ABSTRACT
Background
Although high‐dose biotin interference in automated immunoassays is now considered, there are very few studies showing biotin interference in manually operated research kits, especially with enzyme‐linked immunosorbent assay (ELISA). The aims of our study were to determine the effects of biotin interference on various parameters, including leptin, leptin receptor (LEPR), ghrelin, acylated ghrelin, deacylated ghrelin, ghrelin receptor (GHSR), kisspeptin (KISS1), kisspeptin receptor (KISS1R), preptin, peroxisome proliferator activated receptor gamma (PPARγ), nod‐like receptor pyrin domain‐containing 3 (NLRP3) and interleukin‐18 (IL‐18), which contribute to energy homeostasis in healthy and obese children.
Methods
Serum pools were prepared from healthy and obese individuals, and biotin concentrations in samples containing different amounts of biotin were measured via sandwich and competitive ELISA methods. In addition, possible biotin interactions were investigated by determining the concentrations of all the study parameters in serum pools containing different amounts of biotin.
Results
We found that the biotin‐competitive, ghrelin‐competitive, KISS1‐competitive, GHSR, leptin and LEPR ELISA kits were less affected by biotin interference and the results of these assay kits were more reliable. Unexpectedly, high levels were also measured in the biotin sandwich ELISA kit, indicating that biotin interference can also occur in manually operated assay kits.
Conclusions
Biotin exhibited an interference effect even in well‐functioning, qualified kits, and this negative effect was less common in competitive kits. Biotin interference was closely associated with the quality of the research kit, the parameters studied, and the presence of high biotin concentrations in the blood.
Keywords: biotin interference, ELISA, pediatric obesity
In this study, we aimed for the first time to demonstrate biotin interference in ELISA research kits used in the analysis of pediatric obesity parameters other than routine testing. Our study revealed that biotin interaction may occur in manual ELISA kits and may cause erroneous test results. It will raise awareness among health professionals, researchers, and medical companies on this issue. By encouraging manufacturers developing medical products to take measures to reduce the effect of biotin interaction, it will prevent erroneous results in scientific studies and contribute to more precise measurements.

Abbreviations
- AACC
American Association for Clinical Chemistry
- ELISA
Enzyme‐Linked Immuno Sorbent Assay
- FDA
The United States Food and Drug Administration
- GHSR
ghrelin receptor
- IL‐18
interleukin‐18
- KISS1
kisspeptin
- KISS1R
kisspeptin receptor
- LEPR
leptin receptor
- max
maximum
- min
minimum
- NLRP3
nod‐like receptor pyrin domain‐containing 3
- PPARγ
peroxisome proliferator activated receptor gamma
- WHO
World Health Organization
1. Introduction
Biotin, a water‐soluble vitamin, is used as a coenzyme in reactions catalyzed by carboxylase enzymes in gluconeogenesis, fatty acid biosynthesis, single C‐numbered fatty acid degradation, and catabolism of branched‐chain amino acids. Since biotin is also synthesized by intestinal bacteria and taken up through food and diet, its deficiency is rare. The daily dose required for adults is 30 μg. The daily dosage required for biotin in children varies according to age and gender. Children aged 1–3 should receive 8 μg of biotin per day, children aged 4–8 should receive 12 μg per day, children aged 9–13 should receive 20 μg per day, and children aged 14–18 should receive 25 μg of biotin per day [1, 2].
Although biotin deficiency is not common in healthy individuals, it may occur in some hereditary diseases. Metabolic disorders such as biotinidase deficiency, holocarboxylase synthetase deficiency, or biotin‐thiamine‐sensitive basal ganglia disease, long‐term parenteral nutrition, anticonvulsants, or protein pump inhibitors, Crohn's disease, alcoholism, and excessive consumption of raw eggs (avidin binds biotin) may cause biotin deficiency. Biotin treatment can be administered at doses of 5–20 mg/day for biotinidase deficiency and holocarboxylase synthetase deficiency, and 5–10 mg/kg/day for biotin‐thiamine‐sensitive basal ganglia disease. For examples with multiple sclerosis, 300 mg of biotin therapy is prescribed, which is 10,000 times the recommended daily intake [3, 4]. Biotin doses of 5,10,20 mg and 300 mg can cause blood biotin levels of approximately 200–1200 ng/mL [5, 6, 7].
The relationship between biotin and microbiome diversity in severe obesity is very important. In people with severe obesity, circulating biotin levels are below normal, and the expression of biotin‐related genes in adipose tissue varies [8]. Belda et al. showed that biotin supplementation of mice fed a high‐fat diet increased microbiome diversity and improved the potential for bacterial production of biotin and B vitamins while limiting weight gain and glycemic impairment [9]. Furthermore, another study in obese mice showed that biotin supplementation induced beneficial metabolic effects through modulation of lipid metabolism, inflammation, and fibrosis in adipose tissue and liver [10].
The increase in the prevalence of obesity has become an important health issue throughout the lives of children and adolescents as well as adults [11]. Childhood obesity is the cause of adult obesity and is an important problem especially because of its relationship with morbid obesity, metabolic syndrome, diabetes mellitus, cancer, and cardiovascular diseases [12, 13, 14, 15, 16, 17, 18].
In 2017, the United States Food and Drug Administration (FDA) stated that biotin interference may cause erroneous test results, and this information was updated in 2019. The FDA has reported that biotin, found in many multivitamins, interacts with immunodiagnostic tests, including some tests for troponin, a clinically important biomarker that helps diagnose heart attacks. Despite these warnings, the FDA still receives adverse event reports of biotin interference and warns manufacturers about biotin interference [6, 19]. In clinical studies, biotinylated antibodies and antigens are frequently used in immunologic tests. It has also been shown that biotin taken at doses of 100–300 mg daily may adversely affect immunologic test results, such as cardiac and cancer marker data, and may cause erroneous results in many hormone parameters studied in automated immunoassay systems [20, 21, 22]. All immunoassays that use streptavidin‐biotin binding as part of the test reaction are susceptible to biotin interference from excess biotin in the blood. The mechanism of interference is related to signal loss in the immunoassay because high doses of biotin in the blood bind and block biotinylated reagents that are captured by the biotin binding sites on the streptavidin‐coated solid phase [7].
Despite the warning from the FDA, research on biotin interference is relatively new. Today, many laboratories are still unaware of interference problems or have not taken precautions concerning the extent to which biotin interference affects test results. There are very few studies showing biotin interference, especially in studies performed with research kits other than routine tests [23, 24]. None of these studies have been conducted in pediatric obesity. In our study, we aimed to examine the possible effects of different biotin concentrations on several research parameters in the case of pediatric obesity within the framework of the principles of creating ELISA kits and estimating possible biotin interference via these research tools.
