ABSTRACT
Electrospray ionization (ESI) is often the ionization method of choice, particularly for high‐mass polar molecules and complexes. However, when analyzing mixtures of analytes, charge state ambiguities and overlap in mass‐to‐charge (m/z) can arise from species with different masses and charges. While solution‐phase conditions can sometimes be optimized to produce relatively low charge states—thereby reducing charge‐state ambiguity and m/z overlap—gas‐phase methods offer greater control over charge state reduction. For complex mixtures, however, charge state reduction alone often fails to resolve individual components in the mixture. Incorporating a mass‐selection step prior to charge state manipulation can simplify the mixture and significantly improve the separation of the components. This general tandem mass spectrometry approach is referred to here as precursor resolution via ion z‐state manipulation (PRIZM). Examples of variations of PRIZM experiments date back roughly 25 years and have involved ion/molecule proton transfer reactions, ion/ion proton transfer reactions, ion/ion electron transfer reactions, electron capture reactions, and multiply‐charged ion attachment reactions. This tutorial review describes the PRIZM approach and provides illustrative examples using each of the charge state manipulation approaches mentioned above.
Keywords: charge reduction, electrospray ionization, mixture analysis, tandem mass spectrometry
1. Introduction
Electrospray ionization (ESI) [1] is among the most impactful developments in mass spectrometry by virtue of the wide range of important applications that it has enabled. The ability to generate ions directly from solution provides means for producing ions from polar nonvolatile molecules that greatly facilitates the coupling of condensed‐phase separations, such as liquid chromatography [2, 3, 4] and capillary electrophoresis [5, 6], with mass spectrometry. ESI also provides a soft ionization approach for many classes of polymers [7], including biopolymers [8, 9], and complexes derived therefrom [10]. A hallmark of ESI is its tendency to generate multiply‐charged ions from molecules with multiple polar sights. The multiple‐charging phenomenon has several profound implications for mass spectrometry and tandem mass spectrometry. For example, multiple‐charging reduces mass‐to‐charge (m/z) ratios, which decreases the upper m/z requirement for the mass analyzer. Multiple‐charging allows for higher mass resolution measurements for analyzers that provide better resolution at low m/z (e.g., Fourier transform ion cyclotron resonance [FT‐ICR] [11] and electrostatic ion traps [12], including the Orbitrap™ [ 13], mass analyzers) and improves detection efficiency both for approaches based on image current measurements, such as the FT‐ICR and Orbitrap, as well as for approaches that rely on electron multiplication [14]. The multiple‐charging phenomenon also affects the kinetic stability of the molecule‐ion and influences favored dissociation channels. As a result, tandem mass spectra (MS/MS spectra) and the structural information derived therefrom are charge‐state dependent [15]. The consequences of multiple‐charging listed above generally facilitate the determination of the masses and primary structures of analyte ions. However, the multiple‐charging phenomenon introduces ambiguity in determining mass from the measurement of m/z due to the fact that both mass and charge are, a priori, unknown. For this reason, charge state determination is an essential element in many applications of ESI mass spectrometry and tandem mass spectrometry. In this review, approaches to determining charge states in ESI‐MS and ESI‐MS/MS are briefly reviewed, and a type of MS/MS experiment, referred to herein as precursor resolution via ion Z‐state manipulation (PRIZM), is described that facilitates analyte mass determination of multiply‐charged analytes in mass spectra that otherwise are too convoluted to make mass measurements.
1.1. Charge State Determination in ESI‐MS and ESI‐MS/MS
A variety of approaches have been developed to determine charge states of ions in mass spectra generated via ESI. It was shown early in the application of ESI to proteins that the presence of two or more ions from the same molecule with known differences in mass and charge allows for the determination of ion charge via the solution of a set of simultaneous equations [9, 16]. A variety of algorithms have since been described that can be used to automatically determine the mass of an analyte from the most probable charge state distribution based on Bayesian statistics [17, 18, 19, 20]. The output of such algorithms is commonly referred to as a “zero‐charge deconvolution”. An analogous approach takes advantage of the presence of an adduct ion within a specific charge state, characterized by a known mass difference (e.g., the replacement of a proton with a sodium ion leading to a mass difference of 22 Da). The m/z spacing between the adducted and nonadducted ion reveals the charge state. Moreover, even when an adduct of known mass is not present, the m/z spacings between adjacent isotope peaks can reveal the ion's charge [21]. ESI mass spectra were first collected using low resolution mass analyzers, which generally precluded charge determination via isotope spacings for ions greater in mass than a few kiloDaltons. The coupling of ESI with high‐resolution instruments has significantly extended the range of analyte masses for which isotope spacing measurements can be used to determine charge states [22]. Figure 1 provides an example of charge state determination via (a) m/z measurement of adjacent charge states, (b) m/z spacings between isotope peaks for a given charge state, and (c) via m/z spacings of adduct ions for a given charge state.
FIGURE 1.

ESI/MS illustrating the three common means for determining charge states from peak spacings (i.e., [a] spacing between adjacent charge states, [b] spacings between isotopes, and [c] spacing between adduct peaks).
The preceding approaches rely on differences between m/z values determined from measurements of many ions (i.e., experiments involving ion ensembles) to determine charge and are most relevant to the subject of this tutorial review. An alternative approach relies on the independent measurement of ion m/z and z using image current detection. This approach is referred to as charge detection mass spectrometry (CDMS) [23, 24] and has been implemented with electrostatic linear ion traps as well as Orbitraps™ [25, 26]. This approach relies on the measurement of the charge of individual ions and is generally performed with only one [27] or a few ions [28] present in the ion trap at a time. This one‐at‐a‐time approach is particularly effective for relatively highly charged ions and has therefore been demonstrated to be especially advantageous for the mass measurement of large and heterogeneous species such as virus particles and nanoparticles [29]. Since CDMS is not an ion ensemble measurement, that is, hundreds or thousands of ions measured simultaneously, experiments involving complex heterogenous mixtures will tend to be long. Conversely, PRIZM is intended to extend the utility of classic mass spectrometry approaches to determine charge state information rather than utilizing new instrumentation platforms.
The classic tandem mass spectrometry (MS/MS) experiment, which involves the m/z selection of a precursor ion, activation, and dissociation of the precursor ion, and m/z analysis of the product ions, when applied to a multiply‐charged precursor ion, raise the challenge of determining the charges of the product ions. Provided the analyzer resolution is sufficient to resolve isotope spacings and there is minimal overlap between ions of different masses and charges (but still overlapping m/z ratios), product ion charge states can be determined from the magnitudes of the spacings between the isotope peaks. When the analyzer resolution is insufficient to resolve isotopes or the analyte mixture is so heterogeneous such that spectral overlap causes peaks to remain unresolved, product ion charge state determination becomes more challenging. While the charge state of the precursor ion places a constraint on the possible charge states of the product ions, distributions of product ion charge states with spacings of unit charge are usually absent, as are ions with adducts of known mass. Hence, two of the above approaches (Figure 1a,c) that can be used to determine charge states in mass spectra are generally unavailable in assigning charge states in product ion spectra. The first MS/MS platform coupled with ESI was the triple quadrupole, which was not capable of resolving isotope spacings for high‐mass product ions. Therefore, the initial applications of tandem mass spectrometry to multiply‐charged protein ions were restricted to generating “fingerprints” of precursor ions (i.e., no product ion mass assignment) [30]. It was not until ESI was coupled with the FT‐ICR platform that applications like top–down MS/MS of proteins could be pursued, since the higher resolution afforded the use of isotope spacings of product ions for charge state assignment [31].
