A comparison of the image quality between deep learning reconstruction algorithm and iDose4 using low dose abdominopelvic computed tomography for individuals with normal BMI

Thejas Marike Shivakumar 1, Nitika C Panakkal 1,, Shailesh Nayak 1, Rajagopal Kadavigere 2, Tanushree R Kamath 2, Suresh Sukumar 1
PMCID: PMC12374083  PMID: 40862257

Abstract

Objectives:

Radiation exposure has been a cause of concern in computed tomography imaging. Reducing radiation dose increases the image noise which can be compensated by using reconstruction techniques. Recently artificial intelligence-based reconstruction technique has been introduced. Therefore, the purpose of the study was to prospectively compare the image quality between Idose4 and Precise Image in normal BMI individuals.

Methods:

Sixty-six consecutive patients with a normal body habitus undergoing contrast-enhanced abdomen and pelvis scan were included in the study. All scans were performed using 100 kVp and tube current modulation. The acquired images were reconstructed to iDose4 and precise imaging. Quantitatively images were analyzed by placing regions of interest in different organs to estimate the image noise, signal-to-noise ratio, and contrast-to-noise ratio. Qualitative analysis was done by two radiologists on a five-point Likert scale.

Results:

Image noise was significantly reduced using Precise Image across the plain (9.11 ± 1.43 vs 8.18 ± 1.2), arterial (14.34 ± 2.1 vs 10.21 ± 1.5), and portovenous phase (14.78 ± 2.30 vs 11.97 ± 2.07) with maximum noise reduction in the arterial and portovenous phases. Signal-to-noise ratio and contrast-to-noise ratio was significantly improved in all the organs across the plain, arterial, and portovenous phases. Qualitative analysis showed no significant difference between Idose4 and Precise Image with regards to visualization of large vessels in the arterial and portovenous phases. However, precise image was graded better than Idose4 with respect to visualization/conspicuity, image noise, and artifacts.

Conclusion:

Precise Image can be useful in reducing the image noise and improving the signal-to-noise ratio and contrast-to-noise ratio in low-dose computed tomography protocol among normal BMI individuals.

Keywords: Image reconstruction, image quality, deep learning reconstruction, Precise Image reconstruction, computed tomography

Introduction

Since the 1970s, numerous developments in computed tomography (CT) hardware and software have been built resulting in the increased usage. 1 Due to this increased demand, radiation exposure has become a cause of concern. However, implementing methods to reduce radiation dose often result in increased image noise and filtered back projection (FBP) reconstruction techniques are not suited for low-dose CTs. 2 To maintain image quality at lower doses, iterative reconstruction (IR) technique was introduced. This technique increases the spatial resolution and decreases noise making it possible to reduce radiation dose by 23%–76%.3,4 The model-based IR (MBIR) that uses both forward and backward projection steps are computationally demanding requiring longer reconstruction time. 2 Most of the CT vendors use a combined IR and FBP approach known as hybrid iterative reconstruction (HIR), that uses only a single backward projection phase to provide a compromise between the MBIR and FBP. All major vendors have their variations such as Adaptive statistical iterative reconstruction (ASiR), GE healthcare, Adaptive iterative dose reduction 3D (AIDR), Toshiba, Sinogram affirmed iterative reconstruction (SAFIRE) Siemens and iDose, Phillips. 5 The iDose4 filters noise effectively in both projection and image domain and has reportedly brought about dose reduction of upto 76% without loss in image quality. 6 Most recently, artificial intelligence (AI)-based deep learning reconstruction (DLR) techniques have been introduced to reduce noise levels and stimulate signals to deliver sharp, clear, and distinct images at faster speeds. This DLR makes use of convolutional neural networks (CNN) that is trained to reproduce noise patterns of conventional dose images from raw data of low-dose scans in a supervised learning process using simulation techniques. 7 Different manufacturers have come up with their own AI reconstruction techniques namely AiCE by Cannon, TrueFidelity by GE, and Precise Image by Philips. A few studies have evaluated the usefulness of AI reconstruction techniques for improving image quality in terms of signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) in different body regions.3,8,9 Much of the studies reported in literature have only estimated image noise and CNR in either the pre-contrast(plain), arterial phase or portovenous phase. Since deep learning technique involves training the algorithm using different datasets, evaluating the performance of AI algorithm in reducing noise in both pre- and postcontrast phases is essential. To our knowledge, no studies have been done to assess the image quality in abdomen and pelvis for the three phases including plain, arterial, and portovenous phases using precise image. Also, very few studies have evaluated the image quality of these AI-based reconstruction techniques at lower doses. Moreover, all the studies in literature have included a wide range of body habitus ranging from a lower to higher BMI. Body size can influence how radiation is distributed in body and impact image quality. Following a previous study that assessed the feasibility of using 100 kVp among normal BMI patients using iDose4, the current study sought to assess whether the DLR method could improve image quality as described in the literature. 10 Therefore, the study aims to compare the image quality between Idose4 and AI-based DLR technique (precise image) using a lower kVp of 100 among patients with a normal body habitus for a multiphase abdomen and pelvis CT.

