Abstract
Purpose
To evaluate the feasibility of interleaved 23Na/1H cardiac MRI at 7 T using 1H parallel transmission (pTx) pulses.
Methods
A combined setup consisting of a 23Na volume coil and two 1H transceiver arrays was employed and the transmit and receive characteristics were compared in vitro with those of the uncombined radiofrequency coils. Furthermore, the implemented interleaved 23Na/1H pTx sequence was validated in phantom measurements and applied to four healthy subjects. For the latter, three customized 1H excitation pulses (universal and individual phase shims (UPS/IPS) and individual 4kT pulses (4kT)) were employed in the interleaved 23Na/1H pTx sequence and compared with the vendor‐provided default cardiac phase shim (DPS).
Results
Combining both coils resulted in a reduction of the mean 23Na transmit field (B1 +) efficiency and 23Na signal‐to‐noise ratio by 18.9% and 15.4% for the combined setup, whereas the 1H B1 + efficiency was less influenced (−4.7%). Compared with single‐nuclear acquisitions, interleaved dual‐nuclear 23Na/1H MRI showed negligible influence on 23Na and 1H image quality. For all three customized 1H pTx pulses the B1 + homogeneity was improved (coefficients of variation: CVUPS = 0.30, CVIPS = 0.23, CV4kT = 0.15) and no 1H signal dropouts occurred compared with the vendor‐provided default phase shim (CVDPS = 0.37).
Conclusion
The incorporation of customized 1H pTx pulses in an interleaved 23Na/1H sequence scheme was successfully demonstrated at 7 T and improvements of the 1H B1 + homogeneity within the heart were shown. Combining interleaved 23Na/1H MRI with 1H pTx is an important tool to enable robust quantification of myocardial tissue sodium concentrations at 7 T within clinically acceptable acquisition times.
Keywords: 23Na RF birdcage coil, 7 Tesla, cardiac sodium (23Na) MRI, interleaved acquisition, parallel transmission (pTx)
1. INTRODUCTION
Sodium ions (Na+) play an essential role in many physiological processes and the Na+ concentration is closely linked to cell viability. 1 For example, changes of the myocardial tissue sodium concentration (TSC) were observed in patients with myocardial infarction 2 as well as primary hyperaldosteronism. 3 Compared with interventional methods like myocardial biopsy, sodium (23Na) MRI currently represents the only technique to noninvasively determine the myocardial TSC in vivo. 4
Compared to hydrogen (1H), 23Na shows significantly lower MR sensitivity and in vivo concentrations which, lead to a strongly reduced signal‐to‐noise ratio (SNR) for 23Na MRI (1:6000 in myocardium 4 ). To compensate for this SNR deficit, larger voxel sizes and longer acquisition times are needed for 23Na MRI. 4 In addition, ultrahigh field strengths are beneficial for 23Na MRI to further increase the SNR. 5 , 6
Due to the low spatial resolution of 23Na MRI, additional high‐resolution anatomical 1H MR data are required to enable a precise segmentation of different cardiac compartments (e.g., blood pool, myocardium) for partial volume correction 7 and quantitative evaluation of the 23Na images. 8 Previously, cardiac 23Na and 1H MRI were performed consecutively with different pulse sequences 4 , 9 , 10 or even at different field strengths, 8 which prolonged the total acquisition time. Additionally, the consecutive acquisition scheme is prone to changes in the physiological conditions of the subject (heart rate, respiratory amplitude and frequency) as well as prone to image coregistration errors.
Thus, in this work we used a dual‐nuclear interleaved 23Na/1H acquisition scheme 11 , 12 to receive both 23Na and 1H MR data within a single measurement at 7 T. For this, we used a combined coil setup consisting of a 23Na volume torso coil and two 1H torso arrays. This setup in combination with the interleaved 23Na/1H sequence allows to use the idle time of the 23Na acquisition to acquire additional anatomical 1H images. This reduces the total acquisition time, ensures alignment of the 23Na and 1H images without requiring image coregistration and guarantees that both images depict the quasi‐same physiological states. Simultaneous instead of interleaved 23Na/1H MR acquisition schemes, as demonstrated for 23Na/1H brain MRI, 13 provide similar advantages, but would have required hardware modifications of the MR scanner used.
At 7 T, the excitation wavelength for 1H MRI is relatively short (11 cm in water 14 ). This can lead to an inhomogeneous 1H excitation profile and even complete dropouts of the 1H signal within the heart, 15 making a reliable segmentation difficult. To avoid this, we combined the interleaved 23Na/1H acquisition with the concept of multichannel parallel transmission (pTx) 16 , 17 for 1H excitation.
In the following, we investigated the influences of the combined coil setup and the interleaved 23Na/1H sequence. Subsequently, we applied the interleaved 23Na/1H sequence with different customized pTx pulses in four healthy subjects.
2. METHODS
2.1. Data acquisition and reconstruction
All measurements were conducted on a 7T whole‐body system (Terra.X, Siemens Healthineers, Erlangen, Germany) using 23Na and 1H torso coils 18 (Rapid Biomedical, Rimpar, Germany). For 23Na MRI, a volume coil was used, constructed as a four‐rung birdcage with asymmetric end rings adjusted to the magnet bore to provide maximum space for the subject. 18 The 1H setup 15 consists of two arrays, each with four transmit and eight receive channels, which are placed below the back and on top of the chest of the subject. Specific absorption rate (SAR) limits of the 1H arrays were met by limiting the time‐averaged radiofrequency (RF) power of each transmit channel to 1.195 W. 15 For the 23Na coil, weight‐dependent ( kg) SAR limits were used (1.05/2.1 W/kg times the minimum weight of each weight category for normal/first level operation mode). In addition, the component protection limits the maximum power of the 23Na coil to 120 W for the combined and 140 W for the uncombined setup. In addition to the individual use, both coils are also specifically designed to be operated in a combined setup (Figure 1A). Based on extensive testing of the coils (i.e., tuning, matching, and interactions between the coils at both the 23Na and 1H frequency), a declaration of safety and compatibility of the coil setup was provided by the manufacturer, including the setting of safe SAR limits for both the combined and uncombined coil setups.
FIGURE 1.

