Abstract
Purpose:
Magnetic resonance acoustic radiation force imaging (MR-ARFI) is an exceptionally promising technique to non-invasively confirm targeting accuracy and estimate exposure of low-intensity transcranial focused ultrasound applications. Implementing MR-ARFI in the human brain has been hindered by 1) sensitivity to subject motion, and 2) insufficient signal-to-noise ratio at low (<1.0 MPa) ultrasound pressures. The purpose of this study was to optimize human MR-ARFI to allow reduced ultrasound exposure while at the same time being robust to bulk and physiological motion.
Methods:
We developed a novel timeseries approach to MR-ARFI with a single-shot spiral-out MRI sequence and correction for respiratory and cardiac motion artifacts. An MR-compatible 4-element 500 kHz focused ultrasound transducer was coupled to the head and targeted to 60 mm depth in 5 participants. During spiral scans, two 6 ms focused ultrasound pulses (0.5–0.9 MPa in situ) were delivered in on-off blocks of 25 time frames.
Results:
Our method generates ARFI maps that with correction are largely immune to bulk and pulsatile brain motion with reduced scan time (80 seconds per acquisition). Robust ARFI signals were observed at the expected target in 4 human participants, using low intensity ultrasound that does not produce significant tissue heating, confirmed both by simulation and MR-thermometry.
Conclusion:
Single shot spiral MR-ARFI is motion robust in human applications, provides reduction in ultrasound exposure, and reduced scan time, enabling iteration for image-guided targeting. This provide persuasive proof-of-principle that MR-ARFI can be used as a tool to guide ultrasound-based precision neural circuit therapeutics.
Keywords: MR-guided focused ultrasound, neuromodulation, transcranial ultrasound stimulation, single-shot spiral pulse sequence, motion artifact, MR-ARFI
Introduction
MR-ARFI is a powerful method to confirm precise anatomical targeting of focused ultrasound for therapeutic applications1,2. As opposed to high-intensity focused ultrasound ablation procedures, in which an increase in temperature can be visualized with MR-thermometry3–5, in low-intensity applications such as neuromodulation, targeted drug or other payload release, and blood-brain barrier opening, the temperature rise is generally undetectable by MR6–8. The need for a technique such as ARFI has become more important since the emergence of transcranial ultrasound applications, which implements transcranial focused ultrasound as a tool for noninvasive brain stimulation, blood brain barrier (BBB) opening or ablation. Noninvasive transcranial ultrasound holds great promise for non-invasive treatment of brain disorders and causal manipulation in human neuroscience, but the accuracy and focality of transcranial ultrasound are severely impacted by refraction, reflection, and attenuation of ultrasound waves by the skull. Because the ultrasound focus is small (<1 cm), and the targets are often deep inside the brain (usually >6 cm), even small deviations (5–10°) in the incident angle of the beam as it passes through the skull can lead to completely missing the target. Simulations greatly help to correct for skull aberrations but are limited by uncertainty in assigning acoustic properties to skull images, simplifications inherent in computational modeling, and various experimental factors not included in the simulations. Therefore, without a ground-truth measure of the focus, the reliability and consistency of transcranial ultrasound will remain unknown. MR-ARFI measures the phase perturbations induced by ultrasound waves in a motion-encoding magnetic field gradient. MR-ARFI has been shown to reliably visualize the ultrasound focus in large animal models in vivo, but the 2D Fourier Transform (2DFT) acquisitions that were previously employed required relatively high acoustic pressures (2–3 MPa)9–12 in order to overcome motion artifacts. In this prior work, two 2DFT spin echo pulse acquisitions were utilized to acquire the complex images, from which the phase difference is extracted to create tissue displacement (ARFI) maps13,14. These could be either ultrasound-on/ultrasound-off with the same encoding gradients or ultrasound-on with encoding gradients inverted in the second acquisition15. Either way, while the reconstructed magnitude images were essentially immune to small bulk motions in 2DFT, the corresponding phase images were degraded by bulk motion and pulsatile brain artifacts16,17. This is true because ultrasound-induced tissue displacement is only a few microns, compared to bulk motion that can be substantially larger in human scans. Using accelerated 2DFT acquisitions18 is another approach to reducing the number of separate k-space lines and thereby the ghosting artifacts. While the cited results demonstrate advantages in MR-ARFI of non-human-primates fixed in a stereotactic frame, ghosting could not be evaluated since the bulk motion typical in a human without stereotactic fixation was not examined.
An alternative is to use single-shot acquisitions, with either EPI or spiral imaging, to generate the two complex images19, which may have reduced motion artifacts. However, here we use a timeseries of singe-shot spiral images, while triggering the ultrasound on and off in blocks, analogous to fMRI block task designs20 and compare ARFI results with those of the 2DFT design. Induced pseudo-random motion of a phantom allows accurate comparison of the motion sensitivity of the two methods.
