Presence of Tissue Expanders Does Not Affect Radiotherapy Dose Distribution to Heart and Lungs

Jeewon Chon *,, Peter Laub , Mike Wesolowski *, Audrey K Mustoe *, Safi Bajwa *, Taylor Drew *, Naomi Desai *, Nazanin Azarvash *, Jennifer S Wick , Timothy King
PMCID: PMC12119047  PMID: 40438525

Abstract

Background:

Breast cancer treatment often involves mastectomy and postmastectomy radiotherapy (PMRT). PMRT rates are increasing and can improve outcomes in node-positive cases. Although the risks of PMRT to reconstructed breasts are known, the influence of tissue expanders (TEs) on radiation to nearby organs such as the heart and lungs remains unclear.

Methods:

Patients who underwent total mastectomy and completed a full course of PMRT with 3-dimensional computer tomography planning between January 2014 and August 2022 at Loyola University Medical Center were included. Patient dose statistics for ipsilateral lung, heart, and clinical target volume, as well as demographics, clinical characteristics, PRMT boost, and bolus were collected. Dose statistics for ipsilateral lung and heart were compared between mastectomy versus mastectomy + TE, and dose statistics were compared between dichotomized TE intraoperative fill volumes. Correlations between dose statistics and BMI were analyzed.

Results:

A total of 124 patients were included in the study. There were no significant differences in lung or heart radiotherapy across all dose metrics between patients who underwent mastectomy versus mastectomy + TE, or between patients with TE fill volume 60 mL or less versus 60 mL or more. Correlations between BMI and heart maximum dose (P = 0.03) were significantly different and showed a positive, monoclonal correlation (correlation: 0.20, 95% confidence interval: 0.02–0.37).

Conclusions:

The presence of TE and intraoperative expander fill volume did not affect dose distribution or complications to the organs at risk. Increased BMI correlated with an increased maximum dose to the heart.


Takeaways

Question: Does the presence of tissue expanders (TEs) affect the distribution of radiation to organs at risk during postmastectomy radiotherapy (PMRT)?

Findings: TEs and varying fill volumes do not affect complication rates or radiation dose to the heart and lungs during PMRT. A higher body mass index is associated with increased heart radiation.

Meaning: Immediate reconstruction with TEs with varying intraoperative fill volumes is relatively safe to use during PMRT.

INTRODUCTION

Nearly 300,000 women a year are diagnosed with breast cancer in the United States.1 The standard treatment for high-risk disease is mastectomy and radiation therapy. The rate of postmastectomy radiotherapy (PMRT) is on the rise, with 30.6% of T1-2N1 breast cancer patients receiving PMRT in 2004 and 47.1% in 2012.2 PMRT can be performed before or after alloplastic breast reconstruction surgery with silicone implants or tissue expanders (TEs).3,4 Although the risks of radiation such as capsular contraction and explanation from reconstructed breast radiation are well known, there are conflicting data on the risks of radiation to nearby structures such as the heart and lungs.

The goal of a PMRT plan is to deliver the recommended dose of radiation to the mastectomy cavity and chest wall (CW), known as the clinical target volume (CTV). The radiation beams are delivered at different tangents and weights to reach the recommended CTV dose while minimizing radiation to nearby anatomical structures. In PMRT, the heart and lungs are the main organs at risk (OARs) of collateral radiation.5,6

Data on the impact of PMRT on OARs in patients with TEs are limited and conflicting. Studies by Schechter et al7 and Motwani et al8 found that breast reconstruction significantly compromised treatment planning. Schechter et al7 found that only 4 of the 18 plans resulted in optimal treatment of the entire CW while also avoiding the OARs. Motwani et al8 similarly found that 52% of patients had compromised radiation plans compared with 7% of matched controls who did not have intervening reconstruction, with the greatest compromises seen in those with left-sided cancers.

A recent study by Liljegren et al5 showed no difference in dose distribution in OARs with or without breast implants. However, this study did not include TEs with integrated magnetic ports in their evaluation and the patient population exhibited some key demographic differences when compared with the US population. Ohri et al6 showed decreased lung V20 (the percentage of normal lung receiving at least 20 Gy), mean dose, and maximum dose compared with no implant reconstruction.

Various factors affect the distribution of radiation to the OARs. Implants may displace the soft tissue within the CTV and cause imprecise geometric matching of the tangential fields, affecting the radiation distribution.9 In 3-dimensional computer tomography (3D CT) radiation, the OARs have been shown to be difficult to protect. Radiation may cause dose-dependent changes in the heart, such as conduction disorders, coronary artery disease, and cardiomyopathy. It can also cause dose-dependent radiation pneumonitis, one of the most common radiation therapy–induced adverse events.5,8 Yet, few studies have investigated the effect of TEs on radiation dose to the OARs. To our knowledge, no studies investigate the effect of intraoperative expander fill on PMRT planning. Thus, we investigated the differences in PMRT doses to these OARs in patients with mastectomy only versus immediate reconstruction with a TE.

METHODS

Ethical Approval

Ethical approval was obtained from the institutional review board at Loyola University Chicago under approval number 215824 on August 29, 2022. The study was conducted in compliance with the ethical guidelines outlined by the institutional review board.

