Abstract
Continuous glucose monitoring (CGM) has transformed diabetes mellitus management, evolving from a supportive monitoring tool to a central pillar of care. For people with type 1 diabetes and many insulin-treated individuals with type 2 diabetes, CGM now directly informs treatment decisions, especially when integrated with automated insulin delivery (AID) systems. In these hybrid closed-loop systems, sensor data drives real-time insulin adjustments, meaning that accuracy is not just a matter of measurement quality; it is a matter of patient safety. However, the primary accuracy measure currently used, the mean absolute relative difference (MARD), is increasingly inadequate for guiding clinical decisions. MARD offers a single averaged number under controlled conditions, but it does not capture the timing, direction, or clinical consequences of sensor errors. This is particularly problematic in AID systems, where even minor inaccuracies may lead to inappropriate insulin dosing, increasing the risk of hypoglycemia or hyperglycemia. Given the centrality of CGM in modern diabetes care, a more comprehensive evaluation approach is urgently needed, one that reflects real-world performance, prioritizes patient safety, and addresses the diverse contexts in which CGM devices are used. This editorial presents an opinion-based perspective, proposing a four-dimensional framework for CGM evaluation beyond the traditional reliance on MARD.
Keywords: continuous glucose monitoring (cgm), endocrinology and diabetes, mard, measurement accuracy, safety study
Editorial
Continuous glucose monitoring (CGM) systems have become indispensable in the management of type 1 diabetes and are increasingly used in insulin-treated type 2 diabetes. While the mean absolute relative difference (MARD) remains the standardly used index of sensor accuracy, its limitations are well recognized. MARD originated as a simple statistical index in early validation studies of CGM devices, which is the average percentage error between CGM readings and a reference blood glucose value (e.g., capillary glucometer, venous plasma glucose, or laboratory assays). However, MARD is typically derived under ideal laboratory conditions and does not capture the timing, direction, or clinical consequences of sensor deviations [1].
In automated insulin delivery (AID) and hybrid closed-loop (HCL) systems, even slight inaccuracies can lead to inappropriate insulin dosing decisions. Undetected nocturnal hypoglycemia, overestimated postprandial values, and delayed detection of exercise-induced hypoglycemia have all been documented as safety risks [1].
We propose moving beyond single-number metrics and adopting a four-dimensional evaluation framework that encompasses safety, performance, usability, and equity (which are largely interdependent), dimensions that cannot be adequately captured by any single metric, which together ensure that no critical aspect is overlooked (see CGM metrics in Table 1). Safety reflects the ability to prevent harmful errors such as missed hypoglycemia; performance relates to overall analytical accuracy across glucose ranges; usability addresses patient experience and integration into daily life; and equity ensures that benefits are accessible across diverse populations. Together, these four pillars may provide a more balanced assessment of CGM than MARD alone, or at the very least foster a constructive debate aimed at overcoming current limitations.
Table 1. Classical and proposed CGM metrics.
Summary of classical and emerging continuous glucose monitoring (CGM) metrics. MARD (Mean Absolute Relative Difference), GRI (Glycemic Risk Index), SEG (Surveillance Error Grid), and PARD (Precision Absolute Relative Difference) are used to assess accuracy and clinical risk. At the same time, TIR (Time in Range) measures time within the target glucose range. Factory-calibrated CGM systems, along with optional user calibration, address device accuracy under various conditions. Usability metrics capture patient-reported ease of use and satisfaction with the system.
T1D: type 1 diabetes; T2D: type 2 diabetes; SBGM: self blood glucose monitoring (with a glucometer).
| Metric | Definition | Strengths | Limitations |
| MARD | Mean absolute relative difference between CGM readings and a reference | Easy to calculate; historical benchmark | Ignores timing, direction, and clinical consequences |
| GRI | Composite index quantifying glycemic risk | Integrates accuracy and clinical impact | Less familiar; more complex to calculate |
| SEG | Grid-based classification of errors by clinical severity | Provides clinical context for errors | Requires detailed data and validation |
| PARD | Point-by-point precision, absolute and relative difference | Sensitive to short-term variations | Lacks temporal and clinical context |
| TIR | Percentage of time in target glucose range (at least 70% of readings between 70 and 180 mg/dL [3.9–10 mmol/L] for adults with T1D or T2D) | Strongly linked to long-term clinical outcomes | Does not measure accuracy directly |
| Factory-calibrated CGM | Pre-calibrated by the manufacturer | Reduces user error, improves consistency | Cost and technology considerations |
| Optional user calibration (SBGM) | Manual calibration with capillary glucose when needed | Restores accuracy in specific circumstances; boosts user confidence | Requires correct technique and timing |
| Usability metrics | Patient-reported ease of use and experience | Direct impact on adherence and satisfaction | Subjective and population-dependent |
Safety
Beyond MARD, other indices, introduced through academic and clinical research, include the Glycemic Risk Index (GRI), which provides a composite score of hypo- and hyperglycemia risk; the Surveillance Error Grid (SEG), which maps measurement errors to graded clinical risk zones; and the Precision Absolute Relative Difference (PARD), which captures the dispersion of accuracy. Additionally, hypoglycemia detection rates are recognized as a pragmatic measure of safety. These metrics are derived automatically from CGM datasets, with algorithm development informed by expert input and methodological validation in clinical studies. Routine calculation is fully automated once the algorithm is in place. Despite their stronger clinical meaning, these metrics remain clinically underused because regulatory standards, ISO 15197:2013 (International Organization for Standardization), CLSI (Clinical and Laboratory Standards Institute), FDA (U.S. Food and Drug Administration), and EMA (European Medicines Agency), have maintained MARD as the minimum required metric for CGM commercialization, likely for historical reasons and inertia, as well as because it is easy to calculate and straightforward to communicate as an average percentage error [1-4].
