Skip to main content
Journal of Applied Crystallography logoLink to Journal of Applied Crystallography
. 2025 Jul 29;58(Pt 4):1439–1446. doi: 10.1107/S1600576725005862

Oblique diffraction geometry for the observation of several non-coplanar Bragg reflections under identical illumination

Carsten Detlefs a,*, Axel Henningsson b, Brinthan Kanesalingam c,d, Adam A W Cretton e, Cedric Corley-Wiciak a, Felix T Frankus b, Dayeeta Pal c,d, Sara Irvine c,d, Sina Borgi e, Henning F Poulsen e, Can Yildirim a, Leora E Dresselhaus-Marais c,d
Editor: J Keckesf
PMCID: PMC12321021  PMID: 40765970

We derive a closed analytical expression that allows determination of the strain and lattice rotation from the deviation of experimental observables (e.g. goniometer angles) from their nominal position for an unstrained lattice. We also propose an experimental geometry wherein such measurements can be carried out while keeping the illuminated sample volume constant, facilitating registration of data.

Keywords: X-ray diffraction, strain, dark-field X-ray microscopy

Abstract

A method to determine the strain tensor and local lattice rotation with dark-field X-ray microscopy is presented. Using a set of at least three non-coplanar symmetry-equivalent Bragg reflections, the illuminated volume of the sample can be kept constant for all reflections, facilitating easy registration of the measured lattice variations. This requires an oblique diffraction geometry, i.e. the diffraction plane is neither horizontal nor vertical. We derive a closed analytical expression that allows determination of the strain and lattice rotation from the deviation of experimental observables (e.g. goniometer angles) from their nominal values for an unstrained lattice.

1. Introduction

Hierarchically organized crystalline structures are ubiquitous in technological and natural materials such as metals, semiconductors, ceramics, biominerals, geological materials and many others. The microscopic crystal structure (e.g. grains or domains) and the atomic scale defect networks embedded therein determine many of the macroscopic physical and mechanical properties of these materials. The need to study crystalline microstructures, and in particular the spatial variations in the strain fields generated by defects, is therefore persistent throughout materials science (Callister & Rethwisch, 2000).

Diffraction-contrast microscopy techniques have historically been highly effective at mapping strain in crystalline materials. Electron microscopy techniques have mapped strain by electron backscatter diffraction (Schwartz et al., 2009) and transmission electron microscopy (Williams & Carter, 2009). X-ray techniques are complementary by allowing a larger field of view for bulk strain measurements with a resolution that is more coarse spatially but finer angularly. For example, scanning Laue diffraction microscopy can map the 3D strain tensor (Tamura et al., 2003; Liu et al., 2004). By measuring many diffraction peaks simultaneously, the deviatoric strain tensor and lattice rotation are obtained. In order to determine isotropic strain precisely, the precise photon energy of selected reflections has to be measured as well. Similarly, scanning X-ray diffraction microscopy probes the lattice strain by mapping the distortion of single structure peaks in a spatial raster scan (Zatterin et al., 2025). Both are raster-scanning techniques that are relatively slow when large areas or volumes have to be scanned with high resolution. More recently, Bragg coherent diffraction imaging (BCDI) demonstrated similar advances to solve for the 3D strain tensor field based on the amplitude and phase information from image reconstructions along three non-coplanar lattice planes (Hofmann et al., 2020), capturing high-resolution views of sub-micrometre crystals or grains.

Dark-field X-ray microscopy (DFXM; Simons et al., 2015) can be used to map strain fields within bulk crystalline mater­ials over regions of hundreds of micrometres. As a new technique, however, to date it has been used primarily to measure along only one single Bragg reflection, giving access to one component of the strain tensor and two components that are a mix of strain and local lattice rotation; in order to determine all six independent components of the strain tensor ɛ and the three indpendent components of the local lattice rotation w, at least three non-coplanar Bragg reflections must be measured. In DFXM, measurements often focus on strain fields around isolated structure defects such as dislocations or coherent small-angle (sub)grain boundaries. In this context, the local deviation from the nominal unstrained crystal structure is of interest (Poulsen et al., 2021). Hence the spatial variation in the strain tensor ɛ and the local lattice rotation w relative to a reference has to be measured [in order to avoid confusion with the goniometer rotation ω defined below, we use w to denote the local lattice rotation, whereas the literature often uses ω, e.g. Poulsen et al. (2021)].

