MPRAGE like : A novel approach to generate T1w images from multi‐contrast gradient echo images for brain segmentation

Marc‐Antoine Fortin 1,✉, Rüdiger Stirnberg 2, Yannik Völzke 2, Laurent Lamalle 3, Eberhard Pracht 2, Daniel Löwen 2, Tony Stöcker 2,4, Pål Erik Goa 1,5
PMCID: PMC12021339  PMID: 39902546

Abstract

Purpose

Brain segmentation and multi‐parameter mapping (MPM) are important steps in neurodegenerative disease characterization. However, acquiring both a high‐resolution T1w sequence like MPRAGE (standard input to brain segmentation) and an MPM in the same neuroimaging protocol increases scan time and patient discomfort, making it difficult to combine both in clinical examinations.

Methods

A novel approach to synthesize T1w images from MPM images, named MPRAGE like , is proposed and compared to the standard technique used to produce synthetic MPRAGE images (synMPRAGE). Twenty‐three healthy subjects were scanned with the same imaging protocol at three different 7T sites using universal parallel transmit RF pulses. SNR, CNR, and automatic brain segmentation results from both MPRAGE like and synMPRAGE were compared against an acquired MPRAGE.

Results

The proposed MPRAGE like technique produced higher SNR values than synMPRAGE for all regions evaluated while also having higher CNR values for subcortical structures. MPRAGE was still the image with the highest SNR values overall. For automatic brain segmentation, MPRAGE like outperformed synMPRAGE when compared to MPRAGE (median Dice Similarity Coefficient of 0.90 versus 0.29 and Average Asymmetric Surface Distance of 0.33 versus 2.93 mm, respectively), in addition to being simple, flexible, and considerably more robust to low image quality than synMPRAGE.

Conclusion

The MPRAGE like technique can provide a better and more reliable alternative to synMPRAGE as a substitute for MPRAGE, especially when automatic brain segmentation is of interest and scan time is limited.

Keywords: brain segmentation, multi‐contrast, multi‐parameter mapping, neuroimaging, synthetic MPRAGE, UHF‐MRI

1. INTRODUCTION

Magnetization‐Prepared Rapid Acquisition of Gradient Echoes (MPRAGE) has excellent white and gray matter (WM‐GM) contrast and short acquisition time. 1 It is therefore the preferred anatomical reference image in most neuroimaging studies. FreeSurfer, 2 which provides automated brain segmentation from T1w MPRAGE images at 1.5T and 3T, has been widely used in large‐scale neuroimaging projects 3 and, recently, a Deep Learning (DL)‐based version called FastSurferVINN 4 has been proposed, which drastically reduces computational time.

In the past decade, Ultra‐High Field Magnetic Resonance Imaging (UHF‐MRI) (i.e., ≥ 7T) has become more widely used in neuroimaging studies due to its higher SNR, contrast, and spatial resolution. 5 However, UHF‐MRI suffers from larger transmit RF inhomogeneities than those observed at lower field strengths due to the shorter RF wavelength at 7T. 6 This is especially problematic for the RF field used for excitation (B1+), resulting in considerable inhomogeneity in both signal intensity and tissue contrast across the field of view (FOV). 7 To compensate for this effect, a more time‐consuming MP2RAGE sequence is often utilized at UHF. 8 The MP2RAGE sequence acquires two images, one at a short and one at a long inversion time, which are combined to calculate a third and more homogeneous image often referred to as the UNI image. A regularization term λ can be added to the calculation of the UNI image to suppress background noise at the cost of re‐introducing a small amount of inhomogeneity. 9 Additionally, a simpler alternative to MP2RAGE proposes to use a non‐interleaved GRE acquisition to remove the PD, T2∗, and receive sensitivity weightings on MPRAGE images at 7T. 10

Recent progress in RF pulse design has enabled parallel transmission (pTx) RF pulses that considerably reduce the amount of signal inhomogeneities at UHF. One of the most promising pTx techniques applies calibration‐free, universal pTx‐pulses (UP), 11 which are pre‐computed pTx‐pulses optimized offline across multiple subjects, removing the need for time‐consuming pTx‐pulse optimization during each MR exam.

Advanced neuroimaging protocols such as applied in Alzheimer's 12 and Parkinson's Diseases 13 research has recently been extended to include quantitative parameter mapping methods, such as Multi‐Parameter Mapping (MPM) 14 and Variable Flip Angle (VFA) imaging. 15 These methods acquire multi‐contrast spoiled Gradient Echo (GRE) images in relatively short time for the calculation of quantitative maps (qMaps) 16 and have become valuable tools for multi‐center neuroimaging studies. 17 , 18 , 19 However, MPRAGE is usually still included due to its high importance as an anatomical reference and basis for automatic segmentation.

When combining anatomical multi‐contrast imaging and quantitative parameter mapping, neuroimaging protocols can become prohibitively long. To avoid the MPRAGE acquisition, several techniques to create “synthetic” T1w images have been proposed. 20 , 21 , 22 , 23 One of them proposes to derive synthetic MPRAGE (synMPRAGE) images from a combination of RInline graphic, Proton Density (PD), and RInline graphic maps. 21 A recent study using this technique has shown the clinical relevance of synMPRAGE images in spinal cord neurodegeneration characterization at 3T. 24 Furthermore, techniques using DL have been proposed to synthesize MPRAGE images from multi‐echo (ME) GRE images 22 or even from any MR image. 23 In Reference 22, high correlation of segmentation and susceptibility results produced from using either the synthetic T1w or MPRAGE images was presented.

However, all synthetic T1w techniques proposed in the literature require either (1) the computation of qMaps beforehand or (2) DL training, both of which are time‐consuming and require highly specialized processing pipelines. Additionally, the usage of synthetic T1w contrasts at UHF‐MRI has been quite sparse and has shown unsatisfactory results when tested, 25 necessitating novel approaches to be developed at UHF‐MRI. To the best of our knowledge, no technique has been proposed thus far to synthesize T1w contrast images directly from MPM images without the need for advanced preprocessing.

In this work, we propose MPRAGE like , a novel and straightforward method to synthesize T1w images from MPM images acquired with 3D‐EPI readouts. We test it against the most commonly used technique to generate synMPRAGE images, and we compare the results of automatic brain segmentation based on MPRAGE like images with those obtained from UP‐MPRAGE acquisitions. All results reported were obtained at 7T. The name MPRAGE like is inspired by its visual resemblance to the acquired MPRAGE and its primary intended use for automatic segmentation in neuroimaging pipelines, despite them being produced in entirely different ways.

2. METHODS

2.1. Subjects

A total of 23 healthy volunteers were recruited (male/female: 16/7, age: 30.2 ± 7.5 years old) at three partner sites; nine at site A (NTNU, Trondheim, Norway), ten at site B (DZNE, Bonn, Germany), and four at site C (GIGA‐CRC, Liège, Belgium). The study was approved by the local review boards of each site, and volunteers signed a written informed consent form before scanning.

