Abstract
Purpose
The single reference variable flip angle sequence with a multi‐echo stack of stars acquisition (SR‐VFA‐SoS) simultaneously measures temperature change using proton resonance frequency (PRF) shift and T1‐based thermometry methods. This work evaluates SR‐VFA‐SoS thermometry in MR‐guided focused ultrasound in an in vivo rabbit model.
Methods
Simultaneous PRF shift thermometry and T1‐based thermometry were obtained in a New Zealand white rabbit model (n = 7) during MR‐guided focused ultrasound surgery using the SR‐VFA‐SoS sequence at 3 T. Distinct locations in muscle (n = 16), fat (n = 12), or the interface of both tissues (n = 23) were heated. The T1‐temperature coefficient of fat was determined using least‐squares fitting of inversion recovery‐based T1 maps of untreated fat harvested from the animal and was applied to the in vivo measured heat‐induced T1 changes to create temperature maps.
Results
Using k‐space weighted image contrast reconstruction, temporal resolution of 1.71 s was achieved for simultaneous thermometry at 1.5 × 1.5 × 2 mm voxel resolution. PRF shift thermometry was not sensitive to heating in fat. T1 changes were observed in fat at the ultrasound focus. The mean T1‐temperature coefficient for fat was determined to be 1.9%/°C ± 0.2%/°C. Precision was 0.76°C ± 0.18°C for PRF shift thermometry in muscle and 1.93°C ± 0.60°C for T1‐based thermometry in fat. Sonications in muscle showed an increase in T1 of 2.4%/°C ± 0.9%/°C.
Conclusion
The SR‐VFA‐SoS sequence was shown to simultaneously measure temperature change using PRF shift and T1‐based methods in an in vivo model, providing thermometry for both aqueous and fat tissues.
Keywords: focused ultrasound, proton resonance frequency, T1 , thermometry
1. INTRODUCTION
MR‐guided focused ultrasound surgery (MRgFUS) is a noninvasive treatment used in numerous clinical applications. 1 , 2 , 3 , 4 , 5 , 6 , 7 , 8 For thermal applications, a significant advantage of MR‐guidance is the ability to use magnetic resonance temperature imaging (MRTI) to monitor internal temperature change during treatment, providing accurate thermometry without the need for insertion of temperature probes. MRTI can allow for real‐time temperature mapping during thermal procedures. 9 , 10 , 11 , 12 , 13 , 14 Both 2D 15 , 16 , 17 and 3D 18 , 19 , 20 acquisitions have been used in conjunction with focused ultrasound. Necessary spatial and temporal resolutions are dependent on the ultrasound focal size and exposure time, respectively, 21 , 22 and are influenced by the sequence parameters, available SNR, and clinical need. 23 , 24 Generally, spatial resolutions are on the order of a few millimeters 14 , 25 , 26 , 27 , 28 and temporal resolutions are on the order of seconds 15 , 16 , 29 , 30 , 31 dependent on application. The ultrasound focus position as well as near‐ and far‐field regions is monitored for heating to maintain patient safety and treatment efficacy. Thermometry for MRgFUS thermal therapy in tissue volumes comprised of both aqueous tissue and fat, however, is challenging, as different thermometry methods are best suited for each tissue type.
Although many MR properties are sensitive to temperature, MRTI is most accurately and precisely achieved using the proton resonance frequency (PRF) shift method. PRF MRTI has been successfully used to monitor thermal changes in aqueous tissues in real time with clinically appropriate spatial and temporal resolution. 32 , 33 Indeed, all United States Food and Drug Administration‐approved MRgFUS systems use PRF MRTI to monitor the temperature change during treatment. 13 , 34 The PRF shift method relies on temperature‐dependent changes in the electronic shielding of the nucleus in hydrogen atoms in water molecules. This change in shielding changes the local magnetic field of the protons, which in turn alters the protons' resonance frequency. This can be described by Eq. (1):
| (1) |
where B loc is the local magnetic field of the proton, B0 is the main magnetic field of the scanner, α is the PRF coefficient in ppm/°C, and T is the tissue temperature. The PRF coefficient has been found to be approximately −0.010 ppm/°C for many aqueous tissues over the range of temperatures of interest in thermal therapies. 35 This temperature‐dependent change in resonance frequency is quantified in MRTI by subtracting phase data acquired from a baseline image from the phase data in images acquired during heating, thereby removing non‐temperature induced phase changes. Because fat tissue does not have the same hydrogen bonding as that experienced by water molecules, the PRF coefficient of fat is negligibly small (approximately −0.00018 ppm/°C 36 ) and the PRF shift method of MRTI is, therefore, mostly insensitive to temperature changes in fat. 37 To date, no MR‐based thermometry method has been universally adopted for fat during thermal therapy treatments.
Other MRTI methods, which can measure temperature change in fat exist, but they are currently not used clinically in thermal therapy treatments. For example, the MR relaxation time constant T1 has been shown to be dependent on temperature, following Eq. (2):
| (2) |
where T is the dynamic temperature, T ref is a reference starting temperature, and dT1/dT is the T1‐temperature coefficient. Measurements of T1 over the range 10°C to 70°C have shown a near linear relationship 38 with a temperature dependence on the order of 1%/°C to 3%/°C 39 in many biological tissues, including fat. Several methods exist 40 , 41 , 42 , 43 for quantifying T1, and several T1‐based MRTI methods have been developed 44 , 45 , 46 allowing for temperature mapping of fat and other non‐aqueous tissues. One significant drawback to T1‐based MRTI techniques is that the T1‐temperature coefficient is tissue type specific and must be known for accurate temperature mapping, potentially requiring a priori information, including tissue specific calibration values.
Many hybrid techniques 47 , 48 , 49 , 50 , 51 , 52 , 53 that simultaneously measure PRF and T1 change have been investigated that could be used for MRTI of heterogeneous tissue volumes, but have drawbacks such as low temporal resolution or longer time between T1 measurements compared to PRF measurements, low spatial resolution, required multiple interleaved baseline images, or could only acquire 2D images. A substantial improvement in temporal resolution of hybrid PRF/T1 thermometry was introduced with dual‐ and variable‐flip angle (VFA) approaches; however, multiple acquisitions are required for T1 measurements using these methods. 50 , 51 , 52 , 53 , 54 As a result, T1 thermometry measurements were obtained 2 to 4 times less frequently than PRF measurements, limiting the temporal resolution of T1 to the order of 6 to 15 s for 3D acquisitions. 51 , 54 MRgFUS and other MR‐guided thermal therapies, which generate ablative temperatures within 20 to 40 s 14 , 55 and assess clinical efficacy with time‐dependent dose metrics, require near real‐time monitoring. No hybrid PRF/T1 MRTI techniques have yet been adopted clinically for thermally ablative focused ultrasound surgery (FUS) therapies.
Svedin et al 56 developed a more efficient single reference VFA (SR‐VFA) method with a multi‐echo stack‐of‐stars (SoS, in‐plane radial with through‐plane Cartesian sampling) acquisition and evaluated their simultaneous PRF/T1 approach during MRgFUS ablations in phantoms and cadaver breasts, and in healthy volunteers under no heating conditions. This method calculates the PRF shift at every time point using the image phase. To calculate ΔT1, this method uses sets of images, with baseline images acquired at a lower flip angle and dynamic images acquired at a higher flip angle. The change in T1 is calculated as shown in Eq. (3):
| (3) |
where TR is the repetition time, α and β are the lower and higher flip angles, respectively, , and E 1est is an estimated E1 obtained from baseline and dynamic images and the two flip angles (α and β). The reader is pointed to Svedin et al 32 for the full derivation. Advantages of this method over prior work include that the steady‐state signal remains stable between dynamics because the flip angle is not changed, and therefore, the temporal resolution of the PRF shift and T1 measurements are the same. The sequence achieves good spatial precision of T1 measurements in human breast tissue under non‐heating in vivo conditions; however, the temperature precision was not reported because T1‐temperature coefficient was not determined. Malmberg et al 57 assessed the precision and accuracy of the SR‐VFA‐SoS method for T1‐mapping at various TEs (1.24–7.55 ms) in dynamically variable T1 phantoms in static and dynamic conditions. Zhang et al 54 has reported similar performance of a hybrid VFA sequence using an efficient golden‐angle–ordered 3D SoS technique in vivo in non‐heated human prostate, breast, and brain and ex vivo in FUS ablated porcine fat and muscle.
