Abstract
Purpose
23Na MRI is a functional imaging technique that facilitates measurements of the renal corticomedullary sodium gradient (CMSG). The CMSG can be determined by a region of interest (ROI) in the renal parenchyma along the corticomedullary axis (Line method) or by dividing the renal parenchyma into concentric layers, using the twelve‐layer concentric objects (TLCO) method. The aim of this study was to investigate the differences, strengths, and weaknesses in determining the CMSG using these methods.
Methods
Ten healthy volunteers were examined on a 3 T MRI‐system. 23Na images were acquired using a double‐tuned 23Na/1H surface coil and a golden angle (GA) density‐adapted 3D radial (DA‐3D‐RAD) sequence. The CMSG was determined with the Line and TLCO methods. Intra‐ and inter‐reader analyses were performed by two radiologists.
Results
The evaluated CMSG by the two methods does not differ statistically significantly. Compared to the Line method, the TLCO method provides improved results in terms of reliability, precision, reproducibility, and concordance in intra‐ and inter‐reader analyses. A CMSG of (6.7 ± 2.6) mM/mm was determined using the TLCO method in the segmentation process with the lowest standard deviation.
Conclusion
The TLCO method shows superior performance in determining the CMSG compared to the Line method. Accordingly, the use of the TLCO method is recommended for future renal CMSG studies.
Keywords: corticomedullary 23Na gradient, kidney sodium MRI, sodium MRI23Na MR imaging, twelve‐layer concentric objects (TLCO) method
1. INTRODUCTION
In 2017, 1.2 million deaths were attributed to chronical kidney diseases (CKD), with projections indicating an anticipated increase in incidence because of demographic change and a rise in cardiovascular diseases and diabetes. 1 , 2 Early detection of CKD is of paramount importance for prevention and treatment, given that it is often asymptomatic in its early stages. 2 CKD is often associated with kidney inflammation, changes in the microstructure, insufficient oxygen supply, and altered hemodynamics. 3 These aspects can be analyzed using different morphological or functional MRI methods including T1/T2 mapping, BOLD imaging, QSM, DWI, DTI, and 23Na‐MRI. 4 , 5 , 6 , 7 23Na‐MRI can be used to determine the corticomedullary sodium gradient (CMSG) and has been investigated as a potential indicator of kidney diseases such as tubular necrosis, stenosis, and atrophy. 8 , 9 Furthermore, 23Na plays an essential role in the regulation of the water and electrolyte balance of the body. 10 23Na‐MRI uses endogenous 23Na nuclei as a signal source. The physical properties of 23Na nuclei such as very short relaxation times, low abundance and 800 to 5500 times lower SNR compared to 1H‐MRI make it challenging to achieve sufficient image quality in 23Na‐MRI. 11 Therefore, special acquisition techniques and hardware are required to capture 23Na images. Based on the signal intensity in the 23Na images, 23Na concentrations can be calculated. 12 , 13 , 14 In addition to the 23Na concentration, other effects can influence the image intensity such as partial volume effects, differences in sodium relaxation times and the coil sensitivity. 15 These influences can be minimized by suitable correction methods. Because of the difficulty in accurately assessing the sodium concentration, various studies have introduced the term apparent tissue sodium concentration (aTSC). 15 , 16 , 17 Based on the aTSC map, the CMSG can be determined. The CMSG was previously analyzed using a linear region of interest (ROI) through the renal parenchyma along the corticomedullary axis (Line method). 13 , 14 , 18 , 19 In the field of renal BOLD imaging, the twelve‐layer concentric objects (TLCO) segmentation technique was applied. 20 , 21
We hypothesize that the TLCO method can be transferred to determine the CMSG based on aTSC maps (TLCO method). The aim of this study was to investigate differences, strengths, and weaknesses in determining the CMSG with these methods in an intra‐ and inter‐reader study. We hypothesized, that the CMSG does not depend on the chosen method or the radiologist.