2.2. Image acquisition

Images were acquired on Siemens MAGNETOM Terra (sites A and C) and MAGNETOM 7TPlus (site B) scanners. All sites used identical 8Tx/32Rx head coils (Nova Medical, Wilmington, MA) in pTx mode. The imaging protocol included three contrast acquisitions for MPM, MPRAGE and a flip‐angle map. All scans used UPs created from a database of B0 and B1+ maps acquired from 10/10/9 subjects at sites A/B/C in a preceding study.

The MPM protocol used a multi‐echo (ME) skipped‐CAIPI 3D‐EPI sequence 16 , 26 with universal GRAPE 27 binomial‐121 water excitation (1.54 ms pulse duration) to acquire images with MTw, PDw, and T1w contrasts. The scan parameters were the following: voxel size = (0.6 mm)Inline graphic, TE = 4.7, 11.8, 18.9, 26.0 ms for four TEs, 80% phase partial Fourier sampling, 1×3z1 CAIPIRINHA parallel imaging, 28 EPI factor of 4, MTw/PDw/T1w nominal flip angle = 6°/5°/25°, TR = 50.5/34.0/34.0 ms, TA = 6:00/4:05/4:05 min (total scan time 14:10 min). The MTw scan used PUSHUP pTx saturation pulses, 29 which is a UP variation of previously proposed PUSH saturation 30 composed of two Gaussian‐shaped sub‐pulses with a duration of 5 ms each with a target B1,rms=1.7 μT.

The scan parameters for the MPRAGE were the following: voxel size = (0.6 mm)Inline graphic, TE = 2.14 ms, 1×2z1 CAIPIRINHA parallel imaging, elliptical sampling, nominal flip angle = 5°, TI = 1100 ms, TR = 2700 ms, TA = 7:24 min.

The Actual Flip Angle Imaging (AFI) 31 technique was used to map the local flip angle of the MPM excitation pulse. Scan parameters were the following: voxel size = (4.0 mm)Inline graphic, TE = 2.04 ms, 2×2z1 CAIPIRINHA parallel imaging, nominal flip angle = 45°, TR = 20/200 ms, TA = 1:50 min.

2.3. Data processing

The MPM magnitude images and B1+ map were processed using the hMRI toolbox (v0.6.1) 32 embedded in SPM12 33 to first coregister the MPM images with a rigid registration step, then calculate the B1+‐corrected RInline graphic, PD, MTInline graphic, and RInline graphic maps. Additionally, the MPRAGE images were rigidly coregistered to the 1st TE MPM magnitude images to correct for inter‐scan motion from subjects. More details about the individual steps are available in Reference 32.

2.4. MPRAGE like

The raw T1w MPM images do not provide sufficient morphometric details to be used for visual assessment or automatic brain segmentation. On the other hand, PDw and MTw contrasts in the MPM protocol provide an approximately “inverted” contrast compared to the T1w. The rationale behind calculating MPRAGE like images is (1) to increase the T1w contrast‐to‐noise ratio (CNR) by dividing the T1w image by the approximately “inverted” contrast images acquired in the MPM protocol, (2) thereby reducing receive bias and residual transmit bias, and (3) canceling out T2∗ bias in accordance with Reference 8, without the need to calculate qMaps. The following three equations for generating MPRAGE like images were tested:

MPRAGElike,all=T1w−λ0.5·(PDw+MTw)+λ, (1)
MPRAGElike,MT=T1w−λMTw+λ, (2)
MPRAGElike,PD=T1w−λPDw+λ, (3)

where PDw, MTw, and T1w are the 1st TE magnitude MPM images and λ the regularization term. All intensity values < 0 and > 500 were set to 0.

Following the standard MPM theory and derivations of signal equations as described in References 14 and 34 (and ignoring λ), we can obtain the following receive bias‐free approximations for the three signal equations:

MPRAGElike,all≅αnomT1TRT1αnomT12+2·TRT1·R1app0.5×αnomMTTRMTαnomMT2+2·(δapp+TRMT·R1app)+αnomPDTRPDαnomPD2+2·TRPD·R1app, (4)
MPRAGElike,MT≅αnomT1TRT1αnomMTTRMT·αnomMT2+2·(δapp+TRMT·R1app)αnomT12+2·TRT1·R1app (5)
MPRAGElike,PD≅αnomT1αnomPD·αnomPD2+2·TRPD·R1appαnomT12+2·TRT1·R1app, (6)

where αnomX and TRX are the nominal flip angle and TR values of the T1w, MTw, and PDw contrasts, respectively, R1app, the apparent longitudinal relaxation rate, and δapp, the apparent MT saturation. For more details about the derivation of Equations (4), (5), and (6), we refer the reader to Supporting Information. Note that Equations (4) through (6) are free of receive bias, but they are not free of potential signal inhomogeneities, as both R1app and δapp depend on B1+.

2.5. synMPRAGE

For the calculation of synMPRAGE images, the B1+‐corrected R1 map computed by the hMRI toolbox in addition to the sequence parameters of the acquired MPRAGE protocol (provided in section 2.2) were used. More precisely, these values were substituted in eq. 3a of Reference 21 in addition to the PDw with both receive and transmit bias contributions being removed by assuming a constant value of 1 (M0 = 1, RP = 1, and B1 = 1 in eq. 3a of Reference 21). This version of the synMPRAGE equation was selected to produce a superior optical contrast 21 and to create a “pure T1w”, free of PD, T2∗, and inhomogeneities like for the MPRAGE like .

The complete processing pipeline used in this study is shown in Figure 1.

FIGURE 1.

FIGURE 1

Processing pipeline used in this study to compute synMPRAGE and MPRAGE like images. The 1st TE MPM images from each contrast were first registered and then combined to produce the MPRAGE like,all (i), MPRAGElike, MT (ii), and MPRAGE like,PD (iii) images with different λ values. In order to compute the R1 map necessary for the creation of synMPRAGE images, all 12 magnitude volumes were given as inputs to the hMRI toolbox (v0.6.1). 32 The MPRAGE images were also coregistered through hMRI to correct for inter‐scan motion. The background noise in the synMPRAGE was partially masked by removing negative outliers from the PD map, thresholding it with Otsu's method 35 and applying morphological operations (dilation and closing) to create a mask applied on the R1 map. More details about the synMPRAGE calculation are given in Reference 21.

2.6. Image analysis

First, to assess the impact of the value of the regularization term λ on the MPRAGE like images, the following λ values were tested: 0, 50, 100, 200, 300, 400, and 500. For all λ values and MPRAGE like equations, the voxel‐wise relative difference in signal intensity (Δs/s) between the λ = 0 image and other λ values was computed with the following equation:

Δs/s=100×MPRAGElike(λ)−MPRAGElike(λ=0)MPRAGElike(λ=0), (7)

where MPRAGE like (λ) is one of the regularized images and MPRAGE like (λ = 0) is the non‐regularized image.

