Skip to main content
Wiley Open Access Collection logoLink to Wiley Open Access Collection
. 2025 Jan 29;94(1):119–133. doi: 10.1002/mrm.30442

Rapid quantitative MRI at 46 mT: Accelerated T1 and T2 mapping with low‐rank reconstructions

Yiming Dong 1,, Chloé Najac 1, Matthias J P van Osch 1, Andrew Webb 1, Peter Börnert 1,2, Beatrice Lena 1
PMCID: PMC12021333  PMID: 39887430

Abstract

Purpose

To evaluate accelerated T1‐ and T2‐mapping techniques for ultra–low‐field MRI using low‐rank reconstruction methods.

Methods

Two low‐rank–based algorithms, image‐based locally low‐rank (LLR) and k‐space–based structured low‐rank (SLR), were implemented to accelerate T1 and T2 mapping on a 46 mT Halbach MRI scanner. Data were acquired with 3D turbo spin‐echo sequences using variable‐density poisson‐disk random sampling patterns. For validation, phantom and in vivo experiments were performed on six healthy volunteers to compare the obtained values with literature and to study reconstruction performance at different undersampling factors and spatial resolutions. In addition, the reconstruction performance of the LLR and SLR algorithms for T1 mapping was compared using retrospective undersampling datasets. Total scan times were reduced from 45/38 min (R = 1) to 23/19 min (R = 2) and 11/9 min (R = 4) for a 2.5 × 2.5 × 5 mm3 resolution, and to 18/16 min (R = 4) for a higher in‐plane resolution 1.5 × 1.5 × 5 mm3 for T1/T2 mapping, respectively.

Results

Both LLR and SLR algorithms successfully reconstructed T1 and T2 maps from undersampled data, significantly reducing scan times and eliminating undersampling artifacts. Phantom validation showed that consistent T1 and T2 values were obtained at different undersampling factors up to R = 4. For in vivo experiments, comparable image quality and estimated T1 and T2 values were obtained for fully sampled and undersampled (R = 4) reconstructions, both of which were in line with the literature values.

Conclusions

The use of low‐rank reconstruction allows significant acceleration of T1 and T2 mapping in low‐field MRI while maintaining image quality.

Keywords: low‐field MRI, low‐rank, quantitative MRI, relaxation times, undersampling

1. INTRODUCTION

The advancement of medical diagnostics is frequently dependent on the availability of sophisticated, albeit costly, equipment. However, there is a need for cost‐effective, adaptable, and portable solutions, such that low field (less than 0.1 T) MRI can provide, to serve diverse medical needs worldwide. 1 , 2 , 3 Despite its intrinsically low SNR, 4 low‐field MRI is paving the way into widespread adoption, especially in point‐of‐care settings, due to the recent advancements in image processing and hardware. Currently, one of the main remaining challenges is the long scan time and limited achievable spatial resolution due to the limited gradient strength of the system 5 and lack of parallel imaging capability. This limitation is particularly evident in quantitative imaging because this requires repeated scans for parameter fitting as needed for T1, T2 mapping. 6 , 7 , 8 Quantitative MRI is a key technique for studying brain development, disease progression, and myelination patterns. 9 A recent study reported scan times of more than 30 min for T1 mapping of the whole adult brain at 50 mT. 5

To accelerate MRI scans, k‐space undersampling is often used. At clinical field strengths, the advent of undersampling techniques such as parallel imaging 10 , 11 and compressed sensing 12 enables the reconstruction of MR images from data below the Nyquist–Shannon limit. Recently, research has increasingly focused on exploiting the low‐rank 13 , 14 , 15 , 16 , 17 , 18 , 19 , 20 , 21 , 22 , 23 properties of MRI datasets, which represent another promising linear‐dependent constraint for undersampled MRI reconstruction. Most low‐rank–based reconstructions can be divided into either image‐based– 13 , 14 , 20 , 21 or k‐space–based approaches, 16 , 17 , 18 , 19 , 20 , 21 , 22 , 23 which have been widely used in parallel imaging, 15 , 24 , 25 dynamic imaging, 13 , 20 compressed sensing, 17 and diffusion imaging. 18 , 19 , 21

Previous research has shown successful reconstruction of quantitative parameter mapping with image‐based locally low‐rank (LLR), 14 , 26 exploiting the spatial correlation along the parametric mapping dimensions (e.g., different TIs or different flip angles for T1 and T2 mapping). Moreover, a joint low‐rank modeling of local k‐space neighborhoods (LORAKS) approach can also be used to jointly reconstruct multi‐contrast images by exploring the limited support and smooth phase constraints. 23

In this study, low‐rank–based approaches are investigated to reconstruct undersampled scans for quantitative MRI mapping on a custom‐built 46 mT permanent magnet–based MRI scanner that only has a single‐element receive coil. 5 , 27 , 28 Both the image‐based locally low‐rank 14 and low‐rank modeling of local k‐space neighborhoods 17 , 22 , 23 , 24 formulations are employed to investigate their ability to significantly speed up T1 and T2 mapping by undersampling, while still yielding accurate MR parameter estimates. This is done using phantom, simulation, and in vivo experiments.

2. THEORY

For both T1 and T2 mapping, repeated scans with different acquisition parameters are acquired along a contrast‐encoding dimension to enable the fitting of the MR parameters. Data redundancy between different images along this encoding dimension can be explored and guide the reconstruction. 14 , 29 Previous studies have shown how this redundancy can potentially be used to undersample k‐space by incorporating either spatial LLR regularization 14 , 26 or k‐space–based SLR 16 , 23 into image reconstruction. These approaches are particularly beneficial in improving the reconstruction of undersampled datasets, and possibly even more in the context of low‐field MRI, for which the application of parallel imaging techniques is limited but the need for acceleration is important.

In the image domain, for N datasets acquired for parameter mapping, the corresponding optimization problem of LLR can be written as:

x^1,,N=argminx1,,Nn=1NDnFxnbn22+λLmΩRmx1,x2,,xN*, (1)

where Dn represents the sampling mask for each image frame (n=1,2,,N), F indicates the FFT operator, xn the unknown series of images needed to be reconstructed, λL the regularization factor, and bn the undersampled k‐space data. To set up the LLR constraint, the operator Rm extracts a small spatial patch/block around a pixel‐index m in image‐space from N different images, vectorizes each one, and concatenates them along all vectorized arrays to form the Casorati matrix. Due to the redundant information along the contrast‐encoding dimension, this Casorati matrix is assumed to be low rank as shown in previous studies. 14 By minimizing its nuclear norm * using singular value decomposition with thresholding, the property of spatial low‐rankness along the contrast‐encoding dimension can be exploited to help the reconstruction of each separate undersampled image. 14

Similarly, k‐space–based structured low‐rank reconstruction can be used to exploit the redundant information along the contrast‐encoding dimension and guide the reconstruction:

x^1,,N=argminx1,,nn=1NDnFxnbn22+λSSFxnYF2. (2)

Here, a nonconvex S‐matrix formulation from the LORAKS17,22,24 algorithm and its related approaches were used for structured low‐rank constrained reconstruction. Y is the matrix that approximates the series of S‐matrices SFxn, which are concatenated from different Fourier‐transformed images Fxn. This approximation is made under the constraint that Y provides an optimal low‐rank approximation, and λS is the regularization factor. By minimizing the Frobenius norm F2 of the difference between the series S and Y, this approach effectively enforces a low‐rank constraint and guides the reconstruction process. 23

