Remote analysis and management of sweat biomarkers using a wearable microfluidic sticker in adult cystic fibrosis patients

Rachel S Nelson a,b, Margaret G Grossman a,b, Zasu M Klug a,b, Mia Calamari a,b, Alvaro Donayre a,b, Thaddeus Cybulski a,b, Jared Schooley a,b, Garett J Griffith c, Daniel M Corcos c, Donald E Wright d, Jessica C Wallace d, Da Som Yang e, John A Wright Jr d,f, John A Rogers d,f,g,h,1, Stephen P Lee d,f, Roozbeh Ghaffari d,f,h,1, A J Aranyosi d,f,1, Manu Jain a,b,i,j,1
PMCID: PMC12377746  PMID: 40794828

Significance

This study demonstrates the feasibility and accuracy of a wearable microfluidic sweat patch (CF Patch) for remote monitoring of sweat chloride concentrations in people with cystic fibrosis (PwCF). By validating the CF Patch against gold-standard pilocarpine-induced sweat collection and chloridometry, we establish its potential as a noninvasive, real-time tool for assessing CFTR modulator efficacy, adherence, and pharmacodynamic variability. Furthermore, we provide insights into intraindividual fluctuations in sweat chloride levels among PwCF on CFTR modulators, underscoring the value of serial monitoring compared to single-point measurements. These findings highlight the promise of wearable biosensors in transforming remote disease management and optimizing personalized therapeutic strategies for PwCF.

Keywords: wearable microfluidic biosensors, exercise-induced sweat chloride, cystic fibrosis remote monitoring, CFTR modulator pharmacodynamics, sweat patch colorimetric analysis

Abstract

Sweat parameters such as volume and chloride concentration may offer invaluable clinical insights for people with CF (PwCF). Pilocarpine-induced sweat collection for chloridometry measurement is the gold standard for a CF diagnosis, but this technique is cumbersome and not suitable for remote settings or repeat measurements. We have previously reported the utility of a skin-interfaced microfluidic device (CF Patch) in conjunction with a smartphone image processing platform that enables real-time measurement of sweating rates and sodium chloride loss in laboratory and remote settings. Here, we conducted clinical studies assessing the accuracy of the CF Patch compared to chloridometry when using pilocarpine to induce sweat. We also tested the feasibility and accuracy of exercise-induced sweat chloride measurements in PwCF and healthy volunteers (HV). In the laboratory, using either pilocarpine or exercise to induce sweat, the CF Patch demonstrated strong correlations with sweat chloride measured by pilocarpine-induced chloridometry. In remote settings, exercise-induced sweat chlorides measured using the CF patch were strongly correlated with in-laboratory exercise-induced CF patch sweat chlorides in HV but had a weaker correlation in PwCF. For PwCF on CFTR modulators, there was greater day-to-day variability in sweat chloride compared to HV, which highlights the limitations of assessing CFTR modulator efficacy and pharmacodynamics based on a single in-laboratory chloridometry measurement. Moreover, these findings demonstrate that the CF Patch is suitable as a remote management device capable of measuring serial sweat chloride concentrations and offers the potential of monitoring the efficacy of CF medication regimens but should not replace pilocarpine-based chloridometry for making a CF diagnosis.


Cystic fibrosis (CF) is a genetic disorder caused by mutations in the cystic fibrosis transmembrane conductance regulator (CFTR) gene, leading to impaired chloride (Cl-) transport across cell membranes (1). As a result, people with CF (PwCF) have higher Cl and sodium (Na+) concentrations in their sweat (2). The measurement of [Cl] (chloride concentration) in sweat is a cornerstone in the diagnosis of CF, is associated with clinical outcomes (3), and is also used as a biomarker in clinical trials (4, 5). In addition, sweat [Cl] can serve as a tool to monitor patient health, assess adherence to CFTR modulators (small molecule therapies that restore CFTR protein function), and help adjust treatment regimens (6).

Traditional methods of sweat collection for [Cl] measurement involve the use of absorbent pads, gauzes, and rigid skin-interfacing cavities, like the Macroduct system (Macroduct Sweat Collection System; MSCS, ELITechGroup), followed by laboratory analysis with a chloridometer (7). These collection and postanalysis techniques employ sweat induction via the transdermal delivery of pilocarpine through iontophoresis (8). The multitude of steps and requirements involved in the pristine and reliable capture of eccrine sweat are limited to specialized referral centers. Though pilocarpine-induced chloridometry remains the gold standard for measuring sweat chloride to make a CF diagnosis, using it to make remote sweat collections for serial chloride measurements for ongoing CF management remains challenging.

Recent advances in the development and deployment of wearable microfluidic biosensors have gained significant traction in remote care settings, with the promise to overcome the limitations of existing sweat collection strategies (9). In particular, the clinical utility of a microfluidic wearable platform was demonstrated in a feasibility study across pediatric and newborn CF patients compared to gold standard clinical techniques (9). Moreover, commercialization efforts have led to wearable microfluidic sweat patch solutions tailored for athletes (10), with smartphone mobile application software that enables real-time image processing of sweat loss and sodium chloride concentrations in remote settings (11). Taken together, this new class of wearable microfluidic technology facilitates noninvasive and exercise-driven measurements of sweat [Cl] and regional sweat rates during physical activity and exercise, with potential to help support adult PwCF outside of clinical settings (1113). Even before the advent of CFTR modulators, exercise was historically recommended as an important therapy for PwCF (14). Now with the impact of CFTR modulators on exercise capacity (15, 16) and expected impact on health span and life expectancy, exercise has become an even more important therapy.

