Abstract
Ultrasonic phased array technology enables flexible and controllable beamforming through precise phase delay control of individual array elements in the transducer, facilitating dynamic focusing, beam steering, and beamforming. This study presents a 64-channel system achieving 1 ns delay resolution using FPGA-based phase-locked loops. Through systematic testing and calibration of the delay error in the phased array transmission driving system, the actual delay error was successfully controlled within 1 ns. Furthermore, to address the focal shift issue in multi-layer soft tissues, this research implemented active time reversal and phase compensation methods for focal shift correction. Experimental results demonstrate that the proposed system not only exhibits excellent driving and phase modulation capabilities but also effectively reduces tissue-induced focal shift.
Keywords: Ultrasonic phased array, Delay resolution, Phase calibration, FPGA, Time reversal method
Introduction
High-Intensity Focused Ultrasound (HIFU) has emerged as a clinically validated, minimally invasive therapy that induces localized tissue ablation through two primary mechanisms: (1) thermal-mediated protein coagulation and (2) non-thermal mechanical disruption [1]. The technology enables precise spatial control, generating confined regions of coagulative necrosis (> 55 °C) at target depths while sparing surrounding tissues. Compared to alternative thermal therapies (e.g., radiofrequency/microwave ablation), HIFU offers superior temperature regulation and penetration depth in soft tissues, making it particularly valuable for treating solid tumors and other deep-seated pathologies.
Current HIFU systems utilize two primary transducer configurations: bowl-shaped and phased-array. The bowl-shaped focused transducer, characterized by its simple structure and fixed focal length, adjusts the focal position through mechanical movement. This technology has been widely implemented in clinical practice [2, 3]. In contrast, phased-array focused transducer utilizes software-controlled focusing mechanisms, offering several advantages including flexible focusing patterns, precise and rapid acoustic field control, beam steering capability, and multi-focus generation for enhanced ablation area and rate. These advanced characteristics endow phased-array technology with particular promise for treating solid tumor-related pathologies [4]. Jing et al. [5] employed the k-space method to correct transcranial ultrasound beam focusing, demonstrating that this approach enables precise targeting while completing treatment planning computations within minutes. Li et al. [6] proposed a method to enhance the steering range of HIFU phased arrays through array tilting, which effectively expands the focal steering capability and significantly increases the treatment volume. All these tissue or cranial treatments encounter a common challenge: the significant impact of multilayered media's nonlinear and heterogeneous characteristics on focusing efficacy. In addition, the delay resolution is also an important parameter that determines the focusing accuracy [7].
To address these issues, this study employs an active time-reversal method to correct focal shifts caused by the nonlinear and heterogeneous characteristics of multilayered tissues, proposing an innovative dual-delay approach that combines coarse and fine delay mechanisms to significantly improve delay resolution while enhancing echo sampling resolution through cross-correlation and curve-fitting interpolation methods. The experimental setup used two-layer porcine tissue (fat and muscle) as the propagation medium, with hydrophone measurements characterizing the steered focusing acoustic field of the ultrasound phased array.
Materials and methods
Principle of beam steering in phased-array ultrasound focusing
The excitation of each array element in the transducer software precisely controls, enabling the generation of time-delay-induced phase shifts (
) generate steerable wavefronts. This controlled phase variation creates coherent wavefronts that undergo constructive and destructive interference in the acoustic field, resulting in precise beam focusing and steering. The beamforming process obeys the Huygens-Fresnel principle by modeling each element as a secondary wavelet source, where each array element acts as a point source, and the superposition of these wavefronts determines the resultant beam direction and focal characteristics. The reception process represents the inverse of the transmission mechanism. Due to varying propagation paths, each array element receives echo signals from the target point at different times with distinct phase shifts. This spatial diversity necessitates precise time delay compensation and weighted summation during signal reception. The delay-and-sum beamforming algorithm enables flexible control over the synthetic aperture characteristics through manipulation of phase and amplitude parameters across array elements. This sophisticated control mechanism facilitates various operational modes, including conventional focusing, dynamic focusing, synthetic aperture imaging, and multi-beam formation [6]. The schematic diagram of phased-array focusing and beam steering as showed in Fig. 1. It is mainly composed of phase control system, power amplifier circuit, impedance matching circuit and phased-array transducer. In the HIFU phase control system, the host computer transmits the delay time of each channel to the FPGA via a serial interface. The FPGA then synchronizes the excitation signals of all channels using its internal clock to achieve phase control.
