Abstract
Introduction
[68Ga-DOTA-D-Phe1-Tyr3]octreotide ([68Ga]Ga-DOTA-TOC) is a somatostatin analogue largely used in PET/CT applications for the detection of gastroenteropancreatic neuroendocrine tumors (GEP-NET). Initially, it was obtained using a 68Ge/68Ga generator. The increasing cost of good manufacturing practice-compliant generators has led to the need to find alternative ways of producing Gallium-68 (68Ga). The aim of this work is to show the production optimization of [68Ga]Ga-DOTA-TOC via cyclotron, derived from three years of experience.
Methods
The production of [68Ga]GaCl3 via the 68Zn(p,n)68Ga reaction was optimized using a PETtrace 800 cyclotron (equipped with ZnO liquid target) and the synthesis of [68Ga]Ga-DOTA-TOC was performed by FASTlab2 developer system according to the Guidelines on Good Radiopharmacy Practice (cGRPP). Quality control process was validated according to the current specific monograph (2482) of the European Pharmacopoeia (Ph. Eur.).
Results
[68Ga]Ga-DOTA-TOC was produced in 40 minutes; ten validation batches met the quality criteria expected by the Ph. Eur. The synthesis process has involved many issues due to the use of acidic reagents and related corrosion of some components of cyclotron and developer system, resulting in 12.2% failed syntheses and a target breakdown after 11 months.
Discussion
The main issues, their causes and the strategies used to solve them are reported in the troubleshooting section: thanks to these strategies, the number of failed syntheses has decreased, and today, we have achieved a 0% failure rate.
Conclusion
Liquid target production of [68Ga]Ga-DOTA-TOC, once consolidated, instead of 68Ge/68Ga generator has many advantages.
Keywords: Cyclotron, liquid target, gastroenteropancreatic neuroendocrine tumors, [68Ga]Ga-DOTA-TOC, generator, positron emission tomography, radiopharmaceutical, european pharmacopeia
1. INTRODUCTION
In recent years, positron emission tomography/computed tomography (PET/CT) with Gallium68-peptides has become essential in the characterization, staging, and definition of a therapy approach for gastroenteropancreatic neuroendocrine tumors (GEP-NET) with low or intermediate malignancy grades [1]. The diagnosis of neuroendocrine tumors (NET) often occurs late because traditional imaging with [18F]-FDG ([18F] fluoro-2-deoxy-D-glucose) has limitations, but with the advent of 68Ga peptides, such as [68Ga-DOTA-D-Phe1-Tyr3]octreotide ([68Ga]Ga-DOTA-TOC) [2]/[68Ga-DOTA-D-Phe1-Tyr3]octreotate ([68Ga]Ga-DOTA-TATE) [3]/[68Ga-DOTA-D-Phe1-1NaI3]octreotide ([68Ga]Ga-DOTA-NOC) [4], it has became possible to identify the primary tumor, assess the presence of small distant metastases, and guide patients toward targeted therapies [2].
Somatostatin peptides are labeled with Gallium-68, a positron-emitting isotope with a half-life of 67.629 minutes: the positron (β+) produces energy in the form of two coincident 0.511 MeV γ rays [5]. The peptide binding with the 68Ga isotope is achieved using the organic compound “1,4,7,10-tetraazacyclododecane-1,4,7,10-tetraacetic acid (DOTA)”, a chelator that forms a stable complex with 68Ga. The main peptides conjugated with [68Ga]Ga-DOTA are analogs of octreotide, an octapeptide similar to somatostatin that can bind to somatostatin receptors (SSTRs) with greater affinity than the endogenous hormone.
The main ones used in nuclear medicine are:
[68Ga]Ga-DOTA-TATE
[68Ga]Ga-DOTA-TOC
[68Ga]Ga-DOTA-NOC
They differ from each other in their affinity for a specific receptor. In particular: [68Ga]Ga-DOTA-TATE has an affinity for SSTR2 (highly expressed in NETs); [68Ga]Ga-DOTA-NOC has a good affinity for SSTR3 and SSTR5, while [68Ga]Ga-DOTA-TOC has an affinity for tumors that overexpress somatostatin receptors SSTR2 and SSTR5 [6, 7], such as low-grade gastroenteropancreatic neuroendocrine tumors (GEP-NET), pheochromocytomas, and paragangliomas [1].
This work focuses on the implementation of production and quality control of [68Ga]Ga-DOTA-TOC at the nuclear medicine department of Spedali Civili (Brescia, IT) in accordance with the Guidelines on Good Radiopharmacy Practice (cGRPP) [8] issued by the Radiopharmacy Committee of the European Association of Nuclear Medicine (EANM) [9].
DOTATOC [DOTA(0)-Phe(1)-Tyr(3)] edotreotide is a peptide with significant utility both in therapy and imaging: it consists of the chelator DOTA covalently linked to edotreotide via a peptide bond (the substitution of Phe with Tyr at position 3 increases the stability of the compound) and binds the 68Ga radionuclide (as shown in Fig. 1) [6].
Fig. (1).

[68Ga]Ga-DOTATOC: in green the chelator DOTA, in red Gallium-68 isotope and in black the structure of edotreotide (peptide) in which there is in position 3 a Tyrosine instead of a Phenilalanine. (https://www.mdpi.com/1424-8247/13/3/38).
Gallium-68 radionuclide is a largely used isotope for diagnostic applications in nuclear medicine due to its affinity for binding various biomolecular vectors using bifunctional chelators and various macromolecules with rapid pharmacokinetic profiles, such as peptides and peptidomimetics [10].
68Ga isotope can be obtained in the form of [68Ga]GaCl3 from elution of a 1.85 GBq (50mCi) 68Ge/68Ga generator (Eckert & Ziegler, Germany) using dilute hydrochloric acid (HCl 0.1M - Eckert & Ziegler, Germany). [68Ga]GaCl3 from the generator is used to label somatostatin analogue peptides by different devices and also to label lyophilized products, thus obtaining [68Ga]Ga-DOTA-TOC (edotreotide) for human use [11]. The use of a generator is the only alternative for centers that don’t use a cyclotron. There are many reasons why we have given up using the 68Ge/68Ga generator, which has led to production of [68Ga]GaCl3 using a cyclotron: increasingly high costs, 12 months of shelf life of a 68Ge/68Ga generator, the limited half-life of the 68Ge isotope (270.95 days) and the inclusion in the monographs of the European Pharmacopoeia (Ph. Eur.) of Gallium (68Ga) chloride (accelerator-produced) solution for radiolabelling (3109), Gallium (68Ga) edotreotide injection (2482) and Gallium (68Ga) PSMA-11 injection (3044) obtained by cyclotron [12].
[68Ga]Ga-PSMA-11 was produced for staging and follow-up of prostate cancers [13] (for which we have a large case series), while [68Ga]Ga-DOTA-TOC, as previously mentioned, was used for neuroendocrine tumors; however, in this work we will focus only on the transition of [68Ga]Ga-DOTA-TOC production from the 68Ge/68Ga generator to the cyclotron.
Gallium-68 radionuclide can be obtained by proton irradiation of an enriched Zn68 target solution in an accelerator followed by isolation of Gallium-68 in an acidic solution (pH<2.0).
The production of [68Ga]GaCl3 by cyclotron is possible via the 68Zn(p,n)68Ga reaction using a liquid or a solid target: in this work Gallium-68 was obtained by cyclotron using a liquid target (ZnO 1M in HNO3 0.3M) [14, 15]; a cassette developer system was used for [68Ga]GaCl3 purification and [68Ga]Ga-DOTA-TOC labelling.
Quality control of [68Ga]Ga-DOTA-TOC produced was verified in accordance with the current specific monograph (2482) of Ph. Eur., including pH, half-life, radionuclidic purity, chemical identity and radiochemical purity, bacterial endotoxins and sterility [12].
