ABSTRACT
Background and Objective
Chromocolonoscopy significantly improves polyp/adenoma detection rates (PDR/ADR). However, its integration into routine clinical practice is hindered by its cumbersome mode of application. The objective was to develop an oral nutritional grade delivery system and to assess its efficacy for colonic release, mucosal staining and PDR/ADR as a proof of concept.
Methods
Food‐grade shellac microcapsules releasing 87.5 mg of patent blue V (PBM) pH‐ and time‐dependently were used in 35 volunteers receiving diagnostic colonoscopy either due to a positive fecal occult blood test or surveillance in inflammatory bowel disease. Six capsules were administered p.o. during the bowel preparation (Klean‐Prep). Mucosal staining was assessed in total and per segment using a five‐point grading scale. PDR and ADR were evaluated and compared to a propensity score‐matched comparison cohort in a 1:3 ratio.
Results
In the PBM cohort, 97.1% (34/35) achieved an optimal to acceptable staining quality (SQ) score of ≥ 8, with a mean total score of 13.4 ± 2.9. PDR was significantly higher in the PBM group at 62.8% compared to 42.9% in the comparison group (CG; p = 0.04). ADR showed no significant differences (p = 0.06). The use of PBM resulted in a significantly increased number of detected polyps and adenomas per colonoscopy compared with CG (polyps: PMB = 1.1 ± 1.1 vs. CG = 0.6 ± 0.8, p = 0.02; adenomas: PBM = 0.8 ± 0.9 vs. CG = 0.3 ± 0.5; p = 0.02).
Conclusion
The novel PBM demonstrated uniform mucosal staining when utilized in chromocolonoscopy. Delayed‐release patent blue V appears to be a safe and effective alternative to dye‐spray techniques and existing oral chromoendoscopy modalities.
Keywords: adenoma detection rate, chromocolonoscopy, colorectal carcinoma, inflammatory bowel diseases, microcapsules, patent blue V, pH‐dependent release, polyp detection, shellac coating, ulcerative colitis
Summary
-
Established knowledge on this subject
-
◦
Chromocolonoscopy improves the visualization and differentiation of mucosal structures, hence increases the polyp and adenoma detection rates, and is especially recommended for high‐risk individuals with colorectal carcinoma or patients with UC undergoing surveillance colonoscopy.
-
◦
Dye‐based chromocolonoscopy employing spraying catheters is inconvenient to use in the daily routine; the recently approved oral formulation of methylene blue MMX capsules contains the controversial absorptive dye methylene blue.
-
◦
Patent blue V is a well‐established and safe food‐dye (E131) endorsed by the European Food Safety Authority.
-
◦
Food‐grade shellac can be used for encapsulation.
-
◦
-
The significant and new findings of this case series
-
◦
Development, stability testing and production of microcapsules formulated by spray‐coating cellulose matrix cores containing patent blue V with two layers of food grade shellac and an interlayer of citric acid for pH buffering.
-
◦
Proof of concept of the pH‐dependent burst release.
-
◦
Effective staining of the colonic mucosa with patent blue V microcapsules.
-
◦
Elevated polyp detection rates by chromocolonoscopy with patent blue V microcapsules versus propensity score matching comparison population.
-
◦
No significant rise of adverse events according to the use of patent blue V microcapsules.
-
◦
No prolongation of the examination times during chromocolonoscopy after self‐administered staining with patent blue V microcapsules.
-
◦
Abbreviations
- AC
ascending colon
- ADI
acceptable daily intake
- ADR
adenoma detection rate
- AE
adverse event(s)
- AI
artificial intelligence
- AMR
adenoma miss rate
- ASA
American Society of Anesthesiologists
- BBPS
Boston Bowel Preparation Scale
- CADe
computer assisted detection
- CADx
computer assisted differentiation
- CRC
colorectal cancer
- DC
descending colon
- DGVS
Deutsche Gesellschaft für Verdauungs‐und Stoffwechselerkrankungen
- EFSA
European Food Safety Authority
- ESGE
European Society of Gastrointestinal Endoscopy
- HDWL(E)
High‐definition white light (endoscopy)
- IBD
inflammatory bowel diseases
- MB
methylene blue
- MB‐MMX
methylene blue MMX tablets
- PB
patent blue V
- PBM
patent blue V microcapsules
- PDR
polyp detection rate
- PSM
propensity score matching
- SD
standard deviation
- SMDs
standardized mean differences
- SQ
staining quality
- SR
sigmoid/rectum
- TC
transverse colon
- UC
ulcerative colitis
- WLI
white light imaging
1. Introduction
The European Society of Gastrointestinal Endoscopy (ESGE) has established quality criteria for screening and diagnostic colonoscopy, serving as a benchmark for endoscopic practice and ensuring optimal patient outcomes. The polyp (PDR) and adenoma detection rates (ADR) have been defined as a key criteria for ultimately reducing colorectal cancer (CRC) through early detection and prevention [1]. Despite the procedure's high diagnostic accuracy, a recent meta‐analysis evaluating 43 randomized controlled trials with more than 15,000 tandem colonoscopies exhibited an adenoma miss rate (AMR) of 26%. The respective risk of an interval colorectal cancer was estimated between 1.8% and 4.0% [2].