Methods

The prospective cross-sectional study was conducted from September 2023 to January 2024 and received Institutional Ethics Committee approval (IEC no: 182/2023).

At 5% level of significance, assuming 80% power, the minimum sample size required to compare the means of image quality was calculated using the following formula n=(Zα/2+Zβ)2*2*σ2/d2 , where n represents minimum number of subjects, σ2 represents population variance, and d represents the clinically significant difference.

After obtaining informed written consent, around 66 patients were referred for a contrast-enhanced computed tomography (CECT) of abdomen and pelvis and scanned using a 128-slice Incisive CT scanner. The technical parameter included were 100 kVp, pitch of 1.20, rotation time of 0.50, slice thickness of 5 mm, detector configuration 64*0.625, and tube current modulation with reference ranging from 80 to 250 mA. During the scan all patients were administered 80 ml of low osmolar contrast media Iohexol (Omnipaque 350) at a flow rate of 3.5 ml/s. The imaging protocol included acquisition of three phases: plain, arterial, and portovenous phases. All contrast scans were acquired with a delay of 8 and 45 s postthreshold for the arterial and portovenous phase, respectively. Images were reconstructed using iDose (level 4) and Precise Image (Standard). All the images were evaluated quantitatively and qualitatively.

Patient selection

All patients with a normal BMI referred for a triple phase CECT abdomen and pelvis were included in the study. Patients with conditions like fatty liver, cirrhosis that could change the density of organs were excluded from the study.

Quantitative analysis

For quantitative analysis, CT attenuation numbers were estimated by placing regions of interest (ROIs) in the liver, aorta, spleen, and posterior paraspinal muscle on the arterial, plain, and portovenous phases. Additionally, in portovenous phase, the ROI was drawn on the portal vein. 11 The ROI area for vessels was 25 mm2 and for organs was 100 mm2 (Figure 1). ROIs were placed in the organs in two consecutive slices, and average CT Hounsfield numbers were taken to estimate image quality parameters such as SNR, CNR, and image noise using equations (1)–(3).12,13 Image noise was estimated by taking average of the standard deviation of all organs.

Figure 1.

Figure 1.

ROI placement in different organs.

Figure adapted from “Optimising contrast volume based on body mass index in low tube voltage CT of abdomen and pelvis” by Panakkal N C. EPOS (online) ECR2025.10.26044/ecr2025/C-15098. https://epos.myesr.org/poster/esr/ecr2025/C-15098

CNR=CTnumberROI-CTnumberROIparaspinalMuscleSDofROIparaspinalmuscle (1)
SNR=CTnumberROIOrganSDOrgan (2)
Imagenoise=SDoforgans (3)

