Abstract
Silicone elastomers based on polydimethylsiloxane are biocompatible and non-biodegradable thermosetting polymers used in various drug delivery applications, including subdermal implants, vaginal rings, and intrauterine devices. Without exception, all marketed silicone elastomer drug delivery products provide sustained or controlled release of highly hydrophobic small drug molecules, since drug solubility in the silicone matrix is a prerequisite for molecular diffusion and release. We are interested in developing multipurpose silicone elastomer vaginal rings for local administration of metal ions—such as copper and zinc—for non-hormonal contraception and antimicrobial therapy. However, sustained/controlled release of metal ions from silicone elastomers containing metal nanopowders or metal salts is challenging due to their limited solubility in silicone. In this study, we assess the potential for enhancing the release of copper or zinc ions from silicone elastomer devices by co-formulating copper nanopowder, zinc nanopowder, copper sulfate, or zinc acetate with four common pharmaceutical excipients—gelatin, polyvinylpyrrolidone, sucrose, and hydroxypropyl methylcellulose. The study demonstrates that (i) copper/zinc nanopowders and salts can be successfully incorporated into addition-cure silicone elastomers, (ii) in vitro release of Cu2+/Zn2 ions from silicone elastomers loaded with divalent salts was ∼100 times greater compared to nanopowders; (iii) incorporation of gelatin, polyvinylpyrrolidone, sucrose and hydroxypropyl methylcellulose significantly enhanced the release of Cu2+/Zn2+ ions (up to ∼30-fold), and (iv) enhanced release was due to water absorption into the silicone elastomer devices, causing swelling of the devices to an extent proportional to the excipient loading.
Keywords: Drug delivery, gelatin, polyvinylpyrrolidone, hydroxypropyl methylcellulose, sucrose, silicone rubber, vaginal rings
1. Introduction
Silicone elastomers are used widely in drug delivery devices due to their biocompatibility, chemical stability, and ability to control the rate of drug release (Mashak et al. 2009; Aliyar and Schalau 2015). For example, various marketed vaginal ring products are fabricated from silicone elastomers (Estring®, Femring®, Progering®, Annovera®, and DapiRing®) (Boyd et al. 2019; Murphy et al. 2019); Jadelle® and Sini-implant (II)® (Levoplant®) are reservoir-type silicone elastomer subdermal implants for long-term administration of the contraceptive progestin levonorgestrel (Benagiano et al. 2008); and many contraceptive intrauterine devices (e.g. Mirena®, Kyleena®, and Skyla®) comprise a silicone elastomer levonorgetrel-loaded reservoir and a non-medicated silicone elastomer rate-controlling membrane to control release (Bao et al. 2018). Irrespective of the design of these marketed devices, drug release is invariably achieved bymolecular diffusion of the dissolved drug molecules in the silicone elastomer matrix (Mashak et al. 2009). Therefore, drug release rates from silicone elastomer devices are dependent upon factors such as drug solubility in the silicone elastomer, drug diffusivity in the silicone elastomer, and device surface area (Roseman and Higuchi 1970; Malcolm et al. 2003). Commercial silicone elastomer materials commonly used for drug delivery applications are based on polydimethylsiloxane, which is highly hydrophobic due to the methyl groups orientating toward the surface and creating a low surface energy. Therefore, it is not surprising that drug delivery products constructed from silicone elastomer are almost exclusively used for administration of highly hydrophobic and highly potent drugs (Ghannam et al. 1986; McBride, Boyd, et al. 2019).
Various strategies have been reported for enhancing drug release from silicone elastomers and extending release to more hydrophilic drugs (Table 1) (Hsieh et al. 1985; Carelli et al. 1989; Li and Peck 1989; Lopour et al. 1990; Dahl and Sue 1992; Di Colo 1992; Hu et al. 2000; Kajihara et al. 2000, 2001, 2003; Maeda, Brandon, et al. 2003; Maeda, Ohashi, et al. 2003; Woolfson et al. 2003; Manosroi et al. 2005; Woolfson et al. 2006; Brook et al. 2008; Mashak 2008; McBride et al. 2009; Rajendra et al. 2010; Morrow et al. 2011; Snorradóttir et al. 2011; McConville et al. 2012; Tolia and Li 2012; Forbes et al. 2014; Nemati et al. 2014; Murphy et al. 2016; Ma et al. 2018; Mazurek et al. 2018; Farahmandghavi et al. 2019; McBride, Boyd, et al. 2019; Mikolaszek et al. 2020; Mazurek, Frederiksen, et al. 2021; Mazurek, Yuusuf, et al. 2021; Fanse et al. 2024). Most commonly, these involve incorporation of hydrophilic excipients into the silicone elastomer matrix. Upon implantation or during in vitro release testing, fluid is absorbed into the silicone matrix leading to enhanced drug solubility. When sufficiently high concentrations of excipients are used, aqueous pores and channels are formed in the silicone matrix through which solvated drug molecules can more easily diffuse (Di Colo et al. 1983, 1986; Carelli et al. 1987, 1988, 1989, 1995).
Table 1.
Reported strategies for modulation of drug release from silicone elastomers.
| Strategy | Mechanism | Description | References |
|---|---|---|---|
| Incorporation of hydrophilic excipients | Introduce pores to create water channels | HPMC, sucrose, or glycine dispersed in silicone to increase bovine serum albumin (BSA) release | Carelli et al. 1989; Woolfson et al. 2006; Brook et al. 2008; Di Colo 1992; Morrow et al. 2011; Snorradóttir et al. 2011; Murphy et al. 2016; Mazurek et al. 2018; McBride, Boyd, et al. 2019; Mazurek, Frederiksen, et al. 2021; Mazurek, Yuusuf, et al. 2021 |
| Hydrophilic polymer blending | Blend silicone with polymers to enhance water uptake and drug mobility | Silicone–polyethylene glycol (PEG) copolymer matrices used to enhance the permeability of the silicone elastomer to hydrophilic and ionic species | Li and Peck 1989; Rajendra et al. 2010; Forbes et al. 2014; Mikolaszek et al. 2020 |
| Porous matrix | Introduction of pores to promote diffusion | Silicone containing PEG8000-leached pores to enhance acetaminophen and tartrazine release | Dahl and Sue 1992; Nemati et al. 2014 |
| Use of co-solvents or solubilizers | Enhancing drug mobility by reducing matrix–drug interactions and lowering diffusion activation energy | Glycerol and NaCl incorporated into silicone elastomer implants to enhance release of melatonin and estradiol | Hsieh et al. 1985; Maeda, Brandon, et al. 2003 |
| Incorporation of preformed hydrogel particles | Embedding hydrophilic hydrogel particles that swell upon hydration to form aqueous diffusion paths | Polyacrylamide hydrogel particles incorporated in silicone rubber to enhance progesterone release | Lopour et al. 1990; Mashak 2008 |
| Drug molecular complexation | Modify drug solubility and partitioning via inclusion complexation | Azelaic acid-hydroxypropyl-β-cyclodextrin complex used in silicone membrane to enhance the release of azelaic acid | Manosroi et al. 2005 |
| Surface hydrophilization | Graft hydrophilic groups on silicone surface | PEGylation of silicone elastomer to modulate triclosan release | McBride et al. 2009 |
| Drug particle size engineering | Altering particle size to modify surface area and diffusion dynamics | Larger particle size of interferon/human serum albumin (HSA) powder in silicone elastomer to increase release rate of IFN | Kajihara et al. 2000; Farahmandghavi et al. 2019 |
| Polymer microstructure engineering | Using additives/fillers to modify polymer crystallinity, viscosity, drug affinity, and diffusion paths | Silica, PEG, or silicone oil used to modify polydimethylsiloxanes for optimized levonorgestrel release | McConville et al. 2012; Tolia and Li 2012; Ma et al. 2018; Fanse et al. 2024 |
| Formulation geometry design | Control of release via physical structure to maintain a constant diffusion area | A 1/8 fractional segment core design of vaginal ring to achieve zero-order release of oxybutynin | Kajihara et al. 2001, 2003; Maeda, Ohashi, et al. 2003; Woolfson et al. 2003 |
| Stimuli-responsive additives | Triggering release via stimuli, such as pH, enzymes, or temperature | N-isopropylacrylamide hydrogel particles incorporated into a silicone rubber membrane as temperature-responsive additives that enhance release at temperature > 34 °C | Hu et al. 2000; Nemati et al. 2014 |
Metal ions and inorganic salts are used as active pharmaceutical ingredients: magnesium and aluminum salts as antacids; iron, zinc, calcium, and magnesium supplements to treat deficiencies; and copper ions for intrauterine contraception. We are interested in developing silicone elastomer vaginal rings releasing copper or zinc ions locally to the human vagina for the purpose of non-hormonal contraception (Shen et al. 2025). Mimicking the well-established mechanism of copper intrauterine devices (IUDs), divalent copper and zinc ions can adversely impact sperm motility by disrupting the function of sperm cell membranes and enzymes, such that the sperm struggle to penetrate cervical mucus (Jecht and Bernstein 1973; Holland and White 1988; Roblero et al. 1996; Huang et al. 2001). Specifically, these ions generate reactive oxygen species (ROS) which compromise membrane integrity, impairing sperm movement. They also interfere with enzymes involved in sperm metabolism and motility, such as those related to ATP production, reducing energy availability for movement. Together, these effects reduce sperm viability and motility, contributing to contraceptive efficacy. We are not seeking or anticipating systemic uptake of the copper/zinc ions; given the charge on the ions and their ability to complex with proteins or phosphate ions in the vagina, their passive diffusion across epithelial membranes will be limited. For example, copper IUDs release Cu2+ ions locally into the uterine cavity, but show minimal systemic absorption; instead, the copper acts locally, exerting a spermicidal effect and inducing an inflammatory response in the endometrium.
However, achieving sustained or controlled release of metal salts/ions from silicone elastomers is challenging given the poor solubility of metal ions in silicone. One strategy to overcome the poor solubility is to load the device with relatively large initial concentrations of the metal/metal salt. For example, incorporation of up to 25% w/w copper nanoparticles into silicone elastomer matrices produced concentration-dependent and near steady state release of copper ions after an initial burst (Xu et al. 2013). Yet, inclusion of such high active loadings in a drug product can be prohibitively expensive; for context, most marketed drug-releasing vaginal rings contain significantly <1% w/w active, e.g. NuvaRing® (0.54%), Estring® (0.02%), and DapiRing® (0.31%). An alternative and less expensive strategy would be to include a hydrophilic pharmaceutical excipient into the ring device alongside a much smaller loading of the active. In this study, four pharmaceutical excipients—gelatin, polyvinylpyrrolidone (PVP), hydroxypropyl methylcellulose (HPMC) and sucrose—were selected (Table 2) and assessed for their ability to enhance metal ion release from silicone elastomers containing copper nanoparticles (CN), zinc nanoparticles (ZN), copper sulfate pentahydrate (CSP) and zinc acetate dihydrate (ZAD). Gelatin, PVP, HPMC, and sucrose have previously been incorporated into experimental silicone elastomer devices (Acartürk and Altug 2001; Morrow et al. 2011; McKay et al. 2012; Pattani et al. 2012; Zhang et al. 2013; Du et al. 2018; McBride, Boyd, et al. 2019; McBride, Malcolm, et al. 2019; Shen et al. 2025). Here, we opted to use smaller silicone elastomer rods (40 mm long × 4 mm cross-sectional diameter) rather than full-size vaginal rings. The impact of the four hydrophilic excipients on the swelling, release rates, and mechanical properties of rods were assessed.
