Skip to main content
Wiley Open Access Collection logoLink to Wiley Open Access Collection
. 2024 Sep 17;46(5-6):279–289. doi: 10.1002/elps.202400084

Charge variant analysis of monoclonal antibodies by CZE‐MS using a successive multiple ionic‐polymer layer coating based on diethylaminoethyl‐dextran

Alisa Höchsmann 1,2, Laura Dhellemmes 3, Laurent Leclercq 3, Hervé Cottet 3, Christian Neusüß 1,
PMCID: PMC11952282  PMID: 39287066

Abstract

The characterization of the impurities of pharmaceutical monoclonal antibodies (mAbs) is crucial for their function and safety. Capillary zone electrophoresis (CZE) is one of the most efficient tools to separate charge variants of mAbs; however, peak characterization remains difficult, since the hereby used background electrolytes (BGEs) are not compatible with electrospray ionization‐mass spectrometry (ESI‐MS). Here, a method that allows the separation of intact mAb charge variants is presented using CZE‐ESI‐MS, combining a cationic capillary coating and an acidic BGE. Therefore, a successive multiple ionic‐polymer layer coating was developed based on diethylaminoethyl‐dextran–poly(sodium styrene sulfonate). This coating leads to a relatively low reversed electroosmotic flow (EOF) with an absolute mobility slightly higher than that of antibodies, enabling the separation of variants with slightly different mobilities. The potential of the coating is demonstrated using USP mAb003, where it was possible to separate C‐terminal lysine variants from the main form, as well as several acidic variants and monoglycosylated mAb forms. The presented CZE‐MS method can be applied to separate charge variants of a range of other antibodies such as infliximab, NISTmAB (Reference Material from the National Institute of Standards and Technology), adalimumab, and trastuzumab, demonstrating the general applicability for the separation of proteoforms of mAbs.

Keywords: capillary coatings, capillary electrophoresis, cationic polyelectrolyte multilayers, intact antibody separation, mass spectrometry


Abbreviations

CEX

cation exchange chromatography

DEAED

diethylaminoethyl‐dextran

EACA

ε‐amino‐caproic acid

PDADMAC

poly(diallyldimethylammonium chloride)

PEO

polyethylene oxide

PSS

poly(sodium styrene sulfonate)

SA

sialic acid

SMIL

successive multiple ionic‐polymer layer

UPW

ultrapure water

1. INTRODUCTION

Monoclonal antibodies (mAbs) are important tools for the treatment of various diseases. To ensure their safety and optimize the production process, it is important to monitor the purity of the product and perform a detailed characterization of the variants. Commonly found variations in antibodies are differences in glycosylation along with missing lysine clipping, deamidation, and fragmentation of the antibody [1]. For analysis, the intact antibody is often enzymatically fragmented into peptides, glycans, or other subunits [2]. Although these methods allow for high sensitivity and good separation, some information about the structure and overarching composition of the intact antibody is lost. In addition, artifacts might be introduced. This makes the characterization of intact antibodies a desirable tool for obtaining the utmost information on microheterogeneity. However, MS does not allow to differentiate modifications of mAbs with small mass differences, such as deamidation (+1 Da) or disulfide bridge reduction (+2 Da), due to overlapping isotopic patterns. Therefore, separation prior to MS characterization is mandatory.

To separate charge variants of the intact antibody, three different separation methods are commonly applied: cation exchange chromatography (CEX) [3], capillary isoelectric focusing (CIEF) [4], and capillary zone electrophoresis (CZE) [5]. However, coupling to MS is often challenging due to the use of non‐volatile salts, ampholytes, or additives. To allow coupling of the mentioned separation methods with MS, it is possible to use fraction collection in CEX with subsequent LC‐MS analysis [6]. When the volumes of collected sample are low, such as in CIEF or CZE, two‐dimensional approaches are possible, where interfering substances from the charge variant separation method are removed in the second separation dimension before subsequent MS characterization [7, 8, 9]. Recent advances have allowed the direct coupling of CEX [10, 11] and CIEF [12, 13, 14] with MS. Acidic and basic charge variants like deamidation, sialic acid (SA) variants, and lysine truncations can be separated and detected subsequently by MS. However, there exists hardly any generic method and both CEX and CIEF have to be adapted to the protein since they rely on the pI of the mAbs [10]. Furthermore, in the case of CIEF, high intense ampholytes may interfere with ESI‐MS [12].

The most commonly applied CZE separation of mAbs is based on a background electrolyte (BGE) containing ε‐amino‐caproic acid (EACA), hydroxypropyl methyl cellulose, and triethylenetetramine [5]. This universally applicable method shows high separation efficiency and robustness; however, it cannot be directly coupled to MS. Therefore, numerous other approaches were tried to achieve a similar separation using alternative MS‐compatible BGEs and coatings. When performing CZE‐MS applying a neutral coating (Neutral OptiMS) and an acidic BGE, Haselberg et al. achieved partial separation of the two lysine variants from the main form of ustekinumab [15]. Polyethyleneimine‐coated capillaries combined with an acidic BGE allowed the separation of some acidic mAb variants with masses that differed in the range of 3–10 Da from the main variant, although no basic variants were separated for intact adalimumab [16].

When using microchip electrophoresis (ME)‐MS, it was possible to separate the main form and the two lysine variants from each other when using an acidic BGE [17, 18, 19]. When using a native BGE, the separation could be further increased [20], resulting in a comparable separation performance using this zone electrophoretic approach compared to CEX [11, 21] or CIEF [22]. However, all CZE‐MS or ME‐MS approaches do not reach the same resolution as shown for the EACA‐based CZE‐UV approach. Thus, there is still a strong need for improvement regarding separation selectivity, efficiency, and simplicity of application.