Fig. 1.
Schematic diagram of beam steering and focusing principles in phased-array systems
Transducer geometry
The transducer employed in this study is a 64-element Archimedean spiral phased array fabricated from 1 to 3 piezoelectric composite material. This composite structure embeds piezoelectric ceramic pillars within a polymer matrix, combining the high piezoelectric performance of PZT with the polymer's advantageous characteristics including low acoustic impedance, reduced dielectric constant, and enhanced flexibility. The array demonstrates superior beam steering capabilities with effective sidelobe suppression, which is critical for therapeutic applications requiring precise focal control [8]. Figure 2 shows the physical implementation of this transducer, with the following key geometric parameters: element diameter (2r) = 13 mm, aperture diameter (D) = 150 mm, focal length (F) = 120 mm, and operating frequency (f) = 0.905 MHz.
Fig. 2.
Physical diagram of 64 channel phased array transducer
The concave spherical transducer represents a section of a spherical surface, where the physical focal point coincides with the geometric center of the sphere when no time delays are applied. Considering the transducer's center as the origin
and the sphere's radius as r, the physical focal point coordinates are
in the three-dimensional Cartesian coordinate system. For a given sound velocity c, let the spatial coordinates of the nth array element be
and the desired focal point coordinates be
. The acoustic path length (
) between the nth array element and the designated focal point can be calculated using the three-dimensional Euclidean distance formula [9]:
![]() |
1 |
After calculating the acoustic path difference between each element relative to the set focus, the acoustic path difference of each element is subtracted from the maximum value:
![]() |
2 |
Once you have the relative distance, divide it by the speed of sound:
![]() |
3 |
The time-reversal focusing method is related to the invariance of wave propagation equations under time reversal in lossless, fluid media. In soft tissues, the low attenuation coefficient within clinically used frequency ranges ensures the validity of the time-reversal invariance assumption. If
represents the pressure field recorded on the transducer array (where
denotes the spatial position of array element
) from a pulsed point source at location
,the time-reversal mirror inverts the signal to obtain
, where
is the total delay time required for acoustic wave focusing. The time-reversed waves converge toward their source, effectively implementing a spatiotemporal matched filter that adapts to the propagation transfer function in heterogeneous media [10]. In fact, this method has been proven to optimally retrieve the source position of acoustic signals propagating along the optimal path in
. According to the acoustic reciprocity theorem, even in heterogeneous media, the positions of transmitting and receiving transducers can be interchanged without altering the acoustic signal information. That is, the diffraction impulse response
measured at point
from a source at r0 is equivalent to the response
measured at r0 from a source at
:
![]() |
4 |
Assuming the piezoelectric and inverse piezoelectric effects of the transducer array are identical, then:
![]() |
5 |
where
represents the piezoelectric process and
the inverse piezoelectric process. Consequently, after a complete time-reversal process (recording and transmitting signals), the total acoustic field received at
equals:
![]() |
6 |
where
is the number of transducer elements. According to Eq. (6), each element's contribution can be symmetrically expressed as:
![]() |
7 |
These consecutive convolutions implement matched filtering of diffraction impulse responses and transducer responses. Based on matched filter theory, the acoustic waves from each element reach their maximum at point
at time 
FPGA-based transmission delay design methodology
Delay resolution is the most critical parameter in phased array technology, determining the system's imaging resolution and error sidelobes. Dusa et al. [11] implemented a 3.125 ns receive delay resolution using dedicated integrated chips at 40 MHz sampling frequency, while proposing a modular architecture for complete beamforming that enables 1.25 ns minimum delay resolution per channel. Tang et al. [12] proposed a novel polyphase CIC interpolation filter design, achieving 1 ns delay resolution at 100 MHz sampling frequency using an improved 10 × polyphase interpolation CIC filter. Yuhang et al. [13] utilized two PLL resources in FPGA to double and phase-shift a 50 MHz clock signal, generating eight 250 MHz square wave channels with 45° phase differences, achieving a simulated delay resolution of 0.5 ns. Excessively high delay resolution imposes significant pressure on hardware circuits, causing substantial deviations between actual errors and theoretical values. For practical therapeutic applications, 1 ns delay resolution is entirely sufficient. Digital delay techniques such as sampling delay and dedicated chip delay suffer from high costs and difficulties in modification and integration. Software delay has gradually become the primary delay control method due to its strong flexibility, versatility, and portability. Considering the main frequency of the FPGA chip used in this phased array system and implementation feasibility, the simpler PLL-based method was selected for high delay resolution design. This approach achieves an effective balance between delay precision and implementation complexity while maintaining reasonable power consumption.
