. 2023 Jul 28;1(3):357–365. doi: 10.1016/j.mcpdig.2023.06.012

Validation of a Smart Ring Oximeter in Individuals With Dark Skin Pigment

Miguel Meira e Cruz a,b,, Congcong Zhou c,d, Meir H Kryger e, Hongwei Wang f
PMCID: PMC11975752  PMID: 40206614

Abstract

Objective

To evaluate a ring wearable’s accuracy in the measurement of oxyhemoglobin saturation measured by pulse oximeter (SpO2) compared to a simultaneous SaO2 obtained from arterial blood gases (ABGs) because oximeters may overestimate oxyhemoglobin saturation in darkly pigmented individuals resulting in occult hypoxemia.

Patients and Methods

Circul+ (Bodimetrics), with a form factor of a ring, measures several variables for cardiorespiratory assessment (SpO2, movement, and heart rate). The design ensures that measurements are made on the palmar surface of the digit, where there is generally less pigment. Twenty-four healthy participants (8 Black, 16 non-Black) had catheters inserted into their radial arteries during the timeframe between August 2020 and January 2022. They were administered nitrogen-rich air that made them hypoxic in steps to an SaO2 of 70%. ABGs was sampled at various levels of hypoxia, and simultaneous readings were obtained from the Circul+. We compared SpO2 from Circul with measurement of ABGs by a medical grade blood gas analyzer.

Results

There was excellent correlation between SaO2 measured by the ring oximeter and ABGs in both Black (y = 1.0174x − 1.573; R2 = 0.9414) and Non-Black (y = 1.0209x − 2.5607; R2 = 0.9207) participants. No differences were found in comparing the intercept and slopes of the regressions. At ABG of 70% and 100%, the SaO2 measured by the ring was calculated to be 69.6% and 100.2% for the Black participants and 68.9% and 99.5% for the non-Black participants. Bland-Altman analysis found a bias (deviation from the mean) of 0.0 for the Black participants and −0.7% for the non-Black participants.

Conclusion

Results from this study confirmed that Circul+ oximetry accuracy seems to be independent of skin tone, perhaps because measurements are made on the palmar surface of the digit, where there is generally less pigment.


Accuracy is critical for screening, diagnosis, and follow-up in medicine. In recent years, the development of wearable technology has been providing inexpensive tools for data collection and analysis. These tools, when validated, have potential utility for both research and clinical medicine.

Wearable devices (wrist worn, fingertip, and ring) have been developed that have the ability to continuously monitor relevant respiratory variables such as oxyhemoglobin saturation (SaO2). The major class of currently available pulse oximetry devices compute arterial oxyhemoglobin saturation from the ratio of the pulsatile to the total transmitted red light divided by the same ratio for infrared light transilluminating a finger.

Derived oxygen saturation (SpO2) should not be impacted by skin pigmentation. Although several studies reported no significant pigment-related interferences in portable pulse oximeters,1, 2, 3 some early authors published less accurate results in dark-skinned individuals,4 and now, many articles have indicated that there is a great reason for concern.5 The proliferation and widespread use of pulse oximeters by the public during the COVID pandemic and the recent articles highlighting unintended racial bias in the devices have raised alarms.6, 7, 8, 9, 10, 11, 12, 13, 14 In particular, the inaccuracies (eg, occult hypoxemia12) can negatively impact patient care.7 Occult hypoxemia has been defined as arterial blood oxygen saturation (SaO2) of less than 88% despite a pulse oximetry (SpO2) measurement of 92% or higher.12 False high oximeter measurements can negatively impact patient care.15 There are many oximeters currently marketed to both the public and the medical profession. Clinicians managing patients should have access to the validation data of the devices. Here we present the validation data of the Bodimetrics Circul+ oximeter.

We hypothesize that measuring on the palmar surface of digits, where there is generally less pigment, will result in acceptable SaO2 values without occult hypoxemia in people with dark skin tones.

Patients and Methods

The Circul+ (Bodimetrics) is a device with a form factor of a ring, which continuously measures several variables (SpO2 and heart rate by beat-by-beat reflectance photoplethysmography, and movement by accelerometer). The data are sampled beat by beat and processed into a 10-second moving window. The arterial blood samples and the corresponding SpO2 measurements were obtained using the following protocol. The ring has a patented design that assures that measurements are made on the palmar aspect of the finger (Figure 1). This work was performed at Tongde Hospital of Zhejiang province, China.

Figure 1.

Figure 1

The sensors of the ring are close to the pulsating proper palmar digital artery.