Study Population

We collected all women with breast cancer who underwent total mastectomy and completed ipsilateral PMRT with 3D CT between January 2014 and August 2022 at our institution. Patient body mass index (BMI), mastectomy date, side of mastectomy, resection weight, tumor size, lymph node (LN) involvement, chemotherapy type, and endocrine therapy use were obtained. TE placement and removal date, as well as intraoperative expander fill (mL) with saline, were collected for the mastectomy + TE cohort. Patients were identified in the ARIA oncology database, which is a software platform that presents radiation oncology treatments. It integrates patient-specific clinical data, such as body anthropometrics and treatment-specific data to support the planning and delivery of radiation therapy. EPIC Electronic Medical Record is integrated with ARIA, allowing streamlined patient management.

We collected dates of PMRT, total dose, number of fractions, LNs irradiated, CTV (CW only or CW plus LNs), and use of boost/bolus. Exclusion criteria included patients with implants other than TEs with magnetic ports, direct-to-silicone implant reconstructions, radiotherapy to only LNs, missing dosimetry data, or fractionation other than 50.4 Gy in 28 fractions.

Assessment of Dosage per Organ

Radiation dose distribution to the OARs was assessed using dose-volume histograms, which graphically represent the percentage of an organ’s volume receiving a specific radiation dose. For example, V20Gy indicates the percentage of the lung volume that receives 20 Gy or more. VGy stands for “volume of Gray” and defines the volume of tissue that receives a specific dose of radiation. These histograms are generated by 3D CT-based treatment planning software (Eclipse, Varian Medical Systems, Palo Alto, CA), which calculates the dose delivered to each organ based on the patient’s anatomy and beam parameters. Figure 1 provides a visual representation of dose distribution in ARIA.

Fig. 1.

Fig. 1.

Typical 3D CT and radiation plan on ARIA.

Data Collection and Variables

Through the ARIA oncology database, we collected dose statistics for the ipsilateral lung, heart, and CTV. For the ipsilateral lung and heart, we captured organ volume (cm3), minimum, maximum, and mean dose (Gy). We captured the percent of lung volume receiving a dose of 20 Gy or more (V20Gy, %) and the percent of heart volume receiving a dose of 25 Gy or more (V25Gy, %), which are standardized dosing measurements based on radiation modeling studies that accounted for radiation to the CW and LN while balancing clinically meaningful toxicities.10,11

We also assessed the coverage of the CTV, which is the area that needs to receive a certain radiation dose. We measured the percentage of CTV covered by at least 95% of the prescribed dose (CTV V95%). We evaluated CTV V95% in both absolute measures (volume in cubic centimeters covered by 95% of the prescribed dose) and relative measures (percentage of CTV covered by 95% of the prescribed dose).

We accounted for both the size and shape of the rib cage by capturing internal transverse diameter, anteroposterior diameter, and hemithorax anteroposterior diameter taken at a specific slice of CT scans called the mammillary slice. We then calculated the chest wall index (CWi) as the ratio between the transverse and anteroposterior diameter. Relevant radiation oncology terms can be found in Table 1.

Table 1.

Relevant Radiation Oncology Terminology

Term Simplified Explanation
ARIA oncology database A software platform used in radiation oncology to manage patient treatment plans, dosimetry data, and clinical information, ensuring streamlined and accurate care
Dose-volume histogram A graph that shows the relationship between the amount of radiation (dose) and the percentage of an organ receiving that dose. For example, it indicates what percentage of the lung receives at least 20 Gy (V20Gy)
Gray (Gy) A unit of measurement for radiation dose, representing the amount of energy absorbed per kilogram of tissue
Volume of Gray (VGy) The amount of tissue volume that absorbs a specific radiation dose (eg, V20Gy means 20 Gy absorbed by a percentage of lung tissue)
CTV The specific area (eg, CW or LNs) that needs to receive radiation for effective treatment
OARs Critical organs near the treatment area, such as the heart and lungs, that are vulnerable to unintended radiation exposure
Bolus A material placed on the skin to increase radiation dose to superficial tissues, helping ensure adequate dose delivery
Boost An extra dose of radiation aimed at a specific area, often used for treating residual cancer cells or high-risk regions
Transverse diameter The width of the chest measured across the body in a horizontal plane, often used in treatment planning
Anteroposterior diameter The depth of the chest measured from the front (sternum) to the back (spine), important for understanding CW geometry
CWi A ratio of the transverse diameter to the anteroposterior diameter, used to describe chest shape and guide radiation planning
V20Gy (lung) The percentage of lung tissue that receives at least 20 Gy of radiation. Higher values increase the risk of lung damage
V25Gy (heart) The percentage of heart tissue that receives at least 25 Gy of radiation, with higher values increasing the risk of heart complications
Maximum dose The highest radiation dose absorbed by any part of an organ. Helps assess the risk of tissue damage
Mean dose The average radiation dose absorbed by the entire organ. Indicates overall exposure level
Fractionation Dividing the total radiation dose into smaller doses delivered over multiple sessions to reduce side effects
3D CT planning A method of using 3D CT scans to create detailed maps for precise radiation delivery to the target while sparing healthy tissue

Statistical Analysis

SAS Version 9.4 was used for all statistical analysis. CW-only (n = 8) and CW + LN (n = 116) radiation plans were evaluated separately, although CW-only plans were not evaluated for statistical significance due to the small observed sample size. The Pearson chi-square or Fisher exact test assessed differences in patients with and without immediate breast reconstruction.

