Abstract.
Significance
Optoacoustic tomography systems commonly employ bulky and expensive solid-state laser sources readily capable of generating dozens of millijoules of optical energy per pulse. Light-emitting diodes (LEDs) may offer a significantly more affordable and compact solution with excellent pulse-to-pulse stability. Yet, the optical design must optimize the energy density delivered to the target, given the relatively low per-pulse energy output of LEDs.
Aim
We exploit a full-view LED-based optoacoustic tomography (FLOAT) configuration for in vivo imaging of mice.
Approach
The system features panoramic light illumination delivering 0.48 mJ of total per-pulse energy with an array of 160 LEDs stacked into a cylindrically focused circular ultrasound array transducer.
Results
We characterize the imaging performance of the FLOAT system in tissue-mimicking phantoms, subsequently demonstrating its ability for in vivo cross-sectional mouse imaging.
Conclusions
It is anticipated that the compact, low-cost FLOAT imaging system will open up new venues in resource-limited settings for studying large-scale biodynamics such as pharmacokinetics and biodistribution of molecular agents and drugs on a whole-body level.
Keywords: light-emitting diode, whole-body imaging, optoacoustic/photoacoustic tomography, affordable light sources
1. Introduction
Preclinical imaging plays an important role in studying disease pathophysiology and monitoring the efficacy of potential treatments.1,2 Owing to the development of transgenic and gene knockout/in techniques, small rodents are extensively used as animal models.2 Downscaling of several clinical imaging modalities, such as computed tomography (CT),3,4 magnetic resonance imaging (MRI),5,6 positron emission tomography (PET),7,8 and pulse-echo ultrasound (US),9,10 have greatly impacted preclinical studies. Other approaches based on optical contrast have further been developed for whole-body functional and molecular imaging.11–14 Albeit optical imaging techniques are free of ionizing radiation and have the particular advantage of rich functional and molecular contrast, they suffer from poor resolution when applied beyond the superficial areas in optically opaque mammalian tissues. Recently, optoacoustic tomography (OAT) has been gaining popularity in preclinical and clinical research15–17 due to its unique ability to combine rich optical contrast with high spatial resolution enabled by US beamforming in deep tissues. In addition, OAT systems offer fast imaging capabilities at frame rates of hundreds to thousands of hertz, which is important when observing rapid biodynamics.18
Conventional OAT systems employ -switched Nd:YAG pumped tunable optical parametric oscillator (OPO) or dye lasers that deliver high-energy laser pulses (typically in the range of 10 to 200 mJ) for efficient generation of US waves within biological tissue.19 These Nd:YAG-based lasers are bulky, heavy, and expensive. A vibration–isolation optical table is often needed to house those lasers to deliver stable energy output, making them difficult to move between locations.20 Portable OPO lasers have become available in recent years,21,22 but their cost remains prohibitive for widespread use. Recently, alternative light sources such as pulsed laser diodes (PLDs)23–25 and light-emitting diodes (LEDs)26–28 have gained attention owing to their compact and portable footprint, affordability, and high pulse repetition rates reaching several kHz.29 The non-coherent nature of LEDs offers a much safer and cost-effective alternative for OAT implementations, particularly in resource-limited and clinical settings. Generally, the pulse energies delivered by individual LED light sources are very low in the range of a few , which may result in unacceptable signal-to-noise ratio (SNR) levels. Nevertheless, stacked arrays of LEDs deliver higher pulse energies up to a few 100s of .30,31
Here, we exploit a full-view LED-based optoacoustic tomography (FLOAT) configuration for whole-body imaging of mice. FLOAT has the advantage of panoramic light illumination with full-view optoacoustic signal detection in 360 deg. It is based on custom-made stacked LED arrays delivering a total per-pulse energy of 0.48 mJ around the imaging object placed at the center of the circular full-ring ultrasound array transducer. This transducer plays a crucial role in circumventing limited-view artifacts and enhances the effective field-of-view (FOV) and image quality for cross-sectional imaging.
