Abstract
Objective
Enzyme-linked immunosorbent assay (ELISA) is the established method for detecting antiphospholipid antibodies (aPL) in the diagnosis of antiphospholipid syndrome (APS) but is labor-intensive compared with the newer automated chemiluminescent assay (CLIA). This study aims to evaluate CLIA versus ELISA for aPL, correlate each method with clinical manifestations and perform a comprehensive literature review.
Methods
Patient samples were concurrently tested by ELISA (QUANTA Lite®) and CLIA (ACL AcuStar®) for anti-cardiolipin antibody (aCL) and anti-β2-glycoprotein-I (aβ2GPI) IgG and IgM. Assay results were correlated with any of the revised Sapporo APS clinical criteria.
Results
Of the 107 patients, 67% fulfilled at least one clinical criterion. 38 patients (35.5%) had APS. For aCL IgG, aCL IgM and aβ2GPI IgM, CLIA showed above 77% concordance and fair to excellent agreement (Cohen's kappa 0.39–0.86) with moderate/high positive ELISA of ≥40 units. Both methods showed good correlation (Spearman's r 0.60–0.80, p < 0.0001) that was non-linear over the range of titers. CLIA sensitivity and specificity was 46%–100% and 68%–95%, with AUROC ranging from 0.80–0.93. For aβ2GPI IgG, concordance was 36.7% and agreement was low (kappa −0.23). Correlation with clinical criteria revealed no statistically significant difference in the occurrence of clinical manifestations in ELISA-positive versus CLIA-positive groups.
Conclusions
aPL detection by CLIA showed close but incomplete concordance with ELISA. CLIA positivity correlated well with moderate/high ELISA positivity, but antibody titers should not be directly compared across systems. CLIA is an acceptable alternative to ELISA in the routine non-research setting. Our findings are congruent with the reviewed literature.
Keywords: antiphospholipid antibodies, antiphospholipid syndrome, diagnosis, solid phase assay, chemiluminescent, ELISA
Introduction
Antiphospholipid syndrome (APS) is a systemic autoimmune disease characterized by vascular thrombosis in venous, arterial or microvascular beds and/or obstetric morbidity such as fetal losses, intrauterine growth restriction and severe pre-eclampsia, in the presence and persistence of antiphospholipid antibodies (aPL). A severe syndrome of thrombosis in multiple organs, termed catastrophic APS, as well as non-thrombotic manifestations including thrombocytopenia, hemolytic anemia, valvular heart disease and nephropathy are other well-known manifestations of APS. 1
The diagnosis of APS relies on the laboratory detection of aPL. aPL mediate several procoagulant mechanisms and have a direct pathogenic role in APS. β−2 glycoprotein I (β2GPI) dependent aPL, the most important subset of these antibodies, mediate thrombosis by reactivity with β2GPI bound to the membrane of cells that participate in the coagulation cascade.2–4 It is thought that reactivity of anti-cardiolipin antibodies (aCL) is mediated by β2GPI antibodies (aβ2GPI) that bind β2GPI complexed with cardiolipin.5,6 The lupus anticoagulant, aCL and aβ2GPI of IgM and IgG isotypes are part of the laboratory criteria for APS in the revised Sapporo classification criteria and the recent 2023 American College of Rheumatology/ European Alliance of Associations for Rheumatology (ACR/EULAR) Antiphospholipid Syndrome Classification Criteria.7,8
The enzyme-linked immunosorbent assay (ELISA) is the standard solid phase assay used for detection of aPL since the 1980s. The test involves a plate coated with cardiolipin or β2GPI, the addition of patient's serum or plasma, followed by the application of a secondary labeled antibody to quantify the bound IgG or IgM.
More recently, other methods for antibody detection have been developed, such as the chemiluminescent immunoassays (CLIA). 9 This two-phased assay consists of paramagnetic particles coated with cardiolipin and/or β2GPI, which capture aPL from the patient's serum or plasma. Reagents that trigger the chemiluminescence reaction are added and the emitted light is measured and reported in relative light units (RLUs), which are directly proportional to the concentration of aPL. CLIA has several advantages over ELISA due to its several fold greater sensitivity thereby allowing measurement of extremely low levels of analyte, wider dynamic range allowing measurement of high concentrations, and absence of interfering emission conferring high specificity. 10 By now, CLIA is a well-established methodology in use by many instrument manufacturers and laboratories. 11
CLIA has been shown to have comparable performance to ELISA for aPL detection.12–19 Furthermore, CLIA is fully automated, less labor intensive and can be performed more rapidly. However, the 2023 ACR/EULAR Antiphospholipid Syndrome Classification Criteria consensus recommendation stated that for research purposes, ELISA should be used for APS classification, as there is as yet lack of standardization between equivalent CLIA levels corresponding to moderate and high ELISA titers. This has led to debate over whether CLIA is an acceptable alternative to ELISA for APS diagnosis in routine clinical practice, contributed in part by confusion between the definitions of classification criteria and diagnostic criteria.11,20 As the intention of classification criteria is to identify a homogenous patient cohort for inclusion in trials and observational studies, they tend to be highly specific and capture a narrower range of disease. Hence, classification criteria do not necessarily equate to the only criteria for diagnosis. Diagnosis, on the other hand, is made by evaluating signs, symptoms and supporting tests of the individual patient, and therefore, patients diagnosed with disease may or may not fulfill classification criteria.
On this background, this study aimed to clarify the qualitative and quantitative differences between ELISA and CLIA for the detection of aPL, and to correlate the results of aPL detected by each method to the clinical manifestations of APS.
Methods
This study is a diagnostic study conducted in a single tertiary hospital. The study was conducted in accordance with the Declaration of Helsinki. Ethics approval for the study was granted by the institution review board. Patient samples referred to the laboratory for detection of aPL between April 2022 to May 2023 were included in a performance evaluation of CLIA versus existing ELISA prior to implementation. Consecutive samples positive for at least one aPL by the routine ELISA method were tested concurrently using the CLIA method for aCL IgG and IgM and aβ2GPI IgG and IgM.
Blood samples were collected in serum tubes (BD Vacutainer®, Becton Dickinson and Company, Franklin Lakes, NJ, USA) and centrifuged at 4000 RPM. Serum was stored at −25 °C until ELISA and CLIA testing. ELISA was performed using the QUANTA Lite® assay (Inova Diagnostics, San Diego, CA, USA) on the Evolis® ELISA system (BioRad, Marnes-La-Coquette, France). In the ELISA, antibodies present in serum samples bind to the immobilized purified cardiolipin and bovine β2GPI antigen on the microwell plate. Anti-human IgG or IgM conjugate is added after the first wash, binding to any antibodies attached to the microwells. A chromogenic substrate is added after the second wash, and the intensity of the color is measured and compared to a five-point calibration curve. Results are expressed in phospholipid units (GPL/MPL) or standard IgG β2GPI units (SGU/SMU). The laboratory ELISA positive cut-off was ≥20 units, which was validated within the laboratory to correspond to >99th percentile of a normal population. ELISA moderate/high titers (ie ≥ 40 units) is recommended for classification of APS, hence this threshold was also incorporated into the analysis. 7
CLIA was performed on the ACL AcuStar® platform or (Werfen, Barcelona, Spain) as per manufacturer's instructions. CLIA uses paramagnetic beads coated with bovine cardiolipin and human β2GPI. After incubation of the magnetic beads with serum samples, magnetic separation, and washing of the beads, a tracer consisting of isoluminol-labeled anti-human IgG or IgM antibody is added. The tracer binds to antibodies captured on the beads. After another incubation, magnetic separation and wash cycle, reagents are added to induce luminescence reaction. The emitted light is measured as RLUs, which are directly proportional to the aCL and aβ2GPI present in the sample. Results are expressed in U/mL (IgG or IgM Arbitrary Units). The CLIA cut-off for positive was ≥20 U/ml, as per manufacturer's recommendations.