Univariable and multivariable binary logistic regression models estimate the effects of mastectomy procedure, mastectomy side, and CWi (CWi > 2.32 versus CWi < 2.32) on the logit of a lower lung V20Gy exposure (lung V20Gy < 30%) and lower heart mean dose exposure (heart mean dose < 5 Gy), separately. Odds ratios, along with corresponding Wald 95% confidence intervals and Wald chi-square P values, were reported. Lung, heart, and CTV radiotherapy dose metrics were summarized using medians and interquartile ranges, stratified by radiotherapy target and dichotomous intraoperative expander fill volume (<60 mL versus >60 mL). A 60 mL threshold was chosen as it represents the minimum standard expander fill at our institution. Wilcoxon rank sum tests compared dose metrics between fill volume groups among patients receiving radiotherapy to both the CW and LNs.

Spearman correlation coefficients described the bivariate correlations between lung, heart, and CTV dose metrics, BMI, and resection weight stratified by radiotherapy target. The Fisher z transformation was applied to each correlation coefficient, with 95% confidence intervals and P values reported.

RESULTS

Summary Statistics

This study included 124 unique patients undergoing PMRT. Sixty-six (53.23%) of these patients underwent a mastectomy only, whereas 58 (46.77%) underwent mastectomy and immediate TE placement before radiation. Significant differences in mastectomy side (P < 0.01), tumor size (P = 0.04), adjuvant chemotherapy (P = 0.01), bolus (P < 0.01), and age (P < 0.01) were observed between these 2 groups. Significantly more patients who underwent a bilateral mastectomy subsequently underwent reconstruction as well (65.52% versus 28.79%). Tumor size distribution was significantly different between the 2 mastectomy procedure groups. The bolus rate was significantly higher for patients who underwent mastectomy alone than for patients who underwent mastectomy plus reconstruction (93.94% versus 72.41%). Patients who underwent mastectomy alone were significantly older than patients who underwent mastectomy plus reconstruction (mean: 62.62 versus 48.19 y) (Table 2).

Table 2.

Comparison of Patient Demographics, Clinical Characteristics, and Radiotherapy Treatments Between Mastectomy Alone and Mastectomy + TE Implant for Immediate Breast Reconstruction Before PMRT

Variable Mastectomy (n = 66), n (%) Mastectomy + Reconstruction (n = 58), n (%) Overall (n = 124), n (%) P
Race 0.21
 Asian 8 (12.12) 2 (3.45) 10 (8.06)
 Black 9 (13.64) 13 (22.41) 22 (17.74)
 Other 6 (9.09) 7 (12.07) 13 (10.48)
 White 43 (65.15) 36 (62.07) 79 (63.71)
Ethnicity 0.29*
 Missing 1 (1.52) 0 (0.00) 1 (0.81)
 Hispanic origin 6 (9.09) 9 (15.52) 15 (12.10)
 Non-Hispanic origin 59 (89.39) 49 (84.48) 108 (87.10)
Mastectomy side <0.01
 Bilateral 19 (28.79) 38 (65.52) 57 (45.97)
 Left 23 (34.85) 7 (12.07) 30 (24.19)
 Right 24 (36.36) 13 (22.41) 37 (29.84)
Tumor size 0.04
 pT0 13 (19.70) 9 (15.52) 22 (17.74)
 pT1 16 (24.24) 21 (36.21) 37 (29.84)
 pT2 12 (18.18) 18 (31.03) 30 (24.19)
 pT3/pT4 25 (37.88) 10 (17.24) 35 (28.23)
Lymph nodes 0.51
 pN0 30 (45.45) 19 (32.76) 49 (39.52)
 pN1 21 (31.82) 24 (41.38) 45 (36.29)
 pN2 8 (12.12) 9 (15.52) 17 (13.71)
 pN3 7 (10.61) 6 (10.34) 13 (10.48)
Adjuvant chemotherapy 24 (36.36) 35 (60.34) 59 (47.58) 0.01
Neoadjuvant chemotherapy 37 (56.06) 34 (58.62) 71 (57.26) 0.77
Endocrine therapy 32 (48.48) 20 (34.48) 52 (41.94) 0.11
Radiotherapy target 0.28
 Chest wall 6 (9.09) 2 (3.45) 8 (6.45)
 Chest wall + lymph nodes 60 (90.91) 56 (96.55) 116 (93.55)
Boost 57 (86.36) 49 (84.48) 106 (85.48) 0.77
Bolus 62 (93.94) 42 (72.41) 104 (83.87) <0.01
n Mean (SD) n Mean (SD) n Mean (SD)
Age at mastectomy, y 66 62.62 (12.47) 58 48.19 (10.18) 124 55.87 (13.51) <0.01
BMI, kg/m2 66 28.52 (6.77) 58 30.28 (6.81) 124 29.34 (6.82) 0.15
*

Chi-square test excludes missing values.

Significant at α = 0.05 level.

The Fisher exact test P value.

