AI-enabled CT-guided end-to-end quantification of total cardiac activity in 18FDG cardiac PET/CT for detection of cardiac sarcoidosis

Robert JH Miller 1,2,*, Aakash Shanbhag 1,3,*, Anna M Marcinkiewicz 1,4, Helen Struble 1, Heidi Gransar 1, Waseem Hijazi 1, Hidesato Fujito 1, Evan Kransdorf 1, Paul Kavanagh 1, Joanna X Liang 1, Valerie Builoff 1, Damini Dey 1, Daniel S Berman 1, Piotr J Slomka 1
PMCID: PMC12147654  NIHMSID: NIHMS2067882  PMID: 40127777

Abstract

Background:

[18F]-fluorodeoxyglucose ([18F]FDG) positron emission tomography (PET) plays a central role in diagnosing and managing cardiac sarcoidosis. We propose a fully automated pipeline for quantification of [18F]FDG PET activity using deep learning (DL) segmentation of cardiac chambers on computed tomography (CT) attenuation maps and evaluate quantitative approaches based on this framework.

Methods:

We included consecutive patients undergoing [18F]FDG PET/CT for suspected cardiac sarcoidosis. DL segmented left atrium, left ventricle (LV), right atrium, right ventricle, aorta, LV myocardium, and lungs from CT attenuation scans. CT-defined anatomical regions were applied to [18F]FDG PET images automatically to target to background ratio (TBR), volume of inflammation (VOI) and cardiometabolic activity (CMA) using full sized and shrunk segmentations.

Results:

A total of 69 patients were included, with mean age of 56.1 ± 13.4 and cardiac sarcoidosis present in 29 (42%). CMA had highest prediction performance (area under the receiver operating characteristic curve [AUC] 0.919, 95% confidence interval [CI] 0.858 – 0.980) followed by VOI (AUC 0.903, 95% CI 0.834 – 0.971), TBR (AUC 0.891, 95% CI 0.819 – 0.964), and maximum standardized uptake value (AUC 0.812, 95% CI 0.701 – 0.923). Abnormal CMA (≥1) had a sensitivity of 100% and specificity 65% for cardiac sarcoidosis. Lung quantification was able to identify patients with pulmonary abnormalities.

Conclusion:

We demonstrate that fully automated volumetric quantification of [18F]FDG PET for cardiac sarcoidosis based on CT attenuation map-derived volumetry is feasible, rapid, and has high prediction performance. This approach provides objective measurements of cardiac inflammation with consistent definition of myocardium and background region.

Keywords: cardiac sarcoidosis, positron emission tomography, quantification, deep learning, hypermetabolism

CONDENSED ABSTRACT

We developed a fully automated pipeline for [18F]-fluorodeoxyglucose ([18F]FDG) PET quantification in patients with suspected cardiac sarcoidosis. A previously validated deep learning model segmented cardiac chambers from computed tomography, followed by quantification of target to background ratio, volume of inflammation and cardiometabolic activity (CMA). A total of 69 patients were included, with cardiac sarcoidosis present in 29 (42%). CMA had highest area under the receiver operating characteristic curve (0.919, 95% confidence interval 0.858 – 0.980). Fully automated volumetric quantification of [18F]FDG PET is feasible and has high prediction performance for cardiac sarcoidosis.

Graphical Abstract

Patients undergoing [18F]-fluorodeoxyglucose ([18F]FDG) positron emission tomography (PET) for suspected cardiac sarcoidosis were included. Deep learning (DL) segmented cardiac structures from associated computed tomography (CT) images, were used to define background and target regions. Those segmentations were then applied to quantify PET activity, demonstrating excellent prediction for cardiac sarcoidosis.

graphic file with name nihms-2067882-f0001.jpg

INTRODUCTION

Cardiac sarcoidosis is an inflammatory cardiomyopathy characterized by the presence of non-caseating granulomas (1). The disorder leads to heart failure, ventricular and atrial arrhythmias, and conduction system disease (1). Accurate diagnosis is central to targeting immunosuppressive medications which can decrease myocardial inflammation, and therefore decrease the incidence of adverse events (13). However, diagnosis of cardiac sarcoidosis can be challenging and relies on a combination of clinical and imaging criteria (4, 5). Cardiac positron emission tomography (PET) with [18F]-fluorodeoxyglucose ([18F]FDG) plays a central role in establishing the diagnosis by identifying myocardial inflammation (48).

While PET imaging is widely used to aid in the diagnosis of sarcoidosis, visual identification of abnormal myocardial hypermetabolism requires significant expertise and is inherently subjective, which may be particularly difficult in cases with borderline myocardial uptake. Additionally, it does not provide treating clinicians with detailed information regarding the severity of inflammation. Quantitative analysis of cardiac hypermetabolism can be performed using maximum standardized uptake value (SUVmax) or volumetric measurements (which capture both volume and intensity of [18F]FDG uptake). Patients with more extensive cardiac inflammation are more likely to have cardiac sarcoidosis (9), but are also more likely to benefit from immunosuppression (10). We previously demonstrated that volume of inflammation (VOI) and cardiometabolic activity (CMA), quantified from [18F]FDG PET studies, had high diagnostic accuracy for cardiac sarcoidosis (11). However, these measurements required manual annotations, which take about 10 minutes per study, and are prone to interobserver variability related to placement of both the background region and region of interest. Additionally, there remains uncertainty regarding the optimal method for quantifying myocardial inflammation (8).

