Skip to main content
Journal of Aerosol Medicine and Pulmonary Drug Delivery logoLink to Journal of Aerosol Medicine and Pulmonary Drug Delivery
. 2025 Aug 6;38(4):211–215. doi: 10.1089/jamp.2025.87987.sd

Spacers and Valved Holding Chambers

Sunalene Devadason 1,✉
PMCID: PMC12435187  PMID: 40772815

Abstract

Spacers, primarily valved holding chambers (VHCs), are widely used to overcome some of the problems associated with the use of pressurized metered-dose inhalers (pMDIs). These include the difficulty experienced by patients in trying to coordinate the initiation of inhalation with the actuation of the pMDI. High oropharyngeal deposition of drug, which may result in both local and systemic side effects, is also a problem. Although the variability in output from pMDIs under optimized conditions in the laboratory is low, the variability when used in clinical practice is likely to increase considerably. Hence, the dose introduced into a holding chamber may vary significantly depending on the way in which the pMDI canister is handled before it is actuated. Several studies have shown that various design factors can influence the dose delivered from a holding chamber. These include spacer volume, shape, valve design, using multiple actuations, delay between actuation and inhalation, and construction material, which affects the level of electrostatic charge accumulating on the spacer surfaces. Several spacers which are made from low or anti-static materials are now available. Recommendations for optimal use of spacers, including inhalation techniques are outlined in this chapter, and vary according to patient age and inhalation coordination capability. Efficiency of drug delivery and lung deposition are also dependent on pMDI drug formulation and the patient's anatomical and physiological characteristics.

Keywords: Inhalers and accessory devices, electrostatic charge, face mask versus mouthpiece, spacer volume, extension devices

Introduction

The use of standard pressurized metered-dose inhalers (pMDIs) alone has a number of disadvantages. Due to the high velocity at which drug particles are emitted from the pMDI, patients must consistently be able to closely coordinate the actuation of the pMDI with the start of inhalation to minimize oropharyngeal impaction and improve therapeutic efficiency. This can be difficult for many patients, particularly children1 and the elderly,2 and for parents3 or caregivers attempting to administer therapy via pMDI to patients. In those patients who are able to utilize the pMDI alone effectively, a significant amount of the actuated drug will still be impacted in the oropharynx instead of depositing in the lungs.4 Non-pulmonary impaction of drug may be reduced if the patient can be trained to take a slow, deep inhalation after actuation, but irrespective of technique, significant oropharyngeal impaction is still likely to occur.5

Spacers, both valved and non-valved holding chambers, were developed to address the problems of coordination and extrapulmonary drug deposition. A review of the history of spacer development makes interesting reading.6 Spacers are accessory devices placed between the pMDI and the patient's mouth. The pMDI is actuated into the spacer and the patient inhales the drug available as an aerosol cloud within the spacer. While spacers are less convenient and portable than the use of pMDIs alone, there are multiple advantages to spacer use: 1) reducing the need for coordinating actuation with inhalation, since the actuated drug is held within the chamber and is available for inhalation for a short time post-actuation; 2) the high velocity at which the drug is emitted from the pMDI actuator is reduced due to the increased distance/time between actuation and entry into the patient's mouth; and 3) the larger particles which have the greatest inertia at high velocity and which are most likely to impact in the mouth or throat, impact or sediment out in the spacer instead, while the” respirable” drug is still available to be inhaled and deposited in the lungs.

Non-valved spacers may be used by patients who can be trained to take a single slow, deep inhalation after actuation, but valved spacers (VHCs or valved holding chambers) would normally be recommended for inhalation with tidal breathing, so that the exhaled breath is vented away from the spacer chamber. However, homemade (unvalved) spacers made from plastic bottles or even styrofoam coffee cups have been tested. A spacer made from a 500 ml soft drink bottle has been shown to be efficacious and may be a low-cost alternative in developing countries.7,8 Small unvalved extension devices such as the Microspacer®, which increase the distance between the point of aerosol emission and the back of the patient's throat, have also been commercially developed. These devices must also be used with a single maximal inhalation rather than tidal breathing.

The aerosol dose available for inhalation (i.e. the amount of drug emitted from the spacer mouthpiece) is reduced when spacers are used. Arguably, the majority of the drug loss occurs in the particle size range that would normally deposit in the oropharynx, rather than in the lungs. The magnitude of the drug loss, and the proportion of it that would come from the” respirable” fraction that is most likely to deposit in the lungs, depends on several factors shown in Table 1.

Table 1.

