Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2026 Feb 18.
Published in final edited form as: Anal Chem. 2025 Feb 4;97(6):3387–3394. doi: 10.1021/acs.analchem.4c05372

Continuous Delivery of Hyperpolarized Xenon-129 Gas Using a “Stopped-Flow” Clinical-Scale Cryogen-Free Hyperpolarizer

Md Raduanul H Chowdhury a,*, Clementinah Oladun a, Firoz Ahmed a, Nuwandi M Ariyasingha a, Abubakar Abdurraheem a, Faisal Asif a, Joseph Gyesi a, Panayiotis Nikolaou b, Michael J Barlow c, Anton Shcherbakov b,d, Nathan A Rudman e, Ivan J Dmochowski e, Boyd M Goodson f, Eduard Y Chekmenev a,*
PMCID: PMC11969679  NIHMSID: NIHMS2066094  PMID: 39903267

Abstract

In 2022, the FDA approved hyperpolarized (HP) 129Xe gas as an inhalable contrast agent for functional lung imaging. For clinical imaging, HP 129Xe is usually given as a bolus inhalation. However, for preclinical applications (e.g., pulmonary imaging in small rodents), continuous delivery of HP 129Xe is greatly desired to enable MRI scanning under conditions of physiological continuous animal breathing patterns. Moreover, HP 129Xe gas can be utilized for other applications including materials science and bioanalytical chemistry, where continuous flow of hyperpolarized gas through an NMR sample over several minutes is also desired for sensing of 129Xe inside an NMR spectrometer. 129Xe is often hyperpolarized using continuous-flow spin exchange optical pumping, which employs a lean (1-2%) mixture of Xe and a carrier gas (e.g., He and N2). The low Xe concentration in the produced output reduces the NMR detection sensitivity, and thus Xe cryo-collection is typically employed to achieve near-100% pure gas-phase Xe before administration to the sample or subject. However, the need for cryo-collection undermines a key advantage of continuous-flow production, i.e., the continuously flowing in a hyperpolarizer HP 129Xe gas is trapped inside the hyperpolarizer, and the produced HP 129Xe gas is released at once when the production cycle (30-60 mins) is completed. An alternative HP 129Xe production technology employs a “stopped-flow” approach, where a batch of HP gas is hyperpolarized over time and quickly released from a hyperpolarizer. Here, a clinical-scale “stopped-flow” 129Xe hyperpolarizer was employed to hyperpolarize a 1.3 liter-atm batch of 50:50 Xe:N2 gas mixture inside a glass cell with an ultra-long lifetime of the HP 129Xe state (T1 > 2 hours). The produced HP 129Xe gas was slowly delivered into a 5-mm NMR tube via PEEK tubing under a wide range of gas flow rates: 3-180 standard cubic centimeters per minute (sccm). The polarization of the gas ejected from the hyperpolarizer was quantified using in situ low-field NMR polarimetry and additionally verified using a 0.35 T clinical MRI scanner. Continuous-flow delivery of HP 129Xe was demonstrated for up to 15 minutes with a gas flow rate of 45-150 sccm over a 2.5-meter length of PEEK tubing, suffering only small losses in 129Xe polarization. These observations are additionally supported by 129Xe relaxation measurements inside the PEEK tubing employed for gas delivery and the 5-mm NMR tube employed for polarimetry. 129Xe polarization of 16-19% was obtained in the delivered gas, starting with an “in-polarizer” 129Xe polarization of 19%. We envision that this method can be employed for on-demand cryogen-free delivery of hyperpolarized gas using “stopped-flow” 129Xe hyperpolarizers for a broad range of applications, from pre-clinical imaging, to biosensors, to spectroscopy of materials surfaces.

Graphical Abstract

graphic file with name nihms-2066094-f0001.jpg

INTRODUCTION

Nuclear spin polarization (P), or the degree of nuclear spin alignment with applied magnetic field, can be significantly enhanced beyond its thermal equilibrium level using NMR hyperpolarization.1-2 Since the NMR signal is directly proportional to P,3 the orders-of-magnitude polarization gain obtained through hyperpolarization results in correspondingly substantial gains in NMR and MRI detection sensitivity.1, 4-5

Hyperpolarized (HP) NMR-active noble gases such as 3He, 83Kr, and 129Xe can be utilized as inhalable MRI contrast agents to assess lung function,6-13 probe the metabolic status of brown fat,14 image the brain,15 and perform hyperpolarized Chemical Exchange Saturation Transfer (hyper-CEST)16-21 for sensing22 and imaging the interactions of receptors and drugs.23-25

HP 129Xe gas was recently approved by the FDA as a safe contrast agent for lung imaging in December 2022, making it the leading HP contrast agent of any kind, and certainly the key inhalable HP gas contrast agent.26-30 In this new functional imaging modality, the HP 129Xe gas is typically inhaled to enable in vivo administration30; the corresponding MRI scan is thus recorded after inhalation—typically on a single patient breath hold. Another clinical MRI scanning approach is to perform MR image acquisition during multiple subsequent inhalations of HP 129Xe gas (albeit with lower volumes of HP 129Xe gas administered during each inhalation). This approach offers a number of advantages including providing more detailed gas dynamics information and lower Xe dose per individual inhalation; moreover, free-breathing administration facilitates subject (or patient) participation and is more natural—potentially providing a better reflection of physiological conditions. However, unlike a single-dose protocol, multi-breath protocols require continuous delivery of HP 129Xe gas to a patient or subject. Finally, the development of biosensors enabled by HP 129Xe also benefit substantially from continuous delivery of HP 129Xe, as hyper-CEST16-21 experiments typically require multiple repetitions with signal saturation over a range of frequencies in a pseudo-2D acquisition mode.25 Beyond the realm of medical diagnostics, HP 129Xe has the potential to be used in various materials science applications such as porous materials characterization, surface chemistry analysis, polymer membrane studies, magnetometry,31 etc.32-37 These applications also benefit substantially by having a continuous supply of highly concentrated and highly polarized HP 129Xe gas.