The purpose of the presented work is to evaluate this hybrid PRF/T1 sequence in an in vivo MRgFUS ablation treatment setting to demonstrate the ability to simultaneously measure 3D PRF and T1 changes with spatial and temporal resolution of 1.5 × 1.5 × 2 mm and 1.71 s, respectively. A large animal rabbit model was used to evaluate the sequence. The temperature sensitivity of T1 from adipose tissues from both the hamstring and dorsum were found from water‐bath calibration experiments and used to convert ΔT1 to temperature change and assess the temperature precision of this method.
2. METHODS
2.1. In vivo procedure
All animal studies were carried out under the guidance of the local Institutional Animal Care and Use Committee. In vivo experiments were performed in a 3 T MRI scanner (PrismaFIT, Siemens Healthcare) using a 256 element MRI‐compatible phased‐array transducer, 10 cm radius of curvature, f = 1 MHz, 1.8 × 2.5 × 10.9 mm FWHM of pressure profile as measured in water, (Image Guided Therapy and Imasonic). Female and male New Zealand white rabbits (n = 7) were anesthetized with ketamine and xylazine (25 and 5 mg/kg, intramuscular injection, respectively), intubated, and maintained with isoflurane (1%–3%, inhaled, 100% oxygen). Each animal's leg or back was shaved before treatment and a depilatory agent (Nair, Church and Dwight) was applied to remove any remaining hair.
Initial animals (n = 2) were positioned lateral decubitus with the transducer focus approximately centered in the hamstring; later animals (n = 5), were positioned supine with the transducer geometric focus at the interface of the perirenal fat and paraspinal muscle to treat a larger contiguous volume of fat. The treatment area was acoustically coupled to the transducer with degassed and deionized water (Figure 1). The FUS (22.5–30 s, 45.9–49.6 W acoustic power for animals 3 and 5, 18.0–32.3 W for other animals) was electronically steered to 4 to 11 distinct locations per animal, targeting muscle only (n = 16 total for all subjects), fat only (n = 12), and the fat/muscle interface (n = 23), with the number of sonications in each location for each animal shown in Table 1. Interface sonications were defined to be sonications where the peak T1 change in the ultrasound focus was located within 4 mm of the muscle/fat interface.
FIGURE 1.

Rabbit MR‐guided focused ultrasound surgery (MRgFUS) system. Experimental structure includes a single‐loop imaging coil (A), focused ultrasound transducer housing (B) with multi‐axis adjustable positioning, water coupling reservoir (C), positioning coils (D), and animal platform (E). White arrows indicate specific locations of identified components. Animals were positioned such that the focal zone of the transducer was located either in the hamstring (lower left) or dorsum (lower right), then electronic steering was used to target specific sites during the MRgFUS ablation procedure. Ultrasound transducer face outlined in green. Focal zone outlined in blue. Approximate single reference‐variable flip angle‐stack of stars (SR‐VFA‐SoS) volume outlined in orange. Muscle and fat indicated by red and yellow arrows, respectively. Purple line indicates approximate slice location for images in Figure 6.
TABLE 1.
Subject‐specific treatment parameters and MR temperature imaging results.
| Animal | Sonication location | FUS duration (s) | Successful breath holds | Initial muscle T1 (ms) | Initial fat T1 (ms) | Muscle T1‐temperature coefficient (%/°C) | Fat T1‐temperature coefficient (%/°C) | Precision | |||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Muscle | Fat | Interface | PRF in muscle (°C) | T1 in fat (°C) | |||||||
| 1 | 0 | 0 | 3 | 30 | N/A | 779 | 389 | N/A | 1.67 | 0.49 ± 0.10 | 2.89 ± 0.89 |
| 2 | 2 | 0 | 3 | 22.5 | N/A | 929 | 426 | 2.59 | 2.25 | 0.67 ± 0.13 | 1.64 ± 0.31 |
| 3 | 3 | 2 | 4 | 22.5 | 9/9 | 876 | 377 | 1.26 | 1.79 | 1.19 ± 0.24 | 1.83 ± 0.33 |
| 4 | 1 | 3 | 2 | 22.5 | 6/6 | 918 | 382 | 1.91 | a | 0.87 ± 0.09 | 1.06 ± 0.19 |
| 5 | 3 | 0 | 4 | 25.0 | 0/7 | 875 | 338 | 4.14 | 1.72 | 0.71 ± 0.30 | 2.29 ± 0.71 |
| 6 | 4 | 4 | 4 | 22.5 | 12/12 | 843 | 394 | 2.40 | 1.84 | 0.73 ± 0.12 | 1.86 ± 0.68 |
| 7 | 3 | 3 | 3 | 22.5 | 6/9 | 956 | 364 | 2.25 | a | 0.64 ± 0.16 | 1.98 ± 0.71 |
| Average | N/A | N/A | N/A | N/A | N/A | 882 | 381 | 2.43 | 1.85 | 0.76 ± 0.18 | 1.93 ± 0.60 |
Note: Sonications with peak heating within 4 mm of the tissue interface were classified as interface sonications. Number of sonications with successful breath holds relative to the total number of sonications is shown. Subject‐specific measured T1‐temperature coefficients were used to convert ΔT1 to temperature change. Initial muscle and fat T1 values were calculated via inversion recovery images acquired before treatment. A 5 × 5 × 1 voxel region near the heating was selected and the mean T1 value of the region is reported.
Abbreviations: FUS, focused ultrasound surgery; PRF, proton resonance frequency.
T1‐temperature coefficient not calculated, average value of 1.85%/°C used to determine T1‐based temperatures.
For temperature imaging, PRF shift and ΔT1 data were acquired with a custom single‐loop coil during sonications using the SR‐VFA approach with a SoS sequence with a multi‐echo pseudo‐golden angle acquisition with view angle increment of (1–233/377) × 180° and 377 projections per dynamic image (additional parameters found in Table 2). A baseline low flip angle (5°) image and a baseline high flip angle image (15°) were acquired before each sonication. Because a k‐space weighted image contrast (KWIC) algorithm was used in reconstruction as described in Song et al, 58 an image at each flip angle was taken before the respective baseline image to “prefill” k‐space. Immediately following the “prefill” and baseline images, 1 to 3 total higher flip angle dynamic acquisitions were acquired in succession, with the focused ultrasound heating beginning at the start of the first dynamic acquisition (Figure 2A). The acquisition time for each image (377 views) was 1 min and tissues were allowed to cool for ˜5.5 min between sonications.
TABLE 2.
MRI sequence parameters.
| TR (ms) | TE (ms) | BW (Hz/pixel) | FA (°) | Matrix size | Slices | Voxel size (mm3) | ETL | Acquisition time (s) | |
|---|---|---|---|---|---|---|---|---|---|
| SR‐VFA‐SoS | 20 | 2.46–9.06 | 1500 | 5,15 | 128 × 128 | 12 | 1.5 × 1.5 × 2 | 6 | 60.3 |
| 3D T1‐w VIBE Dixon | 13.7 | 3.7, 9.84 | 250 | 15 | 512 × 384 | 104 | 0.5 isotropic | 2 | 99 |
| 3D T2‐w SPACE | 2000 | 151 | 700 | 120 | 512 × 344 | 72 | 0.5 isotropic | 106 | 282 |
| TurboFLASH | 5700 | 1.72 | 1184 | 12 | 128 × 104 | 11 | 1.5 × 1.5 × 4 | 1 | 5.4 |
| TSE | 8000 | 7.3 | 1000 | 180 | 256 × 208 | 11 | 0.75 × 0.75 × 2 | 14 | 104 |
Abbreviations: TR, repetition time; TE, echo time; BW, bandwidth; ETL, echo train length; FA, flip angle; SPACE, sampling perfection with application optimized contrasts using different flip angle evolution; SR‐VFA‐SoS, single reference‐variable flip angle‐stack of stars; TSE, turbo spin echo; VIBE, volumetric interpolated breath‐hold examination.
FIGURE 2.

