. 2025 Apr 23;14(1):2628. doi: 10.4102/ajlm.v14i1.2628

Mitigating the risk of tube shortages: A blood collection tube validation study conducted in South Africa

Marizna Korf 1,2,3,4,, Jody Rusch 3,4, Aye Aye Khine 5, Nalene Strauss 5, Lourens Jacobsz 5, Annalise E Zemlin 1,2, Helena Vreede 3,4
PMCID: PMC12067025  PMID: 40356694

Abstract

Background

The National Health Laboratory Service was using Becton Dickinson (BD) blood drawing tubes and, in 2021, the supplier notified customers of supply challenges, indicating a risk of global shortages.

Objective

This study aimed to validate candidate blood collection tubes from four brands (VACUCARE, VACUETTE®, VACUTEST®, and V-TUBE) compared to BD Vacutainer® tubes in three National Health Laboratory Service laboratories in Cape Town, South Africa.

Methods

Blood was collected from 300 healthy volunteers between October 2021 and November 2021. The technical validation assessed 11 quality indicators, with a sigma metric greater than 4 deemed acceptable. Usability feedback was gathered from phlebotomists. The clinical validation estimated differences in results across 52 clinical chemistry tests, using desirable bias specified by the European Federation of Clinical Chemistry and Laboratory Medicine Biological Variation Database, or Ricos, as acceptance criteria. Analysis was performed on Roche cobas® 6000 and DiaSorin Liaison® XL analysers.

Results

All VACUCARE tubes exhibited sigma metrics above 4, indicating excellent performance. VACUETTE® and V-TUBE were not uncapped by all Roche pre-analytical systems. VACUTEST® caps had rigid rubber, making it more challenging to puncture and detach the tube, which resulted in needle displacement. Both VACUCARE and V-TUBE were reported as user-friendly. All candidate tube analytes showed acceptable clinical performance.

Conclusion

VACUCARE, VACUETTE®, VACUTEST® and V-TUBE are viable alternatives to BD Vacutainer®. However, based on the results obtained from the technical validation, VACUCARE was identified as the most suitable interim replacement for BD Vacutainer® during the shortage.

What this study adds

This study addresses a gap in the literature on tube validation and provides valuable insights for clinical laboratories considering a replacement. It also presents an alternative approach to technical validation by utilising sigma metrics.

Keywords: blood specimen collections, pre-analytical phase, risk management, validation study, South Africa

Introduction

Blood collection tube technology has advanced, with the use of complex materials and additives designed to enhance convenience and improve turnaround time. These tubes are an important, and often under-recognised, pre-analytical consideration in clinical laboratories, since these devices are used to analyse patient samples on complex state-of-the-art equipment. The tube type, its additives, and analyte stability under various storage conditions, can potentially contribute to significant error. Results of laboratory tests are critical to patient care, and precision and accuracy are essential. Manufacturers and clinical laboratories should, therefore, validate the technical and clinical performance of new blood collection products.1,2 Proper tube validation is crucial to prevent the financial and human consequences related to inaccurate testing. These include unnecessary testing, delayed results, and compromised patient care.

Several factors, including supply chain disruptions driven by transportation delays, constrained supplies, and increased costs, as well as the increased demand during the Coronavirus disease 2019 pandemic, imposed unprecedented challenges on the entire healthcare industry. As a result, Becton, Dickinson and Company (BD, UK Limited, Berkshire, United Kingdom), a prominent manufacturer of blood collection tubes, notified customers of supply challenges with the risk of global product shortages in 2021.3

Following this notice, the National Health Laboratory Service of South Africa, which is mandated by government to provide clinical laboratory services to over 80% of the population, sent a notice to healthcare workers with guidance on the efficient use of blood collection tubes. Clinicians and phlebotomists were urged to refrain from collecting spare tubes, since a properly filled blood collection tube will be sufficient for all tests requested, in most cases. Clinicians were reminded to first review patients’ results before ordering more tests, to avoid unnecessary testing and re-testing, and add-on requests were encouraged. Additionally, legible completion of request forms and good phlebotomy technique were advised to prevent unnecessary rejections. Although the current blood tube shortage has been resolved, these measures should continue to be practised, promoting environmental stewardship and sustainable laboratory practices.

To mitigate tube shortage risk, the research team conducted a comprehensive blood collection tube validation study. This was considered a validation study in accordance with the International Organization for Standardization 9000:2005 specifications, which define validation as the ‘confirmation, through the provision of objective evidence … that the requirements … for a specific intended use or application have been fulfilled’.4 Additionally, the terminology aligns with that used by the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) Working Group for Pre-Analytical Phase in their opinion paper on validating blood collection tubes in clinical laboratories.5 While the validation protocol followed EFLM Working Group for Pre-Analytical Phase guidance, the study’s unique circumstances involved a global tube shortage. Unlike the usual transitions to new tubes following the awarding of a new tender, the urgent aim was to identify and validate alternative options to mitigate potential supply risks for the National Health Laboratory Service. Blood collection tubes from four brands (VACUCARE, VACUETTE®, VACUTEST®, and V-TUBE [Table 1]) were validated to identify the closest match to the BD tubes’ performance, with the intention of returning to BD tubes once supply issues were resolved.

TABLE 1.

Manufacturer details and lot numbers of blood collection tubes used during the study at all three testing sites in Cape Town, South Africa, October 2021 to November 2021.

