. 2025 Jul 11;70(14):145010. doi: 10.1088/1361-6560/ade843

Optimization-based image reconstruction regularized with inter-spectral structural similarity for limited-angle dual-energy cone-beam CT

Junbo Peng 1, Tonghe Wang 2, Huiqiao Xie 2, Richard L J Qiu 1, Chih-Wei Chang 1, Justin Roper 1, David S Yu 1, Xiangyang Tang 3, Xiaofeng Yang 1,*
PMCID: PMC12246749  PMID: 40562072

Abstract

Objective. Limited-angle dual-energy (DE) cone-beam CT (CBCT) is considered as a potential solution to achieve fast and low-dose DE imaging on current CBCT scanners without hardware modification. However, its clinical implementations are hindered by the challenging image reconstruction from limited-angle projections. While optimization-based and deep learning-based methods have been proposed for image reconstruction, their utilization is limited by the requirement for x-ray spectra measurement or paired datasets for model training. This work aims to facilitate the clinical applications of fast and low-dose DE-CBCT by developing a practical solution for image reconstruction in limited-angle DE-CBCT. Approach. An inter-spectral structural similarity-based regularization was integrated into the iterative image reconstruction in limited-angle DE-CBCT. By enforcing the similarity between the DE images, limited-angle artifacts were efficiently reduced in the reconstructed DECBCT images. The proposed method was evaluated using two physical phantoms and three digital phantoms, demonstrating its efficacy in quantitative DECBCT imaging. Main results. In all the studies, the proposed method achieves accurate image reconstruction without visible residual artifacts from limited-angle DE-CBCT projection data. In the digital phantom studies, the proposed method reduces the mean-absolute-error from 309/290 HU to 14/20 HU, increases the peak signal-to-noise ratio from 40/39 dB to 70/67 dB, and improves the structural similarity index measurement from 0.74/0.72–1.00/1.00. Significance. The proposed method can efficiently reduce limited-angle artifacts during the image reconstruction, enabling quantitative DE-CBCT with comparable data acquisition time and radiation dose to that of a single-energy scan on current onboard scanners without hardware modification. This work is of great clinical significance and can boost the clinical application of DE-CBCT in image-guided radiation therapy and surgical interventions.

Keywords: dual-energy CT, cone-beam CT, optimization, structural similarity

1. Introduction

Dual-energy (DE) CT, as an implementation of spectral CT (Tang 2023), has found a variety of clinical applications, including virtual non-contrast imaging (Graser et al 2009, Zhang et al 2010, Toepker et al 2014), virtual monochromatic imaging (Yu et al 2012), automatic bone removal (Buerke et al 2009), and iodine quantification (Chandarana et al 2011). On the other hand, cone-beam CT (CBCT), using a flat-panel detector (FPD), is a volumetric imaging technique that plays a vital role in image-guided radiation therapy (Jaffray et al 2002), surgery (Siewerdsen et al 2005), and interventional procedures (Maybody et al 2013). Recently, integrating DE imaging with CBCT has attracted increasing attention. It has shown promise in various applications, such as contrast-enhanced lesion detection (Roeske et al 2020), angiography (Müller et al 2016), material classification (Zbijewski et al 2014), and dose calculation for image-guided photon and proton therapy (Freixas 2017, Men et al 2017, Skaarup et al 2019).

However, the routine practice of DE-CBCT has been hindered due to the lack of a practical DE-CBCT implementation. Early efforts in DE-CBCT involved performing two separate scans at different tube voltages (Zbijewski et al 2014). Besides the excessive radiation dose compared to a single-energy scan, the utility of this data acquisition scheme is hindered by the motion-induced spatial misregistration between the two scans because of the much slower CBCT gantry rotation than that of diagnostic CT due to patient safety considerations (Cho et al 2020). To achieve DE-CBCT with data acquisition time and radiation dose comparable to the single-energy scans, techniques such as fast kVp-switching and dual-layer FPD have been investigated on CBCT systems (Müller et al 2016, Cassetta et al 2020, Shi et al 2020, Wang et al 2021). Nonetheless, the fast kVp-switching faces technical challenges due to instability in the rapid x-ray tube voltage modulation on current CBCT scanners (Lee et al 2017), and the dual-layer FPD suffers from limited DE spectral separation due to its physical structure (Shi et al 2020). In short, despite the success of these two schemes in diagnostic DECT scanners, both of them require sophisticated hardware upgrades on existing onboard CBCT scanners and have not been commercialized yet.

It has been an active research area to achieve low-dose single-scan DE imaging on current scanners without significant hardware modifications since the introduction of the slow kVp-switching technique with prior image constrained compressed sensing (PICCS)-based image reconstruction (Szczykutowicz and Chen 2010). From the data acquisition perspective, the reported methods can be divided into spatial and temporal downsampling schemes. The spatial downsampling strategies install a reciprocating or static beam filter on the source or detector side to selectively alter the spectrum of x-ray photons in specific paths, (Lee et al 2017, Petrongolo and Zhu 2018, Cho et al 2020, Peng 2022, Xiang et al 2023, Peng et al 2024a) generating spatially alternated high- and low-energy projection data in each projection view. Two common challenges of these methods are partial filtration in the penumbra region and increased noise after the beam filtration. The temporal downsampling methods acquire DE projection data with alternating angular distribution via a slow kVp-switching scheme or a fast-rotating filter installed on the source side (Szczykutowicz and Chen 2010, Jiang et al 2021, Cao et al 2022, Zhang et al 2023, Peng et al 2024b). Among all the proposed data acquisition schemes, limited-angle DE-CBCT is the most cost-effective and feasible-to-implement solution in terms of hardware, and it is readily available for existing scanners (Zhang et al 2023, Chen et al 2024, Peng et al 2024b). As a temporal downsampling strategy, limited-angle DE-CBCT requires no hardware modification and alters the tube voltage only once in the middle of the source trajectory, acquiring DE projection data within two complementary limited-angle arcs. Besides, the spectral separation between the DE projection data in this scheme can be higher than that of the beam filter-based solutions. In brief, limited-angle DE-CBCT is an optimal candidate for practical DE-CBCT solutions on existing scanners.