Sequence acquisition order [e.g., k‐space weighted image contrast (KWIC) reconstruction and resultant magnitude and phase images]. (A) Two 5° flip angle images (377 full views each) were acquired followed by three 15° flip angle images. Focused ultrasound was applied during the third high flip angle image. Animal breath holds were attempted on the baseline and dynamic acquisitions, dependent on animal tolerance. A KWIC reconstruction algorithm was used to reconstruct a dynamic image sequence with high temporal resolution. The low and high flip angle KWIC images before focused ultrasound surgery (FUS) heating are used as the low flip angle and baseline images for the corresponding images during FUS heating. (B) An example of fully sampled k‐space with radial projections, (C) an example asymmetric KWIC window, and (D) an example of k‐space sampled with three center and 13 total projections after the KWIC window was applied. Actual single reference‐variable flip angle‐stack of stars (SR‐VFA‐SoS) measurements was taken with 13 center and 377 total projections. Representative (E) magnitude (single dynamic averaged across echoes) and (F) phase (single dynamic at the first echo time) images after reconstruction with the KWIC algorithm.
For animals treated in the perirenal fat/paraspinal muscle, 20‐s breath holds were attempted during the 5° and 15° baseline acquisitions and the dynamic acquisition acquired during heating for matching during reconstruction. To avoid excessive CO2 buildup, breath holds were not implemented if post‐sonication end tidal CO2 did not return to baseline levels or if irregular respiration patterns were observed before beginning imaging for each sonication. Breath holds, assessed by respiratory monitoring, were implemented and successful for all sonications in three animals, no sonications in one animal, and in six of nine sonications in one animal.
B1 maps were acquired using a vendor‐supplied TurboFLASH‐based B1‐mapping sequence before each treatment. Three dimensional T1‐weighted volumetric interpolated breath‐hold examination Dixon and 3D T2‐weighted sampling perfection with application optimized contrasts using different flip angle evolution images were captured before and after MRgFUS ablation, with additional contrast‐enhanced T1‐weighted imaging (gadoteridol, IV 0.1 mmol/kg) post‐ablation.
Animals were euthanized and tissue immediately harvested following post‐ablation imaging.
2.2. SR‐VFA‐SoS reconstruction
All reconstruction was performed using MATLAB 2022b (The MathWorks) on an Intel Xeon Silver 4210 CPU (2.20 GHz). Reconstruction of the SR‐VFA‐SoS images was performed using a KWIC method 58 for increased temporal resolution, with 13 central projections and 377 total projections per dynamic using a sliding asymmetric window to leverage oversampling at the center of radial projections (Figure 2B–D), advancing 13 projections between KWIC frames. Echo combination was done by weighting the individual echoes by (where is the complex MRI signal) for PRF shift 59 thermometry and by normalized by (where j is the echo number and N e is the total number of echoes) for ΔT1‐based 53 thermometry calculations. If available, breath‐held baseline and dynamic acquisitions were paired in the ΔT1 calculation to reduce motion artifacts. Temporal resolution for both the T1 and PRF shift temperature data was 1.71 s. Temperature and ΔT1 data were zero‐filled interpolated to 0.75 × 0.75 × 2 mm voxel spacing to reduce partial volume artifacts. All temperature reconstruction was performed retrospectively.
All VFA approaches, including the SR‐VFA‐SoS sequence, are highly sensitive to inhomogeneities in the B1 field. In this study, B1 inhomogeneities were corrected by acquiring a series of 2D B1 maps before the treatment and interleaving them to have B1 maps corresponding to all slice locations for the SR‐VFA‐SoS sequence. This B1 correction was used in all calculations.
2.3. Fat ex vivo T1 ‐temperature coefficient measurement
The fat T1‐temperature coefficient was determined ex vivo by measuring the T1 of excised non‐ablated fat from the contralateral side of the animal at several discrete temperature values between 20°C and 65°C. Immediately after animal euthanasia, untreated fat was excised, sliced into ˜1 cm3 segments, and maintained in saline. Within 24 h of excision, fat was packed into seven vials (ø1.4, length = 5.7 cm) of an MR‐compatible custom circulating water bath apparatus (Figure 3A) containing a single‐loop RF imaging coil. Fiber optic temperature sensors (polytetrafluoroethylene Teflon, range −80°C to 250°C, accuracy = ±0.2°C, Qualitrol) were inserted into the center of four fat‐filled vials for accurate temperature monitoring. The circulating water bath apparatus was placed in a 3 T MRI scanner (PrismaFIT). Water was brought to a desired temperature by heating in a heated water pump placed outside the scanner suite, then pumped through the circulating water bath apparatus, heating the fat‐filled vials (Figure 3B). The temperature at the center of the fat‐filled vials was monitored using the inserted temperature sensors. After the temperature sensors in the heated fat reported an approximate constant temperature for 5 min, the circulating heated water was diverted through a bypass line, ceasing water flow through the water bath apparatus. Water circulating through the bypass line continued heating to the next desired temperature. This allowed for heating the water during imaging without transferring heat to the fat or incurring excessive water motion artifacts in the images, increasing the efficiency of the process and the quality of the T1 maps. After a 5‐min waiting period from diverting water flow through the bypass line (to allow for additional temperature equalization of the fat and for water motion inside the apparatus to subside), Turbo spin‐echo (TSE) inversion recovery (IR) images were acquired (TI = 25, 100, 200, 500, 1000, 2000, and 3000 ms) (Table 2) at the center of the vials, with temperature readings from the inserted thermocouples recorded at 1 Hz. After imaging, the bypass line was closed, allowing the fat‐filled vials to be heated to the next temperature. This process was repeated for 5 to 6 increasing temperatures in the 20°C to 65°C range per sample. The ex vivo T1‐temperature coefficient measurement was only completed for five of the seven rabbits because of inadequate contralateral fat volume (animal 4) and data lost during transfer from the scanner (animal 7).
FIGURE 3.

Experimental setup for ex vivo fat calibration. (A) Seven fat‐packed vials were submerged in a water bath and inserted with thermocouples. A single loop imaging coil encircles the water bath. (B) Water was heated external to the scanner, and then pumped through the water bath until fat reached a specified temperature. Water was diverted through a bypass line during imaging, which allowed for continued heating of water without heating fat or inducing water motion artifacts while inversion recovery images were acquired. This process was repeated for the temperature range 20°C to 65°C.
2.4. T1 ‐temperature coefficient calculation
The muscle T1‐temperature coefficient was estimated using the MRgFUS muscle sonications using voxels in a 2.25 × 2.25 × 2 mm3 (3 × 3 × 1 voxel) region of interest centered at each sonication's peak PRF‐shift temperature rise and ΔT1 for all muscle sonications. Linear regression analysis of PRF‐shift temperature and T1 changes for each voxel in the region of interest for all time points was used to calculate the estimated muscle T1‐temperature coefficient.
The fat T1‐temperature coefficient was determined from the IR images taken during the ex vivo T1‐temperature coefficient procedure. 60 Digital masks were made, which only contained voxels interior to the vials in the reconstructed images (Figure 4A). For each vial, voxels whose T1 residual values were determined to be outliers via MATLAB's isoutlier function [outliers calculated as values more than three scaled median absolute deviations (MAD) from the median, where , and where , is the residual vector, and erfc−1 is the inverse complementary error function] were also masked out, retaining 80.0% ± 12.3% of voxels. For each rabbit, a T1‐temperature coefficient was calculated using reduced dimension nonlinear least squares fitting 61 using the resultant T1 data and the measured probe temperature.
FIGURE 4.

Ex vivo fat calibration results. Left: Axial view 3D volumetric interpolated breath‐hold examination (VIBE) of interior of fat calibration setup. Approximate masks for vial fat T1 calculation shown (dashed yellow). Right: Measured fat T1‐temperature curves derived from inversion recovery data. Average measured T1 values for vials of fat at indicated temperatures with standard error between vials shown as bars. The average T1‐temperature coefficient was 7.3 ± 0.7 ms/°C or 1.9%/°C ± 0.2%/°C for the excised fat.
