Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2025 Mar 24.
Published in final edited form as: Magn Reson Med. 2023 May 19;90(3):1121–1129. doi: 10.1002/mrm.29695

Prostate Perfusion Mapping using Fourier-Transform based Velocity-Selective Arterial Spin Labeling: Choice of Cutoff Velocity and Comparison with Brain

Dapeng Liu 1,2, Dan Zhu 1,2, Feng Xu 1,2, Farzad Sedaghat 1, Qin Qin 1,2
PMCID: PMC11932130  NIHMSID: NIHMS2059018  PMID: 37203405

Abstract

Purpose:

To develop velocity selective arterial spin labeling (VSASL) protocols for prostate blood flow (PBF) and prostate blood volume (PBV) mapping.

Methods:

Fourier-transform based velocity-selective inversion and saturation pulse trains were utilized in VSASL sequences to obtain blood flow and blood volume weighted perfusion signal, respectively. Here four cutoff velocities (Vcut = 0.25, 0.50, 1.00 and 1.50 cm/s) for PBF and PBV mapping sequences were evaluated with a parallel implementation in brain for measuring cerebral blood flow (CBF) and cerebral blood volume (CBV) with identical 3D readout. This study was performed at 3T on eight young and middle-aged healthy subjects comparing both perfusion weighted signal (PWS) and temporal SNR (tSNR).

Results:

In contrast to CBF and CBV, the PWS of PBF and PBV were rather unobservable at Vcut of 1.00 or 1.50 cm/s and both PWS and tSNR of PBF and PBV considerably increased at the lower Vcut, indicating that blood moves much slower in prostate than in brain. Similar to the brain results, the tSNR of PBV-weighted signal was about 2-4 times over the corresponding values of PBF-weighted signal. The results also suggested a trend of reduced vascularity within prostate during aging.

Conclusion:

For prostate, a low Vcut of 0.25-0.50 cm/s seemed necessary for both PBF and PBV measurements to obtain adequate perfusion signal. As in brain, PBV mapping yielded a higher tSNR than PBF.

Keywords: prostate perfusion, prostate blood flow, prostate blood volume, velocity-selective arterial spin labeling

Introduction

Prostate cancer is the most commonly occurring malignancy among American men, accounting for 27% of newly diagnosed cancers, and is the second leading cause male cancer mortality, accounting for 11% of cancer related death (1). Angiogenesis plays important roles for invasive proliferation of prostate carcinoma (2), which has been pathologically demonstrated by microvessel density counts (3-5). Tumor vascularity is an important radiologic biomarker utilized in prostate cancer detection, staging, and evaluation of therapeutic effect. The most well-established MRI methodology for quantitative perfusion and permeability measurements of prostate cancer is dynamic contrast-enhanced (DCE) MRI (6-8).

Arterial spin labeling (ASL) offers the ability to assess perfusion without injection of an exogenous contrast agent, potentially reducing imaging time and cost, and obviating risks related to cumulative gadolinium deposition. ASL is ideal for frequent non-invasive longitudinal monitoring and may serve as a viable alternative to commonly utilized DCE-MRI techniques. During the last decade, only the pulsed ASL (PASL) with 2D acquisition was explored for measuring prostate blood flow (PBF) (9-11).The pseudo-continuous ASL technique (12), which may offer higher perfusion sensitivity, has not been evaluated for prostate perfusion mapping, likely due to the requirement of manual positioning of the labeling plane and the susceptibility to long transit time delay.

To circumvent these restrictions, velocity-selective ASL (VSASL) was proposed to label all the blood in the vascular tree that is flowing faster than a certain cutoff velocity (Vcut) (13,14). The emerging Fourier-transfer based velocity-selective (FT-VS) pulse trains are able to saturate (FT-VSS) or invert (FT-VSI) static tissue while preserving spins flowing above a Vcut. FT-VSI has been applied for mapping cerebral blood flow (CBF) (15-19), as well as myocardial (20) and renal (21) blood flow. Additionally, by employing FT-VSS and removing the labeling delay in the VSASL sequence, cerebral blood volume (CBV) could be measured quantitatively with lower Vcut (22-24). Furthermore, the effectiveness of FT-VS labeling has also been demonstrated in cerebral (25,26), abdominal (27), and peripheral (28,29) MR angiography, along with a flowing phantom (30).

As the choice of Vcut determines the sensitivity to perfusion signal, the optimization of the VSASL labeling module for prostate perfusion mapping needs to be investigated first. In previous brain applications, Vcut = 2.0 cm/s was recommended for CBF mapping (14) and Vcut = 0.5 cm/s has been applied for CBV mapping (23,24); for prostate VSASL, lower Vcut values might be needed as slower blood movement in the prostate is expected. Here FT-VS prepared PBF and prostate blood volume (PBV) mapping sequences employing different Vcut and identical 3D acquisition were performed with a parallel implementation in brain for measuring CBF and CBV. This study aims to demonstrate the feasibility of mapping the PBF and PBV by comparing both perfusion weighted signal (PWS) and temporal SNR (tSNR) among different Vcut implementations.