Selection of Participants and Demographic characteristics

Data were obtained for as part of submissions of the Circul+ to the China Food and Drug Administration, currently the National Medical Products Administration. Volunteers were recruited by advertising and were students or hospital staff. Data were obtained from 24 participants (12 men and 12 women; 8 Black and 16 non-Black participants). The mean age was 22.58±2.71 years for men and 26±6.24 years for women. The Black patients were Fitzpatrick scale type VI.16 The others were of Han Chinese, mainly types I to IV. The research was approved by the ethics committee of Tongde Hospital, and all participants signed informed consent forms.

Protocol and Data Acquisition

In a hospital setting, after topical anesthesia, participants had a catheter placed in a radial artery. Testing environment conditions were as follows: temperature 22-25 °C and humidity 50%-65%. The sensors (Circul+; Bodimetrics) for SpO2 and IntelliVue MX500/MX550 (Philips) for pulse rate and SpO2 were placed on fingers the same side as the arterial catheter. Each participant while breathing naturally was connected to an anesthesia breathing circuit with the inspired oxygen concentration adjusted to 21%. Tidal volume and breathing rate were monitored continuously. The relevant physiologic parameters, described below, were recorded. When the SpO2 was stable for at least 30 seconds and between 97% and 100%, 5 samples of blood (1-2 mL) were withdrawn with intervals of at least 20 seconds. The SaO2 of the samples was obtained by blood gas analysis (ABL90 FLEX blood gas analyzer; Radiometer), and the vital signs of the participant were observed and recorded. The participants’ inhaled oxygen concentration was then reduced by the progressive addition of nitrogen into the circuit. When the SpO2 reached 92%-96% and was stable for at least 30 seconds, 5 samples of blood (1-2 mL) were withdrawn during intervals of at least 20 seconds. As mentioned earlier, the SaO2 of the samples was obtained by blood gas analysis, and the vital signs of the participant were observed and recorded. As above, inhaled oxygen concentration was then further reduced sequentially so that measurements of 5 samples were made at each of the following ranges: 85%-91%; 78%-84%; and 70%-77%. Once all the data and samples had been collected, the participant-anesthesia machine interface was removed, and the participant breathed room air. After observing that the participant’s vital signs were normal, the arterial catheter was removed and the site pressurized to stop any bleeding, and the participant left the bed. After confirming that the participant was stable and physiological variables normal, at least 1 hour after leaving bed, the participant was discharged.

Statistical Analyses

Statistical analysis was conducted using IBM SPSS (version 25.0, SPSS).

Results

The data comparing Circul+ SpO2 and arterial blood gas (ABG)-obtained SaO2 in the Black and non-Black participants are reported in Figure 2. There was excellent correlation between SaO2 measured by the ring oximeter and ABGs in both Black (y = 1.0174x − 1.573; R2 = 0.9414) and non-Black (y = 1.0209x − 2.5607; R2 = 0.9207) participants. No relevant differences were found in comparing the intercept and slopes of the regressions. At ABG SaO2 of 70% and 100%, the SpO2 measured by the ring was calculated to be 69.6% and 100.2% for the Black participants and 68.9% and 99.5% for the non-Black participants, respectively. Figure 3 shows the Bland-Altman analyses of the Black and non-Black participants and Figure 4 shows analysis using a modified Bland-Altman technique. This latter analysis was done because the SaO2 measured by ABGs could be considered the gold standard. The Table summarizes the average root mean square error for the decadal ranges of SaO2. Figure 4 shows a box plot of (SpO2 − SaO2) by SpO2 range increments for all participants. This shows that the pulse oximeter readings were slightly lower than SaO2 when SaO2 was less than 80% for both Black and non-Black participants. Thus, severe hypoxemia would not be missed. This is corroborated in Figure 5, which shows the count of SaO2/SpO2 pairs by absolute value of the bias 3% or less, or more than 3% over SaO2 range. There were very few measurements in which SpO2 deviated by more than 3%.

Figure 2.

Figure 2

SpO2 measured by ring oximeter vs SaO2 measured by ABG. Hypoxemia was induced by lowering inhaled oxygen concentration.

Figure 3.

Figure 3

Bland-Altman plots for the Black (top) and non-Black (bottom) participants.

Figure 4.

Figure 4

Modified Bland-Altman plots for the Black (top) and non-Black (bottom) participants.

Table.

Average Root Mean Square Error

SaO2 range
90%-100% 80%-89% 70%-79%
Non-Black participants 1.57 2.66 2.71
Black participants 1.73 3.84 2.84

Values are root mean square errors for the decadal ranges of SaO2

SaO2, oxyhemoglobin saturation.

