A 70-year-old woman with a refractory duodenal ulcer developed a 1.5-cm perforation at the duodenal bulb following two failed attempts at endoscopic hemostasis ( Fig. 1 ). Placement of an over-the-scope clip for closure was unsuccessful, and resulted in sepsis.
Fig. 1.
Endoscopic views showing: a the original duodenal perforation; b a failed closure attempt using over-the-scope clips.
A nitinol duodenal covered stent (22 mm in diameter, 10 cm in length) was deployed endoscopically ( Fig. 2 , Video 1 ). This stent integrates three key innovations: (i) a tripartite anchoring system – an umbrella-shaped gastric component for antral apposition, a cylindrical pyloric segment for radial fixation, and a bulbous duodenal anchor for coaptation; (ii) a papilla-aligned fenestration within the membranous mid-section to preserve physiological pancreatobiliary drainage; and (iii) optimized radial force distribution across segmented zones to minimize migration. Follow-up at 15 days confirmed closure of the complete perforation, and the stent was safely retrieved.
Fig. 2.
Images of treatment with a duodenal stent showing: a the fistula prior to insertion of the stent; b stent deployment; c the anchoring mechanism; d stent retrieval after the perforation had healed; e the anatomical design of the stent; f the healed perforation site.
An anatomically designed duodenal stent that can both obliterate the fistula and provide physiological drainage is used to close a perforation of a refractory duodenal ulcer in an elderly frail woman.
Video 1
Duodenal perforations are challenging owing to anatomical fragility and the enzyme-rich secretions 1 2 . Conventional stents risk migration and obstructing drainage, thereby exacerbating leaks 3 . An anatomically designed duodenal stent uniquely addresses these limitations: its anchoring system ensures stable positioning without tissue injury, while the fenestration aligns precisely with the papilla, allowing continuous biliary/pancreatic flow into the bowel lumen. This design prevents reflux-related complications and promotes fistula healing by isolating the defect from digestive enzymes. The case underscores the stent’s dual role as a mechanical sealant and physiological conduit, which may be particularly critical in elderly patients with impaired healing.
Endoscopy_UCTN_Code_TTT_1AO_2AZ
Footnotes
Conflict of Interest The authors declare that they have no conflict of interest.
Endoscopy E-Videos https://eref.thieme.de/e-videos .
E-Videos is an open access online section of the journal Endoscopy , reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high-quality video and are published with a Creative Commons CC-BY license. Endoscopy E-Videos qualify for HINARI discounts and waivers and eligibility is automatically checked during the submission process. We grant 100% waivers to articles whose corresponding authors are based in Group A countries and 50% waivers to those who are based in Group B countries as classified by Research4Life (see: https://www.research4life.org/access/eligibility/ ). This section has its own submission website at https://mc.manuscriptcentral.com/e-videos .
References
- 1.Saito K, Ishikawa Y, Kitahara M et al. Novel rendezvous technique for covered metal stent placement using balloon-assisted endoscope in duodenal perforation. Endoscopy. 2024;56:E420–E421. doi: 10.1055/a-2313-3265. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Nakagawa Y, Nagai T, Soma W et al. Endoscopic closure of a large ERCP-related lateral duodenal perforation by using endoloops and endoclips. Gastrointest Endosc. 2010;72:216–217. doi: 10.1016/j.gie.2009.10.040. [DOI] [PubMed] [Google Scholar]
- 3.Williams J, Joshi H, Schwartz M et al. Endoscopic repair of duodenal perforations, a scoping review. Surg Endosc. 2024;38:4839–4845. doi: 10.1007/s00464-024-11133-x. [DOI] [PMC free article] [PubMed] [Google Scholar]