilation surrogate markers with 129Xe MRI along with a strong correlation of VDP values and less bias for the dynamic 3D PREFUL FLORET UTE measures. Future work should focus on assessing the repeatability of the proposed method and studying disease‐specific abnormalities.
CONFLICT OF INTEREST
F.K., A.V., and J.V.C. are shareholders of BioVisioneers GmbH, a company that has interest in pulmonary MRI methods. M.W. and L.W. have received consulting fees from Polarean Imaging, PLC.
ACKNOWLEDGMENTS
NIH R01HL166335 (Walkup), 2R01HL126771 (Cleveland), R01HL143011 (Cleveland), R01HL151588 (Cleveland), University of Cincinnati, Office of the Vice President for Research—URC Graduate Student Stipend and Research Cost Program for Faculty; Student Collaboration: D700169. Cincinnati Children's Research Foundation. National Organization for Rare Disorders (Award ID 20003, Walkup). Open Access funding enabled and organized by Projekt DEAL.
Klimeš F, Plummer JW, Willmering MM, et al. Quantifying spatial and dynamic lung abnormalities with 3D PREFUL FLORET UTE imaging: A feasibility study. Magn Reson Med. 2025;93(5):1984‐1998. doi: 10.1002/mrm.30416
Filip Klimeš and Joseph Plummer contributed equally to this work.
DATA AVAILABILITY STATEMENT
Reconstruction scripts (Python) and sample data are located at https://github.com/joeyplum/spatio‐temporal‐recon. We kindly encourage users to ask questions, post issues, and collaborate with us on this repository.
REFERENCES
- 1. Mulkern R, Haker S, Mamata H, et al. Lung parenchymal signal intensity in MRI: a technical review with educational aspirations regarding reversible versus irreversible transverse relaxation effects in common pulse sequences. Concepts Magn Reson Part A. 2014;43A:29‐53. doi: 10.1002/cmr.a.21297 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Willmering MM, Robison RK, Wang H, Pipe JG, Woods JC. Implementation of the FLORET UTE sequence for lung imaging. Magn Reson Med. 2019;82:1091‐1100. doi: 10.1002/mrm.27800 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Willmering MM, Krishnamoorthy G, Robison RK, Rosenberg JT, Woods JC, Pipe JG. High‐quality FLORET UTE imaging for clinical translation. Magn Reson Med. 2025;93:276‐288. doi: 10.1002/mrm.30277 [DOI] [PubMed] [Google Scholar]
- 4. Pipe JG, Zwart NR, Aboussouan EA, Robison RK, Devaraj A, Johnson KO. A new design and rationale for 3D orthogonally oversampled k‐space trajectories. Magn Reson Med. 2011;66:1303‐1311. doi: 10.1002/mrm.22918 [DOI] [PubMed] [Google Scholar]
- 5. Tan F, Zhu X, Chan M, et al. Motion‐compensated low‐rank reconstruction for simultaneous structural and functional UTE lung MRI. Magn Reson Med. 2023;90:1101‐1113. doi: 10.1002/mrm.29703 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Zapke M, Topf H‐G, Zenker M, et al. Magnetic resonance lung function—a breakthrough for lung imaging and functional assessment? A phantom study and clinical trial. Respir Res. 2006;7:106. doi: 10.1186/1465-9921-7-106 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Bauman G, Puderbach M, Deimling M, et al. Non‐contrast‐enhanced perfusion and ventilation assessment of the human lung by means of Fourier decomposition in proton MRI. Magn Reson Med. 2009;62:656‐664. doi: 10.1002/mrm.22031 [DOI] [PubMed] [Google Scholar]
- 8. Fischer A, Weick S, Ritter CO, et al. SElf‐gated non‐contrast‐enhanced FUnctional lung imaging (SENCEFUL) using a quasi‐random fast low‐angle shot (FLASH) sequence and proton MRI. NMR Biomed. 2014;27:907‐917. doi: 10.1002/nbm.3134 [DOI] [PubMed] [Google Scholar]
- 9. Voskrebenzev A, Gutberlet M, Klimeš F, et al. Feasibility of quantitative regional ventilation and perfusion mapping with phase‐resolved functional lung (PREFUL) MRI in healthy volunteers and COPD, CTEPH, and CF patients. Magn Reson Med. 2018;79:2306‐2314. doi: 10.1002/mrm.26893 [DOI] [PubMed] [Google Scholar]
- 10. Bauman G, Bieri O. Matrix pencil decomposition of time‐resolved proton MRI for robust and improved assessment of pulmonary ventilation and perfusion. Magn Reson Med. 2017;77:336‐342. doi: 10.1002/mrm.26096 [DOI] [PubMed] [Google Scholar]