2.5. PRF temperature and T1 precision measurement
The precision of the measured PRF shift‐based temperature and the measured ΔT1 by the SR‐VFA‐SoS sequence was determined for each acquired image using an unheated region of tissue. For the PRF‐shift temperature and ΔT1‐based temperature precision calculations, temperature data from an untreated region of interest in muscle or fat (3.75 × 3.75 × 2 mm3 [5 × 5 × 1 voxels]) was analyzed. For each voxel in the region, the SD of temperature through time was calculated, and then the calculated SD of all voxels in the region was averaged to determine the reported precision value.
To determine if fat may have changed phase during heating, an additional animal was acquired and a slip test 62 was performed. Perirenal fat from both sides of the animal was harvested immediately following euthanasia. Fat was then slowly heated at ˜1°C/min in a beaker on a hot plate until melting occurred, then liquid fat was drawn into three capillary pipettes and immediately solidified by cooling in a −7°C freezer for 1 min. The outer edges of the fat were marked, and then the capillary pipettes were submerged in a water bath and heated at 1°C/min. The fat volumes were carefully observed during heating and the temperature of the water bath was recorded the moment the fat slipped beyond the marks.
3. RESULTS
3.1. T1 ‐temperature coefficient
Fat T1‐temperature calibration experiments demonstrated an average increase of 7.3 ± 0.7 ms/°C in excised fat or approximately a 1.85% ± 0.23% change in T1 per (°C) change for the range of 20°C to 65°C. T1‐temperature curves for each animal are plotted in Figure 4. The error bars represent standard error in mean T1 values between fat‐filled vials.
The in vivo muscle T1‐temperature coefficient was calculated to be 2.59, 1.26, 1.91, 4.14, 2.40, and 2.25%/°C for animals 2 to 7, respectively, with an average of 2.43% ± 0.96% increase in T1 per (°C) above subject body temperature. Correlation between PRF‐shift temperature and ΔT1 data is shown in Figure 5.
FIGURE 5.

Correlation of proton resonance frequency (PRF) shift temperature and ΔT1 in muscle. The muscle T1‐temperature coefficient was estimated for each animal using sonications centered in muscle. The points of maximum heating and maximum T1 change were identified, and linear regression analysis was performed using a 2.25 × 2.25 × 2 mm3 (3 × 3 × 1 voxel) region of interest centered at those voxels for all time points. The animal‐specific linear fit is shown in orange. The average T1‐temperature coefficient in muscle was 2.4%/°C ± 0.9%/°C.
3.2. In vivo temperature
Simultaneous PRF and T1 temperature changes at or near the focal spot were successfully measured for all sonications in muscle. Peak temperature rises from 9°C to 39°C and peak changes of T1 from 39% to 181% were observed. Peak changes in T1 for sonications with the focal spot located in fat ranged from 18% to 73%, corresponding to peak temperature rise of 10°C to 41°C. ΔT1‐based temperatures were calculated with subject‐specific T1‐temperature coefficients when available and the intersubject mean for rabbits without a subject‐specific measured T1‐temperature coefficient. Temperature maps at peak‐temperature time points for PRF shift‐based thermometry for a muscle sonication, ΔT1‐based thermometry for a fat sonication, and PRF shift‐ and ΔT1‐based thermometry for an interface sonication are shown in Figure 6. Qualitatively, the PRF shift‐based temperatures contained less noise than the ΔT1‐based temperature maps. These observations are confirmed quantitatively with the measurement precision values detailed in Table 1. The average PRF shift‐based temperature precision was 0.76°C ± 0.18°C (0.60°C ± 0.12°C for hamstring sonications, 0.86°C ± 0.16°C for successful breath‐hold sonications, and 0.72°C ± 0.27°C for sonications in the back without breath holds). The average ΔT1 precision was 13.5 ± 4.0 ms, resulting in an average ΔT1‐based temperature precision of 1.93°C ± 0.60°C (2.11°C ± 0.6°C for hamstring sonications, 1.69°C ± 0.54°C for successful breath‐hold sonications, and 2.30°C ± 0.73°C for sonications in the back without breath holds). Figure 7 shows temperature and ΔT1 versus time plots for the muscle and fat sonications.
FIGURE 6.

Representative single reference‐variable flip angle‐stack of stars (SR‐VFA‐SoS) thermometry maps. (A) Proton resonance frequency (PRF) shift calculated temperature map for focused ultrasound surgery (FUS) sonication in muscle, (B) calculated temperature map from ΔT1 for FUS sonication in fat, and temperature maps of (C) PRF and (D) ΔT1 for a single sonication targeting the interface. Partial volume effects and susceptibility artifacts likely lead to inaccuracies in PRF temperatures near the interface. For the interface ΔT1 image, temperatures are only valid for fat, because the ΔT1 was scaled by the subject‐specific fat ΔT1‐temperature coefficient. Voxels at the fat‐muscle interface were noisier than single‐tissue voxels, including some over‐saturated voxels with non‐physiological temperature rises. For inset images, the background image was removed to accurately display the temperature color scale. Arrows denote muscle tissue (red) and fat tissue (yellow). Temperature maps are at the time of maximum focal spot temperature. Thermometry maps are from different sonications of animal 3.
FIGURE 7.

Representative temperature versus time plots for proton resonance frequency (PRF)‐ and ΔT1‐based thermometry. Temperatures shown are the average temperature of a 3 × 3 in‐plane set of voxels centered at the peak temperature voxel. PRF shift‐based temperature (blue) and ΔT1 (orange) shown for a sonication centered in muscle (left). PRF shift‐based (blue) and ΔT1‐based (orange) temperatures shown for a sonication in fat (right). Temperature data taken from the muscle and fat sonications displayed in Figure 6.
Peak temperatures versus applied acoustic powers are shown in Figure 8 for sonications centered in muscle and in fat, with bars showing the precision values for the individual sonication. Peak temperatures normalized by acoustic power averaged 0.57°C/W ± 0.14°C/W in muscle and 0.77°C/W ± 0.28°C/W in fat.
FIGURE 8.

Applied acoustic power versus peak temperature for sonications centered in muscle (left) and fat (right). Bars show measured temperature precision of each sonication as derived from a non‐heated region of tissue. Average peak temperatures normalized by acoustic power were 0.57°C/W ± 0.14°C/W for sonications in muscle and 0.77°C/W ± 0.28°C/W for sonications in fat.
During the slip test for determining melting point, melting occurred at 51°C for all three fat volumes.
4. DISCUSSION
This work demonstrates the feasibility of using the SR‐VFA‐SoS method for simultaneous T1 and PRF shift MR temperature measurement in vivo. Current United States Food and Drug Administration approved MRgFUS applications acquire MRTI using the PRF method, resulting in no fat thermometry being available. This method of hybrid thermometry allows for the measuring of temperature change across a volume of tissue containing both aqueous tissue and fat, which will be of great value for applications in, for example, the breast and abdomen. The high spatial and temporal resolutions are similar to existing clinical MRgFUS ablation and hyperthermia PRF shift thermometry methods. 34 This work also determined the T1‐temperature coefficient in rabbit perirenal fat and assessed the variability in the T1‐temperature coefficient between subjects.
Temperature changes calculated from the PRF shift method were observed in aqueous tissue only, with sonications in muscle showing a clear temperature rise at the ultrasound focal spot. PRF shift thermometry did not show heating in fat sonications, as expected. Relative T1 changes were observed in the ultrasound focal region in both muscle and fat. The T1 changes in fat were converted to temperature using an experimentally derived T1‐temperature coefficient (Figure 4B). When normalized by applied acoustic power, calculated temperatures were higher for sonications in fat (0.77°C/W ± 0.28°C/W) than in muscle (0.57°C/W ± 0.14°C/W), although the acoustic attenuation of fat is lower than in muscle (4.4 vs. 7.1 Np/m at 1 MHz, respectively). 63 Perfusion in muscle is higher than fat 64 and has been shown to increase with hyperthermia, 65 , 66 which can result in lower focal tissue heating, especially at high perfusion values. 67
The T1‐temperature coefficient in muscle was estimated by using PRF shift derived temperatures in conjunction with the measured T1 change at the focal spot in muscle sonications. The relationship of T1 changes versus temperature rise for maximum heated voxels in muscle showed increases of 1.26%/°C to 4.14%/°C, in line with the 1.4%/°C value reported in ex vivo bovine muscle at 20°C, 68 with the 4.14%/°C coefficient for Animal 5 being an outlier. Additionally, the measured in vivo T1 for fat without heating for animal 5 was 338 ms, whereas the ex vivo T1 measurement was 450 ms at 37°C. As shown in Figure 5, animals 3 and 5 had notably higher variance in temperature rise versus ΔT1 than other animals, which correlates with the higher sonication power used in these animals.
