Abstract
Study background and aims
Pseudomonas aeruginosa can persistently contaminate endoscopes by forming biofilms within internal channels, complicating both detection and eradication. Current microbiological surveillance methods have limited efficacy and may yield false-negative results. This study aimed to identify proteomic markers of P. aeruginosa biofilms on endoscope channel material.
Methods
Three genetically unrelated P. aeruginosa isolates from contaminated duodenoscopes and two reference strains (ATCC 27853 and PAO1) were used. Biofilms were grown on disinfected endoscope biopsy channel rings and incubated for 24, 48, and 72 h. Matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) was employed to analyze temporal changes in protein spectra. Peaks of interest were further characterized by liquid chromatography–tandem mass spectrometry (LC-MS/MS) and whole-genome sequencing to identify associated proteins. To further confirm the origin of these peaks, strains naturally lacking the corresponding genes were analyzed.
Results
MALDI-TOF MS revealed distinct time- and strain-specific spectral profiles, with two notable peaks at approximately 2723 m/z and 5450 m/z. LC-MS/MS identified the 5450 m/z peak as PA2146, corresponding to a 5449.1 Da protein after in vivo methionine cleavage. The 2723 m/z peak was confirmed as its doubly charged ion. Both peaks were absent in strains naturally lacking PA2146, confirming it as the source.
Conclusion
PA2146 expression increases during P. aeruginosa biofilm development on endoscope channel surfaces, indicating its potential as a biomarker for contamination. MALDI-TOF MS could enhance biofilm detection in endoscope surveillance. Further research should assess the clinical utility of proteomic approaches for improving endoscopic microbiological safety.
Keywords: Pseudomonas aeruginosa, Biofilms, Proteomics, Mass spectrometry, Matrix-assisted laser desorption-ionization (MALDI-TOF MS), Bacterial proteins, Biomarkers, Duodenoscopes, Endoscopy, Gastrointestinal, Disinfection, Microbial contamination, Instrument contamination
1. Introduction
Pseudomonas aeruginosa is a well-known cause of hospital-acquired infections, particularly in immunocompromised patients. Contaminated medical devices, including endoscopes, serve as potential transmission routes, with duodenoscopes implicated in outbreaks involving resistant strains [1]. Contamination rates of up to 30 % have been reported in gastroscopes and colonoscopes [2]. Ready-to-use duodenoscopes show contamination with gut or oral microbiota in 15–19 % of cases [3,4]. Our previous research identified P. aeruginosa as one of the most prevalent contaminants in duodenoscopes [5]. P. aeruginosa is known for its strong ability to form biofilms that persist in the internal channels of an endoscope. Such biofilms act as robust protective barriers, shielding the bacteria from conventional cleaning agents and leading to persistent contamination despite repeated reprocessing efforts [5,6]. Currently, no effective method exists for removing biofilms from endoscope channels, leaving channel replacement the only viable solution.
To minimize patient exposure to contaminated endoscopes, guidelines recommend routine microbiological surveillance [[7], [8], [9]]. However, current sampling and culturing protocols have demonstrated limited efficacy [10]. This limitation is likely due to biofilm formation in endoscope channels, which impedes the sampling of embedded bacteria and may lead to false-negative culture results [11,12]. This underscores the urgent need for novel detection methods capable of reliably identifying biofilm presence.
P. aeruginosa biofilm is structurally complex and dynamic, comprising extracellular DNA, proteins, lipids, and polysaccharides that vary according to strain and environmental conditions [13,14]. Proteomic analyses have revealed that numerous genes and proteins are differentially expressed during biofilm formation, highlighting their roles in biofilm architecture and antimicrobial resistance [15,16]. Among these, a recent study by Kaleta et al. identified PA2146 as a key protein upregulated during biofilm growth in P. aeruginosa isolates associated with clinical infections [16].
Identifying proteins and genes uniquely associated with P. aeruginosa biofilms in duodenoscope channels may enable the development of more sensitive and targeted detection strategies. Matrix assisted laser desorption ionization-time-of-flight mass spectrometry (MALDI-TOF MS), widely used for bacterial identification, has demonstrated the ability to monitor P. aeruginosa biofilm development [17]. Using MALDI-TOF MS, liquid chromatography–tandem mass spectrometry (LC-MS/MS), and whole-genome sequencing analyses, this study aimed to identify proteomic markers of P. aeruginosa biofilm formation in strains which were found in microbial surveillance cultures of duodenoscopes.