Methods
Magnetic resonance imaging
MR images were acquired on a 3T scanner (DV750, GE Healthcare, Milwaukee, WI, USA) with spin echo (SE) sequences. Inverted bipolar motion encoding gradients (MEGs) were employed, with a total duration of 20 or 30ms, corresponding diffusion b-values17 of 29.5 or 100.5 s/mm2 and amplitude of 50 mT/m. An 8-channel head coil (MRI Devices) was used.
2DFT ARFI Image acquisition.
For comparison with the proposed timeseries method, we used a conventional 2DFT SE sequence with TR/TE of 800ms/50ms, resolution of 1.72 mm x 1.72 mm, 4 mm slices, scan time of 102 seconds. For MR-ARFI, two separate scans were obtained, without and with ultrasound, for subsequent processing of the complex images to obtain the phase difference caused by acoustic radiation force (ARF). However, in this study, only the ghosting between 2DFT and single-shot timeseries methods was demonstrated (Fig. 3).
Figure 3. Motion correction.
ARFI images obtained with (A) 2DFT, no motion in phantom; (B) Single-shot timeseries spiral, no motion in phantom; (C) 2DFT, the focal spot with ±75 microns random motion is heavily distorted; (D) Single-shot timeseries spiral, with the same motion, but the spiral method retains utility. Phase images obtained using 2DFT MR- ARFI pulse sequence in (E) and parametric maps of single shot spiral timeseries shown in (F).
MR-ARFI using Spiral Time series method
Introduced in this study as an alternative to conventional 2DFT imaging, we utilized a timeseries approach with a variable-density single-shot spiral-out MRI sequence, acquiring 100 time frames with TR/TE = 800ms/40–50ms, 3 slices of 4 or 5 mm thick, with in plane resolution of 2.75 × 2.75 mm2, reconstructed into 1.72 × 1.72 mm2 pixel size (128 × 128 matrix), scan time 80 seconds. A block design analogous to that used in fMRI20,21 is used to turn on and off the ultrasound during the scan in blocks of 25 time frames, as shown in Fig. 1A. The variable-density spiral readout duration is 16.9 ms, less than half that of a comparable resolution single-shot EPI trajectory (40.2ms), which thereby provides greater immunity to motion22.The alternating on-off blocks allowed detrending of the magnitude and phase images, to correct for drift.
Figure 1. Pulse sequence and processing pipeline.
(A) single-shot (SS) spiral ARFI spin echo pulse sequence, showing repeated bipolar MEGs together with ultrasound pulses to generate phase encoding of tissue displacement. The sequence is repeated an equal number of times with ultrasound on and with ultrasound off. (B) The thermometry gradient echo sequence using the same duration of ultrasound as used for ARFI. (C) The processing pipeline for producing ARFI statistical map using SS spiral pulse sequence, and the 80-second block-design protocol for measuring ARF.
During MR-ARFI, focused ultrasound was applied during the second lobe of the first bipolar and the second lobe of the second inverted bipolar motion encoding gradient (Fig. 1A). The resulting phase shift Δϕ of tissue displaced by the sonication is
| (1) |
where γ is the gyromagnetic ratio for 1H protons, Gmeg(t)is the MEG waveform, Δx(t) is the time-dependent displacement, is defined as an average displacement, G is the maximum amplitude of the MEG, and Tmeg is the total duration of the MEGs.
Generation of ARFI maps
The spiral acquisitions were reconstructed as a complex (magnitude & phase) timeseries. During reconstruction, the phase of the second time frame was subtracted from all subsequent time frames to keep the phase timeseries well within ±π, thereby avoiding phase wraparound. With reference to Fig. 1A, the pixel-wise average of the magnitude timeseries was used to generate a mask of those pixels with magnitude greater than 25% of the peak brain signal. This mask was applied to the phase timeseries to exclude non-brain pixels during subsequent processing with a General Linear Model (GLM) to identify voxels with significant phase differences between ultrasound-on/off blocks.
Bulk motion correction.
To correct for head motion in the longitudinal direction (direction of sonication beam), the average phase of the brain slices collected was calculated for every time frame, as a regressor of no interest in the general linear model. Since the sonication region was a tiny fraction of the brain volume, this average signal demonstrated bulk motion of the brain, but was minimally affected by ARF modulation.
Physiological correction.
In fMRI, it is standard practice to correct timeseries data for quasiperiodic physiological artifacts23. These include cardiovascular-induced brain pulsatility (monitored by photoplethysmography [PPG] on a finger), respiratory-induced changes in blood volume, and resonance frequency shifts caused by diaphragm movement (monitored with a respiration belt). For the purpose of removing physiological noise, we applied RETROICOR24 to the phase timeseries which corrects for cardiac and respiratory-induced phase artifacts (Fig. 1A).
General Linear Model (GLM).
The corrected timeseries of phase images Δ∅(t) was statistically tested for the inclusion of signals that matched the off-on-off-on block design d(t) of the ultrasound stimulation (Fig. 1C) using a GLM21,25. The GLM is defined as
| (2) |
where β is the amplitude of the design contained in the measured timeseries data for every voxel Δ∅(t), N(t) is a nuisance regressor of no interest with βN its amplitude, and ε(t) is residual noise. Then β represents the observed phase change due to ARF.