It is worth noting that a different undersampling pattern should be used for each different image frame (i.e., for each TI in T1 mapping and for each TE in T2 mapping). This helps to generate incoherence between different samples along the contrast‐encoding dimension for optimal reconstruction performance, as shown in previous studies. 14 , 29

Finally, another commonly used undersampling strategy is partial Fourier, which adapts the conjugate symmetry property of real‐valued images to reconstruct partially structured undersampled k‐space. Previous research has shown that SLR reconstruction using the LORAKS S‐matrix can directly reconstruct partial Fourier data without phase information. 22 , 24 In contrast, LLR 22 , 24 requires an additional step in which the conjugate symmetry property is explored using the concept of the so‐called virtual conjugate shots (VCS). 19 , 21 , 30 These additional “shots” can be generated through inverting and taking the conjugate of the original T1‐/T2‐mapping datasets. These additional virtual shots are then concatenated with the original measured data to form the Casorati matrix within each local patch. Finally, nuclear norm minimization is performed on all the double‐sized Casorati matrices.

3. METHODS

3.1. Acquisition

All experiments were performed on a 46 mT (1.97 MHz) Halbach‐based MRI system using a Magritek Kea2 spectrometer (Magritek, Aachen, Germany). A spiral solenoid single‐channel transmit/receive head coil was used for imaging. For both T1 and T2 mapping, different variable‐density poisson disk‐sampling patterns along the two phase‐encoding directions were randomly generated for each undersampled contrast‐encoding frame (with fully sampled regions in the center of the k‐space covering approximately 12.5% of each direction). Conventional frequency encoding was used along the readout direction.

The T1‐mapping protocol was based on a 3D inversion recovery turbo spin echo (TSE) sequence, with the following acquisition parameters: TR/TE: 1200 ms/20 ms (center‐out k‐space trajectory); six TIs logarithmically spaced between 50 and 900 ms; echo train length: 6; slice thickness 5 mm; acquisition bandwidth: 20 kHz; no signal averaging.

The T2‐mapping protocol was performed using a T2‐preparation 31 (T2‐prep) module. This method was used to implement different undersampling masks for each TE and to ensure sufficient incoherence along the contrast‐encoding dimension for low‐rank reconstruction. 3D TSE data were acquired with the following acquisition parameters: TR/TE: 1000 ms/18 ms (center‐out k‐space trajectory); echo train length: 6; slice thickness 5 mm; acquisition bandwidth: 20 kHz; no signal averaging. Scans were acquired with T2‐prep durations of 17, 52, 87, 122, 157, and 182 ms.

The range of inversion delays and T2‐prep durations were optimized for parameter mapping of the white (WM) and gray matter (GM). Effectively, this range precludes the calculation of accurate values for the CSF with its very long T1 and T2 values. Those are challenging to measure in vivo even at clinical MRI field strengths and are of very limited diagnostic value.

For the phantom study, we used a morphometric 3D‐printed brain shape, filled with a mix of copper sulphate and agarose to mimic the relaxation times expected for the brain at 46 mT. Fully sampled (R = 1) data at low in‐plane resolution (2.5 × 2.5 × 5 mm2) and variable‐density undersampled data (R = 4 and R = 8) for both T1 and T2 mapping were acquired.

In vivo experiments were conducted on six healthy volunteers. T1 mapping was performed in three of these volunteers, while T2 mapping was performed in the other three. Informed consent was obtained, and this study has been approved by the local ethics committee under the stipulation that the maximum scan time for a subject is limited to 60 min. To accommodate this constraint, we divided the experimental sets for each parameter mapping as: R = 1 and R = 4 low resolution (2.5 × 2.5 × 5 mm3) for one volunteer, and R = 2 with R = 4 low resolution (2.5 × 2.5 × 5 mm3) for the second volunteer to test the effect of undersampling; R = 4 “low resolution” (2.5 × 2.5 × 5 mm3) and R = 4 “high resolution” (1.5 × 1.5 × 5 mm3) for the third volunteer to test the performance of reconstruction algorithms in high‐resolution cases.

The acquisition matrix size, scan time for different acceleration factors, and resolution are shown in Table 1. A fixed FOV of 235 × 175 × 160 mm3 was used for all experiments.

TABLE 1.

Data acquisition parameters and scan times for fully sampled and undersampled scans for both phantom and in vivo scans.

Resolution Matrix size Acceleration factor (R) T1 mapping (scan time per TI/total scan time) [min] T2 mapping (min per TE/total scan time) [min]
Low (2.5 × 2.5 × 5 mm3) 94 × 70 × 32 1 7.5/45.0 6.4/38.4
Low (2.5 × 2.5 × 5 mm3) 94 × 70 × 32 2 3.9/23.6 3.2/19.5
Low (2.5 × 2.5 × 5 mm3) 94 × 70 × 32 4 1.9/11.4 1.6/9.6
Low (2.5 × 2.5 × 5 mm3) 94 × 70 × 32 8 1.0/6.0 0.8/4.8
High (1.5 × 1.5 × 5 mm3) 156 × 116 × 32 4 3.1/18.6 2.7/16.2

3.2. Image reconstruction

In order to solve Equation 1 for both T1 and T2 mapping, a 3D LLR reconstruction was implemented in Python 3.7 using the Sigpy toolbox. 32 All hyperparameter selections were made with respect to prior simulations of retrospective undersampling from R = 1 data, shown in Figures S1 and S2. For each iteration, a series of 3D nonoverlapping blocks (10 × 10 × 4) were randomly selected from each individual image to form the Casorati matrix. 26 , 33 The LLR regularization was implemented with a proximal gradient descent–associated singular value decomposition for each Casorati matrix, plus soft thresholding, to minimize its nuclear norm with 30 iterations and regularization factor of 20.

For the SLR reconstruction with the S‐matrix from the LORAKS algorithm, a majorize–minimize algorithm is used to iteratively solve Equation 2 in a 2D manner 17 , 22 , 24 , 34 for the 3D dataset. Specifically, a 1D IFFT was first applied along the readout direction to transform the k‐space data to hybrid space (x, ky, kz). Then, for each readout position (x), the SLR reconstruction was performed independently on the 2D ky‐kz plane. Ten fixed outer iterations were performed, with a maximum of 200 inner iterations of conjugate gradients, using a chosen convergence tolerance of 7 × 10−7. The k‐space radius used to construct LORAKS neighborhoods was set to 3. The matrix rank value used to define the non‐convex regularization penalty for Y in Eq. (2) was chosen to be 120.

All image reconstructions and data processing were performed on a workstation equipped with an NVIDIA RTX A6000 GPU (Nvidia, Santa Clara, CA, USA). Reconstruction time for a typical T1‐mapping dataset (matrix size: 6 × 94 × 70 × 32) is 8 s for LLR and 452 s for SLR.

