Abstract
Background
Kidney replacement therapy (KRT) is needed by patients with kidney failure to eliminate metabolic waste. When kidney transplantation is not feasible, hemodialysis (HD) is performed. Hemodiafiltration (HDF) provides a better solute clearance than conventional HD due to its higher convective dose. The clearance performances of HDF also depend on the type of dialysis membrane used. Polysulfone (PS) membranes have been used since 1965 due to their excellent resistance and biocompatibility. However, the newer generation of polymethylmethacrylate (PMMA) membranes offers comparable convective capacity with reduced albumin loss.
Objective
The forthcoming study aims to compare the solute clearance of postdilution HDF using PMMA versus PS dialyzer membranes in patients with kidney failure requiring long-term HD.
Methods
The literature search will use databases including PubMed, Scopus, Web of Science, Embase, and Cochrane. Two independent reviewers will select studies and assess their quality. Pooled estimates of relevant factors will be computed via a random-effects model using Stata/BE 18 (StataCorp) software. The reporting of findings will adhere to the PRISMA (Preferred Reporting Items for Systematic Reviews and Meta-Analyses) guidelines.
Results
The search and screening for the systematic literature review are expected to be completed in May 2026. Data extraction, quality appraisal, and subsequent data synthesis will begin in June 2026. The review should be completed by October 2026, and the study results will be published in 2027.
Conclusions
Dialyzer membrane clearance determines HD quality. Nevertheless, many middle uremic molecules are poorly cleared with conventional HD. Consequently, HDF was invented to improve their clearance. Among the most widely used dialyzer membranes, PS and PMMA are used in the newer generation of HDF. Nevertheless, no available systematic reviews have compared the efficacy of the 2 dialyzer membrane materials. Although no ideal dialyzer can efficiently remove all types of uremic toxins while retaining essential elements, improving solute clearance should benefit patients. This information could help clinicians determine which membrane to use in HDF.
Trial Registration
PROSPERO CRD42024603277; https://www.crd.york.ac.uk/PROSPERO/view/CRD42024603277
International Registered Report Identifier (IRRID)
PRR1-10.2196/71943
Keywords: albumin, beta 2-microglobulin, creatinine, dialysis, end stage kidney disease, polymethyl methacrylate, polysulfone
Introduction
In patients with chronic kidney disease (CKD), signs and symptoms typically develop gradually [1-3]. As kidney function declines progressively, the condition may eventually lead to kidney failure. At this stage, the kidneys can no longer effectively eliminate metabolic waste, resulting in the accumulation of toxins and disturbances in electrolyte balance [4]. The definitive treatment for kidney failure is kidney replacement therapy (KRT) through a kidney transplant. Still, this option is not always feasible or successful, necessitating other KRT modalities, namely hemodialysis (HD) or peritoneal dialysis [5-9]. HD can be divided into conventional HD, hemodiafiltration (HDF), and expanded HD [10].
Conventional HD uses solute diffusion and limited ultrafiltration, using a high-flux membrane inserted in a hollow-fiber dialyzer. Conventional HD is linked to “residual risk” resulting from the accumulation of inadequately removed medium-sized compounds. Meanwhile, HDF uses convection and diffusion to eliminate small- and medium-sized molecules. A significant distinction between HDF and conventional HD is the total ultrafiltered volume per session (dialytic convective dose). HDF can achieve a significantly higher convective dose (10- to 15-fold), leading to a 2-fold increase in middle-molecular-weight solute clearance [11,12].
HDF was initially introduced in the United States by Lee Henderson in 1973 [13]. On-line generation of substitution fluid for HDF was then implemented in the mid-1980s [14]. HDF has since undergone substantial development. See et al (2019) [15] found that HDF was associated with a significantly lower risk of all-cause mortality (adjusted hazard ratio [HR] of 0.79 and 0.88 for patients in Australia and New Zealand, respectively). An association was also found between HDF and reduced cardiovascular mortality, with an adjusted HR of 0.78 in Australian patients. Using the French Renal Epidemiology and Information Network Registry, Mercadal et al (2016) [16] found that the HRs for all-cause and cardiovascular mortality in patients treated exclusively with HDF were 0.77 and 0.66, respectively.
HDF can be conducted with dialysis membranes comprising different raw materials, including cellulose, polysulfone (PS), polyamide, polymethylmethacrylate (PMMA), and polycarbonate [17]. Clinicians must consider several factors to determine which dialyzer membrane is most suitable for dialysis: performance, biocompatibility, cytotoxicity, sterilization, and endotoxin retention [18]. This systematic review and meta-analysis will focus on the efficacy of PMMA dialyzers compared to that of PS in postdilution HDF in patients with kidney failure requiring long-term HD. A PS membrane has an asymmetric structure where the pore size enlarges from inside to outside and a thin skin layer with very small pore sizes to control solute separations. Conversely, a PMMA membrane has a uniform pore size structure, and the entire membrane controls solute separations. Therefore, the separation characteristics of PMMA are broader than those of PS membranes. Furthermore, PS membranes remove uremic toxins from the blood via diffusion and filtration, while PMMA membranes also use adsorption in addition to diffusion and filtration [19-21]. However, PMMA membranes have lower permeability than PS membranes, increasing the risk of fouling and platelet-induced membrane coagulation [22].
A dialyzer’s efficacy is assessed based on the clearance of solutes, including urea, creatinine, phosphate, uric acid, albumin, α1-microglobulin, ß2-microglobulin, myoglobin, prolactin, and κ and λ free light chains [23,24]. Our review aims to answer the following question: in patients with kidney failure requiring long-term HD, does the PMMA dialyzer membrane demonstrate a higher efficacy than the PS dialyzer membrane? The PICO (Population, Intervention, Comparison, and Outcomes) framework will be used during the article search; “P (population)” refers to patients with kidney failure requiring long-term HD, “I (intervention)” refers to HDF using PMMA dialyzer membrane, “C (comparison)” refers to HDF using the PS dialyzer membrane, and “O (outcome)” refers to the efficacy of solute clearance.