HP 129Xe is typically produced via spin-exchange optical pumping (SEOP).38 The SEOP process consists of two steps: Firstly, laser photons of the required wavelength (for example, the D1 transition wavelength of rubidium at ~794.8 nm) supplied by a circularly polarized laser are absorbed by the outer-shell electron of an alkali metal (rubidium in this study), causing a highly spin-polarized electronic ground state of Rb to be achieved over time.39 This electron spin angular momentum is then subsequently transferred to the 129Xe nuclei through gas-phase collisions of Rb and 129Xe via Fermi contact interactions.40 Typically, the production of HP 129Xe using the SEOP process is performed either in a continuous-flow41-43 mode or in a stopped-flow (a.k.a. batch-) mode.29, 44-47 A recirculating variant of the continuous-flow method has also been reported for materials science applications.48

As the name may imply, the continuous-flow SEOP methodology should be ideally suited for continuous delivery of produced HP 129Xe gas exiting the hyperpolarizer device:49 Indeed, the flow rate can be readily adjusted for the desired applications.50 However, this method usually utilizes a lean (ca. 2% Xe and the rest is carrier) gas mixture as a hyperpolarizer feedstock; richer Xe mixtures are possible for some applications51 but generally result in much lower Xe polarization levels. Because NMR signal is directly proportional to the concentration or partial pressure of the HP Xe analyte,3 and the corresponding goal of maximizing the signal-to-noise ratio (SNR) of the HP 129Xe analyte, it usually becomes necessary to increase the concentration of the produced HP 129Xe via cryo-collection49 (while other workarounds have been shown, they have their own additional disadvantages52-53). During cryo-collection, the produced 129Xe is cryogenically trapped, and at completion, the frozen HP 129Xe is rapidly thawed to produce a bolus of pure HP Xe gas.54 The cryo-collection requires the use of liquid N2, which complicates the production process and defies a key intrinsic advantage of continuous-flow production technology as no HP Xe gas exits the hyperpolarizer during the production process. Moreover, once the HP gas is cryogenically trapped, there is no practical way to cancel the experiment—thus requiring careful planning of such studies and leaving virtually no room for error. Of note, 129Xe is expensive ($200-300 per liter*atm in isotopically enriched form and $20-$200 per liter*atm in the non-enriched 26%-natural abundance form).

Batch-mode SEOP production of HP 129Xe usually operates in the Xe-rich regime: indeed, since PXe over 50% has been demonstrated on a clinical scale44, 55-56 in gas mixtures with 50% or greater Xe content, the cryo-collection of HP Xe can be obviated (presenting a key advantage of this method),11, 57 therefore making batch-mode SEOP process completely cryogen-free.11-12, 55, 58-64 However, this method historically was employed to produce a bolus of HP Xe gas, which is typically ejected into a storage Tedlar bag as a single bolus. Although the ejected gas dosing can be adjusted (to tailor the needs to the desired experimental goal), the produced bolus is poorly suited for continuous-HP-Xe delivery experiments. Due to this gap in the field, a number of continuous delivery devices have been designed to utilize a bolus of HP 129Xe gas (as a source) with the goal of performing continuous delivery of HP gas for the desired application.58, 65 While these designs have their merits, they inevitably add additional experimental complexity and cost, and require custom hardware.

In this study, we explored the utility of continuous delivery of HP gas from a hyperpolarizer otherwise operating in a “batch mode” of 129Xe gas hyperpolarization. Specifically, the produced HP gas is stored inside the hyperpolarizer production glassware. This stored HP gas is released continuously for the delivery of HP 129Xe to a 5-mm NMR tube with the flow rate ranging from 3 to 180 standard cubic centimeters per minute (sccm) for up to 15 min. This efficient process becomes possible due to (1) the long lifetime of the HP state in the SEOP cell employed for transient storage inside the hyperpolarizer (T1 ≅ 2 h), and (2) the short travel time of HP 129Xe through the microfluidic PEEK tubing, employed for the continuous gas delivery into standard 5-mm NMR tubes. The utility of the method has been verified by in situ low-field polarimetry performed on the SEOP cell and ex situ polarimetry performed using the 5-mm NMR tube, where the HP 129Xe was delivered from the SEOP cell. The PXe was additionally verified via ex situ polarimetry of the ejected HP 129Xe gas using a 0.35 T clinical MRI scanner. The obtained results are also supported by the HP 129Xe relaxation dynamics performed inside the transfer PEEK tubing, glass NMR tube, and Tedlar bag in low magnetic fields. The ejection process was performed using PXe-safe solenoid valves and a mass-flow controller that can be readily integrated in the automated workflow of high-resolution NMR experiments using transistor-transistor logic (TTL) control lines of the NMR spectrometer and MRI scanners (work in progress in our laboratory). The presented method provides a simple, cost-efficient approach for continuous delivery of HP 129Xe gas produced via the batch-mode SEOP technique.

EXPERIMENTAL SECTION

HP 129Xe gas production and in situ polarimetry.

HP 129Xe gas mixture was produced by using a batch-mode generation-3 (GEN-3) hyperpolarizer.66-67 Briefly, hyperpolarization was performed inside a Pyrex 0.5-L 2-inch-inner diameter SEOP cell at an optimized temperature of 75 °C. Once PXe buildup was completed (in ~30-60 min), an automated sequence was used to cool down the SEOP cell (in ~5 min) to eject the HP gas at a temperature below 42 °C (near the Rb metal melting point of 39.5 °C, where the Rb vapor pressure is very low). The SEOP cell was filled with a 2000 torr gas mixture: 50% Xe and 50% N2. Unfortunately, the LDA laser employed in the reported experiments was no longer within the vendor specifications: FWHM ~0.5 nm (versus 0.2 nm FWHM) and output power of <100 W (versus 170 W), resulting in reduced SEOP efficiency: indeed, PXe was limited to 30-40% upon completion of the buildup procedure (e.g., Figure S13a) versus PXe of over 50% that was routinely achieved previously with an LDA exhibiting normal performance.66-67 Following the ejection of HP gas, the SEOP cell was reloaded with unpolarized gas mixture for the next SEOP cycle.66-67 In situ HP 129Xe polarimetry (i.e., NMR spectroscopic P measurement of HP species) was performed as described previously (see SI for details).68

HP 129Xe gas delivery in 5-mm NMR tube and ex situ low-field NMR polarimetry.

Following the cooldown of the SEOP cell below 42 °C, the pressurized HP gas mixture (2000 torr, 50% Xe and 50% N2) was ejected from the hyperpolarizer via polyetheretherketone (PEEK) tubing line (1/16-in. OD, 0.040-in ID, Idex, P/N 1538XL). The ejection of the HP gas from the hyperpolarizer was done over the course of 15 min of gas delivery into 5-mm NMR tube that was situated inside the polarimetry station operating at ~3.5 mT.68 HP 129Xe polarimetry in the polarimetry station was performed using Eq. S1, and the corresponding-size reference sample of thermally polarized water doped with CuSO4 (SI).

The ejected HP gas from the polarizer flows to the NMR tube through PEEK tubing, and then passes through a one-way check valve (to prevent any gas backflow that could result in SEOP cell poisoning), and through a 1/16-in. OD (1/32-in. ID, 1 ft long) Teflon catheter that guides the incoming HP Xe gas to the bottom of the NMR tube. A non-magnetic push-to-connect “T” shaped reducer allows mating (i) the 5-mm NMR tube (via 1/4-in. end), (ii) the incoming HP Xe gas (via 1/8-in. input port), and (iii) exhausting Xe gas mixture (via 1/8-in. exit port). The 1/8-in. terminals of the push-connect adapter use 1/8-in. Teflon tubing lines. The NMR tube is jacketed with 1/4-in. OD Teflon tubing, with the overall approach that follows closely to our previous designs.69-70 The exit line is then connected to a pressure gauge and to a needle valve (for variable-flow delivery). The exhaust of the setup is open to atmosphere, and it was employed for real-time flowrate measurement of the exiting gas using water displacement in a graduated cylinder, Figure 1.