2. Materials and methods
2.1. Strain selection
Two standard reference strains, ATCC 27853 and PAO1 (designated isolates A and B, respectively), were included due to their widespread use in P. aeruginosa research. Both are well-characterized, with publicly available whole genome sequencing (WGS) data enabling comparative analyses [18]. In addition, three genetically distinct clinical isolates (C, D, and E), confirmed by WGS, were selected from a microbiological surveillance database maintained by the Department of Medical Microbiology and Infectious Diseases. Isolates C and E were recovered from separate incidents of persistent contamination in duodenoscope biopsy channels, likely due to biofilm formation. Isolate D, a Verona Integron-encoded Metallo-β-lactamase (VIM)-2 producing strain, was previously linked to a duodenoscope-related outbreak in 2012 [19]. To further support experimental findings, strains F and G, lacking the PA2146 gene, were later included. These were selected from an in-house strain collection based on their genome sequences.
2.2. Biofilm generation in vitro model
An unused polytetrafluoroethylene (PTFE) biopsy channel from a Pentax ED34-i10T2 duodenoscope was sectioned into 4 mm rings. A biopsy channel was used because it is one of the most frequently contaminated duodenoscope sites [5,20]. The 4 mm size was chosen for compatibility with 48-well plates, facilitating efficient experimental handling. To ensure uniformity, a custom PTFE tube slicer was designed and 3D-printed specifically for this purpose (Supplementary appendix Figure S1). Biopsy channel rings (BCRs) were disinfected by immersion in 70 % ethanol for 24 h, then air-dried in a sterile flow cabinet. Fresh overnight cultures of all isolates were grown on Tryptic Soy Agar (TSA; Becton, Dickinson and Company, Franklin Lakes, NJ, USA) and incubated at 37 °C for 24 h. Disinfected BCRs were placed horizontally in 500 μL of Tryptic Soy Broth (TSB; same supplier), and inoculated with 10 μL of a 0.5 McFarland suspension prepared from a single-strain overnight culture. Incubation was carried out for 24, 48, and 72 h. To enable simultaneous termination, the 72-h condition was initiated on day 0, the 48-h on day 1, and the 24-h on day 2. On day 3, all BCRs were removed and thoroughly washed in 0.9 % NaCl to eliminate loosely attached material.
2.3. Mass spectrometry
For MALDI-TOF (Bruker, Bremen, Germany) analysis, protein extraction was performed using formic acid and acetonitrile [21]. Briefly, BCRs were washed and transferred to Eppendorf tubes containing 300 μL of water, followed by the addition of 900 μL of water. After centrifugation at 22,000g for 5 min, the supernatant was discarded, and BCRs were briefly dried under a nitrogen stream. Subsequently, 20 μL of 70 % formic acid was added, and BCRs were incubated for 5 min, after which 20 μL of acetonitrile was added. Following a second centrifugation at 22,000g for 5 min, 1 μL of the supernatant was spotted in triplicate onto a MALDI Biotarget (Bruker, Bremen, Germany), air-dried, and overlaid with 1 μL of α-cyano-4-hydroxycinnamic acid (CHCA) matrix solution. Once dried, the target plate was loaded into the MALDI-TOF. Spectra were recorded in linear mode within a mass-to-charge (m/z) range of 2K–20K, at a laser intensity of 35 %, using a recording summing 200 shots, to a total of 800 shots per spectrum. This range is widely used for detecting peptides and proteins with masses between 2000 Da and 20,000 Da, including multiple ribosomal proteins (Fig. 1). This approach enables precise genus and species identification and facilitates the detection of biofilm-associated proteins. All reagents were of analytical grade and purchased from Sigma-Aldrich, unless stated otherwise.
Fig. 1.
MALDI-TOF mass spectrum of Pseudomonas aeruginosa from blood agar plate (ATCC 27853). Unique molecular fingerprint of Pseudomonas aeruginosa obtained by MALDI-TOF MS. Each peak represents a specific molecular ion, primarily ribosomal proteins and other abundant biomolecules. The x-axis (m/z) indicates the mass-to-charge ratio, while the y-axis shows relative intensity, reflecting ion abundance. Detected masses correspond to proteins encoded by the bacterial genome, providing insights into molecular composition and structure. ∗ mark known ribosomal proteins, which serve as molecular markers for species identification and as internal references to detect relative changes in other protein abundances.