2. Methodology
All laboratory studies were performed in the Biochemistry Research Laboratory of the university. Before the study, Ethics Committee approval documents numbered 2020.01.01.01 were obtained from the institution where the study was conducted. Families of the children, especially their parents, were informed. After a detailed pediatric physical examination and informing them about their inclusion in the study, signed consent forms from the consenting parents were attained and children were informed verbally. According to the World Health Organization (WHO) criteria with age‐ and sex‐specific body mass index percentile values above 95 and between 15 and 85, respectively, were included in the study. The study included healthy and obese children aged 5–18 years.
In this study, the concentrations of leptin, leptin receptor (LEPR), ghrelin, acylated ghrelin, deacyl ghrelin, ghrelin receptor (GHSR), kisspeptin (KISS1), kisspeptin receptor (KISS1R), preptin, peroxisome proliferator‐activated receptor gamma (PPARγ), Nod‐like receptor pyrin domain‐containing 3 (NLRP3), and interleukin‐18 (IL‐18) were determined using enzyme‐linked immunosorbent assay (ELISA) kits prepared with biotinylated reagents. Table 1 shows the detection ranges and CV values of the ELISA kits. All of the assays are Research Use Only (RUO) assays.
TABLE 1.
Detection range and CV values of the ELISA kits.
| ELISA Kits | Manufacturers | Catalog number | Intra assay CV (%) | Inter assay CV (%) | Detection range |
|---|---|---|---|---|---|
| Leptin sandwich | DRG | EIA‐2395 | 7.3%–7.1%–4.2% | 6.9%–3.7%–9.1% | 0.7–100 ng/mL |
| Ghrelin sandwich | Bioassay Technology Laboratory | E3091Hu | < 8% | < 10% | 0.05–10 ng/mL |
| Acylated ghrelin sandwich | Bioassay Technology Laboratory | E3090Hu | < 8% | < 10% | 0.05–15 ng/mL |
| Deacyl ghrelin sandwich | Bioassay Technology Laboratory | E4218Hu | < 8% | < 10% | 0.5–200 ng/L |
| GHSR sandwich | Cloud‐Clone Corp. | SEC516Hu | < 10% | < 12% | 0.156–10 ng/mL |
| Ghrelin competitive | Elabscience | E‐EL‐H1919 | 6%–4.44%–3.17% | 6%–4.44%–3.51% | 0.16–10 ng/mL |
| LEPR sandwich | Elabscience | E‐EL‐H0093 | 6.48%–5.43%–3.44% | 6.25%–5.21%–3.49% | 0.31–20 ng/mL |
| KISS1 sandwich | Elabscience | E‐EL‐H2129 | 5.1%–4.65%–4.9% | 6.59%–4.94%–3.11% | 125–8000 pg/mL |
| KISS1 competitive | Cloud‐Clone Corp. | CEC559Hu | < 10% | < 12% | 12.35–1000 pg/mL |
| KISS1R sandwich | Cloud‐Clone Corp. | SEF540Hu | < 10% | < 12% | 0.312–20 ng/ml |
| NLRP3 sandwich | Cloud‐Clone Corp. | SEK115Hu | < 10% | < 12% | 0.312–20 ng/ml |
| Preptin sandwich | Elabscience | E‐EL‐H0913 | 5.62%–4.63%–4.65% | 6.99%–5.24%–3.03% | 62.5–4000 pg/mL |
| PPARγ sandwich | Elabscience | E‐EL‐H1361 | 6.38%–6.25%–3.48% | 6.52%–6.09%–5.41% | 0.16–10 ng/mL |
| IL‐18 sandwich | Elabscience | E‐EL‐H0253 | 6.04%–4.67%–4.81% | 6.88%–4.37%–3.59% | 15.63–1000 pg/mL |
| Biotin andwich | Bioassay Technology Laboratory | E3932Hu | < 8% | < 10% | 1–300 ng/mL |
| Biotin Competitive | Cusabio | CSB‐E16209 | < 8% | < 10% | 12.5–200 pg/mL |
Abbreviations: GHSR, ghrelin receptor; IL‐18, interleukin‐18; KISS1, kisspeptin; KISS1R, kisspeptin receptor; LEPR, leptin receptor; NLRP3, nod‐like receptor pyrin domain‐containing 3; PPARγ, peroxisome proliferator activated receptor gamma.
Serum pools were prepared from the blood samples of healthy controls and obese children. The study included 5 individuals in the control group (45 samples) and 5 individuals in the obese group (45 samples). A 1 mg/mL stock of biotin (Sigma‐14,400) solution with 0.1 mol/L NaOH was obtained. Using this stock solution of biotin, working solutions containing biotin at concentrations of 0 (control), 10, 25, 50, 100, 250, 500, 1000, and 1200 ng/mL were prepared by serial dilution with distilled water and were subsequently added to the serum samples. The biotin concentrations in the serum were determined by both the sandwich biotin ELISA kit and the competitive biotin ELISA kit. In biotin sandwich and competitive parameters, subgroups were named as obese 1 (O1), obese 2 (O2), obese 3 (O3), obese 4 (O4), obese 5 (O5), healthy control 1 (C1), healthy control 2 (C2), healthy control 3 (C3), healthy control 4 (C4), healthy control 5 (C5).
In the second phase of the study, working solutions containing biotin at concentrations of 10, 50, 500, and 1000 ng/mL were prepared from a 1 mg/mL stock biotin solution. The study was conducted from the serum pool prepared with samples taken from healthy children and obese children. A total of 28 samples, 3 groups (12 samples) from healthy individuals and 4 groups (16 samples) from obese individuals, were used in the study. The various working solutions were added to the pooled serum samples. Then, leptin, LEPR, ghrelin, acylated ghrelin, deacylated ghrelin, GHSR, KISS1, KISS1R, preptin, PPARγ, NLRP3, and IL‐18 concentrations were measured in each sample via the sandwich method. The ghrelin and KISS1 parameters were determined based upon both sandwich and competitive principles. The difference in Biotin % bias between the groups formed by the addition of biotin at concentrations of 10, 50, 500, and 1000 ng/mL was calculated with the following formula: measured value (concentration of the parameter with 50, 500 and 1000 ng/mL biotin added)—value not affected by biotin interference (concentration of the parameter with 10 ng/mL biotin added)/value not affected by biotin interference (concentration of the parameter with 10 ng/mL biotin added) × 100.
3. Statistics
All the statistical analyses were performed with SPSS (Statistical Package for the Social Sciences) Version 26. Parametric or nonparametric statistical tests were applied according to the distribution characteristics of the data. The Shapiro–Wilk test was used to determine whether the dataset followed a normal distribution. In addition to the independent samples t test, the Mann–Whitney U test and Kruskal–Wallis test were used to evaluate the differences between the groups, and Pearson and Spearman correlation analyses were performed to examine whether the values related to the parameters changed together. In addition, Bonferroni correction was used, and the Mann–Whitney U test was applied in the evaluation of subgroups. p < 0.05 were considered to indicate statistical significance.