Qualitative analysis

A blinded qualitative analysis was conducted by two board-certified radiologists. Image quality was graded on a 5-point Likert scale based on “Visualization/conspicuity (1-cannot identify; 2-suboptimal; 3-acceptable; 4-better than acceptable; 5-excellently visualized). Critical reproduction (1-blurry; 2-suboptimal; 3-acceptable; 4-better than acceptable; 5-sharpest). Visualization of large vessels (1-cannot identify; 2-suboptimal; 3-acceptable; 4-better than acceptable; 5- excellently visualized). Image contrast (1-very poor contrast; 2-suboptimal image contrast; 3-acceptable image contrast; 4-above-average contrast; 5-excellent image contrast). Image noise (1-un acceptable image noise; 2-above-average noise; 3-average image noise; 4-less-than-average noise; 5-minimal image noise).” Artifacts were assessed in a four-point Likert scale. “(1-Artifacts affecting diagnostic information, 2-Major artifacts affecting visualization of major structures but diagnosis still possible, 3-Minor artifacts are not interfering with the diagnostic decision, 4-No artifacts).” 14

Statistical analysis

The data were analyzed using International business machines corporation (IBM) Statistical package for the social sciences Statistics 21.0. Demographic characteristics and image quality parameters (CNR, SNR, and image noise) were summarized using descriptive statistics. To compare image quality between two techniques, paired t-test was used for normally distributed data and skewed data was analyzed using the Wilcoxon signed rank test. Intraclass correlation was done to evaluate rater agreement. A p-value of less than 0.05 was considered statistical significance.

Results

Patient characteristics

There were 66 patients included in the study with a normal body mass index between 18.9 and 24.9 kg/m2 with an average BMI of 22.8 ± 1.19, of which 35 were males and 31 were females. A wide range of age (18–88 years, mean 52 ± 17.5 years), height (150–185 cm, mean, 169.9 ± 8.2 cm), and weight (50–81 kg, mean 65.5 ± 8.0) were noted.

Quantitative image quality

The mean CT numbers, SNR, CNR are presented in Tables 1 and 2. In the normal BMI group, CT numbers didn’t differ significantly between iDose4 and Precise Image across the three phases apart from the liver in plain images. Image noise was significantly reduced with precise image in the plain (9.1 ± 1.4 vs 8.1 ± 1.2; p < 0.001), arterial (14.3 ± 2.1 vs 10.2 ± 5; p < 0.001), and portovenous (14.7 ± 2.3 vs 11.9 ± 2; p < 0.001) phase. SNR was significantly improved in all the organs across the plain, arterial, and portovenous phases using precise image as shown in Table 1. However, the increase in SNR in the aorta in plain series was not significant. CNR was observed to significantly increase in all the organs across the plain, arterial, and portovenous phases using precise image. However, the increase in CNR in the liver in plain series was not significant as shown in Table 2.

Table 1.

Descriptive summary of CT numbers in mean/median (Q1, Q3) for Idose4 and Precise Image across the plain, arterial, and portovenous phase.

Organs Idose4 Precise image p-Value
Plain
 Aorta 42.7 ± 4 42 (39.8, 45.5) 0.4
 Liver 53 (47, 57.6) 53.2 (46.2, 56.7) 0.000
 Spleen 49.7 (47.5, 52.1) 49.2 (47, 52.5) 0.5
Arterial
 Aorta 375.2 (333.6, 442.2) 381.2 (333, 442) 0.1
 Liver 62.5 ± 11.6 62 ± 11.1 0.3
 Spleen 115.1 ± 24.7 114.6 ± 24.7 0.3
Portovenous
 Aorta 175 ± 27.4 174.6 ± 27.3 0.5
 Liver 114.9 ± 20.5 115.4 ± 20.3 0.5
 Spleen 126 (113, 134.7) 126 (113, 138.1) 0.4
 Portal vein 177 ± 27.2 178 ± 27.4 0.6

Table 2.

Descriptive summary of CNR and SNR in mean/median (Q1, Q3) for Idose4 and Precise Image across the plain, arterial, and portovenous phase.