The present study aimed to compare and correlate sweat [Cl] measurements obtained from a wearable CF sweat patch (CF Patch) to those from the gold standard MSCS and chloridometry in people previously diagnosed with CF and healthy volunteers (HV). We also examined the relationship between [Cl] in sweat produced by exercise and that induced by pilocarpine. Measurement of [Cl] by the CF Patch during exercise at remote settings was used to assess the feasibility of measuring sweat [Cl] remotely as well as to determine how well it correlates with sweat [Cl] measured clinically. This wearable technology, though not intended to replace pilocarpine-chloridometry for the diagnosis of CF, could revolutionize remote monitoring of PwCF, allowing for regular, accessible measurements of sweat chloride to assess several clinically useful parameters such as CFTR modulator adherence, pharmacodynamics, and drug–drug interactions, among others. In addition, since regular exercise is a vital contributor to long-term health for PwCF (17), the CF Patch could also empower patients and caregivers with timely and actionable data for individualized hydration and electrolyte replenishment postexercise, thus enhancing its utility.

Results

Soft Wearable CF Patch Design.

The CF Patch consists of a soft and flexible polymeric material stack with an enclosed network of microfluidic channels, bioassay (corresponding to chloride concentrations), and food coloring dye (orange colored) embedded in microreservoirs. The CF Patch is manufactured using a roll-to-roll process, which enables multiple layers of soft polymers, embedded microchannels, patterned hypoallergenic skin adhesive, and an ultrathin graphical capping layer to be constructed rapidly at speeds approaching 2,000 devices per minute. Fig. 1 shows a schematic drawing of the CF Patch being worn on the inner volar forearm during physical exercise. Smartphone algorithms custom developed for image capture and processing enable real-time analysis of sweat loss, sweat rate, chloride concentration, and sodium chloride loss in near real-time following sweat capture upon completion of physical exercise.

Fig. 1.

Fig. 1.

CF Patch. Schematic drawing of the CF patient exercising on treadmill and wearing sweat sensor (CF Patch) on their volar forearm. The CF Patch contains two fluidic channels for i) measuring regional and whole-body sweat loss (orange colored channel) and ii) chloride concentrations (purple colored channel). Color reference markers printed on the CF Patch enable image analysis in different lighting conditions.

CF Patch Sweat Rate and Loss Analysis.

The study combined multiple sweat collection and analysis protocols. An initial lab-based session consisted of three separate measurements:

  • Pilocarpine-induced sweat collected by MSCS and analyzed with a chloridometer (“pilocarpine-MSCS”).

  • Pilocarpine-induced sweat collected by the CF patch and analyzed colorimetrically (“pilocarpine-CF”).

  • Exercise-induced sweat collected by the CF Patch and analyzed colorimetrically (“exercise-CF”).

In addition, a separate set of up to 5 remote sessions using exercise to collect sweat into the CF Patch were performed.

Fig. 2 shows a flow diagram of the sample size for each type of sweat collection. The final on-site sample size analyzed was n = 19 for pilocarpine-MSCS, n = 18 for pilocarpine-CF Patch, and n = 15 for on-site exercise-CF Patch. For the remote exercise part of the study, 18 PwCF completed all 5 exercise sessions and were able to successfully collect at least 1 sweat sample in the CF Patch. In addition, 15 PwCF met the prespecified criteria of at least 2 successful CF Patch collections. All 7 HV successfully completed every phase of the study.

Fig. 2.

Fig. 2.

Flow diagram of the study protocol.

Demographics and selected physiological measures show that PwCF and HV were well matched for the listed parameters (Table 1). The mean percent predicted forced expiratory volume in one second (ppFEV1) for PwCF was 93.3 ± 19%, and 18 of them were being treated with CFTR modulators. As measured by cardiopulmonary exercise testing (CPET) during the on-site visit, the peak oxygen uptake (VO2peak) was similar for HV and PwCF participants (34.4 ± 3.0 mL/kg/min and 30.9 ± 7.0 mL/kg/min, respectively). During the on-site constant wattage exercise session, during which sweat was collected in the CF Patch, the mean exercise time was 36.4 ± 3.8 min for HV and 40.5 ± 8.3 min for PwCF.

Table 1.

Demographics of the enrolled cohort

Variable PwCF (n = 20) HV (n = 7) P value
Physiological characteristics
 Age at enrollment* (years) 32.6 (20 to 64) 30.7 (27 to 34) 0.512
 Female (%) 10 (55%) 3 (42.9%) 0.678
 Weight (lb.) 162.9 (36.6) 176.4 (43.7) 0.481
 Height (in) 68 (4.6) 68.2 (2.8) 0.902
 Peak VO2 (ml/kg/min) 31.2 (6.9) 34.4 (3) 0.105
Ethnicity
 Non-Hispanic (%) 100 100 ns
Race
 White (%) 100 100
Cystic fibrosis medical history
 F508del homozygous 10 (50%)
 F508del heterozygous 9 (45%)
 Heterozygous, other 1 (5%)
 Diagnostic Sweat Chloride#, (mmol/L) 93.3 (24.9)
 FEV1 (L) 37 (1.1)
 ppFEV1 (L) 94.9 (19.6)
 Treated with CFTR modulators§ 19 (95%)

*Average and range.

