Abstract
Ultrasound driven cavitation is widely used to intensify lab and industrial-scale processes. Various studies and experiments demonstrate that the acoustic energy, dissipated through the bubbles collapse, leads to intense physico-chemical effects in the processed liquid. A better understanding of these phenomena is crucial for the optimization of ultrasonic reactors, and their scale-up. In the current literature, the visual characterization of the reactor is mainly carried out with sonoluminescence and sonochemiluminescence. These techniques have limitations in the time resolution since a high camera exposure time is required. In this research, we proposed an alternative method, based on coumarin dosimetry to monitor the hydroxylation activity, and high-speed imaging for the visualization of the vapor field. By this approach, we aim to capture the structure and the dynamics of the vapor field and to correlate this with the chemical effects induced in the ultrasonic reactor. This characterization was carried out for four different ultrasonic probe diameters (3, 7, 14 and 40 mm), displacement amplitudes and processing volumes. Key findings indicate that the probe diameter strongly affects the structure of the vapor field and the chemical effectiveness of the system. The proposed methodology could be applied to characterize other types of ultrasonic reactors with different operating and processing conditions.
Keywords: Acoustic cavitation, Sonochemistry, Coumarin dosimetry, High-speed visualization
1. Introduction
Ultrasound finds a rather unique position in human lives with its utility ranging from human diagnostics to intensifying chemical reactions, and aiding in the detection of pipe corrosion. The passage of sufficiently intense ultrasound waves, into a liquid medium, generates the phenomenon of cavitation, with varying extent of physical and chemical transformations brought about in the bulk of the liquid. Broadly 20 to 40 kHz is termed as conventional power ultrasound used for physical and biological applications, with a great utility to process intensification. 40 to 500 kHz can be termed as the range for observing sonochemistry, with applications in sonolysis-based water treatment and food processing.
The interest in a systematic understanding of ultrasonic reactors has been the subject of research from the late 1990’s, propelled by numerous experimental reports through the 1900’s [1], [2], [3]. Due to this profound interest, systematic efforts in expanding the depth of investigation of ultrasound based processes emerged experimentally [4], [5], [6], [7], [8], [9], [10], [11] and numerically [12], [13], [14], [15], [16], [17], [18], and some to correlate them [19], [20], [21], [22]. This past research was crucial for a comprehensive characterization of the sonochemical reactors, which can lead to the design of optimized reactors. The key aspect is the identification of the phenomenological relation between the structure of the acoustic pressure field, the cavitation activity in the domain, and the physico-chemical effects induced by the dynamics and collapse of the vapor bubbles and clusters [23], [24], [25].
The efficiency of these types of reactors has to be evaluated by relating the physical or chemical effects, for which the reactor was designed, with the acoustic energy introduced into the system by the ultrasound source. Some of the most relevant physical effects induced by the collapse of the cavitation bubbles are: emulsification [26], [27], sonofragmentation [28], [29], and leaching [30], [31]. One of the most advanced tools for the stand-alone investigation of these phenomena is the high-speed imaging, since it is able to provide extremely fine spatial and temporal resolutions [25], [27], [32], [33].
The chemical effects of ultrasound cavitation, in an aqueous solution, comprise the generation of different radical species such as OH, OOH, and so on. These are due to the dissociation of water molecules trapped in the cavitation bubbles during their collapse. The radicals generated could play a significant role in increasing the rate of different chemical reactions in the processed solution [34], [35]. The most used methodology to quantify the radicals, generated by sonochemistry, is the “chemical dosimetry” [36], [37]. These techniques provide average information about the reactivity of the system through ex-situ radicals quantification. Two visual techniques are commonly used for in-situ characterizations of the reactive zones of sonochemical reactors: sonoluminescence (SL) and sonochemiluminescence (SCL). In these techniques, a camera sensor collects the light emitted from the collapse of cavitation bubbles (SL) or the reaction between hydroxyl radicals (OH) with luminol, deliberately dosed into the system (SCL) [38]. Qualitative information can be obtained regarding the 2D distribution of the reactivity in the domain. Regarding these two techniques, it is important to highlight that an extremely high exposure time (15 s or longer) is required to obtain a detectable signal on the camera sensor [38], [39], [40], [41]. This constitutes a significant disadvantage since the information is averaged over a time span of orders of magnitude higher than the characteristic time of the ultrasonic waves and cavitation phenomena.
Considering the shortcomings of SL and SCL, a two-pronged approach of high-speed imaging of the system and an independent chemical estimation could be of better utility to individually quantify the physical and chemical effects of the cavitation system thereof. In the present work, this was achieved by monitoring the hydroxylation product of coumarin and correlating this with the structure and the dynamics of the cavitation vapor field, visualized by high-speed imaging. The high-speed visualization experiments were carried out at a frame rate relevant to capture the dynamics and the structure of the vapor field, generated by cavitation in the liquid domain.
Although ultrasound reactors can be realized in a variety of different configurations (e.g. horn-type transducer or ultrasonic bath) and operational modes (e.g. working frequency, area of the emitting surface, acoustic power etc.) [10], [42], the present case study concerns a lab-scale, horn-type ultrasound device with a working frequency of 24 kHz as it is a type of device commonly used, provides a direct contact of the emitting surface with the liquid medium, and this frequency has the potential to be applied to continuous flow processing [43], [44], [45], [46]. Several studies have investigated the impact of the design parameters on the efficiency of sonochemical reactors [39], [47].
For the horn-type configuration, the parameters of primary importance are: probe diameter, vibration amplitude, and volume processed [40]. These are investigated here. The trends obtained in terms of quantitative hydroxylation will help ascertain the behavior of the vapor structure contributing to effective chemical activity. This approach would allow us to highlight the working principles of the sonochemical reactor and understand the main scaling parameters. Although the analysis has been carried out on a single-frequency ultrasonic device, the intent is to propose a methodology that could be applied, with some variations, to study other types of reactors and transducers characterized by different operating conditions and process objectives.
2. Materials and methods
2.1. Experimental setup
For the generation of the ultrasonic waves in the liquid medium, a Hielscher device UP400S (24 kHz) was used. This device absorbs a maximum power of 400 W and titanium probes having different diameters can be installed on it. The vibration amplitude of the probe can be regulated, on the ultrasonic device, in terms of percentage with respect to the maximum (from 30% to 100%). At every percentage amplitude, it is possible to associate the actual amplitude in micrometers and the calorimetric power (relevant parameters are given in Table 1, details are presented in Sections 2.4, 3.1). In the present study, four different probe diameters () have been tested: 3, 7, 14 and 40 mm. For the coumarin dosimetry experiments (details in Section 2.3), the water solution was placed in a cylindrical glass beaker. Two beaker sizes were used to process two different solution volumes (): 200 and 400 mL. The shape of the container was chosen to have the container walls several diameters away from the emitting surface of the probe. For every configuration, the immersion depth of the probe in the solution was kept constant to 25 mm. The temperature was monitored with a K-type wire thermocouple, immersed in the liquid at the corner of the beaker, to avoid any alterations to the ultrasound cavitation field. The electrical power, absorbed by the ultrasound device, was measured with a socket-type multimeter.
Table 1.
Table containing the most relevant quantities (description in Nomenclature and Acronyms Appendix) related to the experiments performed with different probe diameters and displacement amplitudes. For all the experimental cases, the vibration frequency of the probe was fixed at 24 kHz. The acoustic intensity () is derived from the calorimetric power ().