Table 2.
Molecular weight, source, chemical structural formula, and major applications in pharmaceutics of gelatin, PVP, HPMC, and sucrose in pharmaceutics.
| Excipients | Molecular weight (g/mol) | Source | Chemical structural formula | Pharmaceutical application | References |
|---|---|---|---|---|---|
| HPMC | 50,000–1,250,000 | Semisynthetic |
|
Extended release-matrix former; tablet binder; tablet film coating | Rowe 1977; Chowhan 1980; Shah et al. 1989 |
| PVP | 10,000 (average) | Semisynthetic |
|
Binder of wet-granulation in tableting; solubilizer in oral and parenteral formulations; coating agents or binders of sugar beads | Iwata and Ueda 1996; Stubberud et al. 1996; Becker et al. 1997 |
| Gelatin | >15,000 | Porcine skin |
|
Biodegradable matrix material in an implantable delivery system and capsules | Kommareddy et al. 2003; Mikhailov 2023 |
| Sucrose | 342 | Plants |
|
Used as binder in wet-granulation; bulking agent; sweetener in chewable tablets and lozenges; tablet-coating agent | Mullarney et al. 2003; Sugimoto et al. 2006 |
2. Materials and methods
2.1. Materials
An unrestricted drug delivery grade addition-cure silicone elastomer (DDU-4320, Parts A and B) was supplied by NuSil Silicone Technology Inc. (Carpinteria, CA, USA). Copper nanopowder (65 nm mean particle size; 99.95%; CN) (Nanografi, Ankara, Turkey) and copper sulfate pentahydrate (≥98.0%; CSP) (Sigma-Aldrich, St. Louis, MO, USA) were selected as sources of Cu2+ ions; oxidation/corrosion of copper nanopowder produces cupric ions. Zinc nanopowder (790 nm, 99.95%; Nanografi; 790ZN) and zinc acetate dihydrate powder (Merck KGaA, Darmstadt, Germany; ZAD) were selected as a source of Zn2+ ions. Gelatin (from porcine skin), sucrose (>99.5%), and PVP (MW 10,000 g/mol) were purchased from Sigma-Aldrich. HPMC (MW 1261.45 g/mol) was purchased from Alfa Aesar (MA, USA). HPMC and PVP were supplied in micronized form. CSP and ZAD powders were manually milled using a pestle and mortar and sieved using a 100 μm sieve. Polyvinylchloride (PVC) tubing having 4 mm inner CSD was purchased from RS Components Ltd. (Northants, UK). Deionized water was obtained using a Millipore Direct-Q 3 UV Ultrapure Water System (Watford, UK).
2.2. Manufacture of silicone elastomer rods
Thirty-six different DDU-4320 matrix-type rod formulations—containing 0 or 5% w/w CN, ZN, CSP, or ZAD combined with 0, 15, or 30% w/w gelatin, PVP, HPMC, and sucrose (Table S1)—were manufactured using a laboratory-scale injection molding machine. Briefly, Part A premixes (10 g) were prepared by accurately weighing the required quantities of actives, excipients, and silicone into screw-cap polypropylene containers and mixing at 3000 rpm for 30 s (DAC-150 FVZ-K SpeedMixer™, Hauschild, Germany) followed by manual mixing for 1 min. Part B premixes (10 g) were similarly prepared. Part A and Part B premixes were combined in a 1:1 ratio, manually mixed for 1 min, and speedmixed (30 s, 3000 rpm). The mixtures were transferred into a cartridge system, injected into the PVC tubing, and cured at 50 °C for 20 h. The silicone elastomer rods were removed from the PVC tubing and cut into 40 mm lengths. Formulation codes (Table 3) adhere to the following convention: the first part of code denotes the API (fixed at 5% w/w loading); the second part of the code denotes the percentage loading of the excipient (15 or 30%) followed by the first letter of the excipient (G for gelatin; P for PVP; H for HPMC and S for sucrose). For example, CSP-15H rods contain 5% w/w CSP and 15% w/w HPMC. B-B represents a rod containing no drug and no excipient (B = blank), while codes that include an API assignment and followed by a B contain active but no excipient.
Table 3.
Summary of in vitro Cu2+ ion and Zn2+ ion release data for matrix-type silicone elastomer rods containing metal nanopowders + excipients (mean ± SD, n = 3).
| Rods ID | Initial Cu2+/Zn loading (µg) | Day 1 release (µg) | Day 8 release (µg) | Day 11 release (µg) | Release percent over 14 days (%) | Higuchi model |
|
|---|---|---|---|---|---|---|---|
| Release rate (µg/day1/2) | R 2 | ||||||
| CN-B | 28,540 ± 40 | 3.51 ± 1.00 | 0.33 ± 0.47 | 0.00 ± 0.00 | 0.06 | 4.96 | 0.97 |
| CN-15G | 29,417 ± 612 | 7.86 ± 2.07 | 2.19 ± 1.08 | 0.25 ± 0.19 | 0.17 | 14.95 | 0.96 |
| CN-30G | 31,108 ± 129 | 4.96 ± 1.06 | 6.19 ± 1.35 | 1.05 ± 0.55 | 0.28 | 30.24 | 0.99 |
| CN-15P | 27,940 ± 126 | 3.80 ± 0.74 | 1.12 ± 0.59 | 0.23 ± 0.17 | 0.11 | 9.97 | 0.99 |
| CN-30P | 28,135 ± 412 | 3.09 ± 0.99 | 3.87 ± 0.82 | 0.44 ± 0.07 | 0.15 | 14.56 | 1.00 |
| CN-15H | 28,435 ± 44 | 1.20 ± 0.22 | 1.98 ± 0.41 | 0.13 ± 0.18 | 0.08 | 7.21 | 0.99 |
| CN-30H | 28,835 ± 620 | 1.08 ± 0.18 | 2.51 ± 0.20 | 0.51 ± 0.36 | 0.09 | 8.41 | 0.92 |
| CN-15S | 30,353 ± 355 | 3.76 ± 0.30 | 1.15 ± 0.31 | 0.43 ± 0.12 | 0.08 | 7.14 | 0.98 |
| CN-30S | 31,271 ± 455 | 2.64 ± 0.70 | 1.70 ± 0.76 | 0.61 ± 0.08 | 0.11 | 10.48 | 0.98 |
| ZN-B | 28,095 ± 31 | 0.26 ± 0.37 | 0.00 ± 0.00 | 0.00 ± 0.00 | 0.00 | 0.33 | 0.79 |
| ZN-15G | 28,985 ± 212 | 1.39 ± 0.36 | 0.65 ± 0.08 | 0.22 ± 0.31 | 0.01 | 0.94 | 0.91 |
| ZN-30G | 30,472 ± 406 | 4.36 ± 0.41 | 1.04 ± 0.53 | 0.08 ± 0.11 | 0.03 | 2.23 | 0.94 |
| ZN-15P | 27,778 ± 327 | 5.50 ± 1.37 | 0.19 ± 0.27 | 0.00 ± 0.00 | 0.03 | 0.81 | 0.95 |
| ZN-30P | 57,463 ± 391 | 10.57 ± 1.05 | 2.00 ± 0.33 | 0.70 ± 0.38 | 0.05 | 7.05 | 0.99 |
| ZN-15H | 28,355 ± 252 | 1.29 ± 0.36 | 0.46 ± 0.33 | 0.07 ± 0.10 | 0.01 | 0.86 | 0.93 |
| ZN-30H | 28,243 ± 141 | 3.98 ± 0.54 | 0.91 ± 0.20 | 1.70 ± 1.22 | 0.06 | 4.00 | 0.77 |
| ZN-15S | 29,860 ± 730 | 1.31 ± 0.72 | 0.33 ± 0.46 | 0.10 ± 0.14 | 0.01 | 0.72 | 0.90 |
| ZN-30S | 32,200 ± 514 | 2.74 ± 1.04 | 1.68 ± 0.39 | 0.41 ± 0.38 | 0.05 | 4.76 | 0.93 |
2.3. Swelling testing
Swelling of rods due to absorption of release medium was assessed in parallel with in vitro release testing. On Days 0, 7, 14, 21, 28, and 30 of the release experiment, rods were removed from the tubes (see Section 2.4), and rod dimensions (vernier calipers) and weights (4 dp balance) were measured. Swelling percentages in terms of weight, length, and cross-sectional diameter (CSD) were calculated according to Equations (1)–(3), respectively. For nanopowder formulations, swelling measurements were continued until Day 14.
| (1) |
| (2) |
| (3) |
where Wt = weight at time t, W0 = initial weight, CSDt = CSD at time t, CSD0 = initial CSD, Lt = length at time t, and L0 = initial length.
2.4. In vitro release testing
Three rod samples were randomly selected from each formulation batch for release testing. On day 0, individual rods were placed into 13 mL plastic tubes (Sarstedt, Wexford, Ireland) containing 10 mL deionized water (pH 5.2) and stored in a shaking incubator at 37 °C and 60 rpm. The release medium was sampled each weekday and then completely replaced with fresh 10 mL deionized water. No sampling was performed during weekends. Following completion of the release experiment, the rods were placed in a 40 °C oven for 7 days to dry.
2.5. Mechanical testing
2.5.1. Shore M hardness testing
The Shore M hardness testing of ring formulations was tested according to the material Certificate of Analysis using ASTM D2240 (Type M scale) (‘ASTM D2240-15, Standard Test Method for Rubber Property—Durometer Hardness,’ 2017). For accuracy, the silicone elastomer test materials were required to be flat and of sufficient size (width > 2.5 mm) to ensure edge effects could not influence the measurement. A custom slab mold (dimensions: 54 × 4 mm) was designed and manufactured to enable reproducible sample production (Figure 1(A)). Although rods possess a curved geometry that deviates from ASTM D2240 standard conditions, the Shore M hardness gauge is suitable for such geometries and is frequently employed as an investigational tool in comparative analysis of elastomeric devices (Boyd et al. 2020; Dallal Bashi et al. 2021; Shen et al. 2025).
Figure 1.
