Validity and Agreement of a Portable Dynamometer in Evaluating Hip Strength and Rate of Torque Development: Influence of Fixation Methods

Sungwan Kim 1,, Takhyun Yoo 1, Julie P Burland 1, Neal R Glaviano 1
PMCID: PMC11964701  PMID: 40182909

Abstract

Background

Portable dynamometers (PD) are widely used to assess hip muscle strength; however, few studies have explored their validity and agreement when evaluating the rate of torque development (RTD) of hip muscles, particularly with different fixation methods.

Hypothesis/Purpose

The purpose of this study was to determine the validity and agreement of isometric hip strength and RTD using the PD with two fixation methods: placing the PD on the body (PDpush) or a treatment table (PDpull), compared to an isokinetic dynamometer (IKD). It was hypothesized that an PD would be a valid tool for assessing hip strength and RTD, with PDpull demonstrating strong validity and PDpush showing moderate validity; and its assessments would yield higher values than those obtained from IKD.

Study Design

Cross-sectional study.

Methods

Twenty healthy, physically active individuals (10 females and 10 males) were included in this study. Hip abduction and extension strength (peak torque) and RTD (early phase: RTD0-100 ms and late phase: RTD100-200 ms) were assessed using PDpush, PDpull, and an IKD.

Results

Moderate-to-strong correlations were found between PD (both PDpush and PDpull) and IKD for all hip strength (r=0.46-0.73; p<0.05) and RTD (r=0.45-0.73, ρ=0.80; p<0.05) measures. PDpull demonstrated strong correlations for hip abduction RTD0-100 ms and hip extension strength and RTD100-200 ms (r=0.70-0.73; p<0.001), whereas PDpush showed moderate correlations (r=0.45-0.58; p<0.05). Bland-Altman plots with mean differences ± 95% limits of agreement indicated that the PDpush method overestimated all hip abduction and extension strength and RTD as compared to IKD. The PDpull method overestimated hip abduction and extension strength but had mixed results for RTD, either underestimating hip abduction RTD while overestimating hip extension RTD.

Conclusion

The PD is valid tool for evaluating hip strength and RTD, with PDpull offering strong validity for certain measures; however, PD measurements did not agree with the absolute values obtained from IKD.

Level of Evidence

Level 3

Keywords: dynamometry, hip abduction, hip extension, neuromuscular function, torque-generating capacity

INTRODUCTION

Hip muscles are essential for lower extremity stability and postural control.1,2 Deficiencies in hip strength have been linked to various musculoskeletal pathologies (e.g., anterior cruciate ligament sprain,3,4 patellofemoral pain,5,6 chronic ankle instability7,8); hence, objective assessments of hip strength are widely performed in both clinical and laboratory settings.9,10 While an isokinetic dynamometer (IKD) is regarded as the “gold standard” for assessing hip strength,9,11 it is often impractical for clinical use due to its cost, lack of portability, and the need for substantial space. As an alternative, portable dynamometers (PD) are used in clinical practice as a reliable and valid testing option for evaluating hip strength.9,12

One important aspect of muscle capacity is the ability to generate torque rapidly, known as the rate of torque development (RTD).13 This concept has gained attention in the literature14,15 because RTD, which accounts for the time-dependent characteristics of muscle contraction, shows stronger associations with daily and sport activities than traditional peak isometric strength measures.13 Recent evidence supports the importance of RTD in pathological conditions, demonstrating that lower hip RTD, but not strength, has been associated with greater lower extremity frontal plane projection angles during functional tasks in individuals with patellofemoral pain.16 Given that RTD can be calculated from a maximal voluntary isometric contraction,17,18 assessing RTD using a PD appears feasible in clinical practice. Based on previous evidence,13,19,20 an in-depth analysis of RTD, distinguishing between the early (e.g., RTD0-100_ms) and late (e.g., RTD100-200_ms) phases, may provide insights into more specific neuromuscular impairments, such as motor unit recruitment and muscle fiber type. Clinicians may consider incorporating fast-velocity concentric and eccentric resistance training to address deficiencies in early-phase RTD,21,22 while non-explosive high-intensity resistance training for improvements in late-phase RTD.20

Current literature recommends the belt-stabilized PD method since the assessor’s sex or strength can influence PD assessments.23,24 The belt-stabilized PD can be performed through two different methods: the push (PDpush) and pull (PDpull) methods.25 The PDpush method entails placing the PD directly on the individual’s body, whereas the PDpull method positions the PD against the back of a stationary object (e.g., treatment table) to measure force by the individual exerting pressure against the rigid structure. The PDpull method has previously exhibited greater validity and offered more comfortable performance (i.e., less perceived patient discomfort) for assessing knee extension strength than the PDpush method,25 indicating that the choice of fixation method could affect the assessment outcomes. Some studies26–28 suggest that the PD can be a valid tool for measuring hip strength and RTD; however, there is no research on the validity between the two different fixation methods (PDpush vs. PDpull) and how these measurements compare to the absolute values obtained by IKD. Evidence shows that PD often overestimates hip strength values compared to those measured by the IKD.29,30

The primary purpose of this study was to determine the validity and agreement of hip abduction and extension isometric strength and RTD measurements using an PD with two different fixation methods (PDpush or PDpull) in comparison with an IKD. The secondary purpose of this study was to compare the level of discomfort among the three different assessment methods (PDpush vs. PDpull vs. IKD). It was hypothesized that an PD would be a valid tool for assessing hip strength and RTD, with the PDpull method demonstrating strong validity and the PDpush method showing moderate validity while its assessments would yield higher values than those obtained from IKD. It was also hypothesized that the IKD would be more comfortable than both the PDpush and PDpull methods, with the PDpull method being more comfortable than the PDpush method.

MATERIALS AND METHODS

This was a cross-sectional study reported according to the Standards for Reporting Diagnostic accuracy studies (STARD) guidelines.31 The independent variables were the device and fixation method (PDpush, PDpull, and IKD). The dependent variables were hip (abduction and extension) neuromuscular function (strength and RTD) and level of discomfort.