Methods
Overview
This review will be conducted and reported following the PRISMA-P (Preferred Reporting Items for Systematic Reviews and Meta-Analyses Protocols) guidelines (Multimedia Appendix 1) [25,26]. We will also integrate this review with the Meta-analysis of Observational Studies in Epidemiology guidelines, since the outcome will be a meta-analysis of selected observational studies [27].
Eligibility Criteria
We will include all original articles in English based on the following criteria:
Studies involving adult patients diagnosed with kidney failure and requiring long-term HD.
Studies investigating postdilution HDF using PMMA and/or PS as dialyzer membranes.
Studies reporting dialysis efficacy as the primary or secondary outcome.
Peer-reviewed original articles, including randomized controlled trials (RCTs) and observational studies (cohort, case-control, and cross-sectional studies).
The exclusion criteria are as follows:
Studies involving patients with acute kidney injury, from which long-term dialysis patients’ group data cannot be extracted.
Studies using dialyzer membranes other than PMMA or PS and using mixed or combined membrane types, from which PMMA or PS group data cannot be extracted.
Studies investigating dialysis modalities other than HDF (eg, high-flux and low-flux HD, expanded HDF).
Any review, case study, commentary, qualitative study, or editorial.
Outcome Measures
The review outcomes are efficacy based on solute clearance of urea, creatinine, phosphate, uric acid, α1‑microglobulin, β2-microglobulin, myoglobin, albumin, and prolactin. The efficacy will be assessed based on the solutes’ reduction ratio (RR) measured before and after the HDF session. Quantitative data will be reported as mean (SD) values.
Information Sources
Electronic Search
We will systematically conduct a comprehensive literature search using various literature databases, including MEDLINE (PubMed), Web of Science, Scopus, Embase, and the Cochrane Central Register of Controlled Trials (CENTRAL) to identify eligible studies. Secondary searches will be conducted on Google Scholar and other websites, and the reference section of the included studies will also be hand-searched for additional relevant studies. Studies will be restricted to the English language. The search will be performed from selected electronic databases up to October 2025.
Search Strategy
The proposed search term for the first theme will be “long-term hemodialysis.” The second theme is “polymethylmethacrylate” and “polysulfone.” The third theme is “efficacy.” The exploded versions of the MeSH (Medical Subject Headings) browser for each theme will be included. All 3 search themes will be combined using the Boolean operators “OR” and “AND.” Multimedia Appendix 2 presents the detailed search terms for each database.
Study Selection
We will identify and remove duplicates and collate multiple reports. Two review authors (AKK and NN) will independently screen all the titles and abstracts to examine the potential studies for inclusion and exclude evidently irrelevant studies. We will identify studies from databases and remove duplicate records. We will then perform an initial screening based on the title and abstract of the articles to identify those that are irrelevant or meet the inclusion criteria. Articles deemed irrelevant based on the title and abstract will be excluded from further consideration. For the included study, we will retrieve the full-text study reports and publications, and the review authors will independently screen the full text to identify studies for inclusion and exclusion to determine their eligibility. The reasons for excluding articles, such as having an in vitro design and having different interventions, HD modalities, patient groupings, outcomes, or measurements, will be documented. Articles that meet all inclusion criteria and are not excluded will be included in the review.
We will resolve any disagreement through discussion. If no consensus is reached, the other 2 authors will act as arbiters. We will record the selection process in sufficient detail to complete a PRISMA (Preferred Reporting Items for Systematic Reviews and Meta-Analyses) flow diagram and construct a table describing the characteristics of the excluded studies [25,26]. Rayyan (Qatar Computing Research Institute) will be used to store, organize, and manage all the articles identified from databases.
Data Extraction and Management
We will use a standardized data extraction form created by the Microsoft Excel Spreadsheet software for study characteristics and outcome data. Two review authors (CGA and NN) will independently extract outcome data from the included studies. Discrepancies in determining data usability will be resolved by discussion or by involving the other 2 review authors. The usability of outcome data will be documented in the characteristics of included studies. Data will be considered unusable if any of the following apply: (1) a nonstandard formula or procedure is used to calculate the RR; (2) RR is reported as median (IQR), and conversion to mean (SD) is not feasible; (3) results are presented only in graphical form without sufficient resolution or accompanying numerical data; (4) inconsistent or incomplete reporting of outcome data across comparison groups; (5) the HDF procedure is not adequately described, particularly concerning the dialysis modality (eg, postdilution vs predilution), membrane type, and dialysis settings; (6) the type of membrane used is not reported or is ambiguous; and (7) key demographic or baseline characteristics of participants (eg, age, sex, and dialysis duration) are not reported. Studies meeting the eligibility criteria but excluded due to unusable data will be listed in the final review along with specific reasons for exclusion, to ensure transparency and minimize selective reporting bias.
We will recheck that data are entered correctly by comparing the systematic review data with the study reports. The following study characteristics from the included studies will be extracted:
Title, first author, study country based on the study setting, region, and publication year.
Methods: study design, data source, total study duration, and analysis method.
Participants: number of patients, mean and/or median age, age range, and sex.
Outcome: efficacy based on solute clearance, which is the measurement of RR of each solute such as urea, creatinine, phosphate, uric acid, α1-microglobulin, β2-microglobulin, myoglobin, albumin, and prolactin, which were measured before and after HDF.
Exposures or risks: CKD etiology, dialysis vintage, and comorbidity. However, whenever data are applicable, we will extract the data on other exposure risks or modifiable factors, control conditions, and adjustment variables.