The aim of this work is to develop a generic CZE‐MS method for high‐resolution charge variant analysis. As a key point to increase separation under acidic conditions, a new cationic capillary coating is introduced with an EOF strength similar to the mAb analyte mobility. For the diethylaminoethyl‐dextran (DEAED)–poly(sodium styrene sulfonate) (PSS) coating, the best separation was achieved and optimized regarding resolution, evaluating major parameters, such as BGE and separation voltage. The optimized CZE‐ESI‐MS method is evaluated regarding the characterization of various proteoforms of USP mAb003. Several other mAbs are analyzed to demonstrate the general applicability of the CZE‐MS method with the new coating.

2. MATERIALS AND METHODS

2.1. Materials

The USP mAb003 (lot: F12980) was purchased from Sigma‐Aldrich. Infliximab (charge: 8KMKA90603), adalimumab (charge: 1825216), and trastuzumab (charge: N1050H05) were purchased from Evidentic. NISTmAb (Reference Material 8671, lot: 14HB‐D‐002) was purchased from the National Institute of Standard and Technology (NIST).

For capillary coating, DEAE‐dextran 500 (DEAED) (Mw = 500,000 g/mol, Batch 21046) was purchased from TdBLabs. Polyethylene oxide (PEO) (Mw 1,000,000 g/mol) was purchased from Alfa Aesar. Poly(diallyldimethylammonium chloride) (PDADMAC) (Mw 400,000—500,000 g/mol), PSS (Mw 70,000 g/mol), 4‐(2‐hydroxyethyl)‐1‐piperazineethanesulfonic acid (HEPES), hydrofluoric acid (40% (v/v)), sodium hydroxide, and hydrochloric acid (37%) were purchased from Sigma‐Aldrich. Isopropyl alcohol (IPA), ACN (all LC‐MS grade), formic acid (≥98%, FA), and acetic acid (HAc) were obtained from Carl Roth GmbH & Co. KG. All solutions were prepared with ultrapure water (UPW, 18 MΩ·cm at 25°C, SG Ultra Clear UV from Siemens Water Technologies).

2.2. Capillary electrophoresis

CE‐UV was performed with a CE 7100A system from Agilent Technologies. Bare fused silica capillaries with 50/100 µm inner diameter and 360/240 µm outer diameter (separation capillary/sheath liquid capillary) were obtained from Polymicro Technologies. Separation capillaries had a length of 40 cm (UV detection) or 60 cm (UV and MS detection).

Successive multiple ionic‐polymer layer (SMIL) capillaries were coated according to the protocol of Dhellemmes et al. [23]. Briefly, the capillaries were flushed with 1 M NaOH for 10 min, followed by flushing the capillary with UPW for 5 min and 20 mM aqueous HEPES solution (pH 7.4) for 10 min. To create the SMIL coating, the capillary was alternately flushed with polycation and polyanion solution (3 g/L in 20 mM HEPES solution at pH 7.4) for 7 min with intermediate rinsing with HEPES solution for 3 min until a total number of five layers was reached. After a waiting step of 5 min, the capillary is flushed with UPW for 3 min followed by a flushing step with BGE for 10 min.

PEO capillaries were coated based on a published protocol [24]. Briefly, a PEO stock solution was prepared by dissolving 50 mg PEO in 22.5 mL of UPW at 95°C. The PEO coating solution was prepared daily by mixing 10% 0.1 M HCl and 90% PEO stock solution. Before the first daily measurement, the capillaries were flushed with 1 M NaOH, UPW, and 1 M HCl each for 5 min, PEO coating solution (2 mg/mL in 0.01 M HCl) for 10 min, followed by flushing with UPW and BGE for 5 min. During CE‐MS measurements, when applying the PEO coating, it was necessary to apply 2 mbar of pressure to stabilize the current.

All coating steps were performed at 2 bar when using 60 cm capillaries and 950 mbar when using 40‐cm capillaries. During the measurement sequences, the BGE was exchanged at least every two measurements. Before each measurement, the capillary was flushed for 3 min with BGE. For comparison of the coatings, a voltage of +10 kV was applied for the PEO, and −10 kV was applied for the SMIL coatings. All antibodies were injected with 40 mbar, 5 s using a 1 mg/mL antibody solution in BGE. The SMIL capillaries were stored overnight in BGE. Data processing for CE‐UV data was performed using CEval 0.6h9 [25] (available at: https://echmet.natur.cuni.cz/).

2.3. Mass spectrometry

The CE was coupled to an Orbitrap Fusion Lumos mass spectrometer (Thermo Fisher Scientific) using the nanoCEasy interface [26]. For MS detection, the tip of the CZE capillary was etched with hydrofluoric acid to an outer diameter of roughly 80–100 µm. Emitters with a 30 µm tip were obtained from BioMedical Instruments. IPA:UPW (50:50) + 0.5% FA was used as the sheath liquid. The spray voltage for the nanoCEasy interface was set to 2000 V for all MS measurements. All data were acquired in positive mode using a mass range of m/z 700–4000 and a resolution of 30,000. Normalized Automatic Gain control Target was set to 100% and maximum injection time to 75 ms with five micro scans. The radio frequency was set to 100% and the fragmentation voltage to 80 V. MS data were analyzed using Freestyle 1.8 (Thermo Fisher Scientific) and Intact Mass 3.4 (Protein Metrics). If not otherwise detailed, electropherograms obtained from CE‐MS measurements are smoothed by applying Gaussian 5.