One of the major challenges of limited-angle DE-CBCT is reconstruction artifacts. Optimization-based reconstruction algorithms and deep learning models have been proposed to address this issue. The optimization-based methods exploit constraints on the image directional-total-variation (DTV) of virtual monochromatic or material-specific images during the material decomposition (Chen et al 2021, 2022, 2024, Zhang et al 2022). Such iterative material decomposition algorithms require accurate measurement or estimation of DE x-ray spectra, which is challenging on current CBCT scanners. Furthermore, the optimization in these algorithms is computationally intensive due to the integration of material decomposition. Ignoring imaging physics, deep learning models directly learn the mapping from the acquired data to the desired data via supervised learning. The data restoration can be performed in the projection domain or the image domain. The projection-domain model translates the under-sampled DE projection data to the full-sampled data, which are used for subsequent analytical reconstruction (Peng et al 2024b). The image-domain model translates the directly reconstructed DE-CBCT images contaminated by limited-angle artifacts to high-quality images without artifacts (Zhang et al 2023). Despite the state-of-the-art performance of the deep learning-based methods, their clinical potential is significantly limited by the requirement of paired limited-angle and full-sampled DE-CBCT datasets for the supervised model training, which are unavailable in real patient datasets. In short, the clinical application of limited-angle DE-CBCT is hindered by the lack of a practical solution to the challenging image reconstruction.

In this study, we propose an efficient optimization-based image reconstruction method for limited-angle DE-CBCT to facilitate the clinical application of quantitative DE-CBCT. Recognizing the fact that the complete structural information is preserved in the projection data and the DE images share the same anatomical structures, we introduce an inter-spectral structural similarity-based regularization term into the iterative image reconstruction in limited-angle DE-CBCT. We demonstrate the feasibility of the proposed image reconstruction algorithm and evaluate its performance via a study of four physical phantoms and a study of three digital phantoms.

2. Materials and methods

2.1. Principles of iterative CT reconstruction

Iterative image reconstruction methods formulate the CT projection as a discretized linear model (Hsieh 2003)

b=Fμ

where μ denotes the vectorized CT image, F is the system matrix representing the forward projection operator, and b indicates the acquired projection data. μ has a dimension of Nvox×1 where Nvox is the number of voxels in the CT image, F is in a size of Nvox×NvNd where Nv is the number of projection views and Nd is the number of detector pixels, and b has a dimension of NvNd×1.

With a Gaussian noise assumption on the projection data, the maximum likelihood estimation is equivalent to a least-square optimization problem

minμ{1NvNdFμb22+R(μ)}s.t. μ0

where 22 is the 2 norm measuring the Eculidean distance, and R() is an image prior-based regularization term. For example, the total variation (TV) norm, TV, is a common choice of R() based on the piece-wise constant property of CT images (Rudin et al 1992).

2.2. Inter-spectral structural similarity-based iterative image reconstruction for limited-angle DE-CBCT

In limited-angle DE-CBCT, the high-kVp projection data are acquired within the first half of the view angular range, and the low-kVp projection data is acquired within the other half. Thus, the data acquisition in limited-angle DE-CBCT can be formulated as

{bH=FHμHbL=FLμL

where μH,L are the DE-CBCT images, FH,L represent the forward projection matrices in the high- and low-kVp scans, and bH,L indicate the acquired high- and low-kVp projection data.

FH and FL are the top and bottom half of the system matrix in a conventional single-energy scan in equation (1), i.e.,

F=[FHFL].

The projection data bH,L at each x-ray spectrum are angular-limited, leading to severe limited-angle artifacts in the DE-CBCT images directly reconstructed from the acquired projection data.

The proposed iterative method for limited-angle DE-CBCT reconstruction is based on the fact that the complete anatomical information acquired in the mixed-spectra projection data and the anatomical structures are consistent between DE-CBCT images. Therefore, we introduce a structural similarity-based regularization term to suppress the limited-angle artifacts during the iterative reconstruction:

minμH,μL{1NvNd[FHμHbH22+FLμLbL22]+λ[1SSIM(μH,μL)]}s.t. μH,L0

where λ is the user-defined weighting factor that balances the tradeoff between reconstruction data fidelity and image regularization. In equation (5), SSIM(,) calculates the structural similarity index measurement (SSIM) between two input images by Wang et al (2004)

SSIM(μH,μL)=(2μHμL+C1)(2σH,L+C2)(μH2+μL2+C1)(σH2+σL2+C2)

where μH and μL are the mean values of DE-CBCT, σH and σL are the standard deviations of DE-CBCT, σH,L indicates the covariance between DE-CBCT, and C1,2 are two constants associated with the dynamic range of DE-CBCT.

The incorporation of the SSIM regularization is the key to the success of limited-angle artifact reduction in iterative DE-CBCT reconstruction, which is the major contribution of this work. Of note, conventional TV or DTV regularization can also be involved in the proposed iterative reconstruction for noise suppression.

2.3. Evaluation

The performance of the proposed iterative method for image reconstruction in limited-angle DE-CBCT has been evaluated using data from both physical and digital phantoms. The physical phantom data were acquired on the table-top CBCT system, and the digital phantom data were simulated using the DECT images acquired on a Siemens TwinBeam scanner.

2.3.1. Physical phantom study

Due to the difficulty of raw data acquisition in the commercial scanners, we collected the DE projection data of two physical phantoms from the table-top CBCT system.