Phases Organs CNR p-Value SNR p-Value
Idose4 Precise image Idose4 Precise image
Plain Aorta 0.71 (1.63, 1.26) 0.82 (0.26, 1.40) <0.001 4.41 (5.07, 3.82) 4.51 (5.22, 4.00) 0.34
Liver 0.35 (1.01, 0.33) 0.35 (0.99, 0.44) 0.367 6.30 ± 1.70 7.03 ± 1.65 <0.001
Spleen 0.028 ± 0.923 0.0155 ± 0.94 <0.001 5.80 (6.88, 5.17) 6.75 (7.54, 5.98) <0.001
Arterial phase Aorta 25.72 (31.51, 22.80) 38.57 (47.81, 31.81) <0.001 24.83 ± 8.09 36.24 ± 10.36 <0.001
Liver 0.57 ± 1.10 0.74 ± 1.5 <0.001 5.11 ± 1.02 7.60 ± 1.51 <0.001
Spleen 4.54 (6.15, 3.12) 7.05 (8.89, 4.03) <0.001 4.62 (5.83, 3.66) 9.80 (12.2, 7.33) <0.001
Porto venous Aorta 8.47 (10.58, 6.73) 12.14 (14.42, 9.70) <0.001 11.08 ± 3.38 15.86 ± 4.62 <0.001
Liver 4.05 ± 1.80 5.85 ± 2.66 <0.001 6.12 (9.63, 5.25) 8.65 (9.66, 7.34) 0.039
Spleen 4.94 (5.89, 3.89) 6.96 (8.50, 5.72) <0.001 8.98 (11.35, 7.83) 14 (17.36, 11.97) <0.001
Portal vein 8.99 ± 2.68 12.90 ± 4.11 <0.001 11.14 ± 2.69 15.45 ± 3.61 <0.001

Qualitative image quality

The intraclass correlation agreement between the readers ranged from 0.916 to 0.988 indicating excellent reliability. Most of the images in iDose4 were graded excellent across the plain, arterial, and portovenous phases with respect to critical reproduction (89%, 90%, and 81%), visualization of large vessels (86%), and image contrast (93%, 96%, and 90%) as shown in Table 3. Also, there was no significant difference in image quality between Idose4 and Precise Image with respect to visualization of large vessels in the arterial and portovenous phases. Around 20% of the Idose4 images were graded 3 with respect to critical reproduction and approximately 0.06% graded 3 with regards to image contrast for the plain phase indicating acceptable sharpness and image contrast. Nevertheless, there was a significant improvement in image quality in these categories using Precise Image (Table 4). Precise Image was also graded better than iDose4 with respect to visualization/conspicuity, image noise, and artifacts as shown in Figures 2, 3, and 4. Most of the images in precise were graded 4 and 5 in all image quality criteria indicating better/excellent image quality.

Table 3.

Qualitative assessment of image quality scores for both raters in Idose4 and Precise Image.

Image quality criteria Idose4 Precise image
Plain Arterial PV Plain Arterial PV
Visualization/conspicuity 0/0/6/53/7 0/0/4/59/3 0/0/0/59/7 0/0/0/7/59 0/0/0/6/60 0/0/0/4/62
0/0/6/52/8 0/0/4/58/4 0/0/0/58/6 0/0/0/6/58 0/0/0/7/59 0/0/0/3/63
Critical reproduction 0/0/7/0/59 0/0/0/6/60 0/0/0/12/54 0/0/0/7/59 0/0/0/0/66 0/0/0/3/63
0/0/7/1/58 0/0/0/5/61 0/0/0/13/53 0/0/0/6/58 0/0/0/1/65 0/0/0/4/62
Visualization of large vessels 0/0/0/9/57 0/0/0/9/57 0/0/0/9/57 0/0/0/8/58 0/0/0/2/64 0/0/0/2/64
0/0/0/8/58 0/0/0/8/58 0/0/0/8/58 0/0/0/7/59 0/0/0/1/65 0/0/0/3/63
Image contrast 0/0/4/0/62 0/0/2/0/64 0/0/2/4/60 0/0/1/3/62 0/0/0/2/64 0/0/0/6/60
0/0/3/1/62 0/0/1/0/65 0/0/2/5/59 0/0/0/4/62 0/0/0/3/63 0/0/0/5/61
Image noise 0/0/9/57/0 0/0/9/57/0 0/0/9/57/0 0/0/1/8/57 0/0/0/9/57 0/0/0/13/53
0/0/11/55/0 0/0/10/58/0 0/0/8/58/0 0/0/0/9/57 0/0/0/10/58 0/0/0/12/54
Artifacts 0/4/56/6 0/4/56/6 0/9/57/0 0/0/10/56 0/0/9/57 0/0/13/52
0/3/57/6 0/3/57/6 0/8/56/0 0/0/9/57 0/0/8/58 0/0/12/53