| [mm] | [%] | [m] ( 2.35) | [W] ( 0.5) | [W] ( 0.5) | [W/mm] | () [mm] | |
|---|---|---|---|---|---|---|---|
| 3 | 30 | 23.54 | 10.8 | 1.1 | 0.10 | 0.16 | 0.28 |
| 3 | 50 | 25.89 | 17 | 5.6 | 0.33 | 0.79 | 0.41 |
| 3 | 70 | 47.08 | 23.8 | 9.1 | 0.38 | 1.29 | 0.79 |
| 3 | 90 | 62.38 | 35.5 | 13.1 | 0.37 | 1.85 | 1.49 |
| 3 | 100 | 64.74 | 39.3 | 14.4 | 0.37 | 2.04 | 2.19 |
| 7 | 50 | 36.39 | 29.3 | 13.6 | 0.46 | 0.35 | 5.69 |
| 7 | 70 | 50.47 | 42 | 21.5 | 0.51 | 0.55 | 9.38 |
| 7 | 90 | 64.55 | 60.5 | 31.5 | 0.52 | 0.82 | 11.18 |
| 7 | 100 | 68.08 | 65.2 | 32.9 | 0.50 | 0.85 | 13.99 |
| 14 | 30 | 19.95 | 55 | 32 | 0.58 | 0.21 | 4.03 |
| 14 | 50 | 29.34 | 84 | 54 | 0.64 | 0.35 | 7.27 |
| 14 | 70 | 39.91 | 121 | 80.8 | 0.67 | 0.53 | 9.97 |
| 14 | 90 | 53.99 | 187 | 124.5 | 0.67 | 0.81 | 17.51 |
| 14 | 100 | 55.16 | 202 | 138.5 | 0.69 | 0.90 | 19.42 |
| 40 | 30 | 2.35 | 43 | 21.5 | 0.50 | 0.017 | 20.70 |
| 40 | 50 | 2.35 | 58.9 | 28 | 0.48 | 0.022 | 34.91 |
| 40 | 70 | 4.70 | 70 | 36.7 | 0.52 | 0.029 | 43.91 |
| 40 | 90 | 5.88 | 95 | 56.2 | 0.59 | 0.045 | 47.83 |
| 40 | 100 | 5.88 | 103 | 59.3 | 0.58 | 0.047 | 56.57 |
2.2. Chemicals used and analytical techniques employed
The concentration of radicals, produced in an aqueous solution, can be measured with dosimetry techniques. These rely on the reaction between an initial reactant (solute) as a scavenger for the radicals formed during the cavitation activity. In other words, the solute, by partially reacting with radicals, is converted into detectable products. A lot of focus is placed on hydroxyl radicals as these have the second highest oxidation potential (=2.8 eV), only inferior to that of Fluorine. In this work, the so-called coumarin dosimetry was used to quantify the hydroxyl radicals [48], [49], [50], [51]. In the presence of radicals generated by cavitation activity, coumarin molecules react to form different species such as 3-hydroxycoumarin, 4-hydroxycoumarin, 5-hydroxycoumarin, 6-hydroxycoumarin, 7-hydroxycoumarin, and 8-hydroxycoumarin and further reaction of these lead to formation of di-hydroxycoumarin molecules as well [52], [53]. The reaction chemistry of coumarin was not of primary interest in this study, rather the interest was to employ coumarin as a “radical trap” to quantify cavitation effects. Other methods such as the Weissler reaction [54] and dosimetry with salicylic and terephtalic acids are available, however, they have their limitation either in terms of not being specific to the cavitation activity or the need for operating at an altered pH of the system respectively [36], [37], [55].
It is important to clarify that with the proposed dosimetry technique, just a small part of the total hydroxyl radicals, produced into the system, is detected. Indeed, a consistent part of radicals, after being generated, are reabsorbed by recombination reactions. This is an intrinsic drawback of every dosimetry technique: it is impossible to measure all the radicals produced in the system since they are extremely reactive and ephemeral [36], [37].
Furthermore, some radicals react with the chemical probe by taking alternative reaction pathways and forming products which are not easy to monitor [52], [53]. Regarding this aspect, some authors have attempted to estimate the total hydroxyl radicals. For example, in the context of coumarin dosimetry, Zhang et al. [56] estimated that just 6.1% of all the OH radicals produced are scavenged as 7-hydroxycoumarin (7OHC). However, as suggested by De-Nasri et al. [48], this approximation may not be valid and consistent for other advanced oxidation processes, since with the probe based methods of OH quantification, the hydroxylated product yields are process specific.
Considering the complexity of the chemical problem, the present study does not aim at a deeper investigation of all the different chemical pathways of the radicals generated. Coumarin dosimetry is here used just as a method to characterize the sonochemical reactor, and to estimate the relative chemical efficiency of different experimental configurations. For these reasons, various chemical parameters such as operating pH or initial temperature have not been investigated.
The assumption made is that the amount of OH estimated, by monitoring the fluorescent product (7OHC), is representative of the chemical efficiency of the configurations tested. The same assumption has been used in previous cavitation-based investigations [48], [49], [50].
The 7OHC, in an aqueous solution, can be detected by fluorescence spectroscopy, since the fluorescence spectrum responds to the 7OHC (emission wavelength: 332 nm, excitation wavelength: 450 nm). For a quantitative estimation, a calibration curve was prepared across a range of anticipated 7-hydroxycoumarin concentrations. This curve, where the concentration of 7OHC is a function of the intensity of the fluorescence signal, was used to identify the concentration of the compound in samples collected from the experiments.
The chemicals used for the experimental campaign, coumarin ( 99%, CAS number: 91–64-5) and 7-hydroxycoumarin (99%, CAS number: 93–35-6), were procured from Sigma-Aldrich. A stock solution of 1 g/L of coumarin in DI water was prepared to ensure complete coumarin dissolution. Furthermore, a stock solution of 50 mg/L of 7-hydroxycoumarin was created using DI water, for the preparation of the samples, used to generate the calibration curve. The 7OHC compound was monitored on the fluorometer (Cary Eclipse Fluorescence Spectrometer), with an excitation wavelength of 332 nm and an emission wavelength of 450 nm.
To be consistent with other works present in literature [49], [50], for all the experiments, the initial concentration of coumarin was kept at 15 mg/L. This low concentration was also adopted to not affect the cavitation activity [57]. This validates the assumption that the cavitation field in pure water is representative of the cavitation activity during the coumarin experiments. The dosimetry experiments were carried out for a maximum of 28 min, without actively cooling down the system. To minimize the evaporation of the processed solution, the headspace between the ultrasonic probe and the beaker was covered using an aluminum foil. The evaporation was estimated with a mass balance at the end of every experimental run. Based on the amount of water evaporated, the measured concentrations were corrected. All the experiments have been carried out with an initial temperature of the solution around 23 ±1 °C. Over the processing time, at every 4 min, a 4 mL volume was sampled from the reactor by using a calibrated pipette. The collected samples were analyzed with the fluorometer to quantify the concentration. The uncertainty of the measurements was estimated by repeating the fluorometer analysis three times on the same sample. It was calculated to be within ±2%.
2.3. High-speed imaging
The cavitation vapor field, induced by ultrasonic waves at 24 kHz, can be observed with a high-speed camera in a backlighting setup, where the scene is placed between the light source and the camera. Due to the refraction of light at the interface between the phases, it is possible to visualize shadows of objects and the spatial and temporal evolution of the vapor field within the domain [58], [59]. A schematic illustration of the setup employed is shown in a previous study of the authors [60]. A LED light source (Godox SL-200 W II), emitting white light non-coherently, was placed on the back side of a water container and aligned with the axis of the probe. The scene was captured using the high-speed camera Phantom v2511 placed in front of the container, at the opposite side of the light source. A frame rate of 200,000 fps (200 kHz) at 256 × 256 pixels was used to capture the phenomenon. The camera was equipped with a 105 mm Nikon MICRO Lens, providing a resolution of 0.055 mm/pixel. The focal plane was placed on the axis of the vibrating probe, and the maximum aperture of the lens was used (f/2.8). This allows us to have a narrow depth of field, having in focus just the vapor cavities in the neighborhood of the plane of focus.
A large aperture also reduces the exposure time required, as it increases the light reaching the sensor. An exposure time of 1.5 µs was set. For the visualization experiments, a cuboidal container, made of transparent acrylic and having dimensions of 95 mm × 190 mm × 190 mm, was partially filled with Milli-Q® water at 23 °C. The depth of water in the container was 70 mm, and the tip of the probe was immersed 25 mm below the free surface.
The frames, obtained with high-speed visualization, for some of the experimental configurations, are presented in Fig. 1. The shadow of the cylindrical probe and the vapor cavities are defined by the dark areas of the image, while the light region represents the liquid domain (Fig. 1). The cavitation field presents a complex morphology of vapor structures (Fig. 1). In this regard, the vapor field generated by the 3, 7 and 14 mm probes, can be studied by identifying two different regions: the first one, in the near-field, where a large vapor cluster is attached to the probe tip, the second one, in the far-field, characterized by a cloud of smaller bubbles. On the contrary, in the configuration with the 40 mm probe, there is no evidence of the vapor cluster attached to the tip (Fig. 1d).
Fig. 1.
High-speed sequences for different ultrasonic probes (Videos in the SI). (a-d) are individually labeled by their probe diameter, nominal amplitude and input calorimetric power (definition in Eq. (3)). Time between frames is the same for (a-d) as marked.