A Catphan©700 phantom and an anthropomorphic head phantom were scanned on one system. The geometry of the system was designed to match that of a Varian On-Board Imager (OBI) CBCT scanner on the TrueBeam radiation therapy machine. The source-to-detector distance (SDD) and the source-to-axis distance (SAD) are 1500 mm and 1000 mm, respectively. The FPD consists of 1408 × 1408 pixels with a size of 0.308 × 0.308 mm2. A collimator was installed on the x-ray tube to suppress photon scattering. Therefore, only the central row of detector pixels was used for image reconstruction. In the low-energy data acquisition, the tube voltage was set to 125 kVp. In the high-energy data acquisition, the tube voltage was unchanged, but a 0.2 mm tin (Sn) filter was installed on the exit window of the x-ray tube, leading to a higher mean photon energy after the filtration. For both phantom scans, a total of 600 views were acquired over 360° angular range in these two scans. The limited-angle DE-CBCT data consists of 175 projection views out of 0° to 105° in the high-energy data and 175 views out of 105° to 210° in the low-energy data. The total angular range of 210° is based on the short scan protocol of a Varian OBI system (Parker 1982). The imaging parameters of these two phantom scans are summarized in table 1.

Table 1.

Imaging parameters in the limited-angle DE-CBCT scans of physical and digital phantoms.

Physical phantom study Digital phantom study
Catphan©700 Head phantom
High-energy spectrum 125 kVp + 0.2 mm Sn 120 kVp + 0.6 mm Sn
Low-energy spectrum 125 kVp 120 kVp + 0.05 mm Au
Reconstruction volume 512 × 512 × 5 256 × 256 × 150
Voxel size (mm3) 0.5 × 0.5 × 0.5 1.0 × 1.0 × 1.0
SDD/SAD (mm) 1500/1000
Detector dimension 1408 × 15 1024 × 768
Detector pixel size (mm2) 0.308 × 0.308 0.388 × 0.388
High-energy projection views 175 views from 0° to 105°
Low-energy projection views 175 views from 105° to 210°

Of note, in the scans of physical phantoms, the DE spectral separation was achieved via beam filtration instead of different tube voltages. This is because the collected experimental data were originally acquired for other studies. The spectral separation in the digital phantom study is similar because of the hardware design of the Siemens TwinBeam scanner. In this work, we re-sorted and reused these data, which effectively preserved our research purpose.

2.3.2. Digital phantom study

The proposed method was further evaluated on three digital anthropomorphic phantoms based on DECT images of three H&N patients acquired on a Siemens TwinBeam DECT scanner at our institute. The tube voltage of the scanner was fixed at 120 kVp. The high- and low-energy data were acquired using a 0.6 mm tin filter and a 0.05 mm gold (Au) filter, respectively. For each patient, a volume of 256 × 256 × 150 with a voxel size of 1 × 1 × 1 mm3 was selected for the simulated scan of limited-angle DE-CBCT.

First, the DECT images were converted from HU to linear attenuation coefficient (LAC) maps with the effective LACs of water at the high- and low-energy x-ray spectra, which were simulated using the Spektr toolbox (Punnoose et al 2016). Second, the limited-angle DE-CBCT scan was simulated based on the geometry of the OBI system in the physical phantom study, as listed in table 1. SDD and SID were 1500 and 1000 mm, and the FPD was 1024 × 768 with a pixel size of 0.388 × 0.388 mm2. The first 175 projections within 0° to 105° were simulated using the high-energy CBCT volume, and the second 175 projections within 105° to 210° were simulated using the low-energy CBCT volume. The simulated forward projection was implemented using Siddon’s ray-tracing algorithm (Siddon 1985).

2.3.3. Image quality metrics

In the physical phantom studies, the line profiles and HU measurements within regions-of-interest (ROIs) were used to evaluate the accuracy of DE-CBCT reconstruction. In addition, peak signal-to-noise ratio (PSNR) and SSIM were calculated to evaluate the distortion between the reconstructed images and the ground truth.

In the digital phantom studies, the accuracy of image reconstruction was evaluated using the mean absolute error (MAE) calculated within the body mask between the reconstructed and reference DE-CBCT images. The coherence between reconstructed and ground truth images was evaluated using PSNR and SSIM.

2.4. Comparison studies

The proposed method was compared with three other methods. The first directly performed Feldkamp–Davis–Kress (FDK) reconstruction using the limited-angle DE-CBCT projection data. The second one was DTV-regularized iterative reconstruction

minμH{1Nv,HNdFHμHbH22+ηiμHTV,i+ηjμHTV,j}minμL{1Nv,LNdFLμLbL22+ηiμLTV,i+ηjμLTV,j}s.t. μH,L0

where the subscripts i,j denote the directions for TV regularization, and ηi/j indicates the weighting factor for each DTV penalty. Of note, the implementation of DTV-regularized image reconstruction here was different from that in the series of works (Chen et al 2021, 2022, 2024, Zhang et al 2022), where the authors exploited the correlation between DE projection data via the material decomposition framework. In this work, the reconstructions of high- and low-energy images were independent, and only single-energy data was used for the reconstruction of each image.

The third comparison method was PICCS-based iterative reconstruction (Szczykutowicz and Chen 2010)

minμH{1Nv,HNdFHμHbH22+ημHTV+δμHμPTV}minμL{1Nv,LNdFLμLbL22+ημLTV+δμLμPTV}s.t. μH,L0

where μP denotes the prior image reconstructed from mixed-spectral projection data via the FDK algorithm. η and δ are weighting factors to balance the tradeoff between conventional and prior-based TV regularization terms.