Complication Profile

The rates of common radiation-related complications were analyzed between the mastectomy-only and mastectomy + reconstruction groups, with no statistically significant differences observed. Radiation dermatitis occurred in 59 (89.4%) patients in the mastectomy-only group and in 53 (91.4%) patients in the mastectomy + reconstruction group (P = 0.769). Radiation pneumonitis was observed in 1 (1.5%) patient in the mastectomy-only group and in 2 (3.4%) patients in the mastectomy + reconstruction group, with no statistically significant difference (P = 0.599). Cardiology adverse events were also evaluated. Coronary artery disease was observed in 3 patients in both groups before radiation. After radiation, coronary artery disease was seen in 3 patients in the mastectomy-only group and in 4 patients in the mastectomy + reconstruction group, with no statistically significant difference (P = 0.705). Pericarditis was not observed in any patient before radiation. However, after radiation, no cases of pericarditis were reported in the mastectomy-only group, whereas 1 case was observed in the mastectomy + reconstruction group. This difference was not statistically significant (P = 0.468). All other cardiology complications were not reported (Table 3).

Table 3.

Complications in Mastectomy Alone Versus Mastectomy + TE Implant for Immediate Breast Reconstruction

Complication Mastectomy Only Mastectomy + Reconstruction P
Radiation dermatitis 59 (89.4%) 53 (91.4%) 0.769
Radiation pneumonitis 1 (1.5%) 2 (3.4%) 0.599
Cardiology adverse events Before radiation After radiation Before radiation After radiation
Coronary artery disease 3 3 3 4 0.705
Pericarditis 0 0 0 1 0.468

Dosimetry of OARs

Univariate analysis revealed that immediate reconstruction after mastectomy did not affect lung or heart radiotherapy dose metrics (Table 4). Significant differences in CTV (P < 0.01), V95% (P < 0.01), V95% cm3 (P < 0.01), V105% cm3 (P = 0.01), transverse diameter (P < 0.01), anteroposterior diameter (P < 0.01), and ipsilateral internal diameter (P < 0.01) were observed between mastectomy procedure groups. All these metrics were significantly larger, higher, or longer for patients who underwent mastectomy plus reconstruction, when compared with patients who underwent mastectomy only (Table 5).

Table 4.

OAR Volumes and Radiation Doses, by Radiotherapy Target and Mastectomy Procedure

OAR Metrics, Median (IQR) CW CW + LNs
Mastectomy (n = 6) Mastectomy + Reconstruction (n = 2) Overall (n = 8) Mastectomy (n = 60) Mastectomy + Reconstruction (n = 56) Overall (n = 116) P
Lung
 Volume, cm3 2209.65 (2120.00–2217.70) 2123.95 (2089.29–2158.61) 2183.21 (2104.65–2214.60) 1954.25 (1595.05–2381.97) 1989.33 (1811.25–2482.96) 1979.70 (1659.50–2419.05) 0.47
 Minimum dose, Gy 0.20 (0.16–0.69) 0.07 (0.00–0.15) 0.16 (0.11–0.46) 0.34 (0.20–0.64) 0.36 (0.23–0.75) 0.35 (0.21–0.70) 0.59
 Maximum dose, Gy 51.99 (50.81–54.59) 49.68 (49.14–50.22) 51.39 (49.68–53.31) 52.17 (50.97–55.90) 51.70 (50.87–52.55) 51.89 (50.88–53.47) 0.14
 Mean dose, Gy 12.63 (9.10–13.97) 9.14 (7.12–11.16) 11.63 (8.89–13.56) 15.12 (13.49–16.66) 15.97 (14.55–17.72) 15.72 (13.64–17.27) 0.05
 V20Gy, % 23.91 (22.62–27.08) 17.50 (12.82–22.18) 23.10 (18.88–25.66) 28.35 (24.68–31.29) 29.48 (26.69–33.31) 28.95 (25.26–32.33) 0.11
Heart
 Volume, cm3 416.50 (394.20–552.90) 615.66 (571.21–660.11) 487.15 (402.90–615.66) 544.65 (465.75–611.45) 559.72 (493.80–651.05) 550.90 (476.48–634.10) 0.31
 Minimum dose, Gy 0.27 (0.03–0.54) 0.05 (0.01–0.09) 0.09 (0.02–0.49) 0.22 (0.09–0.32) 0.21 (0.06–0.41) 0.22 (0.07–0.36) 0.99
 Maximum dose, Gy 7.50 (5.42–15.69) 15.86 (6.27–25.45) 7.50 (5.85–20.57) 22.52 (7.11–35.93) 15.70 (6.75–35.54) 19.19 (6.79–35.72) 0.66
 Mean dose, Gy 1.19 (0.69–2.09) 0.79 (0.52–1.06) 1.00 (0.61–1.76) 1.54 (1.00–2.28) 1.71 (0.97–2.46) 1.64 (0.97–2.33) 0.68
 V25Gy, % 0.00 (0.00–0.00) 0.00 (0.00–0.00) 0.00 (0.00–0.00) 0.00 (0.00–0.06) 0.00 (0.00–0.02) 0.00 (0.00–0.04) 0.31

Significant at α = 0.05 level.

Table 5.

