. 2025 Sep 10;60(10):e5178. doi: 10.1002/jms.5178

A Sensitive and Comprehensive LC‐MS/MS Method for the Analysis of Hallucinogens, Synthetic Cathinones, and Synthetic Cannabinoids in Urine Samples

Bruno Pereira dos Santos 1,, Letícia Birk 1, Vítor Camargo Pôrto 1, Sabrina Nunes Nascimento 2, Viviane Cristina Sebben 2, Sarah Eller 1, Tiago Franco de Oliveira 1
PMCID: PMC12423360  PMID: 40931322

ABSTRACT

The laboratory analysis of new psychoactive substances and related drugs is crucial for accurate clinical and forensic diagnosis of poisonings. Given this, a new LC‐MS/MS method for analyzing hallucinogens, synthetic cathinones, and synthetic cannabinoids in urine was developed. Urine samples were extracted using a liquid–liquid extraction protocol optimized via a multivariate experimental design. An aliquot of 400 μL of urine was enzymatically hydrolyzed with β‐glucuronidase and extracted with 700 μL of ethyl acetate. The resulting extracts were analyzed using LC‐MS/MS, with a total chromatographic run time of 8 min. The method was validated according to the ANSI/ASB Standard 036 guideline and was applied to 24 samples from suspected poisoning cases. The lower limits of quantification ranged from 0.1 to 1 ng/mL. Within‐run and between‐run precision (CV) were < 16%, and bias ranged from −12.8% to 19.8%. Nine of the 20 analytes investigated showed significant ionization suppression or enhancement (> 25%). Only two analytes (2C‐E and 2‐oxo‐3‐OH‐LSD) had a recovery rate lower than 70%. Among the 24 analyzed urine samples, one tested positive for 25B‐NBOH, one for LSD, and two for 2‐oxo‐3‐OH‐LSD. The developed method enables the simultaneous quantification of 20 illicit drugs, serving as an efficient diagnostic tool for clinical and forensic laboratories.

Keywords: drug analysis, LC‐MS/MS, new psychoactive substances, urine

1. Introduction

New psychoactive substances (NPS) represent a growing public health concern due to their potentially higher toxicity compared to classic drugs of abuse, their dynamic distribution market and easy accessibility, and their frequent absence from routine drug screening methods [1, 2]. As of June 2025, a total of 1343 NPS have been reported globally [3]. The traditional definition of NPS, as established by the United Nations Office on Drugs and Crime (UNODC), includes all substances of abuse not listed in the 1961 Single Convention on Narcotic Drugs or the 1971 Convention on Psychotropic Substances [1, 3].

Among the main groups of NPS are synthetic cathinones, synthetic phenethylamines, synthetic cannabinoid receptor agonists (SCRAs), and synthetic opioids [2]. These compounds, often designed to mimic the effects of controlled substances, pose significant challenges in clinical and forensic toxicology, including nonspecific clinical presentations, reliance on highly sensitive and comprehensive analytical methods, and limited availability of evidence‐based treatment and management protocols for intoxicated patients [4, 5]. Common clinical manifestations of NPS use include cardiovascular disturbances, central nervous system and respiratory depression, hallucinations, delirium, psychosis, and anxiety [6]. The use of NPS, either alone or in combination with other NPS or additional substances, can lead to life‐threatening or fatal outcomes [7].

Analytical approaches for the detection of NPS have significantly improved in recent years. A diverse range of validated methods for the determination of NPS in whole blood [8, 9, 10], plasma [11, 12], urine [13, 14], and hair [15] has been described in the literature. Typically, NPS are present in biological matrices at low concentrations. In addition, the development of comprehensive analytical methods that enable the simultaneous detection of multiple substances or classes is essential, given the large number of NPS currently available to the population. To overcome these challenges, the use of highly sensitive and selective techniques, such as liquid chromatography–tandem mass spectrometry (LC‐MS/MS), is recommended [16]. The implementation of LC‐MS/MS‐based methods in clinical and forensic laboratories is critical for strengthening NPS surveillance, improving patient management, and supporting forensic investigations [5].

Urine analysis allows a broader detection window, increasing the likelihood of obtaining a positive result. Urine is also considered a cleaner biological matrix, with fewer interferents, ample sample volumes (~20–30 mL), higher analyte concentrations, and frequent availability in clinical and forensic cases [17, 18]. An important aspect to consider in urine analysis is biotransformation, as it may be necessary to target the biotransformation products rather than unchanged substances. Furthermore, a hydrolysis step may be required for drugs that are conjugated with polar groups [17, 19, 20, 21].

Given this, this study aimed to develop and validate a quantitative LC‐MS/MS method for the simultaneous analysis of 20 drugs of abuse, including hallucinogens (2C‐E, 25B‐NBOH, 25B‐NBOMe, 25C‐NBOMe, 25E‐NBOH, 25I‐NBOH, 25I‐NBOMe, 2‐oxo‐3‐OH‐LSD, and LSD), synthetic cathinones (4‐chlor‐dimethylcathinone, dipentylone, ethylone, eutylone, mephedrone, methylone, N‐butylpentylone, N‐ethylheptedrone, and pentylone), and synthetic cannabinoids (ADB‐BUTINACA and ADB‐FUBINACA) in urine samples from clinical poisoning cases.

2. Materials and Methods

2.1. Chemicals and Reagents

LC‐MS‐grade acetonitrile, ethyl acetate, hexane, methyl tert‐butyl ether (MTBE), and 99% formic acid were acquired from Merck (Darmstadt, Germany). Standard solutions of 25B‐NBOMe, 25C‐NBOMe, 25I‐NBOH, 25I‐NBOMe, ethylone, lysergic acid diethylamide (LSD), mephedrone, methylone, and pentylone (1 mg/mL) were purchased from Cerilliant (Round Rock, TX, USA). 25B‐NBOH, 2C‐E, 25E‐NBOH, 4‐chloro‐N,N‐dimethylcathinone (4‐CDC), ADB‐BUTINACA, ADB‐FUBINACA, dipentylone, N‐butylpentylone, and N‐ethylpentedrone (1–5 mg total) were purchased from Cayman Chemical (Ann Arbor, MI, USA). Standard solutions of 2‐oxo‐3‐OH‐LSD (100 μg/mL) and eutylone (1 mg/mL) were acquired from LGC Standards (Manchester, NH, USA). The internal standard (IS) solutions MDEA‐d 6 and PCP‐d 5 (1 mg/mL) were also purchased from LGC Standards (Manchester, NH, USA). Ammonium acetate, sodium hydroxide, and β‐glucuronidase (type B‐1, from bovine liver) were purchased from Sigma‐Aldrich (St. Louis, MO). Ultrapure water was supplied by a Milli‐Q system (Millipore, Billerica, MA, USA).

2.2. Solutions

A working solution containing all analytes was prepared in acetonitrile at a concentration of 250 ng/mL. When not in use, the solutions were stored at −20°C. Blank urine samples were spiked with the working solution for the optimization and validation procedures. Drug‐free samples were collected from the researchers and previously analyzed to confirm the absence of target analytes. Both blank and spiked urine samples were stored in a freezer (−20°C) when not in use. The IS solution, containing MDEA‐d 6 and PCP‐d 5, was prepared in acetonitrile at a concentration of 400 ng/mL.