1.2. Charge State Manipulation to Facilitate Charge State Determination in ESI‐MS(/MS)
As ESI‐MS(/MS) applications have evolved, a number of approaches intended to manipulate charge states have been developed to facilitate charge state assignment using one or more of the m/z spacing strategies described above. Among these approaches in ESI‐MS is the adjustment of solution conditions, often changes to pH, to lead to lower charge states, which are more readily resolved [32, 33, 34, 35, 36, 37]. The extent to which charge states can be reduced via altering ESI solution conditions alone, however, is relatively limited. Charge state distributions have been altered by exposing electrospray droplets to acidic [38], basic [39], or organic [40] vapors. In such cases, both droplet chemistry and gas‐phase ion/molecule chemistry can affect the charge states. For example, changes in droplet pH can result in changes in protein conformational states in the droplets [41], which can play a role in the mechanism by which ions emerge from the droplets [42]. The reduction of charge states generated via ESI has also been achieved using either ion/molecule proton transfer reactions [43, 44] or ion/ion proton transfer reactions [45, 46, 47, 48, 49, 50, 51] prior to admitting ions into the mass spectrometer.
Charge reduction strategies have also been implemented within mass spectrometers, which has the advantage of separating the charge reduction step from the processes associated with initial ion formation, desolvation, and transmission into the mass spectrometer. For product ion charge determination in tandem MS, the charge reduction step must take place within the mass spectrometer. The first examples of charge reduction within a mass spectrometer involved ion/molecule proton transfer reactions [52, 53, 54], including proton transfer for determining product ion charge states in tandem mass spectrometry [55]. Ion/molecule reactions within a mass spectrometer are usually implemented in instruments with the ability to trap ions, which allows for control of the reaction time along with the number density of the neutral reagent. Ion/molecule reactions, however, can be limited in the extent to which charge can be reduced and ion/molecule attachment can compete with proton transfer, which can lead to ambiguities in interpretation [56, 57]. Ion/ion proton transfer reactions, on the other hand, can reduce charge states to arbitrarily low values, although ion attachment can compete with proton transfer, depending upon the nature of the reagent ion [58].
Ion/ion proton transfer reactions in electrodynamic ion traps have been used extensively for the manipulation of charge states of multiply‐charged ions [59, 60, 61], including product ions generated in a tandem mass spectrometry experiment [62, 63, 64, 65, 66]. The reactions can take place under mutual ion polarity storage conditions, which can be established in 3‐D quadrupole ion traps and linear quadrupole ion traps with radio‐frequency voltages applied to the end‐plates [67] or in “transmission mode” where the analyte ions are stored in the ion trap while the reagent ions are transmitted through the ion trap such that there is overlap between the analyte and reagent ions [68, 69]. The electrodynamic ion trap environment also allows for the inhibition of ion/ion reactions on the basis of mass‐to‐charge thereby enabling so‐called “ion parking” experiments [70, 71, 72, 73, 74], which provides additional flexibility in multistage MS experiments [75]. Proton transfer, either via ion/molecule or ion/ion reaction, involves a change in charge (Δz) of ± 1, depending upon the analyte ion polarity, and a mass change (Δm) of ± 1 Da (i.e., the mass of a proton), depending upon whether the proton is lost or gained by the analyte ion. Electron‐based reactions, either via electron capture by a cationic analyte (Δm = +5.486x10−4 Da; Δz = −1) or ion/ion electron transfer (Δm ± 5.486x10−4 Da; Δz ± 1, depending upon the analyte charge), can also achieve charge reduction. However, electron capture and electron transfer reactions involving multiply‐charged analyte ions tend to result in analyte ion fragmentation (viz. electron capture dissociation [76, 77] [ECD] and electron transfer dissociation [78, 79] [ETD]). As the size of the analyte ion increases, however, both electron capture and electron transfer without obvious dissociation (ECnoD and ETnoD, respectively) effectively result in charge reduction. Particularly for large analyte ions of low relative charge, as are often encountered in native mass spectrometry [80, 81, 82], ECnoD [83, 84] and ETnoD tend to dominate over ECD and ETD, respectively. The attachment of oppositely charged ions [85] (Δm = mass of the reagent ion, Δz = charge of the reagent ion) can also be used to reduce precursor ion charges. Ion attachment and proton transfer, for example, can be competitive processes, depending upon the sizes and the acidities/basicities of the charge bearing sites [86]. However, particularly for large analyte ions of low relative charge, ion attachment tends to dominate [87, 88].
2. Precursor Resolution via Ion Z‐State Manipulation (PRIZM)
The PRIZM experiment described here facilitates mass measurements in scenarios in which the ESI‐MS yields a broad unresolved signal, often referred to as a “blob” that precludes mass determination of many or all components that may be present. This scenario can arise from highly heterogeneous mixtures that, at the extremes, can be comprised of (i) a single, typically high mass, component that is itself highly heterogeneous or (ii) many individual components, such as a mixture of many small to moderate size polymers. The former case is often encountered in, for example, native MS applications, whereas the latter applies to polymer or protein mixture analysis. PRIZM involves the selection of a narrow range of m/z values followed by charge reduction. The m/z selection step simplifies the mixture of ions subjected to charge reduction, which facilitates the resolution of the charge states formed via charge reduction. There is an interplay between the width of the m/z selection and the extent of charge reduction. As the charge is reduced, the m/z spacing between charge states increases along with the m/z values of the ions. Provided the upper m/z range of the analyzer can accommodate the charge‐reduced ions and the detection method is capable of responding to the ions, the “peak capacity” of the measurement increases with the extent of charge reduction [89]. Hence, there is an inverse relationship between the complexity of the mixture of ions initially selected for reaction via the m/z selection step and the extent of charge state reduction.
2.1. PRIZM and a “Single” Heterogeneous Component
The first description of a PRIZM experiment, although the term was not used to describe it, involved ion/molecule proton transfer as the means for charge state reduction [90]. Figure 2 is an adaptation of the original description of a PRIZM experiment based on a simulated ion/molecule proton transfer experiment with and without an initial mass‐selection step of a highly heterogeneous hypothetical analyte (see reference 90 for details). Figure 2a shows an initial ESI‐MS of an analyte that shows a range of Gaussian charge state peaks with unresolved isotopes that lead to significant charge state overlap. Figure 2b shows the charge state distribution after an ion/molecule proton transfer reaction period, which leads to a modest shift in the abundances of the charge states to lower charge. Extensive overlap between the charge states remains. Figure 2c shows the isolation of a narrow range of ions near the most intense region of Figure 2a. This step simplifies the mixture of ions that are subjected to charge state reduction. Figure 2d shows the spectrum resulting from the same proton transfer conditions used to generate Figure 2b. The product ions in Figure 2d show near baseline resolution, which leads to a more confident charge state assignment from the spacings of the adjacent charge states (see Figure 1).
FIGURE 2.