Average and standard deviation.

Two subjects did not have diagnostic sweat available.

§The predominant CFTR modulator used in this cohort was ETI: Additional modulators were ivacaftor and open-label vanzacaftor triple, which were used in one participant each.

Comparison of MSCS Chloridometry and CF Patch Colorimetry.

In order to validate the CF Patch for sweat [Cl] measurement in PwCF, we first collected paired pilocarpine-induced sweat prior to exercise sweat collection at the on-site exercise visit. Pilocarpine collections for MSCS chloridometry and the CF Patch using colorimetry for [Cl] measurements were successful in 19 and 18 PwCF, respectively, and both types of collections were successful in all 7 HV. PwCF had a mean sweat [Cl] of 55 (range 17 to 106 mmol/L) by MSCS chloridometry and 66 (range 35 to 104 mmol/L) by the CF Patch. The results for HV were 25 mmol/L (range 10 to 49 mmol/L) by MSCS chloridometry and 29 mmol/L (range 5 to 54 mmol/L) by CF Patch colorimetry. There was a strong correlation between MSCS chloridometry and CF sweat patch-derived chloride concentrations across the combined cohorts with an r = 0.90 (P < 0.0001) (Fig. 3A). The Bland–Altman plot shows a mean bias of 12.7 mmol/L (CI: −11.9 to 37.3) without a systematic error as suggested by the confidence interval crossing 0 (Fig. 3B). Thus, the CF sweat patch performed very well compared to MSCS chloridometry across a range of clinically relevant sweat chloride values and was able to detect CFTR modulator decreases in sweat chloride.

Fig. 3.

Fig. 3.

Pilocarpine-induced CF Patch sweat chloride correlation versus MSCS chloridometry. People with CF and HV had resting paired pilocarpine-induced sweat chlorides collected, one via a Macroduct for chloridometry and the other via a CF sweat patch. A regression line is shown for the paired sweat chloride concentration measurements for the combined CF and HV cohort (A). The Bland–Altman plot for assessing agreement between the CF sweat patch and chloridometry shows good agreement with limited bias. The dashed line represents the mean difference (bias), and the gray shaded area represents the limits of agreement (mean difference ±1.96 standard deviations). The shaded area denotes the 95% confidence interval for the limits of agreement (B).

Comparison of Pilocarpine and Exercise Laboratory Sweat Chlorides.

Using the CF sweat patch affixed to each forearm during the exercise laboratory constant wattage session, we were able to collect sweat from 15 PwCF and 7 HV until the adequacy of sample volume was visually confirmed. The mean sweat volume collected was 35 (±18 µl) and 43 (±16.6 µl) for PwCF and HV, respectively. For PwCF, the mean sweat chloride during exercise was 64 (range 18 to 96 mmol/L), and for HV, it was 43 (range 8 to 79 mmol/L). There was a strong correlation between pilocarpine-induced and exercise-induced CF Patch sweat chloride concentrations with an r = 0.79 (P < 0.0001) (Fig. 4A). The Bland–Altman plot showed a mean bias of 2.1 mmol/L (95% CI: −12.7.9 to 26.8) and no systematic error (Fig. 4B). There was also an excellent correlation between pilocarpine-MSCS chloridometry and exercise CF Patch sweat chloride concentrations, r = 0.81 (P < 0.0001) and no systematic bias (Fig. 5 A and B). This demonstrates that exercise can serve an adequate surrogate to pilocarpine for sweat induction and chloride measurement.

Fig. 4.

Fig. 4.

Exercise-induced CF Patch sweat chloride correlation versus pilocarpine-induced CF Patch sweat chloride. People with CF and HV had sweat collected with the CF patch during constant work exercise, and it was compared to pilocarpine-induced sweat chloride measured by CF patch. The regression line indicates a comparison of sweat chloride concentrations between pilocarpine and exercise for the combined CF and HV cohort (A). The Bland–Altman plot for assessing agreement between the CF sweat patch and chloridometry shows good agreement and almost no bias. The dashed line shows the mean difference (bias), and the gray shaded area represents the limits of agreement (mean difference ±1.96 standard deviations). The shaded area denotes the 95% confidence interval for the limits of agreement (B).

Fig. 5.

Fig. 5.

Exercise-induced CF Patch sweat chloride correlation versus pilocarpine-induced chloridometry sweat chloride. People with CF and HV had sweat collected with the CF patch during constant work exercise, and it was compared to pilocarpine-induced sweat chloride measured by MSCS chloridometry. Shown is the regression line for the comparison of sweat chlorides between pilocarpine and exercise for the combined CF and HV cohort (A). The Bland–Altman plot for assessing agreement between the CF Patch and chloridometry shows agreement and a small bias. The dashed line shows the mean difference (bias), and the gray area represents the limits of agreement (mean difference ± 1.96 standard deviations) (B).

Feasibility and Accuracy of Remote Exercise Sweat Chloride Collection.

