. 2025 Jan 19;17(9):1502–1515. doi: 10.1002/dta.3855

Comprehensive Qualitative Drug Screening in Emergency Toxicology Using an Automated LC–MSn System:Simultaneous Quantification of Relevant Drugs and Metabolites in Blood Plasma

Selina Hemmer 1, Maximilian Ninnig 1, Lea Wagmann 1, Sascha K Manier 1, Markus R Meyer 1,
PMCID: PMC12401653  PMID: 39828535

ABSTRACT

Rapid and comprehensive qualitative and quantitative analytical procedures are crucial in 24/7 emergency toxicology (ET) to support diagnosis and treatment of acute intoxications and to monitor their progression and efficacy of detoxification strategies. This study aimed to develop the simultaneous qualitative and quantitative analysis of 62 drugs, as well as seven active metabolites in human blood plasma using an automated liquid chromatography (LC)‐linear ion trap mass spectrometry (MS) screening system. Sample preparation was conducted by liquid–liquid extraction, and plasma concentrations were determined using an electronically stored 5‐point calibration. Validation was performed according to international guidelines and recommendations for ET including selectivity, carry‐over, accuracy, precision, and matrix effects. Finally, applicability was evaluated using case samples and proficiency tests. The method demonstrated selectivity for all analytes, with no significant carry‐over or matrix effects. Accuracy and precision recommended for ET could be fulfilled for all tested analytes, except for 10 analytes. Patient plasma samples were analyzed and compared with results obtained by reference LC‐tandem MS or gas chromatography‐MS methods. Furthermore, the applicability of the method could be demonstrated. It provides a fast, robust, and reliable blood plasma screening for 69 analytes and an additional quantification of 59 analytes relevant in ET. The use of an electronically stored 5‐point calibration and a simplified “push and print solution” allows for straightforward assessment of blood plasma levels of the analytes.

Keywords: clinical toxicology, emergency toxicology, LC–MSn , method validation


Simultaneous qualitative and quantitative analysis of 62 drugs and seven active metabolites in human blood plasma by automated liquid chromatography linear ion trap mass spectrometry.

graphic file with name DTA-17-1502-g002.jpg

1. Introduction

Reliable, rapid, and cost‐effective analytical procedures are needed for 24/7 emergency toxicology (ET). The outcomes of such analyses can support clinicians in evaluating emergencies and treatment options by allowing thorough differential diagnosis and rule out acute poisoning to prevent unnecessary treatment and expenses [1, 2, 3, 4, 5]. Analyses should first include a qualitative screening procedure with confirmed identification covering a wide range of drugs [2, 3]. Second, blood levels of identified drugs should be quantified in order to assess the severity of intoxication [1, 2]. As these analyses can provide guidance for further treatment, it is essential that the duration for such analyses is kept to a minimum. Several comprehensive qualitative screening methods have been published for both blood and urine using gas chromatography (GC)‐mass spectrometry (MS) or liquid chromatography (LC) coupled to low‐ or high‐resolution MS [6, 7, 8, 9, 10, 11] and are well established in clinical and forensic toxicology. For quantification, fully validated therapeutic drug monitoring (TDM) approaches are often used, among others [12]. However, TDM approaches frequently target a specific class of drugs and are often time‐consuming, rendering them unsuitable for ET, where the analysis of a large number of drugs from different classes is required within a limited period of time. However, there are also methods available for ET that allow for screening and simultaneous or subsequent quantification. Meyer et al. developed and validated a method for blood‐level assessment of 40 relevant drugs based on an established blood‐screening procedure using 1‐point calibration and GC–MS [3]. Michely and Maurer developed a similar method using LC‐tandem MS (MS/MS) to supplement the GC–MS method [2]. Nevertheless, for both methods, the data processing was handled manually potentially entailing mistakes. Therefore, Caspar et al. developed and validated a strategy for automated quantification of several drugs and active metabolites in blood plasma based on a linear ion trap LC–MSn system with standardized LC and MS settings including eluents, column, chromatographic gradient, source conditions, MS settings, and spectral database [1].

The present study aimed to extend the method by Caspar et al. via addition of further 45 analytes often requested in ET. These analytes included classes such as analgesics, anticonvulsants, antidepressants, antiemetics, antihistamines, antihypertensives, benzodiazepines, or neuroleptics. The validation should be based on international guidelines and recommendations for 24/7 emergency toxicology [13, 14]. Finally, the proof of concept and applicability should be demonstrated by analysis of proficiency tests and patient blood samples submitted to the authors' laboratory for regular toxicological analysis.

2. Experimental Section

2.1. Chemicals and Reagents

All reference standards of the drugs and metabolites were obtained from the corresponding pharmaceutical companies. The internal standards (IS) trimipramine‐d3 and diazepam‐d5 were obtained from LGC Standards (Wesel, Germany). Ammonium formate and formic acid were obtained from Merck (Darmstadt, Germany). Acetonitrile, dimethyl sulfoxide (DMSO), and methanol (all LC–MS grade) were from VWR (Darmstadt, Germany). Water was purified with a Millipore filtration unit (18.2 Ω × cm water resistance). Pooled plasma samples were obtained from a local blood bank.

2.2. Stock Solutions, Working Solutions, Calibrators, and Quality Controls

Stock solution at 1 or 10 mg/mL in case of carbamazepine, ketamine, oxcarbazepine, and paracetamol was freshly prepared in duplicates, once for calibrator and once for quality control solutions. All substances were solved in methanol, except for oxcarbazepine that was solved in DMSO. The analytes were split into three different mixtures (see Table S1). Their working solutions were prepared by diluting the stock solutions with methanol to a concentration 10‐fold higher than the final plasma concentration (see Table 1). All solutions were stored at −20°C. Five calibrators, 1‐point calibrator and two different quality control (QC) samples containing low and high concentrations were prepared by spiking pooled human blank plasma with the prepared working solution.

TABLE 1.

Final plasma concentrations in ng/mL of the analytes in calibrators (Cal 1–5) and quality control samples (QC) low and high as well as the used weightings of linear calibration model and the therapeutic and toxic ranges to Schulz et al. [23].