2. METHODS
2.1. Study population
Written informed consent was obtained from all participants and the study was approved by the local ethics committee (Ethics Committee, Medical Faculty of the Heinrich‐Heine‐University Düsseldorf, study number 2021–1393). In this prospective study seven female and three male healthy volunteers were examined (mean age 24 ± 2 years).
2.2. Experimental setup and sequence protocol
All measurements were performed on a 3 T MRI System (Siemens Magnetom Prisma, Siemens Healthineers). To obtain morphological reference images, an 18‐channel body coil and a 32‐channel spine coil (Body 18 SlideConnect and Spine 32 DirectConnect, Siemens Healthineers) were used in combination. Morphological reference images were acquired using a half‐Fourier acquisition single‐shot turbo spin echo (HASTE) sequence with a resolution of 0.7 × 0.7 × 3.0 mm3. A double‐tuned 1H/23Na‐surface‐coil (RAPID Biomedical) was used to acquire 23Na images and additional 1H images for image registration. Four reference phantoms with 4% agarose content by weight (ROTI Garose, Carl ROTH) and 23Na concentrations of 50, 75, 100, and 125 mM were placed behind the coil for aTSC calculations. A sensitivity profile of the double‐tuned coil was determined using homogeneous water phantoms containing 154 mM 23Na, positioned behind and in front of the coil. 16 The acquisition parameters included an isotropic resolution of 3.0 mm, a pulse duration of 0.50 ms, a flip angle of 90°, echo times of 0.3 ms, recovery time of 60.0 ms, a number of spokes of 50.000, and 12 averages. In accordance with the sensitivity profile, a pixel‐based sensitivity correction of reference phantoms and in vivo data was performed by dividing the signal intensity of acquired 23Na images by the normalized sensitivity profile. The SNR was calculated as the ratio of signal mean to noise standard deviation in an ROI behind the coil.
All images, obtained with the double tuned coil were measured with a golden angle (GA) density‐adapted 3D radial (DA‐3D‐RAD) sequence developed by Nagel et al. 22 Images were acquired (respectively for [1H/23Na] with an isotropic resolution of [1.0/3.0] mm, a pulse duration of [0.50/0.25] ms, a flip angle of [5.0/70.0] degrees, echo times of [1.0/0.15] ms, and a number of spokes of [30 000/50 000]). For both sequences, the readout time was 5.0 ms and the repetition time was 10.0 ms. The participants were positioned head‐first in the supine position. After the acquisition of the morphological images, the coil setup was modified and the volunteers were repositioned such that the right kidney was aligned with the isocenter and situated at the center of the 1H/23Na‐surface‐coil.
2.3. Image reconstruction and registration
1H and 23Na images were reconstructed with a custom‐written MATLAB script (The MathWorks). 22 Gibbs ringing was reduced using a radial Hanning filter. To adapt the 23Na images to the reference images, the 23Na images were resized to a matrix size of 180 × 180 × 180 using nearest neighbor interpolation. The interpolated images were used for CMSG determination. The 23Na images were registered with the sensitivity profile and spatial dependent sensitivity correction was performed. Pixels with a signal intensity lower than 10% of the measured maximum signal intensity were excluded. The image registration was performed in ITKsnap 3.8.0 by manually merging the 1H images to the acquired morphological HASTE images. 23 The corresponding registration transformation was used to register the aTSC images.