Lupus anticoagulant assays were performed according to the recommendations of the International Society on Thrombosis and Haemostasis (ISTH), using screening, confirmation and mixing steps. In the screening tests, diluted Russell's viper venom test (dRVVT) (LA1 Screening reagent, Siemens Healthineers, Marburg, Germany) and lupus anticoagulant-sensitive activated partial thromboplastin time (APTT) (Dade Actin FSL, Siemens Healthineers) were measured on CS-5100® (Sysmex Corp. Japan). A positive dRVVT screening test was followed by a confirmatory dRVVT (LA2 Confirmation reagent, Siemens Healthineers) measured on CS-5100®. LA1 and LA2 mixing steps were performed according to the laboratory's workflow. A positive lupus anticoagulant-sensitive APTT screening test was followed by a confirmatory test using hexagonal phase phospholipid neutralization (Staclot® LA, Diagnostica Stago, Asniere-sur-Seine, France) performed on STart® 4 Haemostasis analyzer (Diagnostica Stago). Lupus anticoagulant was present when dRVVT results expressed as normalized ratio was >1.15 or Staclot® LA clotting time difference was ≥ 8 s.
Clinical data for the patients were extracted retrospectively from electronic medical records. These were (1) patient demographics (age, gender, race), (2) history of venous or arterial thrombosis, (3) provoking risk factors for venous thromboembolism (VTE), (4) arterial thrombosis risk factors such as obesity, smoking status, hypertension, hyperlipidemia and diabetes mellitus, (5) history of pregnancy complications ie early (<10 weeks) fetal loss, premature births before the 34th week of gestation because of eclampsia or placental insufficiency and spontaneous abortions before the 10th week of gestation, as well as (6) non-criteria manifestations such as heart valve disease, livedo reticularis, thrombocytopenia, nephropathy, and neurological manifestations. Other clinical information such as the presence of other autoimmune disease, previous diagnosis of APS, type of anticoagulant or antiplatelet agent given, previous aPL results were obtained. The revised Sapporo classification criteria was used to assign whether a patient met clinical criteria for APS. 7
Statistical analysis was performed on SPSS version 28.0 (IBM SPSS, Chicago, USA) and GraphPad Prism version 9.4 (GraphPad, Boston, USA). Normality testing was performed with the Kolmogorov-Smirnov test. Qualitative comparison of concordance between the two methods was performed using 2 × 2 contingency tables. Cohen's kappa value for inter-rater agreement was calculated. Quantitative comparison between titers obtained by the two methods was performed using Spearman rank order correlation for non-parametrically distributed data. Receiver operating characteristics (ROC) was performed to evaluate the sensitivity and specificity of CLIA for a dichotomous ELISA positive/negative result, recognizing that although ELISA is not a true gold standard, it is the incumbent against which the newer method is compared. A related samples McNemar test was performed to determine if the frequency of clinical criteria differed by each method. McNemar test evaluates for any difference in the proportion of clinical outcome in the subset of discordant cases. A p < 0.05 was considered statistically significant.
Results
Demographic and Clinical Data
A total of 107 patients were included in this study, of whom 67 were females (63%). The median age was 48 years (range 11-91 years). Seventy-two patients (67%) fulfilled at least one revised Sapporo clinical criteria. Sixty-six patients (62%) had thrombosis, and nine patients (8%) had obstetric criteria. Twenty-five patients (23%) experienced VTE, for a total of 29 VTE events. Out of the VTE events, 17 (59%) were unprovoked. Forty-six patients (43%) had arterial thrombosis, for a total of 56 arterial thrombosis events. Twenty-eight patients (26%) had primary APS, and 10 (9%) had secondary APS associated with systemic lupus erythematosus and other systemic autoimmune disease.
Concurrent ELISA and CLIA testing were performed in 31 patient samples for aCL IgG, 67 samples for aCL IgM, 30 samples for aβ2GPI IgG and 33 samples for aβ2GPI IgM. The relatively higher number for aCL IgM was due to the higher number of samples for aCL IgM measurement received during the study period.
ELISA Versus CLIA Comparison
(I) Concordance Rate
The concordance rate between ELISA and CLIA was calculated using laboratory positive cut-off (ELISA titer ≥20, CLIA ≥20 U/ml), as well as a moderate/high ELISA titer of ≥40. The results are shown in Table 1. The concordance rate was higher when CLIA was compared with an ELISA titer of ≥40 than with the lower cut-off of ≥20. The concordance rate was above 77% for aCL IgG, aCL IgM and aβ2GPI IgM, but was only 36.7% for aβ2GPI IgG. Cohen's kappa coefficient was fair to excellent for aCL IgG, aCL IgM and aβ2GPI IgM when ELISA cut-off was ≥40, but was poor for aβ2GPI IgG (Kappa coefficients = 0.55, 0.39, 0.86, −0.23 respectively). Cohen's kappa coefficients dropped when ELISA ≥20 was compared with CLIA (Table 1).
Table 1.
Concordance Rate of Antiphospholipid Antibodies Detected by ELISA and CLIA.
| aCL IgG | aCL IgM | aβ2GPI IgG | aβ2GPI IgM | ||||||
|---|---|---|---|---|---|---|---|---|---|
| n = 31 | n = 67 | n = 30 | n = 33 | ||||||
| CLIA+ | CLIA- | CLIA+ | CLIA- | CLIA+ | CLIA- | CLIA+ | CLIA- | ||
| (A) | ELISA + (≥20), n | 14 | 5 | 10 | 26 | 14 | 10 | 11 | 5 |
| ELISA- (<20), n | 2 | 10 | 1 | 30 | 4 | 2 | 0 | 17 | |
| Positive concordance | 73.7% | 27.8% | 58.3% | 68.8% | |||||
| Negative concordance | 83.3% | 96.8% | 33.3% | 100.0% | |||||
| Total concordance | 77.4% | 59.7% | 53.3% | 84.9% | |||||
| Kappa | 0.545 | 0.223 | −0.061 | 0.694 | |||||
| (B) | ELISA + (≥40), n | 9 | 0 | 6 | 7 | 6 | 7 | 10 | 1 |
| ELISA- (<40), n | 7 | 15 | 5 | 49 | 12 | 5 | 1 | 21 | |
| Positive concordance | 100.0% | 46.2% | 46.2% | 90.9% | |||||
| Negative concordance | 68.2% | 96.6% | 29.4% | 95.5% | |||||
| Total concordance | 77.4% | 82.1% | 36.7% | 93.9% | |||||
| Kappa | 0.554 | 0.392 | −0.234 | 0.864 | |||||
Abbreviations: ELISA, enzyme linked immunosorbent assay; CLIA, chemiluminescent assay; aCL, anticardiolipin antibody; aβ2GPI, β2-glycoprotein I antibody; IgG, immunoglobulin G; IgM, immunoglobulin M
(II) Quantitative Correlation Between ELISA and CLIA
The correlation between ELISA titer and CLIA units was investigated using Spearman rank correlation. Spearman correlation showed moderate to strong positive correlation between ELISA titer and CLIA units for aCL IgG (r = 0.806, 95% confidence interval [CI] 0.625-0.905, p < 0.0001), aCL IgM (r = 0.607, 95% CI 0.424-0.743, p < 0.0001) and aβ2GPI IgM (r = 0.808, 95% CI 0.637-0.903, p < 0.0001). However, there was no correlation between ELISA titer and CLIA units for aβ2GPI IgG (r = 0.062, 95% CI −0.316-0.422, p = 0.746). The scatter plots of the four pairwise correlations showed a non-linear relationship, as shown in Figure 1.