Quantitative data, on the vapor phase present in the domain, can be extracted by binarizing the original grayscale images (Fig. 1). A similar approach, for the study of cavitation induced by ultrasonic probes, has previously been used by other authors [58], [59].
The binarization procedure allows for a clear distinction between the phases, with the liquid represented in white (pixel value = 1) and the vapor in black (pixel value = 0). Because of the structural complexity of the vapor field, the image binarization was carried out with a hybrid strategy. This is summarized, for a single frame, in Fig. 2.
Fig. 2.
Image process for a 7 mm probe with a nominal amplitude of 70%: (a) Cropped image (grayscale frame from the recorded video); (b) binarized image using the adaptive thresholding method; (c) binarized image using Otsu’s method; (d) final binary image obtained using a hybrid procedure of the two methods; (e) total vapor amount over time from the high-speed image sequence. The unit of measure of the -axis is mm2 since the backlighting visualization provides a 2D representation of the vapor phase.
Two different functions available on Matlab [61] were used: the adaptive thresholding method and the Otsu’s method [62], [63]. The first one provides a good binary representation of the vapor cloud in the far field, but it fails in the binarization of the vapor cluster attached to the tip of the probe (Fig. 2b). The second one is able to preserve the shape of the cluster in the near-field, but fails with the vapor cloud in the far-field (Fig. 2c).
For both methods, the sensitivity parameters have been calibrated to preserve the shape and the boundaries of the vapor structures, visible in the original grayscale images (0.007 for Otsu’s method, and 0.40 for the adaptive thresholding method). The choice of slightly different sensitivity parameters would imply a non-identical binary image. However, the linearity of dependency arising (Fig. 6, 7) and the global validity of the methodology will still be preserved.
Fig. 6.
Average vapor amount over the calorimetric power (the vapor amount was estimated by binarizing the raw high-speed images). The dashed lines are the linear regressions of the scatter data.
Fig. 7.
Initial reaction rate versus the average vapor amount. (The reaction rate was obtained by deriving the time evolution of the 7OHC concentration, Fig. 3). The dashed lines represent the linear regressions of the scatter data.
After obtaining the binary representation (Fig. 2d) of the high-speed sequence, the total amount of vapor (), as a function of time, was computed by multiplying the number of black pixels by the effective area of a pixel, known by optical calibration. The chart in Fig. 2e shows the vapor amount over time for one experimental case. The vertical solid line indicates the temporal average value (). The same procedure can be applied to extract other information from the binary image, like the time evolution of the area of the vapor cluster alone ().
The dynamics of the vapor structures were analyzed with the DFT (Discrete Fourier Transform) of the pixel intensity of the original high-speed image sequence (Fig. 1). The DFT can be done, through the FFT (Fast Fourier Transform) algorithm, on different portions of the image, depending on the dynamics of the structure of interest. The DFT of the image portion related to the far-field reveals one significant peak at 23.8 kHz. This means that the vapor bubbles, in that specific region, mainly oscillate at the same frequency as the ultrasound probe. Instead, the DFT of the near-field reveals, together with the peak at 23.8 kHz, the presence of other peaks in the low frequency range of the spectrum (Fig. 8). This suggests that the vapor cluster oscillates at frequencies lower than the excitation one. These outcomes are important to characterize the dynamic behavior of the system (Section 3.3).
Fig. 8.
DFT spectra of the pixel intensity related to the near-field of the frame sequence captured with high-speed visualization.
The same backlighting setup used for the visualization of the cavitation field (Fig. 1) was also used for measuring the actual vibration amplitude of the ultrasonic probes. This was crucial since the regulation system of the ultrasound device just provides the percentage amplitude (). Since the displacement is expected to be on the order of magnitude of micrometers, a high level of magnification is required. Thus, the Phantom v251 was combined with a Mitutoyo 10X objective lens to provide a resolution of 2.3 µm/pixel. The sequence of magnified frames, obtained for the 7 mm probe, is provided in Fig. 1 of the Supplementary Information (SI). The displacement amplitude can be visualized only by framing the corner or the lateral surface of the probe because the tip is always covered by vapor cavities for all working conditions. The displacement amplitude of the probe () was measured with a tracking procedure, implemented on Matlab [61]. Starting from the frames sequence, the displacement can be tracked by selecting one column of pixels and taking the difference between adjacent pixels. From the resulting array, the element with the maximum value defines the position of the probe tip at every time step. The procedure is described in more detail in a previous work of the authors [60]. The values obtained, for the different configurations tested, are shown in Table 1. The DFT, of the probe displacement over time, revealed a frequency of kHz (slightly different from the advertised value of the device of 24 kHz).
2.4. Measurement of acoustic spectra
The acoustic signal was recorded by inserting a hydrophone (Aquarian AS-1, receiving sensitivity: 40 V/Pa, linear range: 1 Hz to 100 kHz ±2 dB) in the liquid domain. This device works with a direct connection to an oscilloscope. For the signal acquisition, two different sampling frequencies were set: 12.5 and 31.25 MHz. Due to the oscilloscope setting, these frequencies correspond to two different time intervals acquired, respectively, 16 and 6.4 ms. The signals, measured by the hydrophone, have been used to further characterize the cavitation activity in the ultrasonic reactors by comparing the spectra related to the different configurations.
3. Results and discussion
The results of the 7OHC conversion are presented in Fig. 3 for different probe configurations and amplitudes. The concentration over time for the 14 mm probe was estimated for two different solution volumes: 200 and 400 mL (Fig. 3a, b). Furthermore, for these cases, a repeatability study was conducted (Run 1 and Run 2 in Fig. 3a, b). As can be observed, the data reveal a good repeatability of the experiments. The probes having diameters of 40 mm and 7 mm were tested only with the volume solution of 200 mL (Fig. 3c, d). No data related to the 30% amplitude for the 7 mm probe, and to the 3 mm probe are shown, due to the poor energy input and the consequent low chemical conversion for these configurations (below the detection limit of the fluorometer).
Fig. 3.
Concentration profile of 7-hydroxycoumarin for the experimental configurations tested as titled in (a-d). The different marker symbols indicate experiments done with different nominal amplitudes. The specific electric power (W/g) was measured with a power meter. The uncertainty of the measurements (±2%) was estimated by repeating the fluorometer analysis three times on the same sample.
An alternative experiment was conducted to prove the reliability of the coumarin dosimetry method in producing hydroxylated products only in the presence of cavitation activity. The same initial solution, used for the parametric study, was heated by using a heating plate, equipped with a mechanical stirrer. A thermal power of 330 W was provided to the system and the experiment was stopped at 14 min, at a temperature of 95 °C. The fluorometer analysis of the sample has revealed no coumarin conversion into 7OHC. This result shows no oxidation in the presence of a heat source and atmospheric air, confirming the coumarin dosimetry as an unambiguous method for the OH radicals produced by cavitation activity.
For every configuration tested, the concentration of 7OHC increases with time and with the vibration amplitude of the ultrasonic probe (Fig. 3). This is expected since an increase in the amplitude leads to an increase in the energy introduced into the system, resulting in an increase in energy dissipated in the formation of radicals.
It is interesting to note that some configurations, in Fig. 3d, exhibit a linear trend in time. This is probably due to the dependency between the power dissipated by the probe and the liquid temperature. When there is a low increment of the liquid temperature (Fig. 2 in SI), the variation of the physical property (acoustic impedance and viscosity) is small. As a consequence, the power dissipated by the probe and that absorbed by the device remain almost constant (Fig. 3 in SI) during the entire duration of the process. This implies a constant rate of radicals production through time (Fig. 3). On the contrary, in the configurations characterized by a large increase in temperature (Fig. 2 in SI), the significant variation of the liquid properties impacts the power dissipated by the probe. This entails a decrease, with the processing time, in the electrical power (Fig. 3 in SI), and the production rate of radicals (Fig. 3).
Since the energy contributions (energy absorbed by the ultrasound device and energy introduced into the system) are different for every configuration, the results cannot be compared in the form shown in Fig. 3.
For this reason, to compare the conversion data related to different configurations, an energy balance of the system is required.