Methods

Subjects

Four young males (26-40 years old) and four middle-aged males (59-66 years old) were included in this study after providing informed consent following the local Institutional Review Board guidelines. One middle-aged subject participated in the prostate experiment only, while all the other seven subjects participated in both brain and prostate experiments.

Scanner and sequences

Experiments were performed on a 3T scanner (Prisma, Siemens Healthineers, Erlangen, Germany) with a 20-channel head/neck receiving coil for brain scans and an 18-channel body receiving coil for prostate scans.

Both blood flow and blood volume mapping utilized 96 ms FT-VS label/control modules shown in Figure 1A. Nine excitation pulses (hard pulse, 0.06 ms duration each) with a flip angle of 20° (FT-VSI) or of 10° (FT-VSS) were interleaved with eight velocity encoding steps. Each step contained paired and phase-cycled composite refocusing pulses (90x-180y-90-x,1.0 ms duration), surrounded by four velocity-encoding triangular gradient lobes (1.2 ms duration, 0.6 ms ramp time each). Bipolar gradients were used for label, while unipolar gradients were used for control. Two young subjects participated in initial prostate scans comparing velocity encoding along superior-inferior and left-right directions with Vcut of 0.25 and 1.00 cm/s by varying the strength of gradients. For both brain and prostate scans in all subjects, four Vcut values (0.25, 0.50, 1.00 and 1.50 cm/s) were compared along superior-inferior direction. Note that here Vcut followed the definition in the first VSASL review paper (14).

Figure 1.

Figure 1.

(A) Diagram of Fourier-Transform based velocity-selective saturation (FT-VSS, 10° excitation) or inversion (FT-VSI, 20° excitation) pulse trains with bipolar gradients for label and unipolar gradients for control modules. Sequence diagrams of (B) FT-VSI based blood flow mapping and (C) FT-VSS based blood volume mapping. Both methods include slab-selective saturation with a post saturation delay, FT-VSI or FT-VSS label/control modules, regular fat saturation (not shown) followed by 3D single-shot GRASE readout with reduced FOV. (B) In blood flow mapping, post labeling delay with embedded background suppression pulses and a vascular crushing module were added between the label/control modules and readouts. (B) and (C) For prostate, a slab-selective saturation was added before the readout to eliminate the bladder's signal.

As previously described (16), the sequence for measuring blood flow consists five blocks (Figure 1B): a slab-selective saturation with 3.0 s post saturation delay (PSD), a FT-VSI based label/control module with 1.2 s post labeling delay (PLD), three background suppression pulses at 515/910/1075 ms after the labeling, a vascular crushing module with matched Vcut, and a 3D readout. The sequence for measuring blood volume consists of three blocks (Figure 1C): a slab-selective saturation with 3.5 s PSD, a FT-VSS based label/control module with PLD = 0 s, and a 3D readout. Product fat saturation module was applied before readout for both methods. For prostate, a slab-selective saturation was added before the readout to eliminate the bladder's signal (Figure 1B,C).

For both brain and prostate scans, the identical 3D gradient- and spin-echo (GRASE) readout was used with an imaging volume of 82×82×40 mm3 and resolution of 3.4×3.4×4.0 mm3, with a total of 10 slices. Note that this FOV was enough to cover the whole prostate gland, but only the size for the brain’s superior posterior portion. Inner-volume-excitation (31) was used to achieve a single-shot acquisition: slice oversampling = 20%, fast spin echo (FSE) factor = 12, echo planar imaging (EPI) factor = 13, bandwidth = 2894 Hz, echo time (TE) = 11.8 ms, echo spacing = 11.8 ms, and echo length = 142 ms. TR was set as 4.6 s for blood flow imaging and 3.9 s for blood volume imaging. Acquisition of 20 pairs of dynamics took 3.3 min for blood flow imaging and 2.8 min for blood volume imaging. If severe motion was observed during the scans, the affected sequences were repeated to double the number of dynamics acquired.

Proton density-weighted image (M0) was acquired with the same readout and TR of 10 s. For brain scans, a double inversion recovery (DIR) gray matter (GM) weighted image with two inversion pulses (TR = 10 s; TI1 = 3.58 s; TI2 = 0.48 s) to suppress both CSF and WM was also acquired. For prostate scans, a FSE sequence was performed in the axial orientation to get a structural T2 weighted image with the FOV of 200×200×105 mm3 and a resolution of 0.6×0.6×3.5 mm3: echo spacing = 11.2 ms, TE = 101 ms, TR = 8.7 s; the refocusing flip angles were 160°. With a GRAPPA (GeneRalized Autocalibrating Partial Parallel Acquisition) acceleration factor of 2, the scan time was 4 min.

Data Analysis

Data-processing was performed using Matlab (The Mathworks, Nantick, USA) and ImageJ (Rasband W., National Institute of Health, USA, version 1.51s) (32). To mitigate contamination from motion, subtractions of each label/control pairs were manually investigated. If a sudden, abnormal, and large-scale signal change was observed, this label/control pair would be identified as with motion artifacts and excluded from averaging. PWS were analyzed as (label-control)/M0 and tSNR were computed as the ratio of the mean to standard deviation values.