Figure 1.
Correlation between quantitative units measured by ELISA and CLIA. Abbreviations: ELISA, enzyme linked immunosorbent assay; CLIA, chemiluminescent assay; aCL, anticardiolipin antibody; aβ2GPI, β2-glycoprotein I antibody; IgG, immunoglobulin G; IgM, immunoglobulin M; r, Spearman rank correlation coefficient; CI, confidence interval.
(III) Receiver Operating Characteristics of CLIA
Sensitivity and specificity of CLIA quantitative results for a positive ELISA of ≥40 units were assessed by ROC analysis (Figure 2). The area under the ROC curve (AUROC) was 0.934 for CLIA aCL IgG (95% CI 0.848-1.000, p < 0.001), 0.806 for CLIA aCL IgM (95% CI 0.680-0.933, p < 0.001), and 0.913 for CLIA aβ2GPI IgM (95% CI 0.771-1.000, p < 0.001). For CLIA aβ2GPI IgG, AUROC was 0.543 (95% CI 0.313-0.773, p = 0.691). Using the CLIA manufacturer's positive cut-off of ≥20 U/ml and other potential cut-off levels (10, 30, 40 U/ml), the sensitivity and specificity of CLIA for a positive ELISA were obtained from the ROC table and shown in Table 2.
Figure 2.
Receiver operating characteristics of CLIA levels for moderate/high ELISA titer of 40 units and above. Abbreviations: ELISA, enzyme linked immunosorbent assay; CLIA, chemiluminescent assay; aCL, anticardiolipin antibody; aβ2GPI, β2-glycoprotein I antibody; IgG, immunoglobulin G; IgM, immunoglobulin M; AUC, area under the curve; CI, confidence interval.
Table 2.
Sensitivity and Specificity of CLIA for a Moderate/High Positive ELISA of ≥ 40 Units.
| CLIA cut-off ≥10 U/ml | CLIA cut-off ≥20 U/ml | CLIA cut-off ≥30 U/ml | CLIA cut-off ≥40 U/ml | ||||||
|---|---|---|---|---|---|---|---|---|---|
| n | Sen | Spe | Sen | Spe | Sen | Spe | Sen | Spe | |
| aCL IgG | 31 | 100% | 45.5% | 100% | 68.2% | 100% | 77.3% | 100% | 81.8% |
| aCL IgM | 67 | 61.5% | 85.2% | 46.2% | 90.7% | 38.5% | 98.1% | 30.8% | 100% |
| aβ2GPI IgG | 30 | 69.2% | 23.5% | 46.2% | 29.4% | 38.5% | 29.4% | 38.5% | 35.3% |
| aβ2GPI IgM | 33 | 90.9% | 86.4% | 90.9% | 95.5% | 72.7% | 95.5% | 63.6% | 95.5% |
Abbreviations: ELISA, enzyme linked immunosorbent assay; CLIA, chemiluminescent assay; aCL, anticardiolipin antibody; aβ2GPI, β2-glycoprotein I antibody; IgG, immunoglobulin G; IgM, immunoglobulin M; Sen, sensitivity; Spe, specificity. 20 U/ml is the manufacturer and laboratory cut-off for positivity. Italicized columns indicate sensitivity and specificity at manufacturer's positive cutoff.
Association with APS Clinical Criteria
McNemar test was applied to evaluate for any difference in the incidence of clinical manifestations of thrombosis and obstetric complications between cases with ELISA positive status (≥40 units) versus CLIA positive status (≥20 U/ml). The McNemar test showed that there was no statistically significant difference in the occurrence of combined thrombosis and obstetric manifestations, and that of combined venous and arterial thrombosis, in the ELISA positive group versus the CLIA positive group. The occurrence of obstetric complications alone was not evaluated due to small numbers (<5 cases). The summary of the comparison is presented in Table 3.
Table 3.
Association Between Antiphospholipid Syndrome Clinical Criteria and aPL Detected by ELISA or CLIA.
| aCL IgG | aCL IgM | aβ2GPI IgG | aβ2GPI IgM | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Total number with criteria | n = 20 | n = 46 | n = 24 | n = 25 | ||||||||
| Any APS clinical criterion present#, no. (%) | ELISA+ | CLIA+ | p | ELISA+ | CLIA+ | p | ELISA+ | CLIA+ | p | ELISA+ | CLIA+ | p |
| 8 (40.0%) | 13 (65.0%) | 0.063 | 10 (21.7%) | 10 (21.7%) | 1.000 | 11 (45.8%) | 15 (62.5%) | 0.454 | 7 (28.0%) | 7 (28.0%) | 1.000 | |
| Total number with criteria | n = 18 | n = 40 | n = 24 | n = 24 | ||||||||
| Any combined venous and/or arterial thrombosis present, no. (%) | ELISA+ | CLIA+ | p | ELISA+ | CLIA+ | p | ELISA+ | CLIA+ | p | ELISA+ | CLIA+ | p |
| 8 (44.4%) | 12 (66.7%) | 0.125 | 10 (25.0%) | 9 (22.5%) | 1.000 | 11 (45.8%) | 15 (62.5%) | 0.454 | 7 (29.2%) | 7 (29.2%) | 1.000 | |
Any APS clinical criterion refers to presence of any venous and/or arterial thrombosis and/or obstetric complications according to revised Sapporo classification. Exact p value from related samples McNemar test is computed based on binomial distribution.
Abbreviations: aPL, antiphospholipid antibody; ELISA, enzyme linked immunosorbent assay; CLIA, chemiluminescent assay; aCL, anticardiolipin antibody; aβ2GPI, β2-glycoprotein I antibody; IgG, immunoglobulin G; IgM, immunoglobulin M; ELISA+, ELISA ≥40 units; CLIA+, CLIA ≥20 U/ml
Due to the large discordance in aβ2GPI IgG results, we reviewed the clinical details of these cases, and the clinical correlation is presented in Table 4. The patients who were aβ2GPI IgG CLIA positive had a higher proportion who also had lupus anticoagulant detected, were double aPL positive (any two of the three aPL subtypes – lupus anticoagulant, aCL or aβ2GPI) or triple aPL positive (all three aPL subtypes), had co-existing autoimmune disease or clinical thrombosis, than the group who were positive for aβ2GPI IgG ELISA alone.
Table 4.