The proposed delay module architecture adopts a coarse–fine hybrid delay mechanism, as illustrated in Fig. 3. The system utilizes FPGA-integrated Multi-Clock Management Modules (MCMM) to generate precise clock signals through accurate frequency multiplication and phase shifting, achieving high delay precision at relatively low operating frequencies.
Fig. 3.
Conceptual illustration of the delay module design
The coarse delay section utilizes a 250 MHz clock signal obtained by multiplying a 50 MHz crystal oscillator input by a factor of 5 using a phase-locked loop (PLL). Based on the 250 MHz clock frequency counting, the delay value is an integer multiple of the clock cycle, yielding a resolution of 4 ns. Due to the limited operating frequency of the FPGA, a combined approach of fine delay and coarse delay is adopted to further refine the excitation waveform output from the coarse delay stage.
By employing the PLL to multiply the 50 MHz crystal input by 5 and generate three phase-shifted clock signals with 90° phase differences, the 250 MHz clock corresponds to a 4 ns cycle, resulting in a 1 ns time difference between two clock signals with 90° phase shift. This enables a theoretical fine delay resolution of 1 ns in addition to the coarse delay. The schematic diagram of the PLL frequency multiplication for fine delay implementation is shown in Fig. 4.
Fig. 4.
PLL phase shift timing sequence
Phase correction methodology for tissue-induced errors
The focal point of concave spherical focusing transducers exhibits significant displacement from the geometric center when propagating through multi-layer soft tissues, primarily due to heterogeneous media and nonlinear effects. This focal shift becomes particularly pronounced when encountering high-density media such as skull or rib structures during acoustic propagation. Furthermore, research has demonstrated that nonlinear effects in concave spherical focusing systems induce additional focal displacement, with the focal point migrating away from the transducer as signal amplitude increases [14]. The acoustic parameters of each tissue layer are listed in Table 1, with data sourced from the IT'IS Foundation in Zurich, Switzerland.
Table 1.
Media acoustic parameters table
| Medium | Speed of sound (m/s) | Density (kg/m3) | Attenuation coefficient (Np/m) | Nonlinear parameter (B/A) |
|---|---|---|---|---|
| Water | 1500 | 1000 | 0.025 | 5.0 |
| Skin | 1624 | 1109 | 21.254 | 7.0 |
| Fat | 1478 | 911 | 4.360 | 10.0 |
| Muscle | 1588 | 1090 | 7.100 | 7.1 |
Time-reversal methodology has emerged as an effective solution for compensating focal shifts in heterogeneous media [15]. Extensive studies by domestic and international researchers have validated the efficacy of time-reversal techniques in compensating focal displacements under both linear and nonlinear conditions [15, 16]. Building upon these findings, this study implements an active time-reversal method to compensate for focal shifts occurring through multi-layer tissues and aqueous media. The specific method is shown in Fig. 5.
Fig. 5.
The time reversal method excessively centralizes the correction focus
In the forward propagation phase, the wavefront emitted from a point source propagates through the medium and is subsequently detected by one or more receivers. The detected signals at each receiver undergo temporal inversion and are retransmitted from their respective positions, causing the retransmitted signals to converge at the original source location. This receiver array constitutes the time-reversal mirror (TRM), which in ultrasound systems typically comprises a 64-element phased array transducer capable of sampling, time-reversing, and retransmitting the incident acoustic field [17].
The time-reversal (TR) process can be conceptualized as a wave energy concentration method that exploits wave back propagation to focus energy at specific spatiotemporal locations. This technique fundamentally relies on two physical principles: spatial reciprocity and time-reversal invariance of the linear wave equation. In heterogeneous media, TR focusing demonstrates unique capabilities.