Monochromatic diffraction techniques require the sample to be rotated when changing reflections. Specifically, in a typical DFXM measurement the diffraction plane is fixed to the xz plane, with 1D-focused line beam illumination perpendicular to this plane (section topography). Alternatively, structured illumination can be used (Gürsoy et al., 2025). In both of these cases, aligning multiple non-coplanar Bragg reflections then requires the sample to be rotated about several axes (e.g. μ, χ and ϕ; Poulsen et al., 2017), which changes the virtual section illuminated by the line beam. Registration of the different measurements into a 3D model of the lattice strain in the sample is therefore a challenge.

Chung & Ice (1999) and Abboud et al. (2017) circumvented this challenge by using white-beam Laue diffraction, where several Bragg reflections can be observed without moving the sample. This method allows for precise measurement of the deviatoric strain but relies on the relatively poor energy resolution of the detector for the isotropic strain (change in unit-cell volume).

An alternative approach is, for each reflection, to reconstruct the change in Bragg position within the entire gauge volume and then register the volumes measured at different reflections. This technique was adopted by Chatelier et al. (2024), who used BCDI to measure the 3D displacement field within an isolated 350 nm particle of Ni3Fe. A further alternative approach for macroscopic sample sizes could be the ‘topo–tomo’ technique (Ludwig et al., 2001). Here the sample is rotated about a selected reciprocal-space vector of the unstrained sample and tomographic methods are used to reconstruct a 3D model of the strain components probed by this reflection. Extending this method to three or more non-coplanar Bragg reflections, however, is mechanically challenging, as for each reflection a full 360° rotation about the corresponding reciprocal-lattice vector g is needed.

Here we describe a procedure for the observation of several non-coplanar Bragg reflections for the determination of the complete strain tensor in DFXM and other X-ray diffraction techniques that measure one reflection at a time. The main idea of this procedure is to facilitate the registration of the volumes probed at the different reflections by ensuring that the same sample volume is illuminated for all reflections. In the geometry proposed here, we achieve constant illumination by rotating the sample about an axis perpendicular to the incident beam. This is particularly convenient for the case of line beam illumination, as the 3D registration problem is reduced to a 2D problem within the illuminated section, with a known rotation angle between the different measurements. We choose this rotation axis to be along the laboratory Inline graphic axis (pointing upwards; Fig. 1) and align a reference direction hsym = (hsym, ksym, ℓsym)T of the sample parallel to the rotation axis. Observing this reference reflection is not necessary for the procedure described below, as it only serves as a symmetry reference. Observation may, however, help with alignment of the symmetry axis to the ω rotation axis. While the formalism was developed with DFXM in mind (specifically for measuring the strain field around isolated dislocations and coherent subgrain boundaries), the method can also be used for other experiments performed in the z-axis geometry (Bloch, 1985), such as classical section topography (Caliste et al., 2021; Yoneyama et al., 2023) or scanning X-ray diffraction (Corley-Wiciak et al., 2024), and for other types of samples.

Figure 1.

Figure 1

Diffraction geometry. The incident beam kin travels along the Inline graphic axis. The diffracted beam is first rotated by −2θ about the Inline graphic axis and then by η about the Inline graphic axis [equation (25)].

The paper is organized into two main parts. First, Sections 2 and 3 discuss the general formalism for the reconstruction of the full strain tensor and lattice rotation from a series of Bragg reflections by classical X-ray diffraction (XRD). Our treatment in this section is general and does not rely on a specific diffraction geometry, symmetry or other details of the experiment. We assume that the deviations of each reflection from their known unstrained reference positions are small and we thus treat those deviations as perturbations upon the main diffracting beam. In the second part, Sections 4 and 5 discuss how our general approach can be applied to a DFXM experiment using the geometry outlined above, i.e. by rotating the sample about a single axis which is perpendicular to the incident beam. This paper focuses on the diffraction geometry and reciprocal-space characteristics of the experiment, leaving the imaging details for future work (Henningsson et al., 2025; Kanesalingam et al., 2025).

2. General formalism

Before examining the specific geometry that we use in our experiments, we first share some general observations about measuring the strain tensor with X-ray diffraction, specifically how the strain and lattice rotations can be derived from angular shifts of the peak positions without explicit reconstruction of the distorted unit cell as proposed by Schlenker et al. (1978).

In this section, we assume a generalized diffractometer where the reciprocal-space vector in the diffraction condition is given by a set of angles ξ = (ξ1, ξ2,…, ξm). Possible examples include the classical four-circle diffractometer (Busing & Levy, 1967) where ξ = (2θ, ω, χ, ϕ), the DFXM diffractometer (Poulsen et al., 2017) where ξ = (2θ, η, μ, ω, χ, ϕ) and many others (Bloch, 1985; Lohmeier & Vlieg, 1993; You, 1999; Bunk & Nielsen, 2004). The ξm can be generalized to other observables, e.g. the photon energy.

