Abstract
Propose
Proton therapy of esophageal cancer is beneficial to spare normal tissue in clinical practice. However, intra-fractional and inter-fractional variance of tumor motion during treatment may compromise target coverage. The purpose of this study was to investigate the interplay effect due to intra-fractional motion and the effect of the robust optimization parameters for inter-fractional motion in the intensity modulation proton therapy (IMPT).
Materials and methods
This study retrospectively analyzed 42 patients with esophageal cancer treated at Shandong Cancer Hospital. The patients were divided into two groups. Twenty-one patients had a 4DCT image with 10 respiratory phases reconstructed (Gintra). In addition, twenty-one patients underwent a second 3D CT scan following the initial one (Ginter). The RayStation11B treatment planning system was used to create the IMPT plans for these two groups of patients. All patients were planned with a prescribed dose of 50.4 Gy (RBE) in 28 fractions for clinical target volume (CTV). 4D dynamic dose (4DDD) was then calculated to assess the interplay effect by considering respiratory motion and dynamic beam delivery for Gintra. Seven IMPT plans with different robust optimization parameters were designed for the 21 Ginter patient. The setup uncertainties were set to ± 0–6 mm for Ginter. Plan quality and dose-volume histogram (DVH) parameters for the target and organs at risk (OARs) were then analyzed.
Results
For Gintra, 4DDD was slightly perturbated compared to the nominal plan dose. The mean value of CTV D98% of nominal dose and 4DDD were 49.9 and 49.1 Gy (RBE), respectively, and the CTV D95% were 50.4 and 49.9 Gy (RBE), respectively. For Ginter, the DVH parameters of the target area and the OARs showed a linear relationship with the corresponding robust optimization parameters in IMPT. When the robust optimization parameters were set with a larger value, the dose coverage of the target was improved. However, the dose of OARs increased at the same time. The D98% of the target for the seven plans (setup6, setup5, setup4, setup3, setup2, setup1, setup0 plan) were 49.42 ± 0.75, 48.95 ± 1.21, 48.54 ± 1.48, 47.55 ± 2.31, 47.07 ± 2.71, 44.58 ± 4.20 and 44.02 ± 4.44 Gy(RBE), respectively. The Dmean of heart were 12.18 ± 3.05, 11.28 ± 2.86, 10.90 ± 2.77, 9.76 ± 2.39, 9.73 ± 2.39, 8.33 ± 2.22 and 8.22 ± 2.20 Gy (RBE), respectively.
Conclusion
In this study, the differences in dose distributions between the 4DDD and nominal plans for Gintra can be attributed to the interplay effects. While target coverage remained stable, variations in OAR doses should be evaluated for Gintra. For Ginter, smaller setup uncertainty parameters may not fully mitigate inter-fractional tumor motion, leading to greater variation in target dose and potential inadequate coverage. The linear relationship between setup uncertainty and D98% suggested that improved setup uncertainties can enhance target coverage, while higher setup uncertainties tend to increase OARs doses, particularly to the heart and lungs.
Keywords: Esophageal cancer, Inter-fractional motion, 4D dynamic dose, Intensity modulated proton therapy, Interplay effect
Introduction
Esophageal cancer is a common malignancy of the upper digestive tract with a 5-year survival rate of less than 40% [1, 2]. At present, surgery is the preferred treatment method for early esophageal cancer. However, it is not easy to detect early esophageal cancer, resulting in an inoperable condition for most patients in clinical practice. For patients with advanced esophageal cancer who are not suitable for surgery, chemoradiotherapy is the first choice [3]. However, for esophageal cancer patients with the tumor close to critical organs, it is challenging to balance the delivery of tumoricidal dose and the protection of normal tissue [4].
Proton therapy based on active pencil beam scanning (PBS) has obvious advantages in protecting normal tissue and maintaining dose coverage of the target [5]. Several ongoing clinical trials are investigating its efficacy and safety. For example, NCT01512589 compares the toxicity and progression-free survival between proton beam therapy (PBT) and intensity-modulated radiation therapy (IMRT) to assess the clinical benefits of PBT. NCT05055648 is evaluating the combination of proton therapy and chemotherapy in locally advanced esophageal cancer, aiming to enhance treatment outcomes. Additionally, NCT03801876 is exploring the use of PBT in the neoadjuvant setting, focusing on subsequent surgical outcomes. These studies demonstrate the potential of proton therapy to enhance tumor control and reduce toxicity. However, this indirectly underscores the significant challenge posed by intra- and inter-fraction tumor motion, which remains a key consideration in optimizing proton therapy for esophageal cancer.
Nevertheless, pencil beam scanning proton therapy suffers from greater sensitivity to proton beam range and patient setup uncertainties [6]. Furthermore, due to the esophagus being in close anatomical proximity to the diaphragm, stomach, and other motion organs, the patient’s respiratory movements can cause the esophagus to move. During the intra-fractional treatment process, the motion of the esophageal target, combined with the dynamic delivery of the proton beam, may result in interplay effects. The interplay effect is considered a source of random error. This type of deviation arises from patient-specific random motion or periodic motion such as respiration [7, 8]. Several strategies have been proposed to mitigate the interplay effect, but each has its own limitations. By expanding the margin, range-adapted internal target volume (ITV) could address the respiratory motion [9]. Breath-hold and abdominal compression techniques can reduce the amplitude of motion but decrease patient discomfort during treatments [10]. Breath hold, Gating, and repainting can improve dose distribution at the expense of a longer treatment time [11, 12].
Similarly, inter-fraction motion is the motion seen between images taken on different treatment fractions and has both systematic and random components [13]. The mechanical shortcomings (such as laser misalignment), patient-related factors (like consistent displacement of skin marks) and fixation-related issues may contribute to the errors between inter-fractional treatment [7]. Robust optimization as a way to reduce uncertainty has been widely accepted in clinical practice [14–16]. However, more normal tissue around tumor target will get irradiated as setup and range uncertainties are incorporated in the robust optimization to mitigate the setup and range uncertainties during treatment. Although the development of adaptive proton therapy would be a promising approach to account for the actual errors occur during daily treatment, the use of a suitable robust optimization parameter is beneficial to the robust optimization of esophageal cancer IMPT for the current practice. It can prevent normal tissues from receiving excessive unnecessary irradiation.