Participants

A priori sample size calculation was performed using G*Power software (version 3.1.9.6; Universität Düsseldorf, Düsseldorf, Germany). The effect size was estimated based on a previous study32 that investigated the validity of knee extension RTD assessment using an PD. To achieve a power analysis with an α=.05, β=.20 (80% power), and a one-tailed test, a minimum of 20 participants were necessary for this study.

Twenty healthy, physically active individuals were recruited from a local university and surrounding community via flyers and advertisements between March 2024 and May 2024. To be eligible for participation in this study, participants had to (1) be between the ages of 18 and 35 years and (2) self-report engaging in physical activity of moderate-to-vigorous intensity for at least 30 minutes per day on a minimum of three days per week. Participants were excluded if they had (1) a history of low back or lower extremity surgery; (2) low back or lower extremity injury or pain within the prior six months; or (3) were diagnosed neurological or cognitive impairment. Participants were verbally instructed to refrain from lower extremity resistance training 24 hours before testing. The study was approved by the University of Connecticut Institutional Review Board, and written informed consent was obtained from all participants.

Procedures

Once the eligibility screening and informed consent were completed, the participants’ age, height, and mass were obtained. Current physical activity levels were then assessed using the Tegner Activity Scale, a reliable and valid single-item questionnaire that measures sport and work activity levels.33 The Tegner Activity Scale ranges from 0 to 10, with 10 being the most active.33 Prior to testing, participants were asked to wear comfortable clothing. The dominant limb, defined as the limb used to kick a ball, was evaluated for all testing. A single assessor (XX) with eight years of experience conducting strength and RTD assessments with IKD and PD completed all data collection.

All participants underwent maximal voluntary isometric contractions of hip abduction and extension using an IKD (Biodex System Pro 4; Biodex Medical Systems Inc., Shirley, NY) (Figure 1), followed by assessments using an PD (microFET2; Hoggan Scientific LLC., Salt Lake City, UT) (Figure 2). All devices were calibrated at the beginning of the study following the manufacturer’s operational guidelines to ensure accurate measurements throughout the study. Data was collected between April 2024 to May 2024.

Figure 1. Isokinetic dynamometer assessment, depicting the testing positions for (A) hip abduction and (B) hip extension.

Figure 1.

Figure 2. Portable dynamometer assessment, depicting the testing positions for hip abduction with (A) push and (B) pull methods and hip extension with (C) push and (D) pull methods.

Figure 2.

The testing order for hip muscle groups (extension and abduction) and fixation methods for PD (push and pull) was randomized by a coin toss. A coin toss was conducted in two stages: first to determine the muscle group, followed by a second toss to select the fixation method. To familiarize themselves with the muscle contractions, participants performed a series of gradual warm-up sessions, involving three-second isometric contractions at 50%, 75%, and 100% of their self-perceived effort.18 Each level of exertion was performed only once, with a one-minute rest between each level. No additional warm-up activities were permitted to minimize fatigue. After completing the warm-up, participants were given a rest period of up to three minutes before the testing began. Subsequently, participants were instructed to exert maximal force as hard and fast as possible for actual data collection. Standardized verbal encouragement was also provided during each trial to elicit their maximum effort. The assessments consisted of three trials with five-second maximal contractions, with participants given a one-minute rest period between trials and a five-minute rest period between testing positions. We confirmed verbally that participants were not experiencing perceived fatigue immediately before each trial; and, additional rest periods were provided if necessary to ensure they could perform the subsequent trial with maximum effort. Following each trial, participants were asked to rate their level of discomfort related to the placement of the PD or IKD by drawing a vertical line on a 10 cm visual analog scale (VAS: 0=no discomfort; 10=worst discomfort imaginable). They were carefully instructed to exclude any discomfort related to muscle soreness during maximal force exertion.

IKD Assessment

The prone position for hip extension and the side-lying position for hip abduction were used, which provide reliable and consistent measurements of hip muscles.9

For the hip abduction measurement, participants were placed in a side-lying position, with the testing limb placed on top of the non-testing limb at 20° of hip abduction supported by the IKD (Figure 1A).34 To minimize accessory motions, the non-testing limb and trunk were stabilized with straps. The axis of rotation was aligned anteriorly to the anterior superior iliac spine at the level of the greater trochanter of the testing limb. The center of the lever arm was positioned at a point 5 cm above the lateral femoral condyle. Participants were instructed to raise their testing limb towards the ceiling with maximum force and speed.

For the hip extension measurement, participants were placed in a prone position, with both the hip and knee of the testing limb in 90° of flexion (Figure 1B).34 The non-testing limb was comfortably supported on a stool with a cushion. To minimize accessory motion, participants were instructed to warp their arms around the chair of the dynamometer, and their trunk was stabilized with straps. The axis of rotation was aligned with the greater trochanter on the testing limb. The center of the lever arm was positioned at 5 cm above the popliteal fossa. Participants were instructed to kick their testing limb back with maximum force and speed.

For both measurements, participants were instructed to start from a fully relaxed state. Colored stickers were applied to mark both the axis of rotation and the point of resistance to ensure a consistent lever arm for PD assessments. Considering potential sticker movement, anatomical landmarks were confirmed prior to each trial to precisely detect the axis of rotation. The distance between markers was measured using a standard tape measure to determine the lever arm length (m).

PD Assessment

Participants conducted the PD assessments using a treatment table, and all methods were consistent with those employed during the IKD assessment. For the PD assessments, stabilization straps (i.e., gait belts) were utilized to secure the PD and stabilize participants’ non-testing limb and trunk. The belt length was adjusted to achieve the same joint angles as in the IKD assessment at the participant’s hip during maximal abduction (Figures 2A and 2B) and extension (Figure 2C and 2D), as measured using a goniometer.