The accuracy of TR-based tissue-induced focal shift correction critically depends on the precision and efficiency of delay estimation algorithms. This study implements a normalized cross-correlation (NCC) method for echo signal delay estimation. The NCC algorithm extracts a data segment (window) from the reference echo signal and searches for the most matching pattern in the shifted echo signal by computing the correlation coefficient (pattern matching function) at each search step [18]. The NCC between the reference
and the offset
window is defined as:
![]() |
8 |
In the normalized cross-correlation (NCC) algorithm, r represents the origin of the reference window, N denotes the window length, and τ indicates the search lag (i.e., the offset between the reference window and the shifted window). The numerator in the NCC equation represents the unnormalized cross-correlation between
and
, while the denominator normalizes this correlation by accounting for the energy of both windows. This normalization process provides significant advantages by compensating for local variations in standard deviation and window mean values [19].
The time shift between the reference window and the shifted window is determined by the τ value that maximizes the NCC coefficient (Eq. 8). This maximum correlation point corresponds to the optimal temporal alignment between the two signal segments. Figure 6 illustrates the general workflow of motion estimation based on correlation coefficients.
Fig. 6.
A schematic illustration of a correlation coefficient based motion estimation method. A reference window is compared with a shifted window to find the location where correlation coefficient between them is maximum. The location of maximum correlation coefficient corresponds to the time-shift between the two windows
In practical implementation, the estimation of time delays using discrete-time ultrasonic echo signals is inherently limited to integer multiples of the sampling interval. This discrete-time approach may introduce significant estimation bias. However, when the sampling rate exceeds the Nyquist rate, interpolation techniques can achieve time delay estimation accuracy comparable to continuous-time signal processing.
The echo acquisition system in this study operates at a sampling frequency of 50 MHz, yielding an initial delay estimation resolution of 4 ns. This resolution mismatch with the transmission system's delay resolution (1 ns) could potentially introduce measurement errors. To address this limitation, we implement a three-point peak interpolation method with 40 interpolation points following cross-correlation analysis, effectively enhancing the delay estimation resolution to 1 ns, thereby matching the transmission system's resolution. Figure 7 illustrates the curve-fitting interpolation process of the cross-correlation function.
Fig. 7.
Diagram of the curve fitting to the three sampling points (y0, y1, and y2) around the cross-correlation peak. The dashed line represents the continuous time cross correlation, and the solid line represents the fitting curve
The pre-interpolation delay estimation is constrained to integer multiples of the sampling interval, while post-interpolation resolution improves proportionally with the number of interpolation points between samples. The cosine curve fitting method applies a cosine function to three adjacent sampling points (
,
, and
) near the peak of the cross-correlation function, defined by:
![]() |
9 |
α and β are defined as:
![]() |
10 |
![]() |
11 |
Without curve fitting or interpolation, time delay estimation would be constrained to the sampling point with maximum cross-correlation. However, when considering continuous-time signal processing, the peak of the cross-correlation function would be offset by
from
. By fitting a curve to several sampling points surrounding the cross-correlation peak (specifically
,
, and
), the precise location of the maximum
can be analytically determined.
Experimental setup and configuration
The experimental platform was designed to systematically evaluate the performance of the phased-array system through comprehensive sound field measurements. As shown in Fig. 8, the setup integrates four major components: an ultrasonic phased-array system, a sound field scanning system, a point sound source excitation system, and a precision motion control system. The ultrasonic phased-array system features a 64-channel programmable array driver controlled through a custom LabVIEW interface, combined with an AFE5818 analog front-end and TSW1400 digital processor for echo reception and signal processing. For sound field characterization, an ONDA HGL-0085 hydrophone with 0.2 mm active diameter was mounted on a motorized 3-axis positioning stage offering 0.01 mm resolution. The hydrophone signals were acquired through a 14-bit PCIe digitizer sampling at 100 MS/s, with spatial scanning controlled by a custom MATLAB algorithm. The point source excitation system consisted of an arbitrary waveform generator (20 MHz bandwidth) driving a power amplifier (50 dB gain) to stimulate a 10 MHz focused transducer producing a 0.8 × 2 mm focal zone.
Fig. 8.