4. Results
The biotin concentrations of pooled serum samples containing 0 (control), 10, 25, 50, 100, 250, 500, 1000, or 1200 ng/mL biotin working solution were compared with those of two different ELISA methods, sandwich and competitive. In the biotin study of the healthy control and obese groups, the number of samples and the median, minimum–maximum and p values of the biotin parameter working with the sandwich and competitive methods are given in Table 2. The biotin sandwich parameter was significantly lower in the obese group (***p ≤ 0.001), while the difference between the groups was not significant for the biotin competitive parameter (p = 0.675). The biotin concentrations of healthy controls and obese children are shown in Figures 1 and 2. A significant positive correlation (r = 0.572; ***p < 0.001) was found between the values of the biotin sandwich and biotin competitive parameters in the control group, while no correlation was found in the obese group (r = −0.108; p > 0.05). Figure 3 shows the number of samples and median values of the control and obese groups according to the amount of added biotin. Moreover, there was no significant difference between the obese and control groups according to the amount of added biotin and the biotin sandwich parameters. Biotin concentrations measured by the sandwich ELISA method in children were found to be much greater than those measured by the competitive method. The number of samples, medians, and p values for the biotin parameter subgroups are shown in Table 3 and statistically significant differences among the groups were found for both the biotin sandwich (***p < 0.001) and biotin competitive parameters (**p < 0.01). In addition, Mann–Whitney U test with Bonferroni correction was performed as a post hoc test to reveal the difference between the subgroups of the biotin parameter. With Bonferroni correction, p:0.05/5 = 0.01 and all pairwise comparisons with p < 0.01 were considered significant. A significant difference (**p < 0.01) was found between the C1–O1, C2–O2, C3–O3, and C5–O5 groups in terms of the biotin sandwich parameter. While there was a significant difference (**p < 0.01) only between the C2–O2 groups in terms of the biotin competitive parameter, no significant difference was found in any other group.
TABLE 2.
Samples, medians, minimum‐maximum values, and p values for the biotin parameter.
| Parameters | Groups | Samples | Min. | Max. | Median | IQR | p |
|---|---|---|---|---|---|---|---|
| Biotin sandwich (ng/mL) | Control | 45 | 49.3 | 111.1 | 84.9 | 30.90 | |
| Obese | 45 | 40.7 | 88.2 | 64.4 | 23.75 | 0.001*** | |
| Biotin competitive (ng/mL) | Control | 45 | 0.919 | 13.347 | 6.107 | 6.21 | |
| Obese | 45 | 0.472 | 15.513 | 5.333 | 7.52 | 0.675 |
Note: The unit of the biotin competitive kit has been converted to ng/mL. *P≤ 0.05, **P≤ 0.01, ***P≤ 0.001.
Abbreviations: max, maximum; min, minimum.
FIGURE 1.

Concentration graph of biotin sandwich parameter in healthy control and obese children.
FIGURE 2.

Concentration graph of biotin competitive parameter in healthy control and obese children.
FIGURE 3.

Median values of biotin sandwich and biotin competitive tests for obese and control groups.
TABLE 3.
Samples, medians and p values for subgroups of the biotin parameter.
| Parameters | Groups | Samples | Median | p |
|---|---|---|---|---|
| Biotin sandwich (ng/mL) | Obese 1 (O1) | 9 | 79.9 | |
| Obese 2 (O2) | 9 | 66.3 | ||
| Obese 3 (O3) | 9 | 62.2 | ||
| Obese 4 (O4) | 9 | 60.6 | ||
| Obese 5 (O5) | 9 | 60.6 | ||
| Healthy control 1 (C1) | 9 | 94.4 | ||
| Healthy control 2 (C2) | 9 | 97.2 | ||
| Healthy control 3 (C3) | 9 | 97 | ||
| Healthy control 4 (C4) | 9 | 68.2 | ||
| Healthy control 5 (C5) | 9 | 56.5 | ||
| Total | 90 | 70.2 | 0.000*** | |
| Biotin competitive (ng/mL) | Obese 1 (O1) | 9 | 4.6 | |
| Obese 2 (O2) | 9 | 4.6 | ||
| Obese 3 (O3) | 9 | 6.7 | ||
| Obese 4 (O4) | 9 | 8.8 | ||
| Obese 5 (O5) | 9 | 5.3 | ||
| Healthy control 1 (C1) | 9 | 7.1 | ||
| Healthy control 2 (C2) | 9 | 6.6 | ||
| Healthy control 3 (C3) | 9 | 8.2 | ||
| Healthy control 4 (C4) | 9 | 4.1 | ||
| Healthy control 5 (C5) | 9 | 4.4 | ||
| Total | 90 | 5.4 | 0.008** |
We conducted analyses to determine the interference effect of biotin at concentrations of 10, 50, 500, and 1000 ng/mL for NLRP3, leptin, ghrelin, acylated ghrelin, deacylated ghrelin, KISS1, preptin, PPARγ, LEPR, GHSR, KISS1R, and IL‐18. The number of samples, mean, SD, and median values of ELISA parameters according to the amount of biotin added at 10, 50, 500, and 1000 ng/mL between the healthy and obese groups are given in Tables 4 and 5. The results were stratified based on the distribution characteristics of the data. Table 4 presents the mean ± standard deviation (SD) values for parameters that followed a normal distribution, while Table 5 includes the median values and corresponding non‐parametric analyses for parameters that did not follow a normal distribution. This distinction was made to ensure appropriate statistical interpretation based on the underlying data characteristics. The % bias values of all parameters at 10, 50, 500, and 1000 ng/mL added biotin concentrations are shown in Table 6. See Figure 4A–N for changes in the concentrations of parameters due to the addition of 10, 50, 500, and 1000 ng/mL biotin to all ELISA parameters in the healthy and obese groups. The biotin concentration in obese and healthy subgroups is shown in Figure 5.
TABLE 4.