Analyte Weighting Cal 1 Cal 2 Cal 3 Cal 4 Cal 5 LLOQ QC low QC high Therapeutic range Toxic above
Alprazolam 1/x2 50 125 250 375 500 50 60 400 5–50 100
Amisulpride 1/x2 100 500 1000 1500 2000 100 120 1600 100–400 640
Amitriptyline 1/x2 100 500 1000 1500 2000 100 120 1600 50–300 500
Aripiprazole Equal 250 500 1000 1500 2000 250 300 1600 100–350 1000
Biperiden 1/x2 50 125 250 375 500 50 60 1600 1.0–6.5 13
Bisoprolol Equal 100 500 1000 1500 2000 100 120 1600 10–100 200
Bromazepam Equal 1000 1500 2000 2500 3000 1000 1200 2400 50–200 300
Carbamazepine 1/x2 2500 5000 10,000 15,000 20,000 2500 3000 16,000 2000–12,000 10,000
Chlorprothixene Equal 250 500 1000 1500 2000 250 300 1600 20–300 400
Citalopram 1/x2 100 500 1000 1500 2000 100 120 1600 50–110 220
Clobazam 1/x2 100 500 1000 1500 2000 100 120 1600 30–300 500
Clozapine 1/x2 100 500 1000 1500 2000 100 120 1600 100–600 600
Codeine 1/x2 250 500 1000 1500 2000 250 300 1600 30–250 500
Desipramine Equal 250 500 1000 1500 2000 250 300 1600 10–500 500
Diazepam 1/x2 500 1000 1500 2000 2500 500 600 2000 Sum 100–2500 Sum 5000
Diazepam‐M (nor‐)/Nordiazepam+ Equal 750 1000 1500 2000 2500 750 900 2000 120–800 1500
Dihydrocodeine 1/x2 250 500 1000 1500 2000 250 300 1600 30–250 500
Diltiazem 1/x2 100 500 1000 1500 2000 100 120 1600 30–250 800
Diphenhydramine 1/x2 500 1000 1500 2000 2500 500 600 2000 50–100 1000
Doxepin 1/x2 100 500 1000 1500 2000 100 120 1600 Sum 50–150 Sum 300
Doxepin‐M (nor‐) Equal 50 125 250 375 500 50 60 400
Doxylamine 1/x2 200 500 1000 1500 2000 200 250 1600 50–200 1000
Flupirtine Equal 1000 1500 2000 2500 3000 1000 1200 2400 500–1500 3000
Haloperidol 1/x2 50 125 250 375 500 50 60 400 1–17 50
Hydromorphone 1/x2 50 125 250 375 500 50 60 400 5–30 100
Imipramine 1/x2 250 500 1000 1500 2000 250 300 1600 50–350 500
Ketamine 1/x2 1000 2500 5000 7500 10,000 1000 1200 8000 100–6000 7000
Levomepromazine 1/x2 200 500 1000 1500 2000 200 250 1600 5–200 400
Lorazepam 1/x2 500 1000 1500 2000 2500 500 600 2000 20–250 300
Maprotiline 1/x2 100 500 1000 1500 2000 100 120 1600 75–130 500
Melperone 1/x2 100 500 1000 1500 2000 100 120 1600 30–100 200
Methadone 1/x2 100 500 1000 1500 2000 100 120 1600 50–600 600
Metoclopramide 1/x2 100 500 1000 1500 2000 100 120 1600 10–150 200
Metoprolol 1/x2 100 500 1000 1500 2000 100 120 1600 20–600 7800
Mianserin 1/x2 50 125 250 375 500 50 60 400 15–70 250
Midazolam 1/x2 100 500 1000 1500 2000 100 120 1600 40–250 1000
Mirtazapine 1/x2 150 500 1000 1500 2000 150 180 1600 30–300 1000
Moclobemide Equal 500 1500 2500 4000 5000 500 600 4000 300–3000 2000
Olanzapine 1/x2 50 125 250 375 500 50 60 400 1–80 100
Opipramol 1/x2 100 500 1000 1500 2000 100 120 1600 50–500 1000
Oxazepam Equal 1000 1500 2000 2500 3000 1000 1200 2400 200–1500 2000
Oxcarbazepine 1/x2 2500 10,000 20,000 40,000 80,000 2500 3000 65,000 10,000‐35,000 35,000
Oxycodone 1/x2 250 500 1000 1500 2000 250 300 1600 5–100 200
Paracetamol 1/x2 2500 30,000 60,000 90,000 120,000 2500 3000 96,000 5000‐25,000 100,000
Paroxetine 1/x2 50 125 250 375 500 50 60 400 2–65 400
Perazine 1/x2 100 500 1000 1500 2000 100 120 1600 10–230 460
Pethidine 1/x2 500 1000 1500 2000 2500 500 600 2000 100–800 1000
Pethidine‐M (nor‐) Equal 50 125 250 375 500 50 60 400
Pipamperone Equal 250 500 1000 1500 2000 250 300 1600 100–400 500
Promethazine 1/x2 100 500 1000 1500 2000 100 120 1600 10–200 1000
Prothipendyl Equal 50 125 250 375 500 50 60 400 30–80 500
Quetiapine 1/x2 100 500 1000 1500 2000 100 120 1600 100–500 1800
Ramipril 1/x2 50 125 250 375 500 50 60 400 1–40 80
Risperidone 1/x2 50 125 250 375 500 50 60 400 Sum 20–60 Sum 120
Risperidone‐M (9‐hydroxy‐)/Paliperidone* 1/x2 100 500 1000 1500 2000 100 120 1600 20–60 120
Sertraline Equal 250 500 1000 1500 2000 250 300 1600 10–500 300
Sulpiride Equal 500 1000 1500 2000 2500 500 600 2000 50–1000 1500
Tapentadol 1/x2 100 500 1000 1500 2000 100 120 1600 10–300
Temazepam 1/x2 500 1000 1500 2000 2500 500 600 1600 20–900 1000
Tilidine 1/x2 100 500 1000 1500 2000 100 300 1600 50–300
Tilidine‐M (nor‐) 1/x2 100 500 1000 1500 2000 100 120 1600
Tramadol 1/x2 250 500 1000 1500 2000 250 300 1600 100–1000 1000
Tramadol‐M (O‐demethyl) 1/x2 100 500 1000 1500 2000 100 120 1600
Venlafaxine 1/x2 100 500 1000 1500 2000 100 120 1600 Sum 100–400 Sum 1000
Venlafaxine‐M (O‐demethyl) 1/x2 100 500 1000 1500 2000 100 120 1600
Verapamil Equal 100 500 1000 1500 2000 100 120 1600 10–400 1000
Zolpidem 1/x2 100 500 1000 1500 2000 100 120 1600 80–200 500
Zopiclone Equal 100 500 1000 1500 2000 100 120 1600 10–120 150
Zuclopenthixol 1/x2 50 125 250 375 500 50 60 400 4–100 100

Note: Sum = concentration of drug plus active metabolite; + = Nordiazepam possible parent drug or metabolite of diazepam; * = paliperidone (9‐hydroxyrisperidone) possible parent drug or metabolite of risperidone.