We recently demonstrated that cardiac chambers can be segmented accurately from non-contrast computed tomography (CT) attenuation correction (CTAC) scans (12), utilizing a foundational deep learning CT segmentation model (13). This approach can be translated to “hot spot” imaging such as [18F]FDG PET, to provide fully automated quantification of cardiac activity in selected cardiac chambers. Importantly, this method allows for rapid and consistent evaluation of multiple structures simultaneously, facilitating automatic placement of background regions and rapid comparison between quantitative approaches and alternative region placements. We implemented and evaluated the feasibility and diagnostic accuracy of this approach in patients undergoing [18F]FDG PET to determine the optimal approach for image quantification.

MATERIALS AND METHODS

Study Population

For this study, we utilized our previously described cohort of patients with suspected cardiac sarcoidosis (11). In brief, we included consecutive patients undergoing [18F]FDG PET for suspected cardiac sarcoidosis. Patients who were imaged without adequate dietary preparation (n=1) were excluded. In patients with multiple studies, only the first study was included (n=29). A total of 69 patients were included in the analysis. This study was approved by the institutional review board at Cedars-Sinai Medical Center including waiver of informed consent for use of retrospective data. Analyses were performed on de-identified image sets consistent with Health Insurance Portability and Accountability Act compliance.

Demographic Information

Demographic information including age, sex, height, weight, body mass index and medical history were collected prospectively at the time of PET imaging. Diagnosis of cardiac sarcoidosis was determined retrospectively after reviewing all available clinical, imaging, and pathologic information. Five patients underwent cardiac transplant and diagnosis was based on pathologic examination of the explanted heart (4 with cardiac sarcoidosis, 1 without). Endomyocardial biopsy was performed in an additional 8 patients, two with pathologic evidence of cardiac sarcoidosis, one with lupus cardiomyopathy, one with hypertrophic cardiomyopathy, and 4 with non-specific changes (2 ultimately classified as cardiac sarcoidosis and 2 as not cardiac sarcoidosis). An additional two patients had cardiac sarcoidosis diagnosed on the basis of positive endomyocardial biopsy. The remaining patients were categorized as having definite or probable cardiac sarcoidosis using the Japanese Ministry of Health and Welfare criteria (4), which includes regional [18F]FDG uptake on PET (14). Patients with definite or probable cardiac sarcoidosis were considered to have cardiac sarcoidosis. The presence or absence of abnormal PET activity would have shifted classification of 33 (47.8%) patients, of which 18 were classified as cardiac sarcoidosis and 15 were classified as not cardiac sarcoidosis.

PET Imaging Acquisition

Images were acquired with a whole-body PET/CT scanner (Biograph-64, Siemens, Munich, Germany). All patients were instructed to eat a high-fat/low-carbohydrate diet for 36 hours prior to the scan, with fasting for at least 8 hours prior to the scan. Patients who did not comply with the dietary protocol were re-scheduled for imaging when dietary compliance could be ensured. Scans were not performed if fasting blood glucose was >200 mg/dL. [18F]FDG imaging was performed at median time of 102 minutes (interquartile range [IQR] 93 – 135) after injection of 5–15 mCi of [18F]FDG (based on patient weight) and acquired over 15–20 minutes (15).

CTAC scans were acquired with normal breathing, without ECG gating, spiral mode with pitch 1.5, tube voltage 100 kVp, tube current–time product 11–13 mAs, and a full field-of-view. Reconstruction was performed with a slice thickness of 3mm. Hybrid image sets were evaluated for significant misregistration during processing, and if misregistration was noted during clinical reconstruction the data were re-reconstructed after adjustment, according to standard clinical protocols. Images were reviewed for the presence of right ventricular (RV) and atrial uptake as well as adjacent radiotracer activity (in lymph nodes, lung lesions, or the liver) which may have impacted quantitation. All images were reviewed blinded to quantitative results.

Image Analysis

Semi-quantitative assessments of resting perfusion were assessed using summed rest score (SRS) with dedicated software (QPET, Cedars-Sinai Medical Center, Los Angeles, California) (16, 17). Established thresholds for abnormal perfusion (SRS ≥3) were used as the criteria for perfusion deficit (17). Myocardial hypermetabolism was assessed visually by an experienced cardiologists at the time of clinical reporting. Visual interpretation was used as the criteria for positive PET.

CTAC Heart Segmentation

A previously validated deep learning model (TotalSegmentator), was used to segment structures from CTAC (13). In brief, the model utilizes a no new-net (nnU-Net) architecture (which automatically optimized hyperparameters) to automatically segment anatomic structures from CT images (13). During model development the lungs could be segmented directly from expert annotation of non-contrast images, with Dice scores >0.98 in the testing set (13). For cardiac structures, the model was trained using expert annotations from contrast images which were transferred to registered non-contrast images. We recently demonstrated excellent correlations in an external population between volumes manually segmented from gated, contrast-enhanced CT and automated segmentations from non-contrast, ungated CT (left atrium [LA] volume Spearman r=0.926, left ventricle [LV] myocardium Spearman r=0.947, LV volume Spearman r=0.793, right atrial [RA] volume Spearman r=0.893, and RV volume Spearman r=0.922) (12).

