Hyperparameter-controlled regularized reconstruction method based on object structure and acquisition conditions in SPECT

Tomoya Minagawa 1,2,, Kensuke Hori 3, Takeyuki Hashimoto 2
PMCID: PMC12307274  PMID: 40728756

Abstract

Background

In clinical nuclear medicine, reconstruction methods incorporating regularization terms have been widely investigated. However, searching for optimal hyperparameters for the entire examination is time-consuming and arduous because the optimal hyperparameters need to be determined experimentally and vary depending on factors, including the acquisition condition, reconstruction condition, and so on. In this study, we propose a row-action type automatic regularized expectation maximization method (RAREM). This method considers the acquisition conditions and object structure for determining the hyperparameters and does not require the user to set the hyperparameters experimentally. This study was conducted using numerical simulations and a real SPECT system

Methods

Total variation-expectation maximization (TV-EM) and modified-block sequential regularized EM (BSREM) were compared with RAREM, with the optimal hyperparameters of the two conventional reconstruction methods determined in advance from normalized root mean square error (NRMSE) results. This simulation examination utilized three types of phantoms with the number of counts and projections being examined in six ways each, resulting in a total of 108 conditions. The NRMSE and structural similarity index measure (SSIM) were used to evaluate of the simulation examination, and the Mann–Whitney U test was used for statistical analysis. In the real examination, two types of phantoms were used, and the number of projections was examined in three ways, for a total of six conditions. Contrast recovery coefficient (CRC) and specific binding ratio (SBR) were used to evaluate the real examination

Results

The NRMSE, CRC, and SBR of RAREM were equivalent to those of the conventional methods, and the SSIM of RAREM was equivalent to or better than that of the conventional methods, with significant differences in some cases. The results indicated that RAREM worked well with the evaluated object structure and considered the acquisition conditions

Conclusion

In this study, an automatically controlled regularization reconstruction method was proposed. The proposed method does not require the user to set hyperparameters experimentally and can avoid the investigation of optimal hyperparameters; it is an alternative to conventional regularized methods in clinical

Keywords: SPECT, Regularized reconstruction method, Automatic controlled hyperparameters

Background

SPECT can be used to obtain functional images, which are important in clinical practice. However, obtaining high-quality images is time-consuming, and SPECT images exhibit a higher level of noise than the other modalities such as CT, MRI, and PET. In particular, compared with PET, which does not use a collimator, SPECT is limited by its low sensitivity and is significantly affected by noise. Furthermore, when the administered dose is low, or a nuclide with a long half-life is used, ensuring an adequate total count is challenging. Hence, the selection of appropriate acquisition conditions and reconstruction methods is important.

Reconstruction methods for overcoming high-level noise have been extensively investigated. Row-action type reconstruction methods introduced relaxation parameter to reduce noise [1, 2]. Particularly in DRAMA, the relaxation parameter is automatically controlled for each sub-iteration, making the noise propagation to the reconstructed image constant [2]. This process improves the signal-to-noise ratio compared to conventional iterative reconstruction methods, such as ML-EM and OS-EM. Nevertheless, image quality is degraded when data have a high noise level.

Another approach to suppress noise is the incorporation of a regularization term as a prior information of images into ML-EM, called the MAP-EM algorithm, which is based on Bayes’ theorem [3]. TV-EM is a typical form of the MAP-EM algorithm that uses a TV regularization term [4]. It can reduce noise and is often used for sparse-view reconstruction [58]. In clinical applications, BSREM, which is also a MAP-EM algorithm, used for Q.Clear (GE Healthcare, Milwaukee, WI, USA) is the most commonly used regularized reconstruction method for PET [911]. It was reported that the BSREM improved the quality of reconstructed images and lesion detectability [12, 13]. However, several hyperparameters must be determined experimentally. In the modified-BSREM proposed by Ahn et al., the variable parameters are the number of subsets, the coefficient of the regularization term, the initial value of the relaxation parameter, and its rate of decrease at each iteration [10]. Because the optimal parameters vary depending on various factors, such as the acquisition conditions and objective lesion, etc., searching for all optimal parameters requires considerable time and effort [1215]. Recently, SPECT systems that can use BSREM (or modified-BSREM) have become available [16]. Therefore, in SPECT, the search for optimal parameters for each purpose of examination is expected to increase.

In this study, we propose the automatic reconstruction algorithm, which called the RAREM, to omit the investigation of optimal hyperparameters. This method has two main features. The first is an automatic hyperparameter control system that considers object structure and acquisition conditions. Using spatial filter, the object structure was evaluated based on the edges of the update images. The second is the observance of the non-negativity constraint by considering the value that the regularization term can assume. This study aimed to evaluate whether the performance of RAREM is equivalent to or better than that of conventional regularized reconstruction methods using optimal hyperparameters. The study was conducted using numerical simulations and a real SPECT system.

Methods

TV-EM and modified-BSREM were compared with RAREM. In this study, scatter, attenuation, and collimator aperture corrections were not used to compare the differences between image reconstruction methods.

Conventional reconstruction method

In SPECT, the projection data follows the Poisson distribution Py|x.

P(y|x)=i=1Mj=12Cijxjyiyi!exp-j=1N2Cijxj, 1

where, xj is the radioactivity concentration in the j-th pixel of matrix size N2, yi is the projection data in the i-th projection data of M, and Cij is the detection probability from the j-th image pixel to the i-th projection data. The ML method estimates x maximizing the logarithm of Equation (1), which is the log-likelihood function Lx, as follows [17]:

Lx=i=1Myilnj=1N2Cijxj-j=1N2Cijxj. 2

The MAP-EM algorithm, based on Bayes’ theorem, maximizes the log-likelihood function (2) and the prior function. Bayes’ theorem is as follows:

Px|y=Py|x·PxPy, 3

where, Px is the prior function and is assumed to be a Gibbs distribution [4].