When evaluating success of remote exercise CF Patch use, we assessed whether participants were able to complete at least two interpretable CF Patch sweat collections. Of the 19 PwCF that started the remote exercise part of the study, 18 PwCF completed all 5 workouts as instructed. Of those, 15 (79%) PwCF reached the threshold of 2 interpretable CF sweat collections. Among the HV, all 7 completed the requested five collections (Fig. 6A). For the successful remote exercise sessions, average sweat [Cl] was 56 (range 11 to 115 mmol/L) for PwCF and 31 (range 0.5 to 58 mmol/L) for HV. For the combined cohort, there was a lower correlation between the mean remote exercise CF Patch sweat chloride and the exercise laboratory CF Patch sweat chloride, r = 0.72 (P = 0.0001) (Fig. 6B). To ascertain the potential impact of CFTR modulators on the lower correlation between remote and laboratory exercise sweat chlorides, we analyzed HV and PwCF separately. In HV the correlation between remote and laboratory exercise sweat chlorides was excellent, r = 0.91 (P = 0.004) (Fig. 6C), but in PwCF, there was a weaker correlation, r = 0.59 (P = 0.02) (Fig. 6D). This contrasts with the comparison between exercise laboratory CF Patch and pilocarpine-chloridometry sweat [Cl], which showed that when analyzed separately, HV and PwCF had very similar correlations (SI Appendix, Fig. S5 A and B). Since pilocarpine-chloridometry and exercise laboratory CF Patch sweat [Cl] were measured on the same day, and CFTR modulators impact sweat chloride, this is consistent with the notion that CFTR modulators are contributing to more day-to-day variability of sweat [Cl] in PwCF.

Fig. 6.

Fig. 6.

Remote exercise-induced CF Patch sweat chlorides have stronger correlation with exercise laboratory CF sweat chloride in HV than in PwCF. PwCF (Mod = on modulator, No mod = not on a modulator) and HV had sweat chloride assessed sweat during remote exercise sessions, and the mean of these values (A) for each individual was compared to the exercise laboratory sweat chloride collected during constant work exercise. Shown is the regression line for the comparison of sweat chlorides between remote and laboratory exercise for the combined CF and HV cohort (B), and correlation was lower than the laboratory-only comparisons. When examined as separate cohorts, HV had a much stronger correlation than PwCF (C and D).

Intraindividual Variability in Sweat Chloride Concentrations.

The difference between HV and PwCF in remote and laboratory exercise sweat chloride correlations prompted us to assess the intraindividual variability in remote exercise sweat chlorides in each cohort. We compared the standard deviations in remote exercise sweat chlorides between HV and PwCF. The average of the standard deviations for the mean sweat chlorides in HV and PwCF was 14.1 and 23.4 mmol/L, respectively, and the difference in the standard deviations was statistically significant (P = 0.03) (Fig. 7). We then assessed the relationship between time of CFTR modulator intake and remote exercise CF Patch sweat [Cl] for PwCF and found no significant correlation (data not shown). Similarly, there was no correlation between the volume of sweat collected during an individual exercise session and [Cl] among HV or PwCF (data not shown). SI Appendix, Table S1 presents the four sweat chloride values collected for each participant—single values from pilocarpine-induced sweat using the Macroduct Sweat Collection System (MSCS), pilocarpine-induced sweat using the CF Patch, laboratory-based exercise-induced sweat using the CF Patch, and the mean of remote exercise-induced sweat chloride values using the CF Patch—providing subject-level data that support the analyses of correlation, feasibility, and intraindividual variability described in the main text.

Fig. 7.

Fig. 7.

PwCF have greater variability in serial sweat chloride concentrations compared to HV by CF Patch. Each point represents the mean remote exercise sweat chlorides measured by CF Patch for HV and PwCF. We also calculated standard deviations within individuals for remote exercise sweat chlorides and compared HV and PwCF. PwCF had higher mean sweat chlorides and greater standard deviations (*P = 0.005, @P = 0.03).

Discussion

In this study, we report four findings. We confirm that a microfluidic sweat patch using a colorimetric assay has a very strong correlation with MSCS chloridometry, the gold standard, in PwCF on CFTR modulators to measure reductions in sweat [Cl]. In addition, we show a strong correlation between pilocarpine and exercise-induced sweat chlorides and demonstrate that serial remote sweat chloride measurement is feasible. In HV, there is strong correlation between remote and laboratory exercise sweat chlorides, but this correlation is weaker in PwCF on CFTR modulators. Finally, our data suggest that some PwCF on CFTR modulators show substantial variability in sweat chlorides over a 2-week period.

These results further previous work, which demonstrated that a CF sweat patch and colorimetry compare favorably with MSCS chloridometry and show that it can detect CFTR modulator-mediated decreases in sweat [Cl] (9). Importantly, the CF Patch is accurate across a range of clinically relevant sweat [Cl] values and did not demonstrate any systematic bias. The current CF sweat patch has design enhancements over previous iterations that allow for short collection periods (exercise time as low as 25 min), increased collection area for rapid sweat collection, and optimized color analysis in the Cl channel to function across a wider range of lighting conditions. Thus, the CF Patch offers several advantages over laboratory based chloridometry, such as portability, ease of repeat measurements and provides the possibility of measuring sweat [Cl] in remote real-world settings including clinics or hospitals without access to a sweat lab and in low- and middle-income countries (18). In settings where it is possible pilocarpine-chloridometry should always remain the gold standard for a CF diagnosis. Of note, the CF patch could be adapted by imbedding pilocarpine for sweat induction or by changing the assay in each channel to measure other molecules important for the monitoring of other disease states (i.e., heart failure, renal dysfunction, diabetes mellitus).