PET Lung Quantification

To further evaluate the potential utility of applying CT-based segmentation and quantification, we also evaluated methods for assessing PET lung quantification. We utilized lung segmentations from the TotalSegmentator [13], then quantified lung activity within both lungs. We evaluated SUVmax, as well as target to background ratio (TBR) based on maximal activity relative to mean activity in the lungs as background. Since no standardized thresholds for abnormal activity exist, we utilized mean + 2 standard deviations as the threshold for abnormal lung activity. Lung activity above background was measured using both volume above threshold (lung metabolic volume [LMV]) and activity above background (lung metabolic activity [LMA]). We evaluated prediction performance for pulmonary abnormalities (lymphadenopathy, pulmonary nodules, or consolidations) based on expert interpretation of CT radiographic findings.

PET Cardiac Quantification

The LA, LV, RA, RV and aorta segmentations were used as background regions for cardiac hypermetabolism quantification. The LV myocardial volume was used as the region of interest for cardiac quantification. For background activity, we evaluated full sized segmentations as well as segmentations that were volumetrically reduced by 50% to avoid partial volume effects. These “shrunk” segmentations were centered within the original segmentation. In total, we evaluated six potential region definitions (LA-full, LA-shrunk, RA-full, RA-shrunk, Aorta-full, Aorta-shrunk) for computation of the background activity. These segmentations were then translated to co-registered [18F]FDG PET images. We applied four potential thresholds for abnormal activity (max counts, 1.3 * max counts, 1.5* max counts, and mean counts + 2 standard deviations) based on previous work (9, 11). In our primary analyses, we utilized maximum background counts as the threshold for abnormal activity for VOI and CMA. An overview of the segmentation methods and overall workflow for cardiac quantification are shown in the Graphical Abstract.

SUVmax was quantified as maximum uptake within the LV myocardial volume. TBR, VOI, and CMA were computed utilizing 10 different background definitions. TBR was calculated as SUVmax for LV myocardium divided by mean background activity. VOI is calculated as the volume of myocardium above threshold for abnormal. CMA is calculated as the total activity within the LV myocardium above the threshold for abnormal.

Lastly, we evaluated whether this approach could be extended to atrial and RV uptake quantification. For atrial quantification, the left and right atrial volumes were combined and grown 30% (while maintaining the position of other segmentations). The background was set as a shrunk RV. RV quantification was based on a grown (30%) segmentation of the RV without modification of the LV myocardial (or atrial) segmentation. A shrunk LA segmentation was used as the background activity.

Statistical Analysis

Continuous variables were summarized as mean +/− standard deviation and compared with a Student’s t-test or Mann-Whitney U test as appropriate. Categorical variables were summarized as number (proportion) and compared using a chi-square or Fisher exact test as appropriate. Receiver operating characteristic (ROC) curves were generated to assess prediction performance for cardiac sarcoidosis. Area under the ROC curve (AUC) was compared using the method established by DeLong et al. (18). Optimal thresholds were determined using the Youden index. All statistical tests were two-sided and a p-value <0.05 was considered significant, with analyses performed in Stata version 13 (StataCorp, College Station, Texas).

Independent Data Access and Analysis

RM, AS and PS had full access to all the data in the study and take responsibility for its integrity and the data analysis.

RESULTS

Patient Population

A total of 69 patients were included, with mean age 56.1 ± 13.4 years and 41 (59.4%) male patients. Cardiac sarcoidosis was present in 29 (42%) patients. Patient characteristics stratified by the presence of cardiac sarcoidosis are shown in Table 1. There was no significant difference in median age between patients with cardiac sarcoidosis compared to those without cardiac sarcoidosis (51 vs 60, p=0.175). More patients with cardiac sarcoidosis had atrioventricular block (41% vs 13%, p=0.010) and reduced left ventricular function (86% vs 35%, p<0.001).

Table 1:

Clinical characteristics stratified by diagnosis of cardiac sarcoidosis.

No Cardiac Sarcoidosis
N=40
Cardiac Sarcoidosis
N=29
p-value
Age, median (IQR) 60 (46, 68) 51 (45, 59) 0.175
Male, n (%) 22 (55%) 19 (66%) 0.460
Cardiac Sarcoidosis Features
 Advanced AV block 5 (13%) 12 (41%) 0.010
 Basal Septal Thinning 0 (0%) 3 (10%) 0.071
 Visually Abnormal PET 2 (5%) 24 (83%) <0.001
 Decreased LVEF 14 (35%) 25 (86%) <0.001
 Abnormal electrocardiogram 15 (38%) 14 (48%) 0.461
 Abnormal echocardiogram 1 (3%) 1 (3%) 1.000
 Perfusion deficit 8 (20%) 17 (59%) 0.002
 Late gadolinium enhancement 14 (35%) 8 (28%) 0.605

AV – atrioventricular, IQR – interquartile range, LVEF – left ventricular ejection fraction, PET – positron emission tomography.

Fully automated quantification was performed in ~16 seconds per patient. Imaging characteristics stratified by the presence of cardiac sarcoidosis are summarized in Table 2. Patients with cardiac sarcoidosis had higher values for all quantitative metrics (all p<0.01).

Table 2:

Imaging characteristics stratified by diagnosis of cardiac sarcoidosis.