Pxexp-ηUx, 4

where, Ux is the energy function, and η is a regularization parameter. Using Equations (2), (3), and (4), the reconstruction problem of the MAP-EM algorithm is the logarithm of the posterior probability shown in the following equation:

argmaxxlnPx|y=argmaxxLx+lnPx=argmaxxi=1Myilnj=1N2Cijxj-j=1N2Cijxj-ηUx. 5

The MAP-EM algorithm is derived to solve problem (5) and is shown in the following equation [3]:

xjk+1=xjki=1MCij+ηxjUx|x=xki=1MyiCijn=1N2Cinxnk, 6

where, k is the number of main iterations. In the TV-EM algorithm, the TV norm is used for Ux, and is expressed as follows [4]:

UTVx=s,txs+1,t-xs,t2+xs,t+1-xs,t2, 7

where, s and t are the pixel numbers in the horizontal and vertical directions of the two-dimensional image, respectively. As the TV norm is not a differentiable function, an approximate formula for the partial derivative of the TV norm by xs,t was used.

UTVxxs,t=xs,t-xs-1,txs,t-xs-1,t2+xs-1,t+1-xs-1,t2+ε2+xs,t-xs,t-1xs+1,t-1-xs,t-12+xs,t-xs,t-12+ε2-xs+1,t+xs,t+1-2xs,txs+1,t-xs,t2+xs,t+1-xs,t2+ε2, 8

where, ε is an artificial parameter to ensure the differentiability around a point where Ux=0 and usually set to a small value. A previous study mentioned that ε should be set to 1% or less of the expected maximum value of x, and that TV-EM is not sensitive to ε [4]. Therefore, in this study, ε was set to 0.001.

BSREM is a relaxed type of regularization reconstruction method, which is an extension of RAMLA to the regularized case and is derived by differentiating Lx and Px [9]. The BSREM algorithm is expressed as follows:

xjk,q+1/2=xjk,q+λkxjk,qiSqCijyin=12Cinxjk,q-1, 9
xjk,q+1=xjk,q+1/2+λkxjk,q+1/2ηxjUx|x=xk,q+1/2, 10
xjk+1,0=xjk,Q, 11

where, λ is the relaxation parameter, q is the number of sub-iterations, Sq represents all elements in the q-th sub-iteration, and Q is the total number of sub-iterations. As the name implies, the modified-BSREM is a modified version of the BSREM proposed to converge to global solutions under realistic conditions [10]. In the modified-BSREM, the update formula and relaxation parameters are controlled to not exceed the upper and lower bounds to comply with the constraints, and non-negative value processing is included in the algorithm (Table I and II in [10]). The relaxation parameter of the modified-BSREM is given by:

λk=λ0γk+1, 12

where, λ0 is the initial value of the relaxation parameter, and γ is the rate for decreasing the relaxation parameter λ for each main iteration k. The user is required to experimentally set λ0, γ, Q, and η to use the modified-BSREM appropriately. In the modified-BSREM, since the one of the conditions for convergence to optimal solution is k=0λk=, a reconstruction is almost independent of γ [10]. We verified this in a preliminary experiment, and the influence of γ on reconstruction is small when γ is small enough. Therefore, in this study, γ was set at 0.1 and Q included at least three projections in each subset [10].

Proposed method

The image reconstruction method of RAREM was derived by referring to RAMLA and DRAMA using MAP estimation [14, 9]. In RAMLA,

gqx=iSqyilnj=1N2Cijxj-j=1N2Cijxj, 13
xjk,q+1=xjk,q+λkxjk,qgqxxj, 14

are derived by incorporating the row-action algorithm into the ML estimation in Equation (2), where, gqx is the log-likelihood function Lx in the row-action algorithm. Equation (14) is equivalent to Equation (9) in the BSREM algorithm. The goal of RAREM is to maximize Equation (5), which is derived from the logarithms of posterior probability Px|y. Therefore, gqx can be rewritten as

gqx=iSqyilnj=1N2Cijxj-j=1N2Cijxj-ηUx 15

and differentiated with respect to xj to maximize it, the RAREM algorithm is derived as follows:

xjk,q+1=xjk,q+λk,qxjk,qiSqCijyin=12Cinxjk,q-1-ηkxjUx|x=xk,q, 16
xjk+1,0=xjk,Q. 17

In RAREM, the number of subsets is set to the maximum, as in RAMLA and DRAMA, and each parameter is calculated automatically considering the objective structure and acquisition conditions. In this study, the TV norm in Equation (8) is used for regularization term UTVx/xs,t. The regularization parameter ηk is calculated by the following equation:

ηk=1Ek×0.05×1+Aproj+0.3×Acount, 18
Aproj=maxlog10MNqM,0, 19
Acount=maxlog10N128×107T,0, 20

where, MNq is the number of projections that satisfy the Nyquist-Shannon sampling theorem, and T is the total number of counts. maxa,b indicates the calculation that selects a higher value a or b. Aproj can be thought of as the term relating to the number of projections, and Acount as the term relating to the counts. Ek is the evaluation term of the object structure. A schematic of the calculation of Ek is shown in Fig. 1, Ek is calculated using the following equation:

Fig. 1.