3.3. Data processing and evaluation criteria

For T1 mapping, a simple phase correction was applied using the phase of the last TI to obtain real‐valued data for the fitting. 35 Then, the phase‐corrected data sets were fitted (based on the assumption of perfect saturation from the last readout and with TR > TI):

sTI,M0,T1=M012eTIT1+eTRT1, (3)

and for T2 mapping, the data sets were fitted with:

sTE,M0,T2=M0eTET2, (4)

where M0 represents the proton density (in case the magnetization is fully recovered). Segmentation of the brain‐shaped phantom images was obtained by applying thresholds to the T1 and T2 maps fitted from the fully sampled (R = 1) datasets. 6 For the T1 map, an empirical threshold of 240 ms was set to distinguish between WM and GM, respectively. For T2, given the small difference in T2 values between WM and GM at 46 mT (< 10 ms), a joint segmentation mask was generated for both, excluding only the CSF compartment. For the in vivo experiments, segmentation was performed on each of the parameter maps individually to segment WM/GM/CSF for T1 maps, and (WM & GM)/CSF for T2 maps, using the Advanced normalization tools. 36

In addition, pairwise comparisons of the obtained T1 and T2 values were performed using the paired t‐test with Bonferroni correction, focusing on differences between three reconstruction methods: (1) reference data (R = 1) with FFT, (2) undersampled data with LLR, and (3) undersampled data with SLR.

For the phantom study, these comparisons were conducted separately for WM and GM regions. For both WM and GM, two sets of comparisons were performed: (1) R = 1 with FFT versus R = 4 with LLR, R = 1 with FFT versus R = 4 with SLR, and R = 4 with LLR versus R = 4 with SLR; (2) R = 1 with FFT versus R = 8 with LLR, R = 1 with FFT versus R = 8 with SLR, and R = 8 with LLR versus R = 8 with SLR, with regions of interest (ROIs) generated from masks based on the R = 1 data. To confirm that the implementation of undersampling in our MRI system does not introduce inconsistencies in the k‐space data, the normalized RMS errors (nRMSEs) of the retrospective undersampled (R = 4) and prospective undersampled (R = 4) T1/T2 maps were calculated and compared.

For the in vivo experiments, T1 and T2 maps generated by the three different methods—R = 1 with FFT, R = 4 with LLR, and R = 4 with SLR—were compared through pairwise comparisons among the three. For both T1 and T2 mapping, a ROI was placed within uniform WM regions for each slice, including approximately 20 pixels per ROI, across eight slices in total. ROIs were manually selected for the R = 1 and R = 4 data and colocalized in the same regions for the LLR and SLR reconstructions. The comparison in the GM using manually selected ROIs was not conducted for in vivo data due to partial‐volume effects at the given resolution.

Based on the Bonferroni correction, a p‐value <0.017 was considered statistically significant. All P‐values from the statistical comparisons are provided in Table S1.

Additionally, one numerical simulation was performed by retrospective undersampling the fully sampled (R = 1) T1‐mapping datasets, mimicking R = 2/4/6/8 to assess the performance of both algorithms (LLR and SLR) with respect to undersampling. The normalized Root Mean Square Error (NRMSE) was calculated between reconstructed undersampled images and the fully sampled data for both selected TI images and T1 maps, as nRMSE=q=1Qx^qxq2/Q/x^norm, where x^q is the reference value of R = 1, xq denotes the values of the undersampled results from the two reconstruction approaches, xnorm the normalization factor derived by the range of x^q as x^q,maxx^q,min, and Q the number of voxels. The simulation result is shown in Figure S1.

Furthermore, to test the feasibility of partial Fourier data reconstruction with both SLR and VCS‐LLR to further reduce the scan time, the R = 1 T1‐mapping datasets were retrospectively undersampled using a series of randomly undersampled partial Fourier masks. These masks were generated by applying a partial Fourier mask to the original variable density Poisson‐disc undersampling masks.

4. RESULTS

4.1. Phantom experiments comparing T1 and T2 maps

Figure 1 illustrates the performance of spatial LLR and k‐space SLR algorithm with different undersampling factors in terms of mean T1 and T2 values and their respective SDs. For T1 mapping, both algorithms operating at different undersampling factors (R = 4 and 8) provide T1 values for both WM and GM that are consistent with those obtained at R = 1 (within the range of one SD). However, at R = 8, spatial blurring is visually apparent, especially close to GM/CSF boundary (indicated by blue arrows). Due to the similar T2 values of WM and GM, a single mean T2 value was reported for each algorithm at R = 4/8, all of which are consistent with the reference data (R = 1). Nevertheless, at R = 8, streak‐like oversmoothing artifacts due to undersampling are evident for the 2D‐SLR reconstructions (indicated by the blue arrow). As shown in Table S1, for the statistical comparisons, no significant differences (all p > 0.017 after Bonferroni correction) were observed among the different T1‐mapping measurements. For T2 mapping, no significant differences were observed in the comparison between R = 1 and R = 4 with both LLR and SLR algorithms. However, the R = 8 SLR reconstruction had a statistically significant difference compared to R = 1 FFT (p = 0.005) and R = 8 LLR (p < 0.001). This statistical discrepancy, which is also in line with the previous observation, points to the fact that in such a relatively low SNR T2‐mapping experiment with strong undersampling (R = 8), the SLR algorithm may depict artifacts leading to compromised quantitative calculations. From Figure S3, the T1 and T2 results show both low nRMSE values and good mapping quality between the retrospectively undersampled and prospectively undersampled data, indicating that the undersampling strategy does not introduce noticeable inconsistencies into the k‐space data.

FIGURE 1.

FIGURE 1

Phantom experiments comparing T1 and T2 maps. Underlying data were reconstructed from LLR/SLR algorithms at R = 4/8 with those calculated from fully sampled R = 1. Both algorithms produce T1/T2 values consistent with R = 1 results as reported at the bottom of each image. At the highest undersampling factor R = 8, some spatial blurring (blue arrows) of the T1 mapping is visible for both algorithms, whereas for the T2 mapping some artifacts are visible for SLR (blue arrow). LLR, locally low‐rank; SLR, structured low‐rank.

4.2. In vivo experiments comparing T1 and T2 maps

The estimated T1 and T2 values for the in vivo experiments (three subjects each) are reported in Tables 2 and 3, respectively. These values align with literature reports at 46 mT (1.97 MHz). 6 , 37 The comparisons among the three approaches for T1/T2 mapping show no statistically significant differences (Table S1). However, the T2 preparation module used for T2 mapping resulted in a noticeable degradation in the quality of reconstructed images and T2 maps compared with previous studies that used multi‐echo spin‐echo or turbo spin‐echo protocol, even when the data is fully sampled (see Figure S5).

TABLE 2.

Estimated T1 relaxation times (mean ± SD) of each subject.

T1 values
WM [ms] GM [ms]
Subject 1 R = 1 FFT 246 ± 23 310 ± 26
R = 4 LLR 251 ± 18 309 ± 19
SLR 251 ± 19 309 ± 20
Subject 2 R = 2 LLR 261 ± 16 311 ± 20
SLR 256 ± 17 310 ± 25
R = 4 LLR 262 ± 16 304 ± 19
SLR 258 ± 16 304 ± 20
Subject 3 R = 4 LLR 262 ± 16 313 ± 18
SLR 261 ± 18 312 ± 25
R = 4 (high‐res. a ) LLR 264 ± 23 318 ± 21
SLR 255 ± 25 321 ± 23
a

High‐resolution: 1.5 × 1.5 × 5 mm3, all others were measured with 2.5 × 2.5 × 5 mm3.

TABLE 3.

Estimated T2 relaxation times of each subject.