CTV and Rib Cage Metrics, by Radiotherapy Target and Mastectomy Procedure

CTV and Rib Cage Metrics, Median (IQR) CW CW + LNs
Mastectomy (n = 6) Mastectomy + Reconstruction (n = 2) Overall (n = 8) Mastectomy (n = 60) Mastectomy + Reconstruction (n = 56) Overall (n = 116) P
CTV
 CTV, cm3 289.95 (266.20–765.20) 654.28 (420.70–887.85) 357.15 (276.25–811.70) 488.55 (343.74–674.12) 1015.10 (728.36–1209.54) 688.63 (435.54–1015.10) <0.01*
 CTV mean dose, Gy 52.33 (51.11–54.24) 49.27 (48.95–49.59) 51.18 (50.08–53.83) 51.45 (50.81–52.12) 51.56 (51.09–52.37) 51.50 (50.88–52.14) 0.31
 V95%, % 95.16 (91.43–98.65) 94.06 (93.05–95.06) 94.36 (92.24–97.66) 95.67 (93.59–98.41) 98.04 (95.55–99.46) 96.99 (94.66–99.00) <0.01*
 V95%, cm3 272.61 (263.46–675.29) 613.03 (399.92–826.14) 338.35 (265.95–739.50) 464.85 (338.90–647.57) 980.14 (723.35–1108.40) 647.57 (422.88–980.14) <0.01*
 V105%, % 25.04 (8.81–43.29) 3.80 (0.04–7.56) 15.00 (4.12–36.09) 15.69 (6.67–27.79) 16.30 (5.78–30.80) 16.16 (6.54–28.31) 0.92
 V105%, cm3 72.46 (23.45–331.26) 33.65 (0.17–67.12) 64.67 (12.57–206.99) 61.55 (31.64–143.57) 111.02 (48.84–315.85) 81.83 (37.73–230.80) 0.01*
Rib cage
 Transverse diameter, cm 23.60 (22.39–24.50) 24.43 (23.46–25.39) 23.76 (22.77–24.95) 21.70 (20.34–23.13) 24.26 (23.47–24.88) 23.18 (21.51–24.41) <0.01*
 Anteroposterior diameter, cm 11.24 (10.32–11.86) 11.03 (10.19–11.87) 11.24 (10.26–11.87) 9.33 (8.21–9.96) 10.57 (9.46–11.45) 9.66 (8.73–10.90) <0.01*
 Ipsilateral internal diameter, cm 16.80 (14.81–18.03) 16.81 (16.62–17.00) 16.81 (15.39–17.84) 14.03 (12.81–14.97) 15.92 (14.86–16.66) 14.86 (13.58–16.08) <0.01*
 CWi 2.14 (2.03–2.32) 2.23 (1.98–2.49) 2.14 (2.00–2.41) 2.40 (2.21–2.65) 2.29 (2.15–2.53) 2.34 (2.16–2.63) 0.20
*

Significant at α = 0.05 level.

Regressions

The median CWi was 2.32. Mastectomy procedure (P = 0.77), mastectomy side (P = 0.17), and CWi (P = 0.63) all did not demonstrate a significant effect on lung or heart mean dose (Tables 6, 7). In addition, these variables did not demonstrate a significant effect on lung V20Gy among patients who had both their CWs and their LNs irradiated in adjusted analyses.

Table 6.

Unadjusted Effects of Mastectomy Procedure, Mastectomy Side, and CWi on Lung Radiation Exposure, by Radiotherapy Target

Predictor CW CW + LNs
Lung V20Gy ≤ 30%, n %) Lung V20Gy > 30%, n (%) Lung V20Gy ≤ 30%, n (%) Lung V20Gy > 30%, n (%) OR (95% CI) P
Mastectomy procedure
 Mastectomy + reconstruction 2 (100.00) 0 (0.00) 34 (60.71) 22 (39.29) 0.90 (0.42–1.90) 0.77
 Mastectomy 6 (100.00) 0 (0.00) 38 (63.33) 22 (36.67) 1.00 (REF)
Mastectomy side
 Bilateral 0 (0.00) 0 (0.00) 36 (63.16) 21 (36.84) 1.71 (0.71–4.12) 0.23
 Left 3 (100.00) 0 (0.00) 20 (74.07) 7 (25.93) 2.86 (0.95–8.63) 0.06
 Right 5 (100.00) 0 (0.00) 16 (50.00) 16 (50.00) 1.00 (REF)
CWi
  CWi > 2.32 2 (100.00) 0 (0.00) 36 (60.00) 24 (40.00) 0.83 (0.39–1.77) 0.63
  CWi ≤ 2.32 6 (100.00) 0 (0.00) 36 (64.29) 20 (35.71) 1.00 (REF)

Significant at α = 0.05 level.

CI, confidence interval; OR, odds ratio; REF, reference.

Table 7.

Unadjusted Effects of Mastectomy Procedure, Mastectomy Side, and CWi on Heart Radiation Exposure, by Radiotherapy Target

Predictor CW CW + LNs
Heart Mean, Dose ≤ 5 Gy, n (%) Heart Mean, Dose > 5 Gy, n (%) Heart Mean, Dose ≤ 5 Gy, n (%) Heart Mean, Dose > 5 Gy, n (%) OR (95% CI) P
Mastectomy procedure
 Mastectomy + reconstruction 2 (100.00) 0 (0.00) 55 (98.21) 1 (1.79) 2.89 (0.29–28.68) 0.36
 Mastectomy 6 (100.00) 0 (0.00) 57 (95.00) 3 (5.00) 1.00 (REF)
Mastectomy side
 Bilateral 0 (0.00) 0 (0.00) 55 (96.49) 2 (3.51) 0.89 (0.08–10.18) 0.92
 Left 3 (100.00) 0 (0.00) 26 (96.30) 1 (3.70) 0.84 (0.05–14.08) 0.90
 Right 5 (100.00) 0 (0.00) 31 (96.88) 1 (3.13) 1.00 (REF)
CWi
 CWi > 2.32 2 (100.00) 0 (0.00) 57 (95.00) 3 (5.00) 0.35 (0.04–3.42) 0.36
 CWi ≤ 2.32 6 (100.00) 0 (0.00) 55 (98.21) 1 (1.79) 1.00 (REF)

Significant at α = 0.05 level.