During hip abduction measurements, the assessor supported the participant’s limb to counteract gravity, ensuring participants remained fully relaxed without exerting any effort to lift their limbs. During hip extension measurements, with the hip flexed at 90°, the thigh’s orientation was aligned parallel to gravity, omitting the need for any support from the assessor. However, since the PD’s force threshold was 10 N, participants were instructed to apply a minimum pre-tension before starting the assessment, while still being advised to maintain a relaxed state as much as possible prior to trials.

For the PDpush method (Figure 2A and 2C), the PD was placed directly on the participant’s body and positioned at the location where resistance pads were used during the IKD assessments. During the PDpush method, force was measured by pushing the PD against the gait belt as participants abducted or extended their hips. For the PDpull method (Figure 2B and 2D), the PD was situated against the back of the treatment table, employing the same level of positioning as in the PDpush method. During the PDpull method, force was measured by compressing the PD against the table as participants abducted or extended their hips. A flat attachment was used for both fixation methods.

Data Processing

The IKD was interfaced with a data acquisition system (Noraxon USA Inc., Scottsdale, AZ) and recorded at a sampling rate of 1500 Hz. The PD, with a fixed sampling frequency of 100 Hz, wirelessly transmitted force signal data to a laptop computer via a manufacturer-supplied USB receiver using ergoPAK software (version 2.2.1; Hoggan Scientific LLC., Salt Lake City, UT).

Since the PD measures force (N), torque (Nm) was calculated by multiplying the force (N) by the lever arm (m). For hip strength data, peak torque was obtained from maximal voluntary isometric contractions and then normalized to body mass (Nm/kg). The RTD data were computed using custom MATLAB code (MathWorks Inc., Natick, MA), which calculated the slopes within two different time intervals: (1) RTD0-100_ms (early phase) and (2) RTD100-200_ms (late phase).35,36 Each slope value was then normalized to each time interval and body mass (Nm/s/kg). Three trials of hip abduction and extension strength and RTD were averaged for each testing condition, and the average was used for analysis.

The average level of discomfort (cm) recorded across the three trials of hip abduction and extension RTD for each testing condition was also calculated.

Statistical Analysis

Descriptive statistics (mean ± standard deviation) were calculated for all variables. Data were inspected for normality using Shapiro-Wilk tests. Concurrent validity was assessed using Pearson r or Spearman ρ correlation coefficients where appropriate. The strength of the correlations was interpreted as weak (<0.40), moderate (0.40-0.70), and strong (>0.70).37 Agreement between methods was assessed using Bland-Altman plots with mean differences ± 95% limits of agreement (LOA). Root mean square error (RMSE) and mean difference were also calculated to assess the overall magnitude of error and identify any systematic differences between the measurements. Differences in hip strength, RTD and level of discomfort among the three different testing conditions (PDpush vs. PDpull vs. IKD) were determined using repeated measures analysis of variance (ANOVA) or Friedman tests, based on normality. Relevant post-hoc tests (paired t or Wilcoxon signed-rank tests) were performed in the presence of significant findings, to determine differences between testing conditions. All correlation analyses, repeated measures ANOVA, Friedman tests, and post-hoc tests were conducted in SPSS (version 29.0; SPSS Inc., Chicago, IL), with an a priori alpha of p < 0.05. Bland-Altman plots with mean differences ± 95% LOA were calculated using Microsoft Excel 2018 (Microsoft Corp., Redmond, WA).

RESULTS

Participant demographics are presented in Table 1. All strength and RTD data were normally distributed (p>0.05), except for hip extension RTD0-100_ms (p=0.013). None of the discomfort level data were normally distributed (p<0.05).

Table 1. Participant Demographics.

Variable All participants (n=20)
Gender (number; female/male) 10/10
Age (year) 28.1 ± 2.6
Height (cm) 169.2 ± 8.0
Mass (kg) 65.9 ± 12.5
Dominant limb (number; right/left) 19/1
Tegner Activity Scale (range, 0-10) 4.8 ± 0.8

Data are presented as mean ± standard deviation, except for gender and dominant limb.

Abbreviations: cm, centimeter; kg, kilogram.

Validity

Significant moderate-to-strong correlations were found between PD (both PDpush and PDpull methods) and IKD for all hip strength (r=0.46-0.73; p<0.05) and RTD (r=0.45-0.73, ρ=0.80; p<0.05) measures (Table 2). The PDpull method demonstrated strong correlations for hip abduction RTD0-100_ms and hip extension strength and RTD100-200_ms (r=0.70-0.73; p<0.001), whereas the PDpush method showed moderate correlations (r=0.45-0.58; p<0.05) (Table 2).

Table 2. Hip Strength and Rate of Torque Development.

Variable PDpush PDpull IKD Correlation coefficient Repeated measures ANOVA or Friedman tests Post-hoc tests
Hip abduction
Peak torque (Nm/kg) 1.02 ± 0.25 0.80 ± 0.23 0.67 ± 0.26 PDpush vs. IKD: r=0.46; p=.044 p=.017 PDpush vs. IKD: p<.001
PDpull vs. IKD: r=0.58; p=.007 PDpull vs. IKD: p=.017
RTD0-100 ms (Nm/s/kg) 4.98 ± 1.76 3.52 ± 0.97 4.82 ± 1.37 PDpush vs. IKD: r=0.45; p=.045 p<.001 PDpush vs. IKD: p=.663
PDpull vs. IKD: r=0.73; p<.001 PDpull vs. IKD: p<.001
RTD100-200 ms (Nm/s/kg) 4.51 ± 2.06 3.24 ± 1.07 4.07 ± 1.51 PDpush vs. IKD: r=0.45; p=.045 p=.014 PDpush vs. IKD: p=.320
PDpull vs. IKD: r=0.48; p=.032 PDpull vs. IKD: p=.014
Hip extension
Peak torque (Nm/kg) 1.58 ± 0.35 2.99 ± 0.90 1.10 ± 0.49 PDpush vs. IKD: r=0.56; p=.010 p<.001 PDpush vs. IKD: p<.001
PDpull vs. IKD: r=0.73; p<.001 PDpull vs. IKD: p<.001
RTD0-100 ms (Nm/s/kg) 6.54 ± 2.18 11.70 ± 4.55 5.83 ± 3.15 PDpush vs. IKD: ρ=0.80; p<.001 p<.001 PDpush vs. IKD: p=.131
PDpull vs. IKD: ρ=0.80; p<.001 PDpull vs. IKD: p<.001
RTD100-200 ms (Nm/s/kg) 5.47 ± 1.54 9.78 ± 3.35 4.79 ± 2.46 PDpush vs. IKD: r=0.58; p=.007 p<.001 PDpush vs. IKD: p=.146
PDpull vs. IKD: r=0.70; p<.001 PDpull vs. IKD: p<.001

Data are presented as mean ± standard deviation.