For brain imaging, the GM ROIs were generated by applying an empirical threshold on DIR images. White matter (WM) ROIs were manually drawn on DIR images. Large vessel signal was excluded from those ROIs using blood volume images with the lowest Vcut (0.25 cm/s) as references. CSF signal was also manually excluded based on M0 images.

For prostate imaging, ROIs were first manually drawn on the M0 images to include the prostate tissue without further specifying different regions. Large vessels were manually excluded according to blood volume images with the lowest Vcut (0.25 cm/s).

Results

The PBF and PBV based PWS images at one slice of two subjects indicated markedly higher perfusion signal at Vcut of 0.25 cm/s along superior-inferior velocity-encoding direction compared to results along left-right direction or at Vcut of 1.00 cm/s along either direction (Figure 2). Results of more slices are arrayed in Supporting Information Figure S1. The superior-inferior direction was used as default for the rest of the experiments.

Figure 2.

Figure 2.

The proton density-weighted images (M0) of prostate and the perfusion weighted signal (PWS) of prostate blood flow (PBF) and prostate blood volume (PBV) with Vcut of 0.25 and 1.00 cm/s along superior-inferior (S-I) and left-right (L-R) velocity-encoding directions at one axial slice of (A) subject 2 and (B) subject 4 from the young group, both showing higher perfusion signal at Vcut of 0.25 cm/s along S-I direction than the other three conditions.

Among both young and middle-aged groups, the numbers of dynamics excluded in each subject due to motion artifacts are listed in Supporting Information Table S1. Most exclusions were applied on the PBF weighted images with Vcut = 0.25 cm/s, with 14, 6, 5, 4 out of 40 dynamics for the four young subjects (the only four scans requiring repetition due to visible local motion during the scanning) and 2, 7, 6, 3 out of 20 dynamics for the four middle-aged subjects, respectively. All dynamics in PBV weighted images with different Vcut were included in the final analysis as no large motion artifacts were observed.

The CBF and CBV as well as PBF and PBV based PWS and tSNR images from one middle-aged subject are arrayed with four different Vcut at a single slice in Figure 3 as well as 4 out of 10 slices in Supporting Information Figure S2. The corresponding results at 4 slices of a young subject are displayed in Supporting Information Figure S3. One-slice examples of all young and middle-aged subjects are shown in Supporting Information Figures S4 and S5, respectively. As expected, PWS of blood flow and blood volume increased with lower Vcut for both brain GM, WM, and prostate tissues. tSNR showed a similar trend in PBF, PBV, and CBV, except that CBF’s tSNR decreased with reduced Vcut, indicating that FT-VSI labeled signal at very low Vcut was sensitive to physiological noise in the brain. Supplying arteries and draining veins appeared with hyperintensity in PWS images of both brain and prostate but presented rather low tSNR in prostate. In addition, there was also CSF signal visible through the middle of the brain in CBV weighted images, probably because the slow-flowing CSF got unintentionally labeled under lower Vcut.

Figure 3.

Figure 3.

M0, anatomical images (DIR: double inversion recovery), as well as blood flow and blood volume weighted perfusion weighted signal (PWS) and temporal SNR (tSNR) at different Vcut of (A) brain and (B) prostate of one axial slice of one subject (sub 2) from the middle-aged group. CBF: cerebral blood flow; CBV: cerebral blood volume; PBF: prostate blood flow; PBV: prostate blood volume.

The blood flow and blood volume based PWS and tSNR of GM, WM, and prostate tissues at different Vcut are shown in Figure 4 and listed in Table 1. 1) With Vcut increasing from 0.25 to 1.50 cm/s and within young/middle-aged groups: for brain, the CBF based PWS were relatively stable for GM and WM; the CBF based tSNR increased about 77% for GM and 19/80% for WM; the CBV based PWS decreased around 58% for GM and WM; the CBV based tSNR decreased around 44% for GM and around 52% for WM; for prostate, PWS decreased 69-82% for PBF, 78-83% for PBV, and tSNR decreased 58-68% for PBF, 72-75% for PBV. 2) Comparing blood volume to blood flow across different Vcut of both young and middle-aged groups, the enhanced PWS and tSNR were 1.4-3.4 and 1.7-7.3 times higher for GM, 2.7-8.4 and 2.6-13.0 times higher for WM, 2.0-3.4 and 2.1-4.3 times higher for prostate, respectively. 3) At Vcut of 0.25 cm/s: the PBF and PBV based PWS were 3.1 and 2.3 times of CBF and CBV based PWS in WM for the young-age group, and 2.5 and 1.2 times for the middle-aged group; the ratios of the middle-aged over young group of PWS of blood flow and blood volume were 0.87 and 1.19 times in GM, 0.93 and 1.12 times in WM, 0.73 and 0.60 times in prostate, respectively.

Figure 4.

Figure 4.