CI, confidence interval; OR, odds ratio.

Mastectomy + TE Dosimetry

Among patients who underwent mastectomy plus reconstruction and had both their CWs and LNs targeted by radiotherapy, 21 (37.50%) had an intraoperative expander fill volume of 60 mL or less, and 35 (62.50%) had an intraoperative expander fill volume greater than 60 mL. We found that expander fill volume did not affect radiotherapy metrics among patients who underwent mastectomy plus reconstruction (Table 8).

Table 8.

Radiotherapy Metrics Among Patients Undergoing Mastectomy + Reconstruction, by Radiotherapy Target and Intraoperative Expander Fill Volume

Radiotherapy Metric, Median (IQR) CW CW + LNs
≤ 60 mL Fill Volume (n = 1) > 60 mL Fill Volume (n = 1) Overall (n = 2) ≤ 60 mL Fill Volume (n = 21) > 60 mL Fill Volume (n = 35) Overall (n = 56) P
Lung
 Minimum dose, Gy 0.15 (0.15–0.15) 0.00 (0.00–0.00) 0.07 (0.00–0.15) 0.30 (0.14–0.75) 0.41 (0.23–0.76) 0.36 (0.23–0.75) 0.45
 Maximum dose, Gy 49.14 (49.14–49.14) 50.22 (50.22–50.22) 49.68 (49.14–50.22) 51.60 (51.12–51.90) 51.93 (50.76–53.44) 51.70 (50.87–52.55) 0.51
 Mean dose, Gy 11.16 (11.16–11.16) 7.12 (7.12–7.12) 9.14 (7.12–11.16) 15.87 (14.99–17.70) 16.22 (13.91–17.73) 15.97 (14.55–17.72) 0.93
 V20Gy, % 22.18 (22.18–22.18) 12.82 (12.82–12.82) 17.50 (12.82–22.18) 29.28 (26.97–33.95) 29.52 (26.24–33.29) 29.48 (26.69–33.31) 0.98
Heart
 Minimum dose, Gy 0.01 (0.01–0.01) 0.09 (0.09–0.09) 0.05 (0.01–0.09) 0.16 (0.06–0.39) 0.26 (0.07–0.42) 0.21 (0.06–0.41) 0.78
 Maximum dose, Gy 6.27 (6.27–6.27) 25.45 (25.45–25.45) 15.86 (6.27–25.45) 15.96 (7.32–33.87) 14.55 (6.53–39.87) 15.70 (6.75–35.54) 0.97
 Mean dose, Gy 0.52 (0.52–0.52) 1.06 (1.06–1.06) 0.79 (0.52–1.06) 1.66 (1.02–2.23) 1.87 (0.96–2.46) 1.71 (0.97–2.46) 0.95
 V25Gy, % 0.00 (0.00–0.00) 0.00 (0.00–0.00) 0.00 (0.00–0.00) 0.00 (0.00–0.02) 0.00 (0.00–0.04) 0.00 (0.00–0.02) 0.89
CTV
 CTV, cm3 887.85 (887.85–887.85) 420.70 (420.70–420.70) 654.28 (420.70–887.85) 915.00 (692.51–1083.00) 1097.30 (833.00–1249.90) 1015.10 (728.36–1209.54) 0.09
 CTV mean dose, Gy 48.95 (48.95–48.95) 49.59 (49.59–49.59) 49.27 (48.95–49.59) 51.34 (50.81–51.87) 51.75 (51.24–52.51) 51.56 (51.09–52.37) 0.11

Significant at α = 0.05 level.

Mastectomy Resection Weight Correlations

Correlations between resection weight and lung minimum dose (P = 0.04), and between resection weight and CTV (P = 0.01) were significantly different among patients who had both their CWs and LNs targeted by radiotherapy.

BMI Correlations

Correlations between BMI and lung maximum dose (P = 0.02), and between BMI and heart maximum dose (P = 0.03) were significantly different. The correlation between BMI and maximum heart dose showed a positive, monoclonal correlation (correlation: 0.20, 95% confidence interval: 0.02–0.37). The correlation between BMI and CTV (P < 0.01) was significantly different among patients who had both their CWs and LNs targeted by radiotherapy (P < 0.01).

DISCUSSION

Our study aimed to address the discrepancies in the literature about PMRT risks in patients with reconstructed breasts. Prior studies by Schechter et al7 and Motwani et al8 showed that reconstructed breasts compromise radiation treatment plans. Liljegren et al5 found no difference in radiation to OARs in both reconstructed and nonreconstructed patients, but they did not include commonly used TEs with integrated magnetic ports. Our findings demonstrate no statistically significant differences in rates of radiation dermatitis, radiation pneumonitis, or cardiac complications between the mastectomy-only and mastectomy + reconstruction groups following radiation therapy.