In this study, an increase of 7.3 ± 0.7 ms/°C or 1.85%/°C ± 0.23%/°C was calculated for the T1 of perirenal rabbit fat using the IR measurements during fat heating, which is higher than the 5.8 ± 1.9 ms/°C and 5.8 ± 2.1 ms/°C reported by Hey et al 50 and lower than the 7.78 ± 0.19 ms/°C and 8.11 ± 0.25 reported by Zhang et al 54 for ex vivo porcine fat. It is also higher than the average 1.4% increase per (°C) of temperature rise for ex vivo human breast fat reported by Todd et al, 52 but still falls within the 1%/°C 39 to 3%/°C 39 seen in many biological tissues. Because of the observed increase in standard error of T1 at higher temperatures, particularly above 50°C, a slip test 62 was used to determine the melting point of perirenal fat in an additional animal, as phase‐changing fat could affect the measured T1 values. Melting of the fat occurred at 51°C, which may indicate the higher standard error seen at higher temperatures in Figure 4 is because of phase‐changes during the study.
The average calculated precision for the PRF shift‐based temperatures was 0.76°C ± 0.18°C, similar to precision values reported for other PRF methods. In MRgFUS using a porcine model, Quesson et al 9 reported a SD of less than 2°C in 75% of voxels in liver, with some voxels exceeding a SD of 4°C, and a SD of less than 2.6°C in 90% of voxels in skin using a multi‐slice gradient‐recalled EPI sequence, with some voxels reaching a maximum SD of 38°C. Odéen and Parker 59 compared PRF shift‐based MR thermometry sequences for laser interstitial thermotherapy and reported precision values for 0.94 × 1.88 × 3 mm3 voxels of 0.91°C to 3.33°C for a 3D gradient echo (acquisition time = 5.89 s) sequence and 0.37°C to 1.79°C a 3D EPI (acquisition time = 5.85 s) sequence depending on coil selection for in vivo non‐heating measurements in human brains at 3 T. Adams‐Tew et al 19 reported precisions of 1.1°C and 2.0°C for a 3D EPI sequence (acquisition time = 4.88 s) with voxel size of 1 × 2 × 3 mm3 in non‐heating measurements in human breast with and without drift correction, 69 respectively.
The average calculated precision for the ΔT1‐based temperatures at 1.93°C ± 0.60°C was slightly worse than PRF precision. Breath holds improved sonication precision measurements over free‐breathing sonications (1.69°C ± 0.54°C and 2.30°C ± 0.73°C, respectively), likely because of reduced animal motion as evidenced by less noise in non‐heated regions of the tissue used to calculate precision. Temperature precision is also dependent on the KWIC window parameters, such as the number and portions of projections included in the window. Additionally, low SNR from the use of a body coil for imaging, as is common clinically, 70 may result in poor precision, but SNR can be improved using higher field strengths, specialized coils, 71 MR 72 or ultrasound 73 hardware changes, or optimizing sequence parameters. 74
In vivo accuracy of the SR‐VFA‐SoS sequence was not assessed in this work, largely because of technical challenges of the viscous heating artifacts induced by available MR‐compatible temperature sensors. 19 Using a hybrid VFA SoS sequence, Zhang et al 54 reported mean absolute difference between temperature probes and PRF temperatures were 1.89°C and 1.23°C in ex vivo muscle using one transducer and 0.9°C in ex vivo muscle using another. Malmberg et al 57 used a novel quantitative rotating phantom to assess the accuracy of T1 maps acquired using the SR‐VFA‐SoS sequence with and without T2* correction. 75 Correcting for T2* effects improved both the accuracy and precision of the SR‐VFA‐SoS sequence, with greater improvement seen at longer TEs. Future work will include in vivo validation of both the PRF shift and T2* corrected T1‐based temperatures using temperature probes currently undergoing validation for use in focused ultrasound fields.
A significant advantage of the SR‐VFA‐SoS sequence is its ability to produce 3D thermometry maps with both high spatial and temporal resolution. Three dimensional thermometry in MRgFUS allows for better focal spot localization and monitoring of the near‐ and far‐field, as well as reduced partial volume artifacts resulting in, for example, more accurate determination of accumulated thermal dose. It also overcomes FOV and systemic errors seen in Todd et al 52 because of the 2D sequence used. The acquired spatial resolution of 1.5 × 1.5 × 2, interpolated to 0.75 × 0.75 × 2 mm spacing, in this study allowed for sufficient determination of the size and position of localized heating and T1 changes based on the ultrasound transducer FWHM. 27 Increasing acquisition voxel sizes would lead to increased SNR, which can increase the precision of the thermometry, 76 but could also negatively impact the peak‐temperature accuracy because of increased partial volume effects. 27
Acquiring both PRF shift and T1 data with one sequence and using a single reference image eliminates the need for interleaving sequences or alternating flip angles as done in multiple previous studies 49 , 50 , 51 , 52 and hence, allows for increased temporal resolution. Using KWIC we achieved an in vivo temporal resolution of 1.71 s, higher than in other hybrid techniques. This high temporal resolution provides more accurate treatment monitoring and allows better visualization of the temperature rise and cooling in the treatment volume.
Although we found the T1‐temperature coefficient varied from animal to animal, Baron et al 77 showed that this is not the case for human breast tissue. They measured a mono‐exponential T1‐temperature relationship for ex vivo human breast fat and found a low deviation, less than 0.5 ms/°C, between intersubject breast fat measurements indicating that a table value T1‐temperature coefficient for breast fat is likely sufficient for clinical use and subject‐specific T1‐temperature coefficient measurement may not be necessary. Thermal effects may also include altering tissue characteristics, potentially changing the T1‐temperature coefficient 78 , 79 , 80 over the course of a MRgFUS treatment.
Another challenge with this method of hybrid thermometry is accurate temperature calculation for mixed‐tissue voxels. Although percent change in T1 was successfully measured for sonications located at the interface of muscle and fat, accurate T1‐based thermometry for mixed‐tissue voxels was not possible because of the tissue‐specific nature of the T1‐temperature coefficients. Additionally, the partial‐volume effects for voxels at the tissue interface can cause under‐ or over‐estimation of the temperature change using the PRF shift method. 81 The partial‐volume effect at the tissue interface is somewhat mitigated by the small voxel size afforded by the SR‐VFA‐SoS sequence. Susceptibility changes near the interface of fat and aqueous tissues also can cause inaccurate PRF measurements, 82 nevertheless, for sonications targeting only aqueous tissues, susceptibility changes are small. 36 , 83
Finally, for clinical use, reconstruction of the SR‐VFA‐SoS sequence needs to take place in real time, with the thermometry available to clinicians during the sonication. For this study, reconstruction of the SR‐VFA‐SoS sequence was done retrospectively, with reconstruction time not considered. Multiple actions can be taken to streamline the reconstruction to reduce the temporal footprint, including parallelizing graphics and/or central processing units for gridding, 84 calculating water‐ and fat‐fraction maps before treatment for segmentation (to apply PRF‐based temperature maps for aqueous tissues and T1‐based maps for fat), and reducing the time required for pre‐ and post‐heating image acquisition. Additionally, the KWIC algorithm can be implemented with an asymmetric or symmetric window. In this case, an asymmetric window was used because a symmetric window (where the central slices of k‐space are located at the center of the window) requires a delay between the acquisition of the center of k‐space and the reconstruction of the temperature maps, although the symmetric window may be more accurate in some cases. 85
5. CONCLUSION
This work has provided an in vivo evaluation of the SR‐VFA‐SoS method of simultaneous PRF shift thermometry and T1 measurement along with the measurement of the T1‐temperature coefficient in a rabbit model. The SR‐VFA‐SoS method has the capability to measure relative temperature in aqueous tissue and T1 changes in nonaqueous tissue, which, when combined with a T1‐temperature calibration curve, allows for the monitoring of relative temperature of fat and aqueous tissues in the treatment area, significantly augmenting thermometry acquired during MRgFUS treatments.