The purpose of this study was to investigate the impact of the interplay effect in intra-fractional treatment and assess the target movement between inter-fractional treatment. 33 patients with esophageal cancer treated at Shandong Cancer Hospital were retrospectively analyzed. The patients were divided into two groups. For the intra-fraction group, 4D dynamic dose (4DDD) was calculated to assess the interplay effect by considering respiratory motion and dynamic beam delivery. For the inter-fraction group, IMPT plans with different robust optimization parameters were designed for each patient. The effects of various robust optimization parameters on the alleviation of uncertainty and the protection of OARs were analyzed based on the doses of the target and OARs. This study aims to provide a systematic evaluation for the interplay effect in intra-fractional treatment and the dose variation caused by inter-fractional anatomy for esophageal cancer IMPT.
Materials and methods
Patients and tumor characteristics
This study retrospectively analyzed 42 patients with esophageal cancer treated at Shandong Cancer Hospital. The inclusion criteria were as follows: (1) Patients with unresected esophageal cancer; (2) No previous history of radiotherapy or chemotherapy; (3) No known history of pulmonary or cardiac diseases. The exclusion criteria were as follows: (1) Change of radiotherapy regimen during radiotherapy; (2) Complicated with other malignant tumors.
For each patient, a single CT scan (Brilliance Big Bore, Philips, Netherlands) was performed without breath hold in headfirst supine (HFS) with 3 mm slice thickness. Patients were immobilized using a vacuum cushion (Tengfei Yu Technology Co., Ltd, Shenzhen, China) to ensure reproducibility during CT simulation and treatment delivery. The patients were divided into two groups. Twenty-one patients had a 4D CT image set reconstructed into 10 respiratory phases. The 50% phase was defined as the end of expiratory breathing phase (denoted 0%Ex) which served as the reference phase in this work [17]. Twenty-one patients underwent a second 3D CT scan during the treatment after the first CT scan. The 21 patients were classified as intra-fractional group (Gintra) and 21 patients as inter-fractional group (Ginter).
The gross target volume (GTV), heart, lungs were contoured on all ten phases of 4D CT in the RayStation11B(RaySearch Laboratories AB, Stockholm, Sweden)treatment planning system by the certificated radiation oncologists in our center for this study. The clinical target volume (CTV) was generated by adding a 3 cm longitudinal expansion and 1 cm radial expansion of the GTV on the reference phase [18, 19]. Taking the 50% phase (the end of expiratory breathing phase) as the reference phase, the CTV volume for the Gintra was 130.6–881.9 cc (mean volume 299.5 cc), the CTV volume for the Ginter was 230.2–372.1 cc (mean volume 302.1 cc) during the first scan and 236.9–338.5 cc (mean volume 299.1 cc) during the second scan. Table 1 summarizes patient number, tumor volume, location, motion amplitude (Gintra), and CT interval (Ginter). In this study, tumor motion amplitude was defined as the maximum displacement of the center of the tumor mass in the three coordinate directions throughout all the phases of the 4D CT dataset (S-I, A-P, and R-L).
Table 1.
Patient characterizes for Gintra(patient number, tumor volume, location, motion amplitude) and Ginter(patient number, tumor volume, CT interval)
| Patient Group | Gintra | Ginter | ||||||
|---|---|---|---|---|---|---|---|---|
| Patient Number | Tumor Volume(cc) | Tumor Location | Tumor motion amplitude (mm) | Tumor motion amplitude (mm) | CT interval | Tumor Volume(cc) | ||
| R-L (mm) | I-S (mm) | P-A (mm) | ||||||
| 1 | 357.7 | Cervical | 0.3 | 0.5 | 0.2 | 0.6 | 31d | 246.6/251.8 |
| 2 | 215.7 | Entire esophagus | 0.3 | 1.2 | 0.1 | 1.2 | 28d | 356.3/328.7 |
| 3 | 264.9 | Entire esophagus | 0.1 | 0.4 | 0.4 | 0.6 | 29d | 333.3/310.9 |
| 4 | 196.7 | Cervical | 0.3 | 1.5 | 0.4 | 1.6 | 42d | 305.3/301.6 |
| 5 | 244.5 | Entire esophagus | 0.2 | 0.1 | 0.1 | 0.2 | 21d | 311.4/320.7 |
| 6 | 223.7 | Thoracic | 0.3 | 2.4 | 0.3 | 2.4 | 33d | 230.2/251.6 |
| 7 | 280.5 | Entire esophagus | 0.6 | 2.1 | 0.3 | 2.2 | 41d | 330.1/301.3 |
| 8 | 881.9 | Thoracic | 1.0 | 3.9 | 1.2 | 4.2 | 14d | 284.8/306.6 |
| 9 | 417.0 | Thoracic | 1.2 | 2.4 | 1.2 | 2.9 | 30d | 303.4/307.2 |
| 10 | 130.6 | Thoracic | 0.2 | 3.7 | 1.2 | 3.9 | 38d | 372.1/336.8 |
| 11 | 365.1 | Cervical | 0.2 | 0.6 | 0.2 | 0.7 | 24d | 296.0/299.7 |
| 12 | 589.9 | Entire esophagus | 2.6 | 4.3 | 2.0 | 5.4 | 15d | 287.1/298.6 |
| 13 | 133.76 | Lower esophagus | 0.8 | 5.1 | 1.4 | 5.3 | 15d | 297.7/285.3 |
| 14 | 401.55 | Lower esophagus | 1.1 | 7.6 | 1.5 | 7.8 | 32d | 287.8/311.7 |
| 15 | 326.31 | Entire esophagus | 0.9 | 1.1 | 1.2 | 1.9 | 35d | 319.8/311.0 |
| 16 | 300.65 | Lower esophagus | 0.6 | 2.6 | 0.8 | 2.7 | 32d | 256.3/236.9 |
| 17 | 245.36 | Thoracic | 0.5 | 5.6 | 0.7 | 5.7 | 31d | 256.2/283.4 |
| 18 | 145.43 | Thoracic | 0.8 | 4.8 | 2.0 | 5.3 | 21d | 299.7/300.7 |