Figure 1.

Figure 1.

a) Schematic diagram of continuous HP 129Xe gas delivery from the hyperpolarizer to a 5-mm NMR tube located inside the polarimetry station using a needle valve. PXe sensing was performed in-situ within the 129Xe polarizer and in the 5-mm NMR tube throughout the delivery process. The HP gas flow control is enabled either by a needle valve or by a MFC equipped with additional solenoid valves and a coupler.

The ~8-cm-long 5-mm NMR tube (washed extensively with methanol) was employed for collecting and sensing of HP 129Xe gas. This NMR tube resided inside a homogeneous B0 electromagnet of the 3.5 mT polarimeter.68 The NMR tube was placed next to a surface RF coil that was built similarly to the RF coil for in situ polarimetry, except that it was tuned to 41.9 kHz compared to 40.8 kHz in the hyperpolarizer, Figure 1. This coil was connected to a second XeUS NMR spectrometer68 used for ex situ polarimetry on the NMR tube.

The HP 129Xe gas ejection was realized in two ways: variable-flow rate release using a needle valve and a continuous-flow rate using a mass flow controller (MFC) and the additional valve and a coupler shown in Figure 1. For variable rate ejection, before the hyperpolarizer enters the 15-minute-long phase of gas ejection, the two-way valve shown in Figure 1 was closed, and three purge cycles of vacuum and N2 gas application on the hyperpolarizer effectively purged the gas manifold (all the way to the two-way valve) of residual trapped air. When the polarizer started the HP gas ejection, this two-way manual valve was opened, thus, enabling the continuous delivery of HP gas over the 15-min period. When using the needle valve, the gas flow rate is controlled by a pressure gradient, and the flow rate changed from ~3 standard cubic centimeter per second (sccs) at the start of the gas delivery (the highest pressure gradient) down to 0.05 sccs at the end of the delivery period (the smallest pressure gradient between the SEOP cell and the ambient pressure). The following data was sampled every 20 s during the 15-min delivery process: PXe in the SEOP cell (e.g., Figure 2a) and in the 5 mm NMR tube (e.g., Figure 2b), gas pressure, and gas flow rate (by monitoring the amount of the total delivered gas), Table S9.

Figure 2.

Figure 2.

a) In-situ NMR polarimetry immediately before the gas delivery to 5-mm NMR tube. b) Ex-situ NMR polarimetry of the ejected HP 129Xe gas inside the NMR tube at 41.9 kHz. c) T1 decay of stopped HP 129Xe gas in a 5 mm NMR tube. g) T1 decay of HP 129Xe gas inside PEEK tubing. e) PXe, flow rate (sccs), and gas pressure (P, atm) during the process of HP gas delivery.

After closing the two-way valve of the delivery setup and the eject valve inside the hyperpolarizer, the gas flow was effectively stopped. Acquisition of NMR signals from stopped HP gas allowed us to measure the HP 129Xe T1 decay, primarily due to relaxation via collision with the glass walls of the NMR tube, Figure 2c. In a separate experiment, the setup was shifted by placing a section of PEEK tubing in front of the RF coil (instead of the NMR tube), allowing us to measure 129Xe relaxation inside the PEEK tubing, Figure 2d. Finally, after completion of the HP gas delivery, the SEOP cell was reloaded with Xe:N2 gas mixture, repolarized, and T1 was measured again; this measurement was performed to confirm that no Rb metal poisoning occurred during the gas delivery process (which typically manifests itself as a substantial reduction in 129Xe T1 inside the SEOP cell). A series of T1 data obtained ‘before’ and ‘after’ multiple HP gas ejections is presented in Figure S8.

In a separate series of experiments shown in Figure 3, the experimental setup of Figure 1 was modified as shown in Figures S11 and S12 via two important changes: First, two solenoid valves (Burkert, type 6126) were added before and one after the NMR tube (i.e., both valves denoted by ⊗ were realized by the solenoid valves). An additional PEEK coupler was also added to connect the inlet 1/16” PEEK tube to the 1/16” Teflon catheter: as a result, the total length of PEEK path was increased to 8 ft (versus 6 ft used in the needle-valve configuration described above). Second, an MFC (Sierra Instruments, Monterey, CA, P/N C100L-DD-OV1-SV1-PV2-V1-S0-C0) replaced the needle valve, allowing the setup to maintain a constant HP gas mixture flow during gas delivery from the hyperpolarizer to the NMR polarimetry station.

Figure 3.

Figure 3.

Graphs showing time-dependent plots of gas-phase PXe, flow rate (sccs), and gas pressure (P, atm) during the process of HP gas delivery using continuous HP 129Xe gas-mixture flow rate settings of 150 sccm (a), 90 sccm (b), 60 sccm (c); 1 sccm = 60 sccs.

In a separate experiment—the results of which are presented in the SI—the MFC-based setup was employed (as described above), but ~0.4 mL of HPLC-grade methanol was loaded in the NMR tube, thus allowing HP 129Xe signal to be detected from the dissolved phase. Note that PXe quantification in the dissolved phase was not performed because the 129Xe concentration was not independently determined.

HP 129Xe Tedlar bag delivery and ex situ polarimetry.

The PXe level measured inside our stopped-flow hyperpolarizers was verified by both low-field polarimetry and polarimetry inside an MRI scanner, providing additional validation for the PXe level measured in situ, and 129Xe T1 measurements in Tedlar bag at 3.5 mT and 0.35 T (see SI for details).

3. RESULTS AND DISCUSSION

Continuous delivery of HP 129Xe using variable gas flow.

The efficiency of batch-mode hyperpolarizers relies heavily on the relatively long in-cell 129Xe T1 (≥2 h).57, 71-73 The key SEOP cell degradation pathway is the oxidation of Rb metal by residual oxygen and air moisture that may enter the cell in trace quantities, typically during the gas handling process. Following Rb oxidation, 129Xe in-cell T1 is reduced substantially (to less than 30 min), resulting in reduced PXe. Therefore, the 129Xe T1 provides a “good SEOP-cell health” indicator in batch-mode 129Xe SEOP hyperpolarization.66, 74 Thus, before proceeding with systematic studies, we performed a series of HP 129Xe gas ejections from the SEOP cell under conditions of continuous delivery using the setup shown in Figure 1, with the overall rationale to ensure that the SEOP cell is not being poisoned through this new gas delivery process. Indeed, Figure S14b shows that no significant change in T1 was seen before versus after the HP gas continuous delivery: the T1 values were 146 ± 20 and 144 ± 9 min, respectively (a more rigorous testing (Figure S8) showed no substantial T1 changes after the 11th HP gas ejection from the hyperpolarizer, additionally confirming that this HP gas ejection procedure is safe for the Rb contained in the SEOP cell).

