Abstract
Over the past decades, computed tomography (CT) imaging has profited from various technical innovations. Besides improvements such as higher temporal and spatial resolutions, lower radiation dose, and the introduction of dual- and multi-energy imaging, the development and recent clinical introduction of photon-counting detector CT (PCD-CT) represents a milestone with the potential to substantially change clinical CT imaging and expand its indications. This thematic series of European Radiology Experimental comprises a collection of original research papers and review articles demonstrating the benefits and challenges of this cutting-edge technology. The thematic series includes a wide range of relevant topics spanning from initial clinical experiences using PCD-CT to original research papers covering potential applications in various body regions.
Keywords: Computed tomography angiography, Disruptive technology, Photon-counting detector, Radiation dosage, Tomography (x-ray computed)
The field of medical imaging has witnessed several revolutionary advancements over the past decades. Some of these, such as the latest innovation, photon-counting detector computed tomography (PCD-CT), hold the promise to fundamentally transform clinical practice [1, 2]. PCD-CT represents a significant departure from conventional energy-integrating detector (EID) CT systems, which measure the total energy deposited by incoming x-rays without distinguishing between individual photons. EID-CT systems, therefore, have inherent limitations, such as limited spatial resolution and contrast-to-noise, especially at low radiation doses. In contrast, PCD-CT systems operate on a fundamentally different principle by counting individual x-ray photons and measuring the energy of each of them. This technological landscape paves the way for significant advancements in both experimental and clinical imaging, facilitating the transition of phantom studies into routine clinical applications. This sets the stage for the new thematic series in European Radiology Experimental titled “Photon-counting detector CT: a disrupting innovation in medical imaging,” which we are pleased to introduce with this editorial.
The disruptive potential of PCD-CT lies in its ability to not only address the limitations of EID-based systems but also to introduce new capabilities. By directly counting photons and categorizing them based on their energy, PCD-CT can produce images with higher spatial resolution, better contrast, and reduced noise [1]. Such improvements have the potential to dramatically enhance the diagnostic power of CT imaging across a wide range of clinical applications.
One of the most significant benefits of PCD-CT is its ability to improve image quality while simultaneously reducing radiation dose. By more accurately measuring the energy of incoming photons, PCDs can enhance tissue differentiation, yielding more detailed images in both spatial resolution and contrast resolution compared to conventional dual-energy EID-based systems, which primarily rely on superior spectral imaging capabilities.
The ability to perform energy-resolved imaging allows for the generation of various spectral-based reconstructions, including virtual mono-energetic images, iodine maps, virtual non-iodine, or virtual non-calcium images. This capability is especially valuable in detecting calcifications, distinguishing between types of tissue, and identifying contrast agents, all of which may be critical in accurate diagnosis and treatment planning. While dual-energy EID-CT systems can also deliver spectral information, the dual-source setup of the PCD-CT system with its temporal resolution of 66 ms (compared to 125 ms on a third-generation dual-source system in dual-energy mode) makes the use of spectral imaging feasible even for cardiac imaging. PCD-CT also reduces beam-hardening artifacts, which is achieved because PCDs can more effectively discriminate between low- and high-energy photons, minimizing the confounding effects of beam hardening [3, 4]. Another notable advantage is the potential for contrast media dose reduction. Using virtual mono-energetic image reconstructions, a 50% reduction in contrast media dose can still achieve diagnostic level attenuation for coronary CT angiography, as shown in a dynamic phantom study [5].
The unique benefits of PCD-CT make its use particularly advantageous in several clinical areas. Cardiovascular imaging, for instance, stands to benefit greatly from the enhanced spatial resolution (ultra-high resolution, i.e., 120 × 0.2 mm collimation) and spectral imaging capabilities of PCD-CT. The ability to quantify stenosis more accurately [6] and clearly visualize coronary artery lumen even in the presence of heavy calcification or stents [7] leads to a higher positive predictive value and a lower rate of unnecessary referrals for invasive testing (> 50% of patients reclassified to a lower risk category when using ultra-high resolution in a recent patient study), which ultimately result in the better management of patients with coronary artery disease [8].
Such benefits are also applicable to the various vascular structures across the body [9, 10]. PCD-CT may also be beneficial for the differentiation among various types of plaques, leading to earlier and more accurate characterization of coronary artery disease [11–13]. Finally, using PCD-CT, the ability to characterize the myocardium has also shown improvements [14, 15].