(A) Alignment tools (upper) and 54 × 4 mm holder (lower) for Shore M hardness testing; (B) representative photograph showing a rod placed on the flat for Shore M hardness testing; (C) custom grips for tensile testing of silicone elastomer rods; (D) representative photograph showing a rod placed in the grips for tensile testing.
Rod formulations (after manufacture—D0, post release—D14/D30, and after drying—dried) were tested for Shore hardness using a custom holder (Figure 1(A)) and a Shore M Hardness Tester™ (Checkline RX-DD-M digital durometer Type M) held in an OS-3 stand™ (Checkline Europe, Birmingham, UK; Figure 1(B)) at room temperature (∼20 °C). Before testing, alignment tools (Figure 1(A)) were used to ensure the probe contacted the highest point of the rod sample. Standard test blocks bracketing the hardness range of rings were tested before and after measurement of samples, and the measurements recorded were within the hardness range of the standards, except the hardness value of very few samples dropped <20 after releasing. Four individual measurements were recorded per rod for D0, D14/30, and dried samples. Each hardness value was recorded 5 s after the probe contacted the sample surface.
2.5.2. Tensile testing
Tensile testing was performed on the 36 rod formulations using an Instron Universal Testing System (Instron Model 5564, UK) in tension mode fitted with a 5000 N load cell and Merlin software control. At D0, six rods were randomly selected for each formulation—three were tested immediately, and three were first released into deionized water over 14/30 days and then dried at 40 °C for 7 days before being tested. Rods were placed into the upper and lower grips (Figures 1(C,D)). The initial gauge length was set to 10 mm and then automatically adjusted for each rod to ensure sufficient initial tension. Each rod was stretched using a grip separation speed of 1 N/s until reaching a force of 4 N, and then the speed was changed to 200 mm/min, and the rod stretched until fracture. Elongation (mm) to 4 N (E4N), elongation to break (ELB), and the maximum load (N) at maximum tensile extension were recorded for all rods at D0. The E4N was recorded for all formulations both at D0 and after release/drying. E4N and ELB percentage for individual rod were determined using Equation (4).
| (4) |
2.6. Quantification of copper/zinc ion release
Release of copper and zinc ions was measured using a Shimadzu AA-6300 flame atomic absorption spectrophotometer (Shimadzu, Tokyo, Japan) with a graphite furnace atomizer and an autosampler (ASC-6100, Shimadzu, Tokyo, Japan). Stock solutions of Cu2+ or Zn2+ ions (500 μg/mL) were prepared by dissolving CSP or ZAD powders in deionized water. Working standard solutions were prepared by dilution of the stock solution with deionized water. Standard and sample solutions were aspirated and absorbance values recorded at 324.8 and 213.9 nm for Cu and Zn, respectively. A copper/zinc hollow-cathode lamp (Mason Technology, Dublin, Ireland) was used as the radiation source and air-acetylene was the carrier gas.
2.7. Statistical analysis
Weights and mechanical properties post-manufacture were compared: B-B versus each API-B rod formulation; CN-B versus each CN rod containing excipient; ZN-B versus each ZN rod containing excipient; CSP-B versus each CSP rod containing excipient; and ZAD-B versus each ZAD rod containing excipient. Weight, CSD, OD, and Shore M hardness values of each rod formulation before IVRT were compared with post IVRT. Additionally, the Shore M and E4N values of the dried formulations were compared with those from the initial manufacturing process. Statistical comparisons were performed using one-way ANOVA analysis by GraphPad Prism software. Significant differences are represented in the figures: * = significant (0.01 < p < 0.05), ** = very significant (0.001 < p < 0.01), *** = extremely significant (0.0001 < p < 0.001), **** = extremely significant (p < 0.0001).
3. Results and discussion
3.1. Visual and physical properties of rods
Representative photographs of rod are shown in Figure 2. Figures S1–S4 provide a more extensive gallery of photographs showing all rods after manufacture, after release, and after drying. All rods were uniform in color, smooth, and free of air bubbles/pockets. The color of the rods reflected the color of the active substances, with some variation in color due to active release and absorption/removal of water.
Figure 2.
Representative photographs of (A–E) excipient-free, (F–I) CSP with 30% w/w excipients (CSP-30H failed to manufacture, here replaced with CSP-15H), and (J–M) ZAD with 30% w/w excipients rods initially after manufacture.
Rod weights ranged from 545 to 640 mg due to differences in the types and loadings of both actives and excipients (Figure 3, Table S1). Initial rod dimensions were ∼40 × 4 mm (Table S1). Mean weight of the control B-B rod was 561 mg. The incorporation of 5% w/w ZAD significantly decreased the rod weight, while incorporation of 5% w/w CN, ZN, and CSP showed no difference (Figure 3(A)). The incorporation of 15% w/w or 30% w/w PVP had no significant influence on the weight of any type of rods (Figures 3(B–E)), while inclusion of 15 or 30% w/w sucrose or 30% gelatin significantly increased the rod weights. CSP-30H could not be manufactured successfully, likely due to the high loading of HPMC inhibiting the silicone curing reaction (Malcolm et al. 2016).
Figure 3.
Mean weight for matrix-type (A) excipient-free; (B) CN-excipient; (C) ZN-excipient; (D) CSP-excipient; and (E) ZAD-excipient matrix-type rods post-manufacture. Error bars present the standard deviation of six replicates. (B–E) share the same legend as (C). Dash lines in (B–E) represent the mean weight of rods containing only API. Asterisks (*) above the bars in profile (A) represent the significant difference levels in weight compared with the B-B rod; asterisks (*) above the bars in (B–E) represent the significant difference levels in weight compared with the corresponding excipient-free rods. The absence of an asterisk indicates no significant difference.
3.2. Swelling test
The incorporation of high concentrations of hydrophilic actives and excipients has the potential to cause swelling of silicone elastomer devices due to water/fluid absorption, which is not ideal. For example, swelling of a vaginal ring device could result in involuntary expulsion and/or user discomfort (Efentakis et al. 1997; Johnson et al. 2010; Malcolm et al. 2016; Undie et al. 2020; McLellan-Lemal et al. 2022). The weights, CSDs, and lengths of B-B, CN-B, and ZN-B formulations were largely similar (percentage increases < 1%) before and after in vitro release testing, since the silicone elastomer and copper/zinc nanopowders do not absorb water (Figures 4 and 5(A–C)). However, copper/zinc nanopowders are highly reactive and prone to surface oxidation when exposed to moisture (Delalu et al. 2000; Amelkovich et al. 2015). All other formulations—containing various loadings of copper sulfate, zinc acetate, and hydrophilic excipients—showed swelling, and especially with incorporation of 15 or 30% w/w sucrose (Figures 4 and 5). The mean weight, cross-sectional diameter (CSD), and length of CN-30S rods increased 294, 68, and 68%, respectively, and CN-15S rods 89, 26, and 26%, respectively (Figures 4(A–C)). For the other CN + excipient formulations, percentage weight increases were in the range 21–57%, CSD between 7–19%, and length between 6–15%. Similar values and trends were noted for ZN formulations (Figures 4(D–F)). Among the four excipients, nanopowder formulations with 15% gelatin or 15% HPMC showed least swelling, while sucrose formulations showed greatest swelling. With the exception of HPMC, swelling generally increased with higher excipient loadings and longer immersion times. However, for the CN-30H formulation, the rod matrix reached its maximum extent of swelling by Day 7, but then subsequently shrank. This was also observed for ZAD-30H.
Figure 4.
Percentage increases in weight (left), CSD (middle), and length (right) for matrix-type DDU-4320 (A–C) CN-loaded and (D–F) ZN-loaded rods over 14 days. For ease of reading, only upper error bars representing standard deviations of three replicates are shown.
Figure 5.
Percentage increases in weight (left), CSD (middle), and length (right) for matrix-type DDU-4320 (A–C) blank and CSP-loaded, and (D–F) ZAD-loaded rods over 30 days. For ease of reading, only upper error bars representing standard deviations of three replicates are shown.
The swelling behavior of CSP rod formulations is summarized in Figures 5(A–C). After 30 days, CSP-30S increased significantly in weight, CSD, and length (280, 70, and 193%, respectively), and CSP-15S 186, 54, and 48%, respectively. CSP-B increased ∼27% in weight over 30 days, with minimal changes in dimensions. For all other formulations, weight increased 66–95%, CSD 22–31% and length 16–26%. Similar trends were observed for ZAD formulations (Figures 5(D–F)). Overall, for CSP rods with 15% excipient loadings, rank order of swelling was sucrose > PVP > HPMC > gelatin. For ZAD rods with 30% excipients, swelling rank order was sucrose > gelatin > PVP > HPMC. For CSP and ZAD rods loaded with gelatin and sucrose, the extent of swelling increased with loading. For CSP/ZAD combined with PVP or HPMC, swelling values were similar for both 15 and 30% loadings. In conclusion, among the four excipients, 15% w/w gelatin and 30% w/w PVP showed the least extent of swelling for CSP formulations, while 15% w/w gelatin and 30% w/w HPMC showed the least extent of swelling for ZAD formulations. In the context of silicone elastomer vaginal rings (Boyd et al. 2019), the 10 mL release medium used here was significantly larger than typical vaginal fluid volumes in healthy women (0.5–2 mL) (Owen and Katz 1999), and the rod dimensions were also significantly smaller than normal ring sizes. Without in vivo testing, it is difficult to extrapolate these data to predict the extent of swelling of similarly formulated vaginal ring devices.
3.3. In vitro release of copper and zinc ions
Conventional organic small molecule drug substances—which are usually crystalline and micronized—are dispersed uniformly throughout silicone matrix-type drug delivery devices. Upon placement of the device into the release medium, drug on the device surface dissolves in the release medium causing an initial burst release (Malcolm et al. 2003). However, in this study, results from in vitro release testing of CN and ZN formulations showed no burst release; release was very low (Figure 6 and Table 3) and often below the LOQ. Among CN formulations (Figures 6(A–C)), CN-30G released the highest daily quantity of Cu2+ ions (∼13 µg) on Day 3 and was the only CN formulation able to continually release ions over the two-week test period. For the CN-B formulation, daily Cu2+ release was highest (∼3.9 µg) on Day 14 (over weekend), and release was only quantifiable for seven days during the test. However, coformulation of CN with all four excipients significantly increased the quantity and extended the duration of Cu2+ release. For ZN-loaded rods, ZN-30P showed the highest daily Zn2+ release (∼11 µg) on Day 1 (Figures 6(D–F)); by comparison, the ZN-B formulation maximally released 0.5 µg Zn ions release on Day 9, with no release measured on other days. ZN-30P sustained release of Zn2+ over two weeks, while ZN-15G, ZN-15P, and ZN-15S released Zn ions <5 days even with incorporation of excipients.
Figure 6.
Daily and cumulative release (µg) versus time profiles for release of Cu2+ (above) and Zn2+ (below) ions from matrix type of DDU-4320 (A–C) CN-loaded and (D–F) ZN-loaded rods into deionized water over 14 days, respectively. Upper error bars represent standard deviation of three replicates. (A,B) share the same legend as (C); while (D,E) share the same legend as (F).