CONFLICT OF INTEREST STATEMENT
The authors report there are no competing interests to declare.
ACKNOWLEDGMENTS
We acknowledge the National Institutes of Health (R37CA224141, S10OD018482, S10OD026788, R21EB 033638, and 5R01EB028316) and the Mark H. Huntsman Endowed Chair. We appreciate Dr. Peter Kovesi for making the perceptually linear 86 color map used in Figure 6 available at colorcet.com. The data that support the findings of this study are available from the corresponding author, N.R., on reasonable request.
Richards N, Malmberg M, Odéen H, et al. In vivo simultaneous proton resonance frequency shift thermometry and single reference variable flip angle T1 measurements. Magn Reson Med. 2025;93(5):2070‐2085. doi: 10.1002/mrm.30413
REFERENCES
- 1. Izadifar Z, Izadifar Z, Chapman D, Babyn P. An introduction to high intensity focused ultrasound: systematic review on principles, devices, and clinical applications. J Clin Med. 2020;9:460. doi: 10.3390/jcm9020460 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Tempany CMC, McDannold NJ, Hynynen K, Jolesz FA. Focused ultrasound surgery in oncology: overview and principles. Radiological. 2011;259:39‐56. doi: 10.1148/radiol.11100155 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Hynynen K, McDannold N, Sheikov NA, Jolesz FA, Vykhodtseva N. Local and reversible blood‐brain barrier disruption by noninvasive focused ultrasound at frequencies suitable for trans‐skull sonications. Neuroimage. 2005;24:12‐20. doi: 10.1016/j.neuroimage.2004.06.046 [DOI] [PubMed] [Google Scholar]
- 4. Fomenko A, Lozano AM. Neuromodulation and ablation with focused ultrasound ‐ toward the future of noninvasive brain therapy. Neural Regen Res. 2019;14:1509‐1510. doi: 10.4103/1673-5374.255961 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Abdullah BJJ, Subramaniam RV, Omar SS, et al. Magnetic resonance‐guided focused ultrasound surgery (MRgFUS) treatment for uterine fibroids. Biomed Imaging Interv J. 2010;6:e15. doi: 10.2349/biij.6.2.e15 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Kim M, Jung NY, Park CK, Chang WS, Jung HH, Chang JW. Comparative evaluation of magnetic resonance‐guided focused ultrasound surgery for essential tremor. Stereotact Funct Neurosurg. 2017;95:279‐286. doi: 10.1159/000478866 [DOI] [PubMed] [Google Scholar]
- 7. Petrusca L, Salomir R, Milleret R, et al. Experimental investigation of thermal effects in HIFU‐based external valvuloplasty with a non‐spherical transducer, using high‐resolution MR thermometry. Phys Med Biol. 2009;54:5123‐5138. doi: 10.1088/0031-9155/54/17/004 [DOI] [PubMed] [Google Scholar]
- 8. Weeks EM, Platt MW, Gedroyc W. MRI‐guided focused ultrasound (MRgFUS) to treat facet joint osteoarthritis low back pain‐case series of an innovative new technique. Eur Radiol. 2012;22:2822‐2835. doi: 10.1007/s00330-012-2628-6 [DOI] [PubMed] [Google Scholar]
- 9. Quesson B, Laurent C, Maclair G, et al. Real‐time volumetric MRI thermometry of focused ultrasound ablation in vivo: a feasibility study in pig liver and kidney. NMR Biomed. 2011;24:145‐153. doi: 10.1002/nbm.1563 [DOI] [PubMed] [Google Scholar]
- 10. Bleier AR, Jolesz FA, Cohen MS, et al. Real‐time magnetic resonance imaging of laser heat deposition in tissue. Magn Reson Med. 1991;21:132‐137. doi: 10.1002/mrm.1910210116 [DOI] [PubMed] [Google Scholar]
- 11. Lepetit‐Coiffé M, Quesson B, Seror O, et al. Real‐time monitoring of radiofrequency ablation of rabbit liver by respiratory‐gated quantitative temperature MRI. J Magn Reson Imaging. 2006;24:152‐159. doi: 10.1002/jmri.20605 [DOI] [PubMed] [Google Scholar]
- 12. De Senneville BD, Mougenot C, Moonen CTW. Real‐time adaptive methods for treatment of mobile organs by MRI‐controlled high‐intensity focused ultrasound. Magn Reson Med. 2007;57:319‐330. doi: 10.1002/mrm.21124 [DOI] [PubMed] [Google Scholar]
- 13. Masoom SN, Sundaram KM, Ghanouni P, et al. Real‐time MRI‐guided prostate interventions. Cancers (Basel). 2022;14:1860. doi: 10.3390/cancers14081860 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Mattay RR, Kim K, Shah L, et al. MR thermometry during transcranial MR imaging‐guided focused ultrasound procedures: a review. Am J Neuroradiol. 2023;44:1‐8. doi: 10.3174/ajnr.A8038 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Seasons GM, Mazerolle EL, Sankar T, et al. Predicting high‐intensity focused ultrasound thalamotomy lesions using 2D magnetic resonance thermometry and 3D Gaussian modeling. Med Phys. 2019;46:5722‐5732. doi: 10.1002/mp.13868 [DOI] [PubMed] [Google Scholar]
- 16. Jones RM, Kamps S, Huang Y, et al. Accumulated thermal dose in MRI‐guided focused ultrasound for essential tremor: repeated sonications with low focal temperatures. J Neurosurg. 2020;132:1802‐1809. doi: 10.3171/2019.2.JNS182995 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. McDannold N, Jason White P, Rees CG. MRI‐based thermal dosimetry based on single‐slice imaging during focused ultrasound thalamotomy. Phys Med Biol. 2020;65:235018. doi: 10.1088/1361-6560/abb7c4 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Todd N, Vyas U, De Bever J, Payne A, Parker DL. Reconstruction of fully three‐dimensional high spatial and temporal resolution MR temperature maps for retrospective applications. Magn Reson Med. 2012;67:724‐730. doi: 10.1002/mrm.23055 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Adams‐Tew SI, Johnson S, Odéen H, Parker DL, Payne A. Validation of a drift‐corrected 3D MR temperature imaging sequence for breast MR‐guided focused ultrasound treatments. Magn Reson Imaging. 2023;96:126‐134. doi: 10.1016/j.mri.2022.12.006 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Odéen H, de Bever J, Almquist S, et al. Treatment envelope evaluation in transcranial magnetic resonance‐guided focused ultrasound utilizing 3D MR thermometry. J Ther Ultrasound. 2014;2:19. doi: 10.1186/2050-5736-2-19 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Rivens I, Shaw A, Civale J, Morris H. Treatment monitoring and thermometry for therapeutic focused ultrasound. Int J Hyperthermia. 2007;23:121‐139. doi: 10.1080/02656730701207842 [DOI] [PubMed] [Google Scholar]
- 22. Borman PTS, Bos C, De Boorder T, Raaymakers BW, Moonen CTW, Crijns SPM. Towards real‐time thermometry using simultaneous multislice MRI. Phys Med Biol. 2016;61:N461‐N477. doi: 10.1088/0031-9155/61/17/N461 [DOI] [PubMed] [Google Scholar]
- 23. Ramsay E, Mougenot C, Köhler M, et al. MR thermometry in the human prostate gland at 3.0T for transurethral ultrasound therapy. J Magn Reson Imaging. 2013;38:1564‐1571. doi: 10.1002/jmri.24063 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Grogan DP, Abduhalikov T, Kassell NF, Moosa S. Future directions of MR‐guided focused ultrasound. Magn Reson Imaging Clin N Am. 2024;32:705‐715. doi: 10.1016/j.mric.2024.02.004 [DOI] [PubMed] [Google Scholar]
- 25. Zhou Y‐F. High intensity focused ultrasound in clinical tumor ablation. World J Clin Oncol. 2011;2:8‐27. doi: 10.5306/wjco.v2.i1.8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. De Senneville BD, Roujol S, Moonen C, Ries M. Motion correction in MR thermometry of abdominal organs: a comparison of the referenceless vs. the multibaseline approach. Magn Reson Med. 2010;64:1373‐1381. doi: 10.1002/mrm.22514 [DOI] [PubMed] [Google Scholar]