PCD-CT also holds promise in musculoskeletal imaging, especially for the imaging of large joints at ultra-high spatial resolution, for the quantification of gout deposits and the visualization of bone edema using its spectral capability, and for imaging of orthopedic implants due to its improved metal artifact reduction [16]. PCD-CT has been shown to deliver benefits for the accurate assessment if osseointegration of orthopedic joint replacement implants [17], and for the direct visualization of both cartilage and bone details of the knee using virtual mono-energetic images [18]. It also provides superior objective and subjective image quality and visualization for trabecular bone structure of the elbow and the radius, as well as for fracture depiction, compared to EID-CT [19, 20].
Without providing a comprehensive overview, all other areas of radiology have the potential to gain benefits from this new technology. The growing body of literature demonstrates the increasing research and clinical interest in PCD-CT, spanning from neuroradiology [21] through breast imaging [22] to a wide range of applications in pediatric imaging [23]. Latter especially benefits from the reduced radiation dose [24], combined with high-quality imaging, addressing the critical need for safe and effective imaging techniques in children, who are more sensitive to radiation and have a longer lifespan during which radiation-induced risks could manifest.
As we look into the future, the expansion of potential PCD-CT applications seems realistic. One area of ongoing research is the development of multicontrast imaging; i.e., a single scan imaging after the simultaneous injection of multiple contrast agents, such as iodine, gadolinium, or bismuth [25, 26]. Such multicontrast imaging, for example, can be used to visualize arterial and portal venous enhancement in a single scan to characterize liver lesions [27]. Further technical improvements, however, are expected to enable k-edge imaging tuned for non-iodine contrast agents [28].
Another area where we see the field is moving toward is the integration of PCD-CT with artificial intelligence (AI)-based applications. AI-driven deep learning-based noise reduction is one great example that can tackle the higher image noise especially observed in ultra-high-resolution images, which has been demonstrated in patients with multiple myeloma [29]. Beyond improving image quality, AI may also contribute to more granular quantitative image evaluation. PCD-CT-based radiomics, although not necessarily AI-based, however, often AI-powered, has been extensively investigated with the potential of identifying new imaging biomarkers in various conditions [30–32].
In the long run, as PCD-CT becomes more widely adopted, we may see the development of new clinical protocols and guidelines that take full advantage of the technology’s capabilities. This could lead to a shift in how certain conditions are diagnosed and treated, with PCD-CT playing a central role in the imaging toolbox. However, the widespread adoption of PCD-CT will also require addressing several challenges. These include the high cost of the technology, the need for specialized technologist training, and the development of standardized imaging protocols, which will most likely relevantly differ from conventional EID-CT protocols. As with any disruptive technology, there will be a period of adjustment as the medical community learns to fully integrate PCD-CT into clinical practice.
PCD-CT is more than just an incremental improvement over traditional EID-CT technology; it is a disruptive innovation that has the potential to transform CT imaging. Its ability to produce higher-quality images with lower radiation doses, reduce artifacts, and enhance tissue characterization makes it a powerful tool in the diagnosis and treatment of a wide range of conditions. The concept of “spectral CT” was first introduced by Godfrey N. Hounsfield, one of the two inventors of CT, in 1972 [33]. Now, more than 50 years later, we have only just begun to realize the full potential of this remarkable technology.
With this Editorial, we wish to deliver the invitation to the radiology community to contribute to this thematic series and expand our knowledge about PCD-CT.
Abbreviations
- AI
Artificial intelligence
- CT
Computed tomography
- EID
Energy-integrating detector
- EID-CT
Energy-integrating detector computed tomography
- PCD
Photon-counting detector computed tomography
Author contributions
Both authors drafted and reviewed the manuscript.
Funding
The authors state that this work has not received any funding.
Declarations
Ethics approval and consent to participate
Not applicable.
Consent for publication
Not applicable.
Competing interests
AVS receives institutional research and travel support from Siemens Healthineers and is a consultant and shareholder for Elucid Bioimaging. TE receives travel support, speaker fees, and institutional research support from Siemens Healthineers and is a consultant for Circle Cardiovascular Imaging. AVS is deputy editor of the European Radiology Experimental. He has not taken part in the review or selection process of this article. TE is guest editor of European Radiology Experimental. He has not taken part in the review or selection process of this article.
Footnotes
This article belongs to the thematic series entitled “Photon-counting CT: a disrupting innovation in medical imaging” (Guest Editors: Tilman Emrich (Mainz/Germany) and Akos Varga-Szemes (Charleston/US)).