Encouraged by these pilot findings, we proceeded with systematic HP production studies using needle valve control of the flow rate, Figures 1 and S14. For each production cycle, PXe build-up was established for 1 hour at 75 °C on a freshly reloaded SEOP cell (with 2000 Torr Xe:N2 50:50 gas mixture), Figure S14c: in one illustrative example, PXe of 31±2% was obtained via the optimized process, with a buildup time constant Tb of 24 ± 3 min—corresponding to a buildup rate γSEOP of 0.041 ± 0.0006 min−1. Next, the SEOP cell was cooled down to 42 °C, and an in-situ PXe value was measured one last time immediately before the HP gas ejection from the SEOP cell: Figure 2a illustrates a representative in-situ NMR spectrum, confirming PXe of 26% (a minor reduction of PXe during the SEOP cell cool-down is attributed to T1 relaxation decay of the HP state75). Following the HP 129Xe gas delivery into the NMR tube located inside the nearby polarimeter, the NMR signal recorded at 41.9 kHz (Figure 2b), revealing PXe of 22%. The minor decrease of PXe (from 26% measured inside the hyperpolarizer) is rationalized via an accelerated T1 relaxation due to HP 129Xe collision with the walls of the NMR tube and PEEK tubing that Xe gas experiences during the flow process: T1 = 5.3 ± 0.9 min (Figure 2c) and 5.5 ± 3.3 min (Figure 2d), respectively. The substantial decrease in T1 is rationalized by the substantial increase in the wall collision rate due to an increase in surface-to-volume ratio. These results are important as they provide guidance for the limits of HP 129Xe delivery through narrow-bore tubing: in order to retain the bulk of PXe, the travel time of HP gas must be substantially shorter than T1. This notion is semi-quantitively supported by the continuous HP 129Xe gas delivery through the NMR tube, Figure 1h, where the gas flow rate varied from ~3 sccs (at the beginning) down to 0.05 sccs (at the end of a 900-s-long ejection process). The confined volume of the glass and PEEK tubing (i.e., inside the SEOP cell stem, NMR tube path from the bottom to the detection region, and PEEK path) was estimated to be on the order of 10 cc. Therefore, the HP gas travel time at the beginning of the ejection process (P = 2.5 atm) was estimated to be 8 s, which is relatively inconsequential with respect to the T1 value of approximately 5 min. On the other hand, the HP gas travel time at the end of the ejection process (P = 1 atm) was estimated to be over 3 min, which can no longer be neglected, resulting in substantial PXe decrease at the end of the HP gas delivery procedure. The overall contribution of the in-cell T1 was small, as the 15-min delivery process resulted in PXe decreasing by a factor of 1.1. The HP gas delivery process was reproducible: indeed, Figure S9 exhibited overall the same trends.

Continuous delivery of HP 129Xe gas using mass-flow control (fixed HP gas flow rate).

Although the use of a needle valve allowed testing of the efficiency of HP gas delivery under a wide range of flow rates and pressures, a fixed flow rate may be desired where steady-state delivery is required, e.g., for biosensor applications.18-19, 32, 52-53, 76-77 Thus, we replaced the needle valve by an MFC, and also added two solenoid valves to gate the process of HP gas delivery, Figure 1 and Figure S16a. While the solenoid valves were not required for our experiments, they are greatly desired for the process of automated gas flow cessation during NMR acquisition to minimize susceptibility-induced B0 field gradients at high magnetic fields. The HP gas delivery experiments were performed at three fixed (as opposed to variable described in the above sub-section) flow rate values: 150 sccm (Figure 3a), 90 sccm (Figure 3b) and 60 sccm (Figure 3c)

A flow rate of 150 sccm (Figure 3a) resulted in the delivery of ~800 scc HP gas in ~320 s, after which the flow rate could no longer be controlled due to lack of back pressure in the SEOP cell. The initial PXe oscillation at the beginning of the ejection process is explained by initial oscillation in the MFC output. Following the flow rate stabilization, the PXe (18-19%) and the flow rate (150 sccm) remained steady throughout the process, until the HP gas overpressure in the cell was exhausted. Of note, PXe “paradoxically” exhibited no decrease between t = 100 s to t = 300 s, which is discussed in more detail below.

The experiments using a flow rate of 90 sccm (Figure 3b) resulted in an extended gas flow (up to 600 s) with consistent PXe of 16-17%, and also exhibited a small but “paradoxical” growth in PXe despite T1 relaxation in the SEOP cell. The reader is reminded that a small decrease in PXe (by a factor of 1.1) is anticipated during the 900-second-long ejection process due to the in-cell T1 of ~140 min, Figure S14b. Further reduction of the flow rate to 60 sccm (Figure 3c) enabled a stable flow to be maintained for 800 s, as expected. Moreover, the “paradoxical” trend of PXe “growth” (from 16% to 19%) between t = 0 s and t = 400 s was more pronounced in this dataset. We rationalize this counter-intuitive observation as possibly originating from two potential contributions. First, the PXe throughout the SEOP cell may not be uniform thus resulting in the change of the detected PXe (downstream in the NMR tube after the HP gas delivery) as the gas is released from the SEOP cell. Second, while the flow rate is fixed throughout the process of HP gas delivery (blue curve), the pressure was decreasing (green curve). The pressure decrease is important as it results in the effectively reduced gas transfer time, potentially resulting in reduced T1 relaxation losses during gas passage through the tubing. As a result of this counter-effect, the in-cell T1 decay would be mitigated or in some cases (Figure 3c) would be overpowered by the gains manifested in reduced gas travel time and reduced T1 losses during gas delivery—thereby resulting in a smoother trend of the detected PXe inside the 5 mm NMR tube versus initial estimates.

When the HP gas flow rate was reduced to 45 sccm (Figure S15), and the ejected gas was bubbled through an NMR tube filled with 0.4 mL methanol, a steady flow of HP Xe gas was maintained throughout the 900-s ejection period, revealing a nearly constant signal from the dissolved-phase HP Xe in methanol (Figure S15b). The PXe quantification was not possible because concentration of Xe in methanol was not measured.

All in all, the results reported in Figures 2-3 clearly demonstrated a remarkable utility of the stopped-flow hyperpolarizer to create a continuous delivery of a stream of HP Xe gas for up to 15 min, with small variation in PXe throughout the process. The key disadvantage of the reported approach is the requirement to produce a batch of HP gas prior to delivery. As a result, it may take a 30-60 min before the process can be initiated. The primary advantages of this new approach are: 1) cryogen-free operation; 2) a wide dynamic range of gas flow rates; 3) opportunities for automation and integration in the gated workflow of the NMR spectrometer or MRI scanner (e.g., using TTL control lines to synchronize gas ejection with pulse sequence events using solenoid valves and the MFC). Moreover, in case the user decides to abort the procedure or not use the produced HP Xe batch, there is little wasted Xe gas, as the bulk of the unused Xe gas is retained in the SEOP cell and can be re-used. The latter point is an important consideration because Xe is expensive: up to $300/liter.

Additional PXe validation via 0.35 T MRI.