Moreover, to determine an optimal λ, the Structural Similarity Index Metric (SSIM) 36 was computed for each MPRAGE like equation and λ with the MPRAGE as the reference image for all subjects in this study. The SSIM was calculated with the following equation:

SSIM(x,y)=(2μxμy+C1)(2σxσy+C2)(μx2+μy2+C1)(σx2+σy2+C2)×(covxy+C3)(σxσy+C3), (8)

where x and y are the two compared images, μx and μy are the mean values of the images x and y, σx and σy the standard deviation of images x and y, covxy the covariance between images x and y, and C1, C2, and C3 are constants for stabilization in case of weak denominators automatically set for each image as explained in Reference 36.

Once the value of λ that yielded the largest agreement with the reference MPRAGE image was found, the SNR and CNR in several regions of interest (ROI) inside the brain were computed for the MPRAGE, three MPRAGE like equations, and synMPRAGE for all subjects. The SNR and CNR equations used were the following:

SNRA=SA‾σA, (9)
CNRA−B=|SA‾−SB‾|σA2+σB2, (10)

where SA‾ is the mean and σA the standard deviation within ROI (or tissue) A. For this comparison, the reference MPRAGE‐based segmentation was used for all images to avoid effects of segmentation differences.

To correct for remaining signal inhomogeneities in the MPRAGE, MPRAGE like , and synMPRAGE images, N4 bias field correction 37 , 38 was applied as a separate postprocessing steps and the resulting images were included in the further analysis along with the non‐bias‐field corrected images.

To assess the usability of MPRAGE like images for neuroimaging applications, the quality of automatic brain segmentation from MPRAGE like images was evaluated. Both the N4‐corrected and uncorrected MPRAGE, MPRAGE like , and synMPRAGE images were segmented using the FastSurfer software (v2.3.0) with the “voxel size independent neural network” (FastSurferVINN) 39 feature to produce segmentation of brain regions at native input resolution. The N4‐corrected MPRAGE‐based segmentation acted as the reference. The Dice Similarity Coefficient (DSC) 40 , 41 and Average Symmetric Surface Distance (ASSD) 42 metrics were used and computed with the following equations:

DSC=2×|G∪P|G+P, (11)

where G is the reference segmentation (N4‐corrected MPRAGE), and P, the segmentation from one of the other synthetic T1w images tested in this work, and

ASSD=∑i=1NGdG→P,i+∑i=1NPdP→G,iNG+NP. (12)

Herein, dG→P,i is the distance from point i on the surface of the segment from the N4‐corrected MPRAGE to its nearest point on the surface of the segment from one of the other synthetic T1w images tested in this work; dP→G,i is the distance from point i on the surface of the segment from one of the other synthetic T1w images tested in this work to its nearest point on the surface of the segment from the N4‐corrected MPRAGE; NG and NP are the total number of points on each surface, respectively.

All regions except WM hypointensities (usually negligible in young healthy adults) were used in the calculation of DSC and ASSD. DSC and ASSD values were set to 0 and NaN, respectively for regions where no voxels were included in the given segment.

Finally, the volumes of selected subcortical structures were computed to assess the relative differences in volume against the N4‐corrected MPRAGE (reference) for all three N4‐corrected MPRAGE like images. For statistical evaluation of volume differences, a two‐sided Wilcoxon signed rank test 43 with Bonferroni correction 44 was applied.

3. RESULTS

3.1. Optimization of regularization term λ

The impact of λ on the MPRAGE like images is shown in Figure 2. Qualitatively, one can observe that with increasing value of λ, the background noise is reduced while the spatial signal inhomogeneity is increased.

FIGURE 2.

FIGURE 2

Impact of λ on signal intensity of MPRAGE like images for the λ values tested in this study for one example subject. MPRAGE like,all images are shown in the left column with their respective Δs/s image on the right (equation shown on the top right). Only the images for Equation (1) are shown due to negligible differences in results between Equations (1), (2), and (3).

Quantitative comparisons performed using the average SSIM are shown in Figure 3. The highest SSIM value was reached at λ  = 100 for MPRAGE like,all and MPRAGE like,PD , and λ  = 50 for MPRAGE like,MT .

FIGURE 3.

FIGURE 3

SSIM (maximum: 1.0, minimum: −1.0) computed between the MPRAGE and all λ values tested for the three MPRAGE like equations proposed in this work. The dots represent the mean SSIM calculated across all subjects for each λ with the error bars representing the standard deviation.

Based on these results, λ = 100 was selected as the best regularization term for sufficient background noise suppression while minimally impacting the image contrast.

3.2. Qualitative and quantitative comparisons between the acquired MPRAGE, MPRAGE like and synMPRAGE

The acquired MPRAGE, three MPRAGE like images (at λ  = 100), and synMPRAGE for one representative subject are shown in Figure 4. All MPRAGE like images appeared visually similar to each other and to the acquired MPRAGE. A slight variation in contrast was observed between the three MPRAGE like where better overall WM‐GM visual contrast was observed for the MPRAGE like,MT compared to the two other MPRAGE like . One visible difference between MPRAGE like images and the MPRAGE was the brightness of blood vessels in the latter compared to the MPRAGE like . Moreover, receive bias can be observed (red arrows) on the MPRAGE image, which is not present in the other images. The latter, in turn, exhibit different, broad patterns of slight signal inhomogeneities. synMPRAGE showed less‐defined subcortical GM structures compared to MPRAGE like and MPRAGE with visibly more noise. MPRAGE like and synMPRAGE images exhibited signal brightening in the cortex above the sinuses, which was not observed in the MPRAGE as shown with the red arrows in the bottom row. A significant amount of noise was observed in the skull and skin regions surrounding the brain in the synMPRAGE images compared to the MPRAGE like images and MPRAGE.

FIGURE 4.

FIGURE 4

Qualitative comparison between the acquired MPRAGE, three MPRAGE like , and synMPRAGE (not N4‐corrected) in the three anatomical planes with a zoomed‐in version of the axial plane in the second row for one example subject. Green arrows show well‐defined structures, whereas red arrows show issues in the respective images. The MPRAGE like images shown have a λ value of 100.

SNR values are reported in Table 1. Overall, MPRAGE had the highest SNR, the different versions of MPRAGE like slightly lower, while synMPRAGE had systematically the lowest SNR for all ROIs.

TABLE 1.

Mean SNR (± standard deviation) calculated across all subjects for different regions of interest. The non N4‐corrected images were used for the calculations. The largest SNR value between all images for each region is shown in bold.