2. Methods
Schematic of the FLOAT system is shown in Fig. 1. It utilizes high-power stacked arrays, each consisting of 10 LEDs (SFH4171S, OSRAM, Munich, Germany), for light excitation at 850 nm wavelength. Sixteen such arrays were incorporated, with eight attached to both the top and bottom of a custom-built circular 80-mm-diameter ultrasound transducer (IMASONIC SAS, Voray-sur-l’Ognon, France). These LED arrays provided uniform illumination around the imaged object. A plexiglass membrane (2 mm thickness) isolated the LED circuit from water while allowing light transmission. The LEDs were pulsed using a custom-designed driver unit32 that delivered peak currents of , controlled by an external function generator (Agilent, Santa Clara, California, United States) at 50 Hz and powered by a high-voltage supply (Wemaxpower, Shenzhen, China). Note that the heat generated by the LED arrays in pulsed mode operation is dissipated using heat sinks attached to the back of each LED array [Fig. 1(d)]. The full width at half maximum (FWHM) of the light pulse from LEDs is with rise and fall times of and , respectively.32 Each LED emitted per pulse, as measured with a pyroelectric sensor (J-10MT-10 KHZ, Coherent, Saxonburg, Pennsylvania, United States), resulting in a total of 0.48 mJ with a total of 160 LEDs (16 arrays) that corresponds to a peak energy density of on the skin surface of the mouse.
Fig. 1.
(a) Schematic of full-view light-emitting diode (LED)-based optoacoustic tomography (FLOAT) system for in vivo imaging of mice. Individual sensing elements of the ring array are shown with cylindrical lines on the inside of the surface of the array transducer. DAQ, data acquisition unit; OA, optoacoustic; PC, personal computer. (b) Photograph of circular full-ring array transducer together with LED arrays augmented to the bottom of the array transducer. LED arrays on the top side of the array transducer are not shown. (c) Left: photograph of a single LED array with 10 LEDs connected in series, right: circuit diagram. (d) Photograph of the single LED array with a heat sink attached on the back.
The circular full-ring ultrasound transducer had 512 elements, each measuring area, with an inter-element pitch of 0.47 mm, having a central frequency of 5 MHz and a nominal bandwidth of over 60%. The elements were distributed on two arcs, each covering a 174-deg angle, and focused at a distance of 38 mm in the elevational direction. Optoacoustic signals were digitized at 40 Mega samples per second using a custom DAQ (Falkenstein Mikrosysteme, Taufkirchen, Germany) and transmitted via Ethernet to a PC for storage and processing, controlled through a MATLAB (R2022a) interface.
The spatial resolution of the FLOAT system was characterized by scanning a cloud of spheres distributed randomly in a tissue-mimicking phantom [Fig. 2(a)]. Specifically, we used polyethylene microspheres ( diameter; Cospheric Inc., Santa Barbara, California, United States) embedded in a tissue-mimicking agar phantom (20-mm diameter). The phantom consisted of 1.3% agar powder by weight in water, 1.2% by volume of Intralipid®, and 0.002% by volume of black India ink, simulating a background absorption coefficient of and a reduced scattering coefficient of , typical for biological tissues at an excitation wavelength of 850 nm. The phantom was vertically scanned at 1-mm intervals along the -direction (RCP2-RGD6c-I-56 P-4-150-P1-S-B, IAI Inc., Shizuoka Prefecture, Japan) to capture different sphere distributions.
Fig. 2.
FLOAT system resolution characterization. (a) Photograph of the tissue-mimicking agar phantom (consisting of a mixture of Intralipid and Indian ink to mimic tissue background scattering and absorption) embedding randomly distributed absorbing cloud of spheres. (b) The reconstructed cross-sectional 2D OAT image from the sphere phantom. (c) System resolution characterization along the depth dimension. , radial resolution. , azimuthal resolution.