Figure 5.

Figure 5

Box plot of (SpO2 − SaO2) by SpO2 range increments for all participants. Pulse oximeter readings were slightly lower than SaO2 when SaO2 was <80% for both Black and non-Black participants.

The data comparing the Circul with a medical grade finger oximeter are reported for SpO2 and pulse in Figures 6 and 7, respectively.

Figure 6.

Figure 6

Count of SaO2/SpO2 pairs by absolute value of the bias ≤3% or >3% over the SaO2 range. There were very few measurements in which SpO2 deviated by >3% from measured SaO2 for both Black (green bars) and non-Black (blue bars) participants.

Figure 7.

Figure 7

Comparison of the Circul oximeter with a medical grade oximeter using regression analyses (A and B) for Black and non-Black participants and Bland-Altman plots (C and D).

Discussion

The Circul+ ring oximeter was found to be reasonably accurate in the measurement of SpO2 and pulse rate in both Black and non-Black participants in the range between 100% and 70% SaO2 when compared with ABGs (the gold standard) and a medical grade finger oximeter. Of great interest, there was not a single instance of occult hypoxemia (SpO2 > 92% when SaO2 was <88%). The explanation for its accuracy in Black participants may be related to the patented design that ensures that the readings are obtained from the palmar aspect of the finger, which generally has less pigment than the back of the hand. This device has been reported to be useful in screening for sleep apnea.17

There are many consumer oximeters on the market, and their validation data are generally not available.18 Most such oximeters available in the United States have not been cleared or approved by the US FDA. Food and Drug Administration approval or clearance (based on similarity to a predicate approved device) requires “10 or more healthy participants” and “participants with a range of skin pigmentations, including at least 2 darkly pigmented participants.”19 The inadequacy of this requirement is obvious, especially given the decades of knowledge about potential problems with oximeters.4,20,21 The first widely used oximeter, which was not a pulse oximeter (Hewlett Packard 407201A) was validated in a large number of Black patients and was found to accurate in this population.22,23

During the COVID pandemic, patients were often instructed to self-monitor with consumer oximeters at home rather than coming to hospitals.24 It appears that there was little thought given to potential pitfalls (eg, were the recommended devices accurate?) and at times inexplicable readings.25 The alarms raised first by Sjoding et al10 during the start of the COVID pandemic and more recent studies7,9,11,12,15 suggest that pulse oximeters may miss occult hypoxemia in patients with dark skin pigment. That unintended racial bias in oximetry, which was noted by the media, was an issue that led the Federation of American Scientists to convene a forum to review this issue.26

Why Is the Circul+ Ring Less Affected by Dark Pigment?

There are at least 2 possible explanations. First, as mentioned in Methods and shown in Figure 1, the sensors on the ring are on the palmar aspect of the finger closer to the palmar digital artery. Second, the palmar aspect of the hand usually has less pigment than the dorsum (Figure 8).

Figure 8.

Figure 8

The palmar aspect of the hand is less pigmented than the dorsum. There is often increased pigment on the palmar aspect in the area of interphalangeal joints.

Oxygen saturation is a critical variable in management of patients with suspected or known hypoxemia. We believe that the validation data of all devices that measure SpO2 should be available to the prescriber and the user and that the pitfalls in measurement be understood by both. Even when validation is available, it is important to consider that validation studies are done in young healthy participants in controlled laboratory conditions. Clinicians should be aware of other well-known sources of SpO2 measurement errors and artifacts including those related to motion, medical conditions that might cause venous pulsations in tissue beds, poor peripheral perfusion that might occur in heart failure or shock, sensor placement and device sampling rate and signal filtering,27 nail polish, and abnormal hemoglobin (carboxyhemoglobin and methemoglobin) levels.18 SpO2 often overestimates SaO2 when blood pressure and heart rate decrease.28 Thus, the measurement of SaO2 with an ABG may be appropriate in some clinical situations.28

Technology is moving very quickly, and the boundaries between consumer and medical grade devices sometimes overlap. Wearables are being developed for monitoring in several medical fields.29, 30, 31, 32, 33 It is important that developers ensure that their products do not introduce unintended bias into the devices. It is also important that validation data be available.

Conclusion

Results from this validation study confirmed that Circul+ oximetry accuracy seems to be independent of skin tone. Measurement in the palmar aspect of the digits (where there less pigmentation) may improve the accuracy of such devices in participants with dark skin tones.

Potential Competing Interests

Dr Meira e Cruz is a scientific adviser for Bodimetrics. Dr Kryger is a scientific adviser for Bodimetrics, Dormothech, and Wesper. All other authors report no competing interests.