2.3. LC‐MS/MS Settings

The analyses were performed using a liquid chromatographic system coupled to a triple quadrupole mass spectrometer (LC‐MS/MS), model LCMS‐8045 (Shimadzu, Kyoto, Japan). The chromatographic separation was achieved using a Raptor Biphenyl column (50 × 3.0 mm, 2.7 μm particle size) from Restek (Bellefonte, PA, USA). Ultrapure water (A) and acetonitrile (B), both acidified with 0.1% (v/v) formic acid, were used as mobile phases, with a flow rate of 0.5 mL/min. The column oven was set at 50°C. The elution gradient used is described as follows: 0.0–0.25 min, 5%–5% of B; 0.25–4.8 min, 15%–50% of B%; 4.8–5.5 min, 50%–100% of B; 5.5–6.5 min, 100% of B; 6.5–6.6 min, 100%–5% of B; 6.6–8.0 min, 5% of B. The total chromatographic run time was 8 min.

Mass spectral data were acquired using electrospray ionization (ESI) in positive mode. The mass spectrometer settings used were as follows: heat block temperature, 400°C; capillary voltage, 1.5 kV; nebulizer gas (N2) flow, 3.0 L/min; drying gas (N2) flow, 10 L/min; desolvation line temperature, 250°C; and collision‐induced dissociation (CID) gas pressure (Ar), 230 kPa. Analyses were performed in multiple reaction monitoring (MRM) mode. For each analyte, three MRM transitions were monitored: one for quantification and two for confirmation. MRM transitions, collision energies, and retention time for each substance are presented in Table 1. Figure 1 shows a representative chromatogram of all analytes extracted from a spiked urine sample at 12 ng/mL concentration. All data were processed using the LabSolutions software (Shimadzu, Kyoto, Japan).

TABLE 1.

Analytes and their respective mass spectrometry conditions, retention time, and internal standard (IS).

Analyte MRM transitions (m/z) Dwell time (s) Q1 pre bias (V) ce (eV) Q3 pre bias (V) Retention time (min) IS
2C‐E 210.0 → 193.1 1.0 10 14 21 2.81 PCP‐d 5
210.0 → 178.1 1.0 14 20 17
210.0 → 163.0 1.0 10 29 30
25B‐NBOH 366.1 → 243.0 1.0 13 24 25 3.95 PCP‐d 5
366.1 → 107.0 1.0 13 28 18
366.1 → 260.1 1.0 13 12 17
25B‐NBOMe 380.0 → 121.1 1.0 13 22 22 4.59 PCP‐d 5
380.0 → 91.1 1.0 11 45 16
380.0 → 93.1 1.0 13 32 16
25C‐NBOMe 336.1 → 121.1 1.0 12 21 23 4.42 PCP‐d 5
336.1 → 91.1 1.0 12 48 15
336.1 → 93.1 1.0 12 40 16
25E‐NBOH 316.0 → 193.1 1.0 21 21 18 4.25 PCP‐d 5
316.0 → 107.1 1.0 11 31 18
316.0 → 178.1 1.0 21 31 17
25I‐NBOH 414.0 → 290.9 1.0 15 24 30 4.25 PCP‐d 5
414.0 → 107.1 1.0 15 31 18
414.0 → 308.0 1.0 15 14 20
25I‐NBOMe 428.0 → 121.1 1.0 15 24 21 4.90 PCP‐d 5
428.0 → 91.1 1.0 15 53 15
428.0 → 93.1 1.0 15 46 17
2‐Oxo‐3‐OH‐LSD 356.0 → 237.1 1.0 24 25 23 2.11 MDEA‐d 5
356.0 → 222.0 1.0 24 34 22
356.0 → 265.1 1.0 24 19 28
4‐CDC 212.0 → 139.0 1.0 19 20 21 2.28 MDEA‐d 5
212.0 → 167.0 1.0 19 15 25
212.0 → 103.0 1.0 18 25 14
ADB‐BUTINACA 331.2 → 201.1 1.0 12 27 20 5.67 PCP‐d 5
331.2 → 314.2 1.0 12 10 21
331.2 → 286.1 1.0 12 16 18
ADB‐FUBINACA 383.1 → 338.2 1.0 22 16 21 5.89 PCP‐d 5
383.1 → 366.2 1.0 13 10 24
383.1 → 253.1 1.0 14 26 22
Dipentylone 250.0 → 100.1 1.0 17 23 18 2.62 MDEA‐d 5
250.0 → 205.0 1.0 16 15 21
250.0 → 175.1 1.0 17 20 17
Ethylone 222.0 → 174.1 1.0 20 20 20 2.02 MDEA‐d 5
222.0 → 204.1 1.0 16 14 20
222.0 → 146.1 1.0 16 29 14
Eutylone 236.0 → 188.1 1.0 20 20 20 2.30 MDEA‐d 5
236.0 → 218.0 1.0 20 15 20
236.0 → 189.1 1.0 20 25 20
LSD 324.2 → 223.0 1.0 12 26 22 3.04 MDEA‐d 5
324.2 → 281.1 1.0 12 19 30
324.2 → 207.0 1.0 21 45 20
Mephedrone 178.0 → 160.1 1.0 20 15 20 2.10 MDEA‐d 5
178.0 → 145.1 1.0 20 16 20
178.0 → 144.1 1.0 18 15 20
Methylone 208.1 → 160.1 1.0 20 20 20 1.86 MDEA‐d 5
208.1 → 132.1 1.0 15 22 23
208.1 → 190.1 1.0 15 12 17
N‐Butylpentylone 278.1 → 260.2 1.0 20 15 20 3.46 MDEA‐d 5
278.1 → 230.1 1.0 20 20 20
278.1 → 188.0 1.0 20 20 20
N‐Ethylheptedrone 234.1 → 216.1 1.0 20 15 20 3.48 MDEA‐d 5
234.1 → 91.1 1.0 20 30 20
234.1 → 146.1 1.0 20 20 20
Pentylone 236.1 → 188.1 1.0 20 20 20 2.49 MDEA‐d 5
236.1 → 218.1 1.0 20 14 20
236.1 → 131.1 1.0 18 40 22
MDEA‐d 5 213.1 → 163.1 1.0 20 15 20 2.18
213.1 → 105.1 1.0 20 30 20
213.1 → 135.1 1.0 19 23 20
PCP‐d 5 249.3 → 96.1 1.0 20 32 20 3.61
249.3 → 164.2 1.0 20 18 20
249.3 → 86.1 1.0 20 22 20

Abbreviations: ce, collision energy; IS, internal standard; MRM, multiple reaction monitoring.

FIGURE 1.

FIGURE 1

Chromatogram of all analytes included in this study, at medium QC level (12 ng/mL).