All in vivo animal experiments were conducted in compliance with the Swiss Federal Act on Animal Protection and were approved by the Cantonal Veterinary Office of Zurich. Female athymic nude-Foxn1nu mice were anesthetized with isoflurane (5% volume ratio for induction and 1.5% volume ratio during the experiments; Provet AG, Lyssach, Switzerland) in a mixture of oxygen and air (). The animals were secured in a stable position using a custom-made holder, with both fore and hind paws immobilized. A transparent thin membrane was employed to cover the nose and mouth of the mice, allowing them to breathe gas anesthesia freely, with the head kept above the water surface, while the body was fully submerged in water. The water temperature was maintained at 36°C throughout the experiments using a feedback-controlled heating stick. The mice were scanned in a stop-and-go manner, with 6-mm steps in the vertical direction, using the full-ring ultrasound transducer. The position of the transducer was controlled by motorized stages (RCP2-RGD6c-I-56P-4-150-P1-S-B, IAI Inc., Shizuoka Prefecture, Japan), which allowed translation in the vertical () direction.
Several signal-processing steps were applied before the image reconstruction process. First, the collected signals from the imaged object were averaged over 500 consecutive pulses. Subsequently, the averaged baseline signals, obtained when only water was present in the field of view, were subtracted from the averaged signals of the imaged object. A notch filter was then utilized to reduce the abnormal peak frequencies observed in the Fourier transform of the signal matrix. The filtered signal matrix was then used for the OAT image reconstruction, employing a model-based reconstruction algorithm with statistical weighting to minimize artifacts related to acoustic reflections and scattering.33
3. Results
The reconstructed cross-sectional image of the microspheres phantom for a section containing a large number of spheres is shown in Fig. 2(b). The spatial resolutions along the radial/axial direction () and azimuthal/tangential direction () were estimated at each radial position after deconvolving the actual microsphere diameter from the corresponding full width at half maximum (FWHM) in the image, which was calculated after fitting to Gaussian curves in both directions. The radial and azimuthal resolutions were computed from the mean resolution values of the five brightest spheres at each depth (, 1.5, 3, 4.5, and 6 mm) from the phantom surface [Fig. 2(c)]. The in-plane mean radial resolution and mean azimuthal resolution were in the range of 170 to . The resolution improves toward the center of the phantom attributing to the cylindrically focused detection of the ring array transducer. Note that the spheres are clearly discernible up to a depth of 6 mm, which defines the imaging depth across the entire FOV in 360 deg.
The system’s ability to perform whole-body imaging of mice was subsequently demonstrated (Fig. 3). The immobilized animal was scanned from the abdominal to the tail region [Fig. 3(a)]. The LED arrays together with the circular ring array transducer were translated in steps of 6 mm in direction along the length of the mouse torso. Three representative cross-sectional (2D) images reveal different anatomical features inside the mouse. Major organs such as the spleen, liver, left kidney, right kidney, spinal cord, and intestines, as well as some fine anatomical structures such as the surrounding vasculature (tail veins, femoral veins) and the skin outline, were clearly discernible [Fig. 3(b)].
Fig. 3.
Whole-body in vivo imaging of mice with FLOAT: (a) Different cross-sectional planes at which the mouse was scanned. (b) The reconstructed 2D images. SC, spinal cord; SK, skin; SL, spleen; LVL, liver; RK, right kidney; LK, left kidney; IN, intestine; BV, blood vessel; TV, tail vein; FV, femoral vein.
4. Discussion and Conclusion
Overall, our system provides a safe and compact alternative with a cost of around $1 to 2 k, significantly cheaper than both the traditional Nd: YAG-based laser sources, whose cost often exceeds $100 k, as well as the laser-diode-based solutions. Note that this cost includes only the LED arrays together with the pulse driver modules and does not include the ring array, DAQ, and other receiving electronics. However, LEDs have lower peak power, making them less efficient for optoacoustic signal generation. Yet, the FLOAT system compensates for this by placing LED arrays very close to the imaged object, thus generating sufficient energy density for deep tissue imaging. Previous LED-based systems using linear arrays suffered from severe limited-view artifacts resulting in impaired image quality.34 Instead, FLOAT circumvents the limited-view problem with a custom full-ring ultrasound array and 360-deg light illumination, thus facilitating optimal acquisition of tomographic image data.