Abbreviations: ANOVA, analysis of variance; PDpull, portable dynamometer: pull method; PDpush, portable dynamometer: push method; IKD, isokinetic dynamometer; kg, kilogram; ms, millisecond; Nm, newton-meter; RTD, rate of torque development; s, second.

Agreement

Bland-Altman plots with mean differences ± 95% LOA indicated that the PDpush method overestimated all hip abduction and extension strength (0.35±0.27 to 0.48±0.41 Nm/kg; Figure 3) and RTD (0.17±1.67 to 0.71±1.81 Nm/s/kg; Figures 4 and 5). The PDpull method overestimated hip abduction and extension strength (0.13±0.23 to 1.89±0.64 Nm/kg; Figure 3) but had mixed results for RTD, either underestimating hip abduction RTD (-1.30±0.94 to -0.83±1.37 Nm/s/kg; Figure 4) or overestimating hip extension RTD (4.99±2.39 to 5.87±2.85 Nm/s/kg; Figure 5).

Figure 3. Bland-Altman plots for hip abduction and extension strength.

Figure 3.

The solid line indicates the mean differences and the dotted lines indicate the upper and lower limits of agreement. Abbreviations: PDpull, portable dynamometer: pull method; PDpush, portable dynamometer: push method; IKD, isokinetic dynamometer; SD, standard deviation.

Figure 4. Bland-Altman plots for hip abduction rate of torque development.

Figure 4.

The solid line indicates the mean differences and the dotted lines indicate the upper and lower limits of agreement. Abbreviations: PDpull, portable dynamometer: pull method; PDpush, portable dynamometer: push method; IKD, isokinetic dynamometer; RTD, rate of torque development; SD, standard deviation.

Figure 5. Bland-Altman plots for hip extension rate of torque development.

Figure 5.

The solid line indicates the mean differences and the dotted lines indicate the upper and lower limits of agreement. Abbreviations: PDpull, portable dynamometer: pull method; PDpush, portable dynamometer: push method; IKD, isokinetic dynamometer; RTD, rate of torque development; SD, standard deviation.

There were significant differences in hip strength and RTD among the three testing methods (p≤0.017; Table 2). Both the PDpush and PDpull methods resulted in significantly greater hip abduction and extension strength compared to IKD (p≤0.017; Table 2). The PDpull method resulted in significantly lower hip abduction RTD (RTD0-100_ms: p<0.001; RTD100-200_ms: p=0.014) and greater hip extension RTD (RTD0-100_ms: p<0.001; RTD100-200_ms: p<0.001) compared to IKD (Table 2); yet, there were no significant differences in any hip RTD measures between the PDpush method and IKD (p>0.05; Table 2). RMSE and mean difference between the measurements are provided in Supplementary Material 1.

Level of Discomfort

No significant differences in the level of discomfort were observed for both hip abduction (p=0.18) and extension (p=.027) measures among the three testing methods (Table 3).

Table 3. Discomfort Level.

Variable Discomfort level
Hip abduction measures
Portable dynamometer: push method (VAS; cm) 1.8 ± 2.1
Portable dynamometer: pull method (VAS; cm) 1.2 ± 1.3
Isokinetic dynamometer (VAS; cm) 1.1 ± 1.2
Hip extension measures
Portable dynamometer: push method (VAS; cm) 1.7 ± 1.5
Portable dynamometer: pull method (VAS; cm) 1.8 ± 2.4
Isokinetic dynamometer (VAS; cm) 1.1 ± 1.0

Data are presented as mean ± standard deviation.

Abbreviations: cm, centimeter; VAS, visual analog scale.

DISCUSSION

The purpose of this study was two-fold: (1) to determine the validity and agreement of hip abduction and extension strength and RTD using an PD with two different fixation methods (PDpush or PDpull) in comparison with an IKD and (2) to compare the level of discomfort among the three assessment methods (PDpush vs. PDpull vs. IKD). These findings indicate that the PD provides valid measures of hip strength and RTD, with the PDpull method demonstrating strong validity for specific measures. However, the absolute values did not show concordance between the PDpush or PDpull and the IKD. There were no significant differences in the level of discomfort among the assessment methods.

Moderate-to-strong correlations were found between the PD and IKD for hip abduction and extension strength measures (r=0.46-0.73), consistent with earlier findings (r=0.42-0.84).29,38,39 Similarly, the current study found moderate-to-strong correlations between the PD and IKD for all hip abduction and extension RTD measures (r=0.45-0.73, ρ=0.80), regardless of the PD fixation methods. The validity of our PDpush method, showing moderate-to-strong correlations, aligns with a recent study.26 Another study27 also reported moderate associations between the PD and IKD for hip abduction and extension RTD; although this study27 compared PD methods with belt-stabilization and tension. These data indicate strong correlations with the PDpull method for specific measures, where the PD was directly compressed against a stationary table (i.e., without a lockout cable used in the previous study27). Nevertheless, direct comparisons with these prior studies26–28 are challenging due to methodological differences, such as variations in testing positions (e.g., supine or standing for hip abduction, fully prone for hip extension).

