. 2025 Sep 5;104(36):e44092. doi: 10.1097/MD.0000000000044092

Current status of an implant-based augmentation mammaplasty in Korean women: Implications for risk management and social sustainability of breast implant industry in the context of the first Korean case of a medical device fraud

Robert Kim, Breast Implant Research Groupa,*
PMCID: PMC12419266  PMID: 40924612

Abstract

Stakeholders in the breast implant industry in Korea have recently experienced a crisis from breast implant-associated anaplastic large cell lymphoma and the first Korean case of a medical device fraud. We compared the short-term safety between the microtextured devices that are commercially available after the occurrence of breast implant crisis in Korea. The current study was conducted in a cohort of Korean women who had received an implant-based augmentation mammaplasty for aesthetic purposes between November 14, 2020 and October 13, 2022. We considered risk factors of complications in analyzing the safety of devices for the current study. A total of 801 Korean women (n = 801) were finally assessed. Incidences of capsular contracture were 1.79% (3/168), 3.64% (21/577), 8.11% (3/37), and 10.53% (2/19) in the patients receiving Motiva Ergonomix, Sebbin Sublimity, Sebbin Integrity, and Eurosilicone Round Collection, respectively. These differences reached statistical significance (P < .05). There were 2 women with rupture after receiving Sebbin Sublimity, although there were no cases of rupture in association with other brands of breast implants. Overall capsular contracture-free survival was estimated at 681.470 ± 8.314 (95% confidence interval [CI] 665.174–697.766) days. By breast implants, it was 708.899 ± 8.595 (95% CI 692.053–725.745), 599.327 ± 6.607 (95% CI 586.378–612.277), 584.941 ± 22.965 (95% CI 539.931–629.952), and 572.492 ± 37.374 (95% CI 499.240–645.745) days in the patients receiving Motiva Ergonomix, Sebbin Sublimity, Sebbin Integrity, and Eurosilicone Round Collection, respectively, in the increasing order. In conclusion, our results indicate that Motiva Ergonomix Round SilkSurface is currently a relatively safer device as compared with others in the context of the first Korean case of a medical device fraud. The breast implant industry in Korea should be aware of the importance of the social sustainability in manufacturing a device.

Keywords: anaplastic, breast implants, fraud, large-cell, lymphoma, medical device recalls, safety-based medical device withdrawals

1. Introduction

A medical device is defined as instrument, apparatus, implement, machine, appliance, implant, reagent for in vitro use, software, material or other similar or related article that are manufactured for diagnostic or therapeutic purposes.[1] The medical device industry plays a pivotal role in improving both the health and quality of life of patients.[2] It is therefore mandatory to ensure the safety and efficacy of the medical device used by both healthcare professionals and patients. Over past decades, the medical device has undergone a stringent quality assessment from a reactive practice to a complete quality assurance approach throughout its lifecycle.[3]

A breast implant is a type IV implantable medical device that is used in augmentation mammaplasty, and its shape and safety are of primary concern for both surgeons and patients.[4] It has undergone technological innovation and design improvement over time. From the first to the fifth generation of a breast implant, each generation is characterized by the technological development of its components.[5,6] Most recently, the external surface of the fifth-generation of a breast implant has a contact with viable tissue, and it is characterized by an irregular texture that is expected to stimulate tissue adherence to the implant and subsequently to avoid an anatomical misplacement of a device.[7] Thus, a macrotextured breast implant with varying pore size and depth has been commercially available in the market.[8] During the accumulation of more clinical evidences demonstrating the effects of surface texturization in reducing the occurrence of capsular contracture (CC), a macrotextured breast implant has been reported to have a causal relationship with the onset of breast implant-associated anaplastic large cell lymphoma (BIA-ALCL). Therefore, it has been imperative that a breast implant with a novel surface topography emerge. Manufacturers have commercialized a silicone gel-filled breast implant (SGBI) with a microtextured surface as an innovative technology that may lower risks of both CC and BIA-ALCL.[9] But this has also been contradicted by other studies.[1014]

Although a microtextured breast implant is currently considered a smooth device, it is differentiated from a traditional smooth one, as summarized in Table 1.[1518] Motiva Ergonomix (Establishment Labs Holdings Inc., Alajuela, Costa Rica) and BellaGel SmoothFine (HansBiomed Co. Ltd., Seoul, Korea) have been known as 2 representative brands of a microtextured device in Korea.[16,19] The short-term safety of traditional smooth SGBIs and microtextured devices was evaluated by a recent study.[18] In Korea, there were a crisis from BIA-ALCL and the first Korean case of a medical device fraud.[1618,2022]

Table 1.

Summary of nanotextured or microtextured breast implants in Korea.

Manufacturer Trade name Year of release
Establishment Labs Holdings Inc. (Alajuela, Costa Rica) Motiva Ergonomix 2016
HansBiomed Co. Ltd. (Seoul, Korea) BellaGel SmoothFine 2017
GC Aesthetics PLC (Apt Cedex, France) Eurosilicone Round Collection 2016
Groupe Sebbin SAS (Boissy-l’Aillerie, France) Sebbin Integrity 2018
Sebbin Sublimity (formerly the Naturgel) 2012

Despite great advancements in breast implant technology and surgical technique, a risk of postoperative complications poses a challenge for stakeholders in the breast implant industry (BII).[20,23] Of such complications, both CC and rupture are observed the most prevalently. Still, however, little is known about the exact pathogenesis and pathophysiology of CC, although it may occur through a multifactorial process.[24,25] Rupture is another concern that has a close relationship with the age of a breast implant. A continuous monitoring of patients receiving a device would therefore be mandatory to assess its long-term safety.[26]

Given the above background, we compared incidences of CC and rupture between the microtextured SGBIs after the occurrence of breast implant crisis (BIC) in Korea. Then, we discuss implications of the current results for risk management and sustainable development of BII in Korea in the context of the first Korean case of a medical device fraud.