Experimental platform
The measurement protocol established a Cartesian coordinate system centered on the array aperture, with the hydrophone scanned in 0.1 mm increments across a 50 × 50 × 50 mm volume. Each measurement point recorded full waveform data for subsequent analysis including pressure amplitude, spatial intensity distribution, focal zone dimensions, and sidelobe characteristics. This integrated experimental configuration enabled precise quantification of array performance parameters while maintaining rigorous measurement standards essential for reliable phased-array characterization. The combination of high-resolution motion control, sensitive acoustic detection, and comprehensive data processing provided a robust platform for evaluating delay resolution effects on beamforming quality.
Results and discussion
Comparison of FPGA simulation and actual results
In order to verify the correctness of the system design, Vivado simulator is used for simulation test. The simulation results are shown in Fig. 9, which reflects the output timing sequence of 8 pulse excitation signals. The coarse delay time of each channel in the test file is set to 60 ns, and the remainder of the first four channels is set to 0,1,2,3, respectively. The values set correspond to the first four digits of the O_finally_ultrsound_port signal. At the same time, the data bit width of each channel phase control signal is 15 bits, and the dynamic range after multiplying by 4 ns period is 0–13.1 us. The desired pulse excitation signal is generated at the synchronization start signal and the reset signal for the high level and low level respectively.
Fig. 9.
Delay simulation of frequency doubling and phase shifting
It can be seen from the figure that the difference of channels 1, 2, 3, and 4 is 1 ns, which proves that the theoretical delay resolution reaches 1 ns, thus verifying the correctness of the FPGA delay module design and conforming to the proposed parameter indexes.
Since the underlying control program for each FPGA channel is identical, only four adjacent channels were tested. The delay values for channels 1, 2, 3, and 4 were set to 0, 1, 2, and 3, respectively. The test results of ultrasound frequency and modulated waveforms are shown in Fig. 10a. The delay resolution test results between channels 1 and 2 are presented in Fig. 10b, between channels 1 and 3 in Fig. 10c, and between channels 1 and 4 in Fig. 10d
Fig. 10.
a Ultrasonic frequency and modulation waveform test results; b 1 and 2 channel delay resolution test results; c 1 and 3 channel delay resolution test results; d 1 and 4 channel delay Resolution test result
Figure 10a shows the test results with an output frequency of 1 MHz, modulation frequency of 2000 Hz, and output amplitude of 3.22 V. The on-board measured delay error values for Fig. 10b, c, and d are summarized in Table 2.
Table 2.
FPGA board level test results
| Channel (refer to Channel 1) | Theoretical delay value (ns) | Actual delay value (ns) | Error value (ns) |
|---|---|---|---|
| 2-channel | 1 | 0.444 | − 0.556 |
| 3-channel | 2 | 2.385 | 0.385 |
| 4-channel | 3 | 3.878 | 0.878 |
It can be seen from Table 2 that compared with the actual delay value and the theoretical delay value, the error values of the first three channels are all within 1 ns, and the overall delay resolution is within 2 ns, which conforms to the parameter indexes set by the subject. This design only uses 3 dual-port RAM to achieve the actual 2 ns delay resolution, and has the advantages of small error, less resource consumption, strong scalability, and the use of residual judgment form to solve the problem of cross-clock domain skillfully.
Actual results of physical field focusing deflection
The transmission drive system drives the 64-channel phased array transducer to emit ultrasonic waves for sound field scanning. Through the upper computer, the 64-channel delay is set to 0, the ultrasonic frequency is 905 kHz, the power is 1W, the duty cycle is 50%, and the modulation frequency is 1000 Hz. After setting the relevant parameters, the sound field x, Y and z axis distribution is measured by scanning the hydrophone. The sound field distribution after normalization is shown in Fig. 11, in which Fig. (A), (B) and (C) respectively represent the normalized sound pressure distribution above and below the X axis, Y acoustic axis and left and right direction of the Z axis.
Fig. 11.
a 5 mm normalized sound pressure distribution of X-axis geometric focusing and deflection; b Y-axis geometric focusing and deflection 5 mm normalized sound pressure distribution; c Z-axis geometric focusing and deflection 5 mm normalized sound pressure distribution
Figure 11a–c show the acoustic pressure distribution curves of our time-reversal method focusing at preset coordinate points (0, 120, 0) and (5, 125, 5) in water conditions. The black curves represent (0, 120, 0), which is the transducer's natural focal center point, while the red curves represent (5, 125, 5), indicating a 5 mm deflection in all three directions. The results demonstrate that maximum acoustic pressure achieves good focusing at (0, 120, 0). For the 5 mm deflection experiments in X, Y, and Z directions, the maximum acoustic pressure occurs at (5.2, 125.1, 5.3), with focusing position errors within 0.3 mm, fully proving the accuracy and feasibility of our research. Through three repeated experiments, we confirmed that the errors in X, Y, and Z directions are all within 0.3 mm, further demonstrating the stability of our research system.