Type types SS K2EDTA LiH NaFl/KOx
Tube volume 5 mL 4 mL 4 mL 4 mL
VACUCARE 210314 210505 200618 210611
 Zhejiang GongDong Medical Technology Co., Ltd, Taizhou, China 210608
VACUETTE® A21053HC A21013NY A21083T4 A20093X5
 Greiner Bio-One GmbH, Kremsmünster, Austria A201049T
VACUTEST® KZ1801 WZ1321 KZ2561 KZ1521
 VACUTEST Kima SRL, Padua, Italy KG1371 WZ2501
V-TUBE™ 6319036 2519002 27110023 Tube not available
 AB Medical Inc, Jangseong, South Korea
BD Vacutainer® 1064851§ 1053105 1099594 1040024
 Becton Dickinson and Company, UK Ltd, Berkshire, United Kingdom 1117687§ 1116117
1151158§ 1124196

SS, serum tube with gel separator; K2EDTA, dipotassium ethylenediaminetetraacetic acid; LiH, lithium heparin; NaFl/KOx, sodium fluoride/potassium oxalate.

, technical validation only,

, branded VACUCARE for EREZlabmed, Gauteng, South Africa, trademark pending,

§

, SST II Advance Plus.

Methods

Ethical considerations

The study received full approval from the Health Research Ethics Committee of Stellenbosch University (reference no.: N21/09/095), Cape Town, South Africa, on 12 November 2021. The study was conducted in compliance with the 2013 World Medical Association Declaration of Helsinki. Written informed consent was obtained from all volunteers. Each participant was assigned a unique identifier number during blood collection, which was used for labelling the tubes, registering the samples on the Laboratory Information System, and during data analysis. Only the principal investigator held the key to reidentify participants, allowing the research team to share their blood results with them.

Participants

A total of 300 healthy male and female adult volunteers participated in this blood collection tube validation study. The study was conducted across three laboratories in Cape Town, South Africa, from October 2021 to November 2021 (Figure 1).6 Investigators sent recruitment invitations to division heads, who then forwarded them to National Health Laboratory Service staff and healthcare workers. Additional participants, including friends and relatives of staff members, joined the study through word-of-mouth awareness. The number of 300 volunteers was selected to ensure an adequate number of sample pairs for the clinical validation and to limit the number of tubes collected per participant to a maximum of five tubes each for candidate and comparative tubes (Figure 1). Exclusion criteria: anaemia or receiving treatment for anaemia or iron deficiency, autoimmune diseases, hereditary haematological disorders, bleeding disorders or anticoagulation treatment, heart conditions with low cardiac output, confirmed or possible pregnancy, and being unwell at the time of the study. The exclusion criteria were included in the recruitment invitations, allowing volunteers to self-screen and only enrol if they met the eligibility requirements.

FIGURE 1.

FIGURE 1

This flow diagram illustrates the participation of volunteers across three study sites in Cape Town, South Africa, from October 2021 to November 2021. The blocks highlighted in grey indicate tubes collected by the same phlebotomy team. Only tubes marked with or were included in the reported study. The other tubes were used for haematology6 and therapeutic drug monitoring validations, which are not reported in the current study.

Study design

The design of the study was based on the 2016 opinion paper by the EFLM Working Group for Pre-Analytical Phase and the Clinical and Laboratory Standards Institute GP34A guideline.1,5 Technical and clinical validations were conducted. ‘Comparative tubes’ refer to the tubes currently used in the laboratory (BD), while ‘candidate tubes’ are those under evaluation as alternatives. Details of comparative and candidate tube manufacturers are summarised in Table 1. The V-TUBE glucose tube was excluded as it contained ethylenediaminetetraacetic acid, which was incompatible with certain glucose methods used in the National Health Laboratory Service. Lithium heparin tubes were only included in the technical validation, as Troponin T was the only test performed on lithium heparin plasma, and its results were expected to fall below the measuring range within the healthy study population.

Sample collection

Venous blood samples were collected by experienced phlebotomists using a blood collection device (evacuated tube holder) and 21-gauge needle (or 22-gauge butterfly needle) following Clinical and Laboratory Standards Institute GP41 guidelines.7 The same team performed phlebotomy for most of the study, in close proximity to the laboratories (Figure 1). Beckton Dickinson devices were used for both comparative and candidate tubes. Phlebotomy procedures were standardised bilaterally. Comparative tubes were filled from one arm, and candidate tubes from the other, with consistent sides but alternating order. Blood was drawn first from the arm that the phlebotomist assessed as having more challenging venous access, based on their evaluation of vein characteristics such as size, depth, or visibility. This approach helped determine the appropriate device and the number of tubes that could be collected. Tube combinations varied across sites, adhering to the specified draw order. Samples were processed per manufacturers’ instructions and tested immediately or stored under conditions that ensured analyte stability. All analyses were completed within 3 days of collection.

Technical validation

Pathology registrars and experienced laboratory technologists assessed 11 quality indicators (QIs) during various study stages (Figure 2). During the pre-check, pack and blood collection tube details, including lot number, expiry date, draw volume, and fill line presence were reviewed. The ‘filling distance’, measured as the distance between the top of the fill line and the bottom of the tube, was used to identify underfilling, defined as less than 90% of the filling distance. Each tube pack was assigned a reference number recorded during phlebotomy.

FIGURE 2.

FIGURE 2

Process flow of technical validation of blood collection tubes with quality indicators checked at each stage – Cape Town, South Africa, October 2021 to November 2021.