Illustration of a PRIZM experiment. (a) ESI‐MS of a roughly 15 kDa heterogeneous polymer with extensive charge state overlap. (b) ESI‐MS after brief ion‐molecule reaction time that shifts the abundance‐weighted average charge to a lower value. (c) Selection of a relatively narrow slide of m/z in the spectrum of (a). (d) ESI‐MS after the same ion/molecule reaction conditions of (b) for the ions in (c). Adapted with permission from Reference 90, copyright American Chemical Society.
An experimental demonstration of PRIZM using ion/molecule proton transfer as the charge reduction mechanism was provided by anions generated from negative ESI of Escherichia coli (E. coli) tRNA strain W [90]. The ESI‐MS was generated under conditions that lead to a distribution of metal ion adduction, which resulted in extensive overlap between charge states and an unresolved blob in the spectrum (see Figure 3a). This constitutes an example of a “single” (i.e., tRNA) major component with a high degree of heterogeneity. After an ion isolation step and storage in an electrodynamic ion trap in the presence of trifluoro‐acetic acid, proton transfer from the neutral acid to the tRNA anions resulted in the spectrum of Figure 3b. Based on the spacings of the partially resolved product ions, charge state assignments could be made, which enabled the determination of mass.
FIGURE 3.

(a) Electrospray mass spectrum of the anions derived from E. coli tRNA, strain W. (b) Spectrum resulting from the mass selection of a narrow range of mass‐to‐charge values from the distribution of (a) followed by proton transfer reactions with trifluoro‐acetic acid. Adapted with permission from Reference 90, copyright American Chemical Society.
Ion/ion reactions are a more robust means for reducing charge states than are ion/molecule reactions for both thermodynamic and kinetic reasons (see reference 58). Kaltashov et al. demonstrated PRIZM experiments involving glycosylated [91] and PEGylated [92] proteins using limited charge reduction via ion/ion electron transfer as the means for charge reduction. Figure 4 summarizes a series of PRIZM experiments applied to a broad signal generated via ESI of PEGylated interferon β‐1a (PEG‐IFNβ). The gray trace shows the total ESI signal as a function of m/z, the blue traces show 10 m/z unit‐wide regions that were selected for charge reduction, and the red traces show the respective products following a limited ion/ion electron transfer period. The insert shows a MALDI‐TOF spectrum of the same sample. This figure illustrates the generation of mass information from the ESI of a large heterogeneous analyte that is not available from the “blob” (gray trace) generated directly from the sample. Notably, while the data from Figure 4 were itself collected via direct sample infusion, most of the data reported in the work from which Figure 4 was reproduced were collected during an on‐line chromatographic separation, which demonstrated PRIZM on the chromatographic time‐scale.
FIGURE 4.

PEG‐IFN ESI‐MS (gray trace), ion isolations (blue traces), and respective post‐ion/ion reaction spectra (red‐traces). PEG‐IFN MALDI‐TOF‐MS (insert). Reproduced from Ref. 92 with permission from the Royal Society of Chemistry.
Depending upon the size(s) and heterogeneity of the analytes, limited charge reduction through proton transfer or electron transfer/capture may not suffice to resolve charge states. In such cases, charge reduction might only lead to a slight shift of the unresolved “blob” to higher m/z without significantly improving charge state resolution. To address this, it may be advantageous to use larger step sizes in charge reduction, creating larger spacings between adjacent charge state peaks, and to minimize the range over which the analytical signal is dispersed. The attachment of multiply‐charged reagent ions of known mass and charge to analyte ions of opposite polarity allow for the generation of well‐defined changes in m/z that are much greater than those associated with the single‐charge‐at‐a‐time approaches discussed above. A PRIZM experiment using multiply‐charged ion attachment (MIA) as the charge reduction approach is contrasted with proton transfer experiment in the simulation provided in Figure 5. A 4‐MDa analyte species with a 50‐kDa FWHM with 126–140 charges is shown at the top (a). A 1000 m/z selection at the most abundant part of the top trace is shown directly below followed by up to 90 proton transfers (c) or MIA with up to three attachments of a 30− reagent ion of roughly 67 000 Da (e.g., bovine serum albumin). The MIA approach, which provides flexibility in choosing the magnitudes of Δm and Δz, facilitates the resolution of charge states and allows for the determination of mass using the known values of reagent mass and charge [93]. This reaction is typically carried out by first generating analyte ions, isolating a narrow population either through a RF/DC mass filter or by ejecting unwanted ions through resonant excitation frequency sweeps [94] then cooling the selected population in a higher pressure (2–10 mTorr) ion trap; the oppositely charged reagent ions are then generated, isolated via the RF/DC mass filter tuned for the desired m/z of the reagent, then mutually stored in the same ion trap with the previously formed/isolated analyte ions. The application of an AC waveform, typically 100–300 kHz with 100–300 V0‐p, to orthogonal plates on either end the ion trap allow for the mutual storage of both positive and negative ions, providing trapping in the z‐direction [64, 67]. Mutual storage times can further be tuned to allow more or less reaction to occur.
FIGURE 5.

(a) Simulated ESI‐MS of a 4 MDa analyte (z = 126–140) with a FWHM of 50 kDa. (b) Selection of a m/z 1000 window from the mass spectrum. (c) Post‐ion/ion MS following up to 90 consecutive proton transfer reactions. (d) Post‐ion/ion MS following up to three attachments of a 30− reagent anion of mass = 67 kDa.
Two examples of PRIZM experiments applied to a mixture of negatively charged E. coli ribosome 50S particles using MIA are summarized in Figure 6 [95]. The insert in Figure 6a shows the negative ESI‐MS of the E. coli ribosome under conditions in which ions of the 30S and 50S particles are observed. The green box shows the selected m/z window which contains ions of 50S subjected to the MIA experiment. Figure 6a shows the post‐ion/ion reaction spectrum derived from the MIA of up to six 10+ ubiquitin cations. Figure 6b shows an expanded view of the m/z region with ions generated by the sixth attachment, revealing three charge states for each of five major components within the selected 50S particle ion population. Figure 6c shows the expanded m/z region from another MIA experiment in which two 30+ cations of carbonic anhydrase are reacted with the same selected ribosome 50S precursor ion population (green box). Both experiments provide reduction by 60 charges and facilitate mass determination for the five major components of the 50S particles. This demonstrates the flexibility associated with a MIA experiment, as multiple reagent ions can be selected to achieve a similar extent of charge reduction.
FIGURE 6.

(a) Post‐ion/ion reaction MS after the attachment of up to six 10+ ubiquitin ions to selected anions derived from ESI of E. coli ribosome 50S particles. The green box in the insert shows the isolated m/z region subjected to ion/ion reaction. (b) Expanded m/z region for the sixth attachment of ubiquitin 10+ reagent ions, which reveals the presence of five major components in the 50S population. (c) Expanded m/z region for the second attachment of carbonic anhydrase 30+ ions to the same precursor ion population associated with the green box in the insert of (a) and the experiment leading to (b). Equivalent information is present in the spectra of (b) and (c) (see reference 94). Adapted with permission from Reference 94, copyright American Chemical Society.