No Cardiac Sarcoidosis
N=40
Cardiac Sarcoidosis
N=29
p-value
SUVmax, median (IQR) 2.23 (1.89, 2.62) 4.34 (2.54, 8.80) <0.001
Full left atrium, median (IQR)
 Target to Background Ratio 1.30 (1.22, 1.49) 2.11 (1.75, 4.98) <0.001
 VOI (mL) 0.0 (0.0, 0.6) 7.2 (0.1, 51.7) <0.001
 CMA (mL) 0.0 (0.0, 1.3) 35.7 (0.2, 224) <0.001
Shrunk left atrium, median (IQR)
 Target to Background Ratio 1.28 (1.18, 1.47) 2.18 (1.73, 5.23) <0.001
 VOI (mL) 0.0 (0.0, 1.2) 17.5 (3.3, 49.0) <0.001
 CMA (mL) 0.0 (0.0, 5.1) 98.4 (28.7, 303) <0.001
Full left ventricle, median (IQR)
 Target to Background Ratio 1.42 (1.30, 1.63) 2.16 (1.72, 4.11) <0.001
 VOI (mL) 0.0 (0.0, 0.1) 0.2 (0.0, 1.6) 0.009
 CMA (mL) 0.0 (0.0, 0.2) 0.6 (0.0, 2.3) 0.006
Shrunk left ventricle, median (IQR)
 Target to Background Ratio 1.42 (1.30, 1.62) 2.16 (1.71, 4.11) <0.001
 VOI (mL) 0.0 (0.0, 1.6) 3.8 (0.8, 9.7) <0.001
 CMA (mL) 0.1 (0.0, 3.5) 116 (2.5, 41.9) <0.001
Full right atrium, median (IQR)
 Target to Background Ratio 1.35 (1.26, 1.46) 2.07 (1.83, 4.75) <0.001
 VOI (mL) 0.0 (0.0, 0.0) 13.2 (2.4, 42.5) <0.001
 CMA (mL) 0.0 (0.0, 0.1) 37 (9.9, 216) <0.001
Shrunk right atrium, median (IQR)
 Target to Background Ratio 1.27 (1.20, 1.44) 2.02 (1.74, 4.83) <0.001
 VOI (mL) 0.0 (0.0, 0.4) 37.7 (3.6, 92.6) <0.001
 CMA (mL) 0.0 (0.0, 0.9) 92.2 (20.2, 289) <0.001
Full right ventricle, median (IQR)
 Target to Background Ratio 1.45 (1.35, 1.63) 2.10 (1.90, 4.27) <0.001
 VOI (mL) 0.0 (0.0, 0.1) 3.0 (0.2, 7.1) <0.001
 CMA (mL) 0.0 (0.0, 0.1) 14.0 (1.3, 27.3) <0.001
Shrunk right ventricle, median (IQR)
 Target to Background Ratio 1.35 (1.24, 1.51) 2.08 (1.79, 4.28) <0.001
 VOI (mL) 0.0 (0.0, 0.3) 11.0 (3.3, 20.5) <0.001
 CMA (mL) 0.0 (0.0, 0.5) 31.1 (11.5, 94.1) <0.001
Full Aorta, median (IQR)
 Target to Background Ratio 1.17 (1.08, 1.35) 2.16 (1.51, 4.64) <0.001
 VOI (mL) 0.0 (0.0, 0.0) 0.1 (0.0, 37.6) <0.001
 CMA (mL) 0.0 (0.0, 0.0) 0.1 (0.0, 169) <0.001
Shrunk Aorta, median (IQR)
 Target to Background Ratio 1.12 (1.07, 1.30) 1.92 (1.57, 4.64) <0.001
 VOI (mL) 0.0 (0.0, 0.0) 11.36 (0, 59.6) <0.001
 CMA (mL) 0.0 (0.0, 0.0) 48.1 (0.0, 181) <0.001

CMA – cardiometabolic activity, IQR – interquartile range, SUVmax – maximum standardized uptake value, VOI – volume of inflammation.

Prediction Performance for Cardiac Sarcoidosis

Prediction performance for cardiac sarcoidosis using CMA is shown in Figure 1. LA-shrunk (AUC 0.919, 95% confidence interval [CI] 0.858 – 0.980) had the highest prediction performance, and was significantly better compared to RA-full (AUC 0.837, 95% CI 0.739 – 0.936), LA-full (AUC 0.787, 95% CI 0.679 – 0.895), either LV background (full: AUC 0.688, 95% CI 0.564 – 0.812; shrunk: AUC 0.805, 95% CI 0.699 – 0.911), and either aorta background (full: AUC 0.752, 95% CI 0.656 – 0.847; shrunk: AUC 0.796, 95% CI 0.695 – 0.896) (all p<0.05). For comparison, the AUC for visual interpretation (which is included in the criteria for diagnosis) was 0.872 (95% CI 0.789 – 0.954). Similar results were seen for the prediction performance of VOI (Supplementary Figure 1). There were no significant differences in prediction performance using TBR (Supplementary Figure 2).

Fig. 1.

Fig. 1

Prediction performance for cardiac sarcoidosis using different methods for quantifying cardiometabolic activity. Segments were derived using deep learning from computed tomography attenuation imaging. Cardiac sarcoidosis was present in 29/69 patients. AUC – area under the receiver operating characteristic curve, CI – confidence interval, LA – left atrium, LV – left ventricle, RA – right atrium, RV – right ventricle

When evaluating different quantification methods utilizing LA-shrunk as the background (Figure 2), CMA had the highest performance followed by VOI (AUC 0.903, 95% CI 0.834 – 0.971), TBR (AUC 0.891, 95% CI 0.819 – 0.964), and SUVmax (AUC 0.812, 95% CI 0.701 – 0.923). Thresholds for each metric, with the associated sensitivity, specificity, positive predictive value, and negative predictive value are shown in Table 3. Abnormal CMA (≥1) had a sensitivity of 100% and specificity 65% for cardiac sarcoidosis. Performance in patients with and without cardiac devices is shown in Supplementary Figure 3 and in patients with or without adjacent radiotracer activity in Supplementary Figure 4. Case examples of segmentations in patients with cardiac devices are shown in Figure 3.