Time reversal method corrects focus error
The acoustic beam steering test results in the previous section indicate that focal position may deviate in aquatic environments due to equipment errors and other factors. Meanwhile, the nonlinearity of layered tissues can also cause focal shifts. The time-reversal method was employed to calibrate and compensate for both types of errors. After performing delay estimation on the echo data, 40 interpolation points were generated from 3 sample points near the peak of the cross-correlation curve for curve fitting. The sampling frequency matched the theoretical delay resolution of the excitation system. Using the estimated delay values, the transducers were re-excited by the driving system to emit ultrasound through tissues for acoustic field scanning. The normalized acoustic field distribution after multilayer tissue time-reversal calibration for focal shift correction is shown in Fig. 12. The experimental data from 10 repeated trials of beam steering in the positive X-, Y-, and Z-axis directions are presented in Tables 3, 4, and 5, respectively.
Fig. 12.
a displays the experimental data of the steered acoustic field in the positive X-axis direction (vertical) within the tissue, b presents the steered acoustic field data in the positive Y-axis direction (acoustic axis, away from the transducer), and c shows the steered acoustic field data in the positive Z-axis direction (lateral) within the tissue
Table 3.
The experimental data from 9 beam-steering tests along the X-axis direction in tissues (data are presented as median)
| Preset focal offset in positive X-axis direction (mm) | 0 | 2 | 4 | 6 | 8 |
|---|---|---|---|---|---|
| Actual focal position (mm) | 0.2[0.1, 0.3] | 1.7[1.5, 1.9] | 3.5[3.3, 3.7] | 5.1[4.9, 5.3] | 6.3[6.1, 6.6] |
| Focal sound pressure (p/p0) | 1 | 0.88[0.88, 0.92] | 0.76[0.73, 0.79] | 0.59[0.53, 0.61] | 0.40[0.34, 0.42] |
Table 4.
The experimental data from 9 beam-steering tests along the Y-axis direction in tissues (data are presented as median)
| Preset focal offset in positive Y-axis direction (mm) | 120 | 125 | 130 | 135 |
|---|---|---|---|---|
| Actual focal position (mm) | 119.5 [119.3, 119.8] | 124.1[123.6, 124.9] | 128.6[128.0, 129.1] | 133.2[132.5, 133.9] |
| Focal sound pressure (p/p0) | 1 | 0.91[0.85, 0.93] | 0.79[0.75, 0.81] | 0.65[0.61, 0.73] |
Table 5.
The experimental data from 9 beam-steering tests along the Z-axis direction in tissues (data are presented as median)
| Preset focal offset in positive Z-axis direction (mm) | 0 | 2 | 4 | 6 | 8 |
|---|---|---|---|---|---|
| Actual focal position (mm) | 0.3 [0.2, 0.5] | 1.6[1.4, 1.8] | 3.5[3.4, 3.8] | 5.2[4.7, 5.5] | 6.1[5.7, 6.4] |
| Focal sound pressure (p/p0) | 1 | 0.87[0.84, 0.90] | 0.74[0.71, 0.77] | 0.58[0.53, 0.62] | 0.39[0.35, 0.42] |
The results show that the time-reversal method can achieve accurate focusing at preset positions through multiple biological tissue layers. In the Y-axis (acoustic axis) direction, when offsetting from the geometric focus (non-steered focus) at 120 mm by 5 mm, 10 mm, and 15 mm away from the transducer, nine repeated experiments yielded median actual focus positions of 119.5 mm, 124.1 mm, 128.6 mm, and 133.2 mm, with focusing errors of 0.5 mm, 0.9 mm, 1.4 mm, and 1.8 mm respectively. For the X (vertical) and Z (lateral) axes, with the Y-axis fixed at 120 mm, when steering from 0 to 8 mm in 2 mm increments, nine experiments showed median X-axis errors of 0.2 mm, 1.7 mm, 3.5 mm, 5.1 mm, and 6.3 mm, and Z-axis errors of 0.5 mm, 1.6 mm, 3.5 mm, 5.2 mm, and 6.1 mm.