2. Materials and methods

2.1. Literature review

2.1.1. BIC in Korea

Recently, the BIC was a big issue to those who are involved in the BII in Korea; it can be classified as the first and second BIC.[1618,2022]

2.1.1.1. The first BIC in Korea: BioCell textured device (Allergan Inc., Irvine)

There were 3 Korean cases of BIA-ALCL between 2019 and 2020 in relation to BioCell textured device (Allergan Inc., Irvine). Since August 29, 2019, no textured devices have been used, as mandated by the Korean Ministry of Food and Drug Safety (KMFDS).[17,23]

2.1.1.2. The second BIC in Korea: BellaGel/BellaGel SmoothFine

HansBiomed Co. Ltd. violated the regulatory requirement enforced by the KMFDS. The manufacturer used unapproved, harmful materials, such as 7-9700, Q7-4850, MED2-6300, MED-6400, and MED2-4213 (Fig. 1).[2730]

Figure 1.

Figure 1.

Use of unapproved materials for manufacturing of the BellaGel. A total of 5 constituents were not approved for use for manufacturing of the BellaGel implants; these include 7-9700 (soft skin adhesive), Q7-4850 (liquid silicone rubber), MED2-6300 (silicone gel), MED-6400 (silicone dispersion), and MED2-4213 (skin adhesive).

HansBiomed Co. Ltd. deliberately modified the shell layer of BellaGel SmoothFine in violation of the regulatory requirement. The KMFDS approved the commercialization of BellaGel SmoothFine with a 5-layered shell, although the manufacturer illegally manufactured a 4-layered device to increase its soft feel.[31,32]

2.1.2. The global impact of the first Korean case of a medical device fraud

2.1.2.1. Europe

In France, between 2010 and 2012, HansBiomed Co. Ltd. was involved in the Poly Implant Prothèse (PIP) fraud (Figs. 2 and 3).[3335] PIP was allegedly known as the third-largest global manufacturer of an SGBI from France (Fig. 4).[3638] Therefore, it deserves serious consideration that HansBiomed Co. Ltd. was involved in manufacturing the PIP breast implants under different brand names, such as M-implants.[3942]

Figure 2.

Figure 2.

A snapshot of the website of the Rofil Medical Implants Ltd.

Figure 3.

Figure 3.

Involvement of the HansBiomed Co. Ltd. in the Poly Implant Prothèse fraud.

Figure 4.

Figure 4.

A snapshot of the website of the Poly Implant Prothèse.

2.1.2.2. China

According to the news report dated November 12, 2020, HansBiomed Co. Ltd. avoided being inspected by the National Medical Products Administration (NMPA, formerly China Food and Drug Administration) when entering the Chinese market for the commercialization of the BellaGel breast implants in China. After obtaining the regulatory approval for them from the KMFDS in the late 2015, HansBiomed Co. Ltd. accelerated the entry into the overseas market. The manufacturer finally obtained the regulatory approval of BellaGel Smooth from the NMPA in 2016.[43]

In October of 2018, HansBiomed Co. Ltd. held a press conference celebrating a launch of the BellaGel SmoothFine (formerly BellaGel Micro, BellaGel breast implant with a microtextured surface) in Seoul; the manufacturer had a plan to enter the Chinese market since 2019. Nevertheless, the manufacturer had internal problems; the NMPA performed a rapid inspection of BellaGel breast implants that were commercially available in China in about June of 2018. In November of 2018, the manufacturer received the results from a certification testing institute located in China; a responsible officer of HansBiomed China Inc. (Shanghai, China) urgently sent an e-mail to the head office in Seoul, Korea, who reported that BellaGel breast implants did not pass the test for evaporation residue, trace element, and shell thickness. There were considerable differences in criteria for accepting the quality of breast implants based on impurities on the evaporation residue test between China and Korea. In more detail, only the sample is extracted from the device and then assayed for impurities in Korea. In China, however, the sample is boiled and the residue from the vapor is assayed accordingly. Based on the results showing that BellaGel breast implants did not pass trace element test, it can be inferred that they contain heavy metals at higher levels than the standard limit. But the manufacturer did not identify any problems from raw materials. The manufacturer therefore concluded that it would be impossible to pass the test even if a retest is performed.[43]

Nevertheless, the manufacturer decided to resolve the issue using “Guanxi,” referred to as a direct financial relationship with the Chinese government, by asking officials of the NMPA to falsify the results of the test and estimated the payment at approximately CNY 15,000,000.[43]

2.1.3. Changes in the Korean breast implant market after the occurrence of BIC in Korea

According to the KMFDS statistics, approximately 77,000 breast implants are annually circulated in the Korean market between 2016 and 2020. After the occurrence of BIC in Korea, however, there were notable changes in the market; Mentor Worldwide LLC. was the manufacturer whose products were sold the most prevalently (n = 15,570). This is followed by Establishment Labs Holdings Inc. (n = 9732), Groupe Sebbin SAS (n = 7374), GC Aesthetics PLC (n = 1406), Allergan Inc. (n = 145), and Silimed Inc. (n = 2) in the decreasing order (Table 2 and Fig. 5).[44]

Table 2.

The annual number of breast implants that were circulated in the Korean market over a 6-year period (2016–June 2021).

Manufacturer 2016 2017 2018 2019 2020 June 2021
Allergan Inc. (Irvine, CA) 19,231 23,395 11,782 2251 2641 145
Polytech Health & Aesthetics (Dieburg, Germany) 5261 2169 804 12 No longer available
Silimed Inc. (Rio de Janeiro, Brazil) 1187 2000 744 275 377 2
Groupe Sebbin SAS (Boissy-l’Aillerie, France) 8964 14,585 10,063 10,911 2276 7374
HansBiomed Co. Ltd. (Seoul, Korea) 8573 5396 19,752 27,252 33,067 No longer available
GC Aesthetics PLC (Apt Cedex, France) 3139 3991 6955 3536 586 1406
Establishment Labs Holdings Inc. (Alajuela, Costa Rica) 3848 12,635 18,717 22,565 20,078 9732
Mentor Worldwide LLC (Santa Barbara, CA) 19,994 16,509 9118 11,914 17,237 15,570
Total 70,197 80,680 77,935 78,716 76,262 34,229
Figure 5.

Figure 5.

The annual number of breast implants that were circulated in the Korean market over a 6-year period (2016–June 2021). (A) The amount of sales between 2016 and 2020 and (B) the market share of a device between January and June, 2021.