Mice were imaged away from the heart below the thorax region to minimize motion artifacts. FLOAT was able to render high-resolution cross-sectional images down to up to a depth of . Note that, we have averaged 500 frames at a pulse repetition rate of 50 Hz, acquiring each cross-sectional data within a 10-s duration, which corresponds to an acquisition frame rate of 0.1 Hz for 2D in vivo imaging. A sequence of dedicated signal and image processing steps was followed to mitigate the noise- and motion-related artifacts in the images. The image quality can further be improved by increasing the repetition rate to allow for faster scan times, which in turn reduces motion artifacts and minimizes blurring in the reconstructed images.
Future advancements are expected to include multi-spectral imaging capabilities using LEDs of different wavelengths to assess tissue functional parameters, such as blood oxygen saturation, contrast agent perfusion, and targeting. Other applications beyond small animal imaging can also be envisioned, such as the detection of synovial angiogenesis in the finger joints, a key marker of early rheumatoid arthritis.35 In addition, the circular-ring array transducer can support both transmission and reflection ultrasound imaging, offering complementary data on the elastic and functional properties of finger joints.36,37 The current FLOAT system’s limited SNR may be improved through advanced reconstruction and processing techniques. For instance, deep-learning methods38 could help reduce ring artifacts caused by electrical noise, whereas model-based least squares (LS) minimization techniques could further enhance in vivo image quality and mitigate artifacts.39
In summary, FLOAT opens up new venues for preclinical OAT studies in resource-limited settings, such as monitoring pharmacokinetics and biodistribution of molecular agents and drugs on a whole-body level.
Acknowledgments
The authors acknowledge funding support from ETH Zurich (Grant No. ETH-C-01 21-2).
Funding Statement
The authors acknowledge funding support from ETH Zurich (Grant No. ETH-C-01 21-2).
Contributor Information
Sandeep Kumar Kalva, Email: sandeep.kalva@iitb.ac.in.
Xiang Liu, Email: liuxian@student.ethz.ch.
Xosé Luís Deán-Ben, Email: xl.deanben@pharma.uzh.ch.
Lin Tang, Email: tangli@student.ethz.ch.
Daniel Razansky, Email: daniel.razansky@uzh.ch.
Disclosures
The authors declare that there are no financial interests, commercial affiliations, or other potential conflicts of interest that could have influenced the objectivity of this research or the writing of this paper.
Code and Data Availability
The raw dataset used for this study is available upon request to the corresponding author.