The combined radiofrequency coil setup (A) consists of a 23Na volume RF coil and two 1H arrays, each with four transmit (Tx) and eight receive (Rx) channels. The 1H arrays are positioned on top of the chest and below the back of the subject, and the multiple 1H Tx channels enable the use of parallel transmission (pTx) for the 1H excitation. In the interleaved 23Na/1H pTx sequence scheme (B), the acquisition of a 23Na projection is followed by the acquisition of four 1H projections during the idle time of the 23Na repetition time (TR). The excitation of 23Na is performed in 1Tx mode, while customized pTx pulses can be applied for the excitation of 1H.
For in vitro measurements, a torso‐like phantom (Na+ concentration 300 mM) was used mimicking the dielectric properties of the human torso. 19 In vivo measurements were carried out on four healthy subjects (28 3 years). All measurements were approved by the local ethics committee and each subject provided signed informed consent before the examination.
All MR data were acquired during free breathing and without cardiac triggering using a density‐adapted 3D radial readout 20 and a golden angle 21 projection scheme (Data S1). The reconstruction of all data was performed offline using a custom‐written MATLAB (The MathWorks, Natick, MA, USA) script. After application of a Hamming filter, all images were reconstructed using a nonuniform fast Fourier transform. 22 In addition, the interleaved acquired 23Na images were interpolated to match the nominal spatial resolution of the interleaved acquired 1H images.
2.2. Interleaved 23Na/ 1H pTx sequence with customized 1H pTx pulses
In the dual‐nuclear interleaved 23Na/1H pTx sequence scheme (Figure 1B), one 23Na projection is followed by four 1H projections evenly distributed over the idle time of the 23Na repetition time. 12 For 23Na MRI, we used a center‐out (half projection) radial readout enabling very short echo times to reduce signal loss resulting from the fast T2* decay of 23Na nuclei. Because T2* decay is slower for 1H, 1H projections were acquired in a center‐through (full projection) radial readout to ensure more efficient k‐space sampling. The corresponding sequence parameters of the interleaved 23Na/1H sequence can be found in Table 1. The B0 compensation of the scanner had to be turned off manually to avoid an image corruption of the interleaved acquired 23Na image (Figure S1).
TABLE 1.
Acquisition parameters of the used MR sequences. In vitro absolute B1 + mapping was performed using the double‐angle (DA) method for 23Na and the actual flip angle imaging (AFI) method for 1H, respectively. In vivo relative channel‐wise three‐dimensional 1H B1 + maps were necessary for subject‐specific pTx pulse calculations (individual phase shim and individual 4kT‐points pulse). For the phase shims (default phase shim universal phase shim and individual phase shim), we used a 1H pulse duration of 2 ms, whereas for the individual 4kT‐points pulse the 1H pulse durations were 0.87 ms (Subjects 1–3) and 1.27 ms (Subject 4). For the interleaved 23Na/1H sequence, the excitation and readout parameters can be chosen independently for both nuclei. All in vivo measurements were performed during free breathing and without cardiac triggering.
| Interleaved 23Na/1H | 23Na DA | 1H AFI | Rel. 1H B1 + | ||
|---|---|---|---|---|---|
| Application | In vitro/in vivo | In vitro | In vivo | ||
| Nucleus | 23Na | 1H | 23Na | 1H | 1H |
| Readout scheme | DA‐3D‐RAD (half proj.) | DA‐3D‐RAD (full proj.) | DA‐3D‐RAD (half proj.) | DA‐3D‐RAD (full proj.) | DA‐3D‐RAD (full proj.) |
| Nominal spatial resolution [mm3] | 6 × 6 × 6 | 2 × 2 × 2 | 12 × 12 × 12 | 4 × 4 × 4 | 4 × 4 × 4 |
| Nominal FA [°] | 82 | 10 | 45/90 | 70 | 8 |
| Projections | 15 000 | 60 000 | 4000 | 7500 | 10 000 |
| TR [ms] | 60 | TR1H,a = 13.08, TR1H,b = 20.76 | 250 | TR1 = 75, TR2 = 15 | 4.5 |
| TE [ms] | 1.15 | 2.5 | 1.15 | 3.03 | 2.02 |
| Tpulse [ms] | 2 | DPS/UPS/IPS: 2, 4kT: 0.87/1.27 | 2 | 1 | 0.5 |
| TRO [ms] | 5 | 2 | 5 | 2.5 | 2.5 |
| Acquisition time (min:s) | 15:00 | 2× 16:40 | 11:15 | 6:00 | |
Abbreviations: AFI, actual flip angle imaging; DA, double angle method; DA‐3D‐RAD, density‐adapted 3D radial; FA, flip angle; TE, echo time; TR, repetition time; Tpulse, transmit pulse duration; TRO, readout duration.
The employed interleaved sequence allows the application of pTx for the 1H excitation of the sequence. In this work, we used four different 1H excitation pulses 23 :
Default phase shim (DPS): vendor‐provided fixed phase shim for cardiac MRI
Universal phase shim (UPS): optimization of channel phases based on previously acquired cardiac B1 + maps of 35 volunteers 15
Individual phase shim (IPS): optimization of channel phases based on additionally acquired B1 + maps of the measured subject 24
Individual 4kT‐points pulse (4kT): optimization of amplitudes/phases of four sub‐pulses and intervening gradients based on additionally acquired B1 + maps of the measured subject 25
The individual pulse optimization and the calculation of absolute flip angle (FA) maps were performed as described in Egger et al. 15 The additional acquisition of B1 + maps and the computation of the individually optimized pTx pulses (IPS, 4kT) required about 10 min. A linear calibration fit based on 35 previously measured subjects (Figure S2) was employed for a subject‐specific estimation of the reference voltage based on the weight of the subject. The resulting reference voltage was used for all three phase shims.