Fig. 1

The schematic calculation of Ek. Ek is the ratio of the sum of the absolute pixel value of the updated image xk to that of the edge image xedgek

Ek=100×xedgek1xk1, 21
xedgek=xkGL, 22

where, xedgek1 is the L1 norm of the updated image processed by the Gaussian filter G and Laplacian filter L for the calculation of Ek only. In this study, the standard deviation σ of the Gaussian filter, with a kernel size of 5×5, was set to the following equation:

σ=0.4×1+log10120M×104T/M1/2. 23

All of the pixel values of the initial image x0 were 1.0, but the image reconstructed by DRAMA was used for xedge0 with main iteration=MNqM+1, where · is a floor function.

The relaxation parameter λk,q is calculated by the following equation:

λk,q=β0β0+q+γkM×11+log10MNqM×11+ηkVmax. 24

If the acquisition condition was MNq<M, MNq/M is set to 1.0. Equation (24) can be devided into three factors. The initial factor is the width of the decrease in λ utilized in DRAMA, with γ generally set to 1.0 [2]. β0 is the parameter calculated by the average of geometrical coefficients among the projection lines in PET and approximately balances the noise propagation from all subsets at the end of each main iteration [2]. In other imaging modalities, β0 is calculated by approximate expression [18, 19].

β00.72×1sfwhmN1.4M0.4, 25

where, sfwhm is defined as the FWHM of the post-smoothing filter, however, since this study did not use the post-smoothing filter, the TV norm is assumed to be the Gaussian filter with σ=1.3. sfwhm was calculated by the following equation, which relates FWHM to standard deviation σ:

FWHM=2σ2ln2, 26

resulting in sfwhm=3.06. The second factor in Equation (24) is the factor taking into account the number of projections. The third factor in Equation (24) is to satisfy the non-negativity constraint in the RAREM algorithm, which contributes to the fact that RAREM does not require non-negative constraints and the process of replacing negative value with non-negative value in effect. To derive this factor, Equation (16) is transformed:

xjk,q+1=xjk,q[1+λk,qiSqCijyin=12Cinxjk,q-1-ηkxjUx|x=xk,q. 27

As xjk,q is a non-negative value, we need to consider only the inside of the square brackets in Equation (27). Moreover, since λk,q is also a non-negative value, we consider when the inside of the curly brackets in Equation (27) is a minimum value. The minimum value for the first term inside the curly brackets is -1, because both Cij and yi are also non-negative values. The minimum value for the second term inside that is when xjUx|x=xk,q is a maximum value Vmax. From the above, we can obtain the following inequality to ensure that xjk,q+1 is a non-negative value:

1+λk,q-1-ηkVmax>0. 28

Finally, inequality (28) is transformed, and the third factor in Equation (16) is obtained using the following inequality:

λk,q<11+ηkVmax. 29

The Vmax of the partial derivative of the TV norm (8) is 2+2, and the proof of this is shown in Appendix A.

Experimental methods

Simulation condition

In this simulation, a 3-dimensional mathematic phantom (3D-MAC phantom: Nuclear Medicine Section, Japanese Society of Radiological Technology), IB-20 Advanced (Striatum phantom: Kyoto Kagaku Co., Ltd., Kyoto, Japan), and BrainWeb phantom (https://brainweb.bic.mni.mcgill.ca/brainweb/) were used (Fig. 2) [20]. Only the striatum phantom was constructed using the CT data. The components and set radioactivity concentrations of the three phantoms were shown in Table 1. The matrix size and pixel size were 128×128 and 2.0mm×2.0mm, respectively. The total projection angle was set to 360. The total number of projections was varied from 12 to 120. Poisson noise was added to the projection data, varying from 2.5×103 to 5.0×104 counts/projection [2124].

Fig. 2.

Fig. 2

Schematic diagram of phantoms for the simulation experiments. a 3D-MAC phantom, b IB-20 Advanced (striatum phantom), c BrainWeb phantom

Table 1.

The components and the ratio of radioactivity concentration of simulation phantoms

Radioactivity concentration
3D-MAC Background 1 Cold 0 Hot A 2 Hot B 4
Striatum Parenchyma 1 Ventricles 0 R striatum 8 L Striatum 4 Other 0.75
BrainWeb Background 1 WM 5 GM 20

WM and GM indicate white and gray matter, respectively

Acquisition conditions for real phantom data

In this experiment, the Jaszczak phantom (Data Spectrum Corporation, Durham, North Carolina, USA) and IB-20 Advanced (Striatum phantom) were used (Fig. 3). The Jaszczak phantom is a cylindrical phantom containing six hot spheres, with the remaining space being the background region. The diameters of six hot spheres were 9.5, 12.7, 15.9, 19.1, 25.4 and 31.8 mm. It was filled with 99mTc-pertechnetate, which had an approximately 4:1 radioactivity ratio (hot spheres, 26.2 kBq/mL; background, 7.3 kBq/mL). Striatum phantom was also filled with 99mTc-pertechnetate, which had an approximately 8:4:1 radioactivity ratio (right striatum, 56.0 kBq/mL; left striatum, 27.0 kBq/mL; brain parenchyma, 7.0 kBq/mL).

Fig. 3.

Fig. 3

Images of the phantoms used in the real phantom experiment. a, b show the hot sphere and uniform slices in the Jaszczak phantom, respectively, and c show the striatum phantom. The top and bottom rows show the CT and SPECT images, respectively, showing the arrangement of the ROI and VOI. The ROIs at the bottom of a were set to the same diameter as those of each of the six hot spheres (9.5, 12.7, 15.9, 19.1, 25.4, and 31.8mm). The ROI at the bottom of b was selected as the area containing 85% of the uniform slice. The bottom of c denotes the striatum VOI and reference VOI at the striatum level in the striatum phantom.