2.5. Implementation details

All the experiments were implemented on an NVIDIA A100 GPU with a memory of 80 GB. The optimization of equation (5) was solved using an Adaptive Moment Estimation (Adam) algorithm built in the PyTorch framework (Kingma 2014). In other words, we did not design a specific optimization algorithm to solve the proposed optimization problem. The step size was set to 0.001 in the physical phantom experiments and 0.1 in the digital phantom experiments. The first-order decay rate was 0.9, and the second-order decay rate was 0.999. The weighting factor of the SSIM term (λ) was set to 0.1 in all the experiments. Empirically, the numbers of total iterations were set to 1200 in the physical phantom experiments and 200 in the digital phantom experiments to guarantee the convergence. In this work, the forward and backward CBCT projection operators were implemented using an open-sourced toolbox, LEAP (Kim and Champley 2023), which modeled the finite size of the voxel and detector pixel; very similar to the Separable Footprint method (Long et al 2010). In the implementation of the DTV method, the directions i,j were selected based on the limited-angle artifacts in the FDK results and were indicated in the figures. ηi was set to 0.5 and ηj was set to 0.1 in all the experiments. In the implementation of the PICCS method, η was set to 0.25 and δ was set to 0.75 in all the experiments. The GPU memory consumption was around 2 G in the physical phantom studies and about 25 G in the digital phantom studies. The reconstruction took 10 s in the physical phantom studies and 260 s in the digital phantom studies.

3. Results

3.1. Study of physical phantoms

Figure 1 summarizes reconstructed DE-CBCT images of the Catphan©700 phantom. The reference images were reconstructed from 360° projection data using the FDK algorithm. Obvious limited-angle artifacts appear in the results of FDK and DTV. Limited-angle artifacts are reduced in the images reconstructed using PICCS, but there are some residual artifacts along specific directions. The proposed method achieves efficient limited-angle artifact reduction, and no visible distortion is observed in the results.

Figure 1.

Figure 1.

Reconstructed limited-angle DE-CBCT images of the Catphan©700 phantom using different methods. Display window: [−500 500] HU. The arrows in the FDK results indicate the directions of DTV regularization. The dashed circle and four ROIs in the reference images are used to measure line profile and HU accuracy.

The 1D profiles along the dashed circle indicated in figure 1 are plotted in figure 2, where the results of the proposed method show the best coherence to the ground truth.

Figure 2.

Figure 2.

Line profiles along the dashed circle in figure 1.

Four uniform ROIs in the Catphan©700 phantom were selected to measure HU accuracy, as indicated in figure 1. The results are summarized in table 2, showing that the proposed method can achieve accurate image reconstruction in limited-angle DE-CBCT.

Table 2.

Mean and standard deviation of CT numbers (HU) measured within different ROIs which are indicated in figures 1 and 3.

Catphan©700 Head phantom
#1 #2 #3 #4 #1 #2
High-energy DTV 978 ± 10 −108 ± 5 173 ± 1 −872 ± 9 108 ± 8 179 ± 18
PICCS 965 ± 50 −157 ± 52 151 ± 52 −810 ± 47 108 ± 67 130 ± 59
Proposed 1001 ± 15 −137 ± 12 101 ± 13 −959 ± 11 −15 ± 11 212 ± 9
Reference 1003 ± 65 −136 ± 58 104 ± 70 −967 ± 65 −14 ± 67 209 ± 71

Low-energy DTV 1160 ± 7 −69 ± 7 156 ± 4 −862 ± 5 60 ± 16 240 ± 22
PICCS 1111 ± 50 −120 ± 51 170 ± 52 −780 ± 47 50 ± 65 182 ± 66
Proposed 1206 ± 16 −80 ± 9 172 ± 11 −953 ± 10 45 ± 8 302 ± 9
Reference 1207 ± 46 −75 ± 39 175 ± 44 −947 ± 36 41 ± 52 301 ± 53

Figure 3 summarizes the results of the anthropomorphic head phantom. Consistent with previous results, the proposed method achieves the best performance of limited-angle artifact reduction, while obvious artifacts are observed in the results of other methods.

Figure 3.

Figure 3.

Reconstructed limited-angle DE-CBCT images of the Catphan©700 phantom using different methods. Display window: [−500 500] HU. The arrows in the FDK results indicate the directions of DTV regularization. The dashed line and two ROIs in the reference images are used to measure line profile and HU accuracy.

Figure 4 plots the 1D profiles along the dashed line indicated in figure 3, where the proposed method shows the best coherence to the ground truth.

Figure 4.

Figure 4.

Line profiles along the dashed line in figure 3.

The ROIs for HU accuracy measurement are indicated in the reference low-energy images in figure 3, and the results are listed in table 2, showing great accuracy of reconstructed HU values using the proposed method.

Table 3 lists other quantitative metrics, including PSNR and SSIM, used in the study of physical phantom studies. The results of the proposed method show superior quantitative image quality over other results.

Table 3.

Quantitative metrics in the studies of physical phantom studies. The values are calculated between the reconstructed and reference images.

FDK DTV PICCS Proposed
Catphan©700 PSNR (dB) High-energy 39.34 53.29 57.56 60.46
Low-energy 39.20 53.51 57.29 62.30
SSIM High-energy 0.71 0.98 0.99 1.00
Low-energy 0.72 0.98 0.99 1.00
Head phantom PSNR (dB) High-energy 39.57 50.81 52.83 56.60
Low-energy 39.48 48.97 50.77 56.31
SSIM High-energy 0.71 0.98 0.98 1.00
Low-energy 0.71 0.97 0.97 1.00

3.2. Study of digital phantoms

Figures 57 summarize reconstructed DE-CBCT images in different views of the three digital phantoms. The results using FDK and DTV show severe limited-angle artifacts, while residual artifacts are observed in the PICCS’ results. The proposed method achieves efficient limited-angle artifact reduction, and no visible distortion is observed in the results.

Figure 5.

Figure 5.

Reconstructed limited-angle DE-CBCT images of the digital phantom #1 using different methods. Display window: [−500 500] HU. The arrows in the FDK results indicate the directions of DTV regularization. Zoom-in images of the dashed boxes are provided to better demonstrate the limited-angle artifacts.

Figure 7.

Figure 7.