The Y-axis demonstrates excellent focusing performance, maintaining errors within 2 mm even at 15 mm offset (1.8 mm error), meeting HIFU treatment requirements in human tissue. The X and Z axes also maintain good focusing accuracy, enabling dynamic 3D steering. Their comparable accuracy is likely due to the transducer's near-symmetrical spatial distribution about the Y-axis plane. In addition, focal pressure decreases with increasing offset distance—dropping to 40% of the geometric focus pressure at 8 mm offset in X/Z axes, while the Y-axis shows better pressure maintenance at 65% at 15 mm offset. This suggests focal pressure distribution after steering may become an interesting research direction.
In summary, our study demonstrates that time-reversal focusing through multiple tissue layers can achieve accurate focusing at preset positions. Despite nonlinear tissue effects, thermal effects, and inherent system hardware interference factors, it maintains focusing accuracy within small error ranges, meeting HIFU treatment requirements in human tissues.
Conclusion
Conventional HIFU focusing transducers employ relatively simple focusing mechanisms that cannot alter focal positions without physically moving the transducer. Currently, phased-array ultrasound focusing has emerged as a major research focus, particularly for biomedical applications. Phased-array systems can drive multi-element focusing transducers, electronically control focal positions through software-configurable excitation delays to achieve dynamic focusing automatically, and regulate energy distribution in focal regions to better ablate pathological tissues, thereby improving treatment efficacy and efficiency. When conventional delay-based focusing methods are applied in phased-array systems through multi-layered structures, significant focal shifts occur relative to preset positions due to strong nonlinear effects from high-intensity focusing, thermal effects, and errors from material processing and circuitry. These deviations become more pronounced with increasing steering distances.
In this study, our optimized phased-array system employs an active time-reversal method to precisely determine preset focal delays. The ultrasound energy maintains excellent focusing through multi-layered biological tissues, with maximum focal errors below 2 mm—meeting the ≤ 2 mm precision requirement for non-invasive brain tumor ablation. Three-dimensional focal accuracy satisfies HIFU therapeutic standards.Future phased-array ultrasound focusing may replace traditional single-element spherical focusing systems in biomedicine, particularly for HIFU therapy. Extensive research on precise cranial modulation is already underway, and accurate focal control in abdominal applications will be crucial for clinical translation. Additionally, temperature rise in target regions requires special consideration. Under identical irradiation conditions, deeper tissues exhibit smaller temperature increases, slower heating rates, and more dispersed energy distribution, resulting in progressively reduced effective heating areas (> ~ 40 °C). Variations in acoustic (sound speed, attenuation coefficient) and thermal (thermal conductivity, specific heat capacity) parameters among tissue types also lead to different focal temperatures under identical conditions.
Future studies should address the therapeutic impact of focal temperature variations and develop phased-array strategies for optimal energy control and distribution in focal regions—a key direction for our subsequent research. For clinical implementation of phased-array systems, two major challenges must be resolved: focal positioning accuracy and energy distribution patterns. This study provides essential theoretical and experimental support for advancing phased-array ultrasound systems toward clinical applications.
Author contributions
HaoSheng Xu and Shaohui Yang contributed equally in this work. Qi Lai Did some data analysis. Xue-Mei Gao did part of the writing of the article. Weijuan Chen made the main plans and revisions for the overall structure and content of the article. Xiaojin He provided guidance for the overall experiment and offered innovative ideas for this article.
Funding
No funds, grants, or other support was received.
Data availability
Data will be made available on request.
Declarations
Ethics approval and consent to participate
Not applicable.
Consent for publication
Consent for publication was obtained from all individual participants included in the study.
Competing interests
The authors declare no that they have no competing interests.