2.3. Blood Plasma Preparation

In accordance with minor deviations from previous published procedures [1, 2, 3, 15, 16], a liquid–liquid extraction (LLE) for sample preparation was used. One milliliter of blood plasma was mixed with 100 μL of trimipramine‐d3 (0.1 mg/mL) as IS, 2 mL saturated aqueous sodium sulfate solution and 5 mL diethyl ether‐ethyl acetate mixture (1:1, v/v). The mixture was shaken manually for 20 s and centrifuged for 2 min at 3000× g at 24°C. The upper solvent phase was transferred into a pointed flask. Second, a volume of 500 μL sodium hydroxide (1 mol/L, pH 8–9) and 5 mL diethyl ether‐ethyl acetate mixture (1:1, v/v) was added to the remaining liquid and again mixed manually for 20 s, followed by 2 min of centrifugation at 3000× g at 24°C. Upper solvent phase was again transferred into the same pointed flask and evaporated to dryness at 70°C. The combined residues were reconstituted in 100 μL methanol. A 20 μL aliquot was then diluted with 80 μL of a mixture (9:1, v/v) eluent A (2 mM aqueous ammonium formate containing 0.1% formic acid and 1% acetonitrile) and eluent B (acetonitrile containing 0.1% formic acid with 2 mM ammonium formate and 1% water), followed by 2 min of centrifugation at 21,130× g at −10°C. Five microliters of the supernatant were injected onto the LC–MSn system under the conditions described below.

2.4. LC‐MSn Settings

According to a published procedure [1], the LC‐MSn system consisted of a Dionex UltiMate 3000 LC‐system (Thermo Fisher Scientific, TF, Dreieich, Germany) and an amaZon speed ion trap mass spectrometer (Bruker Daltonik) coupled to an electrospray ionization (ESI) source. Gradient elution was performed on a TF Acclaim 120 C18 column (100 mm × 2.1 mm, 2.2 μm) using 2 mM aqueous ammonium formate containing 0.1% formic acid and 1% acetonitrile (eluent A) and acetonitrile containing 0.1% formic acid with 2 mM ammonium formate and 1% water (eluent B). Gradient and flow rate were set as follows: 0–1 min hold 1% B; 1–8 min linear increase to 95% B; 8–9 min hold at 95% B; 9–11 min hold at 1% B at constant flow rate of 0.5 mL/min. The injection volume was set to 5 μL.

The MS was used in AutoMSn mode using UltraScan in positive and negative ionization switching mode with mass range from m/z 70 to 800 at 32,500 m/z/s; ICC target, 70,000; max. Accu time, 50.00 ms. AutoMSn spectra were generated up to n = 3, if possible. Data‐dependent acquisition was performed to record MS2 and MS3 spectra according to a scheduled precursor list (SPL) containing the retention time of each analyte, and the nominal mass of the respective precursor ions for all MS stages and active exclusion was set to exclude after 1 spectra and release after 0.50 min. The following ESI parameters were used: capillary voltage, 4500 V; end plate offset, 500 V; nebulizer gas, 29.0 psi; dry gas, 10 L/min; dry temp., 320°C. The data were automatically processed, and compounds were identified via the unmodified TT library (version Toxtyper 2_0 Library, Bruker Daltonik). A result report as pdf was automatically generated by the DataAnalysis software (Version 4.4, Bruker Daltonik).

2.5. Ionization Effects of Co‐Eluting Analytes

Ion suppression and enhancement of co‐eluting analytes was tested according to published procedures [1, 2, 17]. One set of all analytes and IS in one mixture at a concentration of 1 mg/L and one set of single analyte solutions including trimipramine‐d3 were analyzed in triplicates and the IS‐normalized peak areas using diazepam‐d5 were compared. The acceptance criteria (AC) for ion suppression and enhancement were set to ± 25% [17].

2.6. Method Validation

Method validation was done according to the “Guideline on bioanalytical method validation” of the European Medicines Agency (EMA) [13] and the “Recommendations of criteria for development and validation of analytical methods for estimating concentrations of drugs in blood to be used in 24/7 clinical toxicology” of the Society of Toxicological and Forensic Chemistry (GTFCh) [14]. Statistical evaluation was performed using Microsoft Excel (version 16.84, Redmond, WA, USA), GraphPad Prism 9.00 (GraphPad Software, La Jolla, CA, USA), and R (version 2022.12.0+353) [18].

2.6.1. Selectivity and Carry‐Over

Selectivity was tested by analyzing drug‐free human blood plasma samples from six donors. Carry‐over was tested after the injection of the highest calibrator (calibrator 5) followed by blank blood plasma sample. According to the EMA guideline, absence of interfering compounds at the corresponding retention times of the analytes and carry‐over in a blank blood plasma sample are accepted where the response is less than 20% of the lower limit of quantification (LLOQ) for analytes and 5% for IS [13].

2.6.2. Linearity of the Calibration Curves and Limits of Quantification

For calibration curve, five calibrators were prepared by spiking the working solutions in pooled blank plasma at a ratio of 1:10 followed by an LLE as described above. Calibrator 1 and 5 were defined as concentrations for LLOQ and upper limit of quantification (ULOQ) for all analytes, respectively.

The determination of linearity was done using a manually programmed R script (https://github.com/sehem/Calibration_Weighting.git). The linearity was tested according to Mandel. Therefore, the respective squares of the residual standard deviations from linear and quadratic regression were used and p‐value was set to 0.05. Linear regression with several weightings (equal, 1/x, 1/x2, 1/y, 1/y2) were tested by fitting three individual curves of each analyte. Suitability of the models was evaluated using the sum of square residuals (SSR) by choosing the model with the lowest SSR and calculation of the residual sum of squares. The lowest sum of squares indicated the best weighting model (F‐test heteroscedasticity). According to the EMA guideline, the acceptance criteria (AC) for back calculation of each calibrator should be within ± 15% and ± 20% for LLOQ, respectively. At least 75% of the calibration standards must fulfill these criteria [13], which means four out of five in case of the current method.

2.6.3. Matrix Effects

Matrix effects were done according to the GTFCh recommendations [14, 19]. Therefore, two different sets of samples were prepared at QC low and QC high in six replicates. Sample set 1 contained neat analyte solution in methanol. Sample set 2 represented blank plasma from six individual donors spiked with analyte solution after extraction. For validation, the matrix factor was calculated by comparing sample set 2 to sample set 1. According to the GTFCh, the AC were set to matrix effects within ± 30% [14].

2.6.4. Accuracy and Precision

Within‐run accuracy and precision were performed by analyzing five samples at three different concentration levels (LLOQ, QC low, and QC high). Between‐run accuracy and precision were performed by analyzing five samples from three different concentration levels mentioned before on at least three different days. The AC for the validation in emergency toxicology according to the GTFCh guideline were set to ± 30% of the nominal value of the QCs, which correspond to the mean concentration of the defined concentration level for both within‐ and between‐run accuracy. For within‐ and between‐run precision, the CV had to be ≤ 30% [14].