2.4. aTSC determination
A linear regression of the phantom signal intensity as a function of the contained concentration was performed. An aTSC map was calculated based on the determined slope m in the linear regression and the corresponding signal intensity S(23Na). Because of relaxation effects, a correction factor is necessary for aTSC determination. 15 The aTSC map was obtained by
with the slope in . The correction factor depended on the properties of the kidney and the agarose Assuming a bi‐exponential, decay was calculated by , where 17 , 24 , 25
The bi‐exponential decay was subdivided into a short and long component ( and respectively). 25 f describes the ratio between both components. We used = 0.6, = 34.0 ms, = (2.2/20.4) ms, = 38.5 ms, and / = (6.0/13.0) ms for aTSC calculations. 16 , 24 , 26 , 27
2.5. Segmentation
All segmentations in the kidney were performed in the coronal layer posterior to the renal pelvis. The cortex and medullary pyramids were segmented by P.J.G. (medical physicist, 2 years experience) under supervision of A.L. (radiologist, 10 years experience) with the software ITKsnap 3.8.0. 23 The TLCO and Line method for CMSG determination were carried out by Y.C., who is mentioned here as the first radiologist, and M.S. as the second radiologist. Both have 2 years of experience. For intra‐reader analysis the dataset was segmented twice by Y.C. The radiologist and segmentation are noted (e.g., as “1/2” for the first radiologist and the second segmentation).
2.6. Line method
The Line method was performed on the 23Na images with the medulla and cortex outlined. For each volunteer, the radiologist draws a line ROI through an arbitrary medulla. The aTSC within the drawn linear ROI was regridded using a nearest neighbor interpolation to a matrix with a width and depth of 3 pixels/9 mm and the corresponding drawn pixel length. Subsequently, the CMSG was determined by a linear regression of the averaged aTSC along the drawn line up to the maximum signal intensity. 12
2.7. TLCO method
The TLCO algorithm was applied to the anatomical image by manually segmenting the renal parenchyma and transferring the automatic generated concentric layers to 23Na images. Afterward, the aTSC was averaged for each concentric ROI and plotted against the distance from the cortex. The CMSG was then calculated from the outer layers up to the layer with the maximum aTSC.
2.8. Statistical analysis
Statistical analysis was conducted in Python 3.10 using the scipy 1.11.1 library. 28 After the data was tested for normal distribution using a Shapiro–Wilk test, significant differences were determined using a paired t test (Ppt) between the averaged aTSCs in medulla and cortex for each volunteer. All statistical testing was conducted with a significance level of p ≤ 0.05. Bland–Altmann analysis were used to quantify the bias between Line and TLCO CMSG methods. A one‐way analysis of variance test was conducted to determine whether there were significant differences on CMSG depending on the radiologist and segmentation method.
Both methods were evaluated regarding agreement and reproducibility in an inter‐reader and intra‐reader analysis using Lin's concordance correlation coefficient (ρ). The agreement and reproducibility were classified based on McBride et al. 29 as follows: for ρ < 0.90 as poor, for 0.90 < ρ < 0.95 as moderate, for 0.95 < ρ < 0.99 as substantial, and for 0.99 < ρ as almost perfect. The precision (as a linear correlation in the inter‐ and intra‐reader analysis) was quantified by the amount of the Pearson's correlation coefficient (r), with r < 0.1 none, 0.1 < r < 0.3 poor, 0.3 < r < 0.5 fair, 0.5 < r < 0.6 moderate, 0.6 < r < 0.8 moderate strong, and 0.8 < r < 0.9 as very strong and 0.9 < r as perfect. 30 , 31 Intra‐reader and inter‐reader reliabilities were both quantified by a two‐way mixed single measure correlation coefficient (ICC(3,1)) (ri). 32 Based on the work of Koo and Li, 33 the reliability was categorized as follows ri < 0.5 poor, 0.5 < ri < 0.75 moderate, 0.75 < ri < 0.9 good, and 0.9 < ri excellent. 95% confidence intervals (CI) were determined using Fisher z‐transformation.
3. RESULTS
In 23Na images, a SNR of 12.12 ± 2.58 in the medulla and 6.90 ± 1.35 in the cortex was determined (exemplary 23Na image in Figure 1A). The averaged aTSCs across all volunteers resulted in an aTSC of (137.2 ± 31.3) mM in the cortex and (202.3 ± 47.3) mM in the medulla (exemplary aTSC map in Figure 1B). A significant difference in the aTSC between medulla and cortex was found (Ppt <0.0001).
FIGURE 1.