Fig. 2.

Fig. 2

Prediction performance for cardiac sarcoidosis using methods based on a shrunk left atrium background. The area under the receiver operating characteristic curve (AUC) for cardiometabolic activity (CMA), volume of inflammation (VOI), target to background ratio, and maximum standardized uptake value (SUVmax) were not significantly different. CI – confidence interval

Table 3:

Thresholds for quantitative measures.

Threshold Sensitivity Specificity PPV NPV
SUVmax ≥2.88 66 88 79 78
Target to Background Ratio ≥1.16 86 80 76 89
VOI (mL) >0 100 53 60 100
CMA (mL) ≥1 100 65 67 100

Diagnostic performance for identification of cardiac sarcoidosis. CMA – cardiometabolic activity, NPV – negative predictive value, PPV – positive predictive value, SUVmax – maximum standardized uptake value, VOI – volume of inflammation

Fig. 3.

Fig. 3

Examples of segmentations in two patients with cardiac devices. There were no significant errors identified in the right atrium segmentation (red) or right ventricular segmentation (pink) resulting from metal artifact.

A comparison of thresholds for different background activity are shown in Supplementary Table 1. Quantification based on maximal background count or mean background count + 2 standard deviations had the highest prediction performance. A comparison of thresholds for VOI and CMA, with corresponding sensitivity and specificity, is shown in Supplementary Table 2.

Prediction performance for atrial and RV quantification are shown in Supplementary Tables 3 and 4.

Prediction Performance for Pulmonary Abnormalities

Pulmonary abnormalities were present in 34 patients, including 21 with isolated lymphadenopathy and 13 patients with lymphadenopathy in combination with pulmonary nodules or consolidations. Prediction performance for pulmonary abnormalities is shown in Supplementary Figure 5. LMA had the highest prediction performance (AUC 0.733, 95% CI 0.613 – 0.852), and had significantly higher AUC compared to LMV (AUC 0.609, 95% CI 0.474 – 0.744, p=0.004).

Case Examples

An example of quantification in a patient with cardiac sarcoidosis is shown in Figure 4. Using LA-shrunk background, the patient had elevated values for SUVmax (9.83), TBR (5.23), VOI (111 mL) and CMA (1052 mL). Diagnosis of cardiac sarcoidosis was established by reduced left ventricular ejection fraction, typical pattern of late gadolinium enhancement, and visually active PET.

Fig. 4.

Fig. 4

Case example in a man with cardiac sarcoidosis. Computed tomography imaging (Panel a) is segmented using deep learning model (Panel b) to derive left atrial (blue), right atrial (red), aorta (purple) and left ventricular myocardium (green). Maximum intensity projection in panel c demonstrating uptake in the septum and lateral wall of the left ventricle. Fused imaging is shown in Panel d, with scale shown in standardized uptake values (SUV). Chamber segmentations are applied to the co-registered image sets (Panel e) in order to quantify radiotracer activity. The study was positive by visual interpretation and had abnormal quantification. CMA – cardiometabolic activity, SUVmax – maximum SUV, TBR – target to background ratio, VOI – volume of inflammation

Figure 5 shows quantification in a patient without cardiac sarcoidosis. In the second case, using LA-shrunk background the SUVmax was 1.84, TBR was 0.99, and both CMA and VOI were 0 mL. The patient had no criteria for cardiac sarcoidosis other than reduced left ventricular ejection fraction.

Fig. 5.

Fig. 5

Case example in a man without cardiac sarcoidosis. The patient had a reduced left ventricular ejection fraction but no other criteria for cardiac sarcoidosis. Computed tomography imaging (Panel a) is segmented using deep learning model (Panel b) to derive left atrial (blue), right atrial (red), aorta (purple) and left ventricular myocardium (green). Maximum intensity projection in panel c demonstrating no significant cardiac uptake. Fused imaging is shown in Panel d, with scale shown in standardized uptake values (SUV). The study was reported as possible uptake in the mid-ventricular septum. Chamber segmentations are applied to the co-registered image sets (Panel e) to quantify radiotracer activity. Deep learning quantification was normal. The patient’s only criteria for sarcoidosis were possible active PET and reduced ventricular function. CMA – cardiometabolic activity, SUVmax – maximum SUV, TBR – target to background ratio, VOI – volume of inflammation

Figure 6 shows a case with cardiac sarcoidosis with less intense PET activity which was confirmed pathologically in the explanted heart. The patient had mildly elevated values for SUVmax (2.24), TBR (1.32), VOI (4 mL) and CMA (8 mL) using LA-shrunk background. An example of pulmonary quantification is shown in Supplementary Figure 6.

Fig. 6.