2.1. Defining the sample and goniometer

In this work, we describe the sample using the UB matrix, as introduced by Busing & Levy (1967) and Poulsen et al. (2017). In this formalism, the 3 × 3 matrix B = (a*, b*, c*) contains the basis vectors a*, b* and c* of the reciprocal lattice. Here we use the convention of Schlenker et al. (1978) and Poulsen et al. (2017), where B0 of the undistorted crystal is a lower triangular matrix. The orthogonal 3 × 3 matrix U describes how the sample is oriented relative to the mounting point of the diffractometer (Busing & Levy, 1967; Poulsen et al., 2017; Poulsen et al., 2021).

In the coordinate system attached to the sample mounting point of the diffractometer, the reciprocal-space vector gs corresponding to the Miller indices h = (h, k, ℓ)T is given by (Poulsen et al., 2017)

2.1.

We refer to this as the sample coordinate system (Poulsen et al., 2021) and decorate vectors in this coordinate system with the subscript s.

For a given reciprocal-space vector to be studied, the diffractometer angles must be set to specific positions ξ. These positions are not unique, due to the redundancy of the goniometer angles; see e.g. Busing & Levy (1967). Taking the reverse view, given the angle settings of the diffractometer ξ, we can calculate the reciprocal-space vector under study (Poulsen et al., 2017). We write this vector as g(ξ). We assume that the corresponding Miller indices (h, k, ℓ)T are known but that this vector deviates from a known reference position g0 due to strain.

In this section, we focus on the general aspects of how strain induces shifts in the goniometer angles at which a given reflection is observed and how this effect can be used to determine the strain tensor.

2.2. Strain

For small deformations, the reciprocal-lattice in the strained state can be described by

2.2.

where F1 + ɛ + w is a 3 × 3 tensor and F−T = (F−1)T [see e.g. Bernier et al. (2011) and Poulsen et al. (2021)]. ɛ describes the strain and w describes a superimposed lattice rotation.

Here, the strain tensor ɛ is symmetric with six free parameters,

2.2.

because ɛxy = ɛyxetc.

In addition there may be lattice rotations, described by the antisymmetric matrix w with three free parameters,

2.2.

Different Bragg reflections h = (h, k, ℓ)T are sensitive to different parts of the strain and rotation tensors, leading to a (small) deviation of the reciprocal-space vector relative to its nominal (unstrained) value, g = g0 + Δg, where g0 = UB0h in the unstrained case, and g = UBh in the strained case.

The shift in reciprocal space due to strain and lattice rotation is then given by (Bernier et al., 2011; Poulsen et al., 2021)

2.2.

In an experiment, the reciprocal-space vector will have a small shift in angular peak positions Δξ = (Δξ1, Δξ2,…, Δξm) compared with the nominal unstrained peak positions ξ = (ξ1, ξ2,…, ξm),

2.2.

To a linear approximation,

2.2.

where p enumerates the diffractometer angles and Inline graphic is the 3 × m matrix of gradients of the observed position in reciprocal space with respect to the goniometer angles.

3. Determining strain and lattice rotation from observed peak shifts

Equations (5) and (8) relate the shift of one reciprocal-space vector as a function of the corresponding goniometer angles. As discussed above, at least three non-coplanar Bragg reflections have to be observed in order to calculate the complete strain tensor and lattice rotation. We therefore generalize these equations by grouping several vectors into (rectangular) matrices:

3.
3.
3.
3.

where n is the number of Bragg reflections under study. Note that this definition of H is different from that of Poulsen et al. (2021).

Then equations (5) and (8) are generalized to

3.

For n > 3 non-coplanar vectors the system is overdetermined and, under the assumption of Gaussian isotropic noise in ΔG, the error in equation (14) is minimized by (Anton & Kaul, 2020)

3.

where the terms in square brackets are 3 × 3 matrices. The strain and rotation components are isolated by being symmetric and antisymmetric, respectively [equations (3) and (4)].

Equation (16) provides a closed non-iterative formula that connects the angular shifts observed in the diffraction experiment, collected in the matrix ΔΞ, to the lattice deformations F−T − 1 and thus material strains ɛ. It is therefore our central result in this work.

It can also provide guidance for planning an experiment, specifically for selecting a set of Bragg reflections. Equation (16) will only yield a valid result if the 3 × 3 matrix [B0H(B0H)T] can be inverted, i.e. when its determinant is non-zero. In the case of three reflections, this implies that the determinant of H does not vanish, i.e. that the chosen reciprocal-space vectors are not coplanar. Furthermore, for numerical stability it is desirable that this determinant is large. In other words, the vectors have to be chosen such that they are sufficiently far from limiting cases of being collinear or coplanar. Measuring more than three reflections will reduce statistical errors through averaging.