SPECT data were acquired using a Symbia Intevo SPECT/CT system (Siemens Healthcare, Erlangen, Germany) equipped with a low-energy high resolution collimator. All projection data were acquired in step and shoot mode with a circular orbit of 360, and the radius of rotation was set at 15cm. The number of projections was varied from 30 to 120. The matrix size and pixel size were 128×128 and 3.3mm×3.3mm (scale for enlargement, 1.45), respectively. The energy window was 140keV±7.5%. The acquisition time was set to be approximately 1.0×103counts/projection×slice, a total count of approximately 3.0×106 counts for a 90mm high striatum phantom [21].

Image evaluation

Normalized root mean square error (NRMSE) and structural similarity index measure (SSIM) were used to evaluate the image quality of the simulation experiment. The NRMSE, which evaluates the accuracy of the reconstructed images, was used to determine the parameters used in conventional reconstruction methods in the simulation and real experiment in terms of the importance of quantification, and calculated as follows:

NRMSE%=100×j=1N2bj-aj2j=1N2aj2, 30

where, a is the reference image, and b is the reconstructed image. In the simulation experiment, the original image was used as the reference image. In the real experiment, the projection data obtained by simulation of an imaging nuclear detector (SIMIND), which is one of the Monte-Calro simualtion, under the same geometry as in the real experiment, with low noise and a sufficient number of projections, were reconstructed using ML-EM for the reference image [25, 26]. A smaller NRMSE, which is the difference between the reference and reconstructed images, represents better results.

SSIM is an evaluation index used to evaluate visual assessment close to human subjectivity and is calculated as follows:

SSIM=1N2j=1N2(2μjaμjb+c1)(2σjab+c2)μja2+μjb2+c1σja2+σjb2+c2, 31

where, μja and μjb are the mean values, σja and σjb are the standard deviations, and σjab is the covariance in the square window centered on the j-th pixel of the images a and b, respectively. The size of the square window is 5×5 [27]. c1 and c2 were calculated as 0.01L2 and 0.03L2, respectively, where L is the dynamic range of the image. Because the SSIM indicates the similarity between the original and reconstructed images, the closer the reconstructed image is to the original image, the closer the SSIM is to 1.

The contrast recovery coefficient (CRC) and specific binding ratio (SBR) were used to evaluate the experimental results. The CRC is a quantitative index used to evaluate the effect of the reconstruction method on spatial resolution and contrast. The CRC of the Jaszczak phantom is calculated as:

CRC=μhotμuni-1/TruehotTrueuni-1, 32

where, μhot is the mean value of each hot sphere ROI, and μuni is that of the uniform region ROI. Truehot and Trueuni are the true activity concentration of each hot sphere ROI and that of the uniform region ROI. The circular ROIs of the hot sphere were set to the same diameter as the sphere diameter, and a uniform circular ROI was selected for the area containing 85% of the uniform slice (Fig. 3). The CRC is the ratio of the true radioactivity contrast to the contrast of the reconstructed image. The closer the contrast ratio of the reconstructed image is to the true contrast ratio, the closer the CRC is to 1.

The SBR (SBRBolt) of the striatum, a quantitative index for diagnosis of dementia was calculated as follows:

SBR=CtstrCref-VVOIVstr, 33

where, Ctstr and Cref are the total count of the striatal VOI and counts per volume (mL) of the reference VOI, respectively (Fig. 3) [28]. VVOI is the volume of the striatum VOI, and Vstr is the standard volume of the striatum (11.2mL). The SBR is defined as the ratio of the radioactivity concentration in the striatum to that in the background region. The true SBR value varies depending on the administered dose. Here, the true SBR values for the right and left striatum were 7.00 and 2.86, respectively. The SBR was used to evaluate the clinical applications of RAREM.

Statistical analysis

The NRMSE and SSIM were compared between TV-EM and RAREM and modified-BSREM and RAREM, respectively, using the Mann–Whitney U test. Differences were considered statistically significant when p-values were less than 0.05.

Hyperparameter selection of conventional reconstruction method

Considering the importance of quantification in nuclear medicine, the parameters with the best NRMSE values for all acquisition conditions were selected for each conventional reconstruction method. λ0 and η were varied in nine steps (0.2–1.0) and 10 steps (0.001–0.01, 0.01-0.1, or 0.1-1.0) in each search, respectively [1]. If the maximum or minimum value of the varied parameter range indicated the best NRMSE, the optimal parameter was determined after three additional steps.

In the real experiment, because the original image could not be obtained, a reference image for the calculation of the NRMSE was created by SIMIND. The projection data of the reference image was obtained under the same geometry as that in the real experiment, with a sufficient number of counts and projections. ML-EM was used to reconstruct the reference images.

Results

Simulation experiment

The reconstructed images are shown in Fig. 4. The reconstructed images of the 3D-MAC phantom retained the shape of all circular regions under almost all acquisition conditions and reconstruction methods. In contrast, the reconstructed images of the striatum and BrainWeb phantoms did not retain the shape of the structures under poor acquisition conditions, mainly when the number of projections was less than 18. The images of all the phantoms reconstructed by TV-EM were noisier than those reconstructed by the other methods. The reconstructed images showed no significant visual difference between modified-BSREM and RAREM under almost all acquisition conditions.

Fig. 4.