One unique and notable finding of the current study was that the PDpull method exhibited strong validity for certain hip strength and RTD measures (r=0.70-0.73), while the PDpush method showed moderate validity (r=0.45-0.58), despite no significant differences in discomfort levels on the VAS between the fixation methods (PDpush: 1.7-1.8 cm vs. PDpull: 1.2-1.8 cm). In contrast, a prior report25 examining RTD in knee extension found strong correlations between IKD and both the PDpush (r=0.87) and PDpull (r=0.93) methods, even though the PDpush method reported significantly higher discomfort levels on the VAS (PDpush: 3.1 cm vs. PDpull: 1.1 cm). These findings suggest that discomfort levels may not be a significant factor affecting hip strength and RTD assessments. One potential reason for the relatively lower validity (moderate) of our PDpush method might be the positioning location of the PD. The PD was positioned at a point 5 cm above the lateral femoral condyle or popliteal fossa, which engages regions more abundant in soft tissue and muscle (iliotibial band/vastus lateralis for hip abduction and hamstrings for hip extension), unlike its placement on the more rigid bone (tibia) for knee extension assessment.25 Such placements on bulkier areas during the PDpush method potentially absorb some of the force and/or cause the device to become unstable when generating force momentarily. This speculation may be supported by the strong validity of the PDpull method, as it provides a more stable environment by exerting pressure against a more rigid object. Still, this is all conjecture, and future research is required to investigate whether different PD positioning locations influence force absorption and/or measurement stability and validity across various muscle groups and body regions.

Although moderate-to-high validity was found across all measures, the Bland-Altman plots revealed a tendency towards systematic bias in hip abduction and extension strength and RTD measurements between PD and IKD. These results were expected, as the disagreement between the PD and IKD has often been reported in previous studies.25,29,30,40 Both our PDpush and PDpull methods consistently overestimated hip strength, which aligns with prior data.29,30 Both fixation methods also tended to overestimate most hip RTD measures, except for hip abduction, where the PDpull method underestimated RTD (-0.79 ± 1.32 to -1.30 ± 0.94 Nm/s/kg). Assessing hip RTD using two distinct fixation methods is relatively novel, with limited literature available for direct comparisons.

One possible contributing factor to the underestimation of hip abduction RTD might be the testing position. Unlike the prone position used for hip extension (Figure 2C and 2D), hip abduction was evaluated at 20° in a side-lying position (Figure 2A and 2B), which requires support against gravity. While the IKD offers full support for limb weight during 20° of hip abduction, PD assessments require the assessor to stabilize the limb with pre-tension on the gait belt. The manual stabilization by the assessor may not be as robust as the fixed IKD apparatus. Despite efforts to ensure limb relaxation before trials, the lack of complete support may have led participants to preemptively tense their muscles, especially compromising their ability to generate rapid torque at the beginning of muscle contraction. Since RTD measurements necessitate starting from a relaxed state, testing positions against gravity may not be ideal when using the PD. Future studies may consider the force at the starting point for RTD in their analysis or could utilize electromyography to verify that trials start from a relaxed condition.

Another potential explanation for the underestimation of hip abduction RTD in the PDpull method could be the difference in interface alignment. In the PDpush method for hip abduction (Figure 2A), the PD interface remains parallel to the body, allowing for direct force application, whereas the PDpull method (Figure 2B) introduces a slope between the gait belt and treatment table, altering the angle of direct force application. Future investigations should examine the influences of various testing positions and identify optimal positions for accurate assessment. Nevertheless, the PD can still be deemed a valid tool, given that it evaluates what it purports to measure and demonstrates moderate-to-strong correlations with the criterion standard.

Clinical Implications

In clinical settings, the PD is preferred for its practicality, especially given its ease of use and cost-effectiveness. Given the growing evidence linking RTD deficits to various orthopedic pathologies, such as patellofemoral pain16 and total knee arthroplasty,41 monitoring hip RTD alongside strength in routine clinical practice could be crucial for preventative measures in healthy individuals. The data supports that the PD provides valid measures for both hip strength and RTD from the same time-torque graph during maximal voluntary isometric contraction.28,42 The simultaneous assessment of strength and RTD streamlines the process, eliminating the need for additional resources or time dedicated to evaluating each measure separately, thereby enhancing clinical efficiency. However, it should be acknowledged that coding for RTD calculation may not be common practice in clinical settings. It is recommended to obtain RTD from a torque-time curve graph by manually identifying specific torque and time points (e.g., 0-100 ms).

Based on our findings, it is also recommended that clinicians prioritize using the PDpull method to obtain more valid measures compared to the PDpush method. These findings have practical implications due to the potential reproducibility of evaluation settings, such as acquiring gait belts and treatment tables that would allow for PD testing. The PDpull method may also reduce the need for clinicians to stabilize the device during measurements, which could simplify the assessment process and minimize potential clinician-induced variability. However, it is important to note that the absolute values derived from PD should not be interpreted as equivalent to those obtained from IKD, and vice versa. Clinicians should refrain from interchangeably using the PD and IKD when monitoring assessments across different time points, individuals, or limbs.

Limitations

The sample in this study comprised healthy, physically active individuals aged between 18 and 35 years limiting the applicability to those outside this age range or those with pathological conditions or sedentary lifestyles. The data is limited to findings for hip abduction with a side-lying position with a 20° abduction and hip extension with a standing position with both hip and knee angles at 90°. Future research should explore hip neuromuscular function across various testing positions and joint angles and examine other muscle groups (e.g., hip external and internal rotation) using similar PD configurations. Another limitation is the potential measurement issue caused by not aligning the PD perpendicularly to the body during the pull method for hip abduction (Figure 2B) and by tissue deformation during the push method for hip extension (Figure 2C), both of which may have led to an underestimation of the actual torque. While a standard testing procedure was attempted across all three methods, clinicians should note this potential discrepancy, especially if using a similar setup in clinical practice. Clinicians should recognize that placing the PD on the hamstrings during hip extension testing could dampen the RTD due to tissue deformation before the force is transmitted to the dynamometer. Due to the PD’s 10 N force threshold, participants were instructed to apply a minimum pre-tension before the assessment while maintaining a relaxed state as much as possible. This lack of standardization across participants highlights the need for further studies to explore how the torque level before the trial influences the strength and RTD values. Also, factors including different co-contraction and counteract movement patterns across participants may have influenced the peak torque or RTD generated,43 despite participants receiving the same instructions and undergoing practice trials prior to the measurements. For example, despite the use of stabilization straps, some participants attempted to erect their trunks during hip extension testing, which may be a movement pattern to perform hip extension contractions effectively. Lastly, performing multiple muscle contractions within a single day may have caused fatigue, potentially influencing the results despite the randomization of the testing order.