3. RESULTS AND DISCUSSION

3.1. Capillary coatings

Capillary coatings are a crucial factor in the CZE separation of proteins by (i) reducing protein–wall interactions and (ii) determining the EOF, thereby influencing the resolution and speed of the separation. Resolution in CZE can be expressed as [27]:

R=0.25ΔμeμEOF+μ¯eN (1)

where Δμe is the effective mobility difference between two analytes, μEOF is the mobility of the EOF, μ¯e is the mean effective mobility of the two analytes, and N is the number of theoretical plates. The resolution increases when μEOF and μ¯e are counter‐directed and their sum is minimal. In theory, neutral coatings should be preferred over a cationic coating for low mobility analytes, that is, as soon as the absolute mobility of the reversed EOF provided by the cationic coating is more than a factor of two larger than the absolute mobility of the proteins. It has been demonstrated that proteins with small mobilities can be better separated by neutral coatings, whereas proteins with high mobilities can be better separated by cationic coatings [28]. However, a cationic coating with reduced EOF should lead to better resolution even for low‐mobility analytes.

For the antibodies described here, we measured mobilities of about 17–18 TU (Tiselius unit, 1 TU = 10−9 m2Vs) for a 4 M acetic acid BGE. This means that an ideal cationic coating is supposed to be in the range of 20–30 TU for it to perform significantly better than a neutral coating. Here, we introduce a cationic SMIL coating based on DEAED that leads to EOF in this range. This coating is compared to a neutral coating, as well as a previously used SMIL coating with a higher EOF. Because the dextran groups of DEAED are only partially modified with diethylaminoethyl groups, the DEAED coating is expected to have a lower charge density compared to the PDADMAC‐PSS coating and, thus, a lower EOF. Indeed, the absolute value of EOF mobility for the five‐layer DEAED‐PSS coating was measured to be 24.0 TU and thus in the desired range for mAb separation. The reference five‐layer PDADMAC‐PSS coating resulted in a higher absolute value of EOF mobility of 35.2 TU. The EOF is highly dependent on the BGE composition and the here given values were calculated for a 4 M HAc (voltage of −10 kV [= 3], which turned out to be the best BGE [see Section 3.2]). The different coatings were tested using USP mAb003 as the model mAb for charge variant separation. This model mAb contains two basic variants with one (1 K) and two (2 K) additional lysine, respectively. Incomplete removal of C‐terminal lysine is a common basic variant observed in mAbs [29]. Due to their characteristic mass shift of +128 Da, these variants are easily detectable with MS even when no prior separation is possible, making these variants an ideal tool for studying charge variant separation with subsequent MS analysis. The mobilities of these three USP mAb003 variants are determined to be 16.99 TU (Main), 17.12 TU (1 K), and 17.22 TU (2 K) for the 4 M HAc and the DEAED‐PSS system. Applying Equation (1), the best resolution should be achieved for the DEAED‐PSS coating, while the PEO coating and PDADMAC‐PSS coatings should have a similar but lower resolution. This is demonstrated by measurements as illustrated in Figure 1, showing the separation of charge variants of USP mAb003 for the three different coatings.

FIGURE 1.

FIGURE 1

Separation of the USP mAb003 at (A) +10 kV for the neutral polyethylene oxide (PEO) coating, (B) −10 kV for the cationic poly(diallyldimethylammonium chloride)‐poly(sodium styrene sulfonate) (PDADMAC‐PSS) coating, and (C) −10 kV for the cationic diethylaminoethyl‐dextran‐PSS (DEAED‐PSS) coating with the extracted ion electropherogram (EIE) of the five most intense charge states (± 50 ppm, Gaussian smoothing 5) of G0F/G0F + 0 K (black), G0F/G0F + 1 K (yellow), and G0F/G0F + 2 K (blue) applying a background electrolyte (BGE) of 4 M HAc, a 60 cm capillary, and hydrodynamic injection of 1 mg/mL mAb in BGE (other parameters as described in Section 2). Marked time slices are integrated to gain the mean mass spectra shown in (D)–(F) for the respective separation figure above.

For the PEO coating (Figure 1A), analytes with high ion mobility arrive first, while for the SMIL coatings (Figure 1B,C), analytes with high ion mobilities arrive last. For both PDADMAC‐PSS and PEO, it is possible to see a partial separation of the lysine forms, but the separation is not sufficient to integrate “clean” mass spectra (Figure 1D,E). In contrast, the DEAED‐PSS coating (Figure 1C) with the lower EOF enables a baseline separation of the lysine variants enabling “clean” mass spectra, that is, that do not include the main variant (Figure 1F). Especially in Figure 1C for the mAb variant with two lysines, the importance of separation becomes apparent as a variant with a similar mass to the two lysine variant was detected below the main peak. This will be discussed in more details later. The measured correlations match the expected theoretical results. However, the high resolution of DEAED‐PSS coating goes along with a longer separation time (75 min) compared to both PDADMAC‐PSS coating (30 min) and PEO (45 min—numbers for 4 M HAc as BGE, 10 kV, and a 60 cm long capillary).

3.2. Background electrolyte

In order to find the most suitable acidic BGE for the separation of antibody charge variants on the DEAED‐PSS coating, different concentrations of formic acid and acetic acid were used (see Supporting Information S1). In general, a better separation could be achieved when using HAc instead of FA, which is in agreement with previous reports [28, 30]. Increasing the acetic acid concentration from 0.5 up to 4 M HAc leads to increasing resolution. When further increasing the acetic acid concentration to 8 M HAc, the resolution decreases again, which might not be expected based on the decreasing EOF and antibody mobility (see Supporting Information S2). One explanation could be solvation effects, since 4 and 8 M HAc correspond to 23% and 46% HAc solutions, respectively, with lower auto dissociation (lower current) for the 8 M HAc.