- 27. Todd N, Vyas U, de Bever J, Payne A, Parker DL. The effects of spatial sampling choices on MR temperature measurements. Magn Reson Med. 2011;65:515‐521. doi: 10.1002/mrm.22636 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. De Senneville BD, Quesson B, Moonen CTW. Magnetic resonance temperature imaging. Int J Hyperthermia. 2005;21:515‐531. doi: 10.1080/02656730500133785 [DOI] [PubMed] [Google Scholar]
- 29. Huang Y, Lipsman N, Schwartz ML, et al. Predicting lesion size by accumulated thermal dose in MR‐guided focused ultrasound for essential tremor. Med Phys. 2018;45:4704‐4710. doi: 10.1002/mp.13126 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Santos MA, Wu SK, Li Z, Goertz DE, Hynynen K. Microbubble‐assisted MRI‐guided focused ultrasound for hyperthermia at reduced power levels. Int J Hyperthermia. 2018;35:599‐611. doi: 10.1080/02656736.2018.1514468 [DOI] [PubMed] [Google Scholar]
- 31. Guillemin PC, Sinden D, M'Rad Y, et al. A novel concept of Transperineal focused ultrasound transducer for prostate cancer local deep hyperthermia treatments. Cancers (Basel). 2023;15:163. doi: 10.3390/cancers15010163 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Hokland SL, Pedersen M, Salomir R, Quesson B, Stødkilde‐Jørgensen H, Moonen CTW. MRI‐guided focused ultrasound: methodology and applications. IEEE Trans Med Imaging. 2006;25:723‐731. doi: 10.1109/TMI.2006.873296 [DOI] [PubMed] [Google Scholar]
- 33. Payne A, Chopra R, Ellens N, et al. AAPM task group 241: a medical physicist's guide to MRI‐guided focused ultrasound body systems. Med Phys. 2021;48:e772‐e806. doi: 10.1002/mp.15076 [DOI] [PubMed] [Google Scholar]
- 34. Blackwell J, Kraśny MJ, O'Brien A, et al. Proton resonance frequency shift thermometry: a review of modern clinical practices. J Magn Reson Imaging. 2022;55:389‐403. doi: 10.1002/jmri.27446 [DOI] [PubMed] [Google Scholar]
- 35. McDannold N. Quantitative MRI‐based temperature mapping based on the proton resonant frequency shift: review of validation studies. Int J Hyperthermia. 2005;21:533‐546. doi: 10.1080/02656730500096073 [DOI] [PubMed] [Google Scholar]
- 36. Stollberger R, Ascher PW, Huber D, Renhart W, Radner H, Ebner F. Temperature monitoring of interstitial thermal tissue coagulation using MR phase images. J Magn Reson Imaging. 1998;8:188‐196. doi: 10.1002/jmri.1880080132 [DOI] [PubMed] [Google Scholar]
- 37. Rieke V, Pauly KB. MR thermometry. J Magn Reson Imaging. 2008;27:376‐390. doi: 10.1002/jmri.21265 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38. Kuroda K, Iwabuchi T, Obara M, Honda M, Saito K, Imai Y. Temperature dependence of relaxation times in proton components of fatty acids. Magn Reson Med Sci. 2011;10:177‐183. doi: 10.2463/mrms.10.177 [DOI] [PubMed] [Google Scholar]
- 39. Odéen H, Parker DL. Magnetic resonance thermometry and its biological applications–physical principles and practical considerations. Prog Nucl Magn Reson Spectrosc. 2019;110:34‐61. doi: 10.1016/j.pnmrs.2019.01.003 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40. Homer J, Beevers MS. Driven‐equilibrium single‐pulse observation of T1 relaxation. A reevaluation of a rapid “new” method for determining NMR spin‐lattice relaxation times. J Magn Reson. 1985;63:287‐297. doi: 10.1016/0022-2364(85)90318-X [DOI] [Google Scholar]
- 41. Messroghli DR, Radjenovic A, Kozerke S, Higgins DM, Sivananthan MU, Ridgway JP. Modified look‐locker inversion recovery (MOLLI) for high‐resolution T1 mapping of the heart. Magn Reson Med. 2004;52:141‐146. doi: 10.1002/mrm.20110 [DOI] [PubMed] [Google Scholar]
- 42. Tong CY, Prato FS. A novel fast T1‐mapping method. J Magn Reson Imaging. 1994;4:701‐708. doi: 10.1002/jmri.1880040513 [DOI] [PubMed] [Google Scholar]
- 43. Deichmann R, Haase A. Quantification of T1 values by SNAPSHOT‐FLASH NMR imaging. J Magn Reson. 1992;96:608‐612. doi: 10.1016/0022-2364(92)90347-A [DOI] [Google Scholar]
- 44. Parker DL. Applications of NMR imaging in hyperthermia: an evaluation of the potential for localized tissue heating and noninvasive temperature monitoring. IEEE Trans Biomed Eng. 1984;BME‐31:161‐167. doi: 10.1109/TBME.1984.325382 [DOI] [PubMed] [Google Scholar]
- 45. Bohris C, Schreiber WG, Jenne J, et al. Quantitative MR temperature monitoring of high‐intensity focused ultrasound therapy. Magn Reson Imaging. 1999;17:603‐610. doi: 10.1016/S0730-725X(98)00196-9 [DOI] [PubMed] [Google Scholar]
- 46. Peller M, Reinl HM, Weigel A, Meininger M, Issels RD, Reiser M. T1 relaxation time at 0.2 tesla for monitoring regional hyperthermia: feasibility study in muscle and adipose tissue. Magn Reson Med. 2002;47:1194‐1201. doi: 10.1002/mrm.10155 [DOI] [PubMed] [Google Scholar]
- 47. Ong JT, d'Arcy JA, Collins DJ, Rivens IH, ter Haar GR, Leach MO. Sliding window dual gradient echo (SW‐dGRE): T1 and proton resonance frequency (PRF) calibration for temperature imaging in polyacrylamide gel. Phys Med Biol. 2003;48:1917‐1931. doi: 10.1088/0031-9155/48/13/305 [DOI] [PubMed] [Google Scholar]
- 48. Bos C, Lepetit‐Coiffé M, Quesson B, Moonen CTW. Simultaneous monitoring of temperature and T1: methods and preliminary results of application to drug delivery using thermosensitive liposomes. Magn Reson Med. 2005;54:1020‐1024. doi: 10.1002/mrm.20635 [DOI] [PubMed] [Google Scholar]
- 49. Kneepkens E, Heijman E, Keupp J, Weiss S, Nicolay K, Grüll H. Interleaved mapping of temperature and longitudinal relaxation rate to monitor drug delivery during magnetic resonance‐guided high‐intensity focused ultrasound‐induced hyperthermia. Invest Radiol. 2017;52:620‐630. doi: 10.1097/RLI.0000000000000392 [DOI] [PubMed] [Google Scholar]
- 50. Hey S, De Smet M, Stehning C, et al. Simultaneous T1 measurements and proton resonance frequency shift based thermometry using variable flip angles. Magn Reson Med. 2012;67:457‐463. doi: 10.1002/mrm.22987 [DOI] [PubMed] [Google Scholar]
- 51. Diakite M, Odéen H, Todd N, Payne A, Parker DL. Toward real‐time temperature monitoring in fat and aqueous tissue during magnetic resonance‐guided high‐intensity focused ultrasound using a three‐dimensional proton resonance frequency T1 method. Magn Reson Med. 2014;72:178‐187. doi: 10.1002/mrm.24900 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52. Todd N, Diakite M, Payne A, Parker DL. Hybrid proton resonance frequency/T1 technique for simultaneous temperature monitoring in adipose and aqueous tissues. Magn Reson Med. 2013;69:62‐70. doi: 10.1002/mrm.24228 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53. Todd N, Diakite M, Payne A, Parker DL. In vivo evaluation of multi‐echo hybrid PRF/T1 approach for temperature monitoring during breast MR‐guided focused ultrasound surgery treatments. Magn Reson Med. 2014;72:793‐799. doi: 10.1002/mrm.24976 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54. Zhang L, Armstrong T, Li X, Wu HH. A variable flip angle golden‐angle‐ordered 3D stack‐of‐radial MRI technique for simultaneous proton resonant frequency shift and T1‐based thermometry. Magn Reson Med. 2019;82:2062‐2076. doi: 10.1002/mrm.27883 [DOI] [PubMed] [Google Scholar]