SNR MPRAGE MPRAGE like,all MPRAGE like,MT MPRAGE like,PD synMPRAGE
Cerebral WM 6.7 ± 0.7 8.0
±
1.1
7.2 ± 0.8 8.0 ± 1.2 5.3 ± 0.9
Cerebral cortex 5.5 ± 0.7 5.9 ± 0.7 5.4 ± 0.5 6.0
±
0.8
3.7 ± 0.5
Cerebellum WM 13.7
±
2.3
8.3 ± 1.5 7.5 ± 1.1 8.3 ± 1.5 4.4 ± 1.0
Cerebellum cortex 6.8
±
1.3
5.3 ± 0.7 4.9 ± 0.6 5.5 ± 0.9 2.7 ± 0.4
Putamen 13.5
±
1.9
9.0 ± 1.1 7.9 ± 0.9 9.4 ± 1.1 3.8 ± 0.8
Caudate 14.0
±
2.1
8.8 ± 0.8 7.7 ± 0.7 9.2 ± 0.9 3.1 ± 0.5

Regarding CNR, shown in Table 2, the highest CNR values were obtained by synMPRAGE for the cortex‐WM, whereas MPRAGE like,MT had the highest values for the remaining regions of interest.

TABLE 2.

Mean CNR (± standard deviation) calculated across all subjects for different gray‐white matter adjacent regions. The non N4‐corrected images were used for the calculations. The largest CNR value between all images for each region is shown in bold.

CNR MPRAGE MPRAGE like,all MPRAGE like,MT MPRAGE like,PD synMPRAGE
Cortex‐WM 0.89 ± 0.24 1.45 ± 0.22 1.60 ± 0.22 1.20 ± 0.22 1.71
±
0.31
Cerebellum Cortex‐WM 0.94 ± 0.16 1.12 ± 0.33 1.29
±
0.35
0.87 ± 0.29 1.14 ± 0.24
Caudate‐WM 1.76 ± 0.15 1.78 ± 0.26 1.89
±
0.23
1.52 ± 0.27 1.15 ± 0.34
Putamen‐WM 1.56 ± 0.17 1.66 ± 0.21 1.72
±
0.22
1.46 ± 0.22 1.01 ± 0.28

3.3. FastSurferVINN segmentation results across different MPRAGE‐images

The automatic brain segmentation produced by FastSurferVINN for N4‐corrected MPRAGE, MPRAGE like , and synMPRAGE in one example subject are shown in Figure 5. Visually, the MPRAGE like ‐based segmentation appears very similar to the MPRAGE‐based one, while the synMPRAGE‐based segmentation appears different in various regions.

FIGURE 5.

FIGURE 5

Input images and their corresponding FastSurferVINN segmentation results in the axial and sagittal planes for the MPRAGE versus MPRAGE like (λ = 100) and synMPRAGE (all N4‐corrected). The cortical region next to the sinuses appears to be challenging for FastSurferVINN even for the N4‐corrected MPRAGE where no signal brightening but signal dropouts were observed compared to the MPRAGE like and synMPRAGE images. Thin cerebellum WM branches are not fully segmented (red arrows, last row) for all images. Segmentation of subcortical nuclei in the middle brain appears slightly different between the MPRAGE and MPRAGE like images (yellow arrow). For synMPRAGE, significant visual differences with MPRAGE like are observed. For instance, segmentation in the frontal lobe was difficult, and, in addition, a significant amount of noisy WM regions were segmented as WM‐hypointensities (both identified with red arrows). Several subcortical structures are not correctly segmented compared to the MPRAGE and MPRAGE like . For example, the caudate, putamen, and pallidum are barely visible on the axial slice, whereas it are clearly visible for MPRAGE and MPRAGE like . Moreover, a significant portion of the WM and GM over the sinuses is lacking.

To assess the segmentation results in more detail, a close‐up view of the axial plane is shown in Figure 6. Fine structures like small cortical sulci benefit from the (0.6 mm)Inline graphic resolution. Overall, the segmentation of subcortical nuclei is very similar between the MPRAGE and MPRAGE like,all , while differences are observed mainly in the frontal cortex.

FIGURE 6.

FIGURE 6

Close‐up view of the N4‐MPRAGE and N4‐MPRAGE like,all images in the axial plane (first row) with their corresponding segmentation results (second row). Green arrows represent some discrepancies in the cortex segmentation between the N4‐MPRAGE and N4‐MPRAGE like,all and yellow arrows show the impact of bright blood vessels on the cortex segmentation for the N4‐MPRAGE. Structures segmented are shown in the legend at the bottom. Very narrow parts of the lateral ventricles (one voxel width) are depicted on both images. Only the N4‐MPRAGE like,all is shown due to negligible differences between MPRAGE like segmentations.

The average DSC and ASSD calculated against the reference N4‐corrected MPRAGE across all subjects are shown with box plots in Figure 7A,C ((B) and (D) provide a zoom‐in version of (A) and (C), respectively). For both metrics, the non N4‐corrected MPRAGE was the closest match to the reference, with all MPRAGE like variations also being highly similar to MPRAGE. All MPRAGE like had a median DSC greater than 0.85. Regarding ASSD, all variations of MPRAGE like images had a median value below 0.6 mm (i.e., < one voxel). For synMPRAGE, the results were substantially worse; below 0.3 for DSC and greater than 2.5 mm for ASSD for both synMPRAGE versions. One subject was excluded from the DSC and ASSD results due to label errors from FastSurferVINN making the output segmentations unusable.

FIGURE 7.

FIGURE 7

Box plots of (A) the DSC (ideal: 1.0) and (C) ASSD (ideal: 0.0) values calculated across all subjects for every input image tested in this work. A zoomed‐in version (light blue rectangle) of plots (A) and (C) are shown in (B) and (D), respectively. The black line in each box corresponds to the median, the boxes are the interquartile ranges (IQR), the whiskers are the 5th and 95th percentiles, with the circles correspond to outliers.

The relative difference in volume for N4‐corrected MPRAGE like images compared to the reference N4‐corrected MPRAGE for selected subcortical nuclei are plotted in Figure 8. Most relative differences were below ±5%. The non‐corrected MPRAGE was the technique with the highest number of significant differences in median volumes with 4. N4‐MPRAGE like,MT had 3, N4‐MPRAGE like,all 2, whereas N4‐MPRAGE like,PD had no volume difference that was statistically significant for all ROIs tested.

FIGURE 8.

FIGURE 8

Median relative difference (%) in the volumes calculated across all subjects for the raw MPRAGE and N4‐corrected MPRAGE like images tested against the N4‐corrected MPRAGE for selected subcortical GM nuclei. The * represents methods with a P‐value < the Bonferroni corrected significance threshold of 0.0025.

4. DISCUSSION

This study explores the feasibility of synthesizing images of pure T1w contrast directly from MPM or VFA source images without the requirement to calculate quantitative parameters maps beforehand. The proposed MPRAGE like images provide a robust surrogate to acquired MPRAGE for automatic brain segmentation with sub‐millimeter resolution at 7T. All MPRAGE like equations with λ = 100 have shown to be more robust and high‐quality substitutes to MPRAGE than the previously proposed synMPRAGE.