T2 values
WM/GM b [ms]
Subject 1 R = 1 FFT 94 ± 16
R = 4 LLR 97 ± 11
SLR 96 ± 9
Subject 2 R = 2 LLR 96 ± 13
SLR 94 ± 16
R = 4 LLR 94 ± 10
SLR 96 ± 12
Subject 3 R = 4 LLR 95 ± 9
SLR 94 ± 10
R = 4 (high‐res. a ) LLR 94 ± 10
SLR 92 ± 10
a

High‐resolution: 1.5 × 1.5 × 5 mm3; all others were measured with 2.5 × 2.5 × 5 mm3.

b

GM/WM T2 values may be not distinguishable at 46 mT.

Abbreviations: GM, gray matter; WM, white matter.

In Figure 2, the fully sampled T1‐mapping data set (R = 1) is compared to prospectively undersampled R = 4 data, reconstructed with FFT, LLR, and SLR. The undersampling artifacts visible in the FFT reconstruction (indicated with red arrows) of the undersampled data are effectively mitigated by both LLR and SLR, resulting in comparable T1 maps compared to R = 1 case (no significant differences; see Table S1). The violin plot for WM of R = 1 shows a distribution centered at 245 ms and for GM at 310 ms, with a slightly larger spread for R = 1 compared to the R = 4 LLR and SLR plots. This may suggest potential smoothing effects, either due to low‐rank regularization or increased partial‐volume effects between different tissues. The comparison between fully sampled T2‐mapping data set (R = 1) with prospectively undersampled R = 4 data is shown in Figure S5.

FIGURE 2.

FIGURE 2

Comparison between R = 1 and R = 4 with LLR and SLR reconstructions for T1 mapping. (A) Individual images with six different TIs for R = 1 and R = 4 using three different reconstructions. Undersampling artifacts (red arrows) from FFT (R = 4) can be mitigated using both LLR and SLR algorithms, producing in (B) T1 maps comparable to R = 1. (C) Violin plots also show consistent T1 fitting results between R = 1 and R = 4 with both methods for both WM and GM. GM, gray matter; WM, white matter.

Figures 3 and 4 show one representative slice of six different TI/TE images with R = 2 and R = 4 from two subjects, as well as results of T1 mapping and T2 mapping of six slices. In addition to the aliasing artifacts due to undersampling, some residual aliasing artifacts in the FFT reconstruction can be seen (red arrow in the T1 datasets), which are mitigated by LLR and SLR reconstructions. Both algorithms can produce reasonable contrast between WM and GM in T1 mapping. In T2 mapping, the values of GM and WM are very close, as reported in earlier studies. 6 , 8

FIGURE 3.

FIGURE 3

Comparison between R = 2 and R = 4 datasets (prospective undersampling) reconstructed with FFT, LLR, and SLR algorithms of two subjects, one for T1 and one for T2 mapping. Undersampling artifacts (red arrows) can be removed by both LLR and SLR algorithms.

FIGURE 4.

FIGURE 4

Results of R = 2 and R = 4 (prospective undersampling) reconstructed with LLR and SLR algorithms for two subjects, one for T1 mapping and one for T2 mapping. Both algorithms at R = 2 and R = 4 show visually comparable contrast in all slices. However, at R = 4, higher undersampling may lead to more blurring effects in some regions around CSF and GM (indicated by green arrows).

To test the performance of both methods for different in‐plane resolutions, low (2.5 × 2.5 mm2) and high (1.5 × 1.5 mm2), T1 and T2 maps were acquired at R = 4 for two subjects (Figure 5). As expected, the gain in sharpness and detail (blue arrows) comes at the price of lower SNR. In addition, the result of k‐space–based SLR appears to be slightly noisier, probably because the spatial LLR reconstruction inherently incorporates smoothing and filtering properties to enable denoising.

FIGURE 5.

FIGURE 5

Comparison between single TI/TE images and corresponding T1/T2 maps at low resolution (2.5 × 2.5 × 5 mm3) and high resolution (1.5 × 1.5 × 5 mm3) from two subjects, respectively. More structures were clearly visible in the TI image and T1 map at high resolution (blue arrows), whereas GW/WM are not distinguishable on single TE image and T2 mapping due to the close T2 values and low SNR.

4.3. Effectiveness of VCS‐LLR and SLR on retrospectively simulated partial Fourier datasets

Figure 6 demonstrates the effectiveness of using VCS‐LLR and SLR to reconstruct retrospectively simulated partial Fourier datasets, showing four out of six TI images. Utilizing the virtual conjugate shot concept, the image sharpness from VCS‐LLR is improved compared to conventional LLR reconstruction. SLR (LORAKS) further reduces the blurring caused by partial Fourier, resulting in difference maps with less residual structure for each TI shown. The T1 maps of comparison of LLR, VCS‐LLR, and SLR are shown in Supporting Information S6. Some other results of combinations of different partial Fourier factors and undersampling factors are shown in Supporting Information S7.

FIGURE 6.

FIGURE 6

A retrospective PF reconstruction experiment with three different algorithms. The retrospective sampling masks for each TI are shown at the top right (green squares). (Left) Selected four TIs (out of six) images with a zoom‐in window (red squares) for better visualization. Blurring due to partial Fourier sampling can be seen (red arrows) with conventional LLR reconstruction but can be improved much more by using either VCS‐LLR or SLR reconstructions. This can also be seen in the corresponding difference maps between individual reconstructions and R = 1 reference images (bottom right), where the loss of brain structures (sharpness) is reduced in the last two reconstructions. PF, partial Fourier; VCS, virtual conjugate shots.

5. DISCUSSION

In this study, two low‐rank–based algorithms (image‐based LLR and k‐space–based SLR) were tested using variable‐density Poisson‐disk random sampling trajectories with a 3D TSE sequence for accelerated quantitative T1 and T2 mapping at 46 mT. For reconstruction, the results showed that undersampling artifacts can be significantly reduced with both algorithms, resulting in comparable image quality and quantitatively parameter mapping accuracy to fully sampled datasets. This was demonstrated through a series of phantom, simulation, and in vivo experiments. Our experiments proved that the scan time can be dramatically reduced when using these reconstruction and acquisition approaches.

The influence of undersampling trajectories on the performance of low‐rank–based algorithms in multi‐contrast reconstruction was a critical aspect of our investigation. Although both algorithms have been extensively tested with higher‐field scanners (>1.5 T), they are often combined with parallel imaging to leverage an additional dimension of data redundancy. However, many low‐field MRI systems, 38 , 39 including the one used in this study, 5 have only a single‐channel solenoid transmit/receive coil, which makes the reconstruction of undersampled data more difficult due to the lack of parallel imaging constraints. 10 , 11 , 40 Testing the performance of such “single‐” constraint reconstruction approaches (LLR/SLR without parallel imaging) was first performed in a phantom study. For T1 mapping, both algorithms were able to produce quantitatively and statistically comparable T1 and T2 values at R = 4 and R = 8 compared to the reference R = 1. For T2 mapping, both algorithms performed comparably in most statistical comparisons; however, at R = 8, the SLR algorithm showed significant discrepancies, likely due to the impact of low‐SNR conditions and strong undersampling, which may have led to compromised quantitative accuracy and the presence of artifacts (marked by red arrow in Figure 1).