Factors Which Can Impact the Efficiency of Drug Delivery from Spacers and Drug Deposition into the Patients' Lungs.

Factors impacting efficiency
Device characteristics  
 Spacer design e.g., volume, shape, valve type, construction material
 Drug (pMDI) formulation e.g., emission velocity, emitted particle size distribution, solution or suspension formulation
 Patient interface e.g., mouthpiece, face mask
 Drug (pMDI) formulation e.g., emission velocity, emitted particle size distribution, solution or suspension formulation
Use by patients or parents  
 Administration technique e.g., shaking pMDI prior to actuation, multiple actuations into chamber prior to inhalation
 Inhalation technique e.g., volume and velocity of inhalation, tidal breathing versus slow maximal inhalation
Physiological and cognitive factors e.g., age, lung volume, severity of airways disease

Despite these variable factors, the efficacy of pMDI-spacer use is indicated by the fact that this combination has replaced the use of nebulizers for bronchodilator treatment of acute, severe asthma exacerbations in emergency departments in many countries. Increased lung delivery of bronchodilator has been shown with pMDI-spacers compared with nebulizers,9 and equivalent or increased clinical efficacy has been found using pMDI-spacers compared with nebulizers in adults and children.10–13

When reviewing the published literature in this field, it is important to distinguish between drug delivery from the pMDI-spacer combination, which is determined by a combination of device characteristics and administration technique (Table 1), and drug deposition, which is determined by a combination of all the factors described in Table 1. Drug delivery is generally an in vitro assessment, measuring the amount of drug emitted from the mouthpiece of the chamber, which can be equated to the dose which would be delivered to the patient's mouth or the total body dose. Drug deposition is a much more complex assessment which measures the amount of drug depositing in the patient's lungs as well as quantifying extrapulmonary deposition.

Spacer Design Characteristics

It is difficult to study the individual spacer design factors in isolation, because each spacer comes as a package which incorporates a given volume, valve design and construction material. Studies with earlier spacer designs produced data indicating that larger spacer volumes tended to result in increased lung delivery.14 The Volumatic™, Nebuhaler® and Fisonair® are examples of large volume spacers (>700 ml in volume) available at that time—of these, only the Volumatic™ is still available commercially these days.

Research conducted on spacers in the early days indicated that accumulation of electrostatic charge on the inner surface of plastic spacers could markedly reduce the amount of drug available for inhalation.15,16 These effects have been minimized in the past by the use of an ionic detergent coating on the inner surface of the spacer17 or by the use of metallic spacers.18 “Priming” or coating the inner spacer surface with multiple pMDI actuations was also a method recommended to minimize electrostatic charge;19 however, the effectiveness of this method may be dependent on the pMDI formulation used.

Nowadays, major advances in valve design and low or anti-static construction materials over time have resulted in the development of a variety of more efficient small volume spacers that appear to offset the effect of spacer volume, such that modern small volume spacers produce equivalent or better drug delivery than the older large volume spacers.20 The AeroChamber Plus® Flow-Vu® (135 ml) and Optichamber Diamond (140 ml) are examples of modern anti-static small volume spacers. Patient preferences for smaller, more portable devices have meant that developmental advances in this area have focused on smaller volume spacers, and almost all large volume spacers are no longer commercially available.

pMDI Formulations

The effect of pMDI formulation on efficiency of drug delivery has been covered in other sections of the textbook. One fact that is worth highlighting in this section is the use of spacers with extrafine pMDI formulations. The reduced particle size emitted from the pMDI together with a slower emission velocity can result in increased lung deposition with reduced oropharyngeal impaction,4,21 making the use of a spacer potentially less critical. Additionally, the impact of miscoordination and errors in inhalation technique on lung deposition is potentially reduced.22

Face Mask Versus Mouthpiece

Spacers may be used with a mouthpiece or face mask. Face masks must be used for infants and young children, and the age at which advice is given to change to a mouthpiece is variable in different countries. Traditionally, it is assumed that drug delivery and lung deposition is much lower when using a face mask than with a mouthpiece, primarily due to studies such as the nebulizer deposition study published by Chua and colleagues.23 The inefficiency of nasal inhalation of conventional-sized aerosol particles combined with the lack of a good seal of the face mask to the face in many cases contributes to this advice. However, recent advances in face mask design and construction material, with improved flexibility potentially resulting in a better fit of the mask to the face, have increased the efficiency of drug delivery with these devices.24 While it is still probable that mouthpiece delivery would still be more efficient, in patients with poor inhalation compliance it is questionable whether a mouthpiece would deliver more drug than a face mask.25 ERS-ISAM recommendations are to use a face mask in children under 3 years of age and to transition to a mouthpiece in children aged 3-6 years.26