If there is any unclear or missing data for extraction, we will contact the corresponding author of the original studies to request clarification or check whether additional data can be made available upon request. If there is no response from the corresponding author within a reasonable time frame, the study will be excluded from the analysis, with the reason for exclusion documented in the review. We will also assess the potential impact of missing data on the overall findings through sensitivity analysis, where feasible. We will also consider the risk of bias due to incomplete data in the overall quality assessment.
Quality Assessment
Two review authors will independently assess the risk of bias depending on the study design (AKK and AW). For RCTs, we will use the Cochrane Risk of Bias 2 (RoB 2) tool, which evaluates five domains: (1) bias due to the randomization process, (2) deviation from intended intervention, (3) missing outcome data, (4) measurement of outcomes, and (5) selection of the reported result. Each domain will be rated as “low risk,” “some concerns,” or “high risk,” and an overall risk of bias judgment will be assigned [28].
For nonrandomized intervention studies, we will use the Risk Of Bias In Non-randomized Studies of Interventions (ROBINS-I) tool. The risk of bias will be assessed across several domains: (1) bias due to confounding, (2) bias in participant selection, (3) bias in classification of the intervention, (4) bias from deviations in the intended interventions, (5) bias due to missing outcome data, (6) bias in outcome measurement, and (7) bias in the selection of reported results. The risk of bias for each domain will be rated as low, moderate, serious, or critical. An overall risk of bias judgment will also be made [29]. In the sensitivity analysis, we will exclude trials assessed as high risk of bias with the RoB 2 tool and those rated as serious or critical risk of bias with the ROBINS-I tool. This will allow us to assess whether the inclusion of these studies significantly influences the pooled effect estimates.
Statistical Analysis
Data Analysis and Statistical Analysis
The primary outcome of our study is the efficacy, which will be based on solute clearance, based on the RR of each solute. RR will be determined using laboratory parameters, which include serum concentrations of urea, creatinine, phosphate, uric acid, α1-microglobulin, β2-microglobulin, myoglobin, albumin, and prolactin, measured before and after HDF. Quantitative data will be reported as mean (SD) values. Since RRs are measured using the same scale across all included studies, we will use the mean difference as the effect measure. For parametric comparisons, paired Student t tests followed by Bonferroni post hoc tests will be used. A P value of <.05 will be considered statistically significant.
Although both RCTs and nonrandomized studies will be included in this systematic review, a meta-analysis will be conducted separately for each study design to account for their inherent methodological differences. Meta-analyses will be conducted using a random-effects model to account for potential variability and heterogeneity among studies. Given the inclusion of nonrandomized studies, we acknowledge the potential for confounding and selection bias. To address this, we will prioritize adjusted effect estimates (eg, those derived from multivariable regression models or propensity score analyses) where available. Additionally, we will conduct subgroup or sensitivity analyses by excluding studies judged to be at overall high risk of bias. If 10 or more studies are included, we will consider performing meta-regression to explore potential effect modifiers. A narrative synthesis will be provided for outcomes not amenable to quantitative synthesis.
Heterogeneity Assessment
Assessing heterogeneity is critical in a meta-analysis, as it enables the evaluation of the degree of inconsistency or variability among the results of individual studies. This review will evaluate heterogeneity using the I2 and Q statistics. The I2 statistic will be used to quantify the impact of heterogeneity, with percentages of around 25%, 50%, and 75% representing a low, moderate, and high degree of heterogeneity, respectively [30]. The statistical Q test determines whether significant heterogeneity exists among the studies. The significance level for the Q test is set at 0.01 in this review [31]. If significant heterogeneity is detected using the Q test or I2 index (>50%), we will explore potential heterogeneity sources using subgroup analyses and meta-regression. We will also explore possible causes (for example, differences in study quality, participants, or outcome assessments) and evaluate the studies based on their methodological characteristics to determine whether differences in these characteristics can explain the degree of heterogeneity and whether a meta-analysis is appropriate.
Assessment of Publication Bias
We will create and examine a funnel plot to explore possible small study biases if we can pool over 10 studies in 1 meta-analysis. The number of studies missing from the funnel plot will be estimated. The effect size after inputting these missing studies will be estimated using the trim-and-fill method. This simple estimation approach proposed by Duval and Tweedie [32] involves trimming off the asymmetric outlying part of the funnel, using the symmetric remainder to estimate the true funnel center, and replacing the trimmed studies and their counterparts around the center. Other methods to assess the publication bias will include the Begg rank correlation and the Egger weighted regression method test [33,34].
Ethical Considerations
We registered this systematic review in PROSPERO (systematic review registration: CRD42024603277). Since this review will use published data, the need for ethical approval is waived in accordance with local guidelines. The systematic review will focus on determining whether the PMMA or the PS dialyzer membrane has a higher efficacy in patients with kidney failure requiring long-term HD.
Results
The search and screening for the systematic literature review is expected to conclude in May 2026. Data extraction, quality appraisal, and subsequent data synthesis will begin in June 2026. The review should be completed by October 2026, and the study results will be published in 2027.
Discussion
PS is an amorphous polymer with excellent thermal and chemical stability [35]. The PS membrane is asymmetric, implying that for a membrane thickness of 30 microns, a 1-micron layer controls the separation process, while the remainder has structural functions. Conversely, the PMMA membrane is symmetric, with the whole thickness involved in the separation process [21]. The PMMA membrane adds adsorption to the existing HD diffusion and convection processes. While diffusion and convection remove small and medium-sized molecules, adsorption enables the removal of medium- and high-molecular-weight molecules [22].
The removal of toxins and waste products from the patient’s blood defines the adequacy of HD. Adequate dialysis can improve patient survival and quality of life [36-39]. Kt/V calculation is the most widely used method to assess dialysis adequacy, where K is the effective dialyzer solute clearance, t is the duration of the dialysis session, and V is the solute distribution volume [40]. Currently, small solute clearance is regarded as the best measure of HD adequacy. Kt/V urea, also known as fractional urea clearance, is a widely used tool to assess dialysis efficacy in daily clinical practice [41].