| 19 | 144.37 | Thoracic | 3.3 | 8.8 | 3.6 | 10.1 | 28d | 348.5/338.5 |
| 20 | 238.62 | Entire esophagus | 0.9 | 3.2 | 0.6 | 3.4 | 27d | 271.2/283.3 |
| 21 | 251.8 | Entire esophagus | 0.4 | 1.8 | 0.7 | 2.0 | 27d | 350.6/314.5 |
R-L Right-left, A-P Anterior–posterior, S-I Superior–inferior
Tumor Volume: primary scan time volume/second scan time volume
Treatment planning
The clinical beam model of the ProBeam (Varian Medical Systems, CA, USA) used in the Proton Center of Shandong Cancer Hospital, with the nominal energies of 70–240 MeV in spot scanning mode was implemented. The RayStation11B treatment planning system was used to create the IMPT plans for the two patient groups in this work. The Monte Carlo based algorithm was applied to calculate dose, and the dose grid used was 3 × 3 × 3 mm3. Two oblique posterior fields were adopted [20]. The relative biologic effectiveness (RBE) weighted dose was determined using a fixed RBE value of 1.1 for proton. A prescription dose of 50.4 Gy (RBE) for 28 fractions was administered to 95% volume of the CTV [21, 22]. Only Ginter patients were optimized with robust optimization.
Robust optimization
The plans were optimized using the multi-field optimization (MFO) method. The robust optimization functions were set as follows: The minimum dose of CTV was 50.4 Gy(RBE) with the weight of 100. The maximum dose of CTV was 53 Gy(RBE) with the weight of 100. The dose limits for OARs were as follows: total lung V20 < 25% (V20 defined as the normalized volume of total lung receiving 20 Gy(RBE)), V5 < 60% (V5 defined as the normalized volume of total lung receiving 5 Gy(RBE) < 60%), heart dose V30 < 40%, V40 < 30%, maximum spinal cord dose (Dmax) < 45 Gy(RBE) [22] and spinal cord (Dmax) was treated as D2% in this work [23].
To evaluate the impact of inter-fractional motion, seven IMPT plans with different robust optimization parameters were designed for each patient in Ginter, and the setup uncertainty was set to ± 0 mm (setup0 plan), ± 1 mm (setup1 plan), ± 2 mm (setup2 plan), ± 3 mm (setup3 plan), ± 4 mm (setup4 plan), ± 5 mm (setup5 plan), ± 6 mm (setup6 plan), range uncertainty was set to ± 3% for all seven plans [24]. In order to reflect the influence of the interplay effect, robust optimization for the IMPT plans of patients in Gintra were not adopted.
Dose calculation
In this study, the interplay effect was evaluated for patients in the Gintra, considering both the dynamic pencil beam delivery and respiratory motion. A proton therapy simulation script [25] was developed in RayStation using the Monte Carlo dose algorithm to calculate the 4D dose distribution (4DDD). The script treatment parameters include: (1) Energy layer switching time: 1 s; (2) Beam spot moving speed: 10 m/s; (3) The beam dose rate was 50,000 clinical-MU/min (approximately 800 MU/s). Total duration of respiratory cycle was assumed to be 5 s. The breathing pattern was represented as a sin curve. All phases were equally distributed in time throughout the respiratory cycle. Additionally, considering the effect of inter-fractional motion, the second CT was registered with the first CT using rigid registration of the bone structures. The proton plan with the first CT was then applied to the second CT set to evaluate the inter-fractional dose variations. A scheme for the process of these two types of dose calculation was shown in Fig. 1.
Fig. 1.
Flowchart for intra-fractional dose, i.e.4DDD calculation and inter-fractional dose calculation
Dose evaluation
The plan dose for each patient in the two groups was normalized so that 100% of the prescription dose covered 95% of the volume of CTV. All patients were evaluated using the same DVH parameters. For CTV, the dose metrics analyzed were D98% (prescription dose received by at least 98% of volume), D95%, D2%, and tumor control probability (TCP). The TCP calculation was based on the equivalent uniform dose (EUD) model. The equations were as follows:
![]() |
![]() |
where TCD50 (i.e., the tumor dose required to produce a 50% TCP), γ50 (i.e., the slope of dose response at 50% TCP) and a (i.e., a unitless parameter which describes the magnitude of the volume effect to tumor or the normal structure) were 49.09 Gy(RBE), 0.26, 0.3 [26, 27], respectively. vi is the relative volume related to dose voxel Di. For OARs, total lung V20, V5, mean heart dose (Dmean), maximum spinal cord dose (Dmax). The DVH parameters of the nominal plans and 4DDD were compared to analyze the interplay effect for tumor target and normal tissue. The DVH parameters of the seven plan strategies were linearly fitted to analyze the influence of different setup uncertainty parameters on the dose distribution of target and OARs.
The nominal plans were used as references in the Gintra, and the setup0 plan was used as reference in the Ginter. SPSS software was used for the statistical analysis. The Wilcoxon rank sum test was used to evaluate statistical differences between DVH parameters. A double-tailed p value < 0.05 was considered statistically significant.