For LC-MS/MS analysis, the incubation and processing of BCRs were performed identically to those analyzed with MALDI-TOF MS, except that protein extraction was replaced by trypsin treatment. Following a water/ethanol wash, the supernatant was discarded, and BCRs were treated with 50 μL of 5 % sodium deoxycholate (SDC) solution (Sigma-Aldrich, St. Louis, MO, USA). Samples were subjected to ultrasonic bathing (42 kHz) for 5 min at room temperature, followed by incubation at 80 °C for 10 min with shaking at 450 rounds per minute (rpm) in an Eppendorf ThermoMixer C (Eppendorf, Hamburg, Germany).
Subsequently, 450 μL of MS-grade water (Fisher Chemical, Waltham, MA, USA) was added to each sample. Trypsin (5 μL of 1 μg/μL; Trypsin Gold, Worthington Biochemical Corporation, NJ, USA) was then added, and the tubes were incubated at 37 °C for 1 h with shaking at 450 rpm. Following incubation, 12.5 μL of 5 % trifluoroacetic acid (TFA; Thermo Fisher Scientific, Waltham, MA, USA) was added to each tube, resulting in immediate protein precipitation. Tubes were briefly vortexed and centrifuged at maximum g-force for 5 min using a 5424 refrigerated centrifuge (Eppendorf, Hamburg, Germany).
Mass spectrometry measurements were performed using a capillary liquid chromatography system (Thermo Fisher Scientific, Germany) online coupled to an Orbitrap Fusion Lumos Mass Spectrometer (Thermo Fisher Scientific, San Jose, CA, USA), as described previously [22]. From the raw data files, MS/MS spectra were converted into mascot generic format (MGF) files using ProteoWizard (version 3.0.22130). MGF peak lists were searched against the uniprot_sprot_v201212 database (restricted to bacteria; 334,772 sequences) using Mascot (version 2.3.01; Matrix Science, London, UK). Mascot search results were further analyzed in Scaffold (version 5.3.2, Portland, OR, USA), with protein and peptide confidence levels set to a 1 % false discovery rate (FDR). FDRs were estimated by including a decoy database search generated by Mascot.
2.4. Whole genome sequencing
To support the mass-spectrometry analysis, and to confirm gene/protein relation, WGS of all strains involved was performed. Genomic DNA was isolated from freshly grown cultures of the study isolates using the Quick-DNA Fungal/Bacterial Miniprep Kit (Zymo Research via Baseclear, Leiden, Netherlands). Libraries of ∼350 base pairs fragments were created and sequenced by Novogene (Cambridge, United Kingdom) on Illumina platforms (Illumina, San Diego, CA, USA) generating 150 base pairs paired end reads. De novo assemblies were created using CLC Genomics Workbench v21 (Qiagen, Hilden, Germany) using default parameters. To examine presence of and variation within the target sequence and to overcome possible assembly issues based on short read sequencing alone (e.g. false negatives), both an assembly-based and an assembly-free protocol were used. In the assembly-based protocol, assemblies were imported into BioNumerics v7.6 software. The sequence extraction module was used to extract the target sequence from the assemblies based on a minimal length coverage of 95 % and minimal sequence identity of 80 % to the reference sequence. Extracted sequences were translated and aligned to identify amino acid variants. In the assembly-free protocol, sequence reads were mapped onto the reference sequence in CLC Genomics Workbench using default parameters. From these mappings, a consensus sequence was generated, translated and aligned to identify amino acid variants.
2.5. Protein identification
To identify MALDI-TOF MS peaks associated with biofilm development, spectra from isolates A to E were acquired at the three time points. Peaks were evaluated for temporal intensity changes relative to ribosomal protein peaks, and those consistently increasing across all isolates were selected for further analysis. To determine the proteins corresponding to these peaks, a search was performed in the UniProt-reviewed Swiss-Prot proteome database (taxonomy ID: 208964) to match detected masses with annotated proteins. In parallel, LC-MS/MS analysis was conducted, and the resulting spectra were compared against a database of known protein and peptide masses. Candidate proteins were cross-referenced with WGS data to confirm their presence in the isolates. Finally, isolates F and G, naturally deficient in the PA2146 gene, were analyzed by repeating the MALDI-TOF MS biofilm protocol to assess the impact of gene deletion on the spectral profile.
3. Results
P. aeruginosa reference strains A and B, along with clinical isolates C, D, and E, were incubated on BCRs for 24, 48, and 72 h. Biofilm formation was observed in all samples. Visual inspection confirmed extensive biofilm development, with the medium transforming into a dense, slime-like matrix (Fig. 2). This was accompanied by blue, green, or brown discoloration, attributed to the production of pyoverdine and pyocyanin, secondary metabolites commonly associated with P. aeruginosa biofilm development [13].