4.1. Optimization of regularization term λ

Selecting the optimal λ depends on the overall scaling of image intensities and is a trade‐off between removing the background noise and preserving signal homogeneity across the brain, primarily in the cerebellum and temporal lobes, as also reported in Reference 9. In this work, a λ value of 100 was adopted due to the fact that (1) the vast majority of the background noise was removed, (2) there was minimal signal inhomogeneity reintroduced in the brain, and (3) two of the three MPRAGE like equations achieved their highest SSIM value at that λ. For MPRAGE like,MT , λ = 100 was the second largest SSIM, which was very close to the value at λ = 50. Whole‐brain image intensities at the first TE of MTw, T1w, and PDw data in this study roughly ranged between 500 and 2500. For differently acquired/scaled images, correspondingly scaled λ values are expected to perform comparably.

The SSIM values reported in Figure 3 are arguably low compared to other techniques to produce synthetic MPRAGE images. 22 SSIM is often used in contexts where one wants to evaluate the impact of a “lossy” compression process on an uncompressed image or the quality of an image restoration process. 45 However, in this work, the SSIM was computed against the acquired MPRAGE, which was a different image altogether, acquired independently from the MPM images. Furthermore, the entire field of view was evaluated, which is why a large fraction of voxels outside the brain presumably dominate the SSIM metric. Considering this, the low SSIM values obtained are not surprising nor concerning. SSIM was chosen because a metric that would convey a sense of perceived similarity over a pure quantitative similarity was desired, especially since the voxel values of an MPRAGE image do not convey any direct quantitative meaning.

4.2. Qualitative and quantitative comparisons between the acquired MPRAGE, MPRAGE like , and synMPRAGE

Qualitatively, all MPRAGE like images looked quite similar to each other. The added MTw contrast, both for the MPRAGE like,all and MPRAGE like,MT , helped improve the visual contrast between subcortical nuclei and WM compared to the MPRAGE like,PD . The synMPRAGE image posed further challenges. As seen from Figure 4, the noisy background of synMPRAGE images was not possible to mask completely. Several combinations of masking techniques as removing outliers and negative values for quantitative parameters were tested in this work but none ended up efficiently masking the background noise without removing relevant anatomy inside the brain. Also, no simple regularization approach can be applied on the synMPRAGE images as done for MPRAGE like . Consequently, the remaining noisy regions next to CSF around the brain made it difficult to properly detect borders of structures as seen with the segmentation results (Figure 5). It is noteworthy that RInline graphic mapping at high resolution in very limited scan time is very challenging, and more sophisticated fitting algorithms compared to the MPM equations, such as Bayesian fitting or other approaches, could potentially have reduced the noise level in the RInline graphic maps and thereby produced higher quality synMPRAGE images. 21

The SNR analysis did not show any large difference between the three proposed MPRAGE like equations. Overall, both the MPRAGE like,all and MPRAGE like,PD had similar SNR values, which were always higher than the values reported for MPRAGE like,MT . Additionally, all MPRAGE like images had higher SNR than the synMPRAGE. For the putamen, caudate, and cerebellar cortex, the SNR values of MPRAGE like images reached around a factor of two higher than those of synMPRAGE. This may largely be due to RInline graphic estimation outliers that increased the signal standard deviation in the selected ROIs. Considering the MPRAGE like,all equation includes three contrasts (instead of two for MPRAGE like,PD and MPRAGE like,MT ), an SNR gain could have been expected, but the lower signal in MTw images used for MPRAGE like,MT resulted in no observable SNR gain for MPRAGE like,all . Conversely, the increased contrast of MTw images improved the CNR values for MPRAGE like,all .

While synMPRAGE showed the highest WM‐GM CNR (driven by particularly high WM‐GM signal difference in the numerator that seemed to outweigh the large noise standard deviation in the denominator of Equation (10)), the low CNR values (and SNR) in the subcortical structures caused problems for FastSurferVINN to properly segment the synMPRAGE images. As a result, at least one region was missing completely in the segmentations for 21 of the 23 subjects, which explains the large difference in DSC and ASSD values compared to the different MPRAGE like images. This makes the segmentation results unfit for further use. For comparison, no region was missing for MPRAGE like,all and MPRAGE like,MT across all subjects, whereas only one region in one subject was missing for MPRAGE like,PD .

All MPRAGE like equations produced highly comparable image quality while all of them being superior to the synMPRAGE in every aspect except CNR between WM‐GM. Ultimately, the fact that all MPRAGE like equations produced highly comparable images gives flexibility to researchers to calculate a MPRAGE like image based on the contrasts available in their MPM/VFA protocols. For instance, acquiring only T1w and PDw contrasts for RInline graphic estimation using the VFA method allows one to compute MPRAGE like,PD images similar to the ones presented here. This should even work in short scan time using single‐TE VFA acquisitions as shown with Figure S1 in the Supporting Information.

One advantage of MPRAGE like is its simplicity. Indeed, neither fitting of quantitative parameters nor DL training are required like in comparable studies. 21 , 22 , 24 However, since an RInline graphic map is commonly computed in neuroimaging pipelines, synMPRAGE is also a straightforward technique to implement, which motivated the choice for comparison against MPRAGE like .

4.3. Automatic brain segmentation from FastSurferVINN versus other segmentation techniques

Automatic brain segmentation algorithms optimized for 7T images are limited. To the best of our knowledge, this study is the first one using FastSurferVINN for 7T images at 0.6 mm isotropic nominal resolution. Indeed, FastSurferVINN was designed for T1w MPRAGE images at ≤ 3T with resolutions between (0.7 and 1.0 mm)Inline graphic, meaning that the (0.6 mm)Inline graphic resolution used in this work was considered an “experimental resolution” by FastSurferVINN. However, previous analyses done with FastSurferVINN showed its robustness to higher resolutions unseen from the training dataset. 39 From Figure 6, we could observe that FastSurferVINN handled well the (0.6 mm)Inline graphic resolution of the images. The high resolution enabled proper segmentation of thin cortical regions, which would not have been possible at 1 mmInline graphic.

It is worth noting that universal pTx pulses were used, and therefore, the level of B1+ inhomogeneities observed in the images was considerably lower than what is typically observed for 7T images using circular‐polarized (CP) excitation only (See Figures S2 and S3 from Supporting Information). Indeed, flip angles between 75% and 125% of the nominal flip angle value for UP excitation were observed across the brain, which corresponds to a substantial improvement over CP excitation, where values as low as 25% were observed (Figure S3). Ultimately, that resulted in the N4‐correction being effective to correct for the remaining receive inhomogeneities in a similar fashion as it would at 3T.

In a pilot study, 46 several alternative segmentation algorithms were considered and evaluated. These included FreeSurfer, SynthSeg 47 and CEREBRUM‐7T. 48 For FreeSurfer, a specific pipeline for sub‐millimeter 7T images has been previously proposed in the literature. 49 However, in our pilot study, the FreeSurfer pipeline showed a high sensitivity to B1+ inhomogeneities in addition to a limited capacity in handling sub‐millimeter resolution. Moreover, published studies about neurological diseases (e.g., Alzheimer's disease) have shown that FastSurferVINN gives significantly better or equivalent results to FreeSurfer while also being, on average, 40 times faster. 50 SynthSeg was excluded from this work due to its limitation in output resolution of only 1 mmInline graphic, while CEREBRUM‐7T required MP2RAGE as the input image and produces a limited number of segmented regions. Hence, FastSurferVINN was selected as the segmentation method in this study, with the N4‐corrected MPRAGE images used as reference.