Fig. 4

The reconstructed images of the simulation experiment when the counts per projection were 104. A, B, and C are the TV-EM, modified-BSREM, and RAREM, respectively. “proj” indicates the number of projections

The NRMSE and SSIM are shown in Figs. 5 and 6, respectively. The NRMSE and SSIM are the mean values of all slices for each acquisition condition. Under almost all acquisition conditions, the NRMSEs of the TV-EM were higher, and the SSIMs were lower than those of the others. This was consistent with the visual assessment presented in Fig. 4. In addition, the NRMSEs were not significantly different between the modified-BSREM and RAREM, whereas the SSIMs of RAREM were slightly higher than those of modified-BSREM, especially when the acquisition conditions were poor. As shown in Fig. 4, modified-BSREM did not outperform RAREM in Figs. 5 and 6. Moreover, the SSIM of the modified-BSREM was low under certain acquisition conditions (such as 18 projection and 2.5×103 counts/projection in the BrainWeb phantom). Contrastingly, those reconstructed using RAREM were stable regardless of the acquisition conditions, even if the number of counts was poor.

Fig. 5.

Fig. 5

NRMSEs for each reconstruction method and phantom. a, b and c are the 3D-MAC phantom, striatum phantom, and BrainWeb phantom, respectively. “kcounts” indicates 103 counts

Fig. 6.

Fig. 6

SSIMs for each reconstruction method and phantom. a, b and c are the 3D-MAC phantom, striatum phantom, and BrainWeb phantom, respectively. “kcounts” indicates 103 counts

The images reconstructed by the modified-BSREM with different parameters that were optimal for the other phantom are shown in Fig. 7. Focusing diagonally from the top left to the bottom right in Fig. 7, the images were reconstructed using the optimal parameters. The reconstructed image of the 3D-MAC phantom using the optimal parameters for the BrainWeb phantom (top right in Fig. 7) was noisier, whereas that of the BrainWeb phantom using the optimal parameters for the 3D-MAC phantom (bottom left in Fig. 7) was excessively smoothed and lost detailed structures. Moreover, in the striatum phantom with optimal parameters for the 3D-MAC phantom (center left in Fig. 7), it was more difficult to visualize the ventricles and showed an overestimation of striatal areas owing to smoothing. Conversely, using the optimal parameters for the BrainWeb phantom (center right in Fig. 7), the image was noisier, and the left striatum structure appeared collapsed.

Fig. 7.

Fig. 7

The images reconstructed by modified-BSREM with a different parameter that is optimal for the other phantoms. The acquisition conditions are 60 projections and 5.0×103counts/projection. The rows indicate the phantom used, and the columns indicate the reconstruction using the optimal parameters for the noted phantoms. The blue frames indicate the images reconstructed with optimal parameters, and the other images indicate the images reconstructed with non-optimal parameters

Real phantom experiment

The reconstructed images are shown in Fig. 8. The smallest sphere of the Jaszczak phantom could not be observed under any of the acquisition conditions. The images of the Jaszczak phantom reconstructed by TV-EM exhibited more noise than those reconstructed by the other methods. However, no significant differences existed between modified-BSREM and RAREM in the Jaszczak phantom and between all reconstruction methods in the striatum phantom. The visual results of the real phantom and simulation experiments were the same.

Fig. 8.

Fig. 8

The reconstructed images of the real experiment. The left side is the Jaszczak phantom, and the right side is the striatum phantom. A, B, and C are the TV-EM, modified-BSREM, and RAREM, respectively. “proj” indicates the number of projections

The CRC of the Jaszczak phantom and SBR of the striatum phantom are shown in Figs. 9 and 10, respectively. The CRC was approximately the same for all acquisition conditions and reconstruction methods; however, the standard deviation (STD) in each ROI using TV-EM was slightly higher than that of the other reconstruction methods. The SBR was the same for all the acquisition conditions. In the low-contrast regions (left striatum), the SBRs of all reconstruction methods were approximately 50% of their true values.

Fig. 9.

Fig. 9

CRCs of the Jaszczack phantom. “proj” and “STD” indicate the number of projections and standard deviation in each ROI, respectively.

Fig. 10.

Fig. 10

SBRs of the striatum phantom. “proj”, “R”, and “L” indicate the number of projections, the SBR of the right striatum, and the SBR of the left striatum, respectively. The dotted lines indicate the true SBR values

Discussion

RAREM, which we proposed in this study, controlled the regularization parameter η and the relaxation parameter λ based on the acquisition condition and the complexity of the update image. We evaluated the RAREM algorithm using simulations and real phantom data.

Almost all results using the RAREM were similar to those in the modified-BSREM, where the hyperparameters were optimized. This is attributable to the identical optimization problem (5), and similar solution methods, i.e., gradient method. The statistical analysis results of the SSIM alone for the striatum phantom, where significant differences were found, are shown in Tables 2 and 3. There were no significant differences between conventional methods and RAREM in the other phantoms and evaluations. Under all the acquisition conditions and phantoms, the NRMSE, CRC, and SBR of RAREM were equivalent to those of the conventional methods, and the SSIM of RAREM was better than those of the conventional methods. The regularization parameter was required to set larger value when the acquired data were insufficient, e.g., the number of counts and projections was small [13, 15]. The hyperparameters of RAREM were optimized to achieve good results under all conditions. These results indicate that RAREM can obtain a quantitative performance equivalent to conventional methods and that RAREM can obtain better visual quality images even if the user does not set hyperparameters.

Table 2.