2.3. B1 + and B1 − mapping methods
For the in vitro FA measurements of the 23Na volume coil, the double angle method 26 was used (Table 1). The receive profile within the phantom was calculated based on the following formula 8 :
| (1) |
with the measured signal . For the 1H arrays, in vitro absolute FA maps were acquired using an actual flip angle imaging 27 approach 15 with the sequence parameters given in Table 1. The efficiency of the transmit field (B1 +) was calculated for both coils according to
| (2) |
with the applied pulse voltage . For the calculation of subject‐specific optimized 1H pTx pulses, additional relative channel‐wise three‐dimensional B1 + maps 28 were acquired with a nominal spatial resolution of (4 mm)3 and an acquisition time of 6 min (see remaining sequence parameters in Table 1) as described by Egger et al. 15 FA simulations were performed based on the measured channel‐wise three‐dimensional B1 + maps to calculate FA estimates for all four 1H excitation pulses. 15 For better comparison of the homogeneity, the FA maps were normalized to the mean FA in the heart. 29
2.4. Influence of combined setup
To quantify the mutual influence of the 23Na volume coil and the 1H arrays, 23Na and 1H B1 + efficiency maps and 23Na B1 − maps were measured in vitro for both the combined as well as uncombined coil setup. To assess the B1 + and B1 − homogeneity of the 23Na volume coil, the coefficient of variation (CV) was calculated within a central region of interest (132 × 228 × 204 mm3). Furthermore, voxelwise 23Na SNR maps were calculated based on the 23Na images, acquired with a nominal FA of 90°, and the standard deviation of additional noise scans within the phantom as follows 30 :
To compare the performance of the combined to the uncombined setup, relative difference maps were calculated for the different coil properties using the formula
and the mean was computed over the whole phantom. External markers were used to ensure reproducible positioning of the phantom for both setups. In addition, the images of both setups were coregistered in the postprocessing using the Elastix toolbox. 31
2.5. Influence of interleaved sequence
To investigate the influence of the interleaved 23Na/1H acquisition on the image quality of both nuclei, 23Na and 1H in vitro images acquired with the dual‐nuclear interleaved and single‐nuclear sequence were compared. For the single‐nuclear 23Na sequence, the 1H transmission and gradients in Figure 1B were switched off and vice versa for the single‐nuclear 1H sequence. Moreover, the mean 23Na SNR values were calculated for both sequences based on the corresponding images and additional noise scans as described above. 30
3. RESULTS
3.1. Influence of the combined 23Na and 1H torso coil setup
Figure 2 shows the mutual influence of the 23Na and 1H coils. Compared with the uncombined application of the 23Na volume coil, the mean 23Na B1 + efficiency was reduced by about 19% for the combined setup and especially dropped off toward the top left. Evaluation within a central region of interest displayed a higher CV and thus a decrease in 23Na B1 + homogeneity for the combined setup (CV = 0.15 vs. 0.09), whereas 23Na B1 − homogeneity remained similar for both setups (CV = 0.20 vs. 0.21). For the combined setup, the 23Na noise level was elevated by 24% compared with the uncombined coil setup, such that the 23Na SNR showed a mean reduction of 15.4%. The additional 23Na body coil had only a minor effect on the 1H B1 + efficiency (−4.7%) and showed no relevant influence on customized 1H pTx pulses (Figure S3).
FIGURE 2.

Comparison of the combined (first row) and uncombined (second row) application of the 23Na and 1H coil setups. To evaluate the B1 + and B1 − homogeneity of the 23Na volume coil, the coefficient of variation (CV) values were calculated within a central region of interest (volume of 132×228×204 mm3; white dotted box). Using the combined setup (CVB1+, comb = 0.15), the 23Na B1 + (A) homogeneity was lower than for the uncombined case (CVB1+, uncomb = 0.09), while homogeneity of the 23Na B1 − (B) distribution was less affected (CV B1‐,comb = 0.20 vs CV B1‐,uncomb = 0.21). The relative difference (third row) for each variable X was calculated by , and the mean difference was computed within the whole phantom. The mean 23Na B1 + efficiency (A) and 23Na SNR (C) were reduced by 18.9% and 15.4% for the combined setup, respectively. In contrast, 1H B1 + efficiency (D) was less influenced by the combination of both coils.
3.2. Influence of the dual‐nuclear interleaved 23Na/ 1H acquisition
Figure 3 depicts the influence of the dual‐nuclear interleaved 23Na/1H acquisition on both images compared with the corresponding single‐nuclear acquisitions. For both 23Na and 1H images, no relevant differences were found. The mean relative signal difference over the whole phantom was for 23Na and for 1H. The SNR of the 23Na image was not affected by the additional acquisition of 1H MR data in the dual‐nuclear interleaved sequence.
FIGURE 3.

Comparison of 23Na and 1H images acquired with the dual‐nuclear 23Na/1H interleaved (A) and single‐nuclear sequences (B) using the combined 23Na/1H coil setup. For the single‐nuclear 23Na sequence, the excitation and gradients for 1H were turned off in the interleaved sequence and vice versa for the single‐nuclear 1H sequence. For 1H excitation, the vendor‐provided default phase shim was used. (C) The relative signal difference was obtained by , and the mean relative signal difference was calculated over the whole phantom. For both nuclei, no relevant differences were visible.
3.3. In vivo application of the interleaved 23Na/ 1H pTx sequence
Figure 4 shows an exemplary in vivo application of the interleaved 23Na/1H pTx sequence for one subject. For the acquisition with the vendor‐provided DPS signal dropouts (red arrow) in the heart were visible in the 1H image. For all three customized pTx pulses (UPS, IPS and 4kT), no 1H signal dropouts occurred (first row), which is also shown in the simulated 1H FA map (third row). Averaged over all four measured subjects (Table S1), the UPS showed enhanced homogeneity with reduced CV (CV = ) compared with the DPS (CV = ). The IPS yielded further improvements (CV = ) and the best homogeneity was achieved with the 4kT pulses (CV = ). The animated overlay of the 23Na and 1H images in Figure S4 demonstrates that both images are aligned without the need for image coregistration.
FIGURE 4.

In vivo interleaved 23Na/1H pTx MR images (acquisition time of 15 min) of one healthy subject for different 1H excitation pulses during free breathing and without cardiac triggering. While for the vendor‐provided DPS (A) 1H signal dropouts (red arrow) were visible within the heart, all three customized pTx pulses (B–D) showed no 1H signal dropouts (first row). The improvements in 1H flip angle (FA) homogeneity are also demonstrated by the reduced coefficient of variation (CV) values calculated within the heart volume (red line) of the corresponding simulated 1H FA maps (third row). For better comparison of the homogeneity, the simulated 1H FA maps were normalized to the mean 1H FA within the heart. 23Na images were unaffected by the choice of the 1H excitation pulse (second row). DPS, default phase shim; IPS, individual phase shim; UPS, universal phase shim; 4kT, 4kT‐points pulse.