Reconstructed limited-angle DE-CBCT images of the digital phantom #3 using different methods. Display window: [−500 500] HU. The directions of DTV regularization are identical to the digital phantom #1. Zoom-in images of the dashed boxes are provided to better demonstrate the limited-angle artifacts.

Figure 6.

Figure 6.

Reconstructed limited-angle DE-CBCT images of the digital phantom #2 using different methods. Display window: [−500 500] HU. The directions of DTV regularization are identical to the digital phantom #1. Zoom-in images of the dashed boxes are provided to better demonstrate the limited-angle artifacts.

Table 4 summarizes the quantitative analysis results of the digital phantom study. The MAE is computed within the body mask. All the metrics are calculated for each slice, and the means and standard deviations are computed from all the 450 values.

Table 4.

Quantitative analysis of the digital phantom study. The results are calculated between the reconstructed and reference images.

FDK DTV PICCS Proposed
MAE (HU) High-energy 308.51 ± 42.2 113.01 ± 26.01 29.49 ± 7.15 14.42 ± 2.37
Low-energy 289.69 ± 71.2 193.17 ± 40.46 45.26 ± 10.45 19.94 ± 3.16
PSNR (dB) High-energy 40.20 ± 0.74 51.64 ± 1.30 65.04 ± 1.14 69.88 ± 0.96
Low-energy 39.15 ± 0.72 47.73 ± 1.54 60.95 ± 1.14 66.93 ± 0.76
SSIM High-ENERGY 0.74 ± 0.02 0.98 ± 0.00 1.00 ± 0.00 1.00 ± 0.00
Low-energy 0.72 ± 0.02 0.97 ± 0.01 0.99 ± 0.00 1.00 ± 0.00

The loss-iteration curves are plotted in figure 8, showing that the Adam reliably converges to high-quality solutions in all the experiments. We also ran Adam from three different initializations—(a) all-zero images, (b) all-0.02 (mm−1) images, and (c) all-0.04 (mm−1) images for all five experiments. The initial values were around the LACs range of CT images. As shown in figure 8, Adam solver consistently converges to the same neighborhood of solutions, i.e. it is robust to initialization.

Figure 8.

Figure 8.

The convergence curve of all the experiments.

4. Discussion

This work presents a practical solution to image reconstruction in limited-angle DE-CBCT. By enforcing the inter-spectral structural similarity constraint between DE-CBCT images during the optimization-based image reconstruction, limited-angle artifacts are efficiently reduced in the reconstructed images. To the best of our knowledge, this is the first successful attempt in optimization-based limited-angle DE-CBCT reconstruction in the image domain.

Optimization-based image reconstruction algorithms, such as PICCS and DTV, have been introduced to relax the data acquisition requirement of DECT for a long time and have achieved great success. The main idea of PICCS-based DECT methods is to obtain a prior image with accurate anatomical structures and enforce the constraint in similarity between the target image and the prior image during the subsequent iterative reconstruction. In Szczykutowicz and Chen (2010), a prior spectrally average CT is reconstructed from the slow kVp-switching projection data and used for target DECT reconstruction from sparse-view data. In Wang and Zhu (2016), a prior high-energy CT is reconstructed from full-sampled data and used for similarity-based regularization in the target low-energy CT reconstruction from sparse-view data. However, PICCS may not be the optimal strategy for image reconstruction with the data acquisition scheme in limited-angle DE-CBCT. In a series of works, DTV has been integrated into the material decomposition framework to achieve accurate DECT reconstruction using convex or nonconvex optimization (Chen et al 2021, 2022, 2024, Zhang et al 2022). In these works, the correlation between DE projection data is fully exploited via the material decomposition model, and the DTV regularization is enforced on the virtual monoenergetic or material-specific images. However, the implementation of such strategies may be challenging in limited-angle DE-CBCT due to the measurement of x-ray spectra and the intensive computation required. In short, both PICCS and DTV are existing arts that have significant impacts in this field. The proposed method is a fundamentally different strategy designed for efficient image reconstruction in limited-angle DE-CBCT, which may bring added value to this field.

The introduced inter-spectral SSIM regularization is a non-convex function, and thus, the total objective function of the DE-CBCT reconstruction problem is no longer convex. Theoretically, there is no optimization algorithm that guarantees the convergence to the global minimum for a non-convex optimization. In practice, the Adam solver has been popularly used in the optimization of highly-non-convex deep neural networks and demonstrated good property in convergence (Kingma 2014, Reddi et al 2019). In addition, the Adam solver has been used to solve non-convex material decomposition in DECT (Jiang et al 2024). In our work, the Adam solver consistently converges at high-quality results (images) and is robust over the initialization in all experiments.

The contamination by the Compton scattering is one of the primary challenges in CBCT, resulting in altered projection data and artifacts in reconstructed images. The Compton scattering is excluded in the simulation study because the objective of this study is focused on image formation, rather than scatter correction, in DE-CBCT. We will incorporate of the Compton scattering correction into the DE-CBCT reconstruction in our follow-up studies, either by removing the scattering in the projection domain before image reconstruction (Cui et al 2021), or by performing synthetic CT generation based on the reconstructed DE-CBCT using deep learning models (Peng et al 2025).

A requirement for the proposed method is that the total angular coverage of limited-angle DE-CBCT must meet the Tuy’s data sufficiency condition (Tang 2023), because the success of the inter-spectral structural similarity regularization in this work relies on the acquisition of complete anatomical structures. In other words, the total coverage of high- and low-energy projection views cannot be less than a short scan in the proposed work. However, a shorter scan is required for some clinical scenarios such as motion management, lower radiation dose, and avoidance of possible collisions between the gantry and the subject to be imaged. This would be the focus of our follow-up studies to extend the proposed method to more challenging data acquisition schemes.