To aid with an eventual implementation in a programming language such as Python, we also provide the explicit summations of the various tensor products above: the index i = x, y, z labels coordinate axes, the index j = 1, 2,…, n labels reciprocal-space vectors or Bragg reflections and the index p = 1, 2,…, m labels diffractometer angles. Equations (9) to (14) then take the form

3.
3.
3.
3.
3.

The product between (ΔΞG) and (B0H)T is summed over the index j which labels the reciprocal-space vectors.

4. Specific geometry for dark field X-ray microscopy

We now turn to discussing the specific experimental geometry used in our experiments. For DFXM, recording the goniometer and associated strain information in 2D images raises the problem of reconstructing a 3D sample volume from 2D projections. Since the crystal orientation differs significantly between different structure peaks, this implies that the voxels of the sample corresponding to each pixel differ significantly, requiring image registration to consolidate the volumes obtained from different Bragg reflections into each specific sample region over all orientations ξi.

In this section, we address the image registration challenge by proposing a measurement geometry that can ensure the same sample volume is illuminated for all non-coplanar orientations. Constant illumination is achieved by reducing the sample movement to a single rotation ω about an axis perpendicular to the incident beam. This diffraction geometry is a 1S+2D geometry [one sample rotation, two detector rotations; see Bloch (1985) and Bunk & Nielsen (2004)], even though two additional sample rotations are needed to align the chosen symmetry axis parallel to the main axis of rotation (ω axis).

The configuration is similar to the z-axis geometry (Bloch, 1985; Bunk & Nielsen, 2004), the 3DXRD geometry (Poulsen et al., 2001; Jakobsen, 2006) and that used in diffraction contrast tomography (King et al., 2008; Ludwig et al., 2009). The DFXM goniometer has additional sample rotations, ϕ and χ (Poulsen et al., 2017). As these additional axes are not moved during this experiment we do not consider them explicitly and assume that they are absorbed into U. Full geometry calculations can be found in the literature, e.g. Poulsen et al. (2017). In this geometry, there are two detector angles, θ and η, and one sample angle, ω, such that ξ = (ω, η, θ)T (Fig. 1).

Contrary to most DFXM experiments, in our proposed geometry the diffracted beam is not confined to the vertical plane. Instead it is rotated out of this plane by an angle η (Poulsen et al., 2017). A similar approach relying instead on a 2S+1D geometry (two sample rotations and one detector rotation) was used in scanning X-ray microscopy experiments where a nano-beam was raster-scanned across the sample (Richter et al., 2022; Corley-Wiciak et al., 2023; Corley-Wiciak et al., 2024).

A unique advantage of the geometry proposed here is that the selected symmetry-equivalent reflections can all be observed at the same detector position and only a single sample rotation is required to switch between them. As we show below, several such detector positions exist. The number of measurements can thus be increased by collecting the same reflections at several possible detector positions. Friedel pairs (h, k, ℓ) and (−h, −k, −ℓ) can be collected to improve the statistics further.

DFXM experiments are often carried out with a 1D (line) focused beam, allowing access to the full 6D data set for mechanical deformation in the sample (three dimensions in reciprocal space + three in real space). The method proposed in this work allows for registration even when using 1D focused illumination, because the measured 3D volume remains constant as the sample rotates about a single axis that is orthogonal to the incident beam ω (Fig. 1). There are downsides to this geometry. Notably, only certain reflections with α > θ can fulfil the diffraction condition, where α is the angle between the reflection under study and the ω axis of rotation [equation (39)].

4.1. Summary of the different coordinate systems

Following Poulsen et al. (2017), we define a sequence of coordinate systems as follows:

(i) The laboratory coordinate system, indicated by the subscript ℓ [e.g.g; see Fig. 1 and Poulsen et al. (2017)], is fixed and does not move with any goniometer rotations. The direction of the incident beam is along the positive Inline graphic axis. The Inline graphic axis is horizontal to port (left when seen along the beam axis). The Inline graphic axis is up.

(ii) The sample coordinate system as defined by Poulsen et al. (2021) is attached to the gonimeter’s innermost sample rotation (here the ω axis), indicated by the subscript s, e.g.gs. The sample’s orientation, defined by the matrix U, is fixed in this coordinate system. For this experiment, only the goniometer rotation ω about the Inline graphic axis is used. All other rotation angles are set to zero. Thus the matrix Γ as used by Poulsen et al. (2021) is given by Γ = Rz(ω).