In November 13, 2020, the KMFDS mandated a 6-month suspension of the manufacture and sales of BellaGel breast implants.[45] According to the new report dated September 22, 2023, however, HansBiomed Co. Ltd. canceled a lawsuit against the KMFDS; the manufacturer announced that it would resume the BellaGel business in March of 2023.[46]

Patients receiving the BellaGel SmoothFine complained of the manufacturer’s fraudulent practices. But both the manufacturer and some plastic surgeons denied any causal relationships between the defective BellaGel SmoothFine and patients’ complaints.[47] Many patients even filed a lawsuit against the HansBiomed Co. Ltd. and thereby sought monetary compensation for damages caused by the manufacturer. The amount of such compensation was known to reach approximately USD 3.7 million.[47] Nevertheless, HansBiomed Co. Ltd. claimed that a finished product of BellaGel SmoothFine is safe because it does not emit toxic substances although the manufacturer admitted the illegal use of unapproved materials.[47] But the manufacturer did not admit that it deliberately modified the shell structure during manufacturing process.[48]

2.2. Study patients and setting

The current multicenter, retrospective observational study was conducted in a cohort of Korean women who had received an implant-based augmentation mammaplasty for aesthetic purposes during a 2-year period between November 14, 2020 and October 13, 2022. Inclusion/exclusion criteria for the current study are summarized in Table 3.

Table 3.

Eligibility criteria.

Eligibility criteria
Inclusion criteria
(1) Women aged 22 years or older
(2) Women who underwent primary augmentation mammaplasty using nanotextured or microtextured SGBIs for aesthetic purposes
(3) Women with an adequate amount of tissue for coverage of the breast implant
(4) Women with availability of follow-up data
Exclusion criteria
(1) Women with unilateral or bilateral presence of premalignant breast lesions
(2) Women with untreated malignancies
(3) Women with an inadequate amount of or inappropriate tissue for coverage of the breast implant because of radiation-induced damage, vascular compromise or impaired wound healing
(4) Women with abscess or infection
(5) Women who had a past history of taking any drugs that may interfere with blood clotting or raise risks of developing postoperative complications
(6) Women with underlying medical conditions that may raise risks of developing postoperative complications (e.g., obesity [BMI ≥ 40], DM, autoimmune disease, chronic lung, severe cardiovascular disease, connective tissue or rheumatoid disease)
(7) Women who are pregnant or breastfeeding
(8) Women with medical conditions that may interfere with wound healing (e.g., active infectious collagen disease)
(9) Women with severe lung disease (e.g., chronic obstructive pulmonary disease)
(10) Women with active cutaneous or systemic infections
(11) Women undergoing radiotherapy or chemotherapy within 6 months preoperatively
(12) Women lost to follow-up

BMI = body mass index, DM = diabetes mellitus, SGBI = silicone gel-filled breast implant.

2.3. Ethics statement

We obtained the ethical approval of the current study from the Internal Institutional Review Board of the Korea National Institute of Bioethics Policy (Institutional Review Board approval #: P01-202101-21-021) and conducted it in compliance with the relevant guidelines and applicable laws. However, we failed to receive a written informed consent from the patients because of retrospective nature of the current study.

2.4. Treatment, follow-up and assessment of the patients

The patients were surgically treated and then regularly followed up, as previously described.[1719,4951]

We analyzed age, body mass index (BMI), follow-up (FU) period and incidences of the postoperative complications, such as CC and rupture.[51,52]

We excluded risk factors of postoperative complications, such as an FU period of <1 year, a BMI ≥ 30 kg/m2 and trans-axillary or peri-areolar incisions, in analyzing the safety of microtextured SGBIs.[5361]

Data was presented as the number of the patients with percentage, mean ± standard deviation or mean ± standard error and then analyzed using the SPSS ver. 18.0 for windows (SPSS Inc., Chicago). Differences in incidences of postoperative complications between the microtextured SGBIs were analyzed using the repeated measures analysis of variance and Duncan post hoc analysis. Moreover, time to CC, CC-free survival and cumulative hazards were estimated. Log-rank test was also performed to assess the statistical significance and 95% confidence intervals (CIs) were provided. Statistical significance was set at P < .05.

3. Results

3.1. Baseline characteristics of the patients

A total of 801 Korean women (n = 801) were finally included in the current study; they received an implant-based augmentation mammaplasty using Sebbin Sublimity (n = 577), Motiva Ergonomix (n = 168), Sebbin Integrity (n = 37), and Eurosilicone Round Collection (n = 19). There were no significant differences in age, BMI and the length of FU period (P > .05) (Table 4).

Table 4.

Baseline characteristics of the patients (n = 801).

Values P-value
Sebbin sublimity
(n = 577)
Motiva ergonomix
(n = 168)
Sebbin integrity
(n = 37)
Eurosilicone round collection
(n = 19)
Age (yr old) 32.14 ± 4.18 31.27 ± 5.47 28.36 ± 4.18 32.77 ± 3.89 >.05
BMI (kg/m2) 24.82 ± 0.04 24.79 ± 1.03 24.88 ± 1.27 25.01 ± 2.01
FU (mo) 12.13 ± 0.11 12.77 ± 0.68 12.25 ± 0.18 12.64 ± 0.52

Values are mean ± standard deviation.

BMI = body mass index, FU = follow-up.

3.2. Differences in incidences of postoperative complications between the breast implants

Incidences of CC were 1.79% (3/168), 3.64% (21/577), 8.11% (3/37), and 10.53% (2/19) in the patients receiving Motiva Ergonomix, Sebbin Sublimity, Sebbin Integrity, and Eurosilicone Round Collection, respectively, in the increasing order. These differences reached statistical significance (P < .05). There were 2 women with rupture after receiving Sebbin Sublimity, although there were no cases of rupture in association with other brands of breast implants (Tables 5 and 6).

Table 5.

Incidences of postoperative complications (n = 801).

Values P-value
Sebbin Sublimity
(n = 577)
Motiva Ergonomix
(n = 168)
Sebbin Integrity
(n = 37)
Eurosilicone Round Collection
(n = 19)
CC 21 (3.64%) 3 (1.79%) 3 (8.11%) 2 (10.53%) <.05*
Rupture 2 (0.35%) 0 (0.00%) 0 (0.00%) 0 (0.00%)

Values are the number of the patients with percentage.

CC = capsular contracture.

*

Statistical significance at P < .05.

Table 6.

Differences in incidences of capsular contracture (n = 801).