Fig. 2.
Pseudomonas aeruginosa biofilm formation on a biopsy channel ring (BCR). The pipette tip holds a BCR from the 72-h time point.
3.1. Temporal and strain-specific variations in MALDI-TOF MS spectra
MALDI-TOF MS spectra were generated for all strains across the three time incubation points (24, 48, and 72 h), and compared both over time, and between samples. Distinct differences in spectral patterns were observed between strains and over time. Two peaks of particular interest, at 2723 m/z and 5450 m/z, showed clear temporal and strain-specific trends. In control strains A and B, these peaks were of low intensity at 24 h but increased significantly by 48 and 72 h. In clinical isolates, particularly isolate D, these peaks were already present at high intensities at 24 h, further intensifying over time (Fig. 3). The consistent spectral trend across replicates underscores their biological relevance.
Fig. 3.
MALDI-TOF mass spectra of six Pseudomonas aeruginosa strains recorded at 24, 48, and 72 h post-incubation. The 5450 m/z peak (highlighted in red) is shown relative to 5739 m/z, to which the y-axis is scaled, unless the 5450 m/z had a higher intensity (Isolate D). Rows represent individual strains, and columns correspond to time points. The x-axis denotes mass-to-charge ratio (m/z), and the y-axis represents signal intensity of the individual graphs.
3.2. Protein identification
Using WGS data, we attempted to identify the 5450 m/z peak. However, no direct candidates were found, as no genes encoding a protein with a final mass of 5450 Da were present in the sequenced genomes. By comparing MALDI-TOF MS spectra with LC-MS/MS analysis, we identified four proteins within the 5000–6100 Da region (Table 1). A representative MALDI-TOF MS spectrum is shown in Fig. 4. Three of the four detected proteins, bL34, bL33, and bL7, could be directly correlated with a MALDI-TOF MS peak. For the fourth protein, PA2146, no direct 1:1 correlation was initially observed based on its mass. However, further analysis of the LC-MS/MS data revealed that PA2146 had undergone in vivo methionine cleavage, a common post-translational modification in bacterial proteins [23]. This modification reduced the actual mass of the in vivo protein to 5449.1 Da, aligning with the detected peak at 5450 m/z, thereby confirming PA2146 as the corresponding protein (Table 1).
Table 1.
Proteins identified within the 5000–6100 Da region of MALDI-TOF MS spectra were based on consistent detection across isolates. Peaks were further analyzed via LC-MS/MS, and proteins were marked as detected when corresponding peptides were identified.
| Measured peak + predicted mass (Da) | Protein name | MALDI-TOF MS | LC-MS/MS |
|---|---|---|---|
| A. 5210 | Large ribosomal subunit protein bL34 | ✓ | ✓ |
| B. 5580/5450a | Biofilm-associated protein PA2146 | ✓ | ✓ |
| C. 5739 | Ribosomal protein bL7 | ✓ | ✗ |
| D. 6045 | Large ribosomal subunit protein bL33 | ✓ | ✓ |
✓ indicates successful detection ✗ denotes no detection.
Represents the mass of PA2146 after in vivo methionine cleavage.
Fig. 4.
Representative MALDI-TOF MS spectrum of Pseudomonas aeruginosa strain B, highlighting the 5000–6100 m/z region. Peaks P (5210 m/z), Q (5450 m/z), R (5739 m/z), and S (6045 m/z) correspond to proteins listed in Table 1, with Peak Q (5450 m/z) identified as the biofilm-associated protein PA2146.
Additionally, the peak at approximately 2723 m/z was identified as the doubly charged ion of PA2146, calculated based on the molecular weight of its singly charged species at 5450 m/z and accounting for ionization mass [24].
3.3. Confirmation of PA2146
To further confirm the identity of peak Q as PA2146, we searched our in-house database of sequenced P. aeruginosa genomes for strains lacking the PA2146 gene. Two such strains were retrieved from storage, cultured on agar, and analyzed. As they do not encode PA2146, these strains are incapable of producing the corresponding protein. Repetition of the time-series experiment with the strains lacking PA2146 (strains F and G) confirmed the specificity of this peak, as neither 5450 m/z nor its doubly charged ion at 2723 m/z was detected under identical conditions at any time-point (Fig. 3). Strains F and G were still able to form biofilm, indicating that absence of PA2146 does not entirely impair biofilm formation.