The addition of small amounts (10%) of IPA or MeOH to the BGE is possible while only marginally decreasing the resolution. If the same amount of ACN is added to the BGE, the coating becomes irreversibly damaged with increasing migration times for each subsequent run.

Due to the EOF, the separation in cationic (SMIL) systems is less influenced by the SL in the outlet vial compared to neutral systems. Only cations with a high mobility can migrate into the capillary from the interface side, therefore the nature of the anion is not critical. Indeed, when exchanging the outlet vial for SL during CE‐UV measurements, no effect on the separation efficiency could be observed (data not shown). Formic acid was chosen as SL due to its high ionization efficiency.

3.3. Repeatability and separation voltage

To determine the repeatability of the DEAED‐PSS coating, the same mAb sample was measured five times at five different voltages using CE‐UV. In Figure 2A–C, the overlays for −10 kV, −20 kV, and −30 kV are shown.

FIGURE 2.

FIGURE 2

Influence of the applied separation voltage on CE‐UV separation. Electrophoretic separation of the USP mAb003 in 4 M HAc at different voltages (A–C) (overlaid, = 5) with a hydrodynamic injection of 1 mg/mL mAb in background electrolyte (BGE) (other parameters as in Figure 1): (A) −10 kV, (B) −20 kV, (C) −30 kV, (D) H versus u curve for UV‐data (experimental points at voltage −10 kV, −15 kV, −20 kV, −25 kV, and −30 kV, = 5 with a model curve based on Equation (2) fitted through the five points) and MS data (experimental points at voltage −10 kV, −20 kV, and −30 kV, = 2 with a linear trendline fitted through the three points) (base peak electopherogram (BPE) and three EIEs), and (E) EIEs of the different glycoforms constructed using the signal for their most intense charge state (±0.3 Da, Gaussian 7).

In CE‐UV, relative standard deviations for the migration times between 0.1% and 0.3% were achieved when the BGE was exchanged after every second measurement. For the different voltages, migration times ranging between 25 min for −30 kV and 80 min for −10 kV were measured. When comparing the separation at −10 kV and −30 kV, a slightly better separation can be achieved for the lower voltages with a small shoulder being resolved in front of the main mAb for −10 kV. This is in agreement with recent studies where lower voltages resulted in higher separation efficiencies for small proteins [31]. Similarly, lower H values are reached for lower voltages (−10 kV at 12.2 µm, −30 kV at 14.3 µm) as seen in Figure 2D. Plate height H was plotted against solute velocity u using previously described protocols [23, 32, 33]. To model the experimental points, the following equation was used [33]:

H=A+Bu+pu (2)

where A is the constant term (here: A = 10.97 µm ± 0.14 µm), Bu is the term for the axial diffusion (here: B = 2D = 2 × 1.81 × 10−11 m2/s, D is determined by Taylor dispersion analysis), and pu is the mass transfer term (here: p (10−3 s) = 8.25 ± 0.37). When comparing the H values for −10 kV for the separation of the USP mAb003 with a 4 M HAc BGE of the DEAED‐PSS coating (H = 12.2 µm) to the PDADMAC‐PSS coating (= 12.3 µm, data not shown), similar values can be achieved meaning that both coatings have separation efficiencies in a similar range. In Figure 2D, in addition to the CE‐UV data, the corresponding CE‐MS values are shown for the same capillary length, coating, and BGE. For the base peak electropherogram (BPE), the achieved H values are similar to the UV data, while when the H values for the extracted ion electropherograms (EIEs) of the three most abundant glycoforms are plotted, the H value decreases from 13.4 µm (BPE, −10 kV) to 4–7 µm (different glycoforms, −10 kV). In Figure 2E, the EIEs of the glycoforms of the main mAb form are shown, clearly demonstrating narrow and partially separated peaks. This partial separation of heterogeneous samples is not detectable in CE‐UV. The value of the slope p is controlled by analyte adsorption and charge inhomogeneity induced electroosmotic fluctuations [33]. As seen in Figure 2D, the slope increases when using MS detection compared to UV detection, presumably influenced by additional peak broadening induced by the CE‐MS interface. However, even the CE‐UV slope for the mAb separation is higher (8.25 × 10−3 s) than what can be typically observed for smaller proteins on a PDADMAC coating (1.92 × 10−3 s) [33]. The discrepancy between the slopes can be, at least partially, explained by differences in protein diffusion coefficient and could be also explained by an increase of protein adsorption for larger molecules like the mAb or caused by difference in coating qualities between the PDADMAC‐PSS and DEAED‐PSS coating.

3.4. Characterization of charge variants for USP mAb

To show the separation potential of the developed method, an in‐depth characterization of USP mAb003 was performed. In Figure 3, an overview of the separation and the corresponding mass spectra are shown for one data set. Furthermore, all detected species as the mean of five measurements are summarized in Supporting Information S3.

FIGURE 3.

FIGURE 3

Overview of the separation and mass spectra for the USP mAb003 at −10 kV separation voltage, 4 M HAc background electrolyte (BGE) for the diethylaminoethyl‐dextran‐poly(sodium styrene sulfonate) (DEAED‐PSS) coating (other parameters as in Figure 1), including (A) the BPE, (B) deconvoluted mass spectra of the peaks annotated in the BPE (intensity (counts) vs. mass (Da)), and (C) charge envelope of selected species (relative intensity (%) vs. m/z).