Averages of blood flow and blood volume perfusion weighted signal (PWS) and temporal SNR (tSNR) of gray matter (GM), white matter (WM) and prostate tissues for both young and middle-aged groups at different Vcut. The error bars are the standard deviation within the groups.

Table 1.

Mean and standard deviation of blood flow and blood volume perfusion weighted signal (PWS) and temporal SNR (tSNR) of gray matter (GM), white matter (WM) and prostate tissues for both young and middle-aged groups at different Vcut.

GM WM Prostate
Young group Middle-age group Young group Middle-age group Young group Middle-age group
Vcut
(cm/s)
PWS
(%)
tSNR PWS
(%)
tSNR PWS
(%)
tSNR PWS
(%)
tSNR PWS
(%)
tSNR PWS
(%)
tSNR
Blood flow 0.25 0.39±0.14 0.56±0.11 0.34±0.12 0.44±0.24 0.14±0.10 0.27±0.12 0.13±0.11 0.15±0.08 0.44±0.10 0.25±0.07 0.32±0.11 0.24±0.18
0.50 0.30±0.03 0.78±0.12 0.28±0.05 0.67±0.34 0.10±0.05 0.36±0.21 0.06±0.02 0.17±0.08 0.18±0.06 0.15±0.07 0.16±0.08 0.12±0.11
1.00 0.30±0.03 1.00±0.24 0.29±0.05 0.70±0.38 0.10±0.02 0.42±0.13 0.10±0.02 0.29±0.13 0.07±0.04 0.07±0.05 0.10±0.09 0.12±0.13
1.50 0.29±0.02 0.99±0.22 0.28±0.05 0.78±0.41 0.07±0.02 0.32±0.15 0.07±0.05 0.27±0.17 0.08±0.06 0.08±0.06 0.10±0.04 0.07±0.05
Blood volume 0.25 0.96±0.09 2.90±0.68 1.14±0.20 3.23±0.64 0.59±0.10 2 25±0.51 0.66±0.13 1.90±0.10 1.33±0.57 1.00±0.29 0.80±0.52 0.85±0.75
0.50 0.72±0.09 2.54±0.58 0.86±0.12 2.94±0.46 0.39±0.07 1.59±0.43 0.48±0.11 1.52±0.15 0.49±0.18 0.49±0.16 0.35±0.20 0.41±0.39
1.00 0.50±0.11 1.83±0.52 0.61±0.04 2.28±0.47 0.27±0.02 1.10±0.10 0.33±0.01 1.11±0.20 0.23±0.07 0.30±0.09 0.18±0.12 0.25±0.21
1.50 0.41±0.08 1.65±0.38 0.49±0.01 1.78±0.53 0.25±0.01 1.08±0.15 0.28±0.01 0.93±0.21 0.22±0.06 0.28±0.08 0.18±0.04 0.21±0.15

Discussion

This is the first work systematically investigating the blood flow and blood volume weighted perfusion mapping using FT-VS based ASL techniques in prostate and comparing them with the results in brain. The dependence on both the direction (Figure 2) and Vcut (Figures 3, 4, Supporting Information Figure S1-S3, Table 1) indicated that blood moves much slower in prostate than in brain. Cerebral applications of VSASL were previously shown to be insensitive to the direction of velocity encoding at Vcut of 1-2 cm/s for CBF (13) and of 0.5 cm/s for CBV (23). For CBF, when Vcut was lower than 1.00-1.50 cm/s, the PWS gain was minimum while tSNR dropped and CSF contamination increased (Figure 3). This is consistent with the current recommendation of 2 cm/s Vcut in the brain (14). For CBV, there was improvement in both PWS and tSNR with lower Vcut (Figures 3,4, Information Figure S1-S3, Table 1). However, the CBV with higher Vcut still preserved GM-to-WM contrast (Figure 3) and may have the benefit of less sensitivity to physiological motion. In contrast, for prostate VSASL, velocity encoding direction even at Vcut of 0.25 cm/s still showed some degree of anisotropy in labeling efficiency (Figure 2). The PWS of PBF and PBV were rather unobservable at Vcut of 1.00-1.50 cm/s and both PWS and tSNR of PBF and PBV considerably increased at the lower Vcut (Figures 3,4, Supporting Information Figure S1-S3, Table 1). Therefore, a Vcut around 0.25-0.50 cm/s along superior-inferior direction was necessary for VS based prostate perfusion imaging.

For both brain and prostate, blood volume weighted signal yielded much higher PWS and tSNR than blood flow weighted results (Figure 4). The tSNR of PBV-weighted signal was about 2-4 times over the corresponding values of PBF-weighted signal (Table 1). This was largely because that blood signal was acquired right after labeling without experiencing any T1 relaxation. With blood T1 for males as about 1.8 s at 3T (33), the labeled blood magnetization would exponentially decay ~50% during 1.2 s PLD employed in the VSASL sequence for blood flow estimation (Figure 1b). Given the inherently low SNR suffered by ASL methods, the markedly enhanced tSNR by blood volume based VSASL could be especially advantageous for cancer detection in brain, prostate, or any other tissue organs. One challenge though was the differentiation between large vessels and tumor in PWS images, especially in organs such as prostate, where the local vasculature anatomy is not routinely imaged and thus may not be familiar to radiologists. This could potentially be resolved by comparing with the corresponding tSNR images (Figure 3, Supporting Information Figure S1-S3) as tSNR of large vessels were much lower than that of the static tissue, likely due to the effect of pulsation. By all accounts the spatially connecting features of vessels would also help correct characterizations.