While the Clinical and Laboratory Standards Institute GP34A guideline recommends visual haemolysis assessment, this method is unreliable.8,9,10 Haemolysis was instead measured automatically using Roche analysers (Roche Diagnostics GmbH, Mannheim, Germany), which determine the haemolysis index through absorbance measurements of diluted samples at 570 nm and 600 nm.11 To improve accuracy, a quantitative haemolysis index of 30 mg/dL or lower was considered acceptable. This threshold was chosen because several routine chemistry analytes, including potassium, lactate dehydrogenase, and conjugated bilirubin, begin to be affected around this level.12,13,14

Tubes showing underfilling or in vitro haemolysis during the technical validation were excluded if phlebotomy difficulties were indicated on the phlebotomy checklist. After QI evaluation, the compatibility of all candidate tubes with the pre-analytical systems at the testing sites, namely the Modular Pre-analytics system (Roche Diagnostics GmbH, Mannheim, Germany) and Probenverteiltechnik cobas® p 612 pre-analytical system (Roche Diagnostics GmbH, Mannheim, Germany), was assessed. Phlebotomists provided feedback on their user experience.

The global tube shortage limited the study from including the recommended 120 candidate and comparative tubes for the technical validation.5

Clinical validation

The clinical validation aimed to determine whether the tube introduced bias, increasing total error and potentially having a clinically significant effect on patient results.

The Clinical and Laboratory Standards Institute EP09 guideline recommends 100 sample pairs for validation and 40 for verification.15 However, because of the tube shortage and other resource constraints, results from routine tests conducted at both Groote Schuur Hospital and Tygerberg Hospital were combined, yielding a total of 50 sample pairs. Additionally, 25 sample pairs from selected routine tests and specialised tests conducted only at a single study site were included.

The choice of analytes was limited by the healthy study population recruited. Analytes expected to have values below the measuring range in healthy participants, such as direct bilirubin, C-reactive protein, free prostate specific antigen, growth hormone, progesterone, and oestradiol, were excluded. In addition, lactate dehydrogenase and aspartate aminotransferase are affected by minor degrees of haemolysis and were also excluded.

Thirty analytes were measured on cobas® c501 (Roche Diagnostics GmbH, Mannheim, Germany) analysers: sodium, potassium, chloride, urea, creatinine, glucose, haemoglobin A1c, calcium, magnesium, phosphate, uric acid, total protein, albumin, total bilirubin, alanine transaminase, alkaline phosphatase, gamma glutamyltransferase, creatine kinase, lipase, total cholesterol, triglycerides, high-density lipoprotein cholesterol, haptoglobin, complement 3 (C3), complement 4 (C4), immunoglobulin G, immunoglobulin A, immunoglobulin M, iron, and transferrin. Serum kappa and lambda free light chains were measured with Freelite® assays (The Binding Site Group Ltd©, Birmingham, United Kingdom) on cobas® c501. Sixteen analytes were measured on cobas® e601 (Roche Diagnostics GmbH, Mannheim, Germany) analysers: ferritin, vitamin B12, serum folate, prostate specific antigen, thyroid stimulating hormone, free thyroxine, free triiodothyronine, cortisol, follicle stimulating hormone, luteinising hormone, testosterone, sex hormone binding globulin, insulin, prolactin, parathyroid hormone, and 25-hydroxy vitamin D. Renin, aldosterone, insulin-like growth factor 1, and C-peptide were measured on the DiaSorin Liaison® XL (DiaSorin, Saluggia, Italy) analyser. Most analytes were tested on serum, except for parathyroid hormone and renin which were tested on K2EDTA plasma, glucose tested on NaFl/KOx plasma, and haemoglobin A1c tested on K2EDTA whole blood.

The analytical coefficient of variation for each analyte was determined from 1 year’s quality control data and have been provided to demonstrate performance, and the potential contribution of coefficient of variation to total error in this study.

Statistical analysis

Technical validation data were captured and analysed in Microsoft© Excel® (Microsoft Corporation, Redmond, Washington, United States). The sigma metric was used as a performance metric to assess the technical quality of blood collection tubes, by calculating the defect rate across all 11 QIs. Each QI per tube was treated as a defect opportunity. Total observed defects were converted into defects per million opportunities using an online calculator.16 To adhere to typical performance standards observed in industry processes, a minimum sigma metric of 4 (6210 defects per million opportunities) was deemed to be the acceptable minimum threshold for the candidate tube.16 Additionally, the sigma metric of the candidate tube was required to be equal to or greater than the sigma metric of the comparative tube.

Clinical validation data were analysed using EP EvaluatorTM (Data Innovations LLC, South Burlington, Vermont, United States) and Excel®. Passing-Bablok regression was used to assess correlation, with acceptability based on the 95% confidence interval (CI) for the slope including 1 (no proportional difference) and the intercept including 0 (no systematic difference).17 Given the anticipated strong correlation between comparisons of the same analyte, the correlation coefficient (R) was likely to be highly influenced by the range of measurement results; R > 0.95 was deemed acceptable.18 Bland-Altman plots were used to assess differences between the tubes, with normality evaluated via histograms in EP Evaluator. For analytes with non-normal distribution, pairwise (Walsh) averages of the differences were calculated in a trial version of Minitab® (Minitab, LCC, State College, Pennsylvania, United States) with subsequent determination of the Hodges-Lehmann point estimator (median) and Tukey two-sided CIs.15

Average percentage bias was deemed acceptable if it was at or below the desirable bias calculated from the EFLM Biological Variation Database19 or Ricos criteria.20 For analytes tested at two laboratories, bias exceeding the desirable bias was acceptable if the 95% CI included the bias specification.15 Analytes that failed both Passing-Bablok regression and Bland-Altman analysis underwent total error calculation (% total error = % bias + 1.65*CVa), compared against EFLM Biological Variation Database or Ricos total allowable error. If the % total error exceeded the total allowable error, the clinical significance of the differences rendered the tube unsuitable for the intended purpose. Outlier detection used the ‘Extreme Studentized Deviate’ method in EP Evaluator.15

Results

Technical validation

During the pre-check, no major physical defects were observed in any candidate or comparative tubes. All candidate tubes were ‘Conformité Européenne’ marked, with expiry dates, lot numbers, and clear fill lines. No tubes were excluded during the pre-check.