Compared with CN/ZN formulations, release of Cu2+ and Zn2+ was significantly increased for rods containing the metal salts CSP and ZAD (Figure 7 and Table 4). Except for CSP-30S, burst release was observed for all formulations. On Day 1, Cu2+ and Zn2+ release ranged from 297–744 µg and 146–1210 µg, respectively, followed by declining release on subsequent days. CSP-30S showed an irregular daily release profile, likely due to swelling of the device, with maximum Cu2+ release on Day 8 (880 µg). CSP-30G and ZAD-30H formulations showed the highest Day 1 release—744 µg Cu2+ and 1210 µg Zn2+ ions, respectively. There were three days when CSP-B released Cu ions quantities less than the LOQ, while CSP-30G released 58 µg Cu2+ on Day 30. ZAD-30H released 17 µg Zn2+ on Day 30, compared to 0.5 µg Zn2+ with ZAD-B. For CSP formulations having 15% w/w excipient loadings, CSP-15H provided the highest Cu2+ release (534 µg Day 1, 20 µg Day 30).
Figure 7.
Daily and cumulative release (µg) versus time profiles for release of Cu2+ (above) and Zn2+ (below) ions from matrix type of DDU-4320 (A–C) CSP-loaded and (D–F) ZAD-loaded rods into deionized water over 30 days, respectively. Upper error bars represent standard deviation of three replicates. (A,B) share the same legend as (C); (D,E) share the same legend as (F).
Table 4.
Summary of in vitro Cu2+ and Zn2+ ion release data for matrix-type silicone elastomer rods containing metal salts + excipients (mean ± SD, n = 3).
| Rods ID | Initial Cu2+/Zn loading (µg) | Day 1 release (µg) | Day 15 release (µg) | Day 30 release (µg) | Release percent over 30 days (%) | Higuchi model |
|
|---|---|---|---|---|---|---|---|
| Release rate (µg/day1/2) | R 2 | ||||||
| CSP-B | 7249 ± 28 | 326 ± 25 | 1.7 ± 0.6 | 0.6 ± 0.9 | 7.9 | 43.16 | 0.83 |
| CSP-15G | 7226 ± 18 | 409 ± 56 | 6.4 ± 0.6 | 5.5 ± 0.7 | 12.6 | 99.38 | 0.96 |
| CSP-30G | 7564 ± 59 | 744 ± 209 | 39.6 ± 11.5 | 58.1 ± 22.7 | 47.3 | 611.2 | 1.00 |
| CSP-15P | 7187 ± 46 | 308 ± 29 | 23.5 ± 8.4 | 11.8 ± 3.8 | 14.1 | 148.0 | 1.00 |
| CSP-30P | 7102 ± 73 | 718 ± 94 | 218.7 ± 100.5 | 38.4 ± 4.3 | 77.2 | 1397 | 1.00 |
| CSP-15H | 7194 ± 116 | 534 ± 105 | 25.2 ± 7.7 | 20.3 ± 8.0 | 22.4 | 236.1 | 1.00 |
| CSP-15S | 7429 ± 34 | 297 ± 43 | 11.9 ± 3.0 | 4.8 ± 2.2 | 10.1 | 90.17 | 0.96 |
| CSP-30S | 8023 ± 145 | 416 ± 63 | 432.8 ± 117.6 | 29.4 ± 19.4 | 90.1 | 1590 | 0.91 |
| ZAD-B | 8207 ± 88 | 147 ± 21 | 3.4 ± 0.2 | 0.5 ± 0.6 | 3.2 | 22.82 | 0.86 |
| ZAD-15G | 8727 ± 41 | 289 ± 22 | 4.9 ± 0.9 | 0.8 ± 0.3 | 6.1 | 47.80 | 0.88 |
| ZAD-30G | 8800 ± 107 | 612 ± 122 | 64.2 ± 43.0 | 9.6 ± 5.5 | 21.8 | 302.0 | 0.97 |
| ZAD-15P | 8166 ± 69 | 250 ± 30 | 11.4 ± 1.5 | 3.3 ± 1.9 | 6.2 | 54.19 | 0.96 |
| ZAD-30P | 7966 ± 36 | 633 ± 190 | 44.2 ± 4.6 | 12.2 ± 0.5 | 31.8 | 400.9 | 0.91 |
| ZAD-15H | 8528 ± 110 | 310 ± 36 | 12.9 ± 4.1 | 3.2 ± 1.1 | 8.5 | 86.66 | 0.96 |
| ZAD-30H | 8579 ± 122 | 1210 ± 151 | 40.8 ± 21.9 | 17.2 ± 0.8 | 58.1 | 1102 | 0.94 |
| ZAD-15S | 8553 ± 74 |
|
3.6 ± 0.8 | 1.4 ± 0.6 | 3.4 | 29.42 | 0.89 |
| ZAD-30S | 9085 ± 237 | 323 ± 4 | 25.9 ± 1.7 | 29.0 ± 1.1 | 12.2 | 166.3 | 1.00 |
Historically, drug release from simple silicone elastomer matrix-type devices containing hydrophobic small-molecule drugs has been modeled using the Higuchi equation (Equation 5), with certain important assumptions and criteria—(i) drug is uniformly dispersed throughout the matrix, (ii) initial drug concentration in the device is much higher than drug solubility in the polymer matrix (A ≫ CSIL), (iii) drug diffusion occurs in one dimension, (iv) drug diffusion is the rate-limiting step, (v) the drug diffusion coefficient is constant, and (vi) sink conditions are maintained (Higuchi 1961). For most rod formulations tested in this study, criteria (iv) and (v) do not hold since the devices swell due to water uptake during in vitro release testing. This is confirmed by coefficient of determination (R2) values for linear regression analysis of cumulative release vs root time (t½) plots for up to 60% of drug release (Figure S5) that often significantly deviate from 1.0 (Tables 3 and 4). For the purposes of this study, we designate R2 ≥ 0.99 as good fit, 0.99 > R2 ≥ 0.97 as moderate fit, and R2 < 0.97 as poor fit. We understood that high R2 values might be achieved while still violating Higuchi model assumptions. Nonetheless, the release rates (µg/day½) derived from application of the Higuchi model are useful for comparing formulations (Tables 3 and 4).
| (5) |
where Q is the cumulative amount of drug released per unit area of device (mg/cm2), A is the drug loading (mg/cm3), CSIL is the solubility of the drug in silicone elastomer (mg/cm3), DSIL is the diffusion coefficient of the drug in the elastomer (cm2/day), and t is time (days).
Cumulative release of Cu2+ from CN formulations ranged from 18 to 87 µg and followed the rank order CN-30G > CN-15G > CN-30P > CN-30S > CN-15P > CN-30H > CN-15S > CN-15H > CN-B (Figure 6(C)). Cumulative Cu2+ release and the release rate for CN-30G (87 µg, 30.24 µg/day½, respectively) were five times greater than CN-B (18 µg, 4.96 µg/day½, respectively). For the Zn nanopowder formulations, rank order of cumulative Zn2+ release was ZN-30P (∼30 µg) > ZN-30H > ZN-30S > ZN-30G > ZN-15P > ZN-15G/ZN-15H/ZN-15S > ZN-B (Figure 6(F)). Zn2+ cumulative release for ZN-30P was 30 times that of ZN-B (∼1 µg). As for Cu2+, Zn2+ release rates were significantly enhanced by incorporating 15% or 30% gelatin, PVP, HPMC, and sucrose. 30% w/w PVP provided the largest enhancement (20-fold).
Over the 30-day period, cumulative release of Cu2+ ions from CSP formulations ranged from 570 to 7230 µg (Figure 7(C)); rank order was CSP-30S > CSP-30P > CSP-30G > CSP-15H > CSP-15P > CSP-15G > CSP-15S > CSP-B. Cumulative Cu2+ release from CSP-30S was more than 12-fold greater than the control CSP-B; further, ∼90% of the original Cu2+ loading was released. Release rate of CSP-30S (1590 µg/day1/2) was more than 36 times that for CSP-B (43 µg/day½) (Table 4). For formulations containing 15% w/w excipient loading, HPMC showed the greatest enhancement of Cu2+ release (1614 µg 30 days; 22% w/w released; release rate 236 µg/day½). The cumulative release of Zn2+ from ZAD formulations ranged from 264–4987 µg, with rank order ZAD-30H > ZAD-30P > ZAD-30G > ZAD-30S > ZAD-15H > ZAD-15G > ZAD-15P > ZAD-15S > ZAD-B (Figure 7(F)). Cumulative Zn2+ release from ZAD-30H (58% Zn2+ released, 1102 µg/day½) was significantly greater than for ZAD-B (3% Zn2+ released, 27 µg/day1/2). With the inclusion of excipients, all metal salts formulations were able to maintain the release of Cu/Zn ions over one month.
Although deionized water was used as the release medium in this study, we acknowledge that it does not fully mimic the physiological environment of the vagina, which typically contains 0.5–2 mL of acidic fluid with pH ∼4.2 and moderate ionic strength (Owen and Katz 1999). The use of 10 mL deionized water was a deliberate simplification, selected to maintain sink. This approach is commonly adopted in early-stage screening studies to facilitate comparative evaluation of excipient performance (Boyd et al. 2019). We anticipate that under physiological conditions, the release of metal ions is expected to be slower than observed in vitro. This is primarily due to the limited fluid volume in the vaginal environment, which reduces water uptake into the matrix and slows the dissolution of metal salts. In addition, the presence of biological components (such as ions and proteins) may lead to complexation with Cu2+ and Zn2+, which can further delay the diffusion of free metal ions. Together, these factors suggest that in vivo, the formulations may exhibit more gradual and sustained ion release, which could be beneficial for prolonged local spermicidal activity while minimizing systemic exposure. The formulation screening conducted here provides an important foundation for subsequent in vitro or in vivo studies under more physiologically relevant conditions, such as simulated vaginal fluid or human cervicovaginal secretions.
3.3.1. Mathematical modeling of in vitro release data
To better accommodate device swelling, in vitro release data was also modeled using the Korsmeyer-Peppas equation (Equation 6) (Peppas 1985). For cylindrical devices, a release exponent value (n) of 0.45 is indicative of drug release that obeys Fickian diffusion, for which the rate of drug diffusion is much smaller than the relaxation time of the polymer chains (Crank 1975; Fu and Kao 2010; Talevi and Ruiz 2021). When using the equation to analyze drug release from porous systems, the value of n may be <0.45. This is referred to as quasi-Fickian diffusion and occurs when additional mechanisms influence the release of actives, such as diffusion through a swollen matrix comprising water-filled pores (as is likely occurring here). n values between 0.45 and 0.89 indicate deviation from Fickian diffusion, are classified as anomalous (non-Fickian), and are common when both swelling and diffusional process influence drug release. According to this model, n values equal to 0.89 are classified as case-II transport (zero-order release), for which the diffusional processes are much faster than the polymer relaxation process, and drug release is mainly controlled by swelling. Super case II transport—where n > 0.89—is rarely observed; it is indicative of accelerated release of the dissolved drug from the matrix (Bruschi 2015).
| (6) |
where Mt/M∞ is a fraction of drug released at time t, K is the release rate constant, and n is the release exponent.