References

  • 1.Adler J.N., Hughes L.A., Vivilecchia R., Camargo C.A., Jr. Effect of skin pigmentation on pulse oximetry accuracy in the emergency department. Acad Emerg Med. 1998;5(10):965–970. doi: 10.1111/j.1553-2712.1998.tb02772.x. [DOI] [PubMed] [Google Scholar]
  • 2.Bothma P.A., Joynt G.M., Lipman J., et al. Accuracy of pulse oximetry in pigmented patients. S Afr Med J. 1996;86(5 Suppl):594–596. [PubMed] [Google Scholar]
  • 3.Barker S.J., Wilson W.C. Racial effects on Masimo pulse oximetry: a laboratory study. J Clin Monit Comput. 2023;37(2):567–574. doi: 10.1007/s10877-022-00927-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Bickler P.E., Feiner J.R., Severinghaus J.W. Effects of skin pigmentation on pulse oximeter accuracy at low saturation. Anesthesiology. 2005;102(4):715–719. doi: 10.1097/00000542-200504000-00004. [DOI] [PubMed] [Google Scholar]
  • 5.Cabanas A.M., Fuentes-Guajardo M., Latorre K., Leon D., Martin-Escudero P. Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis. Sensors (Basel) 2022;22(9):3402. doi: 10.3390/s22093402. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Chesley C.F., Lane-Fall M.B., Panchanadam V., et al. Racial disparities in occult hypoxemia and clinically based mitigation strategies to apply in advance of technological advancements. Respir Care. 2022;67(12):1499–1507. doi: 10.4187/respcare.09769. [DOI] [PubMed] [Google Scholar]
  • 7.Fawzy A., Wu T.D., Wang K., et al. Racial and ethnic discrepancy in pulse oximetry and delayed identification of treatment eligibility among patients with COVID-19. JAMA Intern Med. 2022;182(7):730–738. doi: 10.1001/jamainternmed.2022.1906. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Jamali H., Castillo L.T., Morgan C.C., et al. Racial disparity in oxygen saturation measurements by pulse oximetry: evidence and implications. Ann Am Thorac Soc. 2022;19(12):1951–1964. doi: 10.1513/AnnalsATS.202203-270CME. [DOI] [PubMed] [Google Scholar]
  • 9.Sjoding M.W., Ansari S., Valley T.S. Origins of racial and ethnic bias in pulmonary technologies. Annu Rev Med. 2023;74:401–412. doi: 10.1146/annurev-med-043021-024004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Sjoding M.W., Dickson R.P., Iwashyna T.J., Gay S.E., Valley T.S. Racial bias in pulse oximetry measurement. N Engl J Med. 2020;383(25):2477–2478. doi: 10.1056/NEJMc2029240. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Valbuena V.S.M., Barbaro R.P., Claar D., et al. Racial bias in pulse oximetry measurement among patients about to undergo extracorporeal membrane oxygenation in 2019-2020: a retrospective cohort study. Chest. 2022;161(4):971–978. doi: 10.1016/j.chest.2021.09.025. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Valbuena V.S.M., Seelye S., Sjoding M.W., et al. Racial bias and reproducibility in pulse oximetry among medical and surgical inpatients in general care in the Veterans Health Administration 2013-19: multicenter, retrospective cohort study. BMJ. 2022;378 doi: 10.1136/bmj-2021-069775. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Al-Halawani R., Charlton P.H., Qassem M., Kyriacou P.A. A review of the effect of skin pigmentation on pulse oximeter accuracy. Physiol Meas. 2023;44(5) doi: 10.1088/1361-6579/acd51a. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Kyriacou P.A., Charlton P.H., Al-Halawani R., Shelley K.H. Inaccuracy of pulse oximetry with dark skin pigmentation: clinical implications and need for improvement. Br J Anaesth. 2023;130(1):e33–e36. doi: 10.1016/j.bja.2022.03.011. [DOI] [PubMed] [Google Scholar]
  • 15.Crooks C.J., West J., Morling J.R., et al. Differential pulse oximetry readings between ethnic groups and delayed transfer to intensive care units. QJM. 2023;116(1):63–67. doi: 10.1093/qjmed/hcac218. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Fitzpatrick T.B. The validity and practicality of sun-reactive skin types I through VI. Arch Dermatol. 1988;124(6):869–871. doi: 10.1001/archderm.124.6.869. [DOI] [PubMed] [Google Scholar]