(iii) Miller indices in the crystal coordinate system. Here we use the symbol h = (h, k, ℓ)T without a subscript.

These three systems are related by

4.1.
4.1.

4.2. Defining the incident and diffracted beams

In the laboratory coordinate system, the incident beam travels along the positive Inline graphic axis,

4.2.

where λ is the X-ray wavelength.

Rather than using the angles δ and γ to describe the detector rotations (Bloch, 1985; Lohmeier & Vlieg, 1993; You, 1999; Bunk & Nielsen, 2004), we employ the description used for large area detectors, for example in 3DXRD (Poulsen et al., 2001), and also used for DFXM (Poulsen et al., 2017). Here the beam is first rotated by 2θ about the Inline graphic axis and then by η about the incident beam axis (Fig. 1). The diffracted beam is thus given by

4.2.

Here Rx,y,z(α) are (right-handed) rotations about the x, y, z axes through the angle α, respectively (Poulsen et al., 2017). The scattering vector (in the laboratory coordinate system) is given by

4.2.

In particular,

4.2.

4.3. Simplified sample goniometer

To obtain the scattering vector Qs in the sample coordinate system, this vector is rotated about the Inline graphic axis by −ω [Fig. 1 and equation (23)],

4.3.

Bragg’s law is fulfilled when this scattering vector is equal to the reciprocal-lattice vector of an allowed Bragg reflection,

4.3.

4.4. Choice of reciprocal-lattice vectors

We are interested in a series of reciprocal-space vectors of the undistorted crystal that are related by symmetry, specifically vectors that can be transformed into each other by rota­tion about a symmetry axis hsym = Inline graphic. The corresponding reciprocal-space vector in the sample coordinate system is gsym,s = UB0hsym. This axis is used as a reference only; in general we do not observe diffraction from this reciprocal-space vector.

We assume the undistorted sample to be mounted such that this symmetry axis is parallel to the goniometer ω axis, i.e. the matrix U is chosen such that

4.4.

As the ω axis is parallel to Inline graphic, the Inline graphic axes of the laboratory and sample coordinate systems are identical, Inline graphic axis for all settings of ω. Consequently, ω rotations do not change gsym,ℓ and gsym,s = gsym,ℓ.

As the ω axis is perpendicular to the incident beam direction, the illuminated volume of the sample remains unchanged when the sample is rotated about the ω axis, especially when the incident beam is focused to a line for section topography (Caliste et al., 2021; Yoneyama et al., 2023).

Next, we select a family of symmetry-equivalent reflections hn = (hn, kn, ℓn)T that can be transformed into each other via rotations about the symmetry axis or reflection by mirror planes parallel to the symmetry axis. At least three symmetry-equivalent reflections are required. Therefore the lattice system has to be orthorhombic, tetragonal, rhombohedral, hexagonal or cubic. Examples are listed in Table 1.

Table 1. Examples of groups of symmetry-equivalent reflections (h, k, ℓ)T that can be transformed into each other via rotations about a symmetry axis.

C denotes cubic crystal symmetry, T tetragonal, O orthorhombic, H hexagonal and R rhombohedral.

Crystal symmetry Symmetry axis (hsym, ksym, ℓsym) Reflections (hn, kn, ℓn)
C, T (2, 2, 0) (2, 0, ±2), (0, 2, ±2)
    (3, 1, ±1), (1, 3, ±1)
C, R (1, 1, 1) (1, 1, −1), (1, −1, 1), (−1, 1, 1)
    (4, 0, 0), (0, 4, 0), (0, 0, 4)
    (3, 1, 1), (1, 3, 1), (1, 1, 3)
C, T, O (0, 0, 4) (±1, ±1, 1)
C, T (0, 0, 4) (±3, ±1, 1), (±1, ±3, 1)
H (0, 0, 2) (±1, 0, 1), (0, ±1, 1), (±1, ∓1, 1)
    (±1, ±1, 2), ( 2, ±1, 2), (±1, ∓2, 2)

For simplicity, we consider only one of these vectors, g, which we rotate about the Inline graphic axis such that it lies within the yz plane. Let

4.4.

with Inline graphic and Inline graphic.

The different symmetry-equivalent reciprocal-space vectors then differ only by their respective values of ω0. Therefore the following calculations are carried out only for one representative vector, gs, taking into account its orientation ω0.

4.5. Diffraction condition

The ω axis of the goniometer then rotates this vector about the Inline graphic axis, such that the reciprocal-space vector in the laboratory coordinate system is given by

4.5.