Breast implants N n Censored values
Sebbin Sublimity 577 21 556 (96.4%)
Motiva Ergonomix 168 3 165 (98.2%)
Sebbin Integrity 37 3 34 (91.9%)
Eurosilicone Round Collection 19 2 17 (89.5%)

n = incidences of capsular contracture, N = total number of the patients.

Results of log-rank test showed that incidences of CC were the significantly lowest in the patients receiving Motiva Ergonomix (P < .05) (Table 7).

Table 7.

Results of log-rank test (n = 801).

Comparisons χ 2 P-value
Sebbin Sublimity vs Motiva Ergonomix 4.600 .032*
Sebbin Integrity vs Motiva Ergonomix 5.466 .019*
Eurosilicone Round Collection vs Motiva Ergonomix 4.665 .031*
*

Statistical significance at P < .05.

Overall, CC-free survival was estimated at 681.470 ± 8.314 (95% CI 665.174–697.766) days. By breast implants, it was 708.899 ± 8.595 (95% CI 692.053–725.745), 599.327 ± 6.607 (95% CI 586.378–612.277), 584.941 ± 22.965 (95% CI 539.931–629.952) and 572.492 ± 37.374 (95% CI 499.240–645.745) days in the patients receiving Motiva Ergonomix, Sebbin Sublimity, Sebbin Integrity, and Eurosilicone Round Collection, respectively, in the increasing order. This showed that Motiva Ergonomix achieved the longest survival in our series (Table 8).

Table 8.

Capsular contracture-free survival (n = 801).

Breast implants CC-free survival
Total 681.470 ± 8.314 (665.174–697.766)
Sebbin Sublimity 599.327 ± 6.607 (586.378–612.277)
Motiva Ergonomix 708.899 ± 8.595 (692.053–725.745)
Sebbin Integrity 584.941 ± 22.965 (539.931–629.952)
Eurosilicone Round Collection 572.492 ± 37.374 (499.240–645.745)

Values are mean ± standard error with 95% confidence intervals.

CC = capsular contracture.

Cumulative rates of CC are summarized in Table 9. Kaplan–Meier cumulative survival and hazards are plotted in Figures 6 and 7, respectively.

Table 9.

Cumulative rates of capsular contracture (n = 801).

Breast implants FU N n Cumulative rates of CC
Sebbin Sublimity 14 526 1 0.998 ± 0.002 (0.994–1.000)
42 490 1 0.996 ± 0.003 (0.991–1.000)
80 452 1 0.994 ± 0.004 (0.987–1.000)
122 410 1 0.991 ± 0.004 (0.983–1.000)
202 324 1 0.988 ± 0.005 (0.978–0.999)
223 310 1 0.985 ± 0.006 (0.973–0.997)
250 292 1 0.982 ± 0.007 (0.968–0.996)
271 278 1 0.978 ± 0.008 (0.963–0.994)
322 244 1 0.974 ± 0.009 (0.957–0.992)
342 230 1 0.970 ± 0.010 (0.951–0.989)
400 190 1 0.965 ± 0.011 (0.944–0.986)
430 177 1 0.959 ± 0.012 (0.936–0.984)
445 162 2 0.948 ± 0.015 (0.920–0.977)
451 142 1 0.941 ± 0.016 (0.910–0.973)
454 131 1 0.934 ± 0.017 (0.900–0.968)
457 117 1 0.926 ± 0.019 (0.889–0.964)
469 64 1 0.911 ± 0.024 (0.866–0.959)
494 42 1 0.890 ± 0.031 (0.830–0.953)
584 13 1 0.821 ± 0.072 (0.692–0.975)
623 7 1 0.704 ± 0.125 (0.497–0.997)
Motiva Ergonomix 108 141 1 0.993 ± 0.007 (0.979–1.000)
420 63 1 0.977 ± 0.017 (0.944–1.000)
572 25 1 0.938 ± 0.042 (0.860–1.000)
Sebbin Integrity 320 17 1 0.941 ± 0.057 (0.836–1.000)
500 8 1 0.824 ± 0.121 (0.618–1.000)
560 5 1 0.659 ± 0.176 (0.390–1.000)
Eurosilicone Round Collection 228 13 1 0.923 ± 0.074 (0.789–1.000)
342 10 1 0.831 ± 0.110 (0.641–1.000)

Values are mean ± standard error with 95% confidence intervals.

CC = capsular contracture, FU = follow-up, n = incidences of capsular contracture, N = number of risks.

Figure 6.

Figure 6.

Kaplan–Meier survival (n = 801). CC = capsular contracture.

Figure 7.

Figure 7.

Kaplan–Meier hazards (n = 801). CC = capsular contracture.

4. Discussion

Vulnerability of the BII to crisis has been previously described, thus termed as the BIC.[62] Moreover, the BIC has been classified into the first BIC (Dow Corning), the second BIC (PIP), and the third BIC (BIA-ALCL).[63] The BII in Korea has also experienced BIC; the first and second BIC in Korea. Of note, however, the BIC in Korea showed a great difference from that in other countries. As described earlier, the manufacturer announced that it would resume the BellaGel business in March of 2023; the safety of finished products of BellaGel/BellaGel SmoothFine has not been further questioned by both the manufacturer and the KMFDS since November 3, 2020.[46] Thus, the manufacturer survived the BIC in Korea and was exempted from the safety issue. Other manufacturers than HansBiomed Co. Ltd. also spontaneously received a letter of indulgence because HansBiomed Co. Ltd. informed the public that even the illegally manufactured breast implants could not be questioned for their safety.

A medical device risk is defined as the likelihood of occurrence of a damage combined with its severity according to the Article II of EU MDR 2017/745. All medical devices are inherently susceptible to a certain level of risks. Therefore, the medical device risk management is a continuous, iterative process in which manufacturers should plan, document and implement risk management strategies that may either eliminate the risk or mitigate the overall severity of it. This is essential for preventing specific harm or risk.[64] Despite the medical device risk management strategies of the manufacturer, however, both a whistleblower and news media played a key role in reporting the first Korean case of a medical device fraud to the public.