Previous data have shown that in HV, exercise- and pilocarpine-induced sweat chlorides are similar at moderate intensity exercise (19), but there are no previous data in PwCF. Despite physiologic differences between exercise and pilocarpine induction of sweat, our data confirm that exercise is a satisfactory sweat inducer at higher sweat [Cl] seen in PwCF and that sweat [Cl] induced by exercise can serve as an adequate surrogate to pilocarpine sweat induction. Thus, exercise (or other ways to induce sweat) could be used for serial remote sweat [Cl] measurements and these data could be utilized to inform multiple clinical aspects of care for PwCF. This includes but is not limited to assessing CFTR modulator adherence, impact of drug–drug interactions, pharmacodynamic effects of reduced CFTR modulator dosing or modest clinical responses to CFTR modulators. The data from the CF Patch could also be used to provide individualized recommendations for fluid and electrolyte repletion following exercise (12).

The high success rate of remote sweat collection in PwCF suggests that it is feasible for patients to learn and use the CF Patch adequately. There was a lower percentage of successful collections in PwCF than HV due to the development of an intercurrent illness in 2 PwCF which is inherent in people with a chronic illness. Nevertheless, a large majority of PwCF were able to complete the threshold of 2 successful remote collections, and it is likely that those who could not during the study period could do so once they recovered from their acute illness. This suggests that virtually all PwCF who exercise could successfully utilize the CF Patch. Theoretically, the CF patch could be utilized in PwCF who do not exercise with other inducers of sweat such as hot showers or saunas as we have shown in HV (20), but this needs to be rigorously tested before it can be recommended.

Of note, compared to HV, PwCF had lower correlations between sweat [Cl] measured in the exercise laboratory and those measured during remote exercise. For HV, intraindividual variations in sweat sodium concentration [which is highly correlated to sweat (Cl)] is on the order of 5 to 16%, comparable to the present results (21, 22). PwCF showed a greater day-to-day variability of sweat [Cl] compared to HV. Since all PwCF except 1 were on a CFTR modulator, we speculate that this may have been due to variable effects of modulators, though we cannot exclude the possibility that PwCF generally have more day-to-day variability in sweat [Cl].There were also more PwCF than HV in our cohort which could have contributed to increased variability though we did not see differences in correlation between HV and PwCF when we compared correlations in-laboratory exercise CF Patch and pilocarpine-chloridometry sweat [Cl]. Large variations in repeated tests of sweat [Cl] in PwCF have been observed previously, although it is unclear what fraction of participants in that study were on CFTR modulators (23). Other potential contributors to day-to-day variability include hydration status, salt intake, or exercise intensity (24). Nevertheless, these data suggest that a single in clinic sweat [Cl] measurement may not be an adequate surrogate for assessing chronic CFTR modulator pharmacodynamics in PwCF.

Previous data have shown limited correlation between single, clinic measured CFTR modulator-induced decreases in sweat [Cl] and clinical responses (3, 2527). It is possible that serial remote sweat [Cl] measurements may offer a more accurate assessment of individual CFTR modulator exposure and a better correlation with a clinical response. This study demonstrates the potential for integrating exercise-induced sweat [Cl] testing into clinical monitoring protocols for PwCF on CFTR modulators. The CF sweat patch could provide a less invasive, patient-friendly alternative for longitudinal monitoring, particularly for assessing CFTR modulator efficacy and side effects, identifying potential causes of variability in therapeutic responses, and facilitating research studies in nonclinical settings. This could be studied in a prospective fashion to provide evidence for clinical utility before implementing it into regular clinical use. We would advise against making treatment decisions until such data exist.

Methods

CF Patch Device Design and Fabrication.

A detailed description of the CF Patch and smartphone application has been published previously (12, 13, 22). Briefly, a multilayered stack of thin-film polymeric materials (polyurethane film; Innovize Inc., MN) with laser-cut regions creates the microchannels and inlet ports. Pressure-sensitive adhesive bonds the intermediate fluidic layer with the top graphics and subjacent skin adhesive layers (polyethylene film; 3 M, MN). Laser-cut inlet windows in the adhesive layer create openings to define the two sweat collection regions for microchannel 1 and microchannel 2. For colorimetric analysis of microchannel 1 (volume of approximately 112 µL), a dehydrated colored dye (≈4 µL) was deposited in an assay well region and used to mix with sweat, providing an optical method to measure regional sweat volume and sweating rate. Microchannel 2 (volume approximately 20 µL) was embedded with a mixture of silver chloranilate (Green Room Board Co, NC) and polyhydroxyethylmethacrylate (pHEMA; Sigma-Aldrich, MO) hydrogel placed near the assay well region to mix with sweat for Cl concentration analysis (12). Images of the CF Patch were captured using the CF application (beta) for colorimetric analysis. Images were initially stored on the local smartphone memory, then transmitted to a computer via the cloud services operated by Apple Inc.

Study Design.