We have performed additional studies to validate the obtained PXe levels using a 0.35 T MRI scanner, with the overall goal to enhance the scientific validity of the reported findings and brings this study one step closer to an anticipated real-world application. For these studies, the experimental setup was truncated with the exit path connected directly to the Tedlar bag via PEEK/Tygon line and check valve, Figure S16a. Following the collection of ~0.8 L of HP gas in a one-liter Tedlar bag, PXe assessment at 3.5 mT field of the polarimetry station and at 0.35 T field of MRI scanner were performed. In this experiment, in-situ PXe of 25% was measured in the SEOP cell of the hyperpolarizer (Figure S16b) immediately before HP gas delivery into a Tedlar bag placed inside the polarimetry station. Immediately following the ejection, PXe of 21% was recorded from the Tedlar bag at 41.9 kHz, Figure S16c. Following the HP gas transfer to the MRI scanner facility in the same Tedlar bag (total delivery time of ~12 mins), PXe was reduced to 19% as measured inside the 0.35 T MRI scanner, Figure S16d. The experimental details on the transfer process and the transfer device, including bill of materials (BOM) are provided in the SI.

Outlook.

A number of additional efforts could be employed to improve overall performance. For example, using glass wall sealing with SurfaSil for coating of the detection reservoir could improve the T1 relaxation within the NMR tube. Reduction of T1 transfer losses may also allow one to perform HP 129Xe gas delivery with even lower flow rates, without substantial PXe losses. The reported studies clearly indicate that HP 129Xe can be delivered continuously from a stopped-flow hyperpolarizer with steady PXe levels over significant periods of time, spanning 15 min. We anticipate the 0.8 L volume should be sufficient for gas delivery of up to 60 min, albeit at reduced flow rate and PXe values (although this could be augmented by using higher in-cell pressures).

CONCLUSION

In conclusion, we have developed a robust method for continuous delivery of HP 129Xe gas using an automated cryogen-free batch-mode clinical-scale hyperpolarizer. It was possible to deliver HP gas to a 5-mm NMR tube situated approximately 1.5 m from the hyperpolarizer (over ~2.5-meter total path length), and we envision the length can be substantially increased based on the semi-quantitative T1 relaxation model. PXe of up to 25% was demonstrated after the delivery, which can be further improved using a pristine high-power laser for more optimal SEOP performance, as has been previously demonstrated.56, 66, 74 The continuous gas delivery was demonstrated using a wide range of flow rates: 0.05-3 sccs (albeit with PXe reduction at slower rates), clearly demonstrating the versatility of this approach. Constant-flow durations of up to 15 min were shown using 45 sccm HP gas flow rate, which potentially can be further increased. The capability of controlling the continuously-delivered HP gas flow rate using a MFC with solenoid valve gating of the gas flow was also demonstrated, which is desired for automated multi-scan experiments using the control interface of an NMR spectrometer, MRI scanner, or a stand-alone controller. The obtained PXe values of the delivered HP gas were measured by low-field polarimetry at 3.5 mT and additionally confirmed by polarimetry using a 0.35 T clinical open MRI scanner. We also report pilot feasibility MRI experiments employing a 0.35 T scanner with a large-volume transmit RF coil that was integrated in the poles of the open MRI magnet, paving the way for streamlined studies, employing HP 129Xe gas contrast agents in rodents and in human volunteers. Taken together, the results reported here have the potential to expand the utility of the recently FDA-approved HP 129Xe gas contrast agent for a wide range of complementary pre-clinical and bioanalytical applications.

Supplementary Material

supporting information

ACKNOWLEDGMENT

This work was supported by DOD CDMRP under W81XWH-20-10576 and W81XWH-20-10578, NSF CHE-2404387 and CHE-2404388, and NHLBI R21HL168430-01A1 and F32 HL160108, and WSU Thomas C. Rumble University Graduate Fellowship (A.A. and C.O.). BMG acknowledges a Cottrell SEED grant from Research Corporation for Science Advancement. MJB acknowledges support from The School of Medicine UoN. I.J.D. and N.A.R. were supported by NIH R35-GM-131907. We thank Prof. Zach Cleveland for discussion regarding Tedlar bag sourcing, and Dr. Jonathan R. Birchall for helpful discussions and preliminary efforts. We also thank Dr. Michael Hennessy support with low-field MRI RF coil development and construction.

Footnotes

Supporting Information

Additional experimental details, materials and methods including photos, schematics, figures, tables, and bill of materials (PDF).

PN, BMG, and EYC declare a stake of ownership in XeUS Technologies, LTD. EYC serves on the Scientific Advisory Board (SAB) and declares a stake of ownership in Vizma Life Sciences.