Mean ± SD values of ELISA parameters for normally distributed data, according to biotin concentrations of 10, 50, 500, and 1000 ng/mL.
| Groups | Amount of biotin added (ng/mL) | GHSR (ng/mL) | KISS1 competitive (pg/mL) | KISS1R (ng/mL) | Leptin (ng/mL) | NLRP3 (ng/mL) | |
|---|---|---|---|---|---|---|---|
| Healthy | 10 | Mean | 0.69 | 96.34 | 4.32 | 0.87 | 0.16 |
| Samples | 3 | 3 | 3 | 3 | 3 | ||
| SD | 0.07 | 27.7 | 0.8 | 0.16 | 0.11 | ||
| 50 | Mean | 0.65 | 85.82 | 4.77 | 0.92 | 0.16 | |
| Samples | 3 | 3 | 3 | 3 | 3 | ||
| SD | 0.15 | 12.66 | 0.90 | 0.21 | 0.07 | ||
| 500 | Mean | 0.72 | 91.44 | 4.60 | 0.9 | 0.30 | |
| Samples | 3 | 3 | 3 | 3 | 3 | ||
| SD | 0.13 | 11.78 | 1.09 | 0.20 | 0.36 | ||
| 1000 | Mean | 0.7 | 82.45 | 4.46 | 0.88 | 0.22 | |
| Samples | 3 | 3 | 3 | 3 | 3 | ||
| SD | 0.16 | 15.70 | 1.15 | 0.19 | 0.12 | ||
| Total | Mean | 0.69 | 89.01 | 4.54 | 0.89 | 0.21 | |
| Samples | 12 | 12 | 12 | 12 | 12 | ||
| SD | 0.11 | 16.42 | 0.87 | 0.16 | 0.18 | ||
| Obese | 10 | Mean | 0.51 | 117.25 | 4.20 | 8.25 | 0.30 |
| Samples | 4 | 4 | 4 | 4 | 4 | ||
| SD | 0.14 | 22.33 | 1.43 | 1.94 | 0.16 | ||
| 50 | Mean | 0.53 | 121.88 | 4.59 | 8.02 | 0.26 | |
| Samples | 4 | 4 | 4 | 4 | 4 | ||
| SD | 0.11 | 26.53 | 1.97 | 1.73 | 0.16 | ||
| 500 | Mean | 0.67 | 120.51 | 4.44 | 8.01 | 0.28 | |
| Samples | 4 | 4 | 4 | 4 | 4 | ||
| SD | 0.27 | 44.63 | 1.18 | 1.60 | 0.11 | ||
| 1000 | Mean | 0.51 | 116.09 | 4.51 | 8.10 | 0.27 | |
| Samples | 4 | 4 | 4 | 4 | 4 | ||
| SD | 0.07 | 13.94 | 0.98 | 1.33 | 0.07 | ||
| Total | Mean | 0.55 | 118.93 | 4.44 | 8.09 | 0.28 | |
| Samples | 16 | 16 | 16 | 16 | 16 | ||
| SD | 0.16 | 26.14 | 1.3 | 1.49 | 0.12 | ||
Abbreviations: GHSR, ghrelin receptor; KISS1, kisspeptin; KISS1R, kisspeptin receptor; NLRP3, nod‐like receptor pyrin domain‐containing 3.
TABLE 5.
Median values and sample sizes of ELISA parameters for non‐normally distributed data, according to biotin concentrations of 10, 50, 500, and 1000 ng/mL.
| Groups | Amount of biotin added (ng/mL) | Açylated ghrelin (ng/mL) | Deaçyl ghrelin (ng/L) | Ghrelin (ng/mL) | Ghrelin competitive (ng/mL) | IL‐18 (pg/mL) | KISS1 sandwich (pg/mL) | LEPR (ng/mL) | PPARγ (ng/mL) | Preptin (pg/mL) | |
|---|---|---|---|---|---|---|---|---|---|---|---|
| Healthy | 10 | Samples | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 |
| Median | 3.44 | 69.08 | 1.83 | 1.65 | 96.65 | 1417.28 | 8.64 | 0.20 | 80.83 | ||
| 50 | Samples | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | |
| Median | 3.84 | 65.77 | 2.31 | 1.78 | 59.63 | 1490.34 | 7.64 | 0.16 | 172.63 | ||
| 500 | Samples | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | |
| Median | 3.76 | 66.75 | 2.22 | 1.42 | 54.70 | 1362.13 | 9.06 | 0.19 | 85.17 | ||
| 1000 | Samples | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | |
| Median | 2.37 | 60.91 | 2.12 | 1.18 | 62.48 | 1640.25 | 8.06 | 0.22 | 89.64 | ||
| Total | Samples | 12 | 12 | 12 | 12 | 12 | 12 | 12 | 12 | 12 | |
| Median | 3.60 | 66.26 | 2.12 | 1.64 | 61.05 | 1453.81 | 8.53 | 0.20 | 87.40 | ||
| Obese | 10 | Samples | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 |
| Median | 1.15 | 31.17 | 0.70 | 0.92 | 76.77 | 2228.29 | 23.51 | 0.25 | 162.13 | ||
| 50 | Samples | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | |
| Median | 1.45 | 29.73 | 0.66 | 0.93 | 114.55 | 2045.84 | 24.29 | 0.24 | 165.09 | ||
| 500 | Samples | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | |
| Median | 1.44 | 29.77 | 0.75 | 0.85 | 97.67 | 2322.44 | 23.55 | 0.22 | 163.58 | ||
| 1000 | Samples | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | |
| Median | 1.46 | 30.56 | 0.68 | 0.86 | 94.48 | 2233.3 | 23.04 | 0.27 | 116.92 | ||
| Total | Samples | 16 | 16 | 16 | 16 | 16 | 16 | 16 | 16 | 16 | |
| Median | 1.31 | 30.18 | 0.68 | 0.92 | 97.67 | 2194.26 | 23.55 | 0.24 | 153.05 | ||
Abbreviations: IL‐18, interleukin‐18; KISS1, kisspeptin; LEPR, leptin receptor; PPARγ, peroxisome proliferator‐activated receptor gamma.
TABLE 6.
The % bias values of all parameters at 10, 50, 500 and 1000 ng/mL added biotin.