The proposed limited-angle DE-CBCT is of great importance in the image-guided radiation therapy for two reasons. First, due to the slow gantry rotation, the CBCT imaging is susceptible to patient movement between the two scans at different x-ray tube voltages. With the proposed method, we can achieve DE-CBCT imaging within the time that is comparable to a single scan, which will significantly reduce the artifacts that may be induced by patient movement. Second, with the spectral information, the DE-CBCT can significantly enhance tissue contrast, improve tumor visualization, and enable precise material decomposition—the capabilities that are essential for accurate online adaptive radiation therapy (Zbijewski et al 2014, Sajja et al 2020).

In principle, the deep learning-based image prior can provide more information than the mathematically explicit image prior, e.g., the proposed inter-spectral similarity regularization. However, the major bottleneck of deep learning-based methods is the data pair requirement for supervised model training. Therefore, another focus of our future studies is to develop unsupervised alternatives to the existing models, which would pave the way for clinical applications of the deep learning-based limited-angle DE-CBCT. In our follow-up studies, we will investigate how to incorporate the proposed optimization-based image reconstruction into the deep learning methods for unsupervised enforcement of projection data consistency.

5. Conclusions

In this work, we propose an optimization-based image reconstruction regularized with inter-spectral structural similarity for limited-angle DE-CBCT. The proposed method can efficiently reduce limited-angle artifacts during the image reconstruction, enabling quantitative DE-CBCT with comparable data acquisition time and radiation dose to that of a single-energy scan on current onboard scanners without hardware modification. This work is of great clinical significance and can boost the clinical application of DE-CBCT in image-guided radiation therapy and surgical interventions.

Acknowledgment

This research is supported in part by the National Institutes of Health under Award Number R01CA272991, R01DE033512, P30CA008748 and U54CA274513.

Appendix. Proof of the non-convexity of the objective function

First, we would prove that the SSIM function is neither convex nor concave by the counter examples:

f(μH,μL) = SSIM(μH,μL)=(2μHμL+C1)(2σH,L+C2)(μH2+μL2+C1)(σH2+σL2+C2).

For a>0, we can select three points {(a,0),(0,a),(a2,a2)} in the (μH,μL) plane. Then we have

f(a,0)=f(0,a)=C1a2+C1

and

f(a2,a2)=a22+C1a22+C1=1.

A convex function should satisfy

f(a2,a2)f(a,0) + f(0,a)2=C1a2+C1.

But

f(a2,a2)=1>C1a2+C1.

Therefore, the SSIM is not convex.

To prove that SSIM is not concave, we select three points {(C1,0),(0,10C1),(C12,5C1)} in the (μH,μL) plane. Then

f(C1,0)=C1C1+C1=12f(0,10C1)=C1100C1+C1=1101

and

f(C12,5C1)=5C1+C125.25C1+C1=626.25.

A concave function should satisfy

f(C12,5C1)f(C1,0) + f(0,10C1)2=103404.

But

f(C12,5C1)=626.25<103404.

Therefore, the SSIM is not concave. In summary, the SSIM-based regularization term, [1SSIM(μH,μL)], is neither convex nor concave.

The total objective function (L), is a weighted summation of a convex data fidelity term (g) and a non-convex regularization term (f). Then we can find three points, {(a1,b1),((a2,b2)),(a1+a22,b1+b22)}, such that

Δg=g(a1,b1)+g(a2,b2)2g(a1+a22,b1+b22)0

and

Δf=f(a1+a22,b1+b22)f(a1,b1)+f(a2,b2)2>0.

If we let λ>Δg/Δf, then we have

ΔgλΔf<0
L(a1+a22,b1+b22)>L(a1,b1)+L(a2,b2)2

which means that the total objective function L is a non-convex function.

Data availability statement

The data that support the findings of this study are available upon reasonable request from the authors.