3.1. Energy balance and comparison
The energy introduced in the system, through the work done by the ultrasonic probe (), is partly spent to increase the internal energy of the system () and partly exchanged, as heat, towards the environment (). Under the assumption of ideal incompressible behavior [64], the energy balance can be written in the following form:
| (1) |
where is the mass of the liquid, is the specific heat capacity of the liquid medium (in this case water), is the temperature of the system (Fig. 4a), and is the initial temperature. Formally, also the energy stored by the chemical reactions should be taken into account. However, considering the low concentrations of products that characterize these type of sonochemical reactors (Fig. 3), this contribution can be reasonably neglected in the energy balance. Only during the first few minutes of the process, a further approximation can be made by by considering the system to be adiabatic. Indeed, due to the small volume of the liquid domain, the heat exchanged by the system towards the environment can be neglected. This assumption is confirmed by the linear trend shown by the temperature curve in the first few minutes (dashed line in Fig. 4a). Under this assumption, the energy balance simplifies and the effective power, provided by the ultrasonic probe (), can be expressed as:
| (2) |
where is the rate of change of the internal energy, and is the time derivative of the temperature. Since the temperature was monitored with a thermocouple inserted in the solution (Fig. 4a), Eq. (3) allows us to estimate , at the beginning of the experiment, for every configuration (Table 1). The procedure mentioned constitutes the calorimetric methods (), usually employed to characterize sonochemical reactors [39], [40], [41]. Considering the acoustic power provided by the probe equal to the calorimetric power is a reasonable assumption since most of the energy is dissipated as heat into the system.
Fig. 4.
Temperature (a) and energy profile (b) for the configuration with the 14 mm probe diameter and a volume processed of 400 mL. (a) Evolution in time of the solution temperature, measured with a thermocouple, for three different nominal amplitudes. (b) Evolution in time of the electric energy used by the ultrasound device and the internal energy of the system. The difference between the two contributions is due to the efficiency of the device and the heat exchanged.
The efficiency of the device (), at the beginning of the run, can be computed:
| (3) |
where is the electrical power absorbed by the ultrasound device and measured with a digital power meter (Table 1). The device efficiency, for every configuration tested, is listed in Table 1.
In the following analysis, to normalize the concentration of 7OHC, two different energy contributions are considered: the specific variation of the internal energy () and the electrical energy, absorbed by the device, in specific terms (). The trend over time of these two contributions is shown in Fig. 4b. To obtain a dimensionless parameter, the concentration of 7OHC () can be multiplied by the bond dissociation energy of OH (=497.1 kJ/mol) and divided by . In this way, a sort of chemical efficiency of the system () can be computed:
| (4) |
Since just part of the radicals generated is measured and the enthalpy of formation of 7OHC was not considered, the previous definition (Eq. (4)) is formally incorrect. However, it provides an estimation of the portion of energy that is effectively spent for generating the radicals with respect to the global energy that remains in the system (). For this reason, Eq. (4) allows us to evaluate the relative efficiency of the different configurations tested. Another non-dimensional parameter can be introduced to define a sort of total efficiency of the system:
| (5) |
This provides an indication of the energy used for the generation of radicals, compared to the total electric energy absorbed by the ultrasound device. The two dimensionless parameters (Eq. (4), (5)) can be used for a consistent comparison of the results provided by the different configurations (Fig. 5).
Fig. 5.
(a) Influence of processing volume on total efficiency over time for the 14 mm probe. (b) Influence of processing volume on chemical efficiency for the 14 mm probe [symbol legends are identical to chart (a)]. (c) Total efficiency for different probe diameters. (d) Chemical efficiency for different probe diameters [symbol legends are identical to chart (c)].
It is noteworthy that the trend of the chemical efficiency (Fig. 5b, d) is almost constant with the process time. This means that the coumarin conversion is just proportional to the energy dissipated by the probe and that the bulk temperature trend (different for every configuration, Fig. 2 in the SI) does not significantly modify the reaction rate and the availability of OH* radicals into the system (consequence of the cavitation activity). Charts (a) and (b) are respectively related to the parameters and as a function of the processing time, for the configurations with the 14 mm probe and two different process volumes (200 and 400 mL). As observed, for both cases, the variation of amplitude and the process volume does not significantly impact the system efficiency. More interesting findings can be made by comparing the results coming from experiments with fixed process volume (200 mL) and different probe diameters (7, 14, 40 mm) (Fig. 5c, d). For both parameters, (Fig. 5c) and (Fig. 5d), only the configuration with a 40 mm probe shows sensitivity with respect to the displacement amplitude. Indeed, the efficiency increases as the amplitude increases. Instead, the efficiency for the 7 and 14 mm probes is almost insensitive to the amplitude. In global terms, for both parameters, the 40 mm probe shows a significantly greater efficiency than the 7 and 14 mm ones (Fig. 5c, d). The explanation for this is to be found in the different structures of the vapor field generated by the probes (Fig. 1). The detailed analysis of the vapor field and its connection with the efficiency is presented in the following sections.
3.2. Vapor field, structure and dynamics
The analysis of the cavitation vapor field was conducted as described in Section 2.3. To prove the consistency of the proposed methodology, the dependency between acoustic power, vapor amount, and reaction rate was studied. In Fig. 6, the average vapor amount is plotted against the calorimetric power for the different probe diameters. The regressions and the coefficients of determination (Fig. 6) show reasonably linear trends, with the vapor amount that tends to go to zero when the power goes to zero. For a given power, the 40 mm configuration seems to generate a higher amount of vapor compared to others. However, a direct comparison between the rates obtained from the different diameters could be inaccurate in this case. This is due to the slightly different optical setups, used for the visualization experiments, that could have affected the outcome of the binarization procedure (Fig. 2). Consistent trends were also obtained by plotting the reaction rate versus the average vapor amount (Fig. 7). The regressions suggest a linear trend in the investigated region, with the reaction rate tending to zero when the vapor amount reaches zero.
It is important to elucidate that, in the present case, the system reactivity is evaluated based on the hydroxylation of the coumarin molecules. The trends obtained were expected since the vapor production in the domain is a consequence of the acoustic energy introduced, and the reactivity is due to the presence of the vapor cavities, which oscillate and collapse, generating radicals.
Together with the energy dissipation induced by the violent collapse of the cavities, also the mass transfer and mixing, promoted by cavitation through different phenomena (micro-jets, shock waves, intense pressure waves), play a crucial role in the reactivity of the system. In the present analysis, all these effects are considered in evaluating the chemical efficiency for a given configuration (reaction rate in Fig. 7).
An alternative description of the cavitation cloud has been recently proposed by Dular and Petkovšek, with the formulation of an acoustic cavitation parameter [59], [65].
With the total amount of vapor in the domain, the structure and the dynamics of the vapor field are crucial to study the reactivity of the system. As observable in Fig. 1, the structure of the vapor phase significantly changes from one configuration to another. The field generated by the 3, 7, and 14 mm probe configurations is characterized by the presence of a big cluster of vapor attached to the tip of the probe. This peculiar phenomenon was first observed and studied by Žnidarčič et al. and termed “acoustic supercavitation” [58], [66]. On the contrary, for the 40 mm configuration, there is no evidence of such a cluster, and the liquid is covered, more uniformly, by dispersed vapor bubbles. These cavitation phenomena belong to a non-linear regime [67]. The non-linearity is due to the large oscillation amplitudes of the acoustic pressure (high-intensity ultrasound), which induce changes in the non-spherical shape of the vapor cluster, large variations of the cluster radius during each cycle, and a periodic collapse of the vapor structures (“transient acoustic cavitation”) [68], [69].
To correlate the reactivity of a specific configuration with the particular structure of its vapor field, a deeper investigation of the nature of the vapor cluster (Fig. 1) is required. Starting from the time average of the vapor cluster area (), the equivalent radius can be computed using the formula: . For a given probe configuration, can be estimated for different vibration amplitudes. Fig. 8a-c shows the equivalent radius for different 7 mm probe configurations and the frequency spectra resulting from the DFT of the pixel intensity on the portion of the image that contains the vapor cluster (near-field). The spectra (Fig. 8a, b, c) exhibit, together with the peak at the principal frequency (vibration frequency of the probe: 23.8 kHz), peaks in the low-frequency region. These are not subharmonics of the principal frequency since they are not equal to 23.8/, with equal to a positive integer. The different peaks arising from the spectrum demonstrate that the vapor cluster oscillates with a combination of various frequencies. Furthermore, the frequency associated with the maximum peak, in the low-frequency region of the spectrum, decreases with the equivalent radius of the vapor cluster (Fig. 8a, b, c). This trend resembles the natural frequency of oscillation of a single bubble and could be linked with the collective and coherent motion exhibited by a cloud made of bubbles, characterized by a high void fraction and consequent strong interactions between the constitutive bubbles [67], [70], [71], [71], [72], [73]. The different oscillation frequency between the vapor cluster and the smaller bubble, dispersed in the far-field in the liquid domain, is also evident from the high-speed videos (Videos 1, 2, 3 in the SI).