| Groups | Analyte | Analyte concentration (10 ng/mL added biotin) | % Bias with added biotin (ng/mL) | ||
|---|---|---|---|---|---|
| 50 | 500 | 1000 | |||
| Healty | Acylated ghrelin (ng/mL) | 3.44 | 11.77 | 9.36 | −30.95 |
| Deacyl ghrelin (ng/L) | 69.08 | −4.8 | −3.37 | −11.82 | |
| Ghrelin (ng/mL) | 1.83 | 26.47 | 21.71 | 15.91 | |
| Ghrelin competitive (ng/mL) | 1.66 | 7.59 | −14.17 | −28.5 | |
| IL‐18 (pg/mL) | 96.65 | −38.3 | −43.39 | −35.35 | |
| Leptin (ng/mL) | 0.87 | 5.51 | 3.33 | 1.49 | |
| LEPR (ng/mL) | 8.64 | −11.55 | 4.89 | −6.68 | |
| NLRP3 (ng/mL) | 0.16 | −2.45 | 85.88 | 38.65 | |
| PPARγ (ng/mL) | 0.20 | −23.18 | −5.31 | 8.21 | |
| Preptin (pg/mL) | 80.83 | 113.58 | 5.36 | 10.89 | |
| GHSR (ng/mL) | 0.69 | −4.78 | 5.65 | 1.44 | |
| KISS1 sandwich (pg/mL) | 1417.28 | 5.15 | −3.89 | 15.73 | |
| KISS1 competitive (pg/mL) | 96.35 | −10.93 | −5.09 | −14.42 | |
| KISS1R (ng/mL) | 4.32 | 10.53 | 6.5 | 3.37 | |
| Obese | Acylated ghrelin (ng/mL) | 1.15 | 26.45 | 25.76 | 27.32 |
| Deacyl ghrelin (ng/L) | 31.17 | −4.61 | −4.48 | −1.93 | |
| Ghrelin (ng/mL) | 0.70 | −3.17 | 5.94 | −2.69 | |
| Ghrelin competitive (ng/mL) | 0.92 | 1.18 | −7.67 | −6.37 | |
| IL‐18 (pg/mL) | 76.77 | 49.21 | 27.23 | 23.07 | |
| Leptin (ng/mL) | 8.25 | −2.74 | −2.94 | −1.85 | |
| LEPR (ng/mL) | 23.51 | 3.31 | 0.17 | −2.01 | |
| NLRP3 (ng/mL) | 0.30 | −12.78 | −6.88 | −12.13 | |
| PPARγ (ng/mL) | 0.25 | −39.2 | −12.15 | 7.84 | |
| KISS1 sandwich (pg/mL) | 2228.29 | −8.18 | 4.22 | 0.22 | |
| KISS1 competitive (pg/mL) | 117.25 | 3.94 | 2.78 | −0.98 | |
| KISS1R (ng/mL) | 4.20 | 9.24 | 5.6 | 7.34 | |
| Preptin (pg/mL) | 162.11 | 1.83 | 0.9 | −27.87 | |
| GHSR (ng/mL) | 0.511 | 3.91 | 31.31 | 0.78 | |
Abbreviations: GHSR, ghrelin receptor; IL‐18, interleukin‐18; KISS1, kisspeptin; KISS1R, kisspeptin receptor; LEPR, leptin receptor; NLRP3, nod‐like receptor pyrin domain‐containing 3; PPARγ, peroxisome proliferator‐activated receptor gamma.
FIGURE 4.

The changes in the concentrations of the parameters as a result of adding 10, 50, 500, and 1000 ng/mL biotin to all ELISA parameters in healthy and obese groups are shown in the graph. (A) The graph was shown ghrelin sandwich parameter biotin interference. (B) The graph was shown ghrelin competitive parameter biotin interference. (C) The graph was shown acylated ghrelin parameter biotin interference. (D) The graph was shown deacyl ghrelin parameter biotin interference. (E) The graph was shown GHSR parameter biotin interference. (F) The graph was shown for leptin parameter biotin interference. (G) The graph was shown LEPR parameter biotin interference. (H) The graph was shown KISS1 sandwich parameter biotin interference. (I) The graph was shown KISS1 competitive parameter biotin interference. (J) The graph was shown KISSR parameter biotin interference. (K) The graph was shown NLRP3 parameter biotin interference. (L) The graph was shown preptin parameter biotin interference. (M) The graph was shown PPARγ parameter biotin interference. (N) The graph was shown IL‐18 parameter biotin interference.
FIGURE 5.

Biotin concentration in obese and healthy subgroups graph.
5. Discussion
Biotin, which is often found at high levels in multivitamins, including supplements used for hair, skin, and nail growth, can significantly interfere with certain laboratory tests and cause undetectable false high results [6]. In a study performed in children and adolescents, no relationship was found between dietary biotin intake or obesity and adiposity [25]. In addition, it has been shown that there was no change in blood biotin levels before or after bariatric surgery in obese adults [26]. In our study, no significant difference was found between the control and obese groups in terms of the biotin concentration studied via the competitive method.
Normal biotin levels in blood are lower in children than in adults. The normal range of serum biotin levels in children aged 0–4 years is 0.047–0.22 ng/mL, whereas it is 0.084–0.205 ng/mL in adults [27]. Normal serum biotin concentrations in children aged 6–12 years were found as the mean ± SD (0.14 ± 0.05 ng/mL) with a median of 0.1299 ng/mL [28]. In our study, there was a significant difference between the biotin concentrations measured by the biotin sandwich ELISA kit and the competitive ELISA kit. High biotin concentrations were observed with the sandwich ELISA kit, as shown in Table 2. The biotin values given in the above reports were evaluated. The range of normal biotin values in children's blood was compatible with the biotin values measured with the competitive kit. This finding showed that there may be severe interference in the biotin sandwich ELISA kit and that the biotin concentrations measured by the competitive method can provide more accurate results. The biotin sandwich kit used in our study does not have a pre‐dilution procedure. The fact that the measured parameter was directly biotin and that biotin was already used in the ELISA method caused a serious interference. However, the biotin‐competitive kit has pre‐dilution added to its procedure. This suggests that the biotin‐competitive kit was validated more accurately to reduce the interference effect and is therefore a better‐quality kit.
According to a report by the American Association for Clinical Chemistry (AACC), the results indicating biotin interference in laboratory tests included TSH and other thyroid function tests and endocrine tests such as PTH, ACTH, prolactin, testosterone, and cortisol. This report also described the biotin interference thresholds or biotin‐mediated interference levels of clinical tests for laboratory tests using immunologic methods by many manufacturers, such as Beckman Coulter Inc., Roche Diagnostics, Ortho Clinical Diagnostics, and Siemens [29].
In 2016, according to the evaluation of the 8 most popular immunoassay analyzers used in the United States, it was shown that exogenous biotin in the sample can cause more than ±10% biotin interference in test results from sandwich and competitive methods [30]. In another study, it was shown that high‐dose biotin caused false high results in cortisol, cyclosporine A, digitoxin, tacrolimus, FT3 and FT4 competitive immunological tests and false low results in C‐peptide, insulin, TSH, NT‐proBNP, high‐sensitivity troponin T, HIV, procalcitonin and b‐HCG sandwich immunological tests [31]. False negative results in troponin T tests are vital for patients, and newer cardiac troponin T test formulations are now being sought to have a much higher biotin interference threshold [32]. In a study of acute myocardial infarction patients, the probability of biotin interaction giving false negative results to the Elecsys Troponin T Gen 5 (TnT Gen 5; marketed outside the United States [US] as Elecsys Troponin T‐high sensitive; Roche Diagnostics International Ltd., Rotkreuz, Switzerland) test was found to be very low [33]. Dilution and biotin removal procedures could be used to reduce biotin interference in immunoassays [34].