The use of urea as the sole representation of dialysis adequacy tends to oversimplify this concept. In this context, urea does not represent the kinetic behavior of other solutes, including inorganic solutes such as phosphate, middle molecules such as β2-microglobulin, and protein-bound uremic toxins such as p-cresyl sulfate, whose detrimental effects are linked with inflammation, oxidative stress, and cardiovascular disease [42-47]. Hence, the concept of adequacy is evolving to include nonurea parameters [48].
HD adequacy is partly determined by dialyzer membrane clearance [49]. Moreover, membrane properties, including pore size, wall thickness, and surface area, determine solute clearance. The uremic toxins HD removes are divided into four categories: (1) small water-soluble compounds (eg, urea) with a size of <500 Da, (2) a middle-molecular-weight of 500-15,000 Da, (3) large molecules with a size of >15,000 Da, and (4) protein-bound molecules (eg, p-cresol of 108 Da). However, when describing the in vitro performance of dialyzers, many studies only considered the results obtained with small solutes and middle-molecular-weight molecules [50].
Middle-molecular-weight molecules include several cytokines, adipokines, growth factors, and other signaling proteins significantly elevated in dialysis patients compared to that in people with normal kidney function. The serum levels of interleukin (IL)-1β, IL-6, and IL-18, and κ and λ free light chains are also elevated in patients with advanced CKD [51]. These molecules play key roles in the inflammation-cardiovascular disease pathway and are linked with chronic inflammation, atherosclerosis, structural cardiac disease, and protein-energy wasting [52,53].
While conventional HD can effectively remove small uremic molecules (<0.5 kDa), including urea, it poorly removes middle-molecular-weight molecules in the 0.5-60 kDa range. Thus, HDF was invented to improve the clearance of middle-molecular-weight molecules [54]. The clinical application of on-line generation of replacement fluid HDF has been limited in the United States and has predominantly advanced in Europe. Increasing numbers of patients in Japan and other Asian countries have recently opted for HDF treatment [55,56].
The HDF removal efficiency is mainly determined by the replacement fluid volume, while the removal characteristic is affected by the dilution method. Postdilution HDF, when combined with a protein-nonleakage hemodiafilter, is superior in removing small to middle-molecular-weight molecules. Conversely, predilution HDF with a protein-leakage–type hemodiafilter is preferred for removing middle- to high-molecular weight molecules [57]. Predilution HDF efficiently removes middle-molecular-weight molecules of up to α1-microglobulin (26-30 kDa). However, the biological activity of α1-microglobulin is unclear, and the uremic toxins in its vicinity are also unknown. Meanwhile, postdilution HDF increases the removal efficiency for molecular weights up to those of β2-microglobulin (11-12 kDa), which has been established as a uremic toxin [57].
Views differ on the impact of dialyzer characteristics on patient outcomes and the characteristics that the clinician should prioritize. The impact also relies on the dialyzer’s application, where convective therapies such as HDF outperform diffusive therapies such as HD [58]. Although no ideal dialyzer can efficiently remove all types of uremic toxins while retaining vitamins and other essential elements, expanding the range of uremic toxin clearance should benefit patients. Nevertheless, further research is required to demonstrate the superiority of these new dialyzers [59].
Our study will include stable patients with kidney failure undergoing thrice weekly HD sessions. We recognize that the duration of each dialysis session (in minutes) might differ between studies. Therefore, we will only include studies with similar dialysis prescriptions and the same modality, postdilution HDF, to limit bias. To our knowledge, this study will be the first to conduct a systematic review and meta-analysis of the available evidence regarding comparing clearance between PMMA and PS dialyzer membranes for postdilution HDF in patients with kidney failure.
Our study may have several limitations. Even if the studies used the same membrane material (PS or PMMA), if they used dialyzers from different manufacturers or the same manufacturer but different versions (type, design, or specification), bias could result from different clearance capacities. Second, clearance is subject to methodological heterogeneity, as the time point for each blood sampling collection might differ between studies. We will not specify time points as inclusion criteria due to the risk of eliminating potentially relevant studies.
Third, our study will not assess some aspects potentially affecting the fluid removal rate. We will not evaluate the patients’ primary kidney disease or other health problems potentially affecting the blood protein level (albumin). Moreover, we will not assess the quality of blood vessels or HD lumen catheters used for dialysis, which potentially affect dialysis efficiency through resistance and subsequently influence the fluid removal rate.
In summary, our systematic review and meta-analysis may help compare the solute clearance of PMMA and PS dialyzer membranes for postdilution HDF in people with kidney failure. The results may help clinicians determine the most suitable membrane for HDF for their routine HD practice. The data may also provide a starting point for further research on the impacts of various dialyzer membranes on the HDF solute removal rate.
Abbreviations
- CKD
chronic kidney disease
- HD
hemodialysis
- HDF
hemodiafiltration
- HR
hazard ratio
- IL
interleukin
- KRT
kidney replacement therapy
- MeSH
Medical Subject Headings
- PICO
Population, Intervention, Comparison, and Outcomes
- PMMA
polymethylmethacrylate
- PRISMA
Preferred Reporting Items for Systematic Reviews and Meta-Analyses
- PRISMA-P
Preferred Reporting Items for Systematic Reviews and Meta-Analyses Protocols
- PS
polysulfone
- RCT
randomized controlled trial
- RoB 2
Cochrane Risk of Bias 2
- ROBINS-I
Risk Of Bias In Non-randomized Studies of Interventions
- RR
reduction ratio
PRISMA-P checklist.
Search terms.
Footnotes
Authors' Contributions: Conceptualization: CGA and NN
Data curation: CGA and NN
Formal analysis: CGA and NN
Investigation: CGA, AKK, and NN
Methodology: CGA and NN
Project administration: CGA and NN
Resources: CGA
Supervision: CGA, AKK, and AW
Validation: CGA, AKK, and AW
Visualization: CGA and NN
Writing—original draft: CGA and NN
Writing—review and editing: CGA, AKK, AW, and NN
Conflicts of Interest: None declared.