It would be worthwhile to analyze the first Korean case of a medical device fraud, namely the second BIC in Korea, based on benchmark cases in risk and crisis management. Tylenol (Johnson & Johnson, New Brunswick) and the Dow Corning SGBI are 2 representative benchmark cases; the former and the latter indicate how to and how not to manage a crisis, respectively. That is, both cases are closely associated with public health and safety issues, although they were managed very differently. As a result, the reputation and image of Dow Coming were considerably impaired.[65,66] Therefore, it remains problematic that the manufacturer of BellaGel/BellaGel SmoothFine continued to lie to the public.[48]

To summarize, our results are as follows: first, incidences of CC were 1.79% (3/168), 3.64% (21/577), 8.11% (3/37), and 10.53% (2/19) in the patients receiving Motiva Ergonomix, Sebbin Sublimity, Sebbin Integrity, and Eurosilicone Round Collection, respectively. These differences reached statistical significance (P < .05). Second, there were 2 women with rupture after receiving Sebbin Sublimity, although there were no cases of rupture in association with other brands of breast implants. Third, overall CC-free survival was estimated at 681.470 ± 8.314 (95% CI 665.174–697.766) days. By breast implants, it was 708.899 ± 8.595 (95% CI 692.053–725.745), 599.327 ± 6.607 (95% CI 586.378–612.277), 584.941 ± 22.965 (95% CI 539.931–629.952), and 572.492 ± 37.374 (95% CI 499.240–645.745) days in the patients receiving Motiva Ergonomix, Sebbin Sublimity, Sebbin Integrity, and Eurosilicone Round Collection, respectively, in the increasing order. These results indicate that Motiva Ergonomix is a relatively safer device as compared with others in the context of the first Korean case of a medical device fraud, as supported by previous literatures.[17,22,67,68]

But our results cannot be generalized because we failed to identify a causal relationship of a risk of CC with differences in the surface topography of between the breast implants. But our results are advocated by a recent study. Doloff et al recently conducted an experimental study to compare differences in immune responses depending on the surface topography of an SGBI between the commercially-available products. These include Motiva Ergonomix Round SilkSurface and VelvetSurface (microtextured; Establishment Labs Holdings Inc.), Mentor Smooth (smooth) and Mentor MemoryGel SILTEX (microtextured) (Mentor Worldwide LLC., Santa Barbara), Microcell (microtextured) and Natrelle INSPIRA Biocell (macrotextured) (Allergan Inc., Irvine). Doloff et al also tested a hypothesis that the human-sized, commercial and miniaturized sample of Motiva Ergonomix Round SilkSurface might alter the kinetics and characteristics of foreign body responses, thus showing that it had long-term effects in inhibiting the fibrosis in New Zealand White (NZW) rabbits (≤1 year) and C57BL/6 mice (≤6 months), respectively. These authors also compared the profile of fibrosis between wild-type and T-cell-deficient C57BL/6 mice, thus showing that there was a significant decrease in the number of macrophages only in wild-type mice. Moreover, Doloff et al also showed that the immune responses of the NZW rabbits and C57BL/6 mice were matched to those seen in human clinical specimens that were collected from the luminal surface of scar capsules formed for 7 months to 11 years in patients receiving an implant-based augmentation mammaplasty.[69] More recent studies have supported the efficacy and safety of Motiva breast implants. Mojo and Nele performed an implant-based augmentation mammaplasty for a consecutive series of a total of 325 patients who were followed up during a mean period of 14 months. These authors reported that Motiva breast implants could provide a favorable choice for balancing both safety and aesthetics in a cohort of patients receiving primary augmentation procedures.[70] Randquist et al performed a total of 1053 primary and secondary breast augmentations, thus conducting a 6-year retrospective follow-up study. These authors demonstrated a low complication rate and safety profile of Motiva breast implants.[71] Lee et al conducted a 3-year single-center retrospective study in a total of 1324 patients comprising 1027 primary cases and 297 secondary ones, thus reporting rates of CC of 1.07% and 4.97%, respectively.[72] The Motiva US IDE is a prospective, single-arm, multicenter, 10-year pivotal study; it collected data from patients receiving augmentation mammaplasty, reconstruction and revisional surgery and then submitted it to the United States Food and Drug Administration. Glicksman et al reviewed the 3-year data about the efficacy and safety of Motiva breast implants. According to these authors, there were low rates of implant-related complications and high rates of patients’ and surgeons’ satisfaction with postoperative outcomes.[73]

5. Conclusion

According to the National Council for Advanced Manufacturing in the United States, sustainable manufacturing is defined as the manufacturing of sustainable products and the sustainable manufacturing of all the products.[74] The ISO 13485 specifies requirements for a quality management system for organizations required to demonstrate their ability to provide medical devices that consistently meet regulatory requirements, according to which the manufacturer develops and produces under guidelines mandated by the United States Food and Drug Administration.[75]

According to Ghadimi and Heavey, the social sustainability of a medical device includes both its safety and a patient’s health.[76] Therefore, the first Korean case of a medical device fraud should be considered as a lack of the social sustainability of the manufacturer of BellaGel/BellaGel SmoothFine despite regulatory requirements enforced by the KMFDS.

In conclusion, our results indicate that Motiva Ergonomix Round SilkSurface is currently a relatively safer device as compared with others in the context of the first Korean case of a medical device fraud. The BII in Korea should be aware of the importance of the social sustainability in manufacturing a device.

Author contributions

Conceptualization: Robert Kim.

Data curation: Robert Kim.

Formal analysis: Robert Kim.

Funding acquisition: Robert Kim.

Investigation: Robert Kim.

Methodology: Robert Kim.

Project administration: Robert Kim.

Resources: Robert Kim.

Software: Robert Kim.

Supervision: Robert Kim.

Validation: Robert Kim.

Visualization: Robert Kim.

Writing – original draft: Robert Kim.

Writing – review & editing: Robert Kim.

Abbreviations:

BIA-ALCL
breast implant-associated anaplastic large cell lymphoma
BIC
breast implant crisis
BII
breast implant industry
BMI
body mass index
CC
capsular contracture
CIs
confidence intervals
FU
follow-up
KMFDS
Korean Ministry of Food and Drug Safety
NMPA
National Medical Products Administration
PIP
Poly Implant Prothèse
SGBI
silicone gel-filled breast implant

The authors thank Laon Medi Solution Inc. (Seoul, Republic of Korea) and KDH Medical Inc. (Gwangmyeong, Gyeonggi, Republic of Korea) for additional support of this research. Funding number: KDH-2023.