References
- 1.Baker M., “The whole picture,” Nature 463, 977–979 (2010). 10.1038/463977a [DOI] [PubMed] [Google Scholar]
- 2.Kiessling F., Pichler B. J., Small Animal Imaging: Basics and Practical Guide, Springer Science & Business Media; (2010). [Google Scholar]
- 3.van der Heyden B., et al. , “Virtual monoenergetic micro-CT imaging in mice with artificial intelligence,” Sci. Rep. 12, 2324 (2022). 10.1038/s41598-022-06172-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Coudrillier B., et al. , “Phase-contrast micro-computed tomography measurements of the intraocular pressure-induced deformation of the porcine lamina cribrosa,” IEEE Trans. Med. Imaging 35, 988–999 (2016). 10.1109/TMI.2015.2504440 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Qin R., et al. , “Carbonized paramagnetic complexes of Mn (II) as contrast agents for precise magnetic resonance imaging of sub-millimeter-sized orthotopic tumors,” Nat. Commun. 13, 1938 (2022). 10.1038/s41467-022-29586-w [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Zhan S., et al. , “Targeting NQO1/GPX4-mediated ferroptosis by plumbagin suppresses in vitro and in vivo glioma growth,” Br. J. Cancer 127, 364–376 (2022). 10.1038/s41416-022-01800-y [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Kim D.-Y., et al. , “In vivo imaging of invasive aspergillosis with 18F-fluorodeoxysorbitol positron emission tomography,” Nat. Commun. 13, 1926 (2022). 10.1038/s41467-022-29553-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Li D., et al. , “SARS-CoV-2 receptor binding domain radio-probe: a non-invasive approach for angiotensin-converting enzyme 2 mapping in mice,” Acta Pharmacol. Sin. 43, 1749–1757 (2022). 10.1038/s41401-021-00809-y [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Heiles B., et al. , “Performance benchmarking of microbubble-localization algorithms for ultrasound localization microscopy,” Nat. Biomed. Eng. 6, 605–616 (2022). 10.1038/s41551-021-00824-8 [DOI] [PubMed] [Google Scholar]
- 10.Zhang Y., et al. , “Augmented ultrasonography with implanted CMOS electronic motes,” Nat. Commun. 13, 3521 (2022). 10.1038/s41467-022-31166-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Enninful A., Baysoy A., Fan R., “Unmixing for ultra-high-plex fluorescence imaging,” Nat. Commun. 13, 3473 (2022). 10.1038/s41467-022-31110-z [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Glaser A. K., et al. , “A hybrid open-top light-sheet microscope for versatile multi-scale imaging of cleared tissues,” Nat. Methods 19, 613–619 (2022). 10.1038/s41592-022-01468-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Bouchard M. B., et al. , “Swept confocally-aligned planar excitation (SCAPE) microscopy for high-speed volumetric imaging of behaving organisms,” Nat. Photonics 9, 113–119 (2015). 10.1038/nphoton.2014.323 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Santos‐Coquillat A., et al. , “Goat milk exosomes as natural nanoparticles for detecting inflammatory processes by optical imaging,” Small 18, 2105421 (2022). 10.1002/smll.202105421 [DOI] [PubMed] [Google Scholar]
- 15.Beard P., “Biomedical photoacoustic imaging,” Interf. Focus 1, 602–631 (2011). 10.1098/rsfs.2011.0028 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Manohar S., Razansky D., “Photoacoustics: a historical review,” Adv. Opt. Photonics 8, 586–617 (2016). 10.1364/AOP.8.000586 [DOI] [Google Scholar]
- 17.Deán-Ben X. L., et al. , “Advanced optoacoustic methods for multiscale imaging of in vivo dynamics,” Chem. Soc. Rev. 46, 2158–2198 (2017). 10.1039/C6CS00765A [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Özbek A., Deán-Ben X. L., Razansky D., “Optoacoustic imaging at kilohertz volumetric frame rates,” Optica 5, 857–863 (2018). 10.1364/OPTICA.5.000857 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Xu M., Wang L. V., “Photoacoustic imaging in biomedicine,” Rev. Sci. Instrum. 77, 41101 (2006). 10.1063/1.2195024 [DOI] [Google Scholar]
- 20.Kalva S. K., et al. , “Whole-body visualization of nanoagent kinetics in mice with flash scanning volumetric optoacoustic tomography,” Proc. SPIE 11642, 116420A (2021). 10.1117/12.2577844 [DOI] [Google Scholar]
- 21.Lee C., et al. , “Three-dimensional photoacoustic/ultrasound handheld scanner for clinical translation,” Proc. SPIE 11240, 112400K (2020). 10.1117/12.2544021 [DOI] [Google Scholar]
- 22.Dantuma M., et al. , “Fully three-dimensional sound speed-corrected multi-wavelength photoacoustic breast tomography,” arXiv:2308.06754 (2023).