[68Ga]Ga-DOTA-TOC synthesis has involved many issues due to the use of acidic reagents and related corrosion of some components of the cyclotron and the developer system: the aim of this work is to solve these problems, derived from an experience of over three years.
2. MATERIALS AND METHODS
2.1. The Past: Production of [68Ga]Ga-DOTA-TOC by 68Ge/68Ga Generator
[68Ga]Ga-DOTATOC was initially produced only by 68Ge/68Ga generator from May 2018 (month of purchase of the first generator) to December 2020 (a total of 3 generators - 32 months - 170 productions).
The 1.85 GBq (50 mCi) 68Ge/68Ga GalliaPharm Generator (Eckert & Ziegler, Germany) was a GMP-compliant pharmaceutical-grade generator [16]. This generator consisted of a borosilicate glass column containing a bed of titanium dioxide on which 68Ge (parent radionuclide, half-life of 270.95 days) was adsorbed: 68Ga (daughter radionuclide, half-life of 67.71 minutes) was continuously produced through the decay of its parent and eluted using 0.1 M hydrochloric acid (HCl - Advance Accelerator Applications, Italy), ultra-pure and sterile acid [17]. This system was designed to minimize the breakthrough of 68Ge and metallic impurities in accordance with monograph (2464) of the current Ph. Eur. [12]. Activity eluted at the beginning of generator validity was certified to be not less than 1.11GBq and activity eluted at the end of validity (1 year) was certified to be not less than 0.42GBq, as reported in Table 1.
Table 1.
Activity on generator and activity from elution. Shelf life of the generator: 1 year.
| Activity (GBq) | Activity (GBq) Inside of Generator at the Beginning of Validity | Activity (GBq) Inside of Generator at the End of Validity | Activity (GBq) Eluted at the Beginning of Validity | Activity (GBq) Eluted at the End of Validity |
|---|---|---|---|---|
| 1,85 | 1,85 ± 10% | 0,7 ± 10% | NLT * 1,11 | NLT* 0,42 |
Note: *NLT (not less than).
2.1.1. Labeling Procedure
Radiopharmaceuticals must be prepared by healthcare professionals who have received specific training and have experience in the safe use and handling of radionuclides. Following the Somakit TOC® SmPC (Summary of Product Characteristics), 5 mL of [68Ga]GaCl3 was eluted through metal-free materials into a Somakit TOC® (40 mg of edotreotide) vial (Advance Accelerator Applications, Italy), [18]; the vial was then placed in a thermoblock at 95°C for 7 minutes. At the end of the process, 0.5 mL of reaction buffer (formic acid, sodium hydroxide, H2O - Advance Accelerator Applications, Italy) was added to the vial [17-19]. [68Ga]Ga-DOTATOC activity produced was measured using a dose calibrator (COMECER, Italy) and the radiopharmaceutical was verified with quality control before administration to the patient. All these processes were performed following the cGRPP in a class A shielded laminar flow isolator (COMECER, Italy) to ensure the aseptic process and the radioprotection of the operator.
2.1.2. Quality Control Procedure
Somakit TOC® quality control was performed as described in the related SmPC (Summary of Product Characteristics) [17]: appearance (clear solution, free of visible particulates); pH (3.2-3.8) using pH indicator strips (VWR Chemicals, Italy, range 1.0-4.3) radiochemical purity through two thin-layer chromatographies (glass fiber ITLC - Celltech, Italy). The first chromatography used sodium citrate 0.1 M (Sigma Aldrich, USA) as the mobile phase and the second used a mixture of 77g·L-1 ammonium acetate (Sigma Aldrich, USA) in H2O/methanol (Sigma Aldrich, USA) (50:50 v/v), followed by analysis with a radio-TLC scanner (Elysia-Raytest, Germany) to verify labelling efficiency: presence of colloidal 68Ga ≤3%, presence of free 68Ga ≤2%.
2.2. The Present: Cyclotron production of 68Ga, Purification of [68Ga]GaCl3 and Synthesis of [68Ga]Ga-DOTA-TOC
From January 2021 to present, [68Ga]Ga-DOTATOC was produced only by cyclotron.
The GE PETtrace 800 series cyclotron (Uppsala, Sweden) for medical use produced radionuclide Gallium-68, a 68Ga liquid target with the following characteristics: small target volume (2 mL), water cooling only, 200 mm aluminium degrader foil, stacked niobium (25 mm)/Havar (25 mm) foils. GE PETtrace 800 cyclotron provided a proton beam with a maximum energy of 16.5 MeV and maximum current of 100 mA on target [20]. Preparation of the target solution: Isotopically enriched [68Zn]ZnO (665 mg·vial-1, ≥ 98.2% enriched - ISOFLEX, San Francisco, California USA - Table 2) was handled by qualified personnel to prepare [68Zn]ZnO 1M in HNO3 0.3M ensuring the use of metal-free materials to avoid compromising the final result [20, 21].
Table 2.
Physical characteristics of [68Zn]ZnO (ISOFLEX, San Francisco, California USA).
| Description | Isotopic Distribution | Chemical Admixtures | |||||
|---|---|---|---|---|---|---|---|
| Isotope | Zn-68 | Isotope | Content (%) | Element | Content (ppm) | Element | Content (ppm) |
| Enrichment | 98.2% | Zn-64 | 0.1 | Al | <1.0 | Mg | 0.25 |
| Element weight | 665 mg | Zn-66 | 0.180 | As | <1.0 | Mn | <0.1 |
| Form | Oxide (ZnO) | Zn-67 | 0.963 | Ca | 6.6 | Pb | 2.1 |
| - | - | Zn-68 | 98.2 | Cd | <0.1 | Si | 6.8 |
| - | - | Zn-70 | 0.557 | Co | <0.1 | Sn | 53 |
| - | - | - | - | Cu | 1.5 | Na | <1.0 |
| - | Fe | <1.0 | - | - | |||
Note: Data from Certificate of Analysis of [68Zn]ZnO (ISOFLEX).
2.2.1. Irradiation of [68Zn]ZnO 1M in 0.3M HNO3 and Production of 68Ga
Direct production of 68Ga using a liquid target is based on the reaction 68Zn(p,n)68Ga: optimization was achieved with a [68Zn]ZnO 1M solution in 0.3M HNO3 (HNO3 70% ultrapure, Sigma Aldrich, USA) using a 35 μA proton beam for about 60 minutes. During irradiation, two other nuclear reactions occur, which lead to the synthesis of different gallium isotopes, 66Zn (p,n)66Ga and 68Zn (p,2n)67Ga, producing 66Ga (half-life 9.48 h) and 67Ga (half-life 78 h). In particular, the production of 66Ga is related to the percentage of 66Zn in the isotopically enriched [68Zn]ZnO, while the formation of 67Ga depends on 67Zn in the isotopically enriched [68Zn]ZnO and the beam time: increasing the beam time increased the percentage of impurities [21]. In order to prevent cross contamination between the F-18 and Ga-68 production lines, the transfer lines from their respective targets have been completely separated; this way also avoids the possible presence of metal contamination in the Ga-68 production. For optimal 68Ga target performance, a conditioning run was completed daily with a short irradiation cycle of 0.6M HNO3 to prevent zinc precipitates (proton beam of 15 minutes at a current of 25 μA) [21-23].
2.2.2. [68Ga]GaCl3 Purification and [68Ga]Ga-DOTA-TOC Production
A cassette developer system (GE HealthCare, Chicago, Illinois, USA) was used, in which [68Ga]GaCl3 was purified and, sequentially, [68Ga]Ga-DOTA-TOC was produced. The same cassette was used for purification and production steps (Fig. 2); every cassette system was hand-assembled at the moment using the materials reported in Table 3 [24].
Fig. (2).

(A) Cassette developer system, in particular: (A1) External receiving vial Taddeo, (A2) Zn recovery vial, (A3) Product vial; (B) synthesis cassette: the highlighted part on the right (B2) contains the reagents and cartridges for the purification of [68Ga]GaCl3, while the highlighted part on the left (B1) contains the reagents for the production of [68Ga]Ga-DOTATOC.