Results
Impact of the intra-fractional motion
Figure 2 shows D98%, D95%, and D2% of CTV for the nominal plan and 4DDD of 21 patients in the intra-fraction group. Panel a shows that the nominal plan dose was slightly different from D98% in 4DDD. Specifically, the D98% in the 4DDD were marginally lower than the nominal plan dose (exception for one patient). In 4DDD, the mean value of D98% was 49.9 Gy(RBE). The D98% was49.1 Gy(RBE) of the nominal plan. Panel b shows that the nominal plan dose was almost the same as that of 4DDD. The mean value of D95% for nominal plan dose and 4DDD were 50.4 and 49.9 Gy(RBE), respectively. Among the 21 patients, the maximum disparity between the two type of dose was 1.23 Gy(RBE). The median of 4DDD was only 1% lower than nominal plan dose. Panel c shows D2% of the nominal plan and 4DDD. 4DDDs were mostly higher than the nominal plans. The median of D2% increased to 100.84% in 4DDD (compared to the nominal plan).
Fig. 2.
The DVH parameters of CTV D98%, D95%, and D2% for the nominal plans and 4DDDs of the intra-fraction group. The red dots represent the nominal plan dose, and the blue dots represent 4DDD
Figure 3 shows the ratios of 4DDD to the nominal plan for Dmax in the spinal cord, Dmean in the heart, V20 and V5 in total lungs. For the majority, 4DDD showed a higher dose for Dmean in the heart. The ratio was mostly above 99% (below 99% in three patients). The Dmax of the spinal cord was almost indistinguishable between the two plans, with the ratio above 99% (98.2%, 98.1%in two patients). The performances of V20 and V5 in total lungs were almost consistent. The difference in V20 was more pronounced than that in V5 between the two plans. The nominal plan for V5 differed slightly from the dose for 4DDD, with the ratio concentrated around 100%. In summary, 4DDDs were slightly smaller than the nominal plans for the total lung. The mean values of OARs (spinal cord, heart, V20 and V5 of total lung were 101.3 ± 2%, 99.62 ± 0.56%, 98.69 ± 1.96%, 100.44 ± 1.38%, respectively.
Fig. 3.
The ratios of 4DDD to the nominal plan for each patient’s OARs in Gintra
For the TCP of the CTV, the TCP in 4DDD was 60.99 ± 0.96%, which was similar to that in the nominal plan (61.22 ± 0.83%). There were statistically significant differences (p<0.05) between the two plans.
Impact of the inter-fractional motion
Figure 4 shows the DVH parameter of CTV in the Ginter after the dose calculation on the second CT image. Panel a represents the values of D98%. The smaller the setup uncertainty parameter, the dispersed the D98% values at each point were, and the greater the possibility of insufficient dose coverage of the target. There was a linear relationship between the decrease of setup uncertainty and the decrease of D98% of target. The fitting equation was
![]() |
Fig. 4.
Figure 4 This scatter plot shows the DVH parameters D98%, D95%, and D2% of CTV in twenty-one patients of the inter-fractional group. Each dot in the graph represents the patient’s dose of the target under the corresponding setup uncertainty parameter. Each point of the same color represents a single patient. The red line in the figure is a linear fit generated from the mean of these data (dark red is the 95% confidence interval estimate and light red is the 95% prediction interval estimate). The symbols above the dots are the results referring to the statistical significance analysis of the differences of the dose metrics (** p ≤ 0.001; * p ≤ 0.05; ns, p>0.05)
Where y represents the D98% of the target in unit of Gy(RBE), x represents the setup uncertainties in unit of mm. Panel b represents the values of D95%. Among the five plans with setup uncertainty of 2–6 mm, the D98% were more concentrated around 48 Gy(RBE) compared to the other two plans. This phenomenon suggested that when the setup uncertainty was set within the range of 2 to 6 mm, the probability of insufficient dose coverage may decrease. The fitting equation was
![]() |
Where y represents the D95% of the target unit of Gy(RBE), x represents the setup uncertainties in unit of mm. Panel c represents the value of D2%. It was shown that setup uncertainty parameter had little effect on the value of D2%. In most cases, the doses in the seven treatment plans ranged from 53 Gy(RBE) to 54 Gy(RBE). The fitting equation was.
![]() |
Where y represents the D2% of the target in unit of Gy(RBE), x represents the setup uncertainties in unit of mm. There were statistically significant differences (p<0.05) between the six plans and the setup0 plan for the D98%, D95%, and D2% of CTV.
Figure 5 shows the Dmean of the heart, Dmax of the spinal cord, V20 and V5 of total lungs for each case in the inter-fractional group. Panel a represents the average dose received by the heart. The increase of dose was generally consistent with the increase of setup uncertainty parameters. The fitting equation was
![]() |
Fig. 5.
Figure 5 The DVH parameters of the OARs (Dmax of the spinal cord, Dmean for the heart, V20 and V5 of total lungs) for each case in the inter-fractional group. Each point represents the dose of the OARs with the corresponding setup uncertainty plan. Each point of the same color represents a single patient. The red line in the figure represents the trend of the organ dose with the changing robust parameter settings (obtained by linear fitting of the average organ dose). The dark red is the 95% confidence interval estimate and light red is the 95% prediction interval estimate in the figure. The symbols above the dots are the results referring to the statistical significance analysis of the differences of the dose metrics (** p ≤ 0.001; * p ≤ 0.05; ns, p>0.05)
Where y represents the Dmean of the heart in unit of Gy(RBE), x represents the setup uncertainties in unit of mm. Panel b represents the maximum dose received by the spinal cord. In the twenty-one plans, the dose variance trend of the spinal cord under the seven parameter settings was not uniform for each plan. The dose was mostly concentrated between 40 and 45 Gy (RBE) for the seven setup uncertainties. The dose distribution in the spinal cord seemed to be random and no linear fit was performed. Panel c and d represent V20 and V5 of total lungs, respectively. The two DVH parameters were affected by setup uncertainty parameters in a consistent trend. The downward trends of the 21 patients were also similar. The fitting equation of V20 was
![]() |
Where y represents the V20 of the total lung in unit of Gy(RBE), x represents the setup uncertainties in unit of mm. The fitting equation of V5 was
![]() |
Where y represents the V5 of the total lung in unit of Gy(RBE), x represents the setup uncertainties in unit of mm. In general, when the setup uncertainty in the plan settings increase, the radiation dose received by the OARs may also increase. The six plans resulted in significantly (p < 0.05) higher average values than the setup0 plan for the heart, V20 and V5 of total lungs. However, there was no statistically significant difference (p > 0.05) for the spinal cord, except in the setup1 plan. An increase in the setup uncertainty parameter led to improved dose coverage of the target, resulting in a gradual rise in TCP values. The dosimetric parameters derived from DVH for all treatment plans were summarized in Table 2.