The task is now to find the angles θ, η and ω such that equation (37) equals equation (28). θ is found by comparing the length of the two vectors,1 equation (29):

4.5.

This is Bragg’s law, as d = 2π/|g|.

Left-multiplying equations (36) and (28) with Rz(−ω + ω0)/|g| yields

4.5.

For simplicity we assume 0 < α < π/2. An additional set of solutions can be found for −π/2 < α < 0. This provides an opportunity for additional measurements. Solving this yields

4.5.
4.5.

Note that due to symmetry requirements all reflections share the same values of θ and α, and therefore of ω − ω0 and η. Only the value of ω0 (and therefore ω) varies from reflection to reflection. In other words, all reflections can be measured without moving the detector (except scanning around the nominal position).

Valid solutions are only found for Inline graphic, i.e. α > θ. We use the arctan2 function, which correctly determines the sector and avoids division by zero:

4.5.
4.5.

4.6. Linearized shifts

In our diffraction geometry, Δξ = (Δω, Δη, Δθ)T. Equation (8) is explicitly

4.6.

Note again that θ and η have the same value for all reflections under consideration; only the value of ω varies.

5. Examples

To demonstrate how the method could be put into practice, we present an example specifically chosen for the dark-field microscopy instrument on beamline ID03 at the ESRF (Isern et al., 2025). For the detector geometry on this beamline, we need −π/2 < η < 0 (Poulsen et al., 2017; Isern et al., 2025). A common application of DFXM is the study of dislocations and dislocation structures in metals. One of the metals most commonly studied by DFXM is aluminium (Simons et al., 2015; Dresselhaus-Marais et al., 2021; Yildirim et al., 2023).

Aluminium has a face-centred cubic structure with lattice constant a = 4.0495 Å. We use Inline graphic as the reference direction and measure the reflections h1…4 = (2, 0, ±2)T and Inline graphic. The angle between the reference direction and the reciprocal-lattice vectors of interest is α =arccos(1/2) = 60°.

Let

5.

such that the symmetry axis Inline graphic is parallel to Inline graphic,

5.

The reciprocal-lattice vectors in the laboratory coordinate system are then

5.
5.
5.
5.

At E = 19.1 keV, θ ≃ 13.103°, 2θ ≃ 26.206°. ω − ω0 ≃ 15.175° and η ≃ −59.112°.

6. Conclusion

In conclusion, we have presented an experimental procedure and data analysis process to derive the full strain tensor from a series of dark-field X-ray microscopy measurements. The method relies on measuring small angular deviations between strained and unstrained parts of the sample, e.g. around dislocations in near-perfect crystals. Treating these deviations as perturbations, we have derived a non-iterative closed formula for calculating the corresponding lattice strain and rotation relative to the unstrained reference lattice.

By choosing an oblique diffraction plane and rotating the sample about a symmetry axis, several symmetry-equivalent, but non-coplanar, Bragg reflections can be measured without moving the detector. This greatly facilitates the registration of the gauge volume within the sample, which is necessary for the 3D reconstruction of the strain field throughout the sample volume. We explicitly calculate the sample and detector angles for this oblique geometry.

Future work will extend our approach to take into account the imaging aspect of DFXM, in particular the fact that strained and unstrained regions of the sample can be measured simultaneously.

While this study has focused on a single crystal, integrating DFXM with grain-resolved 3DXRD paves the way for full strain tensor mapping in polycrystalline materials. In such cases, 3DXRD can identify grains of interest and their orientations, enabling targeted DFXM scans (Shukla et al., 2025). Ongoing developments in X-ray optics for DFXM measurements, particularly diamond-based lenses (Seiboth et al., 2017; Celestre et al., 2022) and higher-energy X-rays, will expand the accessible hkl range, supporting broader application of this method to studies with complex industrially relevant polycrystalline materials.

Funding Statement

Can Yildirim acknowledges financial support by the European Research Council Starting Grant ‘D-REX’ No. 101116911. Axel Henningsson, Adam Cretton, Felix Frankus, Sina Borgi and Henning Poulsen acknowledge support from ERC Advanced Grant No. 885022 and from the Danish ESS Lighthouse on Hard Materials in 3D, SOLID. Contributions from Leora Dresselhaus-Marais and Sara Irvine were supported by the US Department of Energy (DOE), Office of Basic Energy Sciences, Division of Materials Sciences and Engineering (contract DE-AC02-76SF00515). This study has received financial support from the European Union AddMorePower scheme via the H2020 European Research Council programme (grant No. 101091621 to Cedric Corley-Wiciak.