Most analytes can be found between 65 and 72 min, with the main peak between 67.8 min and 68.8 min (#9). The four most abundant glycoforms of the main form of USP mAb003 are 145,737.8 Da ± 1.1 Da (G0F/G0F), 145,899.2 Da ± 0.4 Da (G0F/G1F), 146,062.1 Da ± 1.0 Da (G1F/G1F or G0F/G2F), and 146,225.1 Da ± 0.5 Da (G1F/G2F) (numbers given as mean, see Supporting Information S3). A slight separation of these non‐charged glycoforms is observed as shown in Figure 2E. Similarly, the three monoglycosylated mAbs arrive later (#11) and are baseline separated from the main form. The ability of CZE‐MS to separate monoglycosylated mAbs has been shown previously [34].

The lysine variants described above are the main components of peak 10 (+1 K, +128 Da) and peak 12 (+2 K, +256 Da) (Figure 3B). Here, the glycosylation pattern of the lysine variants is different from that of the main form. For the main variant, G0F/G1F is the most abundant variant (#9), while for both lysine variants G0F/G0F is more abundant (#10 and #12). While the reason for this difference remains unknown, this observation underlines the usefulness of analyzing the mAb at the intact level, since such an effect cannot be detected at the peptide level.

As part of the main peak (#9), an unknown variant with a mass difference of about +99 Da can be observed. Adding 162 Da (+1 Gal) to +99 Da results in 261 Da, which differs only 5 Da from the mass of 256 Da (+2 K), a mass difference that cannot be distinguished by the MS at intact mAb level. Thus, the separation of lysine variants is necessary to detect these proteoforms.

Variants before the main peak should have fewer positive charges and are generally considered “acidic variants”. However, since CZE separates according to differences in charge‐to‐size, size effects might occur. In the mass spectra of #4‐6, a shift of the charge envelope to smaller m/z ratios is observed. While the main peak (#9) generally has the most intense signal at = 48, the “acidic variant” (#5) has the most intense signal at = 52. Using the three most abundant glycoforms, a mass difference of +2.1 ± 0.9 Da (= 5) can be calculated between the main form and the most intense “acidic form” (#5). This shift in mass, the higher charge density, and the shift in the direction of larger analytes may be explained by one open disulfide bridge. Unpaired cysteines in IgG1 mAbs have been identified by RPLC‐MS before, where a similar shift to higher charges in MS was observed for the variant with unpaired cysteines compared to the main variant [35]. Similar observations, that is, CE‐MS characterization of variants that showed both a shift to lower mobilities and higher charge state in the ESI‐MS, have been made previously, however, no reliable deconvolution could be obtained in this case [34]. Deamidation also causes a slight mass shift (+1 Da) and would appear as an acidic variant; however, it is not expected to cause a shift of the charge envelope toward six additional charges. The three acidic variants #4, #5, and #6 cannot be further distinguished by their mass as all three show a similar shift around +2 Da. All charge envelopes are shifted to a higher number of charges, which is slightly more pronounced for peak #5 and #6 compared to #4 (Figure 3C). Therefore, it is likely that different disulfide bridges are reduced in peaks #4, #5, and #6. In mass spectrum #7, even a variant containing both +1 K and one reduced disulfide bridge can be found, migrating close to the SA variant (#8). Again, it would not be possible to distinguish between these two variants by mass alone, since the mass of 2 Gal (+324 Da) +1 SA (+291 Da) is similar to the mass of 3 Gal (+486) +1 Lys (+128 Da). Again, a possible annotation was done based on differences in their charge envelope. In addition to intact variants, some fragments were found. In peaks 2 and 3, small fragments with masses between 10 and 50 kDa are present. Some low‐abundant species in the 123 kDa range (#12b and #13) and the 101 kDa range (#14) can be found that migrate slower than the main peak. Here, masses were found that follow the typical difference in glycosylation (+162 Da). Furthermore, the respective lysine variant can be detected (#13; 123,590.6 Da with a mass difference of about +128 Da compared to #12b; 123461.9 Da). Thus, these “basic” variants can be attributed to fragments including the heavy chain.

Overall, about 52 proteoforms are detected with the CZE‐MS approach presented here, mainly due to the achieved separation of charge and size variants. The combination of accurate mass, charge distribution, and separation allows a tentative attribution to certain charge and size variants. For further confirmation, either a 2D approach using peptide analysis of previously separated intact charge variants [36] or a top‐down approach of the intact mAb could be performed [37].

3.5. Application of the method to other mAbs

To show the applicability of the method, further mAbs including infliximab, NISTmAb, adalimumab, and trastuzumab were measured. Here, the separation voltage of −20 kV as a compromise between separation efficiency and separation time was chosen. The achieved separation can be found in Figure 4, while a table including the detected variants is presented in Supporting Information S4–S7.

FIGURE 4.

FIGURE 4

Separation of variants of (A) infliximab, (B) NISTmAb, (C) adalimumab, and (D) trastuzumab at −20 kV using the diethylaminoethyl‐dextran‐poly(sodium styrene sulfonate) (DEAED‐PSS) coating with the 4 M HAc background electrolyte (BGE) (other parameters as in Figure 1), showing the BPE (black) and the EIE (± 0.3 Da) of the partial reduced main variant (orange), main variant (red), 1 K variant (blue), and 2 K variant (yellow). Glycoforms G0F/G0F (NIST G0F/G1F) were used for the given charge state along with the respect deconvoluted mass.

Lysine variants are detected for three mAbs, agreeing with the literature, where lysine variants have been described for infliximab [13, 17], adalimumab [38], and the NISTmAb [19, 39]. Using the DEAED‐based coating, the lysine variants could be separated from the main variant in all cases. For trastuzumab (Figure 4D), another (low‐abundance) basic variant is detected with a mass difference of 13.6 Da from the main variant (using the three most intense glycoforms, = 3). Again, this mass difference is too small to be resolved by MS, therefore, separation is essential.