2.2. PRIZM and Multicomponent Mixtures
Another scenario in which a mass spectrum may become uninterpretable due to extensive peak overlap arises when there is a highly complex mixture of components. While each component may individually give readily resolved charge state distributions, their combination can result in an unresolved “blob”. Even when individual peaks can be resolved, uncertainty in which peaks are related to one another can lead to errors in assigning masses to mixture components using common approaches. A simulation of a 12‐component mixture, each producing overlapping charge state distributions, illustrates the utility of a PRIZM experiment, as shown in Figure 7. Figure 7a shows the simulated ESI‐MS of the 12‐component mixture of masses and charges shown in the insert in Figure 7b. Figure 7b shows the post‐ion/ion proton transfer reaction spectrum of the full mixture. Figure 7c shows a 25 m/z slice of the mixture in Figure 7a, and Figure 7d shows the post‐ion/ion proton transfer spectrum of the 25 m/z slice of the original mass spectrum. Note that the peaks in Figure 7d are somewhat narrower than those of Figure 7b as a result of the m/z selection and that there are fewer peaks in Figure 7d due to the fact that the slice contains signals from fewer analytes than does the full spectrum of Figure 7a.
FIGURE 7.

(a) ESI‐MS of a 12‐component mixture ranging in mass from 12.4 to 93.9 kDa. (b) Post‐ion/ion proton transfer mass spectrum of the mixture of ions in (a). (c) Ion isolation of a m/z 25 slice from (a). (d) Post‐ion/ion proton transfer mass spectrum of the mixture of ions in (c).
Results of zero‐charge deconvolutions by the UniDec [19] program using identical parameters applied to the post‐ion/ion proton transfer data taken to m/z 95 000 for the experiments reflected in Figure 7a,b as well as Figure 7c,d are shown in Figure 8a,b, respectively. The zero‐charge mass spectrum of Figure 7b (see Figure 8a) yields more than 30 peaks with relative intensities greater than 25%, most of which are false positives. These errors arise from significant peak overlap in the simulated data. In contrast, the deconvolution of Figure 7d, corresponding to the 25 m/z slice, shows five major components, all matching expected masses in the mixture. These figures also illustrate how the width of the selected m/z window relates to extent to which charge reduction must be performed in order to minimize false positives. In the absence of ion isolation, the extent of charge reduction was not sufficient to minimize spectral overlap, resulting in a high degree of ambiguity in establishing analyte ion charge states.
FIGURE 8.

(a) Zero‐charge deconvolution of the ion/ion proton transfer experiment applied to the mixture of ions in Figure 7a (see Figure 7b). (b) Zero‐charge deconvolution of the ion/ion proton transfer experiment applied to the mixture of ions in Figure 7c (see Figure 7d).
The simulations in Figures 7 and 8 demonstrate the improved confidence with which charges and masses can be assigned by performing a selection step prior to charge reduction (i.e., a PRIZM procedure), even for a relatively simple mixture. Another possible advantage of charge state reduction of mixtures is the improved ability to detect individual components in the mixture that would otherwise be obscured by extensive signal overlap in the initial ESI‐MS. A particularly noteworthy example of the benefit of the PRIZM procedure for top‐down protein mixture analysis was demonstrated in an experiment termed “targeted proton transfer charge reduction” (tPTCR) [96]. In tPTCR, a 1.5 m/z window/slice from an ESI‐MS is selected for a short proton transfer ion/ion reaction period (e.g., 8 ms) to generate mass information from the selected m/z windows. Figure 9a shows the ESI‐MS of a portion of a chromatographic separation of GELFrEE fraction (see insert), and Figures 9b–d show post‐ion/ion proton transfer spectra for selected regions around m/z 878, 879.5, and 882,5, respectively, at the same retention time. According to the report, deconvolution of the data in Figure 9a returned one proteoform mass, whereas the spectra of Figure 9b–d returned four, three, and three unique masses, respectively. The full work‐flow also included a fragmentation step in alternating fashion to enable protein identification. When applied to GELFrEE fractions of Pseudomonas aeruginosa ( P. aeruginosa) (with proteins 30–60 kDa in mass), the tPTCR approach yielded 192 identifications of UniProt entries (1% false discovery rate) compared to 85 identifications using a previously described medium/high approach for high throughput top‐down proteomics analysis by the same group [97].
FIGURE 9.

(a) ESI‐MS at 52‐min retention time of a GELFrEE fraction of P. aeruginosa . (b) 1.5 m/z selection of m/z 878 followed by proton transfer charge reduction. (c) 1.5 m/z selection of m/z 879.5 followed by proton transfer charge reduction. (b) 1.5 m/z selection of m/z 882.5 followed by proton transfer charge reduction. Adapted with permission from Reference 95, copyright American Chemical Society.
A unique case of signal concentration within narrow m/z regions can occur while analyzing homopolymers via ESI. Components of polydisperse homopolymers with the same monomer‐to‐charge (M/C) ratio exhibit similar m/z values, differing only by the combined mass of their terminal groups. A PRIZM experiment involving the selection of ions with a fixed M/C ratio can serve as an efficient method for characterizing polymer distributions. Figure 10 [98] illustrates this approach using poly (ethylenimine) (PEI), CH3(C2H5N)nOH, with a labeled molecular weight of 10 kDa. Figure 10a shows the mass spectrum from directly infusing of PEI‐10k solution to ESI‐MS in positive ion mode. The signal is predominantly concentrated at low m/z values, with regions of high relative abundance corresponding to expected M/C values. Figure 10b shows the isolation of ions with the range of m/z 258–263, which is expected to contain the ions with M/C = 6 (i.e., one charge for every six monomer units). Figure 10c shows the post‐ion/ion proton transfer spectrum over the region of singly‐charged product ions. Four distributions of ions were found to be present in the isolated m/z region, all of which showed a mass spacing of six monomer units, as expected. It was noted that the multiply‐charged PEI ions are particularly fragile, with losses of H2O and CH3OH being prominent. Also, HCl adducts (the PEI was delivered as the HCl salt) and water adducts were observed. The width of the isolation window was sufficiently wide to encompass such products. Note that CH3OH loss effectively removes the end groups so that all of the M/C = 6 ions that lose methanol fall at the same m/z value.
FIGURE 10.

(a) Positive ESI‐MS of poly (ethylenimine). (b) Ion isolation of m/z 258–263, which encompasses M/C = 6. (c) Proton transfer charge reduction to the 1+ charge state of the mixture of ions isolated in (b). Adapted with permission from Reference 97, copyright American Chemical Society.
3. Summary
PRIZM, a type of tandem mass spectrometry experiment, involves the m/z selection of one or more precursor ions, charge state reduction, and mass analysis. A PRIZM experiment facilitates charge state determination of multiply‐charged ions when conventional approaches, such as utilizing isotope spacings, spacings between adducts, and spacings between adjacent charge states, are compromised due to extensive spectral overlap. The PRIZM approach can also reveal components in a mixture that differ in mass and charge but have similar m/z ratios, as often observed in ESI‐MS. Charge state reduction approaches include ion/molecule proton transfer reactions, ion/ion proton transfer, ion attachment ion/ion reactions, electron transfer ion/ion reactions, and electron capture. PRIZM experiments can be conducted, in principle, on any tandem mass spectrometer platform, provided that a suitable charge reduction approach is available and that the mass analyzers have the upper m/z limit to accommodate the charge‐reduced product ions. The number of components that can be separated by charge reduction depends upon the extent of charge reduction as peak separation generally increases as charge states decrease. Therefore, the width of the ion isolation window, which is a determinant in the initial mixture complexity, is inversely related to the extent of charge state reduction and should be carefully considered to achieve a successful PRIZM experiment. Two general applications have emerged for experiments of this type. The first involves systems with one or a few major components that are themselves heterogeneous, as is often encountered in native mass spectrometry. The second focuses on the application of complex mixtures comprised of many components, such as the ESI‐MS of protein mixtures and homopolymers. These examples demonstrate how PRIZM can aid in both the analysis of heterogeneous systems and the deconvolution of complex mixtures. In conclusion, PRIZM is a powerful strategy to obtain accurate charge state and mass information of individual components in complex mixtures undergoing ESI.