Graphs of log fraction Cu2+/Zn2+ release versus log time from the various rods are presented in Figure 8. Exponent values for CN formulations (Table 5) were larger than 0.45 (anomalous transport), while most ZN, CSP, and ZAD formulations had n < 0.45 (quasi-Fickian diffusion). Release for ZN-B could not be modeled using the Korsmeyer-Peppas equation as cumulative release was too low. For copper sulfate and zinc acetate formulations (Table 5), CSP-30G showed Fickian release (n = 0.45, release proportional to t½). CSP-30P and ZAD-30H showed anomalous (non-Fickian) release with n values between 0.45 and 0.89. CSP-30S was the only metal salt formulation that released according to a super-case II transport mechanism. For other metal salts formulations, quasi-Fickian drug release was observed (n < 0.45).
Figure 8.
Graphs showing log fraction Cu2+/Zn2+ release versus log t (days) for matrix-type rods containing (A) CN, (B) ZN, (C) CSP, and (D) ZAD.
Table 5.
Analysis of Cu2+ and Zn2+ ion release from matrix-type rods with Korsmeyer-Peppas model (limited to the first 60% of drug release).
| Rod ID | n | Drug transport mechanism | R 2 | Rod ID | n | Drug transport mechanism | R 2 |
|---|---|---|---|---|---|---|---|
| CN-B | 0.583 | Anomalous | 0.980 | CSP-B | 0.144 | Quasi-Fickian | 0.928 |
| CN-15G | 0.665 | Anomalous | 0.934 | CSP-15G | 0.221 | Quasi-Fickian | 0.992 |
| CN-30G | 1.012 | Super case-II | 0.942 | CSP-30G | 0.451 | Fickian | 0.999 |
| CN-15P | 0.772 | Anomalous | 0.968 | CSP-15P | 0.328 | Quasi-Fickian | 0.994 |
| CN-30P | 0.948 | Super case-II | 0.951 | CSP-30P | 0.705 | Anomalous | 0.998 |
| CN-15H | 1.086 | Super case-II | 0.983 | CSP-15H | 0.328 | Quasi-Fickian | 0.995 |
| CN-30H | 1.160 | Super case-II | 0.986 | CSP-15S | 0.251 | Quasi-Fickian | 0.988 |
| CN-15S | 0.665 | Anomalous | 0.970 | CSP-30S | 0.993 | Super case-II | 0.923 |
| CN-30S | 0.879 | Anomalous | 0.964 | ZAD-B | 0.165 | Quasi-Fickian | 0.953 |
| ZN-B | N/A | N/A | N/A | ZAD-15G | 0.173 | Quasi-Fickian | 0.965 |
| ZN-15G | 0.407 | Quasi-Fickian | 0.895 | ZAD-30G | 0.328 | Quasi-Fickian | 0.989 |
| ZN-30G | 0.314 | Quasi-Fickian | 0.898 | ZAD-15P | 0.209 | Quasi-Fickian | 0.995 |
| ZN-15P | 0.139 | Quasi-Fickian | 0.961 | ZAD-30P | 0.404 | Quasi-Fickian | 0.954 |
| ZN-30P | 0.400 | Quasi-Fickian | 0.978 | ZAD-15H | 0.243 | Quasi-Fickian | 0.993 |
| ZN-15H | 0.397 | Quasi-Fickian | 0.923 | ZAD-30H | 0.473 | Anomalous | 0.968 |
| ZN-30H | 0.469 | Anomalous | 0.870 | ZAD-15S | 0.200 | Quasi-Fickian | 0.962 |
| ZN-15S | 0.387 | Quasi-Fickian | 0.899 | ZAD-30S | 0.354 | Quasi-Fickian | 0.990 |
| ZN-30S | 0.706 | Anomalous | 0.948 |
n and R2 values are rounded to 3 dp.
The limited release of Cu2+/Zn2+ from nanopowder formulations is a consequence of their poor solubility and diffusion within the hydrophobic silicone matrix (Malcolm et al. 2003) Incorporating hydrophilic excipients, such as gelatin, PVP, HPMC, and sucrose into the silicone elastomer promotes the formation of aqueous pores, channels, and cracks when immersed in the release medium. As these excipients swell and leach out, they progressively increase fluid penetration and expose the embedded nanopowder and salts, thereby enhancing ion diffusion and release. This observation is consistent with the biphasic release mechanism first described by Colo in 1992 (Figure 9) (Di Colo 1992). According to this model, the first stage involves a diffusion-controlled process of Cu2+/Zn2+ ions fully surrounded by the pore fluid and then released; while the second phase involves a combined dissolution–diffusion-controlled mechanism of Cu2+/Zn2+ ions embedded within or partially enclosed by the silicone elastomer. To validate this biphasic behavior, scanning electron microscopy was performed on the CSP-30S formulation, which achieved over 90% release of CSP. Cross-sectional images (Figure S6, Supplementary Material) revealed a porous matrix devoid of residual particles after drying post-release testing, confirming the structural changes in the matrix between before- and post-release.
Figure 9.
Release of copper/zinc ions (whether incorporated as a salt or a nanoparticle) from a silicone elastomer matrix containing a hydrophilic excipient is facilitated by fluid update into the matrix (A) and the subsequent formation of aqueous channels throughout the matrix. This figure has been adapted from a previously published figure (Di Colo 1992).
3.4. Mechanical testing
3.4.1. Shore M hardness test
Since many of the rod formulations swelled when immersed in water/fluid, it was deemed important to assess changes in their hardness. For reference, the Shore M hardness values for four commercial vaginal rings—Annovera®, Progering®, Estring®, and NuvaRing®—have previously been reported as 30, 42, 51, and 87, respectively (Shen et al. 2025). The Shore M hardness values of the various rod formulations—measured soon after manufacture (D0), immediately after completion of in vitro release testing (D14/D30), and after drying post-release (Dried)—ranged between 2.6 and 60.1 (Figure 10). While hardness values for CN, ZN, CSP, and ZAD formulations containing no excipients fell within a relatively narrow range (29.7–41.7; indicating only minor changes in hardness pre/during/post testing), formulations containing excipients invariably showed a V-shaped trend in hardness, with values relatively high post-manufacture, decreasing significantly following measurement immediately after completion of in vitro release testing, and then fully or partially recovering with subsequent oven drying to evaporate the absorbed water (Figures 10(B–E)). Immediately post-release (D14/D30; before drying), only certain CN or ZN formulations—specifically, CN-B, ZN-B, CSP-B, CN-15G, ZN-15G, CN-15H, and ZN-15H—fell within the hardness range of commercial rings; for all other excipient-loaded rod formulations, water uptake and swelling reduced hardness values below 30. Several notable trends were observed—(i) for D0 samples, hardness values increased with increasing excipient loading (0 < 15 < 30%), presumably due to a mechanical filler effect; (ii) for D14/D30 samples (post release testing), hardness values decreased with increasing excipient loading (0 > 15 > 30%); and (iii) for D14/D30 samples, hardness values generally followed the excipient rank order S < P < H/G, correlating with the absolute solubilities and dissolution rates of the excipients.
Figure 10.
Mean Shore M hardness values for matrix-type rod formulations at D0, D14/D30, and Dried: (A) excipient-free; (B) CN; (C) ZN; (D) CSP; and (E) ZAD. Error bars represent standard deviation of three replicates. The dashed line at 30 represents the Shore M hardness value of the Annovera® vaginal ring. Shore M hardness values of all rod formulations were smaller than that of NuvaRing® (87 units). Thus, Shore M hardness values >30 fall within the range of marketed vaginal rings. The asterisks above purple bars present significant difference levels in Shore M hardness values between B-B rod and API-B rod/between API-B and API-excipient rod at D0; asterisks above orange bars show significant differences for each rod between D0 and D14/30; while asterisks above green bars indicate differences for each rod between D0 and Dried. There.
3.4.2. Elongation testing
Elongation testing at 4 N was performed at two timepoints—D0 and Dried; results are presented in Figure 11. For reference, blank silicone elastomer control rods had D0 and Dried elongation values of ∼84%. For formulations containing 5% w/w of the various actives but no excipients (Figure 11(A)), mean percentage elongations (ranging from 83 to 107%) were mostly not significantly different from the B-B control. However, as with Shore hardness measurements, incorporation of excipients significantly modified the elongation characteristics (Figures 11(B–E)). For D0 samples, elongation values for excipient-loaded formulations were generally lower than for the excipient-free controls. For Dried samples, elongation values were dependent upon the excipient type and loading; notably, gelatin produced the lowest values (31% for ZAD-30G Dried), while sucrose produced values higher than controls (maximum elongations ∼200% measured for 30% sucrose).
Figure 11.
Mean elongation at 4 N for (A) excipient-free; (B) CN; (C) ZN; (D) CSP; and (E) ZAD matrix-type rods at D0 and Dried. Error bars represent the standard deviation of three replicates. The asterisks above purple bars present significant difference levels of elongation of 4 N between B-B rod and API-B rods/between API-B and API-excipient rods at D0; asterisks above green bars present significant difference levels of elongation of 4 N of each rod at D0 versus Dried.
‘Maximum load at maximum extension’—a measure of how much force a material or device can withstand at its most stretched state—is a useful indicator of mechanical durability and robustness. Values for mean maximum load at maximum extension at D0 were very significantly lower for all excipient-free active formulations compared to the 42 N measured for the non-active control (Figure 12(A)). Incorporation of excipients generally caused the maximum load to decrease compared to active-only formulations, with the notable exception of CSP-15S (Figures 12(B–E)).
Figure 12.
Mean maximum load at maximum extension for (A) excipient-free; (B) CN-excipient; (C) ZN-excipient; (D) CSP-excipient; and (E) ZAD-excipient matrix-type rods post-manufacture. Error bars represent standard deviations of three replicates. (B–E) share the same legend as (C). Dash lines in (B–E) represent the mean maximum load of rods containing only API. Asterisks above the bars in (A) represent the significant difference levels in maximum load compared with the B-B rod; asterisks above the bars in (B–E) represent the significant difference levels in maximum load compared with the corresponding excipient-free rods.