4. Discussion
This study identifies PA2146 as a biofilm-associated protein upregulated during P. aeruginosa biofilm maturation on PTFE duodenoscope biopsy channel rings. Detection was performed using MALDI-TOF MS, a widely accessible and cost-effective proteomic platform. To our knowledge, this is the first application of proteomic biofilm profiling in the context of endoscope contamination. These findings suggest the potential of PA2146 as a biomarker for biofilm detection and underscore the need to clarify its functional role in biofilm development, which may support targeted strategies to prevent or disrupt biofilms in endoscopic devices.
Multiple studies have illustrated the upregulation of PA2146 in biofilm development of P. aeruginosa clinical strains [14,16]. A recent study identified PA2146 as the most consistent and substantial upregulated gene during P. aeruginosa biofilm development [16]. The same study also illustrated that PA2146 is an important gene involved in biofilm architecture and the resistance of P. aeruginosa in biofilm to tobramycin and hydrogen peroxide. Homologs of PA2146 gene have also been discovered in other Proteobacteria, such as Klebsiella pneumoniae and Escherichia coli, where it similarly influences biofilm formation and antibiotic tolerance. However, the precise role of the PA2146 protein in extracellular matrix production, cell adhesion, or intercellular communication within biofilms remains unclear. Our findings show that PA2146 is also consistently upregulated during biofilm formation on endoscope biopsy channel material in isolates cultured from contaminated endoscopes. Nevertheless, P. aeruginosa isolates lacking PA2146 are still capable of forming biofilm. Therefore, sample collection for culture remains essential for detecting endoscope contamination, even if detection of PA2146 through MALDI-TOF MS were to be implemented in clinical practice. Furthermore, future research should investigate the expression and functional role of PA2146 homologs in other common endoscope-associated pathogens to assess their potential as cross-species biofilm biomarkers.
Currently, the detection of endoscope contamination relies on endoscope sampling and culturing [9]. However, the sensitivity of these cultures is limited, with false-negative results occurring in up to 17.8 % of cases [12,25]. P. aeruginosa is detected in up to 7.2 % of surveillance cultures from duodenoscopes and has been implicated in multiple duodenoscope-associated outbreaks worldwide [5,19,26,27]. Biofilm formation is believed to play a major role in persistent duodenoscope contamination, particularly in the case of build-up biofilms, which are formed as concentric layers over successive reprocessing cycles [6]. In such a build-up biofilm, P. aeruginosa can remain embedded in a viable but nonculturable (VBNC) state for up to 26 weeks of dry storage [11]. Consequently, there is an urgent need for the development and implementation of a biofilm detection method to enhance surveillance efforts and facilitate the early identification of persistent contamination. In future experiments, the BCR model developed in this study could be modified to more closely mimic build-up biofilm accumulation. This could be achieved by incorporating simulated reprocessing steps used for clinically employed endoscopes, including brushing, flushing, and rinsing with detergents and disinfectants such as peracetic acid. This will allow a further assessment of changes in the proteomic profile including the presence of the PA2146 peak.
MALDI-TOF MS offers high specificity and sensitivity for P. aeruginosa detection and identification even in the presence of genetic mutations [28]. Furthermore, it has previously been demonstrated to effectively differentiate between various stages of biofilm development, providing valuable insights into the dynamics of biofilm maturation and persistence [17]. Its relatively low cost and broad availability make MALDI-TOF MS a practical and accessible tool with potential for routine clinical applications, including microbiological surveillance of endoscopes. Notably, the analysis can be completed in under 1 h. However, to enable the practical use of MALDI-TOF MS in surveillance, extraction methods tailored for intact, clinically used endoscopes should be developed. Future studies should assess whether MALDI-TOF MS can be used to reliably detect PA2146 in samples collected from endoscopes persistently contaminated with P. aeruginosa using the flush-brush-flush method, a standard technique for duodenoscope sampling [10]. Moreover, the methods used in this study may help to identify potential biomarkers in other gastrointestinal microbiota, broadening the applicability of our approach to other biofilm-forming organisms.