References

  1. Buerke B, Wittkamp G, Seifarth H, Heindel W, Kloska S P. Dual-energy CTA with bone removal for transcranial arteries: intraindividual comparison with standard CTA without bone removal and TOF-MRA. Acad. Radiol. 2009;16:1348–55. doi: 10.1016/j.acra.2009.05.007. [DOI] [PubMed] [Google Scholar]
  2. Cao W, Shapira N, Maidment A, Daerr H, Noël P B. Hepatic dual-contrast CT imaging: slow triple kVp switching CT with CNN-based sinogram completion and material decomposition. J. Med. Imaging. 2022;9:014003. doi: 10.1117/1.JMI.9.1.014003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Cassetta R, et al. Fast-switching dual energy cone beam computed tomography using the on-board imager of a commercial linear accelerator. Phys. Med. Biol. 2020;65:015013. doi: 10.1088/1361-6560/ab5c35. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Chandarana H, Megibow A J, Cohen B A, Srinivasan R, Kim D, Leidecker C, Macari M. Iodine quantification with dual-energy CT: phantom study and preliminary experience with renal masses. Am. J. Roentgenol. 2011;196:W693–W700. doi: 10.2214/AJR.10.5541. [DOI] [PubMed] [Google Scholar]
  5. Chen B, Zhang Z, Xia D, Sidky E Y, Gilat-Schmidt T, Pan X. Accurate image reconstruction in dual-energy CT with limited-angular-range data using a two-step method. Bioengineering. 2022;9:775. doi: 10.3390/bioengineering9120775. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Chen B, Zhang Z, Xia D, Sidky E Y, Pan X. Dual-energy CT imaging over non-overlapping, orthogonal arcs of limited-angular ranges. J. X-ray Sci. Technol. 2021;29:975–85. doi: 10.3233/XST-210974. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Chen B, Zhang Z, Xia D, Sidky E Y, Pan X. Prototyping optimization-based image reconstructions from limited-angular-range data in dual-energy CT. Med. Image Anal. 2024;91:103025. doi: 10.1016/j.media.2023.103025. [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Cho S, Lee S, Lee J, Lee D, Kim H, Ryu J-H, Jeong K, Kim K-G, Yoon K-H, Cho S. A novel low-dose dual-energy imaging method for a fast-rotating gantry-type CT scanner. IEEE Trans. Med. Imaging. 2020;40:1007–20. doi: 10.1109/TMI.2020.3044357. [DOI] [PubMed] [Google Scholar]
  9. Cui H, Jiang X, Fang C, Zhu L, Yang Y. Planning CT‐guided robust and fast cone‐beam CT scatter correction using a local filtration technique. Med. Phys. 2021;48:6832–43. doi: 10.1002/mp.15299. [DOI] [PubMed] [Google Scholar]
  10. Freixas G V. Dual-energy Cone-beam CT for Proton Therapy. Université de Lyon; 2017. [Google Scholar]
  11. Graser A, et al. Dual-energy CT in patients suspected of having renal masses: can virtual nonenhanced images replace true nonenhanced images? Radiology. 2009;252:433–40. doi: 10.1148/radiol.2522080557. [DOI] [PubMed] [Google Scholar]
  12. Hsieh J. Computed Tomography: Principles, Design, Artifacts, and Recent Advances. Wiley; 2003. [Google Scholar]
  13. Jaffray D A, Siewerdsen J H, Wong J W, Martinez A A. Flat-panel cone-beam computed tomography for image-guided radiation therapy. Int. J. Radiat. Oncol. Biol. Phys. 2002;53:1337–49. doi: 10.1016/S0360-3016(02)02884-5. [DOI] [PubMed] [Google Scholar]
  14. Jiang X, Fang C, Hu P, Cui H, Zhu L, Yang Y. Fast and effective single‐scan dual‐energy cone‐beam CT reconstruction and decomposition denoising based on dual‐energy vectorization. Med. Phys. 2021;48:4843–56. doi: 10.1002/mp.15117. [DOI] [PubMed] [Google Scholar]
  15. Jiang X, Gang G J, Stayman J W. Multi-material decomposition using spectral diffusion posterior sampling. 2024 doi: 10.1109/TBME.2025.3543747. (arXiv: 2408.01519) [DOI] [PMC free article] [PubMed]
  16. Kim H, Champley K. Differentiable forward projector for x-ray computed tomography. 2023 (arXiv: 2307.05801)
  17. Kingma D P. Adam: a method for stochastic optimization. 2014 (arXiv: 1412.6980)
  18. Lee D, Lee J, Kim H, Lee T, Soh J, Park M, Kim C, Lee Y J, Cho S. A feasibility study of low-dose single-scan dual-energy cone-beam CT in many-view under-sampling framework. IEEE Trans. Med. Imaging. 2017;36:2578–87. doi: 10.1109/TMI.2017.2765760. [DOI] [PubMed] [Google Scholar]
  19. Long Y, Fessler J A, Balter J M. 3D forward and back-projection for x-ray CT using separable footprints. IEEE Trans. Med. Imaging. 2010;29:1839–50. doi: 10.1109/TMI.2010.2050898. [DOI] [PMC free article] [PubMed] [Google Scholar]
  20. Maybody M, Stevenson C, Solomon S B. Overview of navigation systems in image-guided interventions. Tech. Vasc. Interv. Radiol. 2013;16:136–43. doi: 10.1053/j.tvir.2013.02.008. [DOI] [PubMed] [Google Scholar]
  21. Men K, Dai J, Chen X, Li M, Zhang K, Huang P. Dual-energy imaging method to improve the image quality and the accuracy of dose calculation for cone-beam computed tomography. Phys. Med. 2017;36:110–8. doi: 10.1016/j.ejmp.2017.03.023. [DOI] [PubMed] [Google Scholar]
  22. Müller K, Datta S, Ahmad M, Choi J-H, Moore T, Pung L, Niebler C, Gold G E, Maier A, Fahrig R. Interventional dual‐energy imaging—Feasibility of rapid kV‐switching on a C‐arm CT system. Med. Phys. 2016;43:5537–46. doi: 10.1118/1.4962929. [DOI] [PMC free article] [PubMed] [Google Scholar]
  23. Parker D L. Optimal short scan convolution reconstruction for fan beam CT. Med. Phys. 1982;9:254–7. doi: 10.1118/1.595078. [DOI] [PubMed] [Google Scholar]
  24. Peng J. Single-scan half-fan DECT using static detector modulation: a preliminary study. Proc. SPIE. 2022;12031:756–63. doi: 10.1117/12.2604880. [DOI] [Google Scholar]
  25. Peng J, et al. Diffusion model-based single-scan dual-energy cone-beam CT using primary beam splitting. Proc. SPIE. 2024a;12925:618–23. doi: 10.1117/12.3005046. [DOI] [Google Scholar]