Representative illustration of the evaluation progress. Based on the acquired 23Na images (A), an apparent tissue sodium concentration (aTSC) map of a healthy volunteer (B) was calculated. An aTSC of (103.5 ± 31.7) mM was determined in the cortex and (166.1 ± 38.3) mM in the medulla. The white regions of interest (ROIs) indicated by * shows the area of pixels excluded by the sensitivity profile threshold of 10%. Based on the evaluated areas (C) defined by the Line (orange line) and twelve‐layer concentric object (TLCO) method (blue concentric circles) the corticomedullary sodium gradient (CMSG) was calculated using a linear regression model (D). The calculated CMSG were (9.7 ± 0.1) mM/mm and (7.2 ± 0.1) mM/mm, respectively, for Line and TLCO method.
The application of the TLCO method permitted the evaluation of a larger cortical distance than that achievable with the Line method (Table 1). Furthermore, a stronger linear relationship was identified with the TLCO method than with the Line method. No significant differences were obtained in the determined CMSG, neither between the radiologists nor the segmentation method (Ppt >0.05).
TABLE 1.
Summarized CMSG [mM/mm] results (mean ± SD) and evaluated distance along the renal parenchyma for each segmentation evaluation with the corresponding r2 of the linear regression of 10 healthy volunteers.
| Method | Radiologist/segmentation | CMSG [mM/mm] | Distance [mm] | r2 |
|---|---|---|---|---|
| TLCO | 1/1 | 7.2 ± 3.1 | 15.1 ± 3.7 | 0.91 ± 0.12 |
| 1/2 | 6.7 ± 2.6 | 14.9 ± 3.0 | 0.91 ± 0.09 | |
| 2/1 | 7.5 ± 3.2 | 14.4 ± 2.9 | 0.92 ± 0.09 | |
| Line | 1/1 | 7.4 ± 2.3 | 11.6 ± 4.5 | 0.88 ± 0.11 |
| 1/2 | 6.6 ± 3.7 | 13.5 ± 3.0 | 0.82 ± 0.22 | |
| 2/1 | 7.8 ± 3.4 | 13.3 ± 4.8 | 0.90 ± 0.10 |
Abbreviation: CMSG, corticomedullary sodium gradient.
The Bland–Altman analysis indicates a slight bias of the TLCO method, with a mean difference of 0.2 mM/mm compared to the Line method. The TLCO method provides moderate reproducibility for intra‐reader analysis (ρ = 0.92) and substantial agreement for inter‐reader analysis (ρ = 0.97), whereas the Line method provides poor reproducibility and agreement (ρ < 0.90) (Table 2 and Figure 2). The Pearson's correlation coefficient (r) indicates a very strong precision of the TLCO method for the inter‐ and intra‐reader analysis (r = 0.97 and r = 0.95, respectively). A poor precision was determined for the Line method in inter‐reader analysis (r = −0.13) whereas the precision was strong in the intra‐reader analysis (r = 0.78). The TLCO method shows for inter‐ and intra‐reader analysis an excellent reliability (ri = 0.95). A poor reliability (ri = 0.25) in both the intra‐ and inter‐reader analysis was obtained for the Line method. There were no significant differences (Ppt >0.05) between Line and TLCO methods for the intra‐ and inter‐reader analysis.
TABLE 2.
Summarized results of the CMSG (mean ± SD) and statistical analysis of the inter‐ and intra‐reader analysis.