2.4. Sample Procedure

An aliquot of 400 μL of urine and 20 μL of the IS solution (400 ng/mL) were added to a polypropylene microtube. The sample was hydrolyzed by the addition of 10 μL of β‐glucuronidase (400 U/μL) and 50 μL of acetate buffer (0.2 mol/L, pH 5.2), and the mixture was heated at 55°C for 60 min. After hydrolysis, the sample was cooled to room temperature. Then, 100 μL of NaOH solution (0.1 mol/L) and 700 μL of ethyl acetate were added. The mixture was vortexed for 10 s and centrifuged for 6 min at 9000 rpm. Subsequently, 650 μL of the organic phase was transferred to a new tube and dried under nitrogen flow at 50°C. The dried extract was reconstituted with 30 μL of ACN:water (3:1, v/v) with 0.1% (v/v) formic acid. Finally, 1 μL was injected into the LC‐MS/MS.

2.5. Optimization of Sample Extraction

Sample preparation is an essential step for obtaining bioanalytical methods with reduced matrix interferents and improved sensitivity. To achieve this, proper optimization must be carried out to ensure that the best conditions for the parameters involved in the extraction process [22]. In this study, a multivariate experimental approach was employed to improve the responses of variables that can influence the extraction efficiency. Initially, a fractional factorial design with five factors (25–1) was conducted to evaluate the following variables: solvent volume, sample volume, NaOH concentration, use of salting‐out, and extraction time. For each variable, two values (low and high) were defined, totaling sixteen runs. The complete experiment matrix is presented in Table S1. Subsequently, variables considered significant, when applicable, were further evaluated. A Doehlert design was employed to assess the optimal combination of solvent and sample volumes. Sample volume was optimized at five levels (100–400 μL), whereas solvent volume was evaluated in three levels (200–800 μL). The complete Doehlert experimental matrix is provided in Table S2. Next, the composition of the extractor solvent was investigated using three different solvents: ethyl acetate, hexane, and MTBE. Extraction was performed with both pure solvents and their mixtures, with ratios detailed in Table S3. All optimization experiments were conducted using drug‐free urine samples spiked with analytes at a concentration of 15 ng/mL. In both the fractional factorial and Doehlert designs, MTBE was fixed as the extraction solvent. The geometric mean of the absolute chromatographic area of all analytes was used to assess the response of the experiments. Optimization data were analyzed using Statistica 10.0 software (Statsoft, Tulsa, OK, USA) and statistically evaluated using the analysis of variance (ANOVA).

2.6. Bioanalytical Validation

Bioanalytical validation was performed in accordance with the Standard Practices for Method Validation in Forensic Toxicology (ANSI/ASB Standard 036 guideline), from the American Academy of Forensic Science (AAFS), as well as other international recommendations [23, 24]. The evaluated parameters included lower limit of quantification (LLOQ), interference studies, linearity, precision, bias, matrix effects, extraction recovery, dilution integrity, and processed sample stability.

The LLOQ was determined by analyzing fortified urine samples at decreasing concentrations (1.0, 0.5, and 0.1 ng/mL). Triplicate samples from three drug‐free individuals were extracted and analyzed, totaling nine evaluations. The LLOQ was defined as the lowest concentration level that presented a coefficient of variation (CV%) below 20%.

Interference studies were conducted to assess potential false signals from different urine matrices and substances that may be present in the sample. First, endogenous interference was evaluated by analyzing 10 negative urine samples from different individuals. Second, exogenous interference was assessed by adding compounds commonly found in urine samples, including cocaine and its metabolites, amphetamine‐type stimulants, cannabinoids, benzodiazepines, opioids, antidepressants, antipsychotics, and antiepileptics. Lastly, potential interference from stable isotope–labeled IS was investigated by analyzing a negative sample containing the IS solution and a fortified sample without IS addition.

To assess the linearity, calibration curves were constructed with six calibrators, ranging from the LLOQ to 25 ng/mL. The calibrators were prepared by spiking drug‐free urine samples with the working solution. All curves were required to present coefficients of determination (R 2) greater than 0.99 to be considered linear. Furthermore, Fisher's F‐test was used to assess the presence of heteroscedasticity. If detected, a weighted least squares model was applied using weight factors as 1/x and 1/x2. The optimal weighting scheme was selected based on the model yielding the lowest sum of residuals. After analysis of the upper limit of quantification (25 ng/mL), carryover was evaluated by injecting three blank samples. The method was considered free of carryover if the analyte signal in the blank samples was lower than the signal at the LLOQ.

To assess the precision and bias of the method, fortified urine samples were evaluated at three quality control (QC) concentrations: low (0.3 or 1.5 ng/mL), medium (12 ng/mL), and high (22 ng/mL). Each QC level was analyzed in triplicate within a single analytical run. Five runs were performed for each QC level, totaling 15 analyses per level. Precision was assessed by calculating within‐run and between‐run coefficients of variation (CV%) using ANOVA. According to the ANSI/ASB guideline, acceptable CV% values should be equal to or less than 20%. The same experiments were used to evaluate method bias, which was calculated as the difference between the measured and theoretical concentrations, with an acceptable range of ±20%. Additionally, dilution integrity was evaluated using a 1:20 (v/v) dilution factor to verify precision and bias when sample dilution is required due to analyte concentrations exceeding 25 ng/mL. For this purpose, fortified samples at 300 ng/mL were diluted with ultrapure water and analyzed. Acceptance criteria for this test were CV% ≤ 20% and bias within ±20%.

Matrix effect (ME) and extraction recovery were evaluated according to the procedure described by Matuszewski et al. [24]. For this purpose, three sets of six samples from different individuals were analyzed: Set A, neat solution; Set B, urine samples spiked before extraction; and Set C, urine samples spiked after extraction. The ME% was calculated by dividing Set C by that of Set A and multiplying by 100. The CV% of ME% across the six individual samples was required to be below 20%. Extraction recovery was calculated by dividing Set B by Set C and multiplying by 100. Both ME and extraction recovery were evaluated at low and high QC concentrations.

Processed sample stability was evaluated by analyzing six samples at time zero and again after 24 h of storage in the LC‐MS/MS autosampler maintained at 15°C. This parameter was assessed at all QC levels, and results were expressed as the percentage of the concentration at 24 h relative to the concentration at time zero.

2.7. Application to Urine Samples From Poisoning Cases

As proof of the method's applicability, 24 urine samples from poisoning cases attended by the Toxicological Information Center of Rio Grande do Sul (CIT‐RS), southern Brazil, were analyzed using the validated method. This study was approved by the Ethics Committee for Human Research at the Federal University of Health Sciences of Porto Alegre (UFCSPA), Brazil, and by the Ethics Committee for Human Research of the School of Public Health, State Health Department of Rio Grande do Sul, Brazil (CAAE 31992920.3.0000.5345 and CAAE 31992920.3.3001.5312).

3. Results

3.1. Extraction Optimization

In the fractional factorial design (25–1), three variables (sample volume, solvent volume, and NaOH concentration) had a significant effect (p < 0.05) on the analytical response. Extraction time and the addition of NaCl were not significant, with p values of 0.6355 and 0.7171, respectively. These results can be visualized in the Pareto chart of effects (see Figure 2). All variables that crossed the red significance threshold line (p = 0.05) were considered statistically significant.

FIGURE 2.