‘Percentage elongation at break’ (%ELB) is a mechanical/tensile property that describes how much a material can stretch before it breaks. This parameter has been widely reported for vaginal rings (Boyd et al. 2020; Dallal Bashi et al. 2021); for example, ELB% values for Femring®, NuvaRing®, and Estring® were previously measured as 263, 824, and 1511%, respectively (McCoy et al. 2019), indicating the very different properties of these rings. Mean %ELB values for all rod formulations are presented in Figure 13. Compared to the 939% ELB value for the B-B control rod, values were decreased significantly following incorporation of actives: ZN-B (451%) > ZAD-B (378%) > CN-B (374%) > CSP-B (277%) (Figure 12(A)). %ELB values for CN and ZN formulations containing excipients ranged from 285 to 385 and 348 to 535%, respectively (Figures 12(B,C)). ELB values of 183–548 and 273–485% were measured for CSP and ZAD formulations with excipients, respectively (Figures 13(D,E)). Overall, ELB values for these formulations fell within the range of those measured previously for commercial rings, except for CSP-15G and CSP-30G.
Figure 13.
Mean elongation at break (%) for (A) excipient-free; (B) CN-excipient; (C) ZN-excipient; (D) CSP-excipient; and (E) ZAD-excipient matrix-type rods at post-manufacture. Error bars represent the standard deviation of three replicates. (B–E) share the same legend as (C). Dash lines in (B–E) represent the %ELB of rods containing only API. Asterisks (*) above the bars in (A) represent the significant difference levels in %ELB compared with the B-B rod; asterisks above the bars in (B–E) represent the significant difference levels in %ELB compared with the corresponding excipient-free rods.
In summary, for D0 rods, mechanical performance is dependent upon the formulation composition. Variations in mechanical properties between D0 and D14/D30 are mostly attributed to water absorption by the rods during in vitro release testing, with further contributions due to post-curing effects. Post drying, the compositions of the formulations and the post-curing effect collectively impact the mechanical properties of the dried formulations. For those rods that swelled excessively during in vitro release testing, mechanical performance deteriorated significantly.
4. Conclusions
The results demonstrate several key findings: (i) copper/zinc nanopowders and salts can be successfully incorporated into addition-cure silicone elastomers; (ii) in vitro release of Cu2+/Zn2 ions from silicone elastomers loaded with divalent salts was ∼100 times greater than when loaded with copper/zinc nanopowders, reflecting the differences in aqueous solubility; (iii) incorporation of four common pharmaceutical excipients—gelatin, polyvinylpyrrolidone, sucrose and hydroxypropyl methylcellulose—at 15 or 30% w/w loadings into silicone elastomer devices containing 5% w/w copper/zinc nanopowders or salts significantly enhanced the release of Cu2+/Zn2 ions (up to ∼30-fold); (iv) the enhanced release is due to water absorption into the silicone elastomer devices, causing swelling of the devices to an extent proportional to the excipient loading; (v) water absorption—as measured by weight increase—was greatest for devices containing sucrose, the most water-soluble of the four excipients tested; (vi) the extent of increase in in vitro release of Cu2+/Zn2+ was highly dependent on the type of excipient—for each of copper nanopowders, zinc nanopowders, copper sulfate and zinc acetate, the greatest increase was measured for 30% gelatin, 30% polyvinylpyrrolidone, 30% sucrose, and 30% hydroxypropyl methylcellulose, respectively; and (vii) the incorporation of excipients greatly impacted the mechanical characteristics of the silicone elastomer devices.
The results are significant in the context of ongoing efforts to extend the utility of silicone elastomers to the formulation and release of hydrophilic actives. Currently, all marketed silicone elastomer drug delivery products contain relatively hydrophobic and potent drug actives; hydrophilic drugs do not generally have sufficient solubility within the silicone matrix to release in quantities that are clinically effective. This study demonstrates that while inclusion of hydrophilic excipients may be effective in enhancing release of hydrophilic actives [and presumably also hydrophobic actives (Woolfson et al. 2006)], other issues and challenges are encountered, not least swelling of the devices and changes in their mechanical performance. Certain routes of drug delivery and clinical indications may be able to accommodate such changes, within limits. For example, although marketed silicone elastomer vaginal rings are not known to swell during clinical use, a small increase in weight/dimensions might be clinically acceptable—there are significant differences in weights and dimensions of marketed rings (Boyd et al. 2020; Carson et al. 2021) and considerable variations in the baseline dimensions of the undistended human vagina (Barnhart et al. 2006).
Finally, release of metal ions from silicone elastomers has potential in a range of industrial and clinical indications, due to their antimicrobial, catalytic, and therapeutic properties. While the work here is primarily set within the content of developing vaginal devices for non-hormonal contraception, we can envisage applications around antimicrobial medical devices, bone regeneration and orthopedic implants, cancer therapy, water purification, catalytic applications in chemical processes, and silicone coatings for fabrics releasing antimicrobial ions.
Supplementary Material
Funding Statement
No funding was received.
Disclosure statement
No potential conflict of interest was reported by the author(s).
Data availability statement
The data that support the findings of this study are available from the corresponding author, [R.K.M.], upon reasonable request.
References
- Acartürk F, Altug N.. 2001. In-vitro and in-vivo evaluation of a matrix-controlled bromocriptine mesilate-releasing vaginal ring. J Pharm Pharmacol. 53(12):1721–1726. doi: 10.1211/0022357011778124. [DOI] [PubMed] [Google Scholar]
- Aliyar H, Schalau G.. 2015. Recent developments in silicones for topical and transdermal drug delivery. Ther Deliv. 6(7):827–839. doi: 10.4155/tde.15.39. [DOI] [PubMed] [Google Scholar]
- Amelkovich YA, Nazarenko OB, Sechin AI, Visakh PM.. 2015. Characterization of copper nanopowders after natural aging. IOP Conf Ser Mater Sci Eng. 81:012072. doi: 10.1088/1757-899X/81/1/012072. [DOI] [Google Scholar]
- Bao Q, Gu B, Price CF, Zou Y, Wang Y, Kozak D, Choi S, Burgess DJ.. 2018. Manufacturing and characterization of long-acting levonorgestrel intrauterine systems. Int J Pharm. 550(1–2):447–454. doi: 10.1016/j.ijpharm.2018.09.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Barnhart KT, Izquierdo A, Pretorius ES, Shera DM, Shabbout M, Shaunik A.. 2006. Baseline dimensions of the human vagina. Hum Reprod. 21(6):1618–1622. doi: 10.1093/humrep/del022. [DOI] [PubMed] [Google Scholar]
- Becker D, Rigassi T, Bauer-Brandl A.. 1997. Effectiveness of binders in wet granulation: a comparison using model formulations of different tabletability. Drug Dev Ind Pharm. 23(8):791–808. doi: 10.3109/03639049709150550. [DOI] [PubMed] [Google Scholar]
- Benagiano G, Gabelnick H, Farris M.. 2008. Contraceptive devices: subcutaneous delivery systems. Expert Rev Med Devices. 5(5):623–637. doi: 10.1586/17434440.5.5.623. [DOI] [PubMed] [Google Scholar]
- Boyd P, Merkatz R, Variano B, Malcolm RK.. 2020. The ins and outs of drug-releasing vaginal rings: a literature review of expulsions and removals. Expert Opin Drug Deliv. 17(11):1519–1540. doi: 10.1080/17425247.2020.1798927. [DOI] [PubMed] [Google Scholar]
- Boyd P, Variano B, Spence P, McCoy CF, Murphy DJ, Dallal Bashi YH, Malcolm RK.. 2019. In vitro release testing methods for drug-releasing vaginal rings. J Control Release. 313:54–69. doi: 10.1016/j.jconrel.2019.10.015. [DOI] [PubMed] [Google Scholar]
- Brook MA, Holloway AC, Ng KK, Hrynyk M, Moore C, Lall R.. 2008. Using a drug to structure its release matrix and release profile. Int J Pharm. 358(1–2):121–127. doi: 10.1016/j.ijpharm.2008.02.029. [DOI] [PubMed] [Google Scholar]
- Bruschi ML. 2015. Mathematical models of drug release. In: Strategies to modify the drug release from pharmaceutical systems. Cambridge, UK: Woodhead Publishing (Elsevier; ); p. 63–86. doi: 10.1016/B978-0-08-100092-2.00005-9. [DOI] [Google Scholar]
- Carelli V, Di Colo G, Guerrini C, Nannipieri E.. 1989. Drug release from silicone elastomer through controlled polymer cracking: an extension to macromolecular drugs. Int J Pharm. 50(3):181–188. doi: 10.1016/0378-5173(89)90120-8. [DOI] [Google Scholar]
- Carelli V, Di Colo G, Nannipieri E, Serafini MF.. 1995. Evaluation of a silicone based matrix containing a crosslinked polyethylene glycol as a controlled drug delivery system for potential oral application. J Control Release. 33(1):153–162. doi: 10.1016/0168-3659(94)00081-5. [DOI] [Google Scholar]
- Carelli V, Di Colo G, Nannipieri E.. 1987. Factors in zero-order release of clonidine hydrochloride from monolithic polydimethylsiloxane matrices. Int J Pharm. 35(1–2):21–28. doi: 10.1016/0378-5173(87)90070-6. [DOI] [Google Scholar]
- Carelli V, Di Colo G, Nannipieri E.. 1988. Zero-order drug release from monolithic polydimethylsiloxane matrices through controlled polymer cracking. Farmaco Prat. 43(4):121–135. http://www.ncbi.nlm.nih.gov/pubmed/3208877. [PubMed] [Google Scholar]