CONCLUSION

The findings of this study suggest that the PD assessments were valid for evaluating hip abduction and extension strength and RTD in healthy, physically active individuals. The PDpull method demonstrated strong validity for specific measures, including hip abduction RTD0-100_ms and hip extension strength and RTD100-200_ms, whereas the PDpush method showed moderate validity. However, caution is warranted when interpreting the PD measurements, as they may be less suited for absolute strength and RTD assessments due to potential systematic bias compared to IKD.

Conflicts of Interest

The authors report no conflicts of interest.

Supplementary Material

Supplementary Material 1

References

  1. Kinesiology of the hip: a focus on muscular actions. Neumann D. A. 2010J Orthop Sports Phys Ther. 40(2):82–94. doi: 10.2519/jospt.2010.3025. https://doi.org/10.2519/jospt.2010.3025 [DOI] [PubMed] [Google Scholar]
  2. The architectural design of the gluteal muscle group: implications for movement and rehabilitation. Ward S.R., Winters T.M., Blemker S.S. 2010J Orthop Sports Phys Ther. 40(2):95–102. doi: 10.2519/jospt.2010.3302. https://doi.org/10.2519/jospt.2010.3302 [DOI] [PubMed] [Google Scholar]
  3. Biomechanical measures of neuromuscular control and valgus loading of the knee predict anterior cruciate ligament injury risk in female athletes: a prospective study. Hewett T.E., Myer G.D., Ford K.R.., et al. 2005Am J Sports Med. 33(4):492–501. doi: 10.1177/0363546504269591. https://doi.org/10.1177/0363546504269591 [DOI] [PubMed] [Google Scholar]
  4. Anterior cruciate ligament injuries in female athletes: Part 1, mechanisms and risk factors. Hewett T. E., Myer G. D., Ford K. R. 2006Am J Sports Med. 34(2):299–311. doi: 10.1177/0363546505284183. https://doi.org/10.1177/0363546505284183 [DOI] [PubMed] [Google Scholar]
  5. Differences in hip kinematics, muscle strength, and muscle activation between subjects with and without patellofemoral pain. Souza R. B., Powers C. M. 2009J Orthop Sports Phys Ther. 39(1):12–19. doi: 10.2519/jospt.2009.2885. https://doi.org/10.2519/jospt.2009.2885 [DOI] [PubMed] [Google Scholar]
  6. Hip strength and hip and knee kinematics during stair descent in females with and without patellofemoral pain syndrome. Bolgla L. A., Malone T. R., Umberger B. R., Uhl T. L. 2008J Orthop Sports Phys Ther. 38(1):12–18. doi: 10.2519/jospt.2008.2462. https://doi.org/10.2519/jospt.2008.2462 [DOI] [PubMed] [Google Scholar]
  7. Is chronic ankle instability associated with impaired muscle strength? Ankle, knee and hip muscle strength in individuals with chronic ankle instability: a systematic review with meta-analysis. Khalaj N., Vicenzino B., Heales L. J., Smith M. D. 2020Br J Sports Med. 54(14):839–847. doi: 10.1136/bjsports-2018-100070. https://doi.org/10.1136/bjsports-2018-100070 [DOI] [PubMed] [Google Scholar]
  8. Hip strength and star excursion balance test deficits of patients with chronic ankle instability. McCann R. S., Crossett I. D., Terada M., Kosik K. B., Bolding B. A., Gribble P. A. 2017J Sci Med Sport. 20(11):992–996. doi: 10.1016/j.jsams.2017.05.005. https://doi.org/10.1016/j.jsams.2017.05.005 [DOI] [PubMed] [Google Scholar]
  9. Can clinicians trust objective measures of hip muscle strength from portable dynamometers? A systematic review with meta-analysis and evidence gap map of 107 studies of reliability and criterion validity using the COSMIN methodology. Waiteman M. C., Garcia M. C., Briani R. V.., et al. 2023J Orthop Sports Phys Ther. 53(11):655–672. doi: 10.2519/jospt.2023.12045. https://doi.org/10.2519/jospt.2023.12045 [DOI] [PubMed] [Google Scholar]
  10. Is hip strength a risk factor for patellofemoral pain? A systematic review and meta-analysis. Rathleff M. S., Rathleff C., Crossley K., Barton C. 2014Br J Sports Med. 48(14):1088. doi: 10.1136/bjsports-2013-093305. https://doi.org/10.1136/bjsports-2013-093305 [DOI] [PubMed] [Google Scholar]
  11. Hand-held dynamometry correlation with the gold standard isokinetic dynamometry: a systematic review. Stark T., Walker B., Phillips J.K., Fejer R., Beck R. 2011PM R. 3(5):472–479. doi: 10.1016/j.pmrj.2010.10.025. https://doi.org/10.1016/j.pmrj.2010.10.025 [DOI] [PubMed] [Google Scholar]
  12. Intra-and inter-rater reliability of strength measurements using a pull hand-held dynamometer fixed to the examiner’s body and comparison with push dynamometry. González-Rosalén J., Benítez-Martínez J.C., Medina-Mirapeix F., Pino C.-D., Cervelló A., Agustín M.-S. 2021Diagnostics. 11(7):1230. doi: 10.3390/diagnostics11071230. https://doi.org/10.3390/diagnostics11071230 [DOI] [PMC free article] [PubMed] [Google Scholar]
  13. Rate of force development: physiological and methodological considerations. Maffiuletti N. A., Aagaard P., Blazevich A. J., Folland J., Tillin N., Duchateau J. 2016Eur J Appl Physiol. 116(6):1091–1116. doi: 10.1007/s00421-016-3346-6. https://doi.org/10.1007/s00421-016-3346-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