Author Contributions
Nicolas J. Pizzala: visualization, writing – original draft preparation. Hsi‐Chun Chao: visualization, writing – original draft preparation. Boukar K.S. Faye: visualization, writing – original draft preparation Scott A. McLuckey: resources, supervision, visualization, writing – review and editing.
Conflicts of Interest
The authors declare no conflicts of interest.
Acknowledgments
This effort to generate this review was supported by the National Institute of Health (NIH) under grant GM R35‐GM153191. H.‐C. Chao would like to additionally thank the Beckman Institute for Advanced Science and Technology for their support.
Funding: This work was supported by the National Institute of General Medical Sciences ( GM R35‐GM153191).
Data Availability Statement
Data sharing is not applicable to this article as no datasets were generated or analyzed during the current study.
References
- 1. Fenn J. B., Mann M., Meng C. K., Wong S. F., and Whitehouse C. M., “Electrospray Ionization for Mass Spectrometry of Large Biomolecules,” Science 246, no. 4926 (1989): 64–71. [DOI] [PubMed] [Google Scholar]
- 2. Whitehouse C. M., Dreyer R. N., Yamashita M., and Fenn J. B., “Electrospray Interface for Liquid Chromatographs and Mass Spectrometers,” Analytical Chemistry 57 (1985): 675–679. [DOI] [PubMed] [Google Scholar]
- 3. Gelpí E., “Biomedical and Biochemical Applications of Liquid Chromatography‐Mass Spectrometry,” Journal of Chromatography A 703 (1995): 59–80. [DOI] [PubMed] [Google Scholar]
- 4. Gelpí E., “Interfaces for Coupled Liquid‐Phase Separation/Mass Spectrometry Techniques. An Update on Recent Developments,” Journal of Mass Spectrometry 37 (2002): 241–253. [DOI] [PubMed] [Google Scholar]
- 5. Smith R. D., Wahl J. H., Goodlett D. R., and Hofstadler S. A., “Capillary Electrophoresis/Mass Spectrometry,” Analytical Chemistry 65, no. 13 (1993): 574A–584A. [Google Scholar]
- 6. Krenkova J. and Foret F., “On‐Line CE/ESI/MS Interfacing: Recent Developments and Applications in Proteomics,” Proteomics 12 (2012): 2978–2990. [DOI] [PubMed] [Google Scholar]
- 7. Wesdemiotis C., Williams‐Pavlantos K. N., Keating A. R., McGee A. S., and Bochenek C., “Mass Spectrometry of Polymers: A Tutorial Review,” Mass Spectrometry Reviews 43, no. 3 (2024): 427–476. [DOI] [PubMed] [Google Scholar]
- 8. Oberacher H., Walcher W., and Huber C. G., “Effect of Instrument Tuning on the Detectability of Biopolymers in Electrospray Ionization Mass Spectrometry,” Journal of Mass Spectrometry 38 (2003): 108–116. [DOI] [PubMed] [Google Scholar]
- 9. Smith R. D., Loo J. A., Loo R. R. O., Busman M., and Udseth H. R., “Principles and Practice of Electrospray Ionization—Mass Spectrometry for Large Polypeptides and Proteins,” Mass Spectrometry Reviews 10, no. 5 (1991): 359–452. [Google Scholar]
- 10. Loo J. A., “Studying Non‐Covalent Protein Complexes by Electrospray Ionization Mass Spectrometry,” Mass Spectrometry Reviews 16 (1997): 1–23. [DOI] [PubMed] [Google Scholar]
- 11. Marshall A. G., Hendrickson C. L., and Jackson G. S., “Fourier Transform Ion Cyclotron Resonance Mass Spectrometry: A Primer,” Mass Spectrometry Reviews 17 (1998): 1–35. [DOI] [PubMed] [Google Scholar]
- 12. Johnson J. T., Lee K. W., Bhanot J. S., and McLuckey S. A., “A Miniaturized Fourier Transform Electrostatic Linear Ion Trap Mass Spectrometer: Mass Range and Resolution,” Journal of the American Society for Mass Spectrometry 30, no. 4 (2019): 588–594. [DOI] [PubMed] [Google Scholar]
- 13. Eliuk S. and Makarov A., “Evolution of Orbitrap Mass Spectrometry Instrumentation,” Annual Review of Analytical Chemistry 8 (2015): 61–80. [DOI] [PubMed] [Google Scholar]
- 14. Liu R., Li Q., and Smith L. M., “Detection of Large Ions in Time‐of‐Flight Mass Spectrometry: Effects of Ion Mass and Accleration Voltage on Microchannel Plate Detector Response,” Journal of the American Society for Mass Spectrometry 25 (2014): 1374–1383. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Reid G. E., Wu J., Chrisman P. A., Wells J. M., and McLuckey S. A., “Charge State Dependent Sequence Analysis of Protonated Ubiquitin Ions via Ion Trap Tandem Mass Spectrometry,” Analytical Chemistry 73 (2001): 3274–3281. [DOI] [PubMed] [Google Scholar]
- 16. Mann M., Meng C. K., and Fenn J. B., “Interpreting Mass Spectra of Multiply‐Charged Ions,” Analytical Chemistry 61 (1989): 1702–1708. [Google Scholar]
- 17. Ferrige A. G., Seddon M. J., Green B. N., Jarvis S. A., and Skilling J., “Disentangling Electrospray Spectra With Maximum‐Entropy,” Rapid Communications in Mass Spectrometry 6 (1992): 707–711. [Google Scholar]
- 18. Cleary S. P., Li H., Bagal D., Loo J. A., Campuzano I. D. G., and Prell J. S., “Extracting Charge and Mass Information From Highly Congested Mass Spectra Using Fourier‐Domain Harmonics,” Journal of the American Society for Mass Spectrometry 29 (2018): 2067–2080. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Marty M. T., Baldwin A. J., Marklund E. G., Hochberg G. K. A., Benesch J. L. P., and Robinson C. V., “Bayesian Deconvolution of Mass and Ion Mobility Spectra: From Binary Interactions to Polydisperse Ensembles,” Analytical Chemistry 87, no. 8 (2015): 4370–4376. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Bern M., Caval T., Kil Y. J., et al., “Parsimonious Charge Deconvolution for Native Mass Spectrometry,” Journal of Proteome Research 17 (2018): 1216–1226. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Henry K. D. and McLafferty F. W., “Electrospray Ionization With Fourier‐Transform Mass Spectrometry‐ Charge State Assignment From Resolved Isotopic Peaks,” Organic Mass Spectrometry 25 (1990): 490–492. [Google Scholar]
- 22. Senko M. W., Beu S. C., and McLafferty F. W., “Automated Assignment of Charge States From Resolved Isotopic Peaks for Multiply Charged Ions,” Journal of the American Society for Mass Spectrometry 6, no. 1 (1995): 52–56. [DOI] [PubMed] [Google Scholar]
- 23. Schultz J. C., Hack C. A., and Benner W. H., “Mass Determination of Megadalton‐DNA Electrospray Ions Using Charge Detection Mass Spectrometry,” Journal of the American Society for Mass Spectrometry 9 (1998): 305–313. [DOI] [PubMed] [Google Scholar]