Therefore, enhancing ADR stands as a crucial imperative in CRC prevention with various technical modalities [3]. Chromoendoscopy improves the visualization and differentiation of mucosal structures by applying either digital contrast and color enhancements (dye‐less, e.g., narrow band imaging (NBI)) or direct mucosal staining with different dyes (dye‐based) [4], which is still considered as the gold standard. Beyond significantly increasing ADR in screening colonoscopy [3, 5], chromocolonoscopy is particularly advocated in the surveillance of CRC in inflammatory bowel diseases (IBD), especially ulcerative colitis (UC) [4]. Direct application of dyes enhances visualization of subtle architectural changes of the mucosa, thereby improving the detection rate of dysplasia in UC. Both clinical trials and real‐world studies show that chromoendoscopy outperforms white‐light endoscopy in the detection of dysplasia, which is consistent regardless of the endoscopist's level of experience or the availability of high‐resolution endoscopic equipment [6, 7].
Although dye‐based chromoendoscopy has demonstrated superior efficacy in dysplasia and cancer detection, its widespread adoption remains limited. This is primarily attributed to the heightened procedural demands in terms of time, sedation, and cleaning, coupled with elevated costs and advanced skills to achieve uniform mucosal staining. To overcome these problems, an oral administration of methylene blue tablets has recently been approved (Methylene Blue MMX; MB‐MMX, Cosmo Pharmaceuticals, Dublin, Ireland) and showed an increase in ADR of approximately 9% compared to placebo in a phase III trial [8]. Former studies provide an indication that the incorporation of MB into epithelial and other cells might have a risk of causing DNA damage on a micro‐level [9, 10]. Nevertheless, after a single use in routine clinical practice, there was no evidence of a short‐term clinical adverse effect from this DNA damage [11].
Considering these potential adverse effects, staining with a food‐grade dye appears to offer a safer alternative. Patent blue V (PB), a well‐established food‐dye (E131), exhibits minimal absorption, lacks metabolic conversion and has been endorsed by the European Food Safety Authority (EFSA) with an acceptable daily intake (ADI) of 5 mg/kg/day [12]. In our previous research, our group developed shellac‐based microcapsules designed for prolonged colonic drug release [13, 14]. A double layer of shellac, which was pH‐buffered by a citric acid intermediate layer, provides a highly reliable mechanism for pH‐dependent, prolonged release of PB into the ileum and colon [13, 14, 15].
Building upon this unique release principle, the aim of this investigation was to assess shellac food‐grade microcapsules suitable for oral administration, incorporating PB as a proof of principle for their performance in ileo‐colonic release, mucosal staining, and their effectiveness in PDR and ADR.
2. Material and Methods
2.1. Study Overview
Microcapsules with a diameter of 0.35 mm were formulated by spray‐coating cellulose matrix cores containing PB with two layers of food grade shellac and an interlayer of citric acid for pH‐buffering (PBM, PB microcapsules, manufactured by the Department of Food Technology, Christian‐Albrechts‐University Kiel, Kiel, Germany). The architecture of three layers provides a sharp pH‐dependent release profile. Between December 2018 and March 2022, 35 patients scheduled for either diagnostic colonoscopy or a UC surveillance colonoscopy volunteered to receive a chromoendoscopy using six PBM alongside their standard bowel preparation. This was given under compassionate use criteria and was classified as a nutritional intervention.
The evaluation of this case series received approval from the Ethics Committee of the Medical Faculty at Christian‐Albrechts‐University, Kiel, Germany (Reference Numbers: D439/15 and D579/23). Written informed consent was obtained from all participants. All procedures were conducted at the Interdisciplinary Endoscopy Unit of the University Hospital Schleswig‐Holstein, Campus Kiel, Germany.
The primary endpoint of the evaluation was defined as the staining score of the respective bowel segments, serving as proof‐of‐concept testing for the new formulation. Secondary endpoints included PDR, ADR, number and size of polyps, adverse events, and examination times. Patients were matched in a 1:3 ratio with individuals undergoing either diagnostic or surveillance colonoscopy using propensity score matching.
2.2. In‐ and Exclusion Criteria
Two patient cohorts were selected: (i) individuals referred for diagnostic colonoscopy following a positive fecal occult blood test (FOBT) [16] and (ii) patients with UC in clinical remission who were scheduled for surveillance colonoscopy [17]. Exclusion criteria included pregnancy, American Society of Anesthesiologists (ASA) risk classes III‐V, inadequate bowel preparation (Boston Bowel Preparation Score [BBPS] < 2 per segment), incomplete colonoscopy (cecum not reached) and inability to provide informed consent.
2.3. Shellac‐Coated Patent Blue V Microcapsules and Application
To deliver PB to the colonic mucosa, we employed microcapsules with a triple coating system and two layers of food‐grade shellac enclosing citric acid to ensure pH‐buffering and subsequent pH‐dependent release within the terminal ileum [14] (Figure 1). Gelatin capsules (size 0) filled with 87.5 mg PBM were used for self‐administration. Dissolution of the PBM and stability under long‐term storage conditions at room temperature were tested as described [14] (Figure 1).
FIGURE 1.
(a) Composition of patent blue V microcapsules (PBM) with 3‐layered coating, two layers of shellac pH‐buffered by a citric acid intermediate; (b) pH‐dependent release profile of PBM and stability assessment immediately after coating and after 1, 3, 6, 9 and 12 months of storage.
Patients in the PBM group underwent bowel preparation using a split‐dose regimen in adherence to the manufacturer's patient information. On the day before colonoscopy, patients consumed 3 L of Macrogol 3350 (Klean‐Prep, Norgine GmbH, Wettenberg, Germany), accompanied by the ingestion of 2 PBM gelatin capsules each after the second and third liter. One liter followed by 2 more capsules was administered the next morning. The interval between the ingestion of the final capsules and the colonoscopy was approximately 4 h (Table S1). The timing and sequence of PBM capsule administration were designed to optimize bowel cleansing and mucosal staining, ensuring effective colonic delivery. After 2 L of macrogol for initial cleansing, PBM capsules were given stepwise during ongoing preparation to achieve gradual uniform dye release. A four‐hour interval between final capsule ingestion and colonoscopy allowed sufficient small bowel transit and colonic release.
Control patients underwent identical preparation without PBM. Klean‐Prep was selected for use with the chromocapsule as the only available isoosmotic solution, preserving the osmotic balance to maintain capsule integrity and prevent premature release, ensuring targeted delivery and diagnostic accuracy [18, 19].
2.3.1. Colonoscopy
High‐definition white light colonoscopies (HDWLE) were performed under conscious sedation using Propofol (Propofol‐Lipuro 10 mg/mL, Braun Melsungen AG, Melsungen, Germany) upon patients’ request, with continuous monitoring throughout the procedure in accordance with current guidelines [20]. All procedures were conducted using Olympus equipment, including the EVIS X1 processor and EXERA III with CF‐190 colonoscopes (Olympus, Tokyo, Japan). All colonoscopies comprised in this study were performed by two highly experienced endoscopists (ME and JB), each with experiences exceeding 3000 diagnostic/surveillance colonoscopies. HDWL mode was used exclusively during withdrawal from the cecum in all procedures. NBI was applied solely for the characterization of polyps following detection under HDWL. Quality criteria for colonoscopy [1] were recorded.
2.4. Staining
The PBM staining quality (SQ) of the colon was evaluated by scoring each specified region of interest: the ascending colon (AC), transverse colon (TC), descending colon (DC), and sigmoid/rectum (SR). The total staining score was determined by summing the scores of all four segments. The staining scores were defined as: 0, no staining; 1, traces of staining (poor traces in colonic mucosa); 2, detectable staining (at least 25% of colonic mucosa stained); 3, acceptable staining (at least 50% of colonic mucosa stained); 4, good staining (at least 75% of colonic mucosa stained); 5, overstained (relevant patches of colonic mucosa overstained and not assessable after suction of colored fluid) [21]. The SQ for each patient was defined as a total score of 8 or higher with no signs of overstaining (score 5). The recorded images of the endoscopies were reviewed by an independent, experienced endoscopist (CCC, experience > 3000 diagnostic/surveillance colonoscopies) blinded to any additional information of the respective patient to verify concordance with the investigator's interpretation regarding the quality of PBM staining in each bowel segment. Additionally, the reviewer assessed whether the areas where polyps were identified were stained or unstained.
2.5. Statistics
A propensity score matching (PSM) analysis was performed to compare patients with positive fecal occult blood test undergoing diagnostic colonoscopy and patients with UC undergoing surveillance colonoscopy, with or without the use of PBM. PSM was conducted separately for the diagnostic and surveillance groups. Patients who underwent colonoscopy with PBM were matched in a 1:3 ratio to those without PBM. The propensity scores were calculated using a logistic regression model based on the following matching criteria: age (continuous variable), sex (categorical: male or female), indication for colonoscopy (categorical: diagnostic/surveillance colonoscopy), and quality of bowel preparation (continuous variable). Nearest‐neighbor matching without replacement was applied, and a caliper width was used to reduce mismatches and optimize balance between the groups. After matching, the balance of baseline characteristics was evaluated using standardized mean differences (SMD) with a threshold of < 0.1 indicating good balance. The final matched cohort demonstrated comparable baseline characteristics between the groups. A 1:3 propensity score was used to optimize statistical power, enhance precision, make full use of available data, and maintain appropriate covariate balance. From a clinical perspective, the 1:3 ratio also reflects real‐world practice, in which chromoendoscopy is performed less frequently due to its greater time and resource demands.
Descriptive statistics are presented as mean ± standard deviation (SD) unless otherwise stated. Comparative analyses were performed using the Mann‐Whitney test for nonparametric comparisons or a contingency table with calculation of Fisher's exact test, as appropriate. A p value of < 0.05 was considered statistically significant. All statistical analyses were performed using GraphPad Prism Version 10 (GraphPad Software, La Jolla, CA, USA).
3. Results
3.1. Quality Control of Patent Blue V Microcapsules
In vitro experiments demonstrated a consistent stability of PBM across designated pH levels of 1.4, 4.5, and 6.8. At pH 7.4, a characteristic burst release pattern of PB was observed that was sustained throughout 12 months at ambient temperature, affirming the formulation's enduring stability (Figure 1).
3.2. Patient Characteristics
41 patients were assessed for eligibility. 37 patients volunteered to participate in the case series under “compassionate use criteria”. Two patients were excluded from further analysis: one due to inadequate preparation for colonoscopy (BBPS = 2), another one withdrew the consent to participate. Ultimately, the PBM group consisted of 35 patients (21 diagnostic colonoscopies; 14 surveillance colonoscopies in UC), who were matched in a 1:3 ratio with 105 control group patients (63 diagnostic and 42 surveillance colonoscopies) based on their respective propensity scores (Figure 2). The key quality criteria showed no significant disparities between the PBM and control groups, as summarized in Table 1.
FIGURE 2.
Flow chart of the study. This flow chart shows the composition of the two analyzed groups. In both arms, NBI was used when required. Both groups received the same bowel preparation and were examined by the same group of endoscopists.
TABLE 1.
Patient characteristics and quality criteria of colonoscopy.
Total |
Diagnostic colonoscopy 1:3 ratio propensity score matching |
Surveillance in ulcerative colitis 1:3 ratio propensity score matching |
|||||||
---|---|---|---|---|---|---|---|---|---|
Patent blue V microcapsules (PBM) | Control | p value | Patent blue V microcapsules (PBM) | Control | p value | Patent blue V microcapsules (PBM) | Control | p value | |
Patient characteristics | |||||||||
Number of patients [N] | 35 | 105 | 21 | 63 | 14 | 42 | |||
Female [N] (%) | 13 (37.1) | 39 (37.1) | > 0.9 | 10 (47.6) | 34 (53.9) | 0.6 | 7 (50) | 22 (52.4) | 0.9 |
Age [years] | 61.2 ± 7.8 | 60.4 ± 6.4 | 0.2 | 63 ± 8.1 | 61.9 ± 5.7 | 0.9 | 58.6 ± 8.1 | 58.1 ± 6.7 | 0.1 |
Indication of colonoscopy [N] (%) | |||||||||
Diagnostic colonoscopy (positive FOBT) | 21 (60) | 63 (60) | > 0.9 | ||||||
Surveillance colonoscopy in ulcerative colitis | 14 (40) | 42 (40) | > 0.9 | ||||||
Quality criteria of colonoscopy | |||||||||
Boston Bowel Preparation Score [median] (25%/75%‐percentile) | 9 (8/9) | 8 (8/9) | > 0.9 | 9 (8/9) | 8 (8/9) | 0.7 | 8 (9/9) | 8 (8/9) | 0.9 |
Net withdrawal time [min] (mean ± SD) | 8.8 ± 1.8 | 8.3 ± 1.9 | 0.1 | 8.9 ± 2.3 | 8.5 ± 1.9 | 0.1 | 8.7 ± 1.3 | 8.1 ± 1.9 | 0.1 |
Total procedure time [min] (mean ± SD) | 17.4 ± 5.4 | 18.2 ± 5.6 | 0.3 | 16.4 ± 6.0 | 17.7 ± 5.2 | 0.3 | 18.4 ± 4.8 | 18.7 ± 6 | 0.3 |
Abbreviation: FOBT, fecal occult blood test.
3.3. Staining Quality
In the PBM cohort, an optimal to acceptable SQ score of ≥ 8 was observed in 34/35 patients (97.1%), with a mean total score of 13.4 ± 2.9 (score range: 4–16). No patients showed overstaining (score = 5) in any segment. The SQ was generally consistent across all anatomical segments (SQAC = 3.4 ± 0.7; SQTC = 3.5 ± 0.7; SQDC = 3.4 ± 0.8; SQSR = 3.2 ± 0.9; p = 0.7 for all comparisons; Figure 3). Polypoid lesions were effectively stained, achieving a mean SQ score of 3.4 ± 0.6 (score range: 2–4). The SQ of the polyps showed a strong correlation with the individual mean staining score (r = 0.74, p < 0.0001; Figure 4).
FIGURE 3.
(a,b) Examples of colonic staining with SQ = 4 in the ascending and descending colon. (c) Comparison of staining quality per segment (AC = ascending colon; DC = descending colon; RS = recto‐sigmoid colon; TC = transverse colon).
FIGURE 4.
(a,b) Examples of polyp staining with SQ = 3 and SQ = 4. (c) Correlation analysis of mean staining quality of the colon and the respective staining quality of detected polyps. Due to an overlay of the correlated data, not all 41 polyps can be displayed.
The Cohen's kappa coefficient for interobserver agreement on SQ between the primary and secondary endoscopists was 0.87, reflecting a high level of consistency.
3.4. PDR/ADR
The PDR was significantly higher in the PBM group at 62.8% compared to 42.9% in the control group (p = 0.04). Similarly, the ADR showed a notable difference, with 45.7% in the PBM group versus 27.6% in the control group, although this did not reach statistical significance (p = 0.06). When evaluating polyps per colonoscopy, the PBM group exhibited a significantly higher mean (1.1 ± 1.1) compared to the control group (0.6 ± 0.8; p = 0.02). A similar pattern was observed for adenomas per colonoscopy, with the PBM group having a mean of 0.8 ± 0.9 versus 0.3 ± 0.5 in the control group (p = 0.02). The PBM group had a significantly higher percentage (41.2%) for medium‐sized polyps (6–10 mm) compared to the control group (15.3%; p = 0.003). No cases of colorectal carcinoma were reported in either group. Results including detailed distribution of the polyp sizes are summarized in Table 2.
TABLE 2.
Comparison of detected polyps and adenomas per group.
Total |
Diagnostic colonoscopy 1:3 ratio propensity score matching |
Surveillance in ulcerative colitis 1:3 ratio propensity score matching |
|||||||
---|---|---|---|---|---|---|---|---|---|
Patent blue V microcapsules | Control | p value | Patent blue V microcapsules | Control | p value | Patent blue V microcapsules | Control | p value | |
Number of patients [N] | 35 | 105 | — | 21 | 63 | — | 14 | 42 | — |
Polyp detection rate (N) [%] | 22 (62.8) | 45 (42.9) | 0.04 | 11 (52.4) | 26 (41.3) | 0.5 | 11 (78.6) | 19 (36.5) | 0.04 |
Adenoma detection rate (N) [%] | 16 (45.7) | 29 (27.6) | 0.06 | 9 (42.9) | 16 (25.4) | 0.2 | 7 (50) | 13 (25) | 0.2 |
Total number of polyps [N] | 51 | 59 | — | 16 | 32 | 35 | 26 | — | |
Polyps per colonoscopy [Mean ± SD] | 1.1 ± 1.1 | 0.6 ± 0.8 | 0.02 | 0.8 ± 0.9 | 0.5 ± 0.7 | 0.1 | 1.4 ± 1.2 | 0.5 ± 0.9 | 0.01 |
Adenomas per colonoscopy [Mean ± SD] | 0.8 ± 0.9 | 0.3 ± 0.5 | 0.02 | 0.7 ± 0.9 | 45 | 0.04 | 0.9 ± 1.1 | 0.3 ± 0.5 | 0.09 |
Polyp sizes [N] (% of all polyps) | |||||||||
1–5 mm | 19 (37.3) | 29 (49.1) | 0.25 | 8 (50) | 15 (46.9) | > 0.9 | 11 (31.4) | 14 (53.8) | 0.1 |
6–10 mm | 21 (41.2) | 9 (15.3) | 0.003 | 4 (25) | 3 (9.4) | 0.2 | 17 (48.6) | 6 (23.1) | 0.06 |
11–15 mm | 9 (16.6) | 19 (32.3) | 0.1 | 3 (18.8) | 14 (43.7) | 0.1 | 6 (17.1) | 5 (19.2) | > 0.9 |
> 15 mm | 1 (1.9) | 2 (3.4) | > 0.9 | 1 (6.2) | 0 | 0.3 | 1 (2.9) | 1 (3.8) | > 0.9 |
Incidence of colorectal carcinoma [%] | 0 | 0 | > 0.9 | 0 | 0 | > 0.9 | 0 | 0 | > 0.9 |
3.5. Adverse Events
The overall incidence of adverse events (AE) was similar between the two groups, with 12 AEs (34.3%) in the PBM group and 44 AEs (41.9%) in the control group (p = 0.5). All recorded AEs were classified as mild. The majority of AEs were related to gastrointestinal disorders (PBM: 4 (11.4%), control: 14 (13.3%), p > 0.9). Specific gastrointestinal symptoms included abdominal discomfort (PBM: 2 (5.7%), control: 6 (5.7%), p > 0.9) and nausea (PBM: 2 (5.7%), control: 7 (6.7%), p > 0.9). Vomiting was rare (PBM: 0, control: 1 (0.9%), p > 0.9). No intervention‐related complications were recorded (Table S2).
3.6. Post Hoc Power Analysis
A post hoc power analysis was performed to assess the statistical power of the study. The primary endpoint was the proportion of patients achieving an acceptable to optimal SQ that was achieved by 97.1% (34/35) of patients. With a non‐inferiority margin of 0.1% and a one‐sided significance level of α = 0.05, the calculated post hoc power was 1.1%. This result confirms that the study is significantly underpowered to clearly demonstrate non‐inferiority.
4. Discussion
Chromoendoscopy plays a vital role in both diagnostic colonoscopy for patients with positive FOBT and surveillance colonoscopy for patients with long‐standing UC.
In the context of patients who present with positive FOBT results, PDR and ADR serve as primary benchmarks for the effectiveness of the procedure [1]. Several technical options to improve the PDR/ADR in diagnostic colonoscopy have already been compared with HDWLE. Real dye‐based chromoendoscopy using a spray catheter has been the gold standard for a long time [22]. In this regard, a risk ratio of 1.16 between HDWLE and dye‐based chromocolonoscopy was reported in a recent meta‐analysis [3]. The administration of dyes to the colonic surface can be categorized into selective and unselective techniques. Selective application targets specific lesions identified initially via white‐light imaging (WLI), whereas unselective application involves broad coverage of the mucosa to enhance lesion detectability. The efficacy of unselective, pan‐chromoendoscopy remains inconclusive; a notable improvement in both PDR and ADR, particularly in the right hemicolon, suggests a potential benefit [5, 23]. Other investigations have not demonstrated significant advantages over standard colonoscopy or HDWLE [24, 25]. On the contrary, chromoendoscopy is valuable for differentiating neoplastic from non‐neoplastic lesions, delineating their margins, assessing surface characteristics and estimating the risk of deeper mucosal invasion [26].
Chromoendoscopy remains a cornerstone in surveillance colonoscopy for patients with long‐standing UC, where the risk of colorectal cancer arises through an inflammation‐dysplasia‐carcinoma sequence. In this context, dye‐based chromoendoscopy improves the detection of subtle, flat dysplastic lesions that are often missed by standard white‐light endoscopy. Numerous studies have demonstrated that dye spray techniques, including the use of methylene blue or indigo carmine, significantly increase dysplasia detection rates in UC patients compared to conventional random biopsies [27, 28]. Reflecting this evidence, both the SCENIC consensus [29] and the European Crohn and Colitis Organization (ECCO) guidelines [7] recommend chromoendoscopy with targeted biopsies as the preferred surveillance approach in UC. While virtual chromoendoscopy techniques such as NBI offer advantages in convenience and enhanced vascular pattern recognition, they have not consistently demonstrated dysplasia detection rates equivalent to dye‐based methods, particularly in the context of chronic inflammation [30, 31]. The present study focused on staining quality as a fundamental performance marker in patients with UC, recognizing that effective mucosal contrast is essential for optimizing dysplasia detection [32].
Despite the proven advantages of chromoendoscopy, a widespread adoption is limited by the cumbersome nature of the dye‐spray technique, requiring additional time for the procedure and sedation as well as extensive cleaning of the equipment.
In contrast, the newly developed PBM formulation offers a simplified and patient‐friendly approach to mucosal staining. In this proof‐of‐concept study, PBM yielded adequate SQ across all colonic segments, with no evidence of overstaining or unstained areas, in both patient groups. These results support the feasibility and effectiveness of this novel PBM formulation in enhancing mucosal contrast. The favorable staining performance, coupled with the ease of administration, suggests that the capsule‐based approach may improve both patient comfort and procedural quality in clinical practice.
To date, only one formulation with a comparable mode of application has received market approval (Methylene blue MMX (MB‐MMX)). In a multicenter Phase III trial, homogeneous SQ was achieved in over 90% of patients, accompanied by a significantly higher ADR in the MB‐MMX group (56.29%) compared to placebo (47.81%) [8]. These findings are partially consistent with our own results. We observed a significant increase in the number of polyps and adenomas detected per colonoscopy, primarily driven by a higher detection rate of polyps measuring 6–10 mm in diameter. The PDR in our cohort was 62.8%, which was significantly higher than the 42.9% observed in the control group (p = 0.04). Although the ADR was also notably higher in the intervention group (45.7% vs. 27.6%), this difference did not reach statistical significance (p = 0.06), likely due to the limited sample size and insufficient statistical power of our proof of concept design. In addition to the trial with MB‐MMX [8], our study demonstrated adequate staining quality in patients with UC.
While MB‐MMX has received regulatory approval following clinical safety evaluations, some preclinical studies have reported that MB may induce DNA damage in colonocytes through oxidative stress mechanisms [10]. Although no clinically relevant AE related to DNA damage has been observed in routine use, it may still be prudent to consider minimizing cumulative exposure to agents with potential genotoxic effects—particularly in high‐risk populations such as patients with UC who may require repeated endoscopies. In contrast, the EFSA approved PB for use in various food products. Extensive evaluations have shown that PB is practically not absorbed into cells, shows only a minimal systemic exposure [33] and no genotoxic potential, supporting its safety as a food additive [12] and reliable choice for chromocolonoscopy.
With the advent of emerging technologies, the clinical role of chromocolonoscopy may diminish in the future. This trend is supported by a recent network meta‐analysis including over 60,000 patients from 94 randomized controlled trials, which evaluated the impact of various endoscopic visualization techniques on ADR [3]. The analysis revealed that newly developed artificial intelligence (AI) algorithms significantly improve ADR compared to traditional methods such as dye‐based chromoendoscopy. However, serrated adenomas are a significant contributor to the development of interval colorectal cancers. These lesions often elude identification by AI algorithms, which typically categorize them as non‐adenomatous. This misclassification can mislead endoscopists, given the high transformation risk of these lesions into colorectal cancer. Currently, chromocolonoscopy remains the gold standard for differentiating this particular subgroup of serrated adenomas [34]. Looking ahead, the integration of AI with advanced chromocolonoscopy could enhance diagnostic accuracy by leveraging the contrast enhancement of chromoendoscopy with the pattern recognition capabilities of AI. In this context, capsule‐based chromoendoscopy may find its most appropriate clinical application in clearly defined patient populations. These include: (i) individuals undergoing diagnostic or screening colonoscopy in whom serrated or atypical adenomas are suspected and require improved mucosal visualization and classification and (ii) patients with long‐standing UC undergoing dysplasia surveillance, where enhanced mucosal contrast is critical for detecting flat or subtle lesions.
4.1. Limitations
A significant limitation of this case series is the small sample size of 35 patients with a calculated power of only 1.1%. While patients were prospectively enrolled under “compassionate use” conditions as a proof concept, this work did not employ randomization or blinding to the respective study groups. Furthermore, this proof‐of‐concept case series did not include a direct control arm using the dye‐spray technique, nor were the endoscopists blinded to the type of staining performed.
5. Conclusion
In conclusion, this proof‐of‐concept demonstrated that orally administered PBM effectively stained the colonic mucosa in patients undergoing diagnostic and surveillance colonoscopy. The use of PBM was associated with a significant improvement in PDR and a notable increase in ADR. Further randomized controlled trials are warranted to validate these findings and to compare PBM with conventional dye‐spray chromoendoscopy, particularly in combination with emerging AI‐based detection algorithms.
Ethics Statement
The evaluation of this case series received approval from the Ethics Committee of the Medical Faculty at Christian‐Albrechts‐University, Kiel, Germany (Reference Numbers: D439/15 and D579/23). Written informed consent was obtained from all participants.
Conflicts of Interest
Patent blue V microcapsules are protected by the European patent no. 3445336 (authors among the inventors, in alphabetical order: M. Ellrichmann, J. K. Keppler, S. Schreiber, K. Schwarz, E.‐M. Theismann, and G. H. Waetzig). G. H. Waetzig is partially employed by CONARIS Research Institute AG (Kiel, Germany). M. Ellrichmann received consultation fees from Olympus and Medtronic. All other authors have nothing to disclose.
Supporting information
Table S1
Table S2
Acknowledgments
We acknowledge the nurses and carers of the Interdisciplinary Endoscopy Unit of the University Hospital Schleswig‐Holstein, Campus Kiel, Germany, for supporting the procedures. Open Access funding enabled and organized by Projekt DEAL.
Funding: The study was supported by internal funding.
Berenice Schulte and Georg H. Waetzig shared first authorship, these authors contributed equally to the work; Stefan Schreiber and Mark Ellrichmann shared last authorship, these authors contributed equally to the work.
Data Availability Statement
The data that support the findings of this study are available from the corresponding author upon reasonable request.
References
- 1. Kaminski M. F., Thomas‐Gibson S., Bugajski M., et al., “Performance Measures for Lower Gastrointestinal Endoscopy: A European Society of Gastrointestinal Endoscopy (ESGE) Quality Improvement Initiative,” United European Gastroenterology Journal 5 (2017): 309–334, 10.1177/2050640617700014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Zhao S., Wang S., Pan P., et al., “Magnitude, Risk Factors, and Factors Associated With Adenoma Miss Rate of Tandem Colonoscopy: A Systematic Review and Meta‐Analysis,” Gastroenterology 156, no. 6 (2019): 1661–1674, 10.1053/j.gastro.2019.01.260. [DOI] [PubMed] [Google Scholar]
- 3. Aziz M., Haghbin H., Sayeh W., et al., “Comparison of Artificial Intelligence With Other Interventions to Improve Adenoma Detection Rate for Colonoscopy: A Network Meta‐Analysis,” Journal of Clinical Gastroenterology 58, no. 2 (2022): 143–155, 10.1097/MCG.0000000000001813. [DOI] [PubMed] [Google Scholar]
- 4. Kucharzik T., Dignass A. U., Atreya R., et al., “Aktualisierte S3‐Leitlinie Colitis Ulcerosa – Living Guideline August 2020 – AWMF‐Registriernummer: 021‐009,” Zeitschrift für Gastroenterologie 58, no. 12 (2020): e241–e326, 10.1055/a-1296-3444. [DOI] [PubMed] [Google Scholar]
- 5. Pohl J., Schneider A., Vogell H., Mayer G., Kaiser G., and Ell C., “Pancolonic Chromoendoscopy With Indigo Carmine Versus Standard Colonoscopy for Detection of Neoplastic Lesions: A Randomised Two‐Centre Trial,” Gut 60, no. 4 (2011): 485–490, 10.1136/gut.2010.229534. [DOI] [PubMed] [Google Scholar]
- 6. Omata F., Sachiko O., Gautam A. D., et al., “Image‐Enhanced, Chromo, and Cap‐Assisted Colonoscopy for Improving Adenoma/Neoplasia Detection Rate: A Systematic Review and Meta‐Analysis,” Scandinavian Journal of Gastroenterology 49 (2014): 222–237, 10.3109/00365521.2013.863964. [DOI] [PubMed] [Google Scholar]
- 7. Maaser C., Sturm A., Vavricka S. R., et al., “ECCO‐ESGAR Guideline for Diagnostic Assessment in IBD Part 1: Initial Diagnosis, Monitoring of Known IBD, Detection of Complications,” Journal of Crohn's and Colitis 13, no. 2 (2019): 144–164, 10.1093/ecco-jcc/jjy113. [DOI] [PubMed] [Google Scholar]
- 8. Repici A., Wallace M. B., East J. E., et al., “Efficacy of Per‐Oral Methylene Blue Formulation for Screening Colonoscopy,” Gastroenterology 156, no. 8 (2019): 2198–2207.e1, 10.1053/j.gastro.2019.02.001. [DOI] [PubMed] [Google Scholar]
- 9. National Library of Medicine (US) NC for BI , PubChem Compound Summary for CID 6099, Methylene Blue PubChem: [Internet], https://pubchem.ncbi.nlm.nih.gov. [Google Scholar]
- 10. Davies J., Burke D., Olliver J. R., Hardie L. J., Wild C. P., and Routledge M. N., “Methylene Blue But Not Indigo Carmine Causes DNA Damage to Colonocytes In Vitro and In Vivo at Concentrations Used in Clinical Chromoendoscopy,” Gut 56, no. 1 (2007): 155–156, 10.1136/gut.2006.107300. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Repici A., Ciscato C., Wallace M., et al., “Evaluation of Genotoxicity Related to Oral Methylene Blue Chromoendoscopy,” Endoscopy 50, no. 10 (2018): 1027–1032, 10.1055/a-0630-1004. [DOI] [PubMed] [Google Scholar]
- 12. EFSA Panel on Food Additives and Nutrient Sources added to Food (ANS) , “Scientific Opinion on the Re‐Evaluation of Patent Blue V (E 131) As a Food Additive,” EFSA Journal 11, no. 3 (2013): 2818, 10.2903/j.efsa.2013.2818. [DOI] [Google Scholar]
- 13. Fangmann D., Theismann E.‐M., Türk K., et al., “Targeted Microbiome Intervention by Microencapsulated Delayed‐Release Niacin Beneficially Affects Insulin Sensitivity in Humans,” Diabetes Care 41, no. 3 (2018): 398–405, 10.2337/dc17-1967. [DOI] [PubMed] [Google Scholar]
- 14. Theismann E.‐M., Keppler J. K., Knipp J.‐R., et al., “Adjustment of Triple Shellac Coating for Precise Release of Bioactive Substances With Different Physico‐Chemical Properties in the Ileocolonic Region,” International Journal of Pharmaceutics 564 (2019): 472–484, 10.1016/j.ijpharm.2019.04.039. [DOI] [PubMed] [Google Scholar]
- 15. Wätzig G., Schwarz K., Keppler J., et al., “Shellac Microcapsule Formulations and Compositions”, European Patent No. 3445336 (2020).
- 16. Winawer S. J., “Screening of Colorectal Cancer,” Surgical Oncology Clinics of North America 14, no. 4 (2005): 699–722, 10.1016/j.soc.2005.05.009. [DOI] [PubMed] [Google Scholar]
- 17. Pouw R. E., Bisschops R., Gecse K. B., et al., “Endoscopic Tissue Sampling ‐ Part 2: Lower Gastrointestinal Tract. European Society of Gastrointestinal Endoscopy (ESGE) Guideline,” Endoscopy 53, no. 12 (2021): 1261–1273, 10.1055/a-1671-6336. [DOI] [PubMed] [Google Scholar]
- 18. Tan J. J. Y. and Tjandra J. J., “Which Is the Optimal Bowel Preparation for Colonoscopy ‐ A Meta‐Analysis,” Colorectal Disease 8, no. 4 (2006): 247–258, 10.1111/j.1463-1318.2006.00970.x. [DOI] [PubMed] [Google Scholar]
- 19. Connor A., Tolan D., Hughes S., Carr N., and Tomson C., “Consensus Guidelines for the Safe Prescription and Administration of Oral Bowel‐Cleansing Agents,” Gut 61, no. 11 (2012): 1525–1532, 10.1136/gutjnl-2011-300861. [DOI] [PubMed] [Google Scholar]
- 20. Wehrmann T., Riphaus A., Eckardt A. J., et al., “Updated S3 Guideline “Sedation for Gastrointestinal Endoscopy” of the German Society of Gastroenterology,” Digestive and Metabolic Diseases (DGVS) ‐ June 2023 ‐ AWMF‐Register‐No. 021/014. Z Gastroenterol 61, no. 11 (2023): e654–e705, 10.1055/a-2165-6388. [DOI] [PubMed] [Google Scholar]
- 21. Di Stefano A. F. D., Radicioni M. M., Vaccani A., et al., “Methylene Blue MMX® Tablets for Chromoendoscopy. Bioavailability, Colon Staining and Safety in Healthy Volunteers Undergoing a Full Colonoscopy,” Contemporary Clinical Trials 71 (2018): 96–102, 10.1016/j.cct.2018.06.001. [DOI] [PubMed] [Google Scholar]
- 22. Tada M., Katoh S., Kohli Y., and Kawai K., “On the Dye Spraying Method in Colonofiberscopy,” Endoscopy 8, no. 2 (1977): 70–74, 10.1055/s-0028-1098379. [DOI] [PubMed] [Google Scholar]
- 23. Paiva R. A., Queiroz F. L., França Neto P. R., et al., “Polyp Detection in the Cecum and Ascending Colon by Dye Based Chromoendoscopy ‐ Is its Routine Use Justified?,” Revista do Colégio Brasileiro de Cirurgiões 50 (2023): e20233562, 10.1590/0100-6991e-20233562-en. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Lapalus M.‐G., Helbert T., Napoleon B., Rey J., Houcke P., and Ponchon T., “Does Chromoendoscopy With Structure Enhancement Improve the Colonoscopic Adenoma Detection Rate?,” Endoscopy 38, no. 5 (2006): 444–448, 10.1055/s-2006-925265. [DOI] [PubMed] [Google Scholar]
- 25. Brown S. R., Baraza W., Din S., and Riley S., “Chromoscopy Versus Conventional Endoscopy for the Detection of Polyps in the Colon and Rectum,” Cochrane Database of Systematic Reviews 4 (2016): CD006439, 10.1002/14651858.CD006439.pub4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Tsuda S., Veress B., Tóth E., et al., “Flat and Depressed Colorectal Tumours in a Southern Swedish Population: A Prospective Chromoendoscopic and Histopathological Study,” Gut 51, no. 4 (2002): 550–555, 10.1136/gut.51.4.550. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Guerrero Vinsard D., Bruining D. H., East J. E., et al., “Interobserver Agreement of the Modified Paris classification and Histology Prediction of Colorectal Lesions in Patients With Inflammatory Bowel Disease,” Gastrointestinal Endoscopy 97, no. 4 (2023): 790–798.e2, 10.1016/j.gie.2022.11.006. [DOI] [PubMed] [Google Scholar]
- 28. Parra‐Izquierdo V., Otero‐Regino W., Juliao‐Baños F., et al., “Dysplasia and Colorectal Cancer Surveillance in Ulcerative Colitis Patients in Latin America: Real‐World Data,” Crohns Colitis 360 7, no. 1 (2025): otae081, 10.1093/crocol/otae081. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Laine L., Kaltenbach T., Barkun A., et al., “SCENIC International Consensus Statement on Surveillance and Management of Dysplasia in Inflammatory Bowel Disease,” Gastroenterology 148, no. 3 (2015): 639–651.e28, 10.1053/j.gastro.2015.01.031. [DOI] [PubMed] [Google Scholar]
- 30. Hata K., Kishikawa J., Anzai H., et al., “Surveillance Colonoscopy for Colitis‐Associated Dysplasia and Cancer in Ulcerative Colitis Patients,” Digestive Endoscopy 28, no. 3 (2016): 260–265, 10.1111/den.12505. [DOI] [PubMed] [Google Scholar]
- 31. Kandiah K., Subramaniam S., Thayalasekaran S., et al., “Multicentre Randomised Controlled Trial on Virtual Chromoendoscopy in the Detection of Neoplasia During Colitis Surveillance High‐Definition Colonoscopy (The VIRTUOSO Trial),” Gut 70, no. 9 (2021): 1684–1690, 10.1136/gutjnl-2020-320980. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Mohamed M. F. H., Marino D., Elfert K., et al., “Dye Chromoendoscopy Outperforms High‐Definition White Light Endoscopy in Dysplasia Detection for Patients With Inflammatory Bowel Disease: An Updated Meta‐Analysis of Randomized Controlled Trials,” American Journal of Gastroenterology 119 (2024): 719–726, 10.14309/ajg.0000000000002595. [DOI] [PubMed] [Google Scholar]
- 33. EFSA Panel on Food Additives and Nutrient Sources added to Food (ANS) , “Scientific Opinion on the Re‐Evaluation of Patent Blue V (E 131) as a Food Additive,” EFSA Journal 11, no. 3 (2013): 2818, 10.2903/j.efsa.2013.2818. [DOI] [Google Scholar]
- 34. Hurt C., Ramaraj R., Farr A., et al., “Feasibility and Economic Assessment of Chromocolonoscopy for Detection of Proximal Serrated Neoplasia Within a Population‐Based Colorectal Cancer Screening Programme (CONSCOP): An Open‐Label, Randomised Controlled Non‐Inferiority Trial,” Lancet Gastroenterology & Hepatology 4, no. 5 (2019): 364–375, 10.1016/S2468-1253(19)30035-4. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Table S1
Table S2
Data Availability Statement
The data that support the findings of this study are available from the corresponding author upon reasonable request.