Clinical Details of Patients with Anti-β2 Glycoprotein-I IgG Detected by ELISA or CLIA.
| Table 4(A): Patients with positive ELISA aβ2GPI IgG but negative CLIA aβ2GPI IgG | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Patient | Age (years) | Gender | ELISA (units) | ELISA Other* (units) | CLIA (U/mL) | Lupus Anticoagulant^ | Other positive aPL (ELISA and/or CLIA)# | Event(s) leading to screening | Sapporo Criteria Events on 1-year follow up |
| 1 | 47 | Male | 54.0 | 62.9 (After) Positive | 10.9 | Not Detected | NIL | I) CVA | Started on aspirin (declined warfarin). No further events. |
| 2 | 52 | Female | 34.7 | 43.5 (Prior) Positive | 9.8 | Not Detected | NIL | I) CVA | Started on aspirin. Developed AMI (NSTEMI) and subsequently passed on. |
| 3 | 59 | Male | 60.3 | NIL | 7.8 | Not Detected | NIL | I) CVA | No follow up information available |
| 4 | 76 | Male | 51.1 | NIL | 11.5 | Not performed |
|
|
Started on warfarin. Developed AMI (NSTEMI). |
| 5 | 40 | Male | 88.3 | 115.7 (After) Positive | <6.4 | Not Detected | NIL |
|
Started on warfarin. No further events. |
| 6 | 66 | Female | 25.8 | NIL | <6.4 | Not Detected | NIL | I) Recurrent dialysis access failure | No further events. |
| 7 | 12 | Female | 101.9 | 91.7 (After) Positive | <6.4 | Not Detected | NIL | I) Underlying SLE | No further events. |
| 8 | 51 | Female | 120.6 | >150 (After) Positive | 10.1 | Not Detected | NIL | I) CVA | No further events. |
| 9 | 84 | Female | 59.4 | NIL | 8.5 | Not Detected | NIL | I) Sagittal sinus thrombosis | Kept on apixaban. No further events. |
| 10 | 20 | Female | 22.9 | 32.8 (After) Positive | 8.2 | Not Detected | NIL | I) Underlying SLE | No further events. |
Abbreviations: ELISA, enzyme linked immunosorbent assay; CLIA, chemiluminescent assay; aCL, anticardiolipin antibody; aβ2GPI, β2-glycoprotein I antibody; IgG, immunoglobulin G; IgM, immunoglobulin M; CVA, cerebrovascular accident; AMI, Acute myocardial infarction; NSTEMI, non-ST elevation myocardial infarction; PE, pulmonary embolism; DVT, deep vein thrombosis; SLE, systemic lupus erythematosus; pAF, paroxysmal atrial fibrillation; TIA, transient ischemic attack
* aβ2GPI IgG by ELISA performed at another setting
aβ2GPI IgM, aCL IgG or aCL IgM by ELISA performed at the same setting
^ performed at the same setting
Discussion
From our study, CLIA for aCL IgG, aCL IgM and aβ2GPI IgM showed acceptable analytical performance compared with ELISA assay. This was evident from the excellent AUROC of CLIA, good concordance rate with ELISA (total concordance rate of 77.4%, 82.1% and 93.9% respectively) and fair to excellent kappa coefficients for agreement. For example, CLIA aCL IgG and aβ2GPI IgM positivity showed excellent sensitivity/specificity for the corresponding ELISA moderate/high titer positivity (100%/68.2% and 90.9%/95.5%). For aCL IgM, CLIA had excellent specificity (90.7%) but lower sensitivity (46.2%) for positive ELISA, but sensitivity could be moderately improved to 61.5% if CLIA positive cut-off was lowered to 10 U/ml, but at the cost of specificity which would decline to 85.2%. Moreover, CLIA positivity had better concordance with ELISA ≥40 units than with ELISA ≥20 units, which supports that CLIA positivity correlated well to moderate/high ELISA positivity.
Compared to the ELISA assay, the CLIA assay is less time-consuming and labor-intensive and allows for more flexibility. The assay set-up time per 20 samples for the CLIA is 20 min as compared with 40 min for the ELISA. According to the manufacturer, calibration is needed only during every lot change for the CLIA as compared to every batch run for the ELISA. For the fully automated CLIA, reagents are packaged in self-contained cartridges which eliminates reagent handling. For the semi-automated ELISA, calibrators and reagents require manual set-up for each run. The CLIA gives room for greater flexibility as it allows random access and continuous loading, whereas ELISA is configured for only batch testing. The time-to-result of 30 min for the CLIA is significantly shorter than the 90–120 min for the ELISA.
It is recommended that laboratories establish in-house cut-off values for ELISA and CLIA using the 99th percentile of at least 50 to 120 healthy normal individuals.21,22 The study by Persijn found that in-house CLIA cut-offs were significantly lower than the manufacturer's cut-off but using in-house cut-offs increased the false positive rate, leading to lower specificity. 19 In a study by Li et al, the in-house cut-off of the ACL AcuStar® CLIA was close to the 20 U/ml manufacturer's cut-off. 14 In another study by De Moerloose et al evaluating the ACL AcuStar® assay, the value of 20 U/ml corresponded well to the 99th percentile of normal and was used in the comparison with three established ELISA assays. 13 Therefore, we used the manufacturer's CLIA cut-off in this study. This underscores the importance of careful establishment of reference ranges as it can affect the diagnosis of APS.
Even though these was moderate quantitative correlation between ELISA and CLIA (r coefficients 0.6–0.8), this correlation was less than the coefficient of 0.95 typically required for goodness of fit in laboratory method comparison, and the degree of variability and non-linear relationship was evident on scatter plots. Therefore, ELISA units are not interchangeable with CLIA units. The thresholds of <40, 40–80 and >80 units for low, moderate and high positivity apply only for ELISA and should not be used interchangeably with CLIA. This conclusion is supported by published literature. For example, the review by Vandevelde and Devreese stated that there is limited numerical agreement between different assay methods and direct comparison of values across assays should not be made. 23 A study by Vandevelde et al demonstrated poor agreement between ELISA versus CLIA or multiplex flow immunoassay for classifying positive samples as low, moderate, or high positive. Specifically, there was no agreement observed for aβ2GPI IgG, weak agreement for aCL IgG and only moderate agreement for aCL IgM and aβ2GPI IgM. Use of standard reference calibration material produced slight improvement in agreement, yet none showed strong agreement. 24 All in, our results suggest that CLIA positivity relates well to moderate/high ELISA positivity, but apart from this, direct translation of antibody titers should not be made.
In terms of test result correlation with clinical manifestations, there were no observed or statistically significant differences between assays in the proportion of cases with criteria manifestations. Hence, CLIA is likely able to provide the same degree of clinical predictive value as ELISA in the diagnosis of APS. This conclusion is concordant with that of extant studies that reported CLIA being equivalent if not having better sensitivity for APS than ELISA. Bettacchioli et al reviewed eight studies in a meta-analysis comparing ELISA with CLIA and found that the odds ratio for clinical APS manifestations when aβ2GPI IgG was present by CLIA was higher than the odds ratio when present by ELISA (Odds ratio for CLIA of 16.4, 95% CI 3.1-34.9 vs odds ratio for ELISA of 8.8, 95% CI 2.9-27.9). The odds ratio for aβ2GPI IgM present by CLIA was also higher (Odds ratio for CLIA of 4.5, 95% CI 1.1-50.9 vs odds ratio for ELISA of 2.4, 95% CI 0.9-8.7). 25
As the diagnosis of APS relies heavily on laboratory criteria, it is crucial that tests have high sensitivity and specificity for the disease. We performed a comprehensive review of the English language literature for studies evaluating CLIA versus ELISA for aPL detection and summarize these in Table 5A and B. These studies suggest that CLIA provides excellent analytical performance in terms of precision, reproducibility and linearity. In addition, these studies and ours demonstrate that CLIA has acceptable concordance and agreement with ELISA and provides strong clinical predictive value for APS. Some studies suggest that CLIA may have greater sensitivity than a number of ELISA platforms, particularly for aβ2GPI.12,16–18,25,26 A compelling explanation is better exposure of cryptic epitopes that are required for binding of pathogenic aβ2GPI such as G40-R43 on domain I.
Table 5.