Performance
Time in range (TIR) and glycated hemoglobin (HbA1c) improvements have a strong evidence base linking them to reduced risk of microvascular and macrovascular complications [5]. Performance and safety are interdependent with each other and with the metrics mentioned above. Moreover, modern factory-calibrated CGMs, pre-calibrated by the manufacturer, have demonstrated improved consistency and reduced user-dependent variability in real-world use. Some models also allow optional user-initiated calibration using a capillary glucose measurement through self blood glucose monitoring (SBGM). This feature can be valuable for restoring accuracy in situations such as the physiological lag between blood and interstitial glucose, sensor compression effects (e.g., during sleep), reduced accuracy during the early wear-in period with progressive improvement after the first day, altered tissue perfusion, drug interferences (e.g., acetaminophen, ascorbic acid), calibration issues (e.g., errors from inaccurate fingerstick references or calibrating during rapid glucose changes), and sensor drift (progressive loss of accuracy over time), thus providing an additional safeguard in daily management [6].
Usability
User-friendly design and reduction of alarm fatigue are crucial determinants of adherence. In a large multicenter survey, CGM users reported high quality-of-life scores (87/100) but only moderate usability (66/100), with persistent unmet needs such as seamless real-time data sharing with healthcare providers. Usability also depends on practical aspects such as ease of sensor insertion, comfort during wear, device size and visibility, battery longevity, and intuitive app interfaces for data visualization and interpretation. As noted above, calibration can be either factory-set, in which case it is explicitly stated by the manufacturer, or manual, requiring fingerstick references. Some devices also allow an optional manual calibration, which, when used appropriately, may not only enhance accuracy but also strengthen user confidence, particularly in scenarios where precise readings are critical, such as insulin dosing decisions [7-9].
Equity
Performance consistency must be demonstrated across diverse skin tones, pediatric and geriatric populations, pregnancy, and comorbid conditions such as chronic kidney disease or anemia, which may interfere with glucose dynamics. An international consensus now emphasizes the importance of including heterogeneous populations in CGM validation studies, not only in terms of biological and clinical features but also socioeconomic status, education level, and access to technology. This broader inclusivity is essential to ensure that CGM devices are reliable and equitable for all patients in real-world diabetes care, and to prevent widening of health disparities in already vulnerable groups [7-10].
Several professional organizations have already expressed support for a multidimensional approach to CGM evaluation. The American Diabetes Association (ADA), in its Standards of Care in Diabetes 2025, recommends assessing CGM systems with a broad set of clinical and technical metrics, including TIR, hypoglycemia detection, and usability across diverse populations. The European Association for the Study of Diabetes (EASD), in joint consensus statements with the ADA, likewise emphasizes reporting a complete panel of CGM-derived metrics for both research and clinical care. Similarly, the Advanced Technologies & Treatments for Diabetes (ATTD) consensus reports call for frameworks that consider device safety, measurement precision, user-friendliness, and equitable access. Notably, while leading professional organizations such as the ADA, EASD, and ATTD have already endorsed the need for multidimensional CGM evaluation, these recommendations remain largely aspirational. To date, no central regulatory body or reimbursement authority has operationalized these frameworks into policy or procurement criteria. This gap between consensus and implementation exposes patients to avoidable risks and limits the impact of technological advances [7-9].
This work is based on our previously posted preprint opinion, "Continuous Glucose Monitoring Metrics Are Failing People With Diabetes Mellitus," dated July 29, 2025, available on the Authorea and Zenodo platforms.