We conducted a prospective, single-center, feasibility study of adult PwCF and adult HV who were self-identified as exercising at home. The study was approved by the Northwestern University Biomedical Institutional Review Board (IRB) (STU00215214). Participants were recruited between April 2022 and May 2024. PwCF were recruited from a single CF center at Canning Thoracic Institute (Northwestern Medicine), and HV were recruited from the community. All participants were informed of the experimental procedures and associated risks before providing consent.

Each participant was asked to complete 2 parts of the study. Participants first completed clinic-based sweat testing, consisting of resting pilocarpine-induced sweat collection, a cardiopulmonary exercise test (CPET), and on-site exercise-induced sweat collection at the exercise physiology laboratory of Department of Physical Therapy & Human Movement Sciences (Northwestern University). After completing the on-site portion, participants were asked to complete 5 remote exercise sessions over 14 days while wearing the CF sweat patch. The location and activity of remote exercise were of their own preference.

A total of 23 PwCF and 7 HV consented to participate in this study. Prior to the on-site visit, 3 PwCF withdrew consent and were excluded from analyses. One PwCF experienced a pulmonary exacerbation after completing 3 remote exercise sessions, and all data collected prior to early termination were included in analyses.

Human Subject Sweat Testing.

Our first objective was to collect pilocarpine-induced sweat and compare MSCS collection and chloridometry with the CF Patch colorimetric analysis for baseline sweat [Cl]. Sweat was induced via pilocarpine iontophoresis on the left and right ventral forearms and collected into a CF Patch and Macroduct coil in separate arms. Samples extracted from the Macroduct coil were analyzed at the CF Therapeutics Development Network (TDN) Center for Sweat Analysis at Children’s Hospital Colorado.

CPET consisted of an incremental protocol on a cycle ergometer (LODE, The Netherlands). Once initiated, work rate increased incrementally every minute until the subject was no longer able to maintain a cycling cadence of ≥60 revolutions per minute. Rating of perceived exertion and exercising blood pressure values were recorded every 2 min during the CPET. VO2peak was determined using 30-s averaging, and the highest single 30-s value was recorded as VO2peak. This value was recorded during the last minute of exercise. Exercising heart rate data throughout the study were monitored by a Medtronic Zephyr biomodule and chest strap (Medtronic, Minneapolis, MN).

Our second objective was to collect exercise-induced sweat into the CF sweat patch to measure regional sweat volume and [Cl]. After completion of the CPET, participants were allowed a dedicated rest period and fluid replenishment (water and Gatorade) until their baseline heart rate was reached and post-CPET sweat induction had ceased. Research personnel then applied two microfluidic sweat patches, one on each arm following which each subject began exercising at a constant wattage which was 50 to 60% of peak wattage derived from the CPET. Exercising heart rate and rating of perceived exertion, as well as environmental conditions including temperature, barometric pressure, and humidity of the exercise physiology laboratory, were recorded every 5 min during on-site exercise. Participants continued exercising until either there was sufficient sweat in each channel of the CF Patch (assessed by visual inspection) or 60 min had elapsed.

Remote Exercise Sessions and Sweat Collection.

Following recovery from exercise testing, participants were provided with CF sweat patches and a Medtronic Zephyr biomodule with chest strap to wear during the 5 remote exercise sessions over a 14-d period. Subjects were trained on use of all devices prior to starting remote workouts. They were also provided with a smartphone with the OmniSense mobile application. There were no specific requirements for the remote exercise sessions, and participants were instructed to perform their normal exercise activities (e.g., running/cycling, resistance training, hot yoga). Participants were instructed to take photos of the CF sweat patches following each of the 5 remote exercise sessions and transmit the images to the study team which were subsequently analyzed.

Data Collection.

Study data were collected and managed using Research Electronic Data Capture (REDCap) tools hosted at Northwestern University (2830). REDCap is a secure, web-based software platform designed to support data capture for research studies, providing 1) an intuitive interface for validated data capture; 2) audit trails for tracking data manipulation and export procedures; 3) automated export procedures for seamless data downloads to common statistical packages; and 4) procedures for data integration and interoperability with external sources.

Smartphone and camera-based colorimetric analysis of sweat volume and Cl concentration.

Details on smartphone colorimetric analysis under controlled and ambient lighting conditions have been published. (11) Briefly, photos of microfluidic patches were captured using a smartphone application (iPhone 11, Apple Inc.) in clinical and remote settings. Images were captured at the time of CF Patch removal, at the end of sweat induction by pilocarpine or exercise. When captured in a clinical setting multiple images were captured by study personnel within 5 min of completion of sweat collection. At home, study participants were instructed to capture multiple CF images within 5 min of exercise completion. To assess feasibility of remote image capture, success was deemed if a participant recorded interpretable images from at least 2 remote exercise sessions. Final mean volume and sweat [Cl] were calculated from manual image analysis. Sweat [Cl] and volume data from CF Patch collections were obtained from each sample from which at least a single image measurement was available. We also performed an additional analysis in which we only included data from which triplicate images were obtained which is shown in supplementary data (SI Appendix, Figs. S1–S4). RAW images were used for processing to eliminate artifacts introduced by normalization, compression, and other preprocessing steps. The sweat volume and [Cl] were calculated by manual image analysis by a trained data scientist (11). The colors of four swatches on the CF Patch (green, purple, orange, white) were measured and used to correct for variations in white balance due to ambient lighting. Local sweat volume was measured by identifying the furthest extent to which colored dye propagated in microchannel 1 and mapping that location to a volume using a CAD model of the patch. Chloride was measured by computing the average color within microchannel 2 relative to the background color of the patch. In CIELAB color space, this color is represented by a 2D vector. This vector was then projected onto the unit vector for the color of chloranilic acid, which was measured empirically. The length of this projection maps to chloride concentration by a power law relationship, as described by the Beer–Lambert law.