FIGURE 2

Pareto chart of effects from the variables evaluated in the fractional factorial design (25–1). All variables that crossed the red line (p = 0.05) were considered statistically significant, that is, these variables are able to influence the analytical response.

The subsequent optimization focused exclusively on sample volume and solvent volume, as low NaOH concentrations had already been evaluated in the fractional factorial design. The standardized effect value for NaOH concentration indicates a trend toward better responses at the lowest tested level, corresponding to 0.1 mol/L NaOH. The Doehlert design demonstrated a more sensitive analytical response when using the highest volumes of both sample and extraction solvent. This result was expected, because the larger the sample volume, the greater the amount of analyte available to be extracted. The surface response plot is shown in Figure 3, where the red region represents the highest response (geometric mean of absolute areas). The R 2 value of the model was 0.9909. Finally, the simplex‐centroid optimization indicates neat ethyl acetate as the most suitable extractor solvent, as shown in Figure 4. The obtained R 2 for the design was 0.7628.

FIGURE 3.

FIGURE 3

Surface response graph from Doehlert design for the evaluation of sample and extractor solvent volumes.

FIGURE 4.

FIGURE 4

Surface response from simplex‐centroid design for the optimization of the type of extractor solvent.

3.2. Bioanalytical Validation

The complete validation data is presented in Table 2. The LLOQ ranged from 0.1 to 1.0 ng/mL, with CV% values below 20% between the nine replicates. All analytes showed acceptable linearity, with R 2 values greater than 0.9918. Except for 2‐oxo‐3‐OH‐LSD, the calibration curves exhibited heteroscedasticity, requiring the application of weight factors. The weighting schemes applied were 1/x and 1/x2, both commonly used in analytical toxicology. No carryover effect was observed following the injection of the upper limit of quantification. During the interference assessment, no peaks were observed at the retention times of the analytes or IS, either in the analyses of urine samples fortified with exogenous compounds or in the analysis of drug‐free urine samples from different sources, confirming the selectivity of the method. Additionally, no interference from the stable isotope–labeled ISs was identified.

TABLE 2.

Results for LLOQ, linearity, precision, bias, matrix effect, and extraction recovery parameters.

Analyte LLOQ (ng/mL) Linearity Low QC Medium QC High QC Matrix effect (%) Extraction recovery (%)
R 2 Weighting factor Within‐run precision (CV%) Between‐run precision (CV%) Bias (%) Within‐run precision (CV%) Between‐run precision (CV%) Bias (%) Within‐run precision (CV%) Between‐run precision (CV%) Bias (%)
Hallucinogens
2C‐E 0.5 0.9999 1/x 11.0 5.7 −12.8 7.0 4.3 −8.5 5.6 3.9 −11.1 147.7 60.2
25B‐NBOH 0.5 0.9999 1/x2 7.9 4.6 2.1 5.8 4.5 1.9 6.0 3.5 −2.4 117.8 86.0
25B‐NBOMe 0.1 0.9939 1/x2 15.8 8.7 11.4 5.7 4.1 10.4 7.7 4.6 12.9 153.0 80.8
25C‐NBOMe 0.1 0.9935 1/x2 7.2 5.8 7.2 13.1 7.3 18.5 5.4 3.0 3.1 172.6 83.0
25E‐NBOH 0.1 0.9996 1/x2 10.6 6.3 −6.7 5.6 3.9 7.2 5.1 2.5 2.1 178.5 89.6
25I‐NBOH 0.5 0.9981 1/x 12.0 6.8 −7.9 5.0 3.5 12.5 9.4 4.6 18.3 157.2 94.0
25I‐NBOMe 0.1 0.9999 1/x2 9.7 5.1 4.7 9.7 5.0 19.8 8.4 4.1 16.7 164.9 83.8
2‐Oxo‐3‐OH‐LSD 0.5 0.9995 1 7.6 4.2 −1.2 7.4 4.7 −6.7 9.3 5.3 −6.0 104.5 38.9
LSD 0.5 0.9918 1/x 11.6 7.0 10.1 10.9 5.6 13.5 6.1 4.2 0.5 277.7 75.9
Synthetic cannabinoids
ADB‐BUTINACA 0.5 0.9977 1/x2 4.0 2.6 0.7 5.9 4.2 4.8 9.7 4.8 3.5 114.7 80.5
ADB‐FUBINACA 0.5 0.9962 1/x 10.6 6.1 7.6 4.4 2.6 14.5 8.2 4.5 9.2 106.8 81.5
Synthetic cathinones
4‐CDC 0.5 0.9930 1/x 12.5 6.7 0.1 9.2 7.0 6.7 8.0 5.4 2.4 53.7 94.7
Dipentylone 0.5 0.9999 1/x2 15.3 7.7 −9.1 8.4 5.4 0.9 6.9 3.7 −2.2 90.7 91.6
Ethylone 0.5 0.9995 1/x2 6.1 4.1 3.8 7.9 4.0 6.6 4.2 3.0 2.2 72.8 75.6
Eutylone 0.5 0.9982 1/x2 5.5 4.1 −3.0 7.1 4.3 −0.7 4.8 2.8 −9.8 96.3 92.7
Mephedrone 1 0.9998 1/x2 14.1 7.9 −10.4 9.4 4.9 6.2 4.2 2.5 3.9 116.5 76.5
Methylone 0.5 0.9998 1/x 7.2 4.4 −8.3 4.7 3.7 0.7 5.3 2.8 1.7 89.4 79.9
N‐Butylpentylone 1 0.9962 1/x2 13.9 9.5 3.3 8.5 4.2 13.8 7.0 4.0 12.8 124.9 91.8
N‐Ethylheptedrone 1 0.9989 1/x2 11.0 5.8 −0.7 8.5 4.5 3.9 4.9 3.8 6.8 87.9 89.5
Pentylone 0.5 0.9998 1/x2 8.1 5.1 −2.1 5.6 3.1 4.9 4.4 2.6 5.6 82.7 88.2

Abbreviations: CV, coefficient of variation; LLOQ, lower limit of quantification; QC, quality control; R 2, coefficient of determination.

The obtained within‐run and between‐run precision values were below 20% for all QC concentrations, in accordance with the ANSI/ASB guideline [23]. Within‐run precision CVs ranged from 4.0% to 15.8%, corresponding to the low QCs of ADB‐BUTINACA and 25B‐NBOMe, respectively. Between‐run precision CVs ranged from 2.5% (high QCs of 25E‐NBOH and mephedrone) to 9.5% (low QC of N‐butylpentylone). The observed biases ranged from −12.8% to 19.8%, associated with the low QC of 2C‐E and the medium QC of 25I‐NBOMe, respectively. Dilution integrity results showed within‐run and between‐run CV% and bias values within the ±20% acceptance range.