The primary reasons for exclusion in assessing the remaining technical QIs were primarily due to missing data and underfilling caused by difficult phlebotomy. No haemolysed samples were attributed to phlebotomy difficulties (Table 2 and Table 3).

TABLE 2.

The number of defects observed and defect opportunities for 11 quality indicators assessed during technical validation – Cape Town, South Africa, October 2021.

Variable SS
K2EDTA
LiH
NaFl/KOx
BD
CT
BD
CT
BD
CT
BD
CT
Observed defects Defect opportunities Observed defects Defect opportunities Observed defects Defect opportunities Observed defects Defect opportunities Observed defects Defect opportunities Observed defects Defect opportunities Observed defects Defect opportunities Observed defects Defect opportunities
VACUCARE versus BD
 Manufacturing defects 0 75 0 75 0 75 0 75 0 25 0 25 0 25 0 25
 Did not properly fit into the blood collection device 0 50 0 50 0 75 0 75 0 25 0 25 0 25 0 25
 Failed to form a vacuum 0 50 0 50 0 75 0 75 0 25 0 25 0 25 0 25
 Cap leakage (specifically where the rubber was pierced) 0 50 0 50 0 75 0 75 0 25 0 25 0 25 0 25
 External surface contamination after venepuncture 0 50 0 50 0 75 0 75 0 25 0 25 0 25 0 25
 Undue clotting in K2EDTA tubes n/a n/a n/a n/a 0 50 0 50 n/a n/a n/a n/a n/a n/a n/a n/a
 Incomplete clotting in serum tubes 0 49 0 49 n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a
 Broke or spilled blood during centrifugation 0 49 0 49 0 25 0 25 0 24 0 24 0 25 0 25
 Underfilling 9 48, 0 49 0 75 0 75 1 24 0 24 0 24 0 25
 Inappropriate gel position after centrifugation 0 49 0 49 n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a
 Haemolysis 0 50 1 50 n/a n/a n/a n/a 0 25 0 25 n/a n/a n/a n/a
 Total 9 520 1 521 525 525 0 525 1 198 0 198 0 174 0 175
 efects per million opportunities 17 308 - 1 919 - 0 - 0 - 5051 - 0 - 0 - 0 -
 Sigma metric 3.7 - 4.4 - >6 - >6 - 4.1 - >6 - >6 - >6 -
VACUETTE® versus BD
 Manufacturing defects 0 75 0 75 0 75 0 75 0 25 0 25 0 25 0 25
 Did not properly fit into the blood collection device 0 50 0 50 0 75 0 75 0 25 0 25 0 25 0 25
 Failed to form a vacuum 0 50 0 50 0 75 0 75 0 25 0 25 0 25 0 25
 Cap leakage (specifically where the rubber was pierced) 0 50 7 50 0 75 18 75 0 25 2 25 0 25 6 25
 External surface contamination after venepuncture 0 50 0 50 0 75 0 75 0 25 0 25 0 25 0 25
 Undue clotting in K2EDTA tubes n/a n/a n/a n/a 0 50 0 50 n/a n/a n/a n/a n/a n/a n/a n/a
 Incomplete clotting in serum tubes 0 50 0 49 n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a
 Broke or spilled blood during centrifugation 0 50 0 49 0 25 0 25 0 25 0 25 0 25 0 24
 Underfilling 1 50 1 49 0 75 7 75 0 25 0 25 0 25 0 24
 Inappropriate gel position after centrifugation 0 50 0 49 n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a
 Haemolysis 0 50 0 50 n/a n/a n/a n/a 1 25 0 25 n/a n/a n/a n/a
 Total 1 525 8 521 0 525 25 525 1 200 2 200 0 175 6 173
 Defects per million opportunities 1905 - 15 355 - 0 - 47 619 - 5000 - 10 000 - 0 - 34 682 -
 Sigma metric 4.4 - 3.7 - >6 - 3.2 - 4.1 - 3.9 - >6 - 3.4 -

BD, BD Vacutainer®; SS, serum tube with gel separator; K2EDTA, dipotassium ethylenediaminetetraacetic acid; LiH, lithium heparin; NaFl/KOx, sodium fluoride/potassium oxalate; CT, candidate tube.

, excluded due to missing centrifugation data (n = 1);

, excluded due to difficult phlebotomy (n = 1).

TABLE 3.

The number of defects observed and defect opportunities for 11 quality indicators assessed during technical validation – Cape Town, South Africa, November 2021.