- 24. Keifer D. Z., Shinholt D. L., and Jarrold M. F., “Charge Detection Mass Spectrometry With Almost Perfect Charge Accuracy,” Analytical Chemistry 87 (2015): 10330–10337. [DOI] [PubMed] [Google Scholar]
- 25. Wörner T. P., Snijder J., Bennett A., Agbandje‐McKenna M., Makarov A. A., and Heck A. J. R., “Resolving Heterogeneous Macromolecular Assemblies by Orbitrap‐Based Single‐Particle Charge Detection Mass Spectrometry,” Nature Methods 17 (2020): 395–398. [DOI] [PubMed] [Google Scholar]
- 26. Kafader J. O., Melani R. D., Senko M. W., Makarov A. A., Kelleher N. L., and Compton P. D., “Measurement of Individual Ions Sharply Increases the Resolution of Orbitrap Mass Spectra of Proteins,” Analytical Chemistry 91 (2019): 2776–2783. [DOI] [PubMed] [Google Scholar]
- 27. Keifer D. Z. and Jarrold M. F., “Single‐Molecule Mass Spectrometry,” Mass Spectrometry Reviews 36 (2017): 715–733. [DOI] [PubMed] [Google Scholar]
- 28. Harper C. C., Elliott A. G., Oltrogge L. M., Savage D. F., and Williams E. R., “Multiplexed Charge Detection Mass Spectrometry for High Throughput Single Ion Analysis of Large Molecules,” Analytical Chemistry 91 (2019): 7458–7465. [DOI] [PubMed] [Google Scholar]
- 29. Kiefer D. Z., Pierson E. E., and Jarrold M. F., “Charge Detection Mass Spectrometry: Weighing Heavier Things,” Analyst 142 (2017): 1654–1671. [DOI] [PubMed] [Google Scholar]
- 30. Smith R. D., Loo J. A., Barinaga C. J., Edmonds C. G., and Udseth H. R., “Collisional Activation and Collision‐Activated Dissociation of Large Multiply Charged Polypeptides and Proteins Producted by Electrospray Ionization,” Journal of the American Society for Mass Spectrometry 1 (1990): 53–65. [DOI] [PubMed] [Google Scholar]
- 31. Horn D. M., Zubarev R. A., and McLafferty F. W., “Automated de Novo Sequencing of Proteins by Tandem High‐Resolution Mass Spectrometry,” Proceedings of the National Academy of Sciences of the United States of America 97 (2000): 10313–10317. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Catalina M. I., Van Den Heuvel R. H. H., van Duijn E., and Heck A. J. R., “Decharging of Globular Proteins and Protein Complexes in Electrospray,” Chemistry ‐ A European Journal 11 (2005): 960–968. [DOI] [PubMed] [Google Scholar]
- 33. Lemaire D., Marie G., Serani L., and Laprévote O., “Stabilization of Gas‐Phase Noncovalent Macromolecular Complexes in Electrospray Mass Spectrometry Using Aqueous Triethylammonnium Bicarbonate Buffer,” Analytical Chemistry 73 (2001): 1699–1706. [DOI] [PubMed] [Google Scholar]
- 34. Pagel K., Hyung S. J., Ruotolo B. T., and Robinson C. V., “Alternate Dissociation Pathways Identified in Charge‐Reduced Protein Complex Ions,” Analytical Chemistry 82 (2010): 5363–5372. [DOI] [PubMed] [Google Scholar]
- 35. Lyu J., McCabe J. W., Schrecke S., Russell D. H., and Laganowsky A., “Discovery of Potent Charge‐Reducing Molecules for Native ion Mobility Mass Spectrometry Studies,” Analytical Chemistry 92 (2020): 11242–11249. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36. Walker T. E., Laganowsky A., and Russell D. H., “Surface Activity of Amines Provides Evidence for the Combined ESI Mechinism of Charge Reduction for Protein Complexes,” Analytical Chemistry 94 (2022): 10824–10831. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37. Lemaire D., Marie G., Serani L., and Laprevote O., "Stabilization of Gas‐Phase Noncovalent Macromolecular Complexes in Electrospray Mass Spectrometry Using Aqueous Triethylammonium Bicarbonate Buffer,” Analytical Chemistry 73 (2001): 1699–1706. [DOI] [PubMed] [Google Scholar]
- 38. Kharlamova A., Prentice B. M., Huang T.‐Y., and McLuckey S. A., “Electrospray Droplet Exposure to Gaseous Acids for the Manipulation of Protein Charge State Distributions,” Analytical Chemistry 82 (2010): 7422–7429. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39. Kharlamova A. and McLuckey S. A., “Negative Electrospray Droplet Exposure to Gaseous Bases for the Manipulation of Protein Charge State Distributions,” Analytical Chemistry 83 (2011): 431–439. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40. Hopper J. T. S., Sokratous K., and Oldham N. J., “Charge State and Adduct Reduction in Electrospray Ionization‐Mass Spectrometry Using Solvent Vapor Exposure,” Analytical Biochemistry 421 (2012): 788–790. [DOI] [PubMed] [Google Scholar]
- 41. Kharlamova A., DeMuth J. C., and McLuckey S. A., “Vapor Treatment of Electrospray Droplets: Evidence for the Folding of Initially Denatured Proteins on the Sub‐Millisecond Time‐Scale,” Journal of the American Society for Mass Spectrometry 23 (2012): 88–101. [DOI] [PubMed] [Google Scholar]
- 42. Konermann L., Ahadi E., Rodriguez A. D., and Vahidi S., “Unraveling the Mechanism of Electrospray Ionization,” Analytical Chemistry 85 (2013): 2–9. [DOI] [PubMed] [Google Scholar]
- 43. Loo R. R. O., Winger B. E., and Smith R. D., “Proton Transfer Reaction Studies of Multiply‐Charged Proteins in a High Mass‐to‐Charge Ratio Quadruple Mass Spectrometer,” Journal of the American Society for Mass Spectrometry 5 (1994): 1064–1071. [DOI] [PubMed] [Google Scholar]
- 44. Loo R. R. O. and Smith R. D., “Proton Transfer Reactions of Multiply‐Charged Peptide and Protein Cations and Anions,” Journal of Mass Spectrometry 30 (1995): 339–347. [Google Scholar]
- 45. Kaufman S. L., Skogen J. W., Dorman F. D., Zarrin F., and Lewis K. C., “Macromolecule Analysis Based on Electrophoretic Mobility in Air: Globular Proteins,” Analytical Chemistry 68 (1996): 1895–1904. [DOI] [PubMed] [Google Scholar]
- 46. Scalf M., Westphall M. S., Krause J., Kaufman S. L., and Smith L. M., “Controlling Charge States of Large Ions,” Science 283 (1999): 194–197. [DOI] [PubMed] [Google Scholar]
- 47. Smith L. M., “Is Charge Reduction in ESI Really Necessary?,” Journal of the American Society for Mass Spectrometry 19 (2008): 629–631. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48. Campuzano I. D. G. and Schnier P. D., “Coupling Electrospray Corona Discharge, Charge Reduction and ion Mobility Mass Spectrometry: From Peptides to Large Macromolecular Protein Complexes,” International Journal of Ion Mobility Spectrometry 16 (2013): 51–60. [Google Scholar]