The ME varied considerably among analytes, ranging from 53.7% to 277.7%. Nine substances exhibited significant ME (defined as < 75% or > 125%). In the ME assessment of 4‐CDC and ethylone, ion suppression was observed, whereas ion enhancement was detected for 2C‐E, 25B‐NBOMe, 25C‐NBOMe, 25E‐NBOH, 25I‐NBOH, 25I‐NBOMe, and LSD. Despite ME values falling outside the 75%–125% range, all CV% values between replicates were below 20%, indicating the consistency and reproducibility of these effects. Extraction recovery rates were generally high, with only two analytes exhibiting values below 70%. The analytes with low or moderate recovery—2‐oxo‐3‐OH‐LSD (38.9%) and 2C‐E (60.2%)—are more polar compared to the other compounds.

Additionally, processed sample stability over 24 h ranged from 73.2% to 126.4%. Only three analytes exhibited notable instability (outside the ±25% range): the medium QC of 2‐oxo‐3‐OH‐LSD and the low QC levels of N‐butylpentylone and pentylone. Complete processed sample stability data are provided in Table S4.

3.3. Analysis of Samples From Poisoning Cases

Of the 24 urine samples analyzed from suspected poisoning cases, three tested positive for at least one target compound. Case 1 was positive for LSD (73.8 ng/mL) and its metabolite, 2‐oxo‐3‐OH‐LSD (> 300 ng/mL). Case 2 was positive for 2‐oxo‐3‐OH‐LSD alone (3.8 ng/mL). Finally, Case 3 was positive for 25B‐NBOH, with a concentration of 8.4 ng/mL. MRM chromatograms of the three positive cases are shown in Figure 5.

FIGURE 5.

FIGURE 5

MRM chromatogram of three urine samples with a positive result for at least one analyte.

4. Discussion

This study presented a sensitive and targeted analytical method for the detection of 20 drugs in urine, including NPS from different chemical classes. The NPS panel was defined based on the most relevant substances circulating in Brazilian territory in recent years. Historically, synthetic cathinones and synthetic phenethylamines have been the most frequently encountered NPS in Brazil, according to national drug seizure data [25]. Since the early 2020s, SCRAs have gained increasing prominence in the Brazilian context [26, 27]. Although LSD and its main urinary metabolite, 2‐oxo‐3‐OH‐LSD, are not classified as NPS, they were included in the method due to the common substitution of LSD with NBOMe and NBOH derivatives, particularly in blotter papers or powders [28, 29]. In the state of Rio de Janeiro, analysis of 3899 blotter papers seized between 2006 and 2019 identified 25I‐NBOH as the most prevalent substance (41.3%), followed by 25I‐NBOMe (16.6%) and LSD (14.2%) [30]. In another Brazilian state, Santa Catarina, blotter papers seized between 2011 and 2017 were analyzed by GC‐MS, and the presence of LSD in positive samples per year ranged from 0.12% to 100% [31].

A major challenge in the simultaneous analysis of diverse compounds such as NPS is their wide chemical variability. This complexity makes it difficult to establish a single sample preparation procedure capable of efficiently extracting structurally distinct analytes while maintaining high sensitivity and precision. The use of multivariate experimental design enabled the simultaneous assessment of multiple variables and the selection of optimal extraction conditions for all target compounds included in the study. Additionally, the structural similarity between certain substances, particularly positional isomers, complicates their detection and quantification. In this study, pentylone and eutylone shared the same MRM transitions. Therefore, chromatographic separation was essential for distinguishing them. The gradient employed successfully achieved baseline separation between the two synthetic cathinones, with retention times of 2.30 min for eutylone and 2.49 min for pentylone. Despite this level of resolution, the chromatographic method remains relatively fast, being shorter than other LC‐MS methods for NPS analysis reported in the literature [10, 13, 32, 33].

In urine analysis, hydrolysis is a fundamental step for the cleavage of the analyte with polar groups, such as glucuronic acid. Therefore, in this study, the samples were incubated with the β‐glucuronidase enzyme to obtain the analytes in their unconjugated form. Several studies indicate that some of the NPS included in this work can be excreted conjugated with glucuronic acid [34, 35, 36, 37, 38]. Pelletier et al. analyzed a urine sample from a poisoning case involving 25E‐NBOH, and the glucuronidated metabolite was the most intense, with a relative response to the unchanged 25E‐NBOH higher than 2000% [39]. Xiang et al. recommended the use of β‐glucuronidase for urine analysis of NBOH and NBOMe compounds to avoid false negative results [38]. Nonetheless, the absence of NPS metabolites in the current method should be acknowledged as an important limitation of this study. This is an important point to consider, as the concentration of metabolites in urine tends to be higher than that of the unchanged molecule, as can be seen in Case 1. The inclusion of metabolites helps avoid false negatives, as well as increasing the drug's detection window.

The developed method demonstrated high sensitivity, with LLOQs equal to or lower than 1 ng/mL. Some NPS, mainly SCRAs and phenethylamines, require enhanced sensitivity, as they are typically present in low concentrations in biological samples [16]. The obtained LLOQs are comparable to or inferior to the reported limits of quantification from other published methods [10, 13, 14, 32, 40, 41, 42, 43, 44].

The ME results indicated an intense ionization enhancement for some analytes, mainly phenethylamines. Previous studies have not reported altered ME values for these substances. Ferrari Júnior and Caldas described ME values for between 75.4% and 109.3% for LSD, NBOMe, and NBOH compounds [10]. Fan et al. reported normal ME values (86% to 119%) for 4‐CDC, ethylone, 25B‐NBOMe, 2C‐E, and 25I‐NBOMe [13, 41]. On the other hand, Fabris et al. verified an unusual ionization enhancement for synthetic cathinones when using a dispersive liquid–liquid microextraction protocol [45]. In this study, the CV% of ME replicates was below 20%, that is, the ME, although significant, is constant. This means that the interindividual variability of urine samples affects the ME homogeneously. In this way, it can be ensured that the blank urine employed in the validation studies modulates the analytical response similarly to the urine samples to be tested and quantified. This allows the developed method to be used in the reliable quantification of the listed analytes. Better chromatographic separation to avoid coelution with endogenous interferents or more extensive sample preparation techniques (e.g., solid‐phase extraction) could result in more suitable ME values. Additionally, the use of stable isotope–labeled IS of each analyte is recommended to minimize ME on quantification methods [24, 46, 47]. However, all these procedures demand a higher analysis time or associated cost.

In the processed sample stability assessment, instability greater than 125% was verified for three analytes (LSD, N‐butylpentylone, and pentylone) in some QC levels. In this sense, the analysis of extracts after 24 h may generate increased concentrations of these substances. Therefore, immediate analysis after the extraction process is recommended, at least when these analytes are suspected to be present.

Regarding the cases with positive results, Case 1 involved a male individual of unknown age who was found unconscious at an electronic music festival. Upon admission, he was in a coma, with malignant hyperthermia, hyporeflexia, mydriasis, tachycardia, and respiratory depression. A urine sample was collected and sent to the CIT‐RS laboratory. Hours later, the patient died due to cardiorespiratory arrest. When screened for classic drugs and medications [17], the urine sample tested positive for MDA (> 10 μg/mL), MDEA (561 ng/mL), MDMA (> 10 μg/mL), ketamine (534 ng/mL), lidocaine (1.7 μg/mL), and diclofenac (81 ng/mL). Typically, LSD and its metabolite 2‐oxo‐3‐OH‐LSD are detected at low concentrations (below 25 ng/mL) in urine samples. However, elevated concentrations may occur following high‐dose LSD use [48, 49, 50]. Moreover, markedly high urinary concentrations of LSD and 2‐oxo‐3‐OH‐LSD can be observed in cases of severe poisoning or fatal outcomes. Favretto et al. reported postmortem urine concentrations of LSD and 2‐oxo‐3‐OH‐LSD of 91.0 and 430.2 ng/mL, respectively [51], which are comparable to those found in Case 1. In this case, the concomitant presence of high urinary levels of MDA and MDMA may have exerted a cumulative effect contributing to death alongside LSD.