Similar to the USP mAb003, for all four mAbs, some partial separation for the glycovariants could be observed, with G0F/G0F being the most abundant form for adalimumab, trastuzumab, and infliximab and G0F/G1F being the most abundant form for the NISTmAb. In the case of the NISTmAb, the glycosylation is especially complex, and after G2F/G2F another form with a mass difference of +162 Da is observed. This could be an additional glycation agreeing with the observation of Chen et al. who observed a mass of mAb+Hex after the removal of the N‐glycans by PNGaseF for the NIST mAb [19], although separation might be expected in the case of some glycation due to the removal of a basic amino acid function. Monoglycosylated variants can be annotated for three of the four mAbs (not for infliximab), all of which have higher migration times. Furthermore, for all four mAbs a separation of at least one “acidic” variant can be achieved. Similarly, to the “acidic” variants found for the USP mAb003, those forms have a mass of approximately 2 Da higher and are shifted to higher charge states in the ESI, again, most likely attributed to partly reduced mAb variants. For trastuzumab, two separated “acidic” peaks were detected that both have the characteristics of partially reduced mAbs, indicating positional isomers. Due to the high abundance of the two lysine variants, it is possible to identify partially reduced forms of the lysine variants for infliximab.

3.6. Comparison with other CZE‐MS approaches

A clear improvement could be reached when applying the DEAED‐PSS coating for CZE‐MS in comparison with other coatings such as the two discussed in this paper (PEO and PDADMAC‐PSS) or for the commercially available neutral coating (Neutral OptiMS) where the lysine variants could be only separated partially from each other and the main variant [15]. The separation of lysine variants achieved here for the DEAED‐PSS coating is similar in quality to the one published for ME‐MS when using an acidic BGE for infliximab [17], however slightly less efficient than the one published for NIST mAb [19]. The CZE‐MS approach demonstrated here, however, can be performed with a standard CZE system by using a simple coating protocol, which is flexible with regard to the length and inner diameter of the capillary.

Further improvement of charge variant separation should be expected when shifting the separation from an acidic system to a higher pH value. As previously investigated for the EACA system, a pH close to the pI value of the mAbs allows for a better separation due to a higher difference in the mobility of the charge variants [5]. Similarly, native separation enables improved charge variant separation, as shown for ME‐MS [20]. However, applying an acidic BGE, the detected m/z range is typically between m/z 2000 and 4000, facilitating mass spectrometry on various instruments. Furthermore, an acidic BGE does not add cations like ammonium, which can lead to undesired adduct formations. Additionally, no separation of size variants such as monoglycosylated antibodies is reported for native BGEs to our best knowledge; indeed, it was reported that monoglycosylated antibodies comigrate with the main forms of the mAb for a native separation applying a neutral coating [40].

4. CONCLUSIONS

In this work, a new SMIL coating is introduced for the separation of intact mAb variants by CZE‐MS under acidic conditions. Due to the denaturing conditions, it is more difficult to separate charge variants; however, it allows the application of standard MS. The cationic properties of the DEAED‐PSS coating allow for an efficient reduction of protein adsorption leading to high separation efficiencies similar to other coatings like the PDADMAC‐PSS coating. Importantly, the low EOF, being only slightly higher than the mobility of the mAb, allows for the separation of closely related mAb variants. Charge variants like lysine and SA variants are separated. Moreover, the approach enables the separation of size variants like the partial separation of the different glycoforms, the complete separation of the monoglycosylated mAbs as well as the separation of variants differing in the reduction state of one disulfide bridge. The method shows similar performance for various mAbs including the two model mAbs USP mAb003 and NISTmAb as well as infliximab, trastuzumab, and adalimumab. Overall, the here presented CZE‐MS method using the new five‐layer DEAED‐PSS SMIL coating is an excellent tool for the characterization of mAb variants on the intact level, supporting the development of new protein‐based pharmaceuticals.

CONFLICT OF INTEREST STATEMENT

The authors declare no conflicts of interest.

Supporting information

Supporting Information

ELPS-46--s001.docx (307.1KB, docx)

ACKNOWLEDGMENTS

This project was funded by the Deutsche Forschungsgemeinschaft (DFG, German Research Foundation; 446330632) as part of an ANR‐DFG program (SMIL E, ANR‐20‐C E 92‐0021‐01) project.

Open access funding enabled and organized by Projekt DEAL.

Höchsmann A, Dhellemmes L, Leclercq L, Cottet H, Neusüß C. Charge variant analysis of monoclonal antibodies by CZE‐MS using a successive multiple ionic‐polymer layer coating based on diethylaminoethyl‐dextran. Electrophoresis. 2025;46:279–289. 10.1002/elps.202400084

DATA AVAILABILITY STATEMENT

The data that support the findings of this study are available from the corresponding author upon reasonable request.