Summary of Studies Evaluating Antiphospholipid Antibody Testing by CLIA Compared with ELISA.
| Table 5(A): Laboratory method comparison | ||||
|---|---|---|---|---|
| Paper (Year) | Comparison | n | Qualitative agreement | Quantitative agreement |
| Zhang, et al (2015) 12 | CLIA (QUANTA Flash) versus ELISA (QUANTA Lite) | 227 | Good overall agreement for IgG/IgM/IgA aCL and IgM/IgA aβ2GPI (>90%) and moderate overall agreement for IgG aβ2GPI (76.7%) | Significant quantitative correlations between ELISA and CLIA assays in IgG/IgM/IgA aCL and IgG/IgM/IgA aβ2GPI (P < 0.001) |
| De Moerloose, et al (2010) 13 | CLIA (AcuStar) versus 3 commercial ELISA assays (QUANTA Lite, REAADS and VARELISA) | 321 | CLIA showed high agreement with all 3 ELISA kits (82%–96%). Kappa coefficient (0.44-0.82), with highest agreement with QUANTA Lite. | Correlation slopes of 0.9–1.1 and r coefficient >0.997. |
| Van Hoecke, et al (2012) 15 | CLIA (Zenit RA and Acustar) versus ELISA (Bindazyme) | 124 | Good total agreement for all assays ranging from between 93.5% and 97.6%. | N/A |
| Wan, et al (2020) 18 | CLIA (QUANTA Flash) versus ELISA (QUANTA Lite) | 505 | Good total agreement between ELISA and CLIA (>80%). In the APS population, total agreement using manufacturer's cut-off values was good. Kappa coefficients for the four antibodies ranged from moderate to substantial (0.61-0.75). | aβ2GPI and aCL IgG assays showed the highest Spearman's rho coefficients (aβ2GPI IgG = 0.742, aCL IgG = 0.715). |
| Iwaniec, et al (2015) 27 | CLIA (QUANTA Flash) versus ELISA (QUANTA Lite) | 220 | Good qualitative agreement ranging 81.8% to 90.9%. Kappa coefficient ranged from 0.56 to 0.81. | Quantitative results showed strong, significant correlation between methods. Spearman's rho ranging from 0.75 to 0.83 (p < 0.0001). |
| Noubouossie, et al (2014) 28 | Detection of aCL IgG using CLIA (Liaison) versus ELISA assay (REAADS) | 87 | Moderate overall agreement of 56.3%, positive agreement 49.3%, negative agreement 77.2%. When CLIA was compared to repeated ELISA, the agreement improved to good agreement (overall agreement 82.1%). Most discordant cases were positive with ELISA and negative with CLIA. | N/A |
| Persijn, et al (2011) 19 | CLIA (Zenit RA) versus ELISA (Bindazyme) using manufacturer's and in-house cut off values | 314 | Overall agreement using manufacturer's cut-offs was high for aCL and aβ2GPI IgG and IgM (93-96%). Kappa values was moderate, ranging from 0.48–0.56. Overall agreement was low when using in-house cut-offs for both IgG (60-76%), with Kappa values of 0.14–0.52. In the APS population, there was substantial agreement in aCL and fair to moderate agreement in aβ2GPI. | N/A |
| Li, et al (2015) 14 | CLIA (AcuStar) versus ELISA (QUANTA Lite) | 204 | Agreement was comparable between in-house and manufacturer's cut-offs. Total agreement ranged from 82.6% to 95.7%. Kappa coefficient ranging from fair to good while using both in-house and manufacturer's cut-offs (0.36-0.65). Overall agreement between AcuStar and ELISA was better for aCL and aβ2GPI IgG than IgM. | N/A |
| Janek, et al (2016) 16 | CLIA (AcuStar) versus ELISA (AIDA) | 122 | Excluding aβ2GPI (Kappa 0.21), there is moderate agreement between CLIA and ELISA methods using manufacturer cut-off (Kappa values ranging 0.53 - 0.94). Increasing the CLIA cut-offs increased agreement for all parameters. | N/A |
Abbreviations: ELISA, enzyme linked immunosorbent assay; CLIA, chemiluminescent assay; aCL, anticardiolipin antibody; aβ2GPI, β2-glycoprotein I antibody; IgG, immunoglobulin G; IgM, immunoglobulin M; IgA, immunoglobulin A; APS, antiphospholipid syndrome; ROC, receiver operating characteristics
To further understand the basis for the discordance between aβ2GPI IgG CLIA and ELISA in our study, we reviewed the underpinnings of aPL. Pathogenic aPL target anionic phospholipids via interaction with phospholipid binding proteins, specifically β2GPI, whereas incidental aPL, found in 1%–5% of asymptomatic individuals, bind directly to neutral and anionic phospholipids.4,6 β2GPI is a serum protein that comprises five domains (I – V). The N-terminal domain (Domain I) has been identified as the major binding site of pathogenic aβ2GPI.29–31 A conformational change in β2GPI is necessary for aPL to recognize domain I. This conformational change is induced by β2GPI binding to a negatively charged anionic surface. When this binding occurs, the cryptic epitope at position 40–43 (G40-R43) in domain I is exposed, allowing aβ2GPI exert its effect on coagulation.30–35 Dimerization of β2GPI induced by antibody binding also increases the affinity towards interaction with anionic phospholipids. 36 In the aCL solid phase assay, cardiolipin is used as the binding surface for β2GPI and therefore the assay is effectively an assay for aβ2GPI. As the affinity of β2GPI to cardiolipin is high, dimerization of β2GPI is not essential and only the exposure of the cryptic epitope is essential in this assay. 37
If ELISA plates are unable to induce the conformation change required to expose the cryptic epitope, or have low affinity for β2GPI, the consequence could be an inability to detect aβ2GPI.33,38,39 Pelkmans et al, using a monoclonal antibody that reacts specifically with G40-R43 in the open conformation and another that reacts with β2GPI irrespective of conformation, showed that five different commercial aβ2GPI IgG ELISA assays differed considerably in the detection of anti-G40-R43 antibodies due to decreased exposure of the epitope. 40 The source of β2GPI may also affect the ability to bind aβ2GPI, as recombinant β2GPI shows better binding than plasma purified β2GPI due to differences in glycosylation. 33
Domain I specific aβ2GPI is believed to be largely responsible for the clinical complications of thrombosis and late pregnancy morbidity in APS, while aβ2GPI that recognizes other domains (domains IV/V) are thought to be non-pathogenic.41–46 Andreoli et al showed that asymptomatic aPL carriers displayed higher frequency of anti-domain IV/V, while anti-domain I antibodies were more frequent and present at higher titers in triple aPL positive APS. 45 In the study by de Laat et al, of patients with antibodies to whole β2GPI, 55% had anti-domain I antibodies which was associated with thrombosis (odds ratio 3.5, 95% CI 2.3-5.4) and obstetric complications (odds ratio 2.4, 95% CI 1.4-4.3). 46 Transient and non-pathogenic aβ2GPI are described in a multitude of infections but the majority do not cause thrombosis or APS. 47 The differences in the ability to detect pathogenic versus non-pathogenic aPL is likely to explain discrepancies in sensitivity and specificity and the clinical predictive value of each method.