References
- 1.Santoso DN, Sinuraya FAG, Ambarsari CG. Distal renal tubular acidosis presenting with an acute hypokalemic paralysis in an older child with severe vesicoureteral reflux and syringomyelia: a case report. BMC Nephrol. 2022 Jul 14;23(1):248. doi: 10.1186/s12882-022-02874-9. https://www.biomedcentral.com/1471-2369/23/248 .10.1186/s12882-022-02874-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Ambarsari CG, Agianda HAP, Saraswati M, Kim JJ. Dent disease 1 presented early with Bartter-Like cyndrome features and rickets: A case report. Case Rep Nephrol Dial. 2025 Jan 21;15(1):15–25. doi: 10.1159/000543719. [DOI] [Google Scholar]
- 3.Ambarsari CG, Tambunan T, Pardede SO, Fazlur Rahman FH, Kadaristiana A. Role of dipstick albuminuria in progression of paediatric chronic kidney disease. J Pak Med Assoc. 2021 Feb;71(Suppl 2)(2):S103–S106. [PubMed] [Google Scholar]
- 4.Zhang L, Guo Y, Ming H. Effects of hemodialysis, peritoneal dialysis, and renal transplantation on the quality of life of patients with end-stage renal disease. Rev Assoc Med Bras (1992) 2020 Sep;66(9):1229–1234. doi: 10.1590/1806-9282.66.9.1229. https://www.scielo.br/scielo.php?script=sci_arttext&pid=S0104-42302020000901229&lng=en&nrm=iso&tlng=en .S0104-42302020000901229 [DOI] [PubMed] [Google Scholar]
- 5.Ambarsari CG, Cho Y, Milanzi E, Francis A, Koh LJ, Lalji R, Johnson DW. Epidemiology and outcomes of children with kidney failure receiving kidney replacement therapy in Australia and New Zealand. Kidney Int Rep. 2023 Oct;8(10):1951–1964. doi: 10.1016/j.ekir.2023.07.006. https://linkinghub.elsevier.com/retrieve/pii/S2468-0249(23)01384-0 .S2468-0249(23)01384-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Robinson BM, Akizawa T, Jager KJ, Kerr PG, Saran R, Pisoni RL. Factors affecting outcomes in patients reaching end-stage kidney disease worldwide: differences in access to renal replacement therapy, modality use, and haemodialysis practices. The Lancet. 2016 Jul;388(10041):294–306. doi: 10.1016/s0140-6736(16)30448-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Ambarsari CG, Hidayati EL, Hasan I, Grace A, Oswari H. Successful treatment of hepatitis C virus infection using direct-acting antiviral agents (DAAs) in adolescents with kidney transplantation: A case series. IJNRD. 2020 Jun;Volume 13:139–146. doi: 10.2147/ijnrd.s248632. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Ambarsari CG, Hidayati EL, Trihono PP, Saraswati M, Rodjani A, Wahyudi I, Situmorang GR, Kim JJ, Mellyana O, Kadaristiana A. Experience of the first 6 years of pediatric kidney transplantation in Indonesia: A multicenter retrospective study. Pediatr Transplant. 2020 Dec 14;24(8):e13812. doi: 10.1111/petr.13812. [DOI] [PubMed] [Google Scholar]
- 9.Ambarsari C, Bermanshah E, Putra M, Rahman F, Pardede S. Effective management of peritoneal dialysis-associated hydrothorax in a child: A case report. Case Rep Nephrol Dial. 2020 Feb 12;10(1):18–25. doi: 10.1159/000506119. https://doi.org/10.1159/000506119 .cnd-0010-0018 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Mitchell CR, Hornig C, Canaud B. Systematic review to compare the outcomes associated with the modalities of expanded hemodialysis (HDx) versus high-flux hemodialysis and/or hemodiafiltration (HDF) in patients with end-stage kidney disease (ESKD) Semin Dial. 2023 Mar 27;36(2):86–106. doi: 10.1111/sdi.13130. [DOI] [PubMed] [Google Scholar]
- 11.Canaud B, Stuard S, Laukhuf F, Yan G, Canabal M, Lim P, Kraus Michael A. Choices in hemodialysis therapies: variants, personalized therapy and application of evidence-based medicine. Clin Kidney J. 2021 Dec;14(Suppl 4):i45–i58. doi: 10.1093/ckj/sfab198. https://europepmc.org/abstract/MED/34987785 .sfab198 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Roumelioti M, Trietley G, Nolin T, Ng Y, Xu Z, Alaini A, Figueroa Rocio, Unruh Mark L, Argyropoulos Christos P. Beta-2 microglobulin clearance in high-flux dialysis and convective dialysis modalities: a meta-analysis of published studies. Nephrol Dial Transplant. 2018 Jun 01;33(6):1025–1039. doi: 10.1093/ndt/gfx311.4666452 [DOI] [PubMed] [Google Scholar]
- 13.Henderson LW, Beans E. Successful production of sterile pyrogen-free electrolyte solution by ultrafiltration. Kidney Int. 1978 Nov;14(5):522–525. doi: 10.1038/ki.1978.157. https://linkinghub.elsevier.com/retrieve/pii/S0085-2538(15)31994-3 .S0085-2538(15)31994-3 [DOI] [PubMed] [Google Scholar]
- 14.Canaud B, Flavier J, Argilés A, Stec F, NGuyen Q V, Bouloux C, Garred L J, Mion C. Contributions to Nephrology. Basel: Karger; 1994. 12 - 22: Hemodiafiltration with On-Line Production of Substitution Fluid: Long-Term Safety and Quantitative Assessment of Efficacy. [DOI] [PubMed] [Google Scholar]