Footnotes
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Mason TJ. Therapeutic ultrasound an overview. Ultrason Sonochem. 2011;18:847–52. [DOI] [PubMed] [Google Scholar]
- 2.Mcdannold N, Clement GT, Black P, et al. Transcranial MRI-guided focused ultrasound surgery of brain tumors: initial findings in three patients. Neurosurgery. 2010;66(2):323. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Neer PV, Danilouchkine MG, Verweij MD, et al. Comparison of fundamental, second harmonic, and superharmonic imaging: a simulation study. J Acoust Soc Am. 2011;130(5):3148–57. [DOI] [PubMed] [Google Scholar]
- 4.Ji X, Shen G, Bai J, Li D, Yu Y, Qiao S, Chen Y. Multi-element ultrasound phased array applicator for the ablation of deep-seated tissue. J Shanghai Jiaotong Univer (Science). 2011;16:55–60. [Google Scholar]
- 5.Jing Y, Meral FC, Clement C. Time-reversal transcranial ultrasound beam focusing using a k-space method. Phys Med Biol. 2012;57(4):90. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Li K, Chen Y, Xu Y, Bai J, Chen Y, Ji X. Tilting high-intensity focused ultrasound phased array to augment the focal steering range for treatment of uterine fibroids. Appl Acoust. 2025. 10.1016/j.apacoust.2020.107342. [Google Scholar]
- 7.Li D, Shen G, Bai J, et al. Focus shift and phase correction in soft tissues during focused ultrasound surgery. IEEE Trans Biomed Eng. 2011;58(6):1621–8. 10.1109/TBME.2011.2106210. [DOI] [PubMed] [Google Scholar]
- 8.Morrison KP, Keilman GW, Kaczkowski PJ. Single archimedean spiral close packed phased array HIFU. IEEE Int Ultrasonics Symp. 2014;99(9):400–4. [Google Scholar]
- 9.Hand JW, Shaw A, Sadhoo N, et al. A random phased array device for delivery of high intensity focused ultrasound. Phys Med Biol. 2009;54(19):5675–93. [DOI] [PubMed] [Google Scholar]
- 10.Treeby BE, Zhang EZ, Cox BT. Photoacoustic tomography in absorbing acoustic media using time reversal. Inverse Prob. 2010;26157(26):115003–20. [Google Scholar]
- 11.Dusa C, Kalalii S, Pachamuthu R, et al. Integrated 16-channel transmit and receive beamforming ASIC for ultrasound imaging. Bangalore, India. In: 2015 28th International Conference on VLSI Design, 2015; 215–220.
- 12.Tang W, Liu G, Lin R. Design of polyphase cascaded integrator comb interpolator filter for ultrasonic phased array time delays. Int J Eng Technol. 2017;9:304–9. [Google Scholar]
- 13.Fan Y, Zhang X, Xu Xi, et al. Design of underwater high-precision ultrasonic phased array transmitting system, Dalian, China. In: 2017 International conference on computer systems, electronics and control (ICCSEC), 2017;1549–1552.
- 14.Makov YN, Sánchez-Morcillo VJ, Camarena F, et al. Nonlinear change of on-axis pressure and intensity maxima positions and its relation with the linear focal shift effect. Ultrasonics. 2008;48(8):678–86. [DOI] [PubMed] [Google Scholar]
- 15.Leduc N, Okita K, Sugiyama K, Takagi S, Matsumoto Y. Focus control in HIFU therapy assisted by time-reversal simulation with an iterative procedure for hot spot elimination. J Biomech Sci Eng. 2012;7:43–56. [Google Scholar]
- 16.Liu L, Lin WJ, Zhang HL. Simulation of focusing high intensity ultrasound by time reversal method. In: Symposium on piezoelectricity, acoustic waves and device applications.
- 17.Ding X, Wang Y, Zhang Q, Zhou W, Wang P, Luo M, Jian X. Modulation of transcranial focusing thermal deposition in nonlinear HIFU brain surgery by numerical simulation. Phys Med Biol. 2015;60(10):3975–98. [DOI] [PubMed] [Google Scholar]
- 18.Zahiri-Azar R, Salcudean S. Time-delay estimation in ultrasound echo signals using individual sample tracking. IEEE Trans Ultrason Ferroelectr Freq Control. 2008;55(12):2640–50. [DOI] [PubMed] [Google Scholar]
- 19.Zahiri-azar R, Salcudean SE. Motion estimation in ultrasound images using time domain cross correlation with prior estimates. IEEE Trans Biomed Eng. 2006;53(10):1990–2000. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Data will be made available on request.