There are several limitations to this study. This was a single center study using a limited number of clinical isolates which may impact generalizability. Additionally, the rings of the biopsy channels in our experiment were not subjected to endoscope reprocessing protocols and thus were not exposed to the detergents and chemicals that might influence biofilm formation [29]. Furthermore, in order to maximally stimulate biofilm formation, the biofilms were grown with ample nutrients, which does not accurately reflect the nutrient-poor environment of a biopsy channel from a fully reprocessed and dried endoscope. As a result, our in vitro model may differ from biofilms typically present in endoscopes, where repeated reprocessing cycles can lead to the development of build-up biofilms with distinct structural properties. The upregulation of proteins specifically associated with such biofilms may have been missed. This study focused on biofilms in the biopsy or suction channels of endoscopes, where a flush-brush-flush method can potentially be used to assess PA2146 expression. However, due to the significantly narrower diameter of air/water channels, the flush-brush-flush method is not feasible. Therefore, alternative sampling techniques will need to be developed to evaluate PA2146 upregulation in these channels. While PA2146 was significantly upregulated during biofilm formation in clinical isolates, its expression may also be influenced by environmental factors or microbial competition within the endoscope. Stress conditions, such as nutrient limitation, oxidative stress, or exposure to antimicrobial agents, could similarly induce PA2146 expression, although this was not observed in a previous study [16]. Further research is needed to assess whether our experimental conditions contributed to the observed trends and to determine how well they replicate the environment of reprocessed and dried duodenoscope channels.
5. Conclusion
The upregulation of PA2146 indicates P. aeruginosa biofilm maturation on endoscope biopsy channel material and may serve as a biomarker for detecting biofilm-associated contamination in endoscopes. Future studies are needed to determine whether targeted detection of PA2146 is feasible in clinical samples from endoscope channels and whether its presence correlates with viable P. aeruginosa isolates.
CRediT authorship contribution statement
Koen van der Ploeg: Writing – review & editing, Writing – original draft, Validation, Project administration, Investigation, Conceptualization. Corné P. de Vogel: Writing – review & editing, Validation, Resources, Investigation. Corné H.W. Klaassen: Writing – review & editing, Validation, Methodology. Theo M. Luider: Writing – review & editing, Resources, Investigation. Lona Zeneyedpour: Investigation. Bibi C.G.C. Mason- Slingerland: Writing – review & editing, Supervision. Margreet C. Vos: Writing – review & editing, Supervision. Marco J. Bruno: Writing – review & editing, Supervision. Michiel L. Bexkens: Writing – review & editing, Writing – original draft, Visualization, Validation, Supervision, Software, Methodology, Investigation, Formal analysis, Data curation, Conceptualization. Juliëtte A. Severin: Writing – review & editing, Supervision, Methodology, Conceptualization.
Declaration of competing interest
The authors declare the following financial interests/personal relationships which may be considered as potential competing interests: Margreet C. Vos reports a relationship with Boston Scientific Corporation that includes: funding grants. Marco J. Bruno reports a relationship with Boston Scientific Corporation that includes: consulting or advisory, funding grants, and speaking and lecture fees. Marco J. Bruno reports a relationship with Cook Medical Inc that includes: consulting or advisory, funding grants, and speaking and lecture fees. Marco J. Bruno reports a relationship with PENTAX Medical that includes: consulting or advisory, funding grants, and speaking and lecture fees. Marco J. Bruno reports a relationship with Mylan Pharmaceuticals Inc that includes: funding grants. Marco J. Bruno reports a relationship with ChiRoStim that includes: funding grants. Marco J. Bruno reports a relationship with Ambu that includes: funding grants. If there are other authors, they declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Acknowledgements
This study was conducted without external funding.
Footnotes
Supplementary data to this article can be found online at https://doi.org/10.1016/j.bioflm.2025.100310.
Appendix A. Supplementary data
The following is/are the supplementary data to this article.
A 3D-printed polytetrafluoroethylene tube slicer, used with a sterile scalpel, to ensure precise sectioning
Data availability
Data will be made available on request.