In vivo experiments further validated the performance of low‐rank algorithms with the prospective undersampled data using the 3D TSE sequence. T1/T2 values from all measurements are reported in Tables 2 and 3 and are consistent with the results of previous studies 6 , 8 at the similar field strength. Jordanova et al. 8 measured a T1 of 294 ± 18 ms for WM and 460 ± 126 ms for GM, and a T2 of 97 ± 2 ms for WM/GM, using a yoke‐based 64 mT scanner; and O'Reilly and Webb 6 measured a T1 of 275 ± 5 ms for WM and 327 ± 10 ms for GM, and a T2 of 102 ± 6 ms for WM/GM, using a Halbach‐based 50 mT scanner. Additionally, according to the empirical equation derived by Bottomley et al., 37 the calculated T1 values for WM and GM at 46 mT are 235 ± 40 ms and 315 ± 54 ms, respectively. In comparison, T2 values are typically considered to be field‐independent, with Bottomley et al. 37 indicating values of 92 ± 20 ms for WM and 101 ± 13 ms for GM. Our estimates from all experiments (Tables 2 and 3) fall within these empirically calculated ranges. In addition, the statistical comparisons between fully sampled (R = 1) and undersampled (R = 4) datasets showed that both algorithms could produce comparable results in terms of quantification of tissue parameter mapping, which is consistent with the simulation result.

The comparison between R = 2 and R = 4 showed that the higher undersampling (R = 4) posed more challenges to the reconstruction, resulting in more partial volume effects and image blur (Figures 3 and 4). On the other hand, R = 4 substantially reduced scan time. Therefore, a trade‐off between image quality and acquisition speed should be seriously considered based on practical applications. This was also the case for the high‐resolution experiments. It is important to note that with the current settings, high‐resolution scans with six TIs can only be acquired in an undersampled fashion, taking already a total of 18 min when using R = 4. Acquiring an R = 1 dataset would take over an hour, which exceeds the maximum scan time per subject allowed by our current medical ethics agreement. Building on these observations, Figure 5 further explored high‐resolution scans (1.5 × 1.5 × 5 mm3) and demonstrated their potential to reveal more brain structures compared to low‐resolution scans (2.5 × 2.5 × 5 mm3) despite a noticeable increase in noise level.

The results demonstrated that both algorithms effectively reduce undersampling artifacts and provide comparable image quality and quantitative mapping accuracy to fully sampled datasets. However, there are differences in their performance characteristics that suggest specific trade‐offs. In general, LLR provided smoother, artifact‐free reconstructions with a denoising effect, 14 especially in regions where the SNR was inherently low. This smoothing effect is beneficial in cases in which noise suppression is critical but can introduce blurring at high undersampling factors, as seen in the T1 maps close to the GM/CSF interface (Figures 4 and 5). In contrast, SLR showed a reduced degree of smoothing resulting in a slight increase of noise in the final images, which may preserve more fine details, such as those present near tissue boundaries. However, the SLR reconstructions at R = 8 (Figure 1) also exhibited minor streaking artifacts in the T2 maps under low SNR conditions, highlighting a limitation for higher undersampling in certain scenarios (e.g., Figure 1, marked by red arrows). From a computational perspective, SLR requires greater processing power and time because it iteratively solves the extended Hankel low‐rank matrix across the entire k‐space via single‐value decomposition. 17 LLR, which operates on localized image blocks in 3D space with even smaller Casorati matrices, is generally faster, making it more suitable for real‐time applications or environments with limited computing resources. These results emphasize that LLR may be preferred when noise reduction and speed are priorities as well as in environments with limited computational power, whereas SLR may be more advantageous for applications requiring higher resolution or finer detail preservation. Further developments, such as 3D SLR approaches, may help to alleviate some of the current limitations and enable even more robust performance. However, the Hankel matrix used in the SLR approach is constructed from the full k‐space of different contrast‐encoding points (e.g., TI, ky, kz in T1 mapping), unlike the “locally” relatively smaller Casorati matrix used in LLR. Enabling 3D SLR with the whole 4D dataset (e.g., TI, kx, ky, kz), including all the large matrix multiplications required in the LORAKS 17 formulation, involves a significantly increased computational burden and is not feasible with the present computational resources in this work.

One challenge is to define the reconstruction hyperparameters (e.g., regularization factors, block‐size) for both algorithms for different resolutions and SNR levels. In this work, the hyperparameters were chosen based on a priori simulations, as shown in the Figures S1 and S2. The use of methods based on Stein's unbiased risk estimate (e.g., Ref. 41) could allow an automatic determination based on the data itself but may increase the complexity of the reconstruction.

Another limitation of the current study is the challenge of validating the in vivo scans when accurate segmentation of WM and GM is so difficult due to the low resolution and SNR of the undersampled data. T2 mapping was found to be particularly challenging due to similar T2 values between the two tissues. Future work could explore more advanced segmentation tools, for example, artificial intelligence‐based approaches. 42 , 43 Another limitation of this work is the low SNR of the currently used TSE T2 sequence: We used an approach based on T2‐preparation module for T2 mapping, which may potentially result in a lower image quality compared to the multi‐echo spin‐echo sequence. 44 This is mainly due to a practical limitation in our current implementation of undersampled random k‐space trajectories. In the future, other methods could be used to improve the image quality of T2 mapping, 14 , 26 , 45 which needs further investigation. Furthermore, whereas the last retrospective undersampling experiment showed the ability of both algorithms to reconstruct partial Fourier data, experimental data should be acquired in the future to verify their reliability.

For future studies, the presented approach with an extensive range of TI values (up to 900 ms) may very well be suited for T1 mapping in neonatal and pediatric patients (e.g., see range in Padormo et al. 7 ), given that T1 has the potential to serve as a reliable biomarker for brain development. In general, the potential of these methods is not limited to T1 and T2 mapping. With recent advances in reconstruction methods, redundant information from multiple scans with different contrasts can be exploited to guide the reconstruction along the contrast‐encoding dimension. Future implementations may include, for example, multi‐shot or multi‐average diffusion scans (without any phase navigators) or multi‐contrast clinical protocols (T1‐weighted, T2‐weighted, and FLAIR).

6. CONCLUSION

This study demonstrates that undersampling combined with both image‐based– and k‐space–based low‐rank reconstruction algorithms is a viable strategy for accelerating T1 and T2 mapping at low magnetic field strengths. This methodology holds potential for improving the efficiency and feasibility of low‐field MRI, ultimately contributing to more widespread adoption of this cost‐effective and portable imaging modality.

FUNDING INFORMATION

This work was funded by the Dutch Science Foundation (NWO) Open Technology, grant 18981.

Conflict of Interest Statement

Peter Börnert is an employee of Philips Research.

Supporting information

Figure S1. Evaluation of Locally Low‐Rank (LLR) reconstruction quality for T1 mapping data across different patch sizes and regularization factors. The plots display the Structural Similarity Index Measure (SSIM) and normalized Root Mean Square Error (nRMSE) as functions of the regularization factor, ranging from 2 to 120 in steps of 2, for patch sizes of [8 × 8 × 4], [10 × 10 × 4], and [12 × 12 × 4] for reconstructing the retrospectively undersampled R = 4 dataset. The SSIM and nRMSE metrics are calculated between the R = 4 LLR‐reconstructed images and the reference R = 1 images to assess structural similarity and reconstruction fidelity, respectively. The optimal parameters were determined to be a patch size of [10 × 10 × 4] with regularization factor of 20, as both metrics (SSIM and nRMSE) exhibited their best values at this configuration. This patch size corresponds well with the spatial resolution used in the study (2.5 × 2.5 × 5 mm3). Since it was not possible to acquire (R = 1) 1.5 × 1.5 × 5 mm3 resolution data for optimizing the simulation, the same optimal parameters were applied to the high‐resolution case.