Table 3.
List of consumables (on the left column) and reagents (on the right column) for developer system.
| Consumables Ready to Use, Handly Assembled on the Developer System | Reagents Prepared in the Laboratory using Metal-Free Materials |
|---|---|
| • Developer system (GE HealthCare, Chicago, Illinois USA): cassette cover, scheleton and tubing system, reactor, spike for water bag and vials for reagents | ➢ 3.1 mL vial of ethanol (ultrapure, Merck Life Science) |
| • Water bag (BBraun for injecting): store at 4°C to ensure efficient C18 purification | ➢ 4 mL vial of HNO3 0.6M (HNO3 70%, ultrapure, Sigma Aldrich) |
| • 2 C18 cartridges (Waters) | ➢ 4 mL vial of HCl 4M (HCl 30%, ultrapure, Merck Life Science) |
| • 1 ZR resin, 2 mL (hydroxamate-based resin, Triskem) • 1 TK200 resin, 2 mL (trioctylphosphine oxide, TOPO-based resin, Triskem) • 1 SAX resin, 1 mL (strong anion exchange resin, Triskem) |
➢ 4 mL vial of NaCl 3M (NaCl, ultrapure, Merck Life Science) |
| • 1 external vial receiving Taddeo (Comecer) | ➢ 1 vial precursor containing: 50mg Ga-DOTATOC acetate (ABX, Advanced Biochemical Compound) dissolved in 1.2 mL solution of sodium acetate (1M, pH 4.5, Merck Life Science) and 100 mL of ascorbic acid (ultrapure, Merck Life Science) |
[68Ga]GaCl3 purification step was based on a three-column approach: 68Ga solution was transferred into an external receiving vial Taddeo, then into a C18 cartridge to purify contaminants derived from irradiation, such as 13N. Subsequently, it was trapped on the first column, a hydroxamate resin (ZR) (previously conditioned with 0.1 M HNO3 - Merck Life Science, Italy) for purification of 68Ga via the irradiation of Zn68 on a cyclotron: the zinc was not trapped on the column and residual zinc was washed off the column with 0.1M HNO3. 68Ga was eluted from ZR resin with HCl 1.75M (Merck Life Science, Italy), passed through a strong anion exchange resin (SAX, previously conditioned with 1.75 M HCl) that removed organic molecules remaining from the manufacturing processes and then trapped onto a third column, a trioctylphosphine oxide, TOPO-based resin (TK200 previously conditioned with 1.75 M HCl), an extractant used in the extraction of various metal ions. This column was washed with 0.1M HCl and 3M NaCl solution: at this step 68Ga remained anchored on the column, subsequently, it was eluted from the column into the reactor with water, followed by diluted HCl to formulate [25]. [68Ga]GaCl3 purified by this system was directly comparable to commercially available 68Ge/68Ga generators and was compatible with formulations required for pharmaceutical cold kit labeling [23].
[68Ga]Ga-DOTA-TOC production step: before the [68Ga]GaCl3 purification, the reactor was pre-loaded with precursor (50 mg Ga-DOTATOC acetate - ABX, Germany, in a solution of sodium acetate and ascorbic acid - Merck Life Science, Italy); the step of labelling was performed at 95°C for 10 minutes, finally followed a purification step by C18, to yield the [68Ga]Ga-DOTA-TOC. The final product (4mL) was filtered using a 0.22mm filter unit (Millex-SG - Merck Life Science, Italy) into a vial containing 8 mL of 10 mM of PBS pH 7.4 (VWR Chemicals, Italy) [24]. [68Ga]GaCl3 purification and [68Ga]Ga-DOTA-TOC synthesis steps took approximately 40 minutes. All these processes were performed in a class A shielded laminar flow isolator (COMECER, Italy) to ensure the aseptic process and the radioprotection of the operator, while the assembly of the components of the cassette and synthesis reagents was prepared in a laminar flow hood placed in a class D environment.
2.2.3. [68Ga]Ga-DOTA-TOC Quality Control
Acceptance criteria, specifications and release timing were chosen in compliance with the current general texts and monograph (2482) of the Ph. Eur. [12]. pH was verified using pH indicator strips (VWR Chemicals, Italy, increment 0.5 pH unit); radionuclidic purity by gamma-ray spectrometry (GammaVision - Ortec, Illinois, USA) with principal gamma photons of 0.511 MeV and 1.077 MeV; liquid chromatography was performed on an Ultimate 3000 system (HPLC: High Performance Liquid Chromatography): equipped with a UV variable wavelength detector RS300 (Thermo Fischer Scientific, Germany) and a radiometric detector (GABI, Raytest, Germany). The system was controlled by Chromeleon software version 7.2 SR5 (Dionex Sunnyvale, CA, USA). The column was a base-deactivated end-capped octadecylsilyl silica gel for chromatography (3 μm), 0.15 m - 3.0 mm (Thermo Fisher Scientific, Germany). An isocratic application was performed using a solvent mixture (trifluoroacetic acid from Carlo Erba Reagents, IT/water from Milli-Q Millipore system/ acetonitrile from Sigma Aldrich, USA - 1:780:220 V/V) with a flow rate set at 0.6 mL·min-1, UV wavelength at 220 nm, column oven at 25°C. Thin-layer chromatography was performed using a TLC silica gel plate (glass-fibre plate 5x10cm, Merck Life Science, Italy). The solvent for development of TLC plate was a 77g·L−1 solution of ammonium acetate in water (Sigma Aldrich, USA), methanol (Sigma Aldrich) (50:50 V/V); plate was developed over 2/3 of its length and was then analyzed with a radio-TLC scanner (Elysia-Raytest, Germany). Determination of residual solvent (ethanol) in the final formulation was carried out by gas chromatography (6850 Series II, Agilent, USA) controlled by Raytest Iberica software (Barcelona, Spain): the column was a GC Column “5cg HP-Fast GC Residual Solvent Column” 30m, 0.53mm, 1.0um (Agilent Technologies, Germany), carrier gas was nitrogen, the flame ionization detector (FID) was set at 300°C and oven temperature was programmed from 33°C to 50°C in 7 min. Bacterial endotoxins were investigated by Endosafe Nexgen-PTS technology (Charles River, Massachusetts, USA) on a 1:40 diluted sample. The sterility instead was verified by an external service (ISZLER: Experimental Zooprophylactic Institute of Lombardia and Emilia-Romagna): the radiopharmaceutical was released for human use before the completion of this test [12].
3. RESULTS
Data presented in this section refer to the production and quality control of [68Ga]Ga-DOTA-TOC, initially derived from [68Ga]GaCl3 from a 68Ge/68Ga generator (170 productions) and later derived from a cyclotron (309 productions). Data were evaluated by descriptive statistical analysis and Student’s t-Test to independent variables.
3.1. [68Ga]Ga-DOTA-TOC by 68Ge/68Ga Generator (1.85 Gbq)
Without dwelling too much on this initial data (the aim of this work is focused on the production of [68Ga]Ga-DOTA-TOC via cyclotron), in Table 4 we show the data of [68Ga]Ga-DOTATOC using [68Ga]GaCl3 derived from a 68Ge/68Ga generator: specifically, the data related to Generator 1 refer to the year 2018 (from May to December), Generator 2 to the year 2019 (from January to December), and Generator 3 to the year 2020 (from January to December), as production shifted to the cyclotron since 2021. As shown in the table on the data of 3 generators used (years 2018, 2019, 2020), the activity (MBq) of [68Ga]DOTATOC ranges from a maximum of 1050 MBq to a minimum of 359 MBq for the first generator, a maximum of 1240 MBq to a minimum of 473 MBq for the second generator, and a maximum of 1340 MBq to a minimum of 457MBq for the third generator. These data comply with the manufacturer's specifications for the generator (Eckert & Ziegler, Germany): activity eluted was a maximum average of 1210 MBq at the beginning of the generator's shelf life and a minimum average of 429 MBq at the end of the shelf life (1 year). Table 5 shows the percentage of failed [68Ga]Ga-DOTA-TOC from generator. In 2018, we observed a 2.9%, in 2019, it was around 2.5%, and in 2020 the percentage was 5.3%.