Magnetic Ports

Patients with breast reconstruction undergoing radiation are at risk for adverse events such as capsular contracture and wound healing complications. A 2-stage reconstruction using temporary TEs (TTEs) is often used.12 Most TTEs have a magnetic port through which saline is injected to expand it during surgery and subsequent filling sessions. At our institution, we used single-port TTEs for our patients. Several studies have investigated the dosimetric impact of single-port TTEs on radiation plans with varying results. Some studies report between 5% and 30% dose perturbations, whereas some reported negligible differences.1315 Magnetic fill ports often contain high-density metal materials, which may cause scattering of the radiation beams and cause a phenomenon called “dose shadowing,” which reduces the dose to the target tissue directly behind the port, or a “hot spot,” which slightly increases the dose at the edges of the port due to the scattering effects.13 No studies have yet investigated the effect of magnetic single-port TTEs on radiation to OARs. Our research suggests that magnetic single-port TTEs can be safely used during radiation, independent of patient morphological characteristics such as BMI and CWi.

Fill Volumes

TTEs are typically used to expand the breast tissue before implantation of permanent implants in breast reconstruction. Studies have shown inconsistent outcomes when investigating TTE fill volume and complication rates. A study by Yazar et al16 analyzed complications of Becker-type expanders and found that intraoperative fill volume seemed to be a predictor of developing capsular contracture. Kadakia et al17 found that larger saline fill volumes greater than 350 mL were not associated with postoperative complications. Although some studies evaluated complication risks with varying TTE fill volumes, their effects on PMRT are limited in the literature. Our research is the first to support that varying fill volume does not impact radiation to the OARS.

CW Geometry

Our study was designed to detect whether certain patient factors that affect the geometry of the chest such as BMI, CWi, and resection weight could affect PMRT delivery in women with TEs. We did not find significant differences with regard to these factors except for BMI. Higher BMI correlated with increased maximum heart dose, which may influence radiotherapy timing decisions for high-BMI patients.

Boost/Bolus

In PMRT, a bolus is a material placed on the skin to modify the distribution of dose delivered to the underlying and surrounding tissues. It is widely recommended that the target area should receive at least 95% of the predesigned PMRT dose. However, superficial targets often cannot receive a sufficient dose due to the skin-sparing effect of high-energy photon beams. Boluses are predominantly used to provide a “dose buildup,” intensifying the radiation delivered to underlying structures and minimizing the risk of underdosing target areas.18 Boluses come at a risk of a higher rate of acute grade 3 radiation dermatitis, which must be considered when evaluating skin integrity.19 In our study, patients with immediate reconstruction with TTEs were significantly less likely to receive a bolus.

Boosts refer to an additional, localized dose of radiation that is delivered to or around the primary target field. They are often used to treat residual areas or to compensate for decreasing dosages as the radiation penetrates tissue.20 In our study, there were no differences in boost use between patients with immediate or delayed reconstruction with TTEs.

Study Limitations

Our study has several limitations. The retrospective nature made tracking specific patient information difficult, specifically the TTE fill volumes and practice patterns. Fill volume determination varied by surgeon discretion, leading to inconsistencies in timing and extent of expansion at the start of PMRT. Although we attempted to track these filling patterns as closely as possible to the radiation start dates, the variability in practice patterns must be considered when evaluating these data.

Our decision to dichotomize TE fill volumes into 60 mL or less and greater than 60 mL, rather than as a continuous variable was driven by the distribution of fill volumes within our patient population. Patients receiving immediate breast reconstruction were typically managed in 1 of 2 categories: (1) 60 mL or less (predominantly 60 or 0 mL in cases where expanders were left unfilled), and (2) greater than 60 mL(expanders were filled beyond 60 mL) (Fig. 2). Dichotomization at this threshold enabled us to evaluate clinically relevant differences in radiation dose metrics between these 2 categories.

Fig. 2.

Fig. 2.

Patient with 600 mL intraoperative tissue expander saline fill. RT, radiation therapy. PTV, planning target volume; RT CW CP, right chest wall clinical plan.

In addition, although radiation dermatitis presents shortly after radiation therapy, radiation pneumonitis or cardiac complications such as coronary artery disease, cardiomyopathy, and arrhythmias may form years to decades after radiation therapy, which would not be captured in our data analysis. Longitudinal studies that capture these complication rates would be a useful adjunct when evaluating the safety of radiation therapy.

The use of bolus and boost techniques in PMRT was not fully explored in this study. Although we found that patients with immediate reconstruction had a lower likelihood of receiving a bolus, the underlying reasons for the finding and clinical implications warrant further investigation. The presence of a TTE may make it challenging to achieve proper dose buildup or may lead to increased tension on the skin and underlying tissue; thus, bolus may be avoided by radiation oncology teams.18,19 These themes should be explicitly explored in future research.

CONCLUSIONS

Our study demonstrates that the presence of TEs with magnetic ports and varying intraoperative fill volumes do not significantly affect radiation dose distribution or complication rates to the heart and lungs during PMRT. Immediate breast reconstruction with TEs can be safely performed without compromising radiation treatment plans. However, increased BMI correlated with a higher maximum heart dose, which may be a consideration for high-BMI patients. These findings support the safety of immediate breast reconstruction with TEs in the context of planned PMRT, potentially allowing more patients to benefit from immediate reconstruction without compromising oncological treatment.