Device Use Parameters

Factors such as the effect of multiple actuations into the spacer prior to inhalation and varying delays between actuation and” inhalation” were elegantly characterized in a series of in vitro studies by Barry and O’Callaghan in the 1990s.15,16 There have been many developments since then in pMDI formulations (including the development of extrafine formulations) as well as improvements to spacer design and construction. However, until evidence is shown to the contrary, their recommendations to minimize the delay between actuation and inhalation and only introduce a single actuation into the spacer prior to each patient inhalation should be adhered to.

Spacers may be used with tidal breathing or slow maximal inhalations. There is evidence to indicate that the maximal inhalation technique increases lung deposition27 but tidal breathing must be used in infants and very young children. ERS-ISAM guidelines recommend tidal breathing in children under 6 years with either facemask (<3 years) or mouthpiece (3–6 years), whereas children aged 6 years and older should be trained to use a single, slow maximal inhalation with a mouthpiece.26 Lower lung deposition is achieved when aerosol is delivered to crying or struggling infants and children.28–31

Patient-Related Physiological and Cognitive Factors

Factors relating to patient groups of different ages as well as differing disease conditions are covered in detail in other sections of this textbook. Despite the fact that the pMDI-spacer is the recommended choice of device in many countries for infants and preschool children with asthma, lung deposition in this age group is lower28 than that seen in the other age groups.32 There is little published data on lung deposition from spacers in the elderly, but it has been shown that older patients exhibited more inhaler use errors and found it more difficult to simultaneously coordinate actuation of a pMDI used without a spacer, with inhalation. In addition, older patients were more likely to continue to use pMDIs incorrectly after training,33 indicating that the spacer use may be indicated in this age group also.

SUMMARY

  • Spacers and valved holding chambers (VHCs) are designed to overcome common problems associated with the use of pMDIs, namely difficulty in coordination of inhalation with the actuation of the pMDI and high oropharyngeal deposition of drug.

  • The design of spacers and VHCs can influence the dose delivered. Various factors such as spacer volume, shape, valve design, using multiple actuations, delay between actuation and inhalation, and pMDI formulation could affect the efficiency of drug delivery.

  • Reduced dose delivery due to build up of electrostatic charge within plastic spacers could be overcome by using spacers/VHCs made of low or anti-static materials.

  • Spacers/VHCs are recommended for patients who are unable to coordinate actuation with inhalation, including:
    • Infants and children under 6–7 years of age
    • Any patient with poor pMDI actuation-inhalation, including elderly patients
    • Patients using drugs with adverse local or systemic effects from oropharyngeal impaction and gastrointestinal absorption
    • Patients requiring bronchodilator therapy for acute asthma exacerbations.
  • The recommended technique for using spacers or VHCs is as follows:
    • Pre-treat spacers/VHCs not constructed from low or anti-static materials to remove electrostatic charge
    • Shake the pMDI prior to attachment to the spacer
    • One actuation should be introduced into the spacer prior to patient inhalation. This can be repeated for the number of pMDI actuations required
    • Minimize the delay between actuation into the spacer and patient inhalation
    • Slow, steady inhalations are recommended, whether using tidal breathing or maximal inhalations.

Abbreviations Used

ERS

European Respiratory Society

ISAM

International Society for Aerosols in Medicine

pMDI

pressurized metered dose inhaler

VHC

valved holding chamber

Author Disclosure Statement

The author declares no conflict of interest exists.