Among ELISA assays, imperfect agreement exists between ELISA kits from different manufacturers, particularly in low and medium positive samples. 48 A study from the APS ACTION International Clinical Database found agreement in only 80%–90% when aPL was measured by ELISA at core versus non-core laboratories, with a small number of high titer samples showing discordance. 49 Hence, slight inter-laboratory variation is unavoidable even with modern day assay design and use of reference calibrators, and can be attributed to differences in assay components such as microtiter plates, phospholipids, blocking agents and source of β2GPI. 39 As a result, the ISTH recommends validation of new assays not only on analytical criteria but also on clinical criteria, ie association between assay results and clinical complications, whenever feasible. 22
Therefore, we hypothesize that a major contributory factor to the discordance between CLIA and ELISA for aβ2GPI IgG is the different configuration of the β2GPI molecule. The ACL AcuStar® CLIA contains the open configuration of β2GPI molecule thereby more efficiently detects pathogenic domain I antibodies, whereas the ELISA has the β2GPI molecule in close configuration (personal communication from manufacturers). We note a study by Chayoua et al that supports this. Using two sets of monoclonal antibodies that either recognize β2GPI G40-R43 in domain I or β2GPI irrespective of conformation, they showed that ACL AcuStar® CLIA and other solid phase assays demonstrated variable sensitivity to different antibody concentrations. ACL AcuStar® CLIA detected both types of monoclonal antibodies at much lower concentrations compared with QUANTA Lite® ELISA. 50 Clinically, the patients in our study who were aβ2GPI IgG CLIA positive were more likely to have had thrombosis than those who were ELISA positive (62.5% vs 45.8%), and more likely to be lupus anticoagulant positive, double or triple aPL positive and have co-existing autoimmune disease. The ability to detect more patients with aβ2GPI has significant implications for the diagnostic accuracy of APS.
Nonetheless, due to incomplete concordance between ELISA and CLIA, there is a possibility that APS may be classified differently in 10%–20% of cases using different platforms. Although higher sensitivity means that CLIA may detect more clinically relevant antibodies, it can also lead to more false positives leading to overdiagnosis. Fortunately, the impact of a single false negative or false positive test result will be lessened given that lupus anticoagulant and other aPL are often performed together, and persistence of aPL has to be demonstrated to diagnose APS. 50 Discrepancies between assays are more likely to be of low/medium titers, whereas higher titer aPL tend to be concordant across platforms. 50 This is within the context that isolated low titer aPL is increasingly recognized to carry less clinical significance, in contrast to the clinical importance of high titer aPL, lupus anticoagulant positivity or triple aPL positivity.1,51 Furthermore, Iwaniec et al found that the ability to identify triple aPL positive patients was similar irrespective of the solid phase assay platform used, and that APS diagnosis was concordant when the full aPL profile, instead of individual antibodies was assessed. 52
The main limitations of our study are the modest number of cases in the cohort, and the retrospective nature of the data collection for clinical outcomes which may mean missing or incomplete data. We used the revised Sapporo APS classification criteria to identify clinical manifestations but acknowledge that future studies could evaluate clinical manifestations defined using the 2023 ACR/EULAR updated classification criteria. The revised Sapporo APS clinical criteria for thrombosis did not distinguish between thrombosis with or without provoking risk factors, and hence provoked and unprovoked thrombosis were considered together in our study. The third limitation is the single time point of evaluation which can over-estimate the clinical significance of a positive result. Finally, this study evaluates only one ELISA system and one CLIA system and may not be fully applicable to other comparators.
Conclusions
Detection of antiphospholipid antibodies by CLIA shows a close but incomplete concordance with ELISA. Nevertheless, CLIA can detect clinically relevant aPL and may even be more sensitive than some ELISA platforms for APS, depending on the ability to detect antibodies against relevant epitopes on the β2GPI molecule. CLIA positivity on the ACL AcuStar® platform correlated well with moderate/high titer on ELISA, but apart from this, antibody titers should not be directly compared across systems. CLIA is an acceptable alternative to ELISA in the routine non-research setting for the diagnosis of APS.
Acknowledgements
The authors thank the staff of the Division of Haematology, Department of Laboratory Medicine, National University Hospital for assistance in the acquisition and collation of data.
Abbreviations
- aPL
Antiphospholipid antibodies
- aβ2GPI
β2GPI antibody
- aCL
Anti-cardiolipin antibody
- APS
Antiphospholipid syndrome
- CI
Confidence interval
- CLIA
chemiluminescent assay
- ELISA
enzyme-linked immunosorbent assay
- IgM
Immunoglobulin M
- IgG
Immunoglobulin G
- ROC
Receiver operating characteristics
Abstract Presentation: This paper has not been presented orally before at a professional meeting.
Author Contributions: Conceptualization: Shir Ying Lee, Jacqueline P.J. Lau; Methodology: Shir Ying Lee, Jacqueline P.J. Lau, Eunice E.N. Lai, Cheryl X.Q. Lim; Formal analysis and investigation: Eunice E.N. Lai, Cheryl X.Q. Lim, Jacqueline P.J. Lau, Shir Ying Lee; Writing - original draft preparation: Eunice E.N. Lai, Cheryl X.Q. Lim, Jacqueline P.J. Lau, Stephrene S.W. Chan; Writing - review and editing: Shir Ying Lee, Yen Lin Chee, Eng Soo Yap, Winnie Z.Y. Teo. All authors were involved in drafting or revising this article and approval of the final version to be published.
Data Availability: Data sets used in the study are available from the corresponding author on reasonable request
The authors declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.
Ethics Approval: This study involves human participants and was conducted in accordance with the Declaration of Helsinki. The Domain Specific Review Board of the National Healthcare Group approved this study and granted exemption from informed consent for this retrospective study (DSRB number 2023/00711).
Funding: The authors received no financial support for the research, authorship, and/or publication of this article.