REFERENCES

  • 1.Nikolaou P; Goodson BM; Chekmenev EY NMR Hyperpolarization Techniques for Biomedicine. Chem. Eur. J 2015, 21 (8), 3156–3166. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Eills J; Budker D; Cavagnero S; Chekmenev EY; Elliott SJ; Jannin S; Lesage A; Matysik J; Meersmann T; Prisner T; Reimer JA; Yang H; Koptyug IV Spin Hyperpolarization in Modern Magnetic Resonance. Chem. Rev 2023, 123 (4), 1417–1551. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Hoult DI; Richards RE The signal-to-noise ratio of the nuclear magnetic resonance experiment. J. Magn. Reson 1976, 24 (1), 71–85. [DOI] [PubMed] [Google Scholar]
  • 4.Goodson BM; Whiting N; Coffey AM; Nikolaou P; Shi F; Gust BM; Gemeinhardt ME; Shchepin RV; Skinner JG; Birchall JR; Barlow MJ; Chekmenev EY Hyperpolarization Methods for MRS. Emagres 2015, 4 (4), 797–810. [Google Scholar]
  • 5.Kovtunov KV; Pokochueva EV; Salnikov OG; Cousin S; Kurzbach D; Vuichoud B; Jannin S; Chekmenev EY; Goodson BM; Barskiy DA; Koptyug IV Hyperpolarized NMR: d-DNP, PHIP, and SABRE. Chem. Asian J 2018, 13 (15), 1857–1871. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Walkup LL; Woods JC Translational applications of hyperpolarized 3He and 129Xe. NMR Biomed. 2014, 27 (12), 1429–1438. [DOI] [PubMed] [Google Scholar]
  • 7.Venkatesh AK; Zhang AX; Mansour J; Kubatina L; Oh CH; Blasche G; Unlu MS; Balamore D; Jolesz FA; Goldberg BB; Albert MS MRI of the lung gas-space at very low-field using hyperpolarized noble gases. Magn. Reson. Imaging 2003, 21 (7), 773–776. [DOI] [PubMed] [Google Scholar]
  • 8.Savukov I; Karaulanov T; Castro A; Volegov P; Matlashov A; Urbatis A; Gomez J; Espy M Non-cryogenic anatomical imaging in ultra-low field regime: Hand MRI demonstration. J. Magn. Reson 2011, 211 (2), 101–108. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Krjukov E; Fichele S; Wild JM; Paley MNJ Design and evaluation of a low field system for hyperpolarized 3He gas imaging of neonatal lungs. Concept Magnetic Res. B 2007, 31B (4), 209–217. [Google Scholar]
  • 10.Six JS; Hughes-Riley T; Lilburn DML; Dorkes AC; Stupic KF; Shaw DE; Morris PG; Hall IP; Pavlovskaya GE; Meersmann T Pulmonary MRI contrast using Surface Quadrupolar Relaxation (SQUARE) of hyperpolarized 83Kr. Magn. Reson. Imaging 2014, 32 (1), 48–53. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Hughes-Riley T; Six JS; Lilburn DML; Stupic KF; Dorkes AC; Shaw DE; Pavlovskaya GE; Meersmann T Cryogenics free production of hyperpolarized 129Xe and 83Kr for biomedical MRI applications. J. Magn. Reson 2013, 237 (0), 23–33. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Six JS; Hughes-Riley T; Stupic KF; Pavlovskaya GE; Meersmann T Pathway to Cryogen Free Production of Hyperpolarized Krypton-83 and Xenon-129. PLoS One 2012, 7 (11), 16. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Albert MS; Cates GD; Driehuys B; Happer W; Saam B; Springer CS; Wishnia A Biological Magnetic-Resonance-Imaging Using Laser Polarized Xe-129. Nature 1994, 370 (6486), 199–201. [DOI] [PubMed] [Google Scholar]
  • 14.Antonacci MA; McHugh C; Kelley M; McCallister A; Degan S; Branca RT Direct detection of brown adipose tissue thermogenesis in UCP1−/− mice by hyperpolarized 129Xe MR thermometry. Sci. Rep 2019, 9 (1), 14865. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Rao MR; Norquay G; Stewart NJ; Hoggard N; Griffiths PD; Wild JM Assessment of brain perfusion using hyperpolarized 129Xe MRI in a subject with established stroke. J. Magn. Reson. Imaging 2019, 50 (3), 1002–1004. [DOI] [PubMed] [Google Scholar]
  • 16.Schroder L. Xenon for NMR biosensing - Inert but alert. Phys. Medica 2013, 29 (1), 3–16. [DOI] [PubMed] [Google Scholar]
  • 17.Hane FT; Li T; Smylie P; Pellizzari RM; Plata JA; DeBoef B; Albert MS In vivo detection of cucurbit[6]uril, a hyperpolarized xenon contrast agent for a xenon magnetic resonance imaging biosensor. Sci. Rep 2017, 7 (1), 41027. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Wang Y; Roose BW; Palovcak EJ; Carnevale V; Dmochowski IJ A Genetically Encoded β-Lactamase Reporter for Ultrasensitive 129Xe NMR in Mammalian Cells. Angew. Chem. Int. Ed 2016, 55 (31), 8984–8987. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Wang Y; Dmochowski IJ An Expanded Palette of Xenon-129 NMR Biosensors. Acc. Chem. Res 2016, 49 (10), 2179–2187. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Goodson BM; Kidd B; Hövener J-B; Schröder L; Theis T; Whiting N; Chekmenev EY, Nuclear Magnetic Resonance Spectroscopy ∣ Hyperpolarization for Sensitivity Enhancement. In Encyclopedia of Analytical Science (Third Edition), Worsfold P; Poole C; Townshend A; Miró M, Eds. Academic Press: Oxford, 2019; pp 168–181. [Google Scholar]
  • 21.Barskiy DA; Coffey AM; Nikolaou P; Mikhaylov DM; Goodson BM; Branca RT; Lu GJ; Shapiro MG; Telkki V-V; Zhivonitko VV, et al. NMR Hyperpolarization Techniques of Gases. Chem. Eur. J 2017, 23 (4), 725–751. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Schroder L; Lowery TJ; Hilty C; Wemmer DE; Pines A Molecular imaging using a targeted magnetic resonance hyperpolarized biosensor. Science 2006, 314 (5798), 446–449. [DOI] [PubMed] [Google Scholar]
  • 23.Zhang X; Yang Y; Yuan Y; Yue S; Zhao X; Yue Q; Zeng Q; Guo Q; Zhou X Hyperpolarized 129Xe Atoms Sense the Presence of Drug Molecules in Nanohosts Revealed by Magnetic Resonance Imaging. Anal. Chem 2024, 96 (25), 10152–10160. [DOI] [PubMed] [Google Scholar]
  • 24.Witte C; Schroder L NMR of hyperpolarised probes. NMR Biomed. 2013, 26 (7), 788–802. [DOI] [PubMed] [Google Scholar]
  • 25.Jayapaul J; Schröder L Molecular Sensing with Host Systems for Hyperpolarized 129Xe. Molecules 2020, 25 (20), 4627. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Mugler III JP; Driehuys B; Brookeman JR; Cates GD; Berr SS; Bryant RG; Daniel TM; De Lange EE; Downs JH; Erickson CJ; Happer W; Hinton DP; Kassel NF; Maier T; Phillips CD; Saam BT; Sauer KL; Wagshul ME MR imaging and spectroscopy using hyperpolarized 129Xe gas: preliminary human results. Magn. Reson. Med 1997, 37, 809–815. [DOI] [PubMed] [Google Scholar]