Bonferroni correction and the Mann–Whitney U test were applied to evaluate the subgroups, as shown in Table 3. According to the biotin sandwich ELISA results, biotin levels were generally greater in the healthy group than in the obese group, while in the 5th group, biotin levels were greater in the obese group than in the healthy group. With respect to the biotin competitive parameter, in some subgroups, the value in the healthy group was greater than that in the obese group, whereas in the other groups, the value was greater than that in the healthy group. There was an apparent discrepancy. This finding suggested that biotin interference in the kits may lead to erroneous results. Kabiri et al. demonstrated that biotin interference in automated systems can lead to both false low and false high results in the sandwich method [35]. Although it is widely reported that subphysiologic doses of biotin typically cause false‐positive results in competitive assays and false‐negative results in sandwich assays [31, 36], our spike‐and‐recovery analyses demonstrated that both false low and false high results may occur in both assay formats when using manual ELISA kits. This finding suggests that biotin interference may not always follow the expected pattern, particularly in manually operated systems, and highlights the importance of assay‐specific validation when interpreting results in the presence of biotin.
Dasgupta et al. demonstrated that in the Cobas E‐411 (Roche Diagnostics) immunoassay IL‐6 test, maximum negative interference was observed with 100 ng/mL biotin, but no interference occurred with 1000 ng/mL biotin [37]. Knudsen et al. demonstrated that aldosterone and renin analyses performed in automated systems showed a very high rate of interference with biotin at doses of 100–500 ng/mL, while insulin‐like growth factor 1, growth hormone, and bone alkaline phosphatase analyses were less sensitive to biotin doses in this range [38]. Choi et al. showed that biotin interference is still present in second and third‐generation Elecsys (Roche Diagnostics) free thyroxine immunoassay tests updated by the company [39]. Biotin levels are not measured in all laboratories, which makes the problem of biotin interference difficult to solve. Dilution tests and neutralization analyses may help to solve this problem [40]. Numerous studies have demonstrated biotin interference in routine laboratory tests. This poses a serious risk for patients in the diagnosis and treatment processes.
In this study, we adhered to standard practices in interference testing by comparing results from spiked samples with those from unspiked baseline samples. For the biotin ELISA kits, we used a true 0 ng/mL biotin concentration as the baseline and assessed the effects of increasing biotin levels (10, 25, 50, 100, 250, 500, 1000, and 1200 ng/mL). In contrast, for the other ELISA‐based assays, we began spiking at 10 ng/mL rather than 0 ng/mL. This decision was influenced by practical limitations, including restricted sample volume and budget constraints. Although the threshold for biotin interference can vary depending on the type of immunoassay used, previous studies have reported that 10 ng/mL is a sufficient concentration to induce measurable interference in automated systems [29, 41]. Based on this evidence, we selected 10 ng/mL as the lowest spiking concentration for these assays. While this approach allowed us to focus on clinically relevant concentrations, we acknowledge that the absence of a true zero baseline in these cases is a limitation. Similar to our study, Katzman et al. evaluated biotin interference in the waste blood of adult patients. They measured biotin concentrations in blood using liquid chromatography tandem mass spectrometry and showed that patients were taking high levels of biotin supplements, which caused severe biotin interference in many immunoassays [41]. One of the limitations of our study is that we did not have the opportunity to use mass spectrometry due to budget constraints. However, conducting the study in children and demonstrating that biotin interference may also occur in research kits, especially in obesity parameters, will make a significant contribution to the literature.
For all the ELISA parameters, biotin % bias values between the groups formed by the addition of biotin at concentrations of 10, 50, 500, and 1000 ng/mL were calculated [4, 33, 35]. In our study, there was biotin interference in all ELISA parameters commonly used in investigations performed in the field of pediatric obesity, and this interference was found to be greater in the healthy group. In addition, the addition of 50, 500, or 1000 ng/mL biotin to the studied parameters created both negative and positive interferences. The % bias values for all parameters ranged between 1.4% and 113.6% for the healthy group and 0.17% and 49.2% for the obese group. For example, for all three biotin concentrations of the leptin kit, the % bias value in both the healthy and obese groups was < 5.6, the % bias value of the GHSR kit in the healthy group was < 5.7, the % bias value of the LEPR kit in the healthy group was < 11.6, and that in the obese group was < 3.4. The limitations of our study are that not all parameters can be compared with both sandwich and competitive methods, and the study cannot be conducted with all ELISA brands on the market, and the sample size is limited. This is because there are no ELISA kits on the market that use the competitive method for all the parameters we examined, and the budget is limited. Because of the relationship between biotin and microbiome diversity in severe obesity, giving biotin supplements to obese individuals may increase the risk of biotin interference in immunoassays. It is very important that our study is the first study showing biotin interference in pediatric obesity. It will contribute to raising awareness among scientists and healthcare professionals on this issue. Specifically, biotin interference is closely related to the quality of the ELISA kit, the parameters studied, and the concentration of biotin in the blood. If manufacturers review the quality control procedures, accuracy, recovery, and reproducibility analyses for interference in ELISA kits during the validation phase and make the necessary applications to prevent interference independently of cost calculations, this will lead to the production of higher quality kits. Thus, researchers can perform more accurate analyses and contribute to science.
6. Conclusion
In addition to ELISA kits used in routine practice, it has been shown that biotin interference can also occur in manual ELISA kits used in scientific research and can cause erroneous test results. Therefore, patients should be asked whether they use biotin supplements during scientific studies or routine analyses, and precautions should be taken to reduce the effect of biotin interference. ELISA kit manufacturers should take biotin interference into account. In order to reduce interference in kits, pre‐dilution, use of biotin‐binding nanoparticles, measurement of biotin concentration in samples with rapid tests, and ELISA kit manufacturers taking precautions by determining the cut‐off values of biotin concentrations in blood that create biotin interference may be useful.
In order to create a handbook to reveal the negative effects of biotin in laboratory studies, manual ELISA kits with many different brands and different parameters should be used, and the results obtained should be compared and supported with advanced techniques such as polymerase chain reaction (PCR) or mass spectrometry. Only in this way will it be possible to make more sensitive and accurate measurements in the analysis of many studies conducted to contribute to science.
Author Contributions
All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.
Ethics Statement
Prior to the study, ethics approval was obtained from the institutional Ethics Committee (Approval No: 2020.01.01.01).
Consent
The families of the participating children, particularly their parents, were informed about the study. Following a detailed pediatric physical examination and explanation of the study procedures, written informed consent was obtained from the parents, and verbal assent was obtained from the children.
Conflicts of Interest
The authors declare no conflicts of interest.
Funding: The present work was supported by the Research Fund of Istanbul University‐Cerrahpasa, Project No: 34574; Beykent University. The funding organization played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.
Data Availability Statement
Due to ethical and privacy considerations, the data are not publicly available.