4.4. FastSurferVINN segmentation results across different MPRAGE‐images

In this work, we found that any MPRAGE like image provided a more robust and reliable input to automatic brain segmentation than synMPRAGE. The segmentation results from MPRAGE like images were highly comparable to those obtained from the reference MPRAGE. Limitations observed in the MPRAGE like ‐based segmentations, such as coarse delineations of fine WM‐GM boundaries in the cerebellum and the cortical region over the sinuses, were also observed in the MPRAGE‐based segmentations. This lets us believe that these limitations are not due to the image quality of MPRAGE like , but from the segmentation algorithm used. Indeed, thin WM branches like the ones observed in the cerebellum are often problematic to segment for DL‐based techniques since very few high‐resolution segmented datasets are available for the training dataset. Even datasets like the Human Connectome Project (HCP), 51 which includes (0.7 mm)Inline graphic segmentations from 3T MPRAGE images, do not have well‐segmented cerebral WM branches due to the limited capacity of FreeSurfer (software used for the HCP) to segment sub‐millimeter images and limited cerebral WM‐GM contrast at 3T.

The high variability in image quality for synMPRAGE between subjects was due to the estimation process of the RInline graphic maps, which negatively affected the overall segmentation quality. Quantitatively, that could be observed with the IQRs of the DSC values computed. Indeed, some subjects had high‐quality synMPRAGE images, which resulted in average DSC values around 0.70, but these represent outliers in the distribution of subjects. Reliable image quality is an important factor, and MPRAGE like showed a considerably superior robustness and reliability in comparison to synMPRAGE.

The fact that MPRAGE like segmentation showed only minor improvement after the N4‐correction may be due to the low λ value used, such that receive inhomogeneities are by design almost completely canceled due to the division of two images with identical receive bias in a very similar fashion as MP2RAGE. 8 For synMPRAGE, the observed decrease in DSC and increase in ASSD after N4‐correction might be due to the fact that synMPRAGE images do not exhibit receive inhomogeneities as expected by the N4 algorithm, potentially resulting in suboptimal effects.

Even if MPRAGE like,MT gave the lowest SNR and SSIM values among MPRAGE like images, the output segmentations were not affected. This lets us believe that the superior WM‐GM contrast of MPRAGE like,MT is important for its segmentability. On the other hand, MPRAGE like,PD , which had the highest SNR and SSIM, both uncorrected and N4‐corrected images produced the lowest median DSC and ASSD among all MPRAGE like images. MPRAGE like,all always held the highest DSC and ASSD values (by a very small margin) while never having the highest SNR, CNR, nor SSIM. In the end, the segmentation results of MPRAGE like were highly similar with negligible differences between them, suggesting that (1) the added MTw contrast helps for segmentation and (2) that any of them can be used as a great surrogate to MPRAGE for T1w‐based segmentation algorithms.

The DSC values found in this work for the MPRAGE like are very similar to other values reported in the literature. 22 In Reference 22, an average DSC of 0.88 ± 0.02 was reported across subjects. Here, the DSC (and ASSD) were computed for all segmented regions except the WM‐hypointensities (i.e., 34 regions in total) for each subject, whereas only seven selected regions were included in the DSC calculation in Reference 22. If only the same seven regions were included, average DSC of 0.93 ± 0.01, 0.92 ± 0.01, and 0.92 ± 0.02 would have been obtained here for N4‐corrected MPRAGE like,all , MPRAGE like,MT , and MPRAGE like,PD , respectively.

Most of the median relative differences in volume values calculated had statistically significant differences for the non‐corrected MPRAGE. This highlights the relevance of the N4‐correction when volumetric analysis is performed. Moreover, it is worth mentioning that even if N4‐MPRAGE like,PD had the lowest DSC and highest ASSD values among MPRAGE like images, it was the only one without a significant difference with the reference.

Ultimately, the new approach proposed in this work is a simple and flexible technique to synthesize T1w images directly from multi‐contrast spoiled GRE images. In addition to serving as an anatomical reference image, MPRAGE like images can serve as a reliable surrogate for automated brain segmentation if MPM or VFA data is acquired as part of a long neuroscience imaging protocol with tight time constraints.

4.5. Limitations

All data presented in this work were acquired at 7T, and both the MPM and MPRAGE protocols were based on non‐product sequences using Universal Pulses pTx. Extrapolation of our findings to non‐pTx MPM images at 7T or 3T is therefore limited. While not investigated in this study, previous works suggested that circularly polarized excitation for MPM at 7T resulted in poor MPRAGE like contrast in regions sensitive to B1+ inhomogeneities, essentially due to insufficient T1w contrast in the numerator of the MPRAGE like equation. 52 This cannot be compensated by N4‐correction (Supporting Information Figure S2). Moreover, the MPM protocol used was designed to be as fast as possible, which might have resulted in suboptimal contrasts. While MPM protocols are used both at 3T and 7T, the new MPRAGE like technique has neither been thoroughly tested at 3T nor with single‐TE data. However, single‐subject data shown in Supporting Information Figure S1 demonstrates that the method can, in principle, be applied at 3T and even using single‐TE data as well. The MPRAGE like approach should be further tested on other data acquired on lower field strengths and at other sites with different sequence implementations.

Moreover, it is also important to mention that all subjects in this study were healthy, young, and able to stay still for long periods of time. The MPM sequence used in this study is intended to be used in imaging protocols characterizing neurodegenerative diseases. Even though each of the three MPM contrast acquisitions had shorter scan times at (0.6 mm)Inline graphic voxel size than the reference MPRAGE acquisition at (0.6 mm)Inline graphic voxel size (TA = 6:00/4:05/4:05 min vs. TA=7:24 min) and inter‐volume motion is accounted for by retrospective registration, intra‐volume motion artifacts may become limiting in patients with symptoms, where excessive motion throughout all MPM scans might result in lower image quality for MPRAGE like images.

The segmentations produced from the N4‐corrected MPRAGE images were considered as the reference in this work to compare MPRAGE like and synMPRAGE segmentations. However, this is considered a “silver standard” since manual segmentations are typically considered the gold standard. Producing manual segmentation of several regions of the complete brain for 23 subjects at 0.6 mm isotropic resolution is a considerable challenge that requires a substantial amount of time and expertise, which was beyond the scope of this study.

5. CONCLUSIONS

In summary, we propose MPRAGE like ; a novel and simple technique producing images with synthetic T1w contrast directly from ME‐GRE images with different contrast weightings that can be used as a surrogate to MPRAGE for automatic brain segmentation. The results showed that MPRAGE like produces images highly resembling MPRAGE images except for the absent arterial blood highlighting typical at 7T. Additionally, employing MPRAGE like instead of synMPRAGE substantially improves image quality and segmentation results. This makes MPRAGE like a superior alternative to synMPRAGE and offers a robust new solution for synthetic T1w imaging techniques at UHF‐MRI, which has been previously lacking. MPRAGE like would benefit from further validation across different field strengths and within non‐healthy cohorts. Ultimately, implementing MPRAGE like would allow researchers to skip acquiring a MPRAGE resulting in a considerable amount of scan time saved while still being able to run the same quantitative neuroimaging pipelines.