Table 4.
Activity of [68Ga]Ga-DOTATOC labelling with 68GaCl3 from 68Ge/ 68Ga generator.
| - | Generator 1 68Ga-DOTATOC | Generator 2 68Ga-DOTATOC | Generator 3 68Ga-DOTATOC |
|---|---|---|---|
| n | 33 | 78 | 53 |
| Mean (MBq) | 686 | 913 | 902 |
| Standard deviation (%) | 23.04 | 21.14 | 24.61 |
| Minimum (MBq) | 359 | 473 | 457 |
| Maximum (MBq) | 1050 | 1240 | 1340 |
Note: use of the first generator started on May 2018.
Table 5.
Percentage of failed [68Ga]Ga-DOTATOC labelling during the years (2018-2020).
| Year | 68Ga-DOTATOC Labelling | Total | ||
|---|---|---|---|---|
| Failed | Completed | % Failed | ||
| 2018 | 1 | 33 | 2.9 | 34 |
| 2019 | 2 | 78 | 2.5 | 80 |
| 2020 | 3 | 53 | 5.4 | 56 |
| Total | 6 | 164 | 3.5 | 170 |
Note: use of the first generator started on May 2018.
3.2. Validation of the [68Ga]Ga-DOTA-TOC Production with [68Ga]GaCl3 from Cyclotron using Target Liquid
The fully automated synthesis of [68Ga]Ga-DOTA-TOC using a developer system was completed in 40 min, from the first step of [68Ga]GaCl3 purification to the [68Ga]Ga-DOTA-TOC production step. Ten validation syntheses were performed to verify the critical steps: activity of 68Ga trapped in ZR resin and residual activity after purification; activity of [68Ga]GaCl3 transferred to the reactor for labeling; residual activity in the reactor; activity of [68Ga]Ga-DOTA-TOC produced and synthesis yield (Table 6). The radioactivity sent by the cyclotron at the end of the beam (EOB), at 35mA for 60 minutes, was 3996 Mbq (minimum: 3458 Mbq, maximun: 4502 MBq), the maximum activity produced in the ten validation syntheses of [68Ga]Ga-DOTA-TOC was 1725 MBq, and the minimum activity was 1206 MBq (Mean: 1448 MBq), with a decay-corrected yield for decay (time of synthesis: 40 minutes) from a minimum of 37.0% to a maximum of 53.0% (Mean: 44.5%). Table 7 shows the activities of [68Ga]Ga-DOTA-TOC obtained after validation’s phase: data refer to the period from 2021 to the first 6 months of 2024 (293 syntheses). As shown, the trend is very consistent and does not experience a negative inflection as it did with the 68Ge/68Ga generator (due to a short half-life of the generator). As shown in Table 7, the radioactivity sent by the cyclotron at the end of the beam (EOB), at 35±2 mA for 60 minutes, was a minimum average activity of 4077 MBq and a maximum of 4148 MBq: the minimum average activity produced of [68Ga]Ga-DOTA-TOC was 1484 MBq, and the maximum was 1697 MBq (Mean: 1597.5 MBq), with a decay-corrected yield (time of synthesis: 40 minutes) from an average minimum of 45.1% to a maximum of 52.5% (Mean: 48.9%). As can be deduced from this data, which shows the trend broken down by years, there is a linearity in the [68Ga]Ga-DOTA-TOC activity produced, which allowed us to ensure doses of 150-200 MBq for 3 patients per production (following the Guidelines of the European Association of Nuclear Medicine - EANM) [26], instead of 2 doses derived from the generator. These results are achieved and consolidated thanks to experience: as detailed in the troubleshooting section, this optimization was possible after three years of work, during which many syntheses failed and several problems related to the use of acids deteriorated the components of the cyclotron and the synthesis module.
Table 6.
EOB, 68Ga trapped and residual in ZR cartridge, 68GaCl3 transferred and residual into the reactor, activity and yield of [68Ga]Ga-DOTATOC produced at EOS. Time of synthesis: 40 min. Data related to the 10 validation syntheses of the process.
| - | 68Ga EOB (MBq) 35 mA 60 min | 68GaoZR Trapped (MBq) | 68Ga3ZR Residual (MBq) | 68GaCl3 Reactor Start Label. (MBq) | 68GaCl3 Reactor residual. (MBq) | EOS 68Ga DOTA-TOC (MBq) | Yield (%) | Decay Corrected Yield (%) |
| Mean | 3996 | 3610 | 476 | 2424 | 1098 | 1448 | 35.3 | 44.5 |
| Minimum | 3458 | 3124 | 176 | 2105 | 820 | 1206 | 29.4 | 37.0 |
| Maximum | 4502 | 4067 | 923 | 2813 | 1500 | 1725 | 42.0 | 53.0 |
Table 7.
Data of 293 [68Ga]Ga-DOTATOC syntheses (year 2021- first 6 months of 2024).
| - | Year | 68Ga EOB (MBq) 35 mA 60 min | 68GaCl3 into Reactor (MBq) | EOS 68Ga DOTA-TOC (MBq) | Yield (%) | Decay Corrected Yield (%) |
|---|---|---|---|---|---|---|
| Mean | 2021 | 4135 | 2311 | 1515 | 36.6 | 46.2 |
| - | 2022 | 4077 | 2376 | 1697 | 41.7 | 52.5 |
| - | 2023 | 4108 | 2461 | 1694 | 41.2 | 52.0 |
| - | 2024 | 4148 | 2211 | 1484 | 35.8 | 45.1 |
| Minimum | 2021 | 4100 | 1519 | 950 | 22.9 | 28.9 |
| - | 2022 | 3515 | 1503 | 1166 | 28.1 | 35.5 |
| - | 2023 | 3330 | 1776 | 1218 | 29.1 | 36.7 |
| - | 2024 | 4107 | 1720 | 1019 | 24.6 | 31.0 |
| Maximum | 2021 | 4144 | 3417 | 1893 | 45.7 | 57.6 |
| - | 2022 | 4181 | 3307 | 2010 | 53.8 | 67.9 |
| - | 2023 | 4292 | 4303 | 2514 | 56.3 | 70.9 |
| - | 2024 | 4366 | 4319 | 2500 | 52.4 | 66.0 |
3.3. [68Ga]Ga-DOTA-TOC Quality Control validation
Quality control process was validated for the first 10 syntheses before clinical use: every single production of the validation process, as evidenced in Table 8, met the requirements set by the current monograph (2482) of the Ph. Eur. [12], so the radiopharmaceutical was approved for human use. The table shows the results of our first 10 validation production batches for clinical use: the radiopharmaceutical appeared clear and colorless; pH fell within the acceptance criteria (3.2-8.0) with an average of 7.05; chemical purity was verified by HPLC (High Performance Liquid Chromatography). Fig. (3) shows the HPLC measurement profiles, referring to a reference standard that contains 50 μg·V−1 (where V is the maximum injectable volume, in our case equal to 5 mL) of Gallium-edotreotide and 60 μg·V−1 of edotreotide. In Fig. (4), graphs represent chemical and radiochemical purity: radiochemical purity analyzed in HPLC with a radiometric detector must be at least ≥ 95% of total radioactivity (average of 10 validation synthesis is 97.98%, Table 8), while to satisfy the requirements of chemical purity, edotreotide and metal complexes of edotreotide analyzed with a UV detector at a wavelength of 200 nm, must not exceed than the area of the peak of Gallium-edotreotide in the standard solution (50 μg·V−1), as shown in Fig. (4). Ethanol residual solvent, verified in gas chromatography, is < 10% V/V and <2.5 g·dose−1 in all 10 syntheses; radionuclidic identity fell within the acceptance criteria (61-75 min) with an average of 66.97 min; the level of bacterial endotoxins is under the limit of 175 IU·V−1 and all preparations were sterile. We also verified and validated the stability of the radiopharmaceutical [68Ga]Ga-DOTA-TOC: quality control was carried out at time 0 (T0), after 1 hour (T1) and after 2 hours (T2) from the end of synthesis, with particular attention to radiochemical purity, which can diminish over and affect the quality of PET/CT diagnostic images, risking subjecting the patient to ionising radiation without diagnostic benefit. The stability validation was verified up to 2 hours after synthesis: each batch of radiopharmaceutical produced had sufficient activity for 3 patients [26], considering the half-life of 68Ga. Here, we present the data from 10 validation radiopharmaceutical preparations, but all 293 preparations were declared suitable for human use, in accordance with the European Pharmacopoeia guidelines.