Table 2.
Summary of DVH parameters (mean ± standard deviation) for the seven plans in Ginter
| Ginter | D98% | D95% | D2% | TCP | Heart | SC | V20 | V5 |
|---|---|---|---|---|---|---|---|---|
| Setup0 plan | 44.02 ± 4.44 | 46.87 ± 3.00 | 53.49 ± 0.36 | 56.80 ± 2.98 | 8.22 ± 2.20 | 43.70 ± 2.88 | 11.90 ± 2.60 | 21.91 ± 3.80 |
| Setup1 plan | 44.58 ± 4.20 | 47.26 ± 2.65 | 53.68 ± 0.30 | 57.36 ± 2.51 | 8.33 ± 2.22 | 42.75 ± 5.40 | 12.20 ± 2.61 | 22.13 ± 3.61 |
| Setup2 plan | 47.07 ± 2.71 | 48.86 ± 1.53 | 53.18 ± 0.24 | 58.47 ± 1.63 | 9.73 ± 2.39 | 43.91 ± 2.40 | 15.10 ± 3.94 | 24.02 ± 5.43 |
| Setup3 plan | 47.55 ± 2.31 | 49.12 ± 1.27 | 53.26 ± 0.21 | 58.73 ± 1.31 | 9.76 ± 2.39 | 43.71 ± 2.21 | 14.89 ± 2.86 | 24.70 ± 4.23 |
| Setup4 plan | 48.54 ± 1.48 | 49.76 ± 0.94 | 52.86 ± 0.50 | 58.84 ± 0.96 | 10.9 ± 2.77 | 43.77 ± 1.96 | 15.71 ± 2.71 | 27.75 ± 4.52 |
| Setup5 plan | 48.95 ± 1.21 | 49.88 ± 0.53 | 52.93 ± 0.19 | 59.12 ± 0.81 | 11.28 ± 2.86 | 43.83 ± 1.82 | 16.28 ± 2.75 | 28.02 ± 4.71 |
| Setup6 plan | 49.42 ± 0.75 | 50.10 ± 0.29 | 52.87 ± 0.17 | 59.34 ± 0.75 | 12.18 ± 3.05 | 43.93 ± 1.59 | 18.08 ± 0.32 | 30.04 ± 5.67 |
SC Spinal cord, V20, V5 the V20, V5 of total lung
TCP tumor control probability
Units for D98%, D95%, D2%, Heart, and SC were Gy(RBE), TCP, V20 and V5 were given in percentages (%)
Figure 6 shows the dose distributions for one representative example of Ginter and one of Gintra. For the first part of the figure (Ginter), the dose coverage of the target area gradually decreased as the setup uncertainties were set smaller and smaller. For Gintra, the dose distributions were similar in both panels, 4DDD was sufficient for target.
Fig. 6.
The dose distributions of one representative example of Ginter and one of Gintra. The yellow lines indicated the outlines of the CTVs
Discussion
This study provides a dual-perspective evaluation of both intra- and inter-fractional motion in esophageal cancer proton therapy. A linear assessment of robust parameters under inter-fractional motion was performed. By incorporating TCP analysis, the biological impact of dose variations was further quantified. Among the 21 patients analyzed in the Gintra, the maximum discrepancies between the 4DDD and nominal doses were approximately 1.83 Gy(RBE) for D98% and 1.23 Gy(RBE) for D95%. These findings show that the interplay effect of intra-fractional motion has little effect on proton therapy for esophageal cancer in this work. The robust posterior-field distribution used in the IMPT strategies likely explains this result. In terms of anatomical structure, the diaphragm is attached to the lumbar spine via the left and the right crus tendons. This posterior fan-shaped area is relatively stationary with regard to the contracting and relaxing motion of the diaphragm. The posterior-field was not sensitive to changes in respiration and anatomy. Therefore, the plans were more robust [20, 28]. At the same time, the posterior fields may reduce the risk of developing radiation dermatitis [29].
Tumor motion amplitude in this study was defined as the largest motion in three directions (S-I, A-P, or R-L). The maximum motion of the center of mass for the CTV was measured and no baseline markers were implanted in this study. Liu et al. showed that the displacement range in the direction of S-I was the largest, which was consistent with the results measured in this work [30]. Furthermore, Seki et al. reported that tumors located in the lower thoracic esophagus exhibit significantly larger motion amplitudes compared to tumors in the upper or middle thoracic regions (p<0.017) [31]. Tumors in the lower esophagus showed more movement in the S-I and A-P direction than higher tumors [32]. Several studies have shown that tumor location is an important factor in radiotherapy planning for esophageal cancer. In this study, the measured tumor motion amplitudes were consistent with those reported in previous studies. The study by Møller et al. indicated that the respiratory motion depends on the position of the tumor and is largest in the distal part of the esophagus [20]. Similarly, our results showed statistically significant differences in tumor motion amplitudes among different tumor locations (p < 0.05).