| Inter‐reader analysis | Intra‐reader analysis | |||
|---|---|---|---|---|
| TLCO method | Line method | TLCO method | Line method | |
| CMSG [mM/mm] | 7.3 ± 3.0 | 7.6 ± 2.8 | 6.9 ± 2.7 | 7.0 ± 3.0 |
| ρ [CI] | 0.97 [0.86–0.99] | −0.12 [−0.70 to 0.55] | 0.92 [0.67–0.98] | 0.67 [0.07–0.91] |
| ri [CI] | 0.95 [0.86–0.99] | 0.25 [−0.12 to 0.68] | 0.95 [0.86–0.99] | 0.25 [−0.12 to 0.68] |
| r [CI] | 0.97 [0.88–0.99] | −0.13 [−0.70 to 0.55] | 0.95 [0.86–0.99] | 0.78 [0.29–0.94] |
| Ppt | 0.27 | 0.79 | 0.16 | 0.32 |
| CV | 41% | 36% | 40% | 42% |
Abbreviations: CI, 95% confidence intervals; CMSG, corticomedullary sodium gradient; CV, coefficient of variation; Ppt paired t test p‐value; ρ, Lin's concordance correlation coefficient; (r), Pearson's correlation coefficient; (ri), two‐way mixed single measure correlation coefficient (ICC(3,1)); TLCO, 12 layer concentric objects.
FIGURE 2.

Concordance plots of inter‐reader (A) and intra‐reader (B) analysis for corticomedullary sodium gradient (CMSG) determined with the twelve‐layer concentric objects (TLCO) (blue dots) and line method evaluation (orange crosses).
4. DISCUSSION
The use of the TLCO algorithm enabled the successful determination of a linear aTSC concentration gradient from the cortex to the medulla in all subjects. A similar linear increase in aTSC was also reported in the work of Grist et al. 34 The article did not provide the slope of the gradient, which precludes a comparison of the CMSG. Because the results of the two segmentation methods in our study do not differ significantly, the results can be compared with previously published CMSGs determined by the Line method and the application of the TLCO method is possible for future studies. However, it is recommended that the TLCO method be used in future CMSG evaluations, as the intra‐ and inter‐reader results demonstrate enhanced reliability, precision, reproducibility and concordance in comparison to the Line method.
It has been shown by Moon et al. 12 that CMSG differs between transplanted and native kidneys. As demonstrated by Haneder et al., 35 the CMSG varies depending on the specific radiotherapeutic approach used. This suggests that the CMSG could potentially be used to evaluate not only the physiology but also the efficacy of different radiotherapeutic modalities in the future. However, to enable the clinical applicability of renal 23Na imaging, it is essential to use a robust and objective methodology for the assessment of CMSG. Our results indicate that the TLCO method is more appropriate for establishing CMSG as a clinical parameter than the Line method.
The TLCO segmentation of the first radiologist in the second segmentation shows the lowest standard deviation (see Table 1, [6.7 ± 2.6] mM/mm). This result is lower than the published results by Moon et al., 12 which found a CMSG of (10.5 ± 0.9) mM/mm in native kidneys and (8.9 ± 1.5) mM/mm in transplanted kidneys with normal function. Our CMSG results are increased compared to the results of Haneder et al., 14 who studied the CMSG under the influence of water loading using Line segmentation methods. They determined a CMSG of (2.56 ± 0.38) mM/mm for pre‐waterload and (3.38 ± 0.35) mM/mm for post‐water load. The methodology between these published results and ours differs in that the CMSG determined by Haneder et al. 14 was only evaluated within the first 10 mm from the cortex, whereas we evaluated the CMSG up to the aTSC maximum, similar to the methodology of Moon et al. 12 (Table 1). The aTSC published by Moon et al. 12 (192.2 ± 9.6) mM are in good agreement with our aTSC results in the medulla ([202.3 ± 47.3] mM).
Our study has limitations that need to be mentioned when comparing our results with published studies. First, the aTSC maps depend on the individual hydration status of the test subjects. 14 , 18 In the present study, hydration monitoring was not performed to determine the aTSC maps used to calculate the CMSG. Nevertheless, it is anticipated that the performance of the two methods will not be influenced to a significant extent by differences in hydration. Second, the current MR acquisition time is long (˜45 min) because of coil change, volunteers positioning, shimming, and the current acquisition protocol, which makes additional relaxation time measurements for aTSC calculations and clinical application difficult to access. Third, we anticipate that partial volume artifacts will have an impact on the aTSC in our 23Na data acquired with a 3‐mm isotropic resolution of the 23Na images, given that they might influence the results. However, an improvement in resolution would be associated with a decline in signal accompanied by a loss of SNR, which is currently the limiting factor.