DISCLOSURE

The authors have no financial interest to declare in relation to the content of this article.

Footnotes

Published online 28 May 2025.

Presented at PTSM 2023, October 26–29, 2023, Austin, TX, MAPS 2024, May 10–11, 2024, Chicago, IL.

Disclosure statements are at the end of this article, following the correspondence information.

REFERENCES

  • 1.Siegel RL, Miller KD, Wagle NS, et al. Cancer statistics, 2023. CA Cancer J Clin. 2023;73:17–48. [DOI] [PubMed] [Google Scholar]
  • 2.Li FY, Lian CL, Lei J, et al. Real-world impact of postmastectomy radiotherapy in T1–2 breast cancer with one to three positive lymph nodes. Ann Transl Med. 2020;8:489–489. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Ribuffo D, Lo Torto F, Giannitelli SM, et al. The effect of post-mastectomy radiation therapy on breast implants: unveiling biomaterial alterations with potential implications on capsular contracture. Mater Sci Eng C Mater Biol Appl. 2015;57:338–343. [DOI] [PubMed] [Google Scholar]
  • 4.Berry MG, Cucchiara V, Davies DM. Breast augmentation: part II--adverse capsular contracture. J Plast Reconstr Aesthet Surg. 2010;63:2098–2107. [DOI] [PubMed] [Google Scholar]
  • 5.Liljegren A, Unukovych D, Gagliardi G, et al. No difference in dose distribution in organs at risk in postmastectomy radiotherapy with or without breast implant reconstruction. Radiat Oncol. 2014;9:14. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Ohri N, Cordeiro PG, Keam J, et al. Quantifying the impact of immediate reconstruction in postmastectomy radiation: a large, dose-volume histogram-based analysis. Int J Radiat Oncol Biol Phys. 2012;84:e153–e159. [DOI] [PubMed] [Google Scholar]
  • 7.Schechter NR, Strom EA, Perkins GH, et al. Immediate breast reconstruction can impact postmastectomy irradiation. Am J Clin Oncol. 2005;28:485–494. [DOI] [PubMed] [Google Scholar]
  • 8.Motwani SB, Strom EA, Schechter NR, et al. The impact of immediate breast reconstruction on the technical delivery of postmastectomy radiotherapy. Int J Radiat Oncol Biol Phys. 2006;66:76–82. [DOI] [PubMed] [Google Scholar]
  • 9.Buchholz TA, Kronowitz SJ, Kuerer HM. Immediate breast reconstruction after skin-sparing mastectomy for the treatment of advanced breast cancer: radiation oncology considerations. Ann Surg Oncol. 2002;9:820–821. [DOI] [PubMed] [Google Scholar]
  • 10.Marks LB, Bentzen SM, Deasy JO, et al. Radiation dose-volume effects in the lung. Int J Radiat Oncol Biol Phys. 2010;76:S70–S76. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Gagliardi G, Constine LS, Moiseenko V, et al. Radiation dose–volume effects in the heart. Int J Radiat Oncol Biol Phys. 2010;76:S77–S85. [DOI] [PubMed] [Google Scholar]
  • 12.Bellini E, Pesce M, Santi P, et al. Two-stage tissue-expander breast reconstruction: a focus on the surgical technique. Biomed Res Int. 2017;2017:1–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Damast S, Beal K, Ballangrud A, et al. Do metallic ports in tissue expanders affect postmastectomy radiation delivery? Int J Radiat Oncol Biol Phys. 2006;66:305–310. [DOI] [PubMed] [Google Scholar]
  • 14.Yoon J, Xie Y, Heins D, et al. Modeling of the metallic port in breast tissue expanders for photon radiotherapy. J Appl Clin Med Phys. 2018;19:205–214. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Trombetta DM, Cardoso SC, Facure A, et al. Influence of the presence of tissue expanders on energy deposition for post-mastectomy radiotherapy. PLoS One. 2013;8:e55430. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Yazar S, Karadag EC, Altinkaya A, et al. Risk factor analysis for survival of Becker-type expander in immediate breast reconstruction. Aesthetic Plast Surg. 2018;42:971–979. [DOI] [PubMed] [Google Scholar]
  • 17.Kadakia N, Swisher AR, Lewis PG, et al. Are large intraoperative fill volumes associated with increased complications after tissue expander placement? Eplasty. 2023;23:e12. [PMC free article] [PubMed] [Google Scholar]
  • 18.Lu Y, Song J, Yao X, et al. 3D printing polymer-based bolus used for radiotherapy. Int J Bioprint. 2021;7:414. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Dahn HM, Boersma LJ, de Ruysscher D, et al. The use of bolus in postmastectomy radiation therapy for breast cancer: a systematic review. Crit Rev Oncol Hematol. 2021;163:103391. [DOI] [PubMed] [Google Scholar]
  • 20.Dzhugashvili M, Veldeman L, Kirby AM. The role of the radiation therapy breast boost in the 2020s. Breast. 2023;69:299–305. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Plastic and Reconstructive Surgery Global Open are provided here courtesy of Wolters Kluwer Health

RESOURCES