References
- 1. Donma M. M., Donma O., Michalke B., Halbach S., and Nischwitz V., Vitamins, Minerals and Metabolic Pathways in Health and Disease (Istanbul University‐Cerrahpaşa, 2012). [Google Scholar]
- 2. Otten J. J., Hellwig J. P., and Meyers L. D., eds., “Biotin.” In Dietary Reference Intakes: The Essential Guide to Nutrient Requirements (The National Academies Press, 2006), 196–201, https://www.nap.edu/read/11537/chapter/20. [Google Scholar]
- 3. Basok B. I., “Understanding and Managing Biotin Interference in Immunoassays,” International Journal of Medical Biochemistry 3, no. 1 (2020): 1–2. [Google Scholar]
- 4. Stieglitz H. M. and Korpi‐Steiner N., “Characterization of Biotin Interference in 21 Vitros 5600 Immunoassays and Risk Mitigation for Patient Safety at a Large Academic Medical Center,” Clinical Biochemistry 75 (2020): 53–61, 10.1016/j.clinbiochem.2019.09.006. [DOI] [PubMed] [Google Scholar]
- 5. Samarasinghe S., Meah F., Singh V., et al., “Biotin Interference With Routine Clinical Immunoassays: Understand the Causes and Mitigate the Risks,” Endocrine Practice 23 (2017): 989–996, https://pubmed.ncbi.nlm.nih.gov/28534685/. [DOI] [PubMed] [Google Scholar]
- 6. FDA , “UPDATE: The FDA Warns That Biotin May Interfere With Lab Tests: FDA Safety Communication,” (2019), 3–5, https://www.fda.gov/medical‐devices/safety‐communications/update‐fda‐warns‐biotin‐may‐interfere‐lab‐tests‐fda‐safety‐communication.
- 7. Avery G., “Biotin Interference in Immunoassay: A Review for the Laboratory Scientist,” Annals of Clinical Biochemistry 56, no. 4 (2019): 424–430, https://journals.sagepub.com/doi/10.1177/0004563219842231. [DOI] [PubMed] [Google Scholar]
- 8. Voland L., Debédat J., Belda E., et al., “Severe Obesity Is Associated With Altered Gut Microbiota Biotin Metabolism and Host Biotin Status,” FASEB Journal 35, no. S1 (2021): 2007, https://onlinelibrary.wiley.com/doi/full/10.1096/fasebj.2021.35.S1.02007. [Google Scholar]
- 9. Belda E., Voland L., Tremaroli V., et al., “Impairment of Gut Microbial Biotin Metabolism and Host Biotin Status in Severe Obesity: Effect of Biotin and Prebiotic Supplementation on Improved Metabolism,” Gut 71 (2022): 2463–2480. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Voland L., Yuan M., Lecoutre S., et al., “Tissue Pleiotropic Effect of Biotin and Prebiotic Supplementation in Established Obesity,” American Journal of Physiology. Endocrinology and Metabolism 325, no. 4 (2023): E390–E405. [DOI] [PubMed] [Google Scholar]
- 11. Engin A., “The Definition and Prevalence of Obesity and Metabolic Syndrome,” in Advances in Experimental Medicine and Biology (Springer New York LLC, 2017), 1–17. [DOI] [PubMed] [Google Scholar]
- 12. Kumar S. and Kelly A. S., “Review of Childhood Obesity: From Epidemiology, Etiology, and Comorbidities to Clinical Assessment and Treatment,” Mayo Clinic Proceedings 92, no. 2 (2017): 251–265, 10.1016/j.mayocp.2016.09.017. [DOI] [PubMed] [Google Scholar]
- 13. Donma O., Donma M. M., Demirkol M., et al., “Laboratory Indices in Late Childhood Obesity: The Importance of DONMA Indices,” (2020), http://publications.waset.org/10005008/laboratory‐indices‐in‐late‐childhood‐obesity‐the‐importance‐of‐donma‐indices.
- 14. Martinho L. C. A. P., “Programming of Obesity and Pediatric Metabolic Syndromes: A Review in Light of Epigenetic Modulation,” International Journal of Nutrology 15, no. 1 (2021): 1–6. [Google Scholar]
- 15. Weihrauch‐Blüher S., Schwarz P., and Klusmann J. H., “Childhood Obesity: Increased Risk for Cardiometabolic Disease and Cancer in Adulthood,” Metabolism 92 (2019): 147–152, https://pubmed.ncbi.nlm.nih.gov/30529454/. [DOI] [PubMed] [Google Scholar]
- 16. Skinner A. C., Perrin E. M., Moss L. A., and Skelton J. A., “Cardiometabolic Risks and Severity of Obesity in Children and Young Adults,” New England Journal of Medicine 373, no. 14 (2015): 1307–1317, https://pubmed.ncbi.nlm.nih.gov/26422721/. [DOI] [PubMed] [Google Scholar]
- 17. WHO , “Obesity and Overweight,” (2021), https://www.who.int/news‐room/fact‐sheets/detail/obesity‐and‐overweight#cms.