Case 2 involved a 35‐year‐old woman, reportedly a drug user in withdrawal, though a psychiatric report raised suspicions of ongoing substance use. In this case, only the main LSD metabolite, 2‐oxo‐3‐OH‐LSD, was detected. Additional urine screening revealed the presence of MDA, MDMA, 11‐nor‐Δ9‐tetrahydrocannabinol carboxylic acid, 7‐aminoclonazepam, citalopram, oxazepam, and temazepam. A positive result exclusively for 2‐oxo‐3‐OH‐LSD without LSD detection can be explained by the longer detection window of this metabolite compared to the parent compound [48].

Case 3 involved a 17‐year‐old male adolescent admitted to the hospital presenting with intense agitation. The patient reported using “N‐bomb,” a common street name for substances containing NBOMe derivatives. LC‐MS/MS analysis for drugs and medications [17] also detected lidocaine, haloperidol, and promethazine, which were administered during hospital care. No other drugs of abuse were identified. Although the patient referred to NBOMe use, NBOH compounds currently represent the majority of circulating phenethylamines in Brazil [25, 52]. Ivory et al. described a case series involving 25B‐NBOH exposure, including a patient with severe agitation, as observed in Case 3. That report also mentioned acute kidney injury and rhabdomyolysis [28]. In Case 3, no renal damage was identified; however, elevated levels of total and indirect bilirubin and creatine phosphokinase (CK) were observed, suggesting hemolysis and muscle injury, likely related to the intense agitation presented by the patient.

5. Conclusion

In this work, a method for the analysis of 20 drugs of abuse, including hallucinogens, synthetic cathinones, and synthetic cannabinoids, was successfully developed and validated. The established protocol is simple, efficient, and suitable for detecting NPS at low concentrations. Validation followed the ANSI/ASB Standard 036 guideline, with all assessed mandatory parameters meeting the established acceptance criteria. As proof of applicability, 24 urine samples from suspected poisoning cases were analyzed, with three yielding positive results for at least one of the target substances. The proposed method demonstrates strong potential as a diagnostic tool for NPS‐related intoxications and can be readily implemented in clinical and forensic toxicology laboratories.

Conflicts of Interest

The authors declare no conflicts of interest.

Supporting information

Table S1: Complete matrix from fractional factorial design (25–1) for screening of significant variables.

Table S2: Complete matrix from Doehlert design for solvent and sample volume optimization.

Table S3: Complete matrix from simplex‐centroid design for extractor solvent optimization.

Table S4: Processed sample stability results for each QC level.

JMS-60-e5178-s001.docx (23.2KB, docx)

Acknowledgments

The authors would like to acknowledge the financial support from Coordination for the Improvement of Higher Education Personnel–Brazil (CAPES—Finance Code 001). The Article Processing Charge for the publication of this research was funded by the Coordenação de Aperfeiçoamento de Pessoal de Nível Superior ‐ Brasil (CAPES) (ROR identifier: 00x0ma614).

dos Santos B., Birk L., Pôrto V., et al., “A Sensitive and Comprehensive LC‐MS/MS Method for the Analysis of Hallucinogens, Synthetic Cathinones, and Synthetic Cannabinoids in Urine Samples,” Journal of Mass Spectrometry 60, no. 10 (2025): e5178, 10.1002/jms.5178.

Funding: This work was supported by the Coordination for the Improvement of Higher Education Personnel.

Contributor Information

Bruno Pereira dos Santos, Email: brunods@ufcspa.edu.br.

Tiago Franco de Oliveira, Email: oliveira@ufcspa.edu.br.

Data Availability Statement

The data that support the findings of this study are available from the corresponding author upon reasonable request.