- 55. Dick EA, Gedroyc WMW. ExAblate® magnetic resonance‐guided focused ultrasound system in multiple body applications. Expert Rev Med Devices. 2010;7:589‐597. doi: 10.1586/erd.10.38 [DOI] [PubMed] [Google Scholar]
- 56. Svedin BT, Payne A, Parker DL. Simultaneous proton resonance frequency shift thermometry and T1 measurements using a single reference variable flip angle T1 method. Magn Reson Med. 2019;81:3138‐3152. doi: 10.1002/mrm.27643 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57. Malmberg MA, Odéen H, Hofstetter LW, Hadley JR, Parker DL. Validation of single reference variable flip angle (SR‐VFA) dynamic T1 mapping with T2* correction using a novel rotating phantom. Magn Reson Med. 2024;91:1419‐1433. doi: 10.1002/mrm.29944 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58. Song HK, Dougherty L. k‐space weighted image contrast (KWIC) for contrast manipulation in projection reconstruction MRI. Magn Reson Med. 2000;44:825‐832. [DOI] [PubMed] [Google Scholar]
- 59. Odéen H, Parker DL. Improved MR thermometry for laser interstitial thermotherapy. Lasers Surg Med. 2019;51:286‐300. doi: 10.1002/lsm.23049 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60. Bydder GM, Young IR. MR imaging: clinical use of the inversion recovery sequence. J Comput Assist Tomogr. 1985;9:659‐675. doi: 10.1097/00004728-198507010-00002 [DOI] [PubMed] [Google Scholar]
- 61. Barral JK, Gudmundson E, Stikov N, Etezadi‐Amoli M, Stoica P, Nishimura DG. A robust methodology for in vivo T1 mapping. Magn Reson Med. 2010;64:1057‐1067. doi: 10.1002/mrm.22497 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62. Deman JM, Deman L, Blackman B. Melting‐point determination of fat products. J Am Oil Chem Soc. 1983;60:91‐94. doi: 10.1007/BF02540900 [DOI] [Google Scholar]
- 63. Hasgall PA, Di Gennaro F, Baumgartner C, et al. IT'IS database for thermal and electromagnetic parameters of biological tissues. Version 2.6. Accessed January 13, 2015. www.itis.ethz.ch/database.com
- 64. Arms AD, Travis CC. Reference physiological parameters in pharmacokinetic modeling. 1988. Accessed June 28, 2024. http://nepis.epa.gov/Exe/ZyPURL.cgi?Dockey=9100R7VE.txt
- 65. Song CW, Rhee JG, Levitt SH. Blood flow in normal tissues and tumors during hyperthermia. J Natl Cancer Inst. 1980;64:119‐124. [PubMed] [Google Scholar]
- 66. Akyürekli D, Gerig LH, Raaphorst GP. Changes in muscle blood flow distribution during hyperthermia. Int J Hyperthermia. 1997;13:481‐496. doi: 10.3109/02656739709023547 [DOI] [PubMed] [Google Scholar]
- 67. Hyvärinen M, Huang Y, David E, Hynynen K. Comparison of computer simulations and clinical treatment results of magnetic resonance‐guided focused ultrasound surgery (MRgFUS) of uterine fibroids. Med Phys. 2022;49:2101‐2119. doi: 10.1002/mp.15263 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68. Cline HE, Hynynen K, Hardy CJ, Watkins RD, Schenck JF, Jolesz FA. MR temperature mapping of focused ultrasound surgery. Magn Reson Med. 1994;31:628‐636. doi: 10.1002/mrm.1910310608 [DOI] [PubMed] [Google Scholar]
- 69. Parker DL, Payne A, Odéen H. A k‐space‐based method to measure and correct for temporal B0 field variations in MR temperature imaging. Magn Reson Med. 2022;88:1098‐1111. doi: 10.1002/mrm.29275 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 70. Marx M, Ghanouni P, Butts PK. Specialized volumetric thermometry for improved guidance of MRgFUS in brain. Magn Reson Med. 2017;78:508‐517. doi: 10.1002/mrm.26385 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 71. Fisher MR, Barker B, Amparo EG, et al. MR imaging using specialized coils. Radiology. 1985;157:443‐447. doi: 10.1148/radiology.157.2.4048453 [DOI] [PubMed] [Google Scholar]
- 72. Webb A, O'Reilly T. Tackling SNR at low‐field: a review of hardware approaches for point‐of‐care systems. Magn Reson Mater Phys Biol Med. 2023;36:375‐393. doi: 10.1007/s10334-023-01100-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73. Payne A, Minalga E, Merrill R, Parker DL, Hadley JR. Technical note: effect of transducer position and ground plane configuration on image quality in MR‐guided focused ultrasound therapies. Med Phys. 2020;47:2350‐2355. doi: 10.1002/mp.14138 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74. Bernstein MA, King KF, Zhou XJ. Handbook of MRI pulse sequences. Burlington, MA: Academic Press; 2004. [Google Scholar]
- 75. Malmberg MA, Odéen H, Parker DL. Effects of T2* on accuracy and precision of dynamic T1 measurements using the single reference variable flip angle method: a simulation study. Med Phys. 2022;49:2396‐2412. doi: 10.1002/mp.15476 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 76. Lutz NW, Bernard M. Contactless thermometry by MRI and MRS: advanced methods for thermotherapy and biomaterials. iScience. 2020;23:101561. doi: 10.1016/j.isci.2020.101561 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 77. Baron P, Deckers R, Knuttel FM, Bartels LW. T1 and T2 temperature dependence of female human breast adipose tissue at 1.5 T: groundwork for monitoring thermal therapies in the breast. NMR Biomed. 2015;28:1463‐1470. doi: 10.1002/nbm.3410 [DOI] [PubMed] [Google Scholar]
- 78. Diakite M, Payne A, Todd N, Parker DL. Irreversible change in the T1 temperature dependence with thermal dose using the proton resonance frequency‐T1 technique. Magn Reson Med. 2013;69:1122‐1130. doi: 10.1002/mrm.24322 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 79. Shaarani SM, Nott KP, Hall LD. Combination of NMR and MRI quantitation of moisture and structure changes for convection cooking of fresh chicken meat. Meat Sci. 2006;72:398‐403. doi: 10.1016/j.meatsci.2005.07.017 [DOI] [PubMed] [Google Scholar]
- 80. Graham SJ, Stanisz GJ, Kecojevic A, Bronskill MJ, Henkelman RM. Analysis of changes in MR properties of tissues after heat treatment. Magn Reson Med. 1999;42:1061‐1071. [DOI] [PubMed] [Google Scholar]
- 81. Rieke V, Pauly KB. Echo combination to reduce proton resonance frequency (PRF) thermometry errors from fat. J Magn Reson Imaging. 2008;27:673‐677. doi: 10.1002/jmri.21238 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82. McLean M, Parker DL, Odéen H, Payne A. A T1‐based correction method for proton resonance frequency shift thermometry in breast tissue. Med Phys. 2021;48:4719‐4729. doi: 10.1002/mp.15085 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 83. De Poorter J. Noninvasive MRI thermometry with the proton resonance frequency method: study of susceptibility effects. Magn Reson Med. 1995;34:359‐367. doi: 10.1002/mrm.1910340313 [DOI] [PubMed] [Google Scholar]
- 84. Murphy M, Alley M, Demmel J, Keutzer K, Vasanawala S, Lustig M. Fast ℓ 1‐SPIRiT compressed sensing parallel imaging MRI: scalable parallel implementation and clinically feasible runtime. IEEE Trans Med Imaging. 2012;31:1250‐1262. doi: 10.1109/TMI.2012.2188039 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 85. Svedin BT, Dillon CR, Parker DL. Effect of k‐space‐weighted image contrast and ultrasound focus size on the accuracy of proton resonance frequency thermometry. Magn Reson Med. 2019;81:247‐257. doi: 10.1002/mrm.27383 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 86. Kovesi P. Good Colour Maps: how to Design Them. arXiv. Preprint posted online September 12, 2015. doi:10.48550/arXiv:1509.03700