4. DISCUSSION
In this work, we demonstrated the feasibility of interleaved 23Na/1H pTx cardiac MRI at 7 T. Apart from interleaving 1H MRI with X‐nuclei MR spectroscopy, 11 , 32 interleaved X‐nuclei/1H MRI in general has only been applied to a few different regions of the human body 11 such as knee, 33 , 34 lung, 35 , 36 breast 37 and brain. 12 , 29 , 38 However, an interleaved X‐nuclei/1H MRI acquisition is especially beneficial for cardiac MRI. The latter is challenged by image artifacts caused by respiratory and cardiac motion. Thus, physiological changes, such as from shallow to heavy breathing, between consecutive 23Na and 1H cardiac measurements may falsify the TSC quantification. Furthermore, for 23Na and 1H cardiac MRI at different field strengths, 8 errors in the image coregistration, which has to be performed to correct for the different positioning of the subject, might also affect the TSC quantification. Using an interleaved 23Na/1H acquisition scheme, these described error sources can be avoided, as 23Na and 1H images are acquired within one measurement. Another main advantage of the interleaved acquisition is the reduced total scan time, which can be particularly important for patient studies. In summary, interleaved 23Na/1H cardiac MRI will be an improvement toward a fast and precise quantification of the myocardial TSC.
However, cardiac 1H MRI at 7 T is challenging due to the B1 + inhomogeneity at the high field strength. 39 So far, interleaved X‐nuclei/1H MRI has only been performed with 1Tx and capabilities of pTx were not exploited. 11 Thus, we incorporated, to our knowledge for the first time, the concept of 1H pTx in an interleaved 23Na/1H sequence at 7 T and included different customized 1H pTx pulses. The implemented interleaved sequence allows for the application of individually or universally optimized 1H phase shims as well as more sophisticated dynamic kT‐points pTx pulses, additionally optimizing the pulse amplitudes and/or gradients between subpulses, depending on the target application. For fast applications, a UPS or even a universal kT‐points pulse 25 would be advisable. If additional 1H B1 + mapping, which could be accelerated 15 , 40 in the future, and pulse calculation are feasible in terms of time, one could use individually optimized pTx pulses. As confirmed in this work, these usually outperform universal pulses. 25 , 41 Compared with the vendor‐provided DPS, no 1H signal dropouts occurred for all three customized 1H pTx pulses used in this work. This ensures a precise segmentation of different cardiac compartments in the 1H images, which is necessary for a reliable quantification of the myocardial TSC. Furthermore, not being limited to a vendor‐provided fixed cardiac phase shim facilitates the transfer of the interleaved 23Na/1H acquisition to other abdominal applications by using customized 1H pTx pulses optimized for the specific target region.
Compared with the uncombined application of the 23Na volume coil, for the combined setup the B1 + efficiency was reduced by an average of 19%, whereas the B1 − sensitivity increased by about 8%. Even the uncombined 23Na coil showed different B1 + and B1 − field distributions, probably due to an asymmetry between the horizontal and vertical (linear) modes of the volume coil, which may be caused, for example by the different proximity to the bore. The additional 1H arrays in the combined setup might further increase this asymmetry, leading to the observed deviations of the B1 + and B1 − field distributions compared with the uncombined setup. The SNR reduction of 15% for the combined coil setup arises primarily from a higher noise level for the combined setup. In contrast to the 23Na coil, the B1 + efficiency of the 1H arrays was less influenced by the combination of both coils, as the very large 23Na coil appears to act more like a ground plane for the 1H arrays. In the work of Lott et al., 8 the 23Na B1 + and B1 − correction had only a minor effect on the quantification of the myocardial TSC. However, for our combined setup (CVB1+ = 0.15, CVB1‐ = 0.20), the 23Na B1 + and B1 − variations may have a greater impact on quantification than for their coil setup (CVB1+ = 0.07, CVB1‐ = 0.08 19 ). Thus, additional acquisition of 23Na B1 + maps might be necessary to correct for B1 + variations in quantitative TSC evaluations. 8 However, this would not be possible for B1 − maps, as these can only be acquired in a sample with homogeneous sodium concentration. In the future, a B1 − correction based on electromagnetic field simulations 8 could be promising. Furthermore, an additional dielectric pad 42 might reduce the drop‐off of the 23Na B1 + and B1 − field in the upper left of the phantom. Regarding the coil setup, the used 23Na volume coil should provide a more homogeneous B1 + and B1 − field compared with multichannel 23Na arrays. In contrast, the latter would result in an increased 23Na SNR as demonstrated by Kaggie et al. 43 Thus, our combined coil setup offers the option to incorporate an additional receive‐only 23Na array in the future, which could potentially be integrated directly in the anterior 1H array. Apart from that, dual‐tuned 23Na/1H transceiver arrays might be an interesting alternative. 44
Despite the additional gradients of the dual‐nuclear sequence, no relevant differences were observed in neither the 23Na nor 1H images compared with the images acquired with the single‐nuclear sequences, which agrees with previous publications. 12 , 33 The SNR of the 23Na images was also not affected by the interleaved acquisition. This is in accordance with previously published results of a similar interleaved 23Na/1H sequence scheme acquired with a dual‐tuned 23Na/1H head coil. 12 The Terra.X scanner version enables interleaved 23Na/1H MRI with customized 1H pTx pulses, unlike previous software versions, which only supported interleaved 23Na/1H MRI with 1H excitation in 1Tx mode. Apart from general X‐nuclei modifications like a broadband amplifier and dedicated coils, the interleaved 23Na/1H pTx sequence was applied without any additional hardware modifications that were required in previous studies for interleaved or simultaneous acquisitions. 13 , 45 However, due to software incompatibilities, similar to the incompatibility of the eddy current compensation for X‐nuclei described in McLean et al., 46 the vendor‐provided B0 compensation had to be turned off manually, but this should be fixed with the next software update.