- Carson L, Merkatz R, Martinelli E, Boyd P, Variano B, Sallent T, Malcolm RK.. 2021. The vaginal microbiota, bacterial biofilms and polymeric drug-releasing vaginal rings. Pharmaceutics. 13(5):751. doi: 10.3390/pharmaceutics13050751. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Chowhan ZT. 1980. Role of binders in moisture-induced hardness increase in compressed tablets and its effect on in vitro disintegration and dissolution. J Pharm Sci. 69(1):1–4. doi: 10.1002/jps.2600690102. [DOI] [PubMed] [Google Scholar]
- Crank J. 1975. The mathematics of diffusion. Oxford: Clarendon Press. 2nd ed. [Google Scholar]
- Dahl TC, Sue IT.. 1992. Mechanisms to control drug release from pellets coated with a silicone elastomer aqueous dispersion. Pharm Res. 9(3):398–405. doi: 10.1023/A:1015855305679. [DOI] [PubMed] [Google Scholar]
- Dallal Bashi YH, Murphy DJ, McCoy CF, Boyd P, Brown L, Kihara M, Martin F, McMullen N, Kleinbeck K, Dangi B, et al. 2021. Silicone elastomer formulations for improved performance of a multipurpose vaginal ring releasing dapivirine and levonorgestrel. Int J Pharm X. 3:100091. doi: 10.1016/j.ijpx.2021.100091. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Delalu H, Vignalou JR, Elkhatib M, Metz R.. 2000. Kinetics and modeling of diffusion phenomena occurring during the complete oxidation of zinc powder: influence of granulometry, temperature and relative humidity of the oxidizing fluid. Solid State Sci. 2(2):229–235. doi: 10.1016/S1293-2558(00)00130-8. [DOI] [Google Scholar]
- Di Colo G, Carelli V, Nannipieri E, Serafini MF, Vitale D, Bottari F.. 1983. Effect of different water-soluble additives on the sustained release of sulfanilamide from silicone rubber matrices. Farmaco Prat. 37(12):377–389. http://www.ncbi.nlm.nih.gov/pubmed/6822285. [PubMed] [Google Scholar]
- Di Colo G, Carelli V, Nannipieri E, Serafini MF, Vitale D.. 1986. Effect of water-soluble additives on drug release from silicone rubber matrices. II. Sustained release of prednisolone from non-swelling devices. Int J Pharm. 30(1):1–7. doi: 10.1016/0378-5173(86)90129-8. [DOI] [Google Scholar]
- Di Colo G. 1992. Controlled drug release from implantable matrices based on hydrophobic polymers. Biomaterials. 13(12):850–856. doi: 10.1016/0142-9612(92)90178-Q. [DOI] [PubMed] [Google Scholar]
- Du W, Zhang Z, Fan W, Gao W, Su H, Li Z.. 2018. Fabrication and evaluation of polydimethylsiloxane modified gelatin/silicone rubber asymmetric bilayer membrane with porous structure. Mater Des. 158:28–38. doi: 10.1016/j.matdes.2018.08.017. [DOI] [Google Scholar]
- Efentakis M, Vlachou M, Choulis NH.. 1997. Effects of excipients on swelling and drug release from compressed matrices. Drug Dev Ind Pharm. 23(1):107–112. doi: 10.3109/03639049709148487. [DOI] [Google Scholar]
- Fanse S, Bao Q, Zou Y, Wang Y, Burgess DJ.. 2024. Tailoring drug release from long-acting contraceptive levonorgestrel intrauterine systems. J Control Release. 370:124–139. doi: 10.1016/j.jconrel.2024.04.027. [DOI] [PubMed] [Google Scholar]
- Farahmandghavi F, Imani M, Hajiesmaeelian F.. 2019. Silicone matrices loaded with levonorgestrel particles: Impact of the particle size on drug release. J Drug Deliv Sci Technol. 49:132–142. doi: 10.1016/j.jddst.2018.10.029. [DOI] [Google Scholar]
- Forbes CJ, Mccoy CF, Murphy DJ, David Woolfson A, Moore JP, Evans A, Shattock RJ, Karl Malcolm R.. 2014. Modified silicone elastomer vaginal gels for sustained release of antiretroviral HIV microbicides. J Pharm Sci. 103(5):1422–1432. doi: 10.1002/jps.23913. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fu Y, Kao WJ.. 2010. Drug release kinetics and transport mechanisms of non-degradable and degradable polymeric delivery systems. Expert Opin Drug Deliv. 7(4):429–444. doi: 10.1517/17425241003602259. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ghannam MM, Tojo K, Chien YW.. 1986. Kinetics and thermodynamics of drug permeation through silicone elastomers (I) effect of penetrant hydrophilicity. Drug Dev Ind Pharm. 12(3):303–325. doi: 10.3109/03639048609026616. [DOI] [Google Scholar]
- Higuchi T. 1961. Rate of release of medicaments from ointment bases containing drugs in suspension. J Pharm Sci. 50(10):874–875. doi: 10.1002/jps.2600501018. [DOI] [PubMed] [Google Scholar]
- Holland MK, White IG.. 1988. Heavy metals and human spermatozoa. III. The toxicity of copper ions for spermatozoa. Contraception. 38(6):685–695. doi: 10.1016/0010-7824(88)90050-9. [DOI] [PubMed] [Google Scholar]
- Hsieh DST, Mann K, Chien YW.. 1985. Enhanced release of drugs from silicone elastomers (I) release kinetics of pineal and steroidal hormones. Drug Dev Ind Pharm. 11(6–7):1391–1410. doi: 10.3109/03639048509059870. [DOI] [Google Scholar]
- Hu Z, Wang C, Nelson KD, Eberhart RC.. 2000. Controlled release from a composite silicone/hydrogel membrane. ASAIO J. 46(4):431–434. doi: 10.1097/00002480-200007000-00012. [DOI] [PubMed] [Google Scholar]
- Huang YL, Tseng WC, Lin TH.. 2001. In vitro effects of metal ions (Fe2+, Mn2+, Pb2+) on sperm motility and lipid peroxidation in human semen. J Toxicol Environ Health A. 62(4):259–267. doi: 10.1080/009841001459414. [DOI] [PubMed] [Google Scholar]
- Iwata M, Ueda H.. 1996. Dissolution properties of glibenclamide in combinations with polyvinylpyrrolidone. Drug Dev Ind Pharm. 22(11):1161–1165. doi: 10.3109/03639049609065953. [DOI] [Google Scholar]
- Jecht EW, Bernstein GS.. 1973. The influence of copper on the motility of human spermatozoa. Contraception. 7(5):381–401. doi: 10.1016/0010-7824(73)90140-6. [DOI] [Google Scholar]
- Johnson TJ, Gupta KM, Fabian J, Albright TH, Kiser PF.. 2010. Segmented polyurethane intravaginal rings for the sustained combined delivery of antiretroviral agents dapivirine and tenofovir. Eur J Pharm Sci. 39(4):203–212. doi: 10.1016/j.ejps.2009.11.007. [DOI] [PubMed] [Google Scholar]
- Kajihara M, Sugie T, Hojo T, Maeda H, Sano A, Fujioka K, Sugawara S, Urabe Y.. 2001. Development of a new drug delivery system for protein drugs using silicone (II). J Control Release. 73(2–3):279–291. doi: 10.1016/S0168-3659(01)00302-9. [DOI] [PubMed] [Google Scholar]
- Kajihara M, Sugie T, Maeda H, Sano A, Fujioka K, Urabe Y, Tanihara M, Imanishi Y.. 2003. Novel drug delivery device using silicone: controlled release of insoluble drugs or two kinds of water-soluble drugs. Chem Pharm Bull. 51(1):15–19. doi: 10.1248/cpb.51.15. [DOI] [PubMed] [Google Scholar]
- Kajihara M, Sugie T, Mizuno M, Tamura N, Sano A, Fujioka K, Kashiwazaki Y, Yamaoka T, Sugawara S, Urabe Y.. 2000. Development of new drug delivery system for protein drugs using silicone (I). J Control Release. 66(1):49–61. doi: 10.1016/S0168-3659(99)00257-6. [DOI] [PubMed] [Google Scholar]
- Kommareddy S, Shenoy DB, Amiji MM.. 2003. Gelatin nanoparticles and their biofunctionalization. In: Nanotechnologies for the life sciences, Vol. 1 Biofunctionalization of Nanomaterials. John Wiley & Sons, Ltd. (Web.) doi: 10.1002/9783527610419.ntls0011. [DOI] [Google Scholar]
- Li LC, Peck GE.. 1989. Water based silicone elastomer controlled release tablet film coating 1: free film evaluation. Drug Dev Ind Pharm. 15(1):65–95. doi: 10.3109/03639048909040197. [DOI] [Google Scholar]
- Lopour P, Vondrácek P, Janatová V, Sulc J, Vacík J.. 1990. Silicone rubber-hydrogel composites as polymeric biomaterials. Biomaterials. 11(6):397–402. doi: 10.1016/0142-9612(90)90094-7. [DOI] [PubMed] [Google Scholar]
- Ma X, Chen J, Yang Y, Su X, Zhang S, Gao B, Li YC.. 2018. Siloxane and polyether dual modification improves hydrophobicity and interpenetrating polymer network of bio-polymer for coated fertilizers with enhanced slow release characteristics. Chem Eng J. 350:1125–1134. doi: 10.1016/j.cej.2018.06.061. [DOI] [Google Scholar]
- Maeda H, Brandon M, Sano A.. 2003. Design of controlled-release formulation for ivermectin using silicone. Int J Pharm. 261(1–2):9–19. doi: 10.1016/S0378-5173(03)00293-X. [DOI] [PubMed] [Google Scholar]
- Maeda H, Ohashi E, Sano A, Kawasaki H, Kurosaki Y.. 2003. Investigation of the release behavior of a covered-rod-type formulation using silicone. J Control Release. 90(1):59–70. doi: 10.1016/S0168-3659(03)00158-5. [DOI] [PubMed] [Google Scholar]
- Malcolm K, Woolfson D, Russell J, Tallon P, McAuley L, Craig D.. 2003. Influence of silicone elastomer solubility and diffusivity on the in vitro release of drugs from intravaginal rings. J Control Release. 90(2):217–225. doi: 10.1016/S0168-3659(03)00178-0. [DOI] [PubMed] [Google Scholar]
- Malcolm RK, Boyd PJ, McCoy CF, Murphy DJ.. 2016. Microbicide vaginal rings: technological challenges and clinical development. Adv Drug Deliv Rev. 103:33–56. doi: 10.1016/j.addr.2016.01.015. [DOI] [PubMed] [Google Scholar]
- Manosroi J, Apriyani MG, Foe K, Manosroi A.. 2005. Enhancement of the release of azelaic acid through the synthetic membranes by inclusion complex formation with hydroxypropyl-β-cyclodextrin. Int J Pharm. 293(1–2):235–240. doi: 10.1016/j.ijpharm.2005.01.009. [DOI] [PubMed] [Google Scholar]
- Mashak A, Rahimi A, Mashmak A, Rahimi A.. 2009. Silicone polymers in controlled drug delivery systems: a review. Polym J. 18:279–295. [Google Scholar]
- Mashak A. 2008. In vitro drug release from silicone rubber–polyacrylamide composite. Silicon Chem. 3(6):295–301. doi: 10.1007/s11201-007-9031-1. [DOI] [Google Scholar]
- Mazurek P, Brook MA, Skov AL.. 2018. Glycerol–silicone elastomers as active matrices with controllable release profiles. Langmuir. 34(38):11559–11566. doi: 10.1021/acs.langmuir.8b02039. [DOI] [PubMed] [Google Scholar]
