As transcatheter aortic valve implantation (TAVI) is increasingly used in younger populations with fewer comorbidities, the long-term durability of transcatheter heart valves (THVs) has become a major issue. Each type of THV features distinct expansion mechanisms, stent frame designs, and leaflet configurations that may influence valve haemodynamic performance, leaflet mechanical stress, and long-term valve durability1. However, direct comparisons of durability beyond 5 years between balloon-expandable (BE) and self-expanding (SE) THVs remain scarce and largely indirect1. Nonetheless, current data suggest that SE THV devices with a supra-annular design may be associated with better durability compared to BE THV devices1,2.
The Myval/Octacor THV (Meril Life Sciences) is a new balloon-expandable valve that has recently demonstrated non-inferiority to both the SAPIEN 3 BE THV (Edwards Lifesciences) and the Evolut SE THV (Medtronic) at 30-day and 1-year follow-up in the head-to-head LANDMARK and COMPARE-TAVI randomised trials3,4,5.
In this issue of EuroIntervention, Jain et al6 assessed the 4-year durability of the Myval/Octacor THV. Although this study shares several limitations with previous studies, including (i) the competing risk of death was not accounted for and (ii) the study was not randomised, it has several important strengths. First, an independent echocardiography core lab was used. Second, bioprosthetic valve deterioration (BVD) and bioprosthetic valve failure (BVF) were defined according to Valve Academic Research Consortium-3 criteria. Lastly, the follow-up period was substantial, extending to 4 years.
A key finding of this study is that moderate or greater paravalvular leak (PVL) - previously identified as an independent predictor of mortality - was relatively frequent, occurring in 5.9% of patients at 30 days. This rate is notably higher than the 2-3% reported with the same device in two recent randomised trials (Figure 1A)3,4,5. While the authors suggest this may be partly due to implantation in patients with large annuli, the majority of cases occurred in patients receiving valves in the standard size range (23-26 mm) and rather suggest that the higher PVL rate is more likely related to the difference in the populations: selected patients in randomised controlled trials versus real-world all-comers populations. Similarly, Silva et al2 recently reported significantly higher rates of PVL with both the SAPIEN 3 BE THV and Evolut SE THV in real-world settings compared to those seen in the PARTNER (n=1,000; ClinicalTrials.gov: NCT02675114) EVOLUT Low Risk (n=1,468; NCT02207569) (Figure 1A).
Figure 1. Rates of aortic regurgitation, PPM, bioprosthetic valve deterioration and failure in the Myval balloon-expandable valve versus other types of transcatheter heart valves in randomised trials and registries.

aClinicalTrials.gov: NCT01314313; bNCT01240902; cNCT01586910. Moderate/severe postprocedural AR (A), severe PPM (B), bioprosthetic valve deterioration (BVD) stage 2/3 at 5 years (C), and bioprosthetic valve failure (BVF) at 5 years (D) in BE THV and SE THV in randomised trials and registries. AR: aortic regurgitation; BE: balloon-expandable; PPM: patient-prosthesis mismatch; SAVR: surgical aortic valve replacement; SE: self-expanding; TAVI: transcatheter aortic valve implantation; THV: transcatheter heart valve
More intriguing is the progressive increase in aortic regurgitation over time, reaching 9.2% at 4 years. This trend is consistent with findings from the COMPARE-TAVI trial, which showed a rise in aortic regurgitation from 2% at 30 days to 4% at 1 year with the Myval/Octacor THV5. However, further investigation is needed to clarify whether this progression results from worsening PVL or the development of new transvalvular regurgitation due to valve thrombosis or early structural valve deterioration (SVD).
While SAPIEN 3 BE THV devices have been associated with a relatively high incidence of moderate or severe patient-prosthesis mismatch (PPM), a notable finding of the present study is the low rate of severe PPM observed with the Myval/Octacor BE THV: 3.6% for moderate PPM and 2.0% for severe PPM (Figure 1B). As with other outcomes, this may be, at least in part, attributed to the inclusion of a substantial proportion of patients with very large annuli.
The main finding of this study is that BVD >=stage 2 at 4 years was high, reaching 10.4% - twice the rate reported at 5 years with the SAPIEN 3 BE THV and 5 times higher than with the Evolut SE THV1 (Figure 1C). This was primarily driven by stage 2 BVD, observed in 9.7% of patients, largely due to SVD, as no cases of endocarditis were reported and overt valve thrombosis occurred in only 0.5%. Despite the elevated incidence of >=stage 2 BVD likely related to SVD, the all-cause BVF rate remains relatively low at 3.3%, which is comparable to rates previously reported with the SAPIEN 3 BE THV and may be considered reassuring (Figure 1D).
Overall, the findings of the study by Jain et al6 have two major implications. First, they underscore the need for long-term outcome and durability data from prospective randomised trials directly comparing SE THVs and BE THVs. Notably, the SMART Trial (n=716; NCT04722250), the BEST trial (n=2,000; NCT05454150), the LANDMARK trial, and the COMPARE-TAVI trial, which are comparing the Evolut, SAPIEN 3, and Myval/Octacor THVs have all completed enrolment, and patients will be followed for up to 10 years. Second, these findings highlight the importance of continued research and development to enhance THV durability, particularly by targeting valve calcification, a key contributor to bioprosthetic valve degeneration.