References

  • 1. Shafi A., Berry A. J., Sumnall H., Wood D. M., and Tracy D. K., “New Psychoactive Substances: A Review and Updates,” Therapeutic Advances in Psychopharmacology 10 (2020): 2045125320967197. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Simão A. Y., Antunes M., Cabral E., et al., “An Update on the Implications of New Psychoactive Substances in Public Health,” International Journal of Environmental Research and Public Health 19 (2022): 4869. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. UNODC , “NPS Data Visualisations,” (2025), https://www.unodc.org/LSS/Page/NPS/DataVisualisations.
  • 4. Santos I. C., Maia D., Dinis‐Oliveira R. J., and Barbosa D. J., “New Psychoactive Substances: Health and Legal Challenges,” Psychoactives 3 (2024): 285–302. [Google Scholar]
  • 5. Miller J. J., Yazdanpanah M., Colantonio D. A., Beriault D. R., and Delaney S. R., “New Psychoactive Substances: A Canadian Perspective on Emerging Trends and Challenges for the Clinical Laboratory,” Clinical Biochemistry 133–134 (2024): 110810. [DOI] [PubMed] [Google Scholar]
  • 6. Peacock A., Bruno R., Gisev N., et al., “New Psychoactive Substances: Challenges for Drug Surveillance, Control, and Public Health Responses,” Lancet 394 (2019): 1668–1684. [DOI] [PubMed] [Google Scholar]
  • 7. Kraemer M., Boehmer A., Madea B., and Maas A., “Death Cases Involving Certain New Psychoactive Substances: A Review of the Literature,” Forensic Science International 298 (2019): 186–267. [DOI] [PubMed] [Google Scholar]
  • 8. Ayala J. and Kerrigan S., “Comprehensive Toxicological Screening of Common Drugs of Abuse, New Psychoactive Substances and Cannabinoids in Blood Using Supported Liquid Extraction and Liquid Chromatography–Quadrupole Time‐of‐Flight Mass Spectrometry,” Journal of Analytical Toxicology 47 (2023): 656–667. [DOI] [PubMed] [Google Scholar]
  • 9. Birk L., Dos Santos B. P., Eller S., and De Oliveira T. F., “A Novel Method for the Determination of Synthetic Cathinones and Related Substances in Postmortem Blood Samples Using Cork‐Based Dispersive Solid‐Phase Microextraction Prior to LC‐MS/MS Analysis,” Analytical and Bioanalytical Chemistry 417, no. 17 (2025): 3845–3855. [DOI] [PubMed] [Google Scholar]
  • 10. Ferrari Júnior E. and Caldas E. D., “Determination of New Psychoactive Substances and Other Drugs in Postmortem Blood and Urine by UHPLC–MS/MS: Method Validation and Analysis of Forensic Samples,” Forensic Toxicology 40 (2022): 88–101. [DOI] [PubMed] [Google Scholar]
  • 11. Hwang J., Han M., An S., Moon J. H., Shim G., and Chung H., “Screening of New Psychoactive Substances in Human Plasma by Magnetic Solid Phase Extraction and LC‐QTOF‐MS,” Forensic Science International 332 (2022): 111176. [DOI] [PubMed] [Google Scholar]
  • 12. Fabris A. L., Pedersen‐Bjergaard S., Øiestad E. L., et al., “Solvent‐Free Parallel Artificial Liquid Membrane Extraction for Drugs of Abuse in Plasma Samples Using LC‐MS/MS,” Analytica Chimica Acta 1301 (2024): 342387. [DOI] [PubMed] [Google Scholar]
  • 13. Fan S.‐Y., Zang C.‐Z., Shih P.‐H., et al., “Simultaneous LC‐MS/MS Screening for Multiple Phenethylamine‐Type Conventional Drugs and New Psychoactive Substances in Urine,” Forensic Science International 325 (2021): 110884. [DOI] [PubMed] [Google Scholar]
  • 14. Czarny J., Musiał J., Powierska‐Czarny J., et al., “Determination of 465 Psychoactive Substances, Drugs and Their Metabolites in Urine by LC‐MS/MS,” Analytical Methods 16 (2024): 5426–5432. [DOI] [PubMed] [Google Scholar]
  • 15. Barone R., Pelletti G., Giorgetti A., et al., “Validation and Application of a Method for the Quantification of 137 Drugs of Abuse and New Psychoactive Substances in Hair,” Journal of Pharmaceutical and Biomedical Analysis 243 (2024): 116054. [DOI] [PubMed] [Google Scholar]
  • 16. Krotulski A. J., Mata D. C., Smith C. R., et al., “Advances in Analytical Methodologies for Detecting Novel Psychoactive Substances: A Review,” Journal of Analytical Toxicology 49 (2025): 152–169. [DOI] [PubMed] [Google Scholar]
  • 17. Dos Santos B. P., Birk L., Schwarz P., et al., “A Validated Dilute‐and‐Shoot LC–MS‐MS Urine Screening for the Analysis of 95 Illicit Drugs and Medicines: Insights From Clinical and Forensic Brazilian Cases,” Journal of Analytical Toxicology 48 (2024): 314–331. [DOI] [PubMed] [Google Scholar]
  • 18. Deventer K., Pozo O. J., Verstraete A. G., and Van Eenoo P., “Dilute‐and‐Shoot‐Liquid Chromatography‐Mass Spectrometry for Urine Analysis in Doping Control and Analytical Toxicology,” TrAC Trends in Analytical Chemistry 55 (2014): 1–13. [Google Scholar]
  • 19. Yang G., Ge S., Singh R., et al., “Glucuronidation: Driving Factors and Their Impact on Glucuronide Disposition,” Drug Metabolism Reviews 49 (2017): 105–138. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Diao X. and Huestis M. A., “New Synthetic Cannabinoids Metabolism and Strategies to Best Identify Optimal Marker Metabolites,” Frontiers in Chemistry 7 (2019): 109. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Szeremeta M., Pietrowska K., Niemcunowicz‐Janica A., Kretowski A., and Ciborowski M., “Applications of Metabolomics in Forensic Toxicology and Forensic Medicine,” International Journal of Molecular Sciences 22 (2021): 3010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Bigão V. L. C. P., Da Costa B. R. B., Da Silva J. J. M., De Martinis B. S., and Tapia‐Blácido D. R., “Use of Statistical Design of Experiments (DoE) in Forensic Analysis: A Tailored Review,” Forensic Chemistry 37 (2024): 100554. [Google Scholar]
  • 23. AAFS , “Standard Practices for Method Validation in Forensic Toxicology, ANSI/ASB Standard 036,” (2019).
  • 24. Matuszewski B. K., Constanzer M. L., and Chavez‐Eng C. M., “Strategies for the Assessment of Matrix Effect in Quantitative Bioanalytical Methods Based on HPLC−MS/MS,” Analytical Chemistry 75 (2003): 3019–3030. [DOI] [PubMed] [Google Scholar]
  • 25. Dos Santos B. P., Birk L., De Souza Schwarz P., Eller S., De Oliveira T. F., and Arbo M. D., “A Comprehensive Analysis of Legislative Strategies for New Psychoactive Substances: The Brazilian Panorama,” Psychoactives 2 (2023): 242–255. [Google Scholar]
  • 26. Rodrigues T. B., Souza M. P., De Melo Barbosa L., et al., “Synthetic Cannabinoid Receptor Agonists Profile in Infused Papers Seized in Brazilian Prisons,” Forensic Toxicology 40 (2022): 119–124. [DOI] [PubMed] [Google Scholar]
  • 27. Bombana H. S., De Paula Meirelles G., De Oliveira R. A., Leyton V., and Yonamine M., “Synthetic Cannabinoid Receptor Agonists in Oral Fluid: Development of a Dispersive Liquid–Liquid Microextraction Method With Liquid Chromatography–Mass Spectrometry Detection,” Journal of Analytical Toxicology (2025): bkaf027, 10.1093/jat/bkaf027. [DOI] [PubMed] [Google Scholar]
  • 28. Ivory S. T., Rotella J.‐A., Schumann J., and Greene S. L., “A Cluster of 25B‐NBOH Poisonings Following Exposure to Powder Sold as Lysergic Acid Diethylamide (LSD),” Clinical Toxicology 60 (2022): 966–969. [DOI] [PubMed] [Google Scholar]