  • 27.Driehuys B; Martinez-Jimenez S; Cleveland ZI; Metz GM; Beaver DM; Nouls JC; Kaushik SS; Firszt R; Willis C; Kelly KT; Wolber J; Kraft M; McAdams HP Chronic Obstructive Pulmonary Disease: Safety and Tolerability of Hyperpolarized Xe-129 MR Imaging in Healthy Volunteers and Patients. Radiology 2012, 262 (1), 279–289. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Kennedy RR; Stokes JW; Downing P Anesthesia And The Inert-Gases With Special Reference To Xenon. Anaesth. Intensive Care 1992, 20 (1), 66–70. [DOI] [PubMed] [Google Scholar]
  • 29.Khan AS; Harvey RL; Birchall JR; Irwin RK; Nikolaou P; Schrank G; Emami K; Dummer A; Barlow MJ; Goodson BM; Chekmenev EY Enabling Clinical Technologies for Hyperpolarized Xenon-129 MRI and Spectroscopy. Angew. Chem. Int. Ed 2021, 60 (41), 22126–22147. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Mugler JP; Altes TA Hyperpolarized 129Xe MRI of the human lung. J. Magn. Reson. Imaging 2013, 37 (2), 313–331. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Shi M. Investigation on magnetic field response of a 87Rb-129Xe atomic spin comagnetometer. Opt. Express 2020, 28 (21), 32033–32041. [DOI] [PubMed] [Google Scholar]
  • 32.Schnurr M; Sloniec-Myszk J; Dopfert J; Schroder L; Hennig A Supramolecular Assays for Mapping Enzyme Activity by Displacement-Triggered Change in Hyperpolarized Xe-129 Magnetization Transfer NMR Spectroscopy. Angew. Chem. Int. Ed 2015, 54 (45), 13444–13447. [DOI] [PubMed] [Google Scholar]
  • 33.Knagge K; Smith JR; Smith LJ; Buriak J; Raftery D Analysis of porosity in porous silicon using hyperpolarized 129Xe two-dimensional exchange experiments. Solid State Nucl. Magn. Reson 2006, 29 (1), 85–89. [DOI] [PubMed] [Google Scholar]
  • 34.Pawsey S; Moudrakovski I; Ripmeester J; Wang LQ; Exarhos GJ; Rowsell JLC; Yaghi OM Hyperpolarized Xe-129 nuclear magnetic resonance studies of isoreticular metal-organic frameworks. J. Phys. Chem. C 2007, 111 (16), 6060–6067. [Google Scholar]
  • 35.Terskikh VV; Moudrakovski IL; Breeze SR; Lang S; Ratcliffe CI; Ripmeester JA; Sayari A A general correlation for the 129Xe NMR chemical shift-pore size relationship in porous silica-based materials. Langmuir 2002, 18, 5653–5656. [Google Scholar]
  • 36.Moudrakovski IL; Ratcliffe CI; Ripmeester JA 131Xe, a new NMR probe of void space in solids. J. Am. Chem. Soc 2001, 123, 2066–2067. [DOI] [PubMed] [Google Scholar]
  • 37.Raftery D; Long H; Meersmann T; Grandinetti PJ; Reven L; Pines A High-field NMR of adsorbed xenon polarized by laser pumping. Phys. Rev. Lett 1991, 66 (5), 584–587. [DOI] [PubMed] [Google Scholar]
  • 38.Walker TG; Happer W Spin-exchange optical pumping of noble-gas nuclei. Rev. Mod. Phys 1997, 69 (2), 629–642. [Google Scholar]
  • 39.Bouchiat MA; Carver TR; Varnum CM Nuclear Polarization in Gas Induced by Optical Pumping and Dipolar Exchange. Phys. Rev. Lett 1960, 5 (8), 373–375. [Google Scholar]
  • 40.Walker TG Fundamentals of Spin-Exchange Optical Pumping. J. Phys. Conf. Ser 2011, 294, 012001. [Google Scholar]
  • 41.Imai H; Yoshimura H; Kimura A; Fujiwara H Continuous flow production of concentrated hyperpolarized xenon gas from a dilute xenon gas mixture by buffer gas condensation. Sci. Rep 2017, 7 (1), 7352. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Zook AL; Adhyaru BB; Bowers CR High capacity production of > 65% spin polarized xenon-129 for NMR spectroscopy and imaging. J. Magn. Reson 2002, 159 (2), 175–182. [DOI] [PubMed] [Google Scholar]
  • 43.Ruset IC; Ketel S; Hersman FW Optical pumping system design for large production of hyperpolarized Xe-129. Phys. Rev. Lett 2006, 96 (5), 053002. [DOI] [PubMed] [Google Scholar]
  • 44.Birchall JR; Nikolaou P; Coffey AM; Kidd BE; Murphy M; Molway M; Bales LB; Goodson BM; Irwin RK; Barlow MJ; Chekmenev EY Batch-Mode Clinical-Scale Optical Hyperpolarization of Xenon-129 Using an Aluminum Jacket with Rapid Temperature Ramping. Anal. Chem 2020, 92 (6), 4309–4316. [DOI] [PubMed] [Google Scholar]
  • 45.Korchak SE; Kilian W; Mitschang L Configuration and Performance of a Mobile 129Xe Polarizer. Appl. Magn. Reson 2013, 44 (1-2), 65–80. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Whiting N; Nikolaou P; Eschmann NA; Barlow MJ; Lammert R; Ungar J; Hu W; Vaissie L; Goodson BM Using frequency-narrowed, tunable laser diode arrays with integrated volume holographic gratings for spin-exchange optical pumping at high resonant fluxes and xenon densities. Appl. Phys. B-Lasers Opt 2012, 106 (4), 775–788. [Google Scholar]
  • 47.Hattori M; Hiraga T; Homma K; Ohtake N Automated hyperpolarized 129Xe gas generator for nuclear magnetic resonance spectroscopy and imaging applications. J. Magn. Reson. Open 2024, 20, 100154. [Google Scholar]
  • 48.Knagge K; Prange J; Raftery D A continuously recirculating optical pumping apparatus for high xenon polarization and surface NMR studies. Chem. Phys. Lett 2004, 397 (1-3), 11–16. [Google Scholar]
  • 49.Driehuys B; Cates GD; Miron E; Sauer K; Walter DK; Happer W High-volume production of laser-polarized Xe-129. Appl. Phys. Lett 1996, 69 (12), 1668–1670. [Google Scholar]
  • 50.Norquay G; Collier GJ; Rao M; Stewart NJ; Wild JM 129Xe-Rb Spin-Exchange Optical Pumping with High Photon Efficiency. Phys. Rev. Lett 2018, 121 (15), 153201. [DOI] [PubMed] [Google Scholar]
  • 51.Mortuza MG; Anala S; Pavlovskaya GE; Dieken TJ; Meersmann T Spin-exchange optical pumping of high-density xenon-129. J. Chem. Phys 2003, 118 (4), 1581–1584. [Google Scholar]
  • 52.Roose BW; Zemerov SD; Dmochowski IJ Nanomolar small-molecule detection using a genetically encoded 129Xe NMR contrast agent. Chem. Sci 2017, 8 (11), 7631–7636. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Zemerov SD; Roose BW; Farenhem KL; Zhao Z; Stringer MA; Goldman AR; Speicher DW; Dmochowski IJ 129Xe NMR-Protein Sensor Reveals Cellular Ribose Concentration. Anal. Chem 2020, 92 (19), 12817–12824. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Norquay G; Parnell SR; Xu XJ; Parra-Robles J; Wild JM Optimized production of hyperpolarized Xe-129 at 2 bars for in vivo lung magnetic resonance imaging. J. Appl. Phys 2013, 113 (4), 9. [Google Scholar]