2.7. Stability of Electronically Stored Calibration

To assess the stability of an electronically stored calibration, freshly prepared QC high extracts were analyzed over a 2‐year period. The electronically stored calibration was based on calibrators 1–5 analyzed in triplicates on the first day. The mean slopes and intercepts of each analyte were calculated using GraphPad Prism 9.00 (GraphPad Software, La Jolla, CA, USA) and transferred to a CSV file (see Table S2). The processing method is conducted using the aforementioned CSV file in order to quantify the analytes included in this study. Subsequently, the stability of the electronically stored calibration was monitored by analyzing one freshly prepared QC high sample every second week. The calculated concentrations via the stored calibration were compared with the nominal concentration levels of the QC high. The AC were set to ± 30% in accordance with the recommendations for emergency toxicology [14]. Additionally, 1‐point calibrators were used, with concentrations aligned with those of calibrator 3 (see Table 1). In the event of a decrease in instrument performance, 1‐point calibrators can be used for time saving reasons to adjust the slope of the stored calibration and thus ensure continuous reliability. For this purpose, freshly prepared 1‐point calibrators were analyzed in triplicates and evaluated data were automatically transferred to calibration information CSV file (see Table S2).

2.8. Proof of Concept and Applicability Studies

A total of 29 blood plasma samples submitted to the authors' laboratory for regular toxicological analysis and several interlaboratory tests were analyzed for proof of concept studies. Results were compared with the acceptance range of the interlaboratory tests or in case of patient samples with results obtained by reference methods using LC‐tandem MS or GC–MS [2, 3].

3. Results and Discussion

The objective of this study was to simultaneously identify and semiquantify selected analytes relevant in a 24/7 ET setting. The selection of analytes within this study was based on their relevance in ET and the frequency of request in the authors' laboratory over the past 10 years and aimed to complement those included in the paper by Caspar et al. [1]. This expansion by adding 45 analytes, deemed the need for a (re‐)validation of the procedure including proof of concept studies as, for example, the EMA mandates re‐validation when substantial alterations are made to a method, including the expansion of analyte panels [13].

The method was thus validated based on recommendations of the GTFCh [14] and EMA [13]. Some parameters were not investigated as they are not of relevance in ET such as stability of processed samples. Samples are analyzed immediately after preparation and are not stored in the autosampler. Furthermore, recovery was not evaluated, as the employed extraction method has been comprehensively validated in prior studies [1, 2, 3]. Additionally, the method focuses on the reliable determination of toxic concentrations, and minor losses during extraction have no significant impact on the method's applicability or the clinical interpretation of the results.

3.1. Method Development and Validation

3.1.1. Ionization Effects of Co‐Eluting Analytes

As the TT system consists of a fixed combination of LC and MS components and settings, eluents, column, chromatographic gradient, MS conditions, and SPL have not been modified. The chromatographic separation within 11 min allows sufficient fast analysis in an ET setting but does in turn not allow baseline separation of all 69 analytes (Figure 1). Therefore, ionization suppression or enhancement effects of co‐eluting analytes were tested by comparing the IS‐normalized peak areas from a set containing all analytes in one mixture to a set of single analyte solution. The results revealed that distributing the analytes into three working solutions was necessary to minimize ionization effects due to co‐elution (Table S1).

FIGURE 1.

FIGURE 1

Reconstructed ion chromatograms of all analytes generated by DataAnalysis after analysis of the high‐concentrated quality control sample (QC High).

3.1.2. Selectivity and Carry‐Over

Analytes and internal standards need to be distinguished from endogenous substances and other drugs. Thus, identification of the analytes was based on their retention times (range ± 0.4 min), their MS2 and/or MS3 spectrum using the TT library and published standard setting [1, 20, 21, 22]. Possible interferences were evaluated using blood plasma samples from six different donators known to be free of the analytes being tested. For all six samples, no interfering signals or false positive results were observed.

In addition, possible analyte carry‐over was investigated by injection of pooled blank plasma samples after the highest calibrator concentration (calibrator 5) to exclude any transfer from one sample to another. No analyte carry‐over was observed, but it should be recommended to analyze blank samples after any intoxication to prevent possible carry‐over from sample concentrations above the tested concentrations.

3.1.3. Linearity of Calibration and Limits of Quantification

The concentration ranges of the drug levels were selected in the method to cover the upper therapeutic and toxic range according to the values reported by Baselt and Schulz et al. [23, 24].

The range of calibration for each analyte was set to be applicable for use in ET, meaning that concentrations below the therapeutic range or within were not of importance, as the purpose of the method should be to identify possible intoxications. Thus, the upper therapeutic range served as reference point for the lower limit of quantification (LLOQ) in most cases, with minor adjustments for low‐dosed analytes, such as biperiden, haloperidol, hydromorphone, ramipril, and risperidone, or such with low MS response, such as bromazepam and oxycodone. For those analytes, quantification was only possible for higher concentration. Regarding the upper limit of quantification (ULOQ), the concentration was set to 2000 ng/mL for most analytes, except for the high‐dosed compounds carbamazepine, ketamine, oxcarbazepine, and paracetamol.

All tested analytes could be fitted into a linear calibration model using different weighting factors (equal, 1/x, 1/x2, see Table 1). The back calculation of the calibrators for all analytes was within the ranges defined by the EMA guideline [13].

3.1.4. Matrix Effects

The presence of co‐eluting matrix components can reduce or enhance the ion intensity of analytes, affecting the reproducibility and accuracy of the method. Therefore, the presence or absence of matrix effects in biological matrix was investigated. Matrix effects and corresponding CVs for each analyte are shown in Table S3. All analytes were within the acceptable range of ± 30% matrix effect.

3.1.5. Accuracy and Precision

Quantification was automatically done by integrating the peak area in full scan of the protonated precursor mass of the analytes and corresponding IS. The peak area ratio of each analyte to the IS was compared with an electronically stored 5‐point calibration to calculated plasma concentrations. Analytes and calibration information were stored in a CSV file (Table S2). The file contains the analytes names, the used IS (trimipramine‐d3), the slope and intercept of the calibrations curve, the limits of quantification (LLOQ and ULOQ), the concentration unit (ng/mL), the calibration concentration (1‐point calibration), and in case the m/z value of the quantifier ion. Results are shared via an automatically generated PDF report.