Variable SS
K2EDTA
LiH
NaFl/KOx
BD
CT
BD
CT
BD
CT
BD
CT
Observed defects Defect opportunities Observed defects Defect opportunities Observed defects Defect opportunities Observed defects Defect opportunities Observed defects Defect opportunities Observed defects Defect opportunities Observed defects Defect opportunities Observed defects Defect opportunities
VACUTEST ® versus BD
 Manufacturing defects 0 100 0 100 0 100 0 100 0 25 0 25 0 25 0 25
 Did not properly fit into the blood collection device 0 73 0 73 0 97,§ 0 97,§ 0 23 0 23 0 25 0 25
 Failed to form a vacuum 0 73 0 73 0 97,§ 0 97,§ 0 23 0 23 0 25 0 25
 Cap leakage (specifically where the rubber was pierced) 0 73 0 73 1 97,§ 1 97,§ 0 23 0 23 0 25 0 25
 External surface contamination after venepuncture 0 73 0 73 0 97,§ 0 97,§ 0 23 0 23 0 25 0 25
 Undue clotting in K2EDTA tubes n/a  n/a n/a n/a 0 50 0 50 n/a n/a n/a n/a n/a n/a n/a n/a
 Incomplete clotting in serum tubes 0 50 0 50 n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a
 Broke or spilled blood during centrifugation 0 50 0 50 0 25 0 25 0 25 0 25 0 25 0 25
 Underfilling 21 47, 8 47, 5 71,,§ 4 71,,§ 1 22, 2 22 0 25 0 25
 Inappropriate gel position after centrifugation 0 50 0 50 n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a
 Haemolysis 0 50 3 50 n/a n/a n/a n/a 0 25 1 25 n/a n/a n/a  n/a
 Total 21 639 11 639 6 634 5 634 1 189 3 189 0 175 0 175
 Defects per million opportunities 32 864 - 17 214 - 9464 - 7886 - 5291 - 15 873 - 0 - 0 -
 Sigma metric 3.4 - 3.7 - 3.9 - 4 - 4.1 - 3.7 - >6 - >6 -
V-TUBE™ versus BD
 Manufacturing defects 0 100 0 100 0 100 0 100 0 25 0 25 - - - -
 Did not properly fit into the blood collection device 0 75 0 75 0 100 0 100 0 25 0 25 - - - -
 Failed to form a vacuum 0 75 0 75 0 100 0 100 0 25 1 25 - - - -
 Cap leakage (specifically where the rubber was pierced) 0 75 0 75 0 100 0 100 0 25 0 25 - - - -
 External surface contamination after venepuncture 0 75 0 75 0 100 0 100 0 25 0 25 - - - -
 Undue clotting in K2EDTA tubes n/a n/a n/a n/a 0 50 0 50 n/a n/a n/a n/a - - - -
 Incomplete clotting in serum tubes 0 50 0 50 n/a n/a n/a n/a n/a n/a n/a n/a - - - -
 Broke or spilled blood during centrifugation 0 50 0 50 0 25 0 25 0 25 0 25 - - - -
 Underfilling 20 49 0 50 7 74 0 74 3 24 1 25 - - - -
 Inappropriate gel position after centrifugation 0 50 0 50 n/a n/a n/a n/a n/a n/a n/a n/a - - - -
 Haemolysis 0 50 0 50 n/a n/a n/a n/a 0 25 0 25 - - - -
 Total 20 649 0 650 7 649 0 649 3 199 2 200 - - - -
 Defects per million opportunities 30 817 - 0 - 10 786 - 0 - 15 075 - 10 000 - - - - -
 Sigma metric 3.4 - >6 - 3.8 - >6 - 3.7 - 3.9 - - - - -

Note: V-TUBE™ versus BD (NaFl/KOx) was not evaluated.

BD, BD Vacutainer®; SS, serum tube with gel separator; K2EDTA, dipotassium ethylenediaminetetraacetic acid; LiH, lithium heparin; NaFl/KOx, sodium fluoride/potassium oxalate; CT, candidate tube.

, excluded due to missing phlebotomy data (n = 2);

, excluded due to difficult phlebotomy (n = 1);

§

, not drawn due to difficult phlebotomy (n = 1);

¶, not evaluated.

VACUCARE tubes achieved sigma metrics of 4 or higher across all comparisons. In contrast, BD Vacutainer® SS had sigma metrics below 4 in three of four comparisons (BD vs. VACUCARE, VACUTEST®, and V-TUBE), mainly due to underfilled tubes in the latter two comparisons. VACUETTE® tubes also fell below a sigma metric of 4, primarily due to cap leakage after needle withdrawal. For VACUTEST SS and lithium heparin tubes, haemolysis and underfilling contributed to their sigma metrics falling below 4.

The Modular Pre-analytics system could not uncap VACUETTE® tubes without hardware modifications, which were unavailable because of the manufacturer’s discontinuation of the Modular Pre-analytics system. Similarly, the Probenverteiltechnik cobas® p 612 pre-analytical system could not uncap V-TUBE tubes, and no application settings were available for achieving compatibility.

Phlebotomists reported usability concerns, noting that the rigid rubber in VACUTEST® caps made tubes difficult to puncture and detach from collection devices, often causing needle displacement during the procedure. VACUETTE® tube caps were shorter in length compared to BD Vacutainer®, resulting in increased angular movement within the collection device. In contrast, VACUCARE and V-TUBE tubes were user-friendly.

Based on sigma metrics, pre-analytical systems compatibility, and usability, VACUETTE®, VACUTEST® and V-TUBE were disqualified.

Clinical validation

The clinical validation results are detailed in the Online Supplementary Table 1, Online Supplementary Table 2, Online Supplementary Table 3, and Online Supplementary Table 4. The study aimed to include 50 or 25 sample pairs per comparison; however, the total sample size was occasionally reduced because of missing results or samples. Additionally, outliers were identified and excluded using EP Evaluator. For V-TUBE, 25 of 50 C3 pairs tested at Groote Schuur Hospital were excluded due to a significant positive shift in internal quality control during the run.