Table 4.

Comparison of qualitative image quality parameters between Idose4 and Precise Image.

Qualitative parameters Plain Arterial Portovenous
Idose 4 Precise p-Value Idose 4 Precise p-Value Idose 4 Precise p-Value
Visualization/conspicuity 4 ± 0.4 4.8 ± 0.3 <0.001 4 ± 0.2 4.9 ± 0.1 <0.001 4.2 ± 0.3 4.9 ± 0.1 <0.001
Critical reproduction 4.7 ± 0.6 4.9 ± 0.3 <0.001 4.7 ± 0.6 4.9 ± 0.1 <0.001 4.7 ± 0.1 4.9 ± 0.1 <0.001
Visualization of large vessels 4.7 ± 0.3 4.9 ± 0.2 <0.001 4.8 ± 0.3 4.8 ± 0.08 0.973 4.8 ± 0.3 4.9 ± 0.3 0.08
Image contrast 4.7 ± 0.4 4.9 ± 0.2 <0.001 4.7 ± 0.1 4.9 ± 0.08 <0.001 4.7 ± 0.2 4.9 ± 0.1 <0.001
Image noise 3.8 ± 0.3 4.8 ± 0.3 <0.001 3.8 ± 0.3 4.8 ± 0.3 <0.001 3.8 ± 0.3 4.8 ± 0.3 <0.001
Artifacts 3 ± 0.3 3.4 ± 0.1 <0.001 2.8 ± 0.3 3.4 ± 0.1 <0.001 2.8 ± 0.3 3.4 ± 0.2 <0.001

Figure 2.

Figure 2.

Idose4 (a) versus Precise Image (b) for unenhanced abdomen and pelvis CT.

Figure 3.

Figure 3.

Idose4 (a) versus Precise Image (b) for arterial phase abdomen and pelvis CT.

Figure 4.

Figure 4.

Idose4 (a) versus Precise Image (b) for portal venous phase abdomen and pelvis CT.

Discussion

A lot of studies in the literature shows improvement in image quality using AI-based reconstruction techniques.8,9,1520 Some of these studies have been done on phantom and results may have to be validated with clinical studies.15,16 Different manufacturers have introduced their own deep learning algorithms that make use of deep CNNs into the image reconstruction process. A majority of studies reported, have evaluated image quality using the TrueFidelity by GE healthcare and AiCE by canon medical systems. The studies reported an improvement in SNR and CNR compared to IR, hybrid reconstruction MBIR techniques. 19 Recently, Philips healthcare introduced their DLR named Precise Image that has been tested on both phantoms and patients to provide improved image quality. Phantom study conducted by Greffier et al. 15 reported a reduction in noise using standard and smoother levels. Similar studies conducted by Greffier et al. 7 reported a significant improvement in the overall image quality subjectively for liver metastasis conspicuity using smooth and smoother levels. The results of the present study also reported a better image quality both subjectively and objectively in terms of SNR, CNR, and reduced noise using DLR technique for multiphase CT abdomen and pelvis studies. However, a unique aspect of this study was that a low kVp was implemented in all three phases of the CECT abdomen and pelvis as opposed to 120 kVp used in all other studies reported in literature.8,9,1519 In this regard, an image quality improvement is possible at low doses when a lower kVp is used in conjunction with precise imaging. Also, in all the studies reported, a wide range of body sizes were included, since body size is an important parameter influencing the absorption and distribution of radiation in the body and can therefore affect the image quality, the present study included only patient with a normal body mass index. Although, Tamura et al. 21 found no differences in noise across the BMI groups, the results cannot be generalized to other vendors since the DLR algorithm used in their study was AiCE, and the present study used precise imaging.