- 15.See E, Hedley J, Agar J, Hawley C, Johnson D, Kelly P, Lee Vincent W, Mac Kathy, Polkinghorne Kevan R, Rabindranath Kannaiyan S, Sud Kamal, Webster Angela C. Patient survival on haemodiafiltration and haemodialysis: a cohort study using the Australia and New Zealand Dialysis and Transplant Registry. Nephrol Dial Transplant. 2019 Feb 01;34(2):326–338. doi: 10.1093/ndt/gfy209.5071933 [DOI] [PubMed] [Google Scholar]
- 16.Mercadal L, Franck J, Metzger M, Urena Torres P, de Cornelissen F, Edet S, Béchade Clémence, Vigneau C, Drüeke Tilman, Jacquelinet C, Stengel B, REIN Registry Hemodiafiltration versus hemodialysis and survival in patients with ESRD: The French Renal Epidemiology and Information Network (REIN) Registry. Am J Kidney Dis. 2016 Aug;68(2):247–255. doi: 10.1053/j.ajkd.2015.11.016.S0272-6386(15)01436-5 [DOI] [PubMed] [Google Scholar]
- 17.Kohlová M, Amorim CG, Araújo A, Santos-Silva A, Solich P, Montenegro MCBSM. The biocompatibility and bioactivity of hemodialysis membranes: their impact in end-stage renal disease. J Artif Organs. 2018 Jul 14;22(1):14–28. doi: 10.1007/s10047-018-1059-9. [DOI] [PubMed] [Google Scholar]
- 18.Bowry S, Chazot C. The scientific principles and technological determinants of haemodialysis membranes. Clin Kidney J. 2021 Dec;14(Suppl 4):i5–i16. doi: 10.1093/ckj/sfab184. https://europepmc.org/abstract/MED/34987782 .sfab184 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Torii Y, Yamada S, Yajima M, Sugata T. Polymethylmethacrylate membrane dialyzer: Historic but modern. Blood Purif. 2023 Oct 20;52 Suppl 1(Suppl 1):8–14. doi: 10.1159/000526996. https://doi.org/10.1159/000526996 .000526996 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Molina P, Peiró J, Martínez-Gómez MA, Vizcaíno B, Esteller C, González-Moya M, García-Valdelvira M, Molina MD, Maduell F. Combining diffusion, convection and absorption: A pilot study of polymethylmethacrylate versus polysulfone membranes in the removal of P-cresyl sulfate by postdilution on-line hemodiafiltration. Kidney and Dialysis. 2021 Nov 09;1(2):121–134. doi: 10.3390/kidneydial1020015. [DOI] [Google Scholar]
- 21.Santoro A, Guadagni G. Dialysis membrane: from convection to adsorption. NDT Plus. 2010 May 15;3(Suppl 1):i36–i39. doi: 10.1093/ndtplus/sfq035. https://europepmc.org/abstract/MED/27045937 .sfq035 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Kudo M, Chihara S, Tatsumi H, Kazuma S, Masuda Y. Comparison of pre-dilution and post-dilution methods on cytokine clearance using polymethylmethacrylate (PMMA) membrane hemofilters in continuous hemodiafiltration. Cureus. 2025 Jan;17(1):e77500. doi: 10.7759/cureus.77500. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Belmouaz M, Bauwens M, Hauet T, Bossard V, Jamet P, Joly F, Chikhi Elise, Joffrion Sandrine, Gand Elise, Bridoux Frank. Comparison of the removal of uraemic toxins with medium cut-off and high-flux dialysers: a randomized clinical trial. Nephrol Dial Transplant. 2020 Feb 01;35(2):328–335. doi: 10.1093/ndt/gfz189.5580323 [DOI] [PubMed] [Google Scholar]
- 24.Krieter DH, Rüth Marieke, Lemke H, Wanner C. Clinical performance comparison of two medium cut-off dialyzers. Ther Apher Dial. 2023 Apr 16;27(2):284–292. doi: 10.1111/1744-9987.13919. [DOI] [PubMed] [Google Scholar]
- 25.Moher D, Liberati A, Tetzlaff J, Altman DG, PRISMA Group Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. Int J Surg. 2010;8(5):336–341. doi: 10.1016/j.ijsu.2010.02.007. https://linkinghub.elsevier.com/retrieve/pii/S1743-9191(10)00040-3 .S1743-9191(10)00040-3 [DOI] [PubMed] [Google Scholar]
- 26.Liberati A, Altman DG, Tetzlaff J, Mulrow C, Gøtzsche Peter C, Ioannidis JPA, Clarke M, Devereaux PJ, Kleijnen J, Moher D. The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate health care interventions: explanation and elaboration. J Clin Epidemiol. 2009 Oct;62(10):e1–34. doi: 10.1016/j.jclinepi.2009.06.006. https://linkinghub.elsevier.com/retrieve/pii/S0895-4356(09)00180-2 .S0895-4356(09)00180-2 [DOI] [PubMed] [Google Scholar]
- 27.Stroup DF, Berlin J A, Morton S C, Olkin I, Williamson G D, Rennie D, Moher D, Becker B J, Sipe T A, Thacker S B. Meta-analysis of observational studies in epidemiology: a proposal for reporting. Meta-analysis Of Observational Studies in Epidemiology (MOOSE) group. JAMA. 2000 Apr 19;283(15):2008–2012. doi: 10.1001/jama.283.15.2008.jst00003 [DOI] [PubMed] [Google Scholar]
- 28.Sterne JAC, Savović J, Page MJ, Elbers RG, Blencowe NS, Boutron I, Cates CJ, Cheng H, Corbett MS, Eldridge SM, Emberson JR, Hernán Miguel A, Hopewell S, Hróbjartsson Asbjørn, Junqueira DR, Jüni Peter, Kirkham JJ, Lasserson T, Li T, McAleenan A, Reeves BC, Shepperd S, Shrier I, Stewart LA, Tilling K, White IR, Whiting PF, Higgins JPT. RoB 2: a revised tool for assessing risk of bias in randomised trials. BMJ. 2019 Aug 28;366:l4898. doi: 10.1136/bmj.l4898. https://eprints.whiterose.ac.uk/150579/ [DOI] [PubMed] [Google Scholar]