Footnotes

1

The generalized, but equivalent, equation system was solved by Henningsson & Hall (2023) for an arbitrary orientation of the ω axis. Here, we provide an alternative derivation, valid for the special case when the ω axis is kept parallel to z.

References

  1. Abboud, A., Kirchlechner, C., Keckes, J., Conka Nurdan, T., Send, S., Micha, J. S., Ulrich, O., Hartmann, R., Strüder, L. & Pietsch, U. (2017). J. Appl. Cryst.50, 901–908. [DOI] [PMC free article] [PubMed]
  2. Anton, H. & Kaul, A. (2020). Elementary linear algebra, 12th ed. John Wiley and Sons.
  3. Bernier, J. V., Barton, N. R., Lienert, U. & Miller, M. P. (2011). J. Strain Anal. Eng. Des.46, 527–547.
  4. Bloch, J. M. (1985). J. Appl. Cryst.18, 33–36.
  5. Bunk, O. & Nielsen, M. M. (2004). J. Appl. Cryst.37, 216–222.
  6. Busing, W. R. & Levy, H. A. (1967). Acta Cryst.22, 457–464.
  7. Caliste, T. N. T., Drouin, A., Caliste, D., Detlefs, C. & Baruchel, J. (2021). Appl. Sci.11, 9054.
  8. Callister, W. D. & Rethwisch, D. G. (2000). Fundamentals of materials science and engineering. Wiley London.
  9. Celestre, R., Antipov, S., Gomez, E., Zinn, T., Barrett, R. & Roth, T. (2022). J. Synchrotron Rad.29, 629–643. [DOI] [PMC free article] [PubMed]
  10. Chatelier, C., Atlan, C., Dupraz, M., Leake, S., Li, N., Schülli, T. U., Levi, M., Rabkin, E., Favre, L., Labat, S., Eymery, J. & Richard, M.-I. (2024). ACS Nano18, 13517–13527. [DOI] [PubMed]
  11. Chung, J.-S. & Ice, G. E. (1999). J. Appl. Phys.86, 5249–5255.
  12. Corley-Wiciak, C., Zoellner, M., Zaitsev, I., Anand, K., Zatterin, E., Yamamoto, Y., Corley-Wiciak, A., Reichmann, F., Langheinrich, W., Schreiber, L., Manganelli, C., Virgilio, M., Richter, C. & Capellini, G. (2023). Phys. Rev. Appl.20, 024056.
  13. Corley-Wiciak, C., Zoellner, M. H., Corley-Wiciak, A. A., Rovaris, F., Zatterin, E., Zaitsev, I., Sfuncia, G., Nicotra, G., Spirito, D., von den Driesch, N., Manganelli, C. L., Marzegalli, A., Schulli, T. U., Buca, D., Montalenti, F., Capellini, G. & Richter, C. (2024). Small Methods8, 2400598. [DOI] [PubMed]
  14. Dresselhaus-Marais, L. E., Winther, G., Howard, M., Gonzalez, A., Breckling, S. R., Yildirim, C., Cook, P. K., Kutsal, M., Simons, H., Detlefs, C., Eggert, J. H. & Poulsen, H. F. (2021). Sci. Adv.7, eabe8311. [DOI] [PMC free article] [PubMed]
  15. Gürsoy, D., Yay, K., Kisiel, E., Wojcik, M., Sheyfer, D., Last, A., Highland, M., Fisher, I. R., Hruszkewycz, S. & Islam, Z. (2025). Commun. Phys.8, 34.
  16. Henningsson, A., Borgi, S., Winther, G., El-Azab, A. & Poulsen, H. F. (2025). J. Mech. Phys. Solids204, 106277.
  17. Henningsson, A. & Hall, S. A. (2023). J. Appl. Cryst.56, 282–292. [DOI] [PMC free article] [PubMed]
  18. Hofmann, F., Phillips, N. W., Das, S., Karamched, P., Hughes, G. M., Douglas, J. O., Cha, W. & Liu, W. (2020). Phys. Rev. Mater.4, 013801.
  19. Isern, H., Brochard, T., Dufrane, T., Brumund, P., Papillon, E., Scortani, D., Hino, R., Yildirim, C., Rodriguez Lamas, R., Li, Y., Sarkis, M. & Detlefs, C. (2025). J. Phys. Conf. Ser.3010, 012163.
  20. Jakobsen, B., Poulsen, H. F., Lienert, U., Almer, J., Shastri, S. D., Sørensen, H. O., Gundlach, C. & Pantleon, W. (2006). Science312, 889–892. [DOI] [PubMed]