- 18. Nittari G., Scuri S., Gamo Sagaro G., Petrelli F., and Grappasonni I., “Epidemiology of Obesity in Children and Adolescents,” In Teamwork in Healthcare (IntechOpen, 2021), 10.5772/intechopen.93604. [DOI] [Google Scholar]
- 19. Johnson L. and Li D., “Strategies to Investigate Biotin Interference in Light of the FDA Safety Communication,” Journal of Applied Laboratory Medicine 3, no. 5 (2019): 914–915. [DOI] [PubMed] [Google Scholar]
- 20. Luong J. H. T., Male K. B., and Glennon J. D., “Biotin Interference in Immunoassays Based on Biotin‐Strept(Avidin) Chemistry: An Emerging Threat,” Biotechnology Advances 37 (2019): 634–641, https://pubmed.ncbi.nlm.nih.gov/30872068/. [DOI] [PubMed] [Google Scholar]
- 21. Piketty M. L., Prie D., Sedel F., et al., “High‐Dose Biotin Therapy Leading to False Biochemical Endocrine Profiles: Validation of a Simple Method to Overcome Biotin Interference,” Clinical Chemistry and Laboratory Medicine 55, no. 6 (2017): 817–825, https://www.degruyter.com/document/doi/10.1515/cclm‐2016‐1183/html. [DOI] [PubMed] [Google Scholar]
- 22. Öncül Ü., Eminoğlu F. T., Köse E., Doğan Ö., Özsu E., and Aycan Z., “Serum Biotin Interference: A Troublemaker in Hormone Immunoassays,” Clinical Biochemistry 99 (2022): 97–102. [DOI] [PubMed] [Google Scholar]
- 23. Balieiro Neto G., Engracia Filho J. R., Budino F. E. L., Freitas A. W. P., and Soares W. V. B., “Effects of High‐Biotin Sample Interference on Antibody Concentrations in Sandwich Immunoassays,” Vaccine 11, no. 11 (2023): 1627, 10.3390/vaccines11111627. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Nelson H. A., La'ulu S. L., Lu J., and Doyle K., “Mitigation of Biotin Interference in Manual and Automated Immunoassays by Preconjugating Biotinylated Antibodies to the Streptavidin Surface as an Alternative to Biotin Depletion,” Journal of Applied Laboratory Medicine 7, no. 3 (2022): 762–775, 10.1093/jalm/jfab169. [DOI] [PubMed] [Google Scholar]
- 25. Taleban R., Heidari‐Beni M., Qorbani M., et al., “Is Dietary Vitamin B Intake Associated With Weight Disorders in Children and Adolescents? The Weight Disorders Survey of the CASPIAN‐IV Study,” Health Promotion Perspective 9, no. 4 (2019): 299–306, https://pubmed.ncbi.nlm.nih.gov/31777710/. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Bodunova N. A., Askerkhanov R. G., Khatkov I. E., et al., “Impact of Bariatric Surgery on Vitamin Metabolisms in Obese Patients,” Terapevticheskiĭ Arkhiv 87, no. 2 (2015): 70–76. [DOI] [PubMed] [Google Scholar]
- 27. Wakabayashi K., Kodama H., Ogawa E., Sato Y., Motoyama K., and Suzuki M., “Serum Biotin in Japanese Children: Enzyme‐Linked Immunosorbent Assay Measurement,” Pediatrics International 58, no. 9 (2016): 872–876, https://www.researchgate.net/publication/296686094_Serum_Biotin_Levels_in_Japanese_Children_Analyzed_by_ELISA. [DOI] [PubMed] [Google Scholar]
- 28. Sakurai‐Yageta M., Mashimo Y., Kuroishi T., et al., “Association Between Serum Biotin Levels and Cedar Pollinosis in Japanese Schoolchildren,” Journal of Nutritional Science and Vitaminology 67, no. 4 (2021): 211–216, https://pubmed.ncbi.nlm.nih.gov/34470995/. [DOI] [PubMed] [Google Scholar]
- 29. Li D., Ferguson A., Cervinski M. A., Lynch K. L., and Kyle P. B., “AACC Guidance Document on Biotin Interference in Laboratory Tests,” Journal of Applied Laboratory Medicine 5, no. 3 (2020): 575–587. [DOI] [PubMed] [Google Scholar]
- 30. Holmes E. W., Samarasinghe S., Emanuele M. A., and Meah F., “Biotin Interference in Clinical Immunoassays: A Cause for Concern,” Archives of Pathology & Laboratory Medicine 141 (2017): 1459–1460, 10.5858/arpa.2017-0107-LE. [DOI] [PubMed] [Google Scholar]
- 31. Mrosewski I., Urbank M., Stauch T., and Switkowski R., “Interference From High‐Dose Biotin Intake in Immunoassays for Potentially Time‐Critical Analytes by Roche Evaluation of a Countermeasure for Worst‐Case Scenarios,” Archives of Pathology & Laboratory Medicine 144, no. 9 (2020): 1108–1117, 10.5858/arpa.2019-0425-OA. [DOI] [PubMed] [Google Scholar]
- 32. Collinson P., “Biotin Interference in Cardiac Troponin Immunoassay ‐ Where the Wild Things Are?,” Clinical Chemistry and Laboratory Medicine 58, no. 11 (2020): 1769–1771, https://pubmed.ncbi.nlm.nih.gov/32764161/. [DOI] [PubMed] [Google Scholar]
- 33. Mumma B., Diercks D., Twerenbold R., et al., “Clinical Risk Assessment of Biotin Interference With a High‐Sensitivity Cardiac Troponin T Assay,” Clinical Chemistry and Laboratory Medicine 58, no. 11 (2020): 1931–1940, https://www.degruyter.com/document/doi/10.1515/cclm‐2019‐0962/html. [DOI] [PubMed] [Google Scholar]
- 34. Zaninotto M. and Plebani M., “Understanding and Managing Interferences in Clinical Laboratory Assays: The Role of Laboratory Professionals,” Clinical Chemistry and Laboratory Medicine 58, no. 3 (2020): 350–356. [DOI] [PubMed] [Google Scholar]
- 35. Kabiri P., Weiskirchen R., and van Helden J., “The Biotin Interference Within Interference Suppressed Immunoassays,” Journal of Clinical Laboratory Analysis 35, no. 9 (2021): e23940, https://onlinelibrary.wiley.com/doi/full/10.1002/jcla.23940. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36. Moerman A. and Delanghe R. J., Sense and Nonsense Concerning Biotin Interference in Laboratory Tests [Internet]. Acta Clinica Belgica: International Journal of Clinical and Laboratory Medicine (Taylor and Francis Ltd., 2020), https://pubmed.ncbi.nlm.nih.gov/32567529/. [DOI] [PubMed] [Google Scholar]
- 37. Dasgupta A., Khalil S., Paddock T., and Wahed A., “Technical Note: Bimodal Negative Interference of Biotin in Roche IL‐6 Immunoassay,” Annals of Clinical and Laboratory Science 52 (2022): 161–163, https://pubmed.ncbi.nlm.nih.gov/35181630/. [PubMed] [Google Scholar]
- 38. Knudsen C. S., Adelborg K., Søndergaard E., and Parkner T., “Biotin Interference in Routine IDS‐iSYS Immunoassays for Aldosterone, Renin, Insulin‐Like Growth Factor 1, Growth Hormone and Bone Alkaline Phosphatase,” 82, no. 1 (2021): 6–11, https://www.tandfonline.com/doi/abs/10.1080/00365513.2021.2003854. [DOI] [PubMed] [Google Scholar]
- 39. Choi J. and Yun S. G., “Comparison of Biotin Interference in Second‐ and Third‐Generation Roche Free Thyroxine Immunoassays,” Annals of Laboratory Medicine 40, no. 3 (2020): 274–276, 10.3343/alm.2020.40.3.274. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40. Ghazal K., Brabant S., Prie D., and Piketty M. L., “Hormone Immunoassay Interference: A 2021 Update,” Annals of Laboratory Medicine 42, no. 1 (2022): 3–23, 10.3343/alm.2022.42.1.3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41. Katzman B. M., Lueke A. J., Donato L. J., Jaffe A. S., and Baumann N. A., “Prevalence of Biotin Supplement Usage in Outpatients and Plasma Biotin Concentrations in Patients Presenting to the Emergency Department,” Clinical Biochemistry 60 (2018): 11–16, https://pubmed.ncbi.nlm.nih.gov/30036510/. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Due to ethical and privacy considerations, the data are not publicly available.