  14. The influence of quadriceps strength and rate of torque development on the recovery of knee biomechanics during running after anterior cruciate ligament reconstruction. Knurr K. A., Cobian D. G., Kliethermes S. A., Stiffler-Joachim M. R., Heiderscheit B. C. 2023Am J Sports Med. 51(12):3171–3178. doi: 10.1177/03635465231194617. https://doi.org/10.1177/03635465231194617 [DOI] [PMC free article] [PubMed] [Google Scholar]
  15. Strength Training Rehabilitation Incorporating Power Exercises (STRIPE) for individuals with patellofemoral pain: a randomised controlled trial protocol. Glaviano N. R., Mangum L. C., Bazett-Jones D. M., DiStefano L. J., Toland M. D., Boling M. 2023BMJ Open Sport Exerc Med. 9(1):e001482. doi: 10.1136/bmjsem-2022-001482. https://doi.org/10.1136/bmjsem-2022-001482 [DOI] [PMC free article] [PubMed] [Google Scholar]
  16. Hip rate of torque development, but not strength, is associated with single-leg squat kinematics in individuals with patellofemoral pain. Jordan C. M., Kim S., Glaviano N. R. 2024Physiother Theory Pract. doi: 10.1080/09593985.2024.2360578. https://doi.org/10.1080/09593985.2024.2360578 [DOI] [PubMed]
  17. Influence of severity and duration of anterior knee pain on quadriceps function and self-reported function. Kim S., Park J. 2022J Athl Train. 57(8):771–779. doi: 10.4085/1062-6050-0647.21. https://doi.org/10.4085/1062-6050-0647.21 [DOI] [PMC free article] [PubMed] [Google Scholar]
  18. Sex differences in knee extensor neuromuscular function in individuals with and without patellofemoral pain. Kim S., Glaviano N. R., Park J. 2023Sports Health. doi: 10.1177/19417381231209318. https://doi.org/10.1177/19417381231209318 [DOI] [PMC free article] [PubMed]
  19. Increased rate of force development and neural drive of human skeletal muscle following resistance training. Aagaard P., Simonsen E. B., Andersen J. L., Magnusson P., Dyhre-Poulsen P. 2002J Appl Physiol. 93(4):1318–1326. doi: 10.1152/japplphysiol.00283.2002. https://doi.org/10.1152/japplphysiol.00283.2002 [DOI] [PubMed] [Google Scholar]
  20. Early and late rate of force development: differential adaptive responses to resistance training? Andersen L. L., Andersen J. L., Zebis M. K., Aagaard P. 2010Scand J Med Sci Sports. 20(1):e162–169. doi: 10.1111/j.1600-0838.2009.00933.x. https://doi.org/10.1111/j.1600-0838.2009.00933.x [DOI] [PubMed] [Google Scholar]
  21. Are early and late rate of force development differently influenced by fast-velocity resistance training? de Oliveira F. B., Rizatto G. F., Denadai B. S. 2013Clin Physiol Funct Imaging. 33(4):282–287. doi: 10.1111/cpf.12025. https://doi.org/10.1111/cpf.12025 [DOI] [PubMed] [Google Scholar]
  22. Effects of fast-velocity eccentric resistance training on early and late rate of force development. Oliveira A. S., Corvino R. B., Caputo F., Aagaard P., Denadai B. S. 2016Eur J Sport Sci. 16(2):199–205. doi: 10.1080/17461391.2015.1010593. https://doi.org/10.1080/17461391.2015.1010593 [DOI] [PubMed] [Google Scholar]
  23. Hip strength assessment using handheld dynamometry is subject to intertester bias when testers are of different sex and strength. Thorborg K., Bandholm T., Schick M., Jensen J., Hölmich P. 2013Scand J Med Sci Sports. 23(4):487–493. doi: 10.1111/j.1600-0838.2011.01405.x. https://doi.org/10.1111/j.1600-0838.2011.01405.x [DOI] [PubMed] [Google Scholar]
  24. Isometric knee extension force measured using a handheld dynamometer with and without belt-stabilization. Bohannon R. W., Kindig J., Sabo G., Duni A. E., Cram P. 2012Physiother Theory Pract. 28(7):562–568. doi: 10.3109/09593985.2011.640385. https://doi.org/10.3109/09593985.2011.640385 [DOI] [PubMed] [Google Scholar]
  25. Hand-held dynamometer positioning impacts discomfort during quadriceps strength testing: a validity and reliability study. Hansen E. M., McCartney C. N., Sweeney R. S., Palimenio M. R., Grindstaff T. L. 2015Int J Sports Phys Ther. 10(1):62–68. [PMC free article] [PubMed] [Google Scholar]
  26. Handheld dynamometry: validity and reliability of measuring hip joint rate of torque development and peak torque. McNabb K., Sánchez M. B., Selfe J., Reeves N. D., Callaghan M. 2024PLoS One. 19(8):e0308956. doi: 10.1371/journal.pone.0308956. https://doi.org/10.1371/journal.pone.0308956 [DOI] [PMC free article] [PubMed] [Google Scholar]
  27. Reliability and validity of belt-stabilized and tension dynamometry for assessing hip strength and power in uninjured adults. Batista N. P., Rachid Z. I., Silva D. D. O., Glaviano N. R., Norte G. E., Bazett-Jones D. M. 2024Phys Ther Sport. 69:59–66. doi: 10.1016/j.ptsp.2024.07.004. https://doi.org/10.1016/j.ptsp.2024.07.004 [DOI] [PubMed] [Google Scholar]