  • 29. Birk L., De Oliveira S. E. F., Mafra G., et al., “A Low‐Voltage Paper Spray Ionization QTOF‐MS Method for the Qualitative Analysis of NPS in Street Drug Blotter Samples,” Forensic Toxicology 38 (2020): 227–231. [Google Scholar]
  • 30. Meira V. L., De Oliveira A. S., Cohen L. S. A., et al., “Chemical and Statistical Analyses of Blotter Paper Matrix Drugs Seized in the State of Rio de Janeiro,” Forensic Science International 318 (2021): 110588. [DOI] [PubMed] [Google Scholar]
  • 31. De Souza Boff B., Silveira Filho J., Nonemacher K., Driessen Schroeder S., Dutra Arbo M., and Rezin K. Z., “New Psychoactive Substances (NPS) Prevalence Over LSD in Blotter Seized in State of Santa Catarina, Brazil: A Six‐Year Retrospective Study,” Forensic Science International 306 (2020): 110002. [DOI] [PubMed] [Google Scholar]
  • 32. Glicksberg L., Bryand K., and Kerrigan S., “Identification and Quantification of Synthetic Cathinones in Blood and Urine Using Liquid Chromatography‐Quadrupole/Time of Flight (LC‐Q/TOF) Mass Spectrometry,” Journal of Chromatography B 1035 (2016): 91–103. [DOI] [PubMed] [Google Scholar]
  • 33. Ambach L., Hernández Redondo A., König S., Angerer V., Schürch S., and Weinmann W., “Detection and Quantification of 56 New Psychoactive Substances in Whole Blood and Urine by LC–MS/MS,” Bioanalysis 7 (2015): 1119–1136. [DOI] [PubMed] [Google Scholar]
  • 34. Kavanagh P., Pechnikov A., Nikolaev I., Dowling G., Kolosova M., and Grigoryev A., “Detection of ADB‐BUTINACA Metabolites in Human Urine, Blood, Kidney and Liver,” Journal of Analytical Toxicology 46 (2022): 641–650. [DOI] [PubMed] [Google Scholar]
  • 35. Godoi A. B., Antunes N. D. J., Rodrigues L. C., Martins A. F., and Costa J. L., “In Vitro Metabolism and Metabolite Identification of Eutylone Using Rat Liver Microsomes,” Journal of Pharmaceutical and Biomedical Analysis 260 (2025): 116827. [DOI] [PubMed] [Google Scholar]
  • 36. Ellefsen K. N., Concheiro M., and Huestis M. A., “Synthetic Cathinone Pharmacokinetics, Analytical Methods, and Toxicological Findings From Human Performance and Postmortem Cases,” Drug Metabolism Reviews 48 (2016): 237–265. [DOI] [PubMed] [Google Scholar]
  • 37. Caspar A. T., Helfer A. G., Michely J. A., et al., “Studies on the Metabolism and Toxicological Detection of the New Psychoactive Designer Drug 2‐(4‐Iodo‐2,5‐dimethoxyphenyl)‐N‐[(2‐methoxyphenyl)methyl]ethanamine (25I‐NBOMe) in Human and Rat Urine Using GC‐MS, LC‐MSn, and LC‐HR‐MS/MS,” Analytical and Bioanalytical Chemistry 407 (2015): 6697–6719. [DOI] [PubMed] [Google Scholar]
  • 38. Xiang J., Wen D., Zhai W., et al., “Metabolic Characterization of 25X‐NBOH and 25X‐NBOMe Phenethylamines Based on UHPLC‐Q‐Exactive Orbitrap MS in Human Liver Microsomes,” Journal of Pharmaceutical and Biomedical Analysis 242 (2024): 116020. [DOI] [PubMed] [Google Scholar]
  • 39. Pelletier R., Gicquel T., Carvelli J., et al., “Severe 25E‐NBOH Intoxication Associated With MDPHP Intake: A Case Report, Metabolism Study, and Literature Review,” International Journal of Legal Medicine 138 (2024): 815–822. [DOI] [PubMed] [Google Scholar]
  • 40. Nieddu M., Baralla E., Sodano F., and Boatto G., “Analysis of 2,5‐Dimethoxy‐Amphetamines and 2,5‐Dimethoxy‐Phenethylamines Aiming Their Determination in Biological Matrices: A Review,” Forensic Toxicology 41 (2023): 1–24. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41. Fan S.‐Y., Zang C.‐Z., Shih P.‐H., et al., “A LC‐MS/MS Method for Determination of 73 Synthetic Cathinones and Related Metabolites in Urine,” Forensic Science International 315 (2020): 110429. [DOI] [PubMed] [Google Scholar]
  • 42. Bouzoukas C., Nikolaou P., Athanaselis S., Dona A., Spiliopoulou C., and Papoutsis I., “Development, Validation and Applications of a GC/MS Method for the Simultaneous Determination of 9 Amphetamines and 12 NPS Analogues in Blood and Urine,” Forensic Science International 370 (2025): 112469. [DOI] [PubMed] [Google Scholar]
  • 43. Pérez‐Alcaraz A., Borrull F., Calull M., and Aguilar C., “Cathinones in Urine Samples: A Review of Recent Advances for Their Determination by Chromatographic and Related Techniques,” TrAC Trends in Analytical Chemistry 143 (2021): 116347. [Google Scholar]
  • 44. Nisbet L. A., Wylie F. M., Logan B. K., and Scott K. S., “Gas Chromatography‐Mass Spectrometry Method for the Quantitative Identification of 23 New Psychoactive Substances in Blood and Urine,” Journal of Analytical Toxicology 43 (2019): 346–352. [DOI] [PubMed] [Google Scholar]
  • 45. Fabris A. L., Lanaro R., Costa J. L., and Yonamine M., “Development of a Dispersive Liquid–Liquid Microextraction for Synthetic Cathinones in Biological Fluids Based on Principles of Green Analytical Toxicology,” Journal of Analytical Toxicology 47 (2023): 353–365. [DOI] [PubMed] [Google Scholar]
  • 46. Panuwet P., R. E. Hunter, Jr. , D'Souza P. E., et al., “Biological Matrix Effects in Quantitative Tandem Mass Spectrometry‐Based Analytical Methods: Advancing Biomonitoring,” Critical Reviews in Analytical Chemistry 46 (2016): 93–105. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47. Eeckhaut A. V., Lanckmans K., Sarre S., Smolders I., and Michotte Y., “Validation of Bioanalytical LC–MS/MS Assays: Evaluation of Matrix Effects,” Journal of Chromatography B 877 (2009): 2198–2207. [DOI] [PubMed] [Google Scholar]
  • 48. Jang M., Kim J., Han I., and Yang W., “Simultaneous Determination of LSD and 2‐Oxo‐3‐Hydroxy LSD in Hair and Urine by LC–MS/MS and Its Application to Forensic Cases,” Journal of Pharmaceutical and Biomedical Analysis 115 (2015): 138–143. [DOI] [PubMed] [Google Scholar]
  • 49. Dolder P. C., Liechti M. E., and Rentsch K. M., “Development and Validation of a Rapid Turboflow LC‐MS/MS Method for the Quantification of LSD and 2‐Oxo‐3‐Hydroxy LSD in Serum and Urine Samples of Emergency Toxicological Cases,” Analytical and Bioanalytical Chemistry 407 (2015): 1577–1584. [DOI] [PubMed] [Google Scholar]
  • 50. Martin R., Schürenkamp J., Gasse A., Pfeiffer H., and Köhler H., “Determination of Psilocin, Bufotenine, LSD and Its Metabolites in Serum, Plasma and Urine by SPE‐LC‐MS/MS,” International Journal of Legal Medicine 127 (2013): 593–601. [DOI] [PubMed] [Google Scholar]
  • 51. Favretto D., Frison G., Maietti S., and Ferrara S. D., “LC‐ESI‐MS/MS on an Ion Trap for the Determination of LSD, iso‐LSD, nor‐LSD and 2‐Oxo‐3‐Hydroxy‐LSD in Blood, Urine and Vitreous Humor,” International Journal of Legal Medicine 121 (2007): 259–265. [DOI] [PubMed] [Google Scholar]
  • 52. Machado Y., Coelho Neto J., Lordeiro R. A., Alves R. B., and Piccin E., “Identification of New NBOH Drugs in Seized Blotter Papers: 25B‐NBOH, 25C‐NBOH, and 25E‐NBOH,” Forensic Toxicology 38 (2020): 203–215. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Table S1: Complete matrix from fractional factorial design (25–1) for screening of significant variables.

Table S2: Complete matrix from Doehlert design for solvent and sample volume optimization.

Table S3: Complete matrix from simplex‐centroid design for extractor solvent optimization.

Table S4: Processed sample stability results for each QC level.