Note that absolute quantification of PBF and PBV were not conducted in this work due to still limited knowledge of labeling efficiency at various vessel segments at different Vcut and lack of literature values for comparison. However, as identical VSASL sequences were performed in both brain and prostate, their PWS could be compared among different tissues within the same subject or between young and middle-age groups. In this preliminary analysis, the PWS of normal prostate among middle-aged subjects had about 2.5 times higher blood flow and 1.2 times higher blood volume than normal white matter, and about 27-40% lower blood flow and blood volume than younger subjects, indicating a trend of reduced vascularity within prostate during aging.

As the diffusion weighting between the pair of velocity-selective labeling and velocity-compensated control modules was only partially matched (Figure 1a), the effect of diffusion-imbalance on the subtracted perfusion signal needs to be estimated. Based on the b-values of a single velocity-encoding step of the label and control modules for each Vcut (Supporting Information Table S2) and the normal T1, T2, and ADC values of GM, WM, and prostate (Supporting Information Table S3), the diffusion-induced signal bias among various cerebral and prostate PBS results (Table 1) were compared in Supporting Information Table S4. Considering the lowest Vcut of 0.25 cm/s bearing the highest unbalanced diffusion weighting, as the employed b-values were quite small for a single velocity-encoding step (label: 1.45 s/mm2; control: 0.56 s/mm2), their diffusion effects contributed relatively 14-18% of blood flow based PWS and 9-11% of blood volume based PWS in GM, 17-23% of blood flow based PWS and 14-15% of blood volume based PWS in WM, as well as 20-30% of blood flow based PWS and 10-17% of blood volume based PWS in prostate. This bias of overestimation reduced considerably with the increasing Vcut. It is worth noting that this diffusion-related bias for the absolute perfusion quantification is not going to affect the perfusion contrast of prostate tumors at such low b-values (< 2 s/mm2), as b-values higher than 600 s/mm2 are typically required to detect significant ADC differences between benign tissue and prostate cancer (34)

Another problem from low Vcut is the higher susceptibility to motion near prostate, which required manual inspection and abandoning failed acquisitions. The much less motion-related artifacts in PBV images (Supporting Information Table S1) may be partly due to its higher PWS and better suppressed tissue signal. For PBF, more aggressive background suppression may be needed to improve its robustness to motion.

In addition, a low resolution of 3.4×3.4×4.0 mm3, which is typical for brain, was used in this study for optimizing the Vcut with a single-shot readout. Such a resolution may not be sufficient for clinical applications on a small organ like prostate. Increasing the resolution without being blurred will require more undersampled readout. The reproducibility of VSASL based perfusion mapping for prostate will need to be established as having been done for brain recently (35).

Conclusion

Compared to in brain, blood moves much slower in prostate and a low Vcut of 0.25-0.50 cm/s along superior-inferior direction seemed necessary to obtain adequate perfusion signal for both PBF and PBV measurements. As in brain, PBV mapping yielded a higher tSNR than PBF. Both methods need to be evaluated in patients to demonstrate their clinical utilities for characterizing prostate cancer.

Supplementary Material

supporting information

Table S1: Number of excluded/acquired pairs of label and control dynamics of each subject by the VSASL sequences applied for brain and prostate blood flow and volume, respectively.

Figure S1: The proton density-weighted images (M0) of prostate and the perfusion weighted signal (PWS) of prostate blood flow (PBF) and prostate blood volume (PBV) with Vcut of 0.25 and 1.00 cm/s along superior-inferior (S-I) and left-right (L-R) velocity-encoding directions at 4 out of 10 axial slices of (A) subject 2 and (B) subject 4 from the young group, both showing higher perfusion signal at Vcut of 0.25 cm/s along S-I direction than the other three conditions.

Figure S2: M0, anatomical images (DIR: double inversion recovery), as well as blood flow and blood volume weighted perfusion weighted signal (PWS) and temporal SNR (tSNR) at different Vcut of (A) brain and (B) prostate of 4 out of 10 axial slices of one subject (sub 2) from the young group. CBF: cerebral blood flow; CBV: cerebral blood volume; PBF: prostate blood flow; PBV: prostate blood volume.

Figure S3: M0, anatomical images (DIR: double inversion recovery), as well as blood flow and blood volume weighted perfusion weighted signal (PWS) and temporal SNR (tSNR) at different Vcut of (A) brain and (B) prostate of 4 out of 10 axial slices of one subject (sub 2) from the middle-age group. CBF: cerebral blood flow; CBV: cerebral blood volume; PBF: prostate blood flow; PBV: prostate blood volume.