- 49. Robb D. B., Brown J. M., Morris M., and Blades M. W., “Method of Atmospheric Pressure Charge Stripping for Electrospray Ionization Mass Spectrometry and Its Application for the Analysis of Large Poly(Ethylene Glycol)s,” Analytical Chemistry 86 (2014): 9644–9652. [DOI] [PubMed] [Google Scholar]
- 50. Stutzman J. R., Crowe M. C., Alexander J. N., Bell B. M., and Dunkle M. N., “Coupling Charge Reduction Mass Spectrometry to Liquid Chromatography for Complex Mixture Analysis,” Analytical Chemistry 88 (2016): 4130–4139. [DOI] [PubMed] [Google Scholar]
- 51. de la Mora J. F., “Mobility Analysis of Proteins by Charge Reduction in a Bipolar Electrospray Source,” Analytical Chemistry 90 (2018): 12187–12190. [DOI] [PubMed] [Google Scholar]
- 52. McLuckey S. A., Van Berkel G. J., and Glish G. L., “Reactions of Dimethylamine With Multiply Charged Ions of Cytochrome c ,” Journal of the American Chemical Society 112 (1990): 5668–5670. [Google Scholar]
- 53. Williams E. R., “Proton Transfer Reactivity of Multiply‐Charged Ions,” Journal of Mass Spectrometry 31 (1996): 831–842. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54. Foreman D. J., Bhanot J., Lee K. W., and McLuckey S. A., “Valet Parking for Protein Ion Charge State Concentration: Ion/Molecule Reactions in Linear Ion Traps,” Analytical Chemistry 92 (2020): 5419–5425. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55. McLuckey S. A., Glish G. L., and Van Berkel G. J., “Charge Determination of Product Ions Formed From Collision‐Induced Dissociation of Multiply Protonated Molecules via Ion/Molecule Reactions,” Analytical Chemistry 63 (1991): 1971–1978. [DOI] [PubMed] [Google Scholar]
- 56. McLuckey S. A., Van Berkel G. J., Glish G. L., and Schwartz J. C., Practical Aspects of Ion Trap Mass Spectrometry, Vol. II: Ion Trap Instrumentation, eds. March R. E. and Todd J. F. J. (CRC Press, 1995): 89–141. [Google Scholar]
- 57. J. L. Stephenson, Jr. and McLuckey S. A., “Counting Basic Sites in Oligopeptides via Gas‐Phase Ion Chemistry,” Analytical Chemistry 69 (1997): 281–285. [DOI] [PubMed] [Google Scholar]
- 58. McLuckey S. A., “Biomolecule Ion–Ion Reactions,” in Principles of Mass Spectrometry Applied to Biomolecules, eds. Laskin J. and Lifshitz C. (Wiley‐Interscience, 2006): 519–564. [Google Scholar]
- 59. J. L. Stephenson, Jr. and McLuckey S. A., “Ion/ion Reactions in the Gas‐Phase: Proton Transfer Reactions Involving Multiply‐Charged Proteins,” Journal of the American Chemical Society 118 (1996): 7390–7397. [Google Scholar]
- 60. J. L. Stephenson, Jr. and McLuckey S. A., “Ion/ion Reactions for Oligopeptide Mixture Analysis: Application to Mixtures Comprised of 0.5–100 kDa Components,” Journal of the American Society for Mass Spectrometry 9 (1998): 585–596. [DOI] [PubMed] [Google Scholar]
- 61. Laszlo K. J. and Bush M. F., “Analysis of Native‐Like Proteins and Protein Complexes Using Cation to Anion Proton Transfer Reactions (CAPTR),” Journal of the American Society for Mass Spectrometry 26 (2015): 2152–2161. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62. Herron W. J., Goeringer D. E., and McLuckey S. A., “Product Ion Charge State Determination via Ion/Ion Proton Transfer Reactions,” Analytical Chemistry 68 (1996): 257–262. [DOI] [PubMed] [Google Scholar]
- 63. J. L. Stephenson, Jr. and McLuckey S. A., “Simplification of Product Ion Spectra Derived From Multiply‐Charged Parent Ions via Ion/Ion Chemistry,” Analytical Chemistry 70 (1998): 3533–3544. [DOI] [PubMed] [Google Scholar]
- 64. Coon J. J., Ueberheide B., Syka J. E. P., et al., “Protein Identification Using Sequential Ion/Ion Reactions and Tandem Mass Spectrometry,” Proceedings. National Academy of Sciences. United States of America 102 (2005): 9463–9468. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65. Sanders J. D., Mullen C., Watts E., et al., “Enhanced Sequence Coverage of Large Proteins by Combining Ultraviolet Photodissociation With Proton Transfer Reactions,” Analytical Chemistry 92, no. 1 (2020): 1041–1049. [DOI] [PubMed] [Google Scholar]
- 66. Walker J. N., Lam R., and Brodbelt J. S., “Enhanced Characterization of Histones Using 193 nm Ultraviolet Photodissociation and Proton Transfer Charge Reduction,” Analytical Chemistry 95 (2023): 5985–5993. [DOI] [PubMed] [Google Scholar]
- 67. Xia Y., Wu J., Londry F. A., Hager J. W., and McLuckey S. A., “Mutual Storage Mode Ion/Ion Reactions in Hybrid Linear Ion Trap,” Journal of the American Society for Mass Spectrometry 16 (2005): 71–81. [DOI] [PubMed] [Google Scholar]
- 68. Wu J., Hager J. W., Xia Y., Londry F. A., and McLuckey S. A., “Positive ion Transmission Mode Ion/Ion Reactions in a Hybrid Linear Ion Trap,” Analytical Chemistry 76 (2004): 5006–5015. [DOI] [PubMed] [Google Scholar]
- 69. Liang X. and McLuckey S. A., “Transmission Mode Ion/Ion Proton Transfer Reactions in a Linear Ion Trap,” Journal of the American Society for Mass Spectrometry 18 (2007): 882–890. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 70. McLuckey S. A., Reid G. E., and Wells J. M., “Ion Parking During Ion/Ion Reactions in Electrodynamic Ion Traps,” Analytical Chemistry 74, no. 2 (2002): 336–346. [DOI] [PubMed] [Google Scholar]
- 71. Chrisman P. A., Pitteri S. J., and McLuckey S. A., “Parallel Ion Parking of Protein Mixtures,” Analytical Chemistry 78 (2006): 310–316. [DOI] [PubMed] [Google Scholar]
- 72. Ugrin S. A., English A. M., Syka J. E. P., et al., “Ion‐Ion Proton Transfer and Parallel Ion Parking for the Analysis of Mixtures of Intact Proteins on a Modified Orbitrap Mass Analyzer,” Journal of the American Society for Mass Spectrometry 30 (2019): 2163–2173. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73. Weisbrod C. R., Anderson L. C., Hendrickson C. L., et al., “Advanced Strategies for Proton‐Transfer Reactions Coupled With Parallel Ion Parking on a 21 T FT‐ICR MS for Intact Protein Analysis,” Analytical Chemistry 93 (2021): 9119–9128. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74. McGee W. M. and Brodbelt J. S., “Integration of Ultraviolet Photodissociation With Proton Transfer Reactions and Ion Parking for Analysis of Intact Proteins,” Analytical Chemistry 88 (2016): 1008–1016. [DOI] [PubMed] [Google Scholar]