  • 17.Zhao R., Xue J., Zhang X., et al. Comparison of ring pulse oximetry using reflective photoplethysmography and PSG in the detection of OSA in Chinese adults: a pilot study. Nat Sci Sleep. 2022;14:1427–1436. doi: 10.2147/NSS.S367400. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Blike G.T., McGrath S.P., McGovern K. Skin pigmentation effects on pulse oximetry accuracy need a prospective study. J Patient Saf. 2023;19(4):221–222. doi: 10.1097/PTS.0000000000001116. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Food and Drug Administration Pulse Oximeters—Premarket Notification Submissions [510(k)s]: guidance for Industry and Food and Drug Administration Staff. https://www.fda.gov/regulatory-information/search-fda-guidance-documents/pulse-oximeters-premarket-notification-submissions-510ks-guidance-industry-and-food-and-drug
  • 20.Ries A.L., Farrow J.T., Clausen J.L. Accuracy of two ear oximeters at rest and during exercise in pulmonary patients. Am Rev Respir Dis. 1985;132(3):685–689. doi: 10.1164/arrd.1985.132.3.685. [DOI] [PubMed] [Google Scholar]
  • 21.Ries A.L., Prewitt L.M., Johnson J.J. Skin color and ear oximetry. Chest. 1989;96(2):287–290. doi: 10.1378/chest.96.2.287. [DOI] [PubMed] [Google Scholar]
  • 22.Saunders N.A., Powles A.C., Rebuck A.S. Ear oximetry: accuracy and practicability in the assessment of arterial oxygenation. Am Rev Respir Dis. 1976;113(6):745–749. doi: 10.1164/arrd.1976.113.6.745. [DOI] [PubMed] [Google Scholar]
  • 23.Moran-Thomas A. Oximeters used to be designed for equity. What happened? WIRED Magazine. https://www.wired.com/story/pulse-oximeters-equity/
  • 24.Luks A.M., Swenson E.R. Pulse oximetry for monitoring patients with COVID-19 at home. Potential pitfalls and practical guidance. Ann Am Thorac Soc. 2020;17(9):1040–1046. doi: 10.1513/AnnalsATS.202005-418FR. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Tobin M.J., Laghi F., Jubran A. Why COVID-19 silent hypoxemia is baffling to physicians. Am J Respir Crit Care Med. 2020;202(3):356–360. doi: 10.1164/rccm.202006-2157CP. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Federation of Americal Scientists Bias in pulse oximetry forum. https://fas.org/event/bias-in-pulse-oximetry-forum/
  • 27.West P., George C.F., Kryger M.H. Dynamic in vivo response characteristics of three oximeters: Hewlett-Packard 47201A, Biox III, and Nellcor N-100. Sleep. 1987;10(3):263–271. doi: 10.1093/sleep/10.3.263. [DOI] [PubMed] [Google Scholar]
  • 28.Crooks C.J., West J., Morling J., et al. Inverse association between blood pressure and pulse oximetry accuracy: an observational study in patients with suspected or confirmed COVID-19 infection. Emerg Med J. 2023;40(3):216–220. doi: 10.1136/emermed-2022-212443. [DOI] [PubMed] [Google Scholar]
  • 29.Baman J.R., Passman R.S. The future of long-term monitoring after catheter and surgical ablation for atrial fibrillation. J Cardiovasc Electrophysiol. 2022;33(8):1911–1918. doi: 10.1111/jce.15375. [DOI] [PubMed] [Google Scholar]
  • 30.Lui G.Y., Loughnane D., Polley C., Jayarathna T., Breen P.P. The Apple watch for monitoring mental health-related physiological symptoms: literature review. JMIR Ment Health. 2022;9(9) doi: 10.2196/37354. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Viderman D., Seri E., Aubakirova M., et al. Remote monitoring of chronic critically ill patients after hospital discharge: a systematic review. J Clin Med. 2022;11(4):1010. doi: 10.3390/jcm11041010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Wang Y., Haick H., Guo S., et al. Skin bioelectronics towards long-term, continuous health monitoring. Chem Soc Rev. 2022;51(9):3759–3793. doi: 10.1039/d2cs00207h. [DOI] [PubMed] [Google Scholar]
  • 33.Zhang Y., Ren X., Liang X., Ye X., Zhou C. A refined blood pressure estimation model based on single channel photoplethysmography. IEEE J Biomed Health Inform. 2022;26(12):5907–5917. doi: 10.1109/JBHI.2022.3206477. [DOI] [PubMed] [Google Scholar]

Articles from Mayo Clinic Proceedings: Digital Health are provided here courtesy of Elsevier