The average phase timeseries is included as a nuisance regressor to diminish bulk motion effects, and a quadratic detrending regressor is included to diminish slow phase drifts in the timeseries. The resulting Statistical Parametric Maps (SPMs)were presented as T-scores that depict probability that the block design was observed in the timeseries.
As the last step in the processing pipeline (Fig 1C), spatial filtering was employed to reduce residual low-spatial frequency bias in the SPMs. This was similar to, but not exactly equal to, the multi-baseline processing often used in MR-thermometry in brain applications3,26. Utilizing a Fourier Transform of the ARFI maps, a high-pass filter was performed by zeroing the Fourier components within a radius of 2 of the center of the 128×128 k-space matrix (where the lowest spatial frequencies reside). In this way, the highest spatial frequencies, including those of the ARFI signal, were not affected. The inverse Fourier transform was then applied. This high-pass filter minimally affects the amplitude of the ARFI signal (98% retention) because of the spatially discrete nature of ARFI, but substantially removes low spatial frequency bias.
Phantom ARFI.
To optimize the design for tissue response with MR-ARFI and examine sensitivity to bulk motion, a 1.0 MHz 128-element phased array US transducer (Image Guided Therapy, Paris, France) was focused to a depth of 60 mm to generate micro-displacement in a soft tissue-mimicking phantom (Computerized Imaging Referencing System, Virginia, USA). To generate controlled motion, the phantom with transducer attached was placed in the head coil on a platform with rollers that was attached by a 2.3m rod to a stepper-motor translation stage at the end of the patient table. The motor was driven in pseudo-random steps of multiples of 25.4 micrometers (microns) asynchronously with the ARFI scan TR.
As part of the design optimization, it was useful to estimate the ARF time constant. In Kaye et al, 201316, rise time response of pig brain post-mortem was measured to be 5.9ms. The same time-constant value was used as a reference to simulate encoded phase with inverted bipolar gradient. Time delays of 1ms and 2ms were applied between the ultrasound pulse and the MEGs for a range of time constant to evaluate the simulated encoded phase. To validate the simulation values, we performed ARFI experiments using our phantom with similar properties to soft tissue (Young’s Modulus: 12 +/− 2 kPa, Speed of sound: 1540 +/− 10 m/s, Attenuation coefficient: 0.5 +/− 0.1 dB/cm/MHz).
Thermometry Pulse sequence
To investigate if there was any temperature increase from the sonication, MR-thermometry was obtained using the proton resonant frequency shift method, with a thermal coefficient of 0.01 ppm/°C27. A gradient-recalled echo (GRE) pulse sequence with readout identical to the single-shot spiral SE sequence employed for ARFI was used to collect a timeseries of 100 time frames, while the ultrasound was applied with identical duration to that with ARFI in the same block design. In contrast to the effect of RF pulses, secondary E-fields and the induction of currents in tissue are both negligible for gradient switching at kilohertz frequencies28. In addition, there is evidence in the literature29 that imaging gradients do not generate appreciable eddy currents (and thereby heat) in tissues, so any thermal effects would be due to diffusion from heated implants (of which there are none in our study). The motion encoding gradients were omitted to eliminate phase contributions from ARF (Fig 1B). A TR/TE of 800/40 ms was used, with identical slice positioning as for the ARFI acquisition. To generate maps of sonication-induced temperature rise, the phase of the last 5 time frames of the first block (i.e., without sonication) was pixel-wise averaged and subtracted from the pixel-wise average of the last 5 time frames in the last (4th) block (i.e., with sonication). This timing allowed the temperature in phantoms to reach a maximum and stabilize during sampling.
Transcranial Ultrasound in Human.
A 500 kHz 4 element annular array (NeuroFUS System, UK) was used with a free-water peak pressure of approximately 1.48 MPa, as measured with a calibrated hydrophone (Onda, Sunnyvale, USA). During spiral scans, two 6 ms pulses with a free field value of 1.48 MPa (Isppa = 68 W/cm2) at 60 mm depth were applied and repeated in on-off blocks of 25 time frames (Fig. 1A). Using conservative derating as described by Attali et al.30 for the skull and 0.5 dB/cm/MHz for brain tissue, this yielded an estimate of pressure of 0.83 MPa and a MItc of 1.1731 and a total exposure of 38 J/cm2. Individual simulations from the four subjects gave peak pressure estimates from 500–800 kPa depending on the participants’ skull properties (see simulations below), with MItc values of 0.7 – 1.13. Ultrasound gel (Aquasonic 100 Ultrasonic gel, Parker Laboratories, NJ, USA) and a 10 mm thick gel-pad (Aquaflex, Parker Laboratories, NJ, USA) were used to couple the transducer to the participants’ temporal bone. No specific target was identified for sonication. However, a 60 mm depth of focus was used, and the transducer was applied to the temporal bone, consistently across the subjects. The details of the ultrasound parameters are provided in the supplemental materials. This study was approved by the Institutional Review Board of Stanford University. All subjects provided informed consent.
Acoustic Pressure and Temperature Simulation
Simulations were used to estimate the ultrasound pressure and the temperature rise at the focus. Biophysical head models were generated with the SimNIBS Charm function32 using each participant’s T1 and T2 scans. The transducer position was determined in Brainsight software (Rogue Research Inc., CAN) using the planning scans from each experiment. The ultrasound beam was then simulated in open-source BabelBrain software33 using each participant’s SIMNIBS head model and ZTE pseudo-CT skull image34. BabelBrain generates 3D volumes of the estimated intensity and temperature rise at each pixel in the beam path.
Results
ARF time constant simulation and measurements
We first used simulations to parameterize how the brain ARF time constant and ultrasound-MEG time delays influence ARFI measurements (quantified as the encoded phase). The ARF time constant for cadaverous pig brain was quantified as ~6 ms in Kaye et., al 201316. Figure 2A shows example simulation results, depicting the simulated encoded phase generated by two 6-ms ultrasound pulses at the measured 6 ms time constant, with a 0 ms, 1 ms and 2 ms time-delay between ultrasound pulse and the MEG. This figure shows greater simulated phase using 2 ms time-delay. We then simulated ARFI measurements for a range of ARF time constants and ultrasound-MEG delays. Figure 2B shows simulation results estimating encoded phase for a range of time constant values and two different time delays applied between ultrasound and MEGs. As can be seen, for time constant values greater than 5 ms, a 2 ms time shift gives the greater encoded phase. We then tested the impact of the ultrasound-MEG time delay on measured encoded phase in a CIRS phantom (Fig. 2C,D). Figure 2C shows MR-ARFI data in CIRS phantom at 1 ms (left) and 2 ms (right) ultrasound-MEG time delays. The ARFI value (t-score) was higher at a 2 ms delay compared to a 1 ms delay (Fig. 2D), confirming our simulation results. Therefore, a ultrasound-MEG time delay of 2 ms was employed to maximize displacement in subsequent reported experiments.
Figure 2. Phantom acoustic radiation force (ARF) time constant measurement and benefit of ultrasound delay.
(A) Simulation of encoded phase for ARF time constant of 6 ms using inverted bipolar gradients. The simulated phase is shown in cyan, pink and black color for delaying ultrasound w.r.t MEGs at 0ms, 1ms and 2ms, respectively. (B) Simulated encoded phase for a range of time constant values from1 to 7ms. This figure shows that applying 2ms delays to ultrasound relative to MEGs resulted in a higher phase when the time constant is more than 5ms. (C) Experimental MR-ARFI data in a tissue-mimicking phantom in the middle slice using 1ms (left) and 2ms (right) time shift. The red circle area is defined by a matrix of 5×5 around the maximum value to calculate average t-score. (D) MR-ARFI signal (t-score value) was doubled with 2-ms ultrasound-MEG delay compared to a 1-ms delay (error bars show SEM over three slices).
Motion Correction using Time series spiral method
We then quantified how motion influences ARFI signal in phantoms. Figure 3 shows ARFI maps acquired with (A) 2DFT, no motion; (B) spiral timeseries, no motion; (C) 2DFT with ±75 micron random motion; (D) spiral timeseries with the same motion. As may be seen, the two methods are roughly equivalent with no intentional motion, but the 2DFT ARFI maps with even the small motion applied fail to show a detectable focal spot, while with the spiral maps the focal spot remains detectable. Panels E shows 2DFT in human in which motion artifact distorted the phase image whereas in panel F we demonstrated motion-robust spiral timeseries. It is notable that the spiral scan time was only 40% that of the 2DFT acquisition, suggesting that even better ARFI could be obtained in the presence of motion, if the same scan time had been used.
Acoustic and Bioheat simulation comparison to MR-thermometry in the brain
As a last step in our analysis, we wanted to determine if our protocol would generate any significant tissue heating. We performed realistic bioheat simulations with Babelbrain for each participant using our ARFI protocol as input. Fig. 4A,B depicts the simulation results for the subject with the least attenuative skull, showing a maximum in-brain intensity of 21 W/cm2 (panel A) and a maximum in-brain temperature rise of 0.5°C (panel B). Next, we looked at the MR-thermometry measures of tissue heating, which were acquired during sonications equivalent to those applied during the ARFI experiment. In this participant (as in others), no detectable temperature rise was present at the target during ultrasound pulses with dose identical to that in the ARFI experiment, with background noise (standard deviation) of 0.081°C. The simulated derated intensity values and temperature rise for the other participants are (15.7 W/cm2, 0.1°C), (8.7 W/cm2, 0.2°C) and (16.4 W/cm2, 0.1°C). We confirmed in phantoms that temperature rises of 0.1°C are detectable with our MR-thermometry protocol, ruling out the possibility that the lack of a temperature rise is due to inadequate SNR. Using the same ARFI sonication parameters in a soft tissue phantom resulted in 0.21°C temperature rise at the focal spot. This measurement indicates that ultrasound bioheat simulations may overestimate tissue heating at least in certain situations. Additionally, these data confirm that ARFI signals can be robustly measured in the human brain in situ with no detectable temperature rise at the target.
Figure 4. Comparing acoustic simulation and MR-thermometry.
This figure shows representative data of the simulated acoustic intensity (A), simulated temperature rise (B) and MR-thermometry (C) for the most transmissive skull amongst participants in this study. The simulated temperature rise for this participant at the focus is < 0.5oC. However, MR-thermometry shows no temperature rise, which is consistent with the cooling effect of the brain’s vasculature.
MR-ARFI in Human
We next attempted to measure ARFI signals in the human brain in situ. As shown in Figure 5, a 500 kHz 4 element annular array (BrainBox System, UK) was coupled to the temporal bone of a human participant. During spiral scans, a free field value of 1.48 MPa (ISPPA = 68 W/cm2) at 60 mm was applied. The derated pressure at the focus was estimated as 0.83 MPa (Attali et al, 202230 for skull attenuation and 0.5dB/cm/MHz for brain attenuation). Panel A and B show the planning MRI and the 3D reconstruction of the MR-ARFI in the brain, respectively. Panel C demonstrates t-score heatmap of the ARFI spot overlaid on T1-weighted image in coronal, sagittal and horizontal views. The maximum displacement for this subject was calculated as 1.47 microns.
Figure 5. MR-ARFI in human.
The planning MRI using three slices centered at 60mm from the transducer (A). 3D rendering of the ARFI spot with the transducer coupled to the left temporal bone (B). Heatmap of the ARFI spot (t-score value) overlaid on T1-weighted image in three planes (C).
We performed ARFI experiments in 5 participants. ARFI maps are shown for 4 participants in Fig. 6A. No ARFI was observed in the fifth participant (not shown), despite experiment repetition with that participant on a different day. Note that there was significant variability in the signal from participant to participant. ARFI T-scores reached significance at p < 0.05 in S1, S3, and S4 with T at least 1.7. In S2, the ARFI signal was observable at the expected focal spot, but the T-score did not reach significance (p>0.28). In S1 we did not collect cardiac and respiratory signal to correct for physiological noise, thus, the high variability of ARFI signal for this participant should be considered with caution. To assess if differences in skull anatomy could predict differences in ARFI signal across participants, we simulated ultrasound propagation in biophysically realistic simulations using ZTE pseudo-CTs acquired from each participant. Simulated intensity varied significantly from participant to participant (8.7 to 23.5 W/cm2). Interestingly the simulated intensity only weakly predicted the ARFI measurement (Fig. 6B, scatter plot of ARFI peak signal, t-score, vs simulated intensity), indicating that differences in phase aberrations, simulation accuracy, brain stiffness, brain motion, hair thickness, and/or variability in transducer coupling play a major role in the measured value of ARFI. Given that ARF is the strongest candidate mechanism of ultrasound neuromodulation effects35, the fact that simulations do not predict ARFI well emphasizes further how important it is to measure ARFI to be able to determine dosing for each experimental participant.
Figure 6. MR-ARFI in human participants.
T-score maps of 4 participants; a minimum T-score of 1.7 corresponds to a statistical probability of p < 0.05, uncorrected for multiple comparisons (A). Peak ARFI phase measurements as a function of simulated derated intensity at the focal spot (B).
Discussion
Here we demonstrate that ARFI can be effectively and efficiently used to 1) confirm the location of the focus, and 2) roughly estimate ultrasound dose at the target, at ultrasound intensities that are less than half of what has previously been reported and that do not produce significant tissue heating (confirmed both by simulation and MR-thermometry). The acoustic parameters used for ARFI in this report are within the safety range of ITRUSST safety consensus31. This work provides the transcranial focused ultrasound community with an extremely powerful tool to give ground truth on 1) where in the brain focused ultrasound is delivered (spatial precision), as well as 2) how much ultrasound is delivered (exposure precision).
There is a consensus in the field that reports of highly variable transcranial focused ultrasound effects in published and ongoing studies could be due in large part to uncontrolled variability in the accuracy of targeting36. The targeting of transcranial ultrasound depends on simulations which have inherent imprecisions, and both neuronavigated and in-scanner beam positioning solutions require substantial technical skill and are therefore highly operator-dependent. Even a small 5–10° deviation from the desired transducer angle will lead to a complete miss of the target in small deep structures that are the most likely targets of focused ultrasound37. ARFI allows experimenters to confirm the focal spot at the desired site, and to reposition the focus (either by manual repositioning of the transducer or electronic beam steering for phase arrays) if the focus is not at the desired location.
Inconsistent ultrasound dosing is also likely a major contributor to variability of reported transcranial ultrasound neuromodulation effects. Although simulations help to an extent, there is still some uncertainty about acoustic properties as a function of imaging parameters, both CT Hounsfield units (HU) and MR values38. In addition, some properties of the skull, such as pore eccentricity, are not captures by ZTE or CT39. The data that we report here are T-score statistical parametric maps, which can be converted to an average microscopic displacement in microns (Eq. 2). This knowledge combined with multifrequency MR elastography and tissue stiffness modeling may result in more accurate estimation of in vivo acoustic intensity40. It is reasonable to predict that ARFI could eventually provide a gold-standard unconfounded measure of individualized in situ ultrasound dose. This will in turn allow a more robust determination of how ultrasound dose relates to neuromodulation effects, which is poorly understood.
The key advance in this paper is our demonstration that the timeseries single-shot MR-ARFI method introduced here has substantial advantages over 2DFT methods, with reduced sensitivity to interference from bulk and pulsatile brain motion. Bulk motion of the head can cause large artifacts in 2DFT ARFI methods because the motion occurs over acquisition of phase-encoding lines of k-space, which takes many seconds or minutes. By contrast, each image of the complex time series single-shot method introduced in this study takes less than 17 ms to acquire, which eliminates intra-image bulk motion effects, while temporal averaging over the time series regains SNR. Furthermore, the short spiral readout (compared with 2x longer EPI readout) reduces susceptibility-drop which increases stability against brain pulsatility19,41. Collecting a time series allows all the processing steps of BOLD fMRI to be exploited for low-SNR ARF phase data. Thus, our processing uses a GLM instead of simple subtraction, which allows the incorporation of nuisance regressors to 1) reduce bulk motion with a regressor made from whole-head phase shifts; 2) cycle the ultrasound on and off in multiple blocks to eliminate phase drift while enhancing sensitivity to ARF displacement and reduce temperature rise; and 3) account for physiological pulsation from cardiac and respiratory functions using fMRI-standard processing methods that utilize simultaneously collected cardiac and respiratory data. It has been observed that 2DFT acquisitions with repeated bipolar gradients are more robust against bulk motion and eddy currents compared to inverted bipolar gradients16,17. However, using time-series methods with physiological and bulk motion correction, allows inverted bipolar gradients to provide higher ARFI SNR with reduced artifacts.
The time series approach allows for a reduction in ultrasound exposure to 15% that in our previous animal study (Mohammadjavadi et al, 2022)9, while improving motion robustness with reduced scan time (80s vs. 205s). The ultrasound parameters used in this study are within the ITRUSST safety recommendations31. Simulations estimated at most a 0.2° C temperature rise at the focus, and we did not observe any temperature rise by MR thermometry in our preparation, with a measured noise of ~0.1° C in the target region. Using the same sonication parameters in a soft tissue mimicking phantom resulted in maximum 0.21° C temperature rise at the focal spot.
The variability in measured ARFI signal across participants suggests that different amounts of ultrasound pressure were reaching the target, despite the fact that experimental conditions including transducer positioning, coupling strategy, ultrasound acoustic pressure, sonication protocols, etc. were identical across all conditions. We expect that the differences across participants are primarily due to variations in skull attenuation and phase aberrations. Other factors include differences in brain stiffness, hair thickness, brain motion, skull structure and variability in the quality of transducer coupling. Presumably some of these factors contributed to the lack of significant ARFI signal in one of the participants.
There are several limitations of our study, which include: 1) only a small number of participants were scanned, 2) there were small changes in the parameters employed in scanning between subjects, as technical optimization was being performed while maintaining a cautious approach to safety, 3) the lack of ARFI signal in one subject, without confirmed understanding of the mechanisms at play (as above). Finally (4), we did not attempt specific targeting of the sonication region as would be needed for neuromodulation. This would require corrections for skull and brain composition variability across the multi-element transducer.
CONCLUSION
We demonstrated that the timeseries single-shot MR-ARFI method has substantial advantages over 2DFT methods, by virtue of its reduced sensitivity to interference from bulk and pulsatile brain motion, a reduction in ultrasound exposure (acoustic pressures <0.7 MPa), and reduced scan time (80 seconds per measurement). Future human studies implementing MR-ARFI may be able to achieve more reliable targeting and dosage of low intensity transcranial focused ultrasound.
Supplementary Material
Acknowledgments
The authors would like to thank Patricia Naomi Limon for help with participant preparation during ultrasound transducer coupling. The authors would also like to thank the study participants. This work was supported by NIH MH131684 and NIH EB032743.
Footnotes
Conflicts of interest
No potential conflict of interest was reported by the authors.
Data availability statement
The data that support the findings of this study are available from the corresponding author upon reasonable request.
References
- 1.Legon W, Strohman A. Low-intensity focused ultrasound for human neuromodulation. Nat Rev Methods Primers. 2024;4(1):91. 10.1038/s43586-024-00368-6 [DOI] [Google Scholar]
- 2.Sarica C, Nankoo JF, Fomenko A, et al. Human Studies of Transcranial Ultrasound neuromodulation: A systematic review of effectiveness and safety. Brain Stimulation. 2022;15(3):737–746. 10.1016/j.brs.2022.05.002 [DOI] [PubMed] [Google Scholar]
- 3.Rieke V, Butts Pauly K. MR thermometry. Magnetic Resonance Imaging. 2008;27(2):376–390. 10.1002/jmri.21265 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Ghanouni P, Pauly KB, Elias WJ, et al. Transcranial MRI-Guided Focused Ultrasound: A Review of the Technologic and Neurologic Applications. American Journal of Roentgenology. 2015;205(1):150–159. 10.2214/AJR.14.13632 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Elias WJ, Lipsman N, Ondo WG, et al. A Randomized Trial of Focused Ultrasound Thalamotomy for Essential Tremor. N Engl J Med. 2016;375(8):730–739. 10.1056/NEJMoa1600159 [DOI] [PubMed] [Google Scholar]
- 6.Rezai AR, D’Haese PF, Finomore V, et al. Ultrasound Blood–Brain Barrier Opening and Aducanumab in Alzheimer’s Disease. N Engl J Med. 2024;390(1):55–62. 10.1056/NEJMoa2308719 [DOI] [PubMed] [Google Scholar]
- 7.Lipsman N, Meng Y, Bethune AJ, et al. Blood–brain barrier opening in Alzheimer’s disease using MR-guided focused ultrasound. Nat Commun. 2018;9(1):2336. 10.1038/s41467-018-04529-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Wang JB, Di Ianni T, Vyas DB, et al. Focused Ultrasound for Noninvasive, Focal Pharmacologic Neurointervention. Front Neurosci. 2020;14:675. 10.3389/fnins.2020.00675 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Mohammadjavadi M, Ash RT, Li N, et al. Transcranial ultrasound neuromodulation of the thalamic visual pathway in a large animal model and the dose-response relationship with MR-ARFI. Sci Rep. 2022;12(1):19588. 10.1038/s41598-022-20554-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Gaur P, Casey KM, Kubanek J, et al. Histologic safety of transcranial focused ultrasound neuromodulation and magnetic resonance acoustic radiation force imaging in rhesus macaques and sheep. Brain Stimul. 2020;13(3):804–814. 10.1016/j.brs.2020.02.017 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Phipps MA, Jonathan SV, Yang PF, et al. Considerations for ultrasound exposure during transcranial MR acoustic radiation force imaging. Sci Rep. 2019;9(1):16235. 10.1038/s41598-019-52443-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Ilovitsh A, Fite BZ, Ilovitsh T, Ferrara KW. Acoustic radiation force imaging using a single-shot spiral readout. Phys Med Biol. 2019;64(12):125004. 10.1088/1361-6560/ab1e21 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.McDannold N, Maier SE. Magnetic resonance acoustic radiation force imaging. Med Phys. 2008;35(8):3748–3758. 10.1118/1.2956712 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Radicke M, Engelbertz A, Habenstein B, et al. New image contrast method in magnetic resonance imaging via ultrasound. Hyperfine Interact. 2008;181(1–3):21–26. 10.1007/s10751-008-9628-6 [DOI] [Google Scholar]
- 15.Kaye EA, Chen J, Pauly KB. Rapid MR‐ARFI method for focal spot localization during focused ultrasound therapy. Magnetic Resonance in Med. 2011;65(3):738–743. 10.1002/mrm.22662 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Kaye EA, Pauly KB. Adapting MRI acoustic radiation force imaging for in vivo human brain focused ultrasound applications. Magnetic Resonance in Med. 2013;69(3):724–733. 10.1002/mrm.24308 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Chen J, Watkins R, Pauly KB. Optimization of encoding gradients for MR-ARFI. Magn Reson Med. 2010;63(4):1050–1058. 10.1002/mrm.22299 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Qiao Y, Li Y, Leng Q, et al. Highly accelerated magnetic resonance acoustic radiation force imaging for in vivo transcranial ultrasound focus localization: A comparison of three reconstruction methods. NMR in Biomedicine. 2021;34(12):e4598. 10.1002/nbm.4598 [DOI] [PubMed] [Google Scholar]
- 19.Glover GH. Spiral imaging in fMRI. NeuroImage. 2012;62(2):706–712. 10.1016/j.neuroimage.2011.10.039 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Friston KJ, Jezzard P, Turner R. Analysis of functional MRI time‐series. Human Brain Mapping. 1994;1(2):153–171. 10.1002/hbm.460010207 [DOI] [Google Scholar]
- 21.Friston KJ, Holmes AP, Poline JB, et al. Analysis of fMRI Time-Series Revisited. NeuroImage. 1995;2(1):45–53. 10.1006/nimg.1995.1007 [DOI] [PubMed] [Google Scholar]
- 22.Chang C, Glover GH. Variable-density spiral-in/out functional magnetic resonance imaging: Variable-Density Spiral fMRI. Magn Reson Med. 2011;65(5):1287–1296. 10.1002/mrm.22722 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Friston KJ, Buechel C, Fink GR, Morris J, Rolls E, Dolan RJ. Psychophysiological and Modulatory Interactions in Neuroimaging. NeuroImage. 1997;6(3):218–229. 10.1006/nimg.1997.0291 [DOI] [PubMed] [Google Scholar]
- 24.Glover GH, Li TQ, Ress D. Image-based method for retrospective correction of physiological motion effects in fMRI: RETROICOR. Magn Reson Med. 2000;44(1):162–167. 10.1002/1522-2594(200007)44:1<162::aid-mrm23>3.0.co;2-e [DOI] [PubMed] [Google Scholar]
- 25.Friston KarlJ, Ashburner J, Frith CD, Poline J‐B., Heather JD, Frackowiak RSJ. Spatial registration and normalization of images. Human Brain Mapping. 1995;3(3):165–189. 10.1002/hbm.460030303 [DOI] [Google Scholar]
- 26.Grissom WA, Rieke V, Holbrook AB, et al. Hybrid referenceless and multibaseline subtraction MR thermometry for monitoring thermal therapies in moving organs. Medical Physics. 2010;37(9):5014–5026. 10.1118/1.3475943 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.McDannold N Quantitative MRI-based temperature mapping based on the proton resonant frequency shift: review of validation studies. Int J Hyperthermia. 2005;21(6):533–546. 10.1080/02656730500096073 [DOI] [PubMed] [Google Scholar]
- 28.Winter L, Seifert F, Zilberti L, Murbach M, Ittermann B. MRI‐Related Heating of Implants and Devices: A Review. Magnetic Resonance Imaging. 2021;53(6):1646–1665. 10.1002/jmri.27194 [DOI] [PubMed] [Google Scholar]
- 29.Arduino A, Zanovello U, Hand J, et al. Heating of hip joint implants in MRI: The combined effect of RF and switched‐gradient fields. Magnetic Resonance in Med. 2021;85(6):3447–3462. 10.1002/mrm.28666 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Attali D, Tiennot T, Schafer M, et al. Three-layer model with absorption for conservative estimation of the maximum acoustic transmission coefficient through the human skull for transcranial ultrasound stimulation. Brain Stimulation. 2023;16(1):48–55. 10.1016/j.brs.2022.12.005 [DOI] [PubMed] [Google Scholar]
- 31.Martin E, Aubry JF, Schafer M, Verhagen L, Treeby B, Pauly KB. ITRUSST consensus on standardised reporting for transcranial ultrasound stimulation. Brain Stimulation. 2024;17(3):607–615. 10.1016/j.brs.2024.04.013 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Thielscher A, Antunes A, Saturnino GB. Field modeling for transcranial magnetic stimulation: A useful tool to understand the physiological effects of TMS? In: 2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC). Milan: IEEE; 2015:222–225. 10.1109/EMBC.2015.7318340 [DOI] [PubMed] [Google Scholar]
- 33.Pichardo S BabelBrain: An Open-Source Application for Prospective Modeling of Transcranial Focused Ultrasound for Neuromodulation Applications. IEEE Trans Ultrason, Ferroelect, Freq Contr. 2023;70(7):587–599. 10.1109/TUFFC.2023.3274046 [DOI] [PubMed] [Google Scholar]
- 34.Miscouridou M, Pineda-Pardo JA, Stagg CJ, Treeby BE, Stanziola A. Classical and Learned MR to Pseudo-CT Mappings for Accurate Transcranial Ultrasound Simulation. IEEE Trans Ultrason, Ferroelect, Freq Contr. 2022;69(10):2896–2905. 10.1109/TUFFC.2022.3198522 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Menz MD, Ye P, Firouzi K, et al. Radiation Force as a Physical Mechanism for Ultrasonic Neurostimulation of the Ex Vivo Retina. J Neurosci. 2019;39(32):6251–6264. 10.1523/JNEUROSCI.2394-18.2019 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Sigona MK, Caskey CF. Ultrasound neuromodulation: planning and validating treatments. Current Opinion in Behavioral Sciences. 2024;59:101430. 10.1016/j.cobeha.2024.101430 [DOI] [Google Scholar]
- 37.Chaplin V, Phipps MA, Jonathan SV, et al. On the accuracy of optically tracked transducers for image-guided transcranial ultrasound. Int J CARS. 2019;14(8):1317–1327. 10.1007/s11548-019-01988-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Riis TS, Webb TD, Kubanek J. Acoustic properties across the human skull. Ultrasonics. 2022;119:106591. 10.1016/j.ultras.2021.106591 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Webb TD, Leung SA, Rosenberg J, et al. Measurements of the Relationship Between CT Hounsfield Units and Acoustic Velocity and How It Changes With Photon Energy and Reconstruction Method. IEEE Trans Ultrason, Ferroelect, Freq Contr. 2018;65(7):1111–1124. 10.1109/TUFFC.2018.2827899 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Li N, Gaur P, Quah K, Butts Pauly K. Improving in situ acoustic intensity estimates using MR acoustic radiation force imaging in combination with multifrequency MR elastography. Magnetic Resonance in Med. 2022;88(4):1673–1689. 10.1002/mrm.29309 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Klarhöfer M, Barth M, Moser E. Comparison of multi-echo spiral and echo planar imaging in functional MRI. Magnetic Resonance Imaging. 2002;20(4):359–364. 10.1016/S0730-725X(02)00505-2 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The data that support the findings of this study are available from the corresponding author upon reasonable request.