The author has no conflicts of interest to disclose.

The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.

How to cite this article: Kim R, Breast Implant Research Group. Current status of an implant-based augmentation mammaplasty in Korean women: Implications for risk management and social sustainability of breast implant industry in the context of the first Korean case of a medical device fraud. Medicine 2025;104:36(e44092).

References

  • [1].Kumaravel TS, Sathya TN, Balaje R, et al. Genotoxicity evaluation of medical devices: a regulatory perspective. Mutat Res Rev Mutat Res. 2022;789:108407. [DOI] [PubMed] [Google Scholar]
  • [2].Keil O, Widmann DE. Assessment of the impact of medical devices on the quality of care. QRB Qual Rev Bull. 1984;10:278–80. [PubMed] [Google Scholar]
  • [3].Sujan MA, Koornneef F, Chozos N, Pozzi S, Kelly T. Safety cases for medical devices and health information technology: involving health-care organizations in the assurance of safety. Health Informatics J. 2013;19:165–82. [DOI] [PubMed] [Google Scholar]
  • [4].Chao AH, Garza R, 3rd, Povoski SP. A review of the use of silicone implants in breast surgery. Expert Rev Med Devices. 2016;13:143–56. [DOI] [PubMed] [Google Scholar]
  • [5].Hong P, Kang JK, Hwang SH, Lee KA. Feasibility of imaging modalities combined with a silicone gel-filled breast implant in Korean women. Gels. 2023;9:232. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [6].Aitzetmüller-Klietz ML, Yang S, Wiebringhaus P, et al. Complication rates after breast surgery with the motiva smooth silk surface silicone gel implants – a systematic review and meta-analysis. J Clin Med. 2023;12:1881. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [7].Maxwell GP, Scheflan M, Spear S, Nava MB, Hedén P. Benefits and limitations of macrotextured breast implants and consensus recommendations for optimizing their effectiveness. Aesthet Surg J. 2014;34:876–81. [DOI] [PubMed] [Google Scholar]
  • [8].Atlan M, Bigerelle M, Larreta-garde V, Hindié M, Hedén P. Characterization of breast implant surfaces, shapes, and biomechanics: a comparison of high cohesive anatomically shaped textured silicone, breast implants from three different manufacturers. Aesthetic Plast Surg. 2016;40:89–97. [DOI] [PubMed] [Google Scholar]
  • [9].Montemurro P, Tay VKS. Transitioning from conventional textured to nanotextured breast implants: our early experience and modifications for optimal breast augmentation outcomes. Aesthet Surg J. 2021;41:189–95. [DOI] [PubMed] [Google Scholar]
  • [10].Nava MB, Catanuto G, Rocco N. Nanotextured breast implants: not a solution for all seasons. Aesthet Surg J. 2021;41:NP137–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [11].Nava MB, Rancati A, De Vita R, Catanuto G, Rocco N. Clarification about “Expert consensus on the use of a new bioengineered, cell-friendly, smooth surface breast implant.”. Aesthet Surg J. 2019;39:NP538–9. [DOI] [PubMed] [Google Scholar]
  • [12].Hamdi M. Nano-surface implants: indications and limitations. Aesthet Surg J. 2021;41:NP1141–2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [13].Nava MB, Catanuto G, De Vita R, Rancati A, Rocco N. Comment on: breast implant surfaces and their impact on current practices: where are we now and where are we going. Plast Reconstr Surg Glob Open. 2020;8:e2639. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [14].Duscher D, Aitzetmüller MM, Wenny R, Huemer GM. Comments on “Preliminary 3-year evaluation of experience with SilkSurface and VelvetSurface Motiva silicone breast implants: a single-center experience with 5813 consecutive breast augmentation cases”. Aesthet Surg J. 2019;39:NP18–9. [DOI] [PubMed] [Google Scholar]
  • [15].Munhoz AM, Clemens MW, Nahabedian MY. Breast implant surfaces and their impact on current practices: where we are now and where are we going? Plast Reconstr Surg Glob Open. 2019;7:e2466. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [16].Kim JH. Association of the BellaGel® breast implant scandal with the poly implant prothèse fraud: a review of literatures. J Surg Open Access. 2020;7:1–10. [Google Scholar]
  • [17].Lee S, Jeong JP, Sung JY, et al. High-resolution ultrasound-assisted assessment of preliminary short-term safety outcomes of an implant-based augmentation mammaplasty using a bioengineered, cell-friendly, smooth-surface device in Korean females. Aesthet Surg J Open Forum. 2021;4:ojab046. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [18].Han S, Kim R, Kim TS, et al. A preliminary retrospective study to assess the short-term safety of traditional smooth or microtextured silicone gel-filled breast implants in Korea. Medicina (Kaunas). 2021;57:1370. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [19].Moon DS, Choi WS, Kim HC, Jeong JP, Sung JY, Kim JH. Short-term treatment outcomes and safety of two representative brands of the fifth-generation silicone gel-filled breast implants in Korea. J Plast Surg Hand Surg. 2021;55:345–53. [DOI] [PubMed] [Google Scholar]
  • [20].Kim JH, Paik NS, Nam SY, Cho Y, Park HK. The emerging crisis of stakeholders in implant-based augmentation mammaplasty in Korea. J Korean Med Sci. 2020;35:e103. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [21].Thornton RK, DeVito CA, Peters W, et al. Surface texturing of silicone breast implants: development and efficacy. Plast Reconstr Surg. 1992;90:862–9. [Google Scholar]
  • [22].James W, Vargo J, Digby C, et al. Characterizing smooth breast implant surfaces: an evaluation of process-related variables. Plast Reconstr Surg Glob Open. 2019;7:e2256.31333975 [Google Scholar]
  • [23].De Vita R, Buccheri EM, Rancati A, et al. A clinical evaluation of microtextured implants: optimizing aesthetic outcomes. Aesthetic Plast Surg. 2019;43:1112–20. [Google Scholar]