- 11. Klimeš F, Voskrebenzev A, Gutberlet M, et al. 3D phase‐resolved functional lung ventilation MR imaging in healthy volunteers and patients with chronic pulmonary disease. Magn Reson Med. 2021;85:912‐925. doi: 10.1002/mrm.28482 [DOI] [PubMed] [Google Scholar]
- 12. Plummer JW, Emami K, Dummer A, Woods JC, Walkup LL, Cleveland ZI. A semi‐empirical model to optimize continuous‐flow hyperpolarized 129Xe production under practical cryogenic‐accumulation conditions. J Magn Reson. 2020;320:106845. doi: 10.1016/j.jmr.2020.106845 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Walkup LL, Thomen RP, Akinyi TG, et al. Feasibility, tolerability and safety of pediatric hyperpolarized 129Xe magnetic resonance imaging in healthy volunteers and children with cystic fibrosis. Pediatr Radiol. 2016;46:1651‐1662. doi: 10.1007/s00247-016-3672-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Roach DJ, Willmering MM, Plummer JW, et al. Hyperpolarized 129Xenon MRI ventilation defect quantification via thresholding and linear binning in multiple pulmonary diseases. Acad Radiol. 2022;29:S145‐S155. doi: 10.1016/j.acra.2021.06.017 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Thomen RP, Walkup LL, Roach DJ, et al. Regional structure‐function in cystic fibrosis lung disease using hyperpolarized 129xe and ultrashort echo magnetic resonance imaging. Am J Respir Crit Care Med. 2020;202:290‐292. doi: 10.1164/rccm.202001-0031LE [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Bdaiwi AS, Willmering MM, Plummer JW, et al. 129Xe image processing pipeline: an open‐source, graphical user interface application for the analysis of hyperpolarized 129Xe MRI. Magn Reson Med. 2025;93:1220‐1237. doi: 10.1002/mrm.30347 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Tustison NJ, Avants BB, Cook PA, et al. N4ITK: improved N3 bias correction. IEEE Trans Med Imaging. 2010;29:1310‐1320. doi: 10.1109/TMI.2010.2046908 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Capaldi DPI, Eddy RL, Svenningsen S, et al. Free‐breathing pulmonary MR imaging to quantify regional ventilation. Radiology. 2018;287:693‐704. doi: 10.1148/radiol.2018171993 [DOI] [PubMed] [Google Scholar]
- 19. Ong F, Uecker M, Lustig M. Accelerating non‐Cartesian MRI reconstruction convergence using k‐space preconditioning. IEEE Trans Med Imaging. 2020;39:1646‐1654. doi: 10.1109/TMI.2019.2954121 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Avants BB, Tustison NJ, Song G, Cook PA, Klein A, Gee JC. A reproducible evaluation of ANTs similarity metric performance in brain image registration. Neuroimage. 2011;54:2033‐2044. doi: 10.1016/j.neuroimage.2010.09.025 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Uecker M, Hohage T, Block KT, Frahm J. Image reconstruction by regularized nonlinear inversion—joint estimation of coil sensitivities and image content. Magn Reson Med. 2008;60:674‐682. doi: 10.1002/mrm.21691 [DOI] [PubMed] [Google Scholar]
- 22. Ying L, Sheng J. Joint image reconstruction and sensitivity estimation in SENSE (JSENSE). Magn Reson Med. 2007;57:1196‐1202. doi: 10.1002/mrm.21245 [DOI] [PubMed] [Google Scholar]
- 23. He K, Sun J, Tang X. Guided image filtering. IEEE Trans Pattern Anal Mach Intell. 2013;35:1397‐1409. doi: 10.1109/TPAMI.2012.213 [DOI] [PubMed] [Google Scholar]
- 24. Wernz MM, Voskrebenzev A, Müller RA, et al. Feasibility, repeatability, and correlation to lung function of phase‐resolved functional lung (PREFUL) MRI‐derived pulmonary artery pulse wave velocity measurements. J Magn Reson Imaging. 2024;60:2216‐2228. doi: 10.1002/jmri.29337 [DOI] [PubMed] [Google Scholar]
- 25. Klimeš F, Voskrebenzev A, Gutberlet M, et al. Free‐breathing quantification of regional ventilation derived by phase‐resolved functional lung (PREFUL) MRI. NMR Biomed. 2019;32:e4088. doi: 10.1002/nbm.4088 [DOI] [PubMed] [Google Scholar]
- 26. Lederlin M, Bauman G, Eichinger M, et al. Functional MRI using Fourier decomposition of lung signal: reproducibility of ventilation‐ and perfusion‐weighted imaging in healthy volunteers. Eur J Radiol. 2013;82:1015‐1022. doi: 10.1016/j.ejrad.2012.12.003 [DOI] [PubMed] [Google Scholar]