- 29.Sterne JA, Hernán Miguel A, Reeves BC, Savović J, Berkman ND, Viswanathan M, Henry D, Altman DG, Ansari MT, Boutron I, Carpenter JR, Chan A, Churchill R, Deeks JJ, Hróbjartsson Asbjørn, Kirkham J, Jüni Peter, Loke YK, Pigott TD, Ramsay CR, Regidor D, Rothstein HR, Sandhu L, Santaguida PL, Schünemann Holger J, Shea B, Shrier I, Tugwell P, Turner L, Valentine JC, Waddington H, Waters E, Wells GA, Whiting PF, Higgins JP. ROBINS-I: a tool for assessing risk of bias in non-randomised studies of interventions. BMJ. 2016 Oct 12;355:i4919. doi: 10.1136/bmj.i4919. https://www.bmj.com/lookup/pmidlookup?view=long&pmid=27733354 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Higgins JPT, Thompson SG. Quantifying heterogeneity in a meta-analysis. Stat Med. 2002 Jun 15;21(11):1539–1558. doi: 10.1002/sim.1186. [DOI] [PubMed] [Google Scholar]
- 31.Borenstein M, Hedges LV, Higgins JP, Rothstein HR. A basic introduction to fixed-effect and random-effects models for meta-analysis. Res Synth Methods. 2010 Apr 21;1(2):97–111. doi: 10.1002/jrsm.12. [DOI] [PubMed] [Google Scholar]
- 32.Duval S, Tweedie R. Trim and fill: A simple funnel-plot-based method of testing and adjusting for publication bias in meta-analysis. Biometrics. 2000 Jun 24;56(2):455–463. doi: 10.1111/j.0006-341x.2000.00455.x. [DOI] [PubMed] [Google Scholar]
- 33.Begg CB, Mazumdar M. Operating characteristics of a rank correlation test for publication bias. Biometrics. 1994 Dec;50(4):1088. doi: 10.2307/2533446. [DOI] [PubMed] [Google Scholar]
- 34.Egger M, Davey Smith G, Schneider M, Minder C. Bias in meta-analysis detected by a simple, graphical test. BMJ. 1997 Sep 13;315(7109):629–634. doi: 10.1136/bmj.315.7109.629. https://europepmc.org/abstract/MED/9310563 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Roy A, De S. State-of-the-art materials and spinning technology for hemodialyzer membranes. Sep Purif Rev. 2016 Nov 03;46(3):216–240. doi: 10.1080/15422119.2016.1256323. [DOI] [Google Scholar]
- 36.Somji SS, Ruggajo P, Moledina S. Adequacy of hemodialysis and its associated factors among patients undergoing chronic hemodialysis in Dar es Salaam, Tanzania. Int J Nephrol. 2020 Feb 10;2020:9863065–8963066. doi: 10.1155/2020/9863065. https://doi.org/10.1155/2020/9863065 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Ambarsari CG, Trihono PP, Kadaristiana A, Rachmadi D, Andriastuti M, Puspitasari HA, Tambunan T, Pardede SO, Mangunatmadja I, Hidayati EL. Low-dose maintenance intravenous iron therapy can prevent anemia in children with end-stage renal disease undergoing chronic hemodialysis. Int J Nephrol. 2020 Jun 01;2020:3067453–3067458. doi: 10.1155/2020/3067453. https://doi.org/10.1155/2020/3067453 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Hasan LM, Shaheen DAH, El Kannishy GAH, Sayed-Ahmed NAH, Abd El Wahab AM. Is health-related quality of life associated with adequacy of hemodialysis in chronic kidney disease patients? BMC Nephrol. 2021 Oct 07;22(1):334. doi: 10.1186/s12882-021-02539-z. https://www.biomedcentral.com/1471-2369/22/334 .10.1186/s12882-021-02539-z [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Canaud B, Kooman JP, Selby NM, Taal MW, Francis S, Maierhofer A, Kopperschmidt P, Collins A, Kotanko P. Dialysis-induced cardiovascular and multiorgan morbidity. Kidney Int Rep. 2020 Nov;5(11):1856–1869. doi: 10.1016/j.ekir.2020.08.031. https://linkinghub.elsevier.com/retrieve/pii/S2468-0249(20)31510-2 .S2468-0249(20)31510-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Casino FG, Mancini E, Santarsia G, Mostacci SD, D'Elia Filomena, Di Carlo M, Iannuzzella F, Rossi L, Vernaglione L, Grimaldi D, Rapanà Renato, Basile C. What volume to choose to assess online Kt/V? J Nephrol. 2020 Feb 7;33(1):137–146. doi: 10.1007/s40620-019-00636-9.10.1007/s40620-019-00636-9 [DOI] [PubMed] [Google Scholar]
- 41.National Kidney Foundation KDOQI clinical practice guideline for hemodialysis adequacy: 2015 update. Am J Kidney Dis. 2015 Nov;66(5):884–930. doi: 10.1053/j.ajkd.2015.07.015.S0272-6386(15)01019-7 [DOI] [PubMed] [Google Scholar]
- 42.Kirkman DL, Scott M, Kidd J, Macdonald JH. The effects of intradialytic exercise on hemodialysis adequacy: A systematic review. Semin Dial. 2019 Jul 09;32(4):368–378. doi: 10.1111/sdi.12785. [DOI] [PubMed] [Google Scholar]
- 43.Vanholder R, Glorieux G, Eloot S. Once upon a time in dialysis: the last days of Kt/V? Kidney Int. 2015 Sep;88(3):460–466. doi: 10.1038/ki.2015.155. https://linkinghub.elsevier.com/retrieve/pii/S2157-1716(15)32213-9 .S2157-1716(15)32213-9 [DOI] [PubMed] [Google Scholar]