  28. Assessment of lower limb muscle strength and power using hand-held and fixed dynamometry: a reliability and validity study. Mentiplay B. F., Perraton L. G., Bower K. J.., et al. 2015PLoS One. 10(10):e0140822. doi: 10.1371/journal.pone.0140822. https://doi.org/10.1371/journal.pone.0140822 [DOI] [PMC free article] [PubMed] [Google Scholar]
  29. Reliability and validity of the belt-stabilized handheld dynamometer in hip-and knee-strength tests. Martins J., Da Silva J. R., Da Silva M. R. B., Bevilaqua-Grossi D. 2017J Athl Train. 52(9):809–819. doi: 10.4085/1062-6050-52.6.04. https://doi.org/10.4085/1062-6050-52.6.04 [DOI] [PMC free article] [PubMed] [Google Scholar]
  30. Clinimetric properties of hip abduction strength measurements obtained using a handheld dynamometer in individuals with a lower extremity amputation. Leijendekkers R. A., Hinte Gv., Sman A. D., Staal J. B., Nijhuis-Van Der Sanden M. W., Hoogeboom T. J. 2017PLoS One. 12(6):e0179887. doi: 10.1371/journal.pone.0179887. https://doi.org/10.1371/journal.pone.0179887 [DOI] [PMC free article] [PubMed] [Google Scholar]
  31. STARD 2015 guidelines for reporting diagnostic accuracy studies: explanation and elaboration. Cohen J. F., Korevaar D. A., Altman D. G.., et al. 2016BMJ Open. 6(11):e012799. doi: 10.1136/bmjopen-2016-012799. https://doi.org/10.1136/bmjopen-2016-012799 [DOI] [PMC free article] [PubMed] [Google Scholar]
  32. Validity of hand-held dynamometry in measuring quadriceps strength and rate of torque development. Lesnak J., Anderson D., Farmer B., Katsavelis D., Grindstaff T.L. 2019Int J Sports Phys Ther. 14(2):180–187. doi: 10.26603/ijspt20190180. [DOI] [PMC free article] [PubMed] [Google Scholar]
  33. The reliability, validity, and responsiveness of the Lysholm score and Tegner activity scale for anterior cruciate ligament injuries of the knee: 25 years later. Briggs K. K., Lysholm J., Tegner Y., Rodkey W. G., Kocher M. S., Steadman J. R. 2009Am J Sports Med. 37(5):890–897. doi: 10.1177/0363546508330143. https://doi.org/10.1177/0363546508330143 [DOI] [PubMed] [Google Scholar]
  34. Concentric and eccentric torque of the hip musculature in individuals with and without patellofemoral pain. Boling M. C., Padua D. A., Alexander Creighton R. 2009J Athl Train. 44(1):7–13. doi: 10.4085/1062-6050-44.1.7. https://doi.org/10.4085/1062-6050-44.1.7 [DOI] [PMC free article] [PubMed] [Google Scholar]
  35. Quadriceps function and gait kinetics after anterior cruciate ligament reconstruction. Blackburn J. T., Pietrosimone B., Harkey M. S., Luc B. A., Pamukoff D. N. 2016Med Sci Sports Exerc. 48(9):1664–1670. doi: 10.1249/MSS.0000000000000963. https://doi.org/10.1249/MSS.0000000000000963 [DOI] [PubMed] [Google Scholar]
  36. Knee joint and quadriceps dysfunction in individuals with anterior knee pain, anterior cruciate ligament reconstruction, and meniscus surgery: a cross-sectional study. Kim S., Kim D., Park J. 2021J Sport Rehabil. 30(1):112–119. doi: 10.1123/jsr.2018-0482. https://doi.org/10.1123/jsr.2018-0482 [DOI] [PubMed] [Google Scholar]
  37. Cohen J. Statistical Power Analysis for the Behavioral Sciences. Routledge; New York: [Google Scholar]
  38. Measurement properties of hip strength measured by handheld dynamometry: reliability and validity across the range of motion. Bazett-Jones D. M., Squier K. 2020Phys Ther Sport. 42:100–106. doi: 10.1016/j.ptsp.2020.01.005. https://doi.org/10.1016/j.ptsp.2020.01.005 [DOI] [PubMed] [Google Scholar]
  39. Validity and reliability of handheld dynametric strength assessment of hip extensor and abductor muscles. Kawaguchi J.K., Babcock G. 2010Athl Train Sports Health Care. 2(1):11–17. doi: 10.3928/19425864-20101221-04. https://doi.org/10.3928/19425864-20101221-04 [DOI] [Google Scholar]
  40. Relative validity and reliability of isometric lower extremity strength assessment in older adults by using a handheld dynamometer. Grootswagers P., Vaes A. M., Hangelbroek R., Tieland M., van Loon L. J., de Groot L. C. 2022Sports Health. 14(6):899–905. doi: 10.1177/19417381211063847. https://doi.org/10.1177/19417381211063847 [DOI] [PMC free article] [PubMed] [Google Scholar]
  41. Rate of torque development is the primary contributor to quadriceps avoidance gait following total knee arthroplasty. Kline P. W., Jacobs C. A., Duncan S. T., Noehren B. 2019Gait Posture. 68:397–402. doi: 10.1016/j.gaitpost.2018.12.019. https://doi.org/10.1016/j.gaitpost.2018.12.019 [DOI] [PMC free article] [PubMed] [Google Scholar]
  42. Quadriceps neuromuscular function during and after exercise-induced fatigue in patients with patellofemoral pain. Kim S., Roh Y., Glaviano N. R., Park J. 2023J Athl Train. 58(6):554–562. doi: 10.4085/1062-6050-0348.22. https://doi.org/10.4085/1062-6050-0348.22 [DOI] [PMC free article] [PubMed] [Google Scholar]
  43. Effect of short-term maximal voluntary co-contraction training on neuromuscular function. Maeo S., Yoshitake Y., Takai Y., Fukunaga T., Kanehisa H. 2014Int J Sports Med. 35(2):125–134. doi: 10.1055/s-0033-1349137. [DOI] [PubMed] [Google Scholar]