Results for within‐run and between‐run accuracy and precision are shown in Tables S4 and S5, respectively. According to the recommendations of the GTFCh, blood level determinations in the context of 24/7 emergency toxicology allow a deviation of ± 30%. Regarding within‐run accuracy and precision, most analytes were within the GTFCh criteria of mean concentration within ± 30% of nominal concentration and precision did not exceed a CV of ± 30% except aripiprazole, diphenhydramine, ketamine, melperone, pethidine, sertraline, tilidine, and tilidine‐M (nor‐). In addition to the previously mentioned analytes, olanzapine and pethidine‐M (nor‐) also exceeded prescribed limits for between‐run accuracy and precision. Moreover, a reliable signal could not be detected for doxylamine at the LLOQ in every run. The method implemented in this study did not consider the integrity of dilution, as surpassing the ULOQ implies severe intoxications in all instances and precise measurements exceeding the ULOQ are neglectable in clinical toxicology.

3.1.6. Stability of the Electronically Stored Calibration Over a 2‐Year Period

As the present method was developed for 24/7 toxicology applications, an electronically stored calibration was used as already described by Caspar et al. [1]. To monitor the stability of the electronically stored calibration and to extend data on accuracy and precision, freshly extracted QC high samples were measured over a two‐year period. Figure 2 shows the deviations of the calculated concentrations from the nominal values.

FIGURE 2.

FIGURE 2

Deviations of the calculated from the nominal values of quality control samples (QC) high 1, high 2, and high 3, divided into three working solutions (1, 2, and 3) calculated based on stored 5‐point calibration over a two‐year period (in total 48 analyses). Acceptance criteria were set to ± 30%.

Among the 10 analytes that initially failed to meet the criteria mentioned above, only melperone still did not meet the criteria. The remaining nine analytes were detected and quantified with varying reliability over the period mentioned. Even among analytes that passed validation, some outliers were observed, highlighting that method validation reflects only a snapshot in time. These findings underscore the importance of regularly verifying calibration using freshly prepared or certified QC samples to ensure method reliability over extended periods.

If an analyte failed the acceptable limits of ± 30% of nominal concentration in two continuous measurements, a freshly prepared 1‐point calibration was performed to adjust the slope. The use of a fast 1‐point calibration reduces the workload and turn‐around time without loss of quality [3]. If the analyte still remains outside the acceptable limits after the slope adjustment, a new 5‐point calibration is prepared and applied to the respective analyte. Therefore, for the electronically sorted calibration, it was shown that a 5‐point calibration remains stable for an average of 6 months if supplemented by a freshly extracted 1‐point calibration every 4–8 weeks. This process, managed via separate processing method, automatically transfer 1‐point evaluation data to a CSV file (Table S2) and averages the slopes with existing entries. When a new 5‐point calibration is created, the older CSV file is archived, and a new one is initiated.

3.2. Proof of Concept and Applicability Studies

Proof of concept and applicability studies were done by analyzing 29 plasma samples and comparing the calculated plasma levels to those obtained by the reference GC–MS method described by Meyer et al. or the reference LC–MS/MS method by Michely and Maurer depending on the analyte [2, 3]. The results are given in Table S6. For most analytes, the quantitative results by the presented method were in accordance with those obtained by the reference methods. Although tilidine‐M (nor‐) did not fulfill all validation criteria, the calculated value in the patient plasma sample matched the reference value.

Applicability was also tested by analyzing several interlaboratory tests. The results in Table 2 were compared with the specified acceptance ranges given by Arvecon (https://www.arvecon.de/en/) and the Referenzinstitut für Bioanalytik (RfB, https://www.rfb.bio/cgi/switchLang?lang=en). Most of the interlaboratory tests refer to quantification in the field of therapeutic drug monitoring during daily routine analysis. However, the present method was primarily conceived to cover the upper therapeutic to toxic range. Thus, although most of the target values were below the LLOQ of the present method, the qualitative detection of the analytes in the therapeutic range could be demonstrated. Nevertheless, for olanzapine, which did not fulfill all validation criteria method, the values obtained were within the acceptance range. This is sufficient for the fast and reliable processing of samples during the 24/7 ET service, as it is often necessary to assess whether the drugs are in the therapeutic, overdose, or toxic range.

TABLE 2.

Results of analyses of proficiency tests (PT) with comments if the calculated concentration passed or failed the acceptance range for PT.