Fig. 6

Case example in a man with cardiac sarcoidosis confirmed in the explanted heart. Computed tomography imaging (Panel a) is segmented using deep learning model (Panel b) to derive left atrial (blue), right atrial (red), aorta (purple) and left ventricular myocardium (green). Maximum intensity projection in panel c. Fused imaging is shown in Panel d, with scale shown in standardized uptake values (SUV). Chamber segmentations are applied to the co-registered image sets (Panel e) in order to quantify radiotracer activity. Deep learning quantification was mildly abnormal, with possible uptake visualized in the distal anterior wall. CMA – cardiometabolic activity, SUVmax – maximum SUV, TBR – target to background ratio, VOI – volume of inflammation

DISCUSSION

We demonstrated the feasibility of fully automated quantification of [18F]FDG PET by leveraging recent developments in deep learning-based segmentation of CT attenuation maps. Quantification could be performed rapidly, with a mean processing time of ~16 second, without the need for manual definition of myocardial or background regions. Since CT imaging is utilized for the anatomic definition, the quality of the segmentation is not dependent on the pattern or intensity of myocardial uptake, as is the case for conventional approaches. Full automation allowed us to evaluate multiple quantitative metrics simultaneously (including cardiac and pulmonary quantification), with perfect reproducibility since no operator interaction was involved. This approach could be integrated into clinical practice to provide physicians with consistent quantitative metrics of cardiac inflammation to guide diagnosis and treatment recommendations in patients with suspected cardiac sarcoidosis.

Quantification of [18F]FDG PET studies may aid physicians in establishing a diagnosis of cardiac sarcoidosis (9, 11, 19). This could be important for centers with less expertise or in cases that are equivocal. In our analyses, the accuracy for [18F]FDG quantification obtained by the fully automatic processing was excellent for most of the quantitative metrics. In fact, the prediction performance for CMA (using shrunk LA background and SUVmax threshold) was numerically higher than expert visual interpretation (in spite of the latter being used as part of the diagnostic criteria). Our approach can also quantify RV radiotracer activity (which has independent prognostic utility(20, 21)) or atrial radiotracer uptake (which predicts incidence of atrial fibrillation(22)). However, these results will still need to be reviewed by a physician since adjacent radiotracer activity (eg: lymph nodes or lung lesions) could be unintentionally quantified. Similarly, although we did not identify significantly lower prediction performance in patients with cardiac devices, physicians should review segmentation results to ensure they were not impacted by beam hardening artifact. Additionally, we demonstrated that our approach could be used to quantify radiotracer activity related to pulmonary abnormalities which may be a helpful aid for interpreting physicians without a radiology background (23). However, a more clinically relevant application may be improving reproducibility of serial measurements of [18F]FDG uptake. Variations in region of interest placement would be minimized with the proposed approach. Additionally, it enables automated region placement for background activity which improves reproducibility (8). Previous studies have shown that improvement in quantitative measures of [18F]FDG PET are associated with improvement in ventricular function (24), and may have prognostic importance (25). Since physicians may use these serial measurements to adjust immunosuppressive therapy (26, 27), it is critically important that they can be performed in a consistent and highly reproducible manner.

In our analysis, abnormal CMA was able to identify patients with cardiac sarcoidosis with a high degree of sensitivity (100%) while maintaining specificity (65%). In contrast, we found that abnormal values for SUVmax were able to identify patients with higher specificity (88%) than sensitivity (66%). Physicians could potentially utilize both measures to maximize both sensitivity and specificity. Additionally, as with any threshold, greater extents of abnormal quantitation should lead physicians to diagnose cardiac sarcoidosis with more confidence. For example, the CMA was much higher in our first presented case (which was interpreted clinically as positive) compared to the third case (which was interpreted clinically as equivocal). The thresholds we identified for maximal background activity are potentially more intuitive clinically compared to thresholds based on mean and standard deviation. The former identifies any measurable activity as abnormal, while with the latter, very small amounts of activity would be considered normal (since ~3% of blood pool activity could fall within the abnormal range).

We evaluated different potential background regions as well as thresholds for abnormal activity. Regarding selection of the background region, the use of shrunk LA, shrunk RV, or shrunk RA backgrounds tended towards higher prediction performance in analyses using maximal background activity as the threshold. However, we also identified a relationship between the threshold for abnormal activity and choice of background region. In particular, we saw less variability in prediction performance by utilizing a threshold based on the mean and standard deviation of background activity and no significant difference in prediction performance using TBR. The two findings in concert suggest that background activity measurements can be influenced by small volumes of adjacent myocardial or lymph node activity. When we evaluated RV quantification, we also noted a trend towards higher prediction performance for thresholds based on mean and standard deviation. Therefore, quantification using thresholds based on maximal activity seem to benefit from smaller regions of interest to avoid this adjacent hypermetabolism. Meanwhile, utilizing thresholds based on mean activity minimizes the impact on threshold values. These observations are potentially relevant for automated quantification of a variety of hot spot imaging tests.

Our study has a few limitations, the most important of which is that the sample size is limited, allowing us only to demonstrate the feasibility of the approach. Future studies in larger cohorts are needed to more robustly evaluate the quantification methods we have developed to ensure that they are applicable in different patient populations with a variety of imaging protocols (28). In particular, these studies could help validate the utility of the derived thresholds and evaluate the utility of identifying the most diseased myocardial segment with respect to clinical presentation or response to therapy. However, it is worth noting that CMA and VOI include per-pixel quantification of activity above background in its calculation and by extension any value above 0 could be considered abnormal, particularly given the high negative predictive value of these measures. However, the fully automated workflow could potentially facilitate the quantification needed for these larger, multi-center studies. Additionally, it could potentially be modified to automate quantification on a segmental basis (using a standard 17-segment model) which can be challenging for conventional software approaches given the patchy or absent uptake present in many [18F]FDG PET studies performed for cardiac sarcoidosis. Additionally, we relied on existing clinical criteria for establishing the diagnosis, since cardiac pathology was not available in all patients. However, this reflects current clinical practice and limits bias that may be related to requiring cardiac pathology. Similarly, the population without cardiac sarcoidosis is reflective of selection bias and includes some patients with hypertrophic cardiomyopathy, amyloidosis, genetic cardiomyopathy, and idiopathic non-ischemic cardiomyopathy. Lastly, our method (just like all hybrid imaging) relies on accurate registration between image sets, and this needs to be assessed by technologists during the reconstruction process.