Wang et al. have indicated that inter-fractional motion of the tumor occurs during radiation treatments for esophageal cancer, and the volume of the tumor also varies throughout the entire radiotherapy process. Therefore, for inter-fractional motion, the insufficient dose coverage might be caused by the setup errors and the change of tumor volume with the increasing fractions [33]. Our results showed that larger setup uncertainty parameters or adaptive planning is required as the anatomy changes significantly after ~ 20 fractions, which is well align with that reported in Wang’s publication. The use of conformal intensity-modulated radiation therapy (IMRT) or volumetrically modulated arc therapy (VMAT) decreases the normal tissue toxicity. However, the presence of motion may compromise the alignment of steep dose gradients, potentially affecting treatment efficacy [31]. In the process of robust optimization, the setting of robust parameters could ensure the dose coverage of the target area during beam delivery. However, to address setup and range uncertainties during treatment, incorporating these uncertainties into the optimization process may result in increased irradiation of normal tissues surrounding the tumor target. For esophageal cancer patient, in this study, the tumor size of half of patients had reduced. This alteration may result in additional irradiation of normal tissues during inter-fractional treatment. Therefore, it is meaningful to evaluate the robustness parameters in IMPT plans to avoid the excessive setup uncertainty causing more unnecessary damage.
As shown by literatures, the main reasons for inter-fractional changes are gastric filling, deformation of the target volume and changes of the diaphragm position [20]. These changes may result in underdosage of the target for both proton and photon treatments [34–36]. The adoption of posterior beams can reduce the effect of inter-fractional anatomical changes for proton therapy and are more robust to respiratory motion [37]. While the advantages of thoracic proton therapy is controversial [38], for esophageal cancer specifically, robustness of proton plans can be better than IMRT if posterior fields are adopted [39]. In clinical optimization of the IMPT, the physicist can choose the appropriate optimization parameters based on the location and size of the patient’s target area. The range of setup uncertainty used in our study could accommodate the inter-fractional motion amplitude of the majority of patients [40, 41]. As Warren et al. discussed that large setup and range uncertainties > 3.5% could compromise the robustness of the treatment plan [39]. To achieve the goal of CTV D95% ≥ 95% of the prescribed dose [42], a setup uncertainty of 3 mm is typically required. In addition, for the linear fitting function, its slopes were similar to the dose variance trend for most patients. The intercept varies according to the prescribed dose in IMPT plans. These quantitate results could be useful to balance the target coverage robustness and normal tissue sparing during proton treatment planning. The work published by Lin et al. presented the results of a randomized phase IIB study showing that the reduced doses to OARs translate to reduced toxicity for patients with locally advanced esophageal cancer [43]. Furthermore, the application of the method combining passive scattering and PBS techniques has been shown to translate to a reduction in toxicities for pre-operative patients in a large retrospective analysis [44]. More clinical trials and large retrospective analysis are needed in the era of robust treatment planning.
This study has certain limitations. The small number of cases in Ginter could not fully represent patients with extreme conditions. Patients of Ginter were scanned about 15 or 30 days apart, resulting in limited tumor variation. However, we strongly recommend that once an anatomical change is identified, the radiation therapist and physician team should launch a repositioning CT scan and evaluate whether a new treatment plan is needed. Currently, all cases were analyzed from a single center. In future studies, collaboration with multiple treatment centers could be considered. In this study, two CT images were used to simulate inter-fractional motion. Due to the limitation of the number of CT, the dose variance trend after multiple radiotherapy sessions could not be investigated. The fact that the recalculation has been performed on a second CT over the treatment course may not be sufficient to have a complete assessment of the robustness of the plan. Multiple CTs at different time points would be beneficial. In this study, breathing patterns were assumed to be regular and reproducible based on the 4D CT dataset. In reality, irregular variations in patients’ respiratory motion may occur during the treatment. 4D CT images may not fully capture all aspects of respiratory motion, which may impact the accuracy of 4D dose reconstruction. In subsequent studies, multiple sets of in-room CT images can be introduced for more fractional dose analysis. Future work could include the application of different RBE models to evaluate the effects of RBE uncertainties.
Conclusion
In this study, evaluation of the impact of intra-fractional and inter-fractional motion for a group of 42 esophageal cancer patients receiving IMPT was performed. For Gintra, the observed differences in dose distributions between the 4DDD and nominal plans, particularly in target coverage and OARs, can be attributed to the interplay effects. While target coverage remained largely consistent, the variations in OAR doses should be evaluated for Gintra. Furthermore, the results of Ginter suggest that smaller setup uncertainty parameters may not be sufficient to mitigate the effects of inter-fractional tumor motion, leading to greater variation in the target dose and potentially inadequate target coverage. The relationship between decreased setup uncertainty and reduced D98% of the target was linear, indicating that better setup precision may improve coverage. For OARs, higher setup uncertainties tended to increase the dose, particularly to the heart and lungs.
Acknowledgements
The authors would like to thank Dr. Xin Fan, Dr. Haijiao Shang, Dr. Weiliang Yan, Dr. Yunkai Zang from RaySearch Laboratories for their guides on the application of RayStation system.
Abbreviations
- IMPT
Intensity modulated proton therapy
- 4DDD
4D dynamic dose
- PBS
Pencil beam scanning
- CT
Computed Tomography
- GTV
Gross target volume
- CTV
Clinical tumor volume
- OARs
Organs at risk
- MC
Monte Carlo
- RBE
Relative biologic effectiveness
- HFS
Head First Supine
- R-L
Right-left
- A-P
Anterior–posterior
- S-I
Superior–inferior
- SFO
Single field optimization -
- DVH
Dose-volume histogram
- MFO
Multi-field optimization
- PBT
Proton beam therapy
- TCP
Tumor control probability
- IMRT
Intensity modulated radiation therapy
- VMAT
Volumetrically modulated arc therapy
Authors’ contributions
XYF participated in the design of the study, data collection and the main manuscript writing. STW, WJL and KNM involved in the acquisition of data and revising the manuscript. WH checked and revised the writing parts of the manuscript. YY and TYD participated in the revising of the manuscript critically for important intellectual content. All authors reviewed and approved the final manuscript.