Case type Analyte LLOQ Calculated conc. Target value Accepted range Comment
PT for vitamins and analgesic Paracetamol 2500 21,531 18,400 12,800–24,000 Passed
66,247 82,100 57,400–107,000 Passed
52,800 44,200 30,900–57,500 Passed
PT for benzodiazepines Alprazolam 50 67 53 26.6–79.4 Passed
136 96.8 52.6–141 Passed
79 62.9 32.3–93.5 Passed
Bromazepam 1000 < 1000 174 100–248 Passed (< LLOQ)
< 1000 168 96–240 Passed (<LLOQ)
Clobazam 100 134 209 123–295 Passed
275 298 182–414 Passed
287 265 161–369 Passed
Diazepam 500 752 430 272–588 Failed
< 500 309 191–427 Passed (< LLOQ)
Diazepam‐M (nor‐) 750 < 750 223 133–313 Passed (< LLOQ)
< 750 337 209–465 Passed (< LLOQ)
< 750 518 334–702 Passed (< LLOQ)
Lorazepam 500 < 500 172 100–244 Passed (< LLOQ)
< 500 150 86–214 Passed (< LLOQ)
< 500 147 83–211 Passed (< LLOQ)
Midazolam 100 164 162 92–232 Passed
182 175 101–249 Passed
259 204 120–288 Passed
Oxazepam 1000 < 1000 360 224–496 Passed (< LLOQ)
1271 345 215–475 Failed
< 1000 336 208–464 Passed (< LLOQ)
Temazepam 500 < 500 108 58–158 Passed (< LLOQ)
< 500 59.5 30.3–88.7 Passed (< LLOQ)
< 500 50.1 24.9–75.3 Passed (< LLOQ)
Zolpidem 100 228 209 123–295 Passed
154 128 72–184 Passed
212 168 96–240 Passed
Zopiclone 100 nd 37.9 17.9–57.9 Failed (< LLOQ)
nd 58.9 29.9–87.9 Failed
PT for neuroleptics 1 Amisulpride 100 103 138 78–198 Passed
241 295 181–409 Passed
135 127 71–183 Passed
274 238 142–334 Passed
Chlorprothixene 250 < 250 93.9 50.9–136.9 Passed (< LLOQ)
< 250 257 155–359 Passed (< LLOQ)
< 250 90.9 49.1–132.7 Passed (< LLOQ)
270 246 148–344 Passed
Clozapine 100 359 445 283–607 Passed
154 188 110–266 Passed
430 435 277–593 Passed
152 215 127–303 Passed
Levomepromazine 200 nd 67.4 35.0–99.8 Failed (< LLOQ)
< 200 177 103–251 Passed (< LLOQ)
< 200 58.7 29.9–87.5 Passed (< LLOQ)
< 200 159 91–227 Passed (< LLOQ)
Olanzapine 50 < 50 39.4 18.8–60 Passed (< LLOQ)
89 82.6 44–121.2 Passed
< 50 49.4 24.4–74.4 Passed (< LLOQ)
87 77.9 41.1–114.7 Passed
Perzine 100 117 85.8 46–125.6 Passed
346 391 245–537 Passed
173 85.4 45.8–125 Failed
575 410 258–562 Failed
Promethazine 100 < 100 70.6 36.8–104.4 Passed (< LLOQ)
150 156 88–224 Passed
135 126 70–182 Passed
229 254 154–354 Passed
Quetiapine 100 < 100 44.2 21.4–67 Passed (< LLOQ)
210 210 124–296 Passed
< 100 59 30–88 Passed (< LLOQ)
221 190 110–270 Passed
Risperidone 50 < 50 7.92 3.16–12.68 Passed (< LLOQ)
< 50 10.3 4.1–16.5 Passed (< LLOQ)
< 50 9.13 3.65–14.61 Passed (< LLOQ)
< 50 14.3 5.7–22.9 Passed (< LLOQ)
Risperidone‐M (9‐hydroxy‐) /Paliperidone 100 < 100 31 14.2–47.8 Passed (< LLOQ)
144 97.6 53.2–142 Failed
< 100 36.8 17.4–56.2 Passed (< LLOQ)
110 106 58–154 Passed
PT for neuroleptics 2 Aripiprazole < 250 364 174 100–248 Failed
311 124 68–180 Failed
< 250 214 126–302 Passed (< LLOQ)
nd 341 211–471 Failed
Haloperidol 50 < 50 4.74 1.88–7.6 Passed (< LLOQ)
< 50 13.4 5.2–21.6 Passed (< LLOQ)
< 50 4.14 1.64–6.64 Passed (< LLOQ)
< 50 12.8 5.0–20.6 Passed (< LLOQ)
Melperone 1000 < 1000 61.3 31.3–91.3 Passed (< LLOQ)
< 1000 283 173–393 Passed (< LLOQ)
< 1000 116 64.0–168 Passed (< LLOQ)
< 1000 242 146–338 Passed (< LLOQ)
Pipamperone 250 < 250 134 74–194 Passed (< LLOQ)
< 250 321 199–443 Passed (< LLOQ)
< 250 171 99.0–243 Passed (< LLOQ)
432 331 205–457 Passed
Prothipendyl 50 nd 2.54 1.00–4.08 Failed (< LLOQ)
< 50 36.8 17.4–56.2 Passed (< LLOQ)
71 48.1 23.7–72.5 Passed
179 119 65.0–173 Failed
Sulpiride 500 < 500 279 111–447 Passed (< LLOQ)
1326 537 347–727 Failed
674 248 150–346 Failed
1513 502 322–682 Failed
Zuclopenthixol 50 < 50 17.3 7.1–27.5 Passed (< LLOQ)
52 38.3 18.1–58.5 Passed
< 50 8.49 3.39–13.59 Failed
nd 30.3 12.1–48.5 Passed (< LLOQ)
PT for antidepressants 1 Citalopram 100 < 100 62.3 31.9–92.7 Passed (< LLOQ)
135 115 63–167 Passed
Mianserin 50 < 50 43.2 21–65.4 Passed (< LLOQ)
113 91.3 49.3–133.3 Passed
Mirtazapine 150 < 150 35 16.4–53.6 Passed (< LLOQ)
< 150 73.9 38.7–109.1 Passed (< LLOQ)
Paroxetine 50 126 76.7 40.5–112.9 Failed
91 100 54–146 Passed
Sertraline 250 < 250 60.3 30.7–89.9 Passed (< LLOQ)
< 250 121 67–175 Passed (< LLOQ)
Venlafaxine 100 < 100 64.7 33.3–96.1 Passed (< LLOQ)
142 169 97–241 Passed
Venlafaxine‐M (O‐demethyl‐) 100 < 100 61.8 31.6–92 Passed (< LLOQ)
< 100 108 58–158 Passed (< LLOQ)

Note: Concentrations are given in ng/mL.

Abbreviation: nd, not detected.

3.3. Advantages and Limitations of the Method

The presented LC–MSn method offers advantages, particularly in the context of emergency toxicology. The method allows for the rapid assessment of potentially toxic substances by combining qualitative identification and quantitative determination within a single run. This is crucial to support rapid clinical decisions in acute poisoning cases [1]. Furthermore, the automated data processing and reporting system optimizes the workflow by minimizing the potential for human error and providing reliable results in a short time frame, even in high‐throughput laboratory settings. Another benefit of the method is its extensive range of analytes that are covered. The method was successfully validated for 59 commonly requested drugs and metabolites, representing a wide range of toxicologically relevant substances frequently encountered in emergency toxicology. This comprehensive scope ensures that laboratories can reliably detect and quantify substances associated with acute intoxications, thus meeting the practical demands of clinical toxicology. Of particular importance is the optimization of the method for toxic concentration ranges, which are of primary concern in emergency situations where identifying overdoses or toxic exposures is more relevant than measuring therapeutic levels [1, 2, 3].

Nevertheless, in addition to these advantages, the method also presents certain limitations that should be discussed. One limitation is its sensitivity, which is inferior to that of high‐sensitivity methods based on, for example, LC–MS/MS. This diminished sensitivity results in elevated lower limits of quantification, which, in certain instances, exceed the therapeutic range of specific analytes. For example, the therapeutic range for biperiden is reported to be 1.0–6.5 ng/mL, with toxic concentrations beginning at 13 ng/mL. However, the lowest calibration point (Cal 1) was set at 50 ng/mL, which exceeds the upper limit of the therapeutic range. Consequently, the method is unable to accurately quantify concentrations within the therapeutic range for such compounds. This limitation restricts the suitability of the LC–MSn method for TDM or applications requiring precise quantification at low concentration levels. Nevertheless, despite these limitations, the method remains suitable for its intended purpose, namely, the rapid and reliable detection of toxic concentrations in acute poisoning cases with some known exceptions. It fulfills the critical requirements of emergency toxicology, where timely identification of overdoses can directly impact patient outcomes and clinical decision‐making.