Statistical Analysis.

Statistical analysis was performed using the R statistical language version 4.4.2. Descriptive statistics are reported as means and standard deviations. Comparisons between HV and PwCF and between different conditions were analyzed using independent samples t-tests. Bivariate comparisons between measurement conditions (i.e., MSCS vs. CF patch, etc.) were made by linear regression, and Pearson correlation coefficients were calculated. Proportional bias was calculated for each Bland–Altman analysis. The Wilcoxon rank-sum test was used for comparison of variance measurements of remote exercise sweat chlorides. Statistical significance was attributed when P < 0.05.

Supplementary Material

Appendix 01 (PDF)

pnas.2506137122.sapp.pdf (272.1KB, pdf)

Acknowledgments

We thank Dr. Susanna McColley and Gabriel Fagans for their support in defining key sensing features and fabrication of CF Patches. This work made use of the Northwestern University’s Querrey Simpson Institute for Bioelectronics for testing and analysis. Research reported in this publication was supported, in part, by the National Institutes of Health’s National Center for Advancing Translational Sciences (Grant Number UL1TR001422) and the National Heart, Lung, and Blood Institute Grant Number (R43HL162246). This material is the result of work supported with resources and the use of facilities at the Jesse Brown VA Hospital. The content in this manuscript is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health, the US. Department of Veterans Affairs, or the US Government.

Author contributions

R.S.N., M.G.G., T.C., G.J.G., D.M.C., J.A.R., S.P.L., R.G., A.J.A., and M.J. designed research; R.S.N., M.G.G., Z.M.K., M.C., A.D., and M.J. performed research; R.S.N., D.E.W., J.C.W., D.S.Y., J.A.W., J.A.R., S.P.L., R.G., A.J.A., and M.J. contributed new reagents/analytic tools; R.S.N., T.C., J.S., J.A.R., R.G., A.J.A., and M.J. analyzed data; and R.S.N., M.G.G., Z.M.K., M.C., A.D., T.C., J.S., G.J.G., D.M.C., D.E.W., J.C.W., D.S.Y., J.A.W., J.A.R., S.P.L., R.G., A.J.A., and M.J. wrote the paper.

Competing interests

A.J.A., R.G., S.P.L., and J.A.R. are co-founders of Epicore Biosystems, which develops and commercializes microfluidic devices for sweat analysis. D.E.W., J.C.W., and J.A.W. are currently employed by Epicore Biosystems. All other authors declare that they have no competing interests.

Footnotes

This article is a PNAS Direct Submission.

Contributor Information

John A. Rogers, Email: jrogers@northwestern.edu.

Roozbeh Ghaffari, Email: rooz@epicorebiosystems.com.

A. J. Aranyosi, Email: aja@epicorebiosystems.com.

Manu Jain, Email: m-jain@northwestern.edu.

Data, Materials, and Software Availability

All data supporting the conclusions of this study, portions of software code that are central to interpreting or replicating the current study data have been deposited in Dryad (https://doi.org/10.5061/dryad.bzkh189pj) (31).