In the present study, image noise was found to be significantly reduced using Precise Image. Similarly, SNR and CNR was significantly improved for all organs in all the phases. It was also observed that the maximum noise reduction and improvement of SNR and CNR occurred in the postcontrast phases (arterial and portovenous) compared to the precontrast phase. This significant improvement in image quality in contrast-enhanced images may be attributed to better differentiation of anatomical structures that allows DLR models to effectively identify and reduce noise. Tables 5 and 6 summarizes the image noise, and SNR and CNR values obtained in the present study and other studies reported in the literature that have used DLR technique for CT abdomen and pelvis. The results of the present study were comparable with the other studies reported. These variations can be explained due to the difference in exposure parameter used, protocol variations like scan delay, slice thickness.

Table 5.

Comparison of image noise values in present and other studies reported in literature.

Technical parameters/scan series Present study Singh et al. 18 Akagi et al. 19 Ichikawa et al. 9 Jensen et al. 17
DLR algorithm Precise image AiCE AiCE TrueFidelity TrueFidelity (DLIR-Low)
kVp 100 120 120 120 120
Plain 9.1 5.2
Arterial 14.3 13.9
Porto venous 14.7 11 14.6 9.9

Table 6.

Comparison of CNR and SNR values in present and other studies reported in literature.

Organs Present study Ichikawa et al. 9 Jensen et al. 17 Singh et al. 18 Akagi et al. 19
Scan Series CNR SNR CNR SNR CNR SNR CNR SNR CNR SNR
Plain Aorta 0.82 4.5
Liver 0.35 7.03 9.6
Arterial Aorta 38.57 36.2 19.9
Liver 0.74 7.6 1 1.2
Porto venous Aorta 12.1 15.8 15 14.3 3.3
Liver 5.8 8.6 6.54 9.9 2
Spleen 6.9 14 9.3
Portal vein 12.9 15.4 3.4

Also, there was also a difference on how image noise was calculated in different studies. The present study estimated the image noise from the mean of standard deviation of all organs (aorta, liver, portal vein, and spleen), whereas study conducted by Akagi et al. 19 estimated noise from standard deviation measured in paraspinal muscle. Similarly, Jensen et al. 17 estimated the image noise from standard deviation of subcutaneous fat. A few studies estimated image noise from the standard deviation in liver parenchyma.9,18 Other factors like type of DLR used and strength of DLR can also influence noise reduction ability and noise texture.

This study has some limitations. First, only the standard filter of Precise Image DLR technique was considered, and other filters such as “smooth” and “smoother” were not compared. Second, although qualitative analysis showed promising results, prospective studies including different pathology as per clinical indication can be done to improve the clinical benefit. Also, evaluation for lesion conspicuity between both reconstruction algorithms were not performed due to limited cases. Therefore, future studies can be done to evaluate the performance of precise image in lesion conspicuity. The present study included only normal BMI individuals, since body size can be important parameter to determine image quality, further studies on different BMI ranges can be explored. The present study could only objectively evaluate the image quality indexes like SNR, CNR, and image noise. A more detailed method would include evaluating noise textures that correlates between adjacent pixels that influence the detectability of certain pathologies ideal for nonlinear algorithms like IR and DLR.22,23 Noise texture in the images can be calculated using the noise power spectrum by plotting the amplitudes with the frequency of image using Fourier transform. 24 Considering this, further studies can be conducted to assess the effectiveness of DLR technique in different body sized individuals. Nevertheless, the use of widely popular image matrices such as SNR, CNR, and Image noise are still used frequently to evaluate the image quality, compare performance of modalities and quantitative evaluation of lesion detectability as it can be measured easily.21,25,26

Conclusion

Compared to Idose4, the Precise Image reconstruction algorithm significantly decreased image noise and improved SNR and CNR. Qualitative analysis also reported favorable to the Precise Image reconstruction. Therefore, precise imaging may be useful in improving the image quality in low-dose abdomen and pelvis CT among normal BMI individuals.