4. Conclusions

The present method offers a reliable and rapid screening of 69 analytes and additional quantification in blood plasma applicable in 24/7 emergency toxicology. It has been successfully integrated into the TT screening solution and has been validated in accordance with international guidelines and recommendations. A total of 59 out of the 69 analytes fulfilled all validation criteria and the remaining analytes can still be estimated in their quantity. The simple and fast sample preparation, short gradient, and automatic and fast quantification within 30 min based on an electronically stored 5‐point calibration contribute to the efficiency and sustainability of the method. However, it should be noted that it is only suitable for the identification of an acute overdosing.

Author Contributions

Selina Hemmer: writing–original draft, formal analysis, data curation, conceptualization. Maximilian Ninnig: writing–review and editing, formal analysis, data curation. Lea Wagmann: writing–review and editing, conceptualization. Sascha Manier: formal analysis, writing–review and editing. Markus R. Meyer: writing–review and editing, resources, project administration, conceptualization.

Conflicts of Interest

The authors declare no conflicts of interest.

Supporting information

Table S1. Ionization effects using 1 mg/L solutions and internal standard diazepam‐d5. The table summarized the retention time (RT) of each analyte in minutes (min), the relative deviations of ionization in pooled solutions to single analyte solution, and the distribution of the analytes into three working solutions (1, 2, or 3). Acceptance criteria (AC) for ionization effects are ±25%. Values out of AC are marked in red.

Table S2. Example table for the electronically stored five‐point calibration. Analytes are listed in alphabetical order. The table contains the name of the analytes according to the Toxtyper library, the internal standard (ISTD), the slope and intercept of the five‐point calibration curves, the limits of quantification (LLOQ and ULOQ), the concentration unit in ng/mL, the calibration concentration of the one‐point calibrator, the quantifier’s m/z value (Quan m/z), and the determined slope from an one‐point run (Cal_Slope_1).

Table S3. Matrix effects (ME) with corresponding coefficient of variation (CV) for quality control sample (QC) Low and High. Acceptance criteria for ME were within ±30%.

Table S4. Within‐run accuracy and precision data. Acceptance criteria (AC) for accuracy, ±30% of nominal value and AC precision, coefficient of variation (CV) < 30%. Values out of AC are marked in red.

Table S5. Between‐run accuracy and precision data. Acceptance criteria (AC) for accuracy, ±30% of nominal value and AC precision, coefficient of variation (CV) < 30%. Values out of AC are marked in red.

Table S6. Results of analyses of plasma samples analyzed by reference LC–MS/MS [1] or GC–MS [2] methods with interpretation of the values according to Schulz et al. [3]. Concentrations are given in ng/mL; not available, −.

DTA-17-1502-s001.pdf (348.3KB, pdf)

Acknowledgements

The authors like to thank Armin A. Weber and Juel Maalouli Schaar for their support. Open Access funding enabled and organized by Projekt DEAL.

Hemmer S., Ninnig M., Wagmann L., Manier S., and Meyer M., “Comprehensive Qualitative Drug Screening in Emergency Toxicology Using an Automated LC–MSn System:Simultaneous Quantification of Relevant Drugs and Metabolites in Blood Plasma,” Drug Testing and Analysis 17, no. 9 (2025): 1502–1515, 10.1002/dta.3855.

Data Availability Statement

The data that support the findings of this study are available from the corresponding author upon reasonable request.