For the sake of comparison and description of the phenomenon, the cluster frequencies (maximum peak in the low-frequency region), arising form the DFT spectrum of the experimental configurations are plotted in Fig. 9, together with the theoretical Minnaert frequency as a function of the equivalent radius (). The qualitative trend exhibited by the data, related to the vapor clusters generated by the different probe diameters, is similar to that provided by the Minnaert solution. The deviation of the experimental data compared to the ideal case (Minnaert frequency in Fig. 9) is due to the non-spherical shape of the vapor cluster and to the non-linear effects that influence the phenomena in high-intensity ultrasound [68], [69]. These results prove that the vapor cluster oscillates at its natural frequency, behaving like one single entity, like a unitary macro-bubble.
Fig. 9.
Frequency versus equivalent radius - theoretical (Minnaert resonance) and calculated (vapor cluster attached to the 3, 7, and 14 mm probes). The dashed curves represent the scatter data fitting in the form of: .
On the contrary, the DFT spectrum of the pixel intensity confirms the absence of the cluster for the 40 mm probe since no relevant peaks are shown in the low-frequency range (Fig. 8d).
The presence of cavitation, on the surface of the vibrating probe, can also be verified by comparing the calorimetric power with the ideal acoustic power dissipated by the probe vibrating in liquid water (Section B in the SI). The value of the calorimetric power is much lower than the ideal one (Fig. 4 in the SI). This is due to the reduction of the acoustic impedance () “felt” by the probe, induced by the presence of the vapor phase on the emitting surface.
3.3. Relation between working regimes and chemical efficiency
The differences in the structure and dynamics of the vapor field are likely to influence the chemical efficiency of sonochemical systems. In this regard, the relevant physical quantity to define two different working regimes is the acoustic intensity at the tip of the probe: , where is the surface area of the probe tip. This quantity is listed in Table 1 for every experimental configuration. A high acoustic intensity characterizes the 3, 7, and 14 mm probes. For these configurations, the high concentration of acoustic energy, near the tip of the probe, leads to the formation of the vapor cluster. After that, the acoustic energy emitted is mainly spent to sustain the oscillation of the vapor cluster, which absorbs all the radiated energy. This does not allow a uniform and deep propagation of the acoustic energy in the liquid domain, leading to a chemically inefficient system (Fig. 5).
From the high-speed videos of these configurations (SI), we observe that the small bubbles, present in the far-field, “escape” from the vapor cluster and are dragged down by the secondary Bjerknes force [74]. On the other side, the chemically efficient configurations (40 mm probe) present an acoustic intensity one order of magnitude lower (Table 1). This leads to a completely different phenomenon, where the acoustic energy density is not sufficient for the nucleation of the cluster attached to the tip. The absence of the cluster allows a more uniform and deep propagation of the acoustic waves through the domain. This triggers the nucleation of small vapor bubbles and pockets in a wider region of the liquid (Fig. 1d). For the 40 mm probe, the high-speed video (Video 4 in the SI) shows that the bubbles oscillate in place at the principal frequency (23.8 kHz) and that the primary Bjerknes force is not strong enough to drag them downward.
The differences in the structure of the vapor field, between the two regimes, are also visible in the histogram of the binary image (Fig. 10). For the chemically inefficient regime (7 and 14 mm probe), the histogram shows a population of vapor bubbles having a small area (0–0.2 mm range in Fig. 10a, b). This represents the bubble cloud present in the far-field (Fig. 10a, b). The vapor cluster entity, characterized by a large area, is outside the chart frame in Fig. 10a, b. Instead, the histogram, related to the 40 mm probe (Fig. 10c), shows a population of bubbles distributed over a wide range of values of the area.
Fig. 10.
Histogram of the binary image for different probe diameters. In the description, between the brackets, the calorimetric power appears.
In summary, the data of the coumarin conversion suggest that the largest probe diameter gave the highest efficiency in forming hydroxyl radicals (Fig. 5). This is due to the particular structure of the vapor field, which is composed of a population of small vapor bubbles and cavities. As known from the literature, the production of radicals increases as the radius of the bubble decreases [75], [76], thanks to a higher energy density dissipated during the collapse [67]. This explains the ineffectiveness in generating of radicals of the big vapor cluster attached to the probe tip in the chemically inefficient configurations (3, 7, 14 mm).
Additional information can be obtained by processing the acoustic signals measured by a hydrophone, immersed in the liquid domain. The frequency spectrum of an acoustic signal contains information related to the cavitation phenomena. Usually, the tonal component of the spectrum is associated with driving acoustic waves and non-inertial cavitation. At the same time, the broadband noise is a signature of inertial or transient cavitation (bubbles collapse) [77]. Since the radicals are generated only when cavitation is inertial, the energy associated with the broadband noise is expected to be significantly larger for the 40 mm configuration than the others. Experiments were carried out by keeping a similar calorimetric power between the different probe configurations (Table 2) to obtain a consistent comparison.
Starting from the raw acoustic signal (Pa), registered by the hydrophone, the Power Spectral Density [PSD (Pa2/Hz)] was estimated with the periodogram method [61], [78], [79]. The PSD spectra obtained for the different configuration comparisons are visible in Fig. 11a, b, c. The average power (Pa2) of the acoustic signal can be computed by integrating the PSD over a given frequency band (from 0 kHz to 600 kHz) [61], [78], [79]. The experimental acquisitions and the integrations were repeated three times for every configuration (Table 2). The integral values, related to the three different acquisitions (Exp. 1–3 in Table 2), show poor repeatability. This could be due to the short time window, acquired through the oscilloscope (section 2.4), which affects the measurements. Furthermore, the average of the three measurements (Average in Table 2) does not scale consistently with the calorimetric power provided by ultrasonic probes. The discrepancy between the calorimetric power and the PSD integral suggests that the acoustic power received by the hydrophone is just a portion of the total power introduced into the system by the probe. This is a consequence of the presence, in the domain, of the vapor phase, which hinders the detection of the acoustic signal by the hydrophone. The acoustic waves, coming from different sources (probe vibration, bubble oscillation, and collapse), are reflected and scattered at the vapor/liquid interfaces. This induced the dissipation of a significant portion of the acoustic power that cannot be detected by the hydrophone. This hypothesis is corroborated by the probe comparisons in Table 2. With a similar calorimetry power, the PSD integral of the 40 mm configuration is significantly greater than that of the others (7 and 14 mm). This is due to the presence of the vapor cluster in the 7 and 14 mm configuration, which reflects and dissipates the sound waves coming from the ultrasonic probe. While less energy is dissipated in the 40 mm configuration, where the vapor cluster is absent. The aforementioned hydrophone detection issues do not make a direct quantitative comparison between the PSD spectra of the different probes possible. However, the hydrophone signals provide qualitative information about characteristic features that differentiate the experimental configurations. The PDS spectrum of the 7 and 14 mm configurations (Fig. 11a, b, c) exhibits peaks in the low-frequency range. These peaks, which are not present in the 40 mm configurations (Fig. 11a, b, c), represent the natural frequencies of the vapor cluster oscillation and confirm the result obtained with the DFT of the pixel intensity of the high-speed sequences (Fig. 9). These frequencies have also been identified by Kozmus et al. [59]. It is interesting to note that the 40 mm configuration, where the natural frequencies peak of the cluster natural frequencies are absent, instead exhibits a sub-harmonic peak at 12 kHz (Fig. 11a, b, c).
Fig. 11.
(a, b, c) Power spectral density (PSD) of the cavitation noise, measured by the hydrophone, for different probe diameters with similar calorimetric power. (d) Comparison of the frequency spectra from the hydrophone signal and pixel intensity of the high-speed sequence for the 7 mm probe. (e) Comparison of the frequency spectra from the hydrophone signal and pixel intensity of the high-speed sequence for the 14 mm probe.
In Fig. 11d, e, the spectra related to the hydrophone signal and the pixel intensity are directly compared for the 7 and 14 mm probes. The two different analyses are consistent since the natural frequencies peaks fall in the same range.