Most analytes had correlation coefficients above 0.95, except sodium, potassium, and magnesium, which consistently fell below 0.95 across all comparisons. Calcium, total protein, and albumin displayed coefficients below 0.95 in most comparisons, except with VACUTEST®.

Passing-Bablok regression and Bland-Altman analysis indicated acceptable performance for all analytes except for aldosterone on V-TUBE. Passing-Bablok regression for aldosterone on V-TUBE showed a slope of 0.927 (95% CI: 0.874, 0.961) and intercept of −13.18 (95% CI: −25.170, –6.060). The mean bias was −15.63 (95% CI: −19.70, −11.55), with a total error of 30%, below the EFLM Biological Variation Database limit of 42.8%, deeming the difference clinically insignificant.

Discussion

In response to an impending BD Vacutainer® tube shortage, technical and clinical validations were conducted to evaluate alternative tubes. A modified approach reduced the number of tubes required while maintaining rigorous standards. A key feature was the application of sigma metrics to assess technical QIs, offering an objective measure of their quality. The technical validation proved critical, as all tubes met standards in the clinical validation phase.

Sigma metrics are often used to monitor QIs in the extra-analytical phases, with a Six Sigma level corresponding to a defect rate of 3.4 defects per million opportunities. While sigma metrics often evaluate the process quality, they can also be used to assess product performance. By treating each QI per tube as a defect opportunity, a sigma metric for each blood collection tube was determined. This approach also addressed the global shortage of blood collection tubes, which prevented inclusion of the full 120-tube sample size recommended in the EFLM opinion paper.5

Few publications address the technical validation of blood collection tubes. To the authors’ knowledge, this is the first study to compare VACUTEST and BD Vacutainer® tubes. During technical validation, VACUTEST SS failed to meet the minimum sigma metric of 4 due to underfilling and haemolysis but outperformed BD Vacutainer® SS tubes overall. Notably, underfilled BD Vacutainer® SS tubes were traced to a single lot (1151158), identified through reference numbers assigned during the pre-check process. The EFLM opinion paper recommends calculating percentage differences between candidate and comparative tubes for each QI, but poorly performing comparative tubes may distort results.5 This was similarly noted in a 2019 Turkish study, where 4 candidate tubes and 16 comparative tubes (BD Vacutainer SS) were found to be underfilled.21 Using sigma metrics with a pre-set acceptance threshold allows independent evaluation of candidate tubes, reducing reliance on comparative tube performance.

To the best of the authors’ knowledge, this study provides the first evaluation of VACUCARE tubes compared to BD Vacutainer®. However, a 2009 Korean study assessed tubes from Zhejiang GongDong Medical Technology Co., Ltd., likely using similar manufacturing technology across all their blood collection tubes.22 However, that study did not incorporate a technical evaluation, and it relied on Clinical Laboratory Improvement Amendments of 1988 limits instead of biological variation data, finding no significant differences between the tubes.

VACUETTE® and BD Vacutainer® serum tubes were compared in a 2011 Brazilian study by Lima-Oliveira et al. using a Roche cobas® c501.23 Like the current study, clinical validation acceptance criteria were based on desirable bias, derived from biological variation. While the previous study found a clinically significant mean difference of 4.9% for magnesium, the current study demonstrated a mean difference of -0.69% (95% CI: -1.64, 0.25), well below the desirable bias of 1.62%.

A 2014 study by Bowen et. al. in the United States assessed VACUETTE® and BD Vacutainer® serum tubes against glass tubes lacking clot activators, separator gels, and surfactants that could potentially impact results.24 Both VACUETTE® and BD Vacutainer® tubes showed no statistically or clinically significant differences in cortisol results using the Siemens Immulite 1000 analyser. The current study found a mean difference of 1.90% (95% CI: -2.00, 5.90) for cortisol, in a direct comparison between VACUETTE® and BD Vacutainer® serum tubes, further supporting comparable performance.

Two Korean studies (2013, 2018) performed comparisons between BD Vacutainer® and V-TUBE tubes for routine analytes in healthy volunteers and patients. One study showed statistically significant differences for sodium, chloride, alkaline phosphatase, high-density lipoprotein cholesterol, whereas the other study showed differences for creatinine, high-density lipoprotein cholesterol, iron, lipase, magnesium, potassium, total cholesterol, and uric acid.25,26 Yet, these findings were not deemed clinically significant in the context of Clinical Laboratory Improvement Amendments of 1988 limits.

In the current study, most analytes showed strong correlation coefficients (R > 0.95), but sodium, potassium, and magnesium consistently fell below this threshold. Similarly, calcium, total protein, and albumin displayed correlation coefficients below 0.95 in all comparisons, except for VACUTEST®. Poor correlation does not necessarily indicate poor agreement between tubes, but may reflect the limited range of results in the healthy study population.18,27

While combining blood collection tubes and devices from different manufacturers is discouraged,5 the unique study circumstances required using BD devices for all candidate tubes. Sigma metrics for VACUETTE® tubes consistently fell below 4 because of cap leakage from the puncture sites after needle withdrawal, posing safety concerns. However, this issue did not affect the clinical validation. Given these findings, it is advised against combining BD devices with VACUETTE® tubes.