References

  • 1. Hidinger KG, Kjellman NI. Childhood asthma: improved efficacy of pressurized terbutaline aerosol by use of a 750-ml spacer. Respiration 1984;45(2):157–160. [DOI] [PubMed] [Google Scholar]
  • 2. van Beerendonk I, Mesters I, Mudde AN, Tan TD. Assessment of the inhalation technique in outpatients with asthma or chronic obstructive pulmonary disease using a metered-dose inhaler or dry powder device. J Asthma 1998;35(3):273–279. [DOI] [PubMed] [Google Scholar]
  • 3. Aziz NA, Norzila MZ, Hamid MZA, Noorlaili MT. Skills amongst parents of children with asthma: a pilot interventional study in primary care setting. Med J Malaysia 2006;61(5):534–539. [PubMed] [Google Scholar]
  • 4. Leach CL, Colice GL. A pilot study to assess lung deposition of HFA-beclomethasone and CFC-beclomethasone from a pressurized metered dose inhaler with and without add-on spacers and using varying breathhold times. J Aerosol Med Pulm Drug Deliv 2010;23(6):355–361. [DOI] [PubMed] [Google Scholar]
  • 5. Newman SP, Clark AR, Talaee N, Clarke SW. Pressurised aerosol deposition in the human lung with and without an “open” spacer device. Thorax 1989;44(9):706–710. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Nikander K, Nicholls C, Denyer J, Pritchard J. The evolution of spacers and valved holding chambers. J Aerosol Med Pulm Drug Deliv 2014;27(Suppl 1):S4–S23. [DOI] [PubMed] [Google Scholar]
  • 7. Zar HJ, Liebenberg M, Weinberg EG, Binns HJ, Mann MD. The efficacy of alternative spacer devices for delivery of aerosol therapy to children with asthma. Ann Trop Paediatr 1998;18(2):75–79. [DOI] [PubMed] [Google Scholar]
  • 8. Zar HJ, Brown G, Donson H, Brathwaite N, Mann MD, Weinberg EG. Home-made spacers for bronchodilator therapy in children with acute asthma: a randomised trial. Lancet (London, England) 1999;354(9183):979–982. [DOI] [PubMed] [Google Scholar]
  • 9. Lipworth BJ, Clark DJ. Lung delivery of non-CFC salbutamol via small volume metal spacer and large volume plastic spacer devices compared with an open vent jet nebulizer. Br J Clin Pharmacol 1998;45(2):160–163. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Chou KJ, Cunningham SJ, Crain EF. Metered-dose inhalers with spacers vs nebulizers for pediatric asthma. Arch Pediatr Adolesc Med 1995;149(2):201–205. [DOI] [PubMed] [Google Scholar]
  • 11. Castro-Rodriguez JA, Rodrigo GJ. Beta-agonists through metered-dose inhaler with valved holding chamber versus nebulizer for acute exacerbation of wheezing or asthma in children under 5 years of age: a systematic review with meta-analysis. J Pediatr 2004;145(2):172–177. [DOI] [PubMed] [Google Scholar]
  • 12. Ploin D, Chapuis FR, Stamm D, Robert J, David L, Chatelain PG, et al. High-dose albuterol by metered-dose inhaler plus a spacer device versus nebulization in preschool children with recurrent wheezing: A double-blind, randomized equivalence trial. Pediatrics 2000:106(2 Pt 1):311–317. Available from: http://www.ncbi.nlm.nih.gov/pubmed/10920157 [DOI] [PubMed] [Google Scholar]
  • 13. Closa RM, Ceballos JM, Gómez-Papí A, Galiana AS, Gutiérrez C, Martí-Henneber C. Efficacy of bronchodilators administered by nebulizers versus spacer devices in infants with acute wheezing. Pediatr Pulmonol 1998;26(5):344–348. [DOI] [PubMed] [Google Scholar]
  • 14. Newman SP, Newhouse MT. Effect of add-on devices for aerosol drug delivery: deposition studies and clinical aspects. J Aerosol Med 1996;9(1):55–70. [DOI] [PubMed] [Google Scholar]
  • 15. Barry PW, O’Callaghan C. Multiple actuations of salbutamol MDI into a spacer device reduce the amount of drug recovered in the respirable range. Eur Respir J 1994;7(9):1707–1709. [DOI] [PubMed] [Google Scholar]
  • 16. Barry PW, O’Callaghan C. The effect of delay, multiple actuations and spacer static charge on the in vitro delivery of budesonide from the Nebuhaler. Br J Clin Pharmacol 1995;40(1):76–78. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Wildhaber JH, Devadason SG, Eber E, Hayden MJ, Everard ML, Summers QA, et al. Effect of electrostatic charge, flow, delay and multiple actuations on the in vitro delivery of salbutamol from different small volume spacers for infants. Thorax 1996;51(10):985–988. Available from: http://www.ncbi.nlm.nih.gov/pubmed/8977597 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Bisgaard H, Anhøj J, Klug B, Berg E. A non-electrostatic spacer for aerosol delivery. Arch Dis Child 1995;73(3):226–230. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Kenyon CJ, Thorsson L, Borgstrom L, Newman SP. The effects of static charge in spacer devices on glucocorticosteroid aerosol deposition in asthmatic patients. Eur Respir J 1998;11(3):606–610. [PubMed] [Google Scholar]