A more rigorous investigation of the noise spectra could be conducted by using the recent standard from the International Electrotechnical Commission: IEC TS 63001:2024 RLV (Measurement of cavitation noise in ultrasonic baths and ultrasonic reactors) [80]. This was not implemented in the current study since it is not part of the primary objectives.
Table 2.
Integral of the PSD for experimental configurations, characterized by different probe diameters and calorimetric powers (limits of integration: kHz, kHz).
| Exp.1 | Exp.2 | Exp.3 | Average | |||
|---|---|---|---|---|---|---|
| [mm] | [%] | [W] | [Pa2 ] | [Pa2 ] | [Pa2 ] | [Pa2 ] |
| 7 | 100 | 35.8 | 3.70 | 3.70 | 3.00 | 3.47 |
| 14 | 30 | 34.8 | 3.67 | 2.66 | 2.04 | 2.79 |
| 40 | 70 | 35.1 | 36.48 | 36.48 | 13.30 | 28.75 |
| 14 | 50 | 57.0 | 6.77 | 14.93 | 13.05 | 11.58 |
| 40 | 100 | 57.0 | 35.04 | 3.38 | 4.65 | 14.36 |
| 7 | 70 | 21.4 | 1.90 | 1.99 | 2.44 | 2.11 |
| 40 | 30 | 25.2 | 22.59 | 40.92 | 45.06 | 36.19 |
4. Conclusions
This study investigates the relationship between the production of hydroxyl radicals and the vapor field resulting from ultrasonically induced cavitation. The method proposed is based on coumarin dosimetry for estimating the hydroxyl radicals and high-speed imaging for the visualization of the cavitation field. The study sheds light on the dependency between the chemical efficiency and the dynamics of vapor field structures. The probe diameter and the acoustic intensity play a crucial role in defining the energy dissipation and the chemical efficiency of the system. Between the four configurations tested, it was observed that the probe with the largest tip (40 mm) produces the most intense chemical activity for a given input power. It was also demonstrated that the rate of radicals formation is a function of the acoustic power introduced in the domain by the ultrasonic device and linearly scales with the vapor amount. The presented approach in this work, can be used to characterize and optimize ultrasonic reactors working at different frequencies and for various applications involving physical or chemical treatments.
Nomenclature and acronyms
- SL
sonoluminescence
- SCL
sonochemiluminescence
- OH
hydroxyl radical
- OOH
hydroperoxyl radical
oxidation potential, [eV]
- 7OHC
7-hydroxycoumarin
- DFT
Discrete Fourier Transform
probe diameter, [mm]
processed volume, [mL]
percentage probe displacement amplitude, [%]
actual probe displacement amplitude, []
time average of the vapor amount, [mm]
time average of the vapor cluster, [mm]
actual probe vibration frequency, [kHz]
energy introduced into the system by the ultrasonic probe, [J]
variation of the internal energy, [J]
heat exchanged towards the environment, [J]
mass of liquid, [g]
specific heat capacity, [Jkg K−1]
solution temperature, [K]
solution initial temperature, [K]
power introduced into the system by the ultrasonic probe, [W]
rate of change of the internal energy, [W]
time derivative of the temperature, [Ks−1]
calorimetric power, [W]
electrical power absorbed by the ultrasound, [W]
efficiency of the ultrasound device, [-]
specific variation of the internal energy, [J/kg]
specific energy absorbed by the device, [J/kg]
concentration of 7-hydroxycoumarin, [mol/kg]
bond dissociation energy of OH, [kJ/mol]
chemical efficiency of the system, [-]
total efficiency of the system, [-]
equivalent radius of the vapor cluster, [mm]
theoretical Minnaert resonance, [kHz]
medium density, [kg/m]
speed of sound in the medium, [m/s]
surface area of the probe tip, [mm]
acoustic intensity, [W/mm]
acoustic impedance, [kgm s−1]
Power Spectral Density, [Pa2/Hz]
CRediT authorship contribution statement
Gianmaria Viciconte: Writing – review & editing, Writing – original draft, Visualization, Validation, Supervision, Methodology, Investigation, Formal analysis, Data curation, Conceptualization. Varaha P. Sarvothaman: Writing – review & editing, Validation, Supervision, Methodology, Investigation, Conceptualization. Paolo Guida: Writing – review & editing, Validation, Supervision, Methodology, Funding acquisition. Tadd T. Truscott: Writing – review & editing, Visualization, Validation, Supervision, Resources. William L. Roberts: Writing – review & editing, Validation, Supervision, Resources, Project administration, Methodology, Funding acquisition.
Declaration of competing interest
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Acknowledgments
The authors would like to thank the KAUST Core Labs for the use of their fluorescence spectroscometer.
Footnotes
Supplementary material related to this article can be found online at https://doi.org/10.1016/j.ultsonch.2025.107362.
Contributor Information
Gianmaria Viciconte, Email: gianmaria.viciconte@kaust.edu.sa.
Varaha P. Sarvothaman, Email: varahaprasad.sarvothaman@kaust.edu.sa.
Appendix A. Supplementary data
The following is the Supplementary material related to this article.
Supplementary information contains videos of different probe diameters used in this work.
.
.
.
.
Data availability
Data will be made available upon request.
References
- 1.Richards W.T., Loomis A.L. The chemical effects of high frequency sound waves I. A preliminary survey. J. Am. Chem. Soc. 1927;49(12):3086–3100. [Google Scholar]
- 2.Chen J., Kalback W.M. Effect of ultrasound on chemical reaction rate. Ind. Eng. Chem. Fundam. 1967;6(2):175–178. [Google Scholar]
- 3.Okada K., Fuseya S., Nishimura Y., Matsubara M. Effect of ultrasound on micromixing. Chem. Eng. Sci. 1972;27(3):529–535. [Google Scholar]
- 4.Ashokkumar M., Grieser F. Ultrasound assisted chemical processes. Rev. Chem. Eng. 1999;15(1):41–83. [Google Scholar]
- 5.Suslick K.S., Price G.J. Applications of ultrasound to materials chemistry. Annu. Rev. Mater. Sci. 1999;29(1):295–326. [Google Scholar]
- 6.Mason T. Industrial sonochemistry: potential and practicality. Ultrasonics. 1992;30(3):192–196. [Google Scholar]
- 7.Gogate P.R., Tatake P.A., Kanthale P.M., Pandit A.B. Mapping of sonochemical reactors: review, analysis, and experimental verification. AIChE J. 2002;48(7):1542–1560. [Google Scholar]
- 8.Gogate P.R., Pandit A.B. Sonophotocatalytic reactors for wastewater treatment: a critical review. AIChE J. 2004;50(5):1051–1079. [Google Scholar]
- 9.Gogate P.R. Cavitational reactors for process intensification of chemical processing applications: a critical review. Chem. Eng. Process.: Process. Intensif. 2008;47(4):515–527. [Google Scholar]
- 10.Asgharzadehahmadi S., Raman A.A.A., Parthasarathy R., Sajjadi B. Sonochemical reactors: Review on features, advantages and limitations. Renew. Sustain. Energy Rev. 2016;63:302–314. [Google Scholar]
- 11.Yao Y., Pan Y., Liu S. Power ultrasound and its applications: A state-of-the-art review. Ultrason. Sonochemistry. 2020;62 doi: 10.1016/j.ultsonch.2019.104722. [DOI] [PubMed] [Google Scholar]
- 12.Dähnke S., et al. Numerical calculation of pressure fields in sonochemical reactors-linear effects in homogeneous phase. Period. Polytech. Chem. Eng. 1997;41(1):41–55. [Google Scholar]
- 13.Laborde J.L., Hita A., Caltagirone J.P., Gerard A. Fluid dynamics phenomena induced by power ultrasounds. Ultrasonics. 2000;38(1–8):297–300. doi: 10.1016/s0041-624x(99)00124-9. [DOI] [PubMed] [Google Scholar]