Figure S2. Evaluation of structured low‐rank (SLR) reconstruction quality for T1 mapping data across different ranks. The plots show the Structural Similarity Index Measure (SSIM) and the Normalized Root Mean Square Error (nRMSE) as a function of the rank selection, highlighting its impact on the reconstruction quality for the SLR algorithm. The optimal rank value was found to be 120, where SSIM reaches its highest peak and nRMSE reaches its second lowest value (0.02134), very close to the absolute minimum (at rank = 110, 0.02125). This balance achieves the best trade‐off between SSIM and nRMSE, indicating high structural fidelity and minimal reconstruction error.

Figure S3. Comparison of phantom T1 mapping and T2 mapping retrospectively undersampled data and prospectively undersampled data (R = 4). One concern for current sequence development is that the implementation of undersampling in the point‐of‐care ultra‐low field MRI system may introduce inconsistencies in the k‐space data. The T1 and T2 results show both low nRMSE values and mapping quality between the retrospectively undersampled and prospectively undersampled data, indicating that the undersampling strategy does not introduce noticeable inconsistencies into the k‐space data, despite the challenges of ultra‐low field MRI hardware.

Figure S4. Retrospective simulation study on in vivo T1 mapping. Comparison of retrospective acceleration factors (R = 2, 4, 6, 8) on images and T1 maps generated from one slice of R = 1 data using LLR and SLR reconstructions. The figure shows images from a sample TI = 302 ms, T1 maps from six different TIs, and the corresponding difference maps between the reference (R = 1) and each reconstructed image/T1 map. As the undersampling factor increases, more visible structures appear on the difference maps of the images and larger discrepancies are observed on the T1 maps, reflecting the more difficult reconstruction conditions at higher undersampling factors. The nRMSE values for the LLR images were 0.037, 0.050, 0.058, and 0.065, and for the SLR images were 0.034, 0.047, 0.056, and 0.064 at the respective undersampling factors. For the T1 maps (excluding CSF regions), the nRMSE values were 0.022, 0.028, 0.032, and 0.038 for LLR and 0.020, 0.027, 0.032, and 0.038 for SLR. This study illustrates how both algorithms handle undersampling artifacts, with both methods effectively mitigating artifacts at lower acceleration factors, particularly at R = 2, where no significant structural loss is observed in the difference maps compared to the R = 1 reference.

Figure S5. In‐vivo prospective T2 mapping of R = 1 and R = 4. (A) An example slice of individual TE images for T2 mapping, reconstructed using FFT for fully sampled (R = 1) and undersampled (R = 4) data, and using LLR and SLR reconstruction methods for undersampled data (R = 4). Undersampling artifacts visible in the FFT reconstruction (R = 4) are effectively reduced in the LLR and SLR reconstructions, resulting in improved image quality and structural clarity (red arrows). (B) T2 maps generated from the corresponding reconstructions. Note that the partial volume effects with the lower SNR compared to the T1 mapping results (Figure 2) result in visible contrast changes, especially around the CSF regions (red arrows). GM and WM T2 values are not readily distinguishable due to their close values, but are consistent with reference R = 1 data (Table 2) and agree with previous observations in the literature. 6 , 8 , 37

Figure S6. T1 maps of the partial Fourier experiment as a supplement to Figure 6. Comparison of different partial Fourier reconstruction methods, including LLR, VCS‐LLR, and SLR, for T1 maps (showing 8 example slices). The results are consistent with Figure 6, where VCS‐LLR and SLR can improve image quality compared to LLR, especially with respect to deblurring.

Figure S7. Partial Fourier simulation. Simulated reconstructions from a dataset with 6 TIs, showing the result for one TI (302 ms), demonstrating the effect of different undersampling factors (R = 2 and R = 4) and partial Fourier factors (pF = 0.75, 0.65, 0.57) using different reconstruction methods as a supplement to Figure 6. The first row (FFT, R = 1) serves as the fully sampled reference. The following rows show reconstructions with LLR, VCS‐LLR and SLR algorithms. As the undersampling factor (R) and the partial Fourier factor (pF) increase, more aliasing and artifacts are observed in the LLR reconstruction, which are progressively mitigated by the VCS‐LLR and SLR methods, with SLR showing superior artifact suppression and structure preservation.

Table S1. Statistical comparisons for T1 and T2 mapping in phantom and in vivo experiments. The p‐values from statistical comparisons of T1 and T2 mapping across different reconstruction methods (reference R = 1 FFT, R = 4 LLR, R = 4 SLR, R = 8 LLR, R = 8 SLR) for both white matter (WM) and gray matter (GM) in phantom studies. The in vivo results for T1 (WM) and T2 mapping comparisons are shown. T1 mapping comparisons of GM are not included due to the difficulty of a proper segmentation. No significant differences were observed in most comparisons (p > 0.017, which is considered the significance threshold after Bonferroni correction). The statistical discrepancies in T2 mapping in the phantom at higher undersampling factors (e.g., R = 8) are noted for the SLR algorithm.

MRM-94-119-s001.docx (11.2MB, docx)

Dong Y, Najac C, van Osch MJP, Webb A, Börnert P, Lena B. Rapid quantitative MRI at 46 mT: Accelerated T1 and T2 mapping with low‐rank reconstructions. Magn Reson Med. 2025;94:119‐133. doi: 10.1002/mrm.30442