Conclusion
CGM systems have significantly improved the prognosis of diabetes. However, current devices remain suboptimal in terms of safety, reliability, and accessibility, as reliance on a single accuracy metric (MARD) is insufficient. In this editorial, we propose a four-dimensional framework, encompassing safety, performance, usability, and equity, as an opinion-based perspective, which aligns with professional society recommendations. This framework does not replace existing metrics but calls regulators, payers, and manufacturers to broaden evaluation standards. We believe such an approach could help overcome current limitations, foster safer and more equitable CGM adoption, and support integration with the next generation of AID systems. The time has come to move beyond MARD and adopt a multidimensional approach that truly reflects the needs of people living with diabetes.
Disclosures
Conflicts of interest: In compliance with the ICMJE uniform disclosure form, all authors declare the following:
Payment/services info: All authors have declared that no financial support was received from any organization for the submitted work.
Financial relationships: All authors have declared that they have no financial relationships at present or within the previous three years with any organizations that might have an interest in the submitted work.
Other relationships: All authors have declared that there are no other relationships or activities that could appear to have influenced the submitted work.
Author Contributions
Concept and design: Matteo Acanfora, Erika Pedone, Luca Acanfora, Francesca Ré, Barbara Presciuttini
Acquisition, analysis, or interpretation of data: Matteo Acanfora, Erika Pedone, Luca Acanfora, Francesca Ré, Barbara Presciuttini
Drafting of the manuscript: Matteo Acanfora, Erika Pedone
Critical review of the manuscript for important intellectual content: Matteo Acanfora, Erika Pedone, Luca Acanfora, Francesca Ré, Barbara Presciuttini
Supervision: Barbara Presciuttini
References
- 1. Benefits and limitations of MARD as a performance parameter for continuous glucose monitoring in the interstitial space. Heinemann L, Schoemaker M, Schmelzeisen-Redecker G, et al. https://doi.org/10.1177/1932296819855670. J Diabetes Sci Technol. 2020;14:135–150. doi: 10.1177/1932296819855670. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.A Glycemia Risk Index (GRI) of hypoglycemia and hyperglycemia for continuous glucose monitoring validated by clinician ratings. Klonoff DC, Wang J, Rodbard D, et al. https://doi.org/10.1177/19322968221085273. J Diabetes Sci Technol. 2023;17:1226–1242. doi: 10.1177/19322968221085273. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.The surveillance error grid. Klonoff DC, Lias C, Vigersky R, et al. https://doi.org/10.1177/1932296814539589. J Diabetes Sci Technol. 2014;8:658–672. doi: 10.1177/1932296814539589. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Performance evaluations of continuous glucose monitoring systems: precision absolute relative deviation is part of the assessment. Obermaier K, Schmelzeisen-Redeker G, Schoemaker M, Klötzer HM, Kirchsteiger H, Eikmeier H, del Re L. https://doi.org/10.1177/193229681300700404. J Diabetes Sci Technol. 2013;7:824–832. doi: 10.1177/193229681300700404. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Validation of time in range as an outcome measure for diabetes clinical trials. Beck RW, Bergenstal RM, Riddlesworth TD, Kollman C, Li Z, Brown AS, Close KL. https://doi.org/10.2337/dc18-1444. Diabetes Care. 2019;42:400–405. doi: 10.2337/dc18-1444. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Performance of a factory-calibrated real-time continuous glucose monitoring system utilizing an automated sensor applicator. Shah VN, Laffel LM, Wadwa RP, Garg SK. https://doi.org/10.1089/dia.2018.0143. Diabetes Technol Ther. 2018;20:428–433. doi: 10.1089/dia.2018.0143. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.7. Diabetes Technology: Standards of Care in Diabetes-2025. https://doi.org/10.2337/dc25-S007. Diabetes Care. 2025;48:0–66. doi: 10.2337/dc25-S007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.The management of type 1 diabetes in adults. A consensus report by the American Diabetes Association (ADA) and the European Association for the Study of Diabetes (EASD) Holt RI, DeVries JH, Hess-Fischl A, et al. https://doi.org/10.1007/s00125-021-05568-3. Diabetologia. 2021;64:2609–2652. doi: 10.1007/s00125-021-05568-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Continuous glucose monitoring and metrics for clinical trials: an international consensus statement. Battelino T, Alexander CM, Amiel SA, et al. Lancet Diabetes Endocrinol. 2023;11:42–57. doi: 10.1016/S2213-8587(22)00319-9. [DOI] [PubMed] [Google Scholar]
- 10.Continuous glucose monitoring in pregnant women with type 1 diabetes (CONCEPTT): a multicentre international randomised controlled trial. Feig DS, Donovan LE, Corcoy R, et al. Lancet. 2017;390:2347–2359. doi: 10.1016/S0140-6736(17)32400-5. [DOI] [PMC free article] [PubMed] [Google Scholar]