Statistical analysis between TV-EM and RAREM for SSIM of the striatum phantom

kcounts/projection
projection 2.5 5.0 7.5 10.0 25.0 50.0
12
14
18
30
60
120

” and “” indicate p<0.05 and p<0.01, respectively

Table 3.

Statistical analysis between modified-BSREM and RAREM for SSIM of the striatum phantom

kcounts/projection
Projection 2.5 5.0 7.5 10.0 25.0 50.0
12
14
18
30
60
120

” and “” indicate p<0.05 and p<0.01, respectively

The optimal regularization parameter decreases when the object structure is small or complex because a lower regularization parameter can achieve higher spatial resolution [12, 29]. In addition, the relaxation parameter can control the contribution of noise to update images; therefore, a lower relaxation parameter can suppress noise [1]. However, there are trade-offs between noise suppression and spatial resolution [13]. The values of Ek are shown in Fig. 11. In the simple construction slices ((a) and (b)-1 in Fig. 11), as can be observed in Equations (18) and (24), the Ek had a low value; therefore, η and λ had high and low values, respectively. In contrast, in the complicated construction slices ((b)-2, (c)-1, and (c)-2 in Fig. 11), the Ek value was high; therefore, η and λ were low and high, respectively. In the iterative reconstruction method, low-frequency components (mainly contributing to contrast) are recovered with few updates, whereas high-frequency components (mainly contributing to edges and noise) are recovered with many updates [30, 31]. Moreover, decreasing the regularization parameter iteratively is more effective for converging to a global solution than using fixed parameter [32]. Therefore, in RAREM, the increase of Ek and decrease of η iteratively contributed to preserving edges in complicated slices and obtaining good image quality. The results indicate that RAREM can automatically optimize the hyperparameters when evaluating the complexity of the object structure. The results of the optimal hyperparameter search using the conventional methods and the results of RAREM are shown in Fig. 12. The NRMSE and SSIM in TV-EM and modified-BSREM had variable values depending on the hyperparameter, and as shown in Fig. 7, using a non-optimal parameters resulted in poor image quality. Optimizing the parameters for regularized reconstruction methods is challenging and has become a field of active research in PET [1216]. These studies required a lot of time and computation. However, the results of each study are only applicable to its specific target. In nuclear medicine, the invention of a new radio isotope (RI) has led to the creation of new clinical examinations, and the optimal parameters of modified-BSREM have been determined each time with extensive computation and effort because of many hyperparameters. In SPECT, because of the large number of examination types and acquisition conditions, it is expected that searching for the optimal hyperparameters in all examinations is too difficult. However, RAREM can produce good-quality images that are equivalent to the best image obtained from the conventional reconstruction method, with only one calculation. Because RAREM has a similar number of parameters as modified-BSREM, it provides the same flexibility as BSREM while providing images of the same high quality as modified-BSREM without requiring users to experimentally set hyperparameters. RAREM can be applied to any target. Therefore, RAREM is easily applicable in clinical applications.

Fig. 11.

Fig. 11

The behavior of Ek reconstructed from 120 projections and 104 counts/projection in RAREM by main iteration

Fig. 12.

Fig. 12

The results of the optimal hyperparameter search under the condition of 60 projections and 5.0×103counts/projection using the TV-EM and modified-BSREM. a, b, and c are the 3D-MAC phantom, Striatum phantom, and BrainWeb phantom, respectively. The range of η in TV-EM was from 0.1 to 1.0. The range of η in modified-BSREM was from 0.1 to 1.0 in the 3D-MAC phantom and from 0.01 to 0.1 in the other phantoms, and that of λ0 was from 0.2 to 1.0

Conclusion

In this study, we proposed an automatically controlled regularized reconstruction method that incorporates acquisition conditions and object structure into the hyperparameters. We evaluated the image quality in the simulation and real phantom experiments, demonstrating that the proposed method performs equivalently or better than conventional regularized methods, with optimized hyperparameters. The proposed method does not require the user to set hyperparameters experimentally and can avoid the investigation of optimal hyperparameters; it is an alternative to conventional regularized methods in clinical.

Acknowledgements

We thank the members of the Department of Radiology at Toho University Ohashi Medical Center for their cooperation.

Abbreviations

SPECT

Single photon emission computed tomography

CT

Computed tomography

MRI

Magnetic resonance imaging

PET

Positron emission tomography

DRAMA

Dynamic row-action type maximization likelihood algorithm

ML-EM

Maximum likelihood-expectation maximization

OS-EM

Ordered subset-expectation maximization

MAP-EM

Maximum a posteriori-expectation maximization

TV-EM

Total variation-expectation maximization

BSREM

Block sequential regularized expectation maximization

RAREM

Row-action type automatic regularized expectation maximization

FWHM

Full width at half maximum

NRMSE

Normalized root mean square error

SSIM

Structural similarity index measure

SIMIND

Simulation of an imaging nuclear detector

CRC

Contrast recovery coefficient

SBR

Specific binding ratio

ROI

Region of interest

VOI

Volume of interest

Appendix A

In this Appendix, we prove how Vmax of the partial derivative of the TV norm (8) is derived. First, ε is ignored to simplify the equation, because it reduces Vmax. Next, the first term in Equation (8) is considered. Since xs,t-xs-1,txs,t-xs-1,t2=1 and xs-1,t+1-xs-1,t20, the maximum value of the first term in Equation (8) is 1 when xs-1,t+1=xs-1,t. The maximum value of the second term is 1, considering it in the same manner as in the first term. To consider the maximum value of the third term, it is transformed.

xs+1,t+xs,t+1-2xs,txs+1,t-xs,t2+xs,t+1-xs,t2=xs+1,t-xs,t+xs,t+1-xs,txs+1,t-xs,t2+xs,t+1-xs,t2 34

Because the sign of the third term in Equation (8) is negative, it is necessary to consider the minimum value in Equation (34) to consider the maximum value in Equation (8). When xs+1,t and xs,t+1 are 0, and xs,t>0, the third term indicates the minimum value because each component of x is a non-negative value. Therefore, the minimum value of third term is -2. From the maximum and minimum values of the three terms shown above, the maximum value Vmax of equation (8) is 2+2.