One limitation of dual‐nuclear interleaved 23Na/1H acquisitions is the increased SAR due to the additional 1H excitation pulses compared with single‐nuclear 23Na acquisitions. This can be considered in the optimization of the 23Na pulse sequence parameters. 47 In addition, more sophisticated SAR models such as virtual observation points 48 might be used in the future. When considering local SAR, the interleaved 23Na/1H sequence can take advantage of the different spatial distributions of the SAR hotspots for each nucleus. The SAR would be determined by the maximum of the combined local SAR distribution, potentially resulting in a lower SAR than the cumulated SAR of both single‐nuclear sequences. 49
Another challenge in cardiac 23Na/1H MRI are artifacts caused by respiratory and cardiac motion. The acquisition of all data in free breathing and without cardiac triggering leads to errors in the quantification of the myocardial TSC. 8 As a potential solution, additional data points in the k‐space center could be acquired for each 23Na/1H projection. This may allow to apply retrospective self‐gating methods 19 to reconstruct respiratory and cardiac sorted 23Na/1H images. Due to the interleaved acquisition scheme, the 1H images might then be used for motion correction of the 23Na data 12 in the future.
5. CONCLUSION
In this work, we successfully demonstrated the application of an interleaved 23Na/1H MR acquisition with customized 1H pTx pulses for cardiac MRI at 7 T. Using customized 1H pTx pulses in the presented interleaved 23Na/1H pTx sequence enabled the time‐efficient acquisition of cardiac 23Na and 1H images within one measurement, while ensuring homogeneous 1H excitation of the heart.
CONFLICT OF INTEREST
One of the co‐authors (Titus Lanz) is an employee of Rapid Biomedical GmbH.
Supporting information
Data S1. Formulas to calculate the polar and azimutal angles and Cartesian unit vectors of the radial projections (n = 1, …, 15 000 for 23Na; and n = 1, …, 60 000 for 1H) based on the two‐dimensional (2D) golden means (, ) presented by Chan et al. 21
Figure S1. Influence of vendor‐provided B0 compensation on interleaved 23Na/1H MRI. In (A) and (B), measurements were performed under the B0 compensation currently provided by the vendor, whereas (C) was performed with a corrected software implementation of the B0 compensation at another Terra.X development device of the vendor. All measurements were performed using a 23Na resolution phantom and a dual‐tuned 23Na/1H head coil. The same interleaved sequence parameters as for the in vitro and in vivo cardiac measurements were used for (A) and (B), whereas for (C) the resolution of the 23Na images was (4 mm)3 instead of (6 mm)3.
(A) For 23Na MRI, we used a single‐nuclear 23Na sequence (I) only containing transmission, readouts, and gradients for 23Na as reference image. Acquiring a 23Na image using the dual‐nuclear interleaved 23Na/1H sequence with turned on B0 compensation (II) leads to a spatial shifting and blurring of the 23Na image (II‐I). By manually turning off the B0 compensation (III), these effects can be avoided, and there are no relevant differences between the 23Na image of the single‐nuclear and dual‐nuclear sequence (III‐I; mean difference over the phantom: 0.62% of the maximum value).
(B) For 1H MRI, B0 compensation worked for single‐nuclear and dual‐nuclear sequences (not shown here). Therefore, we used the interleaved acquired 1H image as reference (I). Because we had to turn off the B0 compensation for interleaved 23Na/1H acquisitions due to the artifacts for the 23Na image, interleaved acquired 1H images were not B0 compensated (II), resulting in minor spatial shifts visible in the difference image (II‐I). However, using image coregistration (III), these shifts could be corrected (III‐I; mean difference over the phantom: −0.02% of the maximum value). Because the effect of the B0 compensation is only determined by the shapes and timings of the applied gradients, it can be assumed that these shifts are constant for repeated measurements with the same gradient scheme. Even though the shifts were in the submillimeter range (), and thus almost negligible, we corrected the interleaved acquired 1H images for these shifts in our reconstruction.
After reporting the problem to the vendor, they identified errors in the software implementation of the B0 compensation that affect X‐nuclei MRI during dual‐nuclear interleaved acquisitions. (C) Test measurements of the interleaved sequence acquired under the latest software version with a corrected software implementation of the B0 compensation on another Terra.X development device of the vendor. There were no differences (II‐I) visible between single (I) and dual‐nuclear (II) measurements acquired with turned on B 0 compensation. As for (A), the differences between the single‐nuclear sequence with turned‐on B0 compensation and the dual‐nuclear interleaved sequence with turned‐off B0 compensation did not show relevant differences (III‐I). In summary, after the next software update of our scanner, the interleaved sequence should be applicable without turning off the B0 compensation. However, until then, we have to use the presented work‐around of manually turning off the B0 compensation for the interleaved measurements.
Figure S2. Linear calibration fit () between the calculated 1H reference voltages within the heart for the default phase shim (DPS) and the body weight of 35 previously measured subjects. 15 The linear fit showed good correlation between the reference voltage within the heart and the weight of the subjects (correlation factor: 0.76). The relative absolute deviation between the measured reference voltage and the corresponding reference voltage based on the calibration fit was calculated by . For the 35 subjects, the mean relative absolute deviation was 6.4% (minimal/maximal: 0.02%/14.5%). Because there are currently no inline adjustments available for 1H body imaging at 7 T, we used this calibration fit to estimate subject‐specific 1H reference voltages for the DPS, universal phase shim (UPS), and individual phase shim (IPS) based on the weight of the measured subjects. However, due to the increased B1 + efficiency of UPS and IPS, 15 an additional calibration fit for the UPS and direct calculation of the reference voltage during the optimization of the IPS could further improve the estimation of the flip angle (FA) in the future.
Figure S3. (A) Comparison of measured in vitro channel‐wise relative 1H B1 + maps for the combined and uncombined coil setup. The channel‐wise relative 1H B1 + maps of the combined setup were image‐coregistered to the uncombined setup, reducing misalignments of the phantom due to the repositioning. Absolute difference maps were calculated by , and the mean difference was calculated within an exemplary heart region (red line, Subject 1). Within the heart region, no relevant differences were visible. (B) Influence of coil setups on optimized 1H excitation pulses. Using the shown in vitro channel‐wise relative 1H B1 + maps in (A), 1H phase shims were optimized for each coil setup (combined, uncombined) within the same three‐dimensional (3D) heart region (red line, Subject 1) and then applied to the channel‐wise relative 1H B1 + maps of both setups. For individually designed pulses, the pulses are usually optimized during the measurement and thus applied on the same coil setup. Coefficient of variation (CV) values showed no relevant differences between 1H pulses designed and applied on the uncombined (I) and combined setup (IV). Even pulses optimized on relative B1 + maps of one of the coil setups and then applied to the other coil setup (II, III) yielded comparable CV values. Thus, for the performance of 1H pulses optimized within the heart region, the influence of the 23Na coil in the combined setup appears to be negligible.