  • [24].Rafn B, Orlando JC, Aitken RJ. Novel insights into biofilm formation on nanotextured implant surfaces. J Biomed Mater Res A. 2020;108:1527–36. [Google Scholar]
  • [25].Piper C, Wilson C, Jones D, et al. The evolution of breast implant texturing and its implications on capsular contracture rates. Eplasty. 2019;19:e25. [Google Scholar]
  • [26].Munhoz AM, Clemens MW, Nahabedian MY. Comprehensive analysis of safety concerns associated with textured breast implants. Clin Plast Surg. 2020;47:213–24. [Google Scholar]
  • [27].Headon HL, Sinnot CJ, Kasem A, Mokbel K. Capsular contracture rates in textured versus smooth implants: a systematic review. Arch Plast Surg. 2015;42:508–18. [Google Scholar]
  • [28].Collis N, Sharpe DT. Textured or smooth implants for breast augmentation? Three-year follow-up of a prospective randomised controlled trial. Br J Plast Surg. 2000;53:481–4. [DOI] [PubMed] [Google Scholar]
  • [29].Marotta JS, Alderman AK, Kim J, et al. Characterizing and optimizing implant surfaces to address implant-related complications: an integrative review. Plast Reconstr Surg. 2021;147:561e–7e. [Google Scholar]
  • [30].Ault A. US institute of medicine panel deliberates on breast-implant safety. Lancet. 1998;352:380. [DOI] [PubMed] [Google Scholar]
  • [31].Albornoz CR, Bach PB, Mehrara BJ, et al. A paradigm shift in U.S. breast reconstruction: increasing implant rates. Plast Reconstr Surg. 2013;131:15–23. [DOI] [PubMed] [Google Scholar]
  • [32].Adams WP, Jr, Conner WC, Rohrich RJ. Optimizing breast aesthetics with smooth versus textured silicone gel implants: a clinical analysis. Plast Reconstr Surg. 2001;107:1612–24. [Google Scholar]
  • [33].Spear SL, Murphy DK, Slicton A, Walker PS; Inamed Silicone Breast Implant U.S. Study Group. Inamed silicone breast implant core study results at 6 years. Plast Reconstr Surg. 2007;120:8S–16S. [DOI] [PubMed] [Google Scholar]
  • [34].Brown MH, Shenker R, Silver SA. Cohesive silicone gel breast implants in aesthetic and reconstructive breast surgery. Plast Reconstr Surg. 2005;116:768–79; discussion 780. [DOI] [PubMed] [Google Scholar]
  • [35].Cunningham B. The mentor core study on silicone memorygel breast implants. Plast Reconstr Surg. 2007;120:19S–29S. [DOI] [PubMed] [Google Scholar]
  • [36].Wixtrom RN, Dannucci GA, Matlaga BR, et al. Comparative analysis of adverse event profiles between smooth and textured implants in a national registry. Aesthetic Plast Surg. 2021;45:1290–300. [Google Scholar]
  • [37].Wong CH, Sahar DE, Garfein ES, et al. Textured implants in reconstructive breast surgery: a multicenter study on long-term outcomes. Ann Plast Surg. 2006;57:22–6. [Google Scholar]
  • [38].Gosain A, Jiang L, Schuler K, et al. Long-term clinical outcomes of nano- and macrotextured implants: a retrospective cohort study. Clin Plast Surg. 2021;48:331–40.33674054 [Google Scholar]
  • [39].Clemens MW, Medeiros LJ, Butler CE, et al. CD30+ T-cell lymphoma associated with breast implants: a histologic review and clinical update. Mod Pathol. 2015;28:1167–73.26111975 [Google Scholar]
  • [40].Doren EL, Miranda RN, Selber JC, et al. U.S. epidemiology of breast implant–associated anaplastic large cell lymphoma. Plast Reconstr Surg. 2017;139:1042–50. [DOI] [PubMed] [Google Scholar]
  • [41].Rosnoblet JF. PIP implants sold to Dutch firm under new name. https://de.reuters.com/article/us-breast-implants-netherlands/pip-implants-sold-to-dutch-firm-under-new-name-idUSTRE7BP0H320111226. Accessed October 13, 2022.
  • [42].Finnish woman got dangerous breast implant in Estonia. https://shaan.typepad.com/shaanou/2011/12/finnish-woman-got-dangerous-breast-implant-in-estonia.html. Accessed October 13, 2022.
  • [43].Hong WR. With Guanxi and sample fabrication, HansBiomed Co. Ltd. entered the Chinese market. https://newstapa.org/article/x1fB5. Accessed December 15, 2022.
  • [44].Bang BS, Jung SH, Lee EK, et al. A surgeon’s empirical perspectives on use of high-resolution ultrasound in preoperatively detecting a rupture in the context of breast implant crisis in Korea. Aesthetic Plast Surg. 2022;46:1668–78. [DOI] [PubMed] [Google Scholar]
  • [45].Koo KY. Is it possible that HansBiomed Co. Ltd. will resume the BellaGel® business? https://www.dailymedi.com/news/news_view.php?wr_id=888349. Accessed January 2, 2023.
  • [46].Song YD. [Exclusive] uncertainty resolved; HansBiomed Co. Ltd. presenting a plan to resume the BellaGel business. https://pharm.edaily.co.kr/news/read?newsId=02302566632458152&mediaCodeNo=257. Accessed January 2, 2023.
  • [47].Paek CW. HansBiomed Co. Ltd. facing a lawsuit. The amount of monetary compensation for possible damages by the “BellaGel®” reaching approximately USD 3.7. https://cnews.pinpointnews.co.kr/view.php?ud=202101111256411192cd1e7f0bdf_45. Accessed January 17, 2021.
  • [48].Park CY. No safety issues with the BellaGel® factual errors in news should be corrected. https://www.medisobizanews.com/news/articleView.html?idxno=74699. Accessed December 27, 2022.
  • [49].Sung JY, Jeong JP, Moon DS, et al. Short-term safety of augmentation mammaplasty using the BellaGel implants in Korean women. Plast Reconstr Surg Glob Open. 2019;7:e2566. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [50].Yoon S, Chang JH. Short-term safety of a silicone gel-filled breast implant: a manufacturer-sponsored, retrospective study. Plast Reconstr Surg Glob Open. 2020;8:e2807. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [51].Hong P, Kim SS, Jeong C, et al. Four-year interim results of the safety of augmentation mammaplasty using the Motiva Ergonomix™ Round SilkSurface: a multicenter, retrospective study. Aesthetic Plast Surg. 2021;45:895–903. [DOI] [PubMed] [Google Scholar]