Acknowledgments
Conflict of interest statement
P. Pibarot has received institutional funding from Edwards Lifesciences, Medtronic, Pi-Cardia, and Cardiac Success for echocardiography core laboratory analyses, blood biomarker analyses, and research studies in the field of interventional and pharmacological treatment of valvular heart diseases, for which he received no personal compensation. E. Van Belle has no conflicts of interest to declare.
Contributor Information
Philippe Pibarot, Department of Medicine, Institut Universitaire de Cardiologie et de Pneumologie de Québec, Laval University, Québec, QC, Canada.
Eric Van Belle, Department of Interventional Cardiology for Coronary, Valves and Structural Heart Diseases, Valve Research and Innovation Committee, Lille University Hospital, Lille University, Inserm, Institut Pasteur de Lille, U1011-EGID, Lille, France.
References
- Ueyama H, Kuno T, Takagi H, Kobayashi A, Misumida N, Pinto DS, Laham RJ, Baeza C, Kini A, Lerakis S, Latib A, Søndergaard L, Attizzani GF. Meta-Analysis Comparing Valve Durability Among Different Transcatheter and Surgical Aortic Valve Bioprosthesis. Am J Cardiol. 2021;158:104–11. doi: 10.1016/j.amjcard.2021.07.046. [DOI] [PubMed] [Google Scholar]
- Silva I, Alperi A, Muñoz A, Cheema A, Nombela L, Veiga-Fernandez G, Tay E, Urena M, Asmarats L, Del Trigo, Lim Y, Gutierrez Alonso, Regueiro A, Campelo-Parada F, Serra V, Del Val, Barbosa Ribeiro, Ternacle J, Vilalta V, Vidal P, Belahnech Y, Alfonso F, Nuche J, Rodes-Cabau J, Pibarot P. Incidence and Impact of Structural Valve Deterioration Following TAVI: a Multicenter Real-World Study. Eur Heart J Cardiovasc Imaging. 2025 doi: 10.1093/ehjci/jeaf083. [DOI] [PubMed] [Google Scholar]
- Baumbach A, van Royen, Amat-Santos IJ, Hudec M, Bunc M, Ijsselmuiden A, Laanmets P, Unic D, Merkely B, Hermanides RS, Ninios V, Protasiewicz M, Rensing BJWM, Martin PL, Feres F, De Sousa, van Belle, Linke A, Ielasi A, Montorfano M, Webster M, Toutouzas K, Teiger E, Bedogni F, Voskuil M, Pan M, Angerås O, Kim WK, Rothe J, Kristić I, Peral V, Garg S, Elzomor H, Tobe A, Morice MC, Onuma Y, Soliman O, Serruys PW LANDMARK trial investigators. LANDMARK comparison of early outcomes of newer-generation Myval transcatheter heart valve series with contemporary valves (Sapien and Evolut) in real-world individuals with severe symptomatic native aortic stenosis: a randomised non-inferiority trial. Lancet. 2024;403:2695–708. doi: 10.1016/S0140-6736(24)00821-3. [DOI] [PubMed] [Google Scholar]
- Royen NV, Amat-Santos IJ, Hudec M, Bunc M, Ijsselmuiden A, Laanmets P, Unic D, Merkely B, Hermanides RS, Ninios V, Protasiewicz M, Rensing BJWM, Martin PL, Feres F, Sousa M, Belle EV, Linke A, Ielasi A, Montorfano M, Webster M, Toutouzas K, Teiger E, Bedogni F, Voskuil M, Pan M, Angerås O, Kim WK, Rothe J, Kristić I, Peral V, Van den, Westermann D, Bellini B, Garcia-Gomez M, Tobe A, Tsai TY, Garg S, Thakkar A, Chandra U, Morice MC, Soliman O, Onuma Y, Serruys PW, Baumbach A. Early outcomes of the novel Myval THV series compared to SAPIEN THV series and Evolut THV series in individuals with severe aortic stenosis. EuroIntervention. 2025;21:e105–18. doi: 10.4244/EIJ-D-24-00951. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Terkelsen CJ, Freeman P, Dahl JS, Thim T, Nørgaard BL, Mogensen NSB, Tang M, Eftekhari A, Povlsen JA, Poulsen SH, Pedersen L, Hjort J, Ellert J, Christiansen EH, Sørensen HT, Nissen H. SAPIEN 3 versus Myval transcatheter heart valves for transcatheter aortic valve implantation (COMPARE-TAVI 1): a multicentre, randomised, non-inferiority trial. Lancet. 2025;405:1362–72. doi: 10.1016/S0140-6736(25)00106-0. [DOI] [PubMed] [Google Scholar]
- Jain A, Jose J, Montorfano M, Nissen H, Martin P, Seth A, Stambuk K, Gunasekaran S, Abdurashid M, García-Gómez M, Fernandez-Cordón C, Rodriguez M, Jorgensen MD, Campo A, Serrador A, Vallinas-Hernandez S, Carrasco-Moraleja M, San Román A, Amat-Santos IJ. Four-year durability of the Myval balloon-expandable transcatheter aortic valve. EuroIntervention. 2025;21:e758–65. doi: 10.4244/EIJ-D-24-01112. [DOI] [PMC free article] [PubMed] [Google Scholar]