ORCID iD: Eunice E.N. Lai https://orcid.org/0009-0009-5632-522X
Jacqueline P.J. Lau https://orcid.org/0009-0007-5332-4577
Yen Lin Chee https://orcid.org/0000-0002-8176-8382
Stephrene S.W. Chan https://orcid.org/0000-0003-4828-6303
Winnie Z.Y. Teo https://orcid.org/0000-0003-4356-6146
Eng Soo Yap https://orcid.org/0000-0002-7910-3154
Shir Ying Lee https://orcid.org/0000-0002-5119-8753
References
- 1.Garcia D, Erkan D. Diagnosis and management of the antiphospholipid syndrome. N Engl J Med. 2018;378(21):2010–2021. 10.1056/NEJMra1705454 [DOI] [PubMed] [Google Scholar]
- 2.de Laat HB, Derksen RH, Urbanus RT, Roest M, de Groot PG. beta2-glycoprotein I-dependent lupus anticoagulant highly correlates with thrombosis in the antiphospholipid syndrome. Blood. 2004;104(12):3598–3602. 10.1182/blood-2004-03-1107 [DOI] [PubMed] [Google Scholar]
- 3.McNeil HP, Simpson RJ, Chesterman CN, Krilis SA. Anti-phospholipid antibodies are directed against a complex antigen that includes a lipid-binding inhibitor of coagulation: Beta 2-glycoprotein I (apolipoprotein H). Proc Natl Acad Sci USA. 1990;87(11):4120–4124. 10.1073/pnas.87.11.4120 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Roubey RA, Pratt CW, Buyon JP, Winfield JB. Lupus anticoagulant activity of autoimmune antiphospholipid antibodies is dependent upon beta 2-glycoprotein I. J Clin Invest. 1992;90(3):1100–1104. 10.1172/JCI115926 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Meroni PL, Borghi MO. Antiphospholipid antibody assays in 2021: Looking for a predictive value in addition to a diagnostic one. Front Immunol. 2021;12:726820. 10.3389/fimmu.2021.726820 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Galli M, Comfurius P, Maassen C, et al. Anticardiolipin antibodies (ACA) directed not to cardiolipin but to a plasma protein cofactor. Lancet (London, England). 1990;335(8705):1544–1547. 10.1016/0140-6736(90)91374-j [DOI] [PubMed] [Google Scholar]
- 7.Miyakis S, Lockshin MD, Atsumi T, et al. International consensus statement on an update of the classification criteria for definite antiphospholipid syndrome (APS). J Thromb Haemostasis. 2006;4(2):295–306. 10.1111/j.1538-7836.2006.01753.x [DOI] [PubMed] [Google Scholar]
- 8.Barbhaiya M, Zuily S, Naden R, et al. The 2023 ACR/EULAR antiphospholipid syndrome classification criteria. Arthr Rheumatol. 2023;75(10):1687–1702. 10.1002/art.42624 [DOI] [PubMed] [Google Scholar]
- 9.Devreese KMJ, Ortel TL, Pengo V, de Laat B. Laboratory criteria for antiphospholipid syndrome: Communication from the SSC of the ISTH. J Thromb Haemostasis. 2018;16(4):809–813. 10.1111/jth.13976 [DOI] [PubMed] [Google Scholar]
- 10.Cinquanta L, Fontana DE, Bizzaro N. Chemiluminescent immunoassay technology: What does it change in autoantibody detection? Auto Immun Highlights. 2017;8(1):9. 10.1007/s13317-017-0097-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Favaloro EJ, Pasalic L, Lippi G. Classification criteria for the antiphospholipid syndrome: Not the same as diagnostic criteria for antiphospholipid syndrome. Semin Thromb Hemost. 2024;50(4):605–608. 10.1055/s-0043-1776318 [DOI] [PubMed] [Google Scholar]
- 12.Zhang S, Wu Z, Li P, Bai Y, Zhang F, Li Y. Evaluation of the clinical performance of a novel chemiluminescent immunoassay for detection of anticardiolipin and anti-Beta2-glycoprotein 1 antibodies in the diagnosis of antiphospholipid syndrome. Medicine (Baltimore). 2015;94(46):e2059. 10.1097/md.0000000000002059 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.De Moerloose P, Reber G, Musial J, Arnout J. Analytical and clinical performance of a new, automated assay panel for the diagnosis of antiphospholipid syndrome. J Thromb Haemostasis. 2010;8(7):1540–1546. 10.1111/j.1538-7836.2010.03857.x [DOI] [PubMed] [Google Scholar]
- 14.Li S, Choi J-L, Kim B-R, et al. Performance of automated chemiluminescence assay for antiphospholipid antibody testing. J Laborat Med Qual Assur. 2015;37(3):134–140. 10.15263/jlmqa.2015.37.3.134 [DOI] [Google Scholar]
- 15.Van Hoecke F, Persijn L, Decavele AS, Devreese K. Performance of two new, automated chemiluminescence assay panels for anticardiolipin and anti-beta2-glycoprotein I antibodies in the laboratory diagnosis of the antiphospholipid syndrome. Int J Lab Hematol. 2012;34(6):630–640. 10.1111/j.1751-553X.2012.01448.x [DOI] [PubMed] [Google Scholar]
- 16.Janek D, Slavik L, Ulehlova J, Krcova V, Hlusi A, Prochazkova J. Validation of a new panel of automated chemiluminescence assays for anticardiolipin antibodies in the screening for antiphospholipid syndrome. Clin Lab. 2016;62(7):1309–1315. 10.7754/Clin.Lab.2015.151129 [DOI] [PubMed] [Google Scholar]
- 17.Meneghel L, Ruffatti A, Gavasso S, et al. The clinical performance of a chemiluminescent immunoassay in detecting anti-cardiolipin and anti-β2 glycoprotein I antibodies. A comparison with a homemade ELISA method. Clin Chem Lab Med. 2015;53(7):1083–1089. 10.1515/cclm-2014-0925 [DOI] [PubMed] [Google Scholar]
- 18.Wan LY, Gu JY, Liu TT, et al. Clinical performance of automated chemiluminescent methods for anticardiolipin and anti-β2-glycoprotein I antibodies detection in a large cohort of Chinese patients with antiphospholipid syndrome. Int J Lab Hematol. 2020;42(2):206–213. 10.1111/ijlh.13156 [DOI] [PubMed] [Google Scholar]
- 19.Persijn L, Decavele AS, Schouwers S, Devreese K. Evaluation of a new set of automated chemiluminescense assays for anticardiolipin and anti-beta2-glycoprotein I antibodies in the laboratory diagnosis of the antiphospholipid syndrome. Thromb Res. 2011;128(6):565–569. 10.1016/j.thromres.2011.04.004 [DOI] [PubMed] [Google Scholar]
- 20.Aggarwal R, Ringold S, Khanna D, et al. Distinctions between diagnostic and classification criteria? Arthritis Care Res (Hoboken). 2015;67(7):891–897. 10.1002/acr.22583 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Tincani A, Allegri F, Balestrieri G, et al. Minimal requirements for antiphospholipid antibodies ELISAs proposed by the European Forum on antiphospholipid antibodies. Thromb Res. 2004;114(5-6):553–558. 10.1016/j.thromres.2004.06.035 [DOI] [PubMed] [Google Scholar]
- 22.Devreese KM, Pierangeli SS, de Laat B, Tripodi A, Atsumi T, Ortel TL. Testing for antiphospholipid antibodies with solid phase assays: Guidance from the SSC of the ISTH. J Thromb Haemostasis. 2014;12(5):792–795. 10.1111/jth.12537 [DOI] [PubMed] [Google Scholar]
- 23.Vandevelde A, Devreese KMJ. Laboratory diagnosis of antiphospholipid syndrome: Insights and hindrances. J Clin Med. 2022;11(8):2164. 10.3390/jcm11082164 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Vandevelde A, Chayoua W, de Laat B, et al. Semiquantitative interpretation of anticardiolipin and antiβ2glycoprotein I antibodies measured with various analytical platforms: Communication from the ISTH SSC Subcommittee on Lupus Anticoagulant/Antiphospholipid Antibodies. J Thromb Haemostasis. 2022;20(2):508–524. 10.1111/jth.15585 [DOI] [PubMed] [Google Scholar]
- 25.Bettacchioli E, Nafai S, Renaudineau Y. News and meta-analysis regarding anti-Beta 2 glycoprotein I antibodies and their determination. Clin Immunol. 2019;205:106–115. 10.1016/j.clim.2019.06.002 [DOI] [PubMed] [Google Scholar]