References
- 1.Ofstead C.L., Buro B.L., Hopkins K.M., Eiland J.E., Wetzler H.P., Lichtenstein D.R. Duodenoscope-associated infection prevention: a call for evidence-based decision making. Endosc Int Open. 2020;8(12):E1769–E1781. doi: 10.1055/a-1264-7173. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Goyal H., Larsen S., Perisetti A., Larsen N.B., Ockert L.K., Adamsen S., et al. Gastrointestinal endoscope contamination rates - elevators are not only to blame: a systematic review and meta-analysis. Endosc Int Open. 2022;10(6):E840–E853. doi: 10.1055/a-1795-8883. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Rauwers A.W., Voor In 'T Holt A.F., Buijs J.G., De Groot W., Hansen B.E., Bruno M.J., Vos M.C. High prevalence rate of digestive tract bacteria in duodenoscopes: a nationwide study. Gut. 2018;67(9):1637–1645. doi: 10.1136/gutjnl-2017-315082. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.van der Ploeg K., Vos M.C., Erler N.S., Bulkmans A.J.C., Mason-Slingerland B., Severin J.A., Bruno M.J. Impact of duodenoscope reprocessing factors on duodenoscope contamination: a retrospective observational study. J Hosp Infect. 2024;154:88–94. doi: 10.1016/j.jhin.2024.09.018. [DOI] [PubMed] [Google Scholar]
- 5.van der Ploeg K., Haanappel C.P., Voor In 't holt A.F., de Groot W., Bulkmans A.J.C., Erler N.S., et al. Unveiling 8 years of duodenoscope contamination: insights from a retrospective analysis in a large tertiary care hospital. Gut. 2024 doi: 10.1136/gutjnl-2023-330355. [DOI] [PubMed] [Google Scholar]
- 6.Alfa M.J., Singh H. Impact of wet storage and other factors on biofilm formation and contamination of patient-ready endoscopes: a narrative review. Gastrointest Endosc. 2020;91(2):236–247. doi: 10.1016/j.gie.2019.08.043. [DOI] [PubMed] [Google Scholar]
- 7.Beilenhoff U., Biering H., Blum R., Brljak J., Cimbro M., Dumonceau J.M., et al. Prevention of multidrug-resistant infections from contaminated duodenoscopes: position statement of the european society of gastrointestinal endoscopy (ESGE) and european society of gastroenterology nurses and Associates (ESGENA) Endoscopy. 2017;49(11):1098–1106. doi: 10.1055/s-0043-120523. [DOI] [PubMed] [Google Scholar]
- 8.Dutch Society for Medical Microbiology . 2018. Guideline for monitoring the microbiological safety of thermolabile flexible high-risk gastrointestinal endoscopes. [Google Scholar]
- 9.The U.S. Food and Drug Administration (FDA) tCfDCaP, (CDC) ASfMA . 2018. Guideline duodenoscope surveillance sampling-and culturing protocols. [Google Scholar]
- 10.Pineau L., Alfa M., Radix C. Endoscope sampling and culturing methods. J Hosp Infect. 2024;149:36–45. doi: 10.1016/j.jhin.2024.03.017. [DOI] [PubMed] [Google Scholar]
- 11.Alfa M.J., Ribeiro M.M., da Costa Luciano C., Franca R., Olson N., DeGagne P., Singh H. A novel polytetrafluoroethylene-channel model, which simulates low levels of culturable bacteria in buildup biofilm after repeated endoscope reprocessing. Gastrointest Endosc. 2017;86(3):442–451 e1. doi: 10.1016/j.gie.2017.05.014. [DOI] [PubMed] [Google Scholar]
- 12.van der Ploeg K., Severin J.A., Klaassen C.H.W., Vos M.C., Bruno M.J., Mason-Slingerland B.C.G.C. Contaminated Duodenoscopes in endoscopic retrograde cholangiopancreatography: sensitivity of detection and risk of underdetection. Gastrointest Endosc. 2024 doi: 10.1016/j.gie.2024.08.004. [DOI] [PubMed] [Google Scholar]
- 13.Thi M.T.T., Wibowo D., Rehm B.H.A. Pseudomonas aeruginosa biofilms. Int J Mol Sci. 2020;21(22) doi: 10.3390/ijms21228671. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Thoming J.G., Tomasch J., Preusse M., Koska M., Grahl N., Pohl S., et al. Parallel evolutionary paths to produce more than one Pseudomonas aeruginosa biofilm phenotype. NPJ Biofilms Microbiomes. 2020;6:2. doi: 10.1038/s41522-019-0113-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Erdmann J., Thoming J.G., Pohl S., Pich A., Lenz C., Haussler S. The core proteome of biofilm-grown clinical Pseudomonas aeruginosa isolates. Cells. 2019;8(10) doi: 10.3390/cells8101129. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Kaleta M.F., Petrova O.E., Zampaloni C., Garcia-Alcalde F., Parker M., Sauer K. A previously uncharacterized gene, PA2146, contributes to biofilm formation and drug tolerance across the ɣ-Proteobacteria. npj Biofilms and Microbiomes. 2022;8(1):54. doi: 10.1038/s41522-022-00314-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Pereira F.D., Bonatto C.C., Lopes C.A., Pereira A.L., Silva L.P. Use of MALDI-TOF mass spectrometry to analyze the molecular profile of Pseudomonas aeruginosa biofilms grown on glass and plastic surfaces. Microb Pathog. 2015;86:32–37. doi: 10.1016/j.micpath.2015.07.005. [DOI] [PubMed] [Google Scholar]