References

  • 1. Caspar A. T., Meyer M. R., and Maurer H. H., “Blood Plasma Level Determination Using an Automated LC‐MS(N) Screening System and Electronically Stored Calibrations Exemplified for 22 Drugs and Two Active Metabolites Often Requested in Emergency Toxicology,” Drug Testing and Analysis 11, no. 1 (2019): 102–111, 10.1002/dta.2466. [DOI] [PubMed] [Google Scholar]
  • 2. Michely J. A. and Maurer H. H., “A Multi‐Analyte Approach to Help in Assessing the Severity of Acute Poisonings ‐ Development and Validation of a Fast LC‐MS/MS Quantification Approach for 45 Drugs and Their Relevant Metabolites With One‐Point Calibration,” Drug Testing and Analysis 10, no. 1 (2018): 164–176, 10.1002/dta.2257. [DOI] [PubMed] [Google Scholar]
  • 3. Meyer G. M., Weber A. A., and Maurer H. H., “Development and Validation of a Fast and Simple Multi‐Analyte Procedure for Quantification of 40 Drugs Relevant to Emergency Toxicology Using GC‐MS and One‐Point Calibration,” Drug Testing and Analysis 6, no. 5 (2014): 472–481, 10.1002/dta.1555. [DOI] [PubMed] [Google Scholar]
  • 4. Maurer H. H., “Demands on Scientific Studies in Clinical Toxicology,” Forensic Science International 165, no. 2–3 (2007): 194–198, 10.1016/j.forsciint.2006.05.019. [DOI] [PubMed] [Google Scholar]
  • 5. Maurer H. H., “How Can Analytical Diagnostics in Clinical Toxicology Be Successfully Performed Today?,” Therapeutic Drug Monitoring 34, no. 5 (2012): 561–564, 10.1097/FTD.0b013e31826a622d. [DOI] [PubMed] [Google Scholar]
  • 6. Meyer M. R., Peters F. T., and Maurer H. H., “Automated Mass Spectral Deconvolution and Identification System for GC‐MS Screening for Drugs, Poisons, and Metabolites in Urine,” Clinical Chemistry 56, no. 4 (2010): 575–584, 10.1373/clinchem.2009.135517. [DOI] [PubMed] [Google Scholar]
  • 7. Wissenbach D. K., Meyer M. R., Remane D., Weber A. A., and Maurer H. H., “Development of the First Metabolite‐Based LC‐MS(N) Urine Drug Screening Procedure‐Exemplified for Antidepressants,” Analytical and Bioanalytical Chemistry 400, no. 1 (2011): 79–88, 10.1007/s00216-010-4398-9. [DOI] [PubMed] [Google Scholar]
  • 8. Helfer A. G., Michely J. A., Weber A. A., Meyer M. R., and Maurer H. H., “Orbitrap Technology for Comprehensive Metabolite‐Based Liquid Chromatographic‐High Resolution‐Tandem Mass Spectrometric Urine Drug Screening ‐ Exemplified for Cardiovascular Drugs,” Analytica Chimica Acta 891 (2015): 221–233, 10.1016/j.aca.2015.08.018. [DOI] [PubMed] [Google Scholar]
  • 9. Mueller D. M., Duretz B., Espourteille F. A., and Rentsch K. M., “Development of a Fully Automated Toxicological LC‐MS(N) Screening System in Urine Using Online Extraction With Turbulent Flow Chromatography,” Analytical and Bioanalytical Chemistry 400, no. 1 (2011): 89–100, 10.1007/s00216-010-4560-4. [DOI] [PubMed] [Google Scholar]
  • 10. Stephenson J. B., Flater M. L., Austin J., Bain L. T., Holt L. A., and Mehan J. M., “Comprehensive Drug Screening of Whole Blood by LC‐HRMS‐MS in a Forensic Laboratory,” Journal of Analytical Toxicology 45, no. 3 (2021): 243–251, 10.1093/jat/bkaa073. [DOI] [PubMed] [Google Scholar]
  • 11. Gottardo R., Murari M., Bertaso A., Bortolotti F., and Tagliaro F., “Drug Screening by Using the Toxtyper LC‐Ion Trap MS: Optimization of Its Application on Serum Samples in a DUID Context,” Clinica Chimica Acta 510 (2020): 537–543, 10.1016/j.cca.2020.08.009. [DOI] [PubMed] [Google Scholar]
  • 12. Fang Z., Zhang H., Guo J., and Guo J., “Overview of Therapeutic Drug Monitoring and Clinical Practice,” Talanta 266, no. Pt 1 (2024): 124996, 10.1016/j.talanta.2023.124996. [DOI] [PubMed] [Google Scholar]
  • 13. EMA, Guideline on Bioanalytical Method Validation (London, UK: European Medicines Agency, 2011). [Google Scholar]
  • 14. GTFCh , “Recommendations of Criteria for Development and Validation of Analytical Methods for Estimating Concentrations of Drugs in Blood to Be Used in 24/7 Clinical Toxicology,” Toxichem Krimtech 85, no. 1 (2018): 35. [Google Scholar]
  • 15. Caspar A. T., Kollas A. B., Maurer H. H., and Meyer M. R., “Development of a Quantitative Approach in Blood Plasma for Low‐Dosed Hallucinogens and Opioids Using LC‐High Resolution Mass Spectrometry,” Talanta 176 (2018): 635–645, 10.1016/j.talanta.2017.08.063. [DOI] [PubMed] [Google Scholar]
  • 16. Maurer H. H., Pfleger K., and Weber A. A., Mass Spectral Data of Drugs, Poisons, Pesticides, Pollutants and Their Metabolites (Weinheim: Wiley‐VCH, 2016). [Google Scholar]
  • 17. Remane D., Meyer M. R., Wissenbach D. K., and Maurer H. H., “Ion Suppression and Enhancement Effects of Co‐Eluting Analytes in Multi‐Analyte Approaches: Systematic Investigation Using Ultra‐High‐Performance Liquid Chromatography/Mass Spectrometry With Atmospheric‐Pressure Chemical Ionization or Electrospray Ionization,” Rapid Communications in Mass Spectrometry 24, no. 21 (2010): 3103–3108, 10.1002/rcm.4736. [DOI] [PubMed] [Google Scholar]
  • 18. Team R. Core, R: A Language and Environment for Statistical Computing (Vienna, Austria: R Foundation for Statistical Computing, 2020), https://www.r‐project.org/. [Google Scholar]
  • 19. Matuszewski B. K., Constanzer M. L., and Chavez‐Eng C. M., “Strategies for the Assessment of Matrix Effect in Quantitative Bioanalytical Methods Based on HPLC‐MS/MS,” Analytical Chemistry 75, no. 13 (2003): 3019–3030, 10.1021/ac020361s. [DOI] [PubMed] [Google Scholar]
  • 20. Kempf J., Traber J., Auwärter V., and Huppertz L. M., “‘Psychotropics Caught in a Trap’ ‐ Adopting a Screening Approach to Specific Needs,” Forensic Science International 243 (2014): 84–89, 10.1016/j.forsciint.2014.04.035. [DOI] [PubMed] [Google Scholar]
  • 21. Huppertz L. M., Kneisel S., Auwärter V., and Kempf J., “A Comprehensive Library‐Based, Automated Screening Procedure for 46 Synthetic Cannabinoids in Serum Employing Liquid Chromatography‐Quadrupole Ion Trap Mass Spectrometry With High‐Temperature Electrospray Ionization,” Journal of Mass Spectrometry 49, no. 2 (2014): 117–127, 10.1002/jms.3328. [DOI] [PubMed] [Google Scholar]
  • 22. Klima M., Altenburger M. J., Kempf J., Auwärter V., and Neukamm M. A., “Determination of Medicinal and Illicit Drugs in Post Mortem Dental Hard Tissues and Comparison With Analytical Results for Body Fluids and Hair Samples,” Forensic Science International 265 (2016): 166–171, 10.1016/j.forsciint.2016.02.021. [DOI] [PubMed] [Google Scholar]
  • 23. Schulz M., Schmoldt A., Andresen‐Streichert H., and Iwersen‐Bergmann S., “Revisited: Therapeutic and Toxic Blood Concentrations of More Than 1100 Drugs and Other Xenobiotics,” Critical Care 24, no. 1 (2020): 195, 10.1186/s13054-020-02915-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Baselt R. C., Disposition of Toxic Drugs and Chemicals in Man, 11th ed. (Seal Beach, CA: Biomedical Publications, 2017). [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Table S1. Ionization effects using 1 mg/L solutions and internal standard diazepam‐d5. The table summarized the retention time (RT) of each analyte in minutes (min), the relative deviations of ionization in pooled solutions to single analyte solution, and the distribution of the analytes into three working solutions (1, 2, or 3). Acceptance criteria (AC) for ionization effects are ±25%. Values out of AC are marked in red.

Table S2. Example table for the electronically stored five‐point calibration. Analytes are listed in alphabetical order. The table contains the name of the analytes according to the Toxtyper library, the internal standard (ISTD), the slope and intercept of the five‐point calibration curves, the limits of quantification (LLOQ and ULOQ), the concentration unit in ng/mL, the calibration concentration of the one‐point calibrator, the quantifier’s m/z value (Quan m/z), and the determined slope from an one‐point run (Cal_Slope_1).

Table S3. Matrix effects (ME) with corresponding coefficient of variation (CV) for quality control sample (QC) Low and High. Acceptance criteria for ME were within ±30%.

Table S4. Within‐run accuracy and precision data. Acceptance criteria (AC) for accuracy, ±30% of nominal value and AC precision, coefficient of variation (CV) < 30%. Values out of AC are marked in red.

Table S5. Between‐run accuracy and precision data. Acceptance criteria (AC) for accuracy, ±30% of nominal value and AC precision, coefficient of variation (CV) < 30%. Values out of AC are marked in red.

Table S6. Results of analyses of plasma samples analyzed by reference LC–MS/MS [1] or GC–MS [2] methods with interpretation of the values according to Schulz et al. [3]. Concentrations are given in ng/mL; not available, −.