REFERENCES

  • 1. Harris RJ, Kabakoff B, Macchi FD, Shen FJ, Kwong M, Andya JD, et al. Identification of multiple sources of charge heterogeneity in a recombinant antibody. J Chromatogr B. 2001;752:233–245. [DOI] [PubMed] [Google Scholar]
  • 2. Dadouch M, Ladner Y, Perrin C. Analysis of monoclonal antibodies by capillary electrophoresis: sample preparation, separation, and detection. Separations. 2021;8:4. [Google Scholar]
  • 3. Fekete S, Beck A, Veuthey J‐L, Guillarme D. Ion‐exchange chromatography for the characterization of biopharmaceuticals. J Pharm Biomed Anal. 2015;113:43–55. [DOI] [PubMed] [Google Scholar]
  • 4. Mack S, Cruzado‐Park I, Chapman J, Ratnayake C, Vigh G. A systematic study in CIEF: defining and optimizing experimental parameters critical to method reproducibility and robustness. Electrophoresis. 2009;30:4049–4058. [DOI] [PubMed] [Google Scholar]
  • 5. He Y, Isele C, Hou W, Ruesch M. Rapid analysis of charge variants of monoclonal antibodies with capillary zone electrophoresis in dynamically coated fused‐silica capillary. J Sep Sci. 2011;34:548–555. [DOI] [PubMed] [Google Scholar]
  • 6. Griaud F, Denefeld B, Lang M, Hensinger H, Haberl P, Berg M. Unbiased in‐depth characterization of CEX fractions from a stressed monoclonal antibody by mass spectrometry. MAbs. 2017;9:820–830. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Hühner J, Neusüß C. CIEF‐CZE‐MS applying a mechanical valve. Anal Bioanal Chem. 2016;408:4055–4061. [DOI] [PubMed] [Google Scholar]
  • 8. Jooß K, Hühner J, Kiessig S, Moritz B, Neusüß C. Two‐dimensional capillary zone electrophoresis‐mass spectrometry for the characterization of intact monoclonal antibody charge variants, including deamidation products. Anal Bioanal Chem. 2017;409:6057–6067. [DOI] [PubMed] [Google Scholar]
  • 9. Schlecht J, Jooß K, Neusüß C. Two‐dimensional capillary electrophoresis‐mass spectrometry (CE‐CE‐MS): Coupling MS‐interfering capillary electromigration methods with mass spectrometry. Anal Bioanal Chem. 2018;410:6353–6359. [DOI] [PubMed] [Google Scholar]
  • 10. Füssl F, Cook K, Scheffler K, Farrell A, Mittermayr S, Bones J. Charge variant analysis of monoclonal antibodies using direct coupled pH gradient cation exchange chromatography to high‐resolution native mass spectrometry. Anal Chem. 2018;90:4669–4676. [DOI] [PubMed] [Google Scholar]
  • 11. Füssl F, Trappe A, Carillo S, Jakes C, Bones J. Comparative elucidation of cetuximab heterogeneity on the intact protein level by cation exchange chromatography and capillary electrophoresis coupled to mass spectrometry. Anal Chem. 2020;92:5431–5438. [DOI] [PubMed] [Google Scholar]
  • 12. Wang L, Bo T, Zhang Z, Wang G, Tong W, Da Yong Chen D. High resolution capillary isoelectric focusing mass spectrometry analysis of peptides, proteins, and monoclonal antibodies with a flow‐through microvial interface. Anal Chem. 2018;90:9495–9503. [DOI] [PubMed] [Google Scholar]
  • 13. Wang L, Chen DDY. Analysis of four therapeutic monoclonal antibodies by online capillary isoelectric focusing directly coupled to quadrupole time‐of‐flight mass spectrometry. Electrophoresis. 2019;40:2899–2907. [DOI] [PubMed] [Google Scholar]
  • 14. Schlecht J, Moritz B, Kiessig S, Neusüß C. Characterization of therapeutic mAb charge heterogeneity by iCIEF coupled to mass spectrometry (iCIEF‐MS). Electrophoresis. 2023;44:540–548. [DOI] [PubMed] [Google Scholar]
  • 15. Haselberg R, de Vijlder T, Heukers R, Smit MJ, Romijn EP, Somsen GW, et al. Heterogeneity assessment of antibody‐derived therapeutics at the intact and middle‐up level by low‐flow sheathless capillary electrophoresis‐mass spectrometry. Anal Chim Acta. 2018;1044:181–190. [DOI] [PubMed] [Google Scholar]
  • 16. Giorgetti J, Beck A, Leize‐Wagner E, François Y‐N. Combination of intact, middle‐up and bottom‐up levels to characterize 7 therapeutic monoclonal antibodies by capillary electrophoresis—Mass spectrometry. J Pharm Biomed Anal. 2020;182:113107. [DOI] [PubMed] [Google Scholar]
  • 17. Redman EA, Batz NG, Mellors JS, Ramsey JM. Integrated microfluidic capillary electrophoresis‐electrospray ionization devices with online MS detection for the separation and characterization of intact monoclonal antibody variants. Anal Chem. 2015;87:2264–2272. [DOI] [PubMed] [Google Scholar]
  • 18. Redman EA, Mellors JS, Starkey JA, Ramsey JM. Characterization of intact antibody drug conjugate variants using microfluidic capillary electrophoresis‐mass spectrometry. Anal Chem. 2016;88:2220–2226. [DOI] [PubMed] [Google Scholar]
  • 19. Chen C‐H, Feng H, Guo R, Li P, Laserna AKC, Ji Y, et al. Intact NIST monoclonal antibody characterization—proteoforms, glycoforms—using CE‐MS and CE‐LIF. Cogent Chem. 2018;4:1480455. [Google Scholar]
  • 20. Carillo S, Jakes C, Bones J. In‐depth analysis of monoclonal antibodies using microfluidic capillary electrophoresis and native mass spectrometry. J Pharm Biomed Anal. 2020;185:113218. [DOI] [PubMed] [Google Scholar]
  • 21. Sun Q, Wang L, Li N, Shi L. Characterization and monitoring of charge variants of a recombinant monoclonal antibody using microfluidic capillary electrophoresis‐mass spectrometry. Anal Biochem. 2021;625:114214. [DOI] [PubMed] [Google Scholar]