- 18.Labaer J., Qiu Q., Anumanthan A., Mar W., Zuo D., Murthy T.V., et al. The pseudomonas aeruginosa PA01 gene collection. Genome Res. 2004;14(10B):2190–2200. doi: 10.1101/gr.2482804. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Verfaillie C.J., Bruno M.J., Holt A.F.V.I.T., Buijs J.G., Poley J.W., Loeve A.J., et al. Withdrawal of a novel-design duodenoscope ends outbreak of a VIM-2-producing Pseudomonas aeruginosa. Endoscopy. 2015;47(6):493–502. doi: 10.1055/s-0034-1391886. [DOI] [PubMed] [Google Scholar]
- 20.Forbes N., Elmunzer B.J., Allain T., Parkins M.D., Sheth P.M., Waddell B.J., et al. Effect of disposable elevator cap duodenoscopes on persistent microbial contamination and technical performance of endoscopic retrograde cholangiopancreatography: the ICECAP randomized clinical trial. JAMA Intern Med. 2023 doi: 10.1001/jamainternmed.2022.6394. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Topić Popović N., Kazazić S.P., Bojanić K., Strunjak-Perović I., Čož-Rakovac R. Sample preparation and culture condition effects on MALDI-TOF MS identification of bacteria: a review. Mass Spectrom Rev. 2023;42(5):1589–1603. doi: 10.1002/mas.21739. [DOI] [PubMed] [Google Scholar]
- 22.Zeneyedpour L., Stingl C., Dekker L.J.M., Mustafa D.A.M., Kros J.M., Luider T.M. Phosphorylation ratio determination in fresh-frozen and formalin-fixed paraffin-embedded tissue with targeted mass spectrometry. J Proteome Res. 2020;19(10):4179–4190. doi: 10.1021/acs.jproteome.0c00354. [DOI] [PubMed] [Google Scholar]
- 23.Gamerdinger M., Jia M., Schloemer R., Rabl L., Jaskolowski M., Khakzar K.M., et al. NAC controls cotranslational N-terminal methionine excision in eukaryotes. Science. 2023;380(6651):1238–1243. doi: 10.1126/science.adg3297. [DOI] [PubMed] [Google Scholar]
- 24.Blumer M.R., Chang C.H., Brayfindley E., Nunez J.R., Colby S.M., Renslow R.S., Metz T.O. Mass spectrometry adduct calculator. J Chem Inf Model. 2021;61(12):5721–5725. doi: 10.1021/acs.jcim.1c00579. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Heuvelmans M., Wunderink H.F., van der Mei H.C., Monkelbaan J.F. A narrative review on current duodenoscope reprocessing techniques and novel developments. Antimicrob Resist Infect Control. 2021;10(1):171. doi: 10.1186/s13756-021-01037-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Fernández-Cuenca F., López-Cerero L., Cabot G., Oliver A., López-Méndez J., Recacha E., et al. Nosocomial outbreak linked to a flexible gastrointestinal endoscope contaminated with an amikacin-resistant ST17 clone of Pseudomonas aeruginosa. Eur J Clin Microbiol Infect Dis. 2020;39(10):1837–1844. doi: 10.1007/s10096-020-03915-7. [DOI] [PubMed] [Google Scholar]
- 27.Qiu L., Zhou Z., Liu Q., Ni Y., Zhao F., Cheng H. Investigating the failure of repeated standard cleaning and disinfection of a pseudomonas aeruginosa-infected pancreatic and biliary endoscope. Am J Infect Control. 2015;43(8):e43–e46. doi: 10.1016/j.ajic.2015.04.204. [DOI] [PubMed] [Google Scholar]
- 28.Oumeraci T., Jensen V., Talbot S.R., Hofmann W., Kostrzewa M., Schlegelberger B., et al. Comprehensive MALDI-TOF biotyping of the non-redundant harvard pseudomonas aeruginosa PA14 transposon insertion mutant library. PLoS One. 2015;10(2) doi: 10.1371/journal.pone.0117144. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Beilenhoff U., Biering H., Blum R., Brljak J., Cimbro M., Dumonceau J.M., et al. Reprocessing of flexible endoscopes and endoscopic accessories used in gastrointestinal endoscopy: position statement of the european society of gastrointestinal endoscopy (ESGE) and european society of gastroenterology nurses and associates (ESGENA) - update 2018. Endoscopy. 2018;50(12):1205–1234. doi: 10.1055/a-0759-1629. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
A 3D-printed polytetrafluoroethylene tube slicer, used with a sterile scalpel, to ensure precise sectioning
Data Availability Statement
Data will be made available on request.