  • [52].Nava MB, Rancati A, Angrigiani C, Catanuto G, Rocco N. How to prevent complications in breast augmentation. Gland Surg. 2017;6:210–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [53].Schaub TA, Ahmad J, Rohrich RJ. Capsular contracture with breast implants in the cosmetic patient: saline versus silicone – a systematic review of the literature. Plast Reconstr Surg. 2010;126:2140–9. [DOI] [PubMed] [Google Scholar]
  • [54].Adams WP, Jr, Rios JL, Smith SJ. Enhancing patient outcomes in aesthetic and reconstructive breast surgery using triple antibiotic breast irrigation: six-year prospective clinical study. Plast Reconstr Surg. 2006;118:46S–52S. [DOI] [PubMed] [Google Scholar]
  • [55].Ribeiro LM, Meireles RP, Brito IM, et al. Impact of body mass index, age and tobacco use on the outcomes of immediate breast reconstruction with implants and acellular dermal matrix. Indian J Plast Surg. 2021;54:350–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [56].Baker NF, Brown O, Hart AM, Danko D, Stewart CM, Thompson PW. Preventing infection in implant-based breast reconstruction: evaluating the evidence for common practices and standardized protocols. Plast Reconstr Surg Glob Open. 2022;10:e4208. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [57].Panayi AC, Agha RA, Sieber BA, Orgill DP. Impact of obesity on outcomes in breast reconstruction: a systematic review and meta-analysis. J Reconstr Microsurg. 2018;34:363–75. [DOI] [PubMed] [Google Scholar]
  • [58].Dancey A, Nassimizadeh A, Levick P. Capsular contracture – what are the risk factors? A 14-year series of 1400 consecutive augmentations. J Plast Reconstr Aesthet Surg. 2012;65:213–8. [DOI] [PubMed] [Google Scholar]
  • [59].Jacobson JM, Gatti ME, Schaffner AD, Hill LM, Spear SL. Effect of incision choice on outcomes in primary breast augmentation. Aesthet Surg J. 2012;32:456–62. [DOI] [PubMed] [Google Scholar]
  • [60].Wiener TC. Relationship of incision choice to capsular contracture. Aesthetic Plast Surg. 2008;32:303–6. [DOI] [PubMed] [Google Scholar]
  • [61].Araco A, Gravante G, Araco F, Delogu D, Cervelli V, Walgenbach K. A retrospective analysis of 3000 primary aesthetic breast augmentations: postoperative complications and associated factors. Aesthetic Plast Surg. 2007;31:532–9. [DOI] [PubMed] [Google Scholar]
  • [62].Deva AK, Cuss A, Magnusson M, Cooter R. The “game of implants”: a perspective on the crisis-prone history of breast implants. Aesthet Surg J. 2019;39:S55–65. [DOI] [PubMed] [Google Scholar]
  • [63].Groth K, Graf R. Breast implant-associated anaplastic large cell lymphoma (BIA-ALCL) and the textured breast implant crisis. Aesthetic Plast Surg. 2020;44:1–12. [DOI] [PubMed] [Google Scholar]
  • [64].Risk management of medical devices under MDR. https://omcmedical.com/risk-management-of-medical-devices-under-mdr/. Accessed April 7, 2023.
  • [65].Trinkhaus J, Nathan J, Beane L, Meltzer B. Acetaminophen (Tylenol): Johnson & Johnson and consumer safety. J Law Med Ethics. 1997;25:49–57, 3. [DOI] [PubMed] [Google Scholar]
  • [66].Di Pompeo FS, Paolini G, Firmani G, Sorotos M. History of breast implants: back to the future. JPRAS Open. 2022;32:166–77. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [67].Kang JJ, Lee H, Park BH, et al. Efficacy of a 4-week nurse-led exercise rehabilitation program in improving the quality of life in women receiving a post-mastectomy reconstruction using the Motiva Ergonomix™ Round SilkSurface. Int J Environ Res Public Health. 2023;20:16. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [68].Song KY, Sung JY, Choi WS, et al. An ultrasound-assisted approach to early detection of complications of an implant-based augmentation mammaplasty using the BellaGel® SmoothFine: preliminary 3-year clinical experience. J Surg Open Access. 2021;7:1–12. [Google Scholar]
  • [69].Doloff JC, Veiseh O, de Mezerville R, et al. The surface topography of silicone breast implants mediates the foreign body response in mice, rabbits, and humans. Nat Biomed Eng. 2021;5:1115–30. [DOI] [PubMed] [Google Scholar]
  • [70].Moio M, Nele G. A 5-year multicenter retrospective study on Motiva Ergonomix® breast implants: minimizing complications and maximizing results. Eur J Plast Surg. 2023;46:1099–111. [Google Scholar]
  • [71].Randquist C, Jaeger M, Stavrou D. Six-year evaluation of Motiva Round and Ergonomix SmoothSilk Surface silicone breast implants: a two-center, two-surgeon outcome analysis of 1053 primary and secondary breast augmentations and augmentation mastopexy. Aesthet Surg J. 2023;43:295–307. [DOI] [PubMed] [Google Scholar]
  • [72].Lee S, Kim M, Kim M, Ryu W, Back S. Capsular contracture rate in augmentation mammoplasty with Motiva breast implant insertion: a single-center experience in Korea. Aesthet Surg J. 2023;43:1248–55. [DOI] [PubMed] [Google Scholar]
  • [73].Glicksman C, Wolfe A, McGuire P. The study of the safety and effectiveness of Motiva SmoothSilk silicone gel-filled breast implants in patients undergoing primary and revisional breast augmentation: three-year clinical data. Aesthet Surg J. 2024;44:1273–85. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [74].Rosen MA, Kishawy HA. Sustainable manufacturing and design: concepts, practices, and needs. Sustainability. 2012;4:154–74. [Google Scholar]
  • [75].Arandia N, Garate JI, Mabe J. Medical devices with embedded sensor systems: design and development methodology for start-ups. Sensors (Basel). 2023;23:2578. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [76].Ghadimi P, Heavey C. Sustainable supplier selection in the medical device industry: toward sustainable manufacturing. Procedia CIRP. 2014;15:165–70. [Google Scholar]

Articles from Medicine are provided here courtesy of Wolters Kluwer Health