  26. Peng J, et al. Dual-energy cone-beam CT using two complementary limited-angle scans with a projection-consistent diffusion model. 2024b (arXiv: 2403.11890)
  27. Peng J, Gao Y, Chang C-W, Qiu R, Wang T, Kesarwala A, Yang K, Scott J, Yu D, Yang X. Unsupervised Bayesian generation of synthetic CT from CBCT using patient‐specific score‐based prior. Med. Phys. 2025;52:2238–46. doi: 10.1002/mp.17572. [DOI] [PMC free article] [PubMed] [Google Scholar]
  28. Petrongolo M, Zhu L. Single-scan dual-energy CT using primary modulation. IEEE Trans. Med. Imaging. 2018;37:1799–808. doi: 10.1109/TMI.2018.2796858. [DOI] [PubMed] [Google Scholar]
  29. Punnoose J, Xu J, Sisniega A, Zbijewski W, Siewerdsen J. spektr 3.0—A computational tool for x‐ray spectrum modeling and analysis. Med. Phys. 2016;43:4711–7. doi: 10.1118/1.4955438. [DOI] [PMC free article] [PubMed] [Google Scholar]
  30. Reddi S J, Kale S, Kumar S. On the convergence of adam and beyond. 2019 (arXiv: 1904.09237)
  31. Roeske J C, et al. Characterization of markerless tumor tracking using the on-board imager of a commercial linear accelerator equipped with fast-kV switching dual-energy imaging. Adv. Radiat. Oncol. 2020;5:1006–13. doi: 10.1016/j.adro.2020.01.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  32. Rudin L I, Osher S, Fatemi E. Nonlinear total variation based noise removal algorithms. Physica D. 1992;60:259–68. doi: 10.1016/0167-2789(92)90242-F. [DOI] [Google Scholar]
  33. Sajja S, Lee Y, Eriksson M, Nordström H, Sahgal A, Hashemi M, Mainprize J G, Ruschin M. Technical principles of dual-energy cone beam computed tomography and clinical applications for radiation therapy. Adv. Radiat. Oncol. 2020;5:1–16. doi: 10.1016/j.adro.2019.07.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  34. Shi L, Lu M, Bennett N R, Shapiro E, Zhang J, Colbeth R, Star‐Lack J, Wang A S. Characterization and potential applications of a dual‐layer flat‐panel detector. Med. Phys. 2020;47:3332–43. doi: 10.1002/mp.14211. [DOI] [PMC free article] [PubMed] [Google Scholar]
  35. Siddon R L. Fast calculation of the exact radiological path for a three‐dimensional CT array. Med. Phys. 1985;12:252–5. doi: 10.1118/1.595715. [DOI] [PubMed] [Google Scholar]
  36. Siewerdsen J, Moseley D J, Burch S, Bisland S K, Bogaards A, Wilson B C, Jaffray D A. Volume CT with a flat‐panel detector on a mobile, isocentric C‐arm: pre‐clinical investigation in guidance of minimally invasive surgery. Med. Phys. 2005;32:241–54. doi: 10.1118/1.1836331. [DOI] [PubMed] [Google Scholar]
  37. Skaarup M, Edmund J M, Dorn S, Kachelriess M, Vogelius I R. Dual-energy material decomposition for cone-beam computed tomography in image-guided radiotherapy. Acta Oncol. 2019;58:1483–8. doi: 10.1080/0284186X.2019.1629010. [DOI] [PubMed] [Google Scholar]
  38. Szczykutowicz T P, Chen G-H. Dual energy CT using slow kVp switching acquisition and prior image constrained compressed sensing. Phys. Med. Biol. 2010;55:6411. doi: 10.1088/0031-9155/55/21/005. [DOI] [PubMed] [Google Scholar]
  39. Tang X. Spectral Multi-detector Computed Tomography (Smdct): Data Acquisition, Image Formation, Quality Assessment and Contrast Enhancement. CRC Press; 2023. [Google Scholar]
  40. Toepker M, Kuehas F, Kienzl D, Herwig R, Spazierer E, Krauss B, Weber M, Seitz C, Ringl H. Dual energy computerized tomography with a split bolus—a 1-stop shop for patients with suspected urinary stones? J. Urol. 2014;191:792–7. doi: 10.1016/j.juro.2013.10.057. [DOI] [PubMed] [Google Scholar]
  41. Wang T, Zhu L. Dual energy CT with one full scan and a second sparse-view scan using structure preserving iterative reconstruction (SPIR) Phys. Med. Biol. 2016;61:6684. doi: 10.1088/0031-9155/61/18/6684. [DOI] [PMC free article] [PubMed] [Google Scholar]
  42. Wang W, et al. High‐resolution model‐based material decomposition in dual‐layer flat‐panel CBCT. Med. Phys. 2021;48:6375–87. doi: 10.1002/mp.14894. [DOI] [PMC free article] [PubMed] [Google Scholar]
  43. Wang Z, Bovik A C, Sheikh H R, Simoncelli E P. Image quality assessment: from error visibility to structural similarity. IEEE Trans. Image Process. 2004;13:600–12. doi: 10.1109/TIP.2003.819861. [DOI] [PubMed] [Google Scholar]
  44. Xiang J, et al. DER-GAN: dual-Energy Recovery GAN for Conebeam CT. IEEE Trans. Comput. Imaging; 2023. pp. pp 28–42. [Google Scholar]
  45. Yu L, Leng S, McCollough C H. Dual-energy CT–based monochromatic imaging. Am. J. Roentgenol. 2012;199:S9–S15. doi: 10.2214/AJR.12.9121. [DOI] [PubMed] [Google Scholar]
  46. Zbijewski W, Gang G J, Xu J, Wang A S, Stayman J W, Taguchi K, Carrino J A, Siewerdsen J H. Dual‐energy cone‐beam CT with a flat‐panel detector: effect of reconstruction algorithm on material classification. Med. Phys. 2014;41:021908. doi: 10.1118/1.4863598. [DOI] [PMC free article] [PubMed] [Google Scholar]
  47. Zhang L-J, Peng J, Wu S-Y, Wang Z J, Wu X-S, Zhou C-S, Ji X-M, Lu G-M. Liver virtual non-enhanced CT with dual-source, dual-energy CT: a preliminary study. Eur. Radiol. 2010;20:2257–64. doi: 10.1007/s00330-010-1778-7. [DOI] [PubMed] [Google Scholar]
  48. Zhang Y, Hu D, Yan Z, Zhao Q, Quan G, Luo S, Zhang Y, Chen Y. TIME-Net: transformer-integrated multi-encoder network for limited-angle artifact removal in dual-energy CBCT. Med. Image Anal. 2023;83:102650. doi: 10.1016/j.media.2022.102650. [DOI] [PubMed] [Google Scholar]
  49. Zhang Z, Chen B, Xia D, Sidky E Y, Pan X. Image reconstruction from data over two orthogonal arcs of limited‐angular ranges. Med. Phys. 2022;49:1468–80. doi: 10.1002/mp.15450. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data that support the findings of this study are available upon reasonable request from the authors.


Articles from Physics in Medicine and Biology are provided here courtesy of IOP Publishing