Table 8.
Results of [68Ga]Ga-DOTATOC quality control (10 validation syntheses).
| No. Synthesis | Appearance | pH | Radiochem. Purity (HPLC) | Radiochem. Purity (TLC) | Ethanol Residual (GC) | Radionuclidic Identity (T1/2) | Radionucl. Identity - Gray Spectrometer | Endotoxins | Sterility Test |
|---|---|---|---|---|---|---|---|---|---|
| 1. | Clear and colorless | 7.0 | 98.88 | 0.97 | <10% V/V <2.5 g·dose−1 |
64.61 | √ | <175 IU·V−1 | Sterile |
| 2. | Clear and colorless | 7.0 | 98.41 | 1.58 | <10% V/V <2.5 g·dose−1 |
66.86 | √ | <175 IU·V−1 | Sterile |
| 3. | Clear and colorless | 7.0 | 98.88 | 1.10 | <10% V/V <2.5 g·dose−1 |
66.20 | √ | <175 IU·V−1 | Sterile |
| 4. | Clear and colorless | 7.0 | 97.76 | 2.21 | <10% V/V <2.5 g·dose−1 |
67.70 | √ | <175 IU·V−1 | Sterile |
| 5. | Clear and colorless | 7.5 | 98.24 | 1.67 | <10% V/V <2.5 g·dose−1 |
66.17 | √ | <175 IU·V−1 | Sterile |
| 6. | Clear and colorless | 7.0 | 96.53 | 2.81 | <10% V/V <2.5 g·dose−1 |
65.84 | √ | <175 IU·V−1 | Sterile |
| 7. | Clear and colorless | 7.0 | 97.05 | 2.84 | <10% V/V <2.5 g·dose−1 |
67.13 | √ | <175 IU·V−1 | Sterile |
| 8. | Clear and colorless | 7.0 | 98.12 | 1.87 | <10% V/V <2.5 g·dose−1 |
71.26 | √ | <175 IU·V−1 | Sterile |
| 9. | Clear and colorless | 7.0 | 98.32 | 1.44 | <10% V/V <2.5 g·dose−1 |
66.25 | √ | <175 IU·V−1 | Sterile |
| 10. | Clear and colorless | 7.0 | 97.51 | 2.76 | <10% V/V <2.5 g·dose−1 |
67.86 | √ | <175 IU·V−1 | Sterile |
| Mean | Clear and colorless | 7.05 | 97.98 | 1.99 | <10% V/V <2.5 g·dose−1 |
66.97 | √ | <175 IU·V−1 | Sterile |
| Acceptance Criteria | Clear and colorless | 3.2-8.0 | ≥ 95% | ≤3% | <10% V/V <2.5 g·dose−1 |
61-75 min | Princ. Photons: 0.511 MeV, 1.077 MeV | <175 IU·V−1 | Sterile |
Fig. (3).

The figure shows the profile of measure of Reference solution verified by isocratic HPLC-Ultimate 3000: 50 mg·V−1 Ga-edotreotide and 60 mg·V−1 edotreotide.
Fig. (4).

The figures show the profiles of measures obtained by isocratict HPLC-Ultimate 3000, in particular: on the left [68Ga]Ga-DOTATOC analized with a radioactivity detector; on the right [68Ga]Ga-DOTATOC analized with UV detector.
3.4. Syntheses Failed and Troubleshooting Section
During the first year of [68Ga]Ga-DOTA-TOC production via cyclotron, we encountered several problems due to the use of acids, which damaged components of the gallium liquid target (foils and target body) and the synthesis module (Figs. 5 and 6), resulting in 12.2% failed syntheses (Table 9) and a target breakdown after 11 months. In Figs. (5 and 6), we can see the synthesis module and target parts corroded by acids during the first year of use, which led to the replacement of the entire target and the synthesis module pump. In Table 9 it shows the percentage of failed run: in 2021, we observed a failure rate of 12.2%, in 2022, failed runs decreased to around 3.4%, in 2023, only 2.2% and in 2024 we have not had any failed synthesis. Table 10 reports the main issues in the first period, their causes and the strategies used to solve them. Thanks to these strategies, in the following years, the number of failed syntheses decreased, and today, we have achieved a 0% failure rate. Table 11 shows the data for productions during the first period (68 syntheses, not including test and validation productions) and the second period (225 syntheses): a statistical evaluation was performed on these data by applying a Student’s t-Test to independent variables, assuming that the EOB (End of Beam), EOS (End of synthesis) and yield of synthesis (Table 12) data from m1 (first period) are lower than the m2 data (second period): the test showed statistical significance with a p-value <.001, so the null hypothesis was rejected. From this, it can be concluded that there is no significant difference in activity produced between the first and second period but as already seen in Table 9, the synthesis process has been consolidated.
Fig. (5).

68Ga Liquid Target (on the left); use of acids damaged components of gallium liquid target (foils in the figure on the right).
Fig. (6).

Damaged components of the synthesis module: (A) Valve actuators; (A1) Valve actuators corroded by acids used for [68Ga]GaCl3 purification (HCl, HNO3). (B) Vacuum pump; (B1 and B2) Particular of vacuum pump in which are visible (red arrows) precipitates due to evaporation of HNO3.
Table 9.
Percentage of failed [68Ga]Ga-DOTATOC production during the years (2021- first six months of 2024).
| Year | Syntheses | Total | ||
|---|---|---|---|---|
| Failed | Completed | % Failed | ||
| 2021 | 11 | 79 | 12.2 | 90 |
| 2022 | 3 | 86 | 3.4 | 89 |
| 2023 | 2 | 87 | 2.2 | 89 |
| 2024 | 0 | 41 | 0 | 41 |
| Total | 16 | 293 | 5.2 | 309 |
Table 10.
Issues encountered in production of 68Ga-DOTATOC, possible causes and related troubleshooting.
| Event | Possible Causes | Troubleshooting |
|---|---|---|
| 68Ga-Target Breakdown | Accumulation of [68Zn]ZnO precipitates into the target | Review in preparation of [68Zn]ZnO solution and Cleaning Solution (HNO3): 1. Target Solution: [68Zn]ZnO 1M in HNO3 0.3M instead of 0.2M (referring to the first period) 2. Cleaning Solution: HNO3 0.6M instead of 0.3 (referring to the first period) 3. Incubation of Cleaning solution into the target for 60-90 minutes, after a beam (15 min at 35 mA), before transfer it to receiving vial Taddeo |
| Beam drawback | Pressure loss due to foil degradation | Target Solution: [68Zn]ZnO 1M in HNO3 0.3M instead of 0.2M |
| 68Ga not transferred from receiving vial Taddeo to Zr resin | Synthesis module: Vacuum pump failure due to evaporation of HNO3 | Constant monitoring of the vacuum pump efficiency during the maintenance after production. If it fails the tests, replacement will be carried out |
| Eluition default of 68Ga activity trapped on resin | Manufacturing defect of the resin / HCl used for conditioning of the cartridge non-compliant | Preparation of new vials of HCL 4M |
| Low activity at EOB or at 68Ga-DOTATOC EOS | 1. Worn target 2. HNO3 degradation (used for target solution and cleaning solution 3. Activity remains anchored into the cartridge and not eluted |
1. Preventive maintenance of the 68Ga liquid target not beyond 3500 mA usually 2. Preparation of a new vial of cleaning solution (HNO3 0.6M) using a new lot of HNO3 70% 3. Preparation of a new vial of target solution ([68Zn]ZnO 1M in HNO3 0.3M) using a new lot of HNO3 70% 4. Preparation of new batch of reagents used during the purification of 68GaCl3 |
| Other unforeseen issues | Cassette developer system: -Fluid loss from the cartridge -Issues during assembly of the cassette |
1. Check the cartridges: if necessary, secure them using sealing film (Parafilm) 2. Qualified personnel through training, constant updates on any issues encountered and strategies for resolution. Check of the solution/cassette preparation by a second operator. |
Table 11.