- 14.Hussain M.N., Samad T., Daqaq M., Janajreh I. Numerical modelling of sonicated, continuous transesterification and evaluation of reaction kinetics for optimizing biodiesel reactor design. Int. J. Therm. Environ. Eng. 2016;11(1):79–86. [Google Scholar]
- 15.Johansson Ö., Pamidi T., Löfqvist T. Design of a high-intensity ultrasound reactor. 2017 IEEE International Ultrasonics Symposium; IUS; IEEE; 2017. pp. 1–4. [Google Scholar]
- 16.Yasui K. Numerical simulations for sonochemistry. Ultrason. Sonochemistry. 2021;78 doi: 10.1016/j.ultsonch.2021.105728. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Ferkous H., Hamdaoui O., Pétrier C. Sonochemical reactor characterization in the presence of cylindrical and conical reflectors. Ultrason. Sonochemistry. 2023;99 doi: 10.1016/j.ultsonch.2023.106556. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Sboev I., Lyubimova T., Rybkin K., Kuchinskiy M. A numerical investigation of the effect of boundary conditions on acoustic pressure distribution in a sonochemical reactor chamber. Fluid Dyn. Mater. Process. 2024;20(6) [Google Scholar]
- 19.Mason T.J., Cobley A., Graves J., Morgan D. New evidence for the inverse dependence of mechanical and chemical effects on the frequency of ultrasound. Ultrason. Sonochemistry. 2011;18(1):226–230. doi: 10.1016/j.ultsonch.2010.05.008. [DOI] [PubMed] [Google Scholar]
- 20.Abid M., Jabbar S., Wu T., Hashim M.M., Hu B., Lei S., Zhang X., Zeng X. Effect of ultrasound on different quality parameters of apple juice. Ultrason. Sonochemistry. 2013;20(5):1182–1187. doi: 10.1016/j.ultsonch.2013.02.010. [DOI] [PubMed] [Google Scholar]
- 21.Jambrak A.R., Mason T.J., Lelas V., Paniwnyk L., Herceg Z. Effect of ultrasound treatment on particle size and molecular weight of whey proteins. J. Food Eng. 2014;121:15–23. [Google Scholar]
- 22.Fu X., Belwal T., Cravotto G., Luo Z. Sono-physical and sono-chemical effects of ultrasound: Primary applications in extraction and freezing operations and influence on food components. Ultrason. Sonochemistry. 2020;60 doi: 10.1016/j.ultsonch.2019.104726. [DOI] [PubMed] [Google Scholar]
- 23.Suslick K.S. Sonochemistry. Science. 1990;247(4949):1439–1445. doi: 10.1126/science.247.4949.1439. [DOI] [PubMed] [Google Scholar]
- 24.Didenko Y.T., McNamara W.B., Suslick K.S. Hot spot conditions during cavitation in water. J. Am. Chem. Soc. 1999;121(24):5817–5818. [Google Scholar]
- 25.Wagterveld R., Boels L., Mayer M., Witkamp G.J. Visualization of acoustic cavitation effects on suspended calcite crystals. Ultrason. Sonochemistry. 2011;18(1):216–225. doi: 10.1016/j.ultsonch.2010.05.006. [DOI] [PubMed] [Google Scholar]
- 26.Canselier J., Delmas H., Wilhelm A., Abismail B. Ultrasound emulsification—an overview. J. Dispers. Sci. Technol. 2002;23(1–3):333–349. [Google Scholar]
- 27.Cucheval A., Chow R. A study on the emulsification of oil by power ultrasound. Ultrason. Sonochemistry. 2008;15(5):916–920. doi: 10.1016/j.ultsonch.2008.02.004. [DOI] [PubMed] [Google Scholar]
- 28.Sander J.R., Zeiger B.W., Suslick K.S. Sonocrystallization and sonofragmentation. Ultrason. Sonochemistry. 2014;21(6):1908–1915. doi: 10.1016/j.ultsonch.2014.02.005. [DOI] [PubMed] [Google Scholar]
- 29.Kim H.N., Suslick K.S. The effects of ultrasound on crystals: Sonocrystallization and sonofragmentation. Crystals. 2018;8(7):280. [Google Scholar]
- 30.Luque-Garcıa J., De Castro M.L. Ultrasound: a powerful tool for leaching. TRAC Trends Anal. Chem. 2003;22(1):41–47. [Google Scholar]
- 31.Narayana K., Swamy K., Rao K.S., Murty J. Leaching of metals from ores with ultrasound. Miner. Process. Extr. Metullargy Rev. 1997;16(4):239–259. [Google Scholar]
- 32.Boček Ž., Petkovšek M., Clark S.J., Fezzaa K., Dular M. Dynamics of oil–water interface at the beginning of the ultrasonic emulsification process. Ultrason. Sonochemistry. 2023;101 doi: 10.1016/j.ultsonch.2023.106657. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Perdih T.S., Zupanc M., Dular M. Revision of the mechanisms behind oil-water (o/w) emulsion preparation by ultrasound and cavitation. Ultrason. Sonochemistry. 2019;51:298–304. doi: 10.1016/j.ultsonch.2018.10.003. [DOI] [PubMed] [Google Scholar]
- 34.Suslick K.S., Mdleleni M.M., Ries J.T. Chemistry induced by hydrodynamic cavitation. J. Am. Chem. Soc. 1997;119(39):9303–9304. [Google Scholar]
- 35.Sarvothaman V.P., Velisoju V.K., Subburaj J., Panithasan M.S., Kulkarni S.R., Castaño P., Turner J., Guida P., Roberts W.L., Nagarajan S. Is cavitation a truly sensible choice for intensifying photocatalytic oxidation processes?–Implications on phenol degradation using ZnO photocatalysts. Ultrason. Sonochemistry. 2023;99 doi: 10.1016/j.ultsonch.2023.106548. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Iida Y., Yasui K., Tuziuti T., Sivakumar M. Sonochemistry and its dosimetry. Microchem. J. 2005;80(2):159–164. [Google Scholar]
- 37.Martínez-Tarifa A., Arrojo S., Ávila-Marín A., Pérez-Jiménez J., Sáez V., Ruiz-Lorenzo M. Salicylic acid dosimetry applied for the statistical determination of significant parameters in a sonochemical reactor. Chem. Eng. J. 2010;157(2–3):420–426. [Google Scholar]
- 38.Rivas D.F., Ashokkumar M., Leong T., Yasui K., Tuziuti T., Kentish S., Lohse D., Gardeniers H.J. Sonoluminescence and sonochemiluminescence from a microreactor. Ultrason. Sonochemistry. 2012;19(6):1252–1259. doi: 10.1016/j.ultsonch.2012.04.008. [DOI] [PubMed] [Google Scholar]
- 39.Bampouli A., Goris Q., Van Olmen J., Solmaz S., Hussain M.N., Stefanidis G.D., Van Gerven T. Understanding the ultrasound field of high viscosity mixtures: Experimental and numerical investigation of a lab scale batch reactor. Ultrason. Sonochemistry. 2023;97 doi: 10.1016/j.ultsonch.2023.106444. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Bampouli A., Goris Q., Hussain M.N., Louisnard O., Stefanidis G.D., Van Gerven T. Importance of design and operating parameters in a sonication system for viscous solutions: effects of input power, horn tip diameter and reactor capacity. Chem. Eng. Process.-Process. Intensif. 2024;198 [Google Scholar]
- 41.Pflieger R., Chave T., Vite G., Jouve L., Nikitenko S.I. Effect of operational conditions on sonoluminescence and kinetics of H2O2 formation during the sonolysis of water in the presence of Ar/O2 gas mixture. Ultrason. Sonochemistry. 2015;26:169–175. doi: 10.1016/j.ultsonch.2015.02.005. [DOI] [PubMed] [Google Scholar]
- 42.Adamou P., Harkou E., Villa A., Constantinou A., Dimitratos N. Ultrasonic reactor set-ups and applications: A review. Ultrason. Sonochemistry. 2024 doi: 10.1016/j.ultsonch.2024.106925. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Zhu X., Das R.S., Bhavya M.L., Garcia-Vaquero M., Tiwari B.K. Acoustic cavitation for agri-food applications: Mechanism of action, design of new systems, challenges and strategies for scale-up. Ultrason. Sonochemistry. 2024 doi: 10.1016/j.ultsonch.2024.106850. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Lakshmi N., Gogate P.R., Pandit A.B. Acoustic cavitation for the process intensification of biological oxidation of CETP effluent containing mainly pharmaceutical compounds: Understanding into effect of parameters and toxicity analysis. Ultrason. Sonochemistry. 2023;98 doi: 10.1016/j.ultsonch.2023.106524. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Malek M.N.F.A., Hussin N.M., Embong N.H., Bhuyar P., Rahim M.H.A., Govindan N., Maniam G.P. Ultrasonication: a process intensification tool for methyl ester synthesis: a mini review. Biomass Convers. Biorefinery. 2020:1–11. [Google Scholar]