Funding
This work was jointly funded by projects 12105160, 82072094 and 12275162 supported by National Natural Science Foundation of China; projects ZR2019LZL017 and ZR2021QA099 supported by Shandong Provincial Natural Science Foundation; project ts201712098 supported by Taishan Scholars Project of Shandong Province; project 2022B03019-5 supported by Xinjiang Uygur Autonomous Region key research and development project; Shandong Province Higher Education Institutions Youth Innovation Team Development Plan(2024KJJ013); Young Talent of Lifting engineering for Science and Technology in Shandong, China(SDAST2024QTB034) and project 2021HZ81 supported by Tumor precision radiotherapy peak plan.
Data availability
All data included in this study are available upon request by contact with the corresponding author.
Declarations
Ethics approval and consent to participate
This study was approved by the Research Ethics Board of the Shandong Cancer Hospital. Written informed consent was waived by the Institutional Review Board. This study adhered to the Declaration of Helsinki.
Consent for publication
Not applicable.
Competing interests
The authors declare no competing interests.
Footnotes
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Contributor Information
Yong Yin, Email: yinyongsd@126.com.
Tianyuan Dai, Email: daitianyuan1992@163.com.
References
- 1.Sung H, Ferlay J, Siegel RL, et al. Global cancer statistics 2020: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin. 2021;71(3):209–49. [DOI] [PubMed] [Google Scholar]
- 2.Wang K, Yuan S. Current status and prospect of particle therapy for esophageal cancer. Precis Radiat Oncol. 2024;8(2):92–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Teoh AYB, Chiu PWY, Yeung WK, et al. Long-term survival outcomes after definitive chemoradiation versus surgery in patients with resectable squamous carcinoma of the esophagus: results from a randomized controlled trial. Ann Oncol. 2013;24(1):165–71. [DOI] [PubMed] [Google Scholar]
- 4.Gharzai L, Verma V, Denniston KA, et al. Radiation therapy and cardiac death in long-term survivors of esophageal cancer: an analysis of the surveillance, epidemiology, and end result database. PLoS One. 2016;11(7):e0158916. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.LaRiviere MJ, Santos PMG, Hill-Kayser CE, et al. Proton therapy. Hematol Oncol Clin North Am. 2019. 10.1016/j.hoc.2019.08.006. [DOI] [PubMed] [Google Scholar]
- 6.Lomax AJ. Intensity modulated proton therapy and its sensitivity to treatment uncertainties 1: the potential effects of calculational uncertainties. Phys Med Biol. 2008;53(4):1027–42. [DOI] [PubMed] [Google Scholar]
- 7.Hurkmans CW, Remeijer P, Lebesque JV, et al. Set-up verification using portal imaging; review of current clinical practice. Radiother Oncol. 2001;58(2):105–20. [DOI] [PubMed] [Google Scholar]
- 8.Li H, Dong L, Bert C, et al. AAPM task group report 290: respiratory motion management for particle therapy. Med Phys. 2022;49(4):e50-81. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Kardar L, Li Y, Li X, et al. Evaluation and mitigation of the interplay effects of intensity modulated proton therapy for lung cancer in a clinical setting. Practical Radiation Oncol. 2014;4(6):e259–68. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Knopf A-C, Boye D, Lomax A, et al. Adequate margin definition for scanned particle therapy in the incidence of intrafractional motion. Phys Med Biol. 2013;58(17):6079–94. [DOI] [PubMed] [Google Scholar]
- 11.Ohara K, Okumura T, Akisada M, et al. Irradiation synchronized with respiration gate. Int J Radiat Oncol Biol Phys. 1989;17(4):853–7. [DOI] [PubMed] [Google Scholar]
- 12.Phillips MH, Pedroni E, Blattmann H, et al. Effects of respiratory motion on dose uniformity with a charged particle scanning method. Phys Med Biol. 1992;37(1):223–34. [DOI] [PubMed] [Google Scholar]
- 13.Michalski A, Atyeo J, Cox J, et al. Inter- and intra-fraction motion during radiation therapy to the whole breast in the supine position: a systematic review. J Med Imaging Radiat Oncol. 2012;56(5):499–509. [DOI] [PubMed] [Google Scholar]
- 14.Pflugfelder D, Wilkens JJ, Oelfke U. Worst case optimization: a method to account for uncertainties in the optimization of intensity modulated proton therapy. Phys Med Biol. 2008;53(6):1689–700. [DOI] [PubMed] [Google Scholar]
- 15.Unkelbach J, Bortfeld T, Martin BC, et al. Reducing the sensitivity of IMPT treatment plans to setup errors and range uncertainties via probabilistic treatment planning. Med Phys. 2009;36(1):149–63. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Fredriksson A, Forsgren A, Hårdemark B. Minimax optimization for handling range and setup uncertainties in proton therapy. Med Phys. 2011;38(3):1672–84. [DOI] [PubMed] [Google Scholar]
- 17.Liu W, Schild SE, Chang JY, et al. Exploratory study of 4D versus 3D robust optimization in intensity modulated proton therapy for lung cancer. Int J Radiat Oncol Biol Phys. 2016;95(1):523–33. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Bhangoo RS, DeWees TA, Yu NY, et al. Acute toxicities and short-term patient outcomes after intensity-modulated proton beam radiation therapy or intensity-modulated photon radiation therapy for esophageal carcinoma: a Mayo clinic experience. Adv Radiat Oncol. 2020;5(5):871–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Feng H, Shan J, Ashman JB, et al. Technical note: 4D robust optimization in small spot intensity-modulated proton therapy (IMPT) for distal esophageal carcinoma. Med Phys. 2021;48(8):4636–47. [DOI] [PubMed] [Google Scholar]
- 20.Møller DS, Poulsen PR, Hagner A, et al. Strategies for motion robust proton therapy with pencil beam scanning for esophageal cancer. Int J Radiat Oncol Biol Phys. 2021;111(2):539–48. [DOI] [PubMed] [Google Scholar]