Acknowledgments

The authors would like to acknowledge, Kasturba Hospital and Manipal College of Health Professions for their support to undertake this study. We also extend our heartfelt gratitude to the Department of biostatistics for their assistance in the study.

Footnotes

ORCID iD: Nitika C. Panakkal Inline graphic https://orcid.org/0000-0003-3725-6774

Ethical considerations: Ethical approval for this study was obtained from *Institutional Ethics Committee, Kasturba Hospital (IEC: 182/2023)*.

Consent to participate: Written informed consent was obtained from all subjects before the study.

Author contributions: Conceptualization: Nitika C. Panakkal. Methodology: Thejas M. S., Nitika C. Panakkal. Formal analysis and investigation: Rajagopal Kadavigere, Tanushree R. Kamath, and Shailesh Nayak. Investigation: Thejas M. S. Writing original Draft: Thejas M. S. Writing Review and editing: Nitika C. Panakkal and Shailesh Nayak. Supervision: Nitika C. Panakkal, Rajagopal Kadavigere, and Shailesh Nayak, and Suresh Sukumar.

Funding: The author(s) received no financial support for the research, authorship, and/or publication of this article.

The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

References

  • 1. Zhang Z, Seeram E. The use of artificial intelligence in computed tomography image reconstruction—a literature review. J Med Imaging Radiat Sci 2020; 51(4): 671–677. [DOI] [PubMed] [Google Scholar]
  • 2. Den Harder AM, Willemink MJ, Budde RP, et al. Hybrid and model-based iterative reconstruction techniques for pediatric CT. AJR Am J Roentgenol 2015; 204(3): 645–653. [DOI] [PubMed] [Google Scholar]
  • 3. Willemink MJ, Noël PB. The evolution of image reconstruction for CT-from filtered back projection to artificial intelligence. Eur Radiol 2019; 29(5): 2185–2195. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Willemink MJ, Leiner T, Jong PA, et al. Iterative reconstruction techniques for computed tomography part 2: initial results in dose reduction and image quality. Eur Radiol 2013; 23(6): 1632–1642. [DOI] [PubMed] [Google Scholar]
  • 5. McLeavy CM, Chunara MH, Gravell RJ, et al. The future of CT: deep learning reconstruction. Clin Radiol 2021; 76(6): 407–415. [DOI] [PubMed] [Google Scholar]
  • 6. Funama Y, Taguchi K, Utsunomiya D, et al. Combination of a low-tube-voltage technique with hybrid iterative reconstruction (iDose) algorithm at coronary computed tomographic angiography. J Comput Assist Tomogr 2011; 35(4): 480–485. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Greffier J, Durand Q, Serrand C, et al. First results of a new deep learning reconstruction algorithm on image quality and liver metastasis conspicuity for abdominal low-dose CT. Diagnostics (Basel) 2023; 13(6): 1182. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Njølstad T, Schulz A, Godt JC, et al. Improved image quality in abdominal computed tomography reconstructed with a novel Deep Learning Image Reconstruction technique—initial clinical experience. Acta Radiol Open 2021; 10(4): 20584601211008391. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Ichikawa Y, Kanii Y, Yamazaki A, et al. Deep learning image reconstruction for improvement of image quality of abdominal computed tomography: comparison with hybrid iterative reconstruction. Jpn J Radiol 2021; 39(6): 598–604. [DOI] [PubMed] [Google Scholar]
  • 10. Panakkal NC, Rajagopal K, Suresh S, et al. Low kilovoltage protocol based on body mass index: feasibility and influence on radiation dose and image quality. J Med Imaging Radiat Sci 2023; 54(3): S23. [Google Scholar]