Figure S4. Animated overlay of 23Na and 1H images (Subject 1) acquired with the interleaved 23Na/1H pTx sequence using the universal phase shims (UPS) for 1H excitation. Due to the interleaved 23Na/1H acquisition scheme, the corresponding 23Na and 1H images are aligned without the need for image coregistration.
Table S1. Overview of coefficient of variation (CV) values. The mean CV values were calculated within the heart volume based on the simulated 1H FA maps of each subject and are shown for the four different 1H excitation pulses (default phase shim (DPS), universal phase shim (UPS), individual phase shim (IPS) and individual 4kT pulses (4kT)) of each subject.
ACKNOWLEDGMENTS
We acknowledge support from the Deutsche Forschungsgemeinschaft (project 449552397). Open Access funding enabled and organized by Projekt DEAL.
Ruck L, Egger N, Wilferth T, et al. Interleaved 23 Na/ 1H MRI of the human heart at 7 T using a combined 23Na/ 1H coil setup and 1H parallel transmission. Magn Reson Med. 2025;94:231‐241. doi: 10.1002/mrm.30426
REFERENCES
- 1. Madelin G, Regatte RR. Biomedical applications of sodium MRI in vivo. J Magn Reson Imaging. 2013;38:511–529. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Ouwerkerk R, Bottomley PA, Solaiyappan M, et al. Tissue sodium concentration in myocardial infarction in humans: a quantitative 23Na MR imaging Study. Radiology. 2008;248(1):88–96. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Christa M, Weng AM, Geier B, et al. Increased myocardial sodium signal intensity in Conn's syndrome detected by 23Na magnetic resonance imaging. Eur Heart J Cardiovasc Imaging. 2019;20(3):263–270. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Bottomley PA. Sodium MRI in human heart: a review. NMR Biomed. 2016;29:187–196. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Le Ster C, Grant A, Van de Moortele PF, et al. Magnetic field strength dependent SNR gain at the center of a spherical phantom and up to 11. 7T. Magn Reson Med. 2022;88:2131–2138. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Pohmann R, Speck O, Scheffler K. Signal‐to‐noise ratio and MR tissue parameters in human brain imaging at 3, 7, and 9.4 tesla using current receive coil arrays. Magn Reson Med. 2016;75:801–809. [DOI] [PubMed] [Google Scholar]
- 7. Niesporek SC, Hoffmann SH, Berger MC, et al. Partial volume correction for in vivo 23Na‐MRI data of the human brain. Neuroimage. 2015;112:353–363. [DOI] [PubMed] [Google Scholar]
- 8. Lott J, Platt T, Niesporek SC, et al. Corrections of myocardial tissue sodium concentration measurements in human cardiac (23) Na MRI at 7 Tesla. Magn Reson Med. 2019;82(1):159–173. [DOI] [PubMed] [Google Scholar]
- 9. Sandstede JJ, Hillenbrand H, Beer M, et al. Time course of 23Na signal intensity after myocardial infarction in humans. Magn Reson Med. 2004;52:545–551. [DOI] [PubMed] [Google Scholar]
- 10. Jerecic R, Bock M, Nielles‐Vallespin S, Wacker C, Bauer W, Schad LR. ECG‐gated 23 Na‐MRI of the human heart using a 3D‐radial projection technique with ultra‐short echo times. Magma. 2004;16:297–302. [DOI] [PubMed] [Google Scholar]
- 11. Lopez Kolkovsky AL, Carlier PG, Marty B, Meyerspeer M. Interleaved and simultaneous multi‐nuclear magnetic resonance in vivo. Review of principles, applications and potential. NMR Biomed. 2022;35:e4735. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Wilferth T, Müller M, Gast LV, et al. Motion‐corrected 23Na MRI of the human brain using interleaved 1H 3D navigator images. Magn Reson Med. 2022;88(1):309–321. [DOI] [PubMed] [Google Scholar]
- 13. Yu Z, Madelin G, Sodickson DK, Cloos MA. Simultaneous proton magnetic resonance fingerprinting and sodium MRI. Magn Reson Med. 2020;83:2232–2242. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Schick F. Whole‐body MRI at high field: technical limits and clinical potential. Eur Radiol. 2005;15:946–959. [DOI] [PubMed] [Google Scholar]
- 15. Egger N, Nagelstraßer S, Wildenberg S, et al. Accelerated B1+ B_1^+ mapping and robust parallel transmit pulse design for heart and prostate imaging at 7 T. Magn Reson Med. 2024;92:1933–1951. [DOI] [PubMed] [Google Scholar]
- 16. Webb AG, Collins CM. Parallel transmit and receive technology in high‐field magnetic resonance neuroimaging. Int J Imaging Syst Technol. 2010;20:2–13. [Google Scholar]
- 17. Ibrahim TS, Lee R, Baertlein BA, Kangarlu A, Robitaille P‐ML. Application of finite difference time domain method for the design of birdcage RF head coils using multi‐port excitations. Magn Reson Med. 2000;18:733–742. [DOI] [PubMed] [Google Scholar]
- 18. Ruck L, Wilferth T, Gast LV, et al. Interleaved 23Na/1H MRI of the human heart at 7 Tesla. In: Proceedings of the 30th Annual Meeting of ISMRM, London, UK, 2022. Abstract 0276.
- 19. Platt T, Umathum R, Fiedler TM, et al. In vivo self‐gated 23Na MRI at 7 T using an oval‐shaped body resonator. Magn Reson Med. 2018;80(3):1005–1019. [DOI] [PubMed] [Google Scholar]
- 20. Nagel AM, Laun FB, Weber MA, Matthies C, Semmler W, Schad LR. Sodium MRI using a density‐adapted 3D radial acquisition technique. Magn Reson Med. 2009;62:1565–1573. [DOI] [PubMed] [Google Scholar]
- 21. Chan RW, Ramsay EA, Cunningham CH, Plewes DB. Temporal stability of adaptive 3D radial MRI using multidimensional golden means. Magn Reson Med. 2009;61:354–363. [DOI] [PubMed] [Google Scholar]
- 22. Fessler JA, Sutton BP. Nonuniform fast Fourier transforms using min‐max interpolation. IEEE Trans Signal Process. 2003;51:560–574. [Google Scholar]
- 23. Ruck L, Egger N, Nagelstraßer S, et al. Interleaved 23Na/1H (pTx) MRI of the human heart at 7 Tesla. In: Proceedings of the 32nd Annual Meeting of ISMRM, Singapore, 2024. Abstract 0926.