  • 20. Schultz A, Le Souëf TJ, Venter A, Zhang G, Devadason SG, Le Souëf PN, et al. Aerosol inhalation from spacers and valved holding chambers requires few tidal breaths for children. Pediatrics American Academy of Pediatrics; 2010;126(6):e1493-e1498. [DOI] [PubMed] [Google Scholar]
  • 21. Leach CL, Davidson PJ, Hasselquist BE, Boudreau RJ. Lung deposition of hydrofluoroalkane-134a beclomethasone is greater than that of chlorofluorocarbon fluticasone and chlorofluorocarbon beclomethasone: a cross-over study in healthy volunteers. Chest 2002;122(2):510–516. Available from: http://www.ncbi.nlm.nih.gov/pubmed/12171824 [DOI] [PubMed] [Google Scholar]
  • 22. Leach CL, Davidson PJ, Hasselquist BE, Boudreau RJ. Influence of particle size and patient dosing technique on lung deposition of HFA-beclomethasone from a metered dose inhaler. J Aerosol Med 2005;18(4):379–385. [DOI] [PubMed] [Google Scholar]
  • 23. Chua HL, Collis GG, Newbury AM, Chan K, Bower GD, Sly PD, et al. The influence of age on aerosol deposition in children with cystic fibrosis. Eur Respir J 1994;7(12):2185–2191. [DOI] [PubMed] [Google Scholar]
  • 24. Ditcham W, Murdzoska J, Zhang G, Roller C, von Hollen D, Nikander K, et al. Lung deposition of 99mTc-radiolabeled albuterol delivered through a pressurized metered dose inhaler and spacer with facemask or mouthpiece in children with asthma. J Aerosol Med Pulm Drug Deliv 2014;27(Suppl 1):S63–S75. [DOI] [PubMed] [Google Scholar]
  • 25. Trend S, Ditcham W, Zhang G, von Hollen D, Nikander K, Devadason S. Analysis of lung deposition of radiolabeled salbutamol compared to breathing parameters in asthmatic children using pressurised metered-dose inhalers with valved holding chambers, with and without face masks. Respirology 20(S2):94; doi: 10.1111/resp.12495_7 [DOI] [Google Scholar]
  • 26. Laube BL, Janssens HM, de Jongh FHC, Devadason SG, Dhand R, Diot P, et al. What the pulmonary specialist should know about the new inhalation therapies. Eur Respir J 2011;37(6):1308–1331. [DOI] [PubMed] [Google Scholar]
  • 27. Roller CM, Zhang G, Troedson RG, Leach CL, Le Souëf PN, Devadason SG. Spacer inhalation technique and deposition of extrafine aerosol in asthmatic children. Eur Respir J 2007;29(2):299–306. [DOI] [PubMed] [Google Scholar]
  • 28. Tal A, Golan H, Grauer N, Aviram M, Albin D, Quastel MR. Deposition pattern of radiolabeled salbutamol inhaled from a metered-dose inhaler by means of a spacer with mask in young children with airway obstruction. J Pediatr 1996;128(4):479–484. [DOI] [PubMed] [Google Scholar]
  • 29. Iles R, Lister P, Edmunds AT. Crying significantly reduces absorption of aerosolised drug in infants. Arch Dis Child 1999;81(2):163–165. Available from: http://www.ncbi.nlm.nih.gov/pubmed/10490528 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. Erzinger S, Schueepp KG, Brooks-Wildhaber J, Devadason SG, Wildhaber JH. Facemasks and aerosol delivery in vivo. J Aerosol Med 2007;20(Suppl 1):S78–S83; discussion S83–S84. Available from: http://www.ncbi.nlm.nih.gov/pubmed/17411409 [DOI] [PubMed] [Google Scholar]
  • 31. Everard ML. Trying to deliver aerosols to upset children is a thankless task. Arch Dis Child 2000;82(5):428. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Wildhaber JH, Janssens HM, Piérart F, Dore ND, Devadason SG, LeSouëf PN. High-percentage lung delivery in children from detergent-treated spacers. Pediatr Pulmonol 2000;29(5):389–393. Available from: http://www.ncbi.nlm.nih.gov/pubmed/10790251 [DOI] [PubMed] [Google Scholar]
  • 33. Aydemir Y. Assessment of the factors affecting the failure to use inhaler devices before and after training. Respir Med 2015;109(4):451–458. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Aerosol Medicine and Pulmonary Drug Delivery are provided here courtesy of SAGE Publications

RESOURCES