CONCLUSIONS

We demonstrate that fully automated quantification of [18F]FDG PET for cardiac sarcoidosis, utilizing chamber segmentation from the CT attenuation maps, is feasible and has high diagnostic accuracy. This approach could be applied to provide objective and reproducible measurements of cardiac hypermetabolism to aid in diagnosis or follow response to therapy.

Supplementary Material

Supplementary Material

NEW KNOWLEDGE GAINED.

  • Deep learning can segment cardiac chambers to facilitate an automated image quantification workflow in cardiac sarcoidosis.

  • Quantification using these segmentations has high prediction performance for cardiac sarcoidosis.

CLINICAL IMPLICATIONS.

  • The proposed workflow could be used by centers to quantify [18F]FDG PET studies to gain adjunctive information when diagnosing cardiac sarcoidosis or to quantify response to therapy.

  • This approach could be adapted to a variety of PET radiotracers which target myocardial disease processes.

SOURCES OF FUNDING

This research was supported in part by grant R35HL161195 from the National Heart, Lung, and Blood Institute/ National Institutes of Health (NHLBI/NIH) (PI: Piotr Slomka). The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health.

ABBREVIATIONS

[18F]FDG

[18F]-fluorodeoxyglucose

PET

positron emission tomography

CT

computed tomography

DL

deep learning

TBR

target to background ratio

VOI

volume of inflammation

CMA

cardiometabolic activity

SRS

summed rest score

IQR

interquartile range

SUV

standardized uptake value

Footnotes

CONFLICT OF INTEREST

Dr. Miller reports research and consulting support from Pfizer. Mr. Kavanagh, Drs. Berman and Slomka participate in software royalties for QPS software at Cedars-Sinai Medical Center. Dr. Slomka has received research grant support from Siemens Medical Systems and consulting fees from Synektik. Dr. Berman has served as a consultant for GE Healthcare. Dr. Hidesato Fujito received research funding support from the Fukuda Foundation for Medical Technology, the Nihon University School of Medicine Alumni Association Research Grant and the Society of Nuclear Medicine and Molecular Imaging Wagner-Torizuka Fellowship grant. The authors have no other relevant disclosures.