- 23.Kalva S. K., Upputuri P. K., Pramanik M., “High-speed, low-cost, pulsed-laser-diode-based second-generation desktop photoacoustic tomography system,” Opt. Lett. 44, 81–84 (2019). 10.1364/OL.44.000081 [DOI] [PubMed] [Google Scholar]
- 24.Allen J. S., Beard P., “Pulsed near-infrared laser diode excitation system for biomedical photoacoustic imaging,” Opt. Lett. 31, 3462–3464 (2006). 10.1364/OL.31.003462 [DOI] [PubMed] [Google Scholar]
- 25.Kalva S. K., Pramanik M., Photoacoustic Tomography with High Energy Pulsed Laser Diodes, SPIE Press, Bellingham, Washington: (2020). [Google Scholar]
- 26.Allen T. J., Beard P. C., “Light emitting diodes as an excitation source for biomedical photoacoustics,” Proc. SPIE 8581, 85811F (2013). 10.1117/12.2004471 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Dai X., Yang H., Jiang H., “In vivo photoacoustic imaging of vasculature with a low-cost miniature light emitting diode excitation,” Opt. Lett. 42, 1456–1459 (2017). 10.1364/OL.42.001456 [DOI] [PubMed] [Google Scholar]
- 28.Hariri A., et al. , “The characterization of an economic and portable LED-based photoacoustic imaging system to facilitate molecular imaging,” Photoacoustics 9, 10–20 (2018). 10.1016/j.pacs.2017.11.001 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Erfanzadeh M., Zhu Q., “Photoacoustic imaging with low-cost sources; a review,” Photoacoustics 14, 1–11 (2019). 10.1016/j.pacs.2019.01.004 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Singh M. K. A., et al. , “Point-of-care functional and molecular imaging using LED-based photoacoustics,” in TENCON 2019–2019 IEEE Region 10 Conf. (TENCON), IEEE, pp. 109–113 (2019). [Google Scholar]
- 31.Xia W., et al. , “Handheld real-time LED-based photoacoustic and ultrasound imaging system for accurate visualization of clinical metal needles and superficial vasculature to guide minimally invasive procedures,” Sensors 18, 1394 (2018). 10.3390/s18051394 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Liu X., et al. , “Full-view LED-based optoacoustic tomography,” Photoacoustics 31, 100521 (2023). 10.1016/j.pacs.2023.100521 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Deán-Ben X. L., Razansky D., Ntziachristos V., “Statistical weighting of model-based optoacoustic reconstruction for minimizing artefacts caused by strong acoustic mismatch,” Proc. SPIE 7899, 789930 (2011). 10.1117/12.874623 [DOI] [Google Scholar]
- 34.Deán-Ben X. L., Razansky D., “On the link between the speckle free nature of optoacoustics and visibility of structures in limited-view tomography,” Photoacoustics 4, 133–140 (2016). 10.1016/j.pacs.2016.10.001 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Walsh D. A., “Angiogenesis and arthritis,” Rheumatology 38, 103–112 (1999). 10.1093/rheumatology/38.2.103 [DOI] [PubMed] [Google Scholar]
- 36.Merčep E., et al. , “Transmission–reflection optoacoustic ultrasound (TROPUS) computed tomography of small animals,” Light Sci. Appl. 8, 18 (2019). 10.1038/s41377-019-0130-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Lafci B., et al. , “Noninvasive multiparametric characterization of mammary tumors with transmission-reflection optoacoustic ultrasound,” Neoplasia 22, 770–777 (2020). 10.1016/j.neo.2020.10.008 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Dehner C., et al. , “Deep-learning-based electrical noise removal enables high spectral optoacoustic contrast in deep tissue,” IEEE Trans. Med. Imaging 41, 3182–3193 (2022). 10.1109/TMI.2022.3180115 [DOI] [PubMed] [Google Scholar]
- 39.Dean-Ben X. L., Razansky D., “A practical guide for model-based reconstruction in optoacoustic imaging,” Front. Phys. 10, 1028258 (2022). 10.3389/fphy.2022.1028258 [DOI] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The raw dataset used for this study is available upon request to the corresponding author.