Strengths

The strength of this validation study lies in its comprehensive assessment of analytes using VACUCARE, VACUETTE®, VACUTEST®, and V-TUBE compared to BD Vacutainer® tubes. It is among the few studies reporting on the technical validation of blood collection tubes and introduces a novel assessment approach. Furthermore, it demonstrates no clinically significant differences in specialised tests between comparative and candidate tubes, using desirable bias specified by the EFLM Biological Variation Database or Ricos as acceptance criteria.

Limitations

A key limitation was the inclusion of only healthy adult volunteers, which excluded several analytes due to low levels in this population and resulted in a narrow testing range for several analytes.

Conclusion

The technical validation component, combined with pre-analytical system compatibility considerations, was essential and led to the exclusion of certain candidate tubes. Comprehensive technical validation should precede the clinical validation.

VACUCARE, VACUETTE®, VACUTEST® and V-TUBE are viable alternatives to BD Vacutainer®. However, based on technical validation results, VACUCARE was identified as the most suitable interim replacement for BD Vacutainer® during the shortage.

Acknowledgements

M.K. and J.R. are joint first authors. Haematopathology co-investigators: Dr Jenique Bailly, Prof. Zivanai Chapanduka, Dr Michael Linström, Dr Erica-Mari Nell, Dr Diana Oelofse, Prof. Jessica Opie. Phlebotomists: Aqeela Abrahams, Jehaan Brown, Nokuzola Gemi, Rayaan Isaacs, Marcellino Johannes, Asiphe Ngqandu, Boraro Sehoshe. Assistants: Gadija Abdullah, Shanaaz Arendse, Morné Bezuidenhout, Wendy Blose, Sylvester Chabunya, Benito Dekela, Cameron Francis, Thando Gcingca, Kulsoem Kasker, Thobile Khanyile, Daleen Kriel, Gail Lawrence, Mariam Mahomed, Lwando Mampunye, Abugyer Mosavel, Thembi Mthembu, Igsaan Noordien, David Richardson, Eva Rienhardt, Kagiso Seakamela, Bianca Southon, Ashuelita Thompson, Dieter van der Westhuizen, Sigrid Vollmer, Laetitia Walters, Owen Wiese.

Competing interests

The authors declare that they have no financial or personal relationships that may have inappropriately influenced them in writing this article.

Authors’ contributions

M.K., J.R., A.A.K., N.S., L.J., A.E.Z., and H.V. made substantial contributions to the concept and design of the study. M.K., J.R., A.A.K., N.S., L.J., A.E.Z., and H.V. were involved in protocol development. M.K. was involved in obtaining ethical approval. M.K., J.R. and H.V. were involved in data analysis. M.K. and J.R. wrote the first draft of the article. M.K., J.R., A.A.K., N.S., L.J., A.E.Z., and H.V. reviewed and edited the article and approved the final version of the article.

Data availability

Data supporting the findings of this study are available from the corresponding author, M.K., upon request.

Disclaimer

The views and opinions expressed in this article are those of the authors and are the product of professional research. The article does not necessarily reflect the official policy or position of any affiliated institution, funder, agency, or that of the publisher. The authors are responsible for this article’s results, findings, and content.

Funding Statement

Sources of support The National Health Laboratory Service funded the analysis and supply of BD Vacutainer blood collection tubes for the validation. ILEX South Africa (Pty) Ltd (Gauteng, South Africa) supplied V-TUBE™, LASEC (Cape Town, South Africa) supplied VACUETTE®, Reliable Diagnostic Supplies (Pty) Ltd (Gauteng, South Africa) supplied VACUTEST® and Synbiolab (Gauteng, South Africa) supplied VACUCARE™ blood collection tubes for the validation.

Footnotes

How to cite this article: Korf M, Rusch J, Khine AA, et al. Mitigating the risk of tube shortages: A blood collection tube validation study conducted in South Africa. Afr J Lab Med. 2025;14(1), a2628. https://doi.org/10.4102/ajlm.v14i1.2628

Note: Additional supporting information is available for download alongside the online version of this article as Online Supplementary Document 1.