CONFLICT OF INTEREST STATEMENT

The authors declare no potential conflict of interests.

Supporting information

Data S1. Supporting Information.

MRM-94-134-s001.pdf (1.4MB, pdf)

ACKNOWLEDGMENTS

This publication is an outcome of SCAIFIELD, an EU Joint Programme–Neurodegenerative Disease Research (JPND) project (https://www.jpnd.eu). The project is supported through the following funding organizations under the aegis of JPND: Belgium, The Fund for Scientific Research (F.R.S.‐FNRS; funding code PINT‐MULTI/BEJ R.8006.20); Germany, Federal Ministry of Education and Research (BMBF; funding codes 01ED2109A/B); Norway, The Research Council of Norway (RCN; funding code 322980); Turkey, Scientific and Technological Research Council of Turkey (TÜBÍTAK; funding code 121N028).

Fortin M‐A, Stirnberg R, Völzke Y, et al. MPRAGE like : A novel approach to generate T1w images from multi‐contrast gradient echo images for brain segmentation. Magn Reson Med. 2025;94:134‐149. doi: 10.1002/mrm.30453

DATA AVAILABILITY STATEMENT

A python implementation of the proposed method can be found at https://github.com/mafortin/mprage‐like.

REFERENCES

  • 1. Mugler III, John P, Brookeman JR. Three‐dimensional magnetization‐prepared rapid gradient‐echo imaging (3D MP RAGE). Magn Reson Med. 1990;15:152‐157. [DOI] [PubMed] [Google Scholar]
  • 2. Bruce F. FreeSurfer. Neuroimage. 2012;62:774‐781. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Donatas S, Didac V‐P, Øystein S, et al. Reliability and sensitivity of two whole‐brain segmentation approaches included in FreeSurfer – ASEG and SAMSEG. NeuroImage. 2021;237:118113. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Leonie H, Sailesh C, Santiago E, Kersten D, Bruce F, Martin R. FastSurfer ‐ A fast and accurate deep learning based neuroimaging pipeline. NeuroImage. 2020;219:117012. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Siegfried T, Elisabeth S, Wolfgang B, et al. Key clinical benefits of neuroimaging at 7 T. NeuroImage. 2018;168:477‐489. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Fritz S. Whole‐body MRI at high field: technical limits and clinical potential. Eur Radiol. 2005;15:946‐959. [DOI] [PubMed] [Google Scholar]
  • 7. Webb Andrew G, Collins CM. Parallel transmit and receive technology in high‐field magnetic resonance neuroimaging. Int J Imaging Syst Technol. 2010;20:2‐13. [Google Scholar]
  • 8. Marques José P, Tobias K, Gunnar K, Wietske Z, Pierre‐François M, Rolf G. MP2RAGE, a self bias‐field corrected sequence for improved segmentation and T1‐mapping at high field. NeuroImage. 2010;49:1271‐1281. [DOI] [PubMed] [Google Scholar]
  • 9. O'Brien Kieran R, Tobias K, Patric H, et al. Robust T1‐weighted structural brain imaging and morphometry at 7T using MP2RAGE. PLoS One. 2014;9:e99676. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Hampus O, Mikael N, Jimmy L, Ronnie W, Gunther H. Radiofrequency bias correction of magnetization prepared rapid gradient echo MRI at 7.0 Tesla using an external reference in a sequential protocol. Tomography. 2021;7:434‐451. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Vincent G, Alexandre V, Alexis A, Denis LB, Nicolas B. Universal pulses: A new concept for calibration‐free parallel transmission. Magn Reson Med. 2017;77:635‐643. [DOI] [PubMed] [Google Scholar]
  • 12. McEvoy LK, Christine F‐N, Cooper RJ, et al. Alzheimer Disease: Quantitative Structural Neuroimaging for Detection and Prediction of Clinical and Structural Changes in Mild Cognitive Impairment. Radiology. 2009;251:195‐205. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Menke RA, Jan S, Miller KL, et al. MRI characteristics of the substantia nigra in Parkinson's disease: A combined quantitative T1 and DTI study. NeuroImage. 2009;47:435‐441. [DOI] [PubMed] [Google Scholar]
  • 14. Gunther H, Henning D, Peter D. Quantitative FLASH MRI at 3T using a rational approximation of the Ernst equation. Mag Reson Med. 2008;59:667‐672. [DOI] [PubMed] [Google Scholar]
  • 15. Deoni Sean CL, Peters Terry M, Rutt BK. High‐resolution T1 and T2 mapping of the brain in a clinically acceptable time with DESPOT1 and DESPOT2. Mag Reson Med. 2005;53:237‐241. [DOI] [PubMed] [Google Scholar]
  • 16. Difei W, Philipp E, Tony S, Rüdiger S. Reproducibility of rapid multi‐parameter mapping at 3T and 7T with highly segmented and accelerated 3D‐EPI. Magn Reson Med. 2022;88:2217‐2232. [DOI] [PubMed] [Google Scholar]
  • 17. Nikolaus W, John S, Guy W, et al. Quantitative multi‐parameter mapping of R1, PD*, MT, and R2* at 3T: a multi‐center validation. Front Neurosci. 2013;7:46379. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Maryam S, Tobias L, Simon S, et al. Reliability of multi‐parameter mapping (MPM) in the cervical cord: A multi‐center multi‐vendor quantitative MRI study. NeuroImage. 2022;264:119751. [DOI] [PubMed] [Google Scholar]
  • 19. Tobias L, Maryam S, Gunther H, et al. Multiparameter mapping of relaxation (R1, R2*), proton density and magnetization transfer saturation at 3 T: A multicenter dual‐vendor reproducibility and repeatability study. Hum Brain Mapp. 2020;41:4232‐4247. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Akifumi H, Marcel W, Masaaki H, et al. SyMRI of the brain: rapid quantification of relaxation rates and proton density, with synthetic MRI, automatic brain segmentation, and myelin measurement. Investig Radiol. 2017;52:647. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Ulrike N, Elke H, Oliver B, Julia T, Ralf D. Improved visibility of brain tumors in synthetic MP‐RAGE anatomies with pure T1 weighting. NMR Biomed. 2015;28:818‐830. [DOI] [PubMed] [Google Scholar]
  • 22. Kanghyun R, Na‐Young S, Dong‐Hyun K, Yoonho N. Synthesizing T1 weighted MPRAGE image from multi echo GRE images via deep neural network. Magn Reson Imaging. 2019;64:13‐20. [DOI] [PubMed] [Google Scholar]
  • 23. Eugenio IJ, Benjamin B, Yaël B, et al. Joint super‐resolution and synthesis of 1 mm isotropic MP‐RAGE volumes from clinical MRI exams with scans of different orientation, resolution and contrast. NeuroImage. 2021;237:118206. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Simon S, Maryam S, Tobias L, et al. Reliability of spinal cord measures based on synthetic T1‐weighted MRI derived from multiparametric mapping (MPM). NeuroImage. 2023;271:120046. [DOI] [PubMed] [Google Scholar]