The statistical evaluation shows that the TLCO method should be preferred to the Line method in further CMSG studies, because of better reliability, precision, reproducibility, and concordance. The improvements of the TLCO method over the Line method have already been demonstrated in BOLD studies evaluating the corticomedullary gradient in renal oxygenation. 20 , 36 We acknowledge that integrating the 23Na protocol from this study into clinical routine is currently limited by time constraints. Our work highlights the advantages of the TLCO algorithm over the commonly used line algorithm. Advances in high‐field MRIs, shorter echo times, and AI‐driven developments are pushing 23Na imaging toward potential clinical applications. 15 Objective measurement methods are essential for clinical implementation. We have demonstrated the objectivity of the TLCO algorithm compared to the established line algorithm. This study, therefore, contributes to establishing CMSG with 23Na as a potential clinical parameter.
5. CONCLUSION
The CMSG results using the TLCO method are comparable to the Line method, but offer better reliability, reproducibility, and precision in intra‐ and inter‐reader analyses. Therefore, we recommend using the TLCO method in future CMSG studies.
ACKNOWLEDGMENTS
P.J.G. and R.M. are supported by the Jürgen Manchot Foundation, Düsseldorf, Germany. Open Access funding enabled and organized by Projekt DEAL.
Gallinnis PJ, Möller R, Ljimani A, et al. Comparison of twelve‐layer concentric object (TLCO) and Line segmentations methods to determine the renal corticomedullary sodium gradient (CMSG) with a 3 T MRI . Magn Reson Med. 2025;94:1318‐1325. doi: 10.1002/mrm.30525
REFERENCES
- 1. Bikbov B, Purcell CA, Levey AS, et al. Global, regional, and national burden of chronic kidney disease, 1990–2017: a systematic analysis for the global burden of disease study 2017. Lancet. 2020;395:709‐733. doi: 10.1016/s0140-6736(20)30045-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Francis A, Harhay MN, Ong ACM, et al. Chronic kidney disease and the global public health agenda: an international consensus. Nat Rev Nephrol. 2024;20:473‐485. doi: 10.1038/s41581-024-00820-6 [DOI] [PubMed] [Google Scholar]
- 3. Selby NM, Francis ST. Assessment of acute kidney injury using MRI. J Magn Reson Imaging. 2024;61:25‐41. doi: 10.1002/jmri.29281 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Bane O, Seeliger E, Cox E, et al. Renal MRI: from nephron to NMR signal. J Magn Reson Imaging. 2023;58:1660‐1679. doi: 10.1002/jmri.28828 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Stabinska J, Wittsack H, Lerman LO, Ljimani A, Sigmund EE. Probing renal microstructure and function with advanced diffusion MRI: concepts, applications, challenges, and future directions. J Magn Reson Imaging. 2023;60:1259‐1277. doi: 10.1002/jmri.29127 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Thiel TA, Schweitzer J, Xia T, et al. Evaluation of radiographic contrast‐induced nephropathy by functional diffusion weighted imaging. J Clin Med. 2021;10:4573. doi: 10.3390/jcm10194573 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Bechler E, Stabinska J, Thiel T, et al. Feasibility of quantitative susceptibility mapping (QSM) of the human kidney. Magn Reson Mater Phys Biol Med. 2020;34:389‐397. doi: 10.1007/s10334-020-00895-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Maril N, Rosen Y, Reynolds GH, Ivanishev A, Ngo L, Lenkinski RE. Sodium MRI of the human kidney at 3 tesla. Magn Reson Med. 2006;56:1229‐1234. doi: 10.1002/mrm.21031 [DOI] [PubMed] [Google Scholar]