Publisher’s Note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Rajendran K, Petersilka M, Henning A et al (2022) First clinical photon-counting detector CT system: technical evaluation. Radiology 303:130–138. 10.1148/radiol.212579 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Willemink MJ, Persson M, Pourmorteza A, Pelc NJ, Fleischmann D (2018) Photon-counting CT: technical principles and clinical prospects. Radiology 289:293–312. 10.1148/radiol.2018172656 [DOI] [PubMed] [Google Scholar]
- 3.McDermott MC, Sartoretti T, Stammen L et al (2024) Countering calcium blooming with personalized contrast media injection protocols: the 1-2-3 rule for photon-counting detector CCTA. Invest Radiol. 10.1097/RLI.0000000000001078 [DOI] [PMC free article] [PubMed]
- 4.Pourmorteza A, Choux AR, Holmes TW et al (2024) Beam hardening of K-edge contrast agents: a phantom study comparing clinical energy-integrating detector and photon-counting detector CT systems. Eur Radiol Exp. 10.1186/s41747-024-00530-5 [DOI] [PMC free article] [PubMed]
- 5.Emrich T, O’Doherty J, Schoepf UJ et al (2023) Reduced iodinated contrast media administration in coronary CT angiography on a clinical photon-counting detector CT system: a phantom study using a dynamic circulation model. Invest Radiol 58:148–155. 10.1097/RLI.0000000000000911 [DOI] [PubMed] [Google Scholar]
- 6.Zsarnoczay E, Fink N, Schoepf UJ et al (2024) Accuracy of ultra-high resolution and virtual non-calcium reconstruction algorithm for stenosis evaluation with photon-counting CT: results from a dynamic phantom study. Eur Radiol Exp 8:102. 10.1186/s41747-024-00482-w [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Verelst E, Buls N, De Mey J et al (2023) Stent appearance in a novel silicon-based photon-counting CT prototype: ex vivo phantom study in head-to-head comparison with conventional energy-integrating CT. Eur Radiol Exp 7:23. 10.1186/s41747-023-00333-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Halfmann MC, Bockius S, Emrich T et al (2024) Ultrahigh-spatial-resolution photon-counting detector CT angiography of coronary artery disease for stenosis assessment. Radiology 310:e231956. 10.1148/radiol.231956 [DOI] [PubMed] [Google Scholar]
- 9.Gruschwitz P, Hartung V, Ergun S et al (2023) Comparison of ultrahigh and standard resolution photon-counting CT angiography of the femoral arteries in a continuously perfused in vitro model. Eur Radiol Exp 7:83. 10.1186/s41747-023-00398-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Graafen D, Bart W, Halfmann MC et al (2024) In vitro and in vivo optimized reconstruction for low-kev virtual monoenergetic photon-counting detector CT angiography of lower legs. Eur Radiol Exp 8:89. 10.1186/s41747-024-00481-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Vattay B, Szilveszter B, Boussoussou M et al (2023) Impact of virtual monoenergetic levels on coronary plaque volume components using photon-counting computed tomography. Eur Radiol 33:8528–8539. 10.1007/s00330-023-09876-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Fink N, Emrich T, Schoepf UJ et al (2024) Improved detection of small and low-density plaques in virtual noncontrast imaging-based calcium scoring at photon-counting detector CT. Radiol Cardiothorac Imaging 6:e230328. 10.1148/ryct.230328 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Shami A, Sun J, Gialeli C et al (2024) Atherosclerotic plaque features relevant to rupture-risk detected by clinical photon-counting CT ex vivo: a proof-of-concept study. Eur Radiol Exp 8:14. 10.1186/s41747-023-00410-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Aquino GJ, O’Doherty J, Schoepf UJ et al (2023) Myocardial characterization with extracellular volume mapping with a first-generation photon-counting detector CT with MRI reference. Radiology 307:e222030. 10.1148/radiol.222030 [DOI] [PubMed] [Google Scholar]
- 15.Gnasso C, Pinos D, Schoepf UJ et al (2024) Impact of reconstruction parameters on the accuracy of myocardial extracellular volume quantification on a first-generation, photon-counting detector CT. Eur Radiol Exp 8:70. 10.1186/s41747-024-00469-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Baffour FI, Glazebrook KN, Ferrero A et al (2023) Photon-counting detector CT for musculoskeletal imaging: a clinical perspective. AJR Am J Roentgenol 220:551–560. 10.2214/AJR.22.28418 [DOI] [PubMed] [Google Scholar]