Data relating to: EOB (MBq), EOS [68Ga]Ga-DOTATOC (MBq), Yield of synthesis (%) by Period 1 and 2.
| - | Period | Mean | Minimum | Maximum |
|---|---|---|---|---|
| EOB (MBq) | 1 | 3933 | 3552 | 4144 |
| - | 2 | 4107 | 3330 | 4366 |
| EOS [68Ga]Ga-DOTA-TOC (MBq) | 1 | 1506.3 | 1004 | 1893 |
| - | 2 | 1633.1 | 950 | 2514 |
| Decay corrected yield (%) | 1 | 46.0 | 28.9 | 57.6 |
| - | 2 | 49.8 | 34.4 | 68.3 |
Table 12.
Independent samples student’s t-test (HA: m 1/period 1< m 2/period 2).
| - | - | Statistic | df | p |
|---|---|---|---|---|
| EOB (MBq) | Student's t | -6.91ᵃ | 286 | < .001 |
| EOS 68Ga DOTA-TOC (MBq) | Student's t | -3.27ᵃ | 286 | < .001 |
| Yield (%) | Student's t | -1.51ᵃ | 286 | 0.066 |
Note: HA μ 1 < μ 2; ᵃ Levene's test is significant (p < .05), suggesting a violation of the assumption of equal variances.
4. DISCUSSION
Production of [68Ga]Ga-DOTATOC (a somatostatin analog used in positron emission tomography/computed tomography, PET/CT, for diagnosis of neuroendocrine tumors) was initially possible, in 2018-2020 (32 months - 170 productions) using a 68Ge/68Ga generator (Eckert & Ziegler, Germany). However, we identified the following negative characteristics of the 68Gallium isotope from generator: limited shelf life (1 year), a decrease in activity eluted (a maximum average of 1210 MBq at the beginning of the generator's shelf life and a minimum average of 457 MBq at the end of the shelf life), and the increasingly high costs and dependence on external suppliers. Having a cyclotron already available at the nuclear medicine department of Spedali Civili (Brescia, IT), combined the inclusion in the monograph of the European Pharmacopoeia of [68Ga]GaCl3 obtained by cyclotron, led us to produce [68Ga]Ga-DOTA-TOC by cyclotron.
In our work, we reported data from the production of [68Ga]Ga-DOTATOC using a 68Ge/68Ga generator for human use: as can be seen from Table 4, the activity eluted gradually decreased in step with the end of the shelf life of the generator (32 months - 170 productions).
We described the production, validation and consolidation process, gained from three years of experience (42 months - 309 productions), of the production of [68Ga]Ga-DOTA-TOC following the Guidelines on Good Radiopharmacy Practice (cGRPP), and relative quality control, following the current specific monograph (2482) of the Ph. Eur. [12]: ten validation batches of the production process were carried out before human use because, following specific monograph, the preparation may be released for human use before completion of the sterility test. All the productions were sterile (Table 8), conforming to the safety of the production process.
Referring to the work by Pandey et al. 2014 [14] and another by Riga et al. 2018 [27] 68Ga isotope was produced by cyclotron in a liquid target using a solution of [68Zn]ZnO 1M in HNO3 0.2M, after a beam of 60 min at 35mA with a proton energy beam degraded using aluminium foil; the purification of [68Ga]GaCl3 and subsequent production of [68Ga]Ga-DOTA-TOC were carried out using a cassette developer system, preparing reagents in house and manually assembling the synthesis cassette, as described by Rodnick et al. 2020 [10]. As can be seen from the results section, the trend in the synthesis of [68Ga]Ga-DOTA-TOC by cyclotron remains constant (Table 7), in contrast to the generator, which has a decreasing trend (Table 4). The quality control showed chemical, radiochemical purity and radionuclidic identity that met the requirements of the monograph specification (Table 8). However, as can be seen in the ‘Synthesis failed and troubleshooting section’, the switch to cyclotron production using a liquid target led to many inconveniences due to the use of strong acids and a production failure of around 12.22% in the first year (Table 9). During the first year, in 2021, we unfortunately experienced target breakdown due to accumulation of [68Zn]ZnO precipitates into the target, vacuum pump failure, eluition default of 68Ga activity trapped on resin, low activity at EOB (end of beam)or at [68Ga]Ga-DOTA-TOC EOS (end of synthesis) and other unforeseen issues (Table 10).
Some of these problems encountered in the target, mainly due to the use of strong acids, were also described by Z. Ashhar et al. [28]; we solved them, or at least mitigated them, by changing the preparation of the 1M 68Zn solution to 0.3M HNO3 (instead of 0.2M), changing the molarity of the target rinsing solution (HNO3 from 0.3M to 0.6M) and leaving it for 60-90 min in the target after a beam of 15 min at 35 mA before transferring it to the receiving vial Taddeo. The increase in HNO3 molarity improved the dissolution of ZnO and other metals and avoided the precipitation of metal salts. Moreover, preventive maintenance and reconditioning of the target were conducted every 3500 mA.
For the problems encountered during purification and synthesis, we mitigated them with constant monitoring of the synthesis process and the synthesis module (as described in the appropriate section - Table 10): constant monitoring of the synthesis module's vacuum pump, checking the synthesis reagents (especially by checking that the nitric acid does not turn yellow), but above all constant updates for strategic resolution.
So, from 12.2% of failed syntheses in 2021, we reduced the failure rate to 3.4% in 2022, 2.2% in 2023 and 0% in the first six months of 2024; 16 failed out of a total of 309 productions (5.2%).
In contrast, in the results section of the productions using a generator, the percentages of failed syntheses are 2.9% in 2018, 2.5% in 2019 and 5.4% in 2020 (Table 5). As can be seen from the data shown, while in cyclotron production, the number of failed syntheses was successfully reduced by constant monitoring of the target, the synthesis module and revision of the reagent preparation, the same type of monitoring was not possible for generator productions: a minimum of 2.5% of failed labelling is reasonably attributable exclusively to random error of the operator (metal contamination) and/or sudden instrumental (thermoblock) malfunction. The nature of the problems encountered during labelling (with generator and with cyclotron) are totally different, so we didn’t consider it useful to make a comparison, as it’s not meaningful. In this study, we were instead able to monitor the critical points of the synthesis process from the cyclotron and find tricks that made it possible to reduce the failed run percentage.
What we were able to do, instead, was a comparative evaluation of the [68Ga]Ga-DOTATOC activity (MBq) produced by generator versus cyclotron: with the generator, we obtained a maximum activity of [68Ga]Ga-DOTATOC of 1340 MBq and a minimum of 359 MBq (Table 4); while with the production of [68Ga]Ga-DOTATOC from the cyclotron (Table 7) a maximum activity of 2514 MBq and a minimum of 950 MBq was seen (with an average synthesis yield corrected for decay of 48.9%). This allowed access to radiopharmaceutical administration for PET/CT imaging evaluation of more patients than the generator: 3 per session instead of 1-2, considering doses of 150-200 MBq [26, 29].