  • 11. Benbow M, Bull RK. Simple weight-based contrast dosing for standardization of portal phase CT liver enhancement. Clin Radiol 2011; 66(10): 940–944. [DOI] [PubMed] [Google Scholar]
  • 12. Perrin E, Jackson M, Grant R, et al. Weight adapted iodinated contrast media administration in abdomino-pelvic CT: can image quality be maintained? Radiography 2018; 24(1): 22–27. [DOI] [PubMed] [Google Scholar]
  • 13. Buls N, Van Gompel G, Van Cauteren T, et al. Contrast agent and radiation dose reduction in abdominal CT by a combination of low tube voltage and advanced image reconstruction algorithms. Eur Radiol 2015; 25(4): 1023–1031. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Jurik AG, Bongartz G, Golding SJ, et al. The quality criteria for computed tomography (invited paper). Radiat Protect Dosim 1998; 80(1–3): 49–53. [Google Scholar]
  • 15. Greffier J, Durand Q, Frandon J, et al. Improved image quality and dose reduction in abdominal CT with deep-learning reconstruction algorithm: a phantom study. Eur Radiol 2023; 33(1): 699–710. [DOI] [PubMed] [Google Scholar]
  • 16. Greffier J, Dabli D, Hamard A, et al. Effect of a new deep learning image reconstruction algorithm for abdominal computed tomography imaging on image quality and dose reduction compared with two iterative reconstruction algorithms: a phantom study. Quant Imaging Med Surg 2022; 12(1): 229–243. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Jensen CT, Liu X, Tamm EP, et al. Image quality assessment of abdominal CT by use of new deep learning image reconstruction: initial experience. AJR Am J Roentgenol 2020; 215(1): 50–57. [DOI] [PubMed] [Google Scholar]
  • 18. Singh R, Digumarthy SR, Muse VV, et al. Image quality and lesion detection on deep learning reconstruction and iterative reconstruction of Submillisievert chest and abdominal CT. AJR Am J Roentgenol 2020; 214(3): 566–573. [DOI] [PubMed] [Google Scholar]
  • 19. Akagi M, Nakamura Y, Higaki T, et al. Correction to: Deep learning reconstruction improves image quality of abdominal ultra-high-resolution CT. Eur Radiol 2019; 29(8): 4526–4527. [DOI] [PubMed] [Google Scholar]
  • 20. Arndt C, Güttler F, Heinrich A, et al. Deep learning CT image reconstruction in clinical practice. Rofo 2021; 193(3): 252–261 [DOI] [PubMed] [Google Scholar]
  • 21. Tamura A, Mukaida E, Ota Y, et al. Superior objective and subjective image quality of deep learning reconstruction for low-dose abdominal CT imaging in comparison with model-based iterative reconstruction and filtered back projection. Br J Radiol 2021; 94(1123): 20201357. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Solomon JB, Christianson O, Samei E. Quantitative comparison of noise texture across CT scanners from different manufacturers. Med Phys 2012; 39(10): 6048–6055. [DOI] [PubMed] [Google Scholar]
  • 23. Boedeker KL, Cooper VN, McNitt-Gray MF. Application of the noise power spectrum in modern diagnostic MDCT: part I. Measurement of noise power spectra and noise equivalent quanta. Phys Med Biol 2007; 52(14): 4027–4046. [DOI] [PubMed] [Google Scholar]
  • 24. Verdun FR, Racine D, Ott JG, et al. Image quality in CT: from physical measurements to model observers. Phys Med 2015; 31(8): 823–843. [DOI] [PubMed] [Google Scholar]
  • 25. Igarashi K, Imai K, Matsushima S, et al. Development and validation of the effective CNR analysis method for evaluating the contrast resolution of CT images. Phys Eng Sci Med 2024; 47(2): 717–727. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Singh S, Kalra K, Hsieh J, et al. Abdominal CT: comparison of adaptive statistical iterative and filtered back projection reconstruction techniques. Radiology 2010; 257: 373–383. [DOI] [PubMed] [Google Scholar]