Figure S4: M0, anatomical images (DIR: double inversion recovery), as well as blood flow and blood volume weighted perfusion weighted signal (PWS) and temporal SNR (tSNR) at different Vcut of (A) brain and (B) prostate of all subjects from the young group. Slice 9 was shown for brains and slice 6 was shown for prostate. CBF: cerebral blood flow; CBV: cerebral blood volume; PBF: prostate blood flow; PBV: prostate blood volume.

Figure S5: M0, anatomical images (DIR: double inversion recovery), as well as blood flow and blood volume weighted perfusion weighted signal (PWS) and temporal SNR (tSNR) at different Vcut of (A) brain and (B) prostate of all subjects from the middle-aged group. Slice 9 was shown for brains and slice 6 was shown for prostate. CBF: cerebral blood flow; CBV: cerebral blood volume; PBF: prostate blood flow; PBV: prostate blood volume.

Figure S6: Fitted relationship between (Tseg/T2+b·ADC) and Mz. Dots are Mz calculated from Equation [S5] with matrix form. Solid lines are the fitted Equation [S6] with empirical four-parameter models. Dotted lines are (Tseg/T2+b·ADC) and Mz of the label module of Vcut = 0.25 cm/s calculated from the empirical four-parameter models.

Table S2: The b-values of single velocity-encoding steps (Figure 1a) in this study calculated using the original integration equation for arbitrary gradient waveforms (Eq. [9.7] of Bernstein MA, King KF, Zhou ZJ. Handbook of MRI pulse sequences. Amsterdam; Boston: Academic Press; 2004.) and assuming perfect 180° refocusing pulses.

Table S3: Parameters used for estimating the diffusion bias.

Table S4: Estimation of diffusion bias for gray matter, white matter, and prostate following Equations [S1-S8]. The factor (Tseg/T2+b·ADC) with Tseg of 12 ms used in this study, the corresponding Mzl of label and control modules for FT-VSS and FT-VSI respectively (Equation [S6]), their difference Mzl_diff (Equation [S7]), the final Mzf_diff (Equation [S8], PSD = 3.0, 3.5 s for blood flow, blood volume estimation as used in this study), and the diffusion-induced bias relative to the blood flow or blood volume based perfusion weighted signal (PWS) results among both the young and middle aged groups listed in Table 1.

Grant support:

NIH: R01 HL138182 (QQ), NIH: R01 HL144751 (QQ);