  • 25. Edwards Luke J, Peter MC, Saskia H, et al. Quantitative MRI maps of human neocortex explored using cell type‐specific gene expression analysis. Cereb Cortex. 2023;33:5704‐5716. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Rüdiger S, Tony S. Segmented K‐space blipped‐controlled aliasing in parallel imaging for high spatiotemporal resolution EPI. Magn Reson Med. 2021;85:1540‐1551. [DOI] [PubMed] [Google Scholar]
  • 27. Léo VD, Mauconduit F, Thomas C, Nicolas B, Vincent G. Universal nonselective excitation and refocusing pulses with improved robustness to off‐resonance for Magnetic Resonance Imaging at 7 Tesla with parallel transmission. Magn Reson Med. 2021;85:678‐693. [DOI] [PubMed] [Google Scholar]
  • 28. Breuer Felix A, Martin B, Mueller Matthias F, et al. Controlled aliasing in volumetric parallel imaging (2D CAIPIRINHA). Magnetic Resonance in Medicine: An Official Journal of the International Society for. Magn Reson Med. 2006;55:549‐556. [DOI] [PubMed] [Google Scholar]
  • 29. Yannik V, Suzan A, Daniel L, et al. Calibration‐free whole‐brain CEST imaging at 7T with parallel transmit pulse design for saturation homogeneity utilizing universal pulses (PUSHUP). Magn Reson Med. 2025;93:630‐642. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. David L, Raphael T‐T, Pip B, et al. Parallel transmit pulse design for saturation homogeneity (PUSH) for magnetization transfer imaging at 7T. Magn Reson Med. 2022;88:180‐194. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31. Yarnykh VL. Actual flip‐angle imaging in the pulsed steady state: a method for rapid three‐dimensional mapping of the transmitted radiofrequency field. Mag Reson Med. 2007;57:192‐200. [DOI] [PubMed] [Google Scholar]
  • 32. Karsten T, Evelyne B, John A, et al. hMRI – A toolbox for quantitative MRI in neuroscience and clinical research. NeuroImage. 2019;194:191‐210. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33. John A. SPM: a history. NeuroImage. 2012;62:791‐800. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34. Gunther H, Henning D, Kai K, Peter D. High‐resolution maps of magnetization transfer with inherent correction for RF inhomogeneity and T1 relaxation obtained from 3D FLASH MRI. Mag Reson Med. 2008;60:1396‐1407. [DOI] [PubMed] [Google Scholar]
  • 35. Nobuyuki O. A threshold selection method from gray‐level histograms. IEEE Trans Syst Man Cybern. 1979;9:62‐66. [Google Scholar]
  • 36. Zhou W, Bovik Alan C, Sheikh Hamid R, Simoncelli EP. Image quality assessment: from error visibility to structural similarity. IEEE Trans Image Process. 2004;13:600‐612. [DOI] [PubMed] [Google Scholar]
  • 37. Tustison Nicholas J, Avants Brian B, Cook Philip A, et al. N4ITK: improved N3 bias correction. IEEE Trans Med Imaging. 2010;29:1310‐1320. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38. Avants Brian B, Nick T, Gang S, et al. Advanced normalization tools (ANTS). Insight J. 2009;2:1‐35. [Google Scholar]
  • 39. Leonie H, David K, Martin R. FastSurferVINN: Building resolution‐independence into deep learning segmentation methods—A solution for HighRes brain MRI. NeuroImage. 2022;251:118933. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40. Dice LR. Measures of the amount of ecologic association between species. Ecology. 1945;26:297‐302. [Google Scholar]
  • 41. Thorvald S. A method of establishing groups of equal amplitude in plant sociology based on similarity of species content and its application to analyses of the vegetation on Danish commons. Biologiske Skrifter. 1948;5:1‐34. [Google Scholar]
  • 42. Tobias H, Bram VG, Styner Martin A, et al. Comparison and evaluation of methods for liver segmentation from CT datasets. IEEE Trans Med Imaging. 2009;28:1251‐1265. [DOI] [PubMed] [Google Scholar]
  • 43. Frank W. Individual comparisons by ranking methods. Springer; 1992:196‐202. [Google Scholar]
  • 44. Carlo B. Teoria statistica delle classi e calcolo delle probabilita. Pubblicazioni del R Istituto Superiore di Scienze Economiche e Commericiali di Firenze. 1936;8:3‐62. [Google Scholar]
  • 45. Wang Z, Bovik AC. Mean squared error: Love it or leave it? A new look at Signal Fidelity Measures. IEEE Signal Process Mag. 2009;26:98‐117. [Google Scholar]
  • 46. Marc‐Antoine F, Yannik V, Rüdiger S, et al. High quality brain segmentation from multi‐parameter mapping at Ultra‐High Field MRI. Proceedings of the International Society of Magnetic Resonance in Medicine, 2023, Toronto, Canada; 2023:1776. [Google Scholar]
  • 47. Benjamin B, Greve Douglas N, Oula P, et al. SynthSeg: Segmentation of brain MRI scans of any contrast and resolution without retraining. Med Image Anal. 2023;86:102789. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48. Michele S, Sergio B, Dennis B, Lars M. CEREBRUM‐7T: Fast and Fully Volumetric Brain Segmentation of 7 Tesla MR Volumes. Hum Brain Mapp. 2021;42:5563‐5580. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49. Natalia Z, Bruce F, Martin R, Ville R, Polimeni JR. Advantages of cortical surface reconstruction using submillimeter 7 T MEMPRAGE. NeuroImage. 2018;165:11‐26. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50. Louise B, Friedrich Christoph M. Comparison of Automated Volume Extraction with FreeSurfer and FastSurfer for Early Alzheimer's Disease Detection with Machine Learning. 2021 IEEE 34th International Symposium on Computer‐Based Medical Systems (CBMS). IEEE; 2021:113‐118. [Google Scholar]
  • 51. Van Essen DC, Kamil U, Edward A, et al. The Human Connectome Project: a data acquisition perspective. NeuroImage. 2012;62:2222‐2231. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52. Rüdiger S, Yannik V, Daniel L, et al. High‐resolution whole‐brain multi‐parameter mapping at 7 Tesla with interleaved fly‐back 3D‐EPI and universal pTX pulses. Proceedings of the International Society of Magnetic Resonance in Medicine, 2023, Toronto, Canada; 2023:272. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Data S1. Supporting Information.

MRM-94-134-s001.pdf (1.4MB, pdf)

Data Availability Statement

A python implementation of the proposed method can be found at https://github.com/mafortin/mprage‐like.


Articles from Magnetic Resonance in Medicine are provided here courtesy of Wiley