- 24. Schmitter S, Wu X, Adriany G, Auerbach EJ, Uğurbil K, Van de Moortele PF. Cerebral TOF angiography at 7T: impact of B1+ shimming with a 16‐channel transceiver array. Magn Reson Med. 2014;71:966–977. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Aigner CS, Dietrich S, Schaeffter T, Schmitter S. Calibration‐free pTx of the human heart at 7T via 3D universal pulses. Magn Reson Med. 2022;87:70–84. [DOI] [PubMed] [Google Scholar]
- 26. Insko E, Bolinger L. Mapping of the radiofrequency field. J Magn Reson Imaging. 1993;103:82–85. [Google Scholar]
- 27. Yarnykh VL. Actual flip‐angle imaging in the pulsed steady state: a method for rapid three‐dimensional mapping of the transmitted radiofrequency field. Magn Reson Med. 2007;57:192–200. [DOI] [PubMed] [Google Scholar]
- 28. Van de Moortele P‐F, Ugurbil K. Very fast multi channel B1 calibration at high field in the small flip angle regime. In: Proceedings of the 17th Annual Meeting of ISMRM, Honolulu, Hawaii, USA, 2009. Abstract 0367.
- 29. Lee S, Hilal SK, Cho Z. A multinuclear magnetic resonance imaging technique‐simultaneous proton and sodium imaging. Magn Reson Imaging. 1986;4:343–350. [DOI] [PubMed] [Google Scholar]
- 30. Dietrich O, Raya JG, Reeder SB, Reiser MF, Schoenberg SO. Measurement of signal‐to‐noise ratios in MR images: influence of multichannel coils, parallel imaging, and reconstruction filters. J Magn Reson Imaging. 2007;26:375–385. [DOI] [PubMed] [Google Scholar]
- 31. Klein S, Staring M, Murphy K, Viergever MA, Pluim JP. Elastix: a toolbox for intensity‐based medical image registration. IEEE Trans Med Imaging. 2009;29:196–205. [DOI] [PubMed] [Google Scholar]
- 32. Karkouri J, Rodgers CT. Sequence building block for magnetic resonance spectroscopy on Siemens VE‐series scanners. NMR Biomed. 2024;37:e5165. [DOI] [PubMed] [Google Scholar]
- 33. de Bruin PW, Koken P, Versluis MJ, et al. Time‐efficient interleaved human 23Na and 1H data acquisition at 7 T. NMR Biomed. 2015;28:1228–1235. [DOI] [PubMed] [Google Scholar]
- 34. Kaggie JD, Sapkota N, Thapa B, et al. Synchronous radial 1H and 23Na dual‐nuclear MRI on a clinical MRI system, equipped with a broadband transmit channel. Concepts Magn Reson Part B Magn Reson Eng. 2016;46:191–201. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Wild JM, Marshall H, Xu X, et al. Simultaneous imaging of lung structure and function with triple‐nuclear hybrid MR imaging. Radiology. 2013;267(1):251–255. [DOI] [PubMed] [Google Scholar]
- 36. Wild JM, Ajraoui S, Deppe MH, et al. Synchronous acquisition of hyperpolarised 3He and 1H MR images of the lungs–maximising mutual anatomical and functional information. NMR Biomed. 2011;24(2):130–134. [DOI] [PubMed] [Google Scholar]
- 37. Kaggie JD, Sapkota N, Jeong K, et al. Synchronous 1H and 23Na dual‐nuclear MRI on a clinical MRI system, equipped with a time‐shared second transmit channel. In: Proceedings of the 22nd Annual Meeting of ISMRM, Milan, Italy, 2014. Abstract 3789.
- 38. Liu Y, De Feyter HM, McIntyre S, Nixon TW, de Graaf RA. Interleaved MRI and DMI on human brain in vivo. In: Proceedings of the 29th Annual Meeting of ISMRM, 2021. Abstract 0005.
- 39. Niendorf T, Graessl A, Thalhammer C, et al. Progress and promises of human cardiac magnetic resonance at ultrahigh fields: a physics perspective. J Magn Reson. 2013;229:208–222. [DOI] [PubMed] [Google Scholar]
- 40. Kent JL, de Buck MH, Dragonu I, Chiew M, Valkovič L, Hess AT. Accelerated 3D multi‐channel B1+ mapping at 7 T for the brain and heart. Magn Reson Med. 2024;92:2007–2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41. Gras V, Vignaud A, Amadon A, Le Bihan D, Boulant N. Universal pulses: a new concept for calibration‐free parallel transmission. Magn Reson Med. 2017;77:635–643. [DOI] [PubMed] [Google Scholar]
- 42. Teeuwisse WM, Brink WM, Webb AG. Quantitative assessment of the effects of high‐permittivity pads in 7 Tesla MRI of the brain. Magn Reson Med. 2012;67:1285–1293. [DOI] [PubMed] [Google Scholar]
- 43. Kaggie JD, Lanz T, McLean MA, et al. Combined 23Na and 13C imaging at 3.0 tesla using a single‐tuned large FOV birdcage coil. Magn Reson Med. 2021;86:1734–1745. [DOI] [PubMed] [Google Scholar]
- 44. Boehmert L, Kuehne A, Waiczies H, et al. Cardiorenal sodium MRI at 7.0 Tesla using a 4/4 channel 1H/23Na radiofrequency antenna array. Magn Reson Med. 2019;82(6):2343–2356. [DOI] [PubMed] [Google Scholar]
- 45. Meyerspeer M, Magill AW, Kuehne A, Gruetter R, Moser E, Schmid AI. Simultaneous and interleaved acquisition of NMR signals from different nuclei with a clinical MRI scanner. Magn Reson Med. 2016;76:1636–1641. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46. McLean MA, Hinks RS, Kaggie JD, et al. Characterization and correction of center‐frequency effects in X‐nuclear eddy current compensations on a clinical MR system. Magn Reson Med. 2021;85:2370–2376. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47. Stobbe R, Beaulieu C. Sodium imaging optimization under specific absorption rate constraint. Magn Reson Med. 2008;59:345–355. [DOI] [PubMed] [Google Scholar]
- 48. Eichfelder G, Gebhardt M. Local specific absorption rate control for parallel transmission by virtual observation points. Magn Reson Med. 2011;66:1468–1476. [DOI] [PubMed] [Google Scholar]
- 49. Schmidt S, Metzger GJ. Improved multinuclear parallel transmit optimization using multinuclear virtual observation points. In: Proceedings of the 32nd Annual Meeting of ISMRM, Singapore, 2024. Abstract 4945.