Reference

  • 1.Siegel RL, Miller KD, Fuchs HE, Jemal A. Cancer statistics, 2022. CA A Cancer J Clinicians 2022;72:7–33. [DOI] [PubMed] [Google Scholar]
  • 2.Russo G, Mischi M, Scheepens W, De la Rosette JJ, Wijkstra H. Angiogenesis in prostate cancer: onset, progression and imaging. BJU International 2012;110:E794–E808. [DOI] [PubMed] [Google Scholar]
  • 3.Brawer MK, Deering RE, Brown M, Preston SD, Bigler SA. Predictors of pathologic stage in prostatic carcinoma. The role of neovascularity. Cancer 1994;73:678–687. [DOI] [PubMed] [Google Scholar]
  • 4.Borre M, Offersen B, Nerstrøm B, Overgaard J. Microvessel density predicts survival in prostate cancer patients subjected to watchful waiting. Br J Cancer 1998;78:940–944. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Bono AV, Celato N, Cova V, Salvadore M, Chinetti S, Novario R. Microvessel density in prostate carcinoma. Prostate Cancer Prostatic Dis 2002;5:123–127. [DOI] [PubMed] [Google Scholar]
  • 6.Bonekamp D, Macura KJ. Dynamic Contrast-Enhanced Magnetic Resonance Imaging in the Evaluation of the Prostate. Topics in Magnetic Resonance Imaging 2008;19:273–284. [DOI] [PubMed] [Google Scholar]
  • 7.Bonekamp D, Jacobs MA, El-Khouli R, Stoianovici D, Macura KJ. Advancements in MR Imaging of the Prostate: From Diagnosis to Interventions. RadioGraphics 2011;31:677–703. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Verma S, Turkbey B, Muradyan N, Rajesh A, Cornud F, Haider MA, et al. Overview of Dynamic Contrast-Enhanced MRI in Prostate Cancer Diagnosis and Management. American Journal of Roentgenology 2012;198:1277–1288. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Li X, Metzger GJ. Feasibility of measuring prostate perfusion with arterial spin labeling. NMR in Biomedicine 2013;26:51–57. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Cai W, Li F, Wang J, Du H, Wang X, Zhang J, et al. A comparison of arterial spin labeling perfusion MRI and DCE-MRI in human prostate cancer. NMR in Biomedicine 2014;27:817–825. [DOI] [PubMed] [Google Scholar]
  • 11.Boschheidgen M, Schimmöller L, Kasprowski L, Abrar D, Arsov C, Albers P, et al. Arterial spin labelling as a gadolinium-free alternative in the detection of prostate cancer. Magnetic Resonance Imaging 2021;80:33–38. [DOI] [PubMed] [Google Scholar]
  • 12.Dai W, Garcia D, de Bazelaire C, Alsop DC. Continuous flow-driven inversion for arterial spin labeling using pulsed radio frequency and gradient fields. Magnetic Resonance in Medicine 2008;60:1488–1497. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Wong EC, Cronin M, Wu WC, Inglis B, Frank LR, Liu TT. Velocity-selective arterial spin labeling. Magnetic Resonance in Medicine 2006;55:1334–1341. [DOI] [PubMed] [Google Scholar]
  • 14.Qin Q, Alsop DC, Bolar DS, Hernandez-Garcia L, Meakin J, Liu D, et al. Velocity-selective arterial spin labeling perfusion MRI: A review of the state of the art and recommendations for clinical implementation. Magnetic resonance in medicine 2022;88:1528–1547. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Qin Q, van Zijl PCM. Velocity-selective-inversion prepared arterial spin labeling. Magnetic Resonance in Medicine 2016;76:1136–1148. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Liu D, Xu F, Li W, Zijl PC, Lin DD, Qin Q. Improved velocity-selective-inversion arterial spin labeling for cerebral blood flow mapping with 3D acquisition. Magn Reson Med 2020;84:2512–2522. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Liu D, Li W, Xu F, Zhu D, Shin T, Qin Q. Ensuring both velocity and spatial responses robust to field inhomogeneities for velocity-selective arterial spin labeling through dynamic phase-cycling. Magn Reson Med 2021;85:2723–2734. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Liu D, Jiang D, Tekes A, Kulikowicz E, Martin LJ, Lee JK, et al. Multi-Parametric Evaluation of Cerebral Hemodynamics in Neonatal Piglets Using Non-Contrast-Enhanced Magnetic Resonance Imaging Methods. J Magn Reson Imaging 2021;54:1053–1065. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Qu Y, Kong D, Wen H, Ou X, Rui Q, Wang X, et al. Perfusion measurement in brain gliomas using velocity-selective arterial spin labeling: comparison with pseudo-continuous arterial spin labeling and dynamic susceptibility contrast MRI. Eur Radiol 2022;32:2976–2987. [DOI] [PubMed] [Google Scholar]
  • 20.Landes V, Javed A, Jao T, Qin Q, Nayak K. Improved velocity-selective labeling pulses for myocardial ASL. Magn Reson Med 2020;84:1909–1918. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Franklin SL, Bones IK, Harteveld AA, Hirschler L, van Stralen M , Qin Q, et al. Multi-organ comparison of flow-based arterial spin labeling techniques: Spatially non-selective labeling for cerebral and renal perfusion imaging. Magn Reson Med 2021;85:2580–2594. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Liu D, Xu F, Lin DD, van Zijl PCM, Qin Q. Quantitative measurement of cerebral blood volume using velocity-selective pulse trains. Magn. Reson. Med 2017;77:92–101. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Qin Q, Qu Y, Li W, Liu D, Shin T, Zhao Y, et al. Cerebral blood volume mapping using Fourier-transform-based velocity-selective saturation pulse trains. Magnetic Resonance in Medicine 2019;81:3544–3554. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Li W, Liu D, Zijl PCM, Qin Q. Three-dimensional whole-brain mapping of cerebral blood volume and venous cerebral blood volume using Fourier transform–based velocity-selective pulse trains. Magnetic Resonance in Med 2021;86:1420–1433. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Qin Q, Shin T, Schär M, Guo H, Chen H, Qiao Y. Velocity-selective magnetization-prepared non-contrast-enhanced cerebral MR angiography at 3 Tesla: Improved immunity to B0/B1 inhomogeneity. Magnetic Resonance in Medicine 2016;75:1232–1241. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Li W, Xu F, Schär M, Liu J, Shin T, Zhao Y, et al. Whole-brain arteriography and venography: Using improved velocity-selective saturation pulse trains. Magnetic Resonance in Medicine 2018;79:2014–2023. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Zhu D, Li W, Liu D, Liu G, Pei Y, Shin T, et al. Non-contrast-enhanced abdominal MRA at 3 T using velocity-selective pulse trains. Magn Reson Med 2020;84:1173–1183. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Shin T, Qin Q, Park J-Y, Crawford RS, Rajagopalan S. Identification and reduction of image artifacts in non-contrast-enhanced velocity-selective peripheral angiography at 3T. Magnetic Resonance in Medicine 2016;76:466–477. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Shin T, Qin Q. Characterization and suppression of stripe artifact in velocity-selective magnetization-prepared unenhanced MR angiography. Magnetic Resonance in Medicine 2018;80:1997–2005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Xu F, Zhu D, Fan H, Lu H, Liu D, Li W, et al. Magnetic resonance angiography and perfusion mapping by arterial spin labeling using Fourier transform–based velocity-selective pulse trains: Examination on a commercial perfusion phantom. Magn. Reson. Med 2021;86:1360–1368. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Feinberg DA, Hoenninger JC, Crooks LE, Kaufman L, Watts JC, Arakawa M. Inner volume MR imaging: technical concepts and their application. Radiology 1985;156:743–747. [DOI] [PubMed] [Google Scholar]
  • 32.Schneider CA, Rasband WS, Eliceiri KW. NIH Image to ImageJ: 25 years of image analysis. Nature methods 2012;9:671–5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Li W, Liu P, Lu H, Strouse JJ, van Zijl PCM, Qin Q. Fast measurement of blood T 1 in the human carotid artery at 3T: Accuracy, precision, and reproducibility. Magnetic Resonance in Medicine 2017;77:2296–2302. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Esen M, Onur MR, Akpolat N, Orhan I, Kocakoc E. Utility of ADC measurement on diffusion-weighted MRI in differentiation of prostate cancer, normal prostate and prostatitis. Quantitative Imaging in Medicine and Surgery 2013;3:210–216. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Xu F, Liu D, Zhu D, Hillis AE, Bakker A, Soldan A, et al. Test-Retest Reliability of 3D Velocity-Selective Arterial Spin Labeling for Detecting Normal Variations of Cerebral Blood Flow. Neuroimage 2023:120039. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