- 75. Reid G. E., Shang H., Hogan J. M., Lee G. U., and McLuckey S. A., “Gas‐Phase Concentration, Purification, and Identification of Whole Proteins From Complex Mixtures,” Journal of the American Chemical Society 124 (2002): 7353–7362. [DOI] [PubMed] [Google Scholar]
- 76. Zubarev R., Kelleher N. L., and McLafferty F. W., “Electron Capture Dissociation of Multiply‐Charged Protein Cations. A Non‐Ergodic Process,” Journal of the American Chemical Society 120 (1998): 3265–3266. [Google Scholar]
- 77. Zubarev R., “Electron‐Capture Dissociation Tandem Mass Spectrometry,” Current Opinion in Biotechnology 15 (2004): 12–16. [DOI] [PubMed] [Google Scholar]
- 78. Syka J. E. P., Coon J. J., Schroeder M. J., Shabanowitz J., and Hunt D. F., “Peptide and Protein Sequence Analysis by Electron Transfer Dissociation Mass Spectrometry,” Proceedings. National Academy of Sciences. United States of America 101 (2004): 9528–9533. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 79. Riley N. M. and Coon J. J., “The Role of Electron Transfer Dissociation in Modern Proteomics,” Analytical Chemistry 90 (2018): 40–64. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 80. Leney A. C. and Heck A. J. R., “Native Mass Spectrometry: What Is in the Name?,” Journal of the American Society for Mass Spectrometry 28 (2017): 5–13. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 81. Rostom A. A. and Robinson C. V., “Detection of the Intact GroEL Chaperonin Assembly by Mass Spectrometry,” Journal of the American Chemical Society 121 (1999): 4718–4719. [Google Scholar]
- 82. Heck A. J. R., “Native Mass Spectrometry: A Bridge Between Interactomics and Structural Biology,” Nature Methods 5 (2008): 927–933. [DOI] [PubMed] [Google Scholar]
- 83. Le Huray K. I. P., Worner T. P., Moreira T., et al., “To 200,000 m/z and Beyond: Native Electron Capture Charge Reduction Mass Spectrometry Deconvolves Heterogeneous Signals in Large Biopharmaceutical Analytes,” ACS Central Science 10 (2024): 1548–1561. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 84. Shaw J. B., Harvey S. R., Du C., et al., “Protein Complex Heterogeneity and Topology Revealed by Electron Capture Charge Reduction and Surface Induced Dissociation,” ACS Central Science 10 (2024): 1537–1547. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 85. Wells J. M., Chrisman P. A., and McLuckey S. A., “Formation and Characterization of Protein‐Protein Complexes in Vacuo,” Journal of the American Chemical Society 125 (2003): 7238–7249. [DOI] [PubMed] [Google Scholar]
- 86. Wu J. and McLuckey S. A., “Ion/ion Reactions of Multiply Charged Nucleic Acid Anions: Electron Transfer, Proton Transfer, and Ion Attachment,” International Journal of Mass Spectrometry 228 (2003): 577–597. [Google Scholar]
- 87. Wells J. M., Chrisman P. A., and McLuckey S. A., “Formation of Protein‐Protein Complexes in Vacuo ,” Journal of the American Chemical Society 123 (2001): 12428–12429. [DOI] [PubMed] [Google Scholar]
- 88. Chao H.‐C., Shih M., and McLuckey S. A., “Generation of Multiply Charged Protein Anions From Multiply Charged Protein Cations via Gas‐Phase Ion/Ion Reactions,” Journal of the American Society for Mass Spectrometry 31 (2020): 1509–1517. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89. Liu J., Chrisman P. A., Erickson D. E., and McLuckey S. A., “Relative Information Content and Top‐Down Proteomics by Mass Spectrometry: The Utility of Ion/Ion Proton‐Transfer Reactions in Electrospray‐Based Approaches,” Analytical Chemistry 79 (2007): 1073–1081. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 90. McLuckey S. A. and Goeringer D. E., “Ion/Molecule Reactions for Improved Effective Mass Resolution in Electrospray Mass Spectrometry,” Analytical Chemistry 67 (1995): 2493–2497. [DOI] [PubMed] [Google Scholar]
- 91. Abzalimov R. R. and Kaltashov I. A., “Electrospray Ionization Mass Spectrometry of Highly Heterogeneous Protein Systems: Protein Ion Charge State Assignment via Incomplete Charge Reduction,” Analytical Chemistry 82, no. 18 (2010): 7523–7526. [DOI] [PubMed] [Google Scholar]
- 92. Muneeruddin K., Bobst C. E., Frenkel R., et al., “Characterization of a PEGylated Protein Therapeutic by Ion Exchange Chromatography With on‐Line Detection by Native ESI MS and MS/MS,” Analyst 142, no. 2 (2017): 336–344. [DOI] [PubMed] [Google Scholar]
- 93. Abdillahi A. M., Lee K. W., and McLuckey S. A., “Mass Analysis of Macro‐Molecular Analytes via Multiply‐Charged Ion Attachment,” Analytical Chemistry 92 (2020): 16301–16306. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 94. McLuckey S. A., Goeringer D. E., and Glish G. L., “Selective Ion Isolation/Rejection Over a Broad Mass Range in the Quadrupole Ion Trap,” Journal of the American Society for Mass Spectrometry 2 (1991): 11–21. [DOI] [PubMed] [Google Scholar]
- 95. Pitts‐McCoy A. M., Abdillahi A. M., Lee K. W., and McLuckey S. A., “Multiply‐Charged Cation Attachment to Facilitate Mass Measurement in Negative Mode Native Mass Spectrometry,” Analytical Chemistry 94 (2022): 2220–2226. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 96. Huguet R., Mullen C., Srzentić K., et al., “Proton Transfer Charge Reduction Enables High‐Throughput Top‐Down Analysis of Large Proteoforms,” Analytical Chemistry 91, no. 24 (2019): 15732–15739. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 97. Fornelli L., Durbin K. R., Fellers R. T., et al., “Advancing Top‐Down Analysis of the Human Proteome Using a Benchtop Quadrupole‐Orbitrap Mass Spectrometer,” Journal of Proteome Research 16 (2017): 609–618. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 98. Chao H.‐C., Lee K. W., Shih M., and McLuckey S. A., “Characterization of Homopolymer Distributions via Direct Infusion ESI‐MS/MS Using Wide Mass‐To‐Charge Windows and Gas‐Phase Ion/Ion Reactions,” Journal of the American Society for Mass Spectrometry 33 (2022): 704–713. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Data sharing is not applicable to this article as no datasets were generated or analyzed during the current study.