Author Contributions

TM developed the algorithm and performed most of the experimental work, data analysis, and writing. KH and TH revised the algorithm, provided suggestions for the experimental procedures and data analysis, and critically revised the manuscript for intellectual content. All the authors have read and approved the final version of the manuscript.

Funding

None.

Availability of data and materials

The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.

Declarations

Ethics approval and consent to participate

This article does not contain any studies with human participants performed by any of the authors.

Consent for publication

Not applicable.

Conflict of interest

The authors have no relevant Conflict of interest to disclose.

Footnotes

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Browne J, De Pierro AR. A row-action alternative to the EM algorithm for maximizing likelihoods in emission tomography. IEEE Trans Med Imaging. 1996;15(5):687–99. 10.1109/42.538946. [DOI] [PubMed] [Google Scholar]
  • 2.Tanaka E, Kudo H. Subset-dependent relaxation in block-iterative algorithms for image reconstruction in emission tomography. Phys Med Biol. 2003;48(10):1405–22. 10.1088/0031-9155/48/10/312. [DOI] [PubMed] [Google Scholar]
  • 3.Green PJ. Bayesian reconstructions from emission tomography data using a modified EM algorithm. IEEE Trans Med Imaging. 1990;9(1):84–93. 10.1109/42.52985. [DOI] [PubMed] [Google Scholar]
  • 4.Panin VY, Zeng GL. Total variation regulated EM algorithm. IEEE Trans Nucl Sci. 1999;46(6):2202–10. 10.1109/23.819305. [Google Scholar]
  • 5.Fukami M, Matsutomo N, Hashimoto T, Yamamoto T, Sasaki M. Compressed sensing reconstruction shortens the acquisition time for myocardial perfusion imaging: a simulation study. Radiol Phys Technol. 2023;16:397–405. 10.1007/s12194-023-00730-5. [DOI] [PubMed] [Google Scholar]
  • 6.Matsutomo N, Fukaya K, Hashimoto T, Yamamoto T, Sato E. Performance of compressed sensing-based iterative reconstruction for single-photon emission computed tomography from undersampled projection data: a simulation study in 123I-N-ω-fluoropropyl-2β-carbomethoxy-3β-(4-iodopheny) nortropane imaging. Nucl Med Commun. 2019;40(2):106–14. 10.1097/MNM.0000000000000938. [DOI] [PubMed] [Google Scholar]
  • 7.Zhu Z, Wahid K, Babyn P, Cooper D, Pratt I, Carter Y. Improved compressed sensing-based algorithm for sparse-view CT image reconstruction. Compt Math Methods Med. 2013;2013: 185750. 10.1155/2013/185750. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Wolf PA, Jørgensen JS, Schmidt TG, Sidky EY. Few-view single photon emission computed tomography (SPECT) reconstruction based on a blurred piecewise constant object model. Phys Med Biol. 2013;58:5629–52. 10.1088/0031-9155/58/16/5629. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.De Pierro AR, Yamagishi MEB. Fast EM-like methods for maximum “a posteriori’’ estimates in emission tomography. IEEE Trans Med Imaging. 2001;20(4):280–8. 10.1109/42.921477. [DOI] [PubMed] [Google Scholar]
  • 10.Ahn S, Fessler J. Globally convergent image reconstruction for emission tomography using relaxed ordered subsets algorithms. IEEE Trans Med Imaging. 2003;22(5):613–26. 10.1109/TMI.2003.812251. [DOI] [PubMed] [Google Scholar]
  • 11.Ross SQ. Q.clear. GE Healthcare, White Paper. 2014.
  • 12.Howard BA, Morgan R, Thorpe MP, Turkington TM, Oldan J, James O, et al. Comparison of Bayesian penalized likelihood reconstruction versus OS-EM for characterization of small pulmonary nodules in oncologic PET/CT. Ann Nucl Med. 2017;31(8):623–8. 10.1007/s12149-017-1192-1. [DOI] [PubMed] [Google Scholar]
  • 13.Trägårdh E, Minarik D, Almquist H, Bitzen U, Garpered S, Hvittfelt E, et al. Impact of acquisition time and penalizing factor in a block-sequential regularized expectation maximization reconstruction algorithm on a Si-photomultiplier-based PET-CT system for Inline graphicF-FDG. EJNMMI Res. 2019;9:64. 10.1186/s13550-019-0535-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Teoh EJ, McGowan DR, Macpherson RE, Bradley KM, Gleeson FV. Phantom and clinical evaluation of the bayesian penalized likelihood reconstruction algorithm Q. Clear on an LYSO PET/CT system. J Nucl Med. 2015;56(9):1447–52. 10.2967/jnumed.115.159301. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Te Riet J, Rijnsdorp S, Roef MJ, Arends AJ. Evaluation of a Bayesian penalized likelihood reconstruction algorithm for low-count clinical Inline graphicF-FDG PET/CT. EJNMMI Phys. 2019;6:32. 10.1186/s40658-019-0262-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.CeriÄ AI, Minarik D, Persson E, MosÃn H, Valind K, TrÃgÃrdh E, et al. First clinical experience of a ring-configured cadmium zinc telluride camera: a comparative study versus conventional gamma camera systems. Clin Physiol Funct Imaging. 2023;44:79–88. 10.1111/cpf.12853. [DOI] [PubMed] [Google Scholar]
  • 17.Shepp L, Vardi Y. Maximum likelihood reconstruction for emission tomography. IEEE Trans Med Imaging. 1982;1(2):113–22. 10.1109/TMI.1982.4307558. [DOI] [PubMed] [Google Scholar]
  • 18.Tanaka E, Kudo H. Acceleration of statistical image reconstruction for transmission CT using DRAMA (dynamic RAMLA) strategy. Med Imaging Technol. 2006;24(2):114–24. 10.11409/mit.24.114. [Google Scholar]
  • 19.Kitamura K, Takahashi S.: Transmission image reconstruction using DRAMA in PET. The Japanese Society of Medical Imaging Technology (JAMIT) Annual Meeting. Proceedings. Japanese. 2004;2004:1–10. [Google Scholar]
  • 20.Onishi H, Motomura N, Takahashi M, Yanagisawa M, Ogawa K. A 3-dimensional mathematic cylinder phantom for the evaluation of the fundamental performance of SPECT. J Nucl Med Technol. 2010;38(1):42–8. 10.2967/jnmt.108.061192. [DOI] [PubMed] [Google Scholar]
  • 21.Darcourt J, Booij J, Tatsch K, Varrone A, Vander Borght T, Kapucu Ã-L, et al. EANM procedure guidelines for brain neurotransmission SPECT using 123I-labelled dopamine transporter ligands, version 2. Eur J Nucl Med Mol Imaging. 2010;37(2):443–50. 10.1007/s00259-009-1267-x. [DOI] [PubMed] [Google Scholar]
  • 22.Djang DSW, Janssen JR, Bohnen N, Booij J, Henderson TA, Herholz K, et al. SNM practice guideline for dopamine transporter imaging with Inline graphicI-Ioflupane SPECT 1.0. J Nucl Med. 2012;53(1):154–63. 10.2967/jnumed.111.100784. [DOI] [PubMed] [Google Scholar]
  • 23.Kapucu ÖL, Nobili F, Varrone A, Booij J, Borght V, Någren K, et al. EANM procedure guideline for brain perfusion SPECT using Inline graphicTc-labelled radiopharmaceuticals, version 2. Eur J Nucl Med Mol Imaging. 2009;36(12):2093–102. 10.1007/s00259-009-1266-y. [DOI] [PubMed] [Google Scholar]
  • 24.Juni JE, Waxman AD, Devous MD, Tikofsky RS, Ichise M, Van Herertum RL, et al. Procedure guideline for brain perfusion SPECT using Inline graphicTc radiopharmaceuticals 3.0. J Nucl Med Technol. 2009;37(3):191–5. 10.2967/jnmt.109.067850. [DOI] [PubMed] [Google Scholar]
  • 25.Ljungberg M, Strand SE. A Monte Carlo program for the simulation of scintillation camera characteristics. Comput Meth Programs Biomed. 1989;29(4):257–72. 10.1007/s12194-021-00631-5. [DOI] [PubMed] [Google Scholar]
  • 26.Ito T, Tsuchikame H, Ichikawa H, Onoguchi M, Okuda K, Shibutani T, et al. Verification of phantom accuracy using a Monte Carlo simulation: bone scintigraphy chest phantom. Radiol Phys Technol. 2021;14:336–44. 10.1007/s12194-021-00631-5. [DOI] [PubMed] [Google Scholar]
  • 27.Hashimoto F, Ohba H, Ote K, Teramoto A, Tsukada H. Dynamic PET image denoising using deep convolutional neural networks without prior training datasets. IEEE Access. 2019;7:96594–603. 10.1109/ACCESS.2019.2929230. [Google Scholar]
  • 28.Tossici-Bolt L, Hoffmann S, Kemp P, Mehta R, Fleming J. Quantification of FP-CIT SPECT brain images: an accurate technique for measurement of the specific binding ratio. Eur J Nucl Med Mol Imaging. 2006;33(12):1491–9. 10.1007/s00259-006-0155-x. [DOI] [PubMed] [Google Scholar]
  • 29.Miwa K, Yoshii T, Wagatsuma K, Nezu S, Kamitaka Y, Yamao T, et al. Impact of Inline graphic factor in the penalty function of Bayesian penalized likelihood reconstruction (Q. Clear) to achieve high-resolution PET images. EJNMMI Phys. 2023;10:4. 10.1186/s40658-023-00527-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Jaskowiak CJ, Bianco JA, Perlman SB, Fine JP. Influence of reconstruction iterations on Inline graphicF-FDG PET/CT standardized uptake values. J Nucl Med. 2005;46(3):424–8. [PubMed] [Google Scholar]
  • 31.Kupitz D, Wissel H, Wuestemann S, Bluemel S, Pech M, Amthauer H, et al. Optimization of SPECT/CT imaging protocols for quantitative and qualitative Inline graphicTc SPECT. EJNMMI Phys. 2021. 10.1186/s40658-021-00405-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Mondal PP, Rajan K, Patnaik LM. Adaptively altering hyper-parameter for improved reconstruction in PET. IEEE Nucl Sci Sympos Conf Rec. 2003;2003:3460–3. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.