We can therefore claim to have succeeded, thanks to 3 years' experience and 309 syntheses, in optimising and consolidating the production process of [68Ga]Ga-DOTATOC, achieving a synthesis yield percentage (corrected for decay) of an average value of 48.9% and an average activity of 1597.5 MBq, while complying with the parameters required by the specific monograph of the European Pharmacopoeia: chemical, radiochemical purity and radionuclidic identity; pH; sterility and pyrogenicity.
LIMITATION OF THE STUDY
In conclusion, while the study demonstrates that cyclotron production of [68Ga]Ga-DOTATOC using a liquid target is an efficient alternative to a generator, economic evaluations will be essential to establish the advantages of this production approach.
CONCLUSION
The aim of this study was to switch from generator to cyclotron labelling of [68Ga]Ga-DOTATOC, ensuring the same radiopharmaceutical quality, safety and efficacy, meeting the requirements of EANM guideline for radiopharmaceuticals, of Summary of Product Characteristics (for generator) and European Pharmacopoeia (for cyclotron).
Liquid target production of 68Ga radiopharmaceuticals, instead of the 68Ge/68Ga generator, is viable only for nuclear medicine centers with a cyclotron: at Spedali Civili (Brescia, Italy), it required the purchase of a target for 68Ga, a dedicated synthesis module, and staff training. During the first year of production, we encountered several problems due to the corrosion of cyclotron components and the synthesis module, but optimization of the preparation of the target solution and its washing, as well as constant monitoring of the synthesis module’s vacuum pump has led to excellent results. The advantages of [68Ga]Ga-DOTATOC production compared to the generator include: a consistently constant activity production, not affected by the short life of the generator, with the possibility of preparing doses for 3 patients (150-200 MBq for each one) per cyclotron run instead of 2 patients per generator production. Importantly the ability, once [68Ga]GaCl3 has been purified, to produce new radiopharmaceuticals indipendent of the availability of commercial kits represents a key advantage. We have also validated the synthesis process of [68Ga]Ga-PSMA-11 [30-31] and, in the future, we could produce more 68Ga-peptides [32-34].
ACKNOWLEDGEMENTS
The authors are particularly grateful to Antonietta Caldinelli, Sofia Confortini, Alice Fracassi, Lidia Mariani, and Roberto Rossini for excellent technical support.
LIST OF ABBREVIATIONS
- 68Ga
68Gallium
- cGRPP
Guidelines on Good Radiopharmacy Practice
- EANM
European Association of Nuclear Medicine
- EOB
End of Beam
- EOS
End of Synthesis
- GMP
Good Manufacturing Practices
- HPLC
High Performance Liquid Chromatography
- Ph. Eur
European Pharmacopoeia
- SmPC
Summary of Product Characteristics
AUTHORS’ CONTRIBUTIONS
MC, MS have drafted the first version of the manuscript. LS, CR and FS have implemented the cyclotron section; GLV supervised instrumentation details; LC has contributed to the images section; GB and EM interpreted and approved the writing about quality control’s process. All authors provided critical review, read, and approved the final manuscript. BF supervised the writing of the manuscript.
ETHICS APPROVAL AND CONSENT TO PARTICIPATE
Not applicable.
HUMAN AND ANIMAL RIGHTS
Not applicable.
CONSENT FOR PUBLICATION
Not applicable.
AVAILABILITY OF DATA AND MATERIALS
The data that support the findings of this study are available from the corresponding author, [MC], on special request.
FUNDING
None.
CONFLICT OF INTEREST
Francesco Bertagna is on the Editorial Advisory Board of the journal CRP.
REFERENCES
- 1.Cives M., Strosberg J.R. Gastroenteropancreatic neuroendocrine tumors. CA Cancer J. Clin. 2018;68(6):471–487. doi: 10.3322/caac.21493. [DOI] [PubMed] [Google Scholar]
- 2.Evangelista L., Ravelli I., Bignotto A., Cecchin D., Zucchetta P. Ga-68 DOTA-peptides and F-18 FDG PET/CT in patients with neuroendocrine tumor: A review. Clin. Imaging. 2020;67:113–116. doi: 10.1016/j.clinimag.2020.05.035. [DOI] [PubMed] [Google Scholar]
- 3.Kandathil A., Subramaniam R.M. Gastroenteropancreatic neuroendocrine tumor diagnosis. PET Clin. 2023;18(2):189–200. doi: 10.1016/j.cpet.2022.11.001. [DOI] [PubMed] [Google Scholar]
- 4.Lugat A., Frampas É., Touchefeu Y., Mirallié É., Bras M.L., Senellart H., Rauscher A., Fleury V., Campion L., Rohmer V., Couturier O.F., Lebtahi R., Rouzet F., Ruszniewski P., Kraeber-Bodéré F., Bourgeois M., Ansquer C. Prospective multicentric assessment of 68Ga-DOTANOC PET/CT in grade 1-2 GEP-NET. Cancers. 2023;15(2):513. doi: 10.3390/cancers15020513. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Database of the national nuclear data center (NNDC) at brookhaven national laboratory. Available from: www.nndc.bnl.gov/nndc/nudat/radform.html.
- 6.Weber J., Haberkorn U., Mier W. Cancer stratification by molecular imaging. Int. J. Mol. Sci. 2015;16(3):4918–4946. doi: 10.3390/ijms16034918. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Fani M., Mansi R., Nicolas G.P., Wild D. Radiolabeled somatostatin analogs—A continuously evolving class of radiopharmaceuticals. Cancers. 2022;14(5):1172. doi: 10.3390/cancers14051172. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Gillings N., Hjelstuen O., Ballinger J., Behe M., Decristoforo C., Elsinga P., Ferrari V., Peitl P.K., Koziorowski J., Laverman P., Mindt T.L., Neels O., Ocak M., Patt M., Todde S. Guideline on current good radiopharmacy practice (cGRPP) for the small-scale preparation of radiopharmaceuticals. EJNMMI Radiopharm. Chem. 2021;6(1):8. doi: 10.1186/s41181-021-00123-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Gillings N., Todde S., Behe M., Decristoforo C., Elsinga P., Ferrari V., Hjelstuen O., Peitl P.K., Koziorowski J., Laverman P., Mindt T.L., Ocak M., Patt M. EANM guideline on the validation of analytical methods for radiopharmaceuticals. EJNMMI Radiopharm. Chem. 2020;5(1):7. doi: 10.1186/s41181-019-0086-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Rodnick M.E., Sollert C., Stark D., Clark M., Katsifis A., Hockley B.G., Parr D.C., Frigell J., Henderson B.D., Abghari-Gerst M., Piert M.R., Fulham M.J., Eberl S., Gagnon K., Scott P.J.H. Cyclotron-based production of 68Ga, [68Ga]GaCl3, and [68Ga]Ga-PSMA-11 from a liquid target. EJNMMI Radiopharm. Chem. 2020;5(1):25. doi: 10.1186/s41181-020-00106-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Manoharan P., Lamarca A., Navalkissoor S., Calero J., Chan P.S., Julyan P., Sierra M., Caplin M., Valle J. Safety, tolerability and clinical implementation of ‘ready-to-use’ 68gallium-DOTA0-Tyr3-octreotide (68Ga-DOTATOC) (SomaKIT TOC) for injection in patients diagnosed with gastroenteropancreatic neuroendocrine tumours (GEP-NETs). ESMO Open. 2020;5(2):e000650. doi: 10.1136/esmoopen-2019-000650. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.European Pharmacopoeia. 2025. Available from: https://pheur.edqm.eu/home.