- Mazurek P, Frederiksen NS, Silau H, Yuusuf NA, Mordhorst H, Pamp SJ, Skov AL.. 2021. Glycerol–silicone membranes for sustained and controlled topical delivery of antimicrobial and pain‐relief drugs. Adv Mater Int. 8(5):2001873. doi: 10.1002/admi.202001873. [DOI] [Google Scholar]
- Mazurek P, Yuusuf NA, Silau H, Mordhorst H, Pamp SJ, Brook MA, Skov AL.. 2021. Simultaneous delivery of several antimicrobial drugs from multi‐compartment glycerol‐silicone membranes. J Appl Polym Sci. 138(32):50780. doi: 10.1002/app.50780. [DOI] [Google Scholar]
- McBride JW, Boyd P, Dias N, Cameron D, Offord RE, Hartley O, Kett VL, Malcolm RK.. 2019. Vaginal rings with exposed cores for sustained delivery of the HIV CCR5 inhibitor 5P12-RANTES. J Contr Release. 298:1–11. doi: 10.1016/j.jconrel.2019.02.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- McBride JW, Malcolm RK, Dias N, Cameron D, Offord RE, Hartley O, Kett VL, Devlin B, Boyd P.. 2019. Development and pharmacokinetics of a combination vaginal ring for sustained release of dapivirine and the protein microbicide 5P12-RANTES. Int J Pharm. 564:207–213. doi: 10.1016/j.ijpharm.2019.04.040. [DOI] [PubMed] [Google Scholar]
- McBride MC, Karl Malcolm R, David Woolfson A, Gorman SP.. 2009. Persistence of antimicrobial activity through sustained release of triclosan from pegylated silicone elastomers. Biomaterials. 30(35):6739–6747. doi: 10.1016/j.biomaterials.2009.08.047. [DOI] [PubMed] [Google Scholar]
- McConville C, Andrews GP, David Woolfson A, Karl Malcolm R.. 2012. Effect of the incorporation of hydroxy‐terminated liquid silicones on the cure characteristics, morphology, and release of a model protein from silicone elastomer‐covered rods. J Appl Polym Sci. 124(1):805–812. doi: 10.1002/app.35080. [DOI] [Google Scholar]
- McCoy CF, Millar BG, Murphy DJ, Blanda W, Hansraj B, Devlin B, Malcolm RK, Boyd P.. 2019. Mechanical testing methods for drug-releasing vaginal rings. Int J Pharm. 559:182–191. doi: 10.1016/j.ijpharm.2019.01.026. [DOI] [PubMed] [Google Scholar]
- McKay PF, Mann J, Pattani A, Kett V, Malcolm K, Shattock RJ.. 2012. Intravaginal immunization using a novel antigen delivery device elicits robust vaccine antigen-specific systemic and mucosal humoral immune responses. Retrovirology. 9(S2):P191. doi: 10.1186/1742-4690-9-S2-P191. [DOI] [PMC free article] [PubMed] [Google Scholar]
- McLellan-Lemal E, Deaton SR, Betts JE, Ondenge K, Mudhune V, O’Connor SM, Nyagol B, Thurman AR, Doncel GF, Allen SA, et al. 2022. Acceptability of an intravaginal ring for simultaneously preventing HIV infection and pregnancy: qualitative findings of the Kisumu combined ring study, 2019. Contemp Clin Trials. 122:106935. doi: 10.1016/j.cct.2022.106935. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mikhailov OV. 2023. Gelatin as it is: history and modernity. Int J Mol Sci. 24(4):3583. doi: 10.3390/ijms24043583. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mikolaszek B, Kazlauske J, Larsson A, Sznitowska M.. 2020. Controlled drug release by the pore structure in polydimethylsiloxane transdermal patches. Polymers. 12(7):1520. doi: 10.3390/polym12071520. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Morrow RJ, Woolfson AD, Donnelly L, Curran R, Andrews G, Katinger D, Malcolm RK.. 2011. Sustained release of proteins from a modified vaginal ring device. Eur J Pharm Biopharm. 77(1):3–10. doi: 10.1016/j.ejpb.2010.10.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mullarney MP, Hancock BC, Carlson GT, Ladipo DD, Langdon BA.. 2003. The powder flow and compact mechanical properties of sucrose and three high-intensity sweeteners used in chewable tablets. Int J Pharm. 257(1–2):227–236. doi: 10.1016/S0378-5173(03)00144-3. [DOI] [PubMed] [Google Scholar]
- Murphy DJ, Amssoms K, Pille G, Clarke A, O’Hara M, van Roey J, Malcolm RK.. 2016. Sustained release of the candidate antiretroviral peptides T-1249 and JNJ54310516-AFP from a rod insert vaginal ring. Drug Deliv Transl Res. 6(3):234–242. doi: 10.1007/s13346-015-0273-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Murphy DJ, McCoy CF, Plagianos M, RamaRao S, Merkatz R, Clark H, Boyd P, Variano B, Malcolm RK.. 2019. Post-use ring weight and residual drug content as potential objective measures of user adherence to a contraceptive progesterone vaginal ring. Contraception. 100(3):241–246. doi: 10.1016/j.contraception.2019.06.013. [DOI] [PubMed] [Google Scholar]
- Nemati P, Imani M, Farahmandghavi F, Mirzadeh H, Marzban-Rad E, Nasrabadi AM.. 2014. Artificial neural networks for bilateral prediction of formulation parameters and drug release profiles from cochlear implant coatings fabricated as porous monolithic devices based on silicone rubber. J Pharm Pharmacol. 66(5):624–638. doi: 10.1111/jphp.12187. [DOI] [PubMed] [Google Scholar]
- Owen DH, Katz DF.. 1999. A vaginal fluid simulant. Contraception. 59(2):91–95. doi: 10.1016/S0010-7824(99)00010-4. [DOI] [PubMed] [Google Scholar]
- Pattani A, Lowry D, Curran RM, McGrath S, Kett VL, Andrews GP, Malcolm RK.. 2012. Characterisation of protein stability in rod-insert vaginal rings. Int J Pharm. 430(1–2):89–97. doi: 10.1016/j.ijpharm.2012.03.036. [DOI] [PubMed] [Google Scholar]
- Peppas NA. 1985. Analysis of Fickian and non-Fickian drug release from polymers. Pharm Acta Helv. 60(4):110–111. https://pubmed.ncbi.nlm.nih.gov/4011621/. [PubMed] [Google Scholar]
- Rajendra V, Chen Y, Brook MA.. 2010. Structured hydrophilic domains on silicone elastomers. Polym Chem. 1(3):312–320. doi: 10.1039/B9PY00220K. [DOI] [Google Scholar]
- Roblero L, Guadarrama A, Lopez T, Zegers-Hochschild F.. 1996. Effect of copper ion on the motility, viability, acrosome reaction and fertilizing capacity of human spermatozoa in vitro. Reprod Fertil Dev. 8(5):871–874. doi: 10.1071/RD9960871. [DOI] [PubMed] [Google Scholar]
- Roseman TJ, Higuchi WI.. 1970. Release of medroxyprogesterone acetate from a silicone polymer. J Pharm Sci. 59(3):353–357. doi: 10.1002/jps.2600590317. [DOI] [PubMed] [Google Scholar]
- Rowe RC. 1977. The adhesion of film coatings to tablet surfaces—the effect of some direct compression excipients and lubricants. J Pharm Pharmacol. 29(12):723–726. doi: 10.1111/j.2042-7158.1977.tb11449.x. [DOI] [PubMed] [Google Scholar]
- Shah AC, Britten NJ, Olanoff LS, Badalamenti JN.. 1989. Gel-matrixsystems exhibiting bimodal controlled release for oral drug delivery. J Control Release. 9(2):169–175. doi: 10.1016/0168-3659(89)90007-2. [DOI] [Google Scholar]
- Shen X, Boyd P, Dallal Bashi YH, McCoy CF, Zhao X, Malcolm RK.. 2025. A multipurpose silicone elastomer vaginal ring releasing dapivirine and Cu2+/Zn2+ ions for HIV prevention and non-hormonal contraception. Int J Pharm. 674:125442. doi: 10.1016/j.ijpharm.2025.125442. [DOI] [PubMed] [Google Scholar]
- Snorradóttir BS, Gudnason PI, Thorsteinsson F, Másson M.. 2011. Experimental design for optimizing drug release from silicone elastomer matrix and investigation of transdermal drug delivery. Eur J Pharm Sci. 42(5):559–567. doi: 10.1016/j.ejps.2011.02.011. [DOI] [PubMed] [Google Scholar]
- Stubberud L, Arwidsson HG, Hjortsberg V, Graffner C.. 1996. Water-solid interactions. III. Effect of glass transition temperature, Tg, and processing on tensile strength of compacts of lactose and lactose/polyvinyl pyrrolidone. Pharm Dev Technol. 1(2):195–204. doi: 10.3109/10837459609029894. [DOI] [PubMed] [Google Scholar]
- Sugimoto M, Narisawa S, Matsubara K, Yoshino H, Nakano M, Handa T.. 2006. Development of manufacturing method for rapidly disintegrating oral tablets using the crystalline transition of amorphous sucrose. Int J Pharm. 320(1–2):71–78. doi: 10.1016/j.ijpharm.2006.04.004. [DOI] [PubMed] [Google Scholar]
- Talevi A, Ruiz ME.. 2021. Korsmeyer-Peppas, Peppas-Sahlin, and Brazel-Peppas: models of drug release. In: The ADME encyclopedia. Cham: Springer International Publishing; p. 1–9. doi: 10.1007/978-3-030-51519-5_35-1. [DOI] [Google Scholar]
- Tolia G, Li SK.. 2012. Study of drug release and tablet characteristics of silicone adhesive matrix tablets. Eur J Pharm Biopharm. 82(3):518–525. doi: 10.1016/j.ejpb.2012.07.006. [DOI] [PubMed] [Google Scholar]
- Undie C-C, RamaRao S, Mbow FB.. 2020. Choosing and using the progesterone vaginal ring: women’s lived experiences in three African cities. Patient Prefer Adherence. 14:1761–1770. doi: 10.2147/PPA.S265503. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Woolfson AD, Malcolm RK, Gallagher RJ.. 2003. Design of a silicone reservoir intravaginal ring for the delivery of oxybutynin. J Control Release. 91:465–476. doi: 10.1016/S0168-3659(03)00277-3. [DOI] [PubMed] [Google Scholar]
- Woolfson AD, Malcolm RK, Morrow RJ, Toner CF, McCullagh SD.. 2006. Intravaginal ring delivery of the reverse transcriptase inhibitor TMC 120 as an HIV microbicide. Int J Pharm. 325(1–2):82–89. doi: 10.1016/j.ijpharm.2006.06.026. [DOI] [PubMed] [Google Scholar]
- Xu XX, Ding MH, Zhang JX, Zheng W, Li L, Zheng YF.. 2013. A novel copper/polydimethiylsiloxane nanocomposite for copper-containing intrauterine contraceptive devices. J Biomed Mater Res B Appl Biomater. 101(8):1428–1436. doi: 10.1002/jbm.b.32962. [DOI] [PubMed] [Google Scholar]
- Zhang Y, Li C-X, Ning M-Y, Duan X-Y, Liu Y.. 2013. Preparation and evaluation of intravaginal ring containing drospirenone. Adv Pharmacol Sci. 2013:192408. doi: 10.1155/2013/192408. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The data that support the findings of this study are available from the corresponding author, [R.K.M.], upon reasonable request.