- 26.Hu C, Li S, Xie Z, et al. Comparison of different test systems for the detection of antiphospholipid antibodies in a Chinese cohort. Front Immunol. 2021;12:648881. 10.3389/fimmu.2021.648881 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Iwaniec T, Kaczor MP, Celińska-Löwenhoff M, Polański S, Musiał J. Clinical utility of automated chemiluminescent antiphospholipid antibody assay. Thromb Res. 2015;136(5):1033–1039. 10.1016/j.thromres.2015.08.021 [DOI] [PubMed] [Google Scholar]
- 28.Noubouossie D, Valsamis J, Corazza F, Rozen L, Debaugnies F, Demulder A. An automated chemiluminescence immunoassay may detect mostly relevant IgG anticardiolipin antibodies according to revised Sydney criteria. Acta Clin Belg. 2012;67(3):184–189. 10.2143/acb.67.3.2062653 [DOI] [PubMed] [Google Scholar]
- 29.Iverson GM, Reddel S, Victoria EJ, et al. Use of single point mutations in domain I of beta 2-glycoprotein I to determine fine antigenic specificity of antiphospholipid autoantibodies. J Immunol. 2002;169(12):7097–7103. 10.4049/jimmunol.169.12.7097 [DOI] [PubMed] [Google Scholar]
- 30.Pengo V, Biasiolo A, Fior MG. Autoimmune antiphospholipid antibodies are directed against a cryptic epitope expressed when beta 2-glycoprotein I is bound to a suitable surface. Thromb Haemost. 1995;73(1):29–34. [PubMed] [Google Scholar]
- 31.McNeeley PA, Dlott JS, Furie RA, et al. Beta2-glycoprotein I-dependent anticardiolipin antibodies preferentially bind the amino terminal domain of beta2-glycoprotein I. Thromb Haemost. 2001;86(2):590–595. [PubMed] [Google Scholar]
- 32.Matsuura E, Igarashi Y, Yasuda T, Triplett DA, Koike T. Anticardiolipin antibodies recognize beta 2-glycoprotein I structure altered by interacting with an oxygen modified solid phase surface. J Exp Med. 1994;179(2):457–462. 10.1084/jem.179.2.457 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.de Laat B, Derksen RH, van Lummel M, Pennings MT, de Groot PG. Pathogenic anti-beta2-glycoprotein I antibodies recognize domain I of beta2-glycoprotein I only after a conformational change. Blood. 2006;107(5):1916–1924. 10.1182/blood-2005-05-1943 [DOI] [PubMed] [Google Scholar]
- 34.Ninivaggi M, Kelchtermans H, Lindhout T, de Laat B. Conformation of beta2glycoprotein I and its effect on coagulation. Thromb Res. 2012;130(Suppl 1):S33–S36. 10.1016/j.thromres.2012.08.269 [DOI] [PubMed] [Google Scholar]
- 35.Agar C, van Os GM, Mörgelin M, et al. Beta2-glycoprotein I can exist in 2 conformations: Implications for our understanding of the antiphospholipid syndrome. Blood. 2010;116(8):1336–1343. 10.1182/blood-2009-12-260976 [DOI] [PubMed] [Google Scholar]
- 36.Sheng Y, Kandiah DA, Krilis SA. Anti-beta 2-glycoprotein I autoantibodies from patients with the “antiphospholipid” syndrome bind to beta 2-glycoprotein I with low affinity: Dimerization of beta 2-glycoprotein I induces a significant increase in anti-beta 2-glycoprotein I antibody affinity. J Immunol. 1998;161(4):2038–2043. [PubMed] [Google Scholar]
- 37.Meroni PL, Borghi MO, Raschi E, Tedesco F. Pathogenesis of antiphospholipid syndrome: Understanding the antibodies. Nat Rev Rheumatol. 2011;7(6):330–339. 10.1038/nrrheum.2011.52 [DOI] [PubMed] [Google Scholar]
- 38.Keil LB, Galazka M, el-Kadi HS, Erickson EN, DeBari J, A V. Binding of beta 2 glycoprotein I to activated polystyrene and its recognition by human IgG autoantibodies. Biotechnol Appl Biochem. 1995;22(3):305–313. [PubMed] [Google Scholar]
- 39.Devreese KM. Antiphospholipid antibody testing and standardization. Int J Lab Hematol. 2014;36(3):352–363. 10.1111/ijlh.12234 [DOI] [PubMed] [Google Scholar]
- 40.Pelkmans L, Kelchtermans H, de Groot PG, et al. Variability in exposure of epitope G40-R43 of domain i in commercial anti-beta2-glycoprotein I IgG ELISAs. PloS One. 2013;8(8):e71402. 10.1371/journal.pone.0071402 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.de Laat B, Derksen RH, Urbanus RT, de Groot PG. Igg antibodies that recognize epitope Gly40-Arg43 in domain I of beta 2-glycoprotein I cause LAC, and their presence correlates strongly with thrombosis. Blood. 2005;105(4):1540–1545. 10.1182/blood-2004-09-3387 [DOI] [PubMed] [Google Scholar]
- 42.Mahler M, Albesa R, Zohoury N, et al. Autoantibodies to domain 1 of beta 2 glycoprotein I determined using a novel chemiluminescence immunoassay demonstrate association with thrombosis in patients with antiphospholipid syndrome. Lupus. 2016;25(8):911–916. 10.1177/0961203316640366 [DOI] [PubMed] [Google Scholar]
- 43.Chighizola CB, Pregnolato F, Andreoli L, et al. Beyond thrombosis: Anti-β2GPI domain 1 antibodies identify late pregnancy morbidity in anti-phospholipid syndrome. J Autoimmun. 2018;90:76–83. 10.1016/j.jaut.2018.02.002 [DOI] [PubMed] [Google Scholar]
- 44.Durigutto P, Grossi C, Borghi MO, et al. New insight into antiphospholipid syndrome: Antibodies to β2glycoprotein I-domain 5 fail to induce thrombi in rats. Haematologica. 2019;104(4):819–826. 10.3324/haematol.2018.198119 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Andreoli L, Chighizola CB, Nalli C, et al. Clinical characterization of antiphospholipid syndrome by detection of IgG antibodies against β2 -glycoprotein i domain 1 and domain 4/5: Ratio of anti-domain 1 to anti-domain 4/5 as a useful new biomarker for antiphospholipid syndrome. Arthritis & Rheumatology. 2015;67(8):2196–2204. 10.1002/art.39187 [DOI] [PubMed] [Google Scholar]
- 46.de Laat B, Pengo V, Pabinger I, et al. The association between circulating antibodies against domain I of beta2-glycoprotein I and thrombosis: An international multicenter study. J Thromb Haemostasis. 2009;7(11):1767–1773. 10.1111/j.1538-7836.2009.03588.x [DOI] [PubMed] [Google Scholar]
- 47.Asherson RA, Cervera R. Antiphospholipid antibodies and infections. Ann Rheum Dis. 2003;62(5):388–393. 10.1136/ard.62.5.388 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Reber G, Tincani A, Sanmarco M, de Moerloose P, Boffa MC. Proposals for the measurement of anti-beta2-glycoprotein I antibodies. Standardization group of the European Forum on Antiphospholipid Antibodies. J Thromb Haemostasis. 2004;2(10):1860–1862. 10.1111/j.1538-7836.2004.00910.x [DOI] [PubMed] [Google Scholar]
- 49.Sciascia S, Willis R, Pengo V, et al. The comparison of real world and core laboratory antiphospholipid antibody ELISA results from antiphospholipid syndrome alliance for clinical trials & international networking (APS ACTION) clinical database and repository analysis. Thromb Res. 2019;175:32–36. 10.1016/j.thromres.2019.01.010 [DOI] [PubMed] [Google Scholar]
- 50.Chayoua W, Kelchtermans H, Moore GW, et al. Detection of anti-cardiolipin and anti-β2glycoprotein I antibodies differs between platforms without influence on association with clinical symptoms. Thromb Haemost. 2019;119(5):797–806. 10.1055/s-0039-1679901 [DOI] [PubMed] [Google Scholar]
- 51.Pengo V, Ruffatti A, Legnani C, et al. Clinical course of high-risk patients diagnosed with antiphospholipid syndrome. J Thromb Haemostasis. 2010;8(2):237–242. 10.1111/j.1538-7836.2009.03674.x [DOI] [PubMed] [Google Scholar]
- 52.Iwaniec T, Kaczor MP, Celińska-Löwenhoff M, Polański S, Musiał J. Identification of patients with triple antiphospholipid antibody positivity is platform and method independent. Pol Arch Med Wewn. 2016;126(1-2):19–24. 10.20452/pamw.3259 [DOI] [PubMed] [Google Scholar]