Supporting Information

References

  • 1.Farrell P. M., et al. , Guidelines for diagnosis of cystic fibrosis in newborns through older adults: Cystic Fibrosis Foundation consensus report. J. Pediatr. 153, S4–S14 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Brown M. B., McCarty N. A., Millard-Stafford M., High-sweat Na+ in cystic fibrosis and healthy individuals does not diminish thirst during exercise in the heat. Am. J. Physiol. Regul. Integr. Comp. Physiol. 301, R1177–R1185 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Espel J. C., et al. , The relationship between sweat chloride levels and mortality in cystic fibrosis varies by individual genotype. J. Cyst. Fibros. 17, 34–42 (2018). [DOI] [PubMed] [Google Scholar]
  • 4.Ramsey B. W., et al. , A CFTR potentiator in patients with cystic fibrosis and the G551D mutation. N. Engl. J. Med. 365, 1663–1672 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Secunda K. E., et al. , Females with cystic fibrosis demonstrate a differential response profile to ivacaftor compared with males. Am. J. Respir. Crit. Care Med. 201, 996–998 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Ibrahim H., et al. , Individualized approach to elexacaftor/tezacaftor/ivacaftor dosing in cystic fibrosis, in response to self-reported anxiety and neurocognitive adverse events: A case series. Front. Pharmacol. 14, 1156621 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Hansen L., Buechele M., Koroshec J., Warwick W. J., Sweat chloride assay for cystic fibrosis. Using pilocarpine iontophoresis stimulation, filter paper collection and cotlove chloridometer analysis. Minn. Med. 50, 1191–1196 (1967). [PubMed] [Google Scholar]
  • 8.Davis P. B., Cystic fibrosis since 1938. Am. J. Respir. Crit. Care Med. 173, 475–482 (2006). [DOI] [PubMed] [Google Scholar]
  • 9.Ray T. R., et al. , Soft, skin-interfaced sweat stickers for cystic fibrosis diagnosis and management. Sci. Transl. Med. 13, eabd8109 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Ghaffari R., Aranyosi A. J., Lee S. P., Model J. B., Baker L. B., The Gx Sweat Patch for personalized hydration management. Nature Rev. Bioeng. 1, 5–7 (2023). [Google Scholar]
  • 11.Baker L. B., et al. , Skin-interfaced microfluidic system with machine learning-enabled image processing of sweat biomarkers in remote settings. Adv. Mater. Technol. 7, 2200249 (2022). [Google Scholar]
  • 12.Baker L. B., et al. , Skin-interfaced microfluidic system with personalized sweating rate and sweat chloride analytics for sports science applications. Sci. Adv. 6, eabe3929 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Baker L. B., et al. , Sweating rate and sweat chloride concentration of elite male basketball players measured with a wearable microfluidic device versus the standard absorbent patch method. Int. J. Sport Nutr. Exerc. Metab. 32, 342–349 (2022). [DOI] [PubMed] [Google Scholar]
  • 14.Hebestreit H., et al. , Statement on exercise testing in cystic fibrosis. Respiration 90, 332–351 (2015). [DOI] [PubMed] [Google Scholar]
  • 15.Philipsen L. K. D., et al. , Changes in exercise capacity in people with Cystic Fibrosis after one year of Elexacaftor/Tezacaftor/Ivacaftor treatment— A Danish prospective cohort. J. Cyst. Fibros. 23, 1080–1086 (2024). [DOI] [PubMed] [Google Scholar]
  • 16.Gruber W., et al. , Effects of an exercise intervention on exercise capacity in adults with cystic fibrosis: A quasi-experimental study comparing individuals treated with and without elexacaftor/tezacaftor/ivacaftor. Pediatr. Pulmonol. 60, e71076 (2025). [DOI] [PubMed] [Google Scholar]
  • 17.Caterini J. E., et al. , Exercise intolerance in cystic fibrosis-the role of CFTR modulator therapies. J. Cyst. Fibros. 21, 282–292 (2022). [DOI] [PubMed] [Google Scholar]
  • 18.Bell S. C., et al. , The future of cystic fibrosis care: A global perspective. Lancet Respir. Med. 8, 65–124 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Choi D. H., Kitchen G. B., Stewart K. J., Searson P. C., The dynamic response of sweat chloride to changes in exercise load measured by a wearable sweat sensor. Sci. Rep. 10, 7699 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Zhang Y., et al. , Passive sweat collection and colorimetric analysis of biomarkers relevant to kidney disorders using a soft microfluidic system. Lab Chip 19, 1545–1555 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Baker L. B., Stofan J. R., Hamilton A. A., Horswill C. A., Comparison of regional patch collection vs. whole body washdown for measuring sweat sodium and potassium loss during exercise. J. Appl. Physiol. (1985) 107, 887–895 (2009). [DOI] [PubMed] [Google Scholar]
  • 22.Baker L. B., Sweating rate and sweat sodium concentration in athletes: A review of methodology and intra/interindividual variability. Sports Med. (Auckland, N.Z.) 47, 111–128. (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Vermeulen F., Lebecque P., De Boeck K., Leal T., Biological variability of the sweat chloride in diagnostic sweat tests: A retrospective analysis. J. Cyst. Fibros. 16, 30–35 (2017). [DOI] [PubMed] [Google Scholar]
  • 24.Baker L. B., Wolfe A. S., Physiological mechanisms determining eccrine sweat composition. Eur. J. Appl. Physiol. 120, 719–752 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Pigliasco F., et al. , Simultaneous quantification of Ivacaftor, Tezacaftor, and Elexacaftor in cystic fibrosis patients’ plasma by a novel LC-MS/MS method. Biomedicines 11, 628 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Davis P. B., Schluchter M. D., Konstan M. W., Relation of sweat chloride concentration to severity of lung disease in cystic fibrosis. Pediatr. Pulmonol. 38, 204–209 (2004). [DOI] [PubMed] [Google Scholar]
  • 27.Zemanick E. T., et al. , Sweat chloride reflects CFTR function and correlates with clinical outcomes following CFTR modulator treatment. J. Cyst. Fibros. 24, 246–254 (2025), 10.1016/j.jcf.2024.12.006. [DOI] [PubMed] [Google Scholar]
  • 28.Obeid J. S., et al. , Procurement of shared data instruments for research electronic data capture (REDCap). J. Biomed. Inform. 46, 259–265 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Harris P. A., et al. , The REDCap consortium: Building an international community of software platform partners. J. Biomed. Inform. 95, 103208 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Harris P. A., et al. , Research electronic data capture (REDCap)–a metadata-driven methodology and workflow process for providing translational research informatics support. J. Biomed. Inform. 42, 377–381 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Nelson R. S., et al. , Remote analysis and management of sweat biomarkers using a microfluidic sticker in adult cystic fibrosis patients. Dryad. 10.5061/dryad.bzkh189pj. Deposited 29 July 2025. [DOI] [PMC free article] [PubMed]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.