supporting information

Table S1: Number of excluded/acquired pairs of label and control dynamics of each subject by the VSASL sequences applied for brain and prostate blood flow and volume, respectively.

Figure S1: The proton density-weighted images (M0) of prostate and the perfusion weighted signal (PWS) of prostate blood flow (PBF) and prostate blood volume (PBV) with Vcut of 0.25 and 1.00 cm/s along superior-inferior (S-I) and left-right (L-R) velocity-encoding directions at 4 out of 10 axial slices of (A) subject 2 and (B) subject 4 from the young group, both showing higher perfusion signal at Vcut of 0.25 cm/s along S-I direction than the other three conditions.

Figure S2: M0, anatomical images (DIR: double inversion recovery), as well as blood flow and blood volume weighted perfusion weighted signal (PWS) and temporal SNR (tSNR) at different Vcut of (A) brain and (B) prostate of 4 out of 10 axial slices of one subject (sub 2) from the young group. CBF: cerebral blood flow; CBV: cerebral blood volume; PBF: prostate blood flow; PBV: prostate blood volume.

Figure S3: M0, anatomical images (DIR: double inversion recovery), as well as blood flow and blood volume weighted perfusion weighted signal (PWS) and temporal SNR (tSNR) at different Vcut of (A) brain and (B) prostate of 4 out of 10 axial slices of one subject (sub 2) from the middle-age group. CBF: cerebral blood flow; CBV: cerebral blood volume; PBF: prostate blood flow; PBV: prostate blood volume.

Figure S4: M0, anatomical images (DIR: double inversion recovery), as well as blood flow and blood volume weighted perfusion weighted signal (PWS) and temporal SNR (tSNR) at different Vcut of (A) brain and (B) prostate of all subjects from the young group. Slice 9 was shown for brains and slice 6 was shown for prostate. CBF: cerebral blood flow; CBV: cerebral blood volume; PBF: prostate blood flow; PBV: prostate blood volume.

Figure S5: M0, anatomical images (DIR: double inversion recovery), as well as blood flow and blood volume weighted perfusion weighted signal (PWS) and temporal SNR (tSNR) at different Vcut of (A) brain and (B) prostate of all subjects from the middle-aged group. Slice 9 was shown for brains and slice 6 was shown for prostate. CBF: cerebral blood flow; CBV: cerebral blood volume; PBF: prostate blood flow; PBV: prostate blood volume.

Figure S6: Fitted relationship between (Tseg/T2+b·ADC) and Mz. Dots are Mz calculated from Equation [S5] with matrix form. Solid lines are the fitted Equation [S6] with empirical four-parameter models. Dotted lines are (Tseg/T2+b·ADC) and Mz of the label module of Vcut = 0.25 cm/s calculated from the empirical four-parameter models.

Table S2: The b-values of single velocity-encoding steps (Figure 1a) in this study calculated using the original integration equation for arbitrary gradient waveforms (Eq. [9.7] of Bernstein MA, King KF, Zhou ZJ. Handbook of MRI pulse sequences. Amsterdam; Boston: Academic Press; 2004.) and assuming perfect 180° refocusing pulses.

Table S3: Parameters used for estimating the diffusion bias.

Table S4: Estimation of diffusion bias for gray matter, white matter, and prostate following Equations [S1-S8]. The factor (Tseg/T2+b·ADC) with Tseg of 12 ms used in this study, the corresponding Mzl of label and control modules for FT-VSS and FT-VSI respectively (Equation [S6]), their difference Mzl_diff (Equation [S7]), the final Mzf_diff (Equation [S8], PSD = 3.0, 3.5 s for blood flow, blood volume estimation as used in this study), and the diffusion-induced bias relative to the blood flow or blood volume based perfusion weighted signal (PWS) results among both the young and middle aged groups listed in Table 1.

RESOURCES