REFERENCES

  • 1. Marques JP, Simonis FFJ, Webb AG. Low‐field MRI: an MR physics perspective. J Magn Reson Imaging. 2019;49:1528‐1542. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Wald LL, McDaniel PC, Witzel T, Stockmann JP, Cooley CZ. Low‐cost and portable MRI. J Magn Reson Imaging. 2020;52:686‐696. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Arnold TC, Freeman CW, Litt B, Stein JM. Low‐field MRI: clinical promise and challenges. J Magn Reson Imaging. 2023;57:25‐44. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Sarracanie M, Salameh N. Low‐field MRI: how low can we go? A fresh view on an old debate. Front Phys. 2020;8:172. [Google Scholar]
  • 5. O'Reilly T, Teeuwisse WM, de Gans D, Koolstra K, Webb AG. In vivo 3D brain and extremity MRI at 50 mT using a permanent magnet Halbach array. Magn Reson Med. 2021;85:495‐505. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. O'Reilly T, Webb AG. In vivo T1 and T2 relaxation time maps of brain tissue, skeletal muscle, and lipid measured in healthy volunteers at 50 mT. Magn Reson Med. 2022;87:884‐895. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Padormo F, Cawley P, Dillon L, et al. In vivo T1 mapping of neonatal brain tissue at 64 mT. Magn Reson Med. 2023;89:1016‐1025. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Jordanova KV, Martin MN, Ogier SE, Poorman ME, Keenan KE. In vivo quantitative MRI: T1 and T2 measurements of the human brain at 0.064 T. MAGMA. 2023;36:487‐498. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Weiskopf N, Edwards LJ, Helms G, Mohammadi S, Kirilina E. Quantitative magnetic resonance imaging of brain anatomy and in vivo histology. Nat Rev Phys. 2021;3:570‐588. [Google Scholar]
  • 10. Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999;42:952‐962. [PubMed] [Google Scholar]
  • 11. Griswold MA, Jakob PM, Heidemann RM, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002;47:1202‐1210. [DOI] [PubMed] [Google Scholar]
  • 12. Lustig M, Donoho D, Pauly JM. Sparse MRI: the application of compressed sensing for rapid MR imaging. Magn Reson Med. 2007;58:1182‐1195. [DOI] [PubMed] [Google Scholar]
  • 13.Zhao B, Haldar JP, Brinegar C, Liang ZP. Low rank matrix recovery for real‐time cardiac MRI. In: 2010 7th IEEE International Symposium on Biomedical Imaging: From Nano to Macro, ISBI 2010 ‐ Proceedings. 2010;996–999. Article 5490156. [Google Scholar]
  • 14. Zhang T, Pauly JM, Levesque IR. Accelerating parameter mapping with a locally low rank constraint. Magn Reson Med. 2015;73:655‐661. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Shin PJ, Larson PEZ, Ohliger MA, et al. Calibrationless parallel imaging reconstruction based on structured low‐rank matrix completion. Magn Reson Med. 2014;72:959‐970. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16. Jacob M, Mani MP, Ye JC. Structured low‐rank algorithms: theory, magnetic resonance applications, and links to machine learning. IEEE Signal Process Mag. 2020;37:54‐68. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Haldar JP. Low‐rank modeling of local k‐space neighborhoods (LORAKS) for constrained MRI. IEEE Trans Med Imaging. 2014;33:668‐681. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Mani M, Jacob M, Kelley D, Magnotta V. Multi‐shot sensitivity‐encoded diffusion data recovery using structured low‐rank matrix completion (MUSSELS). Magn Reson Med. 2017;78:494‐507. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Dong Y, Koolstra K, Li Z, Riedel M, van Osch MJP, Börnert P. Structured low‐rank reconstruction for navigator‐free water/fat separated multi‐shot diffusion‐weighted EPI. Magn Reson Med. 2023;91:205‐220. [DOI] [PubMed] [Google Scholar]
  • 20. Otazo R, Candès E, Sodickson DK. Low‐rank plus sparse matrix decomposition for accelerated dynamic MRI with separation of background and dynamic components. Magn Reson Med. 2015;73:1125‐1136. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Hu Y, Levine EG, Tian Q, et al. Motion‐robust reconstruction of multishot diffusion‐weighted images without phase estimation through locally low‐rank regularization. Magn Reson Med. 2019;81:1181‐1190. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Haldar JP, Zhuo J. P‐LORAKS: low‐rank modeling of local k‐space neighborhoods with parallel imaging data. Magn Reson Med. 2016;75:1499‐1514. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Bilgic B, Kim TH, Liao C, et al. Improving parallel imaging by jointly reconstructing multi‐contrast data. Magn Reson Med. 2018;80:619‐632. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Kim TH, Setsompop K, Haldar JP. LORAKS makes better SENSE: phase‐constrained partial Fourier SENSE reconstruction without phase calibration. Magn Reson Med. 2017;77:1021‐1035. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Uecker M, Lai P, Murphy MJ, et al. ESPIRiT ‐ an eigenvalue approach to autocalibrating parallel MRI: where SENSE meets GRAPPA. Magn Reson Med. 2014;71:990‐1001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Tamir JI, Uecker M, Chen W, et al. T2 shuffling: sharp, multicontrast, volumetric fast spin‐echo imaging. Magn Reson Med. 2017;77:180‐195. doi: 10.1002/mrm.26102 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. De Vos B, Fuchs P, O'Reilly T, Webb A, Remis R. Gradient coil design and realization for a Halbach‐based MRI system. IEEE Trans Magn. 2020;56:1‐8. [Google Scholar]
  • 28.de Vos B, Parsa J, Abdulrazaq Z, et al. Design, characterisation and performance of an improved portable and sustainable low‐field MRI system. Front Phys. 2021;9:701157.
  • 29. Doneva M, Börnert P, Eggers H, Stehning C, Sénégas J, Mertins A. Compressed sensing reconstruction for magnetic resonance parameter mapping. Magn Reson Med. 2010;64:1114‐1120. [DOI] [PubMed] [Google Scholar]
  • 30. Blaimer M, Gutberlet M, Kellman P, Breuer FA, Köstler H, Griswold MA. Virtual coil concept for improved parallel MRI employing conjugate symmetric signals. Magn Reson Med. 2009;61:93‐102. [DOI] [PubMed] [Google Scholar]
  • 31. Giri S, Chung YC, Merchant A, et al. T2 quantification for improved detection of myocardial edema. J Cardiovasc Magn Reson. 2009;11:56. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Ong F, Lustig M. SigPy: A Python Package for High Performance Iterative Reconstruction. Proceedings of the 27th Annual Meeting of ISMRM, Montréal, Québec, Canada; 2019:4819. [Google Scholar]
  • 33. Iyer SS, Ong F, Cao X, et al. Polynomial preconditioners for regularized linear inverse problems. SIAM J Imaging Sci. 2024;17:116‐146. [Google Scholar]
  • 34. Wu W. Dynamic field mapping and distortion correction using single‐shot blip‐rewound EPI and joint multi‐echo reconstruction. Magn Reson Med. 2024;92:82‐97. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35. Prah DE, Paulson ES, Nencka AS, Schmainda KM. A simple method for rectified noise floor suppression: phase‐corrected real data reconstruction with application to diffusion‐weighted imaging. Magn Reson Med. 2010;64:418‐429. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Avants B, Tustison N, Song G. Advanced normalization tools: V1.0. Insight J. 2009. Accessed January 06, 2025. http://hdl.handle.net/10380/3113 [Google Scholar]
  • 37. Bottomley PA, Foster TH, Argersinger RE, Pfeifer LM. A review of normal tissue hydrogen NMR relaxation times and relaxation mechanisms from 1‐100 MHz: dependence on tissue type, NMR frequency, temperature, species, excision, and age. Med Phys. 1984;11:425‐448. [DOI] [PubMed] [Google Scholar]
  • 38. Liu Y, Leong ATL, Zhao Y, et al. A low‐cost and shielding‐free ultra‐low‐field brain MRI scanner. Nat Commun. 2021;12:7238. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39. Cooley CZ, McDaniel PC, Stockmann JP, et al. A portable scanner for brain MRI. Nat Biomed Eng. 2021;5:229‐239. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40. Pruessmann KP, Weiger M, Börnert P, Boesiger P. Advances in sensitivity encoding with arbitrary k‐space trajectories. Magn Reson Med. 2001;46:638‐651. [DOI] [PubMed] [Google Scholar]
  • 41. Ilicak E, Saritas EU, Çukur T. Automated parameter selection for accelerated MRI reconstruction via low‐rank modeling of local k‐space neighborhoods. Z Med Phys. 2023;33:203‐219. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42. Ronneberger O, Fischer P, Brox T. U‐net: convolutional networks for biomedical image segmentation. In: Navab N, Hornegger J, Wells W, Frangi A, eds. Medical Image Computing and Computer‐Assisted Intervention (MICCAI) 2015. Lecture Notes in Computer Science. (Including Subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics). Vol 9351. Springer; 2015. doi: 10.1007/978-3-319-24574-4_28 [DOI] [Google Scholar]
  • 43. Sudre CH, Li W, Vercauteren T, Ourselin S, Jorge CM. Generalised dice overlap as a deep learning loss function for highly unbalanced segmentations. In: Cardoso M et al., eds. Deep Learning in Medical Image Analysis and Multimodal Learning for Clinical Decision Support. DLMIA ML‐CDS 2017. Lecture Notes in Computer Science (Including Subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics). Vol 10553. Springer; 2017:240‐248. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44. Ben‐Eliezer N, Sodickson DK, Block KT. Rapid and accurate T2 mapping from multi‐spin‐echo data using bloch‐simulation‐based reconstruction. Magn Reson Med. 2015;73:809‐817. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45. Daniel G, Meirav G, Noam O, et al. Fast and accurate T2 mapping using Bloch simulations and low‐rank plus sparse matrix decomposition. Magn Reson Imaging. 2023;98:66‐75. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Figure S1. Evaluation of Locally Low‐Rank (LLR) reconstruction quality for T1 mapping data across different patch sizes and regularization factors. The plots display the Structural Similarity Index Measure (SSIM) and normalized Root Mean Square Error (nRMSE) as functions of the regularization factor, ranging from 2 to 120 in steps of 2, for patch sizes of [8 × 8 × 4], [10 × 10 × 4], and [12 × 12 × 4] for reconstructing the retrospectively undersampled R = 4 dataset. The SSIM and nRMSE metrics are calculated between the R = 4 LLR‐reconstructed images and the reference R = 1 images to assess structural similarity and reconstruction fidelity, respectively. The optimal parameters were determined to be a patch size of [10 × 10 × 4] with regularization factor of 20, as both metrics (SSIM and nRMSE) exhibited their best values at this configuration. This patch size corresponds well with the spatial resolution used in the study (2.5 × 2.5 × 5 mm3). Since it was not possible to acquire (R = 1) 1.5 × 1.5 × 5 mm3 resolution data for optimizing the simulation, the same optimal parameters were applied to the high‐resolution case.