  • 55.Nikolaou P; Coffey AM; Walkup LL; Gust BM; Whiting N; Newton H; Barcus S; Muradyan I; Dabaghyan M; Moroz GD; Rosen M; Patz S; Barlow MJ; Chekmenev EY; Goodson BM Near-unity nuclear polarization with an 'open-source' 129Xe hyperpolarizer for NMR and MRI. Proc. Natl. Acad. Sci. U. S. A 2013, 110 (35), 14150–14155. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56.Birchall JR; Irwin RK; Nikolaou P; Coffey AM; Kidd BE; Murphy M; Molway M; Bales LB; Ranta K; Barlow MJ; Goodson BM; Rosen MS; Chekmenev EY XeUS: A second-generation automated open-source batch-mode clinical-scale hyperpolarizer. J. Magn. Reson 2020, 319, 106813. [DOI] [PubMed] [Google Scholar]
  • 57.Nikolaou P; Coffey AM; Walkup LL; Gust BM; Whiting NR; Newton H; Muradyan I; Dabaghyan M; Ranta K; Moroz G; Patz S; Rosen MS; Barlow MJ; Chekmenev EY; Goodson BM XeNA: An automated 'open-source' 129Xe hyperpolarizer for clinical use. Magn. Reson. Imaging 2014, 32 (5), 541–550. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58.Rosen MS; Chupp TE; Coulter KP; Welsh RC; Swanson SD Polarized Xe-129 optical pumping/spin exchange and delivery system for magnetic resonance spectroscopy and imaging studies. Rev. Sci. Instrum 1999, 70 (2), 1546–1552. [Google Scholar]
  • 59.Ruth U; Hof T; Schmidt J; Fick D; Jansch HJ Production of nitrogen-free, hyperpolarized Xe-129 gas. Appl. Phys. B-Lasers Opt 1999, 68 (1), 93–97. [Google Scholar]
  • 60.Desvaux H; Gautier T; Le Goff G; Petro M; Berthault P Direct evidence of a magnetization transfer between laser-polarized xenon and protons of a cage-molecule in water. Eur. Phys. J. D 2000, 12 (2), 289–296. [Google Scholar]
  • 61.Nikolaou P; Whiting N; Eschmann NA; Chaffee KE; Goodson BM Generation of laser-polarized xenon using fiber-coupled laser-diode arrays narrowed with integrated volume holographic gratings. J. Magn. Reson 2009, 197 (2), 249–254. [DOI] [PubMed] [Google Scholar]
  • 62.Parnell SR; Deppe MH; Parra-Robles J; Wild JM Enhancement of 129Xe polarization by off-resonant spin exchange optical pumping. J. Appl. Phys 2010, 108, 064908. [Google Scholar]
  • 63.Yee MK; Janulewicz PA; Seichepine DR; Sullivan KA; Proctor SP; Krengel MH Multiple Mild Traumatic Brain Injuries Are Associated with Increased Rates of Health Symptoms and Gulf War Illness in a Cohort of 1990-1991 Gulf War Veterans. Brain Sci 2017, 7 (7), 79. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 64.Whiting N; Nikolaou P; Eschmann NA; Barlow MJ; Goodson BM Interdependence of Xenon Density and Temperature on Rb/129Xe Optical Pumping Efficiency at High Xenon Densities. J. Magn. Reson 2011, 208 (2), 298–304. [DOI] [PubMed] [Google Scholar]
  • 65.Hamedani H; Kadlecek S; Ruppert K; Xin Y; Duncan I; Rizi RR Ventilation heterogeneity imaged by multibreath wash-ins of hyperpolarized 3He and 129Xe in healthy rabbits. The Journal of physiology 2021, 599 (17), 4197–4223. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 66.Birchall JR; Chowdhury MRH; Nikolaou P; Chekmenev YA; Shcherbakov A; Barlow MJ; Goodson BM; Chekmenev EY Pilot Quality-Assurance Study of a Third-Generation Batch-Mode Clinical-Scale Automated Xenon-129 Hyperpolarizer. Molecules 2022, 27 (4), 1327. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 67.Birchall JR; Nikolaou P; Coffey AM; Kidd BE; Irwin RK; Murphy M; Molway M; Bales LB; Goodson BM; Barlow MJ; Chekmenev EY Batch-Mode Clinical-Scale Optical Hyperpolarization of Xenon-129 Using an Aluminum Jacket with Rapid Temperature Ramping. Anal. Chem 2020, 92 (6), 4309–4316. [DOI] [PubMed] [Google Scholar]
  • 68.Chowdhury MR; Ahmed F; Oladun C; Adelabu I; Abdurraheem A; Nantogma S; Birchall JR; Gafar AT; Chekmenev YA; Nikolaou P; Barlow MJ; Goodson BM; Shcherbakov A; Chekmenev EY Low-Cost Purpose-Built Ultra-Low-Field NMR Spectrometer. Anal. Chem 2024, 96 (42), 16724–16734. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 69.Joalland B; Nantogma S; Chowdhury MRH; Nikolaou P; Chekmenev EY Magnetic Shielding of Parahydrogen Hyperpolarization Experiments for the Masses. Magn. Reson. Chem 2021, 59 (12), 1180–1186. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 70.Nantogma S; Chowdhury MRH; Kabir MSH; Adelabu I; Joshi SM; Samoilenko A; de Maissin H; Schmidt AB; Nikolaou P; Chekmenev YA; Salnikov OG; Chukanov NV; Koptyug IV; Goodson BM; Chekmenev EY MATRESHCA: Microtesla Apparatus for Transfer of Resonance Enhancement of Spin Hyperpolarization via Chemical-exchange and Addition. Anal. Chem 2024, 96 (10), 4171–4179. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 71.Nikolaou P; Coffey AM; Ranta K; Walkup LL; Gust B; Barlow MJ; Rosen MS; Goodson BM; Chekmenev EY Multi-Dimensional Mapping of Spin-Exchange Optical Pumping in Clinical-Scale Batch-Mode 129Xe Hyperpolarizers. J. Phys. Chem. B 2014, 118 (18), 4809–4816. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 72.Nikolaou P; Coffey AM; Walkup LL; Gust B; LaPierre C; Koehnemann E; Barlow MJ; Rosen MS; Goodson BM; Chekmenev EY A 3D-Printed High Power Nuclear Spin Polarizer. J. Am. Chem. Soc 2014, 136 (4), 1636–1642. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 73.Skinner JG; Ranta K; Whiting N; Coffey AM; Nikolaou P; Rosen MS; Chekmenev EY; Morris PG; Barlow MJ; Goodson BM High Xe Density, High Photon Flux, Stopped-Flow Spin-Exchange Optical Pumping: Simulations versus Experiments. J. Magn. Reson 2020, 312, 106686. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 74.Birchall JR; Irwin RK; Nikolaou P; Pokochueva EV; Kovtunov KV; Koptyug IV; Barlow MJ; Goodson BM; Chekmenev EY Pilot multi-site quality assurance study of batch-mode clinical-scale automated xenon-129 hyperpolarizers. J. Magn. Reson 2020, 316, 106755. [DOI] [PubMed] [Google Scholar]
  • 75.Nikolaou P; Coffey AM; Barlow MJ; Rosen M; Goodson BM; Chekmenev EY Temperature-Ramped 129Xe Spin Exchange Optical Pumping. Anal. Chem 2014, 86 (16), 8206–8212. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 76.Zemerov SD; Dmochowski IJ Cryptophane–xenon complexes for 129Xe MRI applications. RSC Adv. 2021, 11 (13), 7693–7703. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 77.Hane FT; Li T; Smylie P; Pellizzari RM; Plata JA; DeBoef B; Albert MS In vivo detection of cucurbit[6]uril, a hyperpolarized xenon contrast agent for a xenon magnetic resonance imaging biosensor. Sci. Rep 2017, 7, 41027. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

supporting information

RESOURCES