- 9. Maril N, Margalit R, Rosen S, Heyman SN, Degani H. Detection of evolving acute tubular necrosis with renal 23Na MRI: studies in rats. Kidney Int. 2006;69:765‐768. doi: 10.1038/sj.ki.5000152 [DOI] [PubMed] [Google Scholar]
- 10. Laustsen C, Bøgh N. Sodium MRI of the renal corticomedullary gradient. Radiology. 2022;303:390‐391. doi: 10.1148/radiol.213007 [DOI] [PubMed] [Google Scholar]
- 11. Ladd ME, Bachert P, Meyerspeer M, et al. Pros and cons of ultra‐high‐field MRI/MRS for human application. Prog Nucl Magn Reson Spectrosc. 2018;109:1‐50. doi: 10.1016/j.pnmrs.2018.06.001 [DOI] [PubMed] [Google Scholar]
- 12. Moon CH, Furlan A, Kim JH, Zhao T, Shapiro R, Bae KT. Quantitative sodium MR imaging of native versus transplanted kidneys using a dual‐tuned proton/sodium (1H /23Na) coil: initial experience. Eur Radiol. 2014;24:1320‐1326. doi: 10.1007/s00330-014-3138-5 [DOI] [PubMed] [Google Scholar]
- 13. Haneder S, Kettnaker P, Konstandin S, et al. Quantitative in vivo 23Na MR imaging of the healthy human kidney: determination of physiological ranges at 3.0T with comparison to DWI and BOLD. Magn Reson Mater Phys Biol Med. 2013;26:501‐509. doi: 10.1007/s10334-013-0369-4 [DOI] [PubMed] [Google Scholar]
- 14. Haneder S, Konstandin S, Morelli JN, et al. Quantitative and qualitative 23Na MR imaging of the human kidneys at 3 T: before and after a water load. Radiology. 2011;260:857‐865. doi: 10.1148/radiol.11102263 [DOI] [PubMed] [Google Scholar]
- 15. Gast LV, Platt T, Nagel AM, Gerhalter T. Recent technical developments and clinical research applications of sodium (23Na) MRI. Prog Nucl Magn Reson Spectrosc. 2023;138–139:1‐51. doi: 10.1016/j.pnmrs.2023.04.002 [DOI] [PubMed] [Google Scholar]
- 16. Kamp B, Frenken M, Klein‐Schmeink L, et al. Evaluation of sodium relaxation times and concentrations in the Achilles tendon using MRI. Int J Mol Sci. 2022;23:10890. doi: 10.3390/ijms231810890 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Gast LV, Völker S, Utzschneider M, et al. Combined imaging of potassium and sodium in human skeletal muscle tissue at 7 T. Magn Reson Med. 2020;85:239‐253. doi: 10.1002/mrm.28428 [DOI] [PubMed] [Google Scholar]
- 18. Haneder S, Konstandin S, Morelli JN, Schad LR, Schoenberg SO, Michaely HJ. Assessment of the renal corticomedullary 23Na gradient using isotropic data sets. Acad Radiol. 2013;20:407‐413. doi: 10.1016/j.acra.2012.10.009 [DOI] [PubMed] [Google Scholar]
- 19. Haneder S, Juras V, Michaely HJ, et al. In vivo sodium (23Na) imaging of the human kidneys at 7 T: preliminary results. Eur Radiol. 2013;24:494‐501. doi: 10.1007/s00330-013-3032-6 [DOI] [PubMed] [Google Scholar]
- 20. Li LP, Milani B, Pruijm M, et al. Renal BOLD MRI in patients with chronic kidney disease: comparison of the semiautomated twelve layer concentric objects (TLCO) and manual ROI methods. Magn Reson Mater Phys Biol Med. 2019;33:113‐120. doi: 10.1007/s10334-019-00808-5 [DOI] [PubMed] [Google Scholar]
- 21. Nemirovsky DR, Gupta P, Hu S, Wong R, Thakor AS. Blood Oxygen Level‐Dependent (BOLD) MRI in Glomerular Disease. Transplantology. 2021;2:109‐117. doi: 10.3390/transplantology2020011 [DOI] [Google Scholar]