  • 22. Cao L, Fabry D, Lan K. Rapid and comprehensive monoclonal antibody characterization using microfluidic CE‐MS. J Pharm Biomed Anal. 2021;204:114251. [DOI] [PubMed] [Google Scholar]
  • 23. Dhellemmes L, Leclercq L, Höchsmann A, Neusüß C, Biron J‐P, Roca S, et al. Critical parameters for highly efficient and reproducible polyelectrolyte multilayer coatings for protein separation by capillary electrophoresis. J Chromatogr A. 2023;1695:463912. [DOI] [PubMed] [Google Scholar]
  • 24. Iki N, Yeung ES. Non‐bonded poly(ethylene oxide) polymer‐coated column for protein separation by capillary electrophoresis. J Chromatogr A. 1996;731:273–282. [Google Scholar]
  • 25. Dubský P, Ördögová M, Malý M, Riesová M. CEval: all‐in‐one software for data processing and statistical evaluations in affinity capillary electrophoresis. J Chromatogr A. 2016;1445:158–165. [DOI] [PubMed] [Google Scholar]
  • 26. Schlecht J, Stolz A, Hofmann A, Gerstung L, Neusüß C. nanoCEasy: an easy, flexible, and robust nanoflow sheath liquid capillary electrophoresis‐mass spectrometry interface based on 3D printed parts. Anal Chem. 2021;93:14593–14598. [DOI] [PubMed] [Google Scholar]
  • 27. Landers JP. Handbook of capillary and microchip electrophoresis and associated microtechniques. 3rd ed. Boca Raton: CRC Press, Taylor & Francis Group; 2007. [Google Scholar]
  • 28. Taichrib A, Pioch M, Neusüss C. Toward a screening method for the analysis of small intact proteins by CE‐ESI‐TOF MS. Electrophoresis. 2012;33:1356–1366. [DOI] [PubMed] [Google Scholar]
  • 29. Du Y, Walsh A, Ehrick R, Xu W, May K, Liu H. Chromatographic analysis of the acidic and basic species of recombinant monoclonal antibodies. MAbs. 2012;4:578–585. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. Giorgetti J, Lechner A, Del Nero E, Beck A, François Y‐N, Leize‐Wagner E. Intact monoclonal antibodies separation and analysis by sheathless capillary electrophoresis‐mass spectrometry. Eur J Mass Spectrom (Chichester). 2019;25:324–332. [DOI] [PubMed] [Google Scholar]
  • 31. Leclercq L, Renard C, Martin M, Cottet H. Quantification of adsorption and optimization of separation of proteins in capillary electrophoresis. Anal Chem. 2020;92:10743–10750. [DOI] [PubMed] [Google Scholar]
  • 32. Leclercq L, Morvan M, Koch J, Neusüß C, Cottet H. Modulation of the electroosmotic mobility using polyelectrolyte multilayer coatings for protein analysis by capillary electrophoresis. Anal Chim Acta. 2019;1057:152–161. [DOI] [PubMed] [Google Scholar]
  • 33. Dhellemmes L, Leclercq L, Frick H, Höchsmann A, Schaschke N, Neusüß C, et al. Investigating cationic and zwitterionic successive multiple ionic‐polymer layer coatings for protein separation by capillary electrophoresis. J Chromatogr A. 2024;1720:464802. [DOI] [PubMed] [Google Scholar]
  • 34. Belov AM, Zang L, Sebastiano R, Santos MR, Bush DR, Karger BL, et al. Complementary middle‐down and intact monoclonal antibody proteoform characterization by capillary zone electrophoresis–mass spectrometry. Electrophoresis. 2018;39:2069–2082. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35. Zhang T, Zhang J, Hewitt D, Tran B, Gao X, Qiu ZJ, et al. Identification and characterization of buried unpaired cysteines in a recombinant monoclonal IgG1 antibody. Anal Chem. 2012;84:7112–7123. [DOI] [PubMed] [Google Scholar]
  • 36. Schlecht J, Jooß K, Moritz B, Kiessig S, Neusüß C. Two‐dimensional capillary zone electrophoresis‐mass spectrometry: intact mAb charge variant separation followed by peptide level analysis using in‐capillary digestion. Anal Chem. 2023;95:4059–4066. [DOI] [PubMed] [Google Scholar]
  • 37. Fornelli L, Damoc E, Thomas PM, Kelleher NL, Aizikov K, Denisov E, et al. Analysis of intact monoclonal antibody IgG1 by electron transfer dissociation orbitrap FTMS. Mol Cell Proteomics. 2012;11:1758–1767. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38. Füssl F, Trappe A, Cook K, Scheffler K, Fitzgerald O, Bones J. Comprehensive characterisation of the heterogeneity of adalimumab via charge variant analysis hyphenated on‐line to native high resolution Orbitrap mass spectrometry. MAbs. 2019;11:116–128. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39. Leblanc Y, Faid V, Lauber MA, Wang Q, Bihoreau N, Chevreux G. A generic method for intact and subunit level characterization of mAb charge variants by native mass spectrometry. J Chromatogr B. 2019;1133:121814. [DOI] [PubMed] [Google Scholar]
  • 40. Belov AM, Viner R, Santos MR, Horn DM, Bern M, Karger BL, et al. Analysis of proteins, protein complexes, and organellar proteomes using sheathless capillary zone electrophoresis–native mass spectrometry. J Am Soc Mass Spectrom. 2017;28:2614–2634. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supporting Information

ELPS-46--s001.docx (307.1KB, docx)

Data Availability Statement

The data that support the findings of this study are available from the corresponding author upon reasonable request.


Articles from Electrophoresis are provided here courtesy of Wiley

RESOURCES