  21. Kanesalingam, B., Chalise, D., Detlefs, C. & Dresselhaus-Marais, L. (2025). arXiv, 2507.17929.
  22. King, A., Johnson, G., Engelberg, D., Ludwig, W. & Marrow, J. (2008). Science321, 382–385. [DOI] [PubMed]
  23. Liu, W., Ice, G. E., Larson, B. C., Yang, W., Tischler, J. Z. & Budai, J. D. (2004). Metall. Mater. Trans. A35, 1963–1967.
  24. Lohmeier, M. & Vlieg, E. (1993). J. Appl. Cryst.26, 706–716.
  25. Ludwig, W., Cloetens, P., Härtwig, J., Baruchel, J., Hamelin, B. & Bastie, P. (2001). J. Appl. Cryst.34, 602–607.
  26. Ludwig, W., Reischig, P., King, A., Herbig, M., Lauridsen, E. M., Johnson, G., Marrow, T. J. & Buffière, J. Y. (2009). Rev. Sci. Instrum.80, 033905. [DOI] [PubMed]
  27. Poulsen, H. F., Dresselhaus-Marais, L. E., Carlsen, M. A., Detlefs, C. & Winther, G. (2021). J. Appl. Cryst.54, 1555–1571.
  28. Poulsen, H. F., Jakobsen, A. C., Simons, H., Ahl, S. R., Cook, P. K. & Detlefs, C. (2017). J. Appl. Cryst.50, 1441–1456.
  29. Poulsen, H. F., Nielsen, S. F., Lauridsen, E. M., Schmidt, S., Suter, R. M., Lienert, U., Margulies, L., Lorentzen, T. & Juul Jensen, D. (2001). J. Appl. Cryst.34, 751–756.
  30. Richter, C., Kaganer, V. M., Even, A., Dussaigne, A., Ferret, P., Barbier, F., Le Vaillant, Y.-M. & Schülli, T. U. (2022). Phys. Rev. Appl.18, 064015.
  31. Schlenker, J. L., Gibbs, G. V. & Boisen, M. B. (1978). Acta Cryst. A34, 52–54.
  32. Schwartz, A. J., Kumar, M., Adams, B. L. & Field, D. P. (2009). Editors. Electron backscatter diffraction in materials science, 2nd ed. New York: Springer.
  33. Seiboth, F., Schropp, A., Scholz, M., Wittwer, F., Rödel, C., Wünsche, M., Ullsperger, T., Nolte, S., Rahomäki, J., Parfeniukas, K., Giakoumidis, S., Vogt, U., Wagner, U., Rau, C., Boesenberg, U., Garrevoet, J., Falkenberg, G., Galtier, E. C., Ja Lee, H., Nagler, B. & Schroer, C. G. (2017). Nat. Commun.8, 14623. [DOI] [PMC free article] [PubMed]
  34. Shukla, A., Yildirim, C., Ball, J., Detlefs, C., Cretton, A. A. W. & Nenningsson, N. A. (2025). In preparation.
  35. Simons, H., King, A., Ludwig, W., Detlefs, C., Pantleon, W., Schmidt, S., Stöhr, F., Snigireva, I., Snigirev, A. & Poulsen, H. F. (2015). Nat. Commun.6, 6098. [DOI] [PMC free article] [PubMed]
  36. Tamura, N., MacDowell, A. A., Spolenak, R., Valek, B. C., Bravman, J. C., Brown, W. L., Celestre, R. S., Padmore, H. A., Batterman, B. W. & Patel, J. R. (2003). J. Synchrotron Rad.10, 137–143. [DOI] [PubMed]
  37. Williams, D. B. & Carter, C. B. (2009). Transmission electron microscopy, 2nd ed. New York: Springer.
  38. Yildirim, C., Poulsen, H. F., Winther, G., Detlefs, C., Huang, P. H. & Dresselhaus-Marais, L. E. (2023). Sci. Rep.13, 3834. [DOI] [PMC free article] [PubMed]
  39. Yoneyama, A., Ishiji, K., Sakaki, A., Kobayashi, Y., Inaba, M., Fukuda, K., Konishi, K., Shima, A. & Takamatsu, D. (2023). Sci. Rep.13, 12381. [DOI] [PMC free article] [PubMed]
  40. You, H. (1999). J. Appl. Cryst.32, 614–623.
  41. Zatterin, E. M., Barbier, F., Torrengo, S., Ferret, P., Veux, G., Martin, B., Lévy, F., Schülli, T. U. & Dussaigne, A. (2025). Physica RRL19, 2400241.

Articles from Journal of Applied Crystallography are provided here courtesy of International Union of Crystallography

RESOURCES