- 13.Pinot F., Le Pennec R., Abgral R., Blanc-Béguin F., Hennebicq S., Schick U., Valeri A., Fournier G., Le Roux P.Y., Salaun P.Y., Robin P. PSMA-11 PET/CT for detection of recurrent prostate cancer in patients with negative choline PET/CT. Clin. Genitourin. Cancer. 2023;21(2):248–257. doi: 10.1016/j.clgc.2022.12.007. [DOI] [PubMed] [Google Scholar]
- 14.Pandey M.K., Byrne J.F., Jiang H., Packard A.B., DeGrado T.R. Cyclotron production of (68)Ga via the (68)Zn(p,n)(68)Ga reaction in aqueous solution. Am. J. Nucl. Med. Mol. Imaging. 2014;4(4):303–310. [PMC free article] [PubMed] [Google Scholar]
- 15.Wang I.E., Brooks A.F., Clark M., Morrissette L.J., Scott P.J.H. Improved purification of cyclotron [68Ga]GaCl3 for the production of 68Ga radiopharmaceuticals. Nucl. Med. Biol. 2024;130-131:108892. doi: 10.1016/j.nucmedbio.2024.108892. [DOI] [PubMed] [Google Scholar]
- 16.Marketing authorization for the medicinal product for human use «Germanium Chloride (68 Ge)/Gallium Chloride (68 Ga) GalliaPharm. Gazzetta Ufficiale. 2024;217 [Google Scholar]
- 17.SomaKit TOC : EPAR - Product Information. 2024. Available from: https://www.ema.europa.eu/en/medicines/human/EPAR/somakit-toc.
- 18.Classification, pursuant to Article 12, paragraph 5, Law No. 189 of 8 November 2012, of medicinal products for human use «Rekovelle» and «Somakit Toc», approved with a centralized procedure. Gazzetta Ufficiale. 2017;91 [Google Scholar]
- 19.Revy A., Hallouard F., Joyeux-Klamber S., Skanjeti A., Rioufol C., Fraysse M. Feasibility and evaluation of automated methods for radiolabeling of radiopharmaceutical kits with Gallium-68. Curr. Radiopharm. 2019;12(3):229–237. doi: 10.2174/1874471012666190110170623. [DOI] [PubMed] [Google Scholar]
- 20.PETtrace™ 800 cyclotron series. 2024. Available from: https://www.gehealthcare.com/-/jssmedia/feature/gehc/products/cyclotrons/pt800-cyclotron-system-data-sheet-rev6.pdf?rev=-1&srsltid=AfmBOopL6g_rcZYy-W2Gm9jJ3SFBzxM7Jq7UXe6V1fjfHRJNyJZXgGD_.
- 21.Pandey M.K., Byrne J.F., Schlasner K.N., Schmit N.R., DeGrado T.R. Cyclotron production of 68Ga in a liquid target: Effects of solution composition and irradiation parameters. Nucl. Med. Biol. 2019;74-75:49–55. doi: 10.1016/j.nucmedbio.2019.03.002. [DOI] [PubMed] [Google Scholar]
- 22.Lin M., Waligorski G.J., Lepera C.G. Production of curie quantities of 68Ga with a medical cyclotron via the 68Zn(p,n)68Ga reaction. Appl. Radiat. Isot. 2018;133:1–3. doi: 10.1016/j.apradiso.2017.12.010. [DOI] [PubMed] [Google Scholar]
- 23.Rodnick M.E., Sollert C., Parr D.C., Frigell J., Gagnon K., Scott P.J.H. Preparation of [68Ga]GaCl3 using a Cyclotron. Methods Mol. Biol. 2024;2729:55–64. doi: 10.1007/978-1-0716-3499-8_5. [DOI] [PubMed] [Google Scholar]
- 24.FASTlab2 Developer manual. 2020. Available from: https://www.gehealthcare.com/products/molecular-imaging/pet-radiochemistry/fastlab?srsltid=AfmBOoobZ12veYax8rCNvNlTioCOfZKYGFasA8OiRcr5VsXr4CiD-Vgz.
- 25.FASTlab Gallium Chloride tracer package 1.7. 2022. Available from: https://www.gehealthcare.com/products/molecular-imaging/pet-radiochemistry/fastlab?srsltid=AfmBOorv6xZyFEtzDvPQ8g8UFzqucrY01ern_HhRHSeH3SXHrLUhRXIb.
- 26.Taïeb D., Hicks R.J., Hindié E., Guillet B.A., Avram A., Ghedini P., Timmers H.J., Scott A.T., Elojeimy S., Rubello D., Virgolini I.J., Fanti S., Balogova S., Pandit-Taskar N., Pacak K. European association of nuclear medicine practice guideline/society of nuclear medicine and molecular imaging procedure standard 2019 for radionuclide imaging of phaeochromocytoma and paraganglioma. Eur. J. Nucl. Med. Mol. Imaging. 2019;46(10):2112–2137. doi: 10.1007/s00259-019-04398-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Riga S., Cicoria G., Pancaldi D., Zagni F., Vichi S., Dassenno M., Mora L., Lodi F., Morigi M.P., Marengo M. Production of Ga-68 with a General Electric PETtrace cyclotron by liquid target. Phys. Med. 2018;55:116–126. doi: 10.1016/j.ejmp.2018.10.018. [DOI] [PubMed] [Google Scholar]
- 28.Ashhar Z., Ahmad Fadzil M.F., Md Safee Z., Aziz F., Ibarhim U.H., Nik Afinde N.M.F., Mat Ail N., Jamal Harizan M.A.H., Halib D., Alek Amran A., Adawiyah R., Abd Hamid M.H.N., Mahamood M., Razali N.I., Said M.A. Performance evaluation of Gallium-68 radiopharmaceuticals production using liquid target PETtrace 800 cyclotron. Appl. Radiat. Isot. 2024;205:111161. doi: 10.1016/j.apradiso.2023.111161. [DOI] [PubMed] [Google Scholar]
- 29.Gillings N., Hjelstuen O., Behe M., Decristoforo C., Elsinga P.H., Ferrari V., Kiss O.C., Kolenc P., Koziorowski J., Laverman P., Mindt T.L., Ocak M., Patt M., Todde S., Walte A. EANM guideline on quality risk management for radiopharmaceuticals. Eur. J. Nucl. Med. Mol. Imaging. 2022;49(10):3353–3364. doi: 10.1007/s00259-022-05738-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Thisgaard H., Kumlin J., Langkjær N., Chua J., Hook B., Jensen M., Kassaian A., Zeisler S., Borjian S., Cross M., Schaffer P., Dam J.H. Multi-curie production of gallium-68 on a biomedical cyclotron and automated radiolabelling of PSMA-11 and DOTATATE. EJNMMI Radiopharm. Chem. 2021;6(1):1. doi: 10.1186/s41181-020-00114-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Ashhar Z., Ahmad Fadzil M.F., Othman M.F., Yusof N.A., Abdul Onny M.A., Mat Ail N., Abd Rahman S.F. Cyclotron production of gallium-68 radiopharmaceuticals Using the 68Zn(p,n)68Ga reaction and their regulatory aspects. Pharmaceutics. 2022;15(1):70. doi: 10.3390/pharmaceutics15010070. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Migliari S., Sammartano A., Scarlattei M., Baldari G., Janota B., Bonadonna R.C., Ruffini L. Feasibility of a scale-down production of [68Ga]Ga-NODAGA-Exendin-4 in a hospital based radiopharmacy. Curr. Radiopharm. 2022;15(1):63–75. doi: 10.2174/1874471014666210309151930. [DOI] [PubMed] [Google Scholar]
- 33.Rosenberg A.J., Cheung Y.Y., Liu F., Sollert C., Peterson T.E., Kropski J.A. Fully automated radiosynthesis of [68Ga]Ga-FAPI-46 with cyclotron produced gallium. EJNMMI Radiopharm. Chem. 2023;8(1):29. doi: 10.1186/s41181-023-00216-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Wang I.E., Cheng K., Brooks A.F., Scott P.J.H., Viglianti B.L. Towards a general method for using cyclotron-produced Ga68 to manufacture clinical and research Ga68 tracers. Molecules. 2024;29(22):5457. doi: 10.3390/molecules29225457. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The data that support the findings of this study are available from the corresponding author, [MC], on special request.