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data S1. Formulas to calculate the polar and azimutal angles and Cartesian unit vectors of the radial projections (n = 1, …, 15 000 for 23Na; and n = 1, …, 60 000 for 1H) based on the two‐dimensional (2D) golden means (, ) presented by Chan et al. 21
Figure S1. Influence of vendor‐provided B0 compensation on interleaved 23Na/1H MRI. In (A) and (B), measurements were performed under the B0 compensation currently provided by the vendor, whereas (C) was performed with a corrected software implementation of the B0 compensation at another Terra.X development device of the vendor. All measurements were performed using a 23Na resolution phantom and a dual‐tuned 23Na/1H head coil. The same interleaved sequence parameters as for the in vitro and in vivo cardiac measurements were used for (A) and (B), whereas for (C) the resolution of the 23Na images was (4 mm)3 instead of (6 mm)3.
(A) For 23Na MRI, we used a single‐nuclear 23Na sequence (I) only containing transmission, readouts, and gradients for 23Na as reference image. Acquiring a 23Na image using the dual‐nuclear interleaved 23Na/1H sequence with turned on B0 compensation (II) leads to a spatial shifting and blurring of the 23Na image (II‐I). By manually turning off the B0 compensation (III), these effects can be avoided, and there are no relevant differences between the 23Na image of the single‐nuclear and dual‐nuclear sequence (III‐I; mean difference over the phantom: 0.62% of the maximum value).
(B) For 1H MRI, B0 compensation worked for single‐nuclear and dual‐nuclear sequences (not shown here). Therefore, we used the interleaved acquired 1H image as reference (I). Because we had to turn off the B0 compensation for interleaved 23Na/1H acquisitions due to the artifacts for the 23Na image, interleaved acquired 1H images were not B0 compensated (II), resulting in minor spatial shifts visible in the difference image (II‐I). However, using image coregistration (III), these shifts could be corrected (III‐I; mean difference over the phantom: −0.02% of the maximum value). Because the effect of the B0 compensation is only determined by the shapes and timings of the applied gradients, it can be assumed that these shifts are constant for repeated measurements with the same gradient scheme. Even though the shifts were in the submillimeter range (), and thus almost negligible, we corrected the interleaved acquired 1H images for these shifts in our reconstruction.
After reporting the problem to the vendor, they identified errors in the software implementation of the B0 compensation that affect X‐nuclei MRI during dual‐nuclear interleaved acquisitions. (C) Test measurements of the interleaved sequence acquired under the latest software version with a corrected software implementation of the B0 compensation on another Terra.X development device of the vendor. There were no differences (II‐I) visible between single (I) and dual‐nuclear (II) measurements acquired with turned on B 0 compensation. As for (A), the differences between the single‐nuclear sequence with turned‐on B0 compensation and the dual‐nuclear interleaved sequence with turned‐off B0 compensation did not show relevant differences (III‐I). In summary, after the next software update of our scanner, the interleaved sequence should be applicable without turning off the B0 compensation. However, until then, we have to use the presented work‐around of manually turning off the B0 compensation for the interleaved measurements.
Figure S2. Linear calibration fit () between the calculated 1H reference voltages within the heart for the default phase shim (DPS) and the body weight of 35 previously measured subjects. 15 The linear fit showed good correlation between the reference voltage within the heart and the weight of the subjects (correlation factor: 0.76). The relative absolute deviation between the measured reference voltage and the corresponding reference voltage based on the calibration fit was calculated by . For the 35 subjects, the mean relative absolute deviation was 6.4% (minimal/maximal: 0.02%/14.5%). Because there are currently no inline adjustments available for 1H body imaging at 7 T, we used this calibration fit to estimate subject‐specific 1H reference voltages for the DPS, universal phase shim (UPS), and individual phase shim (IPS) based on the weight of the measured subjects. However, due to the increased B1 + efficiency of UPS and IPS, 15 an additional calibration fit for the UPS and direct calculation of the reference voltage during the optimization of the IPS could further improve the estimation of the flip angle (FA) in the future.
Figure S3. (A) Comparison of measured in vitro channel‐wise relative 1H B1 + maps for the combined and uncombined coil setup. The channel‐wise relative 1H B1 + maps of the combined setup were image‐coregistered to the uncombined setup, reducing misalignments of the phantom due to the repositioning. Absolute difference maps were calculated by , and the mean difference was calculated within an exemplary heart region (red line, Subject 1). Within the heart region, no relevant differences were visible. (B) Influence of coil setups on optimized 1H excitation pulses. Using the shown in vitro channel‐wise relative 1H B1 + maps in (A), 1H phase shims were optimized for each coil setup (combined, uncombined) within the same three‐dimensional (3D) heart region (red line, Subject 1) and then applied to the channel‐wise relative 1H B1 + maps of both setups. For individually designed pulses, the pulses are usually optimized during the measurement and thus applied on the same coil setup. Coefficient of variation (CV) values showed no relevant differences between 1H pulses designed and applied on the uncombined (I) and combined setup (IV). Even pulses optimized on relative B1 + maps of one of the coil setups and then applied to the other coil setup (II, III) yielded comparable CV values. Thus, for the performance of 1H pulses optimized within the heart region, the influence of the 23Na coil in the combined setup appears to be negligible.
Figure S4. Animated overlay of 23Na and 1H images (Subject 1) acquired with the interleaved 23Na/1H pTx sequence using the universal phase shims (UPS) for 1H excitation. Due to the interleaved 23Na/1H acquisition scheme, the corresponding 23Na and 1H images are aligned without the need for image coregistration.
Table S1. Overview of coefficient of variation (CV) values. The mean CV values were calculated within the heart volume based on the simulated 1H FA maps of each subject and are shown for the four different 1H excitation pulses (default phase shim (DPS), universal phase shim (UPS), individual phase shim (IPS) and individual 4kT pulses (4kT)) of each subject.