- 21.Chun SG, Skinner HD, Minsky BD. Radiation therapy for locally advanced esophageal cancer. Surg Oncol Clin N Am. 2017;26(2):257–76. [DOI] [PubMed] [Google Scholar]
- 22.Zhang Y, Fan B, Sun T, et al. The feasibility of dose escalation using intensity-modulated radiotherapy (IMRT) and intensity-modulated proton therapy (IMPT) with FDG PET/CT guided in esophageal cancer. J Cancer Res Ther. 2022;18(5):1261–7. [DOI] [PubMed] [Google Scholar]
- 23.Hodapp N. Der ICRU-Report 83: verordnung, dokumentation und kommunikation der fluenzmodulierten photonenstrahlentherapie (IMRT). Strahlenther Onkol. 2012;188(1):97–100. [DOI] [PubMed] [Google Scholar]
- 24.Canters R, van der Klugt K, Trier Taasti V, et al. Robustness of intensity modulated proton treatment of esophageal cancer for anatomical changes and breathing motion. Radiother Oncol. 2024;198: 110409. [DOI] [PubMed] [Google Scholar]
- 25.Fan X, Wang S, Li W, et al. Effect of breathing phase number on the 4D robust optimization for pancreatic cancer intensity modulated proton therapy. BMC Cancer. 2024;24(1):1337. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Chen H, Huang Y, Wang H, et al. Dosimetric comparison and biological evaluation of fixed-jaw intensity-modulated radiation therapy for T-shaped esophageal cancer. Radiat Oncol. 2021;16(1):158. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Huang BT, Wu LL, Guo LJ, et al. Radiobiological evaluation of simultaneously dose-escalated versus non-escalated intensity-modulated radiation therapy for patients with upper thoracic esophageal cancer. Onco Targets Ther. 2017;10:2209–17. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Feng H, Sio TT, Rule WG, et al. Beam angle comparison for distal esophageal carcinoma patients treated with intensity-modulated proton therapy. J Appl Clin Med Phys. 2020;21(11):141–52. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Lin SH, Komaki R, Liao Z, et al. Proton beam therapy and concurrent chemotherapy for esophageal cancer. Int J Radiat Oncol Biol Phys. 2012;83(3):e345–51. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Liu F, Ng S, Huguet F, et al. Are fiducial markers useful surrogates when using respiratory gating to reduce motion of gastroesophageal junction tumors? Acta Oncol. 2016;55(8):1040–6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Sekii S, Ito Y, Harada K, et al. Intrafraction esophageal motion in patients with clinical T1N0 esophageal cancer. Rep Pract Oncol Radiother. 2018;23(5):398–401. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Lever FM, Lips IM, Crijns SP, et al. Quantification of esophageal tumor motion on cine-magnetic resonance imaging. Int J Radiat Oncol Biol Phys. 2014;88(2):419–24. [DOI] [PubMed] [Google Scholar]
- 33.Wang JZ, Li JB, Wang W, et al. Detection of interfraction displacement and volume variance during radiotherapy of primary thoracic esophageal cancer based on repeated four-dimensional CT scans. Radiat Oncol. 2013;8:224. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Møller DS, Alber M, Nordsmark M, et al. Validation of a robust strategy for proton spot scanning for oesophageal cancer in the presence of anatomical changes. Radiother Oncol. 2019;131:174–8. [DOI] [PubMed] [Google Scholar]
- 35.Nyeng TB, Nordsmark M, Hoffmann L. Dosimetric evaluation of anatomical changes during treatment to identify criteria for adaptive radiotherapy in oesophageal cancer patients. Acta Oncol. 2015;54(9):1467–73. [DOI] [PubMed] [Google Scholar]
- 36.Yamashita H, Haga A, Hayakawa Y, et al. Patient setup error and day-to-day esophageal motion error analyzed by cone-beam computed tomography in radiation therapy. Acta Oncol. 2010;49(4):485–90. [DOI] [PubMed] [Google Scholar]
- 37.Yu J, Zhang X, Liao L, et al. Motion-robust intensity-modulated proton therapy for distal esophageal cancer. Med Phys. 2016;43(3):1111–8. [DOI] [PubMed] [Google Scholar]
- 38.Liao Z, Gandhi SJ, Lin SH, et al. Does proton therapy offer demonstrable clinical advantages for treating thoracic tumors? Semin Radiat Oncol. 2018;28(2):114–24. [DOI] [PubMed] [Google Scholar]
- 39.Warren S, Partridge M, Bolsi A, et al. An analysis of plan robustness for esophageal tumors: comparing volumetric modulated arc therapy plans and spot scanning proton planning. Int J Radiat Oncol Biol Phys. 2016;95(1):199–207. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Doi Y, Murakami Y, Imano N, et al. Quantifying esophageal motion during free-breathing and breath-hold using fiducial markers in patients with early-stage esophageal cancer. PLoS One. 2018;13(6):e0198844. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Fukada J, Hanada T, Kawaguchi O, et al. Detection of esophageal fiducial marker displacement during radiation therapy with a 2-dimensional on-board imager: analysis of internal margin for esophageal cancer. Int J Radiat Oncol Biol Phys. 2013;85(4):991–8. [DOI] [PubMed] [Google Scholar]
- 42.Hallemeier CL, Merrell KW, Neben-Wittich MA, et al. A prospective pilot study of pencil beam scanning proton radiation therapy as a component of trimodality therapy for esophageal cancer. Adv Radiat Oncol. 2024;9(8): 101547. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Lin SH, Hobbs BP, Verma V, et al. Randomized phase IIB trial of proton beam therapy versus intensity-modulated radiation therapy for locally advanced esophageal cancer. J Clin Oncol. 2020;38(14):1569–79. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Thomas M, Defraene G, Lambrecht M, et al. NTCP model for postoperative complications and one-year mortality after trimodality treatment in oesophageal cancer. Radiother Oncol. 2019;141:33–40. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
All data included in this study are available upon request by contact with the corresponding author.