Figure S2. Evaluation of structured low‐rank (SLR) reconstruction quality for T1 mapping data across different ranks. The plots show the Structural Similarity Index Measure (SSIM) and the Normalized Root Mean Square Error (nRMSE) as a function of the rank selection, highlighting its impact on the reconstruction quality for the SLR algorithm. The optimal rank value was found to be 120, where SSIM reaches its highest peak and nRMSE reaches its second lowest value (0.02134), very close to the absolute minimum (at rank = 110, 0.02125). This balance achieves the best trade‐off between SSIM and nRMSE, indicating high structural fidelity and minimal reconstruction error.

Figure S3. Comparison of phantom T1 mapping and T2 mapping retrospectively undersampled data and prospectively undersampled data (R = 4). One concern for current sequence development is that the implementation of undersampling in the point‐of‐care ultra‐low field MRI system may introduce inconsistencies in the k‐space data. The T1 and T2 results show both low nRMSE values and mapping quality between the retrospectively undersampled and prospectively undersampled data, indicating that the undersampling strategy does not introduce noticeable inconsistencies into the k‐space data, despite the challenges of ultra‐low field MRI hardware.

Figure S4. Retrospective simulation study on in vivo T1 mapping. Comparison of retrospective acceleration factors (R = 2, 4, 6, 8) on images and T1 maps generated from one slice of R = 1 data using LLR and SLR reconstructions. The figure shows images from a sample TI = 302 ms, T1 maps from six different TIs, and the corresponding difference maps between the reference (R = 1) and each reconstructed image/T1 map. As the undersampling factor increases, more visible structures appear on the difference maps of the images and larger discrepancies are observed on the T1 maps, reflecting the more difficult reconstruction conditions at higher undersampling factors. The nRMSE values for the LLR images were 0.037, 0.050, 0.058, and 0.065, and for the SLR images were 0.034, 0.047, 0.056, and 0.064 at the respective undersampling factors. For the T1 maps (excluding CSF regions), the nRMSE values were 0.022, 0.028, 0.032, and 0.038 for LLR and 0.020, 0.027, 0.032, and 0.038 for SLR. This study illustrates how both algorithms handle undersampling artifacts, with both methods effectively mitigating artifacts at lower acceleration factors, particularly at R = 2, where no significant structural loss is observed in the difference maps compared to the R = 1 reference.

Figure S5. In‐vivo prospective T2 mapping of R = 1 and R = 4. (A) An example slice of individual TE images for T2 mapping, reconstructed using FFT for fully sampled (R = 1) and undersampled (R = 4) data, and using LLR and SLR reconstruction methods for undersampled data (R = 4). Undersampling artifacts visible in the FFT reconstruction (R = 4) are effectively reduced in the LLR and SLR reconstructions, resulting in improved image quality and structural clarity (red arrows). (B) T2 maps generated from the corresponding reconstructions. Note that the partial volume effects with the lower SNR compared to the T1 mapping results (Figure 2) result in visible contrast changes, especially around the CSF regions (red arrows). GM and WM T2 values are not readily distinguishable due to their close values, but are consistent with reference R = 1 data (Table 2) and agree with previous observations in the literature. 6 , 8 , 37

Figure S6. T1 maps of the partial Fourier experiment as a supplement to Figure 6. Comparison of different partial Fourier reconstruction methods, including LLR, VCS‐LLR, and SLR, for T1 maps (showing 8 example slices). The results are consistent with Figure 6, where VCS‐LLR and SLR can improve image quality compared to LLR, especially with respect to deblurring.

Figure S7. Partial Fourier simulation. Simulated reconstructions from a dataset with 6 TIs, showing the result for one TI (302 ms), demonstrating the effect of different undersampling factors (R = 2 and R = 4) and partial Fourier factors (pF = 0.75, 0.65, 0.57) using different reconstruction methods as a supplement to Figure 6. The first row (FFT, R = 1) serves as the fully sampled reference. The following rows show reconstructions with LLR, VCS‐LLR and SLR algorithms. As the undersampling factor (R) and the partial Fourier factor (pF) increase, more aliasing and artifacts are observed in the LLR reconstruction, which are progressively mitigated by the VCS‐LLR and SLR methods, with SLR showing superior artifact suppression and structure preservation.

Table S1. Statistical comparisons for T1 and T2 mapping in phantom and in vivo experiments. The p‐values from statistical comparisons of T1 and T2 mapping across different reconstruction methods (reference R = 1 FFT, R = 4 LLR, R = 4 SLR, R = 8 LLR, R = 8 SLR) for both white matter (WM) and gray matter (GM) in phantom studies. The in vivo results for T1 (WM) and T2 mapping comparisons are shown. T1 mapping comparisons of GM are not included due to the difficulty of a proper segmentation. No significant differences were observed in most comparisons (p > 0.017, which is considered the significance threshold after Bonferroni correction). The statistical discrepancies in T2 mapping in the phantom at higher undersampling factors (e.g., R = 8) are noted for the SLR algorithm.

MRM-94-119-s001.docx (11.2MB, docx)

Articles from Magnetic Resonance in Medicine are provided here courtesy of Wiley

RESOURCES