- 44.Eloot S, Van Biesen W, Glorieux G, Neirynck N, Dhondt A, Vanholder R. Does the adequacy parameter Kt/V(urea) reflect uremic toxin concentrations in hemodialysis patients? PLoS One. 2013 Nov 13;8(11):e76838. doi: 10.1371/journal.pone.0076838. https://dx.plos.org/10.1371/journal.pone.0076838 .PONE-D-12-39055 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Clark W, Dehghani N, Narsimhan V, Ronco C. Uremic toxins and their relation to dialysis efficacy. Blood Purif. 2019 Sep 27;48(4):299–314. doi: 10.1159/000502331.000502331 [DOI] [PubMed] [Google Scholar]
- 46.Palupi-Baroto R, Hermawan K, Murni IK, Nurlita T, Prihastuti Y, Puspitawati I, Tandri CC, Ambarsari CG. Carotid intima-media thickness, fibroblast growth factor 23, and mineral bone disorder in children with chronic kidney disease. BMC Nephrol. 2024 Oct 21;25(1):369. doi: 10.1186/s12882-024-03771-z. https://www.biomedcentral.com/1471-2369/25/369 .10.1186/s12882-024-03771-z [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Palupi-Baroto R, Hermawan K, Murni IK, Nurlitasari T, Prihastuti Y, Sekali DRK, Ambarsari CG. High fibroblast growth factor 23 as a biomarker for severe cardiac impairment in children with chronic kidney disease: A single tertiary center study. IJNRD. 2021 Jun;Volume 14:165–171. doi: 10.2147/ijnrd.s304143. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Ding L, Johnston J, Pinsk MN. Monitoring dialysis adequacy: history and current practice. Pediatr Nephrol. 2021 Aug 05;36(8):2265–2277. doi: 10.1007/s00467-020-04816-9.10.1007/s00467-020-04816-9 [DOI] [PubMed] [Google Scholar]
- 49.Bonomini M, Pieroni L, Di Liberato L, Sirolli V, Urbani A. Examining hemodialyzer membrane performance using proteomic technologies. TCRM. 2017 Dec;Volume 14:1–9. doi: 10.2147/tcrm.s150824. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Said N, Lau WJ, Ho Y, Lim SK, Zainol Abidin MN, Ismail AF. A review of commercial developments and recent laboratory research of dialyzers and membranes for hemodialysis application. Membranes (Basel) 2021 Oct 07;11(10):767. doi: 10.3390/membranes11100767. https://www.mdpi.com/resolver?pii=membranes11100767 .membranes11100767 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Fraser SD, Fenton A, Harris S, Shardlow A, Liabeuf S, Massy ZA, Burmeister A, Hutchison CA, Landray M, Emberson J, Kalra P, Ritchie JP, Cockwell P, Taal MW. The association of serum free light chains with mortality and progression to end-stage renal disease in chronic kidney disease: systematic review and individual patient data meta-analysis. Mayo Clin Proc. 2017 Nov;92(11):1671–1681. doi: 10.1016/j.mayocp.2017.08.021.S0025-6196(17)30644-4 [DOI] [PubMed] [Google Scholar]
- 52.Wolley M, Hutchison C. Large uremic toxins: an unsolved problem in end-stage kidney disease. Nephrol Dial Transplant. 2018 Oct 01;33(suppl_3):iii6–iii11. doi: 10.1093/ndt/gfy179. https://europepmc.org/abstract/MED/30281131 .5114425 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Kim TH, Kim S, Kim TY, Park HY, Jung KS, Lee MH, Jhee JH, Lee JE, Choi HY, Park HC. Removal of large middle molecules via haemodialysis with medium cut-off membranes at lower blood flow rates: an observational prospective study. BMC Nephrol. 2019 Dec 31;21(1):2. doi: 10.1186/s12882-019-1669-3. https://www.biomedcentral.com/1471-2369/21/2 .10.1186/s12882-019-1669-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Lindgren A, Fjellstedt E, Christensson A. Comparison of hemodialysis using a medium cutoff dialyzer versus hemodiafiltration: A controlled cross-over study. IJNRD. 2020 Oct;Volume 13:273–280. doi: 10.2147/ijnrd.s263110. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Ward RA, Vienken J, Silverstein DM, Ash S, Canaud B. Regulatory considerations for hemodiafiltration in the United States. CJASN. 2018 Apr 13;13(9):1444–1449. doi: 10.2215/cjn.12641117. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Canaud B, Köhler Katrin, Sichart J, Möller Stefan. Global prevalent use, trends and practices in haemodiafiltration. Nephrol Dial Transplant. 2020 Mar 01;35(3):398–407. doi: 10.1093/ndt/gfz005.5320340 [DOI] [PubMed] [Google Scholar]
- 57.Kawanishi H. Middle molecular uremic toxin and blood purification therapy. J Clin Med. 2024 Jan 23;13(3):647. doi: 10.3390/jcm13030647. https://www.mdpi.com/resolver?pii=jcm13030647 .jcm13030647 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Ward RA. Do clinical outcomes in chronic hemodialysis depend on the choice of a dialyzer? Seminars in Dialysis. 2011 Feb 17;24(1):65–71. doi: 10.1111/j.1525-139x.2010.00807.x. [DOI] [PubMed] [Google Scholar]
- 59.Haroon S, Davenport A. Choosing a dialyzer: What clinicians need to know. Hemodial Int. 2018 Oct 08;22(S2):S65–S74. doi: 10.1111/hdi.12702. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
PRISMA-P checklist.
Search terms.