- 17.Woisetschlager M, Booij R, Tesselaar E, Oei EHG, Schilcher J (2023) Improved visualization of the bone-implant interface and osseointegration in ex vivo acetabular cup implants using photon-counting detector CT. Eur Radiol Exp 7:19. 10.1186/s41747-023-00335-y [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Chappard C, Abascal J, Olivier C et al (2022) Virtual monoenergetic images from photon-counting spectral computed tomography to assess knee osteoarthritis. Eur Radiol Exp 6:10. 10.1186/s41747-021-00261-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Sonnow L, Salimova N, Behrendt L et al (2023) Photon-counting CT of elbow joint fractures: image quality in a simulated post-trauma setting with off-center positioning. Eur Radiol Exp 7:15. 10.1186/s41747-023-00329-w [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Klintstrom B, Henriksson L, Moreno R et al (2022) Photon-counting detector CT and energy-integrating detector CT for trabecular bone microstructure analysis of cubic specimens from human radius. Eur Radiol Exp 6:31. 10.1186/s41747-022-00286-w [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Abel F, Schubert T, Winklhofer S (2023) Advanced neuroimaging with photon-counting detector CT. Invest Radiol 58:472–481. 10.1097/RLI.0000000000000984 [DOI] [PubMed] [Google Scholar]
- 22.Neubauer C, Gebler O, Wilpert C et al (2024) Photon-counting CT of degenerative changes and rupture of silicone breast implants: a pilot study. Eur Radiol Exp 8:36. 10.1186/s41747-024-00434-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Horst KK, Cao JY, McCollough CH et al (2024) Multi-institutional protocol guidance for pediatric photon-counting CT. Radiology 311:e231741. 10.1148/radiol.231741 [DOI] [PubMed] [Google Scholar]
- 24.Tsiflikas I, Thater G, Ayx I et al (2023) Low dose pediatric chest computed tomography on a photon counting detector system—initial clinical experience. Pediatr Radiol 53:1057–1062. 10.1007/s00247-022-05584-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Tao S, Rajendran K, McCollough CH, Leng S (2019) Feasibility of multi-contrast imaging on dual-source photon counting detector (PCD) CT: an initial phantom study. Med Phys 46:4105–4115. 10.1002/mp.13668 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Si-Mohamed S, Bar-Ness D, Sigovan M et al (2018) Multicolour imaging with spectral photon-counting CT: a phantom study. Eur Radiol Exp 2:34. 10.1186/s41747-018-0063-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Muenzel D, Daerr H, Proksa R et al (2017) Simultaneous dual-contrast multi-phase liver imaging using spectral photon-counting computed tomography: a proof-of-concept study. Eur Radiol Exp 1:25. 10.1186/s41747-017-0030-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Kravchenko D, Gnasso C, Schoepf UJ et al (2024) Gadolinium-based coronary CT angiography on a clinical photon-counting-detector system: a dynamic circulating phantom study. Eur Radiol Exp 8:118. 10.1186/s41747-024-00501-w [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Baffour FI, Huber NR, Ferrero A et al (2023) Photon-counting detector CT with deep learning noise reduction to detect multiple myeloma. Radiology 306:229–236. 10.1148/radiol.220311 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Wolf EV, Muller L, Schoepf UJ et al (2023) Photon-counting detector CT-based virtual monoenergetic reconstructions: repeatability and reproducibility of radiomics features of an organic phantom and human myocardium. Eur Radiol Exp 7:59. 10.1186/s41747-023-00371-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Landsmann A, Ruppert C, Wieler J et al (2022) Radiomics in photon-counting dedicated breast CT: potential of texture analysis for breast density classification. Eur Radiol Exp 6:30. 10.1186/s41747-022-00285-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Tremamunno G, Varga-Szemes A, Schoepf UJ et al (2024) Intraindividual reproducibility of myocardial radiomic features between energy-integrating detector and photon-counting detector CT angiography. Eur Radiol Exp 8:101. 10.1186/s41747-024-00493-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Hounsfield GN (1973) Computerized transverse axial scanning (tomography). 1. Description of system. Br J Radiol 46:1016–1022. 10.1259/0007-1285-46-552-1016 [DOI] [PubMed] [Google Scholar]