- 46.Chuah L.F., Klemeš J.J., Bokhari A., Asif S., Cheng Y.W., Chong C.C., Show P.L. A review of intensification technologies for biodiesel production. Biofuels Biorefining. 2022:87–116. [Google Scholar]
- 47.Wood R.J., Lee J., Bussemaker M.J. A parametric review of sonochemistry: Control and augmentation of sonochemical activity in aqueous solutions. Ultrason. Sonochemistry. 2017;38:351–370. doi: 10.1016/j.ultsonch.2017.03.030. [DOI] [PubMed] [Google Scholar]
- 48.De-Nasri S.J., Nagarajan S., Robertson P.K., Ranade V.V. Quantification of hydroxyl radicals in photocatalysis and acoustic cavitation: Utility of coumarin as a chemical probe. Chem. Eng. J. 2021;420 [Google Scholar]
- 49.De-Nasri S.J., Sarvothaman V.P., Nagarajan S., Manesiotis P., Robertson P.K., Ranade V.V. Quantifying OH radical generation in hydrodynamic cavitation via coumarin dosimetry: Influence of operating parameters and cavitation devices. Ultrason. Sonochemistry. 2022;90 doi: 10.1016/j.ultsonch.2022.106207. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Sarvothaman V.P., Kulkarni S.R., Subburaj J., Hariharan S.L., Velisoju V.K., Castaño P., Guida P., Prabhudahnkerwadkar D.M., Roberts W.L. Evaluating performance of vortex-diode based hydrodynamic cavitation device scale and pressure drop using coumarin dosimetry. Chem. Eng. J. 2024;481 [Google Scholar]
- 51.Hirano K., Kobayashi T. Coumarin fluorometry to quantitatively detectable OH radicals in ultrasound aqueous medium. Ultrason. Sonochemistry. 2016;30:18–27. doi: 10.1016/j.ultsonch.2015.11.020. [DOI] [PubMed] [Google Scholar]
- 52.Louit G., Foley S., Cabillic J., Coffigny H., Taran F., Valleix A., Renault J.P., Pin S. The reaction of coumarin with the OH radical revisited: hydroxylation product analysis determined by fluorescence and chromatography. Radiat. Phys. Chem. 2005;72(2–3):119–124. [Google Scholar]
- 53.Nöpel J.A., Ayela F. Experimental evidences of radicals production by hydrodynamic cavitation: a short review. C. R. Chim. 2023;26(G2):157–166. [Google Scholar]
- 54.Morison K., Hutchinson C. Limitations of the weissler reaction as a model reaction for measuring the efficiency of hydrodynamic cavitation. Ultrason. Sonochemistry. 2009;16(1):176–183. doi: 10.1016/j.ultsonch.2008.07.001. [DOI] [PubMed] [Google Scholar]
- 55.Arrojo S., Nerín C., Benito Y. Application of salicylic acid dosimetry to evaluate hydrodynamic cavitation as an advanced oxidation process. Ultrason. Sonochemistry. 2007;14(3):343–349. doi: 10.1016/j.ultsonch.2006.06.007. [DOI] [PubMed] [Google Scholar]
- 56.Zhang J., Nosaka Y. Quantitative detection of OH radicals for investigating the reaction mechanism of various visible-light TiO2 photocatalysts in aqueous suspension. J. Phys. Chem. C. 2013;117(3):1383–1391. [Google Scholar]
- 57.Zupanc M., Petkovšek M., Zevnik J., Kozmus G., Šmid A., Dular M. Anomalies detected during hydrodynamic cavitation when using salicylic acid dosimetry to measure radical production. Chem. Eng. J. 2020;396 [Google Scholar]
- 58.Žnidarčič A., Mettin R., Cairós C., Dular M. Attached cavitation at a small diameter ultrasonic horn tip. Phys. Fluids. 2014;26(2) [Google Scholar]
- 59.Kozmus G., Zevnik J., Hočevar M., Dular M., Petkovšek M. Characterization of cavitation under ultrasonic horn tip–proposition of an acoustic cavitation parameter. Ultrason. Sonochemistry. 2022;89 doi: 10.1016/j.ultsonch.2022.106159. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60.Viciconte G., Guida P., Truscott T.T., Roberts W.L. A general description of the cavitation threshold in acoustic systems. Phys. Fluids. 2025;37(1) [Google Scholar]
- 61.The MathWorks Inc. 2022. MATLAB version: 9.13.0 (R2022b) URL: https://www.mathworks.com. [Google Scholar]
- 62.Otsu N., et al. A threshold selection method from gray-level histograms. Automatica. 1975;11(285–296):23–27. [Google Scholar]
- 63.Bradley D., Roth G. Adaptive thresholding using the integral image. J. Graph. Tools. 2007;12(2):13–21. [Google Scholar]
- 64.Gyftopoulos E.P., Beretta G.P. Dover Publications; Mineola, N.Y: 2005. Thermodynamics: Foundations and Applications. [Google Scholar]
- 65.Dular M., Petkovšek M. Cavitation erosion in liquid nitrogen. Wear. 2018;400:111–118. [Google Scholar]
- 66.Žnidarčič A., Mettin R., Dular M. Modeling cavitation in a rapidly changing pressure field–application to a small ultrasonic horn. Ultrason. Sonochemistry. 2015;22:482–492. doi: 10.1016/j.ultsonch.2014.05.011. [DOI] [PubMed] [Google Scholar]
- 67.Brennen C.E. Cambridge University Press; 2014. Cavitation and Bubble Dynamics. [Google Scholar]
- 68.Lauterborn W. Numerical investigation of nonlinear oscillations of gas bubbles in liquids. J. Acoust. Soc. Am. 1976;59(2):283–293. [Google Scholar]
- 69.Crum L.A., Mason T.J., Reisse J.L., Suslick K.S. Sonochemistry and Sonoluminescence. Springer Science & Business Media; 1998. [Google Scholar]
- 70.d’Agostino L., Brennen C.E. Linearized dynamics of spherical bubble clouds. J. Fluid Mech. 1989;199:155–176. [Google Scholar]
- 71.Maeda K., Colonius T. Bubble cloud dynamics in an ultrasound field. J. Fluid Mech. 2019;862:1105–1134. doi: 10.1017/jfm.2018.968. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72.Carey W.M., Fitzgerald J.W. Natural Physical Sources of Underwater Sound: Sea Surface Sound (2) Springer; 1993. Low frequency noise from breaking waves; pp. 277–304. [Google Scholar]
- 73.Hwang P.A., Teague W.J. Low-frequency resonant scattering of bubble clouds. J. Atmos. Ocean. Technol. 2000;17(6):847–853. [Google Scholar]
- 74.Louisnard O. Analytical expressions for primary bjerknes force on inertial cavitation bubbles. Phys. Rev. E— Stat. Nonlinear Soft Matter Phys. 2008;78(3) doi: 10.1103/PhysRevE.78.036322. [DOI] [PubMed] [Google Scholar]
- 75.Rivas D.F., Stricker L., Zijlstra A.G., Gardeniers H.J., Lohse D., Prosperetti A. Ultrasound artificially nucleated bubbles and their sonochemical radical production. Ultrason. Sonochemistry. 2013;20(1):510–524. doi: 10.1016/j.ultsonch.2012.07.024. [DOI] [PubMed] [Google Scholar]
- 76.Ashokkumar M. The characterization of acoustic cavitation bubbles–an overview. Ultrason. Sonochemistry. 2011;18(4):864–872. doi: 10.1016/j.ultsonch.2010.11.016. [DOI] [PubMed] [Google Scholar]
- 77.Mettin R., Eisener J., Koch M., Holly D., Thiele M. On the sound emission of acoustic cavitation. DAGA Hann. 2024 [Google Scholar]
- 78.Auger F., Flandrin P. Improving the readability of time-frequency and time-scale representations by the reassignment method. IEEE Trans. Signal Process. 1995;43(5):1068–1089. [Google Scholar]
- 79.Fulop S.A., Fitz K. Algorithms for computing the time-corrected instantaneous frequency (reassigned) spectrogram, with applications. J. Acoust. Soc. Am. 2006;119(1):360–371. doi: 10.1121/1.2133000. [DOI] [PubMed] [Google Scholar]
- 80.IECTS, et al. 2024. IEC TS 63001:2024 RLV, Measurement of cavitation noise in ultrasonic baths and ultrasonic reactors. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Supplementary information contains videos of different probe diameters used in this work.
.
.
.
.
Data Availability Statement
Data will be made available upon request.