- 22. Nagel AM, Laun FB, Weber M, Matthies C, Semmler W, Schad LR. Sodium MRI using a density‐adapted 3D radial acquisition technique. Magn Reson Med. 2009;62:1565‐1573. doi: 10.1002/mrm.22157 [DOI] [PubMed] [Google Scholar]
- 23. Yushkevich PA, Piven J, Hazlett HC, et al. User‐guided 3D active contour segmentation of anatomical structures: significantly improved efficiency and reliability. Neuroimage. 2006;31:1116‐1128. doi: 10.1016/j.neuroimage.2006.01.015 [DOI] [PubMed] [Google Scholar]
- 24. Müller‐ Lutz A, Kamp B, Nagel AM, et al. Sodium MRI of human articular cartilage of the wrist: a feasibility study on a clinical 3T MRI scanner. Magn Reson Mater Phys Biol Med. 2020;34:241‐248. doi: 10.1007/s10334-020-00856-2 [DOI] [PubMed] [Google Scholar]
- 25. Madelin G, Regatte RR. Biomedical applications of sodium MRI in vivo. J Magn Reson Imaging. 2013;38:511‐529. doi: 10.1002/jmri.24168 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Maril N, Margalit R, Mispelter J, Degani H. Functional sodium magnetic resonance imaging of the intact rat kidney. Kidney Int. 2004;65:927‐935. doi: 10.1111/j.1523-1755.2004.00475.x [DOI] [PubMed] [Google Scholar]
- 27. Zöllner FG, Konstandin S, Lommen J, et al. Quantitative sodium MRI of kidney. NMR Biomed. 2015;29:197‐205. doi: 10.1002/nbm.3274 [DOI] [PubMed] [Google Scholar]
- 28. Virtanen P, Gommers R, Oliphant TE, et al. Scipy 1.0: fundamental algorithms for scientific computing in python. Nat Methods. 2020;17:261‐272. doi: 10.1038/s41592-019-0686-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. McBride G. A proposal for strength‐of‐agreement criteria for Lin's concordance correlation coefficient. NIWA Client Report: HAM2005‐062. 2005;45:307‐310. [Google Scholar]
- 30. Chan Y. Biostatistics 104: correlational analysis. Singapore Med J. 2004;44:614‐619. [PubMed] [Google Scholar]
- 31. Chai Y, Maes V, Boudali AM, Rackel B, Walter WL. Inadequate annotation and its impact on pelvic tilt measurement in clinical practice. J Clin Med. 2024;13:1394. doi: 10.3390/jcm13051394 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Vallat R. Pingouin: statistics in python. J Open Source Softw. 2018;3:1026. doi: 10.21105/joss.01026 [DOI] [Google Scholar]
- 33. Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med. 2016;15:155‐163. doi: 10.1016/j.jcm.2016.02.012 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34. Grist JT, Riemer F, Hansen ESS, et al. Visualization of sodium dynamics in the kidney by magnetic resonance imaging in a multi‐site study. Kidney Int. 2020;98:1174‐1178. doi: 10.1016/j.kint.2020.04.056 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Haneder S, Michaely HJ, Schoenberg SO, et al. Assessment of renal function after conformal radiotherapy and intensity‐modulated radiotherapy by functional 1H‐MRI and 23Na‐MRI. Strahlenther Onkol. 2012;188:1146‐1154. doi: 10.1007/s00066-012-0254-5 [DOI] [PubMed] [Google Scholar]
- 36. Milani B, Ansaloni A, Sousa‐Guimaraes S, et al. Reduction of cortical oxygenation in chronic kidney disease: evidence obtained with a new analysis method of blood oxygenation level‐dependent magnetic resonance imaging. Nephrol Dial Transplant. 2016;362:gfw362. doi: 10.1093/ndt/gfw362 [DOI] [PubMed] [Google Scholar]