REFERENCES

  • (1).Lehtonen J, Uusitalo V, Pöyhönen P, Mäyränpää MI, Kupari M. Cardiac sarcoidosis: phenotypes, diagnosis, treatment, and prognosis. Eur Heart J 2023;44:1495–510. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (2).Fazelpour S, Sadek MM, Nery PB, Beanlands RS, Tzemos N, Toma M et al. Corticosteroid and Immunosuppressant Therapy for Cardiac Sarcoidosis. J Am Heart Ass 2021;10:e021183. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (3).Roshankar G, Zhang JY, Parekh D, Lyons K, Isaac D, Miller RJH. Steroid Sparing Immunosuppression in Management of Cardiac Sarcoidosis. Health Sci Rev 2022;4:100034. [Google Scholar]
  • (4).Soejima K, Yada H. The work-up and management of patients with apparent or subclinical cardiac sarcoidosis. J Cardiovasc Electrophysiol 2009;20:578–83. [DOI] [PubMed] [Google Scholar]
  • (5).Birnie DH, Sauer WH, Bogun F, Cooper JM, Culver DA, Duvernoy CS et al. HRS expert consensus statement on the diagnosis and management of arrhythmias associated with cardiac sarcoidosis. Heart Rhythm 2014;11:1305–23. [DOI] [PubMed] [Google Scholar]
  • (6).Skali H, Schulman AR, Dorbala S. 18F-FDG PET/CT for the assessment of myocardial sarcoidosis. Curr Cardiol Rep 2013;15:352. [PMC free article] [PubMed] [Google Scholar]
  • (7).Chareonthaitawee P, Beanlands RS, Chen W, Dorbala S, Miller EJ, Murthy VL et al. Joint SNMMI–ASNC expert consensus document on the role of 18 F-FDG PET/CT in cardiac sarcoid detection and therapy monitoring. J Nucl Cardiol 2017;24(5):1741–1758. [DOI] [PubMed] [Google Scholar]
  • (8).Sperry BW, Bateman TM, Akin EA, Bravo PE, Chen W, Dilsizian V et al. Hot Spot Imaging in Cardiovascular Diseases: An Information Statement from SNMMI, ASNC, and EANM. J Nucl Cardiol 2023;30(2):626–652. [DOI] [PubMed] [Google Scholar]
  • (9).Ahmadian A, Brogan A, Berman J, Sverdlov AL, Mercier G, Mazzini M et al. Quantitative interpretation of FDG PET/CT with myocardial perfusion imaging increases diagnostic information in the evaluation of cardiac sarcoidosis. J Nucl Cardiol 2014;21:925–39. [DOI] [PubMed] [Google Scholar]
  • (10).Subramanian M, Swapna N, Ali AZ, Saggu DK, Yalagudri S, Kishore J et al. Pre-treatment myocardial 18FDG uptake predicts response to immunosuppression in patients with cardiac sarcoidosis. JACC Cardiovasc Imaging 2021;14:2008–16. [DOI] [PubMed] [Google Scholar]
  • (11).Miller RJH, Cadet S, Pournazari P, Pope A, Kransdorf E, Hamilton MA et al. Quantitative Assessment of Cardiac Hypermetabolism and Perfusion for Diagnosis of Cardiac Sarcoidosis. J Nucl Cardiol 2022;29:86–96. [DOI] [PubMed] [Google Scholar]
  • (12).MIller R, Shanbhag A, Killekar A, Lemley M, Bednarski B, Kavanagh P et al. AI-defined cardiac anatomy Improves Risk Stratification of Hybrid Perfusion Imaging. JACC Cardiovasc Imaging 2024,17(7):780–791. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (13).Wasserthal J, Breit H-C, Meyer MT, Pradella M, Hinck D, Sauter AW et al. TotalSegmentator: Robust Segmentation of 104 Anatomic Structures in CT Images. Radiol Artific Intell 2023;5:e230024. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (14).Betensky BP, Tschabrunn CM, Zado ES, Goldberg LR, Marchlinski FE, Garcia FC et al. Long-term follow-up of patients with cardiac sarcoidosis and implantable cardioverter-defibrillators. Heart Rhythm 2012;9:884–91. [DOI] [PubMed] [Google Scholar]
  • (15).Dilsizian V, Bacharach SL, Beanlands RS, Bergmann SR, Delbeke D, Dorbala S et al. ASNC imaging guidelines/SNMMI procedure standard for positron emission tomography (PET) nuclear cardiology procedures. J Nucl Cardiol 2016;23:1187–226. [DOI] [PubMed] [Google Scholar]
  • (16).Dekemp RA, Declerck J, Klein R, Pan XB, Nakazato R, Tonge C et al. Multisoftware reproducibility study of stress and rest myocardial blood flow assessed with 3D dynamic PET/CT and a 1-tissue-compartment model of 82Rb kinetics. J Nucl Med 2013;54:571–7. [DOI] [PubMed] [Google Scholar]
  • (17).Nakazato R, Berman DS, Dey D, Le Meunier L, Hayes SW, Fermin JS et al. Automated quantitative Rb-82 3D PET/CT myocardial perfusion imaging: normal limits and correlation with invasive coronary angiography. J Nucl Cardiol 2012;19:265–76. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (18).DeLong ER, DeLong DM, Clarke-Pearson DL. Comparing the areas under two or more correlated receiver operating characteristic curves: a nonparametric approach. Biometrics 1988;44:837–45. [PubMed] [Google Scholar]
  • (19).Ishiyama M, Soine LA, Vesselle HJ. Semi-quantitative metabolic values on FDG PET/CT including extracardiac sites of disease as a predictor of treatment course in patients with cardiac sarcoidosis. EJNMMI Res 2017;7:67. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (20).Tuominen H, Haarala A, Tikkakoski A, Kähönen M, Nikus K, Sipilä K. FDG-PET in possible cardiac sarcoidosis: Right ventricular uptake and high total cardiac metabolic activity predict cardiovascular events. J Nucl Cardiol 2021;28:199–205. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (21).Blankstein R, Osborne M, Naya M, Waller A, Kim CK, Murthy VL et al. Cardiac Positron Emission Tomography Enhances Prognostic Assessments of Patients With Suspected Cardiac Sarcoidosis. J Am Coll Cardiol 2014;63:329–36. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (22).Niemelä M, Uusitalo V, Pöyhönen P, Schildt J, Lehtonen J, Kupari M. Incidence and predictors of atrial fibrillation in cardiac sarcoidosis. JACC Cardiovasc Imaging 2022;15:1622–31. [DOI] [PubMed] [Google Scholar]
  • (23).Al-Mallah MH, Bateman TM, Branch KR, Crean A, Gingold EL, Thompson RC et al. 2022 ASNC/AAPM/SCCT/SNMMI guideline for the use of CT in hybrid nuclear/CT cardiac imaging. J Nucl Cardiol 2022;29:3491–535. [DOI] [PubMed] [Google Scholar]
  • (24).Osborne MT, Hulten EA, Singh A, Waller AH, Bittencourt MS, Stewart GC et al. Reduction in ¹⁸F-fluorodeoxyglucose uptake on serial cardiac positron emission tomography is associated with improved left ventricular ejection fraction in patients with cardiac sarcoidosis. J Nucl Cardiol 2014;21:166–74. [DOI] [PubMed] [Google Scholar]
  • (25).Sperry BW, Vranian MN, Tower-Rader A, Hachamovitch R, Hanna M, Brunken R et al. Regional Variation in Technetium Pyrophosphate Uptake in Transthyretin Cardiac Amyloidosis and Impact on Mortality. JACC Cardiovasc Imaging 2018;11:234–42. [DOI] [PubMed] [Google Scholar]
  • (26).Cabrera R, Ananthasubramaniam K. Diagnosis, therapeutic response assessment, and detection of disease recurrence in cardiac sarcoidosis: Integral role of cardiac PET. J Nucl Cardiol 2016;23:850–3. [DOI] [PubMed] [Google Scholar]
  • (27).Chareonthaitawee P, Bois John P. Quantitative FDG PET/CT to Guide Treatment of Cardiac Sarcoidosis. JACC Cardiovasc Imaging 2021;14:2017–9. [DOI] [PubMed] [Google Scholar]
  • (28).Manabe O, Takahashi K, Kawakami H, Ohtsuka A, Aikawa T, Maki H et al. Diagnostic values of delayed additional FDG PET/CT scan in the evaluation of cardiac sarcoidosis. Ann Nucl Med 2023;37:535–40. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Material

RESOURCES