- 27. Voskrebenzev A, Gutberlet M, Kaireit TF, Wacker F, Vogel‐Claussen J. Low‐pass imaging of dynamic acquisitions (LIDA) with a group‐oriented registration (GOREG) for proton MR imaging of lung ventilation. Magn Reson Med. 2017;78:1496‐1505. doi: 10.1002/mrm.26526 [DOI] [PubMed] [Google Scholar]
- 28. Moher Alsady T, Voskrebenzev A, et al. MRI‐derived regional flow‐volume loop parameters detect early‐stage chronic lung allograft dysfunction. J Magn Reson Imaging. 2019;50:1873‐1882. doi: 10.1002/jmri.26799 [DOI] [PubMed] [Google Scholar]
- 29. Marshall H, Voskrebenzev A, Smith LJ, et al. 129Xe and free‐breathing 1H ventilation MRI in patients with cystic fibrosis: a dual‐center study. J Magn Reson Imaging. 2022;57:1‐14. doi: 10.1002/jmri.28470 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Klimeš F, Obert AJ, Scheller J, et al. Comparison of free‐breathing 3D phase‐resolved functional lung (PREFUL) MRI with dynamic 19F ventilation MRI in patients with obstructive lung disease and healthy volunteers. J Magn Reson Imaging. 2024;60:1416‐1431. doi: 10.1002/jmri.29221 [DOI] [PubMed] [Google Scholar]
- 31. Feng L, Axel L, Chandarana H, Block KT, Sodickson DK, Otazo R. XD‐GRASP: golden‐angle radial MRI with reconstruction of extra motion‐state dimensions using compressed sensing. Magn Reson Med. 2016;75:775‐788. doi: 10.1002/mrm.25665 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Zou Q, Torres LA, Fain SB, Higano NS, Bates AJ, Jacob M. Dynamic imaging using motion‐compensated smoothness regularization on manifolds (MoCo‐SToRM). Phys Med Biol. 2022;67:144001. doi: 10.1088/1361-6560/ac79fc [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Klimeš F, Kern AL, Voskrebenzev A, et al. Free‐breathing 3D phase‐resolved functional lung MRI vs breath‐hold hyperpolarized 129Xe ventilation MRI in patients with chronic obstructive pulmonary disease and healthy volunteers. Eur Radiol. 2024. doi: 10.1007/s00330-024-10893-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34. Munidasa S, Zanette B, Dumas MP, et al. Comparison of 3D UTE free‐breathing lung MRI with hyperpolarized 129Xe MRI in pediatric cystic fibrosis. Magn Reson Med. 2025;93:775‐787. doi: 10.1002/mrm.30299 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. McCallister A, Chung SH, Antonacci M, et al. Comparison of single breath hyperpolarized 129Xe MRI with dynamic 19F MRI in cystic fibrosis lung disease. Magn Reson Med. 2021;85:1028‐1038. doi: 10.1002/mrm.28457 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36. Plummer JW, Hussain R, Bdaiwi AS, et al. A decay‐modeled compressed sensing reconstruction approach for non‐Cartesian hyperpolarized 129Xe MRI. Magn Reson Med. 2024;92:1363‐1375. doi: 10.1002/mrm.30188 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37. Fang Y, Li H, Shen L, et al. Rapid pulmonary 129Xe ventilation MRI of discharged COVID‐19 patients with zigzag sampling. Magn Reson Med. 2024;92:956‐966. doi: 10.1002/mrm.30120 [DOI] [PubMed] [Google Scholar]
- 38. Hallett S, Toro F, Ashurst JV. Physiology, Tidal Volume. StatPearls Publishing; 2024. http://www.ncbi.nlm.nih.gov/pubmed/391255 [PubMed] [Google Scholar]
- 39. Hughes PJC, Smith L, Chan HF, et al. Assessment of the influence of lung inflation state on the quantitative parameters derived from hyperpolarized gas lung ventilation MRI in healthy volunteers. J Appl Physiol. 2019;126:183‐192. doi: 10.1152/japplphysiol.00464.2018 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40. Kaireit TF, Kern A, Voskrebenzev A, et al. Flow volume loop and regional ventilation assessment using phase‐resolved functional lung (PREFUL) MRI: comparison with 129Xenon ventilation MRI and lung function testing. J Magn Reson Imaging. 2021;53:1092‐1105. doi: 10.1002/jmri.27452 [DOI] [PubMed] [Google Scholar]
- 41. Friedlander Y, Munidasa S, Thakar A, et al. Phase‐resolved functional lung (PREFUL) MRI to quantify ventilation: feasibility and physiological relevance in severe asthma. Acad Radiol. 2024;31:3416‐3426. doi: 10.1016/j.acra.2024.01.039 [DOI] [PubMed] [Google Scholar]
- 42. Munidasa S, Zanette B, Couch M, et al. Inter‐ and intravisit repeatability of free‐breathing MRI in pediatric cystic fibrosis lung disease. Magn Reson Med. 2023;89:2048‐2061. doi: 10.1002/mrm.29566 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement