Cavity-enhanced continuous-wave microscopy with potentially unstable cavity length

Oliver Lueghamer 1, Stefan Nimmrichter 2, Clara Conrad-Billroth 3,4, Thomas Juffmann 3,4, Maximilian Prüfer 1,
PMCID: PMC12307652  PMID: 40731072

Abstract

Microscopy gives access to spatially resolved dynamics in different systems, from biological cells to cold atoms. A big challenge is maximizing the information per used probe particle to limit the damage to the probed system. We present a cavity-enhanced continuous-wave microscopy approach that provides enhanced signal-to-noise ratios at fixed damage compared to standard single-pass microscopy. Employing a self-imaging 4f cavity, we show contrast enhancement for controlled test samples as well as biological samples. For thick samples, the imaging cavity leads to a new form of dark-field microscopy, where the separation of scattered and unscattered light is based on optical path length. We theoretically show that enhanced signal, signal-to-noise, and signal-to-noise per damage are also retrieved when the cavity is not length-stabilized. Our results provide an approach to cavity-enhanced microscopy with non-length-stabilized cavities and might be used to enhance the performance of dispersive imaging of ultracold atoms.

Supplementary Information

The online version contains supplementary material available at 10.1038/s41598-025-13589-w.

Subject terms: Microscopy, Ultracold gases


Advances in microscopy have led to new discoveries across scientific disciplines1, from cellular biology2 to quantum physics3 and beyond. Microscopy setups measure certain parameters of interest in a spatially resolved way. Not only the spatial resolution limits the measurement precision but often the limiting factor is the finite number of probe particles detected in a given image. In many applications, the number of probe particles cannot be increased arbitrarily, either due to sample, source, or detector restrictions. Maximizing the information obtained from each detected probe particle is crucial in such applications.

For coherent imaging techniques, it has recently been shown that multi-passing each probe particle through the sample can increase the information per detected probe particle. This has been experimentally demonstrated in optical imaging and diffraction studies4, where self-imaging cavities5 were used to multi-pass a pulsed probe through a sample. Follow-up experiments showed the build-up of orbital angular momentum in a multi-passing experiment7. Notably, multi-passing probe particles m times through a sample can enable a measurement precision per probe-sample interaction similar to a quantum-enhanced measurement with m suitably entangled probe particles. It scales Inline graphic instead of Inline graphic812.

In physics applications, like spectroscopy or the imaging of ultracold atoms, a narrow spectral linewidth is often required. Thus, cavity-enhanced microscopy has to be done with continuous-wave (CW) excitation, which actually has theoretically been shown to outperform the pulsed multi-pass scheme12. However, despite the widespread use of single-mode cavity-enhanced measurements in various scientific and technological fields, applications in imaging remain rare, due to challenges in operating a cavity that is fully degenerate in all transverse modes13. The first progress was recently demonstrated in a 4-pass geometry that showed contrast enhancement in flow-cytommetry14. So far, genuine continuous-wave imaging cavities have been employed in the realization of multi-mode lasers15, multi-mode coherent absorbers16, cavity-enhanced non-linear optics applications17, and imaging of ultracold atoms18.

In this paper, we demonstrate cavity-enhanced continuous-wave microscopy using a self-imaging 4f cavity. We demonstrate the concept experimentally by applying it to the imaging of an artificially fabricated sample, consisting of holes in a Inline graphic thin silicon nitride (Inline graphic) membrane. Imaging human cheek cells, we further demonstrate that cavity-based imaging leads to a novel form of dark-field microscopy, where the separation of scattered and unscattered light is based on optical path length and not on scattering angle. Our experiments are backed by theory which shows that an enhanced contrast can even be observed when scanning the cavity across a free spectral range of the cavity. These results open a path to cavity-enhanced measurements in applications where cavity length stabilization is impossible, i. e. in case the cavity is non-length-stabilized. Our continuous-wave approach will facilitate the usage of cavities for microscopy with the need for small spectral width, for example, the imaging of ultracold atoms.

Degenerate cavity setup For cavity-enhanced microscopy, it is crucial that spatial information is accurately transmitted through the cavity system. Degeneracy can be achieved with multiple cavity arrangements5,13. For example, confocal or concentric cavities feature a certain degree of degeneracy of transversal modes, which leads to re-imaging of the fields after a certain number of roundtrips. For our experiments, we chose a cavity geometry that realizes an 8f transformation after a single roundtrip. It guarantees that the output field is equal to the input field, allowing for a straightforward interpretation of the resulting images.

We employ an effective 4f configuration using two highly reflective planar cavity mirrors and a pair of Inline graphic biconvex lenses; a schematic of the setup is shown in Fig.1a (see Supplementary Material for details and6 for all details on the assembly, and testing of the cavity). The sample is positioned off-axis to prevent overlapping with the mirrored image; as a result, the image is projected onto the sample only after completing one full round trip. Control over the cavity length is achieved via a piezo ring mounted on the first mirror. For monitoring the cavity output, a photodiode is placed after a beamsplitter, and the remaining light is imaged on a charge-coupled device (CCD) camera. Various standard imaging modalities, such as bright-field, darkfield, or Zernicke phase imaging, can be implemented outside the cavity (see Supplementary Material for imaging details).

Fig. 1.

Fig. 1

Cavity setup and characterization. (a) Ray tracing picture showing the self-imaging properties of the 4f cavity. The sample is illuminated with a collimated beam (red); the diffracted light (green) is imaged onto the sample after one full roundtrip. (b) Cavity length scan monitored by the output voltage (normalized) of a photodiode placed at the output of the cavity. The separation between the two resonance peaks yields a free spectral range (FSR) of Inline graphic and a FWHM of Inline graphic in that specific realisation (accordingly the finesse is 13.9; see Methods for Details). (c) We test the degeneracy of transversal modes by projecting the image of a US Air Force (USAF) Target onto the cavity image plane. The recorded light intensity in a conjugate plane after the cavity, and a cross-section thereof, demonstrate its image transmission capabilities.

In Fig. 1b, we present a cavity scan where we show the output intensity measured using a photodiode as a function of the cavity length. The spacing between the two peaks represents the free spectral range (FSR), achieved by moving the mirror over half a wavelength Inline graphic. From the distance between the two peaks, we obtain an FSR of Inline graphic and a FWHM of Inline graphic (accordingly the finesse is 13.9), which aligns well with the theoretical predictions (see Supplementary Material for details).

To demonstrate image transmission capabilities, a US Air Force (USAF) target is positioned in the object plane of the focusing lens in front of the cavity. The resulting image is coupled into the self-imaging cavity and then captured by the CCD camera. As shown in Fig. 1c, the image clearly reveals the structure of the USAF target, confirming that many transverse modes are transmitted through the cavity.

Theoretical results for cavity-enhancement The theoretical performance of our approach is analyzed in a calculation that keeps track of phase shifts along each closed path within the cavity (see Supplementary Material). Our key performance indicators are the sample contrast and the signal-to-noise ratio (SNR) at fixed input power,

graphic file with name d33e445.gif 1

where Inline graphic is the output intensity measured at the sample position, and Inline graphic is the average output intensity measured next to the sample.

While in phase-contrast microscopy the contrast would be Inline graphic19, we find for our cavity-enhanced approach:

graphic file with name d33e467.gif 2

where Inline graphic and Inline graphic denote the reflectivity and transmission of the cavity mirrors Inline graphic, Inline graphic and Inline graphic are the weak phase shifts of the sample and a reference point, respectively, and Inline graphic. Note that the prefactor is Inline graphic if high-reflectivity mirrors are used (Inline graphic).

Strikingly, when averaging over the resonance in the course of the measurement, the contrast is still enhanced compared to a single pass:

graphic file with name d33e507.gif 3

Note that in the shot-noise limited case the signal-to-noise ratio at constant damage SNRInline graphic will also be enhanced12.

This enhanced performance of an non-length-stabilized cavity relies on all modes within the cavity undergoing the same fluctuations. It can also be derived from a textbook-like toy model in which a Fabry-Pérot cavity is used for gas cell spectroscopy. If the input mirror vibrates across an FSR, cavity enhancement is still obtained, if the cavity output is interfered with a reference from an empty cavity that undergoes the same fluctuations in cavity length. We provide this calculation in the Method section, where we also show that the 2x loss in contrast for non-length-stabilized cavities is due to light being reflected from the cavity. Analyzing the reflected light, will regain the lost information.

Demonstration of cavity-enhanced microscopy The enhancement of multi-pass techniques is closely linked to the effective number of round trips N4,12,14. For continuous-wave cavities, we can calculate the effective N by comparing the intensities inside and outside the cavity for a given cavity length,

graphic file with name d33e566.gif 4

Measuring the total input and output intensities Inline graphic thus allows us to evaluate N for a given cavity setting in the experiment.

In order to demonstrate the contrast-enhancing capabilities of the setup, we cut holes with diameters of Inline graphic in a Inline graphic thin silicon nitride (Inline graphic) membrane, which forms a Inline graphic window on a silicon frame; the membrane has a reflectance of approximately Inline graphic. Fig.2a shows a scanning electron microscopy image of the structure. In Fig.2b we show a single-pass defocused image20, where the hole is visible as a bright spot. The inset shows a focused single-pass microscopy image in which the holes are not visible. This is consistent with the contrast levels expected for these techniques. While bright field-contrast would be based on reflection from the membrane and amounts to only Inline graphic, defocus leads to phase contrast which can be as large as Inline graphic, where d is the thickness and Inline graphic denotes the index of refraction of Inline graphic; see Supplementary Material.

Fig. 2.

Fig. 2

Cavity-enhanced microscopy of holes in a Inline graphic membrane. (a) Scanning electron microscopy image of holes cut into a Inline graphic membrane. The red and blue circles indicate holes with a diameter of Inline graphic, and the yellow circle marks a hole with a diameter of Inline graphic. The inset shows a close-up of the hole we use as our main target. (b) Defocused single-pass image of the membrane; the inset shows the region of the image with the target in focus; the target hole is not visible (see main text). (c) Cavity-enhanced image of the membrane; the two circles indicate the targeted hole (solid) and its mirror image (dashed line). The inset shows a close-up of the target; compared to the single pass case we find a contrast enhancement; the two circles show the regions for averaging signal (red) and background (blue without red) intensities, used to estimate contrast and SNR. (d) Signal-to-noise ratio in the cavity-enhanced case plotted against the number of effective round trips. The color-coded data points match those in the inset, where the transmitted intensity is shown as a function of the cavity length.

We find a noticeable contrast enhancement in the cavity-enhanced image on resonance (Fig. 2c). To quantify the contrast, we averaged the intensity of the cavity-enhanced hole and compared it to the nearest averaged radial background. The calculated contrast was approximately Inline graphic, considerably more than the single-pass bright-field contrast. We then evaluated the SNR for each captured image, plotting it against the effective number of roundtrips. Remarkably, we observed an enhanced SNR at the cavity resonance, as depicted in Fig.2d; the SNR overall follows the trend of the transmitted intensity (shown in inset of Fig. 2d). Further characterization of the image reveals a measured hole size of Inline graphic and a resolution of Inline graphic, which aligns well with the expected values (see Supplementary Material), indicating that the cavity does not reduce the spatial resolution of our microscope.

Microscopy of epithelial cells Next, we image epithelial cells obtained from the inner lining of a human cheek, which are fixed onto an anti-reflection coated glass slide using methanol. Fig. 3a displays the morphology of these cells in a single-pass image, with the two black circles in the center likely representing the cell nuclei.

Fig. 3.

Fig. 3

Cavity dark-field imaging. (a) Bright Field image of cheek cell in single pass. (b) Calculated SNR as a function of mirror displacement represented by the driving voltage of the piezo. The blue curve shows the cavity output detected by the photodiode. (c) Yellow: Integrated images over time (see Supplementary Material for all pictures), while the incidence light is resonant and the refracted light is out of resonance. (d) Dark Blue: Integrated images over time while incidence light is out of resonance and refracted light on resonance. This represents a novel approach to dark-field imaging. (e) Integrating over all the detunings, we obtain a picture corresponding to an non-length-stabilized cavity with increased contrast as compared to the single-pass image.

Performing cavity-enhanced microscopy, we first study the SNR as a function of cavity length, defining the whole cell as the signal region and the region around it as the background. We obtain a peak in SNR when the incident, unscattered background light is resonant (indicated by the yellow bar in Fig. 3). Notably, the transparent sample appears completely darkened (see Fig. 3c). Detuning the cavity, we observe a second “resonance” in the SNR when the cavity is off-resonant for the incident light. In Fig.3d, we show the resulting image, which clearly displays the cell content and its structures spatially resolved (see Supplementary Material for all images). The dispersive phase shift that the light undergoes while passing through the sample, along with the laser’s narrow bandwidth, is substantial enough for the refracted light to resonate at a different cavity length. This phenomenon introduces a new method for dark-field imaging; the peak shift can be used as a measure of the sample’s optical thickness.

Microscopy with non-length-stabilized cavities Finally, we show that signal amplification, and enhanced contrast, are also observed in an non-length-stabilized cavity. To demonstrate this effect experimentally, we integrate across the entire measurement, and obtain a combination of bright- and darkfield image, as shown in Fig.3e with enhanced contrast compared to single-pass. This effect is also achieved by integrating with the exposure time over one FSR scan. This makes our cavity-enhanced microscopy method applicable in scenarios without the possibility of stabilizing the cavity length actively.

Conclusion and outlook.— In this work, we demonstrate continuous-wave cavity-enhanced microscopy with a narrow laser bandwidth. We experimentally demonstrate increased contrast and signal-to-noise, and show both theoretically and experimentally that these advantages persist in non-length-stabilized cavities.

Furthermore, we present a new form of dark-field microscopy in which different regions of a sample, corresponding to different optical path lengths, light up at different cavity lengths. This corresponds to an entirely new contrast mechanism in optical microscopy. Contrary to traditional dark-field imaging, our scheme enables dark-field imaging with forward scattered light, which is ideal for quantitatively assessing samples characterized by slowly varying phase shifts.

A potential application of our method is cavity-enhanced imaging of ultracold atoms, which so far has only been done with single mode cavities21. Dispersive imaging, utilizing the phase shifts22, has been demonstrated to allow for multiple repeated images23. We propose using cavity-enhanced imaging to improve the signal-to-noise ratio at fixed feedback to the quantum system to lower the impact of e.g. heating24. The forward scattering leading to the phase signal is coherently amplified by the cavity, in contrast to incoherent scattering processes. This should enable high SNR dispersive imaging of ultracold atoms. At the same time, the imaging cavity provides new ways of applying local control fields within the cavity25.

Methods

Cavity parameters

The cavity length is defined by the 4f requirement to be Inline graphic. Accounting for the optical path length differences introduced by the lenses results in an actual length of approximately Inline graphic. This corresponds to an FSR of Inline graphic, which is in good accordance with the experimentally found value of Inline graphic. We use two highly reflective mirrors with Inline graphic and Inline graphic, yielding a theoretical finesse value of 29.8.

We scan the cavity across the resonance by changing the cavity length with a piezo. We calibrate the applied voltage with a frequency reference. We do so by injecting two laser frequencies. The frequency difference is small compared to the FSR but larger than the FWHM of the cavity resonance and thus can be used as a ruler.

In the actual experimental environment, we observed fluctuating finesse values between 10 and 25. The presence of optical elements within the cavity and the sample carriers likely contributed to this reduction in overall finesse. We observed cavity length drifts of about Inline graphic on the one second scale.

Data analysis and error bars

The shown data points for the SNR are acquired from a single realization. Measuring at exactly the same cavity length would require length-stabilisation; we checked that the features shown in the main text are reproducible (see Supplementary Material figure for repeated scans).

Enhancement with non-length-stabilized cavity

In the main text, we demonstrate cavity-enhanced microscopy of an optically thin lossless sample and show evidence that an enhancement of the phase contrast prevails even in the case of an non-length-stabilized cavity resonance. Here, we corroborate our finding with the help of an instructive toy model based on a Fabry-Pérot cavity - a setup known from textbooks and many applications. This simple example illustrates both the contrast enhancement and its robustness against cavity fluctuations without lengthy formulas.

Consider a Fabry-Pérot cavity of length L with end mirrors of reflectivity Inline graphic that contains a homogeneous, lossless, and optically thin medium. The task is to infer its weak refractive index n from a measurement of the small phase shift, Inline graphic, it imparts on a single-mode light field passing through the cavity. A balanced beam splitter then superimposes a phase-locked reference beam on this field, after which the intensities in the two output ports are detected; see Fig. 4. We will take their difference as the measurement signal.

Fig. 4.

Fig. 4

Set-up for toy model calculation. (a) The output of a Fabry-Pérot cavity setup is homodyned with a local oscillator. (b) Additionally, the local oscillator is transmitted through the same non-length-stabilized cavity picking up the phase shift of the detuned cavity. For both cases, we analyze the possible precision in estimating the phase shift imprinted by a thin lossless sample inside the cavity.

Let us assume for now that the cavity length is stable. Given a monochromatic input field amplitude Inline graphic of wavenumber Inline graphic, the transmitted field reads as26

graphic file with name d33e790.gif 5

Here and in the following, we consistently expand to lowest order in the phase shift Inline graphic we seek to estimate, and we denote the field amplitude inside the empty cavity as

graphic file with name d33e803.gif 6

Notice that all field amplitudes are defined with respect to the position of the right cavity mirror on the optical axis, Inline graphic.

We achieve perfect, resonant transmission if we adjust the cavity length or the light wavelength such that the optical path length 2L over one roundtrip is a multiple of the wavelength, Inline graphic with Inline graphic. This leads to constructive interference enhancing both the intra-cavity field strength, Inline graphic, and the response to the phase shift in (5),

graphic file with name d33e834.gif 7

For phase-sensitive detection, we combine the output with a reference beam of, say, the same input field strength phase-shifted by Inline graphic, Inline graphic, as depicted in Fig. 4(a). The intensities detected after the beam splitter are given by Inline graphic, up to constant prefactors, and the measurement signal by the difference of these two intensities,

graphic file with name d33e855.gif 8

The resonant case yields a gain in phase signal due to the enhancement of the intra-cavity intensity with 1/T,

graphic file with name d33e864.gif 9

Not only the signal is enhanced on resonance, but also the signal-to-noise ratio (SNR): assuming shot-noise-limited detection, the standard deviation Inline graphic of the measurement signal is proportional to the square-root of the sum of the two detected intensities. In the resonant case, Inline graphic, and hence the SNR improves with 1/T.

For sensitive specimens, it is important to characterize the SNR at a fixed intracavity intensity, i.e., a fixed amount of probe-induced specimen damage. To do so, we compare the cavity-enhanced case with a single pass scheme with an input power increased by Inline graphic. We retrieve a signal-to-noise ratio at constant damage (SNRInline graphic) that improves by Inline graphic12, indicating another fundamental advantage of cavity-enhanced measurements over single-pass measurements.

What happens if the cavity length L is unstable for the duration of the experiment? We can describe this by taking uniform averages with respect to L over one spectral range Inline graphic, neglecting the small variation of the phase shift Inline graphic over this range. Analytic expressions for the averaged intracavity intensity and the measurement signal are readily obtained with help of Mathematica,

graphic file with name d33e929.gif 10
graphic file with name d33e933.gif 11

The average intracavity intensity is no longer enhanced, but lower than the input intensity, reducing the signal proportionally. However, even if we increase the input power to match the intra-cavity intensity of the resonant signal (9), the sensitivity to the phase shift is still diminished, lacking the resonant gain by 1/T. In agreement with our intuition, we do not observe an enhanced signal, SNR, and SNRInline graphic with a cavity of unstable length.

However, the situation changes if we let the reference beam pass an empty region of the same cavity and thus be subject to the same unstable length as the probe field; see Fig. 4(b). This is described by the amplitude Inline graphic, which leads to the measurement signal

graphic file with name d33e961.gif 12

On resonance (Inline graphic), the result is the same as before in (9). The length-averaged signal for an non-length-stabilized cavity now reads as

graphic file with name d33e974.gif 13

Once again, the signal suffers from the reduced intra-cavity intensity, but at matching intensity, the gain in phase sensitivity persists. At high cavity finesse, Inline graphic, the sensitivity is reduced merely by a factor two compared to the resonant case (9). Regarding SNR, the detected intensities are diminished by T compared to the input, so that Inline graphic and hence SNR Inline graphic. Compared to a single pass, the intracavity intensity is now reduced by Inline graphic, while the detected intensity is reduced further by T. Hence, at matching damage, SNRInline graphic improves by Inline graphic.

In the Supplementary Material, we analyze an imaging scenario in which many transverse modes are enhanced in a self-imaging cavity. Since they all are transmitted through the same cavity, the assumption of having common noise on the cavity path length is justified for small path length changes on the order of Inline graphic and in the paraxial limit. Spatially distributed phase shifts can be quantified using phase contrast microscopy, which can be analyzed as an interference effect between the scattered and unscattered light. The imaging setup thus plays a role similar to the homodyne detection in our toy model. Also there, we find enhanced signal contrast, which implies, by virtue of the same arguments as made here, that the SNR and SNRInline graphic are enhanced compared to single-pass imaging as well.

Information in reflected field

We have seen that, in the limit of high cavity finesse, the length average leads to a loss in sensitivity by two in case (b) compared to the resonant result. The intuition is that this should come from the signal that is transmitted through the left cavity mirror. The total outgoing field from the left mirror to the left is

graphic file with name d33e1028.gif 14

and one can easily check the conservation of total energy flux, Inline graphic. For weakly refractive media, we again expand to lowest order in Inline graphic and find,

graphic file with name d33e1041.gif 15

Phase-sensitive detection of this field with help of an empty reference cavity of the same varying length is achieved in the same manner as for the outgoing field on the right. Given the phase-shifted reference field Inline graphic of the empty cavity, we have

graphic file with name d33e1051.gif 16

Averaging this over the length variation results in

graphic file with name d33e1057.gif 17

Adding up both average signals, we get Inline graphic. Hence, if we increase the input intensity to match the 1/T-enhanced intracavity intensity of the resonant case (9), then for Inline graphic the total sensitivity to Inline graphic is the same. Indeed, the missing factor of two in the forward signal is recovered by also monitoring the output on the left.

Supplementary Information

Acknowledgements

We thank Helmut Hörner, Julian Léonard, Stefan Rotter, and Jörg Schmiedmayer for discussions. M.P. thanks Jörg Schmiedmayer for scientific guidance and for providing the laboratory infrastructure. The Inline graphic sample preparation was carried out using facilities at the University Service Centre for Transmission Electron Microscopy, Vienna University of Technology, Austria. This work is supported by an ESQ Discovery grant by the Austrian Academy of Sciences (QUIMP). M.P. has received funding from Austrian Science Fund (FWF): ESP 396 (QuOntM; 10.55776/ESP396). T.J. and C.B. acknowledge funding from the European Union’s Horizon 2020 research and innovation programme under grant agreement 758752. This research was funded in whole or in part by the Austrian Science Fund (FWF) [10.55776/ESP396].

Author contributions

M.P. and T.J. conceived the measurement scheme. C.B. took preliminary data. O.L. built the experimental setup and took the experimental data with the help of M.P.. O.L., T.J., and M.P. analyzed the experimental data. S.N. and T.J. developed the theoretical model. All authors contributed to the discussion of the results. O.L., S.N., T.J., and M.P. wrote the manuscript.

Data availability

Source data and all other data that support the plots within this paper and other findings of this study are available from the corresponding author upon reasonable request.

Declarations

Competing interests

The authors declare no competing interests.

Footnotes

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Change history

12/27/2025

The original online version of this Article was revised: The Supplementary Information file of this article was omitted in its original version. The Supplementary Information file now accompanies the original Article.

References

  • 1.Hawkes, P. W. & Spence, J. C. Science of Microscopy Vol. 1 (Springer, 2007). [Google Scholar]
  • 2.Morris, J. D. & Payne, C. K. Microscopy and cell biology: New methods and new questions. Annu. Rev. Phys. Chem.70, 199–218. 10.1146/annurev-physchem-042018-052527 (2019). [DOI] [PubMed] [Google Scholar]
  • 3.Gross, C. & Bakr, W. S. Quantum gas microscopy for single atom and spin detection. Nat. Phys.17, 1316–1323. 10.1038/s41567-021-01370-5 (2021). [Google Scholar]
  • 4.Juffmann, T., Klopfer, B. B., Frankort, T. L., Haslinger, P. & Kasevich, M. A. Multi-pass microscopy. Nat. Commun.7, 12858. 10.1038/ncomms12858 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Arnaud, J. A. Degenerate optical cavities. Appl. Opt.8, 189. 10.1364/AO.8.000189 (1969). [DOI] [PubMed] [Google Scholar]
  • 6.Lueghamer, O. Cavity enhanced microscopy for dispersive imaging of ultracold atoms. Universität Wien Master’s thesis (2024). 10.25365/thesis.76086
  • 7.Klopfer, B. B. B., Juffmann, T. & Kasevich, M. M. A. Iterative creation and sensing of twisted light. Opt. Lett.41, 5744. 10.1364/OL.41.005744 (2016). [DOI] [PubMed] [Google Scholar]
  • 8.Luis, A. Phase-shift amplification for precision measurements without nonclassical states. Phys. Rev. A65, 025802. 10.1103/PhysRevA.65.025802 (2002). [Google Scholar]
  • 9.Giovannetti, V., Lloyd, S. & Maccone, L. Quantum metrology. Phys. Rev. Lett.96, 10401. 10.1103/PhysRevLett.96.010401 (2006). [DOI] [PubMed] [Google Scholar]
  • 10.Higgins, B. L., Berry, D. W., Bartlett, S. D., Wiseman, H. M. & Pryde, G. J. Entanglement-free Heisenberg-limited phase estimation. Nature450, 393–396. 10.1038/nature06257 (2007). [DOI] [PubMed] [Google Scholar]
  • 11.Koppell, S. A., Israel, Y., Bowman, A. J., Klopfer, B. B. & Kasevich, M. A. Transmission electron microscopy at the quantum limit. Appl. Phys. Lett.120, 190502 (2022). [Google Scholar]
  • 12.Nimmrichter, S., Chen, C.-F., Klopfer, B. B., Kasevich, M. A. & Juffmann, T. Full-field cavity enhanced microscopy techniques. J. Phys. Photon.1, 015007. 10.1088/2515-7647/aae228 (2018). [Google Scholar]
  • 13.Gigan, S., Lopez, L., Treps, N., Maître, A. & Fabre, C. Image transmission through a stable paraxial cavity. Phys. Rev. A72, 023804. 10.1103/PhysRevA.72.023804 (2005). [Google Scholar]
  • 14.Israel, Y., Reynolds, J. L., Klopfer, B. B. & Kasevich, M. A. Continuous wave multi-pass imaging flow cytometry. Optica10, 491. 10.1364/OPTICA.482316 (2023). [Google Scholar]
  • 15.Cao, H., Chriki, R., Bittner, S., Friesem, A. A. & Davidson, N. Complex lasers with controllable coherence. Nat. Rev. Phys.1, 156–168. 10.1038/s42254-018-0010-6 (2019). [Google Scholar]
  • 16.Slobodkin, Y. et al. Massively degenerate coherent perfect absorber for arbitrary wavefronts. Science377, 995–998. 10.1126/science.abq8103 (2022). [DOI] [PubMed] [Google Scholar]
  • 17.Kolobov, M. I. Quantum Imaging (Springer, 2007). [Google Scholar]
  • 18.Guo, Y. et al. An optical lattice with sound. Nature599, 211–215. 10.1038/s41586-021-03945-x (2021). [DOI] [PubMed] [Google Scholar]
  • 19.Bouchet, D., Dong, J., Maestre, D. & Juffmann, T. Fundamental bounds on the precision of classical phase microscopes. Phys. Rev. Appl.15, 024047. 10.1103/PhysRevApplied.15.024047 (2021). [Google Scholar]
  • 20.Zuo, C. et al. Transport of intensity equation: A tutorial. Opt. Lasers Eng.135, 106187 (2020) (10.1016/j.optlaseng.2020.106187). [Google Scholar]
  • 21.Mazzinghi, C. et al. Cavity-enhanced polarization rotation measurements for low-disturbance probing of atoms. Opt. Express29, 40854. 10.1364/OE.438675 (2021). [Google Scholar]
  • 22.Ketterle, W., Durfee, D. S. & Stamper-Kurn, D. M. Making, probing and understanding Bose–Einstein condensates. arXiv:cond-mat/9904034 (1999).
  • 23.Andrews, M. R. et al. Direct, nondestructive observation of a bose condensate. Science273, 84–87. 10.1126/science.273.5271.84 (1996). [DOI] [PubMed] [Google Scholar]
  • 24.Altuntaş, E. & Spielman, I. B. Quantum back-action limits in dispersively measured Bose–Einstein condensates. Commun. Phys.6, 66. 10.1038/s42005-023-01181-5 (2023). [Google Scholar]
  • 25.Kroeze, R. M., Marsh, B. P., Lin, K.-Y., Keeling, J. & Lev, B. L. High cooperativity using a confocal-cavity-qed microscope. PRX Quant.4, 020326. 10.1103/PRXQuantum.4.020326 (2023). [Google Scholar]
  • 26.Hodgson, N. & Weber, H. Optical Resonators: Fundamentals, Advanced Concepts, Applications (Springer, 2005). [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Data Availability Statement

Source data and all other data that support the plots within this paper and other findings of this study are available from the corresponding author upon reasonable request.