References

  • 1.Clinical and Laboratory Standards Institute . Validation and verification of tubes for venous and capillary blood specimen collection; Approved guideline. Wayne, PA: CLSI; 2010. [Google Scholar]
  • 2.Clinical and Laboratory Standards Institute . Tubes and additives for venous and capillary blood specimen collection; approved standard. 6th ed. Wayne, PA: CLSI; 2010. [Google Scholar]
  • 3.Becton, Dickinson and Company (BD) . BD Vacutainer® product updates [homepage on the Internet]. 2022. [cited 2022 Jul 15]. Available from: https://go.bd.com/vacutainer-supply-update.html
  • 4.ISO 9000:2005 . Quality management systems – Fundamentals and vocabulary; 2005. Verification, s3.8.5. [Google Scholar]
  • 5.Lippi G, Cornes MP, Grankvist K, et al. EFLM WG-Preanalytical phase opinion paper: Local validation of blood collection tubes in clinical laboratories. Clin Chem Lab Med. 2016;54(5):755–760. 10.1515/cclm-2015-1274 [DOI] [PubMed] [Google Scholar]
  • 6.Nell E-M, Bailly J, Oelofse D, et al. Multicentre verification of haematology laboratory blood collection tubes during a global blood collection tube shortage. Int J Lab Hematol. 2023;45(5):707–716. 10.1111/ijlh.14129 [DOI] [PubMed] [Google Scholar]
  • 7.Clinical and Laboratory Standards Institute . Collection of diagnostic venous blood specimens. Wayne, PA: CLSI; 2017. [Google Scholar]
  • 8.Hawkins R. Discrepancy between visual and spectrophotometric assessment of sample haemolysis. Ann Clin Biochem. 2002;39(5):521–522. 10.1258/000456302320314575 [DOI] [PubMed] [Google Scholar]
  • 9.Atakan Erkal F, Aykal G, Yalçınkaya HM, et al. The effect of automated hemolysis index measurement on sample and test rejection rates. Turkish J Biochem. 2019;44(5):630–634. 10.1515/tjb-2018-0462 [DOI] [Google Scholar]
  • 10.Simundic A-M, Nikolac N, Ivankovic V, et al. Comparison of visual vs. automated detection of lipemic, icteric and hemolyzed specimens: Can we rely on a human eye? Clin Chem Lab Med. 2009;47(11):1361–1365. 10.1515/CCLM.2009.306 [DOI] [PubMed] [Google Scholar]
  • 11.Roche Diagnostics . Cobas® HIL Test (package insert) – version 2 [homepage on the Internet]. Basel: Roche Diagnostics; 2022. [cited 2023 Mar 14]. Available from: https://elabdoc-prod.roche.com/eLD/web/us/en/home [Google Scholar]
  • 12.Roche Diagnostics . Cobas® BILD2 – Bilirubin Direct Gen.2 (package insert) – version 4 [homepage on the Internet]. Basel: Roche Diagnostics; 2022. [cited 2023 Mar 14]. Available from: https://elabdoc-prod.roche.com/eLD/web/us/en/home [Google Scholar]
  • 13.Roche Diagnostics . Cobas® LDHI2 – Lactate dehydrogenase acc. to IFCC ver.2 (package insert) – version 9 [homepage on the Internet]. Basel: Roche Diagnostics; 2022. [cited 2023 Mar 14]. Available from: https://elabdoc-prod.roche.com/eLD/web/us/en/home [Google Scholar]
  • 14.Roche Diagnostics . Cobas® ISE indirect Na-K-Cl for Gen.2 (package insert) – version 9.0 [homepage on the Internet]. Basel: Roche Diagnostics; 2022. [cited 2023 Mar 14]. Available from: https://elabdoc-prod.roche.com/eLD/web/us/en/home [Google Scholar]
  • 15.Clinical and Laboratory Standards Institute . Measurement procedure comparison and bias estimation using patient samples; approved guideline. 3rd ed. Wayne, PA: CLSI; 2013. [Google Scholar]
  • 16.The six sigma calculators [homepage on the Internet]. Madison, WI: Westgard QC Inc.; 2003. [cited 2023 Mar 14]. Available from: https://westgard.com/calculators/calculators [Google Scholar]
  • 17.Bilić-Zulle L. Comparison of methods: Passing and Bablok regression. Biochem Med. 2011;21:49–52. 10.11613/BM.2011.010 [DOI] [PubMed] [Google Scholar]
  • 18.Pum J. A practical guide to validation and verification of analytical methods in the clinical laboratory. In: Makowski GS, editor. Advances in clinical chemistry. Amsterdam: Elsevier; 2019, p. 215–281. [DOI] [PubMed] [Google Scholar]
  • 19.Aarsand AK, Fernandez-Calle P, Webster C, et al. The EFLM biological variation database [homepage on the Internet]. 2023. [cited 2023 Mar 14]. Available from: https://biologicalvariation.eu/
  • 20.Ricos C, Alvarez V, Cava F, et al. Desirable specifications for imprecision, inaccuracy, and total allowable error, calculated from data on within-subject and between-subject biologic variation [homepage on the Internet]. 2014. [cited 2023 Mar 15]. Available from: https://www.westgard.com/biodatabase1.htm
  • 21.Ucar KT, Aksoy N, Erhan B, et al. The local technical validation of new plasma tube with a mechanical separator. Turkish J Biochem. 2020;45(3):329–335. 10.1515/tjb-2019-0354 [DOI] [Google Scholar]
  • 22.Hwang SS, Yoon KR, Yoon KJ. Comparison of three plastic tubes (two Gongdong vacuum blood tubes and improve vacuum blood collection tube) with BD vacutainer tube for 37 laboratory tests. Lab Med Qual Assur. 2009;31:301–308. [Google Scholar]
  • 23.Lima-Oliveira G, Lippi G, Salvagno GL, et al. Preanalytical management: Serum vacuum tubes validation for routine clinical chemistry. Biochem Medica. 2012;22:180–186. 10.11613/BM.2012.021 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Bowen RAR, Sattayapiwat A, Gounden V, et al. Blood collection tube-related alterations in analyte concentrations in quality control material and serum specimens. Clin Biochem. 2014;47(3):150–157. 10.1016/j.clinbiochem.2013.11.003 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Won EJ, Shin M-G, Jang M-J, et al. Comparison between V-tubes and BD vacutainer tubes for use in laboratory tests. Lab Med Online. 2013;3:145–154. 10.3343/lmo.2013.3.3.145 [DOI] [Google Scholar]
  • 26.Jo SJ, Kim YS, Lee S-G, et al. Evaluation of the vacuum blood collection tube, V-tube, a domestic product for clinical chemistry and thyroid hormone assays. Lab Med Qual Assur. 2018;40(4):211–222. 10.15263/jlmqa.2018.40.4.211 [DOI] [Google Scholar]
  • 27.Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res 1999;8(2):135–160. 10.1177/096228029900